A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY FOR OPTICAL INTERROGATION OF NEURAL CIRCUIT DYNAMICS By MAY MANSY A DISSERTATION PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY OF FLORIDA IN PARTIAL FULFILLMENT OF THE REQUIREMENTS FOR THE DEGREE OF DOCTOR OF PHILOSOPHY UNIVERSITY OF FLORIDA 2019

Transcript of A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would...

Page 1: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY FOR OPTICAL

INTERROGATION OF NEURAL CIRCUIT DYNAMICS

By

MAY MANSY

A DISSERTATION PRESENTED TO THE GRADUATE SCHOOL

OF THE UNIVERSITY OF FLORIDA IN PARTIAL FULFILLMENT

OF THE REQUIREMENTS FOR THE DEGREE OF

DOCTOR OF PHILOSOPHY

UNIVERSITY OF FLORIDA

2019

Page 2: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

© 2019 May Mansy

Page 3: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

To my mother, for empowering me

To my father, for supporting me

To my daughter, for bringing out the best in me

To my husband, for always being by my side

To my family, who fill me with pride

Page 4: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

4

ACKNOWLEDGMENTS

The true treasures of life are in the people, who stand by us through thick and thin, and

who never cease to believe in us. I have been tremendously blessed to have the most wonderful

people in my life, who have contributed directly and indirectly to the completion of this

dissertation. I would like to start by giving huge thanks to my academic advisor, Dr. Karim

Oweiss, for granting me the opportunity to pursue a Ph.D. in his lab and for his unwavering

efforts to support and guide me over the past six years. I’m immensely grateful for all the

valuable pieces of advice and teachable lessons that improved my critical and analytical thinking

as well as the guidance Dr. Oweiss endowed me throughout the years of my research.

Additionally, I would like to thank Dr. Oweiss for always giving me the opportunity to attend

various scientific research conferences and for encouraging me to engage in activities that

nurtured my personal and professional development. Sincere gratitude also goes to the members

of my committee, Dr. Kevin Otto, Dr. Thomas Foster, Dr. Minghzou Ding and Dr. Brandi

Ormerod for their valuable feedback, advice and continuous support.

Furthermore, I would like to express my genuine appreciation to the graduate advisor of

my Ph.D. program, Dr. Cherie Stabler, who has always been there for me and who never ceased

to believe in me. I would like to thank Dr. Stabler for all the eye-opening, enlightening and very

up-lifting discussions we had. Next, I would like to express my thanks to Dr. Brandi Ormerod

not just for always empowering me and for the many valuable pieces of advice, but also for

being the first to help me discover my teaching talent.

A token of appreciation also goes to Dr. Marwan Abdellah for the opportunity to

collaborate with the Blue Brain Project at EPFL, which made a major part of this dissertation

possible.

Page 5: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

5

I will now turn to the amazing members of the Oweiss lab. It was a real pleasure to work

and interact with each one of the current and former members of the Oweiss lab. Brandey

Andersen, who was more than just a lab colleague and friend, and who always helped me find

my way. Islam Badreldin who never fell short of providing technical advice for any technical

challenge. Dr. Ali Ibrahim who taught me perseverance. Ben Goolsby who always reminded me

to calm down and who always helped review my writing tasks. Joseph Canzano for the fruitful

scientific chats and brainstorming sessions that always helped me critique my thinking and

identify shortcomings of my research. Dr. Narayan Subramanian, for teaching me about

immuno-histochemical techniques, helping me course-correct when I lost track and get up when

I hit rock-bottom. Rebeca Castro for helping me collect my data during the last phase of my

PhD, Naoki Sawahashi for helping resolve some challenging technical issue I faced towards the

end of my degree, and all the other wonderful members who shaped my experience at the Oweiss

lab: Hong-Jae Kim, Phillip Navarro, Mehrdad Hashemi, Joseph Succar, Michael Brodowski,

Abduraham Siddiqi, Kyle Sheller.

I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

opportunity of an internship at Exactech, for being a constant source of support and

empowerment and for the abundant guidance and advice during my job-hunting phase.

I would also like to acknowledge Dr. Erin Patrick for being a great teaching partner and

mentor. I’d like to thank Dr. Patrick for teaching me about teaching, for her valuable feedback on

my performance and for providing the ultimate guide on how to pursue a career in teaching.

Furthermore, I would like to express my deepest appreciation to Professor William

Mcelroy for teaching me things beyond “Engineering Leadership”, providing precious tips and

advice, and for always checking on me.

Page 6: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

6

I would also like to express my appreciation to all the administrative staff members who

helped me with filing paperwork, placing orders for my research supplies, installing new

equipment, delivering mail and maintaining my workstation: Kathryn Thompson, Jason Kawaja,

Marcy Lee, Ray McClure, Quashawn Durant, Kimberly Depue, Michelle Evern, Myra Edwards,

Amanda Redinger, Victor De La Cruz and Kaitlynn Gravely.

I would like to thank my husband, partner and best friend Islam Badreldin, for being my

safe place, for being by my side and for always taking care of the things that I get on my nerves.

I deeply appreciate his support and back up during my downtimes and during my times of stress.

I would like to thank my daughter, Noor, for teaching me how to value and enjoy every minute

of every day and for bringing out the best in me with the simplest of words. Her endless

questions and curiosity about the world make me learn new things and acquire new ways to teach

her about them.

The final token of sincere gratitude goes to my siblings and my loving parents for their

endless love, care, and support. I can’t thank them enough for instigating high moral standards

and discipline in me, for allowing me to be who I am, and for always having my back.

Page 7: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

7

TABLE OF CONTENTS

page

ACKNOWLEDGMENTS ........................................................................................................... 4

LIST OF TABLES .................................................................................................................... 10

LIST OF FIGURES .................................................................................................................. 11

LIST OF ABBREVIATIONS.................................................................................................... 13

ABSTRACT ............................................................................................................................. 14

CHAPTER

1 INTRODUCTION ............................................................................................................. 16

Centuries of Choreographed Synergy ................................................................................. 16

Dissertation Overview ........................................................................................................ 24

2 BACKGROUND AND MOTIVATION ............................................................................. 25

Let there be Light ............................................................................................................... 25 … and there was Light ....................................................................................................... 25

The Origin of Light ..................................................................................................... 26 The Light Duality ........................................................................................................ 26

Basics of Illumination ........................................................................................................ 29 Parameters of Light ..................................................................................................... 31

Radiometric Quantities ................................................................................................ 32 Basics of Optics .......................................................................................................... 33

Propagation of Light .......................................................................................................... 37 Propagation in an Optic Fiber ...................................................................................... 37

Propagation in Neural Tissue ....................................................................................... 38 Fundamentals of Fluorescence............................................................................................ 41

Optical Interrogation of Neural Circuity ............................................................................. 46 Optical Actuators ........................................................................................................ 47

Optical Sensors ........................................................................................................... 49 Comparison of Fluorescent Imaging Techniques ................................................................ 52

Two-Photon Imaging ................................................................................................... 52 Single-Photon Micro-endoscopy.................................................................................. 54

Single-Photon Fiber Photometry .................................................................................. 54 Motivation and Research Aims ........................................................................................... 58

Significance ................................................................................................................ 60 Research Aims ............................................................................................................ 60

3 SPATIAL CHARACTERIZATION OF THE DETECTION VOLUME ............................. 61

Background ........................................................................................................................ 61

Page 8: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

8

Methods ............................................................................................................................. 62 Fiber Photometry Signal Acquisition ........................................................................... 62

Phantom Brain Preparation .......................................................................................... 63 Acute Brain Slice Preparation ...................................................................................... 63

Fiber Photometry Recording Procedure ....................................................................... 63 Results ............................................................................................................................... 66

Volume of Detection from Phantom Slices .................................................................. 67 Volume of Detection from Acute Brain Slices ............................................................. 69

Discussion .......................................................................................................................... 73

4 VALIDATION OF THE SPATIAL CHARACTERIZATION ............................................ 77

Premise and Hypothesis ..................................................................................................... 77 Methods ............................................................................................................................. 80

Surgical Procedure ...................................................................................................... 80 Fiber Photometry Recording ........................................................................................ 81

Data Acquisition ......................................................................................................... 81 Visual Stimulation ....................................................................................................... 81

Epi-fluorescent Volumetric Scanning .......................................................................... 82 Results ............................................................................................................................... 84

Discussion .......................................................................................................................... 91

5 EMPIRICAL MODELING AND PREDICTION OF THE DETECTION VOLUME ......... 93

Monte-Carlo Simulation ..................................................................................................... 94 Optical Properties of Neural Tissue ............................................................................. 95

Results ........................................................................................................................ 96 A Novel Prediction Tool .................................................................................................. 100

Artificial Neural Network .......................................................................................... 100 Results ...................................................................................................................... 102

Discussion ........................................................................................................................ 106

6 CONCLUSION AND FUTURE WORK .......................................................................... 108

Summary .......................................................................................................................... 108 Future Work ..................................................................................................................... 110

Conclusion ....................................................................................................................... 111

APPENDIX

A SYSTEM DESCRIPTION ................................................................................................... 113

B PILOT EXPERIMENTS ...................................................................................................... 120

C SUPPLEMENTARY MATERIAL ...................................................................................... 130

LIST OF REFERENCES ........................................................................................................ 135

Page 9: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

9

BIOGRAPHICAL SKETCH ................................................................................................... 150

Page 10: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

10

LIST OF TABLES

Table page

2-1 Comparison of the three main fluorescent imaging techniques. ...................................... 57

3-1 Detection volume for different optical fibers. ................................................................. 72

3-2 Summary of recent FP studies. ....................................................................................... 75

4-1 Summary of anatomical locations .................................................................................. 91

5-1 Summary of reported optical properties of the rodent brain. ......................................... 107

Page 11: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

11

LIST OF FIGURES

Figure page

1-1 The first historical report about single cells. ................................................................... 18

1-2 The first description of a single neuron. ......................................................................... 18

1-3 The first published graph of an action current. ............................................................... 20

2-1 Newton’s observation on refraction. .............................................................................. 28

2-2 Electromagnetic spectrum.. ............................................................................................ 30

2-3 Electromagnetic wave. ................................................................................................... 32

2-4 Illustration of Specular and diffuse reflection. ................................................................ 34

2-5 Refraction and reflection................................................................................................ 36

2-6 Cone of acceptance. ....................................................................................................... 36

2-7 The properties of an optical fiber. .................................................................................. 38

2-8 Jablonski diagram. ......................................................................................................... 43

2-9 Example excitation and emission spectrum of a green fluorescent fluorophore. ............. 45

2-10 Different configurations of fluorescent microscopy. ....................................................... 46

2-11 All-optical interrogation of neural circuit dynamics. ...................................................... 47

2-12 Commonly used optogenetic actuators and their excitation spectra. ............................... 48

2-13 Single vs. Two-photon excitation. .................................................................................. 53

2-14 Comparison of fluorescent imaging techniques. ............................................................. 55

2-15 Qualitative comparison chart. ........................................................................................ 57

2-16 Neurometric curve as a function of stimulus intensity. ................................................... 59

3-1 Experimental setup. ....................................................................................................... 65

3-2 Agar-bead setup. ............................................................................................................ 66

3-3 Spatial detection extent in brain phantom. ...................................................................... 68

Page 12: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

12

3-4 Spatial characterization of detection volume as a function of fiber geometry in acute

brain slices..................................................................................................................... 71

3-5 3D Volume of detection for each optical fiber. ............................................................... 72

4-1 Epifluorescent volumetric scanning setup and LMM...................................................... 83

4-2 In-vivo validation of the directly measured detection profile and volume, location E.. ... 87

4-3 In-vivo validation of the directly measured detection profile and volume, location B. .... 88

4-4 In-vivo validation of the directly measured detection profile and volume, location C. .... 89

4-5 In-vivo validation of the directly measured detection profile and volume, location D ..... 90

5-1 Detection maps simulated with the developed FPMCS. ................................................. 98

5-2 Simulated and directly measured detection boundaries and axial detection profiles. ....... 99

5-3 Empirical model of the detection volume. .................................................................... 105

A-1 GCaMP6f excitation spectrum. .................................................................................... 114

A-2 Diagram of the Fiber Photometry system. .................................................................... 115

A-3 Sample Ca2+ trace recorded with FP. ............................................................................ 118

B-1 Setup of the experiment. .............................................................................................. 123

B-2 Results of Experiment 1. .............................................................................................. 124

B-3 Methods and results of the experiment. ........................................................................ 129

C-1 Surgical procedure and evaluation. .............................................................................. 133

C-2 Quantification of the visually evoked responses. .......................................................... 134

Page 13: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

13

LIST OF ABBREVIATIONS

1p Single-photon

2p Two-photon

ANN Artificial neural network

AP Anterior-posterior

Ca2+ Calcium

CCD Charged coupled device

Det. Detection

DV Dorso-ventral

Em. Emission

ephys Electrophysiology

Ex. Excitation

FP Fiber Photometry

FPMCS Fiber Photometry Monte Carlo Simulation

GECI Genetically encoded calcium indicator

GEVI Genetically encoded voltage indicator

MCS Monte Carlo Simulation

ML Medio-lateral

OS Optically scanned

TIR Total internal reflection

TPI Two-photon imaging

Page 14: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

14

Abstract of Dissertation Presented to the Graduate School

of the University of Florida in Partial Fulfillment of the

Requirements for the Degree of Doctor of Philosophy

A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY FOR OPTICAL

INTERROGATION OF NEURAL CIRCUIT DYNAMICS

By

May Mansy

August 2019

Chair: Karim Oweiss

Major: Biomedical Engineering

Identification of the neural signature of behavior is the ultimate goal of neuroscience.

Recent efforts have been devoted to developing all-optical tools for cell-type-specific

interrogation of Ca2+ dynamics in awake behaving subjects. Fiber Photometry (FP) is one such

tool that allows recording and manipulating the aggregate activity of many fluorescing neurons

in vivo. While FP has a good temporal resolution, it has poor spatial resolution compared to

other widely used tools such as microelectrode arrays, single-photon micro-endoscopy, and

multi-photon laser-scanning microscopy.

This dissertation aims to fill a knowledge gap regarding the temporal and spatial

characteristics of the fluorescing tissue volume that the optical fiber can record from as a

function of the fiber geometry. First, the spatial extent of the detected fluorescence in phantom

brain tissue as well in acute brain slices is quantified. The results suggest that two critical device

parameters, namely the numerical aperture and the fiber diameter, affect the possible detection

volume. Second, we demonstrate experimentally that FP can be used to record sufficiently

sensitive signals to characterize the orientation tuning of neurons in the mouse visual cortex.

Validation of the FP signal is achieved in this setting by axial scanning of the hypothesized

detection volume using epi-fluorescent CCD imaging, permitting the weighted reconstruction of

Page 15: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

15

the FP signal from the CCD-derived fluorescing sources. Third, an empirical model consisting of

an artificial neural network (ANN), which is trained to predict the FP detection volume for

different combinations of fiber diameters and numerical apertures, is developed. The model is

trained using a biophysically-plausible Monte Carlo simulation (MCS) of scattering photons in

brain tissue to simulate detection volumes for any fiber geometry. Results suggest a general

agreement between the experimental data (acute and in vivo preparations) and the simulated data

(MCS). In summary, this research contributes knowledge about the use of fiber photometry as a

ubiquitous tool for biological investigations given its ever-increasing use by the scientific

community.

Page 16: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

16

CHAPTER 1

INTRODUCTION

Centuries of Choreographed Synergy

“It takes two to tango” would perfectly describe the progression of the dynamic and

symbiotic relationship between optical microscopy and neuroscience. Modern microscopic

neuroscience is the field where the development of a new optical instrument enables a

fascinating scientific discovery, and that discovery then demands an upgrade to the instrument to

further the findings and vice versa. The scientists’ constant curiosity-driven demands and the

availability of the device, constrained by manufacturing limits, choreographed this ongoing

dance and created eras of different performances. It would be fair to say that it all began when

the first microscope inspired the emergence of the term ‘cell’.

The word ‘cell’ did not exist until 1665 when Robert Hooke published his

groundbreaking book, Micrographia, which is Latin for little pictures. This is the first report in

history on biological micrographs of small organisms and objects including, insects, the tip of a

needle and the cross-section of plants. In his 18th observation (out of a total of 60 observations)

Hooke describes the hollow structures in thin slices of cork. The discrete nature and unified

shape of the structures reminded him of the cells of a monastery, and he termed “cells” (Figure

1-1). This was the first occurrence for the word cell in conjunction with a biological system. In

his experiments, Hooke used a simple single-lens microscope to achieve high magnification and

a rather unrefined compound microscope to investigate larger fields of view [1].

Decades later, when glass grinding and lens polishing techniques were improved, Antony

van Leeuwenhoek picked up where Hooke left off and made the first description of the bovine

optic nerve using a 10x compound microscope. In the time between 1832 and 1834 Jan Purkyně

(Purkinje) and his student, Gabriel Valentin, used a then state-of-the-art 40x compound

Page 17: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

17

achromatic microscope made by the Austrian optical instrument maker Simon Plössl to examine

the structure of nerve tissue in animals and humans [2], [3]. In his publication “Über den Verlauf

Und Die Letzten Enden Der Nerven” (German for: on the course and the terminals of nerves)

Valentin makes the first depiction of a cortical neuron that has a “tail-like process” and

postulates that nerve fibers are not hollow as was the consensus then (Figure 1-2) [4]. A

controversial discovery that surely instigated the interest to dig deeper and see more.

Over the course of the years, optical components were further enhanced to boost

magnification and achieve better visualization of the microscopic structure of nerves. As a result,

more refined anatomical observations became possible like Robert Remak’s description of

myelinated and unmyelinated axons and Jan Purkyne’s (Purkinje) report on cerebellar cells. And

so, the dance goes on until advances in tissue preparation, histology and staining allowed Ramon

y Cajal to visualize different compartments of a neuron and pave the way for the neuron doctrine

in 1875 [5]. Anatomical investigation of the brain and the nervous system was -and still is -

fueled by the agile progression in optical manufacturing and the comprehensive exploration of

light properties, which are the essential elements of optical structural visualization. Anatomical

studies, therefore, continued to be at the forefront. Physiological and functional assessment of the

nervous system, on the other hand, took a slow time course.

