An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the...

29
C An Instr Center fo n Introd rumen or Micros duction ntation, A scopy an n to El , Imag Andres K d Image lectron ging an Kaech Analysis n Micro nd Prep s, Univer oscopy paratio rsity of Z y on Zurich

Transcript of An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the...

Page 1: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

C

AnInstr

Center fo

n Introdrumen

or Micros

ductionntation,

A

scopy an

n to El, Imag

Andres K

d Image

lectronging an

Kaech

Analysis

n Micrond Prep

s, Univer

oscopyparatio

rsity of Z

y on

Zurich

Page 2: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron
Page 3: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introduction to electron microscopy Andres Kaech April 2012

1

Table of content

1.  Introduction ................................................................................................................................................ 2 

2.  Principle of SEM and TEM ........................................................................................................................ 4 

The electron source .......................................................................................................................................... 5 

Electromagnetic lenses .................................................................................................................................... 6 

Specimen holders and stages ........................................................................................................................... 8 

3.  Electron specimen interactions ................................................................................................................. 10 

4.  Contrast formation and imaging in TEM ................................................................................................. 12 

5.  Contrast formation and imaging in SEM .................................................................................................. 15 

Imaging using secondary electrons ................................................................................................................ 15 

Imaging using backscattered electrons .......................................................................................................... 17 

Elemental analysis using x-ray ...................................................................................................................... 18 

Artefacts ........................................................................................................................................................ 19 

6.  Sample preparation ................................................................................................................................... 21 

Preparation of cells and tissues for transmission electron microscopy .......................................................... 21 

Preparation of cells and tissues for scanning electron microscopy ................................................................ 24 

7.  Literature .................................................................................................................................................. 27 

Page 4: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introduction to electron microscopy Andres Kaech April 2012

2

1. Introduction

Microscopy enables a “direct” imaging of organisms, tissues, cells, organelles, molecular assemblies and even individual proteins. Analytical techniques as used in molecular biology provide a data set which represents cellu-lar processes in an abstract way. Therefore, microscopy was and still is an important, complementary technique to visualize the macro and/or microscopic structure and to assign structure to function and vice versa.

Microscopy techniques, their resolution limit, scales and corresponding objects are depicted in figure 1.1. The typical size of eukaryotic cells is 10-20 micrometer and their subcellular components can not be seen by naked eyes. Additionally, cells are colorless and transparent. The organelles or other constituents have to be stained to make them visible in the light microscope. A variety of histological stains and fluorescent markers have been developed.

Imaging the ultrastructure of cellular components requires electron microscopes which have a much higher re-solving power than light microscopes. The physical properties of electron microscopes (e.g. high-vacuum) de-mand specific preparation and staining techniques to reveal the ultrastructure of cells and tissues.

Fig. 1.1: Resolution limit of different imaging techniques, involved radiation and size of biological objects.

The limit of resolution of an optical system depends on the numerical aperture and the wavelength of light (Law of Ernst Abbé). This law holds as well for electrons, of which the speed determines their wavelength: The higher the speed of the electrons, the smaller the wavelength and the better the resolution. The practical resolution of a transmission electron microscope (TEM) run at 100 kV (acceleration voltage that determines the speed) is ap-proximately 0.5 nm and exceeds the resolution of light microscopes by far. In scanning electron microscopes (SEM), the effective resolution is around 1 nm. The resolution of modern electron microscopes is much better than the resolution that a prepared biological specimen can provide due to preparation.

The most important and challenging part of electron microscopy of biological material is the preparation proce-dure. With every preparation step artifacts are introduced. Electron microscopists have put a lot of effort in the preparation of the specimens to reduce the artifacts to a minimum and to image the specimens in a more lifelike state. The image of the specimen can never be better than the entity of the introduced artifacts during prepara-tion.

In combination with immunocytochemical methods, electron microscopy is powerful method to label and high-light single, specific proteins enabling a correlation of the ultra structure and its function.

Note: Preparation of biological specimens for electron microscopy is often a very time-consuming process.

Atoms

1 mm

100 m

10 m

1 m

100 nm

10 nm

1 nm

0.1 nm

Cells

Red blood cells

Bacteria

Mycoplasma

Viruses

Proteins

Amino acids

Radio

Infrared

Visible

Ultraviolet

x,-rays

Human eye

Electron microscope

Light microscope

Resolution Limit Wavelength/Size Object

MRI, CT

Page 5: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introdu

Fig. 1transming moed imaview. Iand fre

uction to electron

.2: Overview ofmission electronodes in transmisage of the sampImages show Heeze-fractured H

Transmiss

microscopy

of the different n microscope onssion and scannple in TEM. A b

Hela cells, thin sHela cell image

sion electron

types of electron the left and aning electron mbulk sample is section of a fixeed in the frozen

n microscopy

Andres

on microscopesa scanning electmicroscopy. A th

scanned with aed, dehydrated,n state in a cryo

y

Kaech

s and their gentron microscophin sample is tra focused beam plastic embedd

o SEM on the rig

Scann

neral mode of ie on the right. Sransmitted by th of electrons inded cell cultureght.

ning electron

imaging. Top rSecond row: Ache electrons to fn an SEM prove (70 nm thickn

n microscopy

April 2012

3

row: Biologicalccording imag-form a project-iding a surfaceess) on the left,

y

2

3

l --e ,

Page 6: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introduction to e

4

2. Prin

The overall dthe light is lenses. The dfigure 2.1.

Fig. 2.1: Similthe tip of a heand are guidethe specimen aand recorded.

