Agilent OpenLAB Chromatography Data System (CDS) ChemStation Edition Supported Instruments and Firmware Guide

Transcript of Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC...

Page 1: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

Agilent OpenLAB

Chromatography Data

System (CDS)

ChemStation Edition

Supported Instruments and

Firmware Guide

Page 2: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

2 Supported Instruments and Firmware Guide

Notices

© Agilent Technologies. 2010, 2011- 2016

No part of this manual may be reproduced

in any form or by any means (including

electronic storage and retrieval or

translation into a foreign language) without

prior agreement and written consent from

Agilent Technologies, Inc. as governed by

United States and international copyright

laws.

Manual Part Number

M8301-90018

Edition

11/2016

Agilent Technologies, Inc.

Software Revision This guide is valid for Agilent OpenLAB CDS

ChemStation Edition Version C.01.07 SR3

Warranty

The material contained in this document is

provided “as is,” and is subject to being

changed, without notice, in future editions.

Further, to the maximum extent permitted by

applicable law, Agilent disclaims all

warranties, either express or implied, with

regard to this manual and any information

contained herein, including but not limited to

the implied warranties of merchantability and

fitness for a particular purpose. Agilent shall

not be liable for errors or for incidental or

consequential damages in connection with

the furnishing, use, or performance of this

document or of any information contained

herein. Should Agilent and the user have a

separate written agreement with warranty

terms covering the material in this document

that conflict with these terms, the warranty

terms in the separate agreement shall

control.

Technology Licenses

The hardware and/or software described in

this document are furnished under a license

and may be used or copied only in accordance

with the terms of such license.

Restricted Rights Legend

If software is for use in the performance of a

U.S. Government prime contractor subcontract,

Software is delivered and licensed as

“Commercial computer software” as defined in

DFAR 252.227-7014 (June 1995), or as a

“commercial item” as defined in FAR 2.101(a)

or as “Restricted computer software” as

defined in FAR 52.227-19 (June 1987) or any

equivalent agency regulation or contract

clause. Use, duplication or disclosure of

Software is subject to Agilent Technologies’

standard commercial license terms, and non-

DOD Departments and Agencies of the U.S.

Government will receive no greater than

Restricted Rights as defined in FAR 52.227-

19(c)(1-2) (June 1987). U.S. Government users

will receive no greater than Limited Rights as

defined in FAR 52.227-14 (June 1987) or DFAR

252.227-7015 (b)(2) (November 1995), as

applicable in any technical data.

Safety Notices

A CAUTION notice denotes a hazard.

It calls attention to an operating

procedure, practice, or the like that, if

not correctly performed or adhered to,

could result in damage to the product

or loss of important data. Do not

proceed beyond a CAUTION notice

until the indicated conditions are fully

understood and met.

A WARNING notice denotes a

hazard. It calls attention to an

operating procedure, practice, or the

like that, if not correctly performed

or adhered to, could result in

personal injury or death. Do not

proceed beyond a WARNING notice

until the indicated conditions are

fully understood and met.

