An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the...
-
Upload
duongtuyen -
Category
Documents
-
view
226 -
download
1
Transcript of An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the...
C
AnInstr
Center fo
n Introdrumen
or Micros
ductionntation,
A
scopy an
n to El, Imag
Andres K
d Image
lectronging an
Kaech
Analysis
n Micrond Prep
s, Univer
oscopyparatio
rsity of Z
y on
Zurich
Introduction to electron microscopy Andres Kaech April 2012
1
Table of content
1. Introduction ................................................................................................................................................ 2
2. Principle of SEM and TEM ........................................................................................................................ 4
The electron source .......................................................................................................................................... 5
Electromagnetic lenses .................................................................................................................................... 6
Specimen holders and stages ........................................................................................................................... 8
3. Electron specimen interactions ................................................................................................................. 10
4. Contrast formation and imaging in TEM ................................................................................................. 12
5. Contrast formation and imaging in SEM .................................................................................................. 15
Imaging using secondary electrons ................................................................................................................ 15
Imaging using backscattered electrons .......................................................................................................... 17
Elemental analysis using x-ray ...................................................................................................................... 18
Artefacts ........................................................................................................................................................ 19
6. Sample preparation ................................................................................................................................... 21
Preparation of cells and tissues for transmission electron microscopy .......................................................... 21
Preparation of cells and tissues for scanning electron microscopy ................................................................ 24
7. Literature .................................................................................................................................................. 27
Introduction to electron microscopy Andres Kaech April 2012
2
1. Introduction
Microscopy enables a “direct” imaging of organisms, tissues, cells, organelles, molecular assemblies and even individual proteins. Analytical techniques as used in molecular biology provide a data set which represents cellu-lar processes in an abstract way. Therefore, microscopy was and still is an important, complementary technique to visualize the macro and/or microscopic structure and to assign structure to function and vice versa.
Microscopy techniques, their resolution limit, scales and corresponding objects are depicted in figure 1.1. The typical size of eukaryotic cells is 10-20 micrometer and their subcellular components can not be seen by naked eyes. Additionally, cells are colorless and transparent. The organelles or other constituents have to be stained to make them visible in the light microscope. A variety of histological stains and fluorescent markers have been developed.
Imaging the ultrastructure of cellular components requires electron microscopes which have a much higher re-solving power than light microscopes. The physical properties of electron microscopes (e.g. high-vacuum) de-mand specific preparation and staining techniques to reveal the ultrastructure of cells and tissues.
Fig. 1.1: Resolution limit of different imaging techniques, involved radiation and size of biological objects.
The limit of resolution of an optical system depends on the numerical aperture and the wavelength of light (Law of Ernst Abbé). This law holds as well for electrons, of which the speed determines their wavelength: The higher the speed of the electrons, the smaller the wavelength and the better the resolution. The practical resolution of a transmission electron microscope (TEM) run at 100 kV (acceleration voltage that determines the speed) is ap-proximately 0.5 nm and exceeds the resolution of light microscopes by far. In scanning electron microscopes (SEM), the effective resolution is around 1 nm. The resolution of modern electron microscopes is much better than the resolution that a prepared biological specimen can provide due to preparation.
The most important and challenging part of electron microscopy of biological material is the preparation proce-dure. With every preparation step artifacts are introduced. Electron microscopists have put a lot of effort in the preparation of the specimens to reduce the artifacts to a minimum and to image the specimens in a more lifelike state. The image of the specimen can never be better than the entity of the introduced artifacts during prepara-tion.
In combination with immunocytochemical methods, electron microscopy is powerful method to label and high-light single, specific proteins enabling a correlation of the ultra structure and its function.
Note: Preparation of biological specimens for electron microscopy is often a very time-consuming process.
Atoms
1 mm
100 m
10 m
1 m
100 nm
10 nm
1 nm
0.1 nm
Cells
Red blood cells
Bacteria
Mycoplasma
Viruses
Proteins
Amino acids
Radio
Infrared
Visible
Ultraviolet
x,-rays
Human eye
Electron microscope
Light microscope
Resolution Limit Wavelength/Size Object
MRI, CT
Introdu
Fig. 1transming moed imaview. Iand fre
uction to electron
.2: Overview ofmission electronodes in transmisage of the sampImages show Heeze-fractured H
Transmiss
microscopy
of the different n microscope onssion and scannple in TEM. A b
Hela cells, thin sHela cell image
sion electron
types of electron the left and aning electron mbulk sample is section of a fixeed in the frozen
n microscopy
Andres
on microscopesa scanning electmicroscopy. A th
scanned with aed, dehydrated,n state in a cryo
y
Kaech
s and their gentron microscophin sample is tra focused beam plastic embedd
o SEM on the rig
Scann
neral mode of ie on the right. Sransmitted by th of electrons inded cell cultureght.
ning electron
imaging. Top rSecond row: Ache electrons to fn an SEM prove (70 nm thickn
n microscopy
April 2012
3
row: Biologicalccording imag-form a project-iding a surfaceess) on the left,
y
2
3
l --e ,
Introduction to e
4
2. Prin
The overall dthe light is lenses. The dfigure 2.1.
Fig. 2.1: Similthe tip of a heand are guidethe specimen aand recorded.
The design opicted in figu
Fig 2.2: Similaat the tip of a scanning elect
electron microsco
nciple of SE
design of an esubstituted wdesign of a tra
larity of a transeated filament. ed through the eand leads to an
of a scanning ure 2.2.
