AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An...

38
AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations September 10, 2014 1 Tim Rice BioColumn Technical Specialist

Transcript of AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An...

Page 1: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

AdvanceBio Peptide Mapping

An HPLC Column Technology for

Faster Protein Biocharacterizations

September 10, 20141

Tim RiceBioColumn Technical Specialist

Page 2: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

What Is Peptide Mapping?

• The chemical or enzymatic treatment of a protein to produce peptide fragments

• Separation and identification of these fragments in a reproducible manner

• For bio-therapeutic proteins and peptides peptide mapping is:

In-depth analysis that can identify minor and even isobaric differences in protein primary structure such as errors in the transcription of complementary DNA, point mutations., and PTMs

September 10, 2014Confidentiality Label

2

Page 3: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

Proteins and Monoclonal Antibodies – Complex Molecules, Complex Manufacturing

• Proteins have multiple layers of structure –primary, secondary, tertiary.., that all must be characterized

• Monoclonal antibodies/proteins have glycosylations/post-translational modifications that can vary

• Proteins have multiple layers of structure –primary, secondary, tertiary.., that all must be characterized

• Monoclonal antibodies/proteins have glycosylations/post-translational modifications that can vary

Complex Molecules

• Manufacturing requires living cells – which can vary, creating variable results

• Variation in process can change structures, post-translational modifications

• These changes need to be characterized

• Manufacturing requires living cells – which can vary, creating variable results

• Variation in process can change structures, post-translational modifications

• These changes need to be characterized

Complex Manufacturing

3

Page 4: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

Why peptide mapping?• Analysis of a single modification on a large protein may not be possible because the

change is too small to be seen in the large molecule

• Results in a collection of peptides of varying length & number depending on the protease used. The resulting peptides can be analyzed where each peak represents a particular peptide

treatment with a protease enzyme

Page 5: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

Peptide mapping is the single most important technique in analytical characterization…

Confirm primary structure by comparison of a product to a reference protein (detect point mutations, mis-translations & confirm genetic stability)

Identify location of disulfide bonds

Characterize & analyze degradation processes such as deamidation & oxidation

Isolate digest fragments for sequencing or further identification

Identify sites of glycosylation

Drug substance identity test

Drug substance purity test

5

Page 6: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

6

NEW: Peptide Mapping ‘How To’ Guide #5991-2348EN

Page 7: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

THE COLUMN

September 10, 20147

Page 8: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

This Talk Will Show That The AdvanceBio Peptide Mapping Column Allows You To:

1. Achieve high efficiency separation performance with rapid run times. Faster than traditional UV and LC/MS analysis times

2. Conserve the bio-characterization information content with accelerated run times

3. Achieve superior UV and LC/MS separation performance to competitive products

4. Achieve UHPLC-like speed and peptide mapping performance at HPLC operating conditions, in direct comparison to competitive uHPLC columns

September 10, 20148

Page 9: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

“Traditional” Peptide MapColumn: 4.6 X 250 MMC18 stationary phase5 micron particle size,Mobile Phase: A= 0.1%TFA in water, B=0.1%TFA in ACNGradient: 10% B to 90 B in 100 minutesRuntime: 115 minutesRe-equilibration time: 20 minutes

Page 10: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

2.7um Superficially porous particle technology delivers UHPLC type column performance but without high column back pressures

September 10, 201410

AdvanceBio Peptide Mapping Column Technology

Decrease the diffusion time for macromolecules and limit the diffusion path!

This results in more efficient mass transfer at increasing flows = sharper peaks during fast run times.

Actual

120A pore sizeC18 phase

Page 11: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

AdvanceBio Peptide Mapping Column

Primary Benefit – Reduce Peptide Mapping Time without Losing ResolutionWhat is It? - A superficially porous column with a 2.7um particle and C18 functionality which enables separation of hydrophilic through hydrophobic peptides to give superior resolution across the gradient range to efficiently resolve peptide fragments.Major Features High Resolution of the Peptide Map Fast Analysis Times Lower Pressure Drops than sub 2um Columns Quality Checked for Peptide Performance

September 10, 201411

Page 12: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

September 10, 201412

AdvanceBio Peptide Mapping Column Technology

min0 2.5 5 7.5 10 12.5 15 17.5 20

mAU

0

10

20

30

40

50

60

70

BSA tryptic digest on 2.1 x 150mm AdvanceBio Peptide Mapping Column

Hydrophilic peptide retention

Narrow Peaks w baseline resolution

Hydrophobic peptide retention

Reduced and fast analysis time

Critical and desired peptide mappingcomponents to achieve fast, selective andhighly efficient peptide separations acrossa wide dynamic range.

