Agilent 5973 MS User Manual

162
Agilent Technologies Agilent 5975 Series MSD Operation Manual

description

Agilent MSD manual for MS5973

Transcript of Agilent 5973 MS User Manual

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 1/162

Agilent Technologies

Agilent 5975

Series MSD

Operation Manual

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 2/162

2 5975 Series MSD Operation Manual

Notices© Agilent Technologies, Inc. 2012

No part of this manual may be reproduced in

any form or by any means (including elec-

tronic storage and retrieval or translation

into a foreign language) without prior agree-

ment and written consent from Agilent

Technologies, Inc. as governed by United

States and international copyright laws.

Manual Part Number

G3170-90036

Edition

Fourth edition, June 2012

Replaces G3170-90030

Printed in USA

Agilent Technologies, Inc.

5301 Stevens Creek Boulevard

Santa Clara, CA 95052

Warranty

The material contained in this docu-ment is provided “as is,” and is sub-ject to being changed, without notice,in future editions. Further, to the max-imum extent permitted by applicablelaw, Agilent disclaims all warranties,

either express or implied, with regardto this manual and any informationcontained herein, including but notlimited to the implied warranties ofmerchantability and fitness for a par-ticular purpose. Agilent shall not beliable for errors or for incidental orconsequential damages in connec-tion with the furnishing, use, or per-formance of this document or of anyinformation contained herein. ShouldAgilent and the user have a separatewritten agreement with warrantyterms covering the material in thisdocument that conflict with theseterms, the warranty terms in the sep-arate agreement shall control.

Safety Notices

CAUTION

A CAUTION notice denotes a

hazard. It calls attention to an

operating procedure, practice, orthe like that, if not correctly

performed or adhered to, could

result in damage to the product or

loss of important data. Do not

proceed beyond a CAUTION notice

until the indicated conditions are

fully understood and met.

WARNING

A WARNING notice denotes a

hazard. It calls attention to an

operating procedure, practice, or

the like that, if not correctly

performed or adhered to, could

result in personal injury or death.

Do not proceed beyond a

WARNING notice until the

indicated conditions are fully

understood and met.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 3/162

5975 Series MSD Operation Manual 3

About This ManualThis manual contains information for operating and

maintaining the Agilent 5975 Series Gas Chromatograph/Mass

Selective Detector (GC/MSD) system.

1 “Introduction”

Chapter 1 describes general information about the 5975 Series

MSDs, including a hardware description, general safety 

warnings, and hydrogen safety information.

2 “Installing GC Columns”

Chapter 2 shows you how to prepare a capillary column for use

with the MSD, install it in the GC oven, and connect it to the

MSD using the GC/MSD interface.

3 “Operating in Electron Impact (EI) Mode”

Chapter 3 describes basic tasks such as setting temperatures,

monitoring pressures, tuning, venting, and pumpdown. Much of 

the information in this chapter also applies to CI operation.

4 “Operating in Chemical Ionization (CI) Mode”

Chapter 4 describes additional tasks necessary to operate in CI

mode.

5 “General Maintenance”

Chapter 5 describes maintenance procedures common to both

EI and CI instruments.

6 “CI Maintenance”

Chapter 6 describes maintenance procedures unique to CI

MSDs.

A “Chemical Ionization Theory”

 Appendix A is an overview of chemical ionization theory.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 4/162

4 5975 Series MSD Operation Manual

Online User InformationNow your Agilent instrument documentation is in one place, at 

 your fingertips.

The Instrument Utilities DVD that ships with your instrument 

provides an extensive collection of online help, videos, and

books for the Agilent 7890A GC, 7820A GC, 6890N GC, 6850 GC, 5975T

LTM GC/MS, 7693A ALS, and the 7683B ALS. Included are localized

 versions of the information you need most, such as:

• Getting Familiar documentation• Safety and Regulatory guides• Site Preparation checklists• Installation information• Operating guides• Maintenance information• Troubleshooting details

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 5/162

5975 Series MSD Operation Manual 5

Contents

1 Introduction

5975 MSD Version 10

Abbreviations Used 11

The 5975 Series MSD 13

CI MSD Hardware Description 15

Important Safety Warnings 17

Hydrogen Safety 19

GC precautions 19

Safety and Regulatory Certifications 24

Cleaning/Recycling the Product 27

Liquid Spillage 27

Moving or Storing the MSD 27

2 Installing GC Columns

Columns 30

To reconfigure a 6850 GC column on its basket 32

To prepare a capillary column for installation 37

To install a capillary column in a split/splitless inlet 39

To condition a capillary column 41

To install a capillary column in the GC/MSD interface 43

Agilent 7890A and 7820A, and 6890 GCs 43

6850 GC 45

3 Operating in Electron Impact (EI) Mode

Operating the MSD from the Data System   51

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 6/162

5975 Series MSD Operation Manual 6

Operating the MSD from the LCP 51Modes of operation 51

LCP Status Messages 53

ChemStation Loading <timestamp> 53

Executing <type>tune 53

Instrument Available <timestamp> 53

Loading Method <method name> 53Loading MSD Firmware 53

Loading OS 54

<method> Complete <timestamp> 54

Method Loaded <method name> 54

MS locked by <computer name> 54

Press Sideplate 54

Run: <method> Acquiring <datafile> 54

To view system status during startup 54

LCP Menus 55

The EI GC/MSD Interface 58

Before You Turn On the MSD 60

Pumping Down 61

Controlling Temperatures 61

Controlling Column Flow 62

Venting the MSD 63

To view MSD analyzer temperature and vacuum status 64

To set monitors for MSD temperature and vacuum status 66

To set the MSD analyzer temperatures 67

To set the GC/MSD interface temperature from the

ChemStation 69

To monitor high vacuum pressure 71

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 7/162

5975 Series MSD Operation Manual 7

To measure column flow linear velocity 73

To confirm column flow 74

To tune the MSD 75

To verify system performance 76

High-Mass Testing (5975 Series MSDs) 77

To remove the MSD covers 80

To vent the MSD 82

To open the analyzer chamber 84

To close the analyzer chamber 87

To pump down the MSD 91

To move or store the MSD 93

To set the interface temperature from the GC 95

4 Operating in Chemical Ionization (CI) Mode

General Guidelines 98

The CI GC/MSD Interface 99To Operate the CI MSD 101

To switch from the EI source to the CI source 102

To pump down the CI MSD 103

To set up the software for CI operation 104

To operate the reagent gas flow control module 106

To set up methane reagent gas flow 109

To use other reagent gases 111

To switch from the CI source to the EI source 115

CI Autotune 116

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 8/162

5975 Series MSD Operation Manual 8

To perform a PCI autotune (methane only) 118

To perform an NCI autotune (methane reagent gas) 120

To verify PCI performance 122

To verify NCI performance 123

To monitor high vacuum pressure 124

5 General Maintenance

Before Starting 128

Maintaining the Vacuum System 133

6 CI Maintenance

General Information 140To Set Up Your MSD for CI Operation 141

A Chemical Ionization Theory

Chemical Ionization Overview 146

Positive CI Theory 148

Negative CI Theory 155

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 9/162

9

Agilent 5975 Series MSD

Operation Manual

Agilent Technologies

1

Introduction

5975 MSD Version 10

Abbreviations Used 11

The 5975 Series MSD 13CI MSD Hardware Description 15

Important Safety Warnings 17

Many internal parts of the MSD carry dangerous voltages 17

Electrostatic discharge is a threat to MSD electronics 17

Many parts are dangerously hot 18

The oil pan under the standard foreline pump can be a fire hazard 18

Hydrogen Safety 19

Dangers unique to GC/MSD operation 20

Hydrogen accumulation in an MSD 20

Precautions 22

Safety and Regulatory Certifications 24

Information 24

Symbols 25

Electromagnetic compatibility 26

Sound emission declaration 26

Cleaning/Recycling the Product 27

Liquid Spillage 27

Moving or Storing the MSD 27

This manual describes the operation, and routine maintenance of the Agilent 

Technologies 5975 Series MSD.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 10/162

10 5975 Series MSD Operation Manual

1 Introduction

5975 MSD Version

5975 Series MSDs are equipped with a diffusion pump or one of two

turbomolecular (turbo) pumps. The serial number label displays a product 

number (Table 1) that indicates what kind of MSD you have.

Table 1 Available high vacuum pumps

Model name Product number Description Ionization modes

5975C TAD VL MSD G3170A Diffusion Pump MSD Electron impact (EI)

5975C TAD inert

MSD

5975C TAD inert XL

MSD

G3171A

G3172A

Standard Turbo MSD

Performance Turbo MSD

Electron impact (EI)

Electron impact (EI)

5975C TAD inert XL

MSD

G3174A CI High Mass Performance

Turbo Pump

Electron impact (EI)

Negative chemical ionization (NCI)

Positive chemical ionization (PCI)

7820 MSD VL G3175A Diffusion Pump MSD Electron impact (EI)

7820 MSD G3176A Standard Turbo MSD Electron impact (EI)

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 11/162

Introduction 1

5975 Series MSD Operation Manual 11

Abbreviations Used

The abbreviations in Table 2 are used in discussing this product. They are

collected here for convenience.

Table 2 Abbreviations

Abbreviation Definition

AC Alternating current

ALS Automatic liquid sampler

BFB Bromofluorobenzene (calibrant)

CI Chemical ionization

DC Direct current

DFTPP Decafluorotriphenylphosphine (calibrant)

DIP Direct insertion probe

DP Diffusion pump

EI Electron impact ionization

EM Electron multiplier (detector)

EMV Electron multiplier voltage

EPC Electronic pneumatic control

eV Electron volt

GC Gas chromatograph

HED High-energy dynode (refers to detector and its power supply)

id Inside diameter

LAN Local Area Network

LCP Local control panel (on the MSD)

LTM Low thermal mass

m/z  Mass to charge ratio

MFC Mass flow controller

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 12/162

12 5975 Series MSD Operation Manual

1 Introduction

MSD Mass Selective Detector

NCI Negative CI

OFN Octafluoronaphthalene (calibrant)

PCI Positive CI

PFDTD Perfluoro-5,8-dimethyl-3,6,9-trioxydodecane (calibrant)

PFHT 2,4,6-tris(perfluoroheptyl)-1,3,5-triazine (calibrant)

PFTBA Perfluorotributylamine (calibrant)

Quad Quadrupole mass filter

RF Radio frequency

RFPA Radio frequency power amplifier

Torr Unit of pressure, 1 mm Hg

Turbo Turbomolecular (pump)

Table 2 Abbreviations (continued)

Abbreviation Definition

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 13/162

Introduction 1

5975 Series MSD Operation Manual 13

The 5975 Series MSD

The 5975 Series MSD is a stand-alone capillary GC detector for use with an

 Agilent Series Gas Chromatograph (Table 3). The MSD features:

• Local Control Panel (LCP) for locally monitoring and operating the MSD

• One of three different high vacuum pumps

• Rotary vane foreline pump

• Independently MSD heated electron-ionization ion source

• Independently MSD heated hyperbolic quadrupole mass filter

• High-energy dynode (HED) electron multiplier detector

• Independently GC heated GC/MSD interface

• Chemical ionization (EI/PCI/NCI) modes available

Physical description

The 5975 Series MSD is a rectangular box, approximately 42 cm high, 26 cm

wide, and 65 cm deep. The weight is 25 kg for the diffusion pump mainframe,

26 kg for the standard turbo pump mainframe, and 29 kg for the performance

turbo pump mainframe. The attached foreline (roughing) pump weighs an

additional 11 kg (standard pump).

The basic components of the instrument are: the frame/cover assemblies, thelocal control panel, the vacuum system, the GC interface, the electronics, and

the analyzer.

Local control panel

The local control panel allows local monitoring and operation of the MSD. You

can tune the MSD, run a method or a sequence, and monitor instrument 

status.

Vacuum gauge

The 5975 Series MSD may be equipped with a Micro-Ion Vacuum Gauge. The

MSD ChemStation can be used to read the pressure (high vacuum) in the

 vacuum manifold. Operation of the gauge controller is described in this

manual.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 14/162

14 5975 Series MSD Operation Manual

1 Introduction

The gauge is required for chemical ionization (CI) operation.

Table 3 5975 series MSD models and features

Model

Feature G3170A

G3175A

G3171A

G3176A

G3172A G3174A

High vacuum pump Diffusion Standard

turbo

Performance

turbo

Performance

turbo

Optimal He column flow mL/min 1 1 1 to 2 1 to 2

Maximum recommended gas flow

mL/min*

* Total gas flow into the MSD: column flow plus reagent gas flow (if applicable).

1.5 2.0 4.0 4

Maximum gas flow, mL/min†

† Expect degradation of spectral performance and sensitivity.

2 2.4 6.5 6.5

Max column id 0.25 mm

(30 m)

0.32 mm

(30 m)

0.53 mm

(30 m)

0.53 mm

(30 m)

CI capability No No No Yes

DIP‡ capability (3rd party)

‡ Direct insertion probe.

Yes Yes Yes Yes

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 15/162

Introduction 1

5975 Series MSD Operation Manual 15

CI MSD Hardware Description

Figure 1 is an overview of a typical 5975 GC/MSD system.

The CI hardware allows the 5975 Series MSD to produce high-quality, classical

CI spectra, which include molecular adduct ions. A variety of reagent gases

can be used.

Figure 1 5975 Series GC/MSD system

ALS

7890A GC

CI gas flow module

Local control panel

5975 Series MSD

MSD power switch

GC power switch

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 16/162

16 5975 Series MSD Operation Manual

1 Introduction

In this manual, the term “CI MSD” refers to the G3174A MSD and upgraded

G3172A MSDs. It also applies, unless otherwise specified, to the flow modulesfor these instruments.

The 5975 Series CI system adds to the 5975 Series MSD:

• EI/CI GC/MSD interface

• CI ion source and interface tip seal

• Reagent gas flow control module

• Bipolar HED power supply for PCI and NCI operation

 A methane/isobutane gas purifier is provided and is required . It removes

oxygen, water, hydrocarbons, and sulfur compounds.

 A high vacuum gauge controller (G3397A) is required for CI MSD and is

recommended for EI also.

The MSD CI system has been optimized to achieve the relatively high source

pressure required for CI while still maintaining high vacuum in thequadrupole and detector. Special seals along the f low path of the reagent gas

and very small openings in the ion source keep the source gases in the

ionization volume long enough for the appropriate reactions to occur.

The CI interface has special plumbing for reagent gas. A spring-loaded

insulating seal fits onto the tip of the interface.

Switching back and forth between CI and EI sources takes less than an hour,

although a 1- to 2-hour wait is required to purge the reagent gas lines and bake

out water and other contaminants. Switching from PCI to NCI requires about 

2 hours for the ion source to cool.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 17/162

Introduction 1

5975 Series MSD Operation Manual 17

Important Safety WarningsThere are several important safety notices to always keep in mind when using

the MSD.

Many internal parts of the MSD carry dangerous voltages

If the MSD is connected to a power source, even if the power switch is off,

potentially dangerous voltages exist on:

• The wiring between the MSD power cord and the AC power supply, the AC

power supply itself, and the wiring from the AC power supply to the power

switch.

With the power switch on, potentially dangerous voltages also exist on:

•  All electronics boards in the instrument.

• The internal wires and cables connected to these boards.• The wires for any heater (oven, detector, inlet, or valve box).

Electrostatic discharge is a threat to MSD electronics

The printed circuit boards in the MSD can be damaged by electrostaticdischarge. Do not touch any of the boards unless it is absolutely necessary. If 

 you must handle them, wear a grounded wrist strap and take other antistatic

precautions. Wear a grounded wrist strap any time you must remove the MSD

right side cover.

WARNINGAll these parts are shielded by covers. With the covers in place, it should be difficult

to accidentally make contact with dangerous voltages. Unless specifically

instructed to, never remove a cover unless the detector, inlet, or oven are turned off.

WARNINGIf the power cord insulation is frayed or worn, the cord must be replaced. Contact

your Agilent service representative.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 18/162

18 5975 Series MSD Operation Manual

1 Introduction

Many parts are dangerously hot

Many parts of the GC/MSD operate at temperatures high enough to cause

serious burns. These parts include but are not limited to:

• The inlets

• The oven and its contents

• The detector

• The column nuts attaching the column to an inlet or detector

• The valve box

• The foreline pump

 Always cool these areas of the system to room temperature before working on

them. They will cool faster if you first set the temperature of the heated zone

to room temperature. Turn the zone off after it has reached the setpoint. If you

must perform maintenance on hot parts, use a wrench and wear gloves.

Whenever possible, cool the part of the instrument that you will bemaintaining before you begin working on it.

The oil pan under the standard foreline pump can be a fire hazard

Oily rags, paper towels, and similar absorbents in the oil pan could ignite and

damage the pump and other parts of the MSD.

WARNINGBe careful when working behind the instrument. During cool-down cycles, the GC

emits hot exhaust which can cause burns.

WARNING The insulation around the inlets, detectors, valve box, and the insulation cups ismade of refractory ceramic fibers. To avoid inhaling fiber particles, we recommend

the following safety procedures: ventilate your work area; wear long sleeves,

gloves, safety glasses, and a disposable dust/mist respirator; dispose of insulation

in a sealed plastic bag; wash your hands with mild soap and cold water after

handling the insulation.

WARNINGCombustible materials (or flammable/non-flammable wicking material) placed

under, over, or around the foreline (roughing) pump constitutes a fire hazard. Keep

the pan clean, but do not leave absorbent material such as paper towels in it.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 19/162

Introduction 1

5975 Series MSD Operation Manual 19

Hydrogen Safety 

Hydrogen is a commonly used GC carrier gas. Hydrogen is potentially 

explosive and has other dangerous characteristics.

• Hydrogen is combustible over a wide range of concentrations. At 

atmospheric pressure, hydrogen is combustible at concentrations from 4%

to 74.2% by volume.

• Hydrogen has the highest burning velocity of any gas.

• Hydrogen has a very low ignition energy.

• Hydrogen that is allowed to expand rapidly from high pressure can

self-ignite.

• Hydrogen burns with a nonluminous flame which can be invisible under

bright light.

GC precautions

When using hydrogen as a carrier gas, remove the large round plastic cover for

the MSD transfer line located on the GC left side panel. In the unlikely event of 

an explosion, this cover may dislodge.