arity of a scannheated filamen
tron microscop
opy
EM and TEM
electron microwith electrons
ansmission el
smission electrThe electrons aelectron micros
n image. The im
electron micr
ning electron mnt. The electronpe column by e
M
oscope is similand the glas
lectron micros
on microscope are acceleratedscope column b
mage is monitore
roscope and it
icroscope with ns are acceleratlectromagnetic
Andres Kaech
lar to that of ass lenses are scope and its
with a wide fied with high voltby electromagnred on a phosph
ts similarity t
a confocal laseted with voltag
c lenses. The be
a light microssubstituted wsimilarity to
eld light microstages (60 – 120etic lenses. The
horescent screen
to a confocal
er scanning micges between 0.2 eam is focused
cope. In the ewith electroma light micros
scope. An electr00 kV dependinge beam penetran or specially d
laser scanning
croscope. An ele– 30 kV and ain the “object
Ap
electron micromagnetic/electr
scope is depic
ron beam is forg on the type of
ates and interacdesigned CCD
ng microscope
ectron beam is fare guided throutive” lens and a
pril 2012
oscope, rostatic cted in
rmed at of TEM) cts with camera
e is de-
formed ugh the a small
Introdu
beam sted eledetecteimagin
Electris necevacucan flresiduum syTEM serial
Fig 2.3
The e
Electrof thea tungthrouggy andbetwewhichtremevery hresolu
uction to electron
spot is moved/sectrons are deted in accordingng device such a
ron microscopcessary to preuated steadily ly before hittinual gas molecuystem of an eis given in fisetup of low
3: Example of a
lectron sourc
rons can eithee electron sourgsten filamengh the electrond, therefore, n
een the electroh the elctrons ely fine tungstehigh yield of ution. These in
microscopy
scanned over thtected in specifg detectors. Eveas a CCD came
pes are high-vvent oxidationto a vacuum
ng another parule in the systlectron microigure 2.3. Higand high vacu
a vacuum system
ce
er be producedrce and its pront, a LaB6 crn source enab
need to be accon source (caare guided anen tip withoutelectrons and
nstruments are
he specimen surffic detectors anery scanned pixera in SEM.
acuum systemn/burning of tof 10-5 to 10-7
rticle, accounttem reduces thscope consist
gh vacuum puuum pumps is
m for a TEM. R
d by thermionoperties are deystal or a Zr
bling the escapelerated to the
athode) and and acceleratedt heating (room
d the very lowe very costly a
Andres
rface using deflnd amplified u
xel on the specim
ms. In the areathe heated fil7 mbar. The mts to about 50he resolution ats of a cascadumps can not
required.
RP…rotary pum
nic emission oepicted in figurO/W Schottkpe of electrone desired speean anode plated. During coldm temperature
w chromatic aband require pa
Kaech
lector coils. Theusing photomultmen represents
a of the electroament. The cmean free path0 meter at a vaand performan
de of low and directly pump
mp, TMP…turbo
r in a process ure 2.4. Durinky emitter is hs. The electro
ed before entere is applied ld field emissioe). The advantberration of tharticularly hig
e beam interacttiplier tubes. Ea pixel on the s
on source a vaolumn and thh, referred to
acuum of 10-6 nce by scatterhigh-vacuum
p against amb
o molecular pum
called cold fig thermionic heated by an ns leaving thering the electrleading to an on, the electrotage of cold fihe electrons ah vacuum.
ts with the specEmitted radiatioscreen. Importa
acuum of 10-7
he area of the as the distancmbar. Any co
ring of electrom pumps. An ebient pressure
mp, IGP…ion g
field emission.emission, a v
n electrical cue filament havron column. Aelectrostatic
ons can escapfield emission allowing imag
April 2012
5
cimen and emit-on as x-rays isant: There is no
to 10-10 mbarspecimen are
ce a moleculeollision with aons. The vacu-example for a. Therefore, a
getter pump
. The materialery fine tip of
urrent flowingve a low ener-A high voltage
field throughe from an ex-sources is the
ging at atomic
2
5
-s o
r e e a -a a
l f g -e h -e c
Introduction to e
6
Fig. 2.4: Diffe
Electromagn
Electromagnpiece (figurethe pole pieca charge pastion of the mthrough the l
Depending oan electron mness of the bform the primtive lens. Thlenses whichphotosensitivtion of the imrected in mod
Material
Heating tem
Normalized
Required vac
∆E (eV)
electron microsco
erent types of el
netic lenses
netic lenses coe 2.5 A). A mce of the lens. sing a magnet
magnetic fieldlens system (fi
on the strengthmicroscope acbeam and guidmary image. Fhe final magnh act in combve film), see amage dependindern TEM by
p. (K)
brightness
cuum (Pa)
opy
lectron sources
onsist of a hugmagnetic field i
The accelerattic field. The
d and the direfigure 2.5 B).
h of the magncts as a condedes it to the spFocusing of thification is deination and p
also figure 2.1ng on the maga set of corre
Tungsten
W
2700
Chrom
high
and their prope
ge bundle of wis induced byted electrons eresultant forc
ection of the
netic field, thenser lens, whipecimen. The
he image is peretermined by
project the fin and 2.2 for le
gnification (thctor coils.
Thermionic
LaB6
LaB6
1800
matic aberration!
Andres Kaech
erties. The imag
windings of iny the current inentering the mce is always pelectrons. In
e focal width oich bundles th
e electrons tranrformed with the following
nal image on tens set up. Th
he lens curren
c
Schottky
ZrO/W
1800
ge shows a Sch
nsulated coppn the winding
magnetic field erpendicular tconclusion, t
of the lens is he electrons, dnsmit the spethe strength o
g projective lthe fluorescenhe circular patt of the projec
Cold field emi
W
300
Ultra high
ottky emitter.
per wire, a sofg and reaches are deviated fto the plane dhe electrons t
changed. Thedetermines thecimen and theof the magnetienses, a systent screen or ath of the electrctive lenses).
ission
Ap
ft iron cast anits main stren
following the defined by thetake a circula
e first lens syse the overall be objective leic field in the ems of two ora CCD cameratrons leads to This rotation
2.5 cm
pril 2012
nd pole ngth at law of
e direc-ar path
stem in bright-ns and objec-r three a (or a a rota-is cor-
Introdu
Fig. 2.changepassinThe bl
Electrtion aent elconfulensesregulacorrec
A
uction to electron
.5: Electromagned by the curre
ng the magneticlack arrows dem
romagnetic lenand astigmatismectrons derivi
usion of the ims and aperturearly, in extremctor coils.
microscopy
netic lens. A) Tent trhough thec field are deviamonstrate the ro
nses show them. The most ping from a po
mage (figure 2es and chargime cases after
The magnetic fiee windings. Theated perpendicuotation of the im
e same aberratprominent abe
oint source do2.6). Astigmating of the sper every move
Andres
eld is strongeste lens is water ular to the planmage.
tions as glass erration in ele
o not match onatism is causedecimen. In pa
ement of the s
B
Kaech
t in the area of cooled to main
ne defined by th
lenses such aectron microscne and the samd by inhomogarticular in SEspecimen. The
B
the pole piece. tain stability d
he magnetic fiel
s chromatic, scopy is the axime point in thgeneities of thEM, the astige correction i
The focal lengduring operationld B and the ve
spherical aberial astigmatismhe image plainhe lenses, congmatism mustis done again
April 2012
7
th of the lens isn. B) Electrons
elocity vector v.
rration, distor-m. The differ-n leading to a
ntamination oft be correctedwith a set of
2
7
s s .