3-10% ACN

10-40% ACN

40-70% ACN

Page 13: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

September 10, 201413

Ensures High quality batch-to-batch reproducibilityConfidence in separation performance (peak shape, Rs, selectivity) Visualization: High performance Peptide Separation Column

Quality Assurance Testing with Agilent Peptide Mix

Peptide Mix Now Commercially Available

Page 14: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

Decreasing Run TimeInitial experiment was to decrease run time from traditional 120 min to 75 min

While faster runtime is the goal, we start here to make sure we do not compromise resolution and provide same level of coverage

This is followed up with gradient, temperature, and column length optimization

September 10, 201414

Page 15: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

Temperature

Initial temp of 30-50 C recommended

Optimal temp will be application dependent, effected by composition and digestion type

Change in temp can change selectivity and result in RT switching

Elevated temp produces narrower peak bands and lower pressure

September 10, 201415

Page 16: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

09-Nov-1216

0

100

200

300

400

500

600

700

800

900

0 10 20 30 40 50 60 70

Pres

sure

(bar

)

Temp, C

2.1 x 250mm Back Pressure Analysis 90*/10 water/ACN linear gradient

0.25mL/min

0.5mL/min.

blue = recommended operation region

Column CharacterizationLinear Flow Analysis (back pressure)

Column dim.: 2.1 x 250mmMobile phase: A-water (0.1%TFA), B- ACN (0.08%TFA)Gradient: 0-15min ,10-90%BInstrument: Agilent Infinity 1290

* Press reading obsv. (max)

Page 17: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

Gradient Optimization

Changes in gradient steepness can improve band spacing and change selectivity

Varied by either changing the %B change over the same run time

Or by keeping the % change in B the same, but changing the run time with flow rate change.

September 10, 201417

Page 18: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

min0 10 20 30 40 50 60

mAU

0

20

40

60

80

100

120

140

160

DAD1 A, Sig=215,4 Ref=off (CSD_LSCA_IGG\0000008.D)

75 Minute Run IgG Tryptic Digest0.2ml/min, 0-40% B

Page 19: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

September 10, 201419

Decreasing Analysis Times Without Sacrificing Separation Performance or Reaching Back Pressure Constraints

IgG Tryptic Digest Peptide Mapping

min0 10 20 30 40 50 60

mAU

0

20

40

60

80

100

120

140

160

min0 2 4 6 8 10 12

mAU

0

20

40

60

80

100

120

140

160

10-40%B, 0.2 mL/min

10-40%B, 0.6 ml/min

75 min run59 peaks

2.1 x 100mm AdvanceBio peptide Mapping Column14 min run57peaks

2.1 x 150mm AdvanceBio peptide Mapping Column

%B

-50

-50

%B

211 bar

433 bar

Keeping gradient change/column volume the same ensures selectivity remains constant

Page 20: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

DO WE LOSE ANY CRITICAL DATA WHEN WE SPEED THINGS UP?

September 10, 201420

Page 21: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

September 10, 201421

8x10

0

0.75

1.75

2.75

3.75

+ESI TIC Scan Frag=200.0V igg023.d

Counts vs. Acquisition Time (min)1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39

40 min. Run, 2.1 x 100mm

Native peptide

Deamidatedform 2

Deamidatedform 1

Critical Post Translational Modification (PTM) Conserved during Reduced Run Times

Native peptide

Deamidatedform 2

Deamidatedform 1

8x10

0

0.6

1.6

2.6

3.6

4.6

433 bar0.6mL/min10-40%B

+ESI TIC Scan Frag=200.0V igg011.d

Counts vs. Acquisition Time (min)0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5 13

140bar0.2mL/min10-40%B

Heavy Chain Peptide 357-366 and its two deamidated forms

conserved

Rapid LC/MS IgG Tryptic Digest Peptide Mapping

14 min. Run, 2.1 x 100mm

Page 22: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

September 10, 201422

Rapid LC/MS IgG Tryptic Digest Peptide Mapping

14 min. analysis, 2.1 x 100mm 99.63% sequence coverage

40 min. analysis, 2.1 x100mm 99.84% sequence coverage

Sequence Coverage Remains Unchanged When Gradients are Shortened to increase Analysis Times

140bar0.2mL/min3-35% B

433 bar0.6mL/min3-35% B

Page 23: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

How Does AdvanceBio Peptide Mapping compare to other options?

