Agilent 1100 Series System Operating principles
description
Transcript of Agilent 1100 Series System Operating principles
04/22/23
Agilent 1100 Series System Operating principles
• Pumping principle• Sampler principles• TCC peltier principle• Detector principle
04/22/23
Agilent 1100 Series Pump Principles
04/22/23
Agilent 1100 Series Pump Principle
04/22/23
Metering Drive
Power Supply1100 Common Part
Fan
Leak Tray
LPM BoardCommon to Iso and Quat Pumps ("Type"? )
Leak Sensor1100 Common Part
Flow Pulse DamperCommon to 1100, HP1050and HP1090 Pumps
Pump Head
Agilent 1100 Series Pump Principle
04/22/23
Agilent 1100 Series Autosampler Principle
04/22/23
Agilent 1100 Series Autosampler Principle
Power SupplyCommon 1100 Part
Transport Unit
Sampling Unit
ASM Board
Vial Trays:100-2ml40-2ml15-ml
Critical Connections, snap down correctly!
04/22/23
Agilent 1100 Series Well Plate Sampler principle
04/22/23
Agilent 1100 Series Thermostatted Column Compartment principle
04/22/23
Agilent 1100 Series Variable Wavelength Detector Principle
Lamp
Optical mirrors
Grating
Detector cell
Slit
Beam splitter
04/22/23
Agilent 1100 Series Variable Wavelength Detector Principle
Lamp housing
Flow cell Leak Sensor1100 Common Part
Optical unit
VWD Board
Fan
Power Supply1100 Common Part
04/22/23
Recommended Flow Cells
Typical Column Length
Typical Peak Width
<= 5 cm 0.025 Micro Flow Cell
10 cm 0.05 Semimicro Flow Cell
20 cm 0.1 Standard Flow Cell
>= 40 cm 0.2
0.05-0.2 mL/min 0.2-0.4 mL/min 0.4-0.8 mL/min 1-2 mL/min 0.05-5 mL/min
1.0 mm 2.1 mm 3.0 mm 4.6 mm
Recommended Flow Cell
Typical Flow Rate
Internal Column Diameter
High-Pressure Flow Cell
For Pressure Above 100 bar
04/22/23
Agilent 1100 Series Diode Array Detector Principle
Tungsten lamp
Lens
Deuterium lamp
Lens system
Holmium filter
Detector cell
Optical slitGrating
Diode Array
04/22/23
Agilent 1100 Series Diode Array Detector Principle
Lamp housing
Flow cell
Leak Sensor1100 Common Part
DAD Board
Fan
Power Supply1100 Common Part
Optical unit
04/22/23
Recommended Flow cells
Typical Column Length
Typical Peak Width
<= 5 cm 0.025 500 nl Flow Cell
10 cm 0.05 Semimicro Flow Cell
20 cm 0.1 Standard Flow Cell
>= 40 cm 0.2
0.01-0.2 mL/min 0.2-0.4 mL/min 0.4-0.8 mL/min 1-2 mL/min 0.05-5 mL/min
0.3-1 mm 2.1 mm 3.0 mm 4.6 mm
Recommended Flow Cell
Typical Flow Rate
Internal Column Diameter
High-Pressure Flow Cell