APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very...

135
DESIGN AND DEVELOPMENT OF A BIOSTRETCH APPARATUS FOR TISSUE ENGINEERING By QIMING PANG A thesis submitted in conformity with the requirements for the degree of Doctor of Philosophy Graduate Department of Mechanical and Industrial Engineering University of Toronto ©Copyright by Qiming Pang 2009

Transcript of APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very...

Page 1: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

DESIGN AND DEVELOPMENT OF A BIOSTRETCH

APPARATUS FOR TISSUE ENGINEERING

By

QIMING PANG

A thesis submitted in conformity with the requirements

for the degree of Doctor of Philosophy

Graduate Department of Mechanical and Industrial Engineering

University of Toronto

©Copyright by Qiming Pang 2009

Page 2: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

DESIGN AND DEVELOPMENT OF A BIOSTRETCH

APPARATUS FOR TISSUE ENGINEERING

DOCTOR OF PHILOSOPHY 2009

QIMING PANG

DEPARTMENT OF MECHANCIDAL AND INDUSTRIAL ENGINEERING

UNIVERSITY OF TORONTO

ABSTRACT

Tissue engineering has emerged as a promising approach to repair, replace or regenerate

damaged tissues using tissue constructs created in vitro. The standard procedure of the strategy

to create a functional tissue is to seed cells on a 3-D biodegradable and biocompatible scaffold,

to grow them under precisely controlled culture conditions provided by a bioreactor system, and

to deliver the matured construct into the patient’s body to induce and direct the growth of the

new and healthy tissue.

In this thesis, a novel bioreactor system is designed and developed, which can provide

uniaxial cyclic stretch to the tissue patch during culture process. The biostretch apparatus

employs non-contact electromagnetic force to cyclically stretch a cell-seeded three-dimensional

scaffold. The non-contact driving force and the specially designed mount allow researchers to

use standard Petri dishes and commercially available CO2 incubators to culture an engineered

ii

Page 3: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

tissue patch with precisely controlled strain. The device greatly simplifies the procedure to

deliver mechanical stimulation during engineering a tissue patch.

Since the applied mechanical stimulus is generated by a magnetic force, the engineered

tissue construct is not only affected by a mechanical force, but also exposed to a magnetic field.

Thus, the effects of the time-varying magnetic field during the culture process are investigated.

The flux density of the field is modeled by COMSOL, and verified by a Gaussmeter. In

addition, one side effect of using electromagnets, that of a temperature increase, is also

investigated. The biomedical experiment results show that neither a weak low frequency

magnetic field (0.1T, 1Hz) nor an increase of 1℃ in temperature has a significant effect on the

cell culture.

The performance of the designed apparatus is verified by the biomedical experiments

from the aspects of cell proliferation and reorganization. Moreover, the mechanical parameters

(strain distribution, strain rate, and stretch force) provided by the apparatus have also been

quantitatively investigated.

iii

Page 4: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Acknowledgments

I would like to thank my supervisors, Professor Jean W. Zu and Professor Ren-ke Li, for

their guidance, support and encouragement. Their knowledge and enthusiasm gave me great

strength throughout my thesis research and have made a profound impact on my future career

endeavors.

I would like to extend my appreciation to my Ph.D. committee, Professor C.A.

Simmons, Professor W.L. Cleghorn, Professor K. Howard, Professor J. Yeow (University of

Waterloo), for their insight, suggestions and time in evaluating my research.

I would also like to thank my colleagues for their scientific discussions and friendship. I

would like to especially thank Hansong Xiao, Peyman Honarmandi, Adebukola Olatunde, and

Ming Jia. They helped to create a comfortable work environment with great enthusiasm and

creativity. Special thanks to Anson Wong for his assistance for testing, to Shuhong Li and Linda

Li for their friendly support and assistance during the years we spent together in the laboratory.

Finally, I dedicate this work to the love of my husband, Hai, for his love, patience, and

unconditional support during these years. To my son Jack who is an unending source of strength

in my life, and to my parents whose love and guidance made this work possible. I am indebted

to them. Thank you so much. I love you all!

iv

Page 5: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Table of Contents

ABSTRACT ................................................................................................................................. ii

Acknowledgments ...................................................................................................................... iv

Table of Contents ........................................................................................................................ v

List of Tables .............................................................................................................................. ix

List of Figures .............................................................................................................................. x

Nomenclature............................................................................................................................ xiii

Chapter 1...................................................................................................................................... 1

Introduction ................................................................................................................................. 1

1.1 Background ........................................................................................................................... 1

1.2 Literature Review.................................................................................................................. 4

1.2.1 Mechanical Stimulation ............................................................................................... 4

1.2.2 Electrical Stimulation................................................................................................... 5

1.2.3 Magnetic Stimulation ................................................................................................... 6

1.2.4 Bioreactors ................................................................................................................... 7

1.3 Objectives of the Thesis ........................................................................................................ 9

1.4 Thesis Overview.................................................................................................................. 10

Chapter 2.................................................................................................................................... 12

Design and Development of a Uniaxial Biostretch Apparatus.............................................. 12

2.1 Introduction ......................................................................................................................... 13

2.1.1 Types of Stretch Systems ........................................................................................... 13

2.1.2 Devices for Uniaxial Stretch ...................................................................................... 15

v

Page 6: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

2.2 Design Principle .................................................................................................................. 20

2.3 Design of Mechanical System............................................................................................. 22

2.3.1 Culture Board ............................................................................................................. 22

2.3.2 Clamps........................................................................................................................ 23

2.3.3 Mounting Tray ........................................................................................................... 24

2.4 Design of Control System ................................................................................................... 26

2.4.1 DAQ PCI-6601........................................................................................................... 26

2.4.2 Definition of the Stretch Plan..................................................................................... 28

2.5 Design of Application Software.......................................................................................... 29

2.5.1 Main Form.................................................................................................................. 30

2.5.2 Setup Form................................................................................................................. 31

2.5.3 Control Flow Chart .................................................................................................... 32

2.6 Conclusions ......................................................................................................................... 33

Chapter 3.................................................................................................................................... 35

Effect of the Time-varying Electromagnetic Field on Cell Biology...................................... 35

3.1 Introduction ......................................................................................................................... 36

3.2 Modeling the Time-varying Electromagnetic Field ............................................................ 38

3.2.1 Electromagnetic Theory ............................................................................................. 38

3.2.2 Electromagnetic Model Analysis ............................................................................... 41

3.3 Experimental Verification of the COMSOL Model............................................................ 47

3.3.1 Flux Density vs. Distance .......................................................................................... 47

3.3.2 Flux Density vs. Exciting Frequency ......................................................................... 50

3.3.3 Flux Density vs. Exciting Current.............................................................................. 51

3.3.4 Comparisons of Simulation and Experiment Results................................................. 52

3.4 Effect on Cell Biology......................................................................................................... 55

3.4.1 Cell Proliferation........................................................................................................ 55

3.4.2 Cell Morphology ........................................................................................................ 58

3.5 Heating Effect of the Electromagnet................................................................................... 60

3.5.1 Experiment Results .................................................................................................... 60

vi

Page 7: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

3.5.2 Bio-experiment Verification ...................................................................................... 63

3.5.3 Effect Factors ............................................................................................................. 65

3.6 Conclusions ......................................................................................................................... 66

Chapter 4.................................................................................................................................... 68

Effect of the Uniaxial Cyclic Stretch on Cell Biology ............................................................ 68

4.1 Introduction ......................................................................................................................... 69

4.2 Validation of the Uniaxial Cyclic Stretch ........................................................................... 71

4.2.1 Cell Proliferation........................................................................................................ 71

4.2.2 Cell Orientation.......................................................................................................... 73

4.2.3 Discussion .................................................................................................................. 76

4.3 Study of Strain Distribution ................................................................................................ 77

4.3.1 Measurement Methodology ....................................................................................... 78

4.3.2 Results ........................................................................................................................ 80

4.4 Study of Strain Rate ............................................................................................................ 85

4.4.1 Measurement Methodology ....................................................................................... 85

4.4.2 Results ........................................................................................................................ 86

4.4.3 Discussion .................................................................................................................. 91

4.5 Study of Stretch Force......................................................................................................... 91

4.5.1 Electromagnetic Force Calculation ............................................................................ 92

4.5.2 Electromagnetic Model Analysis ............................................................................... 93

4.5.3 Results ........................................................................................................................ 95

4.6 Conclusions ....................................................................................................................... 100

Chapter 5.................................................................................................................................. 102

Summary and Future Research............................................................................................. 102

5.1 Summary ........................................................................................................................... 102

5.2 Future Research................................................................................................................. 103

5.2.1 Optimizing the Electromagnet Model...................................................................... 104

vii

Page 8: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

5.2.2 Implementing Bio-experiments................................................................................ 104

5.2.3 Calibrating the stretch force ..................................................................................... 105

5.2.4 Integrating the mechanical and electrical stimuli together ...................................... 105

viii

Page 9: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

List of Tables

Table 2.1 Uniaxial Stretch Devices for Tissue engineering…………………….… 18

Table 3.1 Nomenclature …..….……..….……..….……..….……..….…………… 39

Table 3.2 Geometrical Parameters …..….……..….……..….……..….……………42

Table 3.3 Electrical Parameters..…...……………………….…………….…….… 45

Table 3.4 Magnetic Flux Density vs. Exciting Frequency .….…………….…….…50

Table 3.5 Cell Numbers of Magnetized and Control Samples…………….…….…57

Table 3.6 Cell Numbers at 38℃ and at 37℃ ……………….…………….…….…64

Table 4.1 Cell Numbers of Stretched and Control Samples….…………….…….…72

ix

Page 10: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

List of Figures

Figure 2.1 Motor-driven Uniaxial Stretch Apparatus ………………………….…… 17

Figure 2.2 Electromagnetic-driven Uniaxial Stretch Apparatus ………………….… 20

Figure 2.3 Schematic Diagram of the Biostretch Apparatus …………………….… 21

Figure 2.4 Culture Board ………………….…………………….………………… 23

Figure 2.5 Clamps ………………………….……………………….……………… 24

Figure 2.6 Mounting Tray and the Assembled Culture Dish …………………….… 25

Figure 2.7 Port Allocation of DAQ ……………………………………………….… 27

Figure 2.8 Definition of a Stretch Plan …….…………………….………………… 28

Figure 2.9 Main Form ……………………….…………………….………………… 30

Figure 2.10 Setup Form……………………….……………………….……………… 31

Figure 2.11 Control Flow Chart of the Software……………………….………………32

Figure 2.12 Assembled Bioreator System ………………………………………….… 34

Figure 2.13 Assembled Culture Dish with Electromagnets………….…………………34

Figure 3.1 Meshed Electromagnet Model ……………………………………….……43

Figure 3.2 Distribution of Magnetic Flux Density…………………….………………44

Figure 3.3 Magnetic Flux Density Bx along y Direction ………………………….… 46

Figure 3.4 Simulation Results vs. Testing Results of Magnetic Flux Density ……… 49

Figure 3.5 Exciting Current vs. Flux Density .……………………….……………… 51

Figure 3.6 Magnetic Flux Density Distribution of the Stretch Device…………….… 54

x

Page 11: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Figure 3.7 Cell Morphology of SMC …………………………………………….……59

Figure 3.8 Cell Morphology of BMSC ……………………………….……………… 60

Figure 3.9 Temperature vs. Exciting Time under Room Temperature …………….… 61

Figure 3.10 Temperature vs. Exciting Time in incubator……………………….………62

Figure 4.1 Gross Structure of the Engineered Tissue ………………….………………74

Figure 4.2 Comparison of Cell Numbers…………………………………………….…75

Figure 4.3 Edge Structure of the Engineered Tissue ………………….……………… 76

Figure 4.4 Images of the Gelfoam and GE RTV 6166 …………………………….… 80

Figure 4.5 Strain Distribution of a Typical Sample of Gelfoam……….………………81

Figure 4.6 Distortion of a Marked Dot …………………………………………….…82

Figure 4.7 Strain Distribution for the Middle Rows of the Gelfoam and RTV Silicone

Samples ………………….………………………………………………… 83

Figure 4.8 Stretch Strain Rate at Different Stretch Strength ……………………….… 87

Figure 4.9 Retreat Strain Rate at Different Stretch Strength ……………………….…89

Figure 4.10 Stretch Frequency ………………………………………….……………… 90

Figure 4.11 Meshed Model of an Electromagnet and a Steel Bar…………………….… 94

Figure 4.12 Magnetic Flux Density, x component……………………….………………94

Figure 4.13 Magnetic Flux Density Bn along x Direction…………………………….…96

Figure 4.14 Magnetic Flux Density Bn along y Direction…………………………….…97

Figure 4.15 Comparison of Magnetic Flux Density Bx ……………………………….…97

Figure 4.16 Exciting Current vs. Magnetic Force……………………….……………… 99

Figure 4.17 Distance vs. Magnetic Force……………………………………………….100

xi

Page 12: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Figure 5.1 Schematic Diagram of the Upgraded Device ……………….……………106

xii

Page 13: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Nomenclature

English Characters

A

Magnetic vector potential

zA

Magnetic vector potential along z direction

B

Magnetic flux density

Bx Magnetic flux density, x component

By Magnetic flux density, y component

Bn Magnetic flux density, normal component

D

Electric flux density

d Thickness in z direction

E

Electric field intensity

F

Magnetic force

fext External volume force

H

Magnetic field intensity

J

Volume current density

eJ

External volume current density

zeJ

External volume current density along z direction

M

Magnetization vector

n1 Outward normal direction of material 1

n2 Outward normal direction of material 2

xiii

Page 14: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

P

Electric polarization vector

p Air pressure

Q Heat energy

q Free charge

Rod_force_x Stretching force along x direction

T Stress tensor

T1 Stress tensors of material 1

T2 Stress tensors of material 2

t Temperature

V Electric scalar potential

x Stretching direction

y perpendicular direction of the stretch direction

z perpendicular direction of the stretch plane

Greek Characters

Permittivity

0 Vacuum permittivity

Density

0 Vacuum permeability

r Relative permeability

Conductivity

Angular frequency

xiv

Page 15: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Chapter 1

Introduction

1.1 Background

Organ failure and tissue loss are some of the most devastating and costly problems

affecting current medicine. Organ or tissue transplantation is a well known method to counter

these problems, but it remains an imperfect solution due to donor shortages and potential side

effects of immunosuppression that the recipients may have to endure lifelong [1]. Given these

circumstances, tissue engineering has emerged as a very promising approach to repair, replace

or regenerate damaged tissues using tissue constructs created in vitro.

Tissue engineering, which was formalized in the late 1980s [2], is a multidisciplinary field

combining diverse aspects of engineering, the life sciences, and clinical medicine. It contains

three general strategies to create functional tissue: isolated cells or cell substitutes, tissue-

inducing substances, and cells placed on or within matrices [1, 3]. Among these strategies, the

third strategy is in mainstream use in current tissue engineering. The standard procedure of the

strategy is to seed isolated specific cells on a 3-D biodegradable and biocompatible scaffold, to

grow them under precisely controlled culture conditions provided by a laboratory apparatus, and

1

Page 16: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

to deliver the matured construct into the patient’s body to induce and direct the growth of the

new and healthy tissue [4]. The concept of using the 3-D scaffolds has three major advantages:

(1) The 3-D scaffolds may replace the missing or damaged tissues and provide functional cell-

based substitutes of the native tissues; (2) The size, shape, and mechanical properties of the

engineered tissue patch can be controlled in vitro to treat large-scale defects or damages; (3)

Engineered tissues can provide high-fidelity in vitro models for basic studies of cell functions,

drug screening and responses to physical stimuli [5].

Although the preliminary results of tissue engineering research have helped to develop new

concepts and theories of medical technology and tissue repair, this area is still in its infancy, and

facing significant difficulties and challenges. Firstly, the artificial scaffolds have limited

diffusive ability, preventing nutrients and oxygen from being transported to cells for metabolic

activities. Therefore, acellular tissue-like constructs only have a maximum thickness of 100 µm,

or less than 10 cell layers, and mainly exist at the outer layer of the artificial scaffolds [4].

Additionally, proper mechanical properties of the tissues and organs are critical for them to

achieve their in vivo biomechanical functions, but the cell density and mechanical properties of

engineered tissue patches remain markedly poorer than those of the native tissues [5]. For

example, a lung fuctions under the respiratory pressure, and the vascular system pulses by

hemodynamic shear stress, so the magnitude and frequency of the mechanical loading that

tissues and organs are subject to in vivo can be very large. An engineered tissue patch cultured

under static environment may be unable to withstand the loading at the time of graft, which is a

known potential cause of graft failure in experimental animal studies and preclinical trials [6].

In fact, integrating cells into tissues is a very intricate process. In order to mimic the

morphological, physiological and functional properties of the native tissue, we need to advance

2

Page 17: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

our understanding of the basic principles governing tissue formation, function and failure [7].

Although D’Arcy Thompson proposed that physical forces act as causative agents during tissue

morphogenesis even in the early twentieth century [8], we have still not determined how the

physical forces affect the synthesis and functions of tissues and organs. We are equally

uncertain as to how the living systems (cells, tissues or organs) transfer the dynamical

information and regulate their networks and behaviours. To date, there are still many unknown

puzzles about intercellular interactions, such as the formation of extracellular matrix, as well as

the mechanotransduction in cells.

The above challenges cause significant research in the following areas to develop a

functional engineered tissue: the selection of cell sources, the optimization of scaffolds, and the

creation of bioreactors [1], [7]. The cell sources for implantation include autologous cells (from

a patient), allogeneic cells (from other human donors), and xenogenetic cells (from different

species). In choosing the cell sources, the issues such as ethics, safety and efficacy need to be

considered. Scaffolds may be made of natural or synthetic materials, and must be well matched

with the organ type. For the optimization of the scaffolds, there are many issues that need to be

taken into account, such as the ability for cell adhesion and the degradation time of the scaffold.

The creation of bioreactors means providing the physicochemical environment for the cells to

develop a functioning tissue. Instead of simply pumping the culture medium during culture

process, the next generation of the bioreactors combines more external stimuli during tissue

culture. However, the issues about cell sources and scaffold optimization will not be addressed

further in this thesis. The major objective of the thesis is to design and develop a novel

bioreactor system, which can provide mechanical stimulation (uniaxial stretch) to the tissue

patch during culture process.

3

Page 18: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

1.2 Literature Review

Culturing a tissue patch in a dynamic in vitro microenvironment is important in guiding the

formation of tissue with certain structural and functional characteristics [1], [7]. Scientists

believe that external guides and signals, such as mechanical stress and strain, magnetic, and

electrical stimuli, are important to help cells to grow into functional 3-D implantable tissues [7,

9-11]. In 1988, Vandenburgh et al. [12] and Terracio et al. [13] imposed cyclic stretch on

skeletal and cardiac myocytes, and observed that mechanical stimulation is crucial for the

differentiation and orientation of tissue development. Meanwhile, Radisic et al. [14] found that

electrical field stimulation induced cell alignment and coupling. In this section, a brief literature

review is carried out on the aspects of applying external physical stimuli for tissue engineering.

1.2.1 Mechanical Stimulation

Providing appropriate physiological stresses and strains may help to form correct molecular

and macroscopic architecture during engineering the tissue in vitro, which is crucial for

obtaining proper tissue functions [7, 11]. Therefore, as long as formulating the concept of tissue

engineering, mechanical stimulation has been considered. In 2000, Brown classified the

techniques for mechanical stimulation of cells according to the primary loading modality:

compression loading, longitudinal stretch, bending, axisymmetric substrate bulge, in-plane

substrate distention, and fluid shear stress, etc. [15]. Among these techniques, three loading

modes, compression, stretching, and fluid flow, are the most commonly used methods.

