American Ecotech VOC PAMS Webinar
-
Upload
ae2007 -
Category
Environment
-
view
50 -
download
2
Transcript of American Ecotech VOC PAMS Webinar
![Page 1: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/1.jpg)
Tips on How to Simplify VOC Monitoring and Set Up a Successful Network
![Page 2: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/2.jpg)
Since 2002 American Ecotech has been serving industrial and government customers throughout the USA
• Systems integration & deployment
• Onsite operational training
• Remote maintenance support
Specializing in :
Rack-mountable VOC analyzers designed for PAMS Fenceline monitoring systems
All other types of ambient air monitoring instruments Shelters, trailers, and all support infrastructure associated with ambient air monitoring turn-key systems
![Page 3: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/3.jpg)
VOC Monitoring
•Methane Non-Methane•BTEX•C6 – C12•PAMS
![Page 4: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/4.jpg)
What will you learn from this Webinar
1. What is PAMS?
2. GC pre-concentrator management, maintenance and functions
3. Common mistakes when setting up a calibration system
4. Getting Data from the GC Monitors
![Page 5: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/5.jpg)
Polling Question
What application of VOC monitoring do you most have a need for ?
BTEX
PAMS
C6 – C12
Methane Non-Methane
Total VOC’s
Other
![Page 6: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/6.jpg)
An Overview and Brief History of PAMS
Photochemical Assessment Monitoring Stations
![Page 7: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/7.jpg)
VOC Measurement Technologies
Canisters vs Auto-GCs
• Data averaged over sampling period• Low Capital Cost• Reoccuring lab costs• Need for manual labor
• Hourly data• Higher Capital costs• Higher skill level required to run the system
![Page 8: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/8.jpg)
About PAMSnew auto-GCs have been developed thatprovide near real-time data, designed foruse in ambient air monitoring stations
![Page 9: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/9.jpg)
About PAMSPhotochemical Assessment Monitoring Stations
US Environmental Protection Agency (EPA)
• Revised Ambient AQ Surveillance Regulations
• Effective February 12, 1993
In 1998 the EPA commissioned a study using PAMS to determined if there was a change inambient VOCs as a result of the introduction of Reformulated Gas in 1995. The result werepositive showing a reduction in Benzene related the reduction in the fuel.
The revisions required states and local monitoring agencies to establish PhotochemicalAssessment Monitoring Stations (PAMS) in ozone nonattainment areas classified asserious, severe, or extreme
Monitoring agencies are given options to measure VOCs using either an automated GasChromatograph (auto-GC) or collect samples in the field and analyze them in a laboratory.
When the PAMS program was originally implemented, field rugged auto-GCs were notavailable, so many monitoring agencies had to rely upon canister sampling and/or verycomplex laboratory GCs equipped with automatic samplers.
![Page 10: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/10.jpg)
• April 1998 - EPA's Office of Mobile Sources commissioned a contractor to look at PAMS data to investigate if changes in ambient VOC levels were related to the introduction of reformulated gas in 1995.
• The report concluded that there is a strong case that a reduction in ambient benzene was related to the reduction of benzene in fuel and that other species, such as aromatics, may also have been significantly reduced.
• These findings are consistent with observations in EPA's "National Air Quality and Emissions Trends Reports" for both 1995 and 1996.
Current design calls for up to 5 sites in each PAMS area
Type 1 UpwindType 2 Max emissionsType 3 Max ozoneType 4 Extreme Downwind
•PAMS Season June-August
![Page 11: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/11.jpg)
SIP Control Strategy Evaluation: Evaluation of the effectiveness of State
implemented plans control strategies.
Emissions Tracking:• Corroboration of NOx and VOC inventories Trends of VOC species source profiles• Analysis of air toxics.
Ambient Trend Appraisals:
• O3• NOx• Total and speculated VOC• Adjustments in meteorological conditions.
Exposure Assessment: Estimation of risk levels and the size of effected populations.
The principal objectives of the PAMS program can be summarized as follows:
NAAQS Attainment and Control Strategy Development:• Attainment / Nonattainment determinations• Assessment of the relative contributions of local and upwind sources• boundary conditions for photochemical modeling
![Page 12: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/12.jpg)
What Does a PAMS Consist of?
