Antares Peripheral Instruction manual

150
© Siemens AG The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Print No.: Siemens Medical Solutions Ultrasound Division n.a. lne Antares Peripherals Installation Instructions US 2004 US03-201.812.01.02.02 Replaces: US03-201.812.01.01.02 English Doc. Gen. Date: 07.05 n.a. Part No.: n.a. sd Printer and VCR Configurations 05936518 08653771 10032746 10032747

description

Antares Peripheral Instruction manual

Transcript of Antares Peripheral Instruction manual

Page 1: Antares Peripheral Instruction manual

© Siemens AGThe reproduction, tranof this document or its contents is notpermitted without express writtenauthority. Offenders will be liable fordamages. All rights, including rightscreated by patent grant or registrationof a utility model or design, arereserved.

Siemens Medical SolutionsUltrasound Divisionn.a.lne

Antares

Peripherals

Installation Instructions

US

smission or use

2004

Print No.: US03-201.812.01.02.02Replaces: US03-201.812.01.01.02

EnglishDoc. Gen. Date: 07.05

n.a.

Part No.: n.a.sd

Printer and VCR Configurations

05936518086537711003274610032747

Page 2: Antares Peripheral Instruction manual

2 Revision / Disclaimer

1Revision / Disclaimer

Copyright

Copyright © 2004 by Siemens Corporation. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, stored in retrieval systems, or translated into any language or computer language, in any form or by any means, electronic, mechanical, magnetic, optical, chemical, manual, or otherwise, without the prior written permission of Siemens Corporation. Siemens Corporation reserves the right to change its products and services at any time. In addition, this manual is subject to change without notice. Siemens Corporation welcomes customer input on corrections and suggestions for improvements to this manual.Although Siemens Corporation has attempted to ensure accuracy through-out this manual, Siemens Corporation assumes no liability for any errors or omissions, nor for any damages resulting from the application or use of this information.

Trademarks

ACUSON, Sequoia, Cypress, AEGIS, Aspen, 128XP, XP, AcuNav, CV70, SONOLINE, Adara, Antares, G20, G40, G50, G60S, Omnia and KinetDx are trademarks of Siemens Corporation registered in the U.S. Patent and Trademark Office. CWS3000, DS3000, WS3000, and WebPro are trademarks of Siemens Corporation.

Windows XP, 2000 and NT are registered trademarks of Microsoft Corporation. Internet Explorer is a trademark of Microsoft Corporation. Netscape and Netscape Navigator are registered trademarks of Netscape Communications Corporation. Java and the JavaScript name are registered trademarks of Sun Microsystems, Inc. Adobe Acrobat is a registered trademark of Adobe Systems, Inc.

Other products and brand names are trademarks of their respective owners.

Disclaimer

The service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Sie-mens or one of its affiliates to provide such services.

Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro-cedures.

License Agreement

All computer programs copyright 1990-2004 by Siemens Corporation or its suppliers. Pro-grams are licensed under the following agreement:

Siemens or its suppliers retain(s) ownership of and title to any computer program supplied with the equipment and to the trade secrets embodied in such computer programs. Sub-ject to the Buyer’s acceptance and fulfillment of the obligations in this paragraph, Siemens grants the Buyer a personal, non-transferable, perpetual, non-exclusive license to use any computer program supplied with the Equipment that is necessary to operate the Equip-ment solely on the medium in which such program is delivered for the purpose of operat-ing the equipment in accordance with the instructions set forth in the operator’s manuals supplied with the Equipment and for no other purpose whatsoever. Buyer may not reverse-assemble, reverse-compile or otherwise reverse-engineer such computer pro-grams nor may Buyer make a copy of such program or apply any techniques to derive the trade secrets embodied therein. In the event of a failure by Buyer to comply with the terms

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 2 of 150Medical Solutions USA, Inc.

Page 3: Antares Peripheral Instruction manual

Revision / Disclaimer 3

of this license, the license granted by this paragraph shall terminate. Further, because unauthorized use of such computer programs will leave Siemens without an adequate remedy at law, Buyer agrees that injunctive or other equitable relief will be appropriate to restrain such use, threatened or actual. Buyer further agrees that (i) any of Siemens"s suppliers of software is a direct and intended beneficiary of this end-user sublicense and may enforce it directly against Buyer with respect to software supplied by such supplier, and (ii) No supplier of Siemens shall be liable to buyer for any general, special, direct, indi-rect, consequential, incidental or other damages arising out of the sublicense of the com-puter programs supplied with the equipment.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 3 of 150Medical Solutions USA, Inc.

Page 4: Antares Peripheral Instruction manual

4 Table of Contents

1- 0Table of Contents

1 _______ Overview_______________________________________________________ 6

Report Printers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

Mitigating the Risk of Connecting a Report Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

Patient Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

2 _______ Hardware Configuration _________________________________________ 14

Single Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14CP770 or CP800 Color Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14UP-D23MD Color Printer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17UP-D55MD Color Printer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20P91 D or P93 D B/W Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23HS-MD3000 VCR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26SVO9500 VCR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

Double Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34B/W Printer and Color Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34B/W Printer and VCR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47Color Printer and VCR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62

Peripheral Shelf Mounting Hole Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95

I/O Module Peripheral Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97Printer I/O Module Cable Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97VCR I/O Module Cable Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98

3 _______ Printer Setup and Configuration __________________________________ 99

SW v2.1 and V3.0. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100CP800 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100CP770DW, P91DW, P93DW, UPD23MD, and UPD55MD Installation . . . . . . . . . . 107HP1150, HP2500n, HP4050, HP4100, HP4200, and HP6122 Installation . . . . . . . 108

SW v4.0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110CP800 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110CP770DW, P91DW, P93DW, UPD23MD, and UPD55MD Installation . . . . . . . . . . 118HP4050, HP4100 and HP4200 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120HP1150, HP2500n and HP6122 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121

Reboot System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122

Print/Store Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123Restore Factory Settings and Factory Printer Settings . . . . . . . . . . . . . . . . . . . . . . 123Configure Routing to Print and Store Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125Configure Print . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128Configure Store . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129

Verify Printer Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130

4 _______ VCR Setup and Configuration ___________________________________ 131

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 4 of 150Medical Solutions USA, Inc.

Page 5: Antares Peripheral Instruction manual

Table of Contents 5

HS-MD3000 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132Front Panel Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133

SVO9500 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138Front Panel Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138Main Menu Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139Hidden Menu Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140

Verify the Operation of the VCR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142

5________ Printer Troubleshooting _________________________________________143

Printer Installation Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143

Restoring Plug-and-Play Resources . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145

Re-Install A Printer After Troubleshooting or Restoring Plug-and-Play. . . . . . . . . . . . . 147Report Printers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147Re-install Plug-and-Play Printers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 5 of 150Medical Solutions USA, Inc.

Page 6: Antares Peripheral Instruction manual

6 Overview

2- 1Overview

The Peripheral Installation procedure lists the latest peripheral configurations for installa-tion and describes how to install each configuration for the the following Antares systems:

• ACUSON Antares ultrasound system

• ACUSON Antarespe ultrasound system

• SONOLINE Antares ultrasound system

Install an onboard or offboard single configuration or double configuration on the Antares peripheral shelf. The peripherals consist of color printers, black and white (B/W) printers, and VCRs.

The following table lists the onboard peripheral image printers and VCR and their software version installation types.

Tab. 1 ONboard Peripheral Image Printers and VCRs - Software Installation Types:

The following table lists the offboard peripheral report printers and their software version installation types.

Peripheral SONOLINE Antares ACUSON Antares ACUSON AntaresPre-mier Edition

Software Versions

v2.1 v3.0 v4.0 v4.0 v4.0

P91 D X X X NA NA

P93 D X X X X X

CP770DW X X X X X

CP800DW X X X1

1. CP800 is not supported on systems configured with the RM-165 PCB.

NA NA

UP-D23MD X X X X X

UP-D55MD NA X X NA X

HS-MD3000 X2

2. Valid with combined code installed.

X X X X

SVO9500 X X X NA NA

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 6 of 150Medical Solutions USA, Inc.

Page 7: Antares Peripheral Instruction manual

Overview 7

Tab. 2 Report Printers:

Before installing report printers, comply with the guidellines in the following sections:

• (Report Printers / p. 8)

• (Mitigating the Risk of Connecting a Report Printer / p. 10)

• (Patient Environment / p. 11)

Peripheral SONOLINE Antares ACUSON Antares ACUSON AntaresPre-mier Edition

Software Versions

v2.1 v3.0 v4.0 v4.0 v4.0

HP1150 X X X X X

HP2500n X X X X X

HP4050 X X X X X

HP4100 X X X X X

HP4200 X X X X X

HP6122 DeskJet

X X X X X

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 7 of 150Medical Solutions USA, Inc.

Page 8: Antares Peripheral Instruction manual

8 Overview

Report Printers 0

WARNING Accessory equipment connected to the USB port must be certifiedaccording to the respective EN and IEC standards (for example, EN60950 and IEC 60950 for data processing equipment and EN60601-1 and IEC 60601-1 for medical equipment). Furthermore, allconfigurations shall comply with the system standard EN60601-1-1 and IEC 60601-1-1.

Anyone who connects additional equipment to the signal input or signal output port configures a medical system and is therefore responsible that the system complies with the requirements of the system standard EN 60601-1-1 and IEC 60601-1-1. Siemens can only guaran-tee the performance and safety of the devices listed in the Accessories and Options chapter of the System Ref-erence (shipped with the ultrasound imaging system). If in doubt, consult the Siemens Uptime Service Center.

WARNING Equipment connected to the ultrasound system and in the patientenvironment must be powered from a medically isolated powersource or must be a medically isolated device.

Equipment powered from a non-isolated source can result in chassis leakage currents exceeding safe levels. Chassis leakage current created by an accessory or device connected to a non-isolated outlet may add to the chassis leakage current of the ultrasound system.

WARNING Non-medical grade report printers cannot be used within a patientenvironment.

During use of a non-medical grade report printer or when a non-medical grade report printer is connected to the ultrasound system, the ultrasound system cannot in any way be in contact with a patient.

NOTE Siemens cannot support or service any external devices connect-ed to the USB port. Siemens cannot assume responsibility forfunctionality beyond the scope of this specification.

NOTE Siemens does not specifically recommend any particular commu-nication or analysis software; please contact a local vendor for as-sistance with interpretation and analysis of the data transmitted bythe ultrasound imaging system.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 8 of 150Medical Solutions USA, Inc.

Page 9: Antares Peripheral Instruction manual

Overview 9

Relevant standards for some non-medical electrical equipment may have limits for enclo-sure leakage currents higher than required by medical standards. These higher standards are acceptable only outside the patient environment. It is essential to reduce enclosure leakage currents when non-medical electrical equipment is to be used within the patient environment. Measures for reducing leakage current include use of a medically-approved isolation transformer.

The ultrasound system has printer drivers compatible with non-medical grade report print-ers. Siemens does not guarantee the performance and safety of any non-medical grade report printer. Non-medical grade report printers do not fulfill the following safety require-ments:

• EN 60601-1-1 and IEC 60601-1-1 (Medical Electrical Equipment, Part 1: General Re-quirements for Safety).

• EN 60601-1-2 and IEC 60601-1-2 (Electromagnetic Compatibility of Medical Devices).

If a non-medical grade report printer is to be used with the ultrasound system, then ensure mitigation is provided to meet all safety requirements. It is the responsibility of the user to ensure that the ultrasound system in combination with the non-medical grade report printer complies with safety requirements.

Always use a medically-approved isolation transformer with a non-medical grade report printer.

Mitigate risk with regard to the Electromagnetic Compatibility of Medical Devices require-ment. A non-medical grade device must be designed, manufactured, and certified to meet the same EMC (electromagnetic compatibility) requirements as the ultrasound system, or other means must ensure that the overall EMC requirements are met.

Tab. 3 Manufacturers of Medically Approved Isolation Transformers

Manufacturer Web Site

Tripp Lite www.tripplite.com

Toroid Corporation of Maryland www.toroid.com

Dale Technology www.daletech.com

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 9 of 150Medical Solutions USA, Inc.

Page 10: Antares Peripheral Instruction manual

10 Overview

Mitigating the Risk of Connecting a Report Printer 0

To fulfill EN 60601-1-1 and IEC 60601-1-1 (Medical Electrical Equipment, Part 1: General Requirements for Safety) requirements for non-medical peripheral equipment:

The non-medical peripheral equipment must be approved according to any other EN or IEC standard (EN XXXXX or IEC XXXXX, e.g., equipment complying with EN 60348 and IEC 60348, EN 60950 and IEC 60950, etc.)

The connection of non-medical peripheral equipment to the ultrasound system must adhere to the following conditions:

1. Connect the ultrasound system to an AC power outlet within a medically used room within the patient environment. A patient environment is defined as an area in which medical examination, monitoring, or treatment of the patient takes place. The patient environment is located 1.5 meters (or, in Canada and the U.S.A., 1.8 meters [6 feet]) around the patient location.

2. Connect the peripheral equipment AC power cable to a medically approved isolation transformer.

3. Connect the medically approved isolation transformer to a main AC outlet either inside or outside the patient environment. The peripheral device and medically approved iso-lation transformer combination can be located:

- Within the same room as the ultrasound system, or

- In a non-medically used room.

For additional information and other possible combinations, please refer to the Medical Electrical Equipment Standard EN 60601-1-1 or IEC 60601-1-1, Annex BBB.7, Scenario 3c.

NOTE The above information is based on current EN 60601-1-1 and IEC60601-1-1 standards, dated 2000-12. If a country"s regulatory stan-dards for medical equipment do not correspond to EN 60601-1 andIEC 60601-1, as well as EN 60601-1-1 and IEC 60601-1-1, then thelocal requirements may differ.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 10 of 150Medical Solutions USA, Inc.

Page 11: Antares Peripheral Instruction manual

Overview 11

Patient Environment 0

Fig. 1: Patient Environment (Sheet 1)

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 11 of 150Medical Solutions USA, Inc.

Page 12: Antares Peripheral Instruction manual

12 Overview

Fig. 2: Patient Enviroment (Sheet 2)

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 12 of 150Medical Solutions USA, Inc.

Page 13: Antares Peripheral Instruction manual

Overview 13

Requirements 0

Time required:

• Software setup and configuration: 10 minutes per device

• Single hardware configuration: 30 minutes

• Double hardware configuration: 60 minutes

Tools:

• Pen

• Exacto knife or 1/4-inch hollow punch

• Phillips head screw driver

• Flat head screw driver

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 13 of 150Medical Solutions USA, Inc.

Page 14: Antares Peripheral Instruction manual

14 Hardware Configuration

3- 2Hardware ConfigurationSingle Configuration 0

Tab. 4 Single Configurations

CP770 or CP800 Color Printer 0

1. Cut the fastener holes into the rubber mat of the peripheral shelf.

See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

2. Return the mat to the peripheral shelf.

3. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.

4. Remove the feet and the shipment locking screw from the bottom of the printer.

a) Point the bottom side of the printer face up.

b) Remove the four screws that secure the feet to the printer and remove the feet.

c) Remove the shipment locking screw.

Fig. 3: CP770 or CP800 Printer - Shipping screw removal

Color Printers B/W Printers VCRs

(CP770 or CP800 Color Printer / p. 14)

(UP-D23MD Color Printer / p. 17)

(UP-D55MD Color Printer / p. 20)

(P91 D or P93 D B/W Printer / p. 23)

(HS-MD3000 VCR / p. 26)

(SVO9500 VCR / p. 31)

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 14 of 150Medical Solutions USA, Inc.

Page 15: Antares Peripheral Instruction manual

Hardware Configuration 15

5. Secure the mounting bracket to the color printer.

a) Put the mounting bracket on the printer.

b) Align the mounting holes.

c) Secure the bracket to the bottom of the color printer.

Fig. 4: CP770 or CP800 - Mounting Bracket

WARNING Use fasteners of the correct length.

Using fasteners that are too long may cause the hard-ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.

6. Put the color printer assembly on the peripheral shelf and align the mounting holes of the color printer with the mounting holes in the mat.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 15 of 150Medical Solutions USA, Inc.

Page 16: Antares Peripheral Instruction manual

16 Hardware Configuration

7. Set the color printer dip switches as follows:

8. Connect the color printer cables.

a) Connect the power cable to the printer AC power inlet and to the peripheral AC pow-er outlet.

b) Connect the USB cable to the back of the printer.