Mechanical stimulation has been shown to increase cellular alignment, proliferation, gene

expression and construct morphology in many different cell types [11, 12, 16]. However,

4

Page 19: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

previous studies have demonstrated that cells throughout the human body are exposed to

various forms of mechanical stimuli, which means that different tissues are sensitive to different

mechanical stimuli [17].

Chondrocytes, cells found in cartilage, are sensitive to compressive loading since cartilage

tissues are primarily subjected to this kind of loading in vivo. Compression has also been

utilized for other cell/tissue types such as bone [18-20]. Other cell types, such as endothelial

cells in blood vessel and osteocytes in bone tissue, are subject to shear stress in vivo. Therefore,

fluid flow is often successfully used as the mechanical stimulation in this kind of cell culture

[21-23]. However, ligaments, muscles such as skeletal muscle, cardiac muscle and smooth

muscle are usually under cyclic stretch in vivo. For these types of cells, cyclic stretch is served

as an important regulator [6, 12, 16, 24]. The types of stretching devices and the effects of cell

biology will be reviewed in Chapter 2 introduction section.

Although literature shows that mechanical stimulation is an important regulator for the

development of engineered tissues with similar construction and mechanical properties as native

tissues, significant work is needed to identify the types of mechanical stimulation required to

optimize the formation of a functional engineered tissue. To date, it is still unclear how the

mechanical force is transferred to the individual cells, and what kinds of responses are yielded

at the molecular and cellular level. Therefore, quantitative study of the influence of mechanical

force is necessary for functional tissue engineering.

1.2.2 Electrical Stimulation

Many tissues and organs in vivo are not only subjected to mechanical loading in order to

achieve their functions, but also exposed to some electrical signals. For instance, the nervous

5

Page 20: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

system utilizes neurons and nerve cells to build an electrochemical wiring network, and to

generate and transform the electrochemical impulses. In addition, the contraction of

myocardium (cardiac muscle) is controlled by an electrical signal generated by the sinoatrial

node. These facts suggest that electrical signals can also be used as an external factor to

stimulate the creation of an engineered tissue patch [25].

Previous studies show that when a weak DC electric field was applied to embryonic

fibroblasts cultivated on a 2-D substitute, the fibroblasts became oriented perpendicular to the

electric field lines and migrated towards the cathodal end of the field [26]. The group of Cho et

al. [27] explored the cellular behaviours on the 3-D matrix with the stimulation of a DC electric

field, and demonstrated that fibroblasts were more readily reoriented than rat mesenchymal

stem cells. Radisic et al. [14] reported that a pulsed electrical field stimulus induced cell

alignment and coupling, improved cell contractile function, and resulted in a remarkable level

of ultra-structural organization. They also found that electrical field stimulation can enhance

cell elongation and orientation, which is directly related to the functional properties of the

engineered contractile tissues [28].

1.2.3 Magnetic Stimulation

To facilitate the development of engineered tissue, besides mechanical and electrical

stimulations, researchers are also seeking for other stimulations. Magnetic field (MF) has been

used as a simple and effective means to orient collagen gels, since collagen gels provide the

matrix substrate circumstance which directly influences the cell growth. In 1993, Tranquillo et

al. [29, 30] presented that a strong magnetic field (Tesla order) can be used to orient collagen

fibrils and subsequently orient the entrapped smooth muscle cells. Hashimoto et al. [31]

6

Page 21: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

demonstrated that moderate magnetic field (millitesla order) can enhance cell migration,

depending on the types of cells. Even a weak electromagnetic field (Gauss order) with low

frequency (<300Hz) has been reported to have an effect on enzyme reaction and transcript

levels for specific genes [32]. Thus, magnetic stimulation can positively affect cell growth in a

wide range of the MF magnitude.

Although studies show that both magnetic and electrical fields can influence the tissue

culture process to some degree compared with purely mechanical stimulation, reports of in vitro

culturing under electric or magnetic stimulus are still rare. Thus, intensive investigation is

needed to carry out to study the influence of physical stimuli, such as mechanical, electrical and

magnetic, on cell morphology and behaviours that are important for controlling engineered

tissue constructs.

1.2.4 Bioreactors

During tissue culture process, the use of scaffolds and laboratory apparatus has been

regarded as a crucial technique. Although there is no universally recognized definition of these

laboratory apparatus, they can be called bioreactors. Bioreactors, with respect to Freed et al.

[33], can be defined as “laboratory tissue-culture devices, which provide a controllable,

mechanically active environment that can be used to study and potentially improve engineered

tissue structure, properties, and integration.” Even though bioreactors should be designed and

fabricated following specifications that differ from tissue to tissue, cell-culture parameters such

as temperature, pH, biochemical gradients, and mechanical stimulation must be continuously

controlled during the maturation period because bioreactors should provide an in vitro

environment mimicking the in vivo conditions [34, 35]. Thereby, the design of an appropriate

7

Page 22: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

bioreactor for a specific tissue is important, but also difficult in that bioreactors provide all the

environment control and regulatory factors necessary for cell culture. Compared with

conventional static culture systems, modern bioreactors normally integrate perfusion system

and/or external physical stimuli in order to regenerate a functional engineered tissue.

To improve nutrient and oxygen transportation, researchers have designed several

bioreactors with different perfusion systems [36], [37] [38]. A basic fluid-dynamic cultivation

vessel is the spinner flask that provides a well-mixed environment around cells. The engineered

tissues growing in a spinner flask have better metabolic activity and morphological appearance

than under static conditions [36]. However, the spinner flask may not be the optimal cultivation

vessel for cells. It causes the turbulent fluid flow at the surface of the constructs. The turbulent

fluid is usually characterized by eddies which can destroy the seeded cells [4, 39]. Fluid-

dynamic cultivation environment can only partially compensate the absence of capillaries by

providing nutrients and gas through the entire thickness [36, 40], meaning that merely adding

perfusion system in the bioreactor is not a powerful way to increase mass diffusion.

Some research groups consider integrating multi stimuli in the culture process. The most

common method is to combining the perfusion system and mechanical stimuli together in one

bioreactor. For example, a continuous pulsatile perfusion system is integrated with mechanical

stimuli for vascular tissue [41, 42]; the perfusion system and static/dynamic compression

loading are applied to cartilage tissue [43], and the perfusion flow and fluid flow-induced shear

stress are combined in bone tissue engineering [44]. Moreover, some groups considered

integrating more than one external stimulus into the system. For instance, Feng et al. [45]

developed a device to provide both electrical stimulus and dynamic tensile force during culture

8

Page 23: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

process. This device tries to solve the coordination problem between the dynamic mechanical

loading and the spontaneous beating.

However, the above techniques have a number of shortcomings that limit their applicability

for in vitro applications. First of all, the relationships between external stimuli and cell reactions

are still unclear. Secondly, using most of the above bioreactors to make tissue constructs

requires a complex casting process beyond commercially available standard cell culture dishes

and CO2 incubators. These specifically designed apparatus make the size and quality of tissue

patch variable and the culture process difficult to operate. In addition, applying external

stimulation during the culture process increases the risk of contamination, especially for long-

term in vitro culture, due to the physical connection and the complicated assembly procedure.

Therefore, design a simple bioreactor system is essential for researchers to further their study.

1.3 Objectives of the Thesis

The main objective of the thesis is to design a bioreactor system that can provide

mechanical stimulation under standard culture frames. It includes a biostretch apparatus that can

provide uniaxial cyclic stretch to a three-dimensional cell-seeded scaffold. The apparatus

employs non-contact electromagnetic force as driving force. The non-contact driving force,

together with the specially designed mount, allows researchers to use standard Petri dishes and

commercially available CO2 incubators to culture an engineered tissue patch under well defined

mechanical conditions. Based on this bioreactor system, the effects of the magnetic field and

uniaxial stretch on cell biology have been studied.

The prototype of the apparatus has been utilized in a cardiac tissue research lab at the

University Health Network in Toronto. Although the apparatus is initially designed for cardiac

9

Page 24: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

tissue engineering research, the apparatus has the potential to be used in other tissue engineering

areas such as bone and musculoskeletal tissue. The objectives are pursued through the following

sub-objectives:

Developing a prototype of the apparatus, including the mechanical system, computer-

based control system and application software

Modeling the electromagnetic field, and investigating the magnetic effects on cell

proliferation and morphology

Investigating the effects of uniaxial stretch from the aspects of cell proliferation and

organization, and quantitatively studying the mechanical parameters provided by the

apparatus, such as strain distribution, strain rate and stretch force

1.4 Thesis Overview

The main body of this thesis consists of three chapters, which discuss three tasks mentioned

in the previous section. They are closely concatenated and organized as a whole system.

However, they can also be treated as three different studies in the related corresponding areas.

In Chapter 2, Design and Development of a Uniaxial Biostretch Apparatus, the past uniaxial

stretch devices are first reviewed. The development procedure of the biostretch apparatus is

elaborated from the aspects of the design principles, the mechanical system, the control system,

and the application software.

In Chapter 3, Effect of the Time-varying Electromagnetic Field, A time-varying

electromagnetic field is studied, since the apparatus uses non-contact force as stretch force. The

field is modeled by commercial software COMSOL. The magnetic flux density B and field

10

Page 25: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

distribution are simulated and verified. The effect of the time-varying magnetic field with

temperature fluctuation on cell proliferation and morphology are investigated.

In Chapter 4, Effect of the Uniaxial Stretch, the performance of the apparatus has been

tested by the bio-experiments from the aspects of cell proliferation and organization. The

mechanical parameters (strain distribution, strain rate, stretch force) provided by the device

have been investigated.

In Chapter 5, Summary and future research, the results obtained in this research are

summarized and the future research directions in this area are proposed.

11

Page 26: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Chapter 2

Design and Development of a Uniaxial

Biostretch Apparatus

The objective of this chapter is to design and develop a prototype of the uniaxial biostretch

apparatus for tissue engineering research, especially for cardiac engineered tissue. The

biostretch apparatus employs non-contact electromagnetic force to cyclically stretch a cell-

seeded three-dimensional scaffold. The non-contact driving force and the specially designed

mount allow researchers to use standard Petri dishes and commercially available CO2

incubators to culture an engineered tissue patch with precisely controlled strain. The device

greatly simplifies the procedure to deliver mechanical stimulation during engineering a tissue

patch. This chapter mainly describes the design principles including the mechanical system,

control system and application software.

In this chapter, a brief introduction of the currently used stretch apparatus is presented in

section 2.1. Section 2.2 describes the design principle and criteria; section 2.3, 2.4, and 2.5

12

Page 27: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

introduce the design of mechanical system, control system and application software,

respectively. Concluding remarks are drawn in section 2.6.

2.1 Introduction

It has long been recognized that mechanical signals can improve the organization,

composition, and function of engineered tissue, such that mechanical stimulation is the most

common physical stimulation in a bioreactor system design. These mechanical stimuli can

successfully mimic the in vivo activities of daily living. However, before utilizing these

mechanical signals to “precondition” the tissue-engineered constructs, the magnitudes of the

mechanical parameters in vivo have to be estimated. Literature review shows that the strain of

tendon tissue can reach 2.4% [46]. Meanwhile, previous research demonstrated that the

circumferential strain from a complete set of the planar tagged images of a human left ventricle

ranges from 25% shortening to 10% stretching, while the strain of a paced canine left ventricle

can range from 20% shortening to 20% stretching [47].

Although mechanical stimuli can be categorized into different groups, as we discussed in

chapter one, only three major loading modes, compression, stretching, and fluid flow are

commonly used [15], [17] . Since my initial study object, cardiac tissue, along with

musculoskeletal and vessel tissues, are all sensitive to stretching, the classification and

application of stretch systems are emphasized here.

2.1.1 Types of Stretch Systems

There are two major stretch methods for cell culture, uniaxial stretch and biaxial stretch. In

biaxial stretch devices, cells are usually cultured on a circular membrane and the force is

13

Page 28: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

commonly generated by pneumatic pressure or pistons [48-52], such that the membrane

deforms in both the radical and circumferential directions. These systems have been used for

culturing fibroblast, tendon, and annulus cells. One of the pneumatic driven stretch devices is

commercialized, called Flexcell® culture systems. However, the strain profile of the biaxial

stretch device [48, 50] varies, with the highest strain along the culture walls and lowest strain in

the center. These variations cause difficulties in analysis the relationship between the strain and

cell effects.

Uniaxial stretch is another commonly used stimulus, because it can offer a very precise

control of end-to-end displacement that is important to evaluate the average strain of the

scaffold [53]. This type of stimulation provides a versatile platform for researchers to

extensively study cell response and its mechanical properties. Uniaxial cyclic stretch exhibits

improvements of morphology and organization of engineered tissues by the following groups.

The group of Vandeburgh [12], [54][55] indicated that mechanical stimulation affected protein

accumulation and localization, while mechanical stimulation also improved cellular

proliferation, myofiber organization. The in vitro generated artificial muscle organs contain

parallel networks of long unbranched myofibers organized into fascicle-like structures.

Eschenhagen et al. and Zimmermanne el al. [2][16][24] presented that chronic mechanical

stretch of engineered heart tissue improved organization of cardiac myocytes into parallel arrays

of rod-shaped cells, and increased cell length and width, myofilament length, and mitochondrial

density. The group of Terracio [13] also imposed mechanical stretch on particular cell types

(cardiac myoctyes, endothelial cells, or fibroblasts) and showed that mechanical stimulation had

a favorable effect on cell orientation. Li et al. [6] cultured isolated heart cells on gelatin-matrix

scaffolds, and demonstrated that cyclical mechanical stretch improved the proliferation and

14

Page 29: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

distribution of the seeded human heart cells, and stimulated the formation and organization of

extracellular matrix, which contributed to the improvement in the mechanical strength of the

cardiac graft.

2.1.2 Devices for Uniaxial Stretch

According to the methods of driving force application, the uniaxial stretch systems can be

classified mainly into two groups, motor-driven and magnetic force driven, shown in Table 2.1.

Motor-driven Devices

The first group uses step-motors to provide tension. Vandeburgh et al. [54-56] [12] started

studying mechanical stimulation in the 1980s, and developed a mechanical cell stimulation

device (version four) shown in Fig. 2.1 (a). The cell-seeded construct is combined with two

steel pins. One pin is immobile, and the other pin attaches to a step-motor (SM), which applies

direct mechanical stretch. This device can hold six plates, and apply sensitive force transducers

(LC) to detect the loading force. With this device, users can define uniaxial the amplitude,

velocity, and pattern of the stretch. Eschenhagen et al. and Zimmermann et al. [16, 24][57],

developed a uniaxial cyclic stretch apparatus driven by a stepper motor, shown in Fig. 2.1 (b).

This device provided a circular mold, and phasic mechanical stretch. It was reported that the

circular geometry causes a homogeneous force distribution throughout the tissue. This phasic

stretch induced hypertrophic growth and marked functional improvement. Meanwhile, the

group of Terracio also designed motor-driven cell stretch devices to study the cardiac cell

behaviors [13, 58], shown in Fig. 2.1 (c). This device has a dual-stretch unit. The unit consists

of a TiN-coated stainless steel yoke (C) assembled with dual-rod sliders (E). One side of the

15

Page 30: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

silicone stretch membrane is clamped with the slider, and the other side is combined with a poly

amide-imide base plate (F). The stepper motor (A) drove the yoke through lead screw (B). The

stretching was processed inside the culture dish (D). Another type of motor-drive device is

shown in Fig. 2.1 (d) [59, 60]. This device consisted of a computer equipped with a four-axis

motion control card, customize software, and four stepper motor controllers operating four

independent cycliuc strain units.

These motor-driven devices allow great versatility of input waveform, which provides

researchers more flexibility to define the stretch patterns. However, in order to apply cyclic

stretch, the motor and the engineered tissue constructs must be in physical contact, which

requires both the culture dishes and the bioreactor system to be customized. Therefore, the

assembly procedure of the culture systems is very complex, which may increase the risk of

contamination and the variability of the products.

16

Page 31: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

(a) (b)

(c) (d)

Figure 2.1 Motor-driven Uniaxial Stretch Apparatus

17

Page 32: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Table 2.1 Uniaxial Stretch Devices for Tissue Engineering

Authors Cell Source Methods of Force Application Parameters studied

Terracio et al. [13, 58] Rat cardiac fibroblast Stepper-motor driven strain

Vandeburgh et al. [54-56] [12] Skeletal muscle myoblast Stepper-motor driven Strain, force

Elson et al.[61, 62] Cardiac fibroblasts,

chicken embryo fibroblasts

Stepper-motor driven Strain, force, stiffness,

frequency

Eschenhagen et al. and

Zimmermann et al. [16, 24][57]

Cardiac myocytes Stepper-motor driven Force, strain

Langelier et al.[63] Fibroblasts Stepper-motor driven strain

Joshi et al.[59] Human dermal fibroblasts Stepper-motor driven strain

Mol et al. [64] Human venous myofibroblasts Linear actuator driven strain

Pfister et al. [65] Rat neuroblastoma and glioma cells Voice coil actuator driven Strain, strain rate

Liu et al. [66][6] Rat aorta endothelial cells, cardiac

myocytes, smooth muscle cells

Magnetic force driven strain, deformation

Smalt et al. [67] Osteoblastic cells Magnetic force driven strain

18

Page 33: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Electromagnetic Force Driven Devices

The second group of uniaxial stretch systems uses electromagnetic force. In one design, a

specially designed bio-stretch apparatus with square culture dishes, shown in Fig. 2.2(a), is used

to impose the uniaxial cyclic stretch on the tissue samples. One side of the tissue patch is fixed

on the dish (lower side of the figure), and the opposite side is attached to a magnetic stainless

steel bar (upper side of the figure). The culture dish is placed in front of a programmable

electromagnet, which can provide a dynamically changing magnetic field. Stretch patterns of

variable frequency, duty cycle, and amplitude can be controlled by the magnetic field [66].

Similarly, Smalt et al. [67], designed another kind of customized culture well, which can apply

cyclic strain (500-5000µƐ, 1Hz) to a polystyrene film by using electromagnetic force, shown in

Fig. 2.2(b).

Literature review shows that uniaxial cyclic stretch is essential for the reorganization,

differentiation, and the mechanical properties of different types of tissues, and that the

magnitude and the type of these stresses/strains are specific to cultured tissues. Among these

stretch devices, I prefer the electromagnet-driven apparatus developed by Liu et al. [66]. This

apparatus used non-contact magnetic force as driving force, which allows the culture dishes to

be easily isolated from the environment and thus greatly reduces the contamination risk. With

this apparatus, Liu et al. and Yang et al. [68, 69], [70] investigated the effect of mechanical

stretch on cell proliferation and differentiation; Lewis et al. [71] studied the adaptation of

craniofacial skeletal muscle following different mechanostimulation – rapid ramp stretch or

cyclical ramp stretch with 7.5% and 15% strain; and Freed et al. [72] applied this equipment to

study the mechanical properties of hybrid cardiac constructs made from rat heart cells, fibrin,

and biodegradable elastomeric knitted fabric. Finally, the group of Li et al. [6] used this

19

Page 34: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

apparatus to grow cardiac muscle grafts and found that cyclical mechanical stretch can improve

the mechanical strength of the engineered cardiac grafts.