Functional assessment of the nervous system pivots on two major requisites: 1- sufficient

knowledge of anatomy (structure) and 2- the maturity of the device technology. Observing the

activity of the nervous system requires a device that operates at a speed and a precision

comparable to the temporal dynamics of the neural activity. Hence, the dance between device

technology and functional neural recording is more of a solo dance challenge, where one dancer

Page 18: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

18

must wait (sometimes for decades) for the other to finish their performance before they get a turn

to perform.

A)

B)

Figure 1-1. The first historical report about single cells. A) Front cover of Robert Hooke’s book

“Micrographia: OR SOME Physiological Descriptions of MINUTE BODIES MADE

BY MAGNIFYING GLASSES WITH OBSERVATIONS AND INQUIRIES

thereupon”. B) Excerpt summarizing pages 112-116 of the book [6]

A)

B)

Figure 1-2. The first description of a single neuron. Top: original drawing and German text.

Bottom: digitally enhanced drawing and English translation of text [7].

Page 19: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

19

Galvani’s discovery of the electrical properties of nerve fibers in the frog nerves (1791)

was the first solo on nerve electrophysiology inspiring several scientists to use galvanic currents

to electrically stimulate nerves and measure the effect using Galvani’s galvanometer. However,

the limited speed and sensitivity of the electromechanical galvanometer hampered the

progression of electrophysiology and restricted it to stimulation experiments only. Therefore,

electrophysiology recording from nerve fibers was halted for decades. Alternatively, one could

think of it as the waiting phase in the solo dance challenge when the other is performing their

dance.

In this phase (1845-1902), Dubois Reymond and his students made major improvements

on the side of instrumentation development to allow the detection of small physiological currents

and their amplification. Reymond and his team built the “astatic galvanometer”, which was

relatively more sensitive than the regular galvanometer, yet still fell short on size, sensitivity, and

speed, limiting the investigation to large nerve bundles and muscles while fine nerve fibers were

still not recordable. Nevertheless, their novel tool allowed the observation of the extraordinary

phenomena of “nerve current” and “muscle current” and “negative variation” (negative

Schwankungen) which were proven quantitatively later by his student Julius Bernstein [6], [7].

Julius Bernstein’s ingenious invention of the differential rheotome extended the astatic

galvanometer and allowed plotting the time and magnitude course of the nerve impulse. The first

graphical representation of the “action current” (Aktionsstrom), now known as an action

potential, was published in 1868 [8] and marks the dawn of neurophysiological instrumentation.

Page 20: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

20

Figure 1-3. The first published graph of an action current (action potential)[8]

In 1875 Richard Caton recorded electrical activity from the brain surface of rabbits and

monkeys but his work was only recognized a few decades later in Hans Berger’s study of the

“Elektrenkephalogramm” (electroencephalogram aka EEG) of the human brain. Berger was the

first to describe brain waves of healthy and diseased brains as well as during different

physiological states in 1929. Still limited by the physical size of the recording instrument all

functional recordings, so far, are extracellular and on a rather macroscopic scale, e.g.; nerve

bundles, muscle fibers, and brain surface. Intracellular recordings were not achieved until the

1940s [9] and were done predominantly in in-vitro settings. The first in-vivo intracellular single-

unit recording was achieved by Sir John Eccles, his daughter and student Rosamond Eccles, and

Anders Lundberg when they recorded spikes (intra-cellular action potentials) from motoneurons

in the anesthetized cat in 1957. Their seminal work was enabled only by the arrival of smaller,

sharper electrodes that could penetrate the nervous tissue in vivo, amplifiers with high input

impedance, better current delivery systems, and relatively high-speed display systems to view

and record these minuscule physiological biopotentials [10], [11].

Further advances in microelectrode fabrication and packaging laid the groundwork for

the first multi-electrode bundle (as opposed to a single electrode) used by Marg and Adams to

record activity from many neural cells (multi-unit activity) at the same time from a patient during

brain surgery in 1967 [12]. This milestone can be considered the finale of the solo dance

Page 21: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

21

challenge of neural electrophysiology and the opening of the modern era of fusion dance - the era

of cellular functional imaging.

Cellular functional imaging is the fusion between microscopic imaging and functional

recording. It is the art of imaging neurons in action, as they fire and as they communicate. The

only impediment is that functional recording instrumentation was designed based on the

electrical and biophysical properties of neurons. Hence, the key was to tie the neural activity to

an imageable physical quantity, like light. And to be more specific, light emanating from the

molecules of a dye imaged by a fluorescent microscope. The first fluorescent dye, fluorescein,

was developed by Adolf von Bayer in 1871 but needed to wait for the introduction of the first

fluorescent microscope by Oskar Heimstadt to be utilized on a microscopic level in the

investigation of the autofluorescence of cells and tissue under ultra-violet (UV) light in 1914.

Going beyond the study of UV based autofluorescence seemed farfetched to Heimstadt as he

ended his paper stating: “If and to what degree fluorescence microscopy will widen the

possibilities of microscopic imaging only the future will show” [13], [14]. Nevertheless, the

major challenges of delivering external light, suppressing autofluorescence, eliminating

reflections, and capturing only the desired wavelength were overcome by a team of scientists

including Max Haitinger, who developed fluorochromes to enable secondary fluorescence,

Phillip Ellinger, who built the first prototype of an epifluorescence microscope in 1929, and

Johan Ploem who built dichromatic mirrors and beam-splitters in 1967. After that, fluorescent

imaging was used in various anatomical and immunological applications. However, correlating

fluorescence with function in live cells, i.e.: fluorescence correlation spectroscopy, was first

introduced by Magde et al. [15] when the orange fluorescence of DNA binding with ethidium

bromide was imaged. In the same vein arises the concept of fluorescently tagging calcium

Page 22: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

22

molecules in actively firing neurons. Since it was known by then that neural activity relied

causally and proportionally on intracellular calcium concentration [16]–[27], Roger Tsien set

forth on developing synthetic fluorescent calcium (Ca2+) indicators that would reliably report the

underlying neural dynamics [28]–[31].

Over the past few decades and till this very day, microbiological advances in the genetic

engineering of Ca2+ indicators and fluorescence microscopy have constantly pushed the limits of

optics to enhance the quality of the functional image (spatial resolution) and to convey more

information about the captured neural dynamics (temporal resolution). In stark contrast to

Heimstadt’s doubts and combining the best of the two worlds, this fusion dance granted

neurophysiology an exquisite technique, as neural activity can be precisely pinned, in time and

space, to neural anatomy. It is needless to state how fluorescence microscopy has revolutionized

our understanding of various neurological processes, disorders, and diseases and will continue to

shed more light on the mysteries of the human brain [32]–[45].

The long-standing synergy between fluorescent microscopy and neurophysiology laid the

foundation for more recent advances in microscopic technology, fabrication of miniaturized

optics and the genetic engineering of fluorescent indicators. Hence, further propelling the optical

approach to the functional and structural interrogation of neural circuits. The result was the

emergence of various optical methods, which makes the selection process quite intricate and the

comprehensive awareness of the technical underpinning indispensable, to guarantee proper

interpretation of the recorded dynamics. Furthermore, the inevitable compromise between cost,

resolution, focality, targetability, flexibility in behavioral paradigms and complexity is shared by

all-optical methods of in-vivo neural interrogation. Determining the optical method of choice for

an experiment can thus be a quite arduous task.

Page 23: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

23

Among the many optical methods is Fiber Photometry. Fiber Photometry stands out for

the exceptional flexibility in awake behaving experiments, the ability to reach very deep brain

areas, the ease of operation and the low cost. The Fiber Photometry is a versatile, cost-effective

optical method that records bulk fluorescent Ca2+ dynamics from the brain and reports it as an

aggregate one-dimensional signal, making its interpretation an unsolved challenge. This

dissertation, therefore, pivots on systematically characterizing the Fiber Photometry system to

appraise its underlying mechanisms of action and to offer new means of interpreting the

information acquired during in-vivo interrogation of neural circuit dynamics.

Page 24: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

24

Dissertation Overview

The focus of this dissertation is the systematic characterization of the single-photon Fiber

Photometry (FP), a fluorescence-based recording device of neural activity. The dissertation is

designed to be a self-contained document for both, the general as well as the proficient reader.

Chapter 2 presents background on the fundamental principles of optics as they apply to

fluorescence-based imaging devices. After a brief review of the properties of light and the basics

of single-photon light propagation in biological tissue, a framework for optically interrogating

neurons using fluorescent actuators and sensors will be provided. The experienced reader may

choose to skip to the end of Chapter 2. At the end of Chapter 2, the central problem will be

defined, and the specific research aims of this dissertation will be listed. These aims will then be

addressed one by one in Chapters 3 through 5. The experiments carried out in these chapters are

separate but complement one another to achieve the overall goal. Each chapter will reiterate the

aim of the experiment, provide some experiment-specific background, describe the methods used

and finally present the results and discuss their relevance. Chapter 6 will recapitulate the

findings, put them in perspective to the current state of the field, and discuss the potential

limitations of the experimental approach and provide an outlook on possible future directions.

Additional information is provided in the appendices.

Page 25: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

25

CHAPTER 2

BACKGROUND AND MOTIVATION

Let there be Light

Light, the visible spectrum of electromagnetic radiation, is one of the most intriguing

elements of life. It has fascinated the early human mind and became a chief consideration to

almost all ancient mythologies. Apollo, Eos, Baldr, Horus, Ao, and many other light deities were

conceived to impersonate the unexplained yet mighty phenomena of light. Scientific endeavors

to examine and investigate the nature of light are as old as mankind and have just come to

fruition in the early 20th century. Before that, the postulation that light could be used to record

and control the activity of individual as well as ensembles of nerve cells inside the living brain

would have been immediately convicted for heresy.

Fortunately, the physical and mathematical formulation for light is now well established

and permits the design of devices that precisely control light delivery, manipulation, and

collection. Moreover, recent advances in light-based technology, do indeed allow us to read

(record) and write (control or modulate) the activity of brain cells. The upcoming sections will

expound the technical details behind this de·Light·ful dialogue (read/write) with the brain.

… and there was Light

For thousands of years, mankind witnessed the bedazzling effects of light as it shimmers

in reddish hues on the water at sunset, as it softly spreads a radiant morning glow and as it

manifests in an iridescent rainbow during a rainy day. Puzzled by its mysterious and impalpable

nature early philosophers and scientists asked two very natural questions: “Where does light

come from?” and “What is light?”

Page 26: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

26

The Origin of Light

In answer to the first question came the following philosophy: “It is bright, and we can

see things when our eyes are open, and it is dark, and we can’t see when we close them.

Therefore, light must be emanating from our eyes”. The proper term for this theory is the

‘extramission theory’, which was founded by Plato (428 BC–328 BC) and advocated for by his

followers for almost a thousand years. Like Plato, Euclid (330-275 B.C.), Hero (10–70) believed

that light propagates from our eyes as rays that travel to objects in our field of view. Upon

striking an object, those rays interact with it, allowing the eye (or rather the brain) to perceive

size, shape, and color [46].

Over the span of another thousand years, some scientists argued against the extramission

theory, yet no one was able to provide evidence except for Al-Hazen. Al-Hazen laid down the

foundation to our current understanding of optics, light, and vision with his book, Kitab Al-

Manazir (Alhazen's Book of Optics 1027), that constitutes a cornerstone in the field of optics.

Al-Hazen placed two lanterns at different heights outside of a dark room and made a small hole

in the wall. He then stepped into the room and found that when both lanterns are on, their light

passed through the hole and created two bright spots on the opposite wall of the room. When he

turned one of the lanterns off and stepped back into the darkroom, the bright spot corresponding

to that lantern disappeared. Thereby, he provided the experimental evidence to prove that light

emanates from light sources like lanterns and candles and not from the eye. Al-Hazen’s simple

yet elegant experiment marked the fall of the long-standing extramission theory and set the stage

for the next question: “What is light”?

The Light Duality

“What IS light?” - A simple question if light was palpable or would take on one of the

three physical forms of matter: gas, liquid or solid. But, it doesn’t. Light has a dual nature, called

Page 27: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

27

the wave-particle or wave-corpuscular duality. In other words, light can manifest as a wave

and/or like an energized particle depending on the experimental setup1. It left scientists baffled

until the 1600s when two European scientists tried to explain refraction. Johannes Snell and

René Descartes took different experimental approaches to quantitively describe the phenomenon

of refraction, which is the change in direction of an incident light ray when it passes from one

medium to another, and eventually arrived at the same mathematical formulation, which is now

known as the Snell-Descartes law of refraction [47].

In the 1700s, Francesco Maria Grimaldi and Christiaan Huygens investigated the

diffraction phenomena and could only explain it by attributing a wave-like nature to the light

beam, and drew a one-to-one analogy between sound waves and light waves [48]. Sir Isaac

Newton was Huygens’ contemporary rival and adamantly opposed the wave-theory in favor of

his proposed corpuscular theory, which postulates that light is a beam of minuscule particles that

move at high speed through the ether and hence comply to the laws of gravity and inertia. Based

on Newton’s corpuscular theory, reflection occurs as particles bounce off the surface of a

medium and refraction is the result of the medium’s higher density that creates a stronger

gravitational pull on the particles. However, the corpuscular theory did not lend itself directly to

Newton’s observations on refraction. In one of the pioneering experiments described in his book

“Opticks: or a treatise of the reflections, refractions, inflections, and colours of light” [49],

Newton decomposed a sunbeam into different colors using two prisms and noticed that “Lights

which differ in colour, differ also in degrees of refrangibility” [Part I, page 13]. While his

discovery could give credence to the wave-theory, Newton insisted on the corpuscular nature of

light and argued that the different colors are different types of particles (Figure 2-1).

1 I wonder if the ancient deities could have done that!

Page 28: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

28

Figure 2-1. Newton’s observation on refraction. Excerpt from The First Book of Opticks, Part I,

Page 30, by Sir Isaac Newton (1704).

The wave-particle dispute continued throughout the 17th century, with some experiments

showing the clear wave-like behavior of light, while others could only be rationalized by a

stream of particles traveling in a straight line. Since further delving into the history of the wave-

particle debate would be beyond the scope of this chapter, I will rudely ignore the efforts of

Young, Malus, Fresnel, Poisson, and Arago, and fast forward to the 19th century, the time of

James Clerk Maxwell. Maxwell is credited for the classical theory of electromagnetism and his

mathematical formulation ingeniously integrated the results of several other physicists

(Coulomb, Volta, Ampère, Faraday, and others) to demonstrate that electricity, magnetism, and

light are different manifestations of the same phenomena, namely the electromagnetic wave.

Maxwell postulated that since electromagnetic waves, akin to light, propagated through the ether

by undulations and undergo refraction and interference when passing through different media or

small gratings, then light must be an electromagnetic wave. He further asserted his proposition

by proving that light and electromagnetic waves travel through matter at the same speed [50].

Page 29: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

29

Maxwell revealed the electromagnetic nature of light and by that also ingrained the wave-theory,

which ruled the roost until the rise of quantum mechanics in the early 1900s. Max Planck’s

quantum theory on the quantization of energy was the tacit precursor to the recovery of the

corpuscular theory.

Albert Einstein appreciated Planck’s idea of energy packets and used it to explain the

photoelectric effect, further argued that “According to the assumption considered here, when a

light ray starting from a point is propagated, the energy is not continuously distributed over an

ever-increasing volume, but it consists of a finite number of energy quanta, localised in space,

which move without being divided and which can be absorbed or emitted only as a whole” [51].

Einstein’s revered stature at the time gave his argument immense credence. However, the

wave specific phenomena like interference were still observed and were only explained with the

wave theory. The conundrum was solved in 1927, when Clinton Davisson and Lester Germer

finally proved the dual nature of light by hitting the surface of a nickel target with a beam of

accelerated electrons (a beam of particles) and observed a diffractive pattern (constructive and

destructive interference) on the detector, reminiscent of wave-behavior [52]. The experiment was

repeated by others to confirm that the Davisson-Germer experiment was not a fluke. The dispute

was once and for all resolved: Light is an electromagnetic wave that consists of minuscule

particles, called photons.

The previous section succinctly chronicled centuries of protracted scientific debates that

revealed the nature of light. The next section will present the common terminology that describes

the different physical qualities and quantities of light.

Basics of Illumination

The Electromagnetic (EM) spectrum is the range of all electromagnetic radiation, from

long, low energy radio waves all the way to very short, highly energetic gamma-rays. Optical

Page 30: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

30

radiation is the portion that includes ultraviolet (UV), visible and infrared (IR) radiation. Light is

the very narrow band within the optical radiation spectrum, that is visible to the human eye

depicted by the ROYGBIV in Figure 2-2. ROYGBIV is the acronym that describes the colors

that compose the visible spectrum: Red, Orange, Yellow, Green, Blue, Indigo and Violet. In

contrast to the rest of the optical radiation spectrum, only the visible portion (light) can be

detected by the photoreceptors in the eye and elicits a ‘photometric’ response in the visual

system.

Figure 2-2. Electromagnetic spectrum. Diagram of the electromagnetic spectrum illustrating the

parameters of wavelength and frequency. Depicted are also the range of visible light

as well as that of ionizing and non-ionizing radiation.

There are three approaches to measure the emission and transmission of EM radiation.

The first is Radiometry, which is the most common and general measurement EM radiation.

Radiometry applies to the entire EM spectrum and reports radiative quantities using the SI-

derived units. For instance, the energy of an EM wave is reported in joules [J] and the power in

watts [W=1J/s]. However, those radiative quantities, as well as other quantities, also have a

quantum nature that can be expressed in terms of photons and photon flux to provide a physically

Page 31: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

31

meaningful interpretation of the result in certain experiments. Actinometry, the second approach

to the measurement of EM radiation, emphasizes the quantum effect of light and is, therefore,

more suitable for photobiological, photophysical and photochemical applications that rely on the

particle-behavior of light. In other words, actinometry is the quantized version of radiometry.

The third measurement approach is Photometry. Photometry differs from actinometry and

radiometry in that it uses the human eye as the optical detector. The photoreceptors of the human

eye are sensitive to some colors more than others and our visual perception depends on the eye’s

adaptation to darkness or dimness. Photometry, therefore, measures only the visible portion of

the EM spectrum based on its ability to elicit a specific visual response. For that, it uses the CIE2

standard observer conversion function to convert radiant energy to luminous energy reported by

with units of lumens or candela [53]–[55].