Fig. 7: CP770 or CP800 - Cable connectionsPos. 1 USB Cable (CP800 may have a parrallel cable)

Pos. 2 Power cord

c) Route the other end of the USB cable through the peripheral shelf cable passage to the I/O module.

9. Install the cover.

a) Put the cover over the printer.

b) Align the four captive screws in the cover with the mounting holes.

c) Tighten the four captive screws.

Fig. 5: CP770 Dip Switch Settings to OFF

1-OFF, 2-OFF, 3-OFF, 4-OFF,5-OFF, 6-OFF, 7-OFF, and 8-OFF.

Fig. 6: CP800 Dip Switch Settings

1-ON, 2-OFF, 3-OFF, 4-OFF,5-OFF, 6-OFF, 7-OFF, and 8-ON.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 16 of 150Medical Solutions USA, Inc.

Page 17: Antares Peripheral Instruction manual

Hardware Configuration 17

10. See the Mitsubishi User Instruction to perform the following tasks:

- Load the ink cartridge

- Load the paper tray

- Slide the paper tray into the printer

11. Complete the printer cable connections to the I/O module, see (I/O Module Peripheral Connections / p. 97).

12. To complete the color printer configuration, see (Printer Setup and Configuration / p. 99).

UP-D23MD Color Printer 0

1. Cut the fastener holes into the rubber mat of the peripheral shelf.

See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

2. Return the mat to the peripheral shelf.

3. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 17 of 150Medical Solutions USA, Inc.

Page 18: Antares Peripheral Instruction manual

18 Hardware Configuration

4. Put the printer feet into the mounting bracket cut-outs.

Fig. 8: Printer Foot Placement LocationsPos. 1 Rear mounting plate cutout for placement of the printer’s rear feet

Pos. 2 Front mounting plate cutout for placement of the printer’s front feet

Fig. 9: Printer Placement on Mounting Plate Pos. 1 Cable routing passage through the mat and panel to the I/O module

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 18 of 150Medical Solutions USA, Inc.

Page 19: Antares Peripheral Instruction manual

Hardware Configuration 19

5. Connect the printer cables.

Fig. 10: Cable Connections - AC Power and USBPos. 1 Antares OEM AC power outlet

Pos. 2 Printer AC power inlet

Pos. 3 Printer USB connection

a) Connect the AC power cable to the Antares peripheral AC power outlet and to the printer AC power inlet.

b) Connect the USB cable to the back of the printer.

c) Route the other end of the USB cable through the I/O module cable routing pas-sage.

(See (1/Fig. 9 / p. 18) for the routing passage location.)

NOTE Do not connect the USB cable to the I/O module at this time.

Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the printer cover.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 19 of 150Medical Solutions USA, Inc.

Page 20: Antares Peripheral Instruction manual

2 0 Hardware Configuration

6. Install the printer cover.

Fig. 11: Printer Cover - InstalledPos. 1 Captive screw

Pos. 2 Captive screw locations at base of cover (not in view)

a) Put the printer cover over the printer.

b) Align the three captive screws with the three fastener holes cut into the mat.

c) Tighten the captive screws

7. See the Sony User Instruction to perform the following tasks:

- Load the ink cartridge

NOTE Power on the system prior to loading the ink cartridge. This resetsthe gears to their default location, permitting the ink cartridge to beinserted.

- Load the paper tray

- Slide the paper tray into the printer

8. Complete the printer cable connections to the I/O module, see (I/O Module Peripheral Connections / p. 97).

9. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).

UP-D55MD Color Printer 0

1. Cut the fastener holes into the rubber mat of the peripheral shelf.

See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

2. Return the mat to the peripheral shelf.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 20 of 150Medical Solutions USA, Inc.

Page 21: Antares Peripheral Instruction manual

Hardware Configuration 21

3. Secure the printer bracket to the bottom of the color printer.

Fig. 12: UP-D55M Color Printer Bracket

NOTE In (Fig. 12 / p. 21), the flat side of the bracket is oriented on the leftside of the picture.

a) Point the bottom side of the printer face up.

b) Put the bracket on the bottom of the printer.

c) Tighten the bracket captive screw.

4. Align the mounting holes of the mounting bracket to the holes in the peripheral mat.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 21 of 150Medical Solutions USA, Inc.

Page 22: Antares Peripheral Instruction manual

22 Hardware Configuration

5. Connect the printer cables.

Fig. 13: UP-D55M Cable ConnectionsPos. 1 Power cord between the OEM AC power outlet and the printer AC power inlet

Pos. 2 USB cable

a) Connect the AC power cable to the Antares peripheral AC power outlet and to the printer AC power inlet.

b) Connect the USB cable to the back of the printer.

c) Route the other end of the USB cable through the I/O module cable routing pas-sage.

NOTE Do not connect the USB cable to the I/O module at this time.

Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the printer cover.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 22 of 150Medical Solutions USA, Inc.

Page 23: Antares Peripheral Instruction manual

Hardware Configuration 23

6. Install the printer cover.

Fig. 14: UP-D55 M Cover - Installed Pos. 1 Captive screw location

Pos. 2 Captive screw locations at base of cover (not in veiw)

a) Put the printer cover over the printer.

b) Align the three captive screws with the three fastener holes cut into the mat.

c) Tighten the captive screws

7. See the Sony User Instruction to perform the following tasks:

- Load the ink cartridge

- Load the paper tray

- Slide the paper tray into the printer

8. Complete the printer cable connections to the I/O module, see (I/O Module Peripheral Connections / p. 97).

9. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).

P91 D or P93 D B/W Printer 0

1. Cut the fastener holes into the rubber mat of the peripheral shelf.

See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

2. Return the mat to the peripheral shelf.

3. Put the B/W printer bracket on the B/W printer mounting plate. Use the applicable in-structions that follow.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 23 of 150Medical Solutions USA, Inc.

Page 24: Antares Peripheral Instruction manual

24 Hardware Configuration

WARNING Use fasteners of the correct length.

Using fasteners that are too long may cause the hard-ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.

P91 D Instructions: P93 D Instructions:

• Put the flat side of the bracket face down on the peripheral mat.

• Align the bracket mounting holes to the peripheral mat holes.

• Put the P91 B/W printer on top of the P91 mounting bracket.

• Align the B/W printer feet against the mounting bracket.

Fig. 15: P91 D Positioned on the Bracket

• Point the bottom side of the P93 D print-er up.

• Put the P93 D mounting bracket on the bottom of the P93 D printer.

Fig. 16: P93 D Mounting Bracket

• Secure the P93 D printer mounting bracket to the P93 D printer.

• Point the bottom side of the printer fac-ing down.

• Put the Printer assembly on the periph-eral mat.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 24 of 150Medical Solutions USA, Inc.

Page 25: Antares Peripheral Instruction manual

Hardware Configuration 25

4. Set the printer B/W dip switches:

- P91 D settings: 1-ON, 2-OFF, 3-OFF, and 4-OFF- P93 D settings: 1-ON, 2-OFF, 3-ON, and 4-OFF

Fig. 17: P91 D and P93 D Printer Cables and Dip Switches (P91 D shown)Pos. 1 Dip Switches

Pos. 2 USB Cable

Pos. 3 AC Power Cord

5. Connect the printer cables.

a) Connect the AC power cable to the Antares peripheral AC power outlet and to the printer AC power inlet.

b) Connect the USB cable to the back of the printer.

c) Route the other end of the USB cable through the I/O module cable routing pas-sage.

NOTE Do not connect the USB cable to the I/O module at this time.

Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the printer cover.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 25 of 150Medical Solutions USA, Inc.

Page 26: Antares Peripheral Instruction manual

26 Hardware Configuration

6. Install the printer cover.

Fig. 18: P91 D and P93 D Printer Cover Installed

a) Put the cover over the printer.

b) Tighten the two captive screws.

c) Make sure the back of the printer touches the printer cover stop.

7. Complete the printer cable connections to the I/O module, see (I/O Module Peripheral Connections / p. 97).

8. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).

HS-MD3000 VCR 0

Fig. 19: HS-MD3000 VRC - Single ConfigurationPos. 1 VCR bracket

Pos. 2 VCR

Pos. 3 Cover

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 26 of 150Medical Solutions USA, Inc.

Page 27: Antares Peripheral Instruction manual

Hardware Configuration 27

1. Cut the fastener holes into the rubber mat of the peripheral shelf.

See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

2. Return the mat to the peripheral shelf.

3. Remove the feet from the bottom of the VCR.

Fig. 20: VCR Foot Locations

a) Point the bottom side of the VCR facing up.

b) Remove the four screws that secure the feet to the VCR.

c) Remove the feet.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 27 of 150Medical Solutions USA, Inc.

Page 28: Antares Peripheral Instruction manual

28 Hardware Configuration

4. Secure the mounting bracket to the VCR.

Fig. 21: HS-MD3000 Secured to the Mounting Bracket

WARNING Use fasteners of the correct length.

Using fasteners that are too long may cause the hard-ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.

NOTE In (Fig. 21 / p. 28), the front of the VCR is facing the left side of thepicture.

a) Put the mounting bracket on the VCR.

b) Align the mounting holes.

c) Secure the bracket to the bottom of the VCR.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 28 of 150Medical Solutions USA, Inc.

Page 29: Antares Peripheral Instruction manual

Hardware Configuration 29

WARNING Use fasteners of the correct length.

Using fasteners that are too long may cause the hard-ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.

5. Put the VCR assembly on the peripheral shelf by aligning the mounting holes of the VCR with the mounting holes in the mat.

Fig. 22: HS-MD3000 Positioned on the OEM Bay

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 29 of 150Medical Solutions USA, Inc.

Page 30: Antares Peripheral Instruction manual

30 Hardware Configuration

6. Connect the VCR power cord to the peripheral AC power outlet and to the VCR AC power inlet.

Fig. 23: HS-MD3000 Power Cord Connection Pos. 1 OEM power outlets

Pos. 2 VCR power inlet

7. Connect the VCR cables. See (VCR I/O Module Cable Connections / p. 98) for details.

Fig. 24: HS-MD3000 VCR Connections from the I/O ModulePos. 1 Connect P128 to RS-232C

Pos. 2 Connect VIDEO IN

Pos. 3 Connect VIDEO OUT

Pos. 4 Connect AUDIO IN

Pos. 5 Connect AUDIO OUT

Pos. 6 Set the 75Ω VIDEO and the 75Ω S-VIDEO switches to ON

8. Route the cables through the I/O panel cable routing passage.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 30 of 150Medical Solutions USA, Inc.

Page 31: Antares Peripheral Instruction manual

Hardware Configuration 31

NOTE Do not connect the cables to the I/O module at this time.

Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the VCR cover.

9. Install the cover.

Fig. 25: VCR Cover Captive Screw Locations

a) Put the cover over the VCR.

b) Align the three captive screws in the cover with the mounting holes.

c) Tighten the three captive screws.

10. Complete the VCR cable connections to the I/O module, see (I/O Module Peripheral Connections / p. 97).

11. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).

SVO9500 VCR 0

1. Cut the fastener holes into the rubber mat of the peripheral shelf.

See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

2. Secure the mounting bracket to the peripheral shelf.

3. Put the VCR assembly on the peripheral shelf by aligning the mounting holes of the VCR with the mounting holes in the mat.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 31 of 150Medical Solutions USA, Inc.

Page 32: Antares Peripheral Instruction manual

32 Hardware Configuration

4. Set the VCR switches: 1-OFF, 2-OFF, 3-OFF, 4-OFF, 5- ON, 6-ON

Fig. 26: SVO9500 Dip Switch Settings

- Set both 75 ohm switches to ON.

- For PAL model only, set the voltage to 240V.

NOTE Only the PAL model includes the 240V switch.

5. Connect the VCR power cord to the peripheral AC power outlet and to the VCR AC power inlet.

6. Connect the VCR cables. See (VCR I/O Module Cable Connections / p. 98) for details.

Fig. 27: SVC 9500 VCR Connections from I/O ModulePos. 1 Audio In (left and right)

Pos. 2 Video In (use either connector)

Pos. 3 Audio Out (left and right)

Pos. 4 Video Out

Pos. 5 Dip Switches

Pos. 6 RS-232C (connect to P128)

Pos. 7 75 Ω (set both switches to ON)

7. Route the cables through the I/O panel cable routing passage.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 32 of 150Medical Solutions USA, Inc.

Page 33: Antares Peripheral Instruction manual

Hardware Configuration 33

NOTE Do not connect the cables to the I/O module at this time.

Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the VCR cover.

8. Install the cover.

a) Put the cover over the VCR.

b) Align and tighten the two captive screws in the cover with the mounting holes.

9. Complete the VCR cable connections to the I/O module, see (VCR I/O Module Cable Connections / p. 98).

10. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 33 of 150Medical Solutions USA, Inc.

Page 34: Antares Peripheral Instruction manual

34 Hardware Configuration

Double Configuration 0

Tab. 5 Double Configuration

B/W Printer and Color Printer 0

P91 D or P93 D / CP770 or CP800

Installation Summary

1. Install the color printer on the peripheral shelf.

2. Install the B/W printer on the color printer.

3. Complete the cable I/O module connections.

4. Complete the printer configuration.

NOTE When performing this upgrade on a system that already has a B/Wprinter or a color printer installed, remove the printer from the pe-ripheral shelf.

Do not disconnect the USB cable from the I/O module. The systemhas previously configured the USB port for the printer connected.Therefore, the same USB port must be used for that printer whenthe USB cable is reconnected.

Label the USB cable for the correct reconnection to the printer.

1. Cut the fastener holes into the rubber mat of the peripheral shelf.

See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

2. Return the mat to the peripheral shelf.

3. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.

B/W Printer/Color Printer

Double Configuration

B/W Printer/VCR

Double Configuration

Color Printer/VCR

Double Configuration

(P91 D or P93 D / CP770 or CP800 / p. 34)

(P91 D or P93 D / HS-MD3000 / p. 47)

(CP770 or CP800 / HS-MD3000 / p. 62)

(P91 D or P93 D / UP-D23MD / p. 37)

(P91 D or P93 D / SVO9500 / p. 56)

(UP-D23MD / HS-MD 3000 / p. 75)

(P91 D or P93 D / UP-D55M / p. 42)

(UP-D55MD / HS-MD3000 / p. 86)

(CP770 or CP800 / SVO9500 / p. 67)

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 34 of 150Medical Solutions USA, Inc.

Page 35: Antares Peripheral Instruction manual

Hardware Configuration 35

4. Remove the feet and the shipment locking screw from the bottom of the printer.

a) Point the bottom side of the printer face up.

b) Remove the four screws that secure the feet to the printer and remove the feet.

Fig. 28: CP770 or CP800 Secured to the Mounting Bracketc) Remove the shipment locking screw.

5. Secure the mounting bracket to the color printer.

a) Put the mounting bracket on the printer.

b) Align the mounting holes.

c) Secure the bracket to the bottom of the color printer.

WARNING Use fasteners of the correct length.

Using fasteners that are too long may cause the hard-ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.

6. Put the color printer assembly on the peripheral shelf and align the mounting holes of the color printer with the mounting holes in the mat.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 35 of 150Medical Solutions USA, Inc.

Page 36: Antares Peripheral Instruction manual

36 Hardware Configuration

7. Set the color printer dip switches as follows:

8. Connect the color printer cables.

a) Connect the color printer AC power cable to the color printer AC power inlet and to the lower peripheral AC power outlet.

b) Connect the B/W printer AC power cable to the upper peripheral AC power outlet.

c) Connect the color printer USB cable to the back of the color printer.

d) Route the other end of the color printer USB cable through the peripheral shelf ca-ble passage to the I/O module.

e) Route the I/O module end of the B/W printer USB cable through the peripheral shelf cable passage to the I/O module.

NOTE Do not connect the USB cables to the I/O module at this time, un-less the system previously had one of the printers installed on theperipheral bay. Then, connect the previously installed printer tothe same port it was connected to before.

9. Install the color printer cover/black and white printer mounting plate assembly.

a) Route the B/W printer USB cable and power cable up through the hole in the B/W printer mounting plate.

b) Put the printer cover over the printer.

c) Align the four captive screws with the fasener holes cut into the mat.

d) Tighten the captive screws.

10. Set the B/W printer dip switches to:

- P91 D settings: 1-ON, 2-OFF, 3-OFF, and 4-OFF- P93 D settings: 1-ON, 2-OFF, 3-ON, and 4-OFF

11. Put the B/W printer bracket on the B/W printer mounting plate. Use the applicable in-structions that follow.

Fig. 29: CP770 Dip Switch Settings to OFF

1-OFF, 2-OFF, 3-OFF, 4-OFF,5-OFF, 6-OFF, 7-OFF, and 8-OFF.