September 10, 201423

Page 24: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

09-Nov-1224

min0 2.5 5 7.5 10 12.5 15 17.5 20

mAU

0

10

20

30

40

50

60

70

min0 2.5 5 7.5 10 12.5 15 17.5 20

mAU

0

10

20

30

40

50

60

70

HPLC (UV)Results

104 peaks319 bar

76 peaks166 bar

Competitive Superficially Porous: Peptide Column

Higher Peak Capacity Achieved by AdvanceBio Peptide Column during 20 min. Rapid Run2.1 x 150mm AdvanceBio Peptide Mapping Column, BSA tryptic digest

2.1 x 150mm 3.6um Superficially Porous Peptide Column, BSA tryptic digest

Page 25: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

September 10, 201425

Competitor Superficially Porous Peptide Column Comparison

2.1 x 100mm AdvanceBio Peptide Mapping Column, IgG1 tryptic digest

2.1 x 100mm 1.7um Superficially Porous Peptide Column, IgG1 tryptic digest

Mapping Results

Tryptic peptides i.d: 76 peptides

Sequence coverage.:99.63%

Mobile phase A:water (0.1%FA)Mobile phase B:90% ACN (0.1%FA)

Inj.: 15uL Flow: 0.6 ml/minTemp.: 40 oC

Gradient (14min.)

Time(min.) %B0 310 3512 9014 90

LC/MS:Agilent 1290 coupled to 6530 QTOF

Higher Sequence Coverage Achieved by AdvanceBio Peptide Column during 14min. Rapid Run

Mapping Results

Tryptic peptides i.d:

65 peptides

Sequence coverage.:89.54%

14 min. run

14 min. run

433bar

680bar

Page 26: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

September 10, 201426

Achieving UHPLC-like Speed and Peptide Mapping Performance at HPLC Operating Conditions

Page 27: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

September 10, 201427

min0 2 4 6 8 10 12

mAU

0

20

40

60

80

100

120

140

160

min0 2 4 6 8 10 12

mAU

0

20

40

60

80

100

120

140

160

433 Bar56 peaks

700 Bar52 peaks

2.1 x 100mm Competitive 1.7um Peptide ColumnIgG1 tryptic digest

2.1 x 100mm AdvanceBio Peptide Mapping ColumnIgG1 tryptic digest

Competitive UHPLC Peptide Column ComparisonDelivering UHPLC-type Speed and Separation Performance under HPLC Operation

Mobile phase A:water (0.1%FA)Mobile phase B:90% ACN (0.1%FA)

Inj.: 15uL Flow: 0.6 ml/minTemp.: 40 oC

Gradient (14min.)

Time(min.) %B0 310 3512 9014 90

LC/MS:Agilent 1290 coupled to 6530 QTOF

Page 28: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

September 10, 201428

433bar8x10

0

0.6

1.6

2.6

3.6

4.6+ESI TIC Scan Frag=200.0V igg011.d

Counts vs. Acquisition Time (min)0.51 1.52 2.53 3.54 4.55 5.56 6.57 7.58 8.59 9.51010.511 11.512 12.513

8x108

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

+ESI TIC Scan Frag=200.0V igg010.d

Counts vs. Acquisition Time (min)0.51 1.52 2.53 3.54 4.55 5.56 6.57 7.58 8.59 9.51010.51111.51212.513

700 bar

TOP Competitor UHPLC Peptide Column ComparisonMass Spec BioCharacterization Performance: UHPLC vs HPLC

14 min. run

14 min. run

UHPLC Results

Tryptic peptides i.d: 76 peptides

Sequence coverage.:99.00%

LC Pressure: 700 bar

2.1 x 100mm AdvanceBio Peptide Mapping Column, IgG1 tryptic digest

2.1 x 100mm Waters 1.7um CSH Peptide Column, IgG1 tryptic digest

HPLC Results

Tryptic peptides i.d: 76 peptides

Sequence coverage.:99.63%

LC Pressure: 433 bar

Page 29: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

ROBUSTNESS

September 10, 201429

Page 30: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

pH stability (phase)

09-Nov-1230

Column dim.: 2.1 x 250mmConditions: flow: 0.50ml/min., inj: 1uL, Temp: 55C, det:220nm, Mobile phase: A-water (0.1%TFA), B- ACN (0.08%TFA)Gradient: 0-8min ,10-60%B; 8.1-9min, hold 95%B, re-equilibration time: 8 minsInstrument: Agilent 1260 Bio-Inert

1 2 3 4 5Sample:1-Gly-Tyr2-Val-Tyr-Val3-Met Enk4- Angio II5- Leu Enk

Inj. #1 (395 bar)Inj. #200 (388 bar) Injection #RT1 (min) RT2 (min) RT3 (min) RT4 (min)

1 3.39 4.36 4.59 4.90

200 3.52 4.48 4.70 5.02

Injection # PW1 PW2 PW3 PW41 0.020 0.021 0.020 0.022

200 0.020 0.021 0.019 0.021

200 repeated injections at pH 2.2 at 55C – continuous HPLC operation for 2.5 days

Page 31: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

09-Nov-1231

Column Lifetime Efficiency (packed bed stability)

-50

-40

-30

-20

-10

0

10

0 1000 2000 3000 4000 5000

Plat

es L

oss [

%]

