Conventional instrument: Spatial resolution about 20-50 μm NanoSIMS Spatial resolution about 50-200 nm An ion microprobe is a SIMS technique S econdary I on M ass-S pectroscopy Objective: generate secondary ions Sputtering a sample with beam of primary ions

Transcript of An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of...

Page 1: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

Conventional instrument: Spatial resolution about 20-50 µm

NanoSIMS Spatial resolution about 50-200 nm

An ion microprobe is a SIMS technique Secondary Ion Mass-Spectroscopy Objective: generate secondary ions Sputtering a sample with beam of primary ions

Page 2: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

1 mm

Lophelia deep-sea coral

NanoSIMS Normal SIMS

Page 3: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

5-20 nm

NanoSIMS primary ions: Cs+ or O-

Page 4: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

Mass spectrometry: Lorentz Force: F = q·v×B

Page 5: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

Force = q·v×B = m·a Circular motion: a = v2/r r = (m/q) · (v/B)

Page 6: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (
Page 7: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

I(iM) = db * S * Sy * XM * iAM * YM * T

C/S

primary ions s * area

area

Sputter yield: atoms removed from surface per incoming ion

Atomic fraction

Ionization yield: Mi / M sputtered

Transmission

Isotopic abundance

How many counts per seconds of the isotope iM of the element M?

Page 8: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

I(iM) = db * S * Sy * XM * iAM * YM * T The useful yield τ :

τ = YM * T

τ is the fraction of sputtered atoms M that are ionized and detected.

1/τ is the number of atoms M that must be removed from the sample in order to detect one iM.

Typically 1/τ is on order of 103-105.

Page 9: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

What can τ teach us? How big a volume we have to sputter to get one detection of a trace element: Example (Fontcuberta): 100 nm thick GaAs nanowire with trace-amounts of Si ρ  = 4.4*1022 atoms/cm3

XSi = 4.5*10-4

1/τ = 1000

To detect one Si ion, we need to sputter a volume of:

(1/τ) * (1/XSi) * (1/ρ) ≈ 5*10-5 µm3

(Assume that Si has only one isotope. 28Si is 92 % : iAM ≈ 1)

Page 10: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

To detect one Si ion, we need to sputter a volume of:

(1/τ) * (1/XSi) * (1/ρ) ≈ 5*10-5 µm3

With a beam spot of 50*50 nm2 = 2.5*10-3 µm2 the required sputter depth is: 5*10-5 µm3 / 2.5*10-3 µm2 = 0.02 µm = 20 nm But the GaAs nanowire is only 100 nm thick. Thus, only about 5 counts of Si can be expected before the wire has been sputtered away! To get 100 counts (10% precision): 20 * 20 nm = 400 nm

Page 11: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

Distribution Si dopant in two different YAG (Ytrium, Aluminum Garnet) samples showing two different segregation patterns (grain boundary and triple points) after thermal treatment. Field: 10µm x 10µm. Maximum Silicon concentration at grain boundary: 0.1 at. % => 50 counts

0.1µm lateral resolution

Page 12: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

Distribution Si dopant in two different YAG (Ytrium, Aluminum Garnet) samples showing two different segregation patterns (grain boundary and triple points) after thermal treatment. Field: 10µm x 10µm. Maximum Silicon concentration at grain boundary: 0.1 at. % => 50 counts

0.1µm lateral resolution

Typical beam size in conventional ion miroprobe

Page 13: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

Symbiosis between corals an dinoflagellate algae (zooxanthellae) Essential for coral reef ecosystems

Dinoflagellate cell of the genus Symbiodinium

Page 14: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

P. damicornis coral nubbin (5 cm height) Localization of dinoflagellates in coral tissue

Coral-dinoflagellate metabolic interactions are poorly understood, especially at (sub)-cellular levels Isotopic labeling experiments: 15NH4

+,15NO3-, 15N-aspartic acid and 13C-bicarbonate

Transmission Electron Microscopy (TEM) NanoSIMS isotopic imaging

Page 15: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

TEM 12C14N- 12C15N- 12C15N-/12C14N-

Page 16: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

15N-hotspots (white arrows) are crystalline inclusions of uric acid (EELS and GC-IRMS)

15NH4+ incubation

90 minutes

15N incorporation at the scale of individual zooxanthellae

N storage in uric acid crystals C5H4N4O3

Page 17: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

TEM ultrastructural characterization of uric acid crystals (white arrows) within dinoflagellates

Page 18: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

NanoSIMS map allows δ15N values to be assigned to each subcellular compartment separately

uric acid deposits 15N “hotspots’

Combining TEM and NanoSIMS images

Page 19: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (
Page 20: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

Nitrogen translocation from the dinoflagellates to the coral host - a nitrate experiment

Page 21: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

Direct imaging of N translocation from the symbiotic dinoflagellates to their coral host. This translocation begins ca. 6 hours after nitrogen incorporation by the algal cells (delayed transfer).

Page 22: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

Biomineralization - pearl (nacre) forming oyster:

Page 23: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (
Page 24: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (
Page 25: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (
Page 26: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (
Page 27: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

Time-of-Flight SIMS (TOF-SIMS)

Page 28: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

Bombardment with primary ions (Ga+, In+, Aun+, Bin+, Cs+, Ar+, Xe+, C60+, …) Energy 5-25 keV Secondary Ions(+/-): Depth of origin 1-3 monolayers

Page 29: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

Kinetic energy = qV = ½*v2/m

=> v = (2qV/m)1/2 = L / t Flight-time t = L * (m/2qV)1/2 L = effective length of flight path V = accelerating voltage q = charge of ion m = ion mass v = velocity 2 = two

Page 30: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

TOF-SIMS spectra are generated with very short pulses of primary ions (<1 ns). The ability to distinguish different masses of secondary ions is directly related to the width (in time) of the primary beam pulse. If the pulse is not short, the ejection time is not well-defined and ions not well separated in time at the detector.

Page 31: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

With each pulse of primary ions: a complete mass spectrum is obtained from H to molecules of thousands of daltons

TOF-SIMS gives a chemical/isotopic finger print of what is at the surface, but its hard to quantify

Page 32: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (
Page 33: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

Ekin = EK - EL1 – EL2,3 orbital energies are unique to an atom of a specific element,

Page 34: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (

Cylindrical Mirror Analyzer: Electron energies.

Page 35: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (
Page 36: An ion microprobe is a SIMS technique Secondary Ion Mass ... analysis sims and aes.pdf · of primary ions (