Accumate 2300 Volumetric Infusion Pump
Transcript of Accumate 2300 Volumetric Infusion Pump
Log
o
Accumate® 2300Volumetric Infusion Pump
Operating Manualwww.wooyoungmed.com
2
Important
This Operating Manual is for clinician
use only.
Do not permit patients to have access
to this Operating Manual.
Please read the entire Operating
Manual before operating the Accumate®
2300 volumetric infusion pump.
Failure to properly follow warnings,
cautions and instructions could result in
death or serious injury to the patient.
3
Warnings
If the pump is used to deliver life-sustaining medication, an additional pump must be available.
Prior to use, always check the audible and visual alarms.
The pump is not to be used for delivery of blood or cellular blood products.
Do not use this pump on mains power if the integrity of the protective earth system is suspect.
To prevent a possible explosion hazard do not use this pump in presence of inflammable anesthetic gases.
Do not use this pump in close proximity to equipment such as surgical diathermy with generates high levels of RFI.
Do not permit the use of mobile telephones in wards where this pump is in use.
If the pump is dropped or hit, inspect the pump for damage. Do not use a pump that is damaged or not functioning properly. Contact Customer Service to return a pump for service.
Do not permit this pump to used if standing in a pool of liquid.
System delivery inaccuracies may occur as a result of back pressure or fluid resistance, which depends upon drug viscosity, catheter size, and extension set tubing.
4
Warnings
Always read and follow the manual that accompany the source container and IV administration sets you are using. Carefully follow the manual for loading, removing, and reloading the set, as well as recommended set change intervals.
Change pumping section on tubing every 24 hours. Do not use tubing for more than 72 hours.
Always check pump data prior to starting infusion.
After starting an infusion, make sure drops are falling in the drip chamber. If no drops are falling, make sure the roller clamp is open. If the roller clamp is open and still no drops are falling, replace and dispose of the set.
The air detector cannot recognize the introduction of air at 3-way stopcocks, infusion ports, and other lines/tubes below the pump.
To avoid mechanical or electronic damage, do not steam autoclave or immerse the pump in any fluids or cleaning solutions. Always disconnect electrical power cord from outlet before cleaning to prevent electrical shock.
5
Cautions
Do not operate the pump at temperatures below +10◦C(50 ◦F) or above 40◦C(104◦F) and humidity levels below 30% or above 85%.
Do not store the pump at temperatures below +5◦C(41 ◦F) or above 40◦C(104◦F) and humidity levels below 10% or above 90%. Do not store the pump with the Administration Set attached.
Do not store the pump for prolonged periods with the battery installed (above 3months).
Do not immerse the pump in cleaning fluid or water or allow solution to soak into the pump, accumulate on the key pad, or enter the battery compartment.
Do not clean the pump with acetone, other plastic solvents, or abrasive cleaners.
Do not use the pump to therapeutic levels of ionizing radiation.
Do not use the pump directly to ultrasound.
Do not use the pump in the vicinity of magnetic resonance imaging (MRI) equipment.
Do not use the pump near ECG equipment.
Do not sterilize the pump.
All service and repair of the Accumate® 2300 pumps must be performed by Woo Young Medical or its authorized agents.
6
Table of Contents
In t roduct ion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Features of the Accumate® 2300 – Front View & Key Pad .................................................................8
– Controls and Indicators .................................................................9
– Re a r V ie w . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 0
Directions for Operation– Preparation of the IV-set ..............................................................12
– Turning ON the Power - Rear ............................................................13
– Turning ON the Power - Front ...........................................................14
– Insertion of the IV-set ................................................................15
– Setting the Flow Rate and Delivery Limit ...........................................17
– Clearing the Volume Delivered ..........................................................18
– Operating the Pump .............................................................18
– Purge Infusion (stop mode) ...............................................................19
– Bolus Infusion (start mode) ...............................................................19
– T i t ra t ion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
– Data Lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
– Adjust Sound ...... .... .. ... ... .. ... ... .. ... ... .... .. ... ... .. ... ... .. ... ... .... 22
– Occlusion Level ................................................................ 23
– Near End of Infusion ............................................................24
– End of Infusion ................................................................... 25
Alarms and Troubleshooting .....................................................26
Specif ications ..... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... .... ... . 28
Maintenance– Routine Maintenance Procedures ....................................................30
– Cleaning and Storage ..................................................................31
– Replacing the AC fuses ...................................................................32
– Battery operation ................................................................. 33
Occlusion Pressure Limits ........................................................34
Start-up & Trumpet Curves ......................................................35
Warranty ... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38
Service and Packaging ............................................. 39
Contact . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
7
Introduction
The Accumate® 2300 volumetric infusion pump is a fully
featured infusion pump suitable for critical care and
generation applications.
