An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

53
# 1 # 1 An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

description

Learn about Polytec’s non-intrusive and highly productive vibration measurement technology for automated multi-point scanning of structures. The webinar will explain its application to aerospace, automotive, civil, biological, medical and micro- structures. Topics Include: - Concept and theory of 1D and 3D scanning vibrometry - Technical advantages and limitations - Productivity benefits - Example applications including FEM validation, modal analysis, acoustics, NVH troubleshooting, automated 3D data acquisition with RoboVib and strain measurement

Transcript of An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

Page 1: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 1# 1

An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration

Measurement

Page 2: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 2

Presentation Outline Deflection Shape and Experimental Modal

Analysis used in the product development cycle

Traditional test methods versus Scanning laser vibrometry

Principles of 1-D and 3-D laser vibrometry Applications examples Accessories Conclusions

Page 3: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 3

Why Can’t We Rely 100% on the Computer Model?

Material properties Tolerances in manufacturing

Complex joint dynamics Damping Boundary conditions Changes in stiffness during rotation

→ → → FEM Validation and Updating

Uncertainties in:

Impossible or Difficult to Model:

Page 4: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 408.04.2023 # 4

PSV-500-3D

Page 5: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 5

Other Examples of the Need to Perform an Experimental Modal Test

Build an experimental model Compute structural modifications

Understand the dynamics of a structure (of a competitor for example) for which we do not have a finite element model

Troubleshooting:

Benchmarking:

Page 6: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 6

Full Field - Traditional Approach Issues with Contact Transducers

Mass loading Damping and stiffness errors

Time consuming Limited working environments Complicated cabling Safety Coarse spatial resolution Coordinate correlation to FEM

geometry cumbersome Transducer calibration Limited bandwidth Poor Modal Assurance Criteria

Page 7: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 7

A Better SolutionScanning Laser Vibrometer

Laser has no mass Faster - laser can be scanned Long standoffs, through windows,

fluids, hi temps No cabling on structure Spot dia typically 10s of µm FEM geometry import for easy

point definition One (3) transducer(s) to calibrate

(no book keeping) Wide bandwidth – no accel rolloff or

resonance - to 1.2GHz!! Reduced chance of spatial aliasing

Page 8: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 8

Spatial Aliasing19 MP

-4.00E-08

-3.00E-08

-2.00E-08

-1.00E-08

0.00E+00

1.00E-08

2.00E-08

3.00E-08

0.23

4668

0.18

6155

0.13

7642

0.08

9129

0.04

06156

-0.00

7897

-0.05

641

-0.10

4923

-0.15

3436

-0.20

1949

9 MP

-4.00E-08

-3.00E-08

-2.00E-08

-1.00E-08

0.00E+00

1.00E-08

2.00E-08

3.00E-08

0.23467 0.1619 0.08913 0.01636 -0.05641 -0.12918 -0.20195

5 MP

-2.00E-08

-1.50E-08

-1.00E-08

-5.00E-09

0.00E+00

5.00E-09

1.00E-08

0.234668 0.113385 0.0163594 -0.104923 -0.201949

19 point profile

19 points

9 points

5 points

Low point density:amplitude errordistorted shape

514 Hz

Page 9: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 9

PSV Configurations

1D Scanning

Simple structuresODS

Acoustics – noise source identification

Rotating

08.04.2023 # 9

3D Scanning

Complex structuresStructural Dynamics

Modal analysis

FEM validation

Stress & strain

Robot-integrated

Polytec Scanning Vibrometer (PSV)

Page 10: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 10

Front-end Data ManagementSystem

HD Video

LDVSensor

ScanningMirrors

ScanElectronics

Stitched data

Sensor Head Test Object

Test Object

1-D Scanning Vibrometry (PSV)

Up to 250,000 locations scanned point-by-point – vibration velocity or displacement

Software for exciter signal generation, data acquisition, display & manipulation

Geometry file imported or measured

Animated data visualization

Data export for modal analysis or FEM validation

Page 11: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 11

Operational steps:Define measurement points

using video image and draw programby geometry import

Excite structureusing internal or external function generator signal

Scan to acquire vibration response at each point time history orFFT spectrum

Visualizeanimated operating deflection shapes at selected frequencies of interesttime domain animations

Export for post-processing (e.g. modal analysis or FEM validation)