(a) (b)

Figure 2.2 Eletromagnetic-driven Uniaxial Stretch Apparatus

2.2 Design Principle

The most important advantage of the electromagnet-driven apparatus is the use of non-

contact driving force, which allows researchers to culture the engineered tissue patch in an

isolated environment. However, some limitations of this apparatus [66, 67] hamper its further

utilization. First, the apparatus has a relatively complicated control system that includes a

biostretch manager, a biostretch controller, and a set of magnet boards with customized Petri

dishes. As such, the complicated assembly procedure and customized culture dishes increases

the risk of contamination. Moreover, the deformation of the tissue patch is only controlled by

magnitude of the stretching force, and a precise and controllable strain magnitude is difficult to

obtain by controlling only the electromagnetic force. One of the goals of my thesis study is to

20

Page 35: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

design a uniaxial stretch apparatus, which can apply precisely controlled strain to the tissue

constructs with the minimum risk of contamination during a long-term tissue culture process.

The new apparatus is initially designed to culture cardiac tissue patches in a standard CO2

incubator and impose controllable uniaxial strain on the cell-seeded scaffold during the culture

procedure. However, if the apparatus can provide a wide range of stretch patterns with different

strain magnitude and frequency, it can become a yielding and versatile tool for researchers to

study different tissue cultures. The desired design criteria of the new biostretch apparatus was

summarized as follows:

It can use standard Petri dishes.

It can provide a wide range of stretch patterns with precisely controlled strain

magnitude and frequency applied on the scaffold.

It can reduce the potential for bacterial or fungal contamination during the culture

procedure.

It can be easily installed in and removed from a CO2 incubator.

Figure 2.3 Schematic Diagram of the Biostretch Apparatus

(1 - Culture boards, 2 - Incubator)

21

Page 36: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

The apparatus consists of two parts, as shown in Fig. 2.3. The first part is a computerized

control system used outside of the incubator, and the other part is composed of up to three

culture boards used inside the incubator. Each culture board can hold three standard Petri

dishes. A cell-seeded scaffold with two stainless steel clamps at each end is mounted in the

center of the culture dish. On the two opposite sides of each culture dish are two identical

electromagnets providing forces equal in magnitude and opposite in direction to two ends of

each culture patch. Therefore, when a current is passed through both magnets, force is applied

symmetrically to the tissue patch.

2.3 Design of Mechanical System

2.3.1 Culture Board

The culture board is designed to hold culture dishes and electromagnets, and can be easily

installed in or removed from the incubator, shown in Fig. 2.4. In order to stretch a useful

number of samples in the commercial CO2 incubator, while minimizing interactions between

electromagnets, each culture board is designed to hold three dishes. One culture board consists

of two plates, the base plate (1) and the hold plate (4). The base plate is used to mount the

electromagnets and the hold plate. The hold plate is mounted in the middle of base plate, and is

used to hold the Petri dishes. Each culture board has two different sizes of hold plates that are

designed for the 35-mm and 60-mm Petri dishes, respectively. There are three pairs of V-shape

blocks (3) on the hold plate, which are used to mount the dish in the middle of two opposite

electromagnets. Combined with hold plate, the slots (2) on the base plate can be used to adjust

the distances between the culture dishes and electromagnets.

22

Page 37: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Figure 2.4 Culture Board

(1-base plate, 2-slot, 3-V-shape block, 4-hold plate)

2.3.2 Clamps

Since the apparatus uses non-contact magnetic force as stretch force, two stainless steel

magnetic clamps need to attach to the two ends of the scaffold, one at each end. When a

magnetic field is applied to the clamps, the clamps move under the magnetic force, causing the

scaffold to stretch. Before designing a proper clamp, two elements have to be taken into

account: Firstly, the clamps must be very simple to assemble because the combination of the

clamps and the cell-seeded scaffold has to be finished in the hood. Secondly, the clamps should

have a symmetric structure such that the tissue patch can undergo a symmetric force.

Fig. 2.5 shows the design of the clamps. The clamp (2) is made of polycarbonate, which

can withstand an alcohol disinfectant without cracking, and has some elastic property for the

stainless magnetic bar (1) to be inserted. The pressure enforces the scaffold (3), steel bar (1),

and the clamp (2) to combine together tightly. The material of the bar is Ph-174, a type of steel

that satisfies two characters – magnetic and stainless, when put into medium. The shape of the

steel bar is rectangular, so it is simple to manufacture and assemble. For a 35-mm Petri dish, the

dimensions of the steel bar are 16mm x 3.5mm x 1mm. In the hood, the operator can simply use

23

Page 38: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

tweezers to insert the rectangular-shaped steel bar into the slot of the clamp and tighten the

scaffold.

Figure 2.5 Clamps

(1-steel bar, 2-clamp, 3-scaffold)

2.3.3 Mounting Tray

To insert the assembled tissue construct (the scaffold with clamps) into the culture dish and

to prevent motion other than uniaxial stretch, a mounting tray is used, shown in Fig. 2.6 on the

bottom left. The center groove (2) confines the scaffold to the centre of the dish, while eight

small slots (3) increase the bottom area of the scaffold that is exposed to the culture medium. To

control the deformation of the scaffold, stopper pins (1) of various sizes are selectively inserted

into the holes on each side of the mounting tray (4). The inner stopper pins (10) ensure that the

scaffold returns to its original position after stretch; the outer stopper pins (11) limit the amount

of stretch. The position and distance between the pins determine the deformation of the scaffold

24

Page 39: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

and the stretch methods. After the tissue construct mounted on the mounting tray, the

assembled tray can be fit into the standard culture dish, shown in Fig. 2.6 right side.

Compared to the traditional stretch method (fixing one side and stretching the other side),

this device can provide two kinds of stretch methods, symmetric stretch, and asymmetric

stretch, based on the assembly of the tray. When the two clamps have same distance L

between the inner and outer stopper pins, the scaffold exhibits symmetric stretch; when the

inner and outer stopper pins tightly secure one of the clamps, and only one clamp is movable,

the scaffold performs asymmetric stretch (traditional method).

Figure 2.6 Mounting Tray and the Assembled Culture Dish

(1-stopper pins; 2-center groove; 3-slot; 4-hole; 5-clamp; 6-magnet; 7-culture dish;

8-scaffold; 9-mounting tray; 10-inner stopper; 11-outer stopper)

25

Page 40: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

2.4 Design of Control System

The apparatus is designed to allow two different types of stretch patterns during the tissue

culture process: continuous cyclic stretch and intermittent cyclic stretch [66]. Continuous cyclic

stretch means cyclically stretching the scaffold at a constant frequency without stopping;

intermittent cyclic stretch means alternately stretching and resting the scaffold. For example, the

scaffold first experiences a period of cyclic stretch on the order of 15 minutes, called burst time,

and then sits in a static environment for a period on the order of 45 minutes, called rest time, to

avoid cell injury. The stretch plan parameters, with specific burst and rest times, are determined

by researchers. In order to conduct a designed stretch plan, a controlled current pulse train is

needed to work as the power supply of electromagnets. This pulse train contains all the

information of the stretch plan. Thus, we need to implement pulse train generators and a user-

friendly interface to customize the stretch parameters of a stretch plan.

2.4.1 DAQ PCI-6601

To fulfill the design criteria, a National Instruments data acquisition card (DAQ) NI PCI-

6601 is employed for data acquisition and pulse train generation. This DAQ card has four

up/down 32-bit Counters/Timers with a maximum source frequency of 20MHz, and up to 32

digital I/O lines. The counters can be used as the pulse train generators, while the digital I/O

lines can be used as input/output control signal channels. To obtain three entirely independent

stretch plans for three culture boards, the four stretch parameters of each stretch plan, strength,

frequency, pattern, and duration of the stretch time, must be controlled separately. Thus, three

counters of the DAQ are employed to generate three pulse trains so that the frequency and the

pattern of each pulse train can be controlled separately. The pattern means the proportion of

26

Page 41: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

high-voltage time (ON-time) and low-voltage time (OFF-time) of a pulse period. Both the

pattern and the frequency can be controlled by setting the configuration of the counter. The port

allocation of PCI-6601 is shown as in Fig. 2.7.

For each stretch plan, the PCI-6601 needs to allocate one output channel from a counter to

generate the pulse train with a specific frequency and pattern; and another eight output channels

to an 8-bit D/A converter to control the voltage of the pulse train, so that the stretch strength can

be controlled with the accuracy of 1/128. The pulse train generated by the counter modulates

the frequency of the D/A converter’s output. After amplification, the modulated pulse train

eventually serves as the power supply of the electromagnet. For security reasons, the system

also needs four input ports to check the status of the power supply and test the connections

between the culture boards (inside the incubator) and the control system (outside the incubator).

To build up this control system, 33 out of 68 ports of the PCI-6601 have been used.

Figure 2.7 Port Allocation of DAQ

27

Page 42: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

2.4.2 Definition of the Stretch Plan

There are different methods to add mechanical stimulation during a tissue culture process.

Researchers can design a specific stretch plan according to the specific requirement of a tissue

culture process. For instance, in the first 24 hours, we may let the cells grow in a static

environment without any mechanical stimulation. This period of time is necessary for the cells

to recover from isolation procedure and enable attachment and spreading on the scaffold. In the

next 24 hours, we may add mechanical stimulation with low frequency and strength. After

several days of light stimulation, we may gradually increase the stretch frequency and strength.

Additionally, different parameters may be used for each of the three stretch boards, allowing

researchers to study the influence of specific parameters.

Figure 2.8 Definition of a Stretch Plan

28

Page 43: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Fig. 2.8 shows a detailed stretch plan. A stretch plan can have several durations of stretch

with different stretch parameters (sessions). It can satisfy different stretch patterns, such as

continuous cyclic stretch and intermittent cyclic stretch. The number of the sessions and the

proportion of the burst time and rest time in a session are determined purely by the software.

Each session has its own stretch pattern. Each pattern includes two parts, burst time (stretched

time) and rest time (static time). For a continuous cyclic stretch, the rest time is set to be zero.

In each burst time, three parameters of the pulse train need to be defined: voltage (stretch

strength), frequency (stretch frequency), and ON time (duty cycle). In a period, ON time

controls the stretch holding time of a period; OFF time determines the relaxed state between

two periods. The parameters are set in the Setup Form of the application software, and fulfilled

according to the above port allocation. A stretch pattern can be performed once or

continuously.

2.5 Design of Application Software

The function of the control software is to provide a user-friendly interface where all of the

stretch parameters can be characterized and monitored. As an improvement over Liu’s

apparatus [66], the new apparatus does not limit the number of sessions for each stretch plan. It

gives researchers more flexibility to perform long-term (>2 days) study of tissue culture, which

may need several different stretch sessions. In addition, each stretch plan can have its own

stretch strength instead of three stretch plans sharing the same stretch strength. The application

software has two major forms: Setup Form and Main Form. Users design a stretch plan and

characterize each parameter from the Setup Form, however, trigger, monitor, or stop the

specifically defined stretch plans from Main Form.

29

Page 44: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

2.5.1 Main Form

Main form is the main user interface of the apparatus. It monitors the status of the three

stretch boards. The form is divided into three sections, one for each of the stretch boards, shown

in Fig. 2.9. The left side of each board shows the stretch information, including Sessions, Status,

Strength, ON Time, Burst Frequency, and Cycle Type. Although the information is displayed

here, it cannot be modified from this form, since the Setup Form is the only place to define the

parameters. The right side of the board shows the stretching time and graph with specific

strength and frequency. At the right side of the form, there are four buttons: Run Board1, Run

Board2, Run Board3, and Run All. These buttons are used to trigger each board’s stretch plan.

The three boards can be triggered separately or simultaneously. After starting a stretch plan, the

button’s function changes to Stop Board, so that the four buttons will change to: Stop Board1,

Stop Board2, Stop Board3 and Stop All. The operator can stop any of the boards by clicking the

relevant button.

Figure 2.9 Main Form

30

Page 45: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

2.5.2 Setup Form

The Setup Form is the only entry point where users can modify the stretch parameters. There

are three major list groups in the form: Plan List, Plan Details, and Select Plans. In the Plan List

group, several defined plans are listed. Users can add, delete or modify the plan using the

control bar underneath. In the Plan Details list group, the details of the currently selected Plan

are shown. For example, in Fig. 2.9, the currently selected plan number is 1, and the Plan

Details group lists all the stretch sessions of that plan. Users can set the parameters such as

BurstTime, RestTime, BurstFrequency, Strength, and ONTime in the group. In the Select Plans

list group, users determine the stretch plan for each stretch board by selecting the defined

stretch plans. The Setup Form gives users the maximum flexibility to design a stretch plan with

the specific stretch time, pattern, frequency and strength.

Figure 2.10 Setup Form

31

Page 46: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

2.5.3 Control Flow Chart

Fig. 2.11 demonstrates the control flow of the software. When the application software is

activated, the control thread starts working. First, it controls three boards to load their own

stretch plan. Then periodicly checks the statuses of three boards, Board1, Board2, and Board3.

When the thread finds any of the boards need triggering, it will control the corresponding board

to perform burst or rest action according to the defined stretch plan. When more than two

objects are triggered, the control thread will determine the next action according to its own

timeline. The major function of the control thread is coordinating the action of each stretch plan.

Figure 2.11 Control Flow Chart of the Software

32

Page 47: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

2.6 Conclusions

This chapter describes the development of a uniaxial biostretch apparatus to culture

engineered tissue patches under well defined mechanical conditions. The apparatus provides a

versatile platform to study the influence of the mechanical stimuli on cell culture. The new

apparatus has demonstrated several technical and practical advantages to existing ones:

It allows tissue patches to be cultured in a standard culture dish and provides two

kinds of stretch methods, asymmetric stretch and symmetric stretch.

It can deliver a wide range of precision consistent strains and stretch frequencies.

The use of uniquely designed scaffold mounting trays, with non-contact force, allows

researchers to use standard Petri dishes, which significantly simplifies the assembly

process, and ultimately reduces the potential for contamination.

The device is easy to sterilize and clean, which makes delivering mechanical

stimulation easy and simple.

With this device, various sizes of culture tissues (35 and 60 dishes) are available.

A prototype of the apparatus, shown in Fig. 2.12, was constructed and tested in a research

lab at the University Health Network in Toronto. Fig. 2.13 shows the assembled culture dish

mounted between a pair of electromagnets.

Since the device utilizes non-contact magnetic force as its driving force, the effect of the

magnetic field will be studied in the following chapter. The calibration of the device and the

effects of mechanical stimulus will be discussed in Chapter 4.

33

Page 48: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Figure 2.12 Assembled Bioreactor System

Electromagnet Culture Dish Gelfoam Stopper Steel Bar

Figure2.13 Assembled Culture Dish with Electromagets

34

Page 49: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Chapter 3

Effect of the Time-varying

Electromagnetic Field on Cell Biology

The objective of this chapter is to study the effect of the time-varying electromagnetic field

generated by a horseshoe-shaped electromagnet, which is utilized in the stretch device described

in Chapter 2. The magnitude and distribution of the electromagnetic field is modeled and

simulated with COMSOL Multiphysics 3.3, which also provides an effective tool to

quantitatively study the applied magnetic force in Chapter 4. The magnetic field influence on

cell biology is investigated from the aspects of cell proliferation and morphology.

In this chapter, section 3.1 briefly introduces the current state of the understanding of

magnetic field influence on cell biology. Section 3.2 presents the modeling procedure of the

time-varying electromagnetic field. Section 3.3 verifies the model through physical

experimentation. Section 3.4 describes the study of magnetic influence on cell biology through

two bio-experiments. These experiments are performed in a research lab at the University

35

Page 50: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Health Network in Toronto. Section 3.5 illustrates the heating effects of the electromagnet on

cell proliferation. Conclusions are discussed in section 3.6.

3.1 Introduction

The uniaxial cyclic stretch apparatus that was described in Chapter 2 uses electromagnetic

force as the driving force. The primary advantage of magnetic force is that it can directly apply

controlled force on an engineered tissue patch without introducing a pathway of infection.

Besides acting as the driving force to stretch an engineered tissue patch, the magnetic force has

also been used in other recent studies along with magnetite nanoparticles for cell seeding [73],

gene transfection [74], and constructing multilayered cell sheets with heterotypic cocultured

cells [75] in tissue engineering. This promising technique developed a new branch of

engineered tissue called magnetic force-based tissue engineering [75], [76, 77]. However, the

biological effects of the magnetic field during tissue culture have not yet been fully

investigated.

The most enigmatic external source of stimulation in tissue engineering is the magnetic

field, since all biological systems are exposed to magnetic fields from the earth or other sources.

Institutions such as EMF-RAPID (Electric and Magnetic Fields Research and Public

Information Dissemination) and NIEHS (National Institute of Environmental Health Sciences)

started to study the influence of magnetic fields on the human body in 1992 [78]. It was

reported that although magnetic fields maybe have a negative effect on the human body, that

effect, if any, is considerably small [31]. To date, the effects of magnetic fields to human beings

have been continually studied in a wide range, from the chemical reaction models [79] to

physical therapy [80]. For example, Pulsed Electromagnetic Fields are utilized to heal fracture

36

Page 51: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

non-unions and treat some bone-related diseases [80, 81], although the specific molecular

mechanisms are not fully understood. However, in this dissertation, I will focus on investigating

the magnetic influence on cell biology.

Literature review shows that both static [29-31, 82] and time-varying [32, 83-85] magnetic

fields can influence tissue culture at the cellular level. Miyakoshi [86] concluded that although a

strong static magnetic field can induce orientation phenomena in cell culture, a static magnetic

field alone does not have significant effects on the basic properties of cell growth and survival,

regardless of the magnetic density. However, with the development of magnetic force-based

tissue engineering, researchers recently have drawn a growing attention to the influence of the

electromagnetic field (EMF), especially to the extremely low frequency electromagnetic field

(<300Hz) because the frequencies of most biological processes are under the range of extremely

low frequencies [85]. Scientists found that compared to a static field, a small pulsed radio

frequency field is more effective in the Ca2+/CaM -dependent myosin phosphorylation system

[86, 87]. Nevertheless, the studies of EMF are mainly focused on connective tissues, bone and

cartilage tissues. For example, it has been proven that EMF (75 Hz, 2,3 mT) can promote

anabolic activites and proteoglycan synthesis in bovine articular cartilage explants [84].

Furthermore, Fassina et al. [83] determined that electromagnetic stimulation with 2 mT flux

density and 75 Hz frequency accelerated SAOS-2 cells’ proliferation and increased calcium

deposition. Komazaki and Takano [88] examined the influence of a low frequency EMF (50 Hz,

5-30 mT), and found that EMF specifically increased the [Ca2+]i of gastrula cells, therefore,

accelerating the rate of morphogenetic cell movements during gastrulation.

These facts motivate us to explore the effects of time-varying electromagnetic field

generated by the stretch device. The magnetic field generated by the apparatus should be weak

37

Page 52: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

and at a low frequency (<300Hz) according to the above reviews. But the distribution and

magnitude of the field is unknown, and the effect on cell biology is unclear. The following

sections will illustrate the magnitude and distribution of the flux density B generated by the

electromagnetic field, the factors affecting the field, a potential side effect of the field, and the

cellular response to the field.

3.2 Modeling the Time-varying Electromagnetic Field

In this section, the time-varying electromagnetic field generated by the stretch apparatus is

modelled. The geometrical model is based on a horseshoe-shaped electromagnet MCI-3404

(Mci Limited Co., ON, CA) with the external dimensions 50×48×26mm, which can provide a

time-variant electromagnetic field when excited by a harmonic current. We are interested in not

only the distribution and magnitude of the magnetic flux density, but also the relationship

between the exciting current I and the magnetic flux density B. Theoretically, the exciting

current I determines the distribution and magnitude of density B. On the other hand, density B

controls the magnitude of the magnetic force F. The model is studied in two dimensions,

especially along the horizontal (X) direction because the horizontal magnetic force is served as

longitudinal stretch force, while Y direction is the perpendicular direction on the surface of the

scaffold. Instead of implementing home developed code, which is often time consuming and

requires considerable resources, a commercial finite-element package, COMSOL Multiphysics

3.3 [89], was utilized. All the simulations were performed with the package.

