A TIME OF FLIGHT DETECTOR FOR THERMAL NEUTRONS FROM
RADIOTHERAPY LINACS
V. Conti, G.Bartesaghi, D.Bolognini, M.Prest, S.Scazzi, P.Cappelletti, M.Frigerio, S.Gelosa, A.Monti, A.Ostinelli,
G.Giannini, E.Vallazza, A.Mozzanica
VCI - V. Conti February,19-24 2007 2
OUTLINE
• The Boron Neutron Capture Therapy and its future in a hospital environment
• A novel TOF detector : description and comparison with standard neutron detectors
• Spectra and flux measurements with a Clinac 2100
• Simulation developments and comparison with data
• Conclusion
VCI - V. Conti February,19-24 2007 3
RADIOTHERAPY
The radiotherapy goal local control of the primary tumour
Ideal situation : a large amount of energy deposited in the tumour volume and none in the surrounding healthy tissue
Radiotherapy beams
Electrons for superficial or semideep cancers
Photons are useful to treat deep-seated tumours that is tumours located several
cm below the skin surface
DEPTH IN WATER (cm)
VCI - V. Conti February,19-24 2007 4
Although the percentage of survival is increased in the years, some tumours have a very low five-year survival probability
VCI - V. Conti February,19-24 2007 5
For localized tumours Hadron therapy
BRAGG PEAK
Hadron radiation :
Negative pions, protons and light ions (e.g. helium, carbon, neon, silicon and argon nuclei)
The dose increases with penetration depth from a low dose at the entrance to a sharp maximum at the end of the particle range
Depth in water (cm)
VCI - V. Conti February,19-24 2007 6
PROTONX-RAYS
HADRON THERAPY ADVANTAGE
but this treatment cannot be used for some kinds of
tumour like
Extended tumours (liver, stomach or lung) Radio-resistant ones (melanoma)
Tumours surrounded by vital organs (brain)
VCI - V. Conti February,19-24 2007 7
The BNCT : Boron Neutron Capture therapy
The BCNT is an oncology radiotherapy treatment that exploits the capture of thermal neutrons by 10B and the following emission of an α particle and a nucleus of 7Li
Four years after the discovery of neutrons in 1932 by Chadwick, Locker introduced the concept of BNCT
VCI - V. Conti February,19-24 2007 8
specific drugs (BSH or BPA) are being developed for this purpose
boron is delivered mainly in the tumoral cells
Irradiating tumor cells with a thermal neutron beam thus producing heavy charged particles releasing their whole energy in the cell
High damage only in the target cells
healthy celltumor cell
10B
VCI - V. Conti February,19-24 2007 9
The BNCT requires:
• a high thermal neutron flux (> 5x108 n cm-2 s-1)
• with a low energy neutron spectrum (neutron energy < 10 keV)
Up to now such a beam is produced only in nuclear reactors
a reactor would be too expensive and too dangerous to be put inside a hospital
to develop a neutron source that can be used inside a hospital for BNCT treatments
PHONES PROJECT
AN IDEA
VCI - V. Conti February,19-24 2007 10
PhoNeS (PHOton NEutron Source)
The aim of the PhoNeS project is to provide a competitive thermal neutron source using clinac radiotherapy Linacs, maximizing with a dedicated photo-neutron converter the neutrons produced by Giant Dipole Resonance (GDR) by a high energy photon beam (> 8 MeV)
photo-neutron converter (PhoNeS)
Accelerator head (Clinac 1800, 2100)
VCI - V. Conti February,19-24 2007 11
• high energy photons hit a high Z target
• fast neutrons are produced by GDR
• neutrons are slowed down by a moderating material
• neutron capture inside the device should be kept low
• with as low as possible gamma field
To obtain a Photon Neutron Source useful for medical purposes the following requirements have to be fulfilled:
• a thermal neutron flux greater than 108 n cm-2 s-1
• a fast neutron and gamma dose per thermal neutron lower than 2x10-12 Sv cm2 n-1
Giant Dipole Resonance
VCI - V. Conti February,19-24 2007 12
Different prototypes and simulation have been developed to optimize the following parameters:
Photoconverter lead thicknessModerator shape
and thickness
Box of moderator material (D2O)
VCI - V. Conti February,19-24 2007 13
MCNP-4B GN code
Simulation WITH and WITHOUT PhoNeS prototype
without PhoNeS with PhoNeS
34 % slow 76 % slow
VCI - V. Conti February,19-24 2007 14
In this framework a development of neutron detection methods to evaluate neutron and gamma doses is of primary importance
COMMERCIAL NEUTRON DETECTORS:
TLD• passive detector, it integrates the flux in a certain energy range over the whole measurement duration
• not real time, it needs a long calibration procedure, an annhealing cycle before the use and after the readout
• also sensitive to photons
• high dynamic range
BUBBLE DOSIMETERS• integrated measurement over time and energy
• not real time; they can be used only with low doses; the counting of the bubbles can introduce large systematic errors