CAUTION

WARNING

Page 3: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

Contents

CDS Supported Instruments and Firmware Guide 3

Contents

OpenLAB CDS Driver Information 5

Minimum required Firmware 5

Operating Systems 5

Driver types 5

Instrument Communication Information – OpenLAB CDS ChemStation Editions 7

Instrument LAN Interface Card 7

GPIB Compatibility 7

Agilent HPLC/LC Instrument and Firmware Compatibility 8

Support Statement 8

Minimum required Firmware for Agilent HPLC/LC Modules 9

Agilent LC Sampling Systems 9

CTC PAL Autosamplers (Agilent and CTC) with LC 9

Agilent LC – Pumps 10

Agilent LC – Column Compartments 11

Agilent LC – Detectors 11

Agilent LC – Valves, Valve Drives and Clusters 12

Agilent LC – Other Module types 13

Agilent 218/SD1 PrepLC 14

Agilent 1120 Compact LC Systems 14

Agilent 1220 Infinity LC Systems 14

Supported Agilent LC/MS 15

Supported Agilent CE 16

Agilent Capillary Electrophoresis Driver 16

Agilent Analog-Digital Converter Information 17

Agilent A/D Converter 17

Agilent GC Instrument Compatibility 18

Agilent GC support 18

GC Sampler, Autosampler and Tray Compatibility 19

Page 4: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

Contents

4 Supported Instruments and Firmware Guide

6850 ALS hardware 19

7650 ALS hardware: 19

7683B ALS hardware 19

7683A ALS hardware: 19

7693 ALS hardware 20

Headspace support with GC 20

Control of PAL sampler with GC 20

GC hardware not supported by the current CDS revision 21

Waters Drivers 22

Waters Alliance 22

Waters Acquity 22

OpenLAB CDS ChemStation Edition VL 24

Overview 24

OpenLAB CDS ChemStation Edition VL rev. C.01.07 SR3 24

Supported Configurations for OpenLAB CDS VL Systems 24

A. Control of one Agilent GC 24

B. Control of one 1260 Infinity LC instrument, using instrument type “Agilent LC

Core System”. 25

C. Control of one 1120 Compact LC, or one 1220 Infinity LC system using

Instrument type “Agilent 1220 LC System”. 25

Classic Driver Support Information 27

Agilent HPLC – Sampling Systems 27

Agilent HPLC – Pumps 27

Agilent HPLC – Detectors 28

Agilent HPLC – Column Compartments 28

Agilent HPLC – Quick Change Valves 28

Agilent 1200 Infinity Series LC – Other Module types 29

Agilent A/D Converter 29

In This Book 30

Page 5: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

OpenLAB CDS Driver Information

CDS Supported Instruments and Firmware Guide 5

OpenLAB CDS Driver Information

Minimum required Firmware

Method and Run Control

LC Instruments Instrument > Revisions & Serial#s.

GC Instruments Instrument > Instrument Configuration

Selected Modules configure

NOTE The “minimum required firmware” specified allows operation of the instrument with the data

system, but preserves the minimal functionality of the stated firmware. Not all functionality

introduced by more recent releases of instrument drivers and the associated firmware is

available with previous versions. A Firmware upgrade should be done if you face problems or

want to add system functionality to your system.

Operating Systems

Driver types

Page 6: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

OpenLAB CDS Driver Information

6 Supported Instruments and Firmware Guide

Page 7: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

CDS Supported Instruments and Firmware Guide 7

Instrument Communication Information – OpenLAB CDS

ChemStation Editions

NOTE On Agilent Instrument Controllers (AICs), OpenLAB CDS ChemStation Edition exclusively

supports instruments and modules connected via LAN. GPIB, RS232, USB or any converter

are not supported on AICs.

Instrument LAN Interface Card

Product Hardware Minimum required

Firmware Revision

Notes

G1369C LAN Interface

card

B.06.40 CAN support for hosted modules/CAN slave (e.g. G1170A

Universal Valve Drive)

SW 7 and SW 8 must be in the OFF position

10/100BaseT

G1369B LAN Interface

card

A.01.01 SW 7 and SW 8 must be in the OFF position

10/100BaseT

G1369A LAN Interface

card

A.01.10 2nd version – Board revision C.03.00 – firmware A.01.10 –

Supported with 35900E and GC

1st version – firmware A.01.05 – Not Supported with GC and

35900E

GPIB Compatibility

Product Minimum required

Firmware Revision

Support Statement / Comments

82350B IO Library Suite 16.3

Update 3

ChemStation LC/MS.

Not supported for GC instruments

Page 8: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

8 Supported Instruments and Firmware Guide

Agilent HPLC/LC Instrument and Firmware Compatibility

Support Statement

NOTE LC and CE Firmware revisions are grouped into sets for each module or system. Firmware sets

include just the latest firmware of each module. Modules combined into one LC instrument

always need to have firmware from the same set. Firmware of one set is fully compatible with

your CDS that supports this firmware set. However, if one module uses a firmware revision

lower that the specified Minimum Firmware Revision some functionality might not be

supported.

Do not mix firmware revisions from one set with older or newer sets. Firmware is not tested

across set borders!

NOTE The “minimum required firmware” specified is tested to allow operation of the respective

instrument with the data system, but preserves the minimal functionality of the stated

firmware.

Not all functionality introduced by more recent releases of instrument drivers and the

associated firmware is available with previous versions.

Page 9: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

Agilent HPLC/LC driver information for Agilent OpenLAB CDS ChemStation Edition

CDS Supported Instruments and Firmware Guide 9

Minimum required Firmware for Agilent HPLC/LC Modules

Agilent LC Sampling Systems

Product

Number

Module Name Minimum required Firmware

Revision

G1313A 1100 Series Standard Autosampler A.06.10

G1329A 1100 Series Standard Autosampler A.06.10

G1329B 1260 Infinity Standard Autosampler A.06.10

G1367A 1100 Series Well-Plate Sampler A.06.31

G1367B 1200 Series High Performance Autosampler A.06.31

G1367C 1200 Series High Performance Autosampler SL A.06.31

G1367D 1200 Series High Performance Autosampler SL+ A.06.31

G1367E 1260 Infinity High Performance Autosampler A.06.32

G1373A Purge Kit A.06.56

G1377A 1260 Infinity High Performance Micro Autosampler A.06.12

G1389A 1100 Series Micro Thermostatted Autosampler A.06.10

G2256A 1260 Infinity Sample Capacity Extension n/a

G2258A 1260 Infinity Dual-Loop Autosampler A.06.50

G2260A 1260 Infinity Preparative Autosampler (High flow) A.06.50

G4226A 1290 Infinity Autosampler A.06.31

G4303A 1260 Infinity SFC standard autosampler A.06.54

G5667A 1260 Infinity Bio-inert Autosampler A.06.32

G5668A 1260 Infinity II Bio-Inert Multisampler D.07.01

G7129A 1260 Infinity Autosampler D.06.76

G7129B 1290 Infinity II Vial Sampler D.06.76

G7167A 1260 Infinity Multisampler D.06.75

G7167B 1290 Infinity II Multisampler D.06.75

CTC PAL Autosamplers (Agilent and CTC) with LC

Product Module Name Minimum required Revision

Number CTC Firmware Software

G4277A Agilent 1290 Infinity LC Injector HTS 4.x.x B.01.08

G4278A Agilent 1290 Infinity LC Injector HTC 4.x.x B.01.08

G4270-CTC CTC Analytics HTC PAL Injection System1 2.5.2 (for legacy mainboard)