3.2.1 Electromagnetic Theory

Maxwell’s equations can be written as

38

Page 53: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

0

B

D

t

BE

t

DJH

(3.1)

The first equation is also named as Ampere’s Circuital Law:

t

DJH

(3.2)

Equation (3.2) states that the curl of the magnetic field intensity H

is equal to the sum of

the current density J

due to the flow of charges and the time derivative of the electric flux

density D

[90]; related variables are shown in Table 3.1.

Table 3.1 Nomenclature

H

magnetic field intensity E

electric field intensity

B

magnetic flux density D

electric flux density

A

magnetic vector potential V electric scalar potential

M

magnetization vector P

electric polarization vector

F

magnetic force J

volume current density

q free charge eJ

external volume current density

permeability permittivity

conductivity angular frequency

d thickness in z direction density

T stress tensor p air pressure

39

Page 54: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

In this study case, the electromagnet is supposed to work under low frequency, less than

50Hz. The wavelength for a 50Hz signal is 6000km. On the other hand, the structure size of the

magnets is around 50x50mm, which means the length of the geometry is only a very small

fraction of the exciting wavelength. In this slow variation case, the electromagnetic field is

assumed to be a quasi-static field with 0 tD

. Therefore, Ampere’s Circuital Law (3.2)

simplifies to:

JH

(3.3)

Considering the constitutive relation eJEJ

and the Lorentz force equation

BvEqF

, equation (3.3) for quasi-static electromagnetic field can be extended to:

eJBvEH

)( (3.4)

Combining the constitutive relation, HMHB r 00 )(

, the definition of the

magnetic potential AB

, and electrical potential t

AVE

, Maxwell-Ampere’s law

can be rewritten as:

eJVAvMAt

A

)()( 1

0 (3.5)

Since the exciting current is a harmonic signal, this study case is a time-harmonic case. In

the time-harmonic case, Ampere’s law includes a displacement current Dj , with the

constitutive relation PED 0 , equation (3.5) can be rewritten as following:

PjJVjAvMAAj e

)()()()( 01

002

(3.6)

40

Page 55: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

3.2.2 Electromagnetic Model Analysis

Since our objective is to study the field distribution and the flux density B of a quasi-

static magnetic field, we can ignore the coupling between the electric and magnetic field, and

only consider the induction currents that are relevant to the magnetic potential. Moreover, as we

are mainly interested in the B distribution on a horizontal surface (x-y direction), we can study

the magnetic field in two dimensions to simplify the modeling process. In the two-dimensional

case, with only the consideration of induction current, equation (3.6) can be simplified as:

ezzzr JLVAvA

/)()( 110 (3.7)

Equation (3.7) is utilized in the electromagnet subdomain analysis [89]. To perform the

simulation, we make the following assumptions:

The current source is a time-harmonic sinusoidal current with magnitude of 0.4A and

frequency of 1Hz.

The entire coil is considered as a whole block with a constant external current density

corresponding to the current in each single wire. This assumption is more efficient than

modeling each turn of the coil as a separate wire.

The eddy current between the turns in the coil is negligible.

The material of the core is a solid soft iron, ignoring the laminations among the sheets

of the iron core

Based on the above assumptions, we choose the 2D, time-harmonic, quasi-static,

perpendicular induction current, and vector potential application mode in the COMSOL AC/DC

module. The complete set of geometrical and electrical parameters used for the simulation is

listed in Table 3.2 and 3.3. The meshed geometry of the electromagnet model is shown in Fig.

41

Page 56: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

3.1. In the graph, there are three subdomains: air, coil (copper) and core (soft iron). The bobbins

between the coil and the core are ignored since they do not significantly influence the magnetic

field analysis. The air space surrounding the feet of the magnet is 30mm, just half the size of a

60mm standard culture dish. Therefore, the graph represents half of a stretch sample; the other

side of the stretch sample is symmetrical. The geometrical dimensions and material properties

of the magnet were provided by the product company. The mesh generator partitions the

subdomains into triangular elements. The finite element model has 2716 elements and the

number of degrees of freedom is 5491. The size and density of the elements are allocated based

on the geometrical dimensions. The areas with more boundaries have smaller elements with

higher density, while the outside areas have larger elements with lower density. This allocation

can reduce the computational expenses and facilitate the calculation procedure.

Table 3.2 Geometrical Parameters

Parameters Expression Description

r_coil 0.2mm Radius of the coil wire

N 550 Turns of the winding

Ww 4mm Width of the winding

Lw 28mm Length of the winding

Wc 12.5mm Width of the core

Lc 38mm Length of the core

w 1mm Width between the core and winding

42

Page 57: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Figure 3.1 Meshed Electromagnet Model

(a)

43

Page 58: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

(b)

(c)

Figure 3.2 Distribution of Magnetic Flux Density

(a) Bx component, (b) By component, (c) Bn component

44

Page 59: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Table 3.3 Electrical Parameters

Parameters Expression Description

I_coil 0.4A Current of the coil

J_coil I_coil*N/(Ww*Lw) External current density

5.99e7S/m Conductivity of the copper wire

As the exciting source is a time-harmonic current, the magnetic field distribution of the

study space varies with time. The magnetic poles of the electromagnet alternate between the

two feet during a period. Fig. 3.2 (a), (b), and (c) exhibits the x component, y component and

normal distribution of the maximum flux density in a period at phase zero, respectively. The

units for x and y coordinates are meter. Since the magnetic force along the x direction serves as

the stretch force, the x component of flux density, denoted as Bx, is the main study object. In the

Fig. 3.2 (a), the red arrows act as magnetic field lines representing the magnetic field

distribution. The magnetic south pole (orange) of the horseshoe electromagnet is at the lower

foot and the magnetic north pole (blue) is at the upper foot. The distribution shows that the

magnetic flux density Bx has greater magnitude in the soft iron core than in the air space. The

maximum Bx in the soft iron is 0.204T at the inner corners. However the amplitude of Bx

dramatically reduces to 0.041T at the boundary of the iron core and air space. Except for the

two feet areas, the flux density B in most air space is near zero (green). For example, the

magnitude of the Bx is only 0.015T at 5mm distance of the edge, 0.012T at 10mm distance, and

0.009T at 15mm distance.

45

Page 60: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Figure 3.3 Magnetic Flux Density Bx along y Direction

Since the magnitude of the Bx determines the magnitude of the magnetic force along the x

direction, the effective area of the non-contact magnetic force is limited. In order to apply non-

contact magnetic force, the magnetic rods should be placed within 10mm from the edge of the

electromagnet. Fig. 3.3 shows the distribution of Bx along the y direction (16mm long) at 5mm

distance from the edge of the magnet feet. The plot demonstrates that the magnitude of Bx at

that location is very small, with a maximum value of 0.015T. However, It also shows that the

absolute value of the density Bx is almost symmetrical along the y direction symmetrical axial

at y=0.151, which means if a 16mm long magnetisable rod is put into the field along y

direction, the two sides of the rod will experience symmetric force. This plot will be compared

with Fig. 4.9 in Chapter 4 section 4.6.3. Besides being affected by the flux density, the

magnitude of the force is also determined by the distance between the magnet and the

magnetisable material and the material’s property, which will be discussed in Chapter 4.

46

Page 61: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Although we focus our investigation on the distribution of magnetic flux in the x direction,

we still need to study the normal magnetic flux density B because it is the closest parameter

(2D) to the results of the physical test (3D). For the 2D model, the normal magnetic flux density

B was defined as the total magnitude of the X component and the Y component. The distribution

of the normal magnetic flux density is shown in Fig. 3.2 (c). The simulation results demonstrate

that the normal magnetic flux density B is 0.082T directly beside of the horseshoe magnet. This

result will be compared with the experimental results in the next section.

3.3 Experimental Verification of the COMSOL Model

Serial physical experiments were implemented to verify the COMSOL model of the

horseshoe-shaped electromagnet. In the experiments, the magnet was excited by a pulsed 0.4A

current train with 1Hz frequency, and 50% duty cycle. A Gauss meter with integral Hall Effect

Probe (Model 2010, Magnetic Instruments Inc.) was utilized to test the magnetic flux density B

[91]. The sample time for each test point is one minute. The verification and simulation results

are compared from the aspects of distance, exciting frequency and current.

3.3.1 Flux Density vs. Distance

The first experiment is conducted to verify the relationship between the flux density B and

distance. The corresponding density B was recorded along the x direction every 5mm from each

magnet’s foot. Fig. 3.4 (a) and (b) represent the density B along x direction, at y=0.0143 (south

pole) and y=0.0159 (north pole), respectively. The red curves in the plots are the simulation

results of Bnorm from the COMSOL model. The blue curves are the testing results of flux density

B, which were obtained from the average of twenty peak magnitudes during the sample time.

47

Page 62: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

The y axis of Fig. 3.4 shows the absolute magnitude of B for at the south pole (a) and north pole

(b); while the x axis shows the distance from the test points to the edges of the electromagnet.

The graphs demonstrate that the south pole and the north pole possess the same trends.

The trend of B in these plots shows good correlation between the test and simulation results.

Moreover, the field distribution coincides with the Bio-Savart law. According to the Bio-Savart

law, the magnetic flux density produced at a point from an element of length ld

of a

filamentary wire carrying a steady current I is expressed as:

20

4 R

alIdBd R

(3.8)

where is an element of length in the direction of the current, is the unit vector

pointing from to the point, and R is the distance between the point and the current element

. The equation shows that flux density B is proportional to the current I and the reciprocal of

the distance square [90]. However, most of the testing results are, on average, higher than the

simulated results in the ambient area (distance<0.015m) of the magnet except for the zero

position, and lower that the simulated results in the remote area. This phenomenon will be

discussed in the next section.

ld

ld

Ra

ld

48

Page 63: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Simulation vs Testing Results(South Polar y=0.159)

0

0.01

0.02

0.03

0.04

0.05

0.06

0.07

0.08

0.09

0 0.005 0.01 0.015 0.02 0.025 0.03

Distance [m]

Magnetic Flux Density [T]

simulation

testing

(a)

Simulation vs Testing Results(North Polar y=0.143)

0

0.01

0.02

0.03

0.04

0.05

0.06

0.07

0.08

0.09

0 0.005 0.01 0.015 0.02 0.025 0.03

Distance [m]

Magnetic Flux Density [T]

simulation

testing

(b)

Figure 3.4 Simulation Results vs. Testing Results of Magnetic Flux Density

(a) South Pole at y=0.0159, (b) North Pole at y=0.0143

49

Page 64: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

3.3.2 Flux Density vs. Exciting Frequency

In order to testify the relationship between flux density and the exciting frequency, the

probe of the Gauss meter was fixed at the lower corner of the north pole (around y=0.0159)

0.005m away from the magnet. The flux density was recorded as the average of twenty peak

magnitude during the sample time, with pulse train frequencies from 1Hz to 5Hz. The test

results show that the flux density ranges from 372.0G to 386.5G, with an average of 375.8G,

and a standard deviation of 5.63 (Table 3.4). The results demonstrate that frequency variations

in the lower range (1~5Hz) have little effect on the magnitude of flux density, which coincides

with the simulation assumption that the eddy current between the wires in the coil is negligible.

Due to the assumption of eddy current, the simulation results show that exciting frequency does

not affect the flux density.

Table3.4 Magnetic Flux Density vs. Exciting Frequency

Frequency

[Hz] 1 2 3 4 5

Flux Density

[G] 386.5 374.7 374.5 372.0 377.0

STD DEV 5.63

50

Page 65: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

3.3.3 Flux Density vs. Exciting Current

The relationship between the magnitude of the exciting current and the flux density B was

also investigated in this section. In the interface of the apparatus (Fig. 2.10), only the stretch

related parameters are presented to the users, such as stretch frequency, duty cycle, and strength.

However, all of these mechanical parameters are fulfilled by defining related electrical

parameters. For example, the magnitude of stretch strength is directly controlled by exciting

current. When the stretch strength is set as 95%, the magnitude of the exciting current is 0.4A,

while when the stretch strength is set as 50%, the current is 0.2A.

Exciting Current vs Flux Density

0

50

100

150

200

250

300

350

400

10 20 30 40 50 60 70 80 90

Exciting Current (%)

Magn

etic

Flux

De

nsity

[G]

Figure 3.5 Exciting Current vs. Flux Density

In order to study the relationship between the exciting current and the flux density, the probe

of the Gauss meter was fixed at the lower corner of the north pole at the distance of 0.005m

from the magnet. The flux density B was recorded when the stretch strength increases from10%

51

Page 66: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

to 90% by every 10%. Fig. 3.5 shows the test results. It illustrates that the flux density decreases

linearly as the strength of the exciting current decreases. The standard deviation for each test

point is very small, ranging from 0.09 to 1.40. However, the simulation results demonstrate that

flux density is proportional to the exciting current. The test results agree with the simulation

results that the flux density B is proportional to the exciting current I in the low frequency

range. Therefore, the test results verified the design principle that the magnetic flux density and

force can be controlled by controlling the exciting current.

3.3.4 Comparisons of Simulation and Experiment Results

A series of physical experiments are implemented to test the relationships between the

flux density B and the distance, the exciting current and frequency. Although the experiment

results agree with the simulation results, there are still some differences. These differences will

be discussed in the following:

Waveforms

For testing, the exciting current is a pulse train with 1Hz frequency, 50% duty cycle, and

0.4A amplitude. For simulation, we used a time-harmonic sinusoidal current with the same

frequency, duty cycle, and amplitude, but with different shapes. In experiments, the magnetic

field generates when the pulse train switches from OFF time to ON time, and disappears when

the pulse train switches from ON time to OFF time, but the magnetic polarities of the

electromagnet stay the same during the electrifying time. However, for modeling, as the

exciting current is a time-harmonic sinusoidal signal, the magnetic field exists in the testing

space during the whole period, while the magnetic polarities of the magnet alternate during

52

Page 67: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

every half period. Although the waveforms of the simulation and the testing are different, the

maximum amplitudes of the two waveforms are the same. Therefore, the simulation results

shown in Fig. 3.2 (the maximum flux density in a period) can represent the real distribution.

Dimensions

When we take a close look at Fig 3.3 (a) and (b), we can see although the testing results

and the simulation results share the same trend, there are three clear differences: (1) the testing

result is lower than the simulation result immediately adjacent to the magnet; (2) except for that

first point, the testing results are slightly higher than the simulation results close to the magnet

(distance<0.015m); (3) the testing results are slightly lower than the simulation results further

away from the magnet (0.015m<distance<0.03m).

The first difference may be caused by the thickness of the test probe. Since the probe of

the Gauss meter has a thickness, the first testing result is not the real density B at zero distance

(the boundary of between the core and the air). There is a small gap between the electromagnet

and the test probe. This may be the reason why the first testing result is lower that the

simulation result as density B decreases dramatically outside of the core. For the second

difference, the simulation results are obtained from a two-dimensional model, while the

experiment results are based on a real three-dimensional magnet. When the probe is positioned

at a specific place, it senses the flux density B from three dimensions (including Bz, density

along z direction); however, the simulation model did not consider this parameter. Therefore,

most of the testing results are slightly higher than the simulation results in the ambient area of

the magnet. With the increase of distance, density B radiates toward different directions, and the

difference between the simulation and testing gets smaller. For the third difference, when the

53

Page 68: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

distance from the magnet is greater than 0.02m, the flux density is much lower than 0.01T. In

this situation, the sensitivity of the probe and the test errors greatly influence the experimental

results.

Despite the above differences, the testing results and the simulation results still agree

with each other well. Therefore, the modeling provides a useful tool to optimize the density

distribution and to calculate the non-contact magnetic force.

Figure 3.6 Magnetic Flux Density Distribution of the Stretch Device

(1-Electromagnet, 2-Shadow Area, 3-Scaffold, 4-Steel Bar)

The simulation results show that the spatial distribution of density B is non-uniform. The

ambient area of the magnet (distance<0.01m) has stronger density than other space, shown in

Fig. 3.4. We can apply this distribution feature to the designed apparatus. When a culture dish is

positioned between a pair of the modeled magnets (1), the shadowed areas in Fig. 3.6 (2) have

stronger density. If a pair of magnetic bars (4) is secured in these areas and mounted at the two

ends of an engineered tissue patch, the patch can experience a controllable stretch. Meanwhile,

54

Page 69: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

the tissue patch (3) is exposed to a very weak magnetic field. The density B is less than 0.02T,

over a distance of 10mm (the edge of the shadow area). The uniaxial stretch apparatus described

in Chapter 2 is designed based on the above theory. The stretching force can be calculated

directly by the modeling, since the flux density has been verified by the physical experiment.

Hence, the time-varying magnetic field offers a promising approach to apply controlled

mechanical conditions to the tissue culture process via the non-contact magnetic force.

3.4 Effect on Cell Biology

The above analysis shows that magnetic field distribution along the x direction is non-

uniform. The maximum magnitude of the density B is about 0.08T, while the density reduces to

only 0.04T near where the steel bar is secured. The density B in most of the tissue culture area

(Fig. 3.6 (3)) is very low (B<0.01T). Two bio-experiments were performed to test the influence

of the magnetic field on cell biology. One experiment tests the cell proliferation and the other

investigates the cell morphology.

3.4.1 Cell Proliferation

Cells and Cell Culture

To test the magnetic influence on the cell proliferation of engineered tissue patches, we

cultured rat smooth muscle cells on a gelatine sponge scaffold. The Gelfoam was fully soaked

in culture medium and incubated for three days before cell seeding. The mounting trays were

gas sterilized and placed in 60mm Petri dishes. The soaked scaffold (without clamps) was then

secured in the center groove of the mounting tray. Cell suspension (6×106 cells/250μl culture

55

Page 70: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

medium) was then seeded into each Gelfoam scaffold (40x20x10 mm). The cell seeding

procedure followed previously established methods [6], and the culture medium was gradually

added until the scaffold submerged.

The assembled dishes were placed at 37℃ in a humidified incubator with 5% CO2 and

95% air for twenty-four hours for cell adhesion. Three samples were subsequently placed on the

newly designed bio-stretch apparatus with the modeled magnets. Since no stainless steel bars

were attached at the ends of the scaffold, the engineered tissue patches were only affected by

the time-varying magnetic field. The field was intermittently applied to the patches, alternating

between 15minutes of stretching time and 30minutes of rest time. The exposure time (15

minutes) referred previous studies [92, 93]. The frequency and strength of the exciting current

were set to be 1 Hz and 0.4A, respectively. In order to obtain a stronger magnetic field, the

strength of the current was set very high. However, the control groups were placed in the same

incubator without being exposed to the time-varying magnetic field. The culture medium was

changed every other day.

DNA Assay

To determine the number of cells on each engineered tissue patch, the total DNA was

extracted using the DNeasy Blood&Tissue kit (Qiagen) following the manufacturer’s

instructions on days 3 and 14 after exposure to the magnetic field. The cell-seeded gelatine

sponge was lysed using the lysis buffer and the lysate was loaded onto the DNeasy mini spin

column. Total DNA was then eluted into TE buffer and read at 260 nm. Table 3.5 compares the

average and standard deviation of DNA at a magnetized condition with the control group from

the two time points (Day 3 and Day 14).

