A001-MNC Siemens Chiller

download A001-MNC Siemens Chiller

of 24

  • date post

    08-Nov-2014
  • Category

    Documents

  • view

    220
  • download

    2

Embed Size (px)

Transcript of A001-MNC Siemens Chiller

KKT WaterChiller forMAGNETOM Avanto 1.5T - A Tim System MAGNETOM Espree 1.5T - A Tim System MAGNETOM Trio 3T - A Tim System MAGNETOM Verio 3T - A Tim System

KSC 215-L-U/S KSC 215-L-S

(US Market) (R.O.W. Market)

Siemens AG - Healthcare Sector

KKT KRAUS Klte- und Klimatechnik GmbH

page 1

H:\Marketing\Drucksachen\Prospekte\Geraete\KSC 215_11-2008

The illustrations of products of the Siemens company are shown with friendly permission by Siemens AG - Healthcare Sector.

page 2

KKT KRAUS Klte- und Klimatechnik GmbHH:\Marketing\Drucksachen\Prospekte\Geraete\KSC 215_11-2008

List of Contents

Page 5 Page 6 Page 7 Page 8 Page 9 Page 10 Page 11 Page 12 Page 13 Page 14 Page 15 Page 16 Page 17 Page 18 Page 19 Page 20 Page 21

Description of KKT WaterChiller Nameplate KSC 215-L-U/S Technical Data KSC 215-L-U/S Nameplate KSC 215-L-S Technical Data KSC 215-L-U/S Dimensions KSC 215 (inch) Dimensions KSC 215 (mm) Key: Cold Water/Glycol Diagram Cold Water/Glycol Diagram Key: Refrigerant Diagram/Compact Design Refrigerant Diagram/Compact design WaterChiller Installation I WaterChiller Installation II IFP (Manufacturer Kutting) IFP (Manufacturer Kutting) Overvolting Protection Overview of WaterChiller Transfer Station/Remote Control Panel Transport

Page 22

KKT KRAUS Klte- und Klimatechnik GmbH

page 3

H:\Marketing\Drucksachen\Prospekte\Geraete\KSC 215_11-2008

page 4

KKT KRAUS Klte- und Klimatechnik GmbHH:\Marketing\Drucksachen\Prospekte\Geraete\KSC 215_11-2008

Description of KKT WaterChiller - Outdoor VersionThe KKT WaterChillers are designed, developed and produced as a compact system for the MAGNETOM Avanto/Espree/TaTs/Verio. KKT WaterChillers are suitable for outdoor installations. Power supplies of 380V 480V/50-60Hz. UL approved version is 480V/60Hz. Ambient temperatures up to 118.4 F [48 C]. Produced and testet in accordance with Siemens AG - Healthcare Sector specications.

frontal operation mainly Siemens products in the switch box totally integrated automation with display (Siemens S7) operator panel separate fault alarm switch box heater/fans UPS operation - optional extra, available through our customer service department (low starting current) no fuses (only automatic fuses/circuit breakers)

wired on clampscooling circuit including pump, regulation valve, lter strainer, shut-off ttings, automatic air vent, ll and outlet valves and pressure gauges

expansion receiver, safety valve ow control overheating protection 2 separate cooling circuits 50/50 % partial/emergency cooling possible with one compressor high quality safety and control instruments R 134a Ozon Safe Refrigerant

KKT KRAUS Klte- und Klimatechnik GmbH

page 5

H:\Marketing\Drucksachen\Prospekte\Geraete\KSC 215_11-2008

Nameplate KSC 215-L-U/S

page 6

KKT KRAUS Klte- und Klimatechnik GmbHH:\Marketing\Drucksachen\Prospekte\Geraete\KSC 215_11-2008

Technical DataWaterChiller Typ KSC 215-L-U/S outdoor use designed for MAGNETOM Avanto/Espree/TaTS/VerioStandard Device Standard Construction Note

Nominal cooling capacity (Btu/hr [kW]) Refrigeration medium Quantity of compressors Compressor capacity (HP [kW]) Volume of condenser airow (m/s [m/h]) Number of axial fans (piece) Condenser fan motor (HP [kW]) Max. ambient air temperature (F [C]) Min. ambient air temperature (F [C]) Cooling medium Min. cooling medium out (F [C]) Max. cooling medium out (F [C]) Temperature stabilty (K) Min. volume of cooling medium (gpm [l/min]) Ext. pump pressure (psig [bar]) Pump power (HP [kW]) Noise level (dB[A])

205,000 R 134a 2 18.2

[60.0]

[13.4]

Fabr.: DWM Type: ZR 16

10.0 4 1.54 +118 -4.0 PCW/AFN +66.2 +71.6 0.5 34.34 58.0 4.0 61 55 50 480/3/60 24 58.5 IP 54 4.49 2 3,490 [1,583] 2,425 [1,100]

[36,000] [1.15] [+48] [-20] PKW/AFN [+19.0] [+22.0] AFN = 35 - 38% mono ethylene glycol rate

[130.0] [4.0] [3.0] abt. 32.8 ft [10m] distance/118F[48C] abt. 32.8 ft [10m] distance/90F[32C] abt. 32.8 ft [10m] distance/70F[21C] tolerance -14/+10%V/1 Hz

Power supply (V/Ph/Hz) Control voltage (VDc) Full load amps (A) Form of protection Internal piping volume (gals. [l]) Plumbing Shipping weight (lbs [kg]) Operation weight (lbs [kg])

[17.0]

Warning The 35 - 38% Glycol rate does not depend on the local ambient temperature! Do not use Propylene Glycol or automotive antifreeze. Never mix different types or brands of Glycol.37 .8 96 0 in 0 m ch mh nc 7i 0.4 0 mm 2 1 06 3,