WARNINGThe use of hydrogen as a GC carrier gas is potentially dangerous.

WARNING When using hydrogen (H2) as the carrier gas or fuel gas, be aware that hydrogengas can flow into the GC oven and create an explosion hazard. Therefore, be sure

that the supply is turned off until all connections are made and ensure that the inlet

and detector column fittings are either connected to a column or capped at all times

when hydrogen gas is supplied to the instrument.

Hydrogen is flammable. Leaks, when confined in an enclosed space, may create a

fire or explosion hazard. In any application using hydrogen, leak test all

connections, lines, and valves before operating the instrument. Always turn off thehydrogen supply at its source before working on the instrument.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 20/162

20 5975 Series MSD Operation Manual

1 Introduction

Dangers unique to GC/MSD operation

Hydrogen presents a number of dangers. Some are general, others are unique

to GC or GC/MSD operation. Dangers include, but are not limited to:

• Combustion of leaking hydrogen.

• Combustion due to rapid expansion of hydrogen from a high-pressure

cylinder.

•  Accumulation of hydrogen in the GC oven and subsequent combustion (see

 your GC documentation and the label on the top edge of the GC oven door).

•  Accumulation of hydrogen in the MSD and subsequent combustion.

Hydrogen accumulation in an MSD

 All users should be aware of the mechanisms by which hydrogen can

accumulate (Table 4) and know what precautions to take if they know or

suspect that hydrogen has accumulated. Note that these mechanisms apply to

all mass spectrometers, including the MSD.

WARNINGThe MSD cannot detect leaks in inlet and/or detector gas streams. For this reason,

it is vital that column fittings should always be either connected to a column or havea cap or plug installed.

Table 4 Hydrogen accumulation mechanisms

Mechanism Results

Mass spectrometer turned off A mass spectrometer can be shut down

deliberately. It can also be shut down accidentally

by an internal or external failure. A massspectrometer shutdown does not shut off the flow

of carrier gas. As a result, hydrogen may slowly

accumulate in the mass spectrometer.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 21/162

Introduction 1

5975 Series MSD Operation Manual 21

Mass spectrometer automated shutoff

valves closed

Some mass spectrometers are equipped with

automated diffusion pump shutoff valves. In these

instruments, deliberate operator action or various

failures can cause the shutoff valves to close.

Shutoff valve closure does not shut off the flow of

carrier gas. As a result, hydrogen may slowlyaccumulate in the mass spectrometer.

Mass spectrometer manual shutoff

valves closed

Some mass spectrometers are equipped with

manual diffusion pump shutoff valves. In these

instruments, the operator can close the shutoff

valves. Closing the shutoff valves does not shut off

the flow of carrier gas. As a result, hydrogen mayslowly accumulate in the mass spectrometer.

GC off A GC can be shut down deliberately. It can also be

shut down accidentally by an internal or external

failure. Different GCs react in different ways. If a

6890 GC equipped with Electronic Pressure Control

(EPC) is shut off, the EPC stops the flow of carriergas. If the carrier flow is not under EPC control, the

flow increases to its maximum. This flow may be

more than some mass spectrometers can pump

away, resulting in the accumulation of hydrogen in

the mass spectrometer. If the mass spectrometer is

shut off at the same time, the accumulation can be

fairly rapid.

Power failure If the power fails, both the GC and mass

spectrometer shut down. The carrier gas, however,

is not necessarily shut down. As described

previously, in some GCs a power failure may cause

the carrier gas flow to be set to maximum. As a

result, hydrogen may accumulate in the mass

spectrometer.

Table 4 Hydrogen accumulation mechanisms (continued)

Mechanism Results

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 22/162

22 5975 Series MSD Operation Manual

1 Introduction

Precautions

Take the following precautions when operating a GC/MSD system with

hydrogen carrier gas.

Equipment precaution

 You MUST make sure the front side-plate thumbscrew is fastened finger-tight.

Do not overtighten the thumbscrew; it can cause air leaks.

 You must remove the plastic cover over the glass window on the front of a 5975

MSD. In the unlikely event of an explosion, this cover may dislodge.

General laboratory precautions

•  Avoid leaks in the carrier gas lines. Use leak-checking equipment toperiodically check for hydrogen leaks.

• Eliminate from your laboratory as many ignition sources as possible (open

flames, devices that can spark, sources of static electricity, etc.).

• Do not allow hydrogen from a high pressure cylinder to vent directly to

atmosphere (danger of self-ignition).

• Use a hydrogen generator instead of bottled hydrogen.

WARNING Once hydrogen has accumulated in a mass spectrometer, extreme caution must beused when removing it. Incorrect startup of a mass spectrometer filled with

hydrogen can cause an explosion.

WARNINGAfter a power failure, the mass spectrometer may start up and begin the pumpdown

process by itself. This does not guarantee that all hydrogen has been removed from

the system or that the explosion hazard has been removed.

WARNINGFailure to secure your MSD as described above greatly increases the chance of

personal injury in the event of an explosion.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 23/162

Introduction 1

5975 Series MSD Operation Manual 23

Operating precautions

• Turn off the hydrogen at its source every time you shut down the GC or

MSD.

• Turn off the hydrogen at its source every time you vent the MSD (do not 

heat the capillary column without carrier gas f low).

• Turn off the hydrogen at its source every time shutoff valves in an MSD are

closed (do not heat the capillary column without carrier gas flow).

• Turn off the hydrogen at its source if a power failure occurs.• If a power failure occurs while the GC/MSD system is unattended, even if 

the system has restarted by itself:

1 Immediately turn off the hydrogen at its source.

2 Turn off the GC.

3 Turn off the MSD and allow it to cool for 1 hour.

4 Eliminate all potential sources of ignition in the room.5 Open the vacuum manifold of the MSD to atmosphere.

6 Wait at least 10 minutes to allow any hydrogen to dissipate.

7 Start up the GC and MSD as normal.

When using hydrogen gascheck the system for leaks to prevent possible fire

and explosion hazards based on local Environmental Health and Safety (EHS)

requirements. Always check for leaks after changing a tank or servicing the

gas lines. Always make sure the vent line is vented into a fume hood.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 24/162

24 5975 Series MSD Operation Manual

1 Introduction

Safety and Regulatory Certifications

The 5975 Series MSD conforms to the following safety standards:

• Canadian Standards Association (CSA): CAN/CSA-C222 No. 61010-1-04

• CSA/Nationally Recognized Test Laboratory (NRTL): UL 61010–1

• International Electrotechnical Commission (IEC): 61010–1

• EuroNorm (EN): 61010–1

The 5975 MSD conforms to the following regulations on Electromagnetic

Compatibility (EMC) and Radio Frequency Interference (RFI):

• CISPR 11/EN 55011: Group 1, Class A

• IEC/EN 61326

•  AUS/NZ

This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB—001 du Canada.

The 5975 Series MSD is designed and manufactured under a quality system

registered to ISO 9001.

Information

The Agilent Technologies 5975 Series MSD meets the following IEC

(International Electro-technical Commission) classifications: Equipment Class

I, Laboratory Equipment, Installation Category II, Pollution Degree 2.

This unit has been designed and tested in accordance with recognized safety 

standards and is designed for use indoors. If the instrument is used in a

manner not specified by the manufacturer, the protection provided by theinstrument may be impaired. Whenever the safety protection of the MSD has

been compromised, disconnect the unit from all power sources and secure the

unit against unintended operation.

Refer servicing to qualified service personnel. Substituting parts or

performing any unauthorized modification to the instrument may result in a

safety hazard.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 25/162

Introduction 1

5975 Series MSD Operation Manual 25

Symbols

Warnings in the manual or on the instrument must be observed during all

phases of operation, service, and repair of this instrument. Failure to comply 

with these precautions violates safety standards of design and the intended

use of the instrument. Agilent Technologies assumes no liability for the

customer’s failure to comply with these requirements.

See accompanying instructions for more information.

Indicates a hot surface.

Indicates hazardous voltages.

Indicates earth (ground) terminal.

Indicates potential explosion hazard.

Indicates radioactivity hazard.

Indicates electrostatic discharge hazard.

Indicates that you must not discard this

electrical/electronic product in domestic household

waste.

or

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 26/162

26 5975 Series MSD Operation Manual

1 Introduction

Electromagnetic compatibility

This device complies with the requirements of CISPR 11. Operation is subject 

to the following two conditions:

• This device may not cause harmful interference.

• This device must accept any interference received, including interference

that may cause undesired operation.

If this equipment does cause harmful interference to radio or television

reception, which can be determined by turning the equipment off and on, theuser is encouraged to try one or more of the following measures:

1 Relocate the radio or antenna.

2 Move the device away from the radio or television.

3 Plug the device into a different electrical outlet, so that the device and the

radio or television are on separate electrical circuits.

4 Make sure that all peripheral devices are also certified.5 Make sure that appropriate cables are used to connect the device to

peripheral equipment.

6 Consult your equipment dealer, Agilent Technologies, or an experienced

technician for assistance.

7 Changes or modifications not expressly approved by Agilent Technologies

could void the user’s authority to operate the equipment.

Sound emission declaration

Sound pressure

Sound pressure Lp <70 dB according to EN 27779:1991.

Schalldruckpegel

Schalldruckpegel LP <70 dB am nach EN 27779:1991.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 27/162

Introduction 1

5975 Series MSD Operation Manual 27

Cleaning/Recycling the Product

To clean the unit, disconnect the power and wipe down with a damp, lint-free

cloth. For recycling, contact your local Agilent sales office.

Liquid Spillage

Do not spill liquids on the MSD.

Moving or Storing the MSD

The best way to keep your MSD functioning properly is to keep it pumped

down and hot, with carrier gas flow. If you plan to move or store your MSD, a

few additional precautions are required. The MSD must remain upright at all

times; this requires special caution when moving. The MSD should not be left 

 vented to atmosphere for long periods.

1

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 28/162

28 5975 Series MSD Operation Manual

1 Introduction

A il t 5975 S i MSD

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 29/162

29

Agilent 5975 Series MSD

Operation Manual

Agilent Technologies

2

Installing GC Columns

Columns 30

Conditioning columns 30

Conditioning ferrules 31Tips and hints 31

To reconfigure a 6850 GC column on its basket 32

To prepare a capillary column for installation 37

To install a capillary column in a split/splitless inlet 39

To condition a capillary column 41

To install a capillary column in the GC/MSD interface 43

Before you can operate your GC/MSD system, you must select, install, and

condition a GC column. This chapter will show you how to install and

condition a column. For correct column and flow selection, you must know

what type of vacuum system your MSD has. The serial number tag on the lower

front of the left side panel shows the model number.

2 Installing GC Columns

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 30/162

30 5975 Series MSD Operation Manual

2 Installing GC Columns

Columns

Many types of GC columns can be used with the MSD but there are some

restrictions.

During tuning or data acquisition the rate of column flow into the MSD should

not exceed the maximum recommended flow. Therefore, there are limits to

column length and flow. Exceeding recommended flow will result in

degradation of mass spectral and sensitivity performance.

Remember that column flows vary greatly with oven temperature. See “To

measure column flow linear velocity” for instructions on how to measure

actual flow in your column. Use the Flow Calculation software and Table 5 to

determine whether a given column will give acceptable flow with realistic

head pressure.

Conditioning columns

Conditioning a column before it is connected to the GC/MSD interface is

essential.

Table 5 Gas flows

Feature G3170A

G3175A

G3171A

G3176A

G3172A G3174A

High vacuum pump Diffusion Standard

turbo

Performance

turbo

Performance

turbo

Optimal gas flow, mL/min*

* Total gas flow into the MSD = column flow + reagent gas flow (if applicable)

1 1 1 to 2 1 to 2

Maximum recommended gas flow,mL/min

1.5 2 4 4

Maximum gas flow, mL/min†

† Expect degradation of spectral performance and sensitivity.

2 2.4 6.5 6.5

Maximum column id 0.25 mm

(30 m)

0.32 mm

(30 m)

0.53 mm

(30 m)

0.53 mm

(30 m)

 

Installing GC Columns 2

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 31/162

Installing GC Columns 2

5975 Series MSD Operation Manual 31

 A small portion of the capillary column stationary phase is often carried away 

by the carrier gas. This is called column bleed. Column bleed deposits traces of the stationary phase in the MSD ion source. This decreases MSD sensitivity 

and makes cleaning the ion source necessary.

Column bleed is most common in new or poorly crosslinked columns. It is

much worse if there are traces of oxygen in the carrier gas when the column is

heated. To minimize column bleed, all capillary columns should be

conditioned before they are installed in the GC/MSD interface.

Conditioning ferrules

Heating ferrules to their maximum expected operating temperature a few

times before they are installed can reduce chemical bleed from the ferrules.

Tips and hints

• The column installation procedures for the 5975 Series MSDs is different from that for previous MSDs. Using the procedure from another instrument 

may not work and may damage the column or the MSD.

•  You can remove old ferrules from column nuts with an ordinary push pin.

•  Always use carrier gas that is at least 99.9995% pure.

• Because of thermal expansion, new ferrules may loosen after heating and

cooling a few times. Check for tightness after two or three heating cycles.

•  Always wear clean gloves when handling columns, especially the end that 

will be inserted into the GC/MSD interface.

WARNINGIf you are using hydrogen as a carrier gas, do not start carrier gas flow until the

column is installed in the MSD and the MSD has been pumped down. If the vacuum

pumps are off, hydrogen will accumulate in the MSD and an explosion may occur.

See “Hydrogen Safety” .

WARNINGAlways wear safety glasses when handling capillary columns. Use care to avoid

puncturing your skin with the end of the column.

2 Installing GC Columns

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 32/162

32 5975 Series MSD Operation Manual

2 Installing GC Columns

To reconfigure a 6850 GC column on its basket

Before installing a 6850, first reconfigure it to better position the column ends

for installation in the GC MSD interface.

1 Lay the column (19091S-433E found in the GC ship kit) on a clean surface

with the column label facing the user in the 12 o’clock position. Note that 

the inlet and outlet ends of the column are oriented the same as when a GC

detector is used and the column outlet is positioned at the back (closer to

the fan) of the column cage holder. See Figure 2.

Figure 2 Column

Column inlet

Column outlet

6850 column nut

Installing GC Columns 2

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 33/162

g

5975 Series MSD Operation Manual 33

2 Remove the septum cap from the column OUTLET side and uncoil 2 column

loops. See Figure 3.

3  Attach three column clips (part number G2630-20890) to the column cage

as follows:•  Attach one clip onto the back of the 1 o’clock cross-member piece of the

column cage.

•  Attach two clips onto the front of the 3 o’clock cross-member piece of the

column cage.

These clips will help provide appropriate orientation of column ends for

their insertion into the GC inlet and MSD interface.

Figure 3 Column with 2 uncoiled loops

1 o’clock cross-member

3 o’clock cross-member

2 Installing GC Columns

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 34/162

34 5975 Series MSD Operation Manual

See Figure 4.

4 Feed the outlet side of the column through the 1 o’clock positioned clip sothat the column outlet is pointing toward the front of the column cage. See

Figure 5.

Figure 4 Column with column clips attached

Column clip

(1 o’clock postion)

Column clips

(3 o’clock position)

Column outlet

CAUTIONBe careful not to scratch the column coating.

Installing GC Columns 2

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 35/162

5975 Series MSD Operation Manual 35

5 Next, feed the outlet side of the column through the 3 o’clock positioned

clips so that the column outlet is pointing toward the back of the column

cage. Make sure that the part of the column that is between the two clips

does NOT extend above the column label. See Figure 6.

Figure 5 Column fed through 1 o’clock position

To column outlet

Column clip

(1 o’clock position)

Column clips

(3 o’clock position)

CAUTIONBe careful not to scratch the column coating.

2 Installing GC Columns

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 36/162

36 5975 Series MSD Operation Manual

There should be approximately 50 cm of column extending beyond the

3 o’clock positioned clip.

6 Carefully rewind the remainder of the column outlet end around thecolumn cage.

Figure 6 Column fed through 3 o’clock position

Column clip

(1 o’clock postion)

Column clips

(3 o’clock position)

To column outlet

(at least 50 cm)

Installing GC Columns 2

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 37/162

5975 Series MSD Operation Manual 37

To prepare a capillary column for installation

Materials needed

• Capillary column

• Column cutter, ceramic (5181-8836) or diamond (5183-4620)

• Ferrules

• 0.27-mm id, for 0.10-mm id columns (5062-3518)

• 0.37-mm id, for 0.20-mm id columns (5062-3516)

• 0.40-mm id, for 0.25-mm id columns (5181-3323)

• 0.5-mm id, for 0.32-mm id columns (5062-3514)

• 0.8-mm id, for 0.53-mm id columns (5062-3512)

• Gloves, clean

• Large (8650-0030)

• Small (8650-0029)

• Inlet column nut (5181-8830 for Agilent 7890A, 7820A and 6890, or

5183-4732 for 6850)

• Magnifying loupe

• Septum (may be old, used inlet septum)

Procedure

1 Slide a septum, column nut, and conditioned ferrule onto the free end of the

column (Figure 7). The tapered end of the ferrule should point away from

the column nut.

2 Installing GC Columns

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 38/162

38 5975 Series MSD Operation Manual

2 Use the column cutter to score the column 2 cm from the end.

3 Break off the end of the column. Hold the column against the column cutter

with your thumb. Break the column against the edge of the column cutter.

4 Inspect the end for jagged edges or burrs. If the break is not clean and even,

repeat steps 2 and 3.5 Wipe the outside of the free end of the column with a lint-free cloth

moistened with methanol.

Figure 7 Preparing a capillary column for installation

Capillary column

Column cutter

Ferrule, taper up

Inlet column nut

Septum

Installing GC Columns 2

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 39/162

5975 Series MSD Operation Manual 39

To install a capillary column in a split/splitless inlet

Materials needed

• Gloves, clean

• Large (8650-0030)

• Small (8650-0029)

• Metric ruler

• Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)

To install columns in other types of inlets, refer to your Gas Chromatograph

User Information.

Procedure

1 Prepare the column for installation ( “To prepare a capillary column for

installation” on page 37).

2 Position the column so it extends 4 to 6 mm past the end of the ferrule

(Figure 8).