56

Page 71: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Results

Table 3.5 shows that the control group had a slightly higher number of cells than the

magnetized samples. However, the T-test results for two groups show that there is no significant

difference among these groups because P>>0.05. The statistic power analysis was also

performed by SigmaStat software. The calculation results show that the powers of these two

groups are 1.0 and 0.99, respectively, which means there is no significant difference between

the magnetized group and the control group. Besides the analysis results, condensation was

observed on the lids of the magnetized group, which suggests that the temperature of the

magnetized group was different from that of the control group.

Table 3.5 Cell Numbers of Magnetized and Control Samples

3 Days 14 Days

Magnetized

(n=3)

Control

(n=3)

Magnetized

(n=3)

Control

(n=3)

DNA Ave

(μg/ml) 56.2 68.9 69.7 72.7

DNA Std 1.36 0.93 0.33 0.57

T-Test P=0.25 P=0.47

57

Page 72: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

3.4.2 Cell Morphology

Cells and Cell Culture

To evaluate the magnetic influence on cell morphology, we used rat smooth muscle cells

(SMC) and mouse bone marrow stromal cells (BMSC) as cell sources. The cells were seeded

directly to a 35mm culture dish. Cell suspension with 0.2×106 cells was seeded into each dish.

After culturing 24 hours, three samples were placed into the same defined electromagnetic field

as section 3.4.1. The control dishes were placed into the same incubator without the magnetic

field. The lag period was intended to provide the cell enough time to recover from isolation

procedure and enable attachment.

Results

We learn from the modeling that the magnetic field distribution of the electromagnet is

non-uniform. Therefore, we only observed the cells within 10mm to the magnet, where the

magnetic density is greater than 0.01T. Fig. 3.7 and Fig. 3.8 show the cell morphology of SMC

and BMSC, respectively. The images were taken by a NIKON Ti-S microscope for phase

contrast. The photographs are grouped into two columns. In Fig 3.7 columns (a) and (b),

comparison is made on the morphology of SMC after 48 hours and 72 hours; similarly, in Fig

3.8 columns (a) and (b), comparison is made on the morphology of BMSC after 48 hours and 72

hours, respectively. In each column, the control image is on the top panel while the image of the

magnetized sample is on the bottom panel. However, no significant difference in cell

morphology was observed for either cell type at the two time points by the observers who were

blinded to the study methods, which means that the time-varying magnetic field with the density

of 0.08T at zero distance has little effect on cell morphology.

58

Page 73: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

(a) (b)

Figure 3.7 Cell Morphology of SMC

(a: 48hrs-SMC, b: 72hrs-SMC, magnification: 200x)

59

Page 74: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

(a) (b)

Figure 3.8 Cell Morphology of BMSC

(a: 48hrs-BMSC, b: 72hrs-BMSC, magnification: 200x)

3.5 Heating Effect of the Electromagnet

When performing the bio-experiments, condensation was observed on the lids for the

magnetized group. This drew our attention to a possible side effect of heating when using an

electromagnet, a problem because temperature is an important parameter in tissue culture.

Normally, the temperature remains at 37℃ in the incubator. However, when we simulated the

electromagnet model, the heat dissipation problem was ignored, so that the increase of the

temperature cannot be predicted. Therefore, the variation of the temperature was tested by

experimentation.

3.5.1 Experiment Results

Under Room Temperature

The Hall Effect Probe of the Gauss meter (Model 2010, Magnetic Instruments Inc.),

which we used to verify the flux density B, can also test temperature. When we studied the

60

Page 75: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

relationship between flux density B and exciting frequency f, the probe was fixed at 5mm from

the edge of the magnet. The probe not only sensed the variation of the flux density, but also

recorded the variation of the temperature. Fig. 3.9 illustrates the temperature variation after the

electromagnet was excited. The electrify parameters are set as 1Hz frequency, 95% strength,

and 50% duty cycles. The plot shows that the temperature increased 1.3℃ after 9 minutes

exciting, and remained stable.

Temperature vs Time

25

25.5

26

26.5

27

0:00

:00

0:00

:27

0:01

:38

0:02

:06

0:03

:17

0:03

:44

0:05

:08

0:05

:35

0:06

:02

0:07

:08

0:07

:35

0:08

:46

0:09

:13

0:09

:40

Time [h:mm:ss]

Tem

per

atu

re [

℃ ]

Temperature

Figure 3.9 Temperature vs. Exciting Time under Room Temperature

Under Incubator

The above test was performed under room temperature, while the real working

environment for the electromagnet is in the incubator. Therefore, a precision thermometer UNI-

T (UT328, UNI-TREND technology Inc., Guangdong) was utilized to test the temperature

variation under working conditions. The thermometer has two thermocouple input channels.

One thermocouple (t1) was fixed in the middle of a pair of electromagnets (the center of the

61

Page 76: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

stretch board) to sense the temperature variation caused by the magnets; another thermocouple

(t2) was fixed at the metal shelf in the middle of the incubator to sense the temperature of the

incubator. After the test points fixed, the stretch board was electrified. The stretch plan was set

the same as in the bio-experiments in section 3.4 (electrifying 15 minutes and resting 30

minutes, 1Hz, 95% strength, 50% ON time). The thermometer was placed in the incubator for

11 hours. After three hours and forty minutes (five periods of the stretch plan) for satiability, it

started to record the two test points’ temperature every minute. Fig. 3.10 records seven and half

hours (10 successive periods) data.

Fig. 3.9 and Fig. 3.10 show clearly that the temperature increases when the

electromagnet is electrified. Both experiments verified that the local temperature around the

magnets is about 1.0℃ higher than the environment. Fig. 3.9 emphasizes the short time

increase trend under room temperature. The temperature increased 1.3℃.

Temperature vs Time

35.5

36

36.5

37

37.5

38

38.5

1:0

7:4

7

1:2

6:4

7

1:4

5:4

7

2:0

4:4

7

2:2

3:4

7

2:4

2:4

7

3:0

1:4

7

3:2

0:4

7

3:3

9:4

7

3:5

8:4

7

4:1

7:4

7

4:3

6:4

7

4:5

5:4

7

5:1

4:4

7

5:3

3:4

7

5:5

2:4

7

6:1

1:4

7

6:3

0:4

7

6:4

9:4

7

7:0

8:4

7

7:2

7:4

7

7:4

6:4

7

8:0

5:4

7

8:2

4:4

7

Time [hh:mm:ss]

Tem

pe

ratu

re [

℃ ]

t1

t2

Figure 3.10 Temperature vs. Exciting Time in incubator

62

Page 77: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Fig. 3.10 shows the temperature fluctuation in the incubator. Compared with the baseline

t2, t1 shows more severe fluctuation. During this test period, the temperature for t1 ranges from

37℃ to 37.9℃, with an average of 37.4℃; while the temperature for t2 ranges from 36.℃

to 36.7℃, with an average of 36.4℃. The temperature influence on cell biology was

investigated in the following section.

3.5.2 Bio-experiment Verification

As the experiment results show that the electromagnets causes the local temperature

fluctuated during electrifying, while temperature is an important culture parameter for cell

growth. The biological influence of temperature needs to be clarified.

Cells and Cell Culture

To test the influence of temperature on the cell proliferation of engineered tissue

patches, we again used the gelatine sponge (Gelfoam: Pharmacia & Upjohn Co., Kalamazoo,

MI) as the scaffold and rat smooth muscle cells as the cell source. The experiment followed the

same cell-seeding protocol described in section 3.4.1. However, we used only half the number

of the cells. As the experiment was performed in 35 mm Petri dishes, the size of the scaffold

was only a half of the experiment in section 3.4.1. Hence, the cell suspension (3×106

cells/125μl culture medium) was seeded into each Gelfoam scaffold (20x20x10 mm). The cell-

seeded patches were then assembled into 35 mm culture dishes. All the samples were first

placed into a humidified incubator with 5% CO2 and 95% air at 37℃ for twenty-four hours.

63

Page 78: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

The test and control groups were then placed into two different humidified incubators with 5%

CO2 and 95% air for three days. The temperature for the test group was 38℃, while the control

group’s temperature was 37℃.The number of samples in each group was three.

DNA Assay

A DNA assay was again used to determine the number of cells on each engineered tissue

patch, following the same procedure as in section 3.4.1. Table 3.6 compares the average and

standard deviation of DNA at a high temperature condition (38℃) with the control group

(37℃).

Table 3.6 Cell Numbers at 38℃ and at 37℃

High Temperature 38℃

(n=3)

Control 37℃

(n=3)

DNA Ave

(μg/ml)

48.2 52.0

DNA Std 2.30 5.26

T-Test P=0.32

Results

Table 3.6 shows that the total cell number of the control group is slightly higher than the

group cultured under higher temperature. However, the T-test result for two groups shows that

there is no significant difference among these groups because P>>0.05. This result is in

64

Page 79: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

accordance with the result obtained from section 3.4.1. The phenomenon may explain that one

degree of temperature difference does not significantly affect cell proliferation. In section 3.4.1,

the magnetized group was not only exposed to the time-varying magnetic field, but also affected

by the temperature variation. However, these two parameters also do not significantly affect the

experiment results.

3.5.3 Effect Factors

When an external electric current passes through an electromagnet, the heat conductivity

is known to be coupled with the electromagnetic field simulation. Therefore, the temperature

variation was examined.

There are several physical parameters affecting the heating problem of a pulsed

electromagnet, such as permeability µ, conductivity , and permittivity Ɛ. Heat is generated not

only in the coils but also from the core [65-68]. For example, when the iron core is exposed to a

pulsed magnetic field, the permeability µ of the material will change. The temperature

dependent parameter will introduce a phase shift between the magnetic flux density B and the

magnetic field strength H that causes the hysteresis pneumonia. Joule Heating can be called

resistance heating that is mainly heat loss of the coils. It is determined by the local current flux

and temperature-dependent conductivity. In a Joule Heating study, the heat transfer problem

couples with the electronic current problem. The calculation of the heat source is based on the

equation 2

VQ . However, the thermal energy in turn changes the electrical conductivity

of the coil. The temperature-dependent conductivity can be expressed as

))(1(1

00 tt , where 0 ( m ) is a reference resistivity at a reference temperature

65

Page 80: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

t0(K), and α is proportionality constant (K-1) for the temperature dependence. The temperature-

dependent conductivity causes the variation of the exciting voltage, which results in the

coupling between the temperature and electromagnetic fields.

In future modelling, a multiple physical aspects need to be studied. The model should

consider not only magnetic field, but also heat transfer. The above effect factors should be

considered into the modelling, which is useful to accurately describe the real phenomena. To

reduce the effect of temperature fluctuation in later experiments, there are two methods can take

into account. First, reduce the temperature of the incubator 0.5℃ lower than the previous set

number, which can effectively reduce the average temperature of the stretch group, but the

control group will also be affected. Another way is to place the electromagnets onto a metal

base such as aluminum, which can disparate the heating and not affected by the magnetic field.

Therefore, the local temperature of the electromagnet can lower down.

3.6 Conclusions

The time-varying electromagnetic field generated by the stretch device is quantitatively

studied in this chapter, both through simulation and experimentation. The magnetic field is

modeled by COMSOL multiphysics 3.3 and verified by the Gauss meter Model 2010. The

investigation shows that the distribution of magnetic flux density along the stretch direction (X

direction) Bx is non-uniform; the magnitude of the density is proportional to the exciting current

I; and the magnitude of the density is not related to the exciting frequency at low frequency

range (1Hz to 5Hz). The relevant biological experiments demonstrate that the weak, low

frequency magnetic field generated by the stretch device does not significantly affect cell

proliferation and cell morphology. However, the difference between the exam group and the

66

Page 81: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

control group is caused by the influence of both magnetic field and temperature field as the

electromagnet has heat dissipation problem.

67

Page 82: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Chapter 4

Effect of the Uniaxial Cyclic Stretch on

Cell Biology

After constructing the biostretch apparatus (Chapter 2) and studying its magnetic influence

(Chapter 3), we will investigate the mechanical stimulation provided by the device in this

chapter. We will quantitatively study the influence of the mechanical stimulation generated by

the developed apparatus from the aspects of strain distribution, strain rate, and stretch force.

Relevant bio-experiments are conducted to validate the performances of the apparatus.

This chapter is organized as follows. A brief introduction of the current state of stretch

applications is presented in section 4.1. Section 4.2 validates the apparatus’ performance by

examining cell proliferation and cell orientation. Section 4.3 investigates the strain distribution

on the surface of the scaffold with a nondestructive method. Section 4.4 studies the strain rate of

the apparatus with Photron’s high-speed camera Ultima APX. Section 4.5 calculates the stretch

force provided by the apparatus. Section 4.6 draws conclusions from the above research.

68

Page 83: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

4.1 Introduction

Tissue engineering shows great potential for treating diseases by providing functional cell-

based substitutes for human tissues [6, 94, 95]. In addition, engineered tissues can also serve as

in vitro tissue equivalents for drug testing and development, and even for studying the tissue

functions that can expand our understanding of tissue biology [69, 96, 97]. However, in

conventional three-dimensional tissue culture systems, engineered tissues could not develop the

similar structure as native tissues. Compared with engineered tissues, native tissues live in a

much more complicated environment in vivo. For example, connective tissues such as tendon

and ligaments experience tensile loading, while native cardiac tissue contracts under the control

of electrical pacing signals. The specific constructs of tissues are built to adapt their

environment. Currently, most engineered tissues are cultured under a relevantly static

environment. Their cell density and mechanical properties in tension and compression are

substantially lower than those of native tissues [59]. Therefore, external physical stimuli are

necessary for the culture systems to mimic various aspects of the in vivo environment.

Different types of cells are sensitive and responsive to different mechanical stimuli. As

described in Chapter 1 section 1.2.1, smooth muscle cells usually experience tensile forces [7].

Since the designed apparatus is developed for uniaxial cyclic stretch, smooth muscle cells are

chosen as the cell source for all the relevant bio-experiments. Moreover, the types of the

uniaxial biostretch apparatus are reviewed in Chapter 2 section 2.2. Based on these devices,

related bio-experiments have demonstrated that cyclic stretch is a potent stimulus, which can

improve cell proliferation, stimulate extracellular matrix production and the mechanical

properties of tissues [6, 12, 24, 24, 54, 55, 59, 62, 64, 97, 98].

69

Page 84: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

The group of Vandeburgh [12, 54-56] studied the effects of uniaxial stretch for decades.

Their results show that repetitive stretch/relaxation for eight days increased the elasticity, mean

myofiber diameter, and myofiber area percent of human bioartifical muscles (HBAMs). In

addition, they noted that the muscle cells generated increasing internal forces during formation.

Studies also show that uniaxial cyclic stretch improved the structure and function of cultured

avian and rodent muscle [54, 99]. Eschenhagen et al. and Zimmermann et al. [16, 24]

demonstrated that engineered heart tissue (mixture of neonatal rat cardiac myocytes, collagen I

and matrix factors) displayed important hallmarks of differentiated myocardium when subjected

to unidirectional cyclic stretch. Shimko et al. [98] used murine embryonic stem cells (ESC) to

examine the direct effect of mechanical loading (uniaxial stretch) on the differentiation of ESC-

derived cardiomyocytes, and found that mechanical loading significantly affected gene

expression. Akhyari et al. [6] utilized human heart cells to demonstrate that uniaxial stretch

enhanced the formation of a three-dimensional tissue-engineered cardiac graft.

The above influences were investigated by different custom-built devices. Due to the

diversity of the experimental devices, the comparison of all the mechanical parameters provided

by the devices is difficult. The most commonly studied mechanical parameter for these devices

is strain. In our study, we employed noncontact magnetic force as the cyclic uniaxial stretch

force, and provided two different stretch methods. Compared with the previous device [66], we

can precisely controll the mechanical parameter, strain, by confining the distance between the

stoppers in the mounting tray. However, uniaxial stretch can vary with other parameters, such as

stretch pattern, strain distribution and rate, stretch force and frequency, which will be discussed

in the following sections. These parameters will all contribute to cell orientation. Nevertheless,

70

Page 85: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

before studying these mechanical parameters, the validation of the apparatus will first be

examined by the bio-experiments.

4.2 Validation of the Uniaxial Cyclic Stretch

Although many results of strain-induced cell proliferation, matrix production, and

increased mechanical properties have been widely reported in previous literature review [13,

58], the performance of the designed apparatus still needs to be validated by the biological

experiments. Since cell proliferation has been utilized to study the influence of the magnetic

field and temperature, the same method will be used to test the influence of uniaxial stretch.

4.2.1 Cell Proliferation

Cells and Cell culture

In the experiment, the gelatine sponge (Gelfoam: Pharmacia & Upjohn Co., Kalamazoo,

MI) was again used as the scaffold and rat smooth muscle cells served as the cell source. All the

stretch related accessories, such as clamps, steel bars, and boards, were sterilized. The

experiment followed the same cell-seeding protocol described previously. The cell suspension

for each Gelfoam scaffold (30x10x10 mm) was 3×106 cells/125μl culture medium. The scaffold

and clamps were assembled in the hood with tweezers. The assembled scaffolds were then

mounted into 35mm Petri dishes, and cells were seeded on the top of the Gelfoam. The samples

were then placed at 37℃ in a humidified incubator with 5% CO2 and 95% air for twenty-four

hours. Five sample dishes were subsequently placed on the newly designed bio-stretch

apparatus, and another five samples were used as controls in the same incubator. The stretching

71

Page 86: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

group was exposed to intermittent stretch, alternating between 10 minutes of stretch time and 10

minutes of rest time for three days. The parameters of the mechanical stimulus are defined as

1Hz, 50% strength, 50% ON time and 20% strain. The culture medium was changed every day.

DNA Assay

After three days of stretching, DNA assay was utilized to determine the number of cells

on each engineered tissue patch following the same instruction described before. Table 4.1

compares the average and standard deviation of DNA concentration for each group.

Table 4.1 Cell Numbers of Stretched and Control Samples

Uniaxial Stretched

(n=5)

Control

(n=5)

DNA Average

(μg/ml)

60.0 36.2

DNA Std 5.37 4.99

T-Test 0.0009

Results

Table 4.1 shows that the average of the DNA concentration is much higher in the

stretched group than in the control group. The DNA concentration between the stretched group

and the control group is statistically significant (p<<0.05).The results demonstrate that this

device can provide uniaxial cyclic stretch to the tissue construct, and thus influence cell

proliferation. This result was in accordance with other research groups’ results [6, 55, 97].

72

Page 87: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

However, the experiment results are the response to all the characteristics of a uniaxial cyclic

stretch regime (deformation magnitude, stretch strength, stretch frequency, and even stretch

methods). The independent contributions of each mechanical parameter are unclear.

4.2.2 Cell Orientation

After studying the stretching effect on cell proliferation over a period of in four days,

the influence of stretch on cell orientation was studied for a longer term (2-weeks).

Cells and Cell culture

In this experiment, the gelatine sponge (Gelfoam: Pharmacia & Upjohn Co., Kalamazoo, MI)

was again used as the scaffold and rat smooth muscle cells served as the cell source. The

experiment followed the same procedure described in section 4.3.1., with the only difference

being that the culture period lasted 14 days. The sample number for each group is three.

Hematoxylin and Eosin (HE) Staining

Fourteen days after being seeded with cells, the engineered tissue patches (stretched (n=3);

control (n=3)) were carefully removed from the clamps in the culture dishes. The samples were

then fixed in 10% neutral buffered formalin for 24 hours. Using standard methods, the samples

were embedded in paraffin blocks, with care being taken to maintain their final orientation from

the apparatus.