Fig. 30: CP800 Dip Switch Settings

1-ON, 2-OFF, 3-OFF, 4-OFF,5-OFF, 6-OFF, 7-OFF, and 8-ON.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 36 of 150Medical Solutions USA, Inc.

Page 37: Antares Peripheral Instruction manual

Hardware Configuration 37

12. Install the B/W printer.

a) Put the B/W printer on the bracket and mounting bracket.

b) Connect the AC power cable to the back of the B/W printer.

c) Connect the USB cable to the back of the B/W printer.

NOTE Do not connect the USB cable to the I/O module at this time.

13. Install the B/W printer cover.

a) Put the B/W cover over the B/W printer.

b) Tighten the captive screws (of the B/W printer cover) through the mounting bracket, to the mounting plate.

c) Make sure the back of the B/W printer touches the B/W printer cover stop.

14. See the Sony User Instruction to perform the following tasks for the color printer:

- Load the ink cartridge

NOTE Power on the system prior to loading the ink cartridge. This resetsthe gears to their default location, permitting the ink cartridge to beinserted.

- Load the paper tray

- Slide the paper tray into the printer

15. Complete the printer cable connections to the I/O module, see (I/O Module Peripheral Connections / p. 97).

16. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).

P91 D or P93 D / UP-D23MD

Installation Summary

1. Install the color printer on the peripheral shelf

2. Install the B/W printer on the color printer.

3. Complete the cable I/O module connections.

4. Complete the printer configuration.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 37 of 150Medical Solutions USA, Inc.

Page 38: Antares Peripheral Instruction manual

38 Hardware Configuration

NOTE When performing this upgrade on a system that already has a B/Wprinter or a color printer installed, remove the printer from the pe-ripheral shelf.

Do not disconnect the USB cable from the I/O module. The systemhas previously configured the USB port for the printer connected.Therefore, the same USB port must be used for that printer whenthe USB cable is reconnected.

Label the USB cable for the correct reconnection to the printer.

1. Cut the fastener holes into the rubber mat of the peripheral shelf.

See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

2. Return the mat to the peripheral shelf.

3. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.

4. Put the color printer feet into the mounting bracket cut-outs.

Fig. 31: Printer Foot Placement LocationsPos. 1 Rear mounting plate cutout for placement of the printer’s rear feet

Pos. 2 Front mounting plate cutout for placement of the printer’s front feet

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 38 of 150Medical Solutions USA, Inc.

Page 39: Antares Peripheral Instruction manual

Hardware Configuration 39

5. Connect the color and B/W power cables to the Antares AC outlet.

Fig. 32: UP-D23MD Printer Placement and Cable ConnectionsPos. 1 OEM Bay to I/O Module cable routing passage

Pos. 2 AC power outlet UP-D23MD

Pos. 3 AC power outlet P91 D/P93 D

Pos. 4 Power inlet UP-D23MD

Pos. 5 USB cable connection UP-D23MD

a) Connect the AC power cable to the Antares peripheral AC power outlet and to the color printer AC power inlet.

b) Connect the B/W printer AC power cable to the Antares peripheral AC power outlet and route the cable through the color printer top cover.

c) Connect the USB cable to the back of the color printer.

d) Route both of the USB cables through the I/O module cable routing passage.

NOTE Do not connect the USB cable to the I/O module at this time, unlessthe system previously had this printer installed.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 39 of 150Medical Solutions USA, Inc.

Page 40: Antares Peripheral Instruction manual

40 Hardware Configuration

6. Install the printer cover.

Fig. 33: UP-D23MD Printer Placement and Cable ConnectionsPos. 1 Mounting plate B/W printer

Pos. 2 Cable routing holes

Pos. 3 Captive screw locations

a) Put the printer cover over the printer.

b) Align the three captive screws with the three fastener holes cut into the mat.

c) Tighten the captive screws

NOTE Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the printer cover.

7. Set the B/W printer dip switches to:

- P91 D settings: 1-ON, 2-OFF, 3-OFF, and 4-OFF- P93 D settings: 1-ON, 2-OFF, 3-ON, and 4-OFF

8. Put the B/W printer bracket on the B/W printer mounting plate. Use the applicable in-structions that follow.

WARNING Use fasteners of the correct length.

Using fasteners that are too long may cause the hard-ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 40 of 150Medical Solutions USA, Inc.

Page 41: Antares Peripheral Instruction manual

Hardware Configuration 41

9. Install the B/W printer.

a) Put the B/W printer on the bracket and mounting bracket.

b) Connect the AC power cable to the back of the B/W printer.

c) Connect the USB cable to the back of the B/W printer.

NOTE Do not connect the USB cable to the I/O module at this time.

10. Install the B/W printer cover.

a) Put the B/W cover over the B/W printer.

b) Tighten the captive screws (of the B/W printer cover) through the mounting bracket, to the mounting plate.

c) Make sure the back of the B/W printer touches the B/W printer cover stop.

P91 D Instructions: P93 D Instructions:

• Put the flat side of the bracket face down on the B/W printer mounting plate of the color printer cover.

• Align the bracket mounting holes to the plate mounting holes.

• Put the P91 B/W printer on top of the P91 mounting bracket.

• Align the B/W printer feet against the mounting bracket.

Fig. 34: P91 D Positioned on the Bracket

• Point the bottom side of the P93 D print-er up.

• Put the P93 D mounting bracket on the bottom of the P93 D printer.

Fig. 35: P93 D Mounting Bracket

• Secure the P93 D printer mounting bracket to the P93 D printer.

• Point the bottom side of the printer fac-ing down.

• Put the Printer assembly on the color printer mounting plate.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 41 of 150Medical Solutions USA, Inc.

Page 42: Antares Peripheral Instruction manual

42 Hardware Configuration

11. See the Sony User Instruction to perform the following tasks for the color printer:

- Load the ink cartridge

NOTE Power on the system prior to loading the ink cartridge. This resetsthe gears to their default location, permitting the ink cartridge to beinserted.

- Load the paper tray

- Slide the paper tray into the printer

12. Complete the printer cable connections to the I/O module, see (I/O Module Peripheral Connections / p. 97).

13. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).

P91 D or P93 D / UP-D55M

Installation Summary

1. Install the color printer on the peripheral shelf

2. Install the B/W printer on the color printer.

3. Complete the cable I/O module connections.

4. Complete the printer configuration.

NOTE When performing this upgrade on a system that already has a B/Wprinter or a color printer installed, remove the printer from the pe-ripheral shelf.

Do not disconnect the USB cable from the I/O module. The systemhas previously configured the USB port for the printer connected.Therefore, the same USB port must be used for that printer whenthe USB cable is reconnected.

Label the USB cable for the correct reconnection to the printer.

1. Cut the fastener holes into the rubber mat of the peripheral shelf.

See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

2. Return the mat to the peripheral shelf.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 42 of 150Medical Solutions USA, Inc.

Page 43: Antares Peripheral Instruction manual

Hardware Configuration 43

3. Tighten the printer bracket to the bottom of the color printer.

Fig. 36: UP-D55M Color Printer Bracket

a) Point the bottom side of the printer face up.

b) Put the bracket on the bottom of the printer.

c) Tighten the bracket captive screw.

4. Connect the printer cables.

Fig. 37: UP-D55M Cable ConnectionsPos. 1 Power cord between the OEM AC power outlet and the printer AC power inlet

Pos. 2 USB cable

a) Connect the AC power cable to the Antares peripheral AC power outlet and to the printer AC power inlet.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 43 of 150Medical Solutions USA, Inc.

Page 44: Antares Peripheral Instruction manual

44 Hardware Configuration

b) Connect the USB cable to the back of the printer.

c) Route the other end of the USB cable through the I/O module cable routing pas-sage.

NOTE Do not connect the USB cable to the I/O module at this time, unlessthe system previously had this printer connected.

Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the printer cover.

5. Connect the B/W printer AC power cable into the Antares peripheral AC power outlet. Route the other end of the power cable through the color printer cover.

6. Route the B/W printer USB cable through the I/O panel cable routing passage.

NOTE Do not connect the USB cable to the I/O module at this time.

Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the printer cover.

7. Install the printer cover.

Fig. 38: UP-D55MD Printer Cover with B/W Mounting Plate

a) Put the printer cover over the printer.

b) Align the three captive screws with the three fastener holes cut into the mat.

c) Tighten the captive screws

NOTE Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the printer cover.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 44 of 150Medical Solutions USA, Inc.

Page 45: Antares Peripheral Instruction manual

Hardware Configuration 45

8. Put the B/W printer bracket on the B/W printer mounting plate. Use the applicable in-structions that follow.

WARNING Use fasteners of the correct length.

Using fasteners that are too long may cause the hard-ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.

9. Set the B/W printer dip switches to:

- P91 D settings: 1-ON, 2-OFF, 3-OFF, and 4-OFF- P93 D settings: 1-ON, 2-OFF, 3-ON, and 4-OFF

P91 D Instructions: P93 D Instructions:

• Put the flat side of the bracket face down on the B/W printer mounting plate of the color printer cover.

• Align the bracket mounting holes to the plate mounting holes.

• Put the P91 B/W printer on top of the P91 mounting bracket.

• Align the B/W printer feet against the mounting bracket.

Fig. 39: P91 D Positioned on the Bracket

• Point the bottom side of the P93 D print-er up.

• Put the P93 D mounting bracket on the bottom of the P93 D printer.

Fig. 40: P93 D Mounting Bracket

• Secure the P93 D printer mounting bracket to the P93 D printer.

• Point the bottom side of the printer fac-ing down.

• Put the Printer assembly on the color printer mounting plate.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 45 of 150Medical Solutions USA, Inc.

Page 46: Antares Peripheral Instruction manual

46 Hardware Configuration

10. Connect the B/W printer cables.

Fig. 41: Back of Black and White PrinterPos. 1 Dip switch

Pos. 2 USB cable

Pos. 3 Power cord

a) Connect the AC power cable to the back of the P91 D/P93 D printer.

b) Connect the USB cable to the back of the P91 D/P93 D printer.

NOTE Do not connect the USB cable to the I/O module at this time.

11. Install the B/W printer cover.

Fig. 42: UP-D55M Color Printer and P91 D or P93 D B/W Printer - Double Configuration

a) Put the B/W cover over the P91 D/P93 D printer.

b) Tighten the captive screws (of the B/W printer cover) through the mounting bracket,

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 46 of 150Medical Solutions USA, Inc.

Page 47: Antares Peripheral Instruction manual

Hardware Configuration 47

to the mounting plate.

c) Make sure the back of the P91 D/P93 D printer touches the B/W printer cover stop.

12. See the Sony User Instruction to perform the following tasks:

- Load the ink cartridge

- Load the paper tray

- Slide the paper tray into the printer

13. Complete the printer cable connections to the I/O module, see (I/O Module Peripheral Connections / p. 97).

14. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).

B/W Printer and VCR 0

P91 D or P93 D / HS-MD3000

Installation Summary

1. Install the VCR on the peripheral shelf

2. Install the B/W printer on the VCR.

3. Complete the cable I/O module connections.

4. Complete the VCR configuration.

NOTE When performing this upgrade on a system that already has a B/Wprinter installed, remove the B/W printer from the peripheral shelf.

Do not disconnect the USB cable from the I/O module. The systemhas previously configured the USB port for the B/W printer. There-fore, the same USB port must be used for the B/W printer when theUSB cable is reconnected.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 47 of 150Medical Solutions USA, Inc.

Page 48: Antares Peripheral Instruction manual

48 Hardware Configuration

Fig. 43: HS-MD3000 VRC/P91 Dor P93 D B/W Printer - Double ConfigurationPos. 1 VCR bracket

Pos. 2 VCR

Pos. 3 VCR Cover with B/W printer plate

Pos. 4 B/W printer with bracket attached to the bottom of the printer (Bracket not shown)

Pos. 5 B/W printer cover

1. Cut the fastener holes into the rubber mat of the peripheral shelf.

See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

2. Return the mat to the peripheral shelf.

3. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 48 of 150Medical Solutions USA, Inc.

Page 49: Antares Peripheral Instruction manual

Hardware Configuration 49

4. Remove the feet from the bottom of the VCR.

Fig. 44: VCR Foot Locations

a) Point the bottom side of the VCR facing up.

b) Remove the four screws that secure the feet to the VCR.

c) Remove the feet.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 49 of 150Medical Solutions USA, Inc.

Page 50: Antares Peripheral Instruction manual

50 Hardware Configuration

5. Secure the mounting bracket to the VCR.

Fig. 45: HS-MD3000 Secured to the Mounting Bracket

a) Put the mounting bracket on the VCR.

b) Align the mounting holes.

c) Secure the bracket to the bottom of the VCR.

WARNING Use fasteners of the correct length.

Using fasteners that are too long may cause the hard-ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 50 of 150Medical Solutions USA, Inc.

Page 51: Antares Peripheral Instruction manual

Hardware Configuration 51

6. Put the VCR assembly on the peripheral shelf and align the mounting holes of the VCR bracket with the mounting holes in the mat.

Fig. 46: HS-MD3000 Positioned on the OEM Bay

7. Connect the power cables to the peripheral AC power outlet.

Fig. 47: HS-MD3000 Power Cord Connection Pos. 1 OEM power outlets

Pos. 2 VCR power inlet

a) Connect the power cable to the VCR AC power inlet and to the lower peripheral AC power outlet.

b) Connect the printer power cable to the upper peripheral AC outlet.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 51 of 150Medical Solutions USA, Inc.

Page 52: Antares Peripheral Instruction manual

52 Hardware Configuration

8. Connect the VCR cables. See (VCR I/O Module Cable Connections / p. 98) for details.

Fig. 48: HS-MD3000 VCR Connections from the I/O Module

Pos. 1 Connect P128 to RS-232C

Pos. 2 Connect VIDEO IN

Pos. 3 Connect VIDEO OUT

Pos. 4 Connect AUDIO IN

Pos. 5 Connect AUDIO OUT

Pos. 6 Set the 75Ω VIDEO and the 75Ω S-VIDEO switches to ON

NOTE Do not connect the cables to the I/O module at this time.

Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the VCR cover.

9. Route the B/W printer power cord and USB cable through the cable channel in the VCR cover.

Fig. 49: Cable ChannelsPos. 1 VCR cover to black and white printer

Pos. 2 To I/O module

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 52 of 150Medical Solutions USA, Inc.

Page 53: Antares Peripheral Instruction manual

Hardware Configuration 53

10. Route the other end of the VCR cables through the I/O module cable channel.

11. Route the other end of the B/W printer USB cable through the I/O module cable chan-nel.

12. Put the B/W printer bracket on the B/W printer mounting plate. Use the applicable in-structions that follow.

WARNING Use fasteners of the correct length.

Using fasteners that are too long may cause the hard-ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 53 of 150Medical Solutions USA, Inc.

Page 54: Antares Peripheral Instruction manual

54 Hardware Configuration

P91 D Instructions: P93 D Instructions:

• Put the flat side of the bracket face down on the B/W printer mounting plate of the VCR cover.

• Align the bracket mounting holes to the plate mounting holes.

• Put the P91 B/W printer on top of the P91 mounting bracket.

• Align the B/W printer feet against the mounting bracket.

Fig. 50: P91 D Positioned on the Bracket

• Point the bottom side of the P93 D print-er up.

• Put the P93 D mounting bracket on the bottom of the P93 D printer.

Fig. 51: P93 D Mounting Bracket

• Secure the P93 D printer mounting bracket to the P93 D printer.

• Point the bottom side of the printer fac-ing down.

• Put the Printer assembly on the VCR mounting plate.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 54 of 150Medical Solutions USA, Inc.

Page 55: Antares Peripheral Instruction manual

Hardware Configuration 55

Fig. 52: VCR Cover with Black and White Mounting PlatePos. 1 VCR Cover

Pos. 2 Black and white printer mounting plate

13. Set the B/W printer dip switches:

- P91 D settings: 1-ON, 2-OFF, 3-OFF, and 4-OFF- P93 D settings: 1-ON, 2-OFF, 3-ON, and 4-OFF

Fig. 53: Back of Black and White PrinterPos. 1 Dip switch

Pos. 2 USB cable

Pos. 3 Power cord

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 55 of 150Medical Solutions USA, Inc.