Figure 8 Installing a capillary column for a split/splitless inlet

 

Insulation cup

Reducing nut

Capillary column

Ferrule (inside nut)

Inlet column nut

Septum

4 to 6 mm

2 Installing GC Columns

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 40/162

40 5975 Series MSD Operation Manual

3 Slide the septum to place the nut and ferrule in the correct position.

4 Insert the column in the inlet.

5 Slide the nut up the column to the inlet base and finger-tighten the nut.

6  Adjust the column position so the septum is even with the bottom of the

column nut.

7 Tighten the column nut an additional 1/4 to 1/2 turn. The column should

not slide with a gentle tug.

8 Start carrier gas flow.9  Verify flow by submerging the free end of the column in isopropanol. Look 

for bubbles.

Installing GC Columns 2

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 41/162

5975 Series MSD Operation Manual 41

To condition a capillary column

Materials needed

• Carrier gas, (99.9995% pure or better)

• Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)

Procedure

1 Install the column in the GC inlet ( “To install a capillary column in asplit/splitless inlet” on page 39).

2  Allow the carrier gas to flow through the column for 5 minutes without 

heating the GC oven.

3 Ramp the oven temperature at 5 °C/minute to 10 °C above your highest 

analytical temperature.

4 Once the oven temperature exceeds 80 °C, inject 5 µL methanol into the GC.

Repeat two more times at 5-minute intervals. This helps remove any contamination from the column before it is installed into the GC/MSD

interface.

5 Hold this temperature. Allow the carrier gas to flow for several hours.

6 Return the GC oven temperature to a low standby temperature.

WARNING

Do not condition your capillary column with hydrogen. Hydrogen accumulation in

the GC oven can result in an explosion. If you plan to use hydrogen as your carrier

gas, first condition the column with ultrapure (99.999% or better) inert gas such as

helium, nitrogen, or argon.

 

CAUTIONNever exceed the maximum column temperature, either in the GC/MSD interface, the

GC oven, or the inlet.

2 Installing GC Columns

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 42/162

42 5975 Series MSD Operation Manual

See also

For more information about installing a capillary column, refer to the

application note Optimizing Splitless Injections on Your GC for High 

 Performance MS Analysis, publication number 5988-9944EN.

Installing GC Columns 2

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 43/162

5975 Series MSD Operation Manual 43

To install a capillary column in the GC/MSD interface

Agilent 7890A and 7820A, and 6890 GCs

Materials needed

• Column cutter, ceramic (5181-8836) or diamond (5183-4620)

• Ferrules

• 0.3-mm id, for 0.10-mm id columns (5062-3507)

• 0.4-mm id, for 0.20- and 0.25-mm id columns (5062-3508)

• 0.5-mm id, for 0.32-mm id columns (5062-3506)

• 0.8-mm id, for 0.53-mm id columns (5062-3512)

• Flashlight 

• Hand lens (magnifying loupe)

• Gloves, clean

• Large (8650-0030)

• Small (8650-0029)

• Interface column nut (05988-20066)

• Safety glasses

• Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)

Procedure

1 Condition the column (page 41).

2  Vent the MSD (page 82) and open the analyzer chamber (page 84). Be sure

 you can see the end of the GC/MSD interface.

3 If the CI interface is installed, remove the spring-loaded tip seal from the

MSD end of the interface.

4 Slide an interface nut and conditioned ferrule onto the free end of the GC

column. The tapered end of the ferrule must point towards the nut.

CAUTIONNote that the column installation procedure for the 5975 Series MSDs is different from

that for most previous MSDs. Using the procedure from another instrument may result

in poor sensitivity and possible damage to the MSD.

 

2 Installing GC Columns

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 44/162

44 5975 Series MSD Operation Manual

5 Slide the column into the GC/MSD interface (Figure 9) until you can pull it 

out through the analyzer chamber.

6 Break 1 cm off the end of the column (page 32). Do not let any column

fragments fall into the analyzer chamber. They could damage the high

 vacuum pump.

7 Clean the outside of the free end of the column with a lint-free cloth

moistened with methanol.8  Adjust the column so it projects 1 to 2 mm past the end of the interface.

Use the flashlight and hand lens if necessary to see the end of the column

inside the analyzer chamber. Do not use your finger to feel for the column

end.

Figure 9 Installing a capillary column in the GC/MSD interface

1 to 2 mm

GC/MSD interface

(MSD end)

Analyzer chamber

GC/MSD interface

(GC end)

Interface column nut

Column

MSD GC Oven

Installing GC Columns 2

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 45/162

5975 Series MSD Operation Manual 45

9 Hand-tighten the nut. Make sure the position of the column does not change

as you tighten the nut. Reinstall the spring-loaded tip seal if it was removedearlier.

10 Check the GC oven to be sure that the column does not touch the oven

walls.

11 Tighten the nut 1/4 to 1/2 turn. Check the tightness after one or two heat 

cycles.

6850 GC1 Carefully unwind the outlet end of the GC column until the

3 o’clock clip is reached.

2 Slide an interface column nut (part number 05988-20066) and ferrule (part 

number 5062-3508) onto the outlet end of the GC column.

The tapered end of the ferrule must point towards the nut.

3 Slide the column into the GC/MSD interface until the column protrudesinto the analyzer chamber at least 5 cm.

4  Adjust the length of the column from the 3 o’clock clip to the back of the

interface column nut to be 22–28 cm. See Figure 10.

5 Hand tighten the interface nut.

6 Carefully close the oven door while observing to see that the column does

not develop sharp bends or touch the oven walls/floor. Try this procedure

several times.

 

2 Installing GC Columns

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 46/162

46 5975 Series MSD Operation Manual

7 Loosen the interface nut and push the column an additional 3–5 cm into

the analyzer chamber.

8 Make a clean cut of the column so that now only 3–5 cm protrudes into theanalyzer chamber.

9 Clean the outside of the free end of the column with a lint-free cloth

moistened with methanol.

10 Adjust the column so that it protrudes 1 to 2 mm into the analyzer chamber

past the end of the GC/MSD interface, and hand tighten the nut. See

Figure 11.

Make sure the position of the column does not change as you retighten the

nut.

Figure 10 Oven door opened and closed

22–28 cm from 3 o’clock clip to GC/MSD interface nut

Installing GC Columns 2

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 47/162

5975 Series MSD Operation Manual 47

11 Repeat step 6 to assure column integrity.

12 Tighten the interface nut an additional 1/4 to 1/2 turn with a 1/4-inch

open-end wrench.

Check the tightness after one or two heat cycles.

13 Turn the GC on.

14 Verify that the inlet temperature is set to 25 °C.15 Close the analyzer side plate, then reconnect the source power and side

board control cables.

16 Turn on the MSD power switch to initiate MSD pump down.

Press on the side plate of the MSD to achieve a good seal. Verify that the

foreline pump and front fan turn on and that the foreline pump stops

gurgling within 60 seconds.

Figure 11 MSD - GC column connection

1 to 2 mm

GC/MSD interface

(MSD end)

Analyzer chamber

GC/MSD interface

(GC end)

Interface column nut

Column

MSD GC Oven

2 Installing GC Columns

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 48/162

48 5975 Series MSD Operation Manual

17 Reinstall the MSD analyzer cover.

Agilent 5975 Series MSD

Operation Manual

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 49/162

49Agilent Technologies

3Operating in Electron Impact (EI) Mode

Operating the MSD from the Data System 51

Operating the MSD from the LCP 51

LCP Status Messages 53

LCP Menus 55

The EI GC/MSD Interface 58

Before You Turn On the MSD 60

Pumping Down 61

Controlling Temperatures 61

Controlling Column Flow 62

Venting the MSD 63

To view MSD analyzer temperature and vacuum status 64

To set monitors for MSD temperature and vacuum status 66

To set the MSD analyzer temperatures 67

To set the GC/MSD interface temperature from the ChemStation 69

To monitor high vacuum pressure 71

To measure column flow linear velocity 73

To confirm column flow 74To tune the MSD 75

To verify system performance 76

High-Mass Testing (5975 Series MSDs) 77

To remove the MSD covers 80

To vent the MSD 82

To open the analyzer chamber 84

To close the analyzer chamber 87To pump down the MSD 91

To move or store the MSD 93

To set the interface temperature from the GC 95

3 Operating in Electron Impact (EI) Mode

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 50/162

50 5975 Series MSD Operation Manual

How to perform some basic operating procedures for the MSD.

CAUTIONThe software and firmware are revised periodically. If the steps in these procedures do

not match your MSD ChemStation software, refer to the manuals and online help

supplied with the software for more information.

Operating in Electron Impact (EI) Mode 3

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 51/162

5975 Series MSD Operation Manual 51

Operating the MSD from the Data System

The software performs tasks such as pumping down, monitoring pressures,

setting temperatures, tuning, and preparing to vent. These tasks are described

in this chapter. Data acquisition and data analysis are described in the

manuals and online help supplied with the MSD ChemStation software.

Operating the MSD from the LCP

The local control panel (LCP) shows the status of the MSD or initiates a task 

on the MSD without using the Agilent GC/MSD ChemStation.

The GC/MSD ChemStation may be located anywhere on the site local area

network (LAN), so the GC/MSD ChemStation might not be near the

instrument itself. And because the LCP communicates with the GC/MSD

ChemStation via the LAN, you can access GC/MSD ChemStation software

functions, such as tuning and starting a run, right from the MSD.

Modes of operation

The LCP has two modes of operation: Status and Menu.

 Status mode requires no interaction and simply displays the current status of 

the MSD instrument or its various communication connections. If you select 

[Menu], then [No/Cancel], you will be returned to the Status mode.

 Menu mode allows you to query various aspects of the GC/MSD and to initiate

some actions like running a method or sequence or preparing to vent the

system.

To access a particular menu option:

NOTEOnly certain functions are available from the LCP; the GC/MSD ChemStation is the

full-featured controller for most instrument control operations.

Press [Menu] until the desired menu appears.

Press [Item] until the desired menu item appears.

3 Operating in Electron Impact (EI) Mode

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 52/162

52 5975 Series MSD Operation Manual

Use one or more of the following keys as appropriate to respond to prompts or

select options:

 After you make your selection, or if you cycle through all available menus, the

display automatically returns to Status mode.

Pressing [Menu], then [No/Cancel], will always display the Status mode.

Pressing [No/Cancel] twice will always return to the Status mode.

Use [Up] to increase the displayed value or to scroll up (such as in a message list).

Use [Down] to decrease the displayed value or to scroll down (such as in a message

list).

Use [Yes/Select] to accept the current value.

Use [No/Cancel] to return to the Status mode.

Operating in Electron Impact (EI) Mode 3

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 53/162

5975 Series MSD Operation Manual 53

LCP Status Messages

The following messages may be displayed on the LCP to inform you of the

status of the MSD system. If the LCP is currently in Menu mode, cycle through

the menus to return to Status mode.

ChemStation Loading <timestamp>

The Agilent MSD Productivity ChemStation software is starting up.

Executing <type>tune

 A tuning procedure is in progress (type = QuickTune or Autotune).

Instrument Available <timestamp>

The Agilent MSD Productivity ChemStation software is not running.

Loading Method <method name>

Method parameters are being sent to the MSD.

Loading MSD FirmwareThe MSD’s firmware is being initialized.

The following messages alternately appear on the LCP if the MSD does  NOT  

complete its bootup sequence properly:

Server not Found

Check LAN Connection

Seeking Server

Bootp Query xxx

These messages indicate that the MSD has not received its unique IP address

from the Agilent Bootp Service. If the messages persist after you have logged

onto your account on the GC/MSD ChemStation, consult the Troubleshooting

section of the Software Installation manual.

NOTENo messages will be displayed if an online instrument session is not currently running on

the GC/MSD ChemStation.

3 Operating in Electron Impact (EI) Mode

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 54/162

54 5975 Series MSD Operation Manual

Loading OS

The operating system of the instrument controller is being initialized.

<method> Complete <timestamp>

The run and subsequent data processing are done. The same message appears

even if the run was terminated prematurely.

Method Loaded <method name>Method parameters were sent to the MSD.

MS locked by <computer name>

MS parameters can only be changed from the GC/MSD ChemStation.

Press Sideplate

 A reminder during startup to press the MSD sideplate to ensure an adequate

 vacuum seal.

Run: <method> Acquiring <datafile>

 A run is in progress; data is being acquired to the designated data file.

To view system status during startup

1 The following messages are displayed on the LCP display during startup:

• Press sideplate

• Loading OS

• Press sideplate

• Loading MSD Firmware

2 Continue to press the sideplate of the MSD until the MSD Ready message

appears. This helps the instrument to pump down more quickly.

Operating in Electron Impact (EI) Mode 3

LCP M

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 55/162

5975 Series MSD Operation Manual 55

LCP Menus

To access a particular menu option, press [Menu] until the desired menu

appears, then press [Item] until the desired menu item appears. Table 6 

through Table 11 list the menus and selections.

NOTEMany menu items, especially on the ChemStation, MS Parameters, and Maintenance

menus, have no effect when the instrument is acquiring data.

Table 6 ChemStation menu

Action Description

Run Method Displays the current method name and starts an analysis.

Run Sequence Displays the current sequence and starts a sequence.

Run Current Tune Displays the current tune file and starts an autotune (EI mode

only; CI tune must be started from the GC/MSD ChemStation).

# of Messages Displays the number of messages and the text of the most recent

message. Use the arrow keys to scroll through previous messages

(up to 20).

Release ChemStation Disassociates the GC/MSD ChemStation from the MSD.

Connection Status Displays the LAN connection status for the MSD.

Remote = connected to GC/MSD ChemStation online session

Local = not connected to GC/MSD ChemStation online session

Name of Instrument Displays the name of the instrument if connected to GC/MSD

ChemStation online session. The name of the instrument is the

name assigned to the MSD by the GC/MSD ChemStation

Configuration dialogue.

3 Operating in Electron Impact (EI) Mode

T bl 7 M i

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 56/162

56 5975 Series MSD Operation Manual

Table 7 Maintenance menu

Action Description

Prepare to vent Reminds you to shut down the GC then prepares the instrument for

venting when [Yes/Select] is pressed.

Pumpdown Initiates a pumpdown sequence.

Table 8 MS Parameters menu

Action Description

High Vacuum Pressure Only with Micro-Ion vacuum gauge installed.

Turbo Pump Speed Displays the turbo pump speed.

Foreline Pressure Displays the foreline pressure.

MSD Fault Status Reports a summary fault status code (number) in ‘dec’ (decimal) and

‘hex’ (hexadecimal) format covering all possible fault combinations.

Ion Source Temp, oC Displays and sets the ion source temperature.

Mass Filter Temp, oC Displays and sets the mass filter temperature.

CI Reagent Displays CI reagent gas and flow rate (if installed).

NOTEMS parameters cannot be set from the LCP while an online GC/MSD ChemStation session

is connected to the MSD.

Table 9 Network menu

Action Description

MSD IP via BootP Displays the IP address for the MSD.

Gateway IP Address Displays the gateway IP address for the MSD.

Subnet Mask Displays the subnet mask for the MSD.

ChemStation IP Displays the IP address for the GC/MSD ChemStation.

GC IP Address Displays the IP address for the GC.

Ping gateway Checks communication with the gateway.

Operating in Electron Impact (EI) Mode 3

Table 9 Network menu (continued)

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 57/162

5975 Series MSD Operation Manual 57

Ping ChemStation Checks communication with the GC/MSD ChemStation.

Ping GC Checks communication with the GC.

MS Controller MAC Displays the MAC address of the SmartCard in the MSD.

Table 10 Version menu

Action Description

Control firmware Displays the MSD firmware version.

Operating system Displays the GC/MSD ChemStation operating system version.

Front panel Displays the version of the LCP.

Log amplifier Displays version information.

Sideboard Displays the sideboard type.

Mainboard Displays the mainboard type.

Serial number Is assigned to the MSD by GC/MSD ChemStation Configuration

dialogue.

Table 11 Controller menu

Action Description

Reboot controller Starts the LAN/MS control card.

Test LCP? Initiates a diagnostic test of the two-line display.

Test HTTP link to GC/MSD

ChemStation?

Checks the status of the HTTP server.

Table 9 Network menu (continued)

Action Description

3 Operating in Electron Impact (EI) Mode

The EI GC/MSD Interface

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 58/162

58 5975 Series MSD Operation Manual

The EI GC/MSD Interface

The GC/MSD interface (Figure 12) is a heated conduit into the MSD for the

capillary column. It is bolted onto the right side of the analyzer chamber, with

an O-ring seal. It has a protective cover which should be left in place.

One end of the GC/MSD interface passes through the side of the gas

chromatograph and extends into the GC oven. This end is threaded to allow

connection of the column with a nut and ferrule. The other end of the

interface fits into the ion source. The last 1 to 2 millimeters of the capillary column extend past the end of the guide tube and into the ionization chamber.

The GC/MSD interface is heated by an electric cartridge heater. Normally, the

heater is powered and controlled by Thermal Aux #2 heated zone of the GC.

For 6850 Series GCs, the heater is connected to the auxiliary thermal zone.

For the 7820A Series GC’s, the heater is either connected to the rear inlet 

thermal zone for single inlet models or connected to the manual valve thermal

zone for dual inlet models. The interface temperature can be set from the MSDChemStation or from the gas chromatograph. A sensor (thermocouple) in the

interface monitors the temperature.

The GC/MSD interface should be operated in the 250 to 350 C range. Subject 

to that restriction, the interface temperature should be slightly higher than the

maximum GC oven temperature, but never higher than the maximum column

temperature.

The EI GC/MSD interface can only be used with the EI ion source. However,the CI GC/MSD interface can be used with either source.

See Also

“To install a capillary column in the GC/MSD interface” .

WARNING The GC/MSD interface operates at high temperatures. If you touch it when it is hot,it will burn you.

Operating in Electron Impact (EI) Mode 3

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 59/162

5975 Series MSD Operation Manual 59

Figure 12 The EI GC/MSD interface

Ionization

chamber

Analyzer

chamber

Heater/Sensor

assembly

MSD GC oven

Heater sleeve

Insulation

Column

Column end protrudes 1 to 2 mm into the ionization chamber.