Two paraffin sections were cut out from each sample, parallel to the surface of the samples.

HE staining was used to examine the gross tissue structure. Hematoxylin stains cell nuclei and

polyribosomes, giving nucleic acid a blue color. Eosin stains protein and cytoplasm, giving it a

73

Page 88: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

pink color. The stain images were observed under a Nikon Eclipse 80i fluorescence microscope.

Fig. 4.1 compares the stretched and control sample under the magnification of 100x and 200x,

respectively.

(a)

(b)

Figure 4.1 Gross Structure of the Engineered Tissue

(a) stretched (left: 200x, right: 100x ); (b) control (left: 200x, right: 100x )

74

Page 89: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Results

The images for both the stretched group and control group show that fourteen days later

cells still alive on the scaffold. Further, the visual observation demonstrates that the stretched

constructs have more cells than the control group. Three different images were examined for

each section from vary parts of the scaffold. The number of cells in each image (0.2mm2) was

counted in Adobe Photoshop CS (Adobe Systems, San Jose, California). The stretched group

and control group were averaged and compared (18 images each group) in Fig. 4.2. The

average number of cells was 194 for the stretched group, and 81 for the control group. The T-

test result shows p<<0.05, and marked as * in the graph, which means the stretched group has

significant difference compared with control group.

Cell Proliferation

0

50

100

150

200

250

Ce

ll N

um

be

r a

t 0

.2m

m2

stretched

control

*

Figure 4.2 Comparison of Cell Numbers

(* represent the significant difference between two groups)

Moreover, for the control samples, the cells were scattered among the holes of the

scaffold (Fig. 4.1(b)). For the stretch samples, the cells demonstrated inhomogeneous cell

75

Page 90: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

distribution. The cell clusters were parallel along the stretch direction, longitudinally oriented

(Fig. 4.1(a) arrow direction). Additionally, compared with the control group, the stretched

samples exhibited a good cell orientation at the free edges and two ends. Complexes of

multicellular aggregates and longitudinally oriented cell bundles were also found at the ends

and free edge areas (Fig. 4.3).

Figure 4.3 Edge Structure of the Engineered Tissue

(a) control; (b) stretched (magnification: 100x )

4.2.3 Discussion

The results of these two bio-experiments validated the performances of the apparatus. The

device is capable to provide uniaxial cyclic stretch during tissue culture with well-defined

mechanical parameters (strain amplitude, frequency, strength, and duty cycle), which can affect

cell proliferation and cell orientation. However, the results reflected the influence of all

characteristics of a stretch regime, not a single parameter.

76

Page 91: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Cell Reorganization

The HE images show that without external stimulation, cells clustered among the holes of the

scaffold randomly. However, under cyclic stretch, these tissue-like bundles are not uniformly

distributed along the surface of the scaffold. Cells aggregated at the free edges and ends of the

scaffold, while the centre part had fewer bundles. This phenomenon can be explained as

multiple stimuli existed during stretch. The scaffold was not only subjected to uniaxial stretch,

but also strain-induced fluid flow around and within the porous scaffold. Therefore, this culture

system actually induced not only cyclic stretch but also shear stimulation. Additionally, due to

Poisson’s effects, when the substrate undergoes one dimensional stretch, the other two

dimensions will experience compression. During a stretch/relaxation period, the medium inside

the scaffold may experience a squeeze/inhale period, which in turn improved the mass

transportation condition [17][59]. In future studies, the effect of each mechanical parameter

needs to be isolated to help researchers determine the mechatransduction of the cells.

Variation of Mechanical Property

For the long-term culture, it was observed that the scaffolds maintain a high degree of

elasticity during the first seven days. Later, a stronger force (60%) is needed to obtain the same

strain magnitude, which indicates that the stiffness of the structure changed during culture. This

kind of variation has been observed and recorded by other groups [55, 97].

4.3 Study of Strain Distribution

From the literature review listed in table 2.1, we know that the most frequently studied

parameter of cyclic stretch is strain. The average strain Ɛ provided by the device is defined

77

Page 92: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

as 0/ LL . L0 is the distance between two symmetric inner stoppers, shown in Fig. 2.6. For

symmetric stretch, L is the total deformation of the scaffold, which is twice the distance from

the inner stopper to the outer stopper. For asymmetric stretch, L is the deformation of the

movable end of the construct. Unlike an elastic rigid-body, the scaffold is a vesoelastic material.

Although the average strain of the construct is clear, the strain distribution at the surface of the

scaffold is unknown. Therefore, a nondestructive method [70][100, 101] was utilized to

investigate the full-field surface strain distribution of the three-dimensional scaffold.

4.3.1 Measurement Methodology

To study the strain distribution, samples of both Gelfoam (41 x 20 x 7 mm, n=2) and GE

RTV 6166 silicone (38 x 16 x 3 mm, n=11) were used as scaffolds. Compared with Gelfoam,

which is porous and has a heterogeneous property, the GE RTV 6166 silicone is a homogeneous

material, allowing it to serve as a baseline to compare to. The Gelfoam samples were marked

with dots at 2mm intervals using a fine-tipped permanent marker; while the RTV samples were

marked by dark microbeads. These marks separated each RTV silicone sample and Gelfoam

sample into 16 sections along the stretching axes (Fig. 4.4). The rectangular areas in Fig. 4.4 are

the analysis area. After marking, the scaffolds were affixed to stainless steel clamps with

silicone glue. Scaffolds were then soaked in water for at least 24 hours. The final dimensions of

the scaffolds with clamps were slightly different due to the gap made by the glue and shrinkage

when the scaffolds were inserted in water, ranging from 40x18x7 mm to 40.5x18x7 mm for

Gelfoam, and 40x16x3 mm to 41.5x16x3 mm for RTV silicone. The scaffolds were then placed

in the aforementioned apparatus for cyclic stretch. Two different stretch methods were applied:

symmetric stretch and asymmetric stretch, which were discussed in Chapter 2 section 2.3.2.

78

Page 93: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

In these trials the outer stopper pins were not used. For asymmetric stretch, the inner stopper

pins on the fixed end were placed to secure the scaffold directly against the side of the culture

dish, and on the free end were placed to prevent the scaffold from deforming to less than its

original length. For symmetric stretch, the inner stopper pins were placed only to prevent the

scaffold from deforming to less than its original length, and the scaffold was allowed to stretch

the full length of the culture dish. In this way, the stretch magnitudes of the two methods were

the same. For Gelfoam, the stretch parameters of the biostretch apparatus were set with

frequency of 1Hz, ON time of 50%, and strength of 35% of the maximum value. For RTV, the

parameters were set with frequency of 1Hz, ON time of 50%, and strength of 60% of the

maximum value. The average strain was about 20%.

Digital photographs of the scaffold in the stretched and unstretched states were taken with a

D50 digital camera (Nikon Corporation, Tokyo, Japan) and a Tamron SP 90mm macro lens

(Tamron Co., Ltd., Saitama, Japan). Photographs had a resolution of 60 pixels per millimetre.

The photographs were edited with Adobe Photoshop CS (Adobe Systems, San Jose, California)

using the High Pass and Threshold tools to identify dots. The centroid of each dot was

determined using Scion Image (Scion Corporation, Frederick, Maryland), and distances

between successive centroids along the stretch axis were calculated to determine the distribution

of strain, defined as deformation (∆L) divided by the original distance between dots (L0). For

each distance along the stretch axis, the strains at several equally spaced locations on the

perpendicular axis were calculated and averaged, and the standard distribution of those strains

was determined.

79

Page 94: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

(a) (b)

Figure 4.4 Images of the Gelfoam and GE RTV 6166

(a) up: unstretched RTV, down: stretched RTV

(b) up: unstretched Gelfoam, down: stretched Gelfoam

4.3.2 Results

We investigated two types of strain distributions, global strain and local strain distributions.

Global strain distribution refers to the relationship between the strains of different sections;

local strain refers to only the strain of a specific section.

Global Strain distribution of Gelfoam

Based on the above method, we investigated the global strain distribution in both the stretch

axis (x-axis) and the perpendicular horizontal axis (y-axis). In general, strain distribution was

found to be non-uniform, with a high standard deviation, but exhibiting no significant trends

80

Page 95: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

along the stretch axis. Fig. 4.5 (a) shows the average deformation between each adjacent pair of

data points for 10 rows of one Gelfoam sample along the x-axis. No statistically different results

were found for global strain distribution of Gelfoam, which is due to high standard deviation in

measurements.

(a)

(b)

Figure 4.5 Strain Distribution of a Typical Sample of Gelfoam

(a) Up: the x- axes; (b) Down: the y- axes

81

Page 96: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

A similar graph for strain distribution in the y-axis shows a clearer trend, with the edges of

the Gelfoam deforming least, since they were securely glued to solid metal clamps (Fig. 4.5

(b)). Fig. 4.5 (a) and (b) demonstrates that the scaffold is not only experienced longitudinal

tensile, but also compressed along the lateral direction owing to Poisson’s effects.

Local Strain Distribution of Gelfoam

Compared with the global strain distribution, the deformation of small sections of Gelfoam

(local strain) was far more significant. Photographs of the Gelfoam indicated that shapes of the

dots were visibly distorted during stretching (Fig. 4.6), such that local strain in small regions

(100µm) ranged from negative strains on the order of -5% to large positive strains on the order

of 100%. One explanation for the phenomena is the non-uniform construction of the Gelfoam

sponge, which caused areas with larger surface cavities to exhibit higher strain than areas with

smaller surface cavities. Additionally, parts of the Gelfoam may have torn under tensile stress

and hence exhibited higher strain. In comparison to the global strain (10-20%) and the size of a

typical cell (10-40µm), these local strains are clearly significant. We note that the material

properties of the Gelfoam may be different when the Gelfoam is seeded with cells; this

difference will be discussed in section 4.6.

Figure 4.6 Distortion of a Marked Dot

Left: before stretching; Right: after stretching

82

Page 97: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

The significance of the local strains is demonstrated by comparing Gelfoam to RTV silicone,

which has a visibly more uniform construction. Local strains between dots on Gelfoam spaced

2mm apart exhibit much higher standard deviation than those on RTV silicone, with a standard

deviation of about 30% of the average strain in comparison to less than 10% for RTV silicone.

A visual comparison of typical strain distributions for the two kinds of scaffolds, as in Fig. 4.7,

supports this claim. Since these local strains were calculated by measuring distances between

centroids of dots, they are in effect average values, and the standard deviation for local strains is

actually even higher than calculated.

Figure 4.7 Strain Distribution for the Middle Rows of the Gelfoam and RTV Silicone

Samples

Symmetric and Asymmetric Stretch

Two different stretch methods, asymmetric and symmetric stretch, are compared for two

kinds of samples. However, there are no significant trends along the stretch axis under different

stretch methods. But the symmetric stretch method provides a good chance for real-time strain

monitoring since the sample could stay centered below a camera. Moreover, other simple

83

Page 98: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

physical variables, such as displacement and speed of particles, may be easier to correlate with

differences in cell growth. For example, under one-way stretch, the free end is displaced further

and has a higher average speed than the fixed end, which may correlate with cell growth.

Edge Effects

We note that the strain distribution in the edge sections, where the clamps are attached, is

heavily dependent on the method of application of force, and is subject to high measurement

error. Before utilizing the clamps described in Chapter section 2.3.2, other clamps were tested.

When force was applied to metal bars inserted through the Gelfoam, strain measured at the

edges was less than half of that measured at the centre. Conversely, when clamps were affixed

to the Gelfoam, strain at the edges appeared more than double the strain measured at the centre.

These are genuine differences in strain; however, they are not necessarily the cause of

differences in cell expression, as there are many other differences between the centre and edges

of the Gelfoam. We also note that the substrate undergoes deformation in the vertical direction

near the edges, which causes strain to appear different from its actual value. For these reasons,

we omit the outer 2 millimetres in all of our measurements.

Since the analyzed data was only taken when the Gelfoam was stretched either fully or not at

all, no information was obtained about the partially stretched state of Gelfoam, and hence strain

rate distribution is unknown. Previous biomechanics studies show that biological tissues are not

elastic materials, the living tissues are normally recognized as viscoelastic materials with

nonlinear stress-strain characteristics [102]. Instead of the strain magnitude itself, the history of

84

Page 99: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

strain affects the stress more. Therefore, the study interest was shifted from strain distribution to

strain rate.

4.4 Study of Strain Rate

From the design principle of the apparatus, we know that when the magnet is electrified by a

pulsed current, the magnet will generate a time-varying magnetic field. When the assembled

dish (Fig. 2.6) is positioned under such a field, a magnetic force will act on the magnetizable

bar, which will then stretch the connected tissue construct longitudinally. When the magnetic

field disappears, the tissue construct will restore to its original shape due to the elasticity of the

scaffold. Under a confined strain value, if the stretch force varies, the strain rate will change.

Therefore, the relationship between stretch force and strain rate at the defined strain magnitude

was investigated.

4.4.1 Measurement Methodology

The engineered tissue was prepared according to the protocol provided in section 4.3. After

the tissue patch was cultured for 16 days, the assembled dish was positioned under the scope of

Photron’s high-speed camera ultima APX (Photron, San Diego, California), and continuously

subjected to symmetric stretch. The stretching parameters were set as: 1Hz, 50% ON time, and

the total deformation of 5mm. As the original length of the Gelfoam sample is 20mm, the

maximum strain is 25%. The stretching process was recorded by the digital imaging system.

The high-speed digital imaging system was set to 500 fps at full mega pixel image resolution

(1024 x 1024). The recording time was 4.1 seconds, with 2048 frames, such that each record

85

Page 100: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

includes four successive stretch periods. The recordings were saved in AVI format. Four video

outputs were recorded, with stretch force ranging from 40% to 90%.

4.4.2 Results

The time period of each frame of the video represents 2ms. The images were edited by GIMP

(GNU General Public License). With the help of the calibration picture, the stretch strain rate,

the restore strain rate, and the frequency response of the device for a specific material and strain

were determined from the video.

Stretch Strain Rate

The data show that the stretch strain rate varies under different stretch strengths. Fig. 4.8

(a) and (b) show the stretch rates with different stretch strengths of 90% and 40%, respectively,

and the stretch processes demonstrate very good repetition for the four successive periods.

When the strength of the stretch force reaches 90%, 16ms is needed to reach the total

deformation. When the strength is only 40%, 22ms is needed to reach the same deformation.

Despite the difference in stretch time, the two plots share the trend that the stretch rates vary

and increase nonlinearly during the stretch process. These plots reflect the nonlinear viscoelastic

property of the material or indicate that the stretch force is not constant during the stretch

process. Thus, this stretch force will be calculated in section 4.6.

86

Page 101: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

stretch rate at 90% strength

0

0.5

1

1.5

2

2.5

1 2 3 4 5 6 7

frames

de

form

ati

on

[m

m]

round1

round2

round3

round4

(a)

stretch rate at 40% strength

0

0.5

1

1.5

2

2.5

1 2 3 4 5 6 7 8 9 10

frames

def

orm

atio

n [

mm

]

round1

round2

round3

round4

(b)

Figure 4.8 Stretch Strain Rate at Different Stretch Strength

(a) 90% strength; (b) 40% strength

87

Page 102: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

The strain rate during the stretch process shares the exactly same plots as stretch rate (Fig.

4.8), since the strain is obtained by deformation (L) divided original length (L0). In the test

range, the maximum strain rate for 90% stretch strength is about 107s-1, while the maximum

strain rate for 40% stretch strength is about 102s-1.

Retreat Rate

When the external force (magnetic force) is removed, the construct will retreat back to its

original place due to the elasticity of the material. The analysis data show that the retreat

process takes 48ms, which is much longer than the stretch process. However, these two plots

demonstrate the same trend, which shows that the retreat process is not affected by what’s

happened in the past but the material’s own material properties. The plots demonstrate the creep

property (time dependent strain rate) of the material during the unloading process. These results

are in accordance with the facts that the scaffold is a viscoelestic material. But the plots (Fig.

4.9) do not match the creep functions of the ordinary linear viscoelastic mechanical models (the

Maxwell model, the Voigt model, and the Kelvin model) [102]. Therefore, the material property

of the engineered construct needs further study.

The scaffold of the sample is Gelfoam, an absorbable gelatin sponge. The material

properties of the Gelfoam may change when it is used as a scaffold, due to biodegradation and

cell reorganization. This non-invasive test method provides a new avenue to monitor the change

of the mechanical properties of the engineered tissue during culture process.

88

Page 103: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

retreat rate at 90% strenght

0

0.5

1

1.5

2

2.5

3

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24

frames

def

orm

atio

n [

mm

]

round1

round2

round3

round4

(a)

stretch rate at 40% strength

0

0.5

1

1.5

2

2.5

3

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24

frames

def

orm

atio

n [

mm

]

round1

round2

round3

round4

(b)

Figure 4.9 Retreat Strain Rate at Different Stretch Strength

(a) 90% strength; (b) 40% strength

89

Page 104: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Frequency Response

The above data provides us a method to study the frequency response of the device for a

specific material and strain. The stretch force is generated from the pulsed exciting current,

shown in Fig. 4.10 (a). Theoretically, the frequency of the current pulse can reach very high

values, as the maximum source frequency of the timer is 20MHz. However, according to the

recorded data, the stretch time and the retreat time are 16ms and 48ms, respectively. For a 10Hz

pulse train, the response stretch pulse is shown in Fig. 4.10 (b). The shortest response time that

the construct needs for a period is 64ms (Fig. 4.10 (c)). Therefore, the frequency response of the

scaffold is about 15Hz for 25% strain. The data demonstrate that frequency response is related

to not only the stretch force, but also the strain (deformation) and the material properties.

However, this calculation ignores the delay between the exciting current and the generated

magnetic field.

Figure 4.10 Stretch Frequency

90

Page 105: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

4.4.3 Discussion

All the collected data were recorded under the real culture circumstance. During the process,

the Gelfoam were soaked in medium. The video shows that during stretching, the fluctuation of

the medium distorted the image, especially at the moment that the metal bar hit the wall of the

culture dish or the stoppers of the mounting tray, which is the major measurement error during

the test. Therefore, it was hard to capture the accurate position of the steel bar at the hitting

momentary. Due to this reason, the deformation of the last frame is not plotted in Fig. 4.8.

This test demonstrated that for a specific material the stretch rate is related to the strength of

the stretch force. However, the stretch rate is not proportional to the stretch force. When the

force doubled (from 40% to 90%), the full stretch time did not reduce to half (from 22ms to

16ms). Additionally, the material property affects both the stretch rate and the retreat rate. In

order to obtain the same strain magnitude, different scaffolds need different initial force to

stretch. This claim has been supported in section 4.4.1. To test the strain distributions of

Gelfoam and RTV silicone, different stretch force were applied. Therefore, this nondestructive

method also provides researchers a tool to real time monitor the variation of the material

property during cell orientation.

4.5 Study of Stretch Force

From section 4.5, we noticed that the stretch force may not be a constant force. In this

section, we will estimate the distribution and magnitude of the magnetic force generated

between a magnetisable bar and a horseshoe-shaped electromagnet. The geometrical model of

the magnet is based on MCI-3404 (Mci Limited Co., ON, CA) utilized in the biostretch

91

Page 106: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

apparatus, the magnetizable bar (16x 3.5x 1mm). We focus our model study on two dimensions,

particularly along horizontal (x) direction because the horizontal magnetic force is used as the

uniaxial stretch force. In this model, the two dimensions of the bar are 16 x 1 mm, and s the

distance between the bar and the magnet is 5mm. The simulations were performed using

COMSOL Multiphysics 3.3.