The design opicted in figu

Fig 2.2: Similaat the tip of a scanning elect

electron microsco

nciple of SE

design of an esubstituted wdesign of a tra

larity of a transeated filament. ed through the eand leads to an

of a scanning ure 2.2.

arity of a scannheated filamen

tron microscop

opy

EM and TEM

electron microwith electrons

ansmission el

smission electrThe electrons aelectron micros

n image. The im

electron micr

ning electron mnt. The electronpe column by e

M

oscope is similand the glas

lectron micros

on microscope are acceleratedscope column b

mage is monitore

roscope and it

icroscope with ns are acceleratlectromagnetic

Andres Kaech

lar to that of ass lenses are scope and its

with a wide fied with high voltby electromagnred on a phosph

ts similarity t

a confocal laseted with voltag

c lenses. The be

a light microssubstituted wsimilarity to

eld light microstages (60 – 120etic lenses. The

horescent screen

to a confocal

er scanning micges between 0.2 eam is focused

cope. In the ewith electroma light micros

scope. An electr00 kV dependinge beam penetran or specially d

laser scanning

croscope. An ele– 30 kV and ain the “object

Ap

electron micromagnetic/electr

scope is depic

ron beam is forg on the type of

ates and interacdesigned CCD

ng microscope

ectron beam is fare guided throutive” lens and a

pril 2012

oscope, rostatic cted in

rmed at of TEM) cts with camera

e is de-

formed ugh the a small

Page 7: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introdu

beam sted eledetecteimagin

Electris necevacucan flresiduum syTEM serial

Fig 2.3

The e

Electrof thea tungthrouggy andbetwewhichtremevery hresolu

uction to electron

spot is moved/sectrons are deted in accordingng device such a

ron microscopcessary to preuated steadily ly before hittinual gas molecuystem of an eis given in fisetup of low

3: Example of a

lectron sourc

rons can eithee electron sourgsten filamengh the electrond, therefore, n

een the electroh the elctrons ely fine tungstehigh yield of ution. These in

microscopy

scanned over thtected in specifg detectors. Eveas a CCD came

pes are high-vvent oxidationto a vacuum

ng another parule in the systlectron microigure 2.3. Higand high vacu

a vacuum system

ce

er be producedrce and its pront, a LaB6 crn source enab

need to be accon source (caare guided anen tip withoutelectrons and

nstruments are

he specimen surffic detectors anery scanned pixera in SEM.

acuum systemn/burning of tof 10-5 to 10-7

rticle, accounttem reduces thscope consist

gh vacuum puuum pumps is

m for a TEM. R

d by thermionoperties are deystal or a Zr

bling the escapelerated to the

athode) and and acceleratedt heating (room

d the very lowe very costly a

Andres

rface using deflnd amplified u

xel on the specim

ms. In the areathe heated fil7 mbar. The mts to about 50he resolution ats of a cascadumps can not

required.

RP…rotary pum

nic emission oepicted in figurO/W Schottkpe of electrone desired speean anode plated. During coldm temperature

w chromatic aband require pa

Kaech

lector coils. Theusing photomultmen represents

a of the electroament. The cmean free path0 meter at a vaand performan

de of low and directly pump

mp, TMP…turbo

r in a process ure 2.4. Durinky emitter is hs. The electro

ed before entere is applied ld field emissioe). The advantberration of tharticularly hig

e beam interacttiplier tubes. Ea pixel on the s

on source a vaolumn and thh, referred to

acuum of 10-6 nce by scatterhigh-vacuum

p against amb

o molecular pum

called cold fig thermionic heated by an ns leaving thering the electrleading to an on, the electrotage of cold fihe electrons ah vacuum.

ts with the specEmitted radiatioscreen. Importa

acuum of 10-7

he area of the as the distancmbar. Any co

ring of electrom pumps. An ebient pressure

mp, IGP…ion g

field emission.emission, a v

n electrical cue filament havron column. Aelectrostatic

ons can escapfield emission allowing imag

April 2012

5

cimen and emit-on as x-rays isant: There is no

to 10-10 mbarspecimen are

ce a moleculeollision with aons. The vacu-example for a. Therefore, a

getter pump

. The materialery fine tip of

urrent flowingve a low ener-A high voltage

field throughe from an ex-sources is the

ging at atomic

2

5

-s o

r e e a -a a

l f g -e h -e c

Page 8: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introduction to e

6

Fig. 2.4: Diffe

Electromagn

Electromagnpiece (figurethe pole pieca charge pastion of the mthrough the l

Depending oan electron mness of the bform the primtive lens. Thlenses whichphotosensitivtion of the imrected in mod

Material

Heating tem

Normalized

Required vac

∆E (eV)

electron microsco

erent types of el

netic lenses

netic lenses coe 2.5 A). A mce of the lens. sing a magnet

magnetic fieldlens system (fi

on the strengthmicroscope acbeam and guidmary image. Fhe final magnh act in combve film), see amage dependindern TEM by

p. (K)

brightness

cuum (Pa)

opy

lectron sources

onsist of a hugmagnetic field i

The accelerattic field. The

d and the direfigure 2.5 B).

h of the magncts as a condedes it to the spFocusing of thification is deination and p

also figure 2.1ng on the maga set of corre

Tungsten

W

2700

Chrom

high

and their prope

ge bundle of wis induced byted electrons eresultant forc

ection of the

netic field, thenser lens, whipecimen. The

he image is peretermined by

project the fin and 2.2 for le

gnification (thctor coils.

Thermionic

LaB6

LaB6

1800

matic aberration!

Andres Kaech

erties. The imag

windings of iny the current inentering the mce is always pelectrons. In

e focal width oich bundles th

e electrons tranrformed with the following

nal image on tens set up. Th

he lens curren

c

Schottky

ZrO/W

1800

ge shows a Sch

nsulated coppn the winding

magnetic field erpendicular tconclusion, t

of the lens is he electrons, dnsmit the spethe strength o

g projective lthe fluorescenhe circular patt of the projec

Cold field emi

W

300

Ultra high

ottky emitter.

per wire, a sofg and reaches are deviated fto the plane dhe electrons t

changed. Thedetermines thecimen and theof the magnetienses, a systent screen or ath of the electrctive lenses).

ission

Ap

ft iron cast anits main stren

following the defined by thetake a circula

e first lens syse the overall be objective leic field in the ems of two ora CCD cameratrons leads to This rotation

2.5 cm

pril 2012

nd pole ngth at law of

e direc-ar path

stem in bright-ns and objec-r three a (or a a rota-is cor-

Page 9: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introdu

Fig. 2.changepassinThe bl

Electrtion aent elconfulensesregulacorrec

A

uction to electron

.5: Electromagned by the curre

ng the magneticlack arrows dem

romagnetic lenand astigmatismectrons derivi

usion of the ims and aperturearly, in extremctor coils.

microscopy

netic lens. A) Tent trhough thec field are deviamonstrate the ro

nses show them. The most ping from a po

mage (figure 2es and chargime cases after

The magnetic fiee windings. Theated perpendicuotation of the im

e same aberratprominent abe

oint source do2.6). Astigmating of the sper every move

Andres

eld is strongeste lens is water ular to the planmage.

tions as glass erration in ele

o not match onatism is causedecimen. In pa

ement of the s

B

Kaech

t in the area of cooled to main

ne defined by th

lenses such aectron microscne and the samd by inhomogarticular in SEspecimen. The

B

the pole piece. tain stability d

he magnetic fiel

s chromatic, scopy is the axime point in thgeneities of thEM, the astige correction i

The focal lengduring operationld B and the ve

spherical aberial astigmatismhe image plainhe lenses, congmatism mustis done again

April 2012

7

th of the lens isn. B) Electrons

elocity vector v.

rration, distor-m. The differ-n leading to a

ntamination oft be correctedwith a set of

2

7

s s .