2.6.8 (for legacy mainboard);

4.x.x (for new XT mainboard);

B.01.08

Cycle Parameters in

English G4271-CTC CTC Analytics HTS PAL Injection System1

Page 10: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

Agilent HPLC/LC driver information for Agilent OpenLAB CDS ChemStation Edition

10 Supported Instruments and Firmware Guide

Agilent LC – Pumps

Product

Number

Module Name Minimum required Firmware

Revision

G1310A 1100 Series Isocratic Pump A.06.10

G1310B 1260 Infinity Isocratic Pump A.06.32

G1311A 1100 Series Quaternary Pump 1 A.06.10

G1311B 1260 Infinity Quaternary Pump 1 A.06.32

G1311C 1260 Infinity Quaternary Pump VL 1 A.06.32

G7104A 1290 Infinity II Flexible Pump B.06.71

G7110B 1260 Infinity II Isocratic Pump D.07.01

G7111A 1260 Infinity II Quaternary Pump VL 1 D.07.01

G7111B 1260 Infinity II Quaternary Pump 1 D.07.01

G7112B 1260 Infinity II Binary Pump 1 D.07.01

G7120A 1290 Infinity II High Speed Pump B.06.71

G1312A 1260 Infinity Binary Pump 1 A.06.10

G1312B 1260 Infinity Binary Pump SL1 A.06.10

G1312C 1260 Infinity Binary Pump VL1 A.06.32

G1361A 1260 Infinity Preparative Pump

Cluster with up to 4x G1361A A.06.50

G1376A 1260 Infinity Capillary Pump A.06.10

G2226A 1260 Infinity Nanoflow Pump A.06.10

G4204A 1290 Quaternary Pump 1 B.06.50

G4220A 1290 Infinity Binary Pump 1 B.06.23

G4220B 1290 Infinity Binary Pump VL 1 B.06.43

G4302A 1260 Infinity SFC Binary Pump 1 A.06.10

G5611A 1260 Infinity Bio-inert Quaternary Pump 1 A.06.10

G5654A 1260 Infinity II Bio-inert Pump 1 D.07.01

1 Pump valve clusters are possible for marked pumps with up to 2 valves of type G1160A and/or G1170A

Page 11: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

Agilent HPLC/LC driver information for Agilent OpenLAB CDS ChemStation Edition

CDS Supported Instruments and Firmware Guide 11

Agilent LC – Column Compartments

Product

Number

Module Name Minimum required

Firmware Revision

G1316A 1260 Infinity Column Compartment (Thermostatted) A.06.10

G1316B 1200 Series Column Compartment SL A.06.10

G1316C 1290 Infinity Column Compartment (Thermostatted) 2 A.06.14

G7116A 1260 Infinity II Multicolumn Thermostat (MCT)

(firmware for host module in brackets)

C.07.01

(B.07.01/D.07.01)

G7116B 1290 Infinity II Multicolumn Thermostat (MCT)

(firmware for host module in brackets)

C.06.75

(B.06.75/D.06.75)

G7130A Integrated Column Compartment ICC (option for G7129 A/B) D.06.76

Valve Thermostat Cluster with multiples of these modules: A combination of G7116B, G1170A

and G1316C as valve or column hosts and G1316A/B and G7130A as column host C.06.75 (B.06.75/D.06.75)

2 Cluster with up to three G1316C with integrated 8pos/9port valves (products G4230A/B). Minimum two G1316C TCCs. The third TCC can

be a G1316A, B or C.