4.5.1 Electromagnetic Force Calculation

In COMSOL Multiphysics, Maxwell’s stress tensor T is used to calculate the magnetic

force act on the surface of the object. Cauchy’s equation of continuum mechanics states

extfTdt

rd

2

2

(4.1)

where is the density, r denotes the coordinates of a material point, T is the stress

tensor, and fext is an external volume force. In the stationary case, the equation representing the

force balance is

extfT 0 (4.2)

If the stress tensor is continuous across a stationary boundary between two materials, the

equation can be expressed as:

0)( 121 TTn (4.3)

where T1 and T2 represent the stress tensors of material 1 and 2, and n1 is the outward

normal direction of material 1. The equation shows when material 1 (solid) is surrounded by

material 2 (air/vacuum), the surface force is on the boundary between the materials with the

solid. The user guide of COMSOL [89] deducted the electromagnetic stress tensor expression in

air:

92

Page 107: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

TT HBEDIBHDEpIT )2

1

2

1(2 (4.4)

where p is the air pressure, I is the identity 3*3 tensor, E is the electric field intensity, D

is the electric flux density, H is the magnetic field intensity, B is the magnetic flux density. For

the quasi-static magnetic field, the stress tensor can be simplified as:

TBHnBHnTn )()(2

11121 (4.5)

4.5.2 Electromagnetic Model Analysis

In order to study the magnetic force magnitude and distribution in a specific area

according to the equation (4-5), the magnitude and distribution of the magnetic flux density B in

the study area must first be calculated. To perform the simulation, beside the assumptions we

made for the model discussed in Chapter 3, we added two more assumptions:

The material of the steel bar is soft iron instead of Ph-174

The surrounding area is air instead of medium

The meshed geometry of the electromagnet model has 3096 elements and the number of

degrees of freedom is 6251 (Fig. 4.11). The mesh generator partitions the subdomains into

triangular elements, which are allocated based on the geometrical dimensions.

93

Page 108: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Figure 4.11 Meshed Model of an Electromagnet and a Steel Bar

Figure 4.12 Magnetic Flux Density, x component

94

Page 109: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Fig. 4.12 shows the x component distribution with the maximum flux density in a period

at phase zero. In the graph, the magnetic south pole (orange) of the horseshoe electromagnet is

at the lower foot and the magnetic north pole (blue) is at the upper foot. The distribution shows

that the magnetic flux density Bx has greater magnitude in the soft iron core than in the air

space. The maximum B value in the soft iron is 0.221T located at the inner corners of the core.

However the magnitude of Bx dramatically reduces to 0.086T at the boundary of the iron core

and air space. Compared Fig. 4.12 and Fig. 3.2 (a), we can find that the steel bar was

magnetized in the field. The Bx in the area between the magnet and the steel bare increased.

4.5.3 Results

Flux Density Bx VS Distance

X Direction Fig. 4.7 (a) and (b) shows the cross-section plot of magnetic flux density Bn

normal along the X direction (from x=0.08 to x=0.085) at y=0.144 (south pole), and at y=0.159

(north pole), respectively. Compared to Fig. 3.4, the magnitude of the normal magnetic flux

density in Fig. 4.13 dramatically increased when the distance increased to 4mm. The

phenomenon is caused by the magnetizing of the steel bar.

95

Page 110: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

(a)

(b)

Figure 4.13 Magnetic Flux Density Bn along x Direction

(a) at Y=0.143; (b) at Y=0.159

96

Page 111: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Figure 4.14 Magnetic Flux Density Bx along y Direction

Bx distributions @ x=0.085

0

0.02

0.04

0.06

0.08

0.1

0.12

0.15

90.

1580.

157

0.15

60.

1550.

1540.

1530.

1520.

151

0.15

00.

1490.

1480.

1470.

146

0.14

50.

1450.

144

y [m]

Mag

net

ic F

lux

Den

sity

, x

com

po

nen

t [T

]

without rod

with rod

Figure 4.15 Comparison of Magnetic Flux Density Bx

97

Page 112: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Y Direction Fig. 4.14 shows the cross-section plot of the x component of the magnetic flux

density along y direction (from y=0.143 to y=0.159) at x=0.085. Fig. 4.14 demonstrates the

same trend as Fig. 3.3, but the magnitude is much greater. Fig. 4.15 compares the difference

between these two plots in absolute terms, which shows the flux density at the edge of the bar is

increased six times. This plot also implies that the force is mostly applied on the edge of the bar,

such that the center part will not be subject to magnetic force. Both Fig. 4.13 and Fig 4.14

illustrate the distribution of the flux density along x and y direction, which directly affects the

calculation of the magnetic force.

Magnetic Force Fx vs. Exciting Current I

The magnetic force acting on the steel bar is calculated by the stress tensor method

described in section 4.6.1. The simulation results show that the magnetic force that the bar is

subjected to along the x direction is 5.1N when the electromagnet is electrified by a 0.4A time-

harmonic sinusoidal current. This is the maximum force the device can provide at 5mm distance

far because the maximum current designed for the device is 0.4A. The force varies as the

exciting current (sinusoidal signal) changes during a period. The relationship between the

magnetic force along x direction and the exciting current is illustrated in Fig. 4.16. In the

previous bio-experiments, the Gelfoam worked as the scaffold, and the stretch force was

normally set as 50%, which means the exciting current was 0.2A. From Fig. 4.16, we can see

the stretch force generated would be about 1.25N.

98

Page 113: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Figure 4.16 Exciting Current vs. Magnetic Force

Magnetic Force Fx VS Distance

Fig. 4.17 shows that the distribution of the flux density is non-linear along the x direction,

which means the magnetic force will also vary along the x direction. The plot demonstrates that

when the bar is closer to the electromagnet, it will be subject to a much stronger force. For a

35mm culture dish, the maximum deformation of the substrate is set to be 2.5mm, so the force

ranges form 1.25N to 4N.

99

Page 114: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Rod_force_x versus Distance

0

10

20

30

40

50

60

0.5 1 1.5 2 2.5 3 3.5 4 4.5 5

Distance[mm]

Ro

d_f

orc

e_x

[N]

Figure 4.17 Distance vs. Magnetic Force

In this thesis, the area around the electromagnet was modeled as air instead of medium, and the

steel bar was modeled by soft iron; more realistic modeling wound give more accurate results.

4.6 Conclusions

In order to validate the performance of the designed apparatus, the effects of uniaxial

cyclic stretch on cell biology have been studied from the aspects of cell proliferation and cell

morphology. The DNA assay results show that the apparatus improved cell proliferation in a

similar way as existing uniaxial stretch devices. The HE staining demonstrated the gross tissue

structure, and the results showed that stretching guided the cell reorganization along the stretch

direction. Moreover, the mechanical parameters (strain distribution, strain rate, and stretch

force) provided by the apparatus have also been investigated. A nondestructive method was

employed to investigate the strain distribution on the surface of the porous scaffold. The results

demonstrate that the strain distribution along the stretch direction is non-uniform, with a high

100

Page 115: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

standard deviation. The strain rate was investigated by Photron’s high-speed camera ultima

APX. The results demonstrated the viscoelastic property (time dependent strain) of the scaffold,

and also indicated that the strain rate varies with the strength of the stretch force. The stretch

force was calculated to be 1.25N at 5mm distance from the edge of the electromagnet by

COMSOL Multiphysics, the force was shown to be related to distance and current.

101

Page 116: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Chapter 5

Summary and Future Research

5.1 Summary

Mechanical stimulation is an important regulator in tissue culture for tissue engineering.

Among different forms of mechanical stimuli, uniaxial cyclic stretch is the most commonly

used stimulus for muscle tissues due to its precisely controlled end to end average strain. The

most typical method of applying stretch is motor driven. However, the motor driven method

involves a physical connection between the culture construct and the motor, which requires a

diversity of laboratory devices and increases the risk of contamination.

This thesis proposed a unique idea to design a uniaxial cyclic stretch apparatus based on

non-contact magnetic force. The apparatus developed in this thesis allows researchers to utilize

standard culture dishes, and applies a well-defined strain magnitude to an engineered tissue

patch during culture process, which greatly simplifies the assembly procedure and reduces the

risk of contamination. A prototype of the apparatus has been constructed and tested in a

research lab at the University Health Network in Toronto.

Since the applied mechanical stimulus is generated from magnetic force, the engineered

tissue construct is not only affected by mechanical force, but also exposed to a magnetic field.

102

Page 117: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

Thus, the effects of the time-varying magnetic field during the culture process were

investigated. One side effect of using electromagnets, that of a temperature increase, is also

investigated. However, the biomedical experiment results show that neither a weak low

frequency magnetic field (0.1T, 1Hz) nor an increase of 1℃ in temperature has a significant

effect on the cell culture.

In order to verify the performance of the designed apparatus, the effects of uniaxial

cyclic stretch on cell biology have been studied from the aspects of cell proliferation and cell

orientation. The results show that the apparatus improved cell proliferation and cell

reorganization similarly to other known uniaxial stretch devices. Moreover, the mechanical

parameters (strain distribution, strain rate, and stretch force) provided by the apparatus have

also been quantitatively investigated. The strain distribution on the surface of the porous

scaffold along the stretch direction was shown to be non-uniform, with a high standard

deviation. The results of strain rate test indicated that the strain rate varies during stretching due

to the variation of the stretch force. It also proved the viscoelastic property (time dependent

strain) of the scaffold, and provided a new way to monitor the mechanical property of the

engineered tissue during tissue formation. The stretch force, which was calculated using

COMSOL Multiphysics, was shown to be related to distance and current.

5.2 Future Research

This section discusses the future research work that will be pursued after the completion

of the thesis, from the aspects of modeling, bio-experiments, and improvements.

103

Page 118: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

5.2.1 Optimizing the Electromagnet Model

The model of the electromagnet discussed in this thesis only considered the quasi-static

electromagnetic field itself, while ignoring the influence of the energy dissipation problems

caused by resistance heating and hysteresis. In the future, the study areas need to cover not only

electromagnetics, but also heat transfer coupled with thermal-electric analysis. The multi-

physical model can help us optimize the relationships among the current, force, and

temperature. Additionally, some of the assumptions need to be revised in order to obtain more

accurate results, such as more specifically characterizing the features of the coil. Moreover, the

model should be upgraded from two-dimensional to three-dimensional.

5.2.2 Implementing Bio-experiments

Although the bio-experiments validated the performance of the apparatus, and

demonstrated that uniaxial stretch can influence cell culture from the aspects of cell

proliferation and cell orientation, the relationship between cell characteristics and the

mechanical parameters are still unclear. The results were affected by multiple parameters such

as stretch frequency, strength, strain, and even stretch pattern and shear flow. The independent

contribution of each parameter is not characterized. Therefore, future experiments should isolate

these mechanical parameters, and systematically examine their independent influence. Such an

experiment would provide researchers an effective platform to explore the mechatransduction of

cells. The upcoming experiment is studying tissue formation by leading the human embryonic

stem cells differentiated to cardiac tissue cells with the defined mechanical stimulation.

104

Page 119: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

5.2.3 Calibrating the stretch force

At current stage, the primary controllable parameter of mechanical stimulation in tissue

engineering is strain magnitude, and cultured tissue patches are subjected to predefined

mechanical strain. However, the mechanical force that the patches are subjected to is not fully

studied. According to the simulation results of the COMSOL model, the stretch force is 1.25N

at 5mm away from the magnet, when the magnet is excited by a 0.2A pulsed current. Although

the simulated results of magnetic flux density B are verified by physical experimentation, and

the magnetic force is calculated based on the verified magnetic field, the magnetic force itself is

unverified. Further experiment needs to conduct to determine the stretch force. With the

information of the tested stretch force and filmed deformation, the mechanical properties

(stiffness or elasticity) of the scaffold can be monitored during tissue culture.

5.2.4 Integrating the mechanical and electrical stimuli together

Current bioreactor systems only apply uniaxial cyclic stretch while engineering a tissue

construct. However, the next step of the research will be to combine cyclic stretch and electrical

field stimulation together. The proposed bioreactor is designed to deliver a pulsed electrical

signal to the cell-seeded scaffold with the same frequency as the cyclic stretch. For cardiac

tissue engineering, these external stimuli will mimic the conductive and contractile properties

existing in native heart tissue. Hopefully, these stimuli can lead to better ultrastructural

organization of the engineered tissues.

105

Page 120: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

The proposed research aims to develop a novel, compact electrical and mechanical

stimulation apparatus for 3D tissue culture. As shown in Fig. 5.1, the apparatus will consist of a

computerized controller, an electrified mounting tray, and a pair of magnetic metal bars. The

two ends of the cell-seeded scaffold (1) will be fixed by a pair of silver-coated stainless steel

bars (2). The integrated construct (3) will be secured onto the electrified mounting tray (4). The

tray will contain a battery (5) and relevant circuits (6), which will be coated by a biocompatible

material; it will also contain two pairs of metal stoppers (7) serving as electrodes. The mounting

tray will be designed to fit into a standard culture dish (8). Finally, the assembled culture dish

will be mounted between a pair of electromagnets (9). When the electromagnets are electrified

by the controller, the construct integrated with metal bars will experience uniaxial cyclic stretch

between the stoppers and the wall of the dish. As the metal bars touch the metal stoppers, the

construct will also undergo electrical stimulation.

Figure 5.1 Schematic Diagram of the Upgraded Device

106

Page 121: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

The device will provide simultaneously and synchronously both uniaxial mechanical cyclic

stretch and electrical stimulus to a 3D engineered-tissue construct, which will subtly mimic the

spontaneous excitation-contraction coupling of native cardiac tissue. With the non-contact

magnetic force and encapsulated battery, the tissue patch will still be cultured in a standard

culture dish. Such an isolated environment will greatly limit the risk of contamination. Using

the new device, researchers will be able to study the effects of both mechanical and electrical

stimuli on cell proliferation, the increased expression of cardiogenesis, and the organization of

the extracellular matrix. Biological experiments will be conducted in the Division of

Cardiovascular Surgery at the University of Toronto. The new apparatus will provide an

effective tool to investigate the mechanism of spontaneous beating of cardiac tissues.

107

Page 122: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

References

[1] Fuchs, J. R., Nasseri, B. A., and Vacanti, J. P., 2001, "Tissue Engineering: A 21st Century

Solution to Surgical Reconstruction," Annals of Thoracic Surgery, 72(2) pp. 577-591.

[2] Eschenhagen, T., and Zimmermann, W. H., 2005, "Engineering Myocardial Tissue,"

Circulation Research, 97(12) pp. 1220-1231.

[3] Langer, R., and Vacanti, J. P., 1993, "Tissue Engineering," Science, 260(5110) pp. 920-926.

[4] Leor, J., Amsalem, Y., and Cohen, S., 2005, "Cells, Scaffolds, and Molecules for

Myocardial Tissue Engineering," Pharmacology and Therapeutics, 105(2) pp. 151-163.

[5] Radišić, M., and Vunjak-Novaković, G., 2005, "Cardiac Tissue Engineering," Journal of the

Serbian Chemical Society, 70(3) pp. 541-556.

[6] Akhyari, P., Fedak, P. W. M., Weisel, R. D., 2002, "Mechanical Stretch Regimen Enhances

the Formation of Bioengineered Autologous Cardiac Muscle Grafts," Circulation, 106(13

SUPPL.) pp. I137-I142.

[7] Griffith, L. G., and Naughton, G., 2002, "Tissue Engineering - Current Challenges and

Expanding Opportunities," Science, 295(5557) pp. 1009-1010+1012-1014.

[8] Discher, D., Dong, C., Fredberg, J. J., 2009, "Biomechanics: Cell Research and Applications

for the Next Decade," Annals of Biomedical Engineering, 37(5) pp. 847-859.

108

Page 123: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

[9] Jawad, H., Ali, N. N., Lyon, A. R., 2007, "Myocardial Tissue Engineering: A Review."

Journal of Tissue Engineering and Regenerative Medicine, 1(5) pp. 327-342.

[10] Radisic, M., Park, H., Gerecht, S., 2007, "Biomimetic Approach to Cardiac Tissue

Engineering," Philosophical Transactions of the Royal Society B: Biological Sciences,

362(1484) pp. 1357-1368.

[11] Butler, D. L., Goldstein, S. A., and Guilak, F., 2000, "Functional Tissue Engineering: The

Role of Biomechanics," Journal of Biomechanical Engineering, 122(6) pp. 570-575.

[12] Vandenburgh, H. H., Karlisch, P., and Farr, L., 1988, "Maintenance of Highly Contractile

Tissue-Cultured Avian Skeletal Myotubes in Collagen Gel," In Vitro Cellular and

Developmental Biology - Animal, 24(3 PART I) pp. 166-174.

[13] Terracio, L., Miller, B., and Borg, T. K., 1988, "Effects of Cyclic Mechanical Stimulation

of the Cellular Components of the Heart: In Vitro," In Vitro Cellular and Developmental

Biology - Animal, 24(1) pp. 53-58.

[14] Radisic, M., Park, H., Shing, H., 2004, "Functional Assembly of Engineered Myocardium

by Electrical Stimulation of Cardiac Myocytes Cultured on Scaffolds," Proceedings of the

National Academy of Sciences of the United States of America, 101(52) pp. 18129-18134.

[15] Brown, T. D., 2000, "Techniques for Mechanical Stimulation of Cells in Vitro: A Review,"

Journal of Biomechanics, 33(1) pp. 3-14.

[16] Fink, C., Ergün, S., Kralisch, D., 2000, "Chronic Stretch of Engineered Heart Tissue

Induces Hypertrophy and Functional Improvement," FASEB Journal, 14(5) pp. 669-679.

109

Page 124: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

[17] Ethier, C.R., and Simmons, C.A., 2007, "Introductory Biomechanics From Cells to

Organisms," CAMPRIDGE UNIVERSITY PRESS, New York, pp. 511.

[18] El Haj, A. J., Minter, S. L., Rawlinson, S. C. F., 1990, "Cellular Responses to Mechanical

Loading in Vitro," Journal of Bone and Mineral Research, 5(9) pp. 923-932.

[19] Williams, G. M., Lin, J. W., and Sah, R. L., 2007, "Cartilage Reshaping Via in Vitro

Mechanical Loading," Tissue Engineering, 13(12) pp. 2903-2911.

[20] Hung, C. T., Mauck, R. L., Wang, C. C. -., 2004, "A Paradigm for Functional Tissue

Engineering of Articular Cartilage Via Applied Physiologic Deformational Loading,"

Annals of Biomedical Engineering, 32(1) pp. 35-49.

[21] Knippenberg, M., Helder, M. N., Doulabi, B. Z., 2005, "Adipose Tissue-Derived

Mesenchymal Stem Cells Acquire Bone Cell-Like Responsiveness to Fluid Shear Stress on

Osteogenic Stimulation," Tissue Engineering, 11(11-12) pp. 1780-1788.

[22] Triplett, J. W., O'Riley, R., Tekulve, K., 2007, "Mechanical Loading by Fluid Shear Stress

Enhances IGF-1 Receptor Signaling in Osteoblasts in A PKCζ-Dependent Manner," MCB

Molecular and Cellular Biomechanics, 4(1) pp. 13-25.

[23] DePaola, N., Davies, P. F., Pritchard Jr., W. F., 1999, "Spatial and Temporal Regulation of

Gap Junction connexin43 in Vascular Endothelial Cells Exposed to Controlled Disturbed

Flows in Vitro," Proceedings of the National Academy of Sciences of the United States of

America, 96(6) pp. 3154-3159.

