Acid Tioglicolic
-
Upload
alexandra-alexya -
Category
Documents
-
view
212 -
download
0
Transcript of Acid Tioglicolic
-
8/11/2019 Acid Tioglicolic
1/4
-
8/11/2019 Acid Tioglicolic
2/4
510 N. Xie et al. / Journal of Pharmaceutical and Biomedical Analysis 88 (2014) 509512
Tri-sodium phosphate dodecahydrate (Na3PO412H2O, 98%) was
bought from Beijing Chemical Reagent Co. (Beijing, China). Dode-
cyltrimethylammonium bromide (DTAB, 99%) was purchased from
J.K. Scientific Ltd. (Beijing, China). Tetradecyl trimethyl ammonium
chloride (TTAC, >98%) was obtained from Tokyo Chemical Industry
Co. Ltd. (Tokyo, Japan). Cetyltrimethylammonium bromide (CTAB,
99%)and calcium thioglycolate trihydrate (98%)were purchased
from SigmaAldrich, (St. Louis, MO, USA). Eighty-five samples
including fifty permanent wavings, 16 hair straighteners and 19
depilatory creams were bought from the local market.
2.2. Apparatus and conditions
All CE experiments were performed on a P/ACE MDQ sys-
tem (Beckman Instruments, Fullerton, CA, USA) equipped with
an auto-sampler and a photodiode array (PDA) detector. Data
acquisition and treatment was controlled using 32 Karat soft-
ware, version 8.0 (Beckman Coulter, Brea, CA, USA). The running
buffer was 300mmol L1 Na3PO4 (without pH adjustment) con-
taining 0.5 mmolL1 CTAB. A 1:10 dilution of the running buffer
was used as the sample buffer. An uncoated fused-silica capillary
(Yongnian Ruifeng Sepu Peijian Plant, Heibei Province, China) of
50mi.d.
40.2cm (375m o.d., effective length to the detectionwindow was 30 cm) was used. The new capillary was conditioned
by successively flushing with 1 mol L1 NaOH at 20 psi for 20min,
ultrapure water for 5min, and the running buffer for 5min. To
ensure the repeatability, before each subsequent run, the capillary
was rinsed with 1 mol L1 NaOH, ultrapure water, and the running
buffer for 3, 2, and 2 min, respectively. The separation was car-
ried out at a constant voltage of5 kV (current about 124A).
The detection wavelength was set at 236 nm. The sample was
introduced into the capillary by hydrodynamic injection at 0.5 psi
for 10 s. All electrophoresis runs were performed at a temperature
of 25 C.
2.3. Preparation of calibration curve
A TGAstandardstock solution (10mg mL1) was prepared daily
by accurately weighing 20 mg of calcium thioglycolate trihydrate
into a 1.5 mL of sample vial. 1 mL of water was added and then
vortex-mixed. A series of standard solutions at the final concentra-
tions of 0.006, 0.025, 0.125, 0.250, 0.500 and 1.000 mg mL1 were
made and kept in dark to avoid possible photodissociations[16].
2.4. Sample pretreatment
An appropriate amount of samples were weighed into a 15 mL
centrifuge tube with a cap and then diluted with sample buffer
to the mark of 10 mL. The solutions were thoroughly vortexed to
make the samples homogeneous. The mixedwater-based solutions
could be injected directly, while the mixed oil-based samples mustbe centrifuged at 9000 rpm for 10 min and then the resultant clear
supernatant was directly injected.
3. Results and discussion
3.1. Optimization of CE conditions
3.1.1. Selection of the running buffer
TGA is a weak organic acid with two pKa values, pKa1 = 4.32and
pKa2 = 10.20 which comes from the dissociation of its carboxyl and
thiol groups, respectively [17]. Asa ruleof thumb,thepH oftherun-
ning buffer should be adjusted to at least two units below the pKa1or above the pKa2 of TGA to ensure its complete ionization[18].
Phosphate, with low UV absorption and wide pH buffer capacity
Table 1
Comparison of the results of the TGA in 16 cosmetic samples determined by CE,
HPLC and IC.
Sample CE HPLC IC
Content (%, w/v) Content (%, w/v) Content (%, w/v)
1 / /
2 / /
3 0.7
4 / / /
5 / /
6 4.1 4.0 4.3
7 1.3
8 1.1
9 8.7 8.4 8.2
10 8.1 7.7 8.0
11a 2.7 2.6 2.8
12a 3.4 3.3 2.9
13a 2.3 2.4 2.3
14a 3.7 4.0 3.7
15a 2.7 2.7 2.6
16a 3.4 3.4 3.4
/, Not detected; , cannot be quantified due to interferences.a Oil-based sample.