The central concern of this dissertation is the characterization of the Fiber Photometry, an

optical device that uses visible light to elicit a photobiological effect and measures the response

with a silicon-based photodetector. Since the response of the device is captured using a

photodetector and not human visual perception, radiometry would provide a more suitable

measurement system. This section will, therefore, provide a brief overview of the most important

parameters of light and some of its radiometric quantities with no mention of their photometric

counterparts to avoid confusion3.

Parameters of Light

Speed of light: In a vacuum, the speed of an EM wave, and hence light, is measured to be

C = 3x108 m/s.

2 Commission Internationale de l'Éclairage: International comission on illumination

3 It should be noted here that the term photometry in the name of the device may be misleading, as it is not

associated with human visual perception

Page 32: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

32

Wavelength: The wavelength λ is the distance between two successive crests (or hills) of

the EM wave. It’s measured in units of distance. The wavelength is inversely proportional to the

frequency of the EM wave (Figure 2-2 and Figure 2-3). The visible spectrum extends from

450nm to 650nm. Symbol: λ. Unit: nanometers [nm]

Frequency: The frequency is the number of full cycles (1 cycle = 1 wavelength) the EM

wave completes in one second. It’s measured in units of s-1 or Hz. The frequency is related to the

wavelength through the speed of light υ. The energy of the EM wave is directly proportional to

its frequency (Figure 2-2 and Figure 2-3). Symbol: f. Unit: Hertz or time-1 [Hz or s-1]

Photon: The photon is a single light particle that carries the Energy Ep [J]

𝐸𝑝 = ℎ 𝑓 (2-1)

where h is Planck’s constant = 6.63x10-34 [Js] and f is the frequency of the light in [Hz].

As such, the energy E of a light beam can be only an integer multiple of the energy Ep of a single

photon.

Figure 2-3. Electromagnetic wave. Diagrammatic illustration of frequency f, wavelength λ and

Energy E of an EM wave.

Radiometric Quantities

Radiant energy: The radiant energy is the amount of EM energy emitted, transmitted or

absorbed by a medium or object and is measured in units of Joules [J] or Watt-second. Symbol:

E. Unit: Joules [J]

Page 33: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

33

Radiant power: The radiant power is the amount of radiant energy flowing per unit time.

It is measured in units of Watts [W]. Radiant power is also called radiant flux. Symbol: P. Unit:

Watts [W]

Radiant intensity: The radiant intensity is the radiant power per unit solid angle emitted

by a source. Symbol: I. Unit: Watt/steradian [W/sr]

Irradiance and radiant emittance: The irradiance and emittance are both radiant power per

unit area. However, irradiance is the power per unit area received by an object or medium and

while emittance is the power per unit area emitted by a source. Symbol: R. Unit: Watt/meter2

[W/m2]

Basics of Optics

Light can travel undisturbed for miles, if the medium does not change, with some

attenuation possibly caused by the medium’s absorptivity. Once light passes through an optical

interface, optical phenomena, like reflection and refraction, are observed. The following section

summarizes the basic behavior of light at boundaries as it is relevant for the propagation of light

in the FP system and in neural tissue.

Refractive index: Light travels at slower speeds in media of different densities. The ratio

between the speed of light in vacuum, 𝐶, and the speed of light in a material, υ, is called the

refractive index, n. The refractive index n is dimensionless and material specific. The refractive

index of the rodent brain is 1.36.

Page 34: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

34

𝑛 = 𝐶

υ (2-2)

𝐸𝑝 = hυ

λ, since 𝑓 =

υ

λ (2-3)

Reflection: Reflection happens when light bounces off a surface. If the surface is polished

(like a mirror) the reflection is called specular reflection. In specular reflection, the light beam is

reflected at an angle equal to its angle of incidence only in the opposite direction. If the surface is

rough (or matte) it is called diffuse reflection and the light is reflected in many directions.

Diffuse reflection is also known as Lambertian scattering (Figure 2-4).

Figure 2-4. Illustration of Specular and diffuse reflection.

Refraction: A light beam bends (or refracts) when it passes through an interface between

media of dissimilar refractive indices. Refraction is how the light beam manifests the change in

velocity and wavelength as it passes from one medium to the other. The frequency of the light

beam, however, is unchanged. It’s imperative to point out here that since the refractive index of a

medium is wavelength dependent, refraction is wavelength-dependent as well. This means

different colors will bend with different angles. The amount of bending in the light beam

depends on the angle of incidence and the refractive indices of the two media. Refraction is fully

described by Snell-Descartes law of refraction:

𝑛1𝑠𝑖𝑛𝜃1 = 𝑛2𝑠𝑖𝑛𝜃2 (2-4)

Page 35: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

35

Where n1 and n2 are the refractive indices of medium 1 and 2, respectively. θ1 is the

incident angle and θ2 is the refracted angle with respect to the normal to the surface (Figure 2-5).

Diffraction: The diffraction phenomena is the bending of a light beam after it passes

through a narrow aperture or slit. Diffraction is strongly wavelength-dependent and follows the

following formula:

𝜃𝑑 = 𝜆

𝑑 (2-5)

Where d is the width of the aperture and λ is the wavelength of the light (Figure 2-6B).

Cone of acceptance: The cone of acceptance defines the angular range from which light

can pass through an aperture and be collected by a detector (Figure 2-6C). The cone of

acceptance is measured given by the solid angle Ω of the aperture in [stereo radians or

steradians, sr]. In simple terms, the solid angle Ω is the 3-Dcounterpart of a 2-D angle [radians,

r] that spans a cone rather than a sector. The only catch is that it needs a reference sphere. In

more technical terms, 1 sr solid angle is the angle, which, while its vertex is at the center of a

sphere with radius r, would cut out a spherical surface area equal to the square of the radius of

the circle (Figure 2-6A-B). This means that 4π sr will cover a full sphere. The importance of the

cone of acceptance will become clear in the next section as it relates to the properties of the

optical fiber used in the FP system. Symbol: Ω. Unit: steradians [sr]

Page 36: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

36

A)

B)

Figure 2-5. Refraction and reflection. Diagram illustrating the concepts of A) Refraction of an

incident beam on the surface and B) Diffraction of an incident beam on a slit.

A)

B)

C)

Figure 2-6. Cone of acceptance. A) 2D illustration of an angle. B) Description of the solid angle

(3D). C) Illustration of the cone of acceptance.

Page 37: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

37

Propagation of Light

The FP system is an optical device that manipulates light with the final goal of collecting

Ca2+ fluorescence from neural tissue. The system relies on an optic fiber that delivers and

collects light to and from the brain. This section will describe basic principles that govern the

propagation of light in the optic fiber as well as in the neural tissue.

Propagation in an Optic Fiber

An optic fiber is a light guide made of glass or plastic and transmits light from one end to

the other by total internal reflection (TIR). An optic fiber is illustrated in Figure 2-7. Commonly

optic fibers are defined by the following set of parameters:

Diameter: Diameter d of the core of the optic fiber. The diameter of the cladding, the

protective layer around the fiber is also given sometimes.

Angle of acceptance: The angle of acceptance, θacc, is the 2D equivalent of the cone of

acceptance described previously (Figure 2-4C). It defines the width of the cone that can collect

light and is defined by the refractive indices of the core and the cladding.

sin(𝜃𝑎𝑐𝑐) = 1

𝑛0 √𝑛1

2 − 𝑛22 (2-6)

Numerical aperture: The NA is a unitless number that is indicative of the size of the cone

of acceptance of the fiber. The NA is defined by the angle of acceptance θacc and hence also

depends on the refractive indices of the core and the cladding:

𝑁𝐴 = 𝑛0 sin(𝜃𝑎𝑐𝑐) = √𝑛12 − 𝑛2

2 (2-7)

Page 38: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

38

Total internal reflection: Light propagation through an optic fiber is governed by total

internal reflection (TIR), which relies on the basic optical phenomena of refraction and reflection

described in the previous section. The added constraint is that the incident light beam must arrive

at an angle smaller than θacc to be transmitted to the other end of the fiber via successive total

reflections at the core-cladding interface, i.e.: TIR. A beam of light with an incident angle that is

larger than θacc will be lost due to refraction at the core-cladding interface. Figure 2-7 shows the

concept of TIR as well as the acceptance angle of the fiber.

Figure 2-7. The properties of an optical fiber. Diagram showing the diameter d and the

acceptance angle θacc of an optical fiber. The refractive indices of the medium, core,

and cladding are n0, n1, and n2 respectively. θ1 is the angle of the incident light beam

and θ2 is the angle of refraction at the medium-core interface. The blue dotted line

shows a beam of light propagating through the fiber via TIR at the core-cladding

interface. The orange dotted line shows a beam of light that arrives outside the cone

of acceptance and is lost due to refraction (not transmitted by the fiber).

Propagation in Neural Tissue

For all-optical applications, it is important to differentiate between two kinds of media:

homogenous and inhomogeneous. As the name implies, a homogenous medium has optical

properties that are constant throughout the volume of the medium and is described with optical

parameters that take on constant values. On the other hand, an inhomogeneous (turbid) medium

exhibits optical properties that can change throughout the volume. Neural tissue is one of the

Page 39: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

39

most turbid media due to the differential cellular structures, extracellular matrix density,

myelination intensity, and protein and lipid distribution that changes throughout the entirety of

the brain. The propagation of light in the brain is, therefore, less straightforward and requires the

assumption of homogeneity over infinitesimal volumes where the basic geometric descriptions

can hold. Integrating over all infinitesimal volumes then gives rise to an overall description of

the behavior of light in the brain as dictated by its optical properties.

The previous section described the different phenomena that occur when a beam of

photons (light) traverses a boundary, where the boundary can be any optical interface between

two media of dissimilar refractive indices. Neural tissue4 is a very turbid medium with various

types and sizes of objects that have different refractive indices, like extracellular matrix, blood

vessels, cell membranes, and organelles and other biological structures. Consequently, each

structure that makes up the tissue is an optical interface (boundary), through which the beam of

photons must propagate. At every boundary-crossing, the beam of photons will undergo

reflection, refraction and/or diffraction, as mentioned earlier. However, particle behavior will

now play an important role and has to be considered. As the beam of photons propagates through

the tissue, it will be subject to a lot of collisions and friction against the constituent biological

structures, resulting in either loss of energy, transfer of energy and/or change in direction. The

amount by which a beam of photon loses energy or changes direction as it travels through any

biological tissue is defined by the optical properties of that tissue. In the following, the main

optical properties of neural tissue will be explained.

4 All biological tissue presents a highly scattering medium, but the focus will be kept on neural tissue here.

Page 40: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

40

Transmittance: Transmittance or transmission (T) is a measure of how much energy

remains in the beam of photons (or light beam) after it traverses a medium. It is the ratio between

the initial and final light intensity and can be given in % if multiplied by 100.

𝑇 = 𝐼

𝐼0 (2-8)

Absorption: The energy of a propagating light beam can be lost to the medium as it is

absorbed by the molecules or structures of that medium. The medium’s ability to steal the light

beam’s energy is called absorption coefficient µa and has units of mm-1. Absorption is related to

transmission by Equation 2-7. More on the different types of absorption (radiative and non-

radiative) in later sections.

𝐴 = 𝐼0

𝐼= 𝑙𝑜𝑔10 (

1

𝑇) = 2 − 𝑙𝑜𝑔10(%𝑇) (2-9)

Scattering: Turbid media, like neural tissue, exhibit variations in refractive index as a

function of space, i.e.: n (x, y, z) as opposed to a fixed value for homogenous media.

Consequently, the light beam will be deflected every time it encounters a change in refractive

index. This interaction is called scattering and is measured in terms of the scattering coefficient

µs [mm-1] or its inverse, the mean free path ls [mm]. As such the mean free path ls is the distance

a beam of photons can travel, on average, before it is deflected due to optical non-uniformities

and interfaces in the medium. Therefore, all the microscopic structures that constitute the tissue

are called scattering objects.

Attenuation: Attenuation is the compound effect of scattering and absorption in a

medium. It is characterized by the attenuation coefficient µa [mm-1]. Together, scattering and

absorption exponentially attenuate a light beam according to Beer-Lambert’s law:

Page 41: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

41

𝐼(𝑧)

𝐼0= 𝑒−𝜇𝑧 = 𝑒−(𝜇𝑎+𝜇𝑠)𝑧 (2-10)

Where z is the penetration depth [mm] and I is the intensity of the light beam.

Anisotropy: The Anisotropy of a medium describes its preference for the direction of

scattering. Small objects, like cells, that are of a size comparable to that of the wavelength of the

light tend to scatter in a forward direction. The anisotropy factor 𝑔, therefore, is the average of

the cosine of the scattering angle 𝜃 and takes values from 0 to 1, with 1 being strictly forward

while 0 indicates isotropic scattering.

𝑔 = 𝐸𝑐𝑜𝑠𝜃 (2-11)

Fundamentals of Fluorescence

As mentioned earlier, the energy of a beam of photons can be absorbed by the medium or

tissue it is propagating through. The absorbed energy is transferred to the molecules of the

medium causing them to transition from a low energy level to a higher energy level. After a very

short period of time, the molecule returns to its original energy level and emits the absorbed

energy via radiative and non-radiative processes. The latter process is mainly thermal, i.e.: the

energy is released in the form of heat to the surrounding medium. The former process involves

the emission of another photon and is, therefore, called radiative, as it results in the emission of

one or more photons (the basic unit of an EM wave). While both processes, radiative and non-

radiative, occur upon the absorption of a photon (an energy packet), this section will focus

mainly on radiative emission. Specifically, fluorescent radiative emission.

Radiative emission is the process of emitting a photon (or photons) when transitioning

from a higher energy state (electronically excited) to a lower energy state (ground) and can be

categorized based on how the molecule was brought to the excited state. If a chemical reaction

Page 42: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

42

was used it’s called chemiluminescence. If extreme heat was used it’s called incandescence. And

if UV or visible light is used it’s called photoluminescence. Photoluminescence includes

fluorescence and phosphorescence, which differ in the configuration of the electronic state and

the emission pathway. These processes are best described using a Jablonski diagram which

illustrates the different vibrational energy levels and the possible excitation and emission

pathways in a hypothetical molecule. Figure 2-8 shows a simplified Jablonski diagram that

would serve the scope of this dissertation. Due to the specific configuration of the electronic

states as well as the vibrational and rotational energy levels that allow for fluorescence, some

molecules are more predisposed for fluorescence than others. This kind of molecules is called

fluorophores, fluorescent probes, fluorochromes or fluorescent dyes.

In the Jablonski diagram shown in Figure 2-8, the different energy levels of a molecule

are marked with the letter S. The least energetic state is the singlet ground state S0 while S1 and

S2 are singlet excited states. Within each energy level, there are multiple vibrational energy

levels, which are represented by the thin horizontal lines. When a photon strikes a molecule, the

molecule can react in one of two ways: 1- absorb the entire energy packet, if it is sufficient to

reach an excited state, 2- not absorb the packet, if it does not fit any possible transition. In other

words, the energy packet must be absorbed in its entirety or not at all. Based on Planck’s

quantum theory, no partial absorption of a photon is possible, and the absorption of a photon is

hence an ‘all or none’ process. In some cases, however, the absorption of an energy packet may

simply bump the molecule to a higher vibrational energy level within the same state because a

transition to an excited state requires a larger energy packet. Or the absorbed energy may be

more than what is needed to transition to an excited state, in which case the remainder is

expended as vibrational and/or rotational relaxation (orange wavy lines in Figure 2-8).

Page 43: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

43

Figure 2-8. Jablonski diagram. Showcased is the process of fluorescence through electronic

transitions between the ground state and excited states.

A more efficient way to describe this process is by considering an example. Assuming a

molecule absorbs a photon packet that is sufficient for a transition to an excited state, the

Jablonski diagram in Figure 2-8 depicts two (out of many) possible excitation (absorption)

transitions. The first blue arrow shows a transition from the lowest vibrational level of the

ground state S0 to the third vibrational level of the excited state S2 (S0(0) → S2(3)). The second

blue arrow presents a different excitation path from (S0(1) → S1(5)). In either case, the

absorption process is immediate and takes only ~1x10-15s (1 femtosecond). Also, regardless of

the final excited state and vibrational level, the molecule will release some energy in form of

vibrational relaxation or internal conversion to heat, without the emission of a photon, until it

reaches the first vibrational level of the first excited state S1(0). This non-radiative emission

usually takes ~1x10-12s (1 picosecond). Only a transition from the first vibrational level of the

Page 44: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

44

first excited state (S1(0)) to any of the vibrational levels in the ground state will be accompanied

with the emission of a photon and results in fluorescence, which is relatively slow and takes

~1x10-9s (1 nanosecond). Suffice it to say the fluorescence pathway competes with other possible

relaxation pathways like intersystem crossing, quenching, and non-radiative relaxation (not

shown). The absorption and emission of photons is a cyclic process and fluorophores can repeat

this cycle thousands of times. However, excessive exposure to light can harm the fluorophore

and prevent it from fluorescing. This process is called photobleaching.

There are two important things to note in the fluorescence relaxation process: First, the

possibility of transitioning from any vibrational level within one state to any vibrational level

within another state means that the quantal size of the absorbed (S0(y)→Sx(y)) or emitted

(S1(0)→ S0(y)) energy packet can vary. Subsequently, excitation and emission are described by

spectra of variable width, where the width of the spectra corresponds to the range of possible

energy packets. As stated in Equation 2-3, the energy of a photon is a function of its frequency

and hence its wavelength. Excitation and emission spectra can, therefore, be regarded as the

probability density function for a photon with a specific wavelength to be absorbed and to trigger

the fluorescent emission of another photon. Second, the energy of the emitted photon is always

less than the energy of the absorbed photon as some of the energy is lost in internal conversion or

vibrational relaxation. Hence the emitted photon has a wavelength (red-shifted) that is longer

than the absorbed one (see Figure 2-2). This phenomenon is called Stoke’s shift, in honor of Sir

George G. Stokes, who discovered it and is the reason why the emission spectrum is always

shifted to longer wavelengths (lower energy) with respect to the excitation spectrum (Figure 2-

8). Excitation and emission spectra are usually expressed as the central wavelength λ [nm] at

which excitation/emission is most likely, λex & λem, and a number indicative of the width of the

Page 45: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

45

spectrum on either side of λex and λem. The spectra illustrated in Figure 2-9 would be represented

as λex = 475/45 and λem = 530/45nm.

Figure 2-9. Example excitation and emission spectrum of a green fluorescent fluorophore. The

spectrum was generated using the publicly available Fluorescence SpectraViewer by

ThermoFisher Scientific Inc.