The Advantages
High precision.
Purge, automatic & manual bolus capability.
Easy to operate in only three steps.
– Power On > Select (rate, total, VTBI) > Start
Easy to read, large display.
History function.
– Historical data of infusion records.
(warnings, settings, starts and stops of infusion, etc.)
Micro-infusion for neonatal and high risk application.
None dedicated sets, available from various manufacturers.
Programmable, ultrasonic air bubble detector.
Multi-position clamp for vertical or horizontal fixation to poles.
SAFETY STANDARD
The Accumate® 2300 meets all safety standards for medical electrical devices, corresponding to IEC 60601-1 and IEC 60601-2-24.
8
Features of the Accumate® 2300
-Front View & Key Pad
a
b
c
d
e
fg h
ij
k
l
m
n
opqr
s
t uv
w
x
y
9
Controls and Indicators
NO. Name Function
a Handle To handle for carrying
b Name Plate To display the pump model and name
c Display All essential information is displayed (flow rate,
delivery limit, volume delivered, alarms
information)
d Locker To lock pump door
e Key Pad To set infusion data
f Finger To squeeze the tube with the spring housing by
peristaltic motion
g Occlusion
Sensor
To detect the downstream occlusion of the tube
h Clamp Lever To prevent free flow when the pump door is
opened
i Spring
Housing
To squeeze the tube with the peristaltic fingers
j Door Handle To lock pump door with locker
k Door To protect the tubing set when running
l Door Magnet To detect whether the door is open or not
m Bubble Sensor To detect air in the tube
n LED To indicates pump is infusing or lights up during
an alarm (normal : green, malfunction : red)
10
Controls and Indicators
NO. Name Function
o Volume
Delivered
Displays Volume Delivered (VTBI:Volume To
Be Infused)
p Delivery Limit Displays delivery limit or alarms information
q Flow Rate Displays flow rate (infusion rate)
r SELECT
/SILENCE
To select setting mode (flow rate, delivery limit,
volume delivered)
/ To silence alarm for 2minutes
s BATTERY
LEVEL
To display the charging status of internal
rechargeable battery
t EXTERNAL
POWER
To inform whether the external power is
connected or not
u PURGE To access PURGE or BOLUS key
v POWER
ON/OFF
Press once to switch the pump ON. Press and
hold down for 2 seconds to switch the pump
OFF
w START To start the infusion
x STOP
/CLEAR
To stop the infusion
/To clear the value when setting data
y UP/DOWN To increase or decrease of value shown on
display
11
Features of the Accumate® 2300
-Rear View
a
b
c
d
e
f g h i
j
k
l
NO. Name Function
a Multi-position Clamp To fix the pump to vertical pole stand or horizontal bar of bed
b Rear Support To support to pole stand
c Drop Sensor Port To connect the drop sensor with the pump
d Nurse Call/ Comm. Port To connect the nurse call jack or communication cable
e Front Panel Lock Switch To lock the front key pad
f 12V DC Jack To supply the power of 12V in ambulance car
g Protective Ground To connect the protective earth line
h Spec Plate To inform the product information
i AC Outlet/Fuse Socket To connect the AC 90~245V main power cord where the 1.25A fuse is
j Rechargeable Battery To hold the NIMH rechargeable battery
k Power Switch To switch the main power
l Clamp Knob To fix the clamp
12
Direction for Operation
Preparation of the IV-set
– Check for cleanliness, completeness, and
damage. If unit appears damaged, do not use it.