--a f d f
Introduction to e
8
Fig. 2.6: a) Axthe image planmatism using c
Specimen ho
In TEM, the very thin so tcal specimenhigher the voper grids of 3sections on tsince the sys
a)
b)
electron microsco
xial astigmatismne (focus). Undcorrector coils.
olders and sta
electron coluthat the electr
n should be aroltage, the thi3 mm diametetop are attachestem always st
opy
m in an electronderfocus and ov
ages
umn does not orons can penetround 70 nm fcker specimener, which are aed in a holdertays under hig
n optical systemverfocus images
offer a lot of trate the specimfor a TEM wins can be examavailable in a r and introducgh vacuum. Th
Andres Kaech
m. In focus, a pes appear as ell
space for the men and formith an acceleraamined). Thin wide variety o
ced into the gohe goniometer
oint source applipses orthogon
specimen. Adm an image. Thation voltage sections of th
of materials aoniometer of tr is the mecha
pears as a circlal to each othe
dditionally, thhe average thifor the electro
he specimen and mesh sizesthe TEM throanical setup w
Ap
le of least confuer. b) Corrected
he specimen mickness of a bons of ~100 kare mounted os. The grids w
ough a vacuumwhich enables
pril 2012
usion in d astig-
must be iologi-
kV (the on cop-with the m lock, highly
Introdu
precisimage
Fig. 2.goniom
The sspecim(bulk holderallowby han
Fig. 2.
Electron aired frowith e
uction to electron
se and stable e, in particular
.7: Thin sectionmeter of the TEM
specimen chammen can be atspecimens). Tr able to carrying controllednd requiring v
.8: The inside v
ron microscopr cushions whiom the rest of electromagnet
microscopy
control of ther at high magn
ns of a specimeEM through a va
mber of SEMt least 10 cm The specimeny 8 stubs for d movement inventing of the
view of an SEM
pes are very seich absorbe vithe building ttic coils (induc
e specimen honifications (fig
en on a TEM gracuum lock.
Ms are large anin diameter a
n usually attacexample. Then x, y, z, tiltinwhole chamb
M chamber as see
ensitive to vibibration. Ofteno minimize dictors) mounte
Andres
older during imgure 2.7).
rid, holder tip a
and offer mucand 2 cm thickched on a sme stub holder ng and rotatio
ber or by using
en with an infra
bration and elen, they are evisturbance by
ed to the the w
Kaech
maging. Any
and complete sp
ch more spacek, since it doe
mall aluminiumthen is mount
on. The stub hg a vacuum lo
ared camera mo
ectromagneticen setup on spvibration. Ele
walls, ceiling a
drift or instab
pecimen holder
e for the speces not have tom stub (1 cm ted on a 5 axiholder can be ck.
ounted on the d
c fields. The epecial fundamectromagneticand floor.
bility results i
r, which is intro
cimen. In routo be permeabldiameter) is mis stage of theinserted in th
door of the instr
electron columments which arc fields can be
April 2012
9
in an unsharp
oduced into the
tine SEM thele to electronsmounted on ae microscope,he microscope
rument.
mn is mountedre disconnect-
e compensated
2
9
p
e
e s a , e
d -d
Introduction to electron microscopy Andres Kaech April 2012
10
3. Electron specimen interactions
The electron beam (TEM and SEM) hits the sample and either passes the sample unaffected or interacts with it. Three different interactions of electrons with an atom can be generally discerned (Fig 3.1):
i) An electron of the primary beam is scattered by the electrostatic interaction with the positively charged nucleus of an atom at an angle of more than 90°, yielding backscattered electrons. This type of electrons has practically the same energy as the ones of the primary beam.
ii) An electron of the primary beam is scattered by the electrostatic interaction with the positively charged nucleus of an atom at an angle of less than 90°, yielding elastically scatterd electrons. Al-so these electrons do not loose energy and therefore are referred to as elastically scattered electrons.
iii) Electrons can also loose energy while interacting with the “electron cloud” of the atom. These are called inelastically scattered electrons. This interaction can lead to the following processes (see also http://www.matter.org.uk/tem/default.htm):
Inner-shell ionisation An electron is pushed out of the electron cloud. The electron „hole“ is filled by an electron of an outer shell: Surplus energy is either emitted as characteristic x-ray or transferred to another elec-tron, which is emitted (Auger electron).
Bremsstrahlung (continuum x-rays) Decceleration of electrons in the Coulomb field of the nucleus ⇒ Emission of x-ray carrying the surplus energy ΔE ⇒ Uncharacteristic x-rays
Secondary electrons (SE) Loosely bound electrons (e.g., in the conduction band) can easily be ejected ⇒ low energy (< 50 eV)
Phonons Lattice vibrations (heat) ⇒ beam damage
Plasmons Oscillations of loosely bound electrons in metals
Cathode luminescence
Fig. 3.1: Interaction of electrons with an atom. Note: Inelastically scattered electrons lead to a variety of processes like heating, radiation and electron emission as described in the text.
Inelastic(low angle, E=E0-∆E)
Unscattered(E=E0)
Primary electrons (E0)Backscattered electrons (E=E0)
Elastic(higher angle, E=E0)
Introduction to electron microscopy Andres Kaech April 2012
11
Almost all types of electron interactions can be used to retrieve information about the specimen. Depending on the kind of radiation or emitted electrons which are used for detection, different properties of the specimen such as topography, elemental composition can be deduced.