(pKa1 = 2.12, pKa2 = 7.20 and pKa3 = 12.36), is a popular buffer sys-tem for CE. Considering that TGA salts at pH 7.09.5 or 12.7 and
TGA esters at pH 6.09.5 are usually used in cosmetic formulations
[6]and only phosphate can provide a good buffer capacity at pH
higher than (pKa2 (of TGA)+ 2), phosphate was chosen as a priority
running buffer for the subsequent optimizations.
3.1.2. Optimization of the concentration of Na3PO4and the pH of
the separation buffer
Although the analysis of only a single analyte TGA in cosmetic
samples was studied in this work, however, in such a case, the
researchers were oftendismayedby the interferences coming from
the sample matrix. In the present work, a real sample 12 (listed in
Table 1)was used to optimize the following conditions of the CE
method.The concentration of Na3PO4 played a key role for the sep-
aration of TGA. Keeping the concentration of CTAB (0.5 mmol/L)
unchanged, the concentration of Na3PO4(without pH adjustment)
increasing from 100mmol L1 to350mmol L1, not only decreased
the EOF, but also suppressed the adsorption of TGA to the capillary
wall. As shown in Fig.1, the detection sensitivity was enhancedand
the separation efficiency was increased. However, high concentra-
tions of phosphate would produce large amount of Joule heat due
to its inherently high conductivity,which could make theTGA peak
broadened. As a result, 300mmol L1 Na3PO4was chosen.
The pH of running buffer is another keyfactor for the separation
of TGA from the cosmetic sample matrix. It controls both the TGA
charge state and the level of EOF. Since the running buffer with pH
close to the pKa3value of H3PO4(12.36) can provide better buffer-ing capacity [19], the effectof the running buffer at pH 12.35,12.60
(without pH adjustment) and 12.85 were investigated while the
concentration of 300 mmolL1 Na3PO4 was kept constant. It was
found that under all the three pH values, the TGA could be baseline
separated from the interfering peaks coming from sample matrix.
Consideringthat therunningbuffer at pH 12.60could be easilypre-
paredand hada high buffercapacity,300 mmol L1 Na3PO4 without
pH adjustment was chosen for the following experiments.
3.1.3. The selection of the EOF-reversing agent and its
concentration
Samples are normally injected at the anode and detected at the
cathode. At pH > 7, the EOF is sufficient to ensure most species
to move from the anode to the cathode in the order of cations,
-
8/11/2019 Acid Tioglicolic
3/4
N. Xie et al. / Journal of Pharmaceutical and Biomedical Analysis 88 (2014) 509512 511
Fig. 1. Effect of Na3PO4 concentration on separation. Na3 PO4 concentration
(mmol L1 ): (a) 100, (b) 200, (c) 300, and (d) 350. (1) TGA and (2) unknown peak.
Running buffer: x mmol L1 Na3PO4 and 0.5mmolL1 CTAB. Uncoated capillary:40.2 cm50m, i.d.(30 cm to thedetector). Injectiontime: 10 s. Injectionpressure:
0.5 psi. Separation voltage: 5 kV, detection: UV 236 nm, temperature: 25 C.
neutrals and anions. However, the negatively charged TGA
migrated against the EOF and moved to the anode under the opti-
mized running buffer pH 12.60, which made it difficult to migrate
out of the capillary in a reasonable time (within 30min). To ensure
that the TGA could rapidly move to the detection window, an
EOF-reversing agent was needed[20].Long-chain alkyl trimethyl
quaternary salts are frequently used as EOF modifiers for anionic
analysis. In this study,threeEOF modifiers,namely DTAB, TTAC and
CTAB were investigated. The migration time of TGA was reduced
alongwiththe increasedlength of thealkylchain as shownin Fig.2.
Itmight be related to the surfactant structures, thelonger thechain,the stronger the adsorption and the ability to control EOF[21].As
a result, CTAB was selected.
Once EOF is reversed and reaches a stable value, the EOF no
longer depends on the CTAB concentration when its concentration
is 0.1mmolL1 which corresponds to 10% of its standard critical
micelle concentration. On this occasion, the surface of capillary is
a dynamic structure represented by the bilayer assembly instead
Fig.2. Effectof theEOF-reversingagenttypes onmigration time. (a)CTAB,(b) TTAC
and (c) DTAB. (1) TGA, other CE conditions were the same as in Fig. 1.
