Analysis of corticomuscular coherence
tutorial, coherence, meg, emg, plot, source, connectivity
Analysis of corticomuscular coherence
Introduction
In this tutorial we will analyze cortico-muscular coherence, which reflects functional connectivity
between primary motor cortex and a contralateral effector muscle during isometric contraction.
The dataset used in this example has been recorded in an experiment in which the subject had to lift
her hand and exert a constant force against a lever. The force was monitored by strain gauges on
the lever. The subject performed two blocks of 20 trials in which either the left or the right wrist was
extended for about 10 seconds. A trial started as soon as the subject managed to get his force output
within a specified range from 1 to 2 N. If the force was not kept constant during the course of a trial,
the trial was terminated prematurely.
The bipolar EMG signal was recorded from the right extensor carpi radialis longus muscle in the
lower arm. MEG signals were recorded with a 151 sensor CTF Omega System (Port Coquitlam,
Canada). In addition, the EOG was recorded to later discard trials contaminated by eye movements
and blinks. The ongoing MEG and EOG signals were lowpass filtered at 300 Hz, digitized at 1200 Hz
and stored for off-line analysis. To measure the head position with respect to the sensors, three coils
were placed at anatomical landmarks of the head (nasion, left and right ear canal). While the
subjects were seated under the MEG helmet, the positions of the coils were determined before and
after the experiment by measuring the magnetic signals produced by currents passed through the
coils. Magnetic resonance images (MRIs) were obtained from a 1.5 T Siemens system. During theMRI scan, ear molds containing small containers filled with vitamin E marked the same landmarks.
This allows us, together with the anatomical landmarks, to align source estimates of the MEG with
the MRI.
Background
To study the oscillatory synchrony between two signals, one can compute the coherence. This is
computed in the frequency domain by normalizing the magnitude of the summed cross-spectraldensity between two signals by their respective power. For each frequency bin the coherence value
is a number between 0 and 1. The coherence values reflect the consistency of the phase difference
between the two signals at a given frequency. In this session we will explore the concept of
coherence by investigating a dataset from an experiment in which the subject was required to
maintain an isometric contraction of a forearm muscle. The coherence between the MEG signals and
the acquired EMG will be estimated. First we will explore the coherence between the EMG signal
and all MEG channels. Secondly, we will investigate how the coherence estimate is influenced by the
number of trials, and by the degree of spectral smoothing using multitaper spectral analysis. Even
though the example in this session covers cortico-muscular coherence, coherence between sensors
can be calculated in exactly the same way.
http://fieldtrip.fcdonders.nl/ Printed on 2013/01/27 21:09
7/18/2019 Analysis of Corticomuscular Coherence
http://slidepdf.com/reader/full/analysis-of-corticomuscular-coherence 2/11
Analysis of corticomuscular coherence
Procedure
To compute the coherence between the MEG and EMG signals for the example dataset we will
perform the following steps:
Read the data into Matlab using ft_preprocessingq
Compute the power spectra and cross-spectral densities using the function ft_freqanalysis andq
subsequently compute the coherence using ft_connectivityanalysis
Visualize the results using ft_singleplotER , ft_multiplotER , and ft_topoplotER q
Subsequently it is possible to localise the neuronal sources coherent with the EMG, usingq
ft_sourceanalysis
Preprocessing
We will calculate the coherence between the MEG and the EMG when the subject extended herLEFT wrist, while keeping the right forearm muscle relaxed. The first step is to read the data. In this
section we will apply automatic artifact rejection. Preprocessing requires the original MEG dataset,
which is available from ftp://ftp.fcdonders.nl/pub/fieldtrip/tutorial/SubjectCMC.zip.
The epochs of interest have to be defined according to a custom-written function called
trialfun_left.m. Note that this function is not part of the FieldTrip toolbox: see appendix 2, or
download it from ftp://ftp.fcdonders.nl/pub/fieldtrip/tutorial/coherence/trialfun_left.m. This function
uses the information provided by the triggers which were recorded simultaneously with the data. In
this experiment each trigger corresponds with the start or the end of a contraction. The epochs
which correspond to a contraction of the left forearm muscle are selected. Subsequently these10-second pieces are cut into ten 1-second epochs.