110

Page 125: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

[24] Zimmermann, W. -., Schneiderbanger, K., Schubert, P., 2002, "Tissue Engineering of a

Differentiated Cardiac Muscle Construct," Circulation Research, 90(2) pp. 223-230.

[25] Tandon, N., Cannizzaro, C., Chao, P. -. G., 2009, "Electrical Stimulation Systems for

Cardiac Tissue Engineering," Nature Protocols, 4(2) pp. 155-173.

[26] Erickson, C. A., and Nuccitelli, R., 1984, "Embryonic Fibroblast Motility and Orientation

can be Influenced by Physiological Electric Fields," Journal of Cell Biology, 98(1) pp. 296-

307.

[27] Sun, S., Titushkin, I., and Cho, M., 2006, "Regulation of Mesenchymal Stem Cell

Adhesion and Orientation in 3D Collagen Scaffold by Electrical Stimulus,"

Bioelectrochemistry, 69(2) pp. 133-141.

[28] Au, H. T. H., Cheng, I., Chowdhury, M. F., 2007, "Interactive Effects of Surface

Topography and Pulsatile Electrical Field Stimulation on Orientation and Elongation of

Fibroblasts and Cardiomyocytes," Biomaterials, 28(29) pp. 4277-4293.

[29] Guido, S., and Tranquillo, R. T., 1993, "A Methodology for the Systematic and

Quantitative Study of Cell Contact Guidance in Oriented Collagen Gels. Correlation of

Fibroblast Orientation and Gel Birefringence," Journal of Cell Science, 105(2) pp. 317-331.

[30] Tranquillo, R. T., Girton, T. S., Bromberek, B. A., 1996, "Magnetically Orientated Tissue-

Equivalent Tubes: Application to a Circumferentially Orientated Media-Equivalent,"

Biomaterials, 17(3) pp. 349-357.

111

Page 126: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

[31] Hashimoto, Y., Kawasumi, M., and Saito, M., 2007, "Effect of Static Magnetic Field on

Cell Migration," Electrical Engineering in Japan (English Translation of Denki Gakkai

Ronbunshi), 160(2) pp. 46-52.

[32] Goodman, R., and Blank, M., 2002, "Insights into Electromagnetic Interaction

Mechanisms," Journal of Cellular Physiology, 192(1) pp. 16-22.

[33] Freed, L. E., Guilak, F., Guo, X. E., 2006, "Advanced Tools for Tissue Engineering:

Scaffolds, Bioreactors, and Signaling," Tissue Engineering, 12(12) pp. 3285-3305.

[34] Bilodeau, K., and Mantovani, D., 2006, "Bioreactors for Tissue Engineering: Focus on

Mechanical Constraints. A Comparative Review," Tissue Engineering, 12(8) pp. 2367-2383.

[35] Chen, H. -., and Hu, Y. -., 2006, "Bioreactors for Tissue Engineering," Biotechnology

Letters, 28(18) pp. 1415-1423.

[36] Carrier, R. L., Rupnick, M., Langer, R., 2002, "Perfusion Improves Tissue Architecture of

Engineered Cardiac Muscle," Tissue Engineering, 8(2) pp. 175-188.

[37] Bancroft, G. N., Sikavitsas, V. I., and Mikos, A. G., 2003, "Design of a Flow Perfusion

Bioreactor System for Bone Tissue-Engineering Applications," Tissue Engineering, 9(3) pp.

549-554.

[38] Schwarz, R. P., Goodwin, T. J., and Wolf, D. A., 1992, "Cell Culture for Three-

Dimensional Modeling in Rotating-Wall Vessels: An Application of Simulated

Microgravity," Journal of Tissue Culture Methods, 14(2) pp. 51-57.

112

Page 127: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

[39] Leor, J., and Cohen, S., 2004, "Myocardial Tissue Engineering: Creating a Muscle Patch

for a Wounded Heart," Annals of the New York Academy of Sciences, 1015pp. 312-319.

[40] Carrier, R. L., Papadaki, M., Rupnick, M., 1999, "Cardiac Tissue Engineering: Cell

Seeding, Cultivation Parameters, and Tissue Construct Characterization," Biotechnology

and Bioengineering, 64(5) pp. 580-589.

[41] Sodian, R., Lemke, T., Fritsche, C., 2002, "Tissue-Engineering Bioreactors: A New

Combined Cell-Seeding and Perfusion System for Vascular Tissue Engineering," Tissue

Engineering, 8(5) pp. 863-870.

[42] Dumont, K., Yperman, J., Verbeken, E., 2002, "Design of a New Pulsatile Bioreactor for

Tissue Engineered Aortic Heart Valve Formation," Artificial Organs, 26(8) pp. 710-714.

[43] Seidel, J. O., Pei, M., Gray, M. L., 2004, "Long-Term Culture of Tissue Engineered

Cartilage in a Perfused Chamber with Mechanical Stimulation," Biorheology, 41(3-4) pp.

445-458.

[44] Vance, J., Galley, S., Liu, D. F., 2005, "Mechanical Stimulation of MC3T3 Osteoblastic

Cells in a Bone Tissue-Engineering Bioreactor Enhances Prostaglandin E2 Release," Tissue

Engineering, 11(11-12) pp. 1832-1839.

[45] Feng, Z., Matsumoto, T., Nomura, Y., 2005, "An Electro-Tensile Bioreactor for 3-D

Culturing of Cardiomyocytes," IEEE Engineering in Medicine and Biology Magazine, 24(4)

pp. 73-79.

113

Page 128: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

[46] Butler, D. L., Hunter, S. A., Chokalingam, K., 2009, "Using Functional Tissue Engineering

and Bioreactors to Mechanically Stimulate Tissue-Engineered Constructs," Tissue

Engineering - Part A, 15(4) pp. 741-749.

[47] Osman, N. F., McVeigh, E. R., and Prince, J. L., 2000, "Imaging Heart Motion using

Harmonic Phase MRI," IEEE Transactions on Medical Imaging, 19(3) pp. 186-202.

[48] Garvin, J., Qi, J., Maloney, M., 2003, "Novel System for Engineering Bioartificial Tendons

and Application of Mechanical Load," Tissue Engineering, 9(5) pp. 967-979.

[49] Soma, S., Matsumoto, S., and Takano-Yamamoto, T., 1997, "Enhancement by Conditioned

Medium of Stretched Calvarial Bone Cells of the Osteoclast-Like Cell Formation Induced

by Parathyroid Hormone in Mouse Bone Marrow Cultures," Archives of Oral Biology, 42(3)

pp. 205-211.

[50] Lee, C. H., Shin, H. J., Cho, I. H., 2005, "Nanofiber Alignment and Direction of

Mechanical Strain Affect the ECM Production of Human ACL Fibroblast," Biomaterials,

26(11) pp. 1261-1270.

[51] Gilbert, J. A., 1994, "Strain Profiles for Circular Cell Culture Plates Containing Flexible

Surfaces Employed to Mechanically Deform Cells in Vitro," Journal of Biomechanics, 27(9)

pp. 1169-1177.

[52] Sotoudeh, M., Jalali, S., Usami, S., 1998, "A Strain Device Imposing Dynamic and

Uniform Equi-Biaxial Strain to Cultured Cells," Annals of Biomedical Engineering, 26(2)

pp. 181-189.

114

Page 129: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

[53] Meikle, M. C., Reynolds, J. J., Sellers, A., 1979, "Rabbit Cranial Sutures in Vitro: A New

Experimental Model for Studying the Response of Fibrous Joints to Mechanical Stress,"

Calcified Tissue International, 28(2) pp. 137-144.

[54] Vandenburgh, H. H., Swasdison, S., and Karlisch, P., 1991, "Computer-Aided

Mechanogenesis of Skeletal Muscle Organs from Single Cells in Vitro," FASEB Journal,

5(13) pp. 2860-2867.

[55] Powell, C. A., Smiley, B. L., Mills, J., 2002, "Mechanical Stimulation Improves Tissue-

Engineered Human Skeletal Muscle," American Journal of Physiology - Cell Physiology,

283(5 52-5) pp. C1557-C1565.

[56] Yung, Y. C., Vandenburgh, H., and Mooney, D. J., 2009, "Cellular Strain Assessment Tool

(CSAT): Precision-Controlled Cyclic Uniaxial Tensile Loading," Journal of Biomechanics,

42(2) pp. 178-182.

[57] Zimmermann, W. H., Fink, C., Kralisch, D., 2000, "Three-Dimensional Engineered Heart

Tissue from Neonatal Rat Cardiac Myocytes," Biotechnology and Bioengineering, 68(1) pp.

106-114.

[58] Yost, M. J., Simpson, D., Wrona, K., 2000, "Design and Construction of a Uniaxial Cell

Stretcher," American Journal of Physiology - Heart and Circulatory Physiology, 279(6 48-6)

pp. H3124-H3130.

115

Page 130: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

[59] Joshi, S. D., and Webb, K., 2008, "Variation of Cyclic Strain Parameters Regulates

Development of Elastic Modulus in fibroblast/substrate Constructs," Journal of Orthopaedic

Research, 26(8) pp. 1105-1113.

[60] Webb, K., Hitchcock, R. W., Smeal, R. M., 2006, "Cyclic Strain Increases Fibroblast

Proliferation, Matrix Accumulation, and Elastic Modulus of Fibroblast-Seeded

Polyurethane Constructs," Journal of Biomechanics, 39(6) pp. 1136-1144.

[61] Wakatsuki, T., Kolodney, M. S., Zahalak, G. I., 2000, "Cell Mechanics Studied by a

Reconstituted Model Tissue," Biophysical Journal, 79(5) pp. 2353-2368.

[62] Wille, J. J., Elson, E. L., and Okamoto, R. J., 2006, "Cellular and Matrix Mechanics of

Bioartificial Tissues during Continuous Cyclic Stretch," Annals of Biomedical Engineering,

34(11) pp. 1678-1690.

[63] Langelier, E., Rancourt, D., Bouchard, S., 1999, "Cyclic Traction Machine for Long-Term

Culture of Fibroblast-Populated Collagen Gels," Annals of Biomedical Engineering, 27(1)

pp. 67-72.

[64] Mol, A., Bouten, C. V. C., Zünd, G., 2003, "The Relevance of Large Strains in Functional

Tissue Engineering of Heart Valves," Thoracic and Cardiovascular Surgeon, 51(2) pp. 78-

83.

[65] Pfister, B. J., Weihs, T. P., Betenbaugh, M., 2003, "An in Vitro Uniaxial Stretch Model for

Axonal Injury," Annals of Biomedical Engineering, 31(5) pp. 589-598.

116

Page 131: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

[66] Liu, M., Montazeri, S., Jedlovsky, T., 1999, "Bio-Stretch, a Computerized Cell Strain

Apparatus for Three-Dimensional Organotypic Cultures," In Vitro Cellular and

Developmental Biology - Animal, 35(2) pp. 87-93.

[67] Smalt, R., Mitchell, F. T., Howard, R. L., 1997, "Induction of NO and Prostaglandin E2 in

Osteoblasts by Wall-Shear Stress but Not Mechanical Strain," American Journal of

Physiology - Endocrinology and Metabolism, 273(4 36-4) pp. E751-E758.

[68] Liu, M., and Post, M., 2000, "Invited Review: Mechanochemical Signal Transduction in

the Fetal Lung," Journal of Applied Physiology, 89(5) pp. 2078-2084.

[69] Liu, M., Tanswell, A. K., and Post, M., 1999, "Mechanical Force-Induced Signal

Transduction in Lung Cells," American Journal of Physiology - Lung Cellular and

Molecular Physiology, 277(4 21-4) pp. L667-L683.

[70] Yang, X., Vezeridis, P. S., Nicholas, B., 2006, "Differential Expression of Type X

Collagen in a Mechanically Active 3-D Chondrocyte Culture System: A Quantitative

Study," Journal of Orthopaedic Surgery and Research, 1(1) .

[71] Auluck, A., Mudera, V., Hunt, N. P., 2005, "A Three-Dimensional in Vitro Model System

to Study the Adaptation of Craniofacial Skeletal Muscle Following Mechanostimulation,"

European Journal of Oral Sciences, 113(3) pp. 218-224.

[72] Boublik, J., Park, H., Radisic, M., 2005, "Mechanical Properties and Remodeling of

Hybrid Cardiac Constructs made from Heart Cells, Fibrin, and Biodegradable, Elastomeric

Knitted Fabric," Tissue Engineering, 11(7-8) pp. 1122-1132.

117

Page 132: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

[73] Shimizu, K., Ito, A., and Honda, H., 2007, "Mag-Seeding of Rat Bone Marrow Stromal

Cells into Porous Hydroxyapatite Scaffolds for Bone Tissue Engineering," Journal of

Bioscience and Bioengineering, 104(3) pp. 171-177.

[74] Yang, S. -., Sun, J. -., Liu, C. -., 2008, "Ex Vivo Magnetofection with Magnetic

Nanoparticles: A Novel Platform for Nonviral Tissue Engineering," Artificial Organs, 32(3)

pp. 195-204.

[75] Ito, A., Jitsunobu, H., Kawabe, Y., 2007, "Construction of Heterotypic Cell Sheets by

Magnetic Force-Based 3-D Coculture of HepG2 and NIH3T3 Cells," Journal of Bioscience

and Bioengineering, 104(5) pp. 371-378.

[76] Dobson, J., Cartmell, S. H., Keramane, A., 2006, "Principles and Design of a Novel

Magnetic Force Mechanical Conditioning Bioreactor for Tissue Engineering, Stem Cell

Conditioning, and Dynamic in Vitro Screening," IEEE Transactions on Nanobioscience,

5(3) pp. 173-177.

[77] Dobson, J., 2008, "Remote Control of Cellular Behaviour with Magnetic Nanoparticles,"

Nature Nanotechnology, 3(3) pp. 139-143.

[78] Moulder, J. E., 2000, "The Electric and Magnetic Fields Research and Public Information

Dissemination (EMF-RAPID) Program," Radiation Research, 153(5 II SUPPL.) pp. 613-

616.

[79] Edmonds, D. T., 1993, "Larmor Precession as a Mechanism for the Detection of Static and

Alternating Magnetic Fields," Bioelectrochemistry and Bioenergetics, 30(C) pp. 3-12.

118

Page 133: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

[80] Cane, V., Botti, P., and Soana, S., 1993, "Pulsed Magnetic Fields Improve Osteoblast

Activity during the Repair of an Experimental Osseous Defect," Journal of Orthopaedic

Research, 11(5) pp. 664-670.

[81] Pilla, A. A., 2002, "Low-Intensity Electromagnetic and Mechanical Modulation of Bone

Growth and Repair: Are they Equivalent?" Journal of Orthopaedic Science, 7(3) pp. 420-

428.

[82] Denegre, J. M., Valles Jr., J. M., Lin, K., 1998, "Cleavage Planes in Frog Eggs are Altered

by Strong Magnetic Fields," Proceedings of the National Academy of Sciences of the

United States of America, 95(25) pp. 14729-14732.

[83] Fassina, L., Visai, L., Benazzo, F., 2006, "Effects of Electromagnetic Stimulation on

Calcified Matrix Production by SAOS-2 Cells Over a Polyurethane Porous Scaffold,"

Tissue Engineering, 12(7) pp. 1985-1999.

[84] De Mattei, M., Pasello, M., Pellati, A., 2003, "Effects of Electromagnetic Fields on

Proteoglycan Metabolism of Bovine Articular Cartilage Explants," Connective Tissue

Research, 44(3-4) pp. 154-159.

[85] Funk, R. H. W., Monsees, T., and Özkucur, N., 2009, "Electromagnetic Effects - from Cell

Biology to Medicine," Progress in Histochemistry and Cytochemistry, 43(4) pp. 177-264.

[86] Miyakoshi, J., 2005, "Effects of Static Magnetic Fields at the Cellular Level," Progress in

Biophysics and Molecular Biology, 87(2-3 SPEC. ISS.) pp. 213-223.

[87] Stavroulakis, P., 2003, "Biological Effects of Electromagnetic Radiation,"Springer, Berlin, .

119

Page 134: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

[88] Komazaki, S., and Takano, K., 2007, "Induction of Increase in Intracellular Calcium

Concentration of Embryonic Cells and Acceleration of Morphogenetic Cell Movements

during Amphibian Gastrulation by a 50-Hz Magnetic Field," Journal of Experimental

Zoology Part A: Ecological Genetics and Physiology, 307(3) pp. 156-162.

[89] Anonymous 2008, "COMSOL Multiphysics User’s Guide," .

[90] Bansal, R., 2006, "Fundamentals of Engineering Electromagnetics," Taylor&Francis, .

[91] Magnetic Instruments Inc., "Gaussmeter model 2010 manual," Indianapolis.

[92] Skinner, S. J. M., Sommervell, C. E., and Olson, D. M., 1992, "The Effects of Mechanical

Stretching on Fetal Rat Lung Cell Prostacyclin Production," Prostaglandins, 43(5) pp. 413-

433.

[93] Liu, M., Xu, J., Souza, P., 1995, "The Effect of Mechanical Strain on Fetal Rat Lung Cell

Proliferation: Comparison of Two- and Three-Dimensional Culture Systems," In Vitro

Cellular and Developmental Biology - Animal, 31(11) pp. 858-866.

[94] Shimizu, T., Yamato, M., Isoi, Y., 2002, "Fabrication of Pulsatile Cardiac Tissue Grafts

using a Novel 3-Dimensional Cell Sheet Manipulation Technique and Temperature-

Responsive Cell Culture Surfaces." Circulation Research, 90(3) .

[95] Zimmermann, W. -., Melnychenko, I., Wasmeier, G., 2006, "Engineered Heart Tissue

Grafts Improve Systolic and Diastolic Function in Infarcted Rat Hearts," Nature Medicine,

12(4) pp. 452-458.

120

Page 135: APPARATUS FOR TISSUE ENGINEERING · circumstances, tissue engineering has emerged as a very promising approach to repair, replace or regenerate damaged tissues using tissue constructs

121

[96] Radisic, M., Yang, L., Boublik, J., 2004, "Medium Perfusion Enables Engineering of

Compact and Contractile Cardiac Tissue," American Journal of Physiology - Heart and

Circulatory Physiology, 286(2 55-2) pp. H507-H516.

[97] Kim, B. -., Nikolovski, J., Bonadio, J., 1999, "Cyclic Mechanical Strain Regulates the

Development of Engineered Smooth Muscle Tissue," Nature Biotechnology, 17(10) pp.

979-983.

[98] Shimko, V. F., and Claycomb, W. C., 2008, "Effect of Mechanical Loading on Three-

Dimensional Cultures of Embryonic Stem Cell-Derived Cardiomyocytes," Tissue

Engineering - Part A., 14(1) pp. 49-58.

[99] Okabo, T., and Matsuda, T., 1997, "Hybrid Muscular Tissues: Preparation of Skeletal

Muscle Cell- Incorporated Collagen Gels," Cell Transplantation, 6(2) pp. 109-118.

[100] Clark, C. B., Burkholder, T. J., and Frangos, J. A., 2001, "Uniaxial Strain System to

Investigate Strain Rate Regulation in Vitro," Review of Scientific Instruments, 72(5) pp.

2415-2422.

[101] Vande Geest, J. P., Di Martino, E. S., and Vorp, D. A., 2004, "An Analysis of the

Complete Strain Field within FlexercellTM Membranes," Journal of Biomechanics, 37(12)

pp. 1923-1928.

[102] Fung, Y.C., 1993, "Biomechanics Mechanical Properties of Living Tissues," Springer-

Veriag, New York, .