– Check the IV administration sets. If other IV
administration sets are used, the operating safety
of the pump can no longer be guaranteed.
Patient safety may, as a result, be compromised.
• Recommended IV-set
– BD(BECTON DICKINSON) :A122
– KV(KOREA VACCINE) :143T/243T
– Carefully purge the infusion set, without allowing
any air bubbles to enter, until the drip chamber is
to 1/3 full. If air has entered, repeat purge-
procedure.
– Close the tubing roller clamp.
1/3
chamber
close
13
Direction for Operation
Turning ON the Power - Rear
Attach the pump to an IV
pole by turning the knob
on pole clamp.
Plug the pump's power
cord into an electrical
outlet.
Turn ON the MAIN POWER
switch. MAIN POWER switch
must be ON to charge the
battery.
14
Direction for Operation
Turning ON the Power - Front
Check whether the AC
power LED is turned on.
External Power LED display
will light.
Press the POWER ON key
on the front panel.When turning the pump OFF, the
pump must be at STOP mode in
order for the POWER key to respond.
Wait the self test
procedure.
Check the previously
used data.
15
Direction for Operation
Insertion of the IV-set
Completely insert the IV-set
into the BUBBLE SENSOR.
Insert the IV-set chamber
into the IV bottle/bag. Fill 1/3
of the chamber. Take cautions to avoid any air bubbles.
Open the door by pulling
the DOOR HANDLE.
Push the CLAMP LEVER to right
to open. Place the IV-set into
the clamp. This is a safety feature that
prevents free flow when the door is opened
during infusion.
16
Direction for Operation
Insert the IV-set in the occlusion
guide.
Check if the IV-set is
straightforward.
Close the door by pushing
the tightly DOOR HANDLE.
The roller clamp should be below
the pump. Open the roller clamp
to be ready to infuse. Check that no drops are falling in the drip
chamber.
CAUTION : Make sure that IV tube is NOT loosely fastened into the internal
chamber of the pump. It should be straightforward.
17
Direction for Operation
Setting the Flow Rate and Delivery Limit
Press the SELECT key.
The digits in the LED digital
display will blink in 7seconds.
Use the UP(▲) DOWN(▼) keys to
set the desired rate of infusion.The rate of infusion can be set
from 0.1 to 999.9ml/hr.
Press the SELECT key.
The digits in the LED digital
display will blink in 7seconds.
Use the UP(▲) DOWN(▼) keys to
set the desired volume to be
infused. The volume to be infused can be
set from 0.1 to 9,999ml.
18
Direction for Operation
Clearing the Volume Delivered
Operating the Pump
Press the SELECT key.
The digits in the LED digital
display will blink in 7seconds.
Press the STOP/CLR key.
Press the START key.
Check the three LED turned
on sequentially.
19
Direction for Operation
Purge Infusion (stop mode)
Bolus Infusion (start mode)
Press and hold the PURGE key
when the pump is not infusion. Press and hold the PURGE key until fluid
flows and the purging of the IV-set is complete.
CAUTION : PURGE will infuse at the rate of 999.9ml/hr.The volume used during purging will be displayed, but it is
not added to the volume delivered.
The pump will not purge if the lock has been enabled.
Press the PURGE key when the
pump is infusion.PURGE will infuse at the rate of 999.9ml/hr.
The feature allows a bolus to be given during
an infusion.
Enter the amount of bolus to
be given in ml and Press the
START key. The amount of bolus
can be set from 1.0 to 99.9ml.
20
Direction for Operation
Titration (To change the flow rate while infusing)
Press the SELECT key.
The digits in the LED digital
display will blink in 7seconds.
Use the UP(▲) DOWN(▼) keys to
set the desired rate of infusion.The rate of infusion can be set
from 0.1 to 999.9ml/hr.
Press the START key.
CAUTION : The pump will not titration if the lock has been enabled.