Page 56: Antares Peripheral Instruction manual

56 Hardware Configuration

14. Install the B/W printer cover.

Fig. 54: Black and White Printer and VCR Installed

a) Put the printer cover on the B/W printer.

b) Tighten the two printer cover captive screws.

15. Complete the VCR and printer cable connections to the I/O module, see (I/O Module Peripheral Connections / p. 97).

16. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).

17. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).

P91 D or P93 D / SVO9500

Installation Summary

1. Install the VCR on the peripheral shelf

2. Install the B/W printer on the VCR.

3. Complete the cable I/O module connections.

4. Complete the VCR configuration.

NOTE When performing this upgrade on a system that already has a B/Wprinter installed, remove the B/W printer from the peripheral shelf.

Do not disconnect the USB cable from the I/O module. The systemhas previously configured the USB port for the B/W printer. There-fore, the same USB port must be used for the B/W printer when theUSB cable is reconnected.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 56 of 150Medical Solutions USA, Inc.

Page 57: Antares Peripheral Instruction manual

Hardware Configuration 57

1. Cut the fastener holes into the rubber mat of the peripheral shelf.

See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

2. Secure the mounting bracket to the peripheral shelf.

3. Put the VCR assembly on the peripheral shelf by aligning the mounting holes of the VCR with the mounting holes in the mat.

4. Set the VCR switches:

- Set the dip switches: 1-OFF, 2-OFF, 3-OFF, 4-OFF, 5- ON, 6-ON.

Fig. 55: SVO9500 Dip Switch Settings- Set both 75 ohm switches to ON.

- For PAL model only, set the voltage to 240V.

NOTE Only the PAL model includes the 240V switch.

5. Connect the VCR power cord to the peripheral AC power outlet and to the VCR AC power inlet.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 57 of 150Medical Solutions USA, Inc.

Page 58: Antares Peripheral Instruction manual

58 Hardware Configuration

6. Connect the VCR cables. See (VCR I/O Module Cable Connections / p. 98) for details.

Fig. 56: SVC 9500 VCR Connections from I/O ModulePos. 1 Audio In (left and right)

Pos. 2 Video In (use either connector)

Pos. 3 Audio Out (left and right)

Pos. 4 Video Out

Pos. 5 Dip Switches

Pos. 6 RS-232C (connect to P128)

Pos. 7 75 Ω (set both switches to ON)

7. Route the cables through the I/O panel cable routing passage.

NOTE Do not connect the cables to the I/O module at this time.

Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the VCR cover.

8. Install the cover.

a) Route the B/W printer power cord and USB cable through the cable channel in the VCR cover.

b) Route the other end of the VCR cables through the I/O module cable channel.

c) Route the other end of the B/W printer USB cable through the I/O module cable

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 58 of 150Medical Solutions USA, Inc.

Page 59: Antares Peripheral Instruction manual

Hardware Configuration 59

channel.

d) Put the cover over the VCR.

Fig. 57: Cover Installation - Cables Routed Through the Covere) Align and tighten the two captive screws in the cover with the mounting holes.

Fig. 58: Cover Captive Screw

9. Put the B/W printer bracket on the B/W printer mounting plate. Use the applicable in-structions that follow.

WARNING Use fasteners of the correct length.

Using fasteners that are too long may cause the hard-ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 59 of 150Medical Solutions USA, Inc.

Page 60: Antares Peripheral Instruction manual

60 Hardware Configuration

P91 D Instructions: P93 D Instructions:

• Put the flat side of the bracket face down on the B/W printer mounting plate of the VCR cover.

• Align the bracket mounting holes to the plate mounting holes.

• Put the P91 B/W printer on top of the P91 mounting bracket.

• Align the B/W printer feet against the mounting bracket.

Fig. 59: P91 D Positioned on the Bracket

• Point the bottom side of the P93 D print-er up.

• Put the P93 D mounting bracket on the bottom of the P93 D printer.

Fig. 60: P93 D Mounting Bracket

• Secure the P93 D printer mounting bracket to the P93 D printer.

• Point the bottom side of the printer fac-ing down.

• Put the Printer assembly on the VCR mounting plate.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 60 of 150Medical Solutions USA, Inc.

Page 61: Antares Peripheral Instruction manual

Hardware Configuration 61

10. Set the B/W printer dip switches:

- P91 D settings: 1-ON, 2-OFF, 3-OFF, and 4-OFF- P93 D settings: 1-ON, 2-OFF, 3-ON, and 4-OFF

Fig. 61: Back of Black and White PrinterPos. 1 Dip switch

Pos. 2 USB cable

Pos. 3 Power cord

11. Install the B/W printer cover.

a) Put the printer cover on the B/W printer.

Fig. 62: B/W Printer Cover Installation

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 61 of 150Medical Solutions USA, Inc.

Page 62: Antares Peripheral Instruction manual

62 Hardware Configuration

b) Tighten the two printer cover captive screws.

Fig. 63: B/W and VCR Covers Installed

12. Complete the VCR and printer cable connections to the I/O module, see (I/O Module Peripheral Connections / p. 97).

13. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).

14. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).

Color Printer and VCR 0

CP770 or CP800 / HS-MD3000

Installation Summary

1. Adjust the height of the monitor/control panel.

2. Install the color printer on the peripheral shelf

3. Install the VCR printer on the color printer.

4. Complete the cable I/O module connections.

5. Complete the VCR configuration.

NOTE When performing this upgrade on a system that already has a col-or printer installed, remove the color printer from the peripheralshelf.

Do not disconnect the USB cable from the I/O module. The systemhas previously configured the USB port for the color printer.Therefore, the same USB port must be used for the color printerwhen the USB cable is reconnected.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 62 of 150Medical Solutions USA, Inc.

Page 63: Antares Peripheral Instruction manual

Hardware Configuration 63

1. Adjust the height of the monitor/control panel.

Press the height adjustment knob to set the monitor/control panel to the highest position.

2. Remove the upper, right-front panel.

a) Loosen the 1/4-turn captive screw.

Fig. 64: Locking Bolt in the Storage Loctionb) Grip the transducer port insets of the panel and pull to disengage the ball snaps

3. Remove the left-front panel.

a) Loosen the 1/4-turn captive screw.

Fig. 65: Upper, Left-Front Panel Captive Screwb) Grip the air flow vent insets of the panel and pull the panel to disengage the ball

snaps.

4. Remove the bottom post cover.

a) Loosen the four screws located on either side of the post.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 63 of 150Medical Solutions USA, Inc.

Page 64: Antares Peripheral Instruction manual

64 Hardware Configuration

b) Remove the post cover.

5. Install the locking bolt.

a) Remove the locking bolt from the locking bolt storage location.

b) Tighten the locking bolt to the post to lock the height adjustment on the highest posi-tion

6. Attach the bottom post cover.

a) Place the cover with the painted, chamfered side facing up.

b) Tighten the four post cover screw that secure the cover to the system. Locate two screws on either side of the post.

7. Install the right-front panel.

a) Align the panel ball snap receptacles with the system ball snaps and press the pan-el to engage the ball snaps.

b) Tighten the 1/4-turn captive screw.

8. Install the left-front panel.

a) Align the panel ball snap receptacles with the system ball snaps and press the pan-el to engage the ball snaps.

b) Tighten the 1/4-turn captive screw.

9. Cut the fastener holes into the rubber mat of the peripheral shelf.

See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

10. Return the mat to the peripheral shelf.

11. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.

12. Remove the feet and the shipment locking screw from the bottom of the printer.

a) Point the bottom side of the printer face up.

b) Remove the four screws that secure the feet to the printer and remove the feet.

Fig. 66: UP-D23M Color Printer Connections

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 64 of 150Medical Solutions USA, Inc.

Page 65: Antares Peripheral Instruction manual

Hardware Configuration 65

c) Remove the shipment locking screw.

13. Secure the mounting bracket to the color printer.

a) Put the mounting bracket on the printer.

b) Align the mounting holes.

c) Secure the bracket to the bottom of the color printer.

WARNING Use fasteners of the correct length.

Using fasteners that are too long may cause the hard-ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.

14. Put the color printer assembly on the peripheral shelf and align the mounting holes of the color printer with the mounting holes in the mat.

15. Set the color printer dip switches as follows:

16. Connect the color printer cables.

a) Connect the color printer AC power cable to the color printer AC power inlet and to the lower peripheral AC power outlet.

b) Connect the VCR AC power cable to the upper peripheral AC power outlet.

c) Connect the color printer USB cable to the back of the color printer.

d) Route the other end of the color printer USB cable through the peripheral shelf ca-ble passage to the I/O module.

e) Route the I/O module end of the VCR USB cable through the peripheral shelf cable passage to the I/O module.

Fig. 67: CP770 Dip Switch Settings to OFF

1-OFF, 2-OFF, 3-OFF, 4-OFF,5-OFF, 6-OFF, 7-OFF, and 8-OFF.

Fig. 68: CP800 Dip Switch Settings

1-ON, 2-OFF, 3-OFF, 4-OFF,5-OFF, 6-OFF, 7-OFF, and 8-ON.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 65 of 150Medical Solutions USA, Inc.

Page 66: Antares Peripheral Instruction manual

66 Hardware Configuration

NOTE Do not connect the USB cables to the I/O module at this time, un-less the system previously had one of the printers installed on theperipheral bay. Then, connect the previously installed printer tothe same port it was connected to before.

17. Remove the feet and the shipment locking screw from the bottom of the VCR.

a) Point the bottom side of the VCR facing up.

b) Remove the four screws that secure the feet to the VCR.

c) Remove the feet.

18. Secure the mounting bracket to the VCR.

a) Put the mounting bracket on the VCR.

b) Align the mounting holes.

c) Secure the bracket to the bottom of the VCR.

WARNING Use fasteners of the correct length.

Using fasteners that are too long may cause the hard-ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.

19. Put the VCR assembly on the color printer.

20. Connect the VCR cables.

a) Connect the power cable to the VCR and to the Antares peripheral AC power outlet.

b) Connect the AUDIO IN left and right channel cables to the back of the VCR.

c) Connect the AUDIO OUT left and right channel cables to the back of the VCR.

d) Set the VIDEO and S VIDEO switches to the on position.

e) Connect the VIDEO in/out cables to the back of the VCR.

f) Connect the S VIDEO in/out cables to the back of the VCR.

g) Connect the USB cable to the back of the VCR.

h) Connect the RS232 C cable to the back of the VCR.

21. Install the cover.

a) Put the cover over the VCR.

b) Align the four captive screws in the cover with the mounting holes.

c) Tighten the four captive screws.

22. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).

23. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 66 of 150Medical Solutions USA, Inc.

Page 67: Antares Peripheral Instruction manual

Hardware Configuration 67

24. Complete the VCR and printer cable connections to the I/O module, see (VCR I/O Mod-ule Cable Connections / p. 98).

CP770 or CP800 / SVO9500

Installation Summary

1. Adjust the height of the monitor/control panel.

2. Install the color printer on the peripheral shelf

3. Install the VCR printer on the color printer.

4. Complete the cable I/O module connections.

5. Complete the VCR configuration.

NOTE When performing this upgrade on a system that already has a col-or printer installed, remove the color printer from the peripheralshelf.

Do not disconnect the USB cable from the I/O module. The systemhas previously configured the USB port for the color printer.Therefore, the same USB port must be used for the color printerwhen the USB cable is reconnected.

1. Adjust the height of the monitor/control panel.

Press the height adjustment knob to set the monitor/control panel to the highest position.

2. Remove the upper, right-front panel.

a) Loosen the 1/4-turn captive screw.

Fig. 69: Locking Bolt in the Storage Loctionb) Grip the transducer port insets of the panel and pull to disengage the ball snaps

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 67 of 150Medical Solutions USA, Inc.

Page 68: Antares Peripheral Instruction manual

68 Hardware Configuration

3. Remove the left-front panel.

a) Loosen the 1/4-turn captive screw.

Fig. 70: Upper, Left-Front Panel Captive Screwb) Grip the air flow vent insets of the panel and pull the panel to disengage the ball

snaps.

4. Remove the bottom post cover.

a) Loosen the four screws located on either side of the post.

b) Remove the post cover.

5. Install the locking bolt.

a) Remove the locking bolt from the locking bolt storage location.

b) Tighten the locking bolt to the post to lock the height adjustment on the highest posi-tion

6. Attach the bottom post cover.

a) Place the cover with the painted, chamfered side facing up.

b) Tighten the four post cover screw that secure the cover to the system. Locate two screws on either side of the post.

7. Install the right-front panel.

a) Align the panel ball snap receptacles with the system ball snaps and press the pan-el to engage the ball snaps.

b) Tighten the 1/4-turn captive screw.

8. Install the left-front panel.

a) Align the panel ball snap receptacles with the system ball snaps and press the pan-el to engage the ball snaps.

b) Tighten the 1/4-turn captive screw.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 68 of 150Medical Solutions USA, Inc.

Page 69: Antares Peripheral Instruction manual

Hardware Configuration 69

9. Cut the fastener holes into the rubber mat of the peripheral shelf.

See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

10. Return the mat to the peripheral shelf.

11. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.

12. Put the plastic paper tray on the bracket, aligning the dashed line with the front edge of the bracket.

13. Position the printer on the peripheral shelf.

a) Put the color printer on the plastic paper tray and bracket.

b) Align the printer so the back panel contacts the two sheet-metal tabs. The front sheet-metal tabs should sit in the grooves on the bottom of the printer, a few centi-meters from the front panel.

Fig. 71: Color Printer Foot Placement on Bracket

14. Set the color printer dip switches as follows:

Fig. 72: CP770 Dip Switch Settings to OFF

1-OFF, 2-OFF, 3-OFF, 4-OFF,5-OFF, 6-OFF, 7-OFF, and 8-OFF.

Fig. 73: CP800 Dip Switch Settings

1-ON, 2-OFF, 3-OFF, 4-OFF,5-OFF, 6-OFF, 7-OFF, and 8-ON.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 69 of 150Medical Solutions USA, Inc.

Page 70: Antares Peripheral Instruction manual

70 Hardware Configuration

15. Connect the color printer cables.

Fig. 74: CP770 and CP800 Cable Connections on Bacl of Printer

a) Connect the color printer AC power cable to the color printer AC power inlet and to the lower peripheral AC power outlet.

b) Connect the VCR printer AC power cable to the upper peripheral AC power outlet.

c) Connect the parallel end of the parallel-to-USB cable to the back of the color printer.

d) Route the USB end of the parallel-to-USB cable through the peripheral shelf cable passage to the I/O module.

e) Route the I/O module end of the VCR USB cable through the peripheral shelf cable passage to the I/O module.

NOTE Do not connect the USB cables to the I/O module at this time, un-less the system previously had one of the printers installed on theperipheral bay. Then, connect the previously installed printer tothe same port it was connected to before.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 70 of 150Medical Solutions USA, Inc.

Page 71: Antares Peripheral Instruction manual

Hardware Configuration 71

16. Put the mounting bracket on the printer.

Fig. 75: SVOC9500 Barcket Placement on the CP770 or CP800

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 71 of 150Medical Solutions USA, Inc.

Page 72: Antares Peripheral Instruction manual

72 Hardware Configuration

17. Position the VCR on the VCR bracket and color printer. Align the rear feet of the VCR against the tabs of the retaining bracket.

Fig. 76: SVO9500 Positioned on the CP770 or CP800

18. Install the cover.

a) Put the cover over the VCR.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 72 of 150Medical Solutions USA, Inc.

Page 73: Antares Peripheral Instruction manual

Hardware Configuration 73

b) Align the four captive screws in the cover with the mounting holes.

c) Do not tighten the four captive screws.

Fig. 77: Install the Cover - SVO9500/CP770 or CP800

19. Connect the VCR cables.

Fig. 78: SVC 9500 VCR Connections from I/O ModulePos. 1 Audio In (left and right)

Pos. 2 Video In (use either connector)

Pos. 3 Audio Out (left and right)

Pos. 4 Video Out

Pos. 5 Dip Switches

Pos. 6 RS-232C (connect to P128)

Pos. 7 75 Ω (set both switches to ON)

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 73 of 150Medical Solutions USA, Inc.

Page 74: Antares Peripheral Instruction manual

74 Hardware Configuration

a) Connect the power cable to the VCR and to the Antares peripheral AC power outlet.

b) Connect the AUDIO IN left and right channel cables to the back of the VCR.

c) Connect the AUDIO OUT left and right channel cables to the back of the VCR.

d) Set the VIDEO and S VIDEO switches to the on position.

e) Connect the VIDEO in/out cables to the back of the VCR.

f) Connect the S VIDEO in/out cables to the back of the VCR.

g) Connect the USB cable to the back of the VCR.

h) Connect the RS232 C cable to the back of the VCR.