As discussed earlier, fluorescence-based microscopy originated as a spin-off from UV-

instrumentation used to investigate autofluorescence in small organisms. It was not until the late

1960’s that the first successful fluorescent microscope was reported. Joan S. Ploem, built the first

multiwavelength fluorescent microscope with vertical epi-illumination using 4 dichroic mirrors

mounted on a sliding tray to change the excitation wavelength between UV, violet, blue and

green. His design was based on dichroic mirrors that allowed the excitation beam to reflected

towards the specimen and the emission beam to be transmitted to the observer (Figure 2-10).

Needless to say, the advent of fluorescent-based microscopy revolutionized the field of cell

biology, as the power of live imaging of cellular and subcellular structures was combined with

the highly specific fluorescent labeling of molecular elements achieved with synthetic as well as

genetically encoded fluorophores.

Page 46: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

46

A)

B)

C)

Figure 2-10. Different configurations of fluorescent microscopy. A) Trans-illumination,

Observer and light source are on opposing sides of the objective. B) Epifluorescence,

observer and light source are on the same side of the objective. C) Epifluorescence

with dichroic mirror.

Optical Interrogation of Neural Circuity

In an interrogation, the superior party conducts an interview with a suspect to elicit useful

information. Drawing the analogy, the suspect is the neural circuitry in the brain and the scientist

is the interrogator. And as the name implies, an ‘optical’ interrogation will use fluorescence to

elicit useful information. Now the interview part is when the scientist tells the neurons what to

do (stimulation or write) and then listens to what the same or other neurons have to say (record

or read). Figure 2-11 illustrates the concept of all-optical interrogation of neural circuits.

Optical interrogation of neural circuitry is an extremely powerful concept that emerged

with the turn of the century and was catapulted with the extraordinary advances in fluorescent

reporter and actuator probes. The basic idea is to probe the intertwined neural dynamics of a

brain region by switching cell-type-specific neurons on or off and observe the effect of the

perturbation by recording or imaging the resulting neural dynamics. This can be done in

technically any brain region that may play a relevant role in a cognitive modality like learning,

Page 47: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

47

skill acquisition, habit formation or in neurological diseases, like Parkinson’s, depression,

epileptic seizures … etc. As such, a more time- and space-specific the optical interrogation will

provide more insight about the underpinning of neural disorders and potentially yield to

scientific findings that could pave the way to viable bed-side solutions. This is the aspiration of

the current tour de force in the development and technical refinement of fluorescent probes as we

as optical device technology [56]–[59].

Fluorescent probes can be broadly categorized, based on the mode of action, into

actuators and sensors. The following sections will provide a quick overview and list the most

commonly used fluorescent probes.

Figure 2-11. All-optical interrogation of neural circuit dynamics. Adapted from [60]

Optical Actuators

An optical actuator probe is a molecule that elicits a secondary effect once illuminated

with a certain wavelength. The most common and naturally occurring is rhodopsin, a light-gated

protein that serves as a sensory photoreceptor in algae and controls phototaxis (movement in

response to light). Other light-sensitive proteins, derived from rhodopsin, have been engineered

to control ion flux, cell excitability, and other cellular processes and constitute what is now

called the optogenetic toolkit. Optogenetics is the field concerned with the design of light-

Page 48: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

48

sensitive proteins that express in the membrane of neurons to control the cell’s excitability in

response to the light delivered to the target area via an optical fiber. The two major subtypes are

inhibitory and excitatory optogenetic actuators, where the former pumps ions to hyperpolarize

the cell and the latter pumps ions to depolarize the cell and cause it to fire. As such, optogenetic

actuator probes are tools that allow us to ‘write’ to neurons, telling them when to fire and when

not to.

Optogenetic actuators can be delivered to any target region (deep or superficial) in the

brain via a viral construct that carries the genetic sequence of the light-sensitive protein. Once

inside the cell, it gets transcribed and expresses in the cell’s membrane. This viral construct can

be engineered to target a very specific subtype of neurons and allows for cell-type-specific

targeting. A long array of optogenetic actuators has emerged over the past decade with a spectral

variety that can serve many applications [56]. Figure 2-12 summarizes the most commonly used

optogenetic actuators.

Figure 2-12. Commonly used optogenetic actuators and their excitation spectra. Adapted from

[61]

Page 49: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

49

As illustrated in Figure 2-12, optogenetic actuators require light of a certain wavelength

to operate. Not illustrated, however, is the fact that the light also needs to be of a minimum

strength (>1mW/mm2) [57]. The previous discussion about light propagation in turbid media

described the strongly scattering and attenuating properties of neural tissue that result in an

exponential loss in the magnitude of the traversing light beam as a function of depth. Since

optogenetic actuators require a minimum amount of irradiance (Rmin) then it is pivotal to estimate

the required initial irradiance that would compensate for the attenuation occurring in the tissue

and still be sufficient (above Rmin) to activate the optogenetic actuator at the target region.

Several studies sought to answer this question [57], [62], [63] by using optical fibers to

illuminate brain slices of varying thickness with light of different wavelengths and measure the

amount of light transmitted on the other side of the slice. They concluded that a light beam loses

more than 50% of its initial strength within the first few tens of microns after leaving the

illuminating fiber. This is a very significant conclusion given the fact that a particular amount of

irradiance is necessary to turn on the optogenetic actuator and it means that one needs to start

with a relatively high initial irradiance to make up for the expected loss.

Optical Sensors

While optical actuators initiate a process in response to light, optical sensors report an

ongoing process in response to light. In other words, optical sensors are probes that when

illuminated with light will provide a ‘readout’ on the current state of a neuron or a group of

neurons. That readout is in the form of light as well, an optical readout. Reminiscent of

fluorescent probes, optical sensors rely exclusively on fluorescence and differ only in their

coupling mechanism.

Fluorescent sensors can be coupled to 1- synaptic vesicles to report exocytosis and

endocytosis events occurring during synaptic transmissions, like FM dyes; 2- pH and indicate

Page 50: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

50

changes in pH of synaptic vesicles, like synapto-pHluorins; 3- the membrane of neurons and

report changes in membrane potential, like fluorescent voltage-sensitive dyes (VSDs); and 4-

intracellular calcium to report changes in intracellular calcium concentration, like fluorescent

calcium-indicator dyes. The last two optical sensors have seen immense progress during the past

decade including the ability to become genetically encoded instead of being washed out after a

few hours, as is the case with dyes. Fluorescent genetically encoded voltage indicators (GEVIs)

and calcium indicators (GECIs) have been widely used in conjunction with fluorescent

microscopy to study various aspect of the brain in an all-optical neurophysiological approach

(Figure 2-15) [64]–[74], [74]–[100]. Transmembrane voltage changes can happen over different

timescales and report a rich repertoire of neuronal behavior. Thus, designing a voltage indicator

that can meet all performance requirements is a challenge. Nevertheless, there is a large variety

of voltage indicators that satisfy a subset of the requirements and meet the needs of certain

scientific questions [101]–[106]. Calcium indicators, on the other hand, have grown a lot swifter

and rendered fluorescent calcium imagining the most mature modality for recording neural

activity [107]–[114]. A brief digression will detail the mechanism of calcium indicators as it is

central to the work presented in this dissertation.

Calcium (Ca2+) is a unique metal ion that serves as the second messenger for

neurotransmitter release and is intimately involved in signaling the arrival of an action potential

and the subsequent generation of other cellular processes. It possesses a uniquely large

concentration gradient across the plasma membrane with an extracellular concentration that

ranges from 1.5 to 2.0mM and intracellular level of 50 to 100nM. The result is in an outside-to-

inside chemical gradient of 15,000–40,000:1 in addition to the electrical gradient that points in

Page 51: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

51

the same direction (outside to inside at rest). The opening of a calcium channel thus exposes

calcium ions to an unusually large driving force from the outside to the inside [66], [115]–[117].

There are numerous pathways that allow the influx of Ca2+, through the various types of

Ca2+ channels. However, Ca2+ influx during the conduction of an action potential is

predominately mediated by voltage-gated Ca2+ channels that are present on the neuronal

membrane. Intracellular Ca2+ levels can rise 10-100-fold during the propagation of an action

potential within ~10ms and persists for a few tens of milliseconds.

Ca2+ indicators are fluorescent molecules that increase their fluorescent brightness when

bound to Ca2+. Hence, if they are designed to express in the cytosol, an increase in fluorescence

will be indicative of an increase in intracellular Ca2+ and hence the propagation of an action

potential. Many efforts have been devoted to improving the brightness and kinetics of GECIs

where brightness implies easier detection (higher SNR) and kinetics means faster binding to and

release from Ca2+ to avoid buffering. For instance, a single action potential lasts 3-5 ms at most.

In contrast, it took early GECIs, like GCamP3, ~130 msec to reach the peak of the transient and

a half decay time of ~600 msec, which is orders of magnitude slower than an action potential.

GECIs have evolved over the span of the years providing faster, more robust and sensitive

variants like the widely used GCamp6(f, s) and the recently proposed jGCamp7 series[64],

[110]–[113], [118].

It is essential to reiterate the bidirectional light-based nature of fluorescent Ca2+ imaging

as there are two separate but dependent light paths that play an instrumental role in the imaging

process: 1- the delivery of the excitation light to the region expressing the fluorescent calcium

indicator (GCaMP) and 2- the collection of the fluorescence emission light reported by the

GCaMP. While the basic concept of fluorescent microscopy is at the heart of calcium imaging,

Page 52: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

52

different modalities have been proposed to allow fluorescent microscopy to reach various regions

in the brain and to perform in-vivo recording of neuronal activity. The next section will briefly

compare the most common modalities of fluorescent Ca2+ imaging.

Comparison of Fluorescent Imaging Techniques

Fluorescent microscopy endowed neuroscience with the transformational feature of

imaging or recording neurons in action. The ability to record and/or image electrophysiological

markers and to link them to behavior, cognition, memory formation or disease enabled a wide

range of unique experiments and can be considered a groundbreaking advance in neuroscience.

With neural electrophysiology taking a purely optical approach that relies on light-sensitive

actuator and reporter proteins, more and more scientists were inspired to design and build “all-

optical” device technology. Common among all-optical device technology are two main

functions: 1- the delivery of light with specific wavelength and magnitude to a target region 2-

the collection of light with a specific wavelength and of minuscule magnitude from a target

region. The difference among most optical device technology, on the other hand, allows them to

be loosely categorized into three major groups based on their fluorescence excitation technique

and focality: two-photon imaging (TPI), single-photon micro-endoscopy and single-photon fiber

photometry. The next section will portray and compare these fluorescence-based optical tools.

Two-Photon Imaging

As the name implies, two-photon imaging (TPI) entails the use of two photons, as

opposed to a single photon, to elicit a fluorescent response. Given the quantized nature of

fluorescence, this means that the total energy of the two photons together must be equal to the

energy carried by the single photon. In other words, each of the two photons will carry ½ of the

energy of the single-photon and hence have a longer wavelength. This also means that the low

energy two photons must hit the fluorescent molecule simultaneously (within 1 femtosecond) to

Page 53: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

53

be absorbed. The probability of this quantum event to occur is infinitesimally low. Therefore,

two-photon imaging circumvents this by using exorbitant light powers (5 - 8 Watts) to focus very

fast pulses (femto-trains) of light on an infinitesimally small volume (point-like region) as seen

in Figure 2-13. This non-linear process increases the likelihood of fluorescent excitation if the

sum of the two photons is greater than the energy required to move the molecule from a ground

state to an excited state. The light beam is then scanned across the field of view (x and y-

direction) at very high speeds to sample the entire plane resulting in a 2D image. The recorded

2D image is at the focal plane of the objective and represents a very fine section in the neural

tissue. Changing the height of the objective then allows taking optical sections at different depths

in the z-direction.

While TPI features unprecedented resolving powers and sub-micron spatial resolution,

the major setback is the limited penetration depth, which is bounded to 300-400µm from cortex

due to the severe attenuation of longer wavelengths. Another limitation is the fact that the subject

must be head-fixed under the TPI microscope, which confines the range of behavioral paradigms

that can be investigated.

Figure 2-13. Single vs. Two-photon excitation. Cartoon comparison of single and two-photon

excitation mechanisms as well as the resulting excitation volume.

Page 54: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

54

Single-Photon Micro-endoscopy

Unlike two-photon imaging, single-photon micro-endoscopy confers single energy

quanta (single photons) to elicit fluorescent excitation in a linear process as illustrated in Figure

2-13. This results in a less focused 2D image of coplanar neurons, but the penetration depth

limitation is overcome since shorter wavelengths are used. Single-photon endoscopy relies on

gradient index (GRIN) lenses that collect fluorescent neural activity from a focal plane and can

reach deep brain structures that are not readily accessible using the TPI setup. Usually, GRIN

lenses of 0.5mm - 1.8mm are implanted into the brain and are later connected to a miniaturized

fluorescent microscope and CCD camera. The entire setup is miniaturized and secured to the

subject’s head allowing it to move freely with no head-restraint. Hence, a wider range of

behavioral paradigms can be achieved. One could claim that single-photon endoscopy trades off

resolution for flexibility in the experimental setup and in recording depth, yet on the expense of

introducing reasonable tissue damage due to the large size of the GRIN lens.

Single-Photon Fiber Photometry

Similar to single-photon endoscopy, single-photon fiber photometry (FP) relies on a

linear fluorescent excitation (single photon) scheme. However, the GRIN lens is replaced with a

sleeker optical fiber (core diameter = 50-400 µm). While this reduces the insult to the tissue, the

major consequence is the complete loss of focality. There are two repercussions for the loss of

focality: 1- the readout is a 1D signal instead of a 2D image, and 2- the 1D signal is collected

from the volume beneath the fiber and represents the aggregate fluorescent activity of neurons

residing in that volume (Figure 2-13). Favorably, since the collected information is collapsed

into a 1D analog signal there is no need for any miniaturized instrumentation on the subject’s

Page 55: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

55

head. Instead, the signal can be transmitted via an optical cable to a distant5 single-cell detector

leaving a relatively small footprint on the subject’s cranium. This is very convenient as it allows

simultaneous targeting of other brain regions in the same subject, which can be extremely

challenging with the microendoscope and rather impossible with TPI.

A)

B)

C)

Figure 2-14. Comparison of fluorescent imaging techniques. Comparison of sample data

recordings from different fluorescent imaging modalities. A) Two-photon image of

the mouse visual cortex (V1). B) microendoscope image of the rat’s prefrontal cortex

(PFC). C) fiber photometry signal from the rat’s vibrissal sensory cortex (vS1).

5 A distance within the boundaries of an average lab room.

Page 56: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

56

Data collected with the three imaging techniques is shown in Figure 2-13. The high

resolution of the TPI stands out as individual neurons as well as neuronal processes can be easily

discerned (Figure 2-14A). Despite the loss in spatial resolution, single neurons can still be seen

with the single-photon micro-endoscopy image in Figure 2-14B, yet less defined and neuronal

processes are no longer visible. Data recorded with the FP shows as a single trace as seen in

Figure 2-14C, which echoes the lumped activity of a neural population without any spatial

resolution.

Table 2-1 summarizes the main advantages and disadvantages of the three optical tools

qualitatively. These tools represent the parent nodes from which several other modalities were

derived and developed. Yet, the same trade-off between, fluorescent excitation mechanism,

resolution, focality, cost and ease of operation persists, making the selection process an arduous

task that pivots on the scope of the scientific question. For instance, a study of cell-type-specific

interactions in a small cortical region would probably be best addressed using TPI’s high

resolving capabilities. Mechanisms of coplanar neuronal ensembles in deeper brain areas can be

captured by single-photon micro-endoscopes. However, listening to large, cell-type-specific

neuronal populations and inferring average ensemble activity can only be achieved by Fiber

Photometry or classical multielectrode array-based electrophysiology.

Page 57: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

57

Table 2-1. Comparison of the three main fluorescent imaging techniques.

Two-photon imaging

(TPI)

Single-photon

endoscopy

Single-photon fiber

photometry (FP)

Fluorescence excitation Two-photon Epifluorescence Epifluorescence

Linearity Non-linear Linear Linear

Readout 2D image 2D image 1D aggregate signal

Resolving power +++ + 0

Field of view +++ ++ ++

Light power +++ -- ---

Focality +++ + 0

Penetration depth --- ++ +++

Tissue insult --- +++ +

Surgical footprint +++ ++ +

Cost +++ ++ +

Ease of operation --- + +++

Figure 2-15. Qualitative comparison chart. Comparison of the three fluorescent imaging

techniques in terms of cost, experimental flexibility, tissue damage, and penetration

depth.

Page 58: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

58

Motivation and Research Aims

Fiber photometry offers many features that may not be obvious at first glance or when

contrasted with other fluorescent-based optical tools. The reason being that FP operates on a

different spatial scale, which despite its coarseness can address very significant basic science

questions when coupled to the right experiment design. Appendix A provides an exhaustive

technical description of the FP system.

When I learned about the Fiber Photometry and implemented it, it was the first optical

readout modality used in our lab. Our lab specialized in electrophysiological (ephys) recording of

neural activity using a variety of recording devices like, Michigan Probes, multi-channel

microwire arrays, and tetrodes. Therefore, the most reasonable way to confirm that my

implementation is working as desired was to compare the optical readout to ephys readouts,

which I did in two experiments. For the sake of brevity, I’ll succinctly present only one of the

experiments here. Both experiments are detailed in Appendix B.

The rodent whisker system lent itself as a very apt model to corroborate the viability of

the FP device by virtue of its clear somato-topical organization that allows a one to one mapping

of a single vibrissa (whisker) to a specific cortical column (barrel) in the vibrissal representation

of the somatosensory cortex (vS1). This means that mechanical deflections, which present the

sensory stimulus, of a particular whisker will result in a response in a well-defined cortical area.

Cortical sensory coding of stimulus parameters is abundantly studied, and it has been shown that

changes in sensory stimulus strength are represented as changes in cortical response probability

when multi-unit activity is recorded electro-physiologically [119]–[121]. I was curious to know

how the aggregate FP signal will compare to the multi-unit ephys recoding and whether it will be

sensitive enough to detect changes in sensory stimulus strength. Since this section serves as

motivation and not as experimental results, I’ll present the short answer. The answer was: Yes,

Page 59: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

59

the FP signal was able to detect changes in sensory stimulus strength and, like in the ephys case,

coded this change as variation in cortical response probability. Figure 2-16 shows the

neurometric curve of the FP signal and the previously reported ephys recording. The neurometric

curves show strong agreement and similar saturation to high whisker deflection velocities.