--a f d f

Page 10: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introduction to e

8

Fig. 2.6: a) Axthe image planmatism using c

Specimen ho

In TEM, the very thin so tcal specimenhigher the voper grids of 3sections on tsince the sys

a)

b)

electron microsco

xial astigmatismne (focus). Undcorrector coils.

olders and sta

electron coluthat the electr

n should be aroltage, the thi3 mm diametetop are attachestem always st

opy

m in an electronderfocus and ov

ages

umn does not orons can penetround 70 nm fcker specimener, which are aed in a holdertays under hig

n optical systemverfocus images

offer a lot of trate the specimfor a TEM wins can be examavailable in a r and introducgh vacuum. Th

Andres Kaech

m. In focus, a pes appear as ell

space for the men and formith an acceleraamined). Thin wide variety o

ced into the gohe goniometer

oint source applipses orthogon

specimen. Adm an image. Thation voltage sections of th

of materials aoniometer of tr is the mecha

pears as a circlal to each othe

dditionally, thhe average thifor the electro

he specimen and mesh sizesthe TEM throanical setup w

Ap

le of least confuer. b) Corrected

he specimen mickness of a bons of ~100 kare mounted os. The grids w

ough a vacuumwhich enables

pril 2012

usion in d astig-

must be iologi-

kV (the on cop-with the m lock, highly

Page 11: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introdu

precisimage

Fig. 2.goniom

The sspecim(bulk holderallowby han

Fig. 2.

Electron aired frowith e

uction to electron

se and stable e, in particular

.7: Thin sectionmeter of the TEM

specimen chammen can be atspecimens). Tr able to carrying controllednd requiring v

.8: The inside v

ron microscopr cushions whiom the rest of electromagnet

microscopy

control of ther at high magn

ns of a specimeEM through a va

mber of SEMt least 10 cm The specimeny 8 stubs for d movement inventing of the

view of an SEM

pes are very seich absorbe vithe building ttic coils (induc

e specimen honifications (fig

en on a TEM gracuum lock.

Ms are large anin diameter a

n usually attacexample. Then x, y, z, tiltinwhole chamb

M chamber as see

ensitive to vibibration. Ofteno minimize dictors) mounte

Andres

older during imgure 2.7).

rid, holder tip a

and offer mucand 2 cm thickched on a sme stub holder ng and rotatio

ber or by using

en with an infra

bration and elen, they are evisturbance by

ed to the the w

Kaech

maging. Any

and complete sp

ch more spacek, since it doe

mall aluminiumthen is mount

on. The stub hg a vacuum lo

ared camera mo

ectromagneticen setup on spvibration. Ele

walls, ceiling a

drift or instab

pecimen holder

e for the speces not have tom stub (1 cm ted on a 5 axiholder can be ck.

ounted on the d

c fields. The epecial fundamectromagneticand floor.

bility results i

r, which is intro

cimen. In routo be permeabldiameter) is mis stage of theinserted in th

door of the instr

electron columments which arc fields can be

April 2012

9

in an unsharp

oduced into the

tine SEM thele to electronsmounted on ae microscope,he microscope

rument.

mn is mountedre disconnect-

e compensated

2

9

p

e

e s a , e

d -d

Page 12: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introduction to electron microscopy Andres Kaech April 2012

10

3. Electron specimen interactions

The electron beam (TEM and SEM) hits the sample and either passes the sample unaffected or interacts with it. Three different interactions of electrons with an atom can be generally discerned (Fig 3.1):

i) An electron of the primary beam is scattered by the electrostatic interaction with the positively charged nucleus of an atom at an angle of more than 90°, yielding backscattered electrons. This type of electrons has practically the same energy as the ones of the primary beam.

ii) An electron of the primary beam is scattered by the electrostatic interaction with the positively charged nucleus of an atom at an angle of less than 90°, yielding elastically scatterd electrons. Al-so these electrons do not loose energy and therefore are referred to as elastically scattered electrons.

iii) Electrons can also loose energy while interacting with the “electron cloud” of the atom. These are called inelastically scattered electrons. This interaction can lead to the following processes (see also http://www.matter.org.uk/tem/default.htm):

Inner-shell ionisation An electron is pushed out of the electron cloud. The electron „hole“ is filled by an electron of an outer shell: Surplus energy is either emitted as characteristic x-ray or transferred to another elec-tron, which is emitted (Auger electron).

Bremsstrahlung (continuum x-rays) Decceleration of electrons in the Coulomb field of the nucleus ⇒ Emission of x-ray carrying the surplus energy ΔE ⇒ Uncharacteristic x-rays

Secondary electrons (SE) Loosely bound electrons (e.g., in the conduction band) can easily be ejected ⇒ low energy (< 50 eV)

Phonons Lattice vibrations (heat) ⇒ beam damage

Plasmons Oscillations of loosely bound electrons in metals

Cathode luminescence

Fig. 3.1: Interaction of electrons with an atom. Note: Inelastically scattered electrons lead to a variety of processes like heating, radiation and electron emission as described in the text.

Inelastic(low angle, E=E0-∆E)

Unscattered(E=E0)

Primary electrons (E0)Backscattered electrons (E=E0)

Elastic(higher angle, E=E0)

Page 13: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introduction to electron microscopy Andres Kaech April 2012

11

Almost all types of electron interactions can be used to retrieve information about the specimen. Depending on the kind of radiation or emitted electrons which are used for detection, different properties of the specimen such as topography, elemental composition can be deduced.