![Page 13: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/13.jpg)
A PAMS Monitoring System
![Page 14: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/14.jpg)
A PAMS Monitoring System
![Page 15: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/15.jpg)
A PAMS Monitoring System
![Page 16: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/16.jpg)
A PAMS Monitoring System
![Page 17: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/17.jpg)
A PAMS Monitoring System
![Page 18: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/18.jpg)
A PAMS Monitoring System
![Page 19: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/19.jpg)
Pre - Concentrator Operation and Maintenance
![Page 20: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/20.jpg)
The purpose of the pre concentrator is to sample the ambient and get the most evenand consistent sample. It helps to evens out the air when spikes or other air relatedevent might arise. An example is that there is a spike of Methane when the moreconsistent air is mixing with it and averaging the air for that hour.
The Pre Concentrator works on a three hour sampling cycleThe main purpose of the Pre Concentrator is to:
• Heat the air • Creating a sample that is an average of the hour of sampling
![Page 21: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/21.jpg)
At the end of hour 1 of sampling tube #1 opens and releases the cumulative sampleTo the GC
The ambient air collected in hour 2 begins to flow into tube #2 for heating
![Page 22: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/22.jpg)
Pre Concentrator Operation and Maintenance
The same process that took place in tube#1 in Hour 2 happens in Tube #2
![Page 23: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/23.jpg)
Components of AQM-PAMS-Alert
Pre-Concentrator – diagram 1
![Page 24: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/24.jpg)
Components of AQM-PAMS-Alert
Pre-Concentrator – diagram 2
![Page 25: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/25.jpg)
Components of AQM-PAMS-Alert
Pre-Concentrator – diagram 3
![Page 26: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/26.jpg)
Components of AQM-PAMS-Alert
Pre-Concentrator – diagram 4
![Page 27: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/27.jpg)
Components of AQM-PAMS-Alert
Pre-Concentrator - diagram 5
![Page 28: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/28.jpg)
Components of AQM-PAMS-Alert
Pre-Concentrator - diagram 6
![Page 29: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/29.jpg)
Calibration
Take the time warm up the system ( 30 mins / 1 hour )
Match up the proper gases
Most calibrations can be done with a single point
Pay attention to Column efficiency
![Page 30: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/30.jpg)
FID Chromatogram
![Page 31: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/31.jpg)
FID Analysis
• Ethane
• Ethylene
• Propane
• Propylene
• Isobutane
• N-butane
• Trans-2-butane
• 1-butane
• Cyclopentane
• Cis-2-butene
• Isopentane• N-pentane• Tran-2-Pentane• 1-Pentane• Cis-2-Pentane• 2,2-Dimethylbutane• 2,3-Dimethylbutane• 2-/3-
Methylpentane• Isoprene
Peaks listed in elution order
![Page 32: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/32.jpg)
PID Chromatogram
![Page 33: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/33.jpg)
PID Analysis
• N-Hexane
• Methylcyclopentane
• 2,4-Dimethylpentane
• Benzene
• Cyclohexane
• 2-Methylhexane
• 3-Methylhexane
• 2,2,4-Trimethylpentane
• N-Heptane
• Methylcyclohexane
• 2,3,4-Trimethylpentane
• Toluene
• 2-Methylheptane• 3-Methylheptane• N-octane• Ethylbenzene• M-/P-Xylene• Styrene• O-Xylene• N-Nonane• Isopropylbenzene• N-Propylbenzene• M-Ethyltoluene• P-Ethyltoluene• 1,3,5-
Trimethylbenzene
• O-Ethyltoluene• 1,2,4-
Trimethylbenzene• N-Decane• 1,2,3-
Trimethylbenzene• M-Diethylbenzene• P-Diethylebenzene• N-Undecane• N-Dodecane
Peaks listed in elution order
![Page 34: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/34.jpg)
Getting data from the GC Monitor
Sample takes an hour to gather
The air that is gathered in Hour 1 is available in the second hour
![Page 35: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/35.jpg)
Field Deployment Options
![Page 36: American Ecotech VOC PAMS Webinar](https://reader033.fdocuments.net/reader033/viewer/2022042615/55d0a3d9bb61eb6b498b45c0/html5/thumbnails/36.jpg)
Instrumentation
• Gas Chromatographs for VOC PAMS
• Hydrocarbon Analyzers including NMHC
• Small VOC Sensors
• Fenceline networked sensor systems
• Hand Held VOC Detectors
• Criteria Pollutant Analyzers
• Particulate Samplers (PM2.5, PM10)
• Nephelometers
• Pb Samplers
• Shelters/Trailers/Mobile Vans
• Dataloggers
• Data Validation/Remote Maint software