R 134a

KKT KRAUS Klte- und Klimatechnik GmbH

page 7

H:\Marketing\Drucksachen\Prospekte\Geraete\KSC 215_11-2008

72.05 inch 1,830 mm

Nameplate KSC 215-L-S

page 8

KKT KRAUS Klte- und Klimatechnik GmbHH:\Marketing\Drucksachen\Prospekte\Geraete\KSC 215_11-2008

Technical DataWaterChiller Typ KSC 215-L-S outdoor use designed for MAGNETOM Avanto/Espree/TaTs/VerioStandard Device Standard Construction Note

Nominal cooling capacity (Btu/hr [kW]) Refrigeration medium Quantity of compressors Compressor capacity (HP [kW]) Volume of condenser airow (m/s [m/h]) Number of axial fans (piece) Condenser fan motor (HP [kW]) Max. ambient air temperature (F [C]) Min. ambient air temperature (F [C]) Cooling medium Min. cooling medium out (F [C]) Max. cooling medium out (F [C]) Temperature stabilty (K) Min. volume of cooling medium (gpm [l/min]) Ext. pump pressure (psig [bar]) Pump power (HP [kW]) Noise level (dB[A])

205,000 R 134a 2 21.7

[60.0]

[16.0]

Fabr.: DWM Type: ZR 19

10.0 4 1.54 +118 -4.0 PCW/AFN +66.2 +71.6 0.5 34.34 58.0 4.0 61 55 50 380-480/3/50 24 55.0 IP 54 4.49 2 3,600 [1,633] 2,535 [1,150]

[36,000] [1.15] [+48] [-20] PKW/AFN [+19.0] [+22.0] AFN = 35 - 38% mono ethylene glycol rate

[130.0] [4.0] [3.0] abt. 32.8 ft [10m] distance/118F[48C] abt. 32.8 ft [10m] distance/90F[32C] abt. 32.8 ft [10m] distance/70F[21C] tolerance -14/+10%V/1 Hz

Power supply (V/Ph/Hz) Control voltage (VDc) Full load amps (A) Form of protection Internal piping volume (gals. [l]) Plumbing Shipping weight (lbs [kg]) operation weight (lbs [kg])

[17.0]

Warning The 35 - 38% Glycol rate does not depend on the local ambient temperature! Do not use Propylene Glycol or automotive antifreeze. Never mix different types or brands of Glycol.37 .8 96 0 in 0 m ch mh nc 7i 0.4 0 mm 2 1 06 3,

R 134a

KKT KRAUS Klte- und Klimatechnik GmbH

page 9

H:\Marketing\Drucksachen\Prospekte\Geraete\KSC 215_11-2008

72.05 inch 1,830 mm

Dimensions KSC 215 (inch)

page 10

KKT KRAUS Klte- und Klimatechnik GmbHH:\Marketing\Drucksachen\Prospekte\Geraete\KSC 215_11-2008

Dimensions KSC 215 (mm)

KKT KRAUS Klte- und Klimatechnik GmbH

page 11

H:\Marketing\Drucksachen\Prospekte\Geraete\KSC 215_11-2008

Key: Cold Water/Glycol Diagram

evaporator

high pressure pump

expansion receiver

ACS

advanced cooling system (Siemens)

FS

flow meter

IFP

interface panel

PI

pressure gauge

Hshut-off valve

TT

temperature sensor

HTA regulating valve manual air vent

filter strainer

automatic air vent

filling and draining cock

air receiver with automatic air vent back flow preventer

safety relief valve (water)

overflow valve

page 12

KKT KRAUS Klte- und Klimatechnik GmbHH:\Marketing\Drucksachen\Prospekte\Geraete\KSC 215_11-2008

Cold Water/Glycol Diagram

C6134_08-11-04

KKT KRAUS Klte- und Klimatechnik GmbH

page 13

H:\Marketing\Drucksachen\Prospekte\Geraete\KSC 215_11-2008

Key: Refrigerant Diagram

compressor R134a

Rotalock valve

condenser with fan

hot gas bybass valve

TC

electronic expansion valve evaporator

refrigerant shut-off valve with cap

variable speed control

PTLvibration absorber

PTHlow-/high pressure sensor

no return valve

PZAHH

high pressur cut out

refrigerant dehydrator

inspection gauge with moisture indicator

hot gas line receiver liquid line

Schrader valve

suction gas line

page 14

KKT KRAUS Klte- und Klimatechnik GmbHH:\Marketing\Drucksachen\Prospekte\Geraete\KSC 215_11-2008

Refrigerant Diagram

TC

evaporator 1

PTH

PTL

TT

M

M

M

M

TC

evaporator 2 condenser 2PTH

condenser 1

PTL

TT

System 2

P ZA H H

el.

PTL

PTH

System 1

PZA H H

el.

PTL

PTH

KKT KRAUS Klte- und Klimatechnik GmbH

page 15

H:\Marketing\Drucksachen\Prospekte\Geraete\KSC 215_11-2008

WaterChiller Installation I

Always keep unobstructed min.40.00inch min.1,000mm

keep free for service

min.45.80inch min.1,160mm

37.80inch 960mm

sound absorbant mat

min.40.00inch min.1,000mm

min. 15.75inch min.400mm operator control panel

min. 15.75inch min.400mm

keep free

air outlet

min.4.00inch min.100mm concrete bed 120.47inch 3,060mm min.128.47inch min.3,260mm

Dimensions for the concrete bed should be adjusted according to local codes.

page 16

~72.05inch ~1,830mm min.8.00inch min.200mm

KKT KRAUS Klte- und Klimatechnik GmbHH:\Marketing\Drucksachen\Prospekte\Geraete\KSC 215_11-2008

WaterChiller Installation II

installation e xample A

free access for servicing

installation e