Fig. 3. Effectof samplebuffer on separation:(a) pure water as thesample medium,
(b) one-tenth dilution of the running buffer as the sample buffer, and (c) running
buffer as the sample buffer. (1) TGA and (2) unknown peak. Other CE conditions
were the same as inFig. 1.
of the solid backbone of silica[21].In order to further verify this
observation,CTAB concentrations at 0.2, 0.5 and1.0 mmol L1 were
studied, respectively, with no differences between the separation
efficiency and the migration time observed. Thus, CTAB with con-
centration of 0.5 mmolL1 was chosen because the EOF is stable at
this concentration[14,21].
3.1.4. Choice of sample buffer
For efficient sample stacking, the conductivity of the sample
buffer should be relatively lower than that of the running buffer.
In this study, pure water, one-tenth dilution of the running buffer
and the running buffer were tried as the sample buffer ( Fig. 3).As
expected, one-tenth dilution of the running buffer showedthe best
result.
3.2. CE method validation
3.2.1. Linearity, limit of detection, limit of quantitation, precision
and recovery
The corrected peak areas (Ac) versus the concentrations (,
mgmL1) of TGA showed a good linear relationship (Ac=50.33
+ 40.174) with a correlation coefficient (r) of 0.9998. The limit
of detection (S/N=3) and limit of quantitation (S/N= 10) were
0.002mgmL1 and 0.006 mg mL1, respectively.
Sample 12 was prepared seven times within a day to evaluate
the intra-day precision of the CE method. Samples were accurately
weighed (0.1g for each) and treated as described in Section2.4.To
evaluate the inter-day precision of the CE method, sample 12 waspretreated in triplicate each day on seven consecutive days. Both
the intra- and inter-day precisions of the method were 1.4% and
the average content of TGA in sample 12 was 3.4%.
The average recoveries at the three spiked levels (0.125, 0.250
and 0.500mgmL1) were 96.9%, 102.3% and 94.0% with relative
standard derivations 2.1%, 3.9% and 2.2%, respectively.
3.2.2. Comparison of CE, HPLC and IC results
Sixteen samples of hair-treatment products and depilatory
creams were analyzed by CE, HPLC and IC, respectively. The
results are listed in Table 1. For most of the samples, the
results of the TGA assayed by CE were in good agreement
with the values obtained by HPLC and IC. However, for the
HPLC method, six out of 16 cosmetic samples could not be
-
8/11/2019 Acid Tioglicolic
4/4
512 N. Xie et al. / Journal of Pharmaceutical and Biomedical Analysis 88 (2014) 509512
Fig. 4. Chromatogram and electropherogram of the sample 3. (A) Chromatogram, (B) electropherogram. (1) TGA and (2) interfering peak. Other CE conditions were the
same as inFig. 2. HPLC conditions: Mobile phase: acetonitrile and 0.01mol L1 potassium dihydrogen phosphate solution (pH 2.5) with volume ratio of 10:90. Flow rate:
1.0mLmin1. Detection wavelength: 215nm. Column temperature: 30 C. The sample injected was 20 L.
quantified due to the sample matrix interferences. The chro-
matogram and electropherogram of the sample 3 are shown in
Fig. 4. For the IC method, the TGA could not be quantified due
to the serious interferences suffering from the sample matrix
insamples 1,2, 3,5, 7 and8. Allthedatademonstrated the reliability
and practicality of the new CE method.
3.3. Real sample analysis
Sixty-nine commercially available samples, including forty per-
manent wavings, sixteen hair straighteners and thirteen depilatory
creams were successfully analyzed. None of these samples con-
tained TGA exceed the permitted level.
4. Conclusions
The developed CE method demonstrated the straightforward
preparation, robustness, and accuracy for TGA analysis, and has
been recommended for routine quality control analysis of cosmet-
ics.
Acknowledgement
The project for training high-level medical technical personnel
in health system in Beijing City is gratefully acknowledged.
References
[1] Y.F.Zabashta,A.V. Kasprova, S.P.Senchurov, Y.E.Grabovskii, The location of thethioglycolic acid molecules in intrafibrillar unordered areas of the human hairkeratin structure, Int. J. Cosmet. Sci. 34 (2012) 223225.