% find the interesting epochs of data
cfg = [];
cfg.trialfun = 'trialfun_left';
cfg.dataset = 'SubjectCMC.ds';
cfg = ft_definetrial(cfg);
% detect EOG artifacts in the MEG data
cfg.continuous = 'yes';
cfg.artfctdef.eog.padding = 0;
cfg.artfctdef.eog.bpfilter = 'no';
cfg.artfctdef.eog.detrend = 'yes';
cfg.artfctdef.eog.hilbert = 'no';
cfg.artfctdef.eog.rectify = 'yes';
cfg.artfctdef.eog.cutoff = 2.5;
cfg.artfctdef.eog.interactive = 'no';
cfg = ft_artifact_eog(cfg);
% detect jump artifacts in the MEG data
cfg.artfctdef.jump.interactive = 'no';cfg.padding = 5;
cfg = ft_artifact_jump(cfg);
Printed on 2013/01/27 21:09 http://fieldtrip.fcdonders.nl/
7/18/2019 Analysis of Corticomuscular Coherence
http://slidepdf.com/reader/full/analysis-of-corticomuscular-coherence 3/11
Analysis of corticomuscular coherence
% detect muscle artifacts in the MEG data
cfg.artfctdef.muscle.cutoff = 8;
cfg.artfctdef.muscle.interactive = 'no';
cfg = ft_artifact_muscle(cfg);
% reject the epochs that contain artifacts
cfg.artfctdef.reject = 'complete';
cfg = ft_rejectartifact(cfg);
% preprocess the MEG data
cfg.demean = 'yes';
cfg.dftfilter = 'yes';
cfg.channel = {'MEG'};
cfg.continuous = 'yes';
meg = ft_preprocessing(cfg);
Next, read the left and right EMG data. Note that the settings are different for the EMG and MEG
data. Most importantly, the EMG data are highpass filtered and rectified. This is a standardprocedure when calculating cortico-muscle coherence.
cfg = [];
cfg.dataset = meg.cfg.dataset;
cfg.trl = meg.cfg.trl;
cfg.continuous = 'yes';
cfg.demean = 'yes';
cfg.dftfilter = 'yes';
cfg.channel = {'EMGlft' 'EMGrgt'};
cfg.hpfilter = 'yes';cfg.hpfreq = 10;
cfg.rectify = 'yes';
emg = ft_preprocessing(cfg);
Finally, combine the EMG and MEG trials to a common data structure:
data = ft_appenddata([], meg, emg);
Note, that due to the artifact rejection, this procedure is very slow and we typically would want to
perform this step only once. Therefore you can save the preprocessed data:
save data data
The preprocessed data is available as a mat-file from the FieldTrip ftp server (data.mat) and you can
skip the preprocessing above by loading the data like this
load data
To get a feel for the data, plot a trial from a sensor overlying the left motor-cortex (MRC21) and the
left and right EMG-signals, by selecting the first trial from the data:
figuresubplot(2,1,1);
plot(data.time{1},data.trial{1}(77,:));
http://fieldtrip.fcdonders.nl/ Printed on 2013/01/27 21:09
7/18/2019 Analysis of Corticomuscular Coherence
http://slidepdf.com/reader/full/analysis-of-corticomuscular-coherence 4/11
Analysis of corticomuscular coherence
axis tight;
legend(data.label(77));
subplot(2,1,2);
plot(data.time{1},data.trial{1}(152:153,:));
axis tight;
legend(data.label(152:153));
Figure 1; An example of the raw MEG data from sensor MLC21 (upper frame) and the EMG data
(lower frame). The signals are from the output of ft_preprocessing and plotted using the matlab
plot function. Note that the signal strength of the left EMG is bigger than that of the right EMG.
Exercise 1
Explore the MEG and EMG in figure 1, e.g. by zooming in. How are the signals different from one
another?
Computing the coherence
Using ft_freqanalysis, the characteristics in the frequency domain will be computed. This step
requires the preprocessed MEG and EMG data (see above or download from
ftp://ftp.fcdonders.nl/pub/fieldtrip/tutorial/coherence/data.mat). Load the data with:
load data
After the computation of the frequency domain representation ft_connectivityanalysis will be used
to compute the coherence. There are essentially two ways of achieving the same coherence, and
these will be explained below. The main difference is the way in which the frequency domain
representation is computed. More information about the different options can be found here
Method 1
In this 'method' we will use ft_freqanalysis for the computation of the fourier spectra, which is the
'bare' frequency domain representation of the signal, where both amplitude and phase information
of the oscillations are represented in a complex number for each frequency.