20. Tighten the four captive screws to secure the cover.

Fig. 79: Tighten the Cover -Captive Screws - SVO9500/CP770 or CP800

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 74 of 150Medical Solutions USA, Inc.

Page 75: Antares Peripheral Instruction manual

Hardware Configuration 75

21. Install the back peripheral cover and tighten the four screws.

Fig. 80: Install the Back Cover - SVO9500/CP770 or CP800

22. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).

23. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).

24. Complete the VCR and printer cable connections to the I/O module, see (I/O Module Peripheral Connections / p. 97).

UP-D23MD / HS-MD 3000

Installation Summary

1. Adjust the height of the monitor/control panel.

Press the height adjustment knob to set the monitor/control panel to the highest position.

2. Install the VCR on the peripheral shelf

3. Install the color printer on the VCR.

4. Complete the cable I/O module connections.

5. Complete the VCR configuration.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 75 of 150Medical Solutions USA, Inc.

Page 76: Antares Peripheral Instruction manual

76 Hardware Configuration

NOTE When performing this upgrade on a system that already has a col-or printer installed, remove the color printer from the peripheralshelf.

Do not disconnect the USB cable from the I/O module. The systemhas previously configured the USB port for the color printer.Therefore, the same USB port must be used for the color printerwhen the USB cable is reconnected.

Fig. 81: HS-MD3000 VRC/UPD-23MD Color Printer - Double ConfigurationPos. 1 VCR bracket

Pos. 2 VCR

Pos. 3 Color printer bracket

Pos. 4 Color printer

Pos. 5 Cover

1. Adjust the height of the monitor/control panel.

Press the height adjustment knob to set the monitor/control panel to the highest position.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 76 of 150Medical Solutions USA, Inc.

Page 77: Antares Peripheral Instruction manual

Hardware Configuration 77

2. Remove the upper, right-front panel.

a) Loosen the 1/4-turn capptive screw.

Fig. 82: Upper, Right-Front Panel Captive Screwb) Grip the transducer port insets of the panel and pull to disengage the ball snaps

3. Remove the left-front panel.

a) Loosen the 1/4-turn capptive screw.

Fig. 83: Upper, Left-Front Panel Captive Screwb) Grip the air flow vent insets of the panel and pull the panel to disengage the ball

snaps.

4. Remove the bottom post cover.

a) Loosen the four screws located on either side of the post.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 77 of 150Medical Solutions USA, Inc.

Page 78: Antares Peripheral Instruction manual

78 Hardware Configuration

b) Remove the post cover.

Fig. 84: Bottom Post Cover Screws

5. Install the locking bolt.

a) Remove the locking bolt from the locking bolt storage location.

Fig. 85: Locking Bolt in the Storage Loctionb) Tighten the locking bolt to the post to lock the height adjustment on the highest posi-

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 78 of 150Medical Solutions USA, Inc.

Page 79: Antares Peripheral Instruction manual

Hardware Configuration 79

tion.

Fig. 86: Locking Height Adjustment

6. Attach the bottom post cover.

a) Place the cover with the painted, chamfered side facing up.

b) Tighten the four post cover screw that secure the cover to the system. Locate two screws on either side of the post.

7. Install the right-front panel.

a) Align the panel ball snap recepticals with the system ball snaps and press the panel to engage the ball snaps.

b) Tighten the 1/4-turn captive screw.

8. Install the left-front panel.

a) Align the panel ball snap recepticals with the system ball snaps and press the panel to engage the ball snaps.

b) Tighten the 1/4-turn captive screw.

9. Cut the fastener holes into the rubber mat of the peripheral shelf.

See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

10. Return the mat to the peripheral shelf.

11. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 79 of 150Medical Solutions USA, Inc.

Page 80: Antares Peripheral Instruction manual

80 Hardware Configuration

12. Remove the feet from the bottom of the VCR.

Fig. 87: VCR Foot Locations

a) Point the bottom side of the VCR facing up.

b) Remove the four screws that secure the feet to the VCR.

c) Remove the feet.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 80 of 150Medical Solutions USA, Inc.

Page 81: Antares Peripheral Instruction manual

Hardware Configuration 81

13. Secure the mounting bracket to the VCR.

Fig. 88: HS-MD3000 Secured to the Mounting Bracket

a) Put the mounting bracket on the VCR.

b) Align the mounting holes.

c) Secure the bracket to the bottom of the VCR.

WARNING Use fasteners of the correct length.

Using fasteners that are too long may cause the hard-ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 81 of 150Medical Solutions USA, Inc.

Page 82: Antares Peripheral Instruction manual

82 Hardware Configuration

14. Put the VCR assembly on the peripheral shelf and align the mounting holes of the VCR bracket with the mounting holes in the mat.

Fig. 89: HS-MD3000 Positioned on the OEM Bay

15. Connect the VCR power cable to the VCR and to the lower peripheral AC power outlet.

Fig. 90: HS-MD3000 Power Cord Connection Pos. 1 OEM power outlets

Pos. 2 VCR power inlet

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 82 of 150Medical Solutions USA, Inc.

Page 83: Antares Peripheral Instruction manual

Hardware Configuration 83

16. Connect the VCR cables. See (VCR I/O Module Cable Connections / p. 98) for details.

Fig. 91: HS-MD3000 VCR Connections from the I/O ModulePos. 1 Connect P128 to RS-232C

Pos. 2 Connect VIDEO IN

Pos. 3 Connect VIDEO OUT

Pos. 4 Connect AUDIO IN

Pos. 5 Connect AUDIO OUT

Pos. 6 Set the 75Ω VIDEO and the 75Ω S-VIDEO switches to ON

17. Route the other end of the VCR cables through the I/O module cable channel.

NOTE Do not connect the VCR cables to the I/O module at this time.

Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the VCR cover.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 83 of 150Medical Solutions USA, Inc.

Page 84: Antares Peripheral Instruction manual

84 Hardware Configuration

18. Secure the color printer to the printer mounting bracket.

Fig. 92: UP-D23M Color Printer Positioned on the MD3000 VCR

a) Point the bottom side of the printer face up.

b) Put the printer mounting bracket on the printer.

c) Tighten the bracket captive screws to secure the printer to the bracket.

19. Put the printer assembly on the VCR.

Fig. 93: UP-D23M Color Printer Positioned on the MD3000 VCR

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 84 of 150Medical Solutions USA, Inc.

Page 85: Antares Peripheral Instruction manual

Hardware Configuration 85

20. Connect the color printer cables.

Fig. 94: UP-D23M Color Printer Connections

NOTE (Fig. 94 / p. 85) does not show the VCR with the VCR cables con-nected. However, the VCR cables were connected in a previoussteps, see (VCR I/O Module Cable Connections / p. 98).

a) Connect the printer power cable into the upper peripheral outlet

b) Connect the printer USB cable to the back of the printer

21. Route the other end of the printer USB cable through the I/O module cable channel.

NOTE Do not connect the USB cable to the I/O module at this time.

Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the VCR cover.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 85 of 150Medical Solutions USA, Inc.

Page 86: Antares Peripheral Instruction manual

86 Hardware Configuration

22. Install the cover.

Fig. 95: UP-D23M Color Printer and MD3000 VCR Installed

a) Put the printer cover over the color printer/VCR assembly.

b) Verify the front lip of the cover is against the front of the printer/VCR assembly.

c) Tighten the three printer cover captive screws.

23. See the Sony User Instruction to perform the following tasks:

- Load the ink cartridge

NOTE Power on the system prior to loading the ink cartridge. This resetsthe gears to their default location, permitting the ink cartridge to beinserted.

- Load the paper tray

- Slide the paper tray into the printer

24. Complete the VCR and printer cable connections to the I/O module, see (I/O Module Peripheral Connections / p. 97).

25. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).

26. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).

UP-D55MD / HS-MD3000

Installation Summary

1. Adjust the height of the monitor/control panel.

Press the height adjustment knob to set the monitor/control panel to the highest position.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 86 of 150Medical Solutions USA, Inc.

Page 87: Antares Peripheral Instruction manual

Hardware Configuration 87

2. Install the color printer on the peripheral shelf

3. Install the VCR on the color printer.

4. Complete the cable I/O module connections.

5. Complete the VCR configuration.

NOTE When performing this upgrade on a system that already has a col-or printer installed, remove the color printer from the peripheralshelf.

Do not disconnect the USB cable from the I/O module. The systemhas previously configured the USB port for the color printer.Therefore, the same USB port must be used for the color printerwhen the USB cable is reconnected.

Fig. 96: UP-D55M Color Printer/HS-MD3000 VRC - Double ConfigurationPos. 1 Color printer bracket

Pos. 2 Color printer

Pos. 3 VCR bracket

Pos. 4 VCR

Pos. 5 Cover

1. Adjust the height of the monitor/control panel.

Press the height adjustment knob to set the monitor/control panel to the highest position.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 87 of 150Medical Solutions USA, Inc.

Page 88: Antares Peripheral Instruction manual

88 Hardware Configuration

2. Remove the upper, right-front panel.

a) Loosen the 1/4-turn capptive screw.

Fig. 97: Upper, Right-Front Panel Captive Screwb) Grip the transducer port insets of the panel and pull to disengage the ball snaps

3. Remove the left-front panel.

a) Loosen the 1/4-turn capptive screw.

Fig. 98: Upper, Left-Front Panel Captive Screwb) Grip the air flow vent insets of the panel and pull the panel to disengage the ball

snaps.

4. Remove the bottom post cover.

a) Loosen the four screws located on either side of the post.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 88 of 150Medical Solutions USA, Inc.

Page 89: Antares Peripheral Instruction manual

Hardware Configuration 89

b) Remove the post cover.

Fig. 99: Bottom Post Cover Screws

5. Install the locking bolt.

a) Remove the locking bolt from the locking bolt storage location.

Fig. 100: Locking Bolt in the Storage Loctionb) Tighten the locking bolt to the post to lock the height adjustment on the highest posi-

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 89 of 150Medical Solutions USA, Inc.

Page 90: Antares Peripheral Instruction manual

90 Hardware Configuration

tion.

Fig. 101: Locking Height Adjustment

6. Attach the bottom post cover.

a) Place the cover with the painted, chamfered side facing up.

b) Tighten the four post cover screw that secure the cover to the system. Locate two screws on either side of the post.

7. Install the right-front panel.

a) Align the panel ball snap recepticals with the system ball snaps and press the panel to engage the ball snaps.

b) Tighten the 1/4-turn captive screw.

8. Install the left-front panel.

a) Align the panel ball snap recepticals with the system ball snaps and press the panel to engage the ball snaps.

b) Tighten the 1/4-turn captive screw.

9. Cut the fastener holes into the rubber mat of the peripheral shelf.

See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

10. Return the mat to the peripheral shelf.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 90 of 150Medical Solutions USA, Inc.

Page 91: Antares Peripheral Instruction manual

Hardware Configuration 91

11. Attach the printer bracket to the bottom of the color printer.

Fig. 102: UP-D55M Color Printer Bracket

a) Point the bottom side of the printer face up.

b) Put the bracket on the bottom of the printer.

c) Tighten the bracket captive screw.

12. Connect the printer cables.

Fig. 103: UP-D55M Cable ConnectionsPos. 1 Power cord between the OEM AC power outlet and the printer AC power inlet

Pos. 2 USB cable

a) Connect the AC power cable to the Antares peripheral AC power outlet and to the printer AC power inlet.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 91 of 150Medical Solutions USA, Inc.

Page 92: Antares Peripheral Instruction manual

92 Hardware Configuration

b) Connect the USB cable to the back of the printer.

c) Route the other end of the USB cable through the I/O module cable routing pas-sage.

NOTE Do not connect the USB cable to the I/O module at this time.

Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the printer cover.

13. Secure the mounting bracket to the VCR.

Fig. 104: MD3000 VCR Bracket for the Mounting on the UP55-MD

a) Put the mounting bracket on the VCR.

b) Align the mounting holes.

c) Tighten the captive screws that secure the bracket to the bottom of the VCR.

14. Put the VCR assembly on the color printer.

15. Connect the AC power cord to the upper peripheral AC power outlet and to the VCR AC power inlet. (The color printer is connected to the lower power outlet.)

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 92 of 150Medical Solutions USA, Inc.

Page 93: Antares Peripheral Instruction manual

Hardware Configuration 93

16. Connect the VCR cables. See (VCR I/O Module Cable Connections / p. 98) for details.

Fig. 105: HS-MD3000 VCR Connections from the I/O ModulePos. 1 Connect P128 to RS-232C

Pos. 2 Connect VIDEO IN

Pos. 3 Connect VIDEO OUT

Pos. 4 Connect AUDIO IN

Pos. 5 Connect AUDIO OUT

Pos. 6 Set the 75Ω VIDEO and the 75Ω S-VIDEO switches to ON

17. Route the other end of the VCR cables through the I/O module cable channel.

NOTE Do not connect the cables to the I/O module at this time.

Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the VCR cover.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 93 of 150Medical Solutions USA, Inc.

Page 94: Antares Peripheral Instruction manual

94 Hardware Configuration

18. Install the cover.

Fig. 106: MD3000 VCR/UP-D55M Installed

a) Put the printer cover over the color printer/VCR assembly.

b) Verify the front lip of the cover is flush against the front of the printer/VCR assembly.

c) Tighten the three printer cover captive screws.

19. See the Sony User Instruction to perform the following tasks:

- Load the ink cartridge

- Load the paper tray

- Slide the paper tray into the printer

20. Complete the VCR and printer cable connections to the I/O module, see (I/O Module Peripheral Connections / p. 97).

21. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).

22. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 94 of 150Medical Solutions USA, Inc.

Page 95: Antares Peripheral Instruction manual

Hardware Configuration 95

Peripheral Shelf Mounting Hole Locations 0

Usethe following two tables and “Printer and VCR Mounting Hole Locations” figure to determine which mounting holes to cut out for the peripheral configuration being installed. The two tables list which holes to cut from the mat. The hole locations are numbered for easy identification. The location of the mounting hole cut-outs are shown in the following figure.

Tab. 6 Single Configurations

Tab. 7 Double Configurations

Peripheral Mounted on peripheral Shelf

Mounting Hole Locations

CP770 or CP800 color printer 1-3-10-11

UP-D23MD color printer 1-4-8

UP-D55MD color printer 1-2-9

P91 D or P93 D B/W printer 6-8

HS-MD3000 VCR 1-3-9

SVO9500 2-10

Peripheral Mounted on Peripheral Shelf

Mounting Hole Locations

P91 D or P93 D B/W printer

CP770 or CP800 color printer

1-3-10-11

P91 D or P93 D B/W printer

UP-D23MD color printer

1-2-9

P91 D or P93 D B/W printer

UP-D55MD color printer

1-2-9

P91 D or P93 D B/W printer

HS-MD3000 VCR

1-2-9

CP770 or CP800 color printer

HS-MD3000 VCR

1-3-10-11

UP-D23MD color printer

HS-MD3000 VCR

1-4-8

UP-D55MD color printer

HS-MD3000 VCR

1-2-9

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 95 of 150Medical Solutions USA, Inc.

Page 96: Antares Peripheral Instruction manual

96 Hardware Configuration

Fig. 107: Peripheral Mounting Hole Guide

P91 D or P93 D B/W printer

SVO9500 VCR

1-2-9

CP770 or CP800 color printer

SVO9500 VCR

2-10

UP-D23MD color printer

SVO9500 VCR

1-4-8

UP-D55MD color printer

SVO9500 VCR

1-2-9

Peripheral Mounted on Peripheral Shelf

Mounting Hole Locations

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 96 of 150Medical Solutions USA, Inc.

Page 97: Antares Peripheral Instruction manual

Hardware Configuration 97

I/O Module Peripheral Connections 0

Complete the cable connections to I/O module from the B/W printer, color printer, or VCR. The following figure shows the I/O module cable connections for all peripherals.

Printer I/O Module Cable Connections 0

• For B/W printers: connect the USB cable to J5A.

See (Pos. 10/Fig. 108 / p. 97).

• For color printers: connect the USB cable to J5B.

See (Pos. 9/Fig. 108 / p. 97).