This was a very surprising and intriguing result given the aggregate nature of the FP

signal and the complete absence of spatial resolution, which would suggest the loss of

information or reduced sensitivity compared to a multi-unit ephys recording. While this result

was significant in that it 1- validated my implementation of the FP system and 2- provided

evidence of reliable optically recorded sensory coding, it also instigated the core question of my

research: What IS the FP signal? And what is it made of?

A)

B)

Figure 2-16. Neurometric curve as a function of stimulus intensity. A) Fluorescence-based

neurometric data collected with FP from the rat vS1 showing the average (mean ±

s.e.m., n = 4) for a total of 16 recording sessions over 175 days. B) Neurometric

ephys data reported in [120]

Page 60: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

60

Significance

Unlike, the other optical readouts the FP signal is not a 2D image that represents a well-

defined plane or a section within a volume of neuronal tissue with spatially identifiable neurons.

On the contrary, the FP is a one-dimensional signal that corresponds to the lumped neural

activity of a volume of neural tissue that lies underneath the optical fiber (Figure 2-14C).

Characterizing the volume from which the FP system can detect emitted fluorescence is thus

crucial for the correct interpretation of the information carried by the FP signal about the neural

circuits being investigated. Hence, the question is: How big is this volume, what defines it, and

how is the activity of individual neurons, that reside within this volume, compounded into the 1-

D FP signal.

Research Aims

The overarching goal of this dissertation is to construe the aggregate 1D FP signal by

conducting a systematic characterization of the FP device. This is achieved via three main aims:

• Aim1: Spatial characterization of the detection volume of the FP system in-vitro.

• Aim2: Validation of the in-vitro spatial characterization by reconstructing the ensemble

statistic of the 1D FP signal from its contributing sources in-vivo.

• Aim3: Empirical modeling of the detection extent of the FP system and development of a

tool to predict the detection volume of arbitrary optical fibers.

Each specific research aim will be thoroughly addressed in the next three chapters.

Page 61: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

61

CHAPTER 3

SPATIAL CHARACTERIZATION OF THE DETECTION VOLUME

In this chapter, the work done to fulfill the first Aim will be presented. The intention of

this Aim is to characterize the detection volume of the FP signal by performing direct

measurements of static fluorescing sources with known locations, first in phantom brain slices

then in acute brain slices. The detection volume represents the spatial extent, with respect to the

tip of the fiber, from which the fiber can detect fluorescence. The change in the detection

strength as a function of distance from the fiber tip is the spatial detection profile. The detection

volume and the spatial detection profile of four optical fibers are experimentally measured using

green fluorescent beads.

Background

As described in previously, the FP device relies on an optic fiber to deliver excitation

light and collect fluorescent emission light. Thus, the extent (breadth and depth) from which the

FP system can detect emitted fluorescence relies on three factors: 1) how much of the initial

excitation light power (Pex) propagates from the fiber through the tissue, which is called the

volume of influence, 2) the probability that Pex is sufficient to excite the GCamP6 molecule and

cause it to fluoresce at a certain point (x, y, z) with respect to the fiber tip, which is referred to as

the volume of emission, and 3) how much of the emitted fluorescence arrives back to the face of

the fiber within its angle of acceptance, θacc, without being scattered or absorbed by the tissue,

defined as the volume of detection (Figure 4-1A).

Factor #1 involves characterizing properties of light propagation from the fiber through

biological tissue and has been measured at different wavelengths in fixed brain slices of mice

and rats within the framework of optogenetic actuators [57], [62], [63] as described in Chapter

2.5.1 (Optical actuators). Factors #2 and #3 are concerned with the propagation of the light

Page 62: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

62

emitted by the GCamP6 and its detectability by the fiber. It is important to note that not all

GCamP6 molecules that are influenced (i.e. receive light) are being excited and that not all

fluorescing molecules are being detected. Hence, it is the intersection of these two sets (excited

fluorophores whose emission is detected) that constitute the detection volume.

To quantify the detection volume of the FP system via direct measurements of its extent,

green fluorescent protein (GFP) beads were used as point sources in phantom brain slices, which

presented the turbid media, to demonstrate the principle. In a subsequent experiment, the brain

phantom slices were replaced with live (acute) brain slices to better match the optical properties

of the rodent brain. The thickness of the slices was varied to simulate different depths of the

recorded volume of tissue. Furthermore, four optical fibers of different geometries were used to

investigate possible effects of diameter and numerical aperture (NA) on the detected volume

(Figure 3-1).

Methods

Fiber Photometry Signal Acquisition

A high-power LED (λex = 475 nm, Thorlabs, Newton, NJ) was used to deliver the blue

excitation light to the GFP beads. After exiting the current-controlled LED the light was

collimated, passed through the dichroic mirrors as described previously and transmitted to the tip

of the optical fiber with a patch cord (Figure 3-1). The power of the excitation light coming out

of the fiber was measured with an optical power meter (Thorlabs, Newton, NJ) at the free end of

the fiber. Green fluorescence was collected by the same optical fiber and routed to the femto-

watt photodetector consisting of a single photo-cell (1 mm2) (#2151, Newport, Corporation,

Irvine CA) through a green emission filter (λem = 535 nm) and a convex lens. The photodetector

transduced the detected optical signal to an electric analog signal which is then acquired by the

TDT data acquisition system (Tucker Davis Technologies Inc., FL). The four multimode optical

Page 63: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

63

fibers that were used had the following core diameter [µm] /NA combination of 400/0.50,

400/0.22, 200/0.5, 200/0.22. The recorded signals were analyzed using custom-written

MATLAB® (MathWorks Inc.) scripts.

Phantom Brain Preparation

To resemble the optical properties of the rat brain, a brain phantom was prepared as

described in [122], [123]. A 6.4% v/v skim milk (Millipore Sigma) solution was added to a 1%

Agarose (Sigma-Aldrich) solution and mixed until uniform. Once the solution is set, an agar-

milk block was affixed to the stage of a vibratome using glue. The block was then sliced into

different thicknesses starting at 50 µm until 750 µm thick slices were achieved.

Acute Brain Slice Preparation

All animal care and experimental procedures were approved by the University of Florida

Institutional Animal Care and Use Committee. Long-Evans rats received a large dose of

pentobarbital while deeply anesthetized. When confirmed areflexia, transcardial perfusion with

ice-cold ACSF was performed. Brains were extracted promptly after decapitation and were

placed in ice-cold ACSF for 5 minutes to settle before being transferred to the vibratome. The

slicing tub of the vibratome, as well as the slicing solution (ACSF), were maintained at 0-2 °C.

Two to three sets of 100 µm to 700 µm thick transverse slices (100µm step size) were collected

serially and the slices were maintained viable in a temperature-controlled (0-2°) ACSF bath

which was continuously perfused with O2.

Fiber Photometry Recording Procedure

The same procedure was followed for recording from phantom slices and live (acute)

brain slices (Figure 3-1C). Polystyrene GFP beads (F8844, ThermoFisher) had a diameter of

15µm which in on a scale similar to that of neuronal bodies with a diameter of 10 - 20 µm. The

beads were then placed on a glass microscope slide using a camel-hair brush. Appropriate bead

Page 64: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

64

separation, i.e.: no overlapping beads was assessed under a fluorescent widefield microscope

equipped with a GFP filter. This was done to ensure that the beads are sufficiently separated,

avoiding the possibility to record from overlapping beads. Also, the beads had to be apart from

each other by a distance that is greater than twice the diameter of the used optical fiber to avoid

superposition from multiple sources and to ensure the detected signal comes from a single bead

An acute brain slice was then gently placed on top of the beads and the microscope slide. The

optical fiber was held vertically, as is the case in real-life FP recordings, using a digital

stereotaxic manipulator arm and manipulated in the horizontal plane using micrometer precision.

The micro-manipulator was used to move the fiber along the surface of the slice and to center it

over a fluorescent GFP bead. Fluorescence signals emitted by the beads were then acquired

while the fiber was stereotaxically translated horizontally along the x and y-axis with respect to

the GFP bead. The setup is depicted in Figure 3-2 only using clear for better visualization of the

beads.

A single scan is defined as the process of moving the optical fiber a distance of -500 µm

to +500 µm relative to the GFP bead when the bead is centered at 0 mm. For every slice

thickness, which presents a certain depth, six scans were collected from two non-overlapping

GFP beads. This was repeated twice for each depth using two slices of the same thickness.

Accordingly, there were six scans for every bead, and a total of four beads were recorded for

every depth. Therefore, every recorded data point was repeated 24 times, n = 24 trials. The

instantaneous spatial position (x, y, z) of the optical fiber was displayed on the stereotaxic LED

display and was simultaneously recorded by the data acquisition system to synchronize fiber

position with the detected fluorescent signal. This was achieved by a custom-built interface

board that allowed serial communication between the LED display and the TDT data acquisition

Page 65: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

65

system. Two embedded hardware boards (sbRIO-9216 and NI-9263, National Instruments) were

programmed in LabVIEW software. The built-in serial communication port (RS-232) of the

stereotaxic display provided the current x, y, and z coordinates every 10 ms and sent it to the

sbRIO, which read the coordinates via its RS-232 port. The received coordinates (string data

type) were cast to three analog outputs (integer data type) by an embedded code implemented in

LabVIEW on the sbRIO. The analog output was then mapped using the NI-9263 that conformed

to TDT’s analog input range. This was an indispensable step that ensured firm knowledge of the

exact location of the tip of the optical fiber with respect to the recorded GFP bead at every

instance in time, which is essential for accurate characterization of the detection volume.

Figure 3-1. Experimental setup. A) Enlarged view of the red box in B showing the hypothetical

volume of influence (blue) and emission (green). Notice θacc, axial and off-axial axes.

B) Fiber Photometry system. C) Data collection procedure

Page 66: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

66

A)

B)

C)

Figure 3-2. Agar-bead setup. A) Widefield fluorescent microscope picture of clear agar mixed

with 15µm GFP beads. B) same as A but showing reference grid placed under petri-

dish. C) Full setup, showing the optical fiber held by the stereotaxic arm and shining

blue light. [Photo courtesy of author]

Results

Moving the fiber towards or away from the fluorescent bead and increasing the thickness

of the tissue (brain phantom or acute slice) mimicked the effect of scanning the volume beneath

the tip of the fiber horizontally and vertically to determine the dimensions of the detection

boundary. The quantification of this boundary using a particular optical fiber and 15 µm beads

was associated with a few technical challenges. For example, extreme care had to be taken

during the manipulation of thin acute slices as they tend to fold and unfolding them can’t be

achieved without damaging the tissue. Placing the acute slices over the fluorescent beads also

required very gentle maneuvers to avoid tissue folding or beads to float over the surface of the

slice, which can compromise the accuracy of the localization process. Last but not least, speed of

data collection was critical to ensure the integrity of the acute brain slice remains the same once

removed from the bath, which can compromise the viability of the optical properties of the tissue

over time.

Page 67: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

67

Volume of Detection from Phantom Slices

Brain phantoms can simulate various properties of the brain, like mechanical, optical,

thermal, electrical or magnetic properties. The agar-milk mix mimics the optical properties of the

brain only in a homogenous manner, i.e.: it does not exhibit the small optical inconsistencies

observed in actual biological tissue [122], [123]. The mechanical properties of the agar-milk,

however, are quite different from a rodent brain, especially in terms of texture, consistency and

tensile strength. As such, it provided a good testbed to practice the dexterous manipulations

needed with acute brain slices while eliminating the need for live brains. At each slice thickness,

which represents a different depth, a single fluorescent source (bead) had to be located before the

optical fiber could be centered around it. After localizing the fluorescent bead, the optical fiber

was centered over it by finding the peak in the detected fluorescent signal. The stereotaxic

coordinates of the manipulator's arm were then zeroed, and the optical fiber is positioned -500

µm away from the source center, horizontally, in the x or y-direction. The scanning process

started by translating the optical fiber at ~10 µm/s towards the source center, eventually bringing

it over the source center and passing it by +500 µm. During the scanning process, the volume

beneath the optical fiber is constantly illuminated and any detected fluorescence is recorded with

the TDT system. Hence, the detected FP signal increased monotonically as the optical fiber

approached the source, eventually reached a maximum when the fiber is exactly centered over

the fluorescent source, then started to decline as the fiber was moved away from the source.

Therefore, every scan followed a bell-shaped curve, whose y-abscissa is indicative of the

detection strength as a function of lateral distance, the x-abscissa (Figure 3-1, bottom right).

Characterization of the entire volume from which the optical fiber can detect fluorescent signals

required stacking the recorded curves vertically, along the z-axis, to generate a meridional

section of the fiber’s detection map (Figure 3-3A). Due to the cylindrical nature of optical fibers,

Page 68: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

68

revolving the meridional section for 360° around the z-axis fully describes the volume of

detection for a given fiber.

This scanning process was performed to measure the detection volume in phantom as

well as in the acute slice preparation described in the next section. The process was repeated for

all depths (slice thicknesses) and the detection map of the 400/0.5 optical fiber from the phantom

brain was generated as shown in Figure 3-3A. It was notable that the detection volume extended

relatively further away from the fiber tip compared to optogenetics studies that quantified light

propagation in highly scattering media in fixed brain slices. These studies focused on measuring

the volume of influence (forward light propagation) and did not consider the emission volume

(backward light propagation). The final results of those studies demonstrated more than 50%

reduction in the initial light intensity within 50-100 µm from the tip of the optical fiber [57],

[62], [63]. This rather exponential decay in the profile of influence is predominantly governed by

Beer-Lambert’s law and is depicted by the blue trace in Figure 3-3 B.

Figure 3-3. Spatial detection extent in brain phantom. A) Spatial detection map for a 400/0.5

optical fiber in agar/milk brain phantom slices in the meridional plane of the fiber (x-

z plane). B) Measured axial detection profile and axial influence profile as computed

using Beer-Lambert’s law and the optical properties of neural tissue.

Page 69: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

69

The rapidly declining volume of influence reported by [57], [62], [63] contrasts with that

of the volume of detection measured in our experiment. A substantially different profile was

observed in the gradient of the detection volume along the fiber axis. We found that, while the

influencing light intensity (excitation) is reduced to 50% of its initial value at about 100 µm from

the fiber tip (and even about 50 µm according to [63]), the fluorescent source still received

enough excitation light to fluoresce and be detected with more than 50% probability at depths as

far as ~350 µm below the tip of the optical fiber (Figure 3-3B). This means that the spatial

gradients exhibited by the volumes of influence and detection are significantly different as can be

seen from the overlay of the axial detection profile and the axial influence profile reported

previously (Figure 3-3B).

Volume of Detection from Acute Brain Slices

The discrepancy observed between the detection and influencing profiles could be

explained by the difference in optical properties between the phantom slices and the acute brain

slices. Therefore, we asked whether the detection profile obtained in phantom brains are

similarly observed in acute brain slices that have more biological inhomogeneity and scattering

effects compared to the phantom brain preparation. We also asked whether the detection volume

is a function of the optical fiber geometry, expressed hereafter as the diameter[µm]/ NA

combination. Four optical fiber geometries were compared by varying the core diameter (large =

400 µm and small = 200 µm) and the NA (high = 0.50 and low = 0.22). After collecting six scans

per bead, two beads per slice thickness and two slices per thickness, the resulting 24 trials were

stacked vertically to create a meridional section, which was then revolved around the z-axis. We

found that both the diameter and NA of the optical fiber had differential effects on the size, shape

and spatial gradient of the detection volume (Figure 3-4). In particular, the x-axis represents the

lateral distance from the center of the optical fiber and is normalized to each fiber’s radius, rf.

Page 70: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

70

The y-axis represents the depth from the tip of the fiber, which is usually referred to as the z-axis

during an actual recording. The geometry of the optical fiber (diameter[µm]/ NA) is indicated in

the bottom right corner and the graphed iso-contours indicate the 90%, 75%, 50% and 25% of

the maximum detection strength (red, orange, green and blue respectively).

In particular, it was noticeable that the axial boundary of the detection volume was

significantly larger for low NA fibers compared to high NA fibers, where the 50% and 25% is-

contours were approximately 100-150 µm deeper for the latter (t-test, p < 0.002, α = 0.05, n =

24). We also found that the axial detection profile of low NA fibers was associated with slow

decline compared to high NA fibers that exhibited much faster decay as a function of distance

from the fiber tip. The difference in axial detection profile plays a very important role in how

fluorescent sources along the axial direction contribute differentially to the over-all FP signal. In

other words, it defines the share of each source towards the final lumped 1D FP signal.

Detection depth was not the only parameter that was influenced by the geometry of the

optical fiber. The detection from the periphery or the breadth of the detection volume was

impacted by the diameter of the optical fiber rather than the NA. The detection from the

periphery occurs along the off-axial direction defined by the acceptance angle, θacc, of the fiber

as illustrated in Figure 3-1 and previously described in Chapter 2. Specifically, optical fibers that

have a small diameter fiber detected off-axis fluorescent sources with about 50% more detection

strength compared to large diameter optical fibers as seen in the right panel of Figure 3-4.

In addition to the depth and breadth of the detection volume, its shape also varied

significantly depending on the NA of the used optical fiber. Figure 3-5 shows the 3D shape of

the detection volume that results from revolving the meridional section 360° around the z-axis.

Page 71: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

71

While high NA (0.50) optical fibers detected a rather conic volume, low NA (0.22) fibers

collected fluorescence from a frustum-shaped volume (Figure 3-5).

Finally, the physical volume bounded by each detection iso-contour was quantified

analytically by calculating the volume enclosed by the convex surfaces created by the iso-

contours in 3D. Table 3-1 summarizes the value of the detection volume of each examined

optical fiber and is categorized by iso-contour percentage.

Figure 3-4. Spatial characterization of detection volume as a function of fiber geometry in acute

brain slices. A) Left: Detection extent in the meridional plane the high NA fiber.

Middle: same as A but for the low NA fiber. Right: axial (solid line) and off-axial

(dashed line) detection profiles. B) the same as in A but for small diameter (200µm)

fiber.

Page 72: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

72

Figure 3-5. 3D Volume of detection for each optical fiber. A) Left: Volumes enclosed in 90, 75,

50 and 25% iso-contours for large/high NA fiber. Right: for low NA fiber. B). same

as A but for small (200µm) fiber

Table 3-1. Detection volume for different optical fibers.