3 Operating in Electron Impact (EI) Mode

Before You Turn On the MSD

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 60/162

60 5975 Series MSD Operation Manual

Before You Turn On the MSD

 Verify the following before you turn on or attempt to operate the MSD.

• The vent valve must be closed (the knob turned all the way clockwise).

•  All other vacuum seals and fittings must be in place and fastened correctly.

(The the front side plate screw should not be tightened, unless hazardous

carrier or reagent gasses are being used.

• The MSD is connected to a grounded power source.

• The GC/MSD interface extends into the GC oven.

•  A conditioned capillary column is installed in the GC inlet and in the

GC/MSD interface.

• The GC is on, but the heated zones for the GC/MSD interface, the GC inlet,

and the oven are off.

• Carrier gas of at least 99.9995% purity is plumbed to the GC with the

recommended traps.• If hydrogen is used as carrier gas, carrier gas f low must be off and the front 

sideplate thumbscrew must be loosely fastened.

• The foreline pump exhaust is properly vented.

 

WARNINGThe exhaust from the foreline pump contains solvents and the chemicals you are

analyzing. If using the standard foreline pump, it also contains traces of pump oil. If

you are using toxic solvents or analyzing toxic chemicals, remove the oil trap(standard pump) and install a hose (11-mm id) to take the foreline pump exhaust

outside or to a fume (exhaust) hood. Be sure to comply with local regulations. The

oil trap supplied with the standard pump stops only pump oil. It does not trap or filter

out toxic chemicals.

WARNINGIf you are using hydrogen as a carrier gas, do not start carrier gas flow until the

MSD has been pumped down. If the vacuum pumps are off, hydrogen will

accumulate in the MSD and an explosion may occur. Read “Hydrogen Safety”

before operating the MSD with hydrogen carrier gas.

Operating in Electron Impact (EI) Mode 3

Pumping Down

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 61/162

5975 Series MSD Operation Manual 61

p g

The data system or local control panel helps you pump down the MSD. The

process is mostly automated. Once you close the vent valve and turn on the

main power switch (while pressing on the sideplate), the MSD pumps down by 

itself. The data system software monitors and displays system status during

pumpdown. When the pressure is low enough, the program turns on the ion

source and mass filter heaters and prompts you to turn on the GC/MSD

interface heater. The MSD will shut down if it cannot pump down correctly.

Using the menus or MS monitors, the data system can display:

• Motor speed for turbo pump MSDs (percent spin speed)

• Foreline pressure for diffusion pump MSDs

•  Analyzer chamber pressure (vacuum) for MSDs with the optional G3397A

Micro-Ion Gauge Controller

The LCP can also display these data.

Controlling Temperatures

MSD temperatures are controlled through the data system. The MSD has

independent heaters and temperature sensors for the ion source and

quadrupole mass filter. You can adjust the setpoints and view these

temperatures from the data system or from the local control panel.

Normally, the GC/MSD interface heater is powered and controlled by the

Thermal Aux #2 heated zone of the GC. For the 6850 Series GCs, the heater is

connected to the auxiliary thermal zone. For the 7820 Series GCs, the heater is

either connected to the rear inlet thermal zone for single inlet models or is

connected to the manual valve thermal zone for dual inlet models. The

GC/MSD interface temperature can be set and monitored from the data

system or from the GC.

3 Operating in Electron Impact (EI) Mode

Controlling Column Flow

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 62/162

62 5975 Series MSD Operation Manual

g

Carrier gas flow is controlled by head pressure in the GC. For a given head

pressure, column flow will decrease as the GC oven temperature increases.

With electronic pneumatic control (EPC) and the column mode set to Constant

Flow, the same column flow is maintained regardless of temperature.

The MSD can be used to measure actual column flow. You inject a small 

amount of air or other unretained chemical and time how long it takes to

reach the MSD. With this time measurement, you can calculate the columnflow. See page 73.

Operating in Electron Impact (EI) Mode 3

Venting the MSD

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 63/162

5975 Series MSD Operation Manual 63

 A program in the data system guides you through the venting process. It turns

off the GC and MSD heaters and diffusion pump heater or the turbo pump at 

the correct time. It also lets you monitor temperatures in the MSD and

indicates when to vent the MSD.

The MSD will be damaged by incorrect venting. A diffusion pump will

backstream vaporized pump fluid onto the analyzer if the MDS is vented

before the diffusion pump has fully cooled. A turbo pump will be damaged if it is vented while spinning at more than 50% of its normal operating speed.

WARNINGMake sure the GC/MSD interface and the analyzer zones are cool (below 100 °C)

before you vent the MSD. A temperature of 100 °C is hot enough to burn skin; always

wear cloth gloves when handling analyzer parts.

WARNINGIf you are using hydrogen as a carrier gas, the carrier gas flow must be off before

turning off the MSD power. If the foreline pump is off, hydrogen will accumulate in

the MSD and an explosion may occur. Read “Hydrogen Safety” before operating

the MSD with hydrogen carrier gas.

CAUTION Never vent the MSD by allowing air in through either end of the foreline hose. Use thevent valve or remove the column nut and column.

Do not vent while the turbo pump is still spinning at more than 50%.

Do not exceed the maximum recommended total gas flow. See “5975 series MSD

models and features” .

3 Operating in Electron Impact (EI) Mode

To view MSD analyzer temperature and vacuum status

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 64/162

64 5975 Series MSD Operation Manual

 You can also use the Local Control Panel to perform this task. See the

G1701EA GC/MSD ChemStation Getting Started manual for more

information.

Procedure

1 In Instrument Control view, select Edit Tune Parameters from the Instrument 

menu (Figure 13).

2 Select the tune file you plan to use with your method from the Load MS Tune

File dialog box.

3  Analyzer temperatures and vacuum status are displayed in the Zones field.

Figure 13 Tune parameters

Operating in Electron Impact (EI) Mode 3

Unless you have just begun the pumpdown process, the foreline pressure

should be less than 300 mTorr or the turbo pump should be running at least

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 65/162

5975 Series MSD Operation Manual 65

should be less than 300 mTorr, or the turbo pump should be running at least 

80% speed. MSD heaters remain off as long as the diffusion pump is cold or theturbo pump is operating at less than 80%. Normally, the foreline pressure will

be below 100 mTorr, or the turbo pump speed will be at 100%.

The MSD heaters turn on at the end of the pumpdown cycle and turn off at the

beginning of the vent cycle. The reported setpoints will not change during

 venting or pumpdown, even though both the MSD zones are turned off.

3 Operating in Electron Impact (EI) Mode

To set monitors for MSD temperature and vacuum status

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 66/162

66 5975 Series MSD Operation Manual

 A monitor displays the current value of a single instrument parameter. They 

can be added to the standard instrument control window. Monitors can be set 

to change color if the actual parameter varies beyond a user-determined limit 

from its setpoint.

Procedure

1 Select MS Monitors from the Instrument menu.

2 In the Edit MS Monitors box, under Type, select Zone.3 Under Parameter, select MS Source and click Add.

4 Under Parameter, select MS Quad and click Add.

5 Under Parameter, select Foreline (or TurboSpd) and click Add.

6 Select any other monitors you want and Add them.

7 Click OK. The new monitors will be stacked on top of each other in the lower

right corner of the Instrument Control window. They must be moved for youto see them all.

8 Click and drag each monitor to the desired position. See Figure 14 for one

way of arranging the monitors.

9 To make the new settings part of the method, select Save from the Method

menu.

Figure 14 Arranging monitors

Operating in Electron Impact (EI) Mode 3

To set the MSD analyzer temperatures

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 67/162

5975 Series MSD Operation Manual 67

Setpoints for the MSD ion source and mass filter (quad) temperatures are

stored in the current tune (*.u) file. When a method is loaded, the setpoints in

the tune file associated with that method are downloaded automatically.

Procedure

1 In Instrument Control view, select Edit Tune Parameters from the Instrument 

menu.2 Select Temperatures from the MoreParams menu (Figure 15).

3 Type the desired Source and Quad (mass filter) temperatures in the

setpoint fields. See Table 12 for recommended setpoints.

The GC/MSD interface, ion source, and quadrupole heated zones interact.

The analyzer heaters may not be able to accurately control temperatures if 

the setpoint for one zone is much different from that of an adjacent zone.

Figure 15 Setting temperatures

WARNINGDo not exceed 200 °C for the quadrupole or 350 °C for the source.

3 Operating in Electron Impact (EI) Mode

4 To close the screen, click:

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 68/162

68 5975 Series MSD Operation Manual

• Apply to send the new temperature setpoints to the MSD.• OK to change the currently loaded tune file but not download anything to

the MSD (use Apply).

• Cancel to exit the panel without changing the currently loaded tune file

or downloading anything to the MSD.

5 When the Save MS Tune File dialog box appears, either click OK to save your

changes to the same file or type a new file name and click OK.

Table 12 Recommended temperature settings

EI operation PCI operation NCI operation

MS Source 230 250 150

MS Quad 150 150 150

Operating in Electron Impact (EI) Mode 3

To set the GC/MSD interface temperature from the ChemStation

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 69/162

5975 Series MSD Operation Manual 69

 You can also use the Local Control Panel to perform this task. See “Operating

the MSD from the LCP” .

Procedure

1 Select View>Instrument Control.

2 Select Instrument>GC Edit Parameters.

3 Click the Aux icon to edit the interface temperature (Figure 16).

4 Check the heater On and type the setpoint in the Value °C column.

The typical setpoint is 280 °C. The limits are 0 °C and 350 °C. A setpoint 

below ambient temperature turns off the interface heater.

Figure 16 Setting the interface temperature

3 Operating in Electron Impact (EI) Mode

CAUTION

Never exceed the maximum temperature for your column.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 70/162

70 5975 Series MSD Operation Manual

5 Click Apply to download setpoints or click OK to download setpoints and

close the window.

6 To make the new settings part of the method, select Save from the Method

menu.

CAUTION

CAUTIONMake sure that the carrier gas is turned on and the column has been purged of air

before heating the GC/MSD interface or the GC oven.

Operating in Electron Impact (EI) Mode 3

To monitor high vacuum pressure

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 71/162

5975 Series MSD Operation Manual 71

Pressure monitoring requires an optional G3397A Micro-Ion vacuum gauge.

Materials needed

• Micro-Ion vacuum gauge (G3397A)

Procedure

1 Start up and pump down the MSD (page 91).

2 In the Tune and Vacuum Control view select Turn Vacuum Gauge on/off from

the Vacuum menu.

3 In the Instrument Control view you can set up an MS Monitor for reading.

The vacuum can also be read on the LCP or from the Manual Tune screen.

The largest influence on operating pressure in EI mode is the carrier gas

(column) flow. Table 13 lists typical pressures for various helium carrier gas

flows. These pressures are approximate and will vary from instrument toinstrument by as much as 30%.

WARNINGIf you are using hydrogen as a carrier gas, do not turn on the Micro-Ion vacuum

gauge if there is any possibility that hydrogen has accumulated in the analyzer

chamber. Read “Hydrogen Safety” before operating the MSD with hydrogen carrier

gas.

3 Operating in Electron Impact (EI) Mode

Table 13 Micro-Ion Vacuum Gauge Reading

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 72/162

72 5975 Series MSD Operation Manual

If the pressure is consistently higher than those listed, refer to the online help

in the MSD ChemStation software for information on troubleshooting air leaks

and other vacuum problems.

Column flow rate,

mL/min

Gauge reading, Torr

Performance turbo

pump

Gauge reading, Torr

Standard turbo

pump

Gauge reading, Torr

Diffusion pump

Foreline reading, Torr

Diffusion pump

0.5 3.18E–06 1.3E–05 2.18E–05 34.7

0.7 4.42E–06 1.83E–05 2.59E–05 39.4

1 6.26E–06 2.61E–05 3.66E–05 52.86

1.2 7.33E–06 3.11E–05 4.46E–05 60.866

2 1.24E–05 5.25E–05 7.33E–05 91.784

3 1.86E–05 8.01E–05 1.13E–04 125.76

4 2.48E–05

6 3.75E–05

Operating in Electron Impact (EI) Mode 3

To measure column flow linear velocity

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 73/162

5975 Series MSD Operation Manual 73

With capillary columns, such as those used with the MSD, linear velocity is

often measured rather than volumetric flow rate.

Procedure

1 Set Data Acquisition for splitless manual injection and selected ion

monitoring (SIM) of m/z 28.

2 Press Prep Run on the GC keypad.

3 Inject 1 µL of air into the GC inlet and press Start Run.

4 Wait until a peak elutes at m/z 28. Note the retention time.

5 Calculate the average linear velocity.

 Average linear velocity (cm/s) =

where:

 L = Length of the column in meters

t = Retention time in seconds

Be sure to account for any pieces of column broken off. A 1-meter section

missing from a 25-meter column can yield a 4% error.

6 Use this velocity to verify the MSD ChemStation flow calculations (page 74).

If the numbers disagree, click Change to calibrate the column dimensions.

7 To calculate the volumetric flow rate.

Volumetric flow rate (mL/min) =

where:

 D = Internal column diameter in millimeters

 L = Column length in meters

t = Retention time in minutes

100 L

t --------------

0.785 D2  L

----------------------------

3 Operating in Electron Impact (EI) Mode

To confirm column flow

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 74/162

74 5975 Series MSD Operation Manual

 Volumetric flow can be calculated from the column head pressure if the

column dimensions are known.

Procedure

1 In the Instrument Control view, select Instrument>GC Edit Parameters.

2 Click the Columns icon (Figure 17 shows an example).

3 Select the appropriate column. .

Figure 17 Calculating column flow

Operating in Electron Impact (EI) Mode 3

To tune the MSD

Y l h L l C l P l h h i l

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 75/162

5975 Series MSD Operation Manual 75

 You can also use the Local Control Panel to run the autotune that is currently 

loaded in the PC memory. See “Operating the MSD from the LCP” .

Procedure

1 In the Instrument Control View, verify the correct tune file is loaded. For

most applications, ATUNE.U (Autotune) gives good results. STUNE.U

(Standard Tune) is not recommended as it may reduce sensitivity.Consider Gain autotune (GAIN.U + HiSense.U). This tunes to a target gain

rather than a target abundance. It offers excellent reproducibility, both of 

run-to-run abundance but also between different instruments,

2 Set the system to the same conditions (GC oven temperature and column

flow, and MSD analyzer temperatures) that will be used for data

acquisition.

3 Select Tune MSD to perform a complete tune, or select Quick Tune to adjust peak width, mass assignment, and abundance, without changing ion ratios.

If your system is configured for CI, you will be able to access the CI Tune

panel from this box. The tune will start immediately.

4 Wait for the tune to complete and to generate the report.

Save your tune reports. To view history of tune results, select 

Checkout>View Previous Tunes....

To manually tune your MSD or to perform special autotunes, go to the Tune

and Vacuum Control View.

From this Tune menu, in addition to the tunes available from Instrument 

Control, you can select special autotunes for specific spectral results, such as

DFTPP Tune or BFB Tune.

See the manuals or online help provided with your MSD ChemStation software

for additional information about tuning.

3 Operating in Electron Impact (EI) Mode

To verify system performance

Materials needed

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 76/162

76 5975 Series MSD Operation Manual

Materials needed

• 1 pg/µL (0.001 ppm) OFN sample (5188-5348)

Verify the tune performance

1  Verify that the system has been pumping down for at least 60 minutes.

2 Set the GC oven temperature to 150 °C and the column flow to 1.0 mL/min.3 In the Instrument Control view, select Checkout Tune from the Checkout 

menu. The software will perform an autotune and print the report.

4 When the autotune has completed, save the method and then select 

Evaluate Tune from the Checkout menu.

The software will evaluate the last autotune and print a System Verification

– Tune report.

Verify the sensitivity performance

1 Set up to inject 1 µL of OFN, either with the ALS or manually.

2 In the Instrument Control view, select Sensitivity Check from the Checkout 

menu.

3 Click the appropriate icons in the Instrument Edit window to edit the

method for the type of injection.

4 Click OK to run the method.

When the method is completed, an evaluation report will be printed.

 Verify that rms signal-to-noise ratio meets the published specification.

Please see the Agilent Web site at www.agilent.com/chem for specifications.

Operating in Electron Impact (EI) Mode 3

High-Mass Testing (5975 Series MSDs)

Setup conditions

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 77/162

5975 Series MSD Operation Manual 77

Setup conditions

1 Obtain a sample of PFHT (5188-5357).

2 Load tune file ATUNE.U then auto tune the MSD.

3 Resolve the PFHT.M method under x\5975\PFHT.M where x is instrument 

number being used.

4 Update and save the method.

High-mass checkout

1 Load sample into a vial and place in position 2.

2 Select High Mass Check from the Checkout menu.

3 Follow the instructions on screen.

4 The Run is completed and results are printed within 5 minutes.

3 Operating in Electron Impact (EI) Mode

Results

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 78/162

78 5975 Series MSD Operation Manual

Figure 18 PFHT high mass report

Operating in Electron Impact (EI) Mode 3

Results will indicate the recommended amount to adjust AMU offset for

high-mass. If your results are within 5 units of the targeted amount, there is no

need to make adjustments.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 79/162

5975 Series MSD Operation Manual 79

Adjustments

1  Verify ATUNE.U has been loaded.

2 Select Edit Tune Parameters from the Instrument menu via Instrument 

Control.

3 Click on MoreParams and select DynamicRamping Params...

a Select AMU offset from the drop down box.

b If the values on the right side are greyed out then select the Enable

Dynamic Ramping For This Lens checkbox.

c Enter in the recommend offset and click OK.

4 Click OK on the Edit Parameters box. The Save MS Tune File dialog box

appears.

 You can overwrite the existing ATUNE.U to include high-mass adjustment 

or save this file to a new name, for example, ATUNEHIGH.U.

5 Load the PFHT.M and the saved tune file, then save the method.

6 Rerun test mixture (repeat high-mass checkout). If the correction is within

5 units, no further adjustments are required.

NOTEAnytime an ATUNE.U is performed it will overwrite the AMU offset which was entered. This

is the reason for renaming the tune.

3 Operating in Electron Impact (EI) Mode

To remove the MSD covers

Materials needed

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 80/162

80 5975 Series MSD Operation Manual

• Screwdriver, Torx T-15 (8710-1622)

If you need to remove one of the MSD covers, follow these procedures

(Figure 19):

To remove the analyzer top cover

Remove the five screws and lift the cover off.