The follwing figure 3.2 shows the interaction of the electron beam with the specimen. In biological TEM analy-sis, the elastically scattered electrons are mainly used for imaging. The inelastically scattered electrons will be recorded as well with the imaging device (e.g. CCD camera). They have a slightly lower energy (a longer wave-length) than the primary electrons. In conclusion, they are deviated less strongly in the electron lenses and do not merge at the same spot as the elastically scattered electrons (chromatic aberration). They contribute to unsharp-ness of the image. In high-end instruments, the inelastically scattered electrons can be filtered out with an energy filter.

In SEM, mainly secondary electrons are used for imaging of biological specimens. These electrons have a very low energy (around 50 eV) compared to the energy of the primary electrons (up to 30 keV). Due to the low ener-gy, these electrons can escape only from the surface area of the specimen (for details see chapter 5) and therefore provide information about the surface topography. Also backscattered electrons are frequently used for imaging with an according backscatter detector. The number of electrons which are backscattered from a certain spot of the specimen depends on the local elemental density of the specimen. Hence, backscattered electrons provide a “density image” and information about the elemental composition, respectively. Backscattered electrons are also more resistant to charging due to their high energy.

X-rays are always produced when matter interacts with an electron beam. The energy of a part of the emitted x-rays is specific for different atoms and can be used for elemental analysis in TEM and SEM using special detec-tors. However, the analysis of light elements is rather difficult and the high corrugation of biological specimen surfaces makes the interpretation of elemental maps very demanding. A lot of experience is necessary to prevent pit falls.

Fig 3.2: Interaction of electrons with specimen/matter, induced radiation and emission. Transmitted electrons like elastically scattered electrons are typically used for TEM imaging of biological specimens. Secondary electrons and backscattered electrons are mainly used for imaging in the SEM.

PrimaryPrimary electronselectrons

UnscatteredUnscattered electronselectrons

ElasticallyElastically scatteredscattered electronselectrons

Inelastically scattered electrons

SecondarySecondary electronselectronsBackscatteredBackscattered electronselectrons

AugerAuger electronselectronsHeat

Cathode luminescenseX-rays

Specimen Interaction volume

SEM analysis

TEM analysis

Page 14: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introduction to e

12

4. Con

The image fotrons are verying on the ph

The primary tromagnetic forms the priconsist mainhence, provid

Also the absAbsorbtion oarea of the sp

In conclusionby impregnaand lead ionatomic numbelectrons. Thjective apertumore heavy aperture. Thedense” area trast (more o

Fig. 4.1: Conting on the conthe objective awill yield low intensity (brig

electron microsco

ntrast form

ormation in Ty similar to th

hysical phenom

electron beamwaves. The oimary image (

nly of very ligde almost no c

sorbtion of eleof electrons wpecimen is exp

n, the specimeation of the ths bind specifiber and densihe main contrure located inmetals will leese electrons in the image. or less electron

trast formation nstituents, e.g. aperture and do

signal intensityght spots).

opy

mation and i

TEM is basicalhe ones of lighmenom to be e

m passes throobjective lens (see also lectu

ght elements licontrast.

ectrons by thewould result inposed to high

en (thin sectiohin section wiically to diffeity of the hearast formationn the back-focaead to a highdo not reach Therefore, th

ns will reach t

in biological obmembranes (ph

o not reach the y (electron den

imaging in

lly the same aht. As for lightexplained (wa

ough the thin collates the d

ure Bio 416 Like C, O, H, N

e specimen isn heat damageradiation, bea

on) must be trith heavy metrent constituevy metal ionsis achieved b

al plane of theer number ofthe imaging

he image formthe CCD, scre

bjects: A differehospholipids), rimaging device

nse, dark spots)

Andres Kaech

TEM

as for regulart, electrons caave particle du

specimen anddiffracted and

Light MicroscoN, S, P and le

s very small ae of the speciam damage is

reated to provtals like uranients of biologs leads to a mby trapping the objective lenf diffracted/scdevice (CCD

mation of bioleen), figure 4.

ent number of hribosomes, ande (CCD, viewing) whereas spots

light microscan be observeduality).

d generates nod non-diffracteopy by Urs Ziead to poor in

and does not imen. At highan issue and c

vide enough cium and lead.ical matter an

much higher ihe numerous dns (figure 4.1

cattered electr, screen) and

logical specim2.

heavy metal iond chromatin. Mag screen). Specs with no or few

copy, since thed as particles o

on-diffracted ed waves in thiegler). Biolognteraction wit

contribute to h magnificatiocan easily be o

ontrast for imThe positive

nd to a differeinteraction andiffracted elec). Spots of theons, which wwill cause a

mens is based

ns stick to the saany diffracted/s

cimen spots withw heavy metal

Ap

e properties oor as waves d

and diffractedthe image plangical objects uth the electron

contrast formons at which aobserved.

maging. This iely charged urent extent. Thnd diffraction ctrons using te sample cont

will hit the obj“dark” or “elon amplitud

ample surface d/scattered electrh a lot of heavyions yield high

pril 2012

of elec-epend-

d elec-ne and usually ns and,

mation. a small

is done ranium he high

of the the ob-taining jective lectron

de con-

depend-rons hit y metals h signal

Page 15: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introdu

The d

Fig. 4.

Contrnel rinToo mcorres

Fig. 4.focus.

In parheavydispla

uction to electron

difference of h

.2 Non-stained

rast of biologings (deletion/much underfocsponds to min

.3: Under-focusToo much over

rticular, thin sy metal solutioayed in figure

microscopy

heavy metal st

versurs heavy m

ical specimens/amplification cus or overfoc

nimum contras

s, over-focus anr- or underfocus

sections of frons. The only 4.4.

ained and non

metal stained th

s can be furthof signal), w

cus introducesst.

nd true focus ofs provides an u

ozen soft, hydway to genera

Andres

n-stained thin

hin section of an

her improved bwhich improve

s artifacts and

of a thin sectionunsharp image w

drated organiate enough co

Kaech

section of an

n alga.

by slight unde the visibility

d the image ge

n of a cell. Minwith artefacts.

c matter can ontrast for ima

alga is shown

erfocusing. Unof biological

ets unsharp. T

imum contrast

not be stainedaging is using

n in figure 4.2.

nderfocusing l structures (F

True focus (Ga

corresponds to

d by incubatiunderfocus. A

April 2012

13

.

leads to Fres-Fig 4.3). Note:aussian focus)

o true Gaussian

ng them withAn example is

1 µm

2

3

-: )

n

h s

Page 16: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introduction to e

14

Fig. 4.4: Frozat < – 150° C.