Fiber Detectable Volume (x105 µm3), [µm]

<75% <50% <25%

400/0.50 3.2 [68] 16 [117] 58 [180]

400/0.22 8.7 [95] 47 [168] 156 [250]

200/0.50 1.9 [57] 10 [100] 38 [156]

200/0.22 9.2 [97] 27 [139] 59 [181]

Note: The geometry of each optical fiber is listed in the first column. Each subsequent column

lists the volume enclosed within the 75%, 50%, and 25% detection contours. To develop a better

sense of the physical size of the detection volume, the length of one side of a perfect cube that

would encompass the quantified detection volume is reported in square brackets. In other words,

the values within square brackets denote the cubic root of the detection volume and present the

side of a cube of equivalent volume.

Page 73: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

73

Discussion

The detection volume of the optical fibers was measured directly using fluorescing

sources (GFP beads) in brain phantom and in acute brain slices. The phantom preparation served

as a pilot experiment and revealed a stark difference between the previously reported influencing

profile and the detection profile measured in this experiment. Considering the exponentially

decaying profile of influence, one would presume a similarly shaped detection profile. However,

the detection profile showed a rather smooth spatial decline. We, therefore, conducted the same

experiment using acute brain slices to allow for a better comparison with the volume of influence

reported by optogenetics studies that were carried out in acute and fixed slice preparations. The

results from the acute slice preparation confirmed the smooth spatial decline in the detection

profile, which has direct implications on the extent to which individual neurons, neuropil, and

other structures could contribute to the FP signal.

Furthermore, we considered four optical fiber geometries that are commonly used and

offered good contrast in terms of diameter size (400 vs. 200 µm) and NA (0.5 vs 0.22). Our data

suggest that the fiber diameter and NA play critical roles in defining the size and shape of the

detection volume, in accord with a recent study that conducted a similar experiment to measure

the detection extent of optical fibers [124]. The study investigated the detection volume of three

optical fibers (50/0.22, 200/0.39 and 200/0.5) analytically and numerically. The analytical

approach Pisanello et al. presented in their study is a direct implementation of the theory from

[125], which describes the detection volume of an optical fiber-based on the geometry of the

cone of acceptance only, without the consideration of the optical properties the medium. Hence,

the unreasonably large detection volumes (up to 800 µm) were expected. On the other hand, the

numerical modeling of the detection volume in a homogenous medium yielded results that are

comparable to the detection extent we measured using the phantom-brain, which presents a

Page 74: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

74

homogenous medium as well. At the end of the study, Pisanello et al. quantified the detection

extent of two optical fibers (200/0.39 and 200/0.5) in acute brain slices and the reported

detection volume for the 200.0.5 fiber is comparable to the volume we measured in this study.

One important contribution compared to the work reported in [124] is that we used a

single photon illumination scheme and the same photodetector to accurately replicate the

conditions under which FP recordings are readily conducted. Pisanello et al., on the other hand,

used a combined confocal/two-photon illumination setup and photomultiplier tubes (PMTs) that

are rarely used during FP recordings, possibly hindering the extension of their results to other

setups. Other attempts to quantify the detection volume were performed as an ancillary to the

main study and were either specific to a certain brain region of the mouse brain [126] or used an

unrealistic fluorescent source [127].

Furthermore, the findings illustrated in the previous section instigate a very significant

observation considering the ramping number of recent studies that utilize the FP system to record

neural ensemble activity from various brain regions (Table 3-2). Many FP studies rely on the FP

system for its low cost, versatility, ease of operation and flexibility in awake behavioral

experiments. However, it is often the case that the same optical fiber, that was used in a cited

study, is chosen with little consideration for its suitability [128]–[130]. The previous findings, on

the other hand, suggest that deliberate selection of the geometry of the optical fiber is crucial to

determine the specifications of the detection volume. In other words, the type of study and the

targeted brain region should guide the selection of the diameter and NA combination of the

optical fiber that will be used for an FP-based experiment.

Page 75: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

75

Table 3-2 Summary of recent FP studies listing used animal model, target region(s), type of investigation and geometry

(diameter[µm]/NA) of the optical fiber utilized in the study.

Animal

Model Target Region(s) Investigation Optical Fiber Power Ref

Mouse

PFC, CA1, BLA, LH,

VTA, NAc & BNST

simultaneously

Real-time activity relationship

among many brain regions during

social behavior

400/0.48 2.5-50 µW [131]

Mouse DRN Acute social isolation 400/0.48 10-50 µW [132]

Mouse SON of the hypothalamus Ingestive behavior 200/0.39 100-200 µW [133]

Mouse Lateral hypothalamus Eating behavior 200/0.37 100 µW [127]

Mouse NAc Drug seeking addiction 400/0.48 30-75 µW [134]

Mouse vS1 and OFC Sensory integration 200/0.39 --- [135]

Neonatal

Mouse Temporal lobe

spontaneous activity and early

network oscillations (ENOs) 200/0.48 250 µW [136]

Mouse vS1 Sensory exploration 440/0.22 --- [65]

Mouse Auditory cortex Auditory network calcium

transients’ (NCaTs) 200/0.48 10-100 µW [66]

Mouse Dorsal STR Voluntary action initiation

Hybrid fiber setup exc.

fiber 3.5 µW and det. fiber

105 µm, NA not reported

100-120 µW [67]

Mouse V1 and dLGN thalamus

and VPM

Corticothalamic slow oscillations

of non-Rem sleep 200/0.48 <0.1mW/mm2 [137]

Mouse V1 Sensory integration 200/0.48 30 µW [138]

Non-human

primate M1 Motor movement mapping 200/0.48 --- [68]

Page 76: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

76

Table 3-2 Continued

Animal

Model Target Region(s) Investigation Optical Fiber Power Ref

Mouse striatal projection neurons Operant conditioning

Hybrid fiber setup exc

fiber 3.5µW and det fiber

105µm no NA reported

100 µW [69]

Mouse VTA-Nac projections Social behavior 400/0.48 10-50 µW [70]

Mouse CA1 Locomotion initiation and

velocity 200/0.37 0.6-1.2 mW [71]

Mouse ARC of the hypothalamus and

PVH Feeding circuits and behavior 400/0.48 70 µW [139]

Rat S1FL

Neural networks activation due

to sensory stimulus neuronal

response

200/-- 1.3µW/mm2 [140]

Rat (S1FL)

Relationship between blood

oxygen level–dependent

(Bold)and neural activity

200/0.48 <1 mW [72]

Mouse mPFC and M1 Motor Skill learning 105/0.22 100 µW [126]

Mouse NAc Stress Susceptibility 400/0.4 --- [73]

Mouse PBN–projecting neurons Itch sensation --/0.37 --- [141]

Mouse ARN gonadotropin hormone (GnRH)

release 400/0.48 50 µW [142]

Abbreviations: PFC: Prefrontal cortex. CA1: hippocampal area CA1. BLA: Basolateral amygdala. VTA: Ventral tegmental area. NAc:

Nucleus accumbens, LH: Lateral hypothalamus, BNST: Bed nucleus of stria terminalis, DRN: Dorsal raphe nucleus, SON: Supraoptic

nucleus, OFC: orbitofrontal cortex, vS1: vibrissal primary somatosensory cortex, STR: striatum, V1: Primary visual cortex, dlGN:

dorsal lateral geniculate nucleus, VPM: ventral posteromedial nucleus, M1: Primary motor cortex, ARC: arcuate nucleus, PVH:

paraventricular hypothalamus, S1FL: primary somatosensory cortex, forelimb region, PBN: parabrachial nucleus, ARN: arcuate

nucleus kisspeptin, exc: excitation, det: detection.

Page 77: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

77

CHAPTER 4

VALIDATION OF THE SPATIAL CHARACTERIZATION

The preceding chapter showed how the shape and size of the detection volume depend

heavily on the geometric specification of the optical fiber, specifically the combination of

diameter and numerical aperture (NA). Due to the arduous nature of conducting direct

measurements using 15 µm fluorescent beads, the investigation was carried out in an in-vitro

preparation and was limited to quantifying the detection volume of four optical fibers1. There are

two main limitations of this investigation. First, it was performed in-vitro in acute brain slices

and second, only four optical fibers were considered. This chapter will focus on the first

limitation while the second limitation will be addressed in the next chapter.

The in-vitro acute brain slice preparation offered a good approximation regarding the

inhomogeneous optical properties of the rodent brain, but it cannot be claimed that the optical

properties are identical to their in-vivo counterpart as tissue integrity may be compromised

during the recording process. Therefore, validation of the direct measurements in an in-vivo

setup is the second Aim of this research.

Premise and Hypothesis

The goal of Aim2 is to validate the direct measurements of the detection volume in the

mouse visual cortex (V1) during the presentation of drifting orientation gratings. V1 is a well-

characterized cortical area where sensory-evoked responses from single neurons are referred to

as tuning curves. Turning curves define which visual stimulus, in this case, which orientation of

the drifting gratings a certain neuron is most selective to [143], [144]. Neurons that are selective

to a particular orientation are grouped in a columnar organization in highly visual mammals like

1 Likely the reason why Pisanello et al. only examined two fibers in their in-vitro preparation too.

Page 78: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

78

the cats, ferrets, tree shrews and the non-human primate [145], [146]. In the rodent, however, this

columnar organization is not evident. Recent studies in the mouse visual cortex showed that

neurons that respond to the same orientation have a topographical organization of fuzzy patches

rather than columns [107], [147]–[149].

The difference in the topographical organization doesn’t have a consequence on the

tuning or selectivity when single cells are considered. However, when considering a group of

cells or an ensemble, as is the case with the FP, then the topographical organization of patches

will have a unique effect on the recorded signal. The proportion of differently tuned neurons that

compose a patch and where the optical fiber falls relative to the different patches will result in

different FP signals. Simply stated, the recorded signal will represent the combined ensemble

tuning of the neurons within the detection volume of the optical fiber. The ensemble tuning

recorded by the optical fiber differs from one location to another throughout V1 as each location

falls within a different patch and has a different composition of tuned neurons. This is a desired

feature, which allowed the validation of the direct measurements of the detection volume in

different regions.

The premise is that the combined tuning of the constituent sources will give rise to the

overall ensemble tuning, where the tuning of the sources and the ensemble is defined by the

tuning statistic 𝑇𝑛,𝑑(𝜃𝑘), and 𝑇𝐹𝑃(𝜃𝑘), respectively. Therefore, volumetric optical scanning (OS)

of the detection volume using epi-fluorescent CCD-imaging was performed to record from single

sources and to reconstruct the overall ensemble statistic using a linear mixture model (LMM)

(Figure 4-1A). Optical sectioning was one of the possibilities but was ruled out as it relies on

two-photon illumination, which is nonlinear and fundamentally different from the single-photon

illumination scheme used in FP.

Page 79: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

79

The LMM derives a pseudo-population2 statistic, 𝑂𝑆(𝜃𝑘) from the constituting single

sources’ statistic. Taking advantage of the linearity of the problem, the LMM assigns a weight

wd(x, y) to each source statistic 𝑇𝑛,𝑑(𝜃𝑘), that falls within the boundary of the detection volume,

at every depth d, based on its spatial location x ,y with respect to the fiber. Accounting for all

depths then yields a composite pseudo-population statistic 𝑂𝑆(𝜃𝑘) that corresponds to the

recorded population FP statistic 𝑇𝐹𝑃(𝜃𝑘), (Equation 4-1 and Figure 4-1E). The pseudo-

population statistic, 𝑂𝑆(𝜃𝑘) was contrasted to the observed population statistic, 𝑇𝐹𝑃(𝜃𝑘),

recorded with the FP (Equation 4-2), where 𝜃𝑘represents the eight different visual stimuli (k

=1,..8 corresponding to 8 orientations 0° ,45°, …, 315°). The values of the weights and

boundaries are adapted from the direct measurements reported previously [150] (see Figure 3-3).

A more detailed mathematical formulation of the tuning statistic, as derived from the recorded

calcium trace, and of the LMM can be found in Appendix C.

Figure 4-1E illustrates a hypothetical volume of tissue, where neuronal sources are

distributed within the 90, 75, 50 and 25 % detection contours of a fiber. Taking an optical scan at

depth d = 1 will result in circular shaped boundaries in the x-y plane as opposed to the parabolic

contours in the y-z plane. Accordingly, sources within the red boundary will be weighted by 0.9

while sources inside the blue boundary will contribute by 0.25. The LMM directly implements

the detection boundaries of the in-vitro study. As such, if the in-vitro direct measurements are

valid in-vivo, then the pseudo-population statistic, 𝑂𝑆(𝜃𝑘) should approximate the FP- recorded

population statistic 𝐹𝑃(𝜃).

2 The terms ensemble and population will be used interchangeably.

Page 80: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

80

𝑂𝑆(𝜃𝑘) = ∑

𝐷

𝑑=1

∑ 𝑤𝑛,𝑑(𝑥, 𝑦)𝑇𝑛,𝑑(𝜃𝑘)

𝑁

𝑛=1

(4-1)

𝑇𝐹𝑃(𝜃𝑘) ≈ 𝑂𝑆 (𝜃𝑘) (4-2)

The proposed hypothesis mandates that the FP and OS responses must be recorded from

the exact same location in V1. To achieve this, a coordinate system, whose zero is centered at a

reference point marked on the subject’s head plate, was implemented (Figure 4-1C). Figures 4-

1D-E summarizes the proposed hypothesis and illustrates the experimental setup.

Methods

Surgical Procedure

All animal care and experimental procedures were approved by the University of Florida

Institutional Animal Care and Use Committee. Wild type C57B6/J mice (n = 2) received bilateral

craniotomies of 5mm radius centered over V1 (ML: +/-2.5 mm, AP: -3.5 mm, DV: -0.3 - -0.5

mm). 1 µL of AAV1.Syn.GCaMP6f.WPRE.SV40 (titer: ~ 2.5x1012 genomes/mL) was injected

in each hemisphere at a rate of 75 nl/min. Half of the volume was infused at -0.5 mm before the

tip of the needle was retracted to -0.3 mm to inject the rest of the volume. The needle was fully

retracted after a 10-15 minute wait period. Finally, each craniotomy was covered with an

optically clear cranial window (#1943 5 mm, Bellco Glass, Inc.), which was secured to the skull

and the metal head plate with dental cement. The metal head plate was designed in house using

SolidWorks®. The design was fabricated with stainless steel 304 (304SS) material using

professional laser cutting (https://www.lasercuttinginc.com/). Subjects were monitored post-

operatively to ensure a healthy recovery. Expression spread was monitored weekly and after the

third week, a reference point was marked on the metal head plate (Figure 4-1B-C and Figure C-

1). Coordinates to different locations in V1 were measured with respect to this reference point

which allowed the exact spot to be revisited between sessions of FP and OS data collection.

Page 81: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

81

Fiber Photometry Recording

A current-controlled blue LED (λex = 475 nm, Thorlabs, Newton, NJ) was used for GFP

excitation. The light was collimated and routed through a patch cord to the tip of an optical fiber

(Figure 1B). The power of the light emanating from the fiber was measured with an optical

power meter (Thorlabs, Newton, NJ) and set to 80 µW before every recording session. Emitted

green fluorescence (λem = 535 nm) was collected by the same optical fiber and routed to a single

cell (1 mm2) femto-watt photodetector (Newport, Corporation, Irvine CA) via a green emission

filter and a convex lens. The photodetector converted the optical signal to an analog signal that is

then acquired by a TDT data acquisition system (Tucker Davis Technologies Inc., FL) for

storage and later analysis. Custom MATLAB (MathWorks Inc.) scripts were written to analyze

the recorded signal.

Data Acquisition

The recording setup involved three separate components:1- Ca2+ signal acquisition in

form of an FP signal or a CCD camera image, 2- Visual stimuli display and 3- Delivery of the

stimuli by switching the screen on and off. These three components had to be precisely

synchronized and could not be controlled by different PCs. The TDT system was therefore

programmed to send control signals to trigger each one of those components and to receive

acknowledgment signals in response. All data were collected on the computer that runs the TDT

system. The subjects were head-fixed under the imaging modality (FP, 1p or 2p) by clamping the

head posts.

Visual Stimulation

Drifting sinusoidal orientation gratings were displayed on a presented on a 9.7-inch LCD

screen (LG LP097QX1, Adafruit) that was positioned at eye-level about 15-20cm away from the

eye to ensure coverage of the entire visual field. The screen contrast was set to 100% and the

Page 82: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

82

spatial frequency was set to 0.4 cycles per degree. The gratings were presented at eight

orientations with a 45° step. The eight stimuli were presented randomly for 4 seconds and

repeated ten times. The inter-stimulus-interval was 4 seconds. The screen was programmed using

the Psychophysics Toolbox in Python and was controlled to turn on and off with a TTL signal

generated by the TDT system.

Epi-fluorescent Volumetric Scanning

White light from a high-power mercury lamp (X-Cite 120Q Excelitas Technologies

Corp.) was used for the epi-florescent imaging. The light was first passed through an excitation

to filter out all wavelength except blue (λex = 475 nm) to excite GCamP6f. The blue light was

then attenuated from ~8mW to ~80µW with an optical density filter (NE20A-A, OD:2, Thorlabs

Inc.) to match the power used during FP recording. A 16x/0.8NA objective (Nikon) was used to

focus the light on the brain. The focal plane was then adjusted to different depths by moving the

objective along the z-axis. Emitted fluorescence was collected by the same objective and filtered

by an emission filter (λem = 535 nm) before it was collected by a CCD camera (Electro Retiga,

QImaging Inc.). Images captured were acquired by the CCD camera’s proprietary software

Ocular at an exposure of 60 ms and 2x2bining. Image acquisition was triggered via a TTL signal

from the TDT and every frame acquisition henceforth was reported via another TTL signal to the

TDT system.

