Agilent Technologies Agilent 5977B Series MSD Laboratory Operator Quick Reference Guide Ion Source Routine Maintenance

Transcript of Agilent 5977B Series MSD...9 Ceramic insulator for extractor 10 Extractor lens (5) 11 Post extractor...

Page 1: Agilent 5977B Series MSD...9 Ceramic insulator for extractor 10 Extractor lens (5) 11 Post extractor lens 1 (4) 12 Post extractor lens 2 (3) 13 Ion focus lens (2) 14 Entrance lens

Agilent 5977B Series MSD

Laboratory Operator Quick Reference Guide

Ion Source Routine Maintenance

Agilent Technologies

Page 2: Agilent 5977B Series MSD...9 Ceramic insulator for extractor 10 Extractor lens (5) 11 Post extractor lens 1 (4) 12 Post extractor lens 2 (3) 13 Ion focus lens (2) 14 Entrance lens

Ion Source Cleaning

2

Clean the parts that contact the sample or ion beam. The other parts normally should not require cleaning.

If the contamination is serious, such as an oil backflow into the analyzer, seriously consider replacing the contaminated parts.

Abrasively clean the surfaces that contact the sample or ion beam.

Use an abrasive slurry of alumina powder and reagent-grade methanol on a cotton swab. Use enough force to remove all discolorations. Polishing the parts is not necessary; small scratches will not harm performance. Also, abrasively clean the discolorations where electrons from the filaments enter the source body.

Rinse away all abrasive residue with reagent-grade methanol.

Take care to avoid recontaminating cleaned and dried parts. Put on new, clean gloves before handling the parts. Do not set the cleaned parts on a dirty surface. Set them only on clean, lint-free cloths.

NOTE The main effect of operating the MSD in CI mode is the need for more frequent ion source cleaning. In CI operation, the ion source chamber is subject to more rapid contamination than in EI operation because of the higher source pressures required for CI.

Do not overtighten the repeller nut or the ceramic repeller insulators will break when

CAUTIONthe source heats up. The nut should only be finger-tight.

If insulators are dirty, clean them with a cotton swab dampened with reagent-grade

CAUTIONmethanol. If that does not clean the insulators, replace them. Do not abrasively or ultrasonically clean the insulators.

The filaments, source heater assembly, and insulators cannot be cleaned

CAUTIONultrasonically. Replace these components if major contamination occurs.

5977B Series MSD Lab Operator Quick Reference Guide

Page 3: Agilent 5977B Series MSD...9 Ceramic insulator for extractor 10 Extractor lens (5) 11 Post extractor lens 1 (4) 12 Post extractor lens 2 (3) 13 Ion focus lens (2) 14 Entrance lens

1

2

8*

15151515 141414

2*

2*

161616

8*

12*12*12*

10

2*8*

8*7*

1718

3

88

7

1213

46

511

9

EI Ion Source - Stainless or Inert

Clean the parts highlighted in blue.

1318 3 4 6 5 11 10

1 Gold plated set screw

2 Gold plated screw

3 Interface socket

4 Source body

5 Drawout cylinder

6 Drawout plate

7 4-turn filament

8 Flat and spring washers

9 Lens insulator

10 Entrance lens

11 Ion focus lens

12 Repeller insulator 18 Repeller block insert

13 Repeller

14 Flat washer

15 Belleville spring washer

16 Repeller nut

17 Source heater block assembly

5977B Series MSD Lab Operator Quick Reference Guide 3

Page 4: Agilent 5977B Series MSD...9 Ceramic insulator for extractor 10 Extractor lens (5) 11 Post extractor lens 1 (4) 12 Post extractor lens 2 (3) 13 Ion focus lens (2) 14 Entrance lens

4

2

45

6

7

8

910

1112

1314

15

161

3

4

1

8

910

11

HES Ion Source

Clean the parts highlighted in blue.

8

6

14

2

13 1112 10

1 M2 × 6 mm gold plated screw 2 Filament block 3 High efficiency dual filament4 M2 × 0.4 screw × 12 mm5 Source mount 1.5 mm6 Repeller assembly

7 Ring heater/sensor assembly8 Source body9 Ceramic insulator for extractor10 Extractor lens (5)11 Post extractor lens 1 (4)12 Post extractor lens 2 (3)

13 Ion focus lens (2)14 Entrance lens assembly (1)15 Lens insulator/holder16 Lens stack locking ring

5977B Series MSD Lab Operator Quick Reference Guide

Page 5: Agilent 5977B Series MSD...9 Ceramic insulator for extractor 10 Extractor lens (5) 11 Post extractor lens 1 (4) 12 Post extractor lens 2 (3) 13 Ion focus lens (2) 14 Entrance lens

1

1*

2

7*

14141414 131313

2*

2*

151515

7*

11*11*11*

9

2*7* 7*

6*

1617

77

6

11

12

3

54

10

8

1217 91043

EI Ion Source - Extractor

Clean the parts highlighted in blue.

1 Set screws

2 Screws

3 Source body

4 Extractor lens

5 Extractor lens insulator

6 Filaments

7 Flat and spring washers

8 Lens insulator

9 Entrance lens

10 Ion focus lens

11 Repeller insulator

12 Repeller

13 Flat washer

14 Belleville spring washer

15 Repeller nut

16 Source heater block assembly

17 Repeller block insert

5977B Series MSD Lab Operator Quick Reference Guide 5

Page 6: Agilent 5977B Series MSD...9 Ceramic insulator for extractor 10 Extractor lens (5) 11 Post extractor lens 1 (4) 12 Post extractor lens 2 (3) 13 Ion focus lens (2) 14 Entrance lens

T

1

14

13

15

1514

142

3

4

5

6

7

8

9

11

10

2

CI Ion Source

Clean the parts highlighted in blue.

11 9 6 5 10 8

1 Set screw2 Filament screw3 CI repeller insulator4 CI lens insulator5 CI drawout cylinder6 CI drawout plate

7 CI source heater block assembly8 Entrace lens9 CI source body10 CI ion focus lens11 CI repeller12 CI filament

13 Dummy filament14 Washer spring curved15 Flat washer

6 5977B Series MSD Lab Operator Quick Reference Guide

Page 7: Agilent 5977B Series MSD...9 Ceramic insulator for extractor 10 Extractor lens (5) 11 Post extractor lens 1 (4) 12 Post extractor lens 2 (3) 13 Ion focus lens (2) 14 Entrance lens

Scheduled maintenance

5977B Series MSD L

Task Every week Every 6 months Every year As needed

Tune the MSD X

Check the foreline pump oil level X

Check the calibration vial(s) X

Replace the foreline pump oil* X

Replace the diffusion pump fluid X

Check the dry foreline pump X

Change the dry foreline pump tip seal X

Change the foreline pump exhaust filter X

Clean the ion source X

Check the carrier gas trap(s) on the GC and MSD X

Replace the worn out parts X

Lubricate sideplate or vent valve O-rings† X

Replace CI Reagent gas supply X

Replace GC gas supplies X

* Every 3 months for CI MSDs using ammonia reagent gas.

† Vacuum seals other than the side plate O-ring and vent valve O-ring do not need to be lubricated. Lubricating other seals can interfere with their correct function.

ab Operator Quick Reference Guide 7

Page 8: Agilent 5977B Series MSD...9 Ceramic insulator for extractor 10 Extractor lens (5) 11 Post extractor lens 1 (4) 12 Post extractor lens 2 (3) 13 Ion focus lens (2) 14 Entrance lens

Agilent Technologies

© Agilent Technologies, Inc.

Printed in USA, June 2016

*G7077-90009*G7077-90009