To remove the analyzer window cover

1 Press down on the rounded area on the top of the window.

2 Lift the window forward and off the MSD.

 

 

WARNINGDo not remove any other covers. Dangerous voltages are present under other covers.

Operating in Electron Impact (EI) Mode 3

Analyzer window cover

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 81/162

5975 Series MSD Operation Manual 81

Figure 19 Removing covers

Latch

Analyzer cover

Left side cover

CAUTION Do not use excessive force or the plastic tabs that hold the cover to the mainframe willbreak off.

3 Operating in Electron Impact (EI) Mode

To vent the MSD

Procedure

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 82/162

82 5975 Series MSD Operation Manual

1 Select Vent from the Vacuum menu in the software. Follow the instructions

presented.

2 Set the GC/MSD interface heater and the GC oven temperatures to ambient 

(room temperature).

3 When prompted, turn off the MSD power switch.

4 Unplug the MSD power cord.

WARNINGIf you are using hydrogen as a carrier gas, the carrier gas flow must be off before

turning off the MSD power. If the foreline pump is off, hydrogen will accumulate in

the MSD and an explosion may occur. Read “Hydrogen Safety” before operating

the MSD with hydrogen carrier gas.

CAUTION Be sure the GC oven and the GC/MSD interface are cool before turning off carrier gasflow.

WARNINGWhen the MSD is vented, do not put the ChemStation into Instrument Control view.

Doing so will turn on the interface heater.

Operating in Electron Impact (EI) Mode 3

5 Remove the analyzer window cover (page 80)

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 83/162

5975 Series MSD Operation Manual 83

6 Turn the vent valve knob (Figure 20) counterclockwise only 3/4 turns or

until you hear the hissing sound of air flowing into the analyzer chamber.

Do not turn the knob too far or the O-ring may fall out of its groove. Be sure

to retighten the knob before pumping down.

Figure 20 Venting the MSD

Vent valve knob YES NO

 

WARNINGAllow the analyzer to cool to near room temperature before touching it.

CAUTION Always wear clean gloves while handling any parts that go inside the analyzerchamber.

WARNINGWhen the MSD is vented, do not put the ChemStation into Instrument Control view.

Doing so will turn on the interface heater.

3 Operating in Electron Impact (EI) Mode

To open the analyzer chamber

Materials needed

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 84/162

84 5975 Series MSD Operation Manual

• Gloves, clean, lint-free

• Large (8650-0030)

• Small (8650-0029)

• Wrist strap, antistatic

• Small (9300-0969)• Medium (9300-1257)

• Large (9300-0970)

 

Procedure

1  Vent the MSD (page 82).

2 Disconnect the side board control cable and the source power cable from

the side board.

3 Loosen the side plate thumbscrews (Figure 21) if they are fastened.

The rear side plate thumbscrew should be unfastened during normal use. It 

is only fastened during shipping. The front side plate thumbscrew should

only be fastened for CI operation or if hydrogen or other flammable or toxic

substances are used for carrier gas.

4 Gently swing the side plate out.

 

CAUTIONElectrostatic discharges to analyzer components are conducted to the side board

where they can damage sensitive components. Wear a grounded antistatic wrist strap

and take other antistatic precautions (see page 131) before you open the analyzer

chamber.

 

CAUTION In the next step, if you feel resistance, stop. Do not try to force the side plate open.Verify that MSD is vented. Verify that both the front and rear side plate screws are

completely loose.

Operating in Electron Impact (EI) Mode 3

WARNINGThe analyzer, GC/MSD interface, and other components in the analyzer chamber

operate at very high temperatures. Do not touch any part until you are sure it is cool.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 85/162

5975 Series MSD Operation Manual 85

CAUTIONAlways wear clean gloves to prevent contamination when working in the analyzer

chamber.

3 Operating in Electron Impact (EI) Mode

Thumbscrews

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 86/162

86 5975 Series MSD Operation Manual

Figure 21 The analyzer chamber

Side plate

Analyzer cover

Detector

Feedthrough board

Ion source

Analyzer

Side plate

CHAMBER CLOSED

CHAMBER OPEN

Operating in Electron Impact (EI) Mode 3

To close the analyzer chamber

Materials needed

Gl l li f

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 87/162

5975 Series MSD Operation Manual 87

• Gloves, clean, lint-free

• Large (8650-0030)

• Small (8650-0029)

Procedure

1 Make sure all the internal analyzer electrical leads are correctly attached.

Wiring is the same for both the EI and CI sources.

The wiring is described in Table 14 and illustrated in Figure 22 and

Figure 23. The term “Board” in the table refers to the feedthrough board

located next to the ion source.

Table 14 Analyzer wiring

Wire description Attached to Connects to

Green beaded (2) Quad heater Board, top left (HTR)

White with braided cover (2) Quad sensor Board, top (RTD)

White (2) Board, center (FILAMENT-1) Filament 1 (top)

Red (1) Board, center left (REP) Repeller

Black (2) Board, center (FILAMENT-2) Filament 2 (bottom)

Orange (1) Board, top right (ION FOC) Ion focus lens

Blue (1) Board, top right (ENT LENS) Entrance lens

Green beaded (2) Ion source heater Board, bottom left (HTR)

White (2) Ion source sensor Board, bottom (RTD)

3 Operating in Electron Impact (EI) Mode

HTR

QUADRUPOLE

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 88/162

88 5975 Series MSD Operation Manual

Figure 22 Feedthrough board wiring

HTRRTS

SOURCE

FILAMENT - 2

FILAMENT - 1

REP

IONFOC

ENTRLENS

RTS

White wires to

filament 1

Red wire

to repeller

Black wires to

filament 2

Ion source heater

wires (green)

Blue wire to

entrance lens

Orange wire to

ion focus lens

Ion source sensor

wires (white)

Operating in Electron Impact (EI) Mode 3

FB = Feedthrough Board

Repeller

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 89/162

5975 Series MSD Operation Manual 89

2 Check the side plate O-ring.

Make sure the O-ring has a very light coat of Apiezon L high vacuum grease.

If the O-ring is very dry, it may not seal well. If the O-ring looks shiny, it has

too much grease on it. (Refer to the 5975 Series MSD Troubleshooting and

Maintenance Manual for lubricating instructions.)

Figure 23 Ion source wiring

Ion source

heater wires

Ion source

sensor wires

p

(red wire

from FB)

Filament 1

(white wires

from FB)

Filament 2

(black wires

from FB)

Ion focus lens

(orange wire

from FB)

Entrance lens

(blue wire

from FB)

3 Operating in Electron Impact (EI) Mode

3 Close the side plate.

4 Reconnect the side board control cable and source power cable to the side

board.

5 Make sure the vent valve is closed.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 90/162

90 5975 Series MSD Operation Manual

6 Pump down the MSD (page 91).

7 If you are operating in CI mode or if hydrogen or other flammable or toxic

substance is used for carrier gas, gently hand tighten the front side plate

thumbscrew.

8 Once the MSD has pumped down, close the analyzer cover.

WARNINGThe front thumbscrew must be fastened for CI operation or if hydrogen (or other

hazardous gas) is being used as the GC carrier gas. In the unlikely event of an

explosion, it may prevent the side plate from opening.

CAUTIONDo not overtighten the thumbscrew; it can cause air leaks or prevent successful

pumpdown. Do not use a screwdriver to tighten the thumbscrew.

Operating in Electron Impact (EI) Mode 3

To pump down the MSD

 You can also use the Local Control Panel to perform this task. See “Operating

the MSD from the LCP” .

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 91/162

5975 Series MSD Operation Manual 91

Procedure

1 Close the vent valve.

2 Plug in the MSD power cord.

3 Open the MSD Analyzer top cover.

4 Turn on the MSD while engaging the side plate to the manifold using hand

pressure.

5 Press lightly on the side board to ensure a correct seal. Press on the metal

box on the side board.

The foreline pump will make a gurgling noise. This noise should stop within

a minute. If the noise continues, there is a large air leak in your system,

probably at the side plate seal, the interface column nut, or the vent valve.

6 Start the ChemStation and select Tune and Vacuum Control from the View

menu.

7 Select Pump Down from the Vacuum menu.

WARNINGMake sure your MSD meets all the conditions listed in the introduction to this

chapter (page 60) before starting up and pumping down the MSD. Failure to do so

can result in personal injury.

WARNINGIf you are using hydrogen as a carrier gas, do not start carrier gas flow until the

MSD has been pumped down. If the vacuum pumps are off, hydrogen will

accumulate in the MSD and an explosion may occur. Read “Hydrogen Safety”

before operating the MSD with hydrogen carrier gas.

 

3 Operating in Electron Impact (EI) Mode

8 Once communication with the PC has been established, click OK.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 92/162

92 5975 Series MSD Operation Manual

9 When prompted, turn on the GC/MSD interface heater and GC oven. Click 

OK when you have done so.

The software will turn on the ion source and mass filter (quad) heaters. The

temperature setpoints are stored in the current autotune (*.u) file.

10 After the message Okay to run appears, wait 2 hours for the MSD to reach

thermal equilibrium. Data acquired before the MSD has reached thermal

equilibrium may not be reproducible.

Figure 24 Pumping down

CAUTIONWithin 10 to 15 minutes the diffusion pump should be hot, or the turbo pump speed

should be up to 80% (Figure 24). The pump speed should eventually reach 95%. If these

conditions are not met, the MSD electronics will shut off the foreline pump. In order torecover from this condition, you must power cycle the MSD. If the MSD does not pump

down correctly, see the manual or online help for information on troubleshooting air

leaks and other vacuum problems.

CAUTIONDo not turn on any GC heated zones until carrier gas flow is on. Heating a column with

no carrier gas flow will damage the column.

Operating in Electron Impact (EI) Mode 3

To move or store the MSD

Materials needed

• Ferrule, blank (5181-3308)

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 93/162

5975 Series MSD Operation Manual 93

• Interface column nut (05988-20066)

• Wrench, open-end, 1/4-inch × 5/16-inch (8710-0510)

Procedure

1  Vent the MSD (page 82).

2 Remove the column and install a blank ferrule and interface nut.

3 Tighten the vent valve.

4 Move the MSD away from the GC (see the 5975 Series MSD Troubleshooting

and Maintenance Manual).

5 Unplug the GC/MSD interface heater cable from the GC.

6 Install the interface nut with the blank ferrule.

7 Open the analyzer cover (page 80).

8 Finger-tighten the side plate thumbscrews (Figure 25).

9 Plug the MSD power cord in.

10 Switch the MSD on to establish a rough vacuum. Verify that the turbo pump

speed is greater than 50% or that the foreline pressure is 1 Torr.

11 Switch the MSD off.

12Close the analyzer cover.

13 Disconnect the LAN, remote, and power cables.

CAUTIONDo not overtighten the side plate thumbscrews. Overtightening will strip the threads in

the analyzer chamber. It will also warp the side plate and cause leaks.

3 Operating in Electron Impact (EI) Mode

Front thumbscrew

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 94/162

94 5975 Series MSD Operation Manual

The MSD can now be stored or moved. The foreline pump cannot be

disconnected; it must be moved with the MSD. Make sure the MSD remains

upright and is never tipped on its side or inverted.

Figure 25 Side plate thumbscrews

Rear thumbscrew

CAUTION The MSD must remain upright at all times. If you need to ship your MSD to anotherlocation, contact your Agilent Technologies service representative for advice about

packing and shipping.

Operating in Electron Impact (EI) Mode 3

To set the interface temperature from the GC

If desired, the interface temperature can be set directly at the GC. For the

 Agilent 7890A and 6890, set the Aux #2 temperature. For the 6850, use the

optional handheld controller to set the thermal aux temperature. Refer to the

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 95/162

5975 Series MSD Operation Manual 95

GC User documentation for details.

If you want the new setpoint to become part of the current method, click Save under the Method menu. Otherwise, the first time a method is loaded, all the

setpoints in the method will overwrite those set from the GC keyboard.

CAUTIONNever exceed the maximum temperature of your column.

CAUTIONMake sure that the carrier gas is turned on and the column has been purged of air

before heating the GC/MSD interface or the GC oven.

3 Operating in Electron Impact (EI) Mode

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 96/162

96 5975 Series MSD Operation Manual

Agilent 5975 Series MSD

Operation Manual

4Operating in Chemical Ionization (CI)

Mode

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 97/162

97Agilent Technologies

Mode

General Guidelines 98

The CI GC/MSD Interface 99

To Operate the CI MSD 101To switch from the EI source to the CI source 102

To pump down the CI MSD 103

To set up the software for CI operation 104

To operate the reagent gas flow control module 106

To set up methane reagent gas flow 109

To use other reagent gases 111

To switch from the CI source to the EI source 115

CI Autotune 116

To perform a PCI autotune (methane only) 118

To perform an NCI autotune (methane reagent gas) 120

To verify PCI performance 122

To verify NCI performance 123

To monitor high vacuum pressure 124

This chapter provides information and instructions for operating the

5975 Series CI MSDs in Chemical Ionization (CI) mode. Most of the

information in the preceding chapter is also relevant.

Most of the material is related to methane chemical ionization but one sectiondiscusses the use of other reagent gases.

The software contains instructions for setting the reagent gas flow and for

performing CI autotunes. Autotunes are provided for positive CI (PCI) with

methane reagent gas and for negative CI (NCI) with any reagent gas.

4 Operating in Chemical Ionization (CI) Mode

General Guidelines

•  Always use the highest purity methane (and other reagent gases, if 

applicable.) Methane must be at least 99.9995% pure.

•  Always verify the MSD is performing well in EI mode before switching to CI.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 98/162

98 5975 Series MSD Operation Manual

See “To verify system performance” .

• Make sure the CI ion source and GC/MSD interface tip seal are installed.

• Make sure the reagent gas plumbing has no air leaks. This is determined in

PCI mode, checking for m/z 32 after the methane pretune.

Operating in Chemical Ionization (CI) Mode 4

The CI GC/MSD Interface

The CI GC/MSD interface (Figure 26) is a heated conduit into the MSD for the

capillary column. It is bolted onto the right side of the analyzer chamber, with

an O-ring seal and has a protective cover which should be left in place.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 99/162

5975 Series MSD Operation Manual 99

One end of the interface passes through the side of the GC and extends into

the oven. It is threaded to allow connection of the column with a nut and

ferrule. The other end of the interface fits into the ion source. The last 1 to

2 millimeters of the capillary column extend past the end of the guide tube

and into the ionization chamber.

Reagent gas is plumbed into the interface. The tip of the interface assembly 

extends into the ionization chamber. A spring-loaded seal keeps reagent gases

from leaking out around the tip. The reagent gas enters the interface body and

mixes with carrier gas and sample in the ion source.

The GC/MSD interface is heated by an electric cartridge heater. Normally, the

heater is powered and controlled by Thermal Aux #2 heated zone of the GC.

For 6850 Series GCs, the heater is connected to the auxiliary thermal zone.

The interface temperature can be set from the MSD ChemStation or from the

gas chromatograph. A sensor (thermocouple) in the interface monitors the

temperature.

This interface is also used for EI operation in CI MSDs.

The interface should be operated in the 250 to 350 C range. Subject to that 

restriction, the interface temperature should be slightly higher than themaximum GC oven temperature, but never higher than the maximum column

temperature.

See Also

“To install a capillary column in the GC/MSD interface” .

CAUTIONNever exceed the maximum column temperature, either in the GC/MSD interface, the

GC oven, or the inlet.

4 Operating in Chemical Ionization (CI) Mode

WARNINGThe GC/MSD interface operates at high temperatures. If you touch it when it is hot,

it will burn you.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 100/162

100 5975 Series MSD Operation Manual

Figure 26 The CI GC/MSD interface

Spring-loadedseal

Reagent gas in

MSD GC oven

Column end protrudes 1 to 2 mm into the ionization chamber.

Operating in Chemical Ionization (CI) Mode 4

To Operate the CI MSD

Operating your MSD in the CI mode is slightly more complicated than

operating in the EI mode. After tuning, gas flow, source temperature

(Table 15), and electron energy may need to be optimized for your specific

analyte.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 101/162

5975 Series MSD Operation Manual 101

y

Start the system in PCI mode

By bringing the system up in PCI mode first, you will be able to do the

following:

• Set up the MSD with methane first, even if you are going to use another

reagent gas.

• Check the interface tip seal by looking at the m/z 28 to 27 ratio (in the

methane flow adjust panel).

• Tell if a gross air leak is present by monitoring the ions at m/z 19

(protonated water) and 32.

• Confirm that the MS is generating “real” ions and not just background

noise.

It is nearly impossible to perform any diagnostics on the system in NCI. In

NCI, there are no reagent gas ions to monitor. It is difficult to diagnose an air

leak and difficult to tell whether a good seal is being created between the

interface and the ion volume.

Table 15 Temperatures for CI operation

Ion source Quadrupole GC/MSDinterface

PCI 250 °C 150 °C 280 °C

NCI 150 °C 150 °C 280 °C

4 Operating in Chemical Ionization (CI) Mode

To switch from the EI source to the CI source

CAUTIONAlways verify MSD performance in EI before switching to CI operation.

Always set up the CI MSD in PCI first, even if you are going to run NCI.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 102/162

102 5975 Series MSD Operation Manual

Procedure

1  Vent the MSD. See page 82.2 Open the analyzer.

3 Remove the EI ion source. See page 134.

4 Install the CI ion source. See page 142.

5 Install the interface tip seal. See page 143.

6 Close the analyzer.

7 Pump down the MSD. See page 103.

CAUTIONElectrostatic discharges to analyzer components are conducted to the side board

where they can damage sensitive components. Wear a grounded antistatic wrist strap.

See “Electrostatic discharge” . Take antistatic precautions before you open the

analyzer chamber.

Operating in Chemical Ionization (CI) Mode 4

To pump down the CI MSD

 You can also use the Local Control Panel to perform this task. See “Operatingthe MSD from the LCP” .

Procedure

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 103/162

5975 Series MSD Operation Manual 103

1 Follow the instructions for the EI MSD. See “To pump down the MSD” .

 After the software prompts you to turn on the interface heater and GC

oven, perform the following steps.2 Check the vacuum gauge, if present, to verify that the pressure is

decreasing.