Page 83: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

83

A)

B)

C)

D)

E)

Figure 4-1. Epifluorescent volumetric scanning setup and LMM. A) Illustration of the theory of

volumetric optical scanning. Left: initial position of the objective with maximum

fluorescence collection occurring from the focal plane. Right: Displacement of the

objective in the +/z-direction causes the focal plane, and thus the plane from which

fluorescence is maximally collected, to move up/down. B) Viral injection of

GCaMP6f in V1 followed by placement of a cranial window and the metal head-

plate. 2p, 1p Epifluorescence, and FP data are recorded through the cranial window at

the same anatomical location. C) Anatomical locations of the recorded data within V1

with respect to the reference point. Blue and red dots correspond to the subject1 and 2

(n = 2). Scale bar = 1 mm. D) Visual stimulation setup. Moving orientation gratings

are displayed on the LCD screen while FP or 1p Epifluorescence responses are

recorded. E) Hypothetical decomposition of the aggregate FP signal. Left: Sources

that contribute to the FP response and their spatial location with respect to the fiber’s

detection volume and contours in the y-z plane (red, orange, green, blue correspond to

the 90,75, 50 and 25% detection boundary respectively). Middle: optical planes

corresponding to sections d=1 and d=D in the left panel. Circles are x-y sections of

the contours shown in the left panel. Right: Mathematical formulation of the LMM

Page 84: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

84

Results

As described before, V1 recordings usually show tuning of single cells to one orientation

while the aggregate signal recorded by the FP results in a tuning curve that is representative of

all neurons within the boundary of the detection volume. The tuning curve of this aggregate

signal was called ‘ensemble tuning’, 𝑇𝐹𝑃(𝜃𝑘). In an ensemble tuning curve, the tuning of the

loudest sources, i.e.: the ones closest to the fiber, will dominate the tuning curve while the other

sources will still contribute to the rest of the curve. Given the topographic organization of

patches and the lack of clear columns in the mouse V1 [149], two scenarios are possible during

the recording of the ensemble tuning. One such scenario would be if the optical fiber landed in

the center of an orientation patch, where a group of neurons is selective to the same stimulus,

resulting in a peak in the ensemble tuning at the orientation of this stimulus. The other scenario is

if the optical fiber landed rather at the periphery of one patch or where one or more patches

intersect. In that case, the ensemble tuning curve would be less likely to show prominent tuning

to one orientation and will rather show uniform responsiveness to most stimuli.

In Figure 4-1C five locations (A, B, C, D, and E) in V1 are shown (Table 4-1). These are

the locations from which FP and OS signals were recorded. As described in the Methods section

recording the ensemble tuning curve of a certain location is rather straight forward using the FP.

To recap, the fiber was positioned at the desired location, the eight visual stimuli were presented

randomly and repeated ten times and the visually evoked responses (VER) were collected

simultaneously, with no further manipulation needed. Recording the VERs while optically

scanning the detection volume, on the other hand, is a quite elaborate process, which we’ll refer

to as the OS (optically scanned) signals from now on.

During the OS recording session, each of the eight visual stimuli was presented randomly

and repeated ten times at every depth, starting from the pia at depth = 0 and at every 20 µm steps

Page 85: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

85

until a depth of 250 µm was reached. VERs are recorded simultaneously in the form of a series

of images, a video. At every transition, the objective was moved down to position its focal plane

at the desired depth. This results in a stack of video files that correspond to every depth and that

shows the visually evoked fluorescent sources. These sources were considered regions of interest

(ROIs) and were extracted using the publicly available CaImAn toolbox [151], [152]. Since the

CaImAn was developed to handle high resolution, two-photon images some customization was

needed to adapt it to the less defined, low-resolution images acquired with OS. In OS-ed images,

individual neurons can only be seen if their soma falls exactly within the objective’s focal plane.

Otherwise, their activity is picked up in out-of-focus ROIs that do not necessarily resemble the

shape of neurons. Therefore, some parameters like the ROI size and the spatial correlation

threshold had to be adjusted. After extraction of the ROIs and the location of their centroids

using the modified version of the CaImAn, a weight wn,d(x, y) (score) is assigned to each ROI

(Equation 5-1). Those weights come from the direct measurements reported in Chapter 3 and

depend on the ROI’s spatial location and depth as seen in Figure 4-1E. As such, ROIs that fall

within the red contour will receive a weight of 0.9 while ROIs inside the blue contour will be

weighted by 0.2. The fidelity of the OS-ROIs was validated by comparing them to ROIs in a

two-photon image taken at the same depth. OS-ROIs that didn’t coincide with a two-photon ROI

and that fell outside of the boundary of the detection volume (i.e.: outside the blue contour) were

not included in the construction of the pseudo-ensemble statistic,𝑂𝑆(𝜃𝑘), recorded with OS.

Figure 4-2 shows the data collected from location E as well as the weight assignment

approach that was taken by the LMM based on the spatial location of the OS-ROI. The visually

evoked response (VER) traces of a few OS-ROIs are presented in Figure 4-2C-E along with their

individual tuning profile. While some OS-ROIS were selective to a certain orientation and

Page 86: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

86

exhibited sharp tuning, others seemed unselective to the presented stimulus. The overall

ensemble tuning statistic 𝑇𝐹𝑃(𝜃𝑘) observed with the FP is shown in the left panel of Figure 4-2A.

The middle panel demonstrates the pseudo-tuning statistic 𝑂𝑆(𝜃𝑘)constructed from the OS-

ROIs with the LMM using the weights and boundary constraints from the direct measurements

[150].A visual comparison of the observed and pseudo-tuning statistic makes it clear that both

share common features, with strong ensemble tuning to 90-180° that peaks at 135°, and near

chance tuning to other orientations. This can be explained by the fact that some of the shallower

OS-ROIs were more selective to 90-180°, while deeper and/or more peripheral ROIs were

selective to the other orientations. Furthermore, the statistical evaluation of the observed and

pseudo-tuning statistic proved that they come from the same distribution (two-sample

Kolmogorov-Smirnov test, p = 0.92, α = 0.05). In contrast, the left panel in Figure 4-2A shows

the control-pseudo tuning statistic resulting from the process of eliminating the spatial boundary

and profile. The control-pseudo tuning statistic was constructed without the spatial constraints

and characteristics derived from the direct measurements, i.e.: all OS-ROIs were included with

equal weights regardless of their spatial location. As expected, the result is a near-uniform tuning

statistic that is significantly different from the observed FP statistic and the constructed OS

tuning statistic (two-sample Kolmogorov-Smirnov test, p = 0.0014, α = 0.05). These results

suggest that the direct measurement of the detection volume and the spatial profile are valid and

applicable in-vivo as without them the information coded in the tuning statistic is lost. This

procedure was repeated in three other regions in V1 and similar results were observed for all

locations (Figure 4-3, 4-4 and 4-5) except one where the threshold for statistical significance was

not met.

Page 87: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

87

A)

B)

C)

D)

E)

Figure 4-2. In-vivo validation of the directly measured detection profile and volume. A) Left:

Ensemble tuning statistic, 𝑇𝐹𝑃(𝜃𝑘), recorded through FP from location E. Middle:

pseudo-ensemble tuning statistic, 𝑂𝑆(𝜃𝑘), estimated using individual sources

detected from volumetric OS data. Right: control-pseudo tuning statistic estimated

using all sources in volumetric OS data in the absence of the spatial constraints

imposed by the measured detection volume. (p<0.01 = **, ns = not significantly

different) B) FP and OS recorded calcium traces showing average ± s.e.m VERs to

the eight visual stimuli. Scale bars are 0.5s and 1 z-score unit. C) Left: Sample OS

image from a superficial range of depths (0-80µm). Scale bar is 100µm. The overlay

shows detection boundaries used by the LMM. Source 1 corresponds to a weight of

0.25 while sources 2-4 correspond to 0.75. Right: Corresponding 2p image at the

same depth showing the contributing cell bodies. Scale bar is 100µm. Bottom:

calcium traces of the sources during multiple trials of visual stimuli presentation (red

dashed lines indicate stimulus onset) and their individual tuning profiles 𝑇𝑛,𝑑(𝜃𝑘).

Scale bars are 4s and 1 z-score unit. D) same as C but from an intermediate section

(80-150µm). E) same as C but from a deeper section (150-250µm).

Page 88: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

88

A)

B)

C)

D)

E)

Figure 4-3. In-vivo validation of the directly measured detection profile and volume. A) Left:

Ensemble tuning statistic, 𝑇𝐹𝑃(𝜃𝑘), recorded through FP from location B. Middle:

pseudo-ensemble tuning statistic, 𝑂𝑆(𝜃𝑘), estimated using individual sources

detected from volumetric OS data. Right: control-pseudo tuning statistic estimated

using all sources in volumetric OS data in the absence of the spatial constraints

imposed by the measured detection volume. (p<0.01 = **, ns = not significantly

different) B) FP and OS recorded calcium showing average ± s.e.m VERs to the eight

visual stimuli. Scale bars are 0.5s and 1 z-score unit. C) Left: Sample OS image from

a superficial range of depths (0-80µm). Scale bar is 100µm. The overlay shows

detection boundaries used by the LMM. Sources 2 and 3 corresponds to a weight of

0.75 while sources 1-4 correspond to 0.5. Right: Corresponding 2p image at the same

depth showing the contributing cell bodies. Scale bar is 100µm. Bottom: calcium

traces of the sources during multiple trials of visual stimuli presentation (red dashed

lines indicate stimulus onset) and their individual tuning profiles 𝑇𝑛,𝑑(𝜃𝑘). Scale bars

are 4s and 1 z-score unit. D) same as C but from an intermediate section (80-150µm).

E) same as C but from a deeper section (150-250µm).

Page 89: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

89

A)

B)

C)

D)

E)

Figure 4-4. In-vivo validation of the directly measured detection profile and volume. A) Left:

Ensemble tuning statistic, 𝑇𝐹𝑃(𝜃𝑘), recorded through FP from location C. Middle:

pseudo-ensemble tuning statistic, 𝑂𝑆(𝜃𝑘), estimated using individual sources

detected from volumetric OS data. Right: control-pseudo tuning statistic estimated

using all sources in volumetric OS data in the absence of the spatial constraints

imposed by the measured detection volume. (p<0.01 = **, ns = not significantly

different) B) FP and OS recorded calcium traces showing average ± s.e.m VERs to

the eight visual stimuli. Scale bars are 0.5s and 1 z-score unit. C) Left: Sample OS

image from a superficial range of depths (0-80µm). Scale bar is 100µm. The overlay

shows detection boundaries used by the LMM. Source 2 corresponds to a weight of

0.75 while source 1 corresponds to a weight of 0.5 and sources 3-4 to 0.25. Right:

Corresponding 2p image at the same depth showing the contributing cell bodies.

Scale bar is 100µm. Bottom: calcium traces of the sources during multiple trials of

visual stimuli presentation (red dashed lines indicate stimulus onset) and their

individual tuning profiles 𝑇𝑛,𝑑(𝜃𝑘). Scale bars are 4s and 1 z-score unit. D) same as C

but from an intermediate section (80-150µm). E) same as C but from a deeper section

(150-250µm).

Page 90: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

90

A)

B)

C)

D)

E)

Figure 4-5. In-vivo validation of the directly measured detection profile and volume. A) Left:

Ensemble tuning statistic, 𝑇𝐹𝑃(𝜃𝑘), recorded through FP from location D. Middle:

pseudo-ensemble tuning statistic, 𝑂𝑆(𝜃𝑘), estimated using individual sources

detected from volumetric OS data. Right: control-pseudo tuning statistic estimated

using all sources in volumetric OS data in the absence of the spatial constraints

imposed by the measured detection volume. (p<0.01 = **, ns = not significantly

different) B) FP and OS recorded calcium traces showing average ± s.e.m VERs to

the eight visual stimuli. Scale bars are 0.5s and 1 z-score unit. C) Left: Sample OS

image from a superficial range of depths (0-80µm). Scale bar is 100µm. The overlay

shows detection boundaries used by the LMM. Source 1 corresponds to a weight of

0.75 while sources 2-4 correspond to 0.25. Right: Corresponding 2p image at the

same depth showing the contributing cell bodies. Scale bar is 100µm. Bottom:

calcium traces of the sources during multiple trials of visual stimuli presentation (red

dashed lines indicate stimulus onset) and their individual tuning profiles 𝑇𝑛,𝑑(𝜃𝑘).

Scale bars are 4s and 1 z-score unit. D) same as C but from an intermediate section

(80-150µm). E) same as C but from a deeper section (150-250µm).

Page 91: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

91

Table 4-1. Summary of anatomical locations

Location AP [mm] ML [mm]

A -3.58 2.28

B -3.17 2.92

C -3.57 2.35

D -3.2 2.93

E -4.04 3.1

Discussion

The goal of Aim2 was the validation of the directly measured volume of detection in an

in-vivo setup as opposed to in-vitro. The hypothesis that the LMM of the individual sources that

constitute the ensemble FP tuning statistic would result in a pseudo-ensemble tuning statistic that

resembles the actual FP statistic is a valid test of the reliability of the spatial characterization and

direct measurements reported in Chapter 3. The reason being that the LMM used the weights and

boundary, which are based on the in-vitro spatial characterization. As such, if the previously

reported direct measurements and spatial profiles were poor then the reconstruction of the

ensemble statistic would not have been possible. However, the results presented in the previous

section show that the weights and boundaries that constrained the LMM were accurate as they

yielded a pseudo-ensemble statistic almost identical to the original ensemble statistic. Omitting

the spatial weights and boundary resulted in the loss of the information conveyed by the

ensemble statistics, implying that the detection volume and the spatial profile that were measured

in-vitro also apply and are valid in-vivo.

Furthermore, the in-vivo investigation threw some light onto the nature of the aggregate

1D FP signal. Few interpretations of the FP signal exist but were not tested experimentally.

Among them is the opinion that describes the FP signal as the bulk fluorescence activity of the

neural ensemble [74], [153] within the volume of detection. The investigation presented here

simplified the time-dependent FP signal as an ensemble statistic that is indicative of the

Page 92: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

92

collective neural activity. The findings shown above provide experimental proof that the

ensemble FP statistic is a spatially weighted sum of the sources that fall within the detection

volume. This means that the position of the tip of the fiber with respect to the cells or region of

interest plays a critical role in interpreting the recorded dynamics and in drawing conclusions

about the observed neural activity.

Another opinion states that the FP signal is a measure of synchrony [154] where peaks

are indicative of the simultaneous firing of more than one neuron. The current design of this

experiment, however, doesn’t provide an optimum framework to appraise this opinion.

This chapter addressed the first limitation of the in-vitro investigation of the spatial

detection profile and showed evidence of the validity of the direct measurements in-vivo,

satisfying Aim2. The next chapter will attend to the second limitation, which is the restricted

number of optical fibers, addressing the third and last aim of this dissertation.

Page 93: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

93

CHAPTER 5

EMPIRICAL MODELING AND PREDICTION OF THE DETECTION VOLUME

The findings presented in Chapter 3 showed that the geometry of the optical fiber, as

given by the diameter/NA combination, determined the detection boundary and spatial profile

and buttressed the importance of the deliberate selection of the optical fiber geometry to match

the desired target region. The direct measurements of the detection volume and spatial detection

profile were demonstrated for four specific optical fibers. Limiting the investigation to only four

fibers stemmed from the fact the collecting direct measurements is a very arduous and time-

consuming task. However, the selection of these particular fibers was not arbitrary. It resulted

from an exhaustive literature search that showed large variability in the diameter and NA of the

optical fibers used in Fiber Photometry studies[68]–[70], [72], [73], [75], [76], [127], [131]–

[135], [137]–[139], [142], [153], [155]–[157]. Some studies also chose optical fibers similar to

those used in optogenetic stimulation [74] or used the same optical fiber to stimulate

optogenetically and record fluorescent Ca2+ dynamics with an FP system [140]. The four fibers

were therefore selected to be representative of the range of diameters (largest and smallest) and

NAs (highest and lowest) that are most commonly used. Specifically, the diameters and NAs

were selected to create four combinations that encompass large diameter/ high NA, large-

diameter/low NA, small-diameter/ high NA and small diameter/low NA.

Nonetheless, there was still a clear need to extend the direct measurements of the

detection volume to other optical fibers. Therefore, the intent of Aim3 is to generalize the

findings to arbitrary optical fibers. Since conducting direct measurements is very laborious, a

different solution had to be sought.

Page 94: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

94

Monte-Carlo Simulation

Computer-based simulations of light propagation is a long-standing technique to predict

the behavior of light as it penetrates biological tissue [158]. Many Monte-Carlo simulations

(MCS) have been proposed in conjunction with different models of light propagation to describe

the interaction of light in scattering media. Therefore, implementing an MCS for the FP system

was the solution of choice.

I was honored to be granted the opportunity to collaborate with Dr. Marwan Abdellah,

the Scientific Visualization Engineer in the Blue Brain Project1, led by Prof. Henry Markram.

This fruitful collaboration resulted in the development of a fluorescent, wavelength-dependent,

bi-directional MCS that precisely replicates the optical conditions under which FP recordings are

conducted, the Fiber Photometry Monte-Carlo Simulation (FPMCS).

The FPMCS simulation relied on the implementation reported in [159], [160]. Briefly, a

forward Monte Carlo simulation was built to shoot rays from the source, here the GFP beads,

towards the detector, which in this case was the facet of the optical fiber. A per-photon event

approach was used to capture the scattering and attenuating effects of the tissue on the

propagating photons, rather than a transmittance-based approach where the light beam

interaction with the tissue is simplified using Beer-Lambert’s law (see Chapter 2.4). The

implementation consisted of two steps: 1- calculating the probability of excitation upon

absorption, and 2- the probability of emission as a function of wavelength to accurately simulate

real-life conditions (see Chapter 2.5). The angle of acceptance of the optical fiber was accounted

for in the transmission and reception paths. Furthermore, the code was parallelized and

1 The Blue Brain Project is a research initiative located in the Brain Mind Institute at the Swiss Federal Institute of Technology (EPFL) and is dedicated to digitally reconstruct the brain anatomy of rodents by reverse-engineering

neural circuitry.

Page 95: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

95

distributed on large scale visualization clusters as an extension of [160] to simulate light

propagation in fluorescent tissue models using backward ray tracing.