Agilent LC – Detectors

Product

Number

Module Name Minimum required

Firmware Revision

VWD

G1314A 1100 Series Variable Wavelength Detector A.06.10

G1314B 1200 Series Variable Wavelength Detector A.06.10

G1314C 1200 Series Variable Wavelength Detector A.06.10

G1314D 1200 Series Variable Wavelength Detector B.06.32

G1314E 1290 Infinity Variable Wavelength Detector B.06.32

G1314F 1260 Infinity Variable Wavelength Detector B.06.32

G7114A 1260 Infinity II Variable Wavelength Detector D.07.01

G7114B 1290 Infinity II Variable Wavelength Detector D.06.70

MWD

G1365A 1100 Series Multiple Wavelength Detector A.06.10

G1365B 1100 Series Multiple Wavelength Detector A.06.10

G1365C 1260 Infinity Multiple Wavelength Detector B.06.30

G1365D 1260 Infinity Multiple Wavelength Detector VL B.06.30

G7165A 1260 Infinity II MWD D.07.01

FLD

G1321A 1100 Series Fluorescence Detector (FLD) A.06.10

G1321B 1260 Infinity Fluorescence Detector A.06.32

G1321C 1260 Infinity Fluorescence Detector A.06.54

G7121A 1260 Infinity II Fluorescence Detector D.07.01

G7121B 1260 Infinity II Fluorescence Detector Spectra D.07.01

Page 12: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

Agilent HPLC/LC driver information for Agilent OpenLAB CDS ChemStation Edition

12 Supported Instruments and Firmware Guide

DAD

G1315A 1100 Series Diode Array Detector A.06.10

G1315B 1200 Series Diode Array Detector A.06.10

G1315C 1200 Series Diode Array Detector VL+ B.06.30

G1315D 1200 Series Diode Array Detector VL B.06.30

G4212A 1290 Infinity Diode Array Detector B.06.30

G4212B 1260 Infinity Diode Array Detector B.06.30

G7115A 1260 Infinity II DAD WR D.07.01

G7117A 1290 Infinity II Diode Array Detector (with fixed slit) D.06.70

G7117B 1290 Infinity II Diode Array Detector (with variable slit) D.06.70

G7117C 1260 Infinity II Diode Array Detector (with fixed slit) D.07.01

HDR-DAD Cluster 2x G4212A, 2x G4212B, or a combination of 1x G4212A and 1x G4212B;

or G7117A/B Cluster

B.06.70

RID

G1362A 1260 Infinity Refractive Index Detector A.06.10

G7162A 1260 Infinity II Refractive Index Detector D.06.76

G7162B 1290 Infinity II Refractive Index Detector D.06.76

ELSD

G4218A 1260 Infinity Evaporative Light Scattering Detector 1.4 Classic Plus

(English Driver only) 3

G4260A 1260 Infinity 380 Evaporative Light Scattering Detector V 25.00 3

G4261A 1290 Infinity 385 Evaporative Light Scattering Detector V 25.00 3

G4260B 1260 Infinity Evaporative Light Scattering Detector V 31.06 3

G4261B 1290 Infinity Evaporative Light Scattering Detector V31.06 3

G7102A 1290 Infinity II Evaporative Light Scattering Detector V 31.06 3

3 English only, RS232 not supported on AIC

Agilent LC – Valves, Valve Drives and Clusters

Product

Number

Module Name Minimum required Firmware Revision

G1156A 6Pos/7Port Valve The G1156A valve is not configurable

in ChemStation, neither Classic, nor

RC.NET. It requires an ALS

G1157A 1200 Series 2 Position / 10 Port Valve A.06.02

G1158A 1200 Series 2 Position / 6 Port Valve A.06.02

G1158B 1200 Series 2 Position / 6 Port Valve (600bar) A.06.02

G1159A 1200 Series 6 Position Selection Valve A.06.02

G1160A 1100 Series Multiple Purpose Switching Valve (12Position / 13 Port) A.06.02

G1170A 1290 Universal Actuator (Valve Drive)

(firmware for host module in brackets)

C.06.40

(B.06.40/D.06.60)

G4734A 6-Column-selector: requires valve host- G1170A or G7116A/B

Page 13: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

Agilent HPLC/LC driver information for Agilent OpenLAB CDS ChemStation Edition

CDS Supported Instruments and Firmware Guide 13

Low Flow Valves

G1162A 1200 Series 2 Position/ 6 Port Micro Valve A.06.02

G1163A 1200 Series 2 Position/ 10 Port Micro Valve A.06.02

Valve

Thermostat

Cluster

The Valve Thermostat Cluster is a combination of G7116B, G1170A and G1316C

as valve or column hosts and G1316A/B and G7130A as column hosts.

(firmware for host module in brackets)

C.06.75

(B.06.75/D.06.75)

Agilent LC – Fraction Collectors

Product

Number

Module Name Minimum required Firmware

Revision

G1364A 1100 Series Automatic Fraction Collector 4 A.06.53

G1364B 1260 Infinity Fraction Collector (preparative-scale) 4 A.06.53

G1364C 1260 Infinity Fraction Collector (analytical-scale) 4 A.06.53

G1364D 1100 Series Micro Fraction Collector A.06.53

G5664A 1260 Infinity Bio-inert fraction collector AS 4 A.06.53

G7159B 1290 Infinity II Prep. Open-Bed Fraction Collector D.07.10

G7166A 1260 Infinity II Prep. Valve-based Fraction Collector

(firmware for host module in brackets)

C.07.10

(B.07.10/D.07.10)

Fraction

coll. cluster

Any combination of up to three G1364A/B/C or G5664A plus a fourth G1364A/B/C or G5664A for recovery can be clustered.

Multiple single Fraction Collectors are not supported.