First, a configuration structure (cfg) must be defined. The FFT-algorithm will be used to compute the
fourier representation of each signal. To optimize the estimation, spectral smoothing using
‘multitapers’ will be applied. In this context, the degree of ‘smoothing’ (as defined in the parameter
cfg.tapsmofrq) is critical. We will return to this parameter later.
cfg = [];
cfg.output = 'fourier';
cfg.method = 'mtmfft';
cfg.foilim = [5 100];
Printed on 2013/01/27 21:09 http://fieldtrip.fcdonders.nl/
7/18/2019 Analysis of Corticomuscular Coherence
http://slidepdf.com/reader/full/analysis-of-corticomuscular-coherence 5/11
Analysis of corticomuscular coherence
cfg.tapsmofrq = 5;
cfg.keeptrials = 'yes';
cfg.channel = {'MEG' 'EMGlft' 'EMGrgt'};
freqfourier = ft_freqanalysis(cfg, data);
Method 2
In this 'method' we will use ft_freqanalysis for the computation of the cross- and power spectra,
which are mathematically constructed from the multiplication of a complex-valued fourier spectrum
with the complex conjugate of another fourier spectrum. If the two fourier spectra are derived from
the same channel, the cross spectrum is called the auto spectrum, and this is exactly the same as the
power spectrum. Rather than in method 1 (see above), where the phase in the fourier spectra
represented the phase of the oscillation, the phase in the cross spectra represent the phase
difference between the oscillations of a specific channel pair. For this reason, we need to specify in
the cfg between which pairs we want to compute the cross spectra.
cfg = [];
cfg.output = 'powandcsd';
cfg.method = 'mtmfft';
cfg.foilim = [5 100];
cfg.tapsmofrq = 5;
cfg.keeptrials = 'yes';
cfg.channel = {'MEG' 'EMGlft' 'EMGrgt'};
cfg.channelcmb = {'MEG' 'EMGlft'; 'MEG' 'EMGrgt'};
freq = ft_freqanalysis(cfg, data);
To calculate the coherence between the EMG and the MEG signals from the fourier spectra, or fromthe power- and cross spectra use the following function. This function does not care whether the
input data contains fourier spectra, or power/cross spectra.
cfg = [];
cfg.method = 'coh';
cfg.channelcmb = {'MEG' 'EMG'};
fd = ft_connectivityanalysis(cfg, freq);
fdfourier = ft_connectivityanalysis(cfg, freqfourier);
The field fd.cohspctrm/fdfourier.cohspctrm now contains the coherence for all MEG sensors with
respect to the EMG signals.
Displaying the coherence
Visualize the coherence between the EMG and all the MEG sensors:
cfg = [];
cfg.parameter = 'cohspctrm';
cfg.xlim = [5 80];cfg.refchannel = 'EMGlft';
cfg.layout = 'CTF151.lay';
cfg.showlabels = 'yes';
http://fieldtrip.fcdonders.nl/ Printed on 2013/01/27 21:09
7/18/2019 Analysis of Corticomuscular Coherence
http://slidepdf.com/reader/full/analysis-of-corticomuscular-coherence 6/11
Analysis of corticomuscular coherence
figure; ft_multiplotER(cfg, fd)
Figure 2; The coherence between the left EMG and all the MEG sensors calculated using
ft_freqanalysis and ft_connectivityanalysis. Plotting was done with ft_multiplotER.
Plot the coherence for sensor MRC21 (using the same settings as in ft_multiplotER ):
cfg.channel = 'MRC21';
figure; ft_singleplotER(cfg, fd);
Figure 3; The coherence spectrum between the EMG and sensor MRC21.