The follwing figure 3.2 shows the interaction of the electron beam with the specimen. In biological TEM analy-sis, the elastically scattered electrons are mainly used for imaging. The inelastically scattered electrons will be recorded as well with the imaging device (e.g. CCD camera). They have a slightly lower energy (a longer wave-length) than the primary electrons. In conclusion, they are deviated less strongly in the electron lenses and do not merge at the same spot as the elastically scattered electrons (chromatic aberration). They contribute to unsharp-ness of the image. In high-end instruments, the inelastically scattered electrons can be filtered out with an energy filter.
In SEM, mainly secondary electrons are used for imaging of biological specimens. These electrons have a very low energy (around 50 eV) compared to the energy of the primary electrons (up to 30 keV). Due to the low ener-gy, these electrons can escape only from the surface area of the specimen (for details see chapter 5) and therefore provide information about the surface topography. Also backscattered electrons are frequently used for imaging with an according backscatter detector. The number of electrons which are backscattered from a certain spot of the specimen depends on the local elemental density of the specimen. Hence, backscattered electrons provide a “density image” and information about the elemental composition, respectively. Backscattered electrons are also more resistant to charging due to their high energy.
X-rays are always produced when matter interacts with an electron beam. The energy of a part of the emitted x-rays is specific for different atoms and can be used for elemental analysis in TEM and SEM using special detec-tors. However, the analysis of light elements is rather difficult and the high corrugation of biological specimen surfaces makes the interpretation of elemental maps very demanding. A lot of experience is necessary to prevent pit falls.
Fig 3.2: Interaction of electrons with specimen/matter, induced radiation and emission. Transmitted electrons like elastically scattered electrons are typically used for TEM imaging of biological specimens. Secondary electrons and backscattered electrons are mainly used for imaging in the SEM.
PrimaryPrimary electronselectrons
UnscatteredUnscattered electronselectrons
ElasticallyElastically scatteredscattered electronselectrons
Inelastically scattered electrons
SecondarySecondary electronselectronsBackscatteredBackscattered electronselectrons
AugerAuger electronselectronsHeat
Cathode luminescenseX-rays
Specimen Interaction volume
SEM analysis
TEM analysis
Introduction to e
12
4. Con
The image fotrons are verying on the ph
The primary tromagnetic forms the priconsist mainhence, provid
Also the absAbsorbtion oarea of the sp
In conclusionby impregnaand lead ionatomic numbelectrons. Thjective apertumore heavy aperture. Thedense” area trast (more o
Fig. 4.1: Conting on the conthe objective awill yield low intensity (brig
electron microsco
ntrast form
ormation in Ty similar to th
hysical phenom
electron beamwaves. The oimary image (
nly of very ligde almost no c
sorbtion of eleof electrons wpecimen is exp
n, the specimeation of the ths bind specifiber and densihe main contrure located inmetals will leese electrons in the image. or less electron
trast formation nstituents, e.g. aperture and do
signal intensityght spots).
opy
mation and i
TEM is basicalhe ones of lighmenom to be e
m passes throobjective lens (see also lectu
ght elements licontrast.
ectrons by thewould result inposed to high
en (thin sectiohin section wiically to diffeity of the hearast formationn the back-focaead to a highdo not reach Therefore, th
ns will reach t
in biological obmembranes (ph
o not reach the y (electron den
imaging in
lly the same aht. As for lightexplained (wa
ough the thin collates the d
ure Bio 416 Like C, O, H, N
e specimen isn heat damageradiation, bea
on) must be trith heavy metrent constituevy metal ionsis achieved b
al plane of theer number ofthe imaging
he image formthe CCD, scre
bjects: A differehospholipids), rimaging device
nse, dark spots)
Andres Kaech
TEM
as for regulart, electrons caave particle du
specimen anddiffracted and
Light MicroscoN, S, P and le
s very small ae of the speciam damage is
reated to provtals like uranients of biologs leads to a mby trapping the objective lenf diffracted/scdevice (CCD
mation of bioleen), figure 4.
ent number of hribosomes, ande (CCD, viewing) whereas spots
light microscan be observeduality).
d generates nod non-diffracteopy by Urs Ziead to poor in
and does not imen. At highan issue and c
vide enough cium and lead.ical matter an
much higher ihe numerous dns (figure 4.1
cattered electr, screen) and
logical specim2.
heavy metal iond chromatin. Mag screen). Specs with no or few
copy, since thed as particles o
on-diffracted ed waves in thiegler). Biolognteraction wit
contribute to h magnificatiocan easily be o
ontrast for imThe positive
nd to a differeinteraction andiffracted elec). Spots of theons, which wwill cause a
mens is based
ns stick to the saany diffracted/s
cimen spots withw heavy metal
Ap
e properties oor as waves d
and diffractedthe image plangical objects uth the electron
contrast formons at which aobserved.
maging. This iely charged urent extent. Thnd diffraction ctrons using te sample cont
will hit the obj“dark” or “elon amplitud
ample surface d/scattered electrh a lot of heavyions yield high
pril 2012
of elec-epend-
d elec-ne and usually ns and,
mation. a small
is done ranium he high
of the the ob-taining jective lectron
de con-
depend-rons hit y metals h signal
Introdu
The d
Fig. 4.
Contrnel rinToo mcorres
Fig. 4.focus.
In parheavydispla
uction to electron
difference of h
.2 Non-stained
rast of biologings (deletion/much underfocsponds to min
.3: Under-focusToo much over
rticular, thin sy metal solutioayed in figure
microscopy
heavy metal st
versurs heavy m
ical specimens/amplification cus or overfoc
nimum contras
s, over-focus anr- or underfocus
sections of frons. The only 4.4.
ained and non
metal stained th
s can be furthof signal), w
cus introducesst.
nd true focus ofs provides an u
ozen soft, hydway to genera
Andres
n-stained thin
hin section of an
her improved bwhich improve
s artifacts and
of a thin sectionunsharp image w
drated organiate enough co
Kaech
section of an
n alga.
by slight unde the visibility
d the image ge
n of a cell. Minwith artefacts.
c matter can ontrast for ima
alga is shown
erfocusing. Unof biological
ets unsharp. T
imum contrast
not be stainedaging is using
n in figure 4.2.
nderfocusing l structures (F
True focus (Ga
corresponds to
d by incubatiunderfocus. A
April 2012
13
.
leads to Fres-Fig 4.3). Note:aussian focus)
o true Gaussian
ng them withAn example is
1 µm
2
3
-: )
n
h s
Introduction to e
14
Fig. 4.4: Frozat < – 150° C.