3 Press Shutoff Valve to close the gas supply and shutoff valves.

4  Verify that PCICH4.U is loaded and accept the temperature setpoints.

 Always start up and verify system performance in PCI mode before

switching to NCI.

5 Set the GC/MSD interface to 280 °C.

6 Set Gas A to 20%.

7 Let the system bake out and purge for at least 2 hours. If you will be

running NCI, best sensitivity, bake out the MSD overnight.

4 Operating in Chemical Ionization (CI) Mode

To set up the software for CI operation

Procedure

1 Switch to the Tune and Vacuum Control view.

2 Select Load Tune Values from the File menu.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 104/162

104 5975 Series MSD Operation Manual

3 Select the tune file PCICH4.U.

4 If CI autotune has never been run for this tune file, the software will

prompt you through a series of dialog boxes. Accept the default values

unless you have a very good reason for changing anything.

The tune values have a dramatic effect on MSD performance. Always start 

with the default values when first setting up for CI, and then make

adjustments for your specific application. See Table 16 for default values

for the Tune Control Limits box.

NOTEThese limits are used by Autotune only. They should not be confused with the parameters

set in Edit MS Parameters or with those appearing on the tune report.

Operating in Chemical Ionization (CI) Mode 4

Table 16 Default Tune Control Limits, used by CI autotune only

Reagent gas Methane Isobutane Ammonia

Ion polarity Positive Negative Positive Negative Positive Negative

Abundance target 1x106 1x106 N/A 1x106 N/A 1x106

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 105/162

5975 Series MSD Operation Manual 105

Notes for Table 16:

• N/A Not available. There are no PFDTD ions formed in PCI with any 

reagent gas but methane, hence, CI autotune is not available with theseconfigurations.

• Ion polarity Always set up in PCI with methane first, then switch to your

desired ion polarity and reagent gas.

• Abundance target Adjust higher or lower to get desired signal abundance.

Higher signal abundance also gives higher noise abundance. This is

adjusted for data acquisition by setting the EMV in the method.

• Peakwidth target Higher peakwidth values give better sensitivity, lower values give better resolution.

• Maximum emission current Optimum emission current maximum for NCI

is very compound-specific and must be selected empirically. Optimum

emission current for pesticides, for example, may be about 200 µA.

Peakwidth target 0.6 0.6 N/A 0.6 N/A 0.6

Maximum repeller 4 4 N/A 4 N/A 4

Maximum emissioncurrent, µA

240 50 N/A 50 N/A 50

Max electron energy, eV 240 240 N/A 240 N/A 240

4 Operating in Chemical Ionization (CI) Mode

To operate the reagent gas flow control module

Reagent gas flows are controlled in software (Figure 27).

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 106/162

106 5975 Series MSD Operation Manual

The Valve Settings have the following effects:

Gas A (or B) Valve The present gas flow, if any, is turned off. The system

evacuates the gas lines for 6 minutes, then turns on the selected gas (A or B).

This is to reduce cross-mixing of the gases in the lines.

Shutoff Valve When Shutoff Valve is selected, the system turns off the present 

gas flow while leaving the shutoff valve (Figure 28) open. This is to remove any 

residual gas in the lines. Typical evacuation time is 6 minutes and then the

shutoff valve is closed.

The flow control hardware remembers the flow setting for each gas. When

either gas is selected, the control board automatically sets the same flow that 

was used for that gas last time.

Figure 27 CI flow control

Operating in Chemical Ionization (CI) Mode 4

The flow control module

The CI reagent gas flow control module (Figure 28 and Table 17) regulates the

flow of reagent gas into the CI GC/MSD interface. The flow module consists of a mass flow controller (MFC), gas select valves, CI calibration valve, shutoff 

 valve, control electronics, and plumbing.

The back panel provides Swagelok inlet fittings for methane (CH4) and one

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 107/162

5975 Series MSD Operation Manual 107

OTHER reagent gas. The software refers to them as Gas A and Gas B, 

respectively. If you are not using a second reagent gas, cap the OTHER fitting to

prevent accidental admission of air to the analyzer. Supply reagent gases at 

25 to 30 psi (170 to 205 kPa).

The shutoff valve prevents contamination of the flow control module by 

atmosphere while the MSD is vented or by PFTBA during EI operation. The

MSD monitors will reflect On as 1 and Off as 0 (see Table 17).

Figure 28 Reagent gas flow control module schematic

Gas A

(methane)

supply

Gas B

(other)

supply

Gas A

select valve

Gas B

select valve

Mass

flowcontroller

Calibration

valve

Restrictor

Calibration

vial

Shutoff

valve

GC/MSD

interface

GC column

CI ion

source

4 Operating in Chemical Ionization (CI) Mode

Table 17 Flow control module state diagram

Result Gas A flow Gas B flow Purge

with Gas A

Purge

with Gas B

Pump out

flow module

Standby,

vented, or EI

mode

Gas A Open Closed Open Closed Closed Closed

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 108/162

108 5975 Series MSD Operation Manual

The Open and Closed states are shown in the monitors as 1 and 0 respectively.

Gas B Closed Open Closed Open Closed Closed

MFC On setpoint On setpoint On 100% On 100% On 100% Off 0%

Shutoff valve Open Open Open Open Open Closed

Operating in Chemical Ionization (CI) Mode 4

To set up methane reagent gas flow

The reagent gas flow must be adjusted for maximum stability before tuning theCI system. Do the initial setup with methane in positive chemical ionization

(PCI) mode. No flow adjustment procedure is available for NCI, as no negative

reagent ions are formed.

Adj ti th th t fl i th t tti th

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 109/162

5975 Series MSD Operation Manual 109

 Adjusting the methane reagent gas flow is a three-step process: setting the

flow control, pretuning on the reagent gas ions, and adjusting the flow for

stable reagent ion ratios, for methane, m/z 28/27.

 Your data system will prompt you through the flow adjustment procedure.

Procedure

1 Select Gas A. Follow the instructions and prompts from the Tune Wizard.

2 Set the flow to 20% for PCI/NCI MSDs.

3 Check the vacuum gauge controller to verify correct pressure. See page 124.

4 Select Methane Pretune from the Setup menu.

The methane pretune tunes the instrument for optimum monitoring of the

ratio of methane reagent ions m/z 28/27.

5 Examine the displayed profile scan of the reagent ions (Figure 29).

• Make sure there is no visible peak at m/z 32. A peak there indicates an

air leak. If such a peak is present, find and repair the leak before

proceeding. Operating in the CI mode with an air leak will rapidly 

contaminate the ion source.

• Make sure that the peak at m/z 19 (protonated water) is less than 50% of 

the peak at m/z 17.

6 Perform the Methane Flow Adjust.

CAUTIONAfter the system has been switched from EI to CI mode, or vented for any other reason,

the MSD must be baked out for at least 2 hours before tuning.

4 Operating in Chemical Ionization (CI) Mode

CAUTIONContinuing with CI autotune if the MSD has an air leak or large amounts of water will

result in severe ion source contamination. If this happens, you will need to vent the 

MSD and clean the ion source .

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 110/162

110 5975 Series MSD Operation Manual

Methane pretune after more than a day of baking out

Note the low abundance of m/z 19 and absence of any visible peak at m/z 32.

 Your MSD will probably show more water at first, but the abundance of m/z 19

should still be less than 50% of m/z 17.

Figure 29 Reagent ion scans

Operating in Chemical Ionization (CI) Mode 4

To use other reagent gases

This section describes the use of isobutane or ammonia as the reagent gas. Youshould be familiar with operating the CI-equipped 5975 Series MSD with

methane reagent gas before attempting to use other reagent gases.

Do not use nitrous oxide as a reagent gas It radically shortens the life span of the

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 111/162

5975 Series MSD Operation Manual 111

Changing the reagent gas from methane to either isobutane or ammonia

changes the chemistry of the ionization process and yields different ions. The

principal chemical ionization reactions encountered are described in general

in Appendix A, “Chemical Ionization Theory . If you are not experienced with

chemical ionization, we suggest reviewing that material before you proceed.

CAUTIONDo not use nitrous oxide as a reagent gas. It radically shortens the life span of the

filament.

CAUTION Not all setup operations can be performed in all modes with all reagent gases. SeeTable 18 for details.

4 Operating in Chemical Ionization (CI) Mode

Table 18 Reagent gases

Reagent gas/mode Reagent ion masses PFDTDCalibrant ions

Flow adj ions: RatioEI/PCI/NCI MSD

Performance turbo

pump

Recommended flow:

20% PCI

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 112/162

112 5975 Series MSD Operation Manual

Isobutane CIIsobutane (C4H10) is commonly used for chemical ionization when less

fragmentation is desired in the chemical ionization spectrum. This is because

the proton affinity of isobutane is higher than that of methane; hence less

energy is transferred in the ionization reaction.

 Addition and proton transfer are the ionization mechanisms most often

associated with isobutane. The sample itself influences which mechanism

dominates.

40% NCI

Methane/PCI 17, 29, 41*

* There are no PFDTD ions formed with any reagent gas but methane. Tune with methane and usethe same parameters for the other gas.

41, 267, 599 28/27: 1.5 – 5.0

Methane/NCI 17, 35, 235†

† There are no negative reagent gas ions formed. To pretune in negative mode, use backgroundions: 17 (OH-), 35 (Cl-), and 235 (ReO3-). These ions can not be used for reagent gas flow adjust-ment. Set flow to 40% for NCI and adjust as necessary to get acceptable results for your appli-cation.

185, 351, 449 N/A

Isobutane/PCI 39, 43, 57 N/A 57/43: 5.0 – 30.0

Isobutane/NCI 17, 35, 235 185, 351, 449 N/A

Ammonia/PCI 18, 35, 52 N/A 35/18: 0.1 – 1.0

Ammonia/NCI 17, 35, 235 185, 351, 517 N/A

Operating in Chemical Ionization (CI) Mode 4

Ammonia CI

 Ammonia (NH3) is commonly used for chemical ionization when less

fragmentation is desired in the chemical ionization spectrum. This is becausethe proton affinity of ammonia is higher than that of methane; hence less

energy is transferred in the ionization reaction.

Because many compounds of interest have insufficient proton affinities,

ammonia chemical-ionization spectra often result from the addition of NH4+ 

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 113/162

5975 Series MSD Operation Manual 113

a o a c e ca o at o spect a o te esu t o t e add t o o N 4

and then, in some cases, from the subsequent loss of water. Ammonia reagent 

ion spectra have principal ions at m/z 18, 35, and 52, corresponding to NH4+,

NH4(NH3)+, and NH4(NH3)2+.

To adjust your MSD for isobutane or ammonia chemical ionization, use the

following procedure:

Procedure

1 From the Tune and Vacuum Control view, perform a standard Positive CI

autotune with methane and PFDTD.

2 Under the Setup menu, click CI Tune Wizard and when prompted select 

Isobutane or Ammonia. This will change the menus to use the selected gas and

select appropriate default tune parameters.

3 Select Gas B. Follow the instructions and prompts from the Tune Wizard

and set the gas flow to 20%.

If you use an existing tune file, be sure to save it with a new name if you

don’t want to overwrite the existing values. Accept the default temperatureand other settings.

4 Click Isobutane (or Ammonia) Flow Adjust on the Setup menu.

There is no CI autotune for isobutane or ammonia in PCI.

If you wish to run NCI with isobutane or ammonia, load NCICH4.U or an

existing NCI tune file for the specific gas.

NOTEBe sure to read the following application note: Implementation of Ammonia Reagent 

Gas for Chemical Ionization on the Agilent 5975 Series MSDs (5989-5170EN).

4 Operating in Chemical Ionization (CI) Mode

CAUTIONUse of ammonia affects the maintenance requirements of the MSD. See “CI

Maintenance” for more information.

CAUTIONThe pressure of the ammonia supply must be less than 5 psig. Higher pressures can

result in ammonia condensing from a gas to a liquid.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 114/162

114 5975 Series MSD Operation Manual

 Ammonia tends to break down vacuum pump fluids and seals. Ammonia CI

makes more frequent vacuum system maintenance necessary. (See the 5975

Series MSD Troubleshooting and Maintenance Manual.)

Frequently, a mixture of 5% ammonia and 95% helium or 5% ammonia and 95%

methane is used as a CI reagent gas. This is enough ammonia to achieve good

chemical ionization while minimizing its negative effects.

Carbon dioxide CI

Carbon dioxide is often used as a reagent gas for CI. It has obvious advantages

of availability and safety.

Always keep the ammonia tank in an upright position, below the level of the flow

module. Coil the ammonia supply tubing into several vertical loops by wrapping the

tubing around a can or bottle. This will help keep any liquid ammonia out of the flowmodule.

CAUTIONWhen running ammonia for 5 or more hours a day, the foreline pump must be ballasted

(flushed with air) for at least 1 hour a day to minimize damage to pump seals. Always

purge the MSD with methane after flowing ammonia.

Operating in Chemical Ionization (CI) Mode 4

To switch from the CI source to the EI source

Procedure

1 From the Tune and Vacuum Control view, vent the MSD. See page 82. The

software will prompt you for the appropriate actions.

2 Open the analyzer.

3 Remo e the CI interface tip seal See page 143

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 115/162

5975 Series MSD Operation Manual 115

3 Remove the CI interface tip seal. See page 143.

4 Remove the CI ion source. See page 142.

5 Install the EI ion source. See page 136.

6 Place the CI ion source and interface tip seal in the ion source storage box.

7 Pump down the MSD. See page 91.

8 Load your EI tune file.

CAUTIONAlways wear clean gloves while touching the analyzer or any other parts that go inside

the analyzer chamber.

CAUTIONElectrostatic discharges to analyzer components are conducted to the side board

where they can damage sensitive components. Wear a grounded antistatic wrist strap

and take other antistatic precautions before you open the analyzer chamber.

See page 131.

4 Operating in Chemical Ionization (CI) Mode

CI Autotune

 After the reagent gas flow is adjusted, the lenses and electronics of the MSDshould be tuned (Table 19). Perfluoro-5,8-dimethyl-3,6,9-trioxidodecane

(PFDTD) is used as the calibrant. Instead of flooding the entire vacuum

chamber, the PFDTD is introduced directly into the ionization chamber

through the GC/MSD interface by means of the gas flow control module.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 116/162

116 5975 Series MSD Operation Manual

There is a PCI autotune for methane only, as there are no PFDTD ions

produced by other gases in positive mode. PFDTD ions are visible in NCI for

any reagent gas. Always tune for methane PCI first regardless of which mode

or reagent gas you wish to use for your analysis.

There are no tune performance criteria. If CI autotune completes, it passes.

EMVolts (electron multiplier voltage) at or above 2600 V, however, indicates a

problem. If your method requires EMVolts set at +400, you may not have

adequate sensitivity in your data acquisition.

CAUTIONAfter the source is changed from EI to CI or vented for any other reason, the MSD must

be purged and baked out for at least 2 hours before tuning. Longer bakeout isrecommended before running samples requiring optimal sensitivity.

CAUTIONAlways verify MSD performance in EI before switching to CI operation. See page 76. 

Always set up the CI MSD in PCI first, even if you are going to run NCI.

Operating in Chemical Ionization (CI) Mode 4

Table 19 Reagent gas settings

Reagent gas Methane Isobutane Ammonia EI

Ion polarity Positive Negative Positive Negative Positive Negative N/A

Emission 150 A 50 A 150 A 50 A 150 A 50 A 35 A

Electron

energy

150 eV 150 eV 150 eV 150 eV 150 eV 150 eV 70 eV

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 117/162

5975 Series MSD Operation Manual 117

energy

Filament 1 1 1 1 1 1 1 or 2

Repeller 3 V 3 V 3 V 3 V 3 V 3 V 30 V

Ion focus 130 V 130 V 130 V 130 V 130 V 130 V 90 V

Entrance

lens offset

20 V 20 V 20 V 20 V 20 V 20 V 25 V

EM volts 1200 1400 1200 1400 1200 1400 1300

Shutoff valve Open Open Open Open Open Open Closed

Gas select A A B B B B None

Suggested

flow

20% 40% 20% 40% 20% 40% N/A

Source temp 250 °C 150 °C 250 °C 150 °C 250 °C 150 °C 230 °C

Quad temp 150 °C 150 °C 150 °C 150 °C 150 °C 150 °C 150 °C

Interfacetemp

280 °C 280 °C 280 °C 280 °C 280 °C 280 °C 280 °C

Autotune Yes Yes No Yes No Yes Yes

N/A Not available

4 Operating in Chemical Ionization (CI) Mode

To perform a PCI autotune (methane only)

Procedure

CAUTIONAlways verify MSD performance in EI before switching to CI operation. See page 76. 

Always set up the CI MSD in PCI first, even if you are going to run NCI.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 118/162

118 5975 Series MSD Operation Manual

Procedure

1  Verify that the MSD performs correctly in EI mode first. See page 76.

2 Load the PCICH4.U tune file (or an existing tune file for the reagent gas you

are using).

If you use an existing tune file, be sure to save it with a new name if you

don’t want to overwrite the existing values.

3  Accept the default settings.

4 Perform methane setup. See page 109.

5 Under the Tune menu, click CI Autotune.

There are no tune performance criteria. If autotune completes, it passes(Figure 30). If the tune sets the electron multiplier voltage (EMVolts) at or

above 2600 V, however, you may not be able to acquire data successfully if your

method sets EMVolts to “+400” or higher.

The autotune report contains information about air and water in the system.

The 19/29 ratio shows the abundance of water.

The 32/29 ratio shows the abundance of oxygen.

CAUTIONAvoid tuning more often than is absolutely necessary; this will minimize PFDTD

background noise and help prevent ion source contamination.

Operating in Chemical Ionization (CI) Mode 4

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 119/162

5975 Series MSD Operation Manual 119

Figure 30 PCI autotune

4 Operating in Chemical Ionization (CI) Mode

To perform an NCI autotune (methane reagent gas)

P d

CAUTIONAlways verify MSD performance in EI before switching to CI operation. See page 76. 