Exercise 2
a) What determines the frequency resolution of the spectrum, as displayed in figure 3? How can it be
increased or decreased? Answer the same question for smoothing.
b) Plot a topographical distribution of the coherence in the beta band. The variable cfg.xlim defines
the edges of the frequency band.
cfg = [];
cfg.parameter = 'cohspctrm';
cfg.xlim = [15 20];cfg.zlim = [0 0.1];
cfg.refchannel = 'EMGlft';
cfg.layout = 'CTF151.lay';
figure; ft_topoplotER(cfg, fd)
Figure 4; A topographic representation of the coherence between the left EMG and the sensors. The
plot was created with ft_topoplotER.
Exercise 3
a) Explain the pattern of activation in Figure 4.
b) Plot the topographic representation for other frequencies that might be of interest.
Exercise 4
Explore the consequence of changing the smoothing in the frequency domain. Do this by
recomputing the cortico-muscular coherence between the EMG signal and MEG sensor MRC21 for
Printed on 2013/01/27 21:09 http://fieldtrip.fcdonders.nl/
7/18/2019 Analysis of Corticomuscular Coherence
http://slidepdf.com/reader/full/analysis-of-corticomuscular-coherence 7/11
Analysis of corticomuscular coherence
different degrees of smoothing. Compute the powerspectra and the cross-spectra, and the
corresponding coherence using different degrees of smoothing.
a) 2 Hz smoothing (cfg.tapsmofrq = 2 Hz)
cfg = [];
cfg.output = 'powandcsd';
cfg.method = 'mtmfft';cfg.foilim = [5 100];
cfg.tapsmofrq = 2;
cfg.keeptrials = 'yes';
cfg.channel = {'MEG' 'EMGlft'};
cfg.channelcmb = {'MEG' 'EMGlft'};
freq2 = ft_freqanalysis(cfg,data);
cfg = [];
cfg.method = 'coh';
cfg.channelcmb = {'MEG' 'EMG'};fd2 = ft_connectivityanalysis(cfg,freq2);
Plot the results of the 5 and 2Hz smoothing:
cfg = [];
cfg.parameter = 'cohspctrm';
cfg.refchannel = 'EMGlft';
cfg.xlim = [5 80];
cfg.channel = 'MRC21';
figure; ft_singleplotER(cfg, fd, fd2);
b) 10 Hz smoothing (e.g. cfg.tapsmofrq = 10 Hz)
cfg = [];
cfg.output = 'powandcsd';
cfg.method = 'mtmfft';
cfg.foilim = [5 100];
cfg.keeptrials = 'yes';
cfg.channel = {'MEG' 'EMGlft'};
cfg.channelcmb = {'MEG' 'EMGlft'};
cfg.tapsmofrq = 10;freq10 = ft_freqanalysis(cfg,data);
cfg = [];
cfg.method = 'coh';
cfg.channelcmb = {'MEG' 'EMG'};
fd10 = ft_connectivityanalysis(cfg,freq10);
Plot the results of the 5, 2, and 10 Hz smoothing:
cfg = [];
cfg.parameter = 'cohspctrm';
cfg.xlim = [5 80];
cfg.ylim = [0 0.2];
http://fieldtrip.fcdonders.nl/ Printed on 2013/01/27 21:09
7/18/2019 Analysis of Corticomuscular Coherence
http://slidepdf.com/reader/full/analysis-of-corticomuscular-coherence 8/11
Analysis of corticomuscular coherence
cfg.refchannel = 'EMGlft';
cfg.channel = 'MRC21';
figure;ft_singleplotER(cfg, fd, fd2, fd10);
Which degree of smoothing do you consider optimal in the calculations above?
Exercise 5
Another question pertains to how the estimate of coherence is affected by the number of trials. We
will compare the cortico-muscular coherence at two MEG sensors for different amount of data.
Create the following configuration, and compute the coherence.
cfg = [];
cfg.output = 'powandcsd';
cfg.method = 'mtmfft';cfg.foilim = [5 100];
cfg.tapsmofrq = 5;
cfg.keeptrials = 'yes';
cfg.channel = {'MEG' 'EMGlft'};
cfg.channelcmb = {'MEG' 'EMGlft'};
cfg.trials = 1:50;
freq50 = ft_freqanalysis(cfg,data);
cfg = [];
cfg.method = 'coh';cfg.channelcmb = {'MEG' 'EMG'};
fd50 = ft_connectivityanalysis(cfg,freq50);
Plot the results:
cfg = [];
cfg.parameter = 'cohspctrm';
cfg.xlim = [5 100];
cfg.ylim = [0 0.2];
cfg.refchannel = 'EMGlft';
cfg.channel = 'MRC21';
figure; ft_singleplotER(cfg, fd, fd50);
Compare the results with figure 3. Pay special attention to the noise bias.