Agilent LC – Other Module types

Product

Number

Module Name Minimum required Firmware

Revision

G1390A 1100 Series Universal Interface Box (UIB) A.06.02

G1390B 1200 Infinity Series Universal Interface Box II

(firmware for host module in brackets)

C.06.53

(B.06.53/D.06.60)

G4227A 1290 Infinity Flexible Cube

(firmware for host module in brackets)

C.06.52

(B.06.52/D.06.60)

G2229A 1260 Infinity Nanoflow LC System:

Includes Nanoflow pump, micro degasser, micro well-plate sampler, handheld

controller (G1377A + G2226A)

A.06.02

G2254A 1260 Automation Interface A.06.02

G4301A Superfluid Chromatograph (SFC) Control Module A.03.08,

Windows 10 PC requires FW

update to A.03.09

G1323B 1100 Series Handheld Control Module n/a

G4208A 1200 Infinity Series Instant Pilot n/a

Page 14: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

Agilent HPLC/LC driver information for Agilent OpenLAB CDS ChemStation Edition

14 Supported Instruments and Firmware Guide

Agilent 218/SD1 PrepLC (formerly Varian Prep LC)

Product

Number

Module Name Communications Minimum required Firmware

Revision

Agilent 210 Pump RS232 1.73 rev C

Agilent 218 Pump RS232 1.1 rev B

Agilent SD1 Pump RS232 2.2

Agilent 325 UV/VIS Detector LAN 2.07

Agilent 410 Autosampler RS232 2.07

Agilent 440 Fraction Collector RS232 1.31

Agilent 1120 Compact LC Systems

Product

Number

Module Name Communications Minimum required Firmware

Revision

G4286A 1120 Compact LC, Isocratic pump TCP/IP B.06.50

G4287A 1120 compact LC. Isocratic with Oven and ALS TCP/IP B.06.50

G4288A 1120 Compact LC, Gradient TCP/IP B.06.50

G4289A 1120 Compact LC, Gradient with oven TCP/IP B.06.50

G4290A 1120 Compact LC, Gradient with oven and ALS TCP/IP B.06.50

Agilent 1220 Infinity LC Systems

Product

Number

Module Name Minimum required Firmware

Revision

G4286B 1220 Infinity LC System Isocratic, Man. Inj., VWD, 600 bar B.06.50

G4287B 1220 Infinity LC Isocratic, ALS, TCC, VWD, 600 bar B.06.50

G4288B 1220 Infinity LC Gradient, Man. Inj., VWD, 600 bar B.06.50

G4289B 1220 Infinity LC Gradient, ALS, TCC, VWD, 600 bar B.06.50

G4290B 1220 Infinity LC Gradient, ALS, Man. Inj., TCC, VWD, 600 bar B.06.50

G4291B 1220 Infinity LC Isocratic, Man. Inj., TCC, VWD, 600 bar B.06.50

G4292B 1220 Infinity LC Isocratic, ALS, VWD, 600 bar B.06.50

G4293B 1220 Infinity LC Gradient, ALS, VWD, 600 bar B.06.50

G4294B 1220 Infinity LC Gradient, ALS, TCC, DAD, 600 bar B.06.50

G4288C 1220 Infinity LC System VL, Gradient, Man. Inj. VWD, 400 bar B.06.50

G4289C 1220 Infinity LC System VL, Gradient, TCC, Man. Inj. VWD, 400 bar B.06.50

G4290C 1220 Infinity LC System VL, Gradient, ALS, TCC, VWD, 400 bar B.06.50

G4293C 1220 Infinity LC System VL, Gradient, ALS, VWD, 400 bar B.06.50

G4294B 1220 LC Infinity Gradient System B.06.50

Page 15: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

Agilent CE driver information for Agilent OpenLAB CDS ChemStation

CDS Supported Instruments and Firmware Guide 15

Supported Agilent LC/MS

Model Module Type Supported Firmware Supported Driver Type Communications

G6110A 6100 Series Single Quadrupole

LC/MS System

3.02.48 Classic Plus LAN

G6120A

G6120B

6100 Series Single Quadrupole

LC/MS System

3.02.48 Classic Plus LAN

G6130A

G6130B

6100 Series Single Quadrupole

LC/MS System

3.02.48 Classic Plus LAN

G6140A 6100 Series Single Quadrupole

LC/MS System

3.02.48 Classic Plus LAN

G6150B 6100 Series Single Quadrupole

LC/MS System

3.02.48 Classic Plus LAN

G7100 CE/MS 5 B.06.26 Only when configured as CE-MS system,

workstation and networked workstation only.

4 If Agilent LC/MS is controlled in conjunction with the Agilent 3D Capillary Electrophoresis system, no additional instrument is supported on

that PC system

NOTE

G6110A, G6120A/B, G6130A/B, G6140A, and G6150B are not supported on Russian

ChemStation

Page 16: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

Agilent LC/MS driver information for Agilent OpenLAB CDS ChemStation Edition

16 Supported Instruments and Firmware Guide

Supported Agilent CE

NOTE Configure using Instrument Type “Agilent 7100 CE System”.