21
Direction for Operation
Data Lock (This features prevents unauthorized
use by locking out all key except SILENCE key.)
Turn ON [KEY LOCK] button
to lock the key pad on the
front panel.
Press the any key and
check the displayed in '----'.
Check the set up and pump
status again.
22
Direction for Operation
Adjust Sound (This feature allows the volume of
the alarm tone to be changed)
Turn ON [KEY LOCK] button
to lock the key pad on the
front panel.
Press and hold the SELECT
key and UP(▲) /DOWN(▼)
keys to increase or decrease
the sound level.
Select the volume of the alarm
tone to be changed.
caution : The volume of the alarm to change can be set from
1(45dB/1m) to 5(65dB/1m)
23
Direction for Operation
Occlusion Level (The pressure at which a
downstream occlusion alarm will be triggered
varied. The higher the pressure limit, the less
sensitive the pump is changes in fluid
resistance.)
Turn ON [KEY LOCK] button
to lock the key pad on the
front panel.
Press and hold the SELECT
key and UP(▲)/DOWN(▼) keys
to increase or decrease the
occlusion level.
Select the occlusion level
to be changed.
caution : The occlusion level to change can be set low(7± 2psi),
center(11± 3psi), or high(15± 3psi).
24
Direction for Operation
Near End of Infusion (Sets the near end of
infusion warning time, as time left to End of
Infusion.)
Turn ON [KEY LOCK] button
to lock the key pad on the
front panel.
Press and hold the SELECT
key and UP(▲)/DOWN(▼) keys
to increase or decrease.
Select the near end of infusion
to be changed.
caution : The end of infusion to change can be set from 0(no set)
to 15minutes.
25
Direction for Operation
End of Infusion (The pump has reached the end
of the infusion)
In certain conditions the Accumate® 2300
automatically switches to the KVO (Keep-Vein-
Open) rate. The KVO-rate is set to 1.0ml/hr.
However, if the flow-rate set by the user is lower
than 1.0ml/hr, the KVO-rate is equal to flow-rate.
This alarm occurs 5 minutes
before infusion is completed.
Press the SILENCE key.
This alarm occurs when infusion
is completed. (KVO : Keep Vein
Open)
Press the STOP key to reset.
26
Alarms and Troubleshooting
Alarms are indicated by a combination of an
audible alarm, flashing alarm indicator and
abbreviated message in the display.
First press the SILENCE key to silence the
alarm for a maximum of 2 minutes, then check
the display for an alarm message. Press the
STOP key to cancel the alarm message.
If the infusion has stopped, rectify the cause of
the alarm then press the START key resume
the infusion.
27
Alarms and Troubleshooting
Alarm
Indication &
Sound
Possible Cause Action
Air-in-line
Air
bb-bb-bb
Air bubble air-in-line sensor
IV tube not properly inserted
into the Bubble Sensor
Remove the air from the tube
Completely insert the IV
tube into the Bubble Sensor
Door Open
OPEn
bb-bb-bb
Door opened during infusion Close the door
Occlusion
OccL
bb-bb-bb
Downstream tubing kinked
Filter blocked
Closed roller clamp
IV positional or infiltrated
Catheter or vein too small
Occlusion limit too low
Correct kinks
Correct blocked filter
Open roller clamp
Check IV site
Increase occlusion limit
Drive Error
Err
bb-bb-bb
Motor malfunction
Mechanical error
Request A/S
Request A/S
Low battery
BAtt
b-b-b
30 minutes until battery is fully
discharged
Connect pump to AC
power
End of infusion
End, conP
b-b-b
Programmed infusion is
complete
Stop infusion and reset
No Good
ng-?