• poor spatial resolution
VCI - V. Conti February,19-24 2007 15
BF3 COUNTERS
• real time readout
• not tissue equivalent
• geometry constraints
MATERIAL ACTIVATION
• integrated measurement over time and energy
• a Ge or NaI detector nearby is needed
• Usually one sample per time. Almost real time
VCI - V. Conti February,19-24 2007 16
BEAM TIME STRUCTURE
Electrons (as well as secondary photons) are emitted in a 3-5 µs wide bunch at a 150-300 Hz repetition rate (dose rate dependent)
It means up to tens ms “empty” time between two following bunches
In this time interval there are no more electrons or photons in the environment, only thermal neutrons survive
a 0.025 eV neutron has a speed of 2.2 mm/µs
VCI - V. Conti February,19-24 2007 17
We are developing a real time neutron detector made of a scintillator coupled to a dedicated readout electronics
Plastic scintillators are low cost and flexible detectors, historically used for fast neutrons but can be employed also for slow ones in this particular configuration of the beam
For slow neutrons we can take advantage of the neutron capture by the hydrogen nuclei detecting the 2.2 MeV γ emitted in the reaction
n + H D + γ (2.2 MeV)
VCI - V. Conti February,19-24 2007 18
THE FIRST DETECTOR
A polystyrene plastic scintillator 2x2x1 cm3
Two P30CW5 photomultipliers by Electron Tubes with integrated high voltage power supply facing the two opposite sides of the scintillator to avoid dark counts and improve reliability
The detector assembly
VCI - V. Conti February,19-24 2007 19
Signals on the oscilloscope
Our detection system should be able to return the number of captured neutrons and their arrival time
The read out electronics
The signals from the scintillator are digitized by a NIM discriminator, threshold 30 mV
The coincidence between the two PMs is shifted to LVDS and then sent to sampler
sampling the discriminated signal with a 12.5MHz clock
VCI - V. Conti February,19-24 2007 20
This is an output example:
each “1” corresponds to an impinging neutron
For each experimental condition several bunches have been collected, obtaining a time arrival profile
VCI - V. Conti February,19-24 2007 21
WE HAVE TWO GOALS:
TO MEASURE
The neutron number The neutron energy spectrum
VCI - V. Conti February,19-24 2007 22
The comparison between the Al sample activation and the real time detector measurements in the scan along a diagonal from the
accelerator head
Measurement setup in S.Anna Hospital (Como, Italy)
VCI - V. Conti February,19-24 2007 23
This detector can be used to measure the energy spectrum
The TOF spectrum has been fitted with a double exponential and the result is :
Two different slopes:
• 135 µs
• 1.567 ms
These constants represent the contribution of a slower and a faster part of the neutron spectrum.
is it true?
VCI - V. Conti February,19-24 2007 24
Scan of PMMAAccelerator head rotated of 90° detector
PMMA layer
• the slow constant from the accelerator head material (lead)
• the fast constant from a plastic material in the room
VCI - V. Conti February,19-24 2007 25
The fast component is affected by the PMMA, since
The slow component doesn’t feel the effect of PMMA
Ratio fast/slow
More PMMA=more thermalization
From the TOF of each point ….
VCI - V. Conti February,19-24 2007 26
THE SECOND PROTOTYPE using a boron loaded plastic scintillator
to evaluate an α and 7Li contributions
BC-454(Bicron) a boron loaded plastic scintillator
diameter = 2 cm thickness=0.5cm
P30CW5 photomultiplier by Electron Tubes
The assembled detector
VCI - V. Conti February,19-24 2007 27
Comparison of the scintillator data with the activation method and the simulation data
DETECTOR
DOSIMETERIt counts the photons
produced by the neutrons in the scintillator plus in all surrounding materials
VCI - V. Conti February,19-24 2007 28
the neutron energy spectrum is extracted from the simulation TOF
The TOF simulation comparison with data
This is a very important result
for a BNCT treatment you need the information on the spectrum
VCI - V. Conti February,19-24 2007 29
CONCLUSION• BNCT in hospitals? … a dream becoming reality …
• A real time detector made of standard plastic scintillator that is both a relative DOSIMETER and a SPECTROMETER
• Can be used as a dosimeter at reactors?
• Flux measurement = real time equivalent of attenuation
Boron doped scintillating fibers or very thin scintillators
that will allow to disentangle the contribution due to the photons produced by neutron capture in the surrounding materials.
OUTLOOKS
•Absolute dose computation both with Linac and reactors
• Fiber system development
• Measurements with Phones with different Linacs
VCI - V. Conti February,19-24 2007 30
A TIME OF FLIGHT DETECTOR FOR THERMAL NEUTRONS FROM
RADIOTHERAPY LINACS