The primary final image cscreen consisusually perfotaken by expal film develcameras are Standard CCtions for bioarea of intere

The assemblywith electronwith a scintilthe scinillatocially availab

Fig. 4.5: Desig

electron microsco

zen hydrated thi. Contrast is tre

image is magcan be viewests of a phospormed in the dposing electronlopment in a d

available whCD cameras fo

logical imaginest and acquis

y of a CCD cans. The chip wllator. On top or to the pixelble recently (v

gn of a CCD ca

opy

in section of a Gemendously enh

gnified or demed by eye on phorescent laydark to receivn sensitive filmdark room. Nohich reach theor TEM have ng. Pixel binnition of image

amera for TEMwould be damof every pixe

l of the CCD.very expensive

amera for electr

Golden deliciouhanced by under

magnified by ta large screen

yer which starte a good contms to the elecowadays, spece resolution oa number of 2ning (e.g. 2, 4es can easily b

M is explainedmaged. Therefoel of the CCD . Cameras acqe).

ron microscopy

Andres Kaech

us apple leaf asrfocusing.

the projectiveen in the micrts to emit lightrast on the victron beam andcialized CCDof a conventi2k x 2k to 4k 4, 8 times) enbe performed w

d in figure 4.5ore, the electrcamera, a sm

quiring image

y

s imaged in a cr

e lenses arrangroscope (apprht when electrewing screen.d the negative cameras are ional chemicax 4k pixels, b

nables fast reawith the CCD

5. The CCD chrons have to b

mall fiberoptic es directly wit

500

ryo transmission

ged below theoximately 20 ons hit it. Ele. For many dees were develoused for imag

al film (very being sufficienal-time viewin

D camera on co

hip cannot be be primarily co

element guidth electrons h

0 nm

Ap

on electron micr

e objective len0 cm diameterectron microscecades, imageoped by convege acquisitionexpensive th

ent for most apng. Searching omputer scree

directly bombonverted to p

des the photonhave become c

pril 2012

roscope

ns. The r). The copy is es were ention-

n. CCD hough). pplica-of the

ens.

barded hotons

ns from comer-

Page 17: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introduction to electron microscopy Andres Kaech April 2012

15

5. Contrast formation and imaging in SEM

In SEM, an electron beam with a tiny spot is guided pixel by pixel over the surface of a specimen, e.g. 1024 x 768 pixels. At every pixel, the beam stays for a defined time and generates a signal (e.g. secondary electrons) which are detected, amplified and displayed on a computer screen. The scanning of the beam is synchronized with the display of the computer screen. The signal deriving from every single pixel on the sample is simultane-ously displayed on the corresponding pixel of the computer screen and finally forms the image. Note: There is no CCD camera for image acquisition in SEM! The image consists of displayed grey levels, e.g. 256 gray levels in an 8 bit image. The magnification is changed by scanning a smaller area with the same number of pixels. The pixel size therefore gets smaller and the resolution is increased.

The beam of electrons wich hits a spot of the surface of the specimen interacts with a whole volume of the spec-imen. The interaction volume looks like a pear (figure 5.1). Secondary electrons are produced in the whole vo-lume but can escape only from a small volume close to the surface of the specimen due to their low energy (around 50 eV). Backscattered electrons manage to escape from a larger depth since their energy is the same as the primary electrons. On their way through the sample, backscattered electrons can again interact with atoms and produce secondary electrons (called SE 2 electrons). X-rays can escape from the whole interaction volume.

The size of the interaction volume is dependent on the acceleration voltage of the primary electrons and the atomic number of the atoms close to the surface. Heavy atoms decelerate the beam more than light elements and therefore reduce the interaction volume.

Fig. 5.1: Interaction volume of the electron beam with a bulk sample. Escape depths of SE, BSE, and x-rays are indicated. Note: The SE escape depth is independent on the energy of the primary electrons (PE).

Imaging using secondary electrons

Secondary electrons (SE) are mainly used in SEM for imaging the surface topography of biological specimens. They are detected with an Everhart Thornley detector. The principle of detection is shown in figure 5.2. The SE are collected by a collector grid. A voltage of + 200 to 500 V is applied to the collector grid which attracts the low energy electrons. The SE then hit a scintillator which converts the electrons to photons. The photons are guided by a light conducting tube on the photomultiplier tube (figure 5.3), where the photons again are converted to electrons that are amplified finally leading to an electrical signal. The current is translated into a gray value and displayed on the screen of the monitor.

…interaction volume RPE

SE escape depth (E < 50 eV)

BSE escape depth(E = E0)

X-ray escape depth

“high voltage” “low voltage”

Page 18: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introduction to e

16

Fig 5.2: Princ

Fig. 5.3: Work

The SE escaptrons. Howevaction volumuppermost lavoltages is hitrate deeply i(atomic numbnum. Since bSE escaping volume. The surface, biolothe signal is llower areas cfigure 5.4 an

electron microsco

ciple of SE detec

king principle of

pe depth (λ) aver, the percen

me R is larger aayer. In concluigher and provinto the specim

mber) of the spebiological specthe specimen result is a fain

ogical specimlocalized to thcan not escape

nd figure 5.5.

Primary ele

SE

opy

ction with an E

of a photomultip

as depicted in fntage of the Sat lower accelusion, the numvides a strongmen and the pecimen. It is 1cimen consist is very low. Mnt, unsharp sig

mens need to behe surface, mae through the h

+200

Electron

ectrons

verhardt Thorn

plier tube. The a

figure 5.1. is iE produced inleration voltag

mber of the emger signal. Theproduced SE c10 – 100 nm fo

of light elemeMoreover, thegnal with verye coated with any more SE aheavy metal la

+7-1

0-500V – Co

ns

Andres Kaech

nley detector. Ex

amplification is

independent on this area comges because m

mitted secondae higher the accannot escape for carbon, 2 –ents only, the

e SE escaping y poor contrasa thin layer o

are produced iayer and obsc

2kV HV

ollector volta

Photons

Explanations see

s monitored by

on the accelerampared to the

more electrons ary electrons fcceleration voanymore. λ is

– 3 nm for Chrinteraction vothe specimen

st. To receive f a heavy metin the surface cure the signal

Phot

age

s E

e text.