The primary final image cscreen consisusually perfotaken by expal film develcameras are Standard CCtions for bioarea of intere
The assemblywith electronwith a scintilthe scinillatocially availab
Fig. 4.5: Desig
electron microsco
zen hydrated thi. Contrast is tre
image is magcan be viewests of a phospormed in the dposing electronlopment in a d
available whCD cameras fo
logical imaginest and acquis
y of a CCD cans. The chip wllator. On top or to the pixelble recently (v
gn of a CCD ca
opy
in section of a Gemendously enh
gnified or demed by eye on phorescent laydark to receivn sensitive filmdark room. Nohich reach theor TEM have ng. Pixel binnition of image
amera for TEMwould be damof every pixe
l of the CCD.very expensive
amera for electr
Golden deliciouhanced by under
magnified by ta large screen
yer which starte a good contms to the elecowadays, spece resolution oa number of 2ning (e.g. 2, 4es can easily b
M is explainedmaged. Therefoel of the CCD . Cameras acqe).
ron microscopy
Andres Kaech
us apple leaf asrfocusing.
the projectiveen in the micrts to emit lightrast on the victron beam andcialized CCDof a conventi2k x 2k to 4k 4, 8 times) enbe performed w
d in figure 4.5ore, the electrcamera, a sm
quiring image
y
s imaged in a cr
e lenses arrangroscope (apprht when electrewing screen.d the negative cameras are ional chemicax 4k pixels, b
nables fast reawith the CCD
5. The CCD chrons have to b
mall fiberoptic es directly wit
500
ryo transmission
ged below theoximately 20 ons hit it. Ele. For many dees were develoused for imag
al film (very being sufficienal-time viewin
D camera on co
hip cannot be be primarily co
element guidth electrons h
0 nm
Ap
on electron micr
e objective len0 cm diameterectron microscecades, imageoped by convege acquisitionexpensive th
ent for most apng. Searching omputer scree
directly bombonverted to p
des the photonhave become c
pril 2012
roscope
ns. The r). The copy is es were ention-
n. CCD hough). pplica-of the
ens.
barded hotons
ns from comer-
Introduction to electron microscopy Andres Kaech April 2012
15
5. Contrast formation and imaging in SEM
In SEM, an electron beam with a tiny spot is guided pixel by pixel over the surface of a specimen, e.g. 1024 x 768 pixels. At every pixel, the beam stays for a defined time and generates a signal (e.g. secondary electrons) which are detected, amplified and displayed on a computer screen. The scanning of the beam is synchronized with the display of the computer screen. The signal deriving from every single pixel on the sample is simultane-ously displayed on the corresponding pixel of the computer screen and finally forms the image. Note: There is no CCD camera for image acquisition in SEM! The image consists of displayed grey levels, e.g. 256 gray levels in an 8 bit image. The magnification is changed by scanning a smaller area with the same number of pixels. The pixel size therefore gets smaller and the resolution is increased.
The beam of electrons wich hits a spot of the surface of the specimen interacts with a whole volume of the spec-imen. The interaction volume looks like a pear (figure 5.1). Secondary electrons are produced in the whole vo-lume but can escape only from a small volume close to the surface of the specimen due to their low energy (around 50 eV). Backscattered electrons manage to escape from a larger depth since their energy is the same as the primary electrons. On their way through the sample, backscattered electrons can again interact with atoms and produce secondary electrons (called SE 2 electrons). X-rays can escape from the whole interaction volume.
The size of the interaction volume is dependent on the acceleration voltage of the primary electrons and the atomic number of the atoms close to the surface. Heavy atoms decelerate the beam more than light elements and therefore reduce the interaction volume.
Fig. 5.1: Interaction volume of the electron beam with a bulk sample. Escape depths of SE, BSE, and x-rays are indicated. Note: The SE escape depth is independent on the energy of the primary electrons (PE).
Imaging using secondary electrons
Secondary electrons (SE) are mainly used in SEM for imaging the surface topography of biological specimens. They are detected with an Everhart Thornley detector. The principle of detection is shown in figure 5.2. The SE are collected by a collector grid. A voltage of + 200 to 500 V is applied to the collector grid which attracts the low energy electrons. The SE then hit a scintillator which converts the electrons to photons. The photons are guided by a light conducting tube on the photomultiplier tube (figure 5.3), where the photons again are converted to electrons that are amplified finally leading to an electrical signal. The current is translated into a gray value and displayed on the screen of the monitor.
…interaction volume RPE
SE escape depth (E < 50 eV)
BSE escape depth(E = E0)
X-ray escape depth
“high voltage” “low voltage”
Introduction to e
16
Fig 5.2: Princ
Fig. 5.3: Work
The SE escaptrons. Howevaction volumuppermost lavoltages is hitrate deeply i(atomic numbnum. Since bSE escaping volume. The surface, biolothe signal is llower areas cfigure 5.4 an
electron microsco
ciple of SE detec
king principle of
pe depth (λ) aver, the percen
me R is larger aayer. In concluigher and provinto the specim
mber) of the spebiological specthe specimen result is a fain
ogical specimlocalized to thcan not escape
nd figure 5.5.
Primary ele
SE
opy
ction with an E
of a photomultip
as depicted in fntage of the Sat lower accelusion, the numvides a strongmen and the pecimen. It is 1cimen consist is very low. Mnt, unsharp sig
mens need to behe surface, mae through the h
+200
Electron
ectrons
verhardt Thorn
plier tube. The a
figure 5.1. is iE produced inleration voltag
mber of the emger signal. Theproduced SE c10 – 100 nm fo
of light elemeMoreover, thegnal with verye coated with any more SE aheavy metal la
+7-1
0-500V – Co
ns
Andres Kaech
nley detector. Ex
amplification is
independent on this area comges because m
mitted secondae higher the accannot escape for carbon, 2 –ents only, the
e SE escaping y poor contrasa thin layer o
are produced iayer and obsc
2kV HV
ollector volta
Photons
Explanations see
s monitored by
on the accelerampared to the
more electrons ary electrons fcceleration voanymore. λ is
– 3 nm for Chrinteraction vothe specimen
st. To receive f a heavy metin the surface cure the signal
Phot
age
s E
e text.