Fig. 108: I/O Module Connections

1. USB-A port (modem), J1

2. Left (Channel 1) audio in from VCR

3. Right (Channel 2) audio in from Right (Channel 2) audio in from VCR

4. S-Video In from VCR

5. Composite Video Out

6. GB/s video and limited COM 2

7. VCR control (RS-232 COM 1)

8. Ethernet (Network)

9. USB-A ports (printers): J5B (color printers)

10. USB-A ports (printers): J5A (black and white)

11. System status LEDs

12. Parallel port (printer)

13. Physio Trigger Out

14. S-Video Out to VCR

15. Right Audio Out to VCR

16. Left Audio Out to VCR

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 97 of 150Medical Solutions USA, Inc.

Page 98: Antares Peripheral Instruction manual

98 Hardware Configuration

VCR I/O Module Cable Connections 0

Fig. 109: HS-MD3000 VCR to I/O Module Connections

Fig. 110: SVO9500 VCR to I/O Module Connections

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 98 of 150Medical Solutions USA, Inc.

Page 99: Antares Peripheral Instruction manual

Printer Setup and Configuration 99

4- 3Printer Setup and Configuration

This procedure assumes that the printers were installed on the system before installing software. Software version 4.0 has installation and configuration requirements that differ from previous software releases. There are two different installation and configuration pro-cedures. Use the following table to identify which procedure to use.

Tab. 8 Software Version 4.0 Printers

Use the peripheral software setup instructions that match the software version running on the Antares system.

• Software version 2.1 release runs on Windows 2000

• Software version 3.0 release runs on Windows 2000

• Software version 4.0 release runs on Windows XP OS

Printers Manufacturer Printer Type SW v2.1 SW v3.0 SW v4.0

CP770DW Mitsubishi Image X X X

CP800DW Mitsubishi Image X X X 1, 2

1. When installing multiple printers on a system with v4.0, install the CP800 first.

2. CP800 is not supported on systems configured with the RM-165 PCB.

HP1150 HP Report X X

HP2500n HP Report X X X

HP4050 HP Report X X X

HP4100 HP Report X X X

HP4200 HP Report X X X

HP6122 HP Report X X X

P91DW Mitsubishi Image X X X

P93DW Mitsubishi Image X X

UPD23MD Sony Image X X X

UPD55MD Sony Image X X

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 99 of 150Medical Solutions USA, Inc.

Page 100: Antares Peripheral Instruction manual

100 Printer Setup and Configuration

SW v2.1 and V3.0 0

NOTE When installing two printers and one of them is the CP800, installthe CP800 printer first.

CP800 Installation 0

NOTE When installing two printers and one of the printers is the CP800,install the CP800 first.

1. Boot the Antares system to imaging mode.

2. Press the printer Power button. Allow the printer to initialize (approximately ten sec-onds).

3. Connect the printer USB cable to J5B on the I/O panel. Reference (9/Fig. 108 / p. 97).

NOTE The system briefly displays a message indicating that the SmartCable is installing.

4. Press the Presets button. The system displays the Presets menu.

5. Select Service. The system displays the System Information, Service Options, Re-mote Access Options, and Video panes.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 100 of 150Medical Solutions USA, Inc.

Page 101: Antares Peripheral Instruction manual

Printer Setup and Configuration 101

6. Select Configure Local Printer.

Fig. 111: Presets > Service > Configure Local Printer

The system displays the Add Printer screen.

7. Double-click the Add Printer icon.

Fig. 112: Add Printer Icon (SW v2.1 and v3.0)

The system displays the “Add Printer Wizard”.

8. Select Next. The “Local or Network” window displays.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 101 of 150Medical Solutions USA, Inc.

Page 102: Antares Peripheral Instruction manual

102 Printer Setup and Configuration

9. If not already selected, select the Local printer check box.

Verify Automatically detect and install my Plug-and-play printer is not selected.

Fig. 113: Local or Network Printer (SW v2.1 and v3.0)

10. Select Next. The “Select a Printer Port” screen displays.

11. Select an unused USB virtual printer port from the Use the following Port list.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 102 of 150Medical Solutions USA, Inc.

Page 103: Antares Peripheral Instruction manual

Printer Setup and Configuration 103

NOTE Select a USB port that has not yet been assigned to a printer (usu-ally USB001, but could be USB002 or USB003). Once a USB porthas been assigned to a printer, it cannot be re-assigned to a differ-ent printer without first uninstalling and re-installing the USB portresource.

Fig. 114: Select a Printer Port (SW v2.1and v3.0)

12. Select Next. The “Add Printer Wizard” screen displays.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 103 of 150Medical Solutions USA, Inc.

Page 104: Antares Peripheral Instruction manual

104 Printer Setup and Configuration

13. If not already selected, select Mitsubishi from the “Manufacturers” list and Mitsubishi CP800DW (LPT) from the “Printers” list.

Fig. 115: Add Printer Wizard (SW v2.1 and v3.0)

14. Select Next.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 104 of 150Medical Solutions USA, Inc.

Page 105: Antares Peripheral Instruction manual

Printer Setup and Configuration 105

15. If the “Use Existing Driver” window displays, select Keep existing driver (recom-mended).

Fig. 116: Use Existing Driver (SW v2.1 and v3.0)

16. Select Next. The “Name Your Printer” screen displays.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 105 of 150Medical Solutions USA, Inc.

Page 106: Antares Peripheral Instruction manual

106 Printer Setup and Configuration

17. Verify the name displayed in the Printer Name field is Mitsubishi CP800DW (LPT).

Fig. 117: Name Your Printer (SW v2.1 and v3.0)

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 106 of 150Medical Solutions USA, Inc.

Page 107: Antares Peripheral Instruction manual

Printer Setup and Configuration 107

18. IF not already selected, select Do not share this printer.

Fig. 118: Printer Sharing (SW v2.1 and v3.0)

19. Select Next. The “Print Test Page” screen displays.

20. Select Yes to print a test page.

21. Select Next.

22. Select Finish. A dialog box displays to verify the Windows Test Page printed.

Verify the page printed.

23. Select Ok.

NOTE If the teat page did not print, reference (PrinterTroubleshooting / p. 143).

24. If there are no more printers to install, continue to (Reboot System / p. 122).

CP770DW, P91DW, P93DW, UPD23MD, and UPD55MD Installation 0

NOTE When installing two printers and one of them is the CP800, installthe CP800 printer first.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 107 of 150Medical Solutions USA, Inc.

Page 108: Antares Peripheral Instruction manual

108 Printer Setup and Configuration

1. Boot the Antares system to imaging mode.

2. Press the printer Power button. Allow the printer to initialize (approximately ten sec-onds).

3. Connect the printer USB cable to the I/O panel.

NOTE The system briefly displays a message verifying that the printer isinstalling.

- For B/W printers: connect the printer USB cable to J5B on the I/O panel. Reference (9/Fig. 108 / p. 97).

- For color printers: connect the printer USB cable to J5A on the I/O panel. Reference (10/Fig. 108 / p. 97).

NOTE The system briefly displays a message indicating that the printeris installing. If there are problems trying to install the printer, ref-erence (Printer Troubleshooting / p. 143).

4. If there are no more printers to install, continue to (Reboot System / p. 122).

HP1150, HP2500n, HP4050, HP4100, HP4200, and HP6122 Installation 0

The report printers automatically install and configure during the software installation. The following steps allow the user to install a printer manually (if problems are encountered during the software installation) or if the user is adding the printer to an existing, working system.

1. Power on the Antares system.

2. Power on the printer.

Allow the Antares system to complete the boot sequence to imaging mode.

Wait for the printer to indicate it is ready (indicator light or display message).

3. Connect the printer USB cable to the I/O module.

- For B/W printers: connect the USB cable to J5A. Reference (10/Fig. 108 / p. 97).

- For color printers: connect the USB cable to J5B. Reference (9/Fig. 108 / p. 97).

The system checks for new hardware and displays the following message:

New Hardware has been detected.

Allow the new hardware installation to complete.

4. Register a patient in the Patient Registration screen.

a) Press the Patient key or select the Patient Registration icon. The system displays the Patient Registration screen.

b) Enter patient information into the screen fields and select Ok.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 108 of 150Medical Solutions USA, Inc.

Page 109: Antares Peripheral Instruction manual

Printer Setup and Configuration 109

5. Press the Report key or select the Report icon. The system displays the Patient Re-port screen.

6. Select Print Report. The system prints the report.

- Compare the printed report to the information on the screen.

- If the report did not print properly, see (Printer Installation Troubleshooting / p. 143) and if necessary (Restoring Plug-and-Play Resources / p. 145).

7. If there are no more printers to install, continue to (Reboot System / p. 122).

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 109 of 150Medical Solutions USA, Inc.

Page 110: Antares Peripheral Instruction manual

110 Printer Setup and Configuration

SW v4.0 0

CP800 Installation 0

NOTE The CP800 is not supported on systems with the RM-165 hardware.

NOTE When installing two printers and one of them is the CP800, installthe CP800 printer first.

1. Boot the Antares system to imaging mode.

2. Press the printer Power button. Allow the printer to initialize (approximately ten sec-onds).

3. Connect the printer USB cable to J5B on the I/O panel. Reference (9/Fig. 108 / p. 97).

NOTE The system briefly displays a message indicating that the SmartCable is installing.

4. The system detects the printer and displays the “Found New Hardware Wizard” screen.

a) If not already selected, select the Install the software automatically (recom-mended) check box.

b) Select Next. The system installs the USB software automatically.

c) Verify the installation was successful.

The system displays the following screen when the installation was successful:

Fig. 119: Successful Installation Screen

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 110 of 150Medical Solutions USA, Inc.

Page 111: Antares Peripheral Instruction manual

Printer Setup and Configuration 111

The system displays the following screen when the installation was not successful:

Fig. 120: Installation Failure Screen

NOTE If the installation fails, complete the following steps:

1. Select Finish.

2. Disconnect the Smart Cable from J5B on the I/O module.

3. Reconnect the Smart Cable to J5B on the I/O module.

4. Repeat step 4.

d) Select Finish.

The “Found New Software Wizard” closes.

5. Press the Presets button. The system displays the Presets menu.

6. Select Service. The system displays the System Information, Service Options, Re-mote Access Options, and Video panes.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 111 of 150Medical Solutions USA, Inc.

Page 112: Antares Peripheral Instruction manual

112 Printer Setup and Configuration

7. Select Configure Local Printer.

Fig. 121: Presets > Service > Configure Local Printer

The system displays the Add Printer screen.

8. Double-click the Add Printer icon.

Fig. 122: Add Printer Icon (SW v4.0)

The system initiates the “Add Printer Wizard”.

9. Select Next. The “Local or Network” window displays.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 112 of 150Medical Solutions USA, Inc.

Page 113: Antares Peripheral Instruction manual

Printer Setup and Configuration 113

10. If not already selected, select the Local printer attached to this computer check box.

Verify Automatically detect and install my Plug-and-play printer is not selected.

Fig. 123: Local or Network Printer (SW v4.0)

11. Select Next. The “Select a Printer Port” screen displays.

12. Select an unused USB virtual printer port from the Use the following Port drop-down menu.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 113 of 150Medical Solutions USA, Inc.

Page 114: Antares Peripheral Instruction manual

114 Printer Setup and Configuration

NOTE Select a USB port that has not yet been assigned to a printer (usu-ally USB001, but could be USB002 or USB003). Once a USB porthas been assigned to a printer, it cannot be re-assigned to a differ-ent printer without first uninstalling and re-installing the USB portresource.

Fig. 124: Select a Printer Port (SW v4.0)

13. Select Next. The “Install Printer Software” screen displays.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 114 of 150Medical Solutions USA, Inc.

Page 115: Antares Peripheral Instruction manual

Printer Setup and Configuration 115

14. If not already selected, select the MITSUBISHI from the Manufacturer list, CP800DW (LPT) from the Printers list and select Next. The “Use Existing Driver” win-dow displays.

Fig. 125: Install Printer Software (SW v4.0)

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 115 of 150Medical Solutions USA, Inc.

Page 116: Antares Peripheral Instruction manual

116 Printer Setup and Configuration

15. If the Use Existing Driver screen displays, select Keep existing driver (recommend-ed).

Fig. 126: Use Existing Driver (SW v4.0)

16. Select Next. The “Name Your Printer” screen displays.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 116 of 150Medical Solutions USA, Inc.

Page 117: Antares Peripheral Instruction manual

Printer Setup and Configuration 117

17. Verify the name displayed in the Printer Name field is Mitsubishi CP800DW (LPT).

Fig. 127: Name Your Printer (SW V4.0)

18. Select No for “Do you want to use this printer as the default printer”.

19. Select Next. The “Print Test Page” screen displays.

20. Select Yes for “Do you want to print a test page”.

21. Select Next. The “Completing the Add Printer Wizard” screen displays.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 117 of 150Medical Solutions USA, Inc.

Page 118: Antares Peripheral Instruction manual

118 Printer Setup and Configuration

22. Verify the installation was successful.

Fig. 128: Install Printer Software (SW v4.0)

23. Select Finish. A dialog box prompts to verify the Windows Test Page printed.

24. Verify the test page printed.

NOTE If the test page did not print or If errors occurred during the instal-lation, see (Printer Troubleshooting / p. 143).

25. Select Ok.

26. If there are no more printers to install, continue to the next section, (Reboot System / p. 122).

CP770DW, P91DW, P93DW, UPD23MD, and UPD55MD Installation 0

NOTE When installing two printers and one of them is the CP800, installthe CP800 printer first.

1. Boot the Antares system to imaging mode.

2. Press the printer Power button. Allow the printer to initialize (approximately ten sec-onds).

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 118 of 150Medical Solutions USA, Inc.

Page 119: Antares Peripheral Instruction manual

Printer Setup and Configuration 119

3. Connect the printer USB cable to the I/O panel.

NOTE The system briefly displays a message verifying that the printer isinstalling.

- For B/W printers: connect the printer USB cable to J5A on the I/O panel. Reference (10/Fig. 108 / p. 97).

- For color printers: connect the printer USB cable to J5B on the I/O panel. Reference (9/Fig. 108 / p. 97).

NOTE The system may briefly displays a message indicating that theprinter is installing. Select Next and/or Finnish (depending on theprinter).

Some printers install the USB support automatically.

If there are problems trying to install the printer, reference (PrinterTroubleshooting / p. 143).

4. For CP770 and P91DW only:

During the first installation of a USB device using a Smart Cable, the USB installation fails as shown in (Fig. 120 / p. 111).

If the installation fails, complete the following steps:

a) Select Finish.b) Disconnect the Smart Cable from the J5A (P91DW) or J5B (CP770) I/O module.

c) Reconnect the Smart Cable to J5A (P91DW) or J5B (CP770) on the I/O module.

d) Repeat step 3.

If the installation fails again, reference (Printer Troubleshooting / p. 143).

5. The system detects the printer. The system displays the “Found New Hardware Wiz-ard” for USB printing support.

a) If not already selected, select the Install the software automatically check box.

b) Select Next. The system installs the USB software automatically and then displays a window indicating the install is complete.

c) Select Finish.

The “Found New Hardware Wizard” closes.

NOTE If the first installation attempt of the USB printer will fail. Discon-nect the printer(s) USB cable from the I/O panel and start the pro-cedure from step 2.

6. For P91DW only:

Switch off the Bi-directional support.

a) Press Presets.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 119 of 150Medical Solutions USA, Inc.

Page 120: Antares Peripheral Instruction manual

120 Printer Setup and Configuration

b) Select Service.

c) Select Configure Local Printer. The system displays the printer screen.

d) Right-click Mitsubishi P91DW (LPT) and select Properties.

e) Select the Ports tab.

f) Make sure Enable bidirectional support is not checked.

g) Select Ok.

7. If there are no more printers to install, continue to (Reboot System / p. 122).

NOTE If errors occurred during the installation, see (Printer InstallationTroubleshooting / p. 143).

HP4050, HP4100 and HP4200 Installation 0

1. With the Antares system powered off, connect the parallel cable to J1 of the I/O module parallel printer port and the back of the printer. See (Fig. 108 / p. 97) to locate the paral-lel printer port.

2. Power on the printer.

Allow the Antares system to complete the boot sequence to imaging mode.

Wait for the printer to indicate it is ready (indicator light or display message).

3. Power on the Antares system.

4. The system checks for new hardware and displays the following message:

New Hardware has been detected.

Allow the new hardware installation to complete.

5. Select Next.

Allow the installation to complete and then select Ok.