the voltage of t

ation voltage oSE produced interact with

from the SE esltage, the mor

s also dependeromium, and 1olume is largederive from aa sharp and clal (few nm of layer, and SE

l. This situatio

omultiplier

Electrons

Ap

the photomultip

of the primaryin the whole ithe specimen scape depth atre electrons peent on the den1 – 2 nm for Pe and the numba relatively larlear signal fro

f Au, Pt etc.). NE deriving fromon is explained

pril 2012

plier.

y elec-inter-in the

t lower ene-sity

Plati-ber of rge om the Now,

m d in

Page 19: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introdu

Fig. 5.from threpresproper

Fig. 5.surfacmembring a sare ice

Imagi

BacksferencNo covoltagBSE d

uction to electron

.4: The coatinghe spot of the p

sents the contrar acceleration v

.5: The images e was not coatrane are recognsharp signal ate contamination

ing using bac

scattered electces in the speollector grid isges than 50 eVdetector is loc

SE

microscopy

g of the specimeprimary electroast. The contrasvoltage is select

represent a vieted and therefornizable. Right. t a high signal ns).

ckscattered el

trons (BSE) acimen compos necessary, sV of the SE). cated just bene

E

R

PE

Φ

en with heavy mn beam (PE) isst is a functionted.

ew on the inner re appears noisThe surface wato noise ratio.

lectrons

are also used tosition. For desince the BSEThe BSE pref

eath the object

PE

SE

Andres

metals localizess high. The diffen of the angle b

r surface layer osy, unsharp an

as coated with 4Detail structur

to image the setection of BS

E have the samferably are dirtive lens (figu

Kaech

s the signal to tferent number ofbetween primar

of the plasma md detail structu4 nm of platinumres like hexagon

surface of the SE, a differentme energy as trected back to

ure 5.6).

P

the surface. Thef SE escaping try beam and th

membrane of freures are obscurm localizing thenal protein pat

specimens but arrangementthe primary elowards the obj

SE esca

PE PSE SR Eλ S

PE

e yield of SE dethe specimen athe specimen su

eeze-fractured yred. Only invage signal to the stterns are visibl

ut also for dett of the detectlectron beam

bjective lens. T

ape depth λ

Primary electSecondary elExcited volumSE escape de

April 2012

17

eriving directlyt different spotsrface when the

yeast. Left: Theginations of thesurface provid-le (bright spots

tection of dif-tor is applied.(much higher

Therefore, the

tronslectronsmeepth

2

7

y s e

e e -s

-. r e

Page 20: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introduction to electron microscopy Andres Kaech April 2012

18

Fig. 5.6: Arrangement and working principle of a BSE detector. The scintillator is arranged circular around the primary beam just beneath the objective lens. BSE hit the scintillator and produce photons, which are guided to the photomultiplier tube. As for the SE detector, the signal is amplified, converted to a gray level and dispayed on the computer screen.

Figure 5.7 is an example, how the BSE detector can be used to get additional information of the specimen, e.g. detection of gold particles on the surface of red blood cells.

Fig. 5.7. Red blood cells labelled against surface proteins. The antibody is connected to a gold particle (heavy metal aggre-gation). Whereas the SE signal provides the surface topography, the BSE image distinctly shows the gold particles. The images were recorded simultaneously at 25 kV acceleration voltage. The whole sample is coated with a thin layer of heavy metal (ca. 2 nm). SE derive mainly from this layer and therefore do not clearly accentuate the gold particles, which have a size of around 20 nm. On the contrary, a lot of elctrons are backscattered where the gold particles are, but only very few are backscattered from the thin metal coat.

Elemental analysis using x-ray

X-rays produced in the specimen can be used for elemental analysis of the specimen (Fig 5.8). The x-ray energy is specific for different elements. A dedicated x-ray detector detects the x-rays deriving from every spot of the sample and assigns the according element. Elemental mapping is possible for any area of interest. In particular the interpretation of x-ray spectra is very challenging and needs a lot of experience. A highly corrugated surface of a specimen – typical biological specimen - can lead to misinterpretation. Quantitative analysis with biological specimens is usually not possible.

Scintillator Light pipe Photomultiplier

Primary electrons

BSE

SE BSE

Page 21: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introdu

Fig 5.8

Artefa

Imagimen, ing arlocate

Fig 5.lotus l

Beamdestro

uction to electron

8: Elemental an

acts

ing in the scancharge is intrortefact is showes the signal to

9: Charging arleaf (right)

m damage is anoy the surface

microscopy

nalysis using x-

nning electronoduced into thwn in figure 5o the surface b

rtefacts on the

nother artefactin many diffe

-rays. The eleme

n microscope he material an5.9. Note: Thbut also adds a

surface of a fr

t which often erent ways, fig

Andres

ental spectrum

can lead to a nd must be dishe coating of ta conducting l

freeze-fractured

is encounteregure 5.10.

Kaech

derives from th

lot of artefacssipated in ordthe specimen layer on non-c

d plasma memb

ed. The electro

he red region in

cts. When the der to prevent

with a thin hconducting sp

rane of yeast (

on beam can l

n the image.

beam penetrat charging. A theavy metal lpecimens.

(left) and a wax

locally heat th

April 2012

19

ates the speci-typical charg-ayer not only

xy surface of a

he sample and

2

9

--y

a

d

Page 22: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introduction to e

20

Fig. 5.10: Bea

electron microsco

am damage on t

opy

the surface of a freeze-fracture

Andres Kaech

ed, frozen-hydraated yeast surfa

ace after one sc

Ap

can. Scale bar 5

pril 2012

500 nm.

Page 23: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introduction to electron microscopy Andres Kaech April 2012

21

6. Sample preparation

As we have discussed in the earlier sections, electron microscopes are run under high vacuum conditions. A high vacuum, however, is a very unfriendly environment for biological specimens like cells and tissues (or condensed organic hydrated matter in general). Electron microscopes pose the following prerequisites to the specimen for investigation.