the voltage of t
ation voltage oSE produced interact with
from the SE esltage, the mor
s also dependeromium, and 1olume is largederive from aa sharp and clal (few nm of layer, and SE
l. This situatio
omultiplier
Electrons
Ap
the photomultip
of the primaryin the whole ithe specimen scape depth atre electrons peent on the den1 – 2 nm for Pe and the numba relatively larlear signal fro
f Au, Pt etc.). NE deriving fromon is explained
pril 2012
plier.
y elec-inter-in the
t lower ene-sity
Plati-ber of rge om the Now,
m d in
Introdu
Fig. 5.from threpresproper
Fig. 5.surfacmembring a sare ice
Imagi
BacksferencNo covoltagBSE d
uction to electron
.4: The coatinghe spot of the p
sents the contrar acceleration v
.5: The images e was not coatrane are recognsharp signal ate contamination
ing using bac
scattered electces in the speollector grid isges than 50 eVdetector is loc
SE
microscopy
g of the specimeprimary electroast. The contrasvoltage is select
represent a vieted and therefornizable. Right. t a high signal ns).
ckscattered el
trons (BSE) acimen compos necessary, sV of the SE). cated just bene
E
R
PE
Φ
en with heavy mn beam (PE) isst is a functionted.
ew on the inner re appears noisThe surface wato noise ratio.
lectrons
are also used tosition. For desince the BSEThe BSE pref
eath the object
PE
SE
Andres
metals localizess high. The diffen of the angle b
r surface layer osy, unsharp an
as coated with 4Detail structur
to image the setection of BS
E have the samferably are dirtive lens (figu
Kaech
s the signal to tferent number ofbetween primar
of the plasma md detail structu4 nm of platinumres like hexagon
surface of the SE, a differentme energy as trected back to
ure 5.6).
P
the surface. Thef SE escaping try beam and th
membrane of freures are obscurm localizing thenal protein pat
specimens but arrangementthe primary elowards the obj
SE esca
PE PSE SR Eλ S
PE
e yield of SE dethe specimen athe specimen su
eeze-fractured yred. Only invage signal to the stterns are visibl
ut also for dett of the detectlectron beam
bjective lens. T
ape depth λ
Primary electSecondary elExcited volumSE escape de
April 2012
17
eriving directlyt different spotsrface when the
yeast. Left: Theginations of thesurface provid-le (bright spots
tection of dif-tor is applied.(much higher
Therefore, the
tronslectronsmeepth
2
7
y s e
e e -s
-. r e
Introduction to electron microscopy Andres Kaech April 2012
18
Fig. 5.6: Arrangement and working principle of a BSE detector. The scintillator is arranged circular around the primary beam just beneath the objective lens. BSE hit the scintillator and produce photons, which are guided to the photomultiplier tube. As for the SE detector, the signal is amplified, converted to a gray level and dispayed on the computer screen.
Figure 5.7 is an example, how the BSE detector can be used to get additional information of the specimen, e.g. detection of gold particles on the surface of red blood cells.
Fig. 5.7. Red blood cells labelled against surface proteins. The antibody is connected to a gold particle (heavy metal aggre-gation). Whereas the SE signal provides the surface topography, the BSE image distinctly shows the gold particles. The images were recorded simultaneously at 25 kV acceleration voltage. The whole sample is coated with a thin layer of heavy metal (ca. 2 nm). SE derive mainly from this layer and therefore do not clearly accentuate the gold particles, which have a size of around 20 nm. On the contrary, a lot of elctrons are backscattered where the gold particles are, but only very few are backscattered from the thin metal coat.
Elemental analysis using x-ray
X-rays produced in the specimen can be used for elemental analysis of the specimen (Fig 5.8). The x-ray energy is specific for different elements. A dedicated x-ray detector detects the x-rays deriving from every spot of the sample and assigns the according element. Elemental mapping is possible for any area of interest. In particular the interpretation of x-ray spectra is very challenging and needs a lot of experience. A highly corrugated surface of a specimen – typical biological specimen - can lead to misinterpretation. Quantitative analysis with biological specimens is usually not possible.
Scintillator Light pipe Photomultiplier
Primary electrons
BSE
SE BSE
Introdu
Fig 5.8
Artefa
Imagimen, ing arlocate
Fig 5.lotus l
Beamdestro
uction to electron
8: Elemental an
acts
ing in the scancharge is intrortefact is showes the signal to
9: Charging arleaf (right)
m damage is anoy the surface
microscopy
nalysis using x-
nning electronoduced into thwn in figure 5o the surface b
rtefacts on the
nother artefactin many diffe
-rays. The eleme
n microscope he material an5.9. Note: Thbut also adds a
surface of a fr
t which often erent ways, fig
Andres
ental spectrum
can lead to a nd must be dishe coating of ta conducting l
freeze-fractured
is encounteregure 5.10.
Kaech
derives from th
lot of artefacssipated in ordthe specimen layer on non-c
d plasma memb
ed. The electro
he red region in
cts. When the der to prevent
with a thin hconducting sp
rane of yeast (
on beam can l
n the image.
beam penetrat charging. A theavy metal lpecimens.
(left) and a wax
locally heat th
April 2012
19
ates the speci-typical charg-ayer not only
xy surface of a
he sample and
2
9
--y
a
d
Introduction to e
20
Fig. 5.10: Bea
electron microsco
am damage on t
opy
the surface of a freeze-fracture
Andres Kaech
ed, frozen-hydraated yeast surfa
ace after one sc
Ap
can. Scale bar 5
pril 2012
500 nm.
Introduction to electron microscopy Andres Kaech April 2012
21
6. Sample preparation
As we have discussed in the earlier sections, electron microscopes are run under high vacuum conditions. A high vacuum, however, is a very unfriendly environment for biological specimens like cells and tissues (or condensed organic hydrated matter in general). Electron microscopes pose the following prerequisites to the specimen for investigation.