6. Register a patient in the Patient Registration screen.

a) Press the Patient key or select the Patient Registration icon. The system displays the Patient Registration screen.

b) Enter patient information into the screen fields and select Ok.

7. Press the Report key or select the Report icon. The system displays the Patient Re-port screen.

8. Select Print Report. The system prints the report.

- Compare the printed report to the information on the screen.

- If the report did not print properly, see (Printer Installation Troubleshooting / p. 143) and if necessary (Restoring Plug-and-Play Resources / p. 145).

9. If there are no more printers to install, continue to (Reboot System / p. 122).

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 120 of 150Medical Solutions USA, Inc.

Page 121: Antares Peripheral Instruction manual

Printer Setup and Configuration 121

HP1150, HP2500n and HP6122 Installation 0

1. Complete the printer hardware installation.

2. Power on the printer.

Wait for the printer to indicate it is ready (indicator light or display message).

3. Connect the parallel cable to J1 of the I/O module parallel printer port and the back of the printer. See (Fig. 108 / p. 97) to locate the parallel printer port.

4. Power on the Antares system.

Allow the Antares system to complete the boot sequence to imaging mode.

5. The system checks for new hardware and displays the following message:

New Hardware has been detected.

Allow the new hardware installation to complete.

6. Select Next.

Allow the installation to complete and then select Ok.

7. Register a patient in the Patient Registration screen.

a) Press the Patient key or select the Patient Registration icon. The system displays the Patient Registration screen.

b) Enter patient information into the screen fields and select Ok.

8. Press the Report key or select the Report icon. The system displays the Patient Re-port screen.

9. Select Print Report. The system prints the report.

- Compare the printed report to the information on the screen.

- If the report did not print properly, see (Printer Installation Troubleshooting / p. 143) and if necessary (Restoring Plug-and-Play Resources / p. 145).

10. If there are no more printers to install, continue to (Reboot System / p. 122).

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 121 of 150Medical Solutions USA, Inc.

Page 122: Antares Peripheral Instruction manual

122 Printer Setup and Configuration

Reboot System 0

Perform a system reboot only after completing the installation of all printers.

1. Reboot the system.

a) Press the Standby button.

b) Select Shut Down from the displayed menu.

c) Wait ten seconds after the system completely shuts down. Then press Standby again to power on the system.

2. Continue to (Print/Store Setup / p. 123).

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 122 of 150Medical Solutions USA, Inc.

Page 123: Antares Peripheral Instruction manual

Printer Setup and Configuration 123

Print/Store Setup 0

After the DICOM store and print devices are configured and the system has been reboo-ted, it is necessary to configure the Print and Store key assignments. The Print/Store Setup permits the following configuration of the Antares Print and Store keys:

• Automatic transfer of static images, either at the “End of exam” or “During exam”.

• Automatic transfer of static images, either at the “End of exam” or “During exam”.

• Automatic print routing, based on the image type.

• Storing images as Secondary Capture (global setting).

• Burning graphics into pixel data (global setting).

The Print and Store keys on the control panel can be configured to do print only, print and store, or store only, regardless of the graphic on the button. The graphics located at the top of the print/store presets screen represent the buttons on the control panel, not the function of the button.

Workflow for configuring Print and Store:

1. Restore the factory settings and factory printer settings.

2. Configure routing to Print and Store devices.

3. Configure the Print key.

4. Configure the Store key.

5. Configure Advanced Store Transfer (optional).

Restore Factory Settings and Factory Printer Settings 0

Perform the Restore Factory Printer Settings function when configuring print settings for the first time. Perform the Restore Factory Settings function for configuring print and store settings for the first time.

Fig. 129: Print and Store KeysPos. 1 Print Key

Pos. 2 Store Key

Fig. 130: Print Key

Fig. 131: Store Key

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 123 of 150Medical Solutions USA, Inc.

Page 124: Antares Peripheral Instruction manual

124 Printer Setup and Configuration

1. Restore factory printer settings:

a) Select Presets > Service.

b) Select Restore Factory Printer Settings.

Fig. 132: Presets > Service > Restore Printer Factory Defauts

2. Restore factory settings:

a) Select Print/Store.

b) Select Restore Factory Settings.

Fig. 133: Presets > Print/Store > Restore Factory Settings

Selecting Restore Factory Settings populates the different lists with the installed print and store devices.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 124 of 150Medical Solutions USA, Inc.

Page 125: Antares Peripheral Instruction manual

Printer Setup and Configuration 125

3. Continue to the (US03-103.843.01 / Configure Print) section.

Configure Routing to Print and Store Devices 0

Configure the routing for different image types to specific printers and storage devices.

1. Display the Print/Store Screen.

a) Select the Presets icon.

b) Select Print/Store from the Presets menu.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 125 of 150Medical Solutions USA, Inc.

Page 126: Antares Peripheral Instruction manual

126 Printer Setup and Configuration

2. Assign the routing for specific ultrasound image types to the Print and/or Store keys. Use the following table to determine assignments:

Tab. 9 Image Type Routing

BW Selecting a printer from the drop-down menu sends black and white images to the printer selected.

Selecting a storage device from the drop-down menu sends black and white images to the printer selected.

Color Doppler Selecting a printer from the drop-down menu sends color images to the printer selected.

Selecting a storage device from the drop-down menu sends color images to the printer selected.

Tinted Selecting a printer from the drop-down menu sends tinted images to the printer selected.

Selecting a storage device from the drop-down menu sends tinted images to the printer selected.

2D Ref Selecting a printer from the drop-down menu sends images with color in the 2D portion of PW Doppler to the printer selected.

Selecting a storage device from the drop-down menu sends images with color in the 2D portion of PW Doppler to the printer selected.

Misc. Selecting a printer from the drop-down menu sends secondary cap-tures (screen captures) images to the printer selected.

Selecting a storage device from thre drop-down menu sends sec-ondary captures (screen captures) images to the printer selected.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 126 of 150Medical Solutions USA, Inc.

Page 127: Antares Peripheral Instruction manual

Printer Setup and Configuration 127

Fig. 134: Print Routing SetupPos. 1 Print key routing configuration for printing or storing specific image types

Pos. 2 Store key routing configuration for storing or printing specific image types

Pos. 3 Store function assigned to the Print key

Pos. 4 Store function assigned to the Store key

3. Assign the store function to the Print or the Store key.

- Select the Print icon Store checkbox (3/Fig. 134 / p. 127).

Select to perform a dual function: print the image to the related image printer and store the image to the Antares default storage device when the Print key is pressed.

Uncheck to print the image to the related image printer only.

NOTE Images are always stored to the local database when either thePrint or Store key is pressed, regardless of the Print/Store config-uration.

- Select the Store icon Store checkbox.

Select when printers are added to the store device column, to perform a dual function: print the image to the related image printer and store the image to the Antares default storage device when the Store key is pressed.

Uncheck when no printers are assigned to the store device column: to store the images when the Store key is pressed.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 127 of 150Medical Solutions USA, Inc.

Page 128: Antares Peripheral Instruction manual

128 Printer Setup and Configuration

Configure Print 0

After configuring a DICOM or local peripheral print device, assign image types to a desti-nation printer, configure transfer types, and set the printed page composition.

1. Display the Print/Store Screen.

a) Select the Presets icon.

b) Select Print/Store from the Presets menu.

2. Select the Configure Printer tab.

Fig. 135: Print/Store > Configure Printers TabPos. 1 Printer column: list of installed printers

Pos. 2 Formatted by selection: System or Printer

3. For each device in the Printer column (1/Fig. 135 / p. 128), make the following selec-tions from the related drop-down menus:

- Auto Transfer: defines when to transfer static images.

- Layout: defines the number of images per film sheet.

- Media Size: defines the size of the media.

- Orientation: defines the orientation of the media.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 128 of 150Medical Solutions USA, Inc.

Page 129: Antares Peripheral Instruction manual

Printer Setup and Configuration 129

NOTE Siemens recommends that transfer at the “End of Exam” be select-ed for archive servers. Transfer “During Exam” should only be se-lected for image review workstations that support In-Progressreview.

Servers configured for Storage Commitment should not be config-ured for transfer “During Exam”.

Auto transfer to CD-R is disabled by default.

NOTE Do not restore factory settings if this screen is already set up foronboard printers and it is necessary to add a DICOM printer. If thisis the case, skip step 2 and proceed to step 3.

4. Select Printer or System for the Formatted by option (2/Fig. 135 / p. 128).

a) Printer: Assigns the DICOM printers to perform composition and formatting. This is the default setting. This setting is typically more efficient, allowing smaller blocks of data to be transferred over the network, and consuming less ultrasound system CPU resources.

b) System: Assigns the ultrasound system to perform page composition and format-ting. This provides uniformity across different printers, and in some cases results in very large blocks of data being transferred to the printer.

5. After setting the print routing, select Save to save the settings. All changes are discard-ed unless the Save button is selected.

NOTE If the Print or Store key is not configured for printing, then all thedrop-down menus under that key icon should be set to empty.

Configure Store 0

To configure local printers to the Store key, reference the (US03-103.843.01 / Configure Store) section of System Configuration.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 129 of 150Medical Solutions USA, Inc.

Page 130: Antares Peripheral Instruction manual

130 Printer Setup and Configuration

Verify Printer Operation 0

Verify the operation of the printers.

The instructions provided in the following steps are a condensed version of the instruc-tions found in the “Antares Instructions for Use”. See the “Antares Instructions for Use” for a complete set of all printer instructions.

1. Print an image during an active study.

a) Press the Print/Store key assigned to the printing function.

The system sends a request to the printer that is configured for the type of image printed. The image is also stored to the local database.

b) Verify the print function assigned to the Print/Store key was completed.

2. Store an image during an active study:

a) Press the Print/Store key assigned to the storing function.

b) Verify the system stores the image to the local database.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 130 of 150Medical Solutions USA, Inc.

Page 131: Antares Peripheral Instruction manual

VCR Setup and Configuration 131

5- 4VCR Setup and Configuration

Complete the following VCR configurations sections:

• VCR Configuration

• VCR Controls

• Verify the Operation of the VCR

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 131 of 150Medical Solutions USA, Inc.

Page 132: Antares Peripheral Instruction manual

132 VCR Setup and Configuration

HS-MD3000 Configuration 0

When installing a Mitsubishi HS-MD3000 VCR for use with an Antares, it is necessary to optimize the presets for the best playback image quality. Select these presets using the front panel display and controls on the VCR.

Fig. 136: HS-MD3000 VCR Front PanelPos. 1 Power

Pos. 2 Monitor - Mix

Pos. 3 Audio - Mix

Pos. 4 Key Lock - All

Pos. 5 Stop

Pos. 6 Play

Pos. 7 Frame Field/Enter

Pos. 8 Menu

Pos. 9 Jog Dial

The configuration instructions describe how to set up the following preset menus on the HS-MD3000 VCR:

• <Menu 1>

• <Menu 2>

• <Menu 3>

• <Menu 4>

• Hidden Menu

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 132 of 150Medical Solutions USA, Inc.

Page 133: Antares Peripheral Instruction manual

VCR Setup and Configuration 133

Front Panel Controls 0

Tab. 10 VCR Controls

Menu 1 Settings

NOTE This procedure assumes the VCR displays default settings. If thesetting has been changed, the VCR will display the newest set-tings.

1. Press MENU for approximately two seconds. The VCR displays 101:00.

2. Turn the JOG dial to the left or right to select each value within <Menu 1>.

3. Verify that each menu item value displays correctly. The following table lists the correct settings:

Tab. 11 <Menu 1>

Control Function

Menu To access the VCR preset menus, press the Menu button on the front panel of the VCR. The first press of the Menu button displays <MENU 1>. Each press of the Menu button changes the menu sequentially to the next menu. The sequential order of menus is: <MENU 1>, <MENU 2>, <MENU 3>, <MENU4>, <ELAPSED TIME>; and exit the VCR menus.

Stop + Play To enter the “Hidden Menu”, simultaneously press and hold STOP and PLAY for at least ten seconds.

Jog/Adjust Turn the JOG dial to the left or right to select an item within a menu. Turn to the right for a higher numbered setting. Turned to the left for a lower numbered setting.

Frame/ Field - Enter

Press the Enter button to edit a menu value. Then turn the Jog dial to select a different value. Press Enter again to save the change.

Menu Item Antares

Configuration

VCR Front Panel Display

REC TAPE END STOP 101:00

TAPE END STOP 102:00

TAPE IN STOP 103:00

TAPE LENGTH PAL = T-120

NTSC = E-180

104:00

VISS REC 105:01

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 133 of 150Medical Solutions USA, Inc.

Page 134: Antares Peripheral Instruction manual

134 VCR Setup and Configuration

4. If an item does not match:

a) Press the ENTER button. The value after the colon (:) will flash.

b) Turn the JOG dial until the correct value displays.

c) Press the ENTER button to set.

d) Turn the JOG dial to the next item as needed to change menu item values.

e) When the last <Menu 1> item is displayed, press the MENU button. The VCR dis-plays <Menu 2>.

5. Continue to <Menu 2>.

Menu 2 Settings

NOTE This procedure assumes the VCR displays default settings. If thesetting has been changed, the VCR will display the newest set-tings.

When <Menu 2> is selected the VCR displays 201:01.

1. Turn the JOG dial to the left or right to select each value within <Menu 2>.

2. Verify that each menu item value displays correctly. The following table lists the correct settings:

Tab. 12 <Menu 2>

TIME CODE REC OFF 106:00

TIME DATE DISP OFF 107:00

FOOT SWITCH L.E. 108:00

DIMMER OFF 109:00

Menu Item Antares

Configuration

VCR Front Panel Display

S-VHS/S-ET (NTSC)

SVHS (PAL)

ON/OFF

AUTO

201:01

Y-NR ON 202:01

C-NR ON 203:01

Y-ENHANCER ON 204:01

TBC ON 205:01

Menu Item Antares

Configuration

VCR Front Panel Display

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 134 of 150Medical Solutions USA, Inc.

Page 135: Antares Peripheral Instruction manual

VCR Setup and Configuration 135

3. If an item does not match:

a) Press the ENTER button. The value after the colon (:) will flash.

b) Turn the JOG dial until the correct value displays.

c) Press the ENTER button to set.

d) Turn the JOG dial to the next item as needed to change menu item values.

e) When the last <Menu 2> item is displayed, press the MENU button. The VCR dis-plays <Menu 3>.

4. Continue to <Menu 3>.

Menu 3 Settings

NOTE This procedure assumes the VCR displays default settings. If thesetting has been changed, the VCR will display the newest set-tings.

When <Menu 3> is selected the VCR displays 301:00. <Menu 3> is used to set the clock; it is not necessary to set the clock.Press the MENU button to continue to <Menu 4>.

Menu 4 Settings

NOTE This procedure assumes the VCR displays default settings. If thesetting has been changed, the VCR will display the newest set-tings.

When <Menu 4> is selected the VCR displays 401:04.

1. Turn the JOG dial to the left or right to select each value within <Menu 4>.

COLOR FILTER OFF 206:00

FREEZE MODE FRAME 207:00

HI-FI AUDIO ON 208:01

NORMAL AUDIO IN MIX 209:00

Menu Item Antares

Configuration

VCR Front Panel Display

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 135 of 150Medical Solutions USA, Inc.

Page 136: Antares Peripheral Instruction manual

136 VCR Setup and Configuration

2. Verify that each menu item value displays correctly. The following table lists the correct settings:

Tab. 13 <Menu 4>

3. If an item does not match:

a) Press the ENTER button. The value after the colon (:) will flash.

b) Turn the JOG dial until the correct value displays.

c) Press the ENTER button to set the change and go to the next menu item.

d) When the last <Menu 4> item is displayed, press the MENU button twice to exit.

4. Continue to the Hidden Menu.

Hidden Menu Settings

NOTE This procedure assumes the VCR displays default settings. If thesetting has been changed, the VCR will display the newest set-tings.

1. Simultaneously press and hold STOP and PLAY for at least ten seconds. The VCR dis-plays PC: 2.

If PC: 2 is not displayed, turn the JOG dial until PC: 2 is displayed.