[2] K. Suzuta, S. Ogawa, Y. Takeda, K. Kaneyama, K. Arai, Intermolecular disul-fide cross-linked structural change induced by permanent wave treatment ofhuman hair with thioglycolic acid, J. Cosmet. Sci. 63 (2012) 177196.
[3] D. Oikawa, W. Takeuchi, S. Murata, K. Takahashi, Y. Sekine, Measurement ofconcentrationsof thioglycolicacid, dithiodiglycolicacid andammoniain indoorair of a beauty salon, J. Occup. Health 54 (2013) 370375.
[4] Z.Wang, X.Ren, D.Wang, Y.Guan, L.Xia, Effectsof thioglycolicacid onpartheno-genetic activation ofxenopusoocytes, PLoS ONE 6 (2011) e16220.
[5] L. Xia, S. Hou, X. Ren, Z. Wang, Effects of thioglycolic acid on in vivo oocytesmaturation in mice, PLoS ONE 6 (2011) e23996.
[6] T.Zhao, HygienicStandardforCosmeticsHealth, MilitaryMedicalPress,Beijing,China, 2007, pp. 239241.
[7] Z. Zhong, D. Du, C. Liang, J. Yao, Study on analysis of thioglycolic acid in cos-metics by ion chromatography, Wei Sheng Yan Jiu 33 (2004) 491493.
[8] V. Cavrini, V. Andrisano, R. Gatti, G. Scapini, HPLC determination of thiogly-colic acid and other aliphatic thiols in cosmetic formulations using ethacrynicacid as precolumn derivatization reagent, Int. J. Cosmet. Sci. 12 (1990)141150.
[9] J.M. Zen, H.H. Yang, M.H. Chiu, Y.J. Chen, Y. Shih, Determination of thioglycolicacid in hair-waving products by disposable electrochemical sensor coupledwithhigh-performanceliquidchromatography,J. AOACInt. 92 (2009)574579.
[10] Methods for Determination of Thioglycolic Acid in Cosmetics, Documents ofChinas Food and Drug Administration, Accessory Four, 2012, pp. 1719.
[11] L. Wang, Z. Chen, Determination of thioglycolic acid and cysteine in hair-
treatment products at ceramic carbon composite electrodes, Electroanalysis9 (1997) 12941297.[12] N. Bansal, A kinetic method for the determination of organosulfur compounds
by inhibition: determinationof cysteine,2,3-dimercaptopropanol and thiogly-colic acid, Transition Met. Chem. 35 (2010) 483490.
[13] M. Sierra-Rodero,J.M. Fernndez-Romero, A. Gmez-Hens, Photometric deter-mination of thioglycolic acid in cosmetics by using a stopped-flow reverseflow-injection system and the formation of gold nanoparticles, Microchem. J.97 (2011) 243248.
[14] E.A. Dutra, M.I.R.M. Santoro, G.A. Micke, M.F.M. Tavares, E.R.M. Kedor-Hackmann,Determination of-hydroxyacidsin cosmeticproductsby capillaryelectrophoresis, J. Pharm. Biomed. Anal. 40 (2006) 242248.
[15] P. Wang, X. Ding, Y. Li, Y. Yang, Simultaneous determination of eleven preser-vatives in cosmetics by micellar electrokineticchromatography, J. AOAC Int.95(2012) 10691073.
[16] A.R. Attar, D.E. Blumling, K.J. Knappenberger, Photodissociation of thioglycolicacid studied by femtosecond time-resolvedtransientabsorption spectroscopy,
J. Chem. Phys. 134 (2011) 024514.[17] P. Ke, X. Sun, Preparation of CdS self-assembled monolayers and interfacial
sensing for DNA, Chin. J. Huaqiao Univ. (Nat. Sci.) 32 (1) (2011) 4852.[18] Beckman Coulter, Introduction to Capillary Electrophoresis, Beckman Instru-
ment Inc., 1996, pp. 18.[19] K. Altria, Buffer preparation hints, tipsand common errors,LCGC AsiaPacific
10 (2) (2007) 2529.[20] J.E. Melanson, N.E. Baryla, C.A. Lucy, Dynamic capillary coatings for electroos-
motic flow control in capillary electrophoresis, Trends Anal. Chem. 20 (2001)365374.
[21] M.F.M. Tavares, R. Colombara, S. Massaro, Modified electroosmotic flow bycationic surfactant additives in capillary electrophoresis: evaluation of elec-trolyte systems for anion analysis, J. Chromatogr. A 772 (1997) 171178.