silence
Self-Test NG
Internal Malfunction
Request A/S
28
Specifications
Name Volumetric Infusion Pump
Model Accumate® 2300
Pumping System Linear peristaltic
Flow Rate Range 0.1-999.9ml/hr (0.1ml/hr increments)
Delivery Limit Range0.1-999.9ml (0.1 ml increments), 1,000-9,999ml(1ml
increments)
Total Volume Equals to the upper limit of the delivery limit range
Purge Rate Range Equals to the upper limit of the flow-rate range
Accuracy +/- 5% nominal
KVO( Keep-Vein-
Open)0.1-5.0ml/hr(0.1ml/hr increments)
Adjustable occlusion[5.8-17.4psi(3 level : low, center, high)
Adjustable alarm volume (level up or down)
Infusion completed
Alarms Low battery(30 min. before infusion stop)
Battery depleted(3 min. before power off)
Air-in-line
Door open
Drive error
Near End of Infusion
Others
Titration while running
Self testing (System check)
One key bolus
Front panel lock
29
Specifications
EMS DC 12V
Electrical Safety Class I, Type BF
Dimensions 140(W)*225(H)*140(D)mm
Weight Approximately 2.5Kg
Power Requirements AC 90-245V, 50/60Hz
Battery Built-in rechargeable dc 12V(NI-MH 2.3Ah)
Battery Operation Ca. 4hr @ 100ml/hr
Battery Charging
Time7hrs
Operation Conditions 10-40℃(Relative humidity 30-85%)
Storage Conditions 5-40℃(Relative humidity 10-90%)
Accessories AC power cord for 220V, Operating manual
Warranty 2 years from the date of delivery
Options RS-485, nurse call, drop sensor
CAUTION : Specifications are subject to change without
notice.
30
Maintenance
Routine Maintenance Procedures
To ensure that pump remains in good operating
condition, it is important to keep it clean and carry out
routine maintenance procedures described below. All
servicing should only be performed by a qualified
service engineer with reference to the Technical
Service Manual for this product.
– As required
Thoroughly clean external surfaces of the pump
before and after prolonged period of storage.
– At least once per year
Inspect AC power supply plug and cable for
damage. Perform functional tests as outlined in
the Technical Service Manual. Operate the pump
on battery power until the battery low alarm then
charge the battery to confirm battery operation
and charging.
31
Maintenance
Cleaning and Storage
Before the transfer of the pump to a new patient and periodically during the use, clean the pump by wiping over with a lint-free cloth lightly dampened with warm water and a standard disinfectant/detergent solution.
Disinfectants which are known to be corrosive to metals must not be used. Concentrated Isopropyl alcohol based cleaners will degrade plastic parts.
The IV-set and extension lines are disposable single use items and should be discarded after use according to their manufacturers' instructions.
If the pump is to be stored for an extended period it should be first cleaned and the internal battery fully charged. Store in a clean, dry atmosphere at room temperature and, if available, employ the original packing for protection.
Once every 3 months during storage, carry out functional tests as described in the Technical Service Manual and ensure that the internal battery is fully charged.
Caution : Before cleaning always switch OFF disconnect from the AC power supply. Never allow liquid to enter the casing and excess fluid build up on the pump. Do not steam autoclave, ethylene oxide sterilize or immerse this pump in any fluid. Keep the pump plugged into electrical power during storage.
32
Maintenance
Replacing the AC fuses
If the pump continually illuminates the battery
indicator and the AC power indicator light does not
illuminate when the pump is connected to the AC
power supply and switched ON, suspect that either
the power supply fuse in the AC plug or the internal
fuses have blown.
First check the power supply fuse in the AC mains
plug. If the AC power indicator light does not
illuminate remove the pump from service.
It is recommended that only a qualified service
engineer replaces the AC fuses. For further
information regarding the replacement of internal AC
fuses refer to the Technical Service Manual.
33
Maintenance
Battery Operation
The internal rechargeable battery allows continued
operation when the AC power is unavailable, for
example during patient transfer or AC power failure.
A fully charged battery will provide over 4 hours of
operation at typical infusion rates.
A low battery alert is activated approximately 30
minutes before the battery depleted alarm.
The battery is maintenance free, sealed Nickel Metal
Hydride and requires no routine servicing. However,
to achieve optimum operation, ensure that the battery
is fully recharged after full discharge, before storage,
and at regular 3 month intervals during storage.