2. Press the ENTER button to set the change and continue to the next Hidden Menu item.

3. Verify that each menu item value displays correctly. The following table lists the correct settings:

Tab. 14 Hidden Menu

Menu Item Antares

Configuration

VCR Front Panel Display

TRANSMISSION RATE 19200 401:04

DATA BIT LENGTH 8 BITS 402:01

STOP BIT LENGTH 1 BIT 403:01

PARITY BIT NONE 404:00

Menu Item Antares

Configuration

VCR Front Panel Display

PICTURE CONTROL (PC) 2 PC: 2

SWITCHING POINT (SP) 45 SP:45

POSITION UP/DOWN (Ud) 0 Ud: 0

POSITION L/R (Lr) 0 Lr: 0

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 136 of 150Medical Solutions USA, Inc.

Page 137: Antares Peripheral Instruction manual

VCR Setup and Configuration 137

4. If an item does not match:

a) Press the ENTER button. The value after the colon (:) will flash.

b) Turn the JOG dial until the correct value displays.

c) Press the ENTER button to set.

d) Turn the JOG dial to the next item as needed to change menu item values.

e) When the last item is displayed, press the Stop button to exit.

NOTE A menu item cannot be exited if the menu item is flashing. Pressthe Enter button as needed to stop the menu item from flashing.

CHROMA AMPLI-TUDE (CA) 0 CA: 0

HUE (CP) 0 CP: 0

COLOR DELAY (CD) 1 Cd: 1

BRIGHT-NESS CONTROL (bC)

-1 bC:-1

CONTRAST CONTROL (CC) -1 CC:-1

BLACK DISPLAY (bd) OFF bd:oF

Menu Item Antares

Configuration

VCR Front Panel Display

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 137 of 150Medical Solutions USA, Inc.

Page 138: Antares Peripheral Instruction manual

138 VCR Setup and Configuration

SVO9500 Configuration 0

When installing a Sony SVO9500 VCR for use with an Antares, it is necessary to optimize the presets for the best playback image quality. Select these presets using the front panel display and controls on the VCR.

Front Panel Settings 0

Fig. 137: SVO 9500 VCR Front Panel ControlsPos. 1 Audio switch

Pos. 2 HiFi switch

Pos. 3 Program switch

Pos. 4 CTL/Menu switch

Pos. 5 Menu display

Pos. 6 Power button

Pos. 7 Data setting selection button

Pos. 8 Item menu selection button

Pos. 9 Play button

Tab. 15 SVO 9500 Main Menu Controls

Control Setting Function

Audio Hi-Fi Selects the signal to be output from the Audio Out CH-1 and CH-2 connectors for hi-fi audio to be output.

Hi-Fi ON Records hi-fi audio.

Program OFF The VCR is controled from the Antares system.

CTL/Menu CTL

Menu

Displays the time counter during taping.

Displays the Hidden menu.

Power ON Initiates power to the VCR.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 138 of 150Medical Solutions USA, Inc.

Page 139: Antares Peripheral Instruction manual

VCR Setup and Configuration 139

Main Menu Settings 0

NOTE This procedure assumes the VCR displays default settings. If thesetting has been changed, the VCR will display the newest set-tings.

To display the Main Menu:

1. Set the CTL/Menu switch to Menu.

2. Press the Item button to display each menu and its associated value.

3. Press the Data button to change the item value.

4. Verify the values match the settings in (Tab. 16 / p. 139).

Tab. 16 SVO 9500 Main Menu Items

Data N.A. When the CTL/Menu control is set to Menu, press Data to scroll through the settings of each hidden menu item.

Item N.A. When the CTL/Menu control is set to Menu, press Item to advance to each of the main menu items.

Play N.A. Press Play to play back recorded video.

Control Setting Function

VCR Front Panel Display

Antares Setting

Option

vIA- ON VISS operation mode.

FOt-- LE Footswitch operation mode.

tPE- Et Tape end operation mode. Automatically ejects tape when the tape reaches the end.

sVH- At S-VHS recoreing selection. Automatically detects tape type.

Crn- ON Digital CRN On/Off. Switches on the digital chroma noise reducer.

Yrn- ON Digital YRN On/Off. Switches on the digital luminance noise reducer.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 139 of 150Medical Solutions USA, Inc.

Page 140: Antares Peripheral Instruction manual

140 VCR Setup and Configuration

Hidden Menu Settings 0

NOTE This procedure assumes the VCR displays default settings. If thesetting has been changed, the VCR will display the newest set-tings.

To access the Hidden menu:

1. Set the CTL/Menu switch to Menu.

2. Press the Item button until FrE-Fr displays in the menu window.

3. Simultaneously press Item and Play.

4. Press the Item button to display each menu and its associated value.

5. Press the Data button to change the item value.

6. Verify that each menu item value displays correctly. The following table lists the correct settings:

Tab. 17 SVO 9500 Hidden Menu Items

Yen- ON Digital Y enhancer On/Off. Switches on the digital Y enhancer.

FrE- Fr Freeze moe selection. Displays a frame-freeze picture.

VCR Front Panel Display

Antares Setting

Option

VCR Front Panel Display

Antares Setting

Option

SOP- OFF Pause status on the 34P 1/F becomes “true” when the speed is “zero” in Search mode, asa well as Pause mode.

34P- PA When the host sends a pause command through the 34P I/F, the VCR goes into Freeze instead of Pause (default is PA). It does not function when the VCR is in Jog/Shuttle or search mode. It is available in Play mode only.

FrS- PA After the Auto Freeze function completes, the VCR goes into “speed ero” of the Shuttle mode, instead of Pause mode.

1PH- 65 The switching position loch phase can be shifted to either 6.5H, 5.5H, or 4.5H before Vsync. The phase in REC stays at 6.5H always. 6.5H must be selected when adjusting the switching postition.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 140 of 150Medical Solutions USA, Inc.

Page 141: Antares Peripheral Instruction manual

VCR Setup and Configuration 141

7. If an item does not match:

a) Press the ENTER button. The value after the colon (:) will flash.

b) Turn the JOG dial until the correct value displays.

c) Press the ENTER button to set.

d) Turn the JOG dial to the next item as needed to change menu item values.

e) When the last item is displayed, press the Stop button to exit.

NOTE A menu item cannot be exited if the menu item is flashing. Pressthe Enter button as needed to stop the menu item from flashing.

L1n- ON Defeats the line resizing function in the Reverse Search mode. It reduces the “image remaining” phenomenon. The image is stretched verticallly as a side effect.

nOn- OFF Changes to “Non-Interlaced” mode om Pause and Field Freeze. It is valid only if the VM-12 baord is installed, other-wise no image comes out.

FrA- ON Switches the framing servo off. Field Memory mode can be selected. It is valid only if the “Non-Interlaced” mode is on.

H s- OFF Shifts the image to the left approximately 350 ns whenit is on. The default is off.

v s- OFF Shifts the image down four lines during Jog, Shuttle, or Pause modes. The default is off.

VCR Front Panel Display

Antares Setting

Option

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 141 of 150Medical Solutions USA, Inc.

Page 142: Antares Peripheral Instruction manual

142 VCR Setup and Configuration

Verify the Operation of the VCR 0

1. Power on the Antares system (if not already on).

2. Power on the VCR (if not already on).

3. Insert a VCR tape into the VCR.

NOTE See the VCR manual for detailed steps.

4. Press the VCR key to toggle between the record and the pause functions.

The VCR key illuminates when the Record function is active.

Fig. 138: VCR - Record/ Pause Key

5. Record the active screen to video tape. Use one of the following methods:

- Use the VCR panel controls to record the active screen.

- Use the VCR group box to control the VCR directly from the Image task card (if this option is installed on the system).

6. Playback the recording. Use one of the following methods:

- Use the VCR panel controls to record the active screen.

- Use the VCR group box to control the VCR directly from the Image task card (if this option is installed on the system).

The playback should display the active screen as viewed during the recording period.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 142 of 150Medical Solutions USA, Inc.

Page 143: Antares Peripheral Instruction manual

Printer Troubleshooting 143

6- 5Printer TroubleshootingPrinter Installation Troubleshooting 0

If the test page does not print, refer to the following list for possible causes:

• The printer is connected to AC power.

• The printer is powered on.

• The printer cable is securely connected to the system. When using a parallel cable, tighten the locking screws.

• The printer is not out of paper and/or ribbon.

• The printer is On Line (On Line light is illuminated).

If the printer is still not functioning, consult the following table:

Tab. 18 Printer Installation Troubleshooting

Symptom Cause Solution

Plug-and-play printers do not install when connected to the sys-tem.

Plug-and-play resources are used and must be restored.

1. Follow steps in the “Restoring Plug-and-play Resources”.

2. Using the Add/Remove Hardware Wizard, remove hardware device.

3. Repeat the instructions to install the printer.

Unable to print.

Either nothing happens or an error displays.

The printer was deleted from the Win-dows Printer page or is not correctly config-ured in the system Print/Store routing page.

Repeat the instructions to install the printer or check entries in Print/Store routing page. Printers are not dis-played as selections in the Print/ Store routing page until completing the installation and a system reboot.

Print out does not match screen.

Printer defaults may not be correct.

Select Presets > Service > Restore Printer Factory Defaults. Close Pre-sets screen.

The CP800DW printer does not print.

SCSI mode may be enabled. Check the SCSI light on front panel.

SW1 on rear panel must be on for parallel mode.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 143 of 150Medical Solutions USA, Inc.

Page 144: Antares Peripheral Instruction manual

144 Printer Troubleshooting

Report contains prints that are shrunken or expanded.

Printer defaults may be incorrect.

1. Select Presets > Service > Con-figure Local Printer.

2. Right-click on the report printer and select Properties > Printing Preferences > Layout > Ad-vanced > Graphic > Print Quality > 600 DPI

3. Select Ok to close the Advanced Options screen.

4. Select Ok to close the Printing Preferences screen.

5. Select Ok to close the Properties screen.

6. Close the Printer window.

7. Close the Presets screen.

An error message dis-plays on the Printer screen.

The Printer Error indi-cator lamp illuminates.

Check printer manual for list.

Follow instructions on the printer screen

Identify the error indicator resolution in the printer manual.

Symptom Cause Solution

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 144 of 150Medical Solutions USA, Inc.

Page 145: Antares Peripheral Instruction manual

Printer Troubleshooting 145

Restoring Plug-and-Play Resources 0

A driver can be installed for a plug-and-play resource only once. At which time Windows marks that resource as used.(A used resource displays as a grayed-out device on the hardware list.) If a printer is not recognized and cannot be installed, the resource may need to be restored. The device must be deleted and the plug-and-play function re-initi-ated.

NOTE A service level 4 password or higher is required to restoreplug-and-play resources.

NOTE After restoring the plug-and-play resources, the USB cable mustbe disconnected and reconnected to initiate the plug-and-play in-stallation.

To restore the Plug-and-play resources:

1. Delete the printer from the Configure Local Printer screen:

a) Disconnect from the I/O panel, all USB cables connected to thermal printers.

It is not necessary to disconnect the LPT cable for a Report Printer if one is installed.

b) Select Presets > Service > Configure Local Printer.c) Right-click the Printer icon (if present) and select Delete. Repeat for each printer

that is to be deleted.

d) Select Yes.

If printer deleted was the default printer, the system displays a warning message.

e) Select Ok. Close Printer folder.

2. Access NT Command Interpreter:

a) Select Local Service.

b) Enter the password in the password field and select Ok.

c) Select Utilities.

d) Select Escape to OS from the Source drop-down list.

e) Select NT Command Interpreter from the Command drop-down list.

3. Access and delete the printer through Explorer:

a) Type explorer in the Parameter field.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 145 of 150Medical Solutions USA, Inc.

Page 146: Antares Peripheral Instruction manual

146 Printer Troubleshooting

b) Select Go.

c) Select Next.d) Right-click My Computer, then select Properties > Hardware > Device Manager.e) Select the View pull-down menu.

f) Select Show hidden devices and expand Printers.

g) Select the printer for deletion. Right-click the selected printer and select Uninstall.h) Confirm deletion of printer.

i) Expand Universal Serial Bus controllers, then select the appropriate printing sup-port for the printer to be deleted.

j) Confirm deletion of USB Printing Support.

Repeat steps 3 for each USB Printing Support device entry from the list in turn. There is a ‘USB Printing Support’ device for the USB to parallel cable, which is in-stalled as a separate device. All USB Printing Support devices should be removed to free up resources.

4. Close the Service screen.

Printers may now be installed according to the specific installation procedure in this document.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 146 of 150Medical Solutions USA, Inc.

Page 147: Antares Peripheral Instruction manual

Printer Troubleshooting 147

Re-Install A Printer After Troubleshooting or Restoring Plug-and-Play 0

Use the following table to determine which installation procedure to use.

Tab. 19 Printer Re-installation Instructions

Report Printers 0

The following steps manual printer installation if problems are encountered during the software install process or if add a printer to an existing, working system.

NOTE If the printer is unplugged or removed from the system, refer to the(Restoring Plug-and-Play Resources / p. 145) for the printer unin-stall procedure.

HP1150, HP2500n, HP 6122

The Antares system must be powered off before beginning the printer installation.

1. Power on the report printer and permit printer to initialize, illuminating the “Ready” light.

2. Connect the USB cable between the printer and the Antares:

a) Connect one end of the USB cable into the USB port on the back of the printer.

b) Route the cable through the peripheral cable passage.

c) Connect the other end of the USB cable into a USB port on the I/O panel.

3. The system will detect the printer and display the “Found New Hardware Wizard” win-dow.

a) If not already selected, select the Install the software automatically (recom-mended) check box.

Printers Manufacturer Use Instruction

HP1150, HP2500n, HP6122

HP4050, HP4100, HP4200,

HP (Report Printers / p. 147)

CP800WD Mitsubishi SW v2.1 or v3.0: (CP800 Installation / p. 100)

SW v4.0: (CP800 Installation / p. 110)

P91DW, P93DW, CP770DW

UPD23MD, UPD55

Mitsubishi

Sony

(Re-install Plug-and-Play Printers / p. 148)

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 147 of 150Medical Solutions USA, Inc.

Page 148: Antares Peripheral Instruction manual

148 Printer Troubleshooting

b) Select Next. The system installs the USB software automatically.

c) Select Finish.

The “Found New Hardware Wizard” closes and the system returns to imaging.

4. Press the Patient button. The system displays the Patient Registration screen.

5. Register a patient and select Ok.

6. Press the Report button. The system displays the Patient Report screen.

7. Select Print Report.

8. Compare the printed report to the information on the screen.

HP 4050, HP4100, HP4200

The Antares system must be powered off before beginning the printer installation.

1. Connect the parallel cable between the printer and the Antares:

a) Connect the parallel cable into the parallel printer connector on the I/O panel and secure with the locking screws.

b) Route the cable through the peripheral cable passage.

c) Connect the other end of the parallel cable into the parallel connector on the back of the printer and lock it down with the clips.

2. Power on the Report Printer on.

Wait until the printer displays Ready on the front panel.

The system detects the printer and displays the “Found New Hardware Wizard” window.

a) If not already selected, select the Install the software automatically (recom-mended) check box.

b) Select Next. The system installs the USB software automatically.

c) Select Finish.

The “Found New Hardware Wizard” closes and the system returns to imaging.

3. Press the Patient button. The system displays the Patient Registration screen.

4. Register a patient and select Ok.

5. Press the Report button. The system displays the Patient Report screen.

6. Select Print Report.

7. Compare the printed report to the information on the screen.

Re-install Plug-and-Play Printers 0

The Plug-and-play procedure assumes the system is booted to imaging. Perform the fol-lowing steps for each printer to install; one printer at a time.

1. Connect the AC power cable between the printer and the Antares peripheral outlet.

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 148 of 150Medical Solutions USA, Inc.

Page 149: Antares Peripheral Instruction manual

Printer Troubleshooting 149

2. Press the printer Power button to power on the printer.

3. Connect the USB cable between the printer and the I/O panel.

a) Connect the square-shaped end of the USB cable to the rear panel of the printer.

b) Route the “USB cable through the peripheral cable passage.

c) Connect the rectangular-shaped end of the USB cable to J5B for color printers, or J5A for black and white printers.

4. The system detects the plug-and-play printers. The system displays the “Found New Hardware Wizard” for USB printing support.

a) If not already selected, select the Install the software automatically (recom-mended) check box.

b) Select Next. The system installs the USB software automatically.

c) Select Finish.

The “Found New Hardware Wizard” closes.

Siemens US03-201.812.01.02.02 Antares07.05 sd

Page 149 of 150Medical Solutions USA, Inc.

Page 150: Antares Peripheral Instruction manual

150 Printer Troubleshooting

Antares US03-201.812.01.02.02 Siemens07.05 sd

Page 150 of 150Medical Solutions USA, Inc.