Resistant in the vacuum

Electron beam resistant

Providing contrast

Penetrable for electrons (transmission electron microscope)

Large biological specimens (cells, tissues) do not fulfill any of these requirements and have to be treated in order to be investigated in the electron microscope. Typically, the specimen is turned into a solid state to make it suita-ble for vacuum and resistant in the electron beam. Moreover, it has to be cut to very thin sections of ca. 100 nm for transmission electron microscopy. Such a specimen still does not provide contrast. Biological material is basically composed of light elements like C, O, H, N, S, P, which do not interact with electrons. Selective stain-ing or contrast enhancement with heavy metals is required to form an image as already pointed out in chapters 4 and 5 (contrast formation and imaging in TEM and SEM).

Preparation of cells and tissues for transmission electron microscopy

Typical preparation procedures for cells or tissues for transmission electron microscopy include the steps depict-ed in figure 6.1.

Figure 6.1: Left: Classical preparation pathway for transmission electron microscopy. The whole process is performed at room temperature. Right: Cryo-preparation pathway for transmission electron microscopy. The specimen is frozen and slowly warmed up during freeze-substitution (low temperature dehydration).

The classical preparation is performed at room temperature. During fixation, the native biological specimen is chemically stabilized with chemical fixatives like glutaraldehyde and osmium tetroxide. All biological processes are arrested during fixation. The fixed specimen is ready for dehydration with a sequence of different ethanol concentrations until it is completely dehydrated. Subsequently, the ethanol is exchanged with a monomer solu-tion of a plastic formulation (e.g. Epon/Araldite) in a sequence of rising concentrations of plastic components dissolved in ethanol or acetone. The monomers are polymerized by heat or UV light and provide a hard plastic block containing the embedded specimen. This enables cutting of thin sections (100 nm) in an ultramicrotome. The sections are stained with uranyl-acetate and lead citrate. Uranium and lead ions selectively bind to lipids, proteins and nucleic acid in the specimen and provide a distribution of electron dense material and finally the image in the electron microscope (Figure 6.2).

Embedding

Fixation

Dehydration

Staining (uranyl-acetate, lead-citrate)

Thin sectioning

TEM

Classical preparation(chemical fixation at RT)

Cryo-preparation(cryo-fixation)

Page 24: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introduction to e

22

Figure 6.2: TE

A lot of artifto minutes) apreparation pdehydration a

Cryo preparastep, the specing is very fcause of the ultra structurproperties of 6.3) is achiev

Once the speperatures (e.gplastic, sectio

electron microsco

EM micrograph

facts are introdand the specimprocedure is cand embeddin

ation methodscimen is genefast and can bnature of wat

re. High-pressf the water at 2ved only for b

ecimen is frozg., -90°C, graoning, staining

opy

h of a thin sectio

duced during men has enougconducted at rng.

s were introdurally frozen w

be performed ter to form icesure freezing w2100 bar just b

biological spec

zen, it is dehydadient to roomg of the specim

on of classically

classical TEMgh time to rearoom tempera

uced and devewithout chemic

within millisee crystals duriwas developebefore freezincimens with a

drated and fixm temperaturemen is typical

Andres Kaech

y prepared cere

M preparationact to the chemature leading t

eloped in ordecal pretreatmeeconds. Howeing freezing. Ied to prevent ing. Despite thi

thickness of l

xed with cheme) in a processlly performed

ebellar slice cu

n. The fixationmical attack anto extraction

er to prevent oent (pure physever, the freezIce crystal segice crystal foris technique, sless than 200

mical fixativess called freezethe same way

lture of mouse.

n with chemicand undergoes and shrinkage

or reduce thessical process) zing is a chalgregations distrmation by chsatisfying freeµm.

s simultaneouse-substitutiony as during cla

Ap

Scale bar 1 µm

als is slow (sechanges. The e of material

se artifacts. In– cryo-fixed.

llenging procetort and dama

hanging the phezing quality (

sly at very lown. The embeddassical prepara

pril 2012

m.

econds whole during

n a first Freez-

ess be-age the hysical (Figure

w tem-ding in ation.

Page 25: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introdu

FigureScale b

Negat

Singletissuetaininafter acarbonions) of secmicroraphyding c

Figureformat

uction to electron

e 6.3: TEM micbar 1 µm.

tive staining

e particles likees by a procesng the particlea short incuban layer of theis added on thconds or minuoscope. The hey and provide can be omitted

e 6.4: TEM miction is shown in

microscopy

crograph of a th

e viruses, protss called negaes are pipettedation time (sece TEM grid. She grid. Againutes and the geavy metal iona “negative c

d, making this

crograph of adn the drawing o

hin section of a

teins or other tive staining.

d on a carbonconds to minuSubsequently,n, the heavy mgrid is dried ins aggregate a

contrast” (Figus technique a p

denoviruses prepon the right side

Andres

a high-pressure

small particleA few micro

n coated TEMutes) and wash, a droplet of

metal solution in the air andaround and onure 6.4). The powerful tool

epared by negae.

Kaech

frozen, freeze-s

es can be prep liters of an a

M grid. The suhed with a droan aqueous his removed w

d ready for inn the surface owhole procesfor investigat

tive staining. T

substituted cere

ared in a mucadequately conuspension is reoplet of water.heavy metal sith a filter pap

nvestigation inof the particleses of fixation

tion of single p

The principle of

ebellar slice cu

ch simpler wayncentrated susemoved with . The particles

solution (tungper after an inn the transmises according tn, dehydrationparticles at hig

f negative stain

April 2012

23

ulture of mouse.

y than cells orspension con-a filter paper

s attach to thesten, uranium

ncubation timession electrono their topog-n and embed-gh resolution.

ning and image

2

3

.

r -r e

m e n --

e

Page 26: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introduction to electron microscopy Andres Kaech April 2012

24

Preparation of cells and tissues for scanning electron microscopy

Typical preparation procedures for cells or tissues for scanning electron microscopy include the steps depicted in figure 6.5.

Figure 6.5: Left: Classical preparation pathway for scanning electron microscopy. The specimen is chemically fixed and dehydrated as described for classical preparation for TEM, followed by critical point drying and coating. Right: Cryo prepa-ration pathway. The specimen is frozen, freeze-fractured, partially freeze-dried, coated and imaged in the cold state in the SEM (Cryo-SEM). The specimen always stays at low temperature and undergoes physical treatment only.