PSV Operational Steps

Page 12: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 12

PSV Working Principle

Measure V and at many points relative to a reference signal Amplitude and relative phase at each scan point

Spatial vibration pattern of measurement surface

Scan signal

Ref signal

Page 13: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 13

Operating Deflection Shapes

Page 14: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 14

Wide Ranging Applications

Page 15: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 15

Typical 1-D Scanning Vibrometer

PSV-500-H 4 (or 8) channel data

acquisition with 80 kHz bandwidth

4 or 8 reference channels 50° x 40° scan angle 13 velocity ranges up to 10

m/s Resolution to <0.01 µm/s/√Hz Autofocus MIMO with multi-shaker

excitation, 4 independent generator channels; Principal Component Analysis

Ethernet connection to DMS Coherence Optimizer

Page 16: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 16

Example measurement on black loudspeaker cone

480 pm deflection1775 Hz

8 averages

New Technology - New Quality

08.04.2023 # 16

Digital Broadband Decoder

Page 17: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 17

Scan Modes

Time

Frequency

FFT

Zoom FFT

Time Domain

FastScan

Time

Frequency

Page 18: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 18

Time Domain Animations

Page 19: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 19

Acoustics

Acoustics Configuration50 kHz vibrometer bandwidth digital

7 or 10 measurement ranges

1 channel for reference sensor (digital)

VibroLink Ethernet connection

Optional signal generator

Geometry Scan Unit optional

Software for acoustic measurements – live or stored data – headphones include

Configuration Examples

08.04.2023 # 19

Page 20: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 20

Requirements for Quiet Products

Sound propagationFrom source to receiver

Transmission (structure-borne noise)

Leakage (airborne noise)

Secondary noise (structure-borne noise)

08.04.2023 # 20

PSV-500 and its Applications

F

Structural dynamics

EmissionSurface velocity

Sound pressure

Fluid-structure-coupling Courtesy : P. Zeller; Handbuch

Fahrzeugakustik

Operating Load

Page 21: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 21

Laser vibrometers measure vibration in laser direction only

Data needs careful interpretation due to different vibration directions, surface shape and varying scan angles

Modal analysis software expects 3-D data

FE models produce 3-D data

Limitations of a 1-D Scanning Vibrometer

Test ObjectLaser Vibrometer Θ

Page 22: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 22# 22

3-D Scanning Laser Vibrometry

Page 23: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 23

x

yz

3-D object coordinate system defined by means of a minimum of 3 known reference points

Laser beam unit vectors for all scan heads are determined using the transformation matrix:

Principles of 3-D Laser Vibrometry

3

2

1

1

333

222

111

v

v

v

lll

lll

lll

v

v

v

zyx

zyx

zyx

z

y

x

Page 24: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 24

If the object geometry is known, all 3 lasers intersect for all (accessible) scan points

3 simultaneously acquired vibration signals are transformed into the object‘s coordinate system by software

x

y z

Principles of 3-D Laser Vibrometry

Page 25: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 25

Data import (via UFF):

a) Other test software packages

b) Finite Element Model Geometry File

c) Extracted from other methods such as CAT scan

What if there is no geometry file?

Options for Obtaining Geometry File

Page 26: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 26

Options for Obtaining Geometry File

When there is no geometry file:

a) Alignment of all 3 laser beams manually for every scan point

b) Measure geometry for every scan point

GEOMETRY SCAN UNIT

Page 27: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 27

PSV-500 Geometry Scan Unit

More compact, more stable, no moving parts, fasterFor eg: 844 points in 3mins 30secs

Higher sensitivity and wider optical dynamic range Handles larger range of surfaces and variations in surface finish

Page 28: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 28

Beam Positioning Accuracy

Reasons for poor overlap: Inaccurate “3D-Alignment” during set up procedureImperfect geometry data

Why must beams overlap?Lasers must measure at same point for optimal lateral resolutionMore critical for strain measurements - geometry must be known for transformation to in-plane and out-of-plane data

standard

with triangulation during scan

Page 29: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 29

Modifies the 3D coordinates by overlapping the 3 beams at one point.