# of injections

Lifetime: 5000 Injections at 550bar

2.1 x 150mm2.1 x 100mm2.1 x 75 mm2.1 X 50 mm

Conditions: flow: varied for column length., inj: 1uL, Temp: 25C, det:280nm, isocratic runMobile phase: 70%-water, 30%- MeOH, Sample probe: NapthaleneInstrument: Agilent 1200

**

* stocked part #

Page 32: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

September 10, 201432

Rapid LC/MS IgG Tryptic Digest Peptide Mapping Reproducibility

min2 4 6 8 10 12

mAU

300

350

400

450

500

550

Peptide 1 Peptide 2 Peptide 3

Retention Time 1hydrophilic

Peptide 1 Peak Area

Retention Time 2Mid hydrophobic

Peptide 2Peak Area

Retention Time 3High hydrophobic

Peptide 3Peak Area

Mean(min)

% RSD Mean (mAU/s)

% RSD Mean(min)

% RSD Mean (mAU/s)

% RSD Mean(min)

% RSD Mean (mAU/s)

% RSD

1.54 0.51 38.46 3.77 7.06 0.06 172.74 2.62 10.41 0.04 38.460 8.65

run 4

run 6

run 10

run 8

run 2

10 Consecutive IgG injections on a 2.1 x 150mm AdvanceBio Peptide Mapping Column

Page 33: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

CAN THIS COLUMN BE USED FOR OTHER PEPTIDE APPLICATIONS?

September 10, 201433

Page 34: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

April 2012, Seminar34

Quantitative Peptide AnalysisMelittin Analysis

min6 6.5 7 7.5 8 8.5

mAU

5

10

15

20

25

30

35

40

45

6.87

0 6.93

1

7.04

0

7.78

1 7.83

1

8.04

1

min0 1 2 3 4 5 6 7 8 9

mAU

0

50

100

150

200

250

0.50

80.

564

0.59

50.

656

0.75

10.

934

0.94

6

3.12

53.

243

3.29

2

6.87

06.

931

7.04

0

7.78

17.

831

8.04

1

melittin

impurities

Expanded View

Melittin89.1%

Impurity 12.43%

Impurity 42.36%Impurity 3

1.17%

Impurity 24.16%

Impurity 51.04%

Full scale view

Melittin from bee honey (sigma) separated on a 2.1 x 150mm AdvanceBio Peptide Mapping ColumnMobile phase: A-water (0.1%TFA), B- ACN (0.08%TFA), temp: 55C, Flow: 0.50mL/min

Delivers excellent resolving power, sensitivity and selectivity for peptide impurity profiling

Increased flow rate capability maintains resolution and increase analysis times while column backpressures remain below 500 bar .

Page 35: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

35

Complex Peptide Mixtures and Small Proteins

min0 10 20 30 40 50 60

mAU

0

10

20

30

40

50

E. Coli lysate separated on a 2.1 x 250mm AdvanceBio Ppetide Mapping ColumnMobile phase: A-water (0.1%TFA), B- ACN (0.08%TFA), temp: 65C, Flow: 0.25mL/min

E. coli lysateCellular debris separation

Provides compatibility with TFA and Formic acid mobile phases for LC/MS analyses

Delivers a wide range of separation capability to selectively and efficiently resolve peptides and small proteins with high resolution.

Page 36: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

Summary:Peptide Mapping Separation Performance does not deteriorate when gradients are shortened and run times decreased

Critical PTM information is conserved with reduced LC/MS analysis times

Higher Sequence Coverage during rapid analysis in comparison to competitive Peptide column

UHPLC like speed and performance was achieved at traditional HPLC pressures and in comparison to competitive uHPLC Peptide Column.

High run to run reproducibility achieved on a complex biological

September 10, 201436

AdvanceBio Peptide Mapping Column

min0 5 10 15 20 25 30

mAU

0

50

100

150

200

250

300

Page 37: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

September 10, 201437

AdvanceBio Peptide Mapping ColumnLiterature & Reference Materials

Agilent Peptide Mapping “How-To” GuideFeatures AdvanceBio Peptide Mapping Column and How-to techniques/methods. pub# 5991-2348EN

Agilent Biocolumns Selection Guide: AdvanceBio Peptide Mapping Column configurations, part #s, etc

4 Application Notes (Martosella et. al.) Glycoprotein EPO mapping

(1) pub# 5991-2085EN(2) pub# 5991-1813EN

IgG1 mapping (1) pub# 5991-3585EN(2) in-print

Page 38: AdvanceBio Peptide An HPLC Column Technology for … mapping... · AdvanceBio Peptide Mapping An HPLC Column Technology for Faster Protein Biocharacterizations 1 September 10, 2014

Thank you!

38

QUESTIONS?