Agilent Capillary Electrophoresis Driver

Product

Number

Module Name Minimum required Firmware Revision

G7150A G7100 Capillary Electrophoresis II B.07.01 or higher

Agilent G7100 CE/MS2 B.06.26

G7151A Diode Array Detector for CE II B.06.26

35900E A/D converter (LAN only!) E.01.02

Special Notes:

Page 17: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

Agilent A/D information for Agilent OpenLAB CDS

CDS Supported Instruments and Firmware Guide 17

Agilent Analog-Digital Converter Information

Agilent A/D Converter

Instrument Supported Firmware Version

(minimum)

Communications

Agilent 35900E Dual Channel Interface E.01.02 TCP/IP , LAN only

35900E Series II E.02.04.40

3rd and 4th Signal, supported configurations:

Instrument 35900E Classic Driver 35900E RC.NET Driver

7890A

7820A

Not supported yes

68xx RC.net Not supported Yes

68xx classic Not supported Not supported

LC classic Yes Not supported

LC RC.net Yes Yes

CE classic Yes Not supported

CE RC.net Not supported Yes

Page 18: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

Agilent GC driver information for Agilent OpenLAB CDS

18 Supported Instruments and Firmware Guide

Agilent GC Instrument Compatibility

NOTE Communication is handled via LAN

Agilent GC support

Product Number Description Tested Firmware

Notes

G3950A

G3952A

G3953A

Intuvo 9000 GC system A.01.01 7683 ALS family not supported

G3440A

G3445A

7890A A.01.16

G3440B

G3445B

7890B B.02.04.3

G4350A 7820A A.01.17.004 Backflush and Easy SamplePrep are not

supported with 7820

G1530N

G1540N

6890N N.06.07 manual inlets and manual detectors are not

supported

G1530A

G1540A

6890A

6890Plus

A.03.08 manual inlets and manual detectors are not

supported

G2629A A.05.06

G2630A 6850 serial #<= US00003200

6850 serial # >= US10243001

A.06.02

A.03.07

manual inlets and manual detectors are not

supported

G3581A 490 Micro GC 4.01 TCP-IP.

For legacy mainboard: minimum required FW

is 1.40

G3582A 490 Micro GC Analyzer 4.01 TCP-IP

Page 19: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

Agilent GC driver information for Agilent OpenLAB CDS

CDS Supported Instruments and Firmware Guide 19

GC Sampler, Autosampler and Tray Compatibility

6850 ALS hardware

Instrument Model Firmware Notes

Injector G2880A A.10.05

7650 ALS hardware:

Instrument Model Firmware Notes

Injector G4567A A.10.02

7683B ALS hardware

Instrument Model Firmware Notes

Controller G2912A A.02.01 7683B ALS Controller for 6890

7683B Injector G2913A A.11.03

Tray G2614A A.02.01

BCR/Mixer G2615A n/a

7683A ALS hardware:

Instrument Model Firmware Notes

7683A Injector G2613A A.10.07

Tray G2614A A.02.01

BCR/Mixer G2615A n/a

ALS Controller G2612A A.02.01 7683/6890 ALS interface board

Page 20: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

Agilent GC driver information for Agilent OpenLAB CDS

20 Supported Instruments and Firmware Guide

7693 ALS hardware

Instrument Model Firmware Notes

7693 Injector G4513A A.10.09

7693 Tray G4514A A.11.01

BCR/Mixer G4515A A.10.05

ALS Controller Board G4516A A.01.07 6890 Plus only

ALS Controller G4517A A.01.07 6890A only

Headspace support with GC

NOTE Communication is handled via LAN.

Instrument Model Firmware Software Notes

7697A Headspace G4556A,

G4557A A.01.07.1 B.01.07.2 License number no longer required for

installation.

Note: The software registration packet will

still include a license number. G1888 Headspace G1888A A.01.10 B.01.07.2

Control of PAL sampler with GC

Instrument Model Firmware Software Notes

Agilent PAL-xt

CTC Sampler

G6501B,

G6502B,

G6509B

4.3.0 B.01.08

PAL-xt method parameters are available

to Intelligent Reporting beginning in PAL

Software Rev. B.01.07.

Note: License number no longer required

for installation. The software registration

packet will still include a license number.

Agilent PAL-3

CTC Autosampler

G7366A,

G7367A,

G7368A,

G7370A

2.3.1 A.01.03

The CTC PAL-3 method parameters are

available to Intelligent Reporting since PAL

Software rev. A.01.01.

Note: License number no longer required

for installation. The software registration

packet will still include a license number.

Page 21: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

Agilent GC driver information for Agilent OpenLAB CDS

CDS Supported Instruments and Firmware Guide 21

GC hardware not supported by the current CDS revision

Instrument Notes

5890A

5890 Series II, 4890D The 5890 driver is not tested. There will be no new

updates or hotfixes for the 5890 GC driver.

35900C/D

19405A/B Sampler Event Control Module (SECM)

G1289B,

G1290B, (7694B);

G1883A (7694E) Headspace

The 7694B headspace was last supported with

OpenLAB CDS C.01.05 with the Headspace software

rev B.01.04.