Always set up the CI MSD in PCI with methane as the reagent gas first, even if you are

going to be using a different reagent gas or going to run NCI.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 120/162

120 5975 Series MSD Operation Manual

Procedure

1 From the Tune and Vacuum Control view, load NCICH4.U (or an existing tunefile for the reagent gas you are using).

2 From the Setup menu select the CI Tune Wizard and follow the system

prompts.

 Accept the default temperature and other settings.

If you use an existing tune file, be sure to save it with a new name if you

don’t want to overwrite the existing values.

3 Under the Tune menu, click CI Autotune.

There are no tune performance criteria. If autotune completes, it passes(Figure 31). If the tune sets the electron multiplier voltage (EMVolts) at or

above 2600 V, however, you may not be able to acquire data successfully if your

method sets EMVolts to “+400” or higher.

CAUTIONAvoid tuning unless absolutely necessary; this will minimize PFDTD background noise

and help prevent ion source contamination.

Operating in Chemical Ionization (CI) Mode 4

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 121/162

5975 Series MSD Operation Manual 121

Figure 31 NCI autotune

4 Operating in Chemical Ionization (CI) Mode

To verify PCI performance

Materials needed

• Benzophenone, 100 pg/L (8500-5440)

CAUTIONAlways verify MSD performance in EI before switching to CI operation. See page 76. 

Always set up the CI MSD in PCI first, even if you are going to run NCI.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 122/162

122 5975 Series MSD Operation Manual

Procedure

1  Verify that the MSD performs correctly in E1 mode.

2  Verify that the PCICH4.U tune file is loaded.

3 Select Gas A and set flow to 20%.

4 In Tune and Vacuum Control view, perform CI setup. See page 116.

5 Run CI Autotune. See page 116.

6 Run the PCI sensitivity method BENZ_PCI.M using 1 µL of 100 pg/µL

benzophenone.

7  Verify that the system conforms to the published sensitivity specification.

Please see the Agilent Web site at www.agilent.com/chem for specifications.

Operating in Chemical Ionization (CI) Mode 4

To verify NCI performance

This procedure is for EI/PCI/NCI MSDs only.

Materials needed

• Octafluoronaphthalene (OFN), 100 fg/µL (5188-5347)

Always verify MSD performance in EI before switching to CI operation See page 76

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 123/162

5975 Series MSD Operation Manual 123

Procedure

1  Verify that the MSD performs correctly in EI mode.

2 Load the NCICH4.U tune file, and accept the temperature setpoints.

3 Select Gas A and set flow to 40%.

4 In Tune and Vacuum Control view, run CI Autotune. See page 120.

Note that there are no criteria for a “passing” Autotune in CI. If the

 Autotune completes, it passes.

5 Run the NCI sensitivity method: OFN_NCI.M using 2 µL of 100 fg/µL OFN.

6  Verify that the system conforms to the published sensitivity specification.

Please see the Agilent Web site at www.agilent.com/chem for specifications.

CAUTIONAlways verify MSD performance in EI before switching to CI operation. See page 76. 

Always set up the CI MSD in PCI first, even if you are going to run NCI.

4 Operating in Chemical Ionization (CI) Mode

To monitor high vacuum pressure

Procedure

WARNINGIf you are using hydrogen as a carrier gas, do not turn on the Micro-Ion vacuum

gauge if there is any possibility that hydrogen has accumulated in the manifold.

Read “Hydrogen Safety” before operating the MSD with hydrogen carrier gas.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 124/162

124 5975 Series MSD Operation Manual

Procedure

1 Start up and pump down the MSD. See page 103.2 In the Tune and Vacuum Control view select Turn Vacuum Gauge on/off from

the Vacuum menu.

3 In the Instrument Control view you can set up an MS Monitor for reading.

The vacuum can also be read on the LCP or from the Manual Tune screen.

The gauge controller will not turn on if the pressure in the MSD is above

approximately 8 × 10-3 Torr. The gauge controller is calibrated for nitrogen, but 

all pressures listed in this manual are for helium.

The largest influence on operating pressure is the carrier gas (column) flow.

Table 20 lists typical pressures for various helium carrier gas flows. These

pressures are approximate and will vary from instrument to instrument.

Operating in Chemical Ionization (CI) Mode 4

Typical pressure readings

Use the G3397A Micro-Ion vacuum gauge. Note that the mass flow controller is

calibrated for methane and the vacuum gauge is calibrated for nitrogen, sothese measurements are not accurate, but are intended as a guide to typical

observed readings (Table 20). They were taken with the following set of 

conditions. Note that these are typical PCI temperatures:

Source temperature 250 °CQuad temperature 150 °CI f 280 °C

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 125/162

5975 Series MSD Operation Manual 125

Familiarize yourself with the measurements on your system under operating

conditions and watch for changes that may indicate a vacuum or gas f low

problem. Measurements will vary by as much as 30% from one MSD and gauge

controller to the next.

Interface temperature 280 °CHelium carrier gas flow 1 mL/min

Table 20 Flow and pressure readings

Pressure (Torr)

Methane Ammonia

MFC(%)

EI/PCI/NCI MSD(Performance turbo pump)

EI/PCI/NCI MSD(Performance turbo pump)

10 5.5 ×10 –5 5.0 ×10 –5

15 8.0 ×10 –5 7.0 ×10 –5

20 1.0 ×10 –4 8.5 ×10 –5

25 1.2 ×10 –4 1.0 ×10 –4

30 1.5 ×1 –4 1.2 ×10 –4

35 2.0 ×10 –4 1.5 ×10 –4

40 2.5 ×10 –4 2.0 ×10 –4

4 Operating in Chemical Ionization (CI) Mode

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 126/162

126 5975 Series MSD Operation Manual

Agilent 5975 Series MSD

Operation Manual

5

General Maintenance

Before Starting 128

Maintaining the Vacuum System 133

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 127/162

127Agilent Technologies

5 General Maintenance

Before Starting

 You can perform much of the maintenance required by your MSD. For yoursafety, read all of the information in this introduction before performing any 

maintenance tasks.

Scheduled maintenance

Common maintenance tasks are listed in Table 21. Performing these tasks

when scheduled can reduce operating problems, prolong system life, and

d ll ti t

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 128/162

128 5975 Series MSD Operation Manual

reduce overall operating costs.

Keep a record of system performance (tune reports) and maintenance

operations performed. This makes it easier to identify variations from normal

operation and to take corrective action.

Table 21 Maintenance schedule

Task Every week Every 6 months Every year As needed

Tune the MSD X

Check the foreline pump oil level X

Check the calibration vial(s) X

Replace the foreline pump oil* X

Replace the diffusion pump fluid X

Check the dry foreline pump X

Clean the ion source X

Check the carrier gas trap(s) on the GC and

MSD

X

Replace the worn out parts X

Lubricate sideplate or vent valve O-rings† X

Replace CI Reagent gas supply X

Replace GC gas supplies X

* Every 3 months for CI MSDs using ammonia reagent gas.

† Vacuum seals other than the side plate O-ring and vent valve O-ring do not need to be lubricated. Lubricating other sealscan interfere with their correct function.

General Maintenance 5

Tools, spare parts, and supplies

Some of the required tools, spare parts, and supplies are included in the GC

shipping kit, MSD shipping kit, or MSD tool kit. You must supply others yourself. Each maintenance procedure includes a list of the materials required

for that procedure.

High voltage precautions

Whenever the MSD is plugged in, even if the power switch is off, potentially 

dangerous voltage (120 VAC or 200/240 VAC) exists on:

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 129/162

5975 Series MSD Operation Manual 129

dangerous voltage (120 VAC or 200/240 VAC) exists on:

• The wiring and fuses between where the power cord enters the instrument 

and the power switch

When the power switch is on, potentially dangerous voltages exist on:

• Electronic circuit boards

• Toroidal transformer

• Wires and cables between these boards

• Wires and cables between these boards and the connectors on the back 

panel of the MSD

• Some connectors on the back panel (for example, the foreline power

receptacle)

Normally, all of these parts are shielded by safety covers. As long as the safety 

covers are in place, it should be difficult to accidentally make contact with

dangerous voltages.

Some procedures in this chapter require access to the inside of the MSD while

the power switch is on. Do not remove any of the electronics safety covers inany of these procedures. To reduce the risk of electric shock, follow the

procedures carefully.

WARNINGPerform no maintenance with the MSD turned on or plugged into its power source

unless you are instructed to by one of the procedures in this chapter.

5 General Maintenance

Dangerous temperatures

Many parts in the MSD operate at, or reach, temperatures high enough to

cause serious burns. These parts include, but are not limited to:• GC/MSD interface

•  Analyzer parts

•  Vacuum pumps

WARNINGNever touch these parts while your MSD is on. After the MSD is turned off, give

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 130/162

130 5975 Series MSD Operation Manual

The GC inlets and GC oven also operate at very high temperatures. Use the

same caution around these parts. See the documentation supplied with your

GC for more information.

Chemical residue

Only a small portion of your sample is ionized by the ion source. The majority 

of any sample passes through the ion source without being ionized. It is

pumped away by the vacuum system. As a result, the exhaust from the forelinepump will contain traces of the carrier gas and your samples. Exhaust from

the standard foreline pump also contains tiny droplets of foreline pump oil.

 An oil trap is supplied with the standard foreline pump. This trap stops only 

pump oil droplets. It does not trap any other chemicals. If you are using toxic

solvents or analyzing toxic chemicals, do not use this oil trap. For all foreline

WARNING

these parts enough time to cool before handling them.

WARNINGThe GC/MSD interface heater is powered by a thermal zone on the GC. The interface

heater can be on, and at a dangerously high temperature, even though the MSD is

off. The GC/MSD interface is well insulated. Even after it is turned off, it cools very

slowly.

WARNINGThe foreline pump can cause burns if touched when operating. It has a safety shield

to prevent the user from touching it.

General Maintenance 5

pumps, install a hose to take the exhaust from the foreline pump outdoors or

into a fume hood vented to the outdoors. For the standard foreline pump, this

requires removing the oil trap. Be sure to comply with your local air quality 

regulations.

WARNINGThe oil trap supplied with the standard foreline pump stops only foreline pump oil. It

does not trap or filter out toxic chemicals. If you are using toxic solvents or

analyzing toxic chemicals, remove the oil trap. Do not use the trap if you have a CI

MSD. Install a hose to take the foreline pump exhaust outside or to a fume hood.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 131/162

5975 Series MSD Operation Manual 131

The fluids in the diffusion pump and standard foreline pump also collect 

traces of the samples being analyzed. All used pump fluid should be

considered hazardous and handled accordingly. Dispose of used fluid

correctly, as specified by your local regulations.

Electrostatic discharge

 All of the printed circuit boards in the MSD contain components that can be

damaged by electrostatic discharge (ESD). Do not handle or touch these

boards unless absolutely necessary. In addition, wires, contacts, and cables

can conduct ESD to the electronics boards to which they are connected. This is

especially true of the mass filter (quadrupole) contact wires which can carry 

ESD to sensitive components on the side board. ESD damage may not cause

immediate failure but it will gradually degrade the performance and stability 

of your MSD.

When you work on or near printed circuit boards or when you work on

components with wires, contacts, or cables connected to printed circuit 

boards, always use a grounded antistatic wrist strap and take other antistaticprecautions. The wrist strap should be connected to a known good earth

ground. If that is not possible, it should be connected to a conductive (metal)

part of the assembly being worked on, but not to electronic components,

exposed wires or traces, or pins on connectors.

WARNINGWhen replacing pump fluid, use appropriate chemical-resistant gloves and safety

glasses. Avoid all contact with the fluid.

5 General Maintenance

Take extra precautions, such as a grounded antistatic mat, if you must work 

on components or assemblies that have been removed from the MSD. This

includes the analyzer.

CAUTIONTo be effective, an antistatic wrist strap must fit snugly (not tight). A loose strap

provides little or no protection.

Antistatic precautions are not 100% effective. Handle electronic circuit boards as little

as possible and then only by the edges. Never touch components, exposed traces, or

pins on connectors and cables.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 132/162

132 5975 Series MSD Operation Manual

General Maintenance 5

Maintaining the Vacuum System

Periodic maintenance As listed earlier in Table 21, some maintenance tasks for the vacuum system

must be performed periodically. These include:

• Checking the foreline pump fluid (every week)

• Checking the calibration vial(s) (every 6 months)

• Ballasting the foreline pump (daily in MSDs using ammonia reagent gas)

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 133/162

5975 Series MSD Operation Manual 133

• Replacing the foreline pump oil (every 6 months; every 3 months for CIMSDs using ammonia reagent gas)

• Tightening the foreline pump oil box screws (first oil change after

installation)

• Replacing the diffusion pump fluid (once a year)

• Replacing the dry foreline pump (typically every 3 years)

Failure to perform these tasks as scheduled can result in decreasedinstrument performance. It can also result in damage to your instrument.

Other procedures

Tasks such as replacing a foreline vacuum gauge or Micro-Ion vacuum gauge

should be performed only when needed. See the 5975 Series MSD

Troubleshooting and Maintenance manual and see the online help in the MSD

ChemStation software for symptoms that indicate this type of maintenance isrequired.

More information is available

If you need more information about the locations or functions of vacuum

system components, see the 5975 Series MSD Troubleshooting and

 Maintenance manual.

Most of the procedures in this chapter are illustrated with video clips on the

 Agilent GC/GCMSD Hardware User Information & Instrument Utilities and

5975 Series MSD User Information disks.

5 General Maintenance

To remove the EI ion source

Materials needed

• Gloves, clean, lint-free

• Large (8650-0030)

• Small (8650-0029)

• Pliers, long-nose (8710-1094)

Procedure

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 134/162

134 5975 Series MSD Operation Manual

1  Vent the MSD. See page 82.

2 Open the analyzer chamber. See page 84.

Make sure you use an antistatic wrist strap and take other antistatic

precautions before touching analyzer components.

3 Disconnect the seven wires from the ion source. Do not bend the wires any 

more than necessary (Figure 32 and Table 22).

Table 22 Ion source wires

Wire color Connects to Number of leads

Blue Entrance lens 1

Orange Ion focus 1

White Filament 1 (top

filament)

2

Red Repeller 1

Black Filament 2 (bottom

filament)

2

 

CAUTIONPull on the connectors, not on the wires.

General Maintenance 5

4 Trace the wires for the ion source heater and temperature sensor to the

feedthrough board. Disconnect them there.

5 Remove the thumbscrews that hold the ion source in place.

6 Pull the ion source out of the source radiator.

WARNINGThe analyzer operates at high temperatures. Do not touch any part until you are sure

it is cool.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 135/162

5975 Series MSD Operation Manual 135

Figure 32 Removing the ion source

Source feedthrough board

Ion source

Thumbscrews

Source heater and temperature sensor wires

Source radiator

5 General Maintenance

To reinstall the EI ion source

Materials needed

• Gloves, clean, lint-free

• Large (8650-0030)

• Small (8650-0029)

• Pliers, long-nose (8710-1094)

Procedure

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 136/162

136 5975 Series MSD Operation Manual

1 Slide the ion source into the source radiator (Figure 33).

2 Install and hand tighten the source thumbscrews. Do not overtighten the

thumbscrews.

3 Connect the ion source wires as shown in “To close the analyzer chamber” .

Close the analyzer chamber.

 

General Maintenance 5

4 Pump down the MSD. See page 91.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 137/162

5975 Series MSD Operation Manual 137

Figure 33 Installing the EI ion source

Ion source

Thumbscrews

Source radiator

5 General Maintenance

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 138/162

138 5975 Series MSD Operation Manual

Agilent 5975 Series MSD

Operation Manual

6

CI Maintenance

General Information 140

Ion source cleaning 140

Ammonia 140

To Set Up Your MSD for CI Operation 141

G id li 141

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 139/162

139Agilent Technologies

Guidelines 141To install the CI ion source 142

To install the CI interface tip seal 143

This chapter describes maintenance procedures and requirements that are

unique to 5975 Series MSDs equipped with the Chemical Ionization hardware.

6 CI Maintenance

General Information

Ion source cleaningThe main effect of operating the MSD in CI mode is the need for more frequent 

ion source cleaning. In CI operation, the ion source chamber is subject to more

rapid contamination than in EI operation because of the higher source

pressures required for CI.

WARNING

Always perform any maintenance procedures using hazardous solvents under a

fume hood Be sure to operate the MSD in a well ventilated room

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 140/162

140 5975 Series MSD Operation Manual

Ammonia

 Ammonia, used as a reagent gas, increases the need for foreline pump

maintenance. Ammonia causes foreline pump oil to break down more quickly.

Therefore, the oil in the standard foreline vacuum pump must be checked andreplaced more frequently.

 Always purge the MSD with methane after using ammonia.

Be sure to install the ammonia so the tank is in an upright position. This will

help prevent liquid ammonia from getting into the flow module.

WARNING fume hood. Be sure to operate the MSD in a well-ventilated room.

CI Maintenance 6

To Set Up Your MSD for CI Operation

Setting up your MSD for operation in CI mode requires special care to avoid

contamination and air leaks.

Guidelines

• Before venting in EI mode, verify that the GC/MSD system is performing

correctly. See “To verify system performance” .

• Make sure the reagent gas inlet line(s) are equipped with gas purifiers (not 

applicable for ammonia)

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 141/162

5975 Series MSD Operation Manual 141

applicable for ammonia).

• Use extra-high purity reagent gases; 99.99% or better for methane and as

pure as is available for other reagent gases.

6 CI Maintenance

To install the CI ion source

Procedure

1  Vent the MSD and open the analyzer. See page 84.

2 Remove the EI ion source See page 134

CAUTION Electrostatic discharges to analyzer components are conducted to the side boardwhere they can damage sensitive components. Wear a grounded antistatic wrist strap

and take other antistatic precautions before  you open the analyzer chamber.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 142/162

142 5975 Series MSD Operation Manual

2 Remove the EI ion source. See page 134.3 Remove the CI ion source from its storage box and insert the ion source into

the radiator.

4 Reinstall the thumbscrews (Figure 34).

5 Connect the wiring as described in “To close the analyzer chamber” .

 

Figure 34 Installing the CI ion source

Ion source

Thumbscrews

Source radiator

CI Maintenance 6

To install the CI interface tip seal

Materials needed

• Interface tip seal (G1099-60412)

The interface tip seal must be in place for CI operation. It is necessary to

achieve adequate ion source pressure for CI.