The classical preparation procedure for scanning electron microscopy involves critical point drying (left path in figure 6.5). Air drying of specimens at room temperature leads to a collapse of the sample structure caused by the high surface tension of water. Critical point drying was developed in order to prevent the destroying forces of the surface tension. The first steps including fixation and dehydration are performed the same way as described for the classical preparation for TEM. After dehydration, the ethanol is exchanged with acetone. In a dedicated device (critical point dryer) the acetone is exchanged with liquid pressurized carbon dioxide. Subsequently, the temperature and the pressure in the chamber of the critical point dryer are increased until they reach the critical point of CO2 (31°C, 74 bar). The CO2 in the critical state (neither gas nor liquid) is released from the chamber very slowly providing a gently dried specimen. The process is shown in figure 6.6.

Fixation

Exchange of acetone with pressurized liquid

CO2

Dehydration(ethanol, acetone)

Critical point drying

Chem. fixationChem. fixation Cryo-fixationCryo-fixation

Coating (Contrast)

SEM

Dry specimen

Partial freeze-drying

Freeze-fracturing

Frozen hydrated specimen

Cryo-SEM

No ch

emica

l treatm

ent!

Page 27: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introdu

Figure

The celectrthe scing (F

Note: aqueo

Figure

Similachemicryo tfor cr120°Cprovid

uction to electron

e 6.6: Phase dia

critical point don beam evap

canning electrFigure 6.7).

The critical pous specimens

e 6.7: Critical p

ar artifacts as ical fixation atechniques forryo preprationC) in a dedicding cross-fra

microscopy

agram of carbo

dried specimeporation. The on microscop

point of waters since it woul

point dried liver

during classiand dehydratior SEM (right p

n for TEM. Onated freeze-fr

acture planes a

Pres

on dioxide. The

en is coated wheavy metal

pe to the surfa

r is at 374°C ld be destroye

r tissue, coated

cal TEM prepon at room tempath in figure nce the specimracturing macas well as fract

solid

ssure

Andres

red line in the p

with a thin laylayer renders

ace providing

and 221 bar. ed at the critica

d with platinum.

paration are inmperature. Ma6.5). The free

men is adequachine. The frature planes th

liquid

gas

SS

EE

Phase diagram o

Kaech

panel shows the

yer of heavy s the specimenthe required y

Critical pointal point condi

Blood vessel w

ntroduced duriany of these aezing or cryo ately frozen, iacturing throu

hrough biologi

SS StartingEE End poinCC Critical p

CC

of CO2

e process of cri

metal (e.g. 2 n conductive yield of electr

t drying is nottions required

with erythrocyte

ing this proceartifacts can bimmobilizatiot is fractured ugh the speciical membrane

pointntpoint

Temperature

itical point dryi

nm) by sputtand localizes

rons and contr

t possible dired for water.

es and leukocyte

ess since it invbe prevented aon is the sameat low tempeimen is a ranes (figure 6.8)

April 2012

25

ing.

ter coating ors the signal inrast for imag-

ectly with the

e.

volves as wellagain by usinge as describederatures (e.g. -ndom process).

2

5

r n -

e

l g d -s

Page 28: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introduction to e

26

Figure 6.8: Frfracture face o

After fracturzen water (palow temperatcedure excluperatures thro

Figure 6.9: Crwith platinum.

electron microsco

Freeze-fracturingof the lipid bilay

ring, the visibiartial freeze dture for imagi

usively involvough the who

ryo-SEM image.

opy

g of biologicalyer. Cyt…Cytos

ility of the ultdrying). The fring in the cry

ves physical ple process of p

e of high-pressu

l specimens. PFsol, the cytosol

trastructure cafracture face oyo scanning elprocesses. Thepreparation an

ure frozen mou

Andres Kaech

F…plasmatic fris cross-fractur

an be enhanceof the specimelectron microse specimen isnd imaging.

use brain tissue

racture face of red in the cell o

ed by sublimaen finally is coscope (figure

not treated c

e, freeze-fractur

Ice

the lipid bilayeon the right.

ation of small oated with a h6.9). Note: Th

chemically an

red, partially fr

Ap

er. EF…Exopla

amounts of thheavy metal la

This preparationd stays at low

reeze dried and

pril 2012

asmatic

he fro-ayer at on pro-w tem-

d coated

Page 29: An Introduction to Electron Microscopy Instrumentation ... · PDF filesten filamen h the electron, therefore, n en the electro the elctrons ely fine tungste ... rO/W Schottk e of electron

Introduction to electron microscopy Andres Kaech April 2012

27

7. Literature

Physical Principles of Electron Microscopy, Ray F. Egerton, Springer Verlag, 2007.

Griffith G. (1993). Fine Structure Immunocytochemistry. New York, Berlin, Heidelberg. Springer Verlag. ISBN 0-387-54805-X.

Electron microscopy methods and protocols / ed. by M.A. Nasser Hajibagheri. - Totowa, N.J. : Humana Press, cop. 1999. (Methods in molecular biology ; vol. 117)

Electron microscopy : methods and protocols. - 2nd ed. / ed. by John Kuo - Totowa, N.J. : Humana Press, 2007. (Methods in molecular biology ; 369)

Electron microscopy : principles and techniques for biologists / John J. Bozzola, Lonnie D. Russell. - Boston : Jones and Bartlett, 1991. (The Jones and Bartlett series in biology)

Funktionelle Ultrastruktur : Atlas der Biologie und Pathologie von Geweben / Margit Pavelka, Jürgen Roth. - Wien : Springer, 2005.

Cavalier A., Spehner D., Humbel B. M. (2008). Handbook of Cryo-Preparation Methods for Electron Microsco-py, CRC Press, Boca Raton

Harris, J. R (1997). Negative Staining and Cryoelectron Microscopy. Oxford, BIOS Scientific Publishers Lim-ited

Echlin P. (1992). Low-Temperature Microscopy and Analysis. New York, Plenum Press. ISBN 0-306-43984-0.

Robards A. W., Sleytr U. B (1985). Low Temperature Methods in Biological Electron Microscopy. Practical Methods in Electron Microscopy, Volume 10. Edited by Glauert A. M. Amsterdam, Elsevier Science Publishers B. V. (Biomedical Division)

Steinbrecht R. A. and Zierold K. Cryotechniques in Biological Electron Microscopy (1987). New York, Berlin, Heidelberg. Springer-Verlag. ISBN 0-387-18046-X

http://www.matter.org.uk/tem/default.htm