7673A family (18593A, 18594A, 18596A)

7673B family (18593B, 18594B, 18596B)

7673C family (G1512A, G1513A, 18596C)

G2403A LAN/RS232 converter

G1176A, G1180A (6820 GC)

G1926A

3000 micro GC, M200, M400

G7360A Archon Purge and Trap Auto sampler for Water only

G7361A The Archon Purge and Trap Auto sampler for Needle Sparge

WorkBench Tray (G8130A)

Page 22: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

Non-Agilent GC Devices

22 Supported Instruments and Firmware Guide

Waters Drivers

Waters Alliance

M8239AA Waters Alliance Control for OpenLAB CDS

Module Type Module Names Communications Supported FW

Version

Integrated LC (Auto

sampler + Pump)

Waters Alliance 2690 GPIB 1.20

Waters Alliance 2695 GPIB 2.04

Waters Alliance e2695 Ethernet / GPIB 3.00

UV Detector Waters UV 2487 GPIB 1.01 or 1.02

Waters UV 2489 Ethernet 1.41.2272

PDA Detector

Waters PDA 996 GPIB 1.8

Waters PDA 2996 GPIB R3.5

Waters PDA 2998 Ethernet 1.41.6164

Waters Acquity

M8505AA Waters Acquity Control for OpenLAB CDS

Modules Specific to Waters Acquity

Module Communications Min. Firmware Version

Sample Manager LAN 1.60

Sample Organizer LAN 1.60

Binary Solvent Manager LAN 1.60

Column Manager (CM) LAN 1.60.74

Modules Specific to Waters Acquity H-Class

Module Communications Firmware Version

Sample Manager FTN LAN 1.60

Sample Organizer LAN 1.60

Quaternary Solvent Manager LAN 1.60

Column Manager (CM-A) LAN 1.60

Column Manager Aux LAN N/A

Page 23: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

Non-Agilent Devices no longer supported with this OpenLAB CDS revision

CDS Supported Instruments and Firmware Guide 23

Modules for Waters Acquity and Waters Acquity H-Class

Module Communications Firmware Version

TUV Detector LAN 1.60

PDA Detector LAN 1.60

PDA Detector eLambda LAN 1.60

Fluorescence Detector LAN 1.42

ELS Detector LAN 1.30

Page 24: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

OpenLAB CDS VL – Support Information

24 Supported Instruments and Firmware Guide

OpenLAB CDS ChemStation Edition VL

Overview

OpenLAB CDS ChemStation Edition VL rev. C.01.07 SR3

Allows control of one of the below:

One 1260 Isocratic Infinity LC system, or one 1260 Quaternary Infinity LC system with selected 1260 Infinity LC modules (use instrument type “Agilent LC Core System”).

One Agilent 1220 Infinity LC system or one 1120 Compact LC system (use “Agilent 1220 LC System”). Control of supplemental Infinity LC modules is supported for all LC modules with the exception of modular pumps.

One 7820A GC, or one 490 Micro GC.

Supported Configurations for OpenLAB CDS VL Systems

A. Control of one Agilent GC

Table 1 Control supported with OpenLAB CDS VL ChemStation Edition

Module Name Product Number

7820A G4350B

Agilent 490 Micro GC G3581A

Page 25: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

OpenLAB CDS VL – Support Information

CDS Supported Instruments and Firmware Guide 25

B. Control of one 1260 Infinity LC instrument,

use instrument type “Agilent LC Core System”.

Table 3 Agilent LC Core System modules

Module Name Product Number

1260 Isocratic Pump G1310B

1260 Quaternary Pump G1311B,

1260 Quaternary Pump VL G1311C

1260 Column Compartment G1316A

1260 Autosampler G1328C, G1329B (may add G1330B Thermostat),

1260 Infinity II Autosampler G7129A (including ICC: G7130A)

1260 Diode Array Detectors G1315C, G1315D, G4212B

1260 Multiple Wavelength Det. G1365C, G1365D

1260 Variable Wavelength Det. G1314B, G1314C, G1314F

1260 RID G1362A, G7162A

1260 FLD G1321B, G1321C

1200 UIB II G1390B

C. Control of one 1120 Compact LC, or one 1220 Infinity LC system

use Instrument type “Agilent 1220 LC System”.