Resistant in the vacuum
Electron beam resistant
Providing contrast
Penetrable for electrons (transmission electron microscope)
Large biological specimens (cells, tissues) do not fulfill any of these requirements and have to be treated in order to be investigated in the electron microscope. Typically, the specimen is turned into a solid state to make it suita-ble for vacuum and resistant in the electron beam. Moreover, it has to be cut to very thin sections of ca. 100 nm for transmission electron microscopy. Such a specimen still does not provide contrast. Biological material is basically composed of light elements like C, O, H, N, S, P, which do not interact with electrons. Selective stain-ing or contrast enhancement with heavy metals is required to form an image as already pointed out in chapters 4 and 5 (contrast formation and imaging in TEM and SEM).
Preparation of cells and tissues for transmission electron microscopy
Typical preparation procedures for cells or tissues for transmission electron microscopy include the steps depict-ed in figure 6.1.
Figure 6.1: Left: Classical preparation pathway for transmission electron microscopy. The whole process is performed at room temperature. Right: Cryo-preparation pathway for transmission electron microscopy. The specimen is frozen and slowly warmed up during freeze-substitution (low temperature dehydration).
The classical preparation is performed at room temperature. During fixation, the native biological specimen is chemically stabilized with chemical fixatives like glutaraldehyde and osmium tetroxide. All biological processes are arrested during fixation. The fixed specimen is ready for dehydration with a sequence of different ethanol concentrations until it is completely dehydrated. Subsequently, the ethanol is exchanged with a monomer solu-tion of a plastic formulation (e.g. Epon/Araldite) in a sequence of rising concentrations of plastic components dissolved in ethanol or acetone. The monomers are polymerized by heat or UV light and provide a hard plastic block containing the embedded specimen. This enables cutting of thin sections (100 nm) in an ultramicrotome. The sections are stained with uranyl-acetate and lead citrate. Uranium and lead ions selectively bind to lipids, proteins and nucleic acid in the specimen and provide a distribution of electron dense material and finally the image in the electron microscope (Figure 6.2).
Embedding
Fixation
Dehydration
Staining (uranyl-acetate, lead-citrate)
Thin sectioning
TEM
Classical preparation(chemical fixation at RT)
Cryo-preparation(cryo-fixation)
Introduction to e
22
Figure 6.2: TE
A lot of artifto minutes) apreparation pdehydration a
Cryo preparastep, the specing is very fcause of the ultra structurproperties of 6.3) is achiev
Once the speperatures (e.gplastic, sectio
electron microsco
EM micrograph
facts are introdand the specimprocedure is cand embeddin
ation methodscimen is genefast and can bnature of wat
re. High-pressf the water at 2ved only for b
ecimen is frozg., -90°C, graoning, staining
opy
h of a thin sectio
duced during men has enougconducted at rng.
s were introdurally frozen w
be performed ter to form icesure freezing w2100 bar just b
biological spec
zen, it is dehydadient to roomg of the specim
on of classically
classical TEMgh time to rearoom tempera
uced and devewithout chemic
within millisee crystals duriwas developebefore freezincimens with a
drated and fixm temperaturemen is typical
Andres Kaech
y prepared cere
M preparationact to the chemature leading t
eloped in ordecal pretreatmeeconds. Howeing freezing. Ied to prevent ing. Despite thi
thickness of l
xed with cheme) in a processlly performed
ebellar slice cu
n. The fixationmical attack anto extraction
er to prevent oent (pure physever, the freezIce crystal segice crystal foris technique, sless than 200
mical fixativess called freezethe same way
lture of mouse.
n with chemicand undergoes and shrinkage
or reduce thessical process) zing is a chalgregations distrmation by chsatisfying freeµm.
s simultaneouse-substitutiony as during cla
Ap
Scale bar 1 µm
als is slow (sechanges. The e of material
se artifacts. In– cryo-fixed.
llenging procetort and dama
hanging the phezing quality (
sly at very lown. The embeddassical prepara
pril 2012
m.
econds whole during
n a first Freez-
ess be-age the hysical (Figure
w tem-ding in ation.
Introdu
FigureScale b
Negat
Singletissuetaininafter acarbonions) of secmicroraphyding c
Figureformat
uction to electron
e 6.3: TEM micbar 1 µm.
tive staining
e particles likees by a procesng the particlea short incuban layer of theis added on thconds or minuoscope. The hey and provide can be omitted
e 6.4: TEM miction is shown in
microscopy
crograph of a th
e viruses, protss called negaes are pipettedation time (sece TEM grid. She grid. Againutes and the geavy metal iona “negative c
d, making this
crograph of adn the drawing o
hin section of a
teins or other tive staining.
d on a carbonconds to minuSubsequently,n, the heavy mgrid is dried ins aggregate a
contrast” (Figus technique a p
denoviruses prepon the right side
Andres
a high-pressure
small particleA few micro
n coated TEMutes) and wash, a droplet of
metal solution in the air andaround and onure 6.4). The powerful tool
epared by negae.
Kaech
frozen, freeze-s
es can be prep liters of an a
M grid. The suhed with a droan aqueous his removed w
d ready for inn the surface owhole procesfor investigat
tive staining. T
substituted cere
ared in a mucadequately conuspension is reoplet of water.heavy metal sith a filter pap
nvestigation inof the particleses of fixation
tion of single p
The principle of
ebellar slice cu
ch simpler wayncentrated susemoved with . The particles
solution (tungper after an inn the transmises according tn, dehydrationparticles at hig
f negative stain
April 2012
23
ulture of mouse.
y than cells orspension con-a filter paper
s attach to thesten, uranium
ncubation timession electrono their topog-n and embed-gh resolution.
ning and image
2
3
.
r -r e
m e n --
e
Introduction to electron microscopy Andres Kaech April 2012
24
Preparation of cells and tissues for scanning electron microscopy
Typical preparation procedures for cells or tissues for scanning electron microscopy include the steps depicted in figure 6.5.
Figure 6.5: Left: Classical preparation pathway for scanning electron microscopy. The specimen is chemically fixed and dehydrated as described for classical preparation for TEM, followed by critical point drying and coating. Right: Cryo prepa-ration pathway. The specimen is frozen, freeze-fractured, partially freeze-dried, coated and imaged in the cold state in the SEM (Cryo-SEM). The specimen always stays at low temperature and undergoes physical treatment only.