Charge retention will eventually degrade. Where
retention is critical the internal battery should be
replaced every 3 years.
It is recommended that only a qualified service
engineer replaces the Battery. For further information
regarding the replacement of batteries refer to the
Technical Service Manual.
34
Occlusion Pressure Limits
Time to alarm following occlusion is achieved in less
than 5minutes at rates of 25ml/hr and higher by the
appropriate selection of occlusion levels.
The following graphs show the typical values for time
to alarm and bolus volume that can be expected in
the event of an occlusion when the IV-set (KV 243T)
is selected.
Time to Alarm - 25ml/hr
0
2
4
6
8
10
12
14
16
0 1:00 2:00 3:00 4:00min : sec
Occ
lusi
on
Le
vel(
psi
)
Bolus Volume
0
0.2
0.4
0.6
0.8
1
1.2
1.4
1.6
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
Occlusion Level(psi)
ml
35
Start-up Curves
& Trumpet Curves
In this pump, as with all infusion systems, the action
of the pumping mechanism and variations in
individual IV-sets cause short-term fluctuations in
rate accuracy.
The following curves show typical performance of the
system in two ways : 1) the delay in onset of fluid
flow when infusion commences (start-up curves)...
and 2) the accuracy of fluid delivery over various time
periods is measured (trumpet curves).
The start-up curves represent continuous flow versus
operating time from the start of the infusion. They
exhibit the delay in onset of delivery due to
mechanical compliance and provide a visual
representation of uniformity. Trumpet curves are
derived from the second hour of this data. Tests
performed per IEC60601-2-24 standard.
36
Start-up Curves
& Trumpet Curves
Trumpet curves are named for their characteristic
shape. They display discrete data averaged over
particular time periods or 'observation windows', not
continuous data versus operating time. Over long
observation windows, short term fluctuations have ㅣlittle effect on accuracy as represented by the flat
part of the curve. As the observation window is
reduced, short term fluctuations have greater effects
as represented by the "mouth" of the trumpet.
Knowledge of system accuracy over various
observation windows may be of interest when certain
drugs are being administered. Short term fluctuations
in rate accuracy may have impact depending on the
half-life of the drug being infused, therefore the
clinical effect cannot be determined from the trumpet
curves alone.
Caution : Start-up and trumpet curves may not be
indicative of operation under negative pressure.
37
Start-up Curves
& Trumpet Curves
Start-up (25ml/hr)
05
1015
2025
30
17s 2
m5m
9m
13m
17m
31m
Time
Rate
(m
l/hr)
Trumpet Curve (25ml/hr)
-15.00%
-10.00%
-5.00%
0.00%
5.00%
10.00%
15.00%
17s 1m 5m 11m 19m 31m
Observation Window(mins)
Error
(%)
38
Warranty
Woo Young Medical Co., Ltd. offers a 24-month
warranty on every Accumate® 2300 unit (except
for the battery), effective from the date of
delivery.
The battery is free from defects in material and
workmanship under normal use and service for
a period one hundred eighty(180) days from the
date of delivery by Woo Young Medical to the
original purchaser.
This includes the repair or replacement of
damaged parts as a result of design/
manufacturing errors or material defects.
This warranty do not apply if Accumate® 2300
has been altered, misused, or repaired by
persons not authorized by Woo Young Medical.
The warranty do not cover the elimination of
problems caused by incorrect operation,
inappropriate handling normal wear and tear.
39
Service & Packaging
Service
– If the pump fails to respond to the operating or
troubleshooting procedures listed in this manual
and the cause cannot be determined, discontinue
use and forward it to an authorized Woo Young
Medical's agency.
– With each complaint, please include pump serial
number and full description of difficulty, including
all setting, types of fluids, time, and alarm
messages.
– Return IV-set used if possible.
– Woo Young Medical cannot assume
responsibility for loss or damage to returned
pumps while they are in transit.
Packaging
– The pump packaging is reusable.
– Package Contents : Accumate® 2300 Infusion
Pump, power cable, operation manual