Appendix 1: Localisation of neuronal sources coherent with
the EMG using beamformers
In order to localise the neuronal sources which are coherent with the EMG, we can apply
beamformers to the data. For a more extensive background in beamforming, in particular
beamforming with frequency-domain data, please consult the beamformer tutorial. In this example,
Printed on 2013/01/27 21:09 http://fieldtrip.fcdonders.nl/
7/18/2019 Analysis of Corticomuscular Coherence
http://slidepdf.com/reader/full/analysis-of-corticomuscular-coherence 9/11
Analysis of corticomuscular coherence
we are going to use an algorithm, known as DICS, to estimate the activity of the neuronal sources
and to subsequently estimate the coherence with the EMG. In order to achieve this, we first need an
estimate of the cross-spectral density between all MEG-channel combinations, and between the
MEG-channels and the EMG, at a frequency of interest. This requires the preprocessed data, see
above, or download from ftp://ftp.fcdonders.nl/pub/fieldtrip/tutorial/coherence/data.mat. Load with:
load data
Compute the cross-spectral density matrix for 18 Hz:
cfg = [];
cfg.method = 'mtmfft';
cfg.output = 'powandcsd';
cfg.foilim = [18 18];
cfg.tapsmofrq = 5;
cfg.keeptrials = 'yes';
cfg.channelcmb = {'MEG' 'MEG';'MEG' 'EMGlft'};
freq = ft_freqanalysis(cfg, data);
Once we computed this, we can use ft_sourceanalysis using the following configuration.
This step requires the subject's headmodel, which is available from
ftp://ftp.fcdonders.nl/pub/fieldtrip/tutorial/SubjectCMC.zip.
cfg = [];
cfg.method = 'dics';
cfg.refchan = 'EMGlft';
cfg.frequency = 18;
cfg.hdmfile = 'SubjectCMC.hdm';
cfg.inwardshift = 1;
cfg.grid.resolution = 1;
source = ft_sourceanalysis(cfg, freq);
The resulting source-structure is a volumetric reconstruction which is specified in head-coordinates.
In order to be able to visualise the result with respect to the subject's MRI, we have to interpolate
the functional data to the anatomical MRI.
For this, we need the subject's MRI, which is available from
ftp://ftp.fcdonders.nl/pub/fieldtrip/tutorial/SubjectCMC.zip. After reading the anatomical MRI, we
reslice it along the axes of the head coordinate system for improved visualization.
mri = ft_read_mri('SubjectCMC.mri');
cfg = [];
mri = ft_volumereslice(cfg, mri);
Next, we can proceed with the interpolation.
cfg = [];
cfg.parameter = 'coh';
cfg.downsample = 2;
interp = ft_sourceinterpolate(cfg, source, mri);
There are various ways to visualise the volumetric interpolated data. The most straightforward way
http://fieldtrip.fcdonders.nl/ Printed on 2013/01/27 21:09
7/18/2019 Analysis of Corticomuscular Coherence
http://slidepdf.com/reader/full/analysis-of-corticomuscular-coherence 10/11
7/18/2019 Analysis of Corticomuscular Coherence
http://slidepdf.com/reader/full/analysis-of-corticomuscular-coherence 11/11
Analysis of corticomuscular coherence
indx = indx(sel);
trl = [];
for j = 1:length(trig)-1
trg1 = trig(j);
trg2 = trig(j+1);
if trg1<=100 & trg2==2080,
trlok = [[indx(j)+1:1200:indx(j+1)-1200]'
[indx(j)+1200:1200:indx(j+1)]'];
trlok(:,3) = [0:-1200:-1200*(size(trlok,1)-1)]';
trl = [trl; trlok];
end
end
This tutorial was last tested by Jan-Mathijs with revision 4363 of FieldTrip (~20111006) on a 64-bit
linux machine using Matlab 2009b.