Table 4 OpenLAB CDS VL compatibility with Agilent 1120 Compact LC

Product Number Module Type

G4286A 1120 Compact LC, Isocratic

G4287A 1120 Compact LC Isocratic with Oven and ALS

G4288A 1120 Compact LC, Gradient

G4289A 1120 Compact LC, Gradient with oven

G4290A 1120 Compact LC Gradient with oven and ALS

G1321A 1100/1200 FLD

G1362A 1100/1200 RID

G4218A 1100/1200 ELSD

G4260A ELSD 380 Detector

G4261A ELSD 385 Detector

G4260B ELSD 1260 Infinity Detector

G4261B ELSD 1290 Detector

Page 26: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

OpenLAB CDS VL – Support Information

26 Supported Instruments and Firmware Guide

Table 5 OpenLAB CDS VL compatibility with Agilent 1220 Infinity LC Systems

Product Number Module Name

G4286B 1220 Infinity LC - Isocratic, Man. Inj., VWD, 600 bar

G4287B 1220 Infinity LC - Isocratic, ALS, TCC, VWD, 600 bar

G4288B 1220 Infinity LC - Gradient, Man. Inj., VWD, 600 bar

G4289B 1220 Infinity LC - Gradient, ALS, TCC, VWD, 600 bar

G4290B 1220 Infinity LC - Gradient, ALS, Man. Inj., TCC, VWD, 600 bar

G4291B 1220 Infinity LC - Isocratic, Man. Inj., TCC, VWD, 600 bar

G4292B 1220 Infinity LC - Isocratic, ALS, VWD, 600 bar

G4293B 1220 Infinity LC - Gradient, ALS, VWD, 600 bar

G4294B 1220 Infinity LC - Gradient, ALS, TCC, DAD, 600 bar

G4288C 1220 Infinity LC - System VL, Gradient, Man. Inj. VWD, 400 bar

G4289C 1220 Infinity LC - System VL, Gradient, Man. Inj. VWD, 400 bar

G4290C 1220 Infinity LC - System VL, Gradient, ALS, TCC, VWD, 400 bar

G4293C 1220 Infinity LC - System VL, Gradient, ALS, VWD, 400 bar

Page 27: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

OpenLAB CDS ChemStation Edition: Classic Driver Support Information

CDS Supported Instruments and Firmware Guide 27

Classic Driver Support Information

Agilent HPLC – Sampling Systems

Product Number Module Name Minimum required

Firmware Revision

G1313A Standard Autosampler A.06.10

G1329A Standard Autosampler A.06.10

G1329B Standard Autosampler A.06.10

G1367A Well-plate sampler A.06.10

G1367B High Performance Autosampler A.06.10

G1367C High Performance Autosampler SL A.06.10

G1367D High Performance Autosampler SL+ A.06.12

G1373A Purge Kit A.06.56

G1377A High Performance Micro Autosampler A.06.10

G1389A Micro ALS A.06.10

G2258A Dual-Loop Autosampler A.06.10

G2260A Preparative Autosampler (High flow) A.06.10

G2261A Therm. High Flow ALS A.06.10

Agilent HPLC – Pumps

Product

Number

Module Name Minimum required

Firmware Revision

G1310A Isocratic Pump A.06.10

G1311A Quaternary Pump A.06.10

G1312A Binary Pump A.06.10

G1312B Binary Pump SL A.06.10

G1361A Preparative Pump A.06.10

G1376A Capillary Pump A.06.10

G2225A

G2226A

Nanoflow Pump A.06.10

Page 28: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

OpenLAB CDS ChemStation Edition: Classic Driver Support Information

CDS Supported Instruments and Firmware Guide 28

Agilent HPLC – Detectors

Product

Number

Module Name Minimum required

Firmware Revision

Notes

G1314A Variable wavelength Detector (VWD) A.06.10

G1314B Variable wavelength Detector A.06.10

G1314C Variable wavelength Detector A.06.10

G1314D Variable wavelength Detector B.06.20

G1314E Variable wavelength Detector B.06.20

G1315A Diode Array Detector (DAD) A.06.10

G1315B Diode Array Detector (DAD) A.06.10

G1315C Diode Array Detector VL+ B.06.30

G1315D Diode Array Detector VL B.06.30

G1321A Fluorescence Detector (FLD) A.06.10 No 3D option

G1362A Refractive Index Detector (RID) A.06.10

G1365A Multiple Wavelength Detector (MWD) A.06.10

G1365B Multiple Wavelength Detector A.06.10

G1365C Multiple Wavelength Detector B.06.30

G1365D Multiple Wavelength Detector VL B.06.30

Agilent HPLC – Column Compartments

Product

Number

Module Name Minimum required Firmware Revision

G1316A Thermostatted Column Compartment A.06.10

G1316B Thermostatted Column Compartment SL A.06.10

G1316C Thermostatted Column Compartment A.06.30

Agilent HPLC – Quick Change Valves

Product

Number

Module Name Minimum required Firmware Revision

G1156A 6Pos/7Port Valve Requires an ALS -

not configurable neither Classic, nor RC.NET.

G1157A 2Pos/10Port Valve A.06.10

G1158A 2Pos/6Port Valve (Standard) A.06.10

G1159A 6 Position Column Selector A.06.10

G1160A 12Pos/13Port Selection Valve A.06.10

G1162A 2Pos/6Port Valve (Micro) A.06.10

G1163A 2Pos/10Port Valve (Micro) A.06.10

Page 29: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

OpenLAB CDS ChemStation Edition: Classic Driver Support Information

CDS Supported Instruments and Firmware Guide 29

Agilent 1200 Infinity Series LC – Other Module types

Product

Number

Module Name Minimum required

Firmware Revision

G1390B 1200 Universal Interface Box 2 A.06.10

G4240A 1260 Infinity Chip Cube MS Interface A.06.10

G2254A 1260 Automation Interface A.06.10

Agilent A/D Converter

Instrument Supported Firmware Version (minimum) Communications

Agilent 35900E E.01.02 TCP/IP, LAN only

Page 30: Agilent OpenLAB Chromatography Data System (CDS ......G4270-CTC 2.5.2 (for legacy mainboard)CTC Analytics HTC PAL Injection System1 2.6.8 (for legacy mainboard); 4.x.x (for new XT

www.agilent.com

Manual Part Number Edition

M8300-90018 11/2016

In This Book