CAUTIONElectrostatic discharges to analyzer components are conducted to the side board

where they can damage sensitive components. Wear a grounded antistatic wrist strap

and take other antistatic precautions before you open the analyzer chamber.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 143/162

5975 Series MSD Operation Manual 143

Procedure

1 Remove the seal from the ion source storage box.

2  Verify that the CI ion source is installed.

3 Place the seal over the end of the interface. To remove the seal, reverse theabove steps.

4 Gently check the alignment of the analyzer and the interface.

When the analyzer is aligned correctly, the analyzer can be closed all the

way with no resistance except the spring tension from the interface tip seal.

5  You can align the analyzer and interface by wiggling the side plate on its

hinge. If the analyzer still will not close, contact your Agilent Technologies

service representative.

a d ta e ot e a t stat c p ecaut o s you ope t e a a y e c a be

 

CAUTION

Forcing the analyzer closed if these parts are misaligned will damage the seal or the

interface or the ion source, or will keep the sideplate from sealing.

6 CI Maintenance

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 144/162

144 5975 Series MSD Operation Manual

Agilent 5975 Series MSD

Operation Manual

A

Chemical Ionization Theory

Chemical Ionization Overview 146

Positive CI Theory 148

Negative CI Theory 155

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 145/162

145Agilent Technologies

A Chemical Ionization Theory

Chemical Ionization Overview

Chemical ionization (CI) is a technique for creating ions used in mass

spectrometric analyses. There are significant differences between CI and

electron ionization (EI). This section describes the most common chemical

ionization mechanisms.

In EI, relatively high-energy electrons (70 eV) collide with molecules of the

sample to be analyzed. These collisions produce (primarily) positive ions.

Upon ionization, the molecules of a given substance fragment in fairly 

predictable patterns. EI is a direct process; energy is transferred by collision

from electrons to the sample molecules.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 146/162

146 5975 Series MSD Operation Manual

For CI, in addition to the sample and carrier gas, large amounts of reagent gas

are introduced into the ionization chamber. Since there is so much more

reagent gas than sample, most of the emitted electrons collide with reagent gas

molecules, forming reagent ions. These reagent-gas ions react with each other

in primary and secondary reaction processes that establish an equilibrium.

They also react in various ways with sample molecules to form sample ions. CI

ion formation involves much lower energy and is much more “gentle” thanelectron ionization. Since CI results in much less fragmentation, CI spectra

usually show high abundance of the molecular ion. For this reason, CI is often

used to determine the molecular weights of sample compounds.

Methane is the most common CI reagent gas. It yields certain characteristic

ionization patterns. Other reagent gases yield different patterns and may 

result in better sensitivity for some samples. Common alternative reagent 

gases are isobutane and ammonia. Carbon dioxide is often used in negative CI.

Less common reagent gases are carbon dioxide, hydrogen, Freon,

trimethylsilane, nitric oxide, and methylamine. Different ionization reactions

occur with each reagent gas.

Water contamination in reagent gases will decrease CI sensitivity 

dramatically. A large peak at m/z 19 (H30+) in positive CI is a diagnostic

symptom of water contamination. In high enough concentrations, especially 

when combined with calibrant, water contamination will result in a heavily 

WARNINGAmmonia is toxic and corrosive. Use of ammonia requires special maintenance and

safety precautions.

Chemical Ionization Theory A

contaminated ion source. Water contamination is most common immediately 

after new reagent gas tubing or reagent gas cylinders are connected. This

contamination will often decrease if the reagent gas is allowed to flow for a

few hours, purging the system.

References on chemical ionization

 A. G. Harrison, Chemical Ionization Mass Spectrometry, 2nd Edition, CRC

Press, INC. Boca Raton, FL (1992) ISBN 0-8493-4254-6.

W. B. Knighton, L. J. Sears, E. P. Grimsrud, “High Pressure Electron Capture

Mass Spectrometry”, Mass Spectrometry Reviews (1996), 14, 327-343.

E A Stemmler R A Hites Electron Capture Negative Ion Mass Spectra of

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 147/162

5975 Series MSD Operation Manual 147

E. A. Stemmler, R. A. Hites, Electron Capture Negative Ion Mass Spectra of 

 Environmental Contaminants and Related Compounds, VCH Publishers,

New York, NY (1988) ISBN 0-89573-708-6.

A Chemical Ionization Theory

Positive CI Theory

Positive CI (PCI) occurs with the same analyzer voltage polarities as EI. For

PCI, the reagent gas is ionized by collision with emitted electrons. The reagent 

gas ions react chemically with sample molecules (as proton donors) to form

sample ions. PCI ion formation is more “gentle” than electron ionization,

producing less fragmentation. This reaction usually yields high abundance of 

the molecular ion and is therefore often used for determining molecular

weights of samples.

The most common reagent gas is methane. Methane PCI produces ions with

almost any sample molecule. Other reagent gases, such as isobutane orammonia are more selective and cause even less fragmentation Because of

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 148/162

148 5975 Series MSD Operation Manual

ammonia, are more selective and cause even less fragmentation. Because of 

the high background from the reagent gas ions, PCI is not especially sensitive

and detection limits are generally high.

There are four fundamental ionization processes that take place during

positive chemical ionization at ion source pressures in the 0.8 to 2.0 Torr

range. These are:

• Proton transfer

• Hydride abstraction

•  Addition

• Charge exchange

Depending on the reagent gas used, one or more of these four processes can be

used to explain the ionization products observed in the resulting mass spectra.

EI, methane PCI, and ammonia PCI spectra of methyl stearate are shown in

Figure 35. The simple fragmentation pattern, large abundance of the [MH]+ 

ion, and the presence of the two adduct ions are characteristic of positive

chemical ionization using methane as a reagent gas.

The presence of air or water in the system, especially in the presence of 

PFDTD calibrant, quickly contaminates the ion source.

Chemical Ionization Theory A

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 149/162

5975 Series MSD Operation Manual 149

Figure 35 Methyl stearate (MW = 298): EI, methane PCI, and ammonia PCI

A Chemical Ionization Theory

Proton transfer

Proton transfer can be expressed as

BH+ + M MH+ + B

where the reagent gas B has undergone ionization resulting in protonation. If 

the proton affinity of the analyte (sample) M is greater than that of the reagent 

gas, then the protonated reagent gas will transfer its proton to the analyte

forming a positively charged analyte ion.

The most frequently used example is the proton transfer from CH5+ to the

molecular analyte, which results in the protonated molecular ion MH+.

The relative proton affinities of the reagent gas and the analyte govern the

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 150/162

150 5975 Series MSD Operation Manual

proton transfer reaction. If the analyte has a greater proton affinity than the

reagent gas, then proton transfer can take place. Methane (CH4) is the most 

common reagent gas because its proton affinity is very low.

Proton affinities can be defined according to the reaction:

B + H+ BH+

where the proton affinities are expressed in kcal/mole. Methane's proton

affinity is 127 kcal/mole. Tables 23 and 24 list the proton affinities of several

possible reagent gases and of several small organic compounds with various

functional groups.

The mass spectrum generated by a proton-transfer reaction depends on

several criteria. If the difference in proton affinities is large (as with methane),

substantial excess energy may be present in the protonated molecular ion.

This can result in subsequent fragmentation. For this reason, isobutane with a

proton affinity of 195 kcal/mole may be preferred to methane for some

analyses. Ammonia has a proton affinity of 207 kcal/mole, making it less likely 

to protonate most analytes. Proton-transfer chemical ionization is usually 

considered to be “soft” ionization, but the degree of the softness depends on

the proton affinities of both the analyte and the reagent gas, as well as on

other factors including ion source temperature.

Chemical Ionization Theory A

Table 23 Reagent gas proton affinities

Species Proton affinity

kcal/mole

Reactant ion

formed

H2 100 H3+ (m/z 3)

CH4 127 CH5+ (m/z 17)

C2H4 160 C2H5+ (m/z 29)

H2O 165 H3O+ (m/z 19)

H2S 170 H3S+ (m/z 35)

CH3OH 182 CH3OH2+ (m/z 33)

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 151/162

5975 Series MSD Operation Manual 151

CH3OH 182 CH3OH2 (m/z 33)

t-C4H10 195 t-C4H9+ (m/z 57)

NH3 207 NH4+ (m/z 18)

Table 24 Proton affinities of selected organic compounds for PCI

Molecule Proton affinity

(kcal/mole)

Molecule Proton affinity

(kcal/mole)

Acetaldehyde 185 Methyl amine 211

Acetic acid 188 Methyl chloride 165

Acetone 202 Methyl cyanide 186Benzene 178 Methyl sulfide 185

2-Butanol 197 Methyl cyclopropane l80

Cyclopropane 179 Nitroethane 185

Dimethyl ether 190 Nitromethane 180

Ethane 121 n-Propyl acetate 207

Ethyl formate 198 Propylene 179

Formic acid 175 Toluene 187

Hydrobromic acid 140 trans-2-Butene 180

Hydrochloric acid 141 Trifluoroacetic acid 167

A Chemical Ionization Theory

Isopropyl alcohol 190 Xylene 187

Methanol 182

Table 24 Proton affinities of selected organic compounds for PCI (continued)

Molecule Proton affinity

(kcal/mole)

Molecule Proton affinity

(kcal/mole)

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 152/162

152 5975 Series MSD Operation Manual

Chemical Ionization Theory A

Hydride abstraction

In the formation of reagent ions, various reactant ions can be formed that have

high hydride-ion (H–) affinities. If the hydride-ion affinity of a reactant ion is

higher than the hydride-ion affinity of the ion formed by the analyte's loss of 

H–, then the thermodynamics are favorable for this chemical ionization

process. Examples include the hydride abstraction of alkanes in methane

chemical ionization. In methane CI, both CH5+ and C2H5

+ are capable of 

hydride abstraction. These species have large hydride-ion affinities, which

results in the loss of H– for long-chain alkanes, according to the general

reaction

R+ + M [M–H]+ + RH

For methane R+ is CH5+ and C2H5

+ and M is a long-chain alkane In the case

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 153/162

5975 Series MSD Operation Manual 153

For methane, R is CH5 and C2H5 , and M is a long chain alkane. In the case

of CH5+, the reaction proceeds to form [M–H]+ + CH 4+ H2. The spectra

resulting from hydride abstraction will show an M–1 m/z peak resulting from

the loss of H–. This reaction is exothermic so fragmentation of the [M–H]+ ion

is often observed.

Often, both hydride-abstraction and proton-transfer ionization can be evident 

in the sample spectrum. One example is the methane CI spectrum of 

long-chain methyl esters, where both hydride abstraction from the

hydrocarbon chain and proton transfer to the ester function occur. In the

methane PCI spectrum of methyl stearate, for example, the MH+ peak at 

m/z 299 is created by proton transfer and the [M–1]+ peak at m/z 297 is

created by hydride abstraction.

AdditionFor many analytes, proton-transfer and hydride-abstraction chemical

ionization reactions are not thermodynamically favorable. In these cases,

reagent gas ions are often reactive enough to combine with the analyte

molecules by condensation or association (addition reactions). The resulting

ions are called adduct ions. Adduct ions are observed in methane chemical

ionization by the presence of [M+C2H5]+ and [M+C3H5]+ ions, which result in

M+29 and M+41 m/z mass peaks. Addition reactions are particularly important in ammonia CI. Because the

NH3 has a high proton affinity, few organic compounds will undergo proton

transfer with ammonia reagent gas. In ammonia CI, a series of ion-molecule

reactions takes place, resulting in the formation of NH4+, [NH4NH3]+, and

[NH4(NH3)2]+. In particular, the ammonium ion, NH4+, will give rise to an

A Chemical Ionization Theory

intense [M+NH4]+ ion observed at M+18 m/z , either through condensation or

association. If this resulting ion is unstable, subsequent fragmentation may be

observed. The neutral loss of H2O or NH3, observed as a subsequent loss of 18

or 17 m/z , respectively, is also common.

Charge exchange

Charge-exchange ionization can be described by the reaction:

X+· + M M+· + X

where X+ is the ionized reagent gas and M is the analyte of interest. Examples

of reagent gases used for charge exchange ionization include the noble gases(helium, neon, argon, krypton, xenon, and radon), nitrogen, carbon dioxide,

carbon monoxide hydrogen and other gases that do not react “chemically”

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 154/162

154 5975 Series MSD Operation Manual

carbon monoxide, hydrogen, and other gases that do not react chemically

with the analyte. Each of these reagent gases, once ionized, has a

recombination energy expressed as:

X+· + e–  X

or simply the recombination of the ionized reagent with an electron to form a

neutral species. If this energy is greater than the energy required to remove anelectron from the analyte, then the first reaction above is exothermic and

thermodynamically allowed.

Charge-exchange chemical ionization is not widely used for general analytical

applications. It can, however, be used in some cases when other chemical

ionization processes are not thermodynamically favored.

Chemical Ionization Theory A

Negative CI Theory

Negative chemical ionization (NCI) is performed with analyzer voltage

polarities reversed to select negative ions. There are several chemicalmechanisms for NCI. Not all mechanisms provide the dramatic increases in

sensitivity often associated with NCI. The four most common mechanisms

(reactions) are:

• Electron capture

• Dissociative electron capture

• Ion pair formation

• Ion-molecule reactions

I ll f th t th i l l ti th t

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 155/162

5975 Series MSD Operation Manual 155

In all of the cases except the ion-molecule reactions, the reagent gas serves a

function different from the function it serves in PCI. In NCI, the reagent gas is

often referred to as the buffer gas. When the reagent gas is bombarded with

high energy electrons from the filament, the following reaction occurs:

Reagent gas + e– (230eV)  Reagent ions + e– (thermal)

If the reagent gas is methane (Figure 36), the reaction is:

CH4 + e– (230eV)  CH4+ + 2e–

(thermal)

The thermal electrons have lower energy levels than the electrons from the

filament. It is these thermal electrons that react with the sample molecules.

There are no negative reagent gas ions formed. This prevents the kind of 

background that is seen in PCI mode and is the reason for the much lower

detection limits of NCI. The products of NCI can only be detected when the

MSD is operating in negative ion mode. This operating mode reverses the

polarity of all the analyzer voltages.

Carbon dioxide is often used as a buffer gas in NCI. It has obvious cost,

availability, and safety advantages over other gases.

A Chemical Ionization Theory

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 156/162

156 5975 Series MSD Operation Manual

Figure 36 Endosulfan I (MW = 404): EI and methane NCI

Chemical Ionization Theory A

Electron capture

Electron capture is the primary mechanism of interest in NCI. Electron

capture (often referred to as high-pressure electron capture mass

spectrometry or HPECMS) provides the high sensitivity for which NCI isknown. For some samples under ideal conditions, electron capture can provide

sensitivity as much as 10 to 1000 times higher than positive ionization.

Note that all the reactions associated with positive CI will also occur in NCI

mode, usually with contaminants. The positive ions formed do not leave the

ion source because of the reversed lens voltages, and their presence can

quench the electron capture reaction.

The electron capture reaction is described by:

MX + e–(th l) MX–·

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 157/162

5975 Series MSD Operation Manual 157

MX + e  (thermal)  MX

where MX is the sample molecule and the electron is a thermal (slow) electron

generated by the interaction between high energy electrons and the reagent 

gas.

In some cases, the MX–

·radical anion is not stable. In those cases the reverse

reaction can occur:

MX–·  MX + e–

The reverse reaction is sometimes called autodetachment. This reverse

reaction generally occurs very quickly. Thus, there is little time for the

unstable anion to be stabilized through collisions or other reactions.

Electron capture is most favorable for molecules that have hetero-atoms. For

example: nitrogen, oxygen, phosphorus, sulfur, silicon, and especially thehalogens: fluorine, chlorine, bromine, and iodine.

The presence of oxygen, water, or almost any other contaminant interferes

with the electron-attachment reaction. Contaminants cause the negative ion to

be formed by the slower ion-molecule reaction. This generally results in less

sensitivity. All potential contamination sources, especially oxygen (air) and

water sources, must be minimized.

A Chemical Ionization Theory

Dissociative electron capture

Dissociative electron capture is also known as dissociative resonance capture.

It is a process similar to electron capture. The difference is that during the

reaction, the sample molecule fragments or dissociates. The result is typically an anion and a neutral radical. Dissociative electron capture is illustrated by 

the reaction equation:

MX + e–(thermal)  M· + X–

This reaction does not yield the same sensitivity as electron capture, and the

mass spectra generated typically have lower abundance of the molecular ion.

 As with electron capture, the products of dissociative electron capture are not always stable. The reverse reaction sometimes occurs. This reverse reaction is

sometimes called an associative detachment reaction. The equation for the

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 158/162

158 5975 Series MSD Operation Manual

sometimes called an associative detachment reaction. The equation for the

reverse reaction is:

M· + X–  MX + e–

Ion pair formation

Ion pair formation is superficially similar to dissociative electron capture. The

ion pair formation reaction is represented by the equation:

MX + e–(thermal)  M+ + X¯ + e–

 As with dissociative electron capture, the sample molecule fragments. Unlike

dissociative electron capture however, the electron is not captured by the

fragments. Instead, the sample molecule fragments in such a way that the

electrons are distributed unevenly and positive and negative ions aregenerated.

Chemical Ionization Theory A

Ion-molecule reactions

Ion-molecule reactions occur when oxygen, water, and other contaminants are

present in the CI ion source. Ion-molecule reactions are two to four times

slower than electron-attachment reactions and do not provide the highsensitivity associated with electron capture reactions. Ion-molecule reactions

can be described by the general equation:

M + X–  MX–

where X– is most often a halogen or hydroxyl group that was created by 

ionization of contaminants by electrons from the filament. Ion-molecule

reactions compete with electron capture reactions. The more ion-molecule

reactions that occur, the fewer electron capture reactions occur.

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 159/162

5975 Series MSD Operation Manual 159

A Chemical Ionization Theory

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 160/162

160 5975 Series MSD Operation Manual

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 161/162

Agilent Technologies

7/14/2019 Agilent 5973 MS User Manual

http://slidepdf.com/reader/full/agilent-5973-ms-user-manual 162/162

© Agilent Technologies, Inc.

Printed in USA, June 2012