The classical preparation procedure for scanning electron microscopy involves critical point drying (left path in figure 6.5). Air drying of specimens at room temperature leads to a collapse of the sample structure caused by the high surface tension of water. Critical point drying was developed in order to prevent the destroying forces of the surface tension. The first steps including fixation and dehydration are performed the same way as described for the classical preparation for TEM. After dehydration, the ethanol is exchanged with acetone. In a dedicated device (critical point dryer) the acetone is exchanged with liquid pressurized carbon dioxide. Subsequently, the temperature and the pressure in the chamber of the critical point dryer are increased until they reach the critical point of CO2 (31°C, 74 bar). The CO2 in the critical state (neither gas nor liquid) is released from the chamber very slowly providing a gently dried specimen. The process is shown in figure 6.6.
Fixation
Exchange of acetone with pressurized liquid
CO2
Dehydration(ethanol, acetone)
Critical point drying
Chem. fixationChem. fixation Cryo-fixationCryo-fixation
Coating (Contrast)
SEM
Dry specimen
Partial freeze-drying
Freeze-fracturing
Frozen hydrated specimen
Cryo-SEM
No ch
emica
l treatm
ent!
Introdu
Figure
The celectrthe scing (F
Note: aqueo
Figure
Similachemicryo tfor cr120°Cprovid
uction to electron
e 6.6: Phase dia
critical point don beam evap
canning electrFigure 6.7).
The critical pous specimens
e 6.7: Critical p
ar artifacts as ical fixation atechniques forryo preprationC) in a dedicding cross-fra
microscopy
agram of carbo
dried specimeporation. The on microscop
point of waters since it woul
point dried liver
during classiand dehydratior SEM (right p
n for TEM. Onated freeze-fr
acture planes a
Pres
on dioxide. The
en is coated wheavy metal
pe to the surfa
r is at 374°C ld be destroye
r tissue, coated
cal TEM prepon at room tempath in figure nce the specimracturing macas well as fract
solid
ssure
Andres
red line in the p
with a thin laylayer renders
ace providing
and 221 bar. ed at the critica
d with platinum.
paration are inmperature. Ma6.5). The free
men is adequachine. The frature planes th
liquid
gas
SS
EE
Phase diagram o
Kaech
panel shows the
yer of heavy s the specimenthe required y
Critical pointal point condi
Blood vessel w
ntroduced duriany of these aezing or cryo ately frozen, iacturing throu
hrough biologi
SS StartingEE End poinCC Critical p
CC
of CO2
e process of cri
metal (e.g. 2 n conductive yield of electr
t drying is nottions required
with erythrocyte
ing this proceartifacts can bimmobilizatiot is fractured ugh the speciical membrane
pointntpoint
Temperature
itical point dryi
nm) by sputtand localizes
rons and contr
t possible dired for water.
es and leukocyte
ess since it invbe prevented aon is the sameat low tempeimen is a ranes (figure 6.8)
April 2012
25
ing.
ter coating ors the signal inrast for imag-
ectly with the
e.
volves as wellagain by usinge as describederatures (e.g. -ndom process).
2
5
r n -
e
l g d -s
Introduction to e
26
Figure 6.8: Frfracture face o
After fracturzen water (palow temperatcedure excluperatures thro
Figure 6.9: Crwith platinum.
electron microsco
Freeze-fracturingof the lipid bilay
ring, the visibiartial freeze dture for imagi
usively involvough the who
ryo-SEM image.
opy
g of biologicalyer. Cyt…Cytos
ility of the ultdrying). The fring in the cry
ves physical ple process of p
e of high-pressu
l specimens. PFsol, the cytosol
trastructure cafracture face oyo scanning elprocesses. Thepreparation an
ure frozen mou
Andres Kaech
F…plasmatic fris cross-fractur
an be enhanceof the specimelectron microse specimen isnd imaging.
use brain tissue
racture face of red in the cell o
ed by sublimaen finally is coscope (figure
not treated c
e, freeze-fractur
Ice
the lipid bilayeon the right.
ation of small oated with a h6.9). Note: Th
chemically an
red, partially fr
Ap
er. EF…Exopla
amounts of thheavy metal la
This preparationd stays at low
reeze dried and
pril 2012
asmatic
he fro-ayer at on pro-w tem-
d coated
Introduction to electron microscopy Andres Kaech April 2012
27
7. Literature
Physical Principles of Electron Microscopy, Ray F. Egerton, Springer Verlag, 2007.
Griffith G. (1993). Fine Structure Immunocytochemistry. New York, Berlin, Heidelberg. Springer Verlag. ISBN 0-387-54805-X.
Electron microscopy methods and protocols / ed. by M.A. Nasser Hajibagheri. - Totowa, N.J. : Humana Press, cop. 1999. (Methods in molecular biology ; vol. 117)
Electron microscopy : methods and protocols. - 2nd ed. / ed. by John Kuo - Totowa, N.J. : Humana Press, 2007. (Methods in molecular biology ; 369)
Electron microscopy : principles and techniques for biologists / John J. Bozzola, Lonnie D. Russell. - Boston : Jones and Bartlett, 1991. (The Jones and Bartlett series in biology)
Funktionelle Ultrastruktur : Atlas der Biologie und Pathologie von Geweben / Margit Pavelka, Jürgen Roth. - Wien : Springer, 2005.
Cavalier A., Spehner D., Humbel B. M. (2008). Handbook of Cryo-Preparation Methods for Electron Microsco-py, CRC Press, Boca Raton
Harris, J. R (1997). Negative Staining and Cryoelectron Microscopy. Oxford, BIOS Scientific Publishers Lim-ited
Echlin P. (1992). Low-Temperature Microscopy and Analysis. New York, Plenum Press. ISBN 0-306-43984-0.
Robards A. W., Sleytr U. B (1985). Low Temperature Methods in Biological Electron Microscopy. Practical Methods in Electron Microscopy, Volume 10. Edited by Glauert A. M. Amsterdam, Elsevier Science Publishers B. V. (Biomedical Division)
Steinbrecht R. A. and Zierold K. Cryotechniques in Biological Electron Microscopy (1987). New York, Berlin, Heidelberg. Springer-Verlag. ISBN 0-387-18046-X
http://www.matter.org.uk/tem/default.htm