Result: Impoved geometry as input data L for strain calculation

Result: accurate geometry

PSV-S-TRIA Video Triangulation

][m

m

L

Page 30: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 30

PSV-S-TRIAVideo Triangulation during Scan

before triangulation

Page 31: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 31

PSV-S-TRIAVideo Triangulation during Scan

after triangulation

Page 32: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 32

Structural Dynamics (3D)

3D Structural Dynamics Configuration100 kHz vibrometer bandwidth (digital)

13 measuring ranges

8 reference channels (integrated)

MIMO with multi-shaker excitation up to 4 shakers (integrated)

Principal Component Analysis (PCA)

Geometry Scan Unit

Coherence Optimizer

VibroLink Ethernet connection

StrainProcessor optional

Configuration Examples

08.04.2023 # 32

Page 33: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 33

Deflection shape at 82 Hz showing individual measurement points

Station Wagon 3-D Vibration Data

Page 34: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 34

Faster set-up According to a European car manufacturer just the mounting and cabling of the accels and dummy

masses plus the coordinate definitions for a car body usually takes around 4 – 5 days

FEM mesh (coarsened) can be imported Manual identification of points relative to FEM nodes not necessary All points correlate exactly with FE model

Faster data acquisition No moving of dummy masses, accels and retesting

Spatially more detailed data Important for FEM correlation, model updating, and acoustics

Benefits compared to Accelerometers Insensitive to crosstalk when rotational vibration is present Calibration not affected by temperature Flat frequency response

Software is easy to learn and system easy to use

Other time savings for design (FEM) department No modeling of dummy masses necessary

A major European car manufacturer estimates total time savings of > 50%

Benefits of Scanning Vibrometry for Modal Analysis

Page 35: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 35

RoboVib

CA

E w

ork

flow

Page 36: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 37

Results Exampledeflection shape at 38 Hz

Page 37: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 38

Case Study – Transmission Case

Page 38: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 39

Element type Tetrahedral (solid 185)

Number of nodes 19994

Number of elements 69019

Material Aluminum casting alloy

Case Study – Finite Element Model

Page 39: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 40

Results

Example: deflection shape at 592 Hz

Case Study – Polytec Data

Page 40: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 41

Frequency Response Functions

Case Study – Test Data

Page 41: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 42

Circle Fitting

Resonant Frequency

Damping

Case Study – Curve Fitting

Page 42: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 43

Case Study – Polytec MAC Data

Accel Data < 0.64

Page 43: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 44

Am

plit

ude (

dB

)

Frequency (Hz)

Frequency Shift

Density (% Change) = -1.3 %Elasticity Modulus (% Change) = -5.7 %

EMA

Case Study – Model Update

Page 44: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 45

Traditional - Fixed CostsAerospace GVT Example

Large Size >400 Channels 2-8 Reference Channels (MIMO)

Medium Size 100-400 Channels 2 Reference Channels (MIMO)

Small Size <100 Channels 1 or 2 Reference Channels

Page 45: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 46

Number of Channels

Time & Complexity

$$$

Small Medium Large

Total Cost of Modal Test Add in the variable cost of personnel on a daily rate

Add in time delay, impacts on production schedule!

Page 46: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 47

Many Accessories Enable New Applications

Rotating Structures

Page 47: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 48

Bladed Disk: 10,900 rpm

725 Hz 800 Hz

845 Hz 895 Hz

Page 48: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 49

Many Accessories Enable New Applications

Small parts at high spatial resolution, high frequencies, strain

Page 49: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 50

Stress distribution across the surface of a turbine blade in the X direction

Dynamic Stress and Strain

Page 50: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 51

FEA / Experimental Correlation

Strain in x direction

Strain in y direction

Shear Strain

Results courtesy of the University of Adelaide ME DepartmentPSV data

PSV data

PSV data

Page 52: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 53

More comprehensive vibration testing in a fraction of the time and effort with enhanced detail and accuracy compared to traditional techniques easily integrated with CAE/FEM

FE Model Validations can be made using non-contact 3-D (and 1-D) laser measurement

Many accessories are available to extend the range of applications

Engineering services and rentals are available

Conclusions

Page 53: An Introduction to Scanning Laser Vibrometry for Non-Contact Vibration Measurement

# 54

Engineering Services and Rental Program

Advanced non-contact vibration and surface metrology measurements available for every budget

Measurements using Polytec’s latest, non-contact measurement technology

Skilled and experienced applications engineers to operate the measurement system to its fullest potential

Convenience of testing at the customer’s facility or in a Polytec lab Short-notice, critical measurements Scheduled, occasional measurements Build justification prior to purchase Save cost by renting instead of buying Budget flexibility, rent-to-buy