Optical Properties of Neural Tissue

The FPMCS required prudent selection of the optical properties of the rodent brain

including, the attenuation coefficient µa, the scattering coefficient µs and the anisotropy factor g

(see Chapter 2.4). While µa and g are well documented the scattering coefficient µs is still under

constant investigation. The scattering coefficient, µs, is the inverse of the distance a photon may

travel without disturbance (mean free path), i.e.: it’s the distance the photon travels before being

scattered. The neural tissue of the rodent brain is a very turbid medium with various types and

sizes of scattering objects, like extracellular matrix, blood vessels, cell membranes, and

organelles and other biological structures. Thus, the scattering coefficient is technically varying

at the scale of the smallest structure, making measurements of exact scattering coefficients

impossible. Earlier studies, therefore, resorted to the measurement of absolute scattering

coefficients which report the overall scattering value for a certain type of tissue in-vitro. In-vivo

reports of scattering coeffects are harder due to the complex measuring setups, the manifold of

available calibration techniques, the assumed light propagation model and the used wavelength

[161], [162]. Therefore, the values of in-vivo scattering coefficients reported in the literature,

span a wide range and vary significantly from 8 mm-1 to 32 mm-1 [62], [63], [124], [125], [161],

[163]–[166] (Table 5-1). For instance, for blue light, µs values of 8 - 13 mm-1 correspond to

mean free paths of ~77 - 128 µm, which means a photon can travel through the neural tissue for

77 - 128µm before it interacts with an optical boundary [161]. Knowing that most structures in

neural tissue are smaller than one-tenth this distance, it is clear that these values for the scattering

coefficient are not realistic. One of the smaller reported values for the scattering coefficient at

~475 nm is the value of 21.1 mm-1 indicating a mean free path of 47 µm [63]. This value,

Page 96: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

96

although it presents the shortest reported mean free path, is still larger than the average neuronal

structure.

The goal, however, was to build an MCS that resembles a real-life, in-vivo FP recording,

where photons are transmitted and collected from neuronal tissue with structures whose

dimension range from sub- µm to 10-20µm, considering large neuronal cell bodies. Also,

considering the wavelength dependence of the scattering coefficient [54], [161], [167], means the

FPMCS needs to use a different µs for the forward path (blue, 475 nm from the fiber to the

tissue) and the return path (green, 530 nm from the fluorescent sources to the fiber). All reported

values for the scattering coefficient from the literature were harvested and summarized in (Table

5-1).

Results

The goal was to build an MCS that resembles a realistic in-vivo FP recording, where

photons are transmitted and collected from neuronal tissue with structures whose dimension

range from sub- µm to 10 - 20 µm, considering large neuronal cell bodies. Therefore, an iterative

approach was adopted, and multiple scattering coefficients were simulated. The result for each

simulation was contrasted to the direct measurements conducted previously. It was no surprise

that scattering coefficients that corresponded to mean free paths values that are 5 - 10 multiple

times larger than a single neuron resulted in extremely large detection maps, where an optical

fiber could detect fluorescence from depths farther than 700 µm as reported in [124]. After

multiple iterations, a scattering coefficient that would result in a mean free path comparable to

the size of a neuron (10 - 20 µm) was used. A value of µs = 60 mm-1 for the forward path at 473

nm and µs = 46 mm-1 for the return path at 535 nm presented realistic values and resulted in a

simulated detection profile that strongly resembled the detection maps from the direct

measurements previously shown in Chapter 3 (Figure 5-1).

Page 97: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

97

Figure 5-1 shows the MCS simulated detection maps for the four optical fibers used in

the direct measurement and the directly measured detection maps side-by-side. It is evident that

both simulated and directly measured detection maps are concordant. Furthermore, the 80%,

50% and 20% detection contours depicted in Figure 5-2A show how the shape of the simulated

and directly measured detection boundaries are in strong agreement as indicated by the

significantly positive association measured by Pearson’s correlation (r > 0.96, p < 0.001, α =

0.05) and the failure of a two-sample t-test (t-test, 0.93 < p > 0.12, α = 0.05), confirming the

validity of the developed FPMCS. Another important factor to consider, when comparing

simulated and directly measured data, is the spatial detection profile along the z-axis, as it

determines the weights or shares of the individual sources that constitute the overall FP signal.

Figure 5-2B shows that the axial detection profiles of the FPMCS and those of the direct

measurements are comparable with no significant statistical difference (t-test, p > 0.3, α = 0.05).

These results substantiate the reliability of the developed FPMCS to simulate data that match the

direct measurements and hence can be used to provide actual data in the development of the

prediction tool.

So far, the detection maps were not yet extended to optical fibers other than the one used

in the in-vitro investigation. However, the developed FPMCS paved the way to generalize the

detection maps to arbitrary optical fibers.

Page 98: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

98

A)

B)

Figure 5-1. Detection maps simulated with the developed FPMCS. A) Directly measured detection maps for the original four optical

fibers. B) FPMCS simulated detection maps the same four optical fibers. Optical fiber geometry is indicated in the lower

right corner of each detection map.

Page 99: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

99

Figure 5-2. Simulated and directly measured detection boundaries and axial detection profiles.

A) 80, 50 and 20% detection boundary for each optical fiber. Solid: average of the

directly measured boundary. Shaded: standard error of mean (s.e.m.) of the directly

measured boundary. Dashed: detection boundaries simulated with FPMCS.

Diameter/NA of each optical fiber is indicated in the lower left corner. Lateral

distance is normalized by the radius (rf) of the optical fiber. Pearson’s correlation

coefficient r400/0.5(80%) = 0.98, r400/0.5(50%) = 0.97, r400/0.5(20%) = 0.99, r400/0.22(80%)

= 0.96, r400/0.22(50%) = 0.96, r400/0.22(20%) = 0.98, r200/0.5(80%) = 0.98, r200/0.5(50%) =

0.99, r200/0.5(20%) = 0.96, r200/0.22(80%) = 0.99, r200/0.22(50%) = 0.99, r200/0.22(20%) =

0.76, p<0.000, n=24. Two-sample t-test p400/0.5(80%) = 0.34, p400/0.5(50%) = 0.16,

p400/0.5(20%) = 0.27, p400/0.22(80%) = 0.11, p400/0.22(50%) = 0.35, p400/0.22(20%) = 0.68,

p200/0.5(80%) = 0.92, p200/0.5(50%) = 0.36, p200/0.5(20%) = 0.11, p200/0.22(80%) = 0.7,

p200/0.22(50%) = 0.13, p200/0.22(20%) = 0.12, n=24. B) Axial detection profiles along

rf=0. Fiber diameter is indicated in lower left corner. Solid: average of the directly

measured boundary. Shaded: standard error of mean (s.e.m.) of the directly measured

boundary. Dashed: detection boundaries simulated with FPMCS. Two-sample t-test

p400/0.5= 0.57, p400/0.5 = 0.78, p200/0.5= 0.56, p200/0.22= 0.31, n=24.

Page 100: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

100

A Novel Prediction Tool

The beginning of this chapter discussed the importance of selecting the diameter and the

NA of the optical fiber to match the targeted brain region and to ensure its confinement to the

fiber’s detection volume. So far, only the four original optical fibers that were used for the direct

measurements were considered [150]. Numerous recent FP studies relied on optical fibers similar

to the ones characterized by Mansy et al and Pisanello et al. [124], [150], as well as other optical

fibers (Table 3-2) to record in-vivo neural correlates of social interaction [70], [132], anticipatory

regulation of ingestion [133], sensory and motor skill learning [72], [126], [157], eating-

associated orexin neurons [127], and cocaine addiction [134] from different deep brain areas.

The diameter[um]/NA combination (geometry) of an optical fiber plays a fundamental role in

determining the dimensions of the tissue from which fluorescence will be detected. This implies

that, depending on the shape and size of the target brain region, some optical fiber might be

better suited to record from this region that another fiber. However, to the best of our knowledge,

there is no current means to determine which optical fiber is optimum for a given brain region. In

other words, the detection map of a given optical fiber is not readily available for people to select

the optimum optical fiber. Reporting on the detection boundary for other optical fibers is thus

instrumental.

To fill this gap, the implemented FPMCS was used in conjunction with a machine-

learning algorithm to build a novel tool that would predict the detection boundary for any

arbitrary optical fiber, satisfying the goal of the last Aim (Aim3) of this research.

Artificial Neural Network

The deep learning algorithm was based on a feed-forward artificial neural network

(ANN). Simply stated an ANN is an empirical model that tries to learn a relationship between a

Page 101: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

101

set of inputs and their corresponding output. In machine learning terms, the inputs are usually

called featured and the output is called labels. If the labels take on discrete values from a finite

set (usually 2 or 3) then the ANN is a classification predictive model. If, however, the labels are

continuous values then it is a regression predictive model. Since the goal is to predict the shape

of the detection boundary, which is defined by a set of continuous points, a regression predictive

model was selected.

The ANN consisted of a nine-node input layer, two 64-node hidden layers, and an 11-

node output layer. The detection boundaries of the four optical fibers used in the direct

measurements and their FPMCS simulated peers composed the training dataset of the ANN.

Each node in the input layer corresponded to one out of nine features that were derived from

each optical fiber’s diameter and NA. The 11 nodes in the output layer (the labels) presented 11

points on the 80% detection contour. That is, each detection contour was described with 11

points as illustrated in Figure 5-3A. Thus, nine geometric features of the optical fiber are mapped

to 11 points that defined the fibers 80% detection contour or boundary. As mentioned earlier,

there were 24 trials for each optical fiber in addition to FPMCS simulated trial (i.e. 25 trials per

optical fiber). Since there were 4 optical fibers, then the total number of examples that were used

to train the ANN was 100. The network is built and trained in the TensorFlow platform using the

Keras API and Python programming language. The ANN was optimized by an RMSprop

optimizer with a learning rate of 0.001, which is a variant of stochastic gradient descent (SGD).

Efficient training of the ANN was achieved by making use of the patience parameter to abort the

training if performance is not improving after a set number of training iterations. Following the

convention of the machine learning community, the training dataset (100 samples) was split 80-

20 between the train and test sets. 20% of the training set was used for cross-validation.

Page 102: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

102

Results

The gold standard in machine learning is the to assess the trained ANN’s performance on

the test set, which is a measure of how well the ANN’s has learned the relationship between the

input features and output labels. A common measure for the ANN’s performance on the test set

is the difference between the actual and predicted output in the form of the mean absolute error

(MAE). Figure 5-3 demonstrates the ANN’s performance on the test set. With an MAE as low as

0.01, the predicted and actual boundaries are almost identical (r > 0.98, p < 0.001, α = 0.05),

providing evidence that the ANN has reliably learned the relationship between the geometrical

features of the optical fiber and the associated 80% detection boundary.

However, a more robust test is to assess the ANN’s ability to extrapolate to new samples.

Said differently, it is desired to measure the ANN’s predictive power on geometrical features it

has not seen in the training dataset. The importance of this lies in the fact, that some FP studies

may choose to use an optical fiber whose geometry was not included in the training data set yet

would like to be informed about the fiber’s detection boundary.

The ANN’s extrapolative prediction power was assessed in two steps and both use the

developed FPMCS. First, the detection maps of four new fibers were simulated using the

FPMCS. The diameter[µm]/NA combination of the new optical fibers was 300/39, 300/0.22,

100/39 and 100/22. The 80% detection contours were extracted from the simulated detection

maps and were considered the ground truth. Next, the ANN was used to predict the 80%

detection contour for the new fibers and the predicted contours were compared against the

simulated ones. In Figure 5-3C, it can be clearly seen that the ANN’s predicted detection

boundaries for the new (un-seen) optical fibers are very similar to the actual detection

boundaries, with an average MAE as low as 0.02 and a significant positive association as

measured by the Pearson’s correlation (r > 0.95, p < 0.0001, α = 0.05).

Page 103: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

103

The second step that was taken, to further assess ANN’s extrapolation power, relied on an

exhaustive approach that tested the asymptotic performance of the ANN on new geometric

features. The developed FPMCS was used again to simulate the ground truth detection

boundaries. This time, however, the FPMCS simulated the detection maps for a continuous range

of diameter/NA combinations, ranging from 50µm - 700µm / 0.1 - 0.8, which was a very

computationally expensive and time-consuming operation. The trained ANN was used to predict

to detection boundary of the same range of geometric feature combinations. The predicted and

simulated detection contours were then contrasted, and the MAE was calculated for every

combination. In Figure 5-3D the x-axis presents the range of NAs and the y-axis is the range of

diameters. The color-coded value of each pixel in the heatmap is indicative of the amount of

MAE between the predicted and actual 80% detection boundary. As such blue values indicate

low MAE values and high predictive power while orange and red values mean high MAE and

low prediction performance. The asymptotic performance of the ANN is summarized in Figure

5-D which informs when the proposed prediction tool will provide reliable estimates of the

detection boundary. Figure 6-5D provides compelling evidence that the prediction tool operates

with high performance for geometries that coincide with the optical fibers that are commonly

used in neuroscience applications. In neuroscience studies, valid concerns about tissue damage

usually lead to the selection of optical fibers with diameters < 500µm. Regarding the NA, the

ANN’s prediction power seems to roll off at about 0.7. However, since the largest commercially

available NA is ~0.66, with larger NAs possibly subject to fabrication limits, it is safe to

consider the ANN for optical fibers with NA’s up to the largest commercially available NA.

The ANN was developed in the scientific Python environment Spyder using the

TensorFlow/Keras API. These platforms are not commonly used in the neuroscience community

Page 104: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

104

and could be hard to navigate. Hence, the final element of this dissertation was to package the

ANN in a more common framework, like MATLAB®. For this, the Deep Learning Toolbox™ in

MATLAB® was used to import the ANN’s structure and parameters from TensorFlow/Keras and

to allow predictions to be made in a single MATLAB® script. The MATLAB® script takes the

diameter and NA of an optical fiber as input arguments and returns the predicted 80% detection

boundary associated with the fiber.

Page 105: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

105

A)

B)

C)

D)

Figure 5-3 Empirical model of the detection volume. A) ANN architecture. X1-9 are input nodes

that correspond to the geometrical features of the optical fiber. P1-11 are output nodes

that correspond to samples defining the 80% detection contour. B) ANN performance

on the test set. Diameter/NA of each optical fiber is indicated in the lower-left corner.

MAE and Pearson’s correlation coefficient are MAE400/0.5=0.008, r400/0.5=0.99;

MAE400/0.22=0.010, r400/0.22=0.98; MAE200/0.5=0.009, r200/0.5=0.99; MAE200/0.22=0.007,

r200/0.22=0.99. Lateral distance is normalized by the radius (rf) of the optical fiber. C)

ANN prediction performance using a set of four optical fiber parameters that are not

part of the training set. Diameter/NA of each optical fiber is indicated in the lower-

left corner. MAE and Pearson’s correlation coefficient are MAE300/0.39=0.02,

r300/0.39=0.97; MAE300/0.22=0.011, r300/0.22=0.98; MAE100/0.39=0.009, r100/0.39=0.97;

MAE100/0.22=0.052, r100/0.22=0.95. Lateral distance is normalized by the radius (rf) of

the optical fiber. D) Asymptotic performance of the ANN for a continuous range of

diameters and NAs. The color-coded value of each pixel in the heatmap corresponds

to the amount of MAE between the predicted and actual 80% detection boundary.

Page 106: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

106

Discussion

Fiber photometry has gained immense popularity by merit of its experimental flexibility,

ease of operation and low cost. However, unlike other methods of optical interrogation, the

nature of the recorded 1D signal and the volume of tissue from which it is detected were unclear.

The findings presented in Chapter 4 shed light on the composite nature of the FP signal

demonstrating its dependency on the spatial detection profile of the used optical fiber. Hence

deliberate selection of the optical fiber is key in interpreting the recorded signal and in ensuring

that the collected fluorescence is confined to the proclaimed region of interest. This means that a

brain region-optical fiber-matching process is necessary for meaningful interpretation of the

recorded signal. As such, a small diameter, high NA fiber may offer detection breadth while a

large diameter, low NA fiber would provide detection depth. The results presented in this chapter

aid this selection process and automate it. The bi-directional, fluorescent FPMCS developed in

collaboration with Dr. Abdellah has been the cornerstone for the success of the novel, ANN-

based prediction tool as it relied on the data generated by the FPMCS. The proposed prediction

tool serves as a powerful resource as it provides an accurate estimate of the detection boundary

associated with a given optical fiber and hence will inform the brain region - optical fiber -

matching process.

This chapter concludes the contributions of this dissertation by proposing a novel

MATLAB®-based tool that predicts the detection boundary of any optical fiber-based on its

geometrical features, the diameter and NA, and was developed using the custom-designed

FPMCS and an ANN.

Page 107: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

107

Table 5-1. Summary of reported optical properties of the rodent brain.

Animal

Model Preparation

µs

[mm-1]

µa

[cm-1] g

λ

[nm] Illumination Measurement technique Reference

Mouse

32

23

19

0.9

0.3

0.1

0.9

0.9

0.9

405

532

635

1p Simulation based on mouse brain atlas [164]

Rat

In-vitro-

whole

brain

17

13

0.5

0.9

--

--

480

530 1p Contact spatially resolved spectroscopy [162]

20

12

--

--

0.93

0.93

480

530 1p Mathematical model [165]

Rat

32

23

19

0.9

0.3

0.1

0.9

0.9

0.9

405

532

635

1p Double-integrating sphere setup, and

optical coherence tomography [168]

Mouse In-vitro,

slice 20 Ignored 0.925 920nm 2p --- [124]

Mouse In-vivo 13.3 1 0.9 ---- 2p --- [125]

--- --- 8-12 -- 0.9 650-

950 1p --- [161]

Mouse Slice 21.1 0.62 0.86 473 1p Light transmission using optical power-

meter [63]

Rat/mouse Cortical

slice 10.3/11.2 0.69 0.88 473 1p [62]

Mouse Slice,

Subcortical

16.1

16.5

14.08

--- >0.9

453

528

940

1p Light transmission using optical power-

meter [166]

Rat Cortex 17 0.25 0.9 532 1p

Double-integrating-sphere to measure

diffuse transmittance, diffuse

reflectance, and ballistic transmittance

followed by inverse adding doubling

method (IAD)

[163]

Page 108: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

108

CHAPTER 6

CONCLUSION AND FUTURE WORK

Summary

Fiber photometry is an optical method for in-vivo interrogation of neural circuit dynamics

that depends on an optical fiber to relay neural dynamics from deep brains structures. Since

optical fibers lack a focal plane, the FP collects aggregate fluorescent Ca2+ dynamics from an

excitable volume of tissue as a one-dimensional signal. The aggregate and surrogate nature of the

FP signal make its interpretation a challenge. Hence, the central aim of this dissertation was to

construe the lumped 1D FP signal by conducting a systematic characterization of the FP

modality through three main steps. The three steps were motivated when the FP’s sensitivity was

assessed and compared to classical electrophysiological multi-unit recordings by applying

mechanical stimuli of varying intensity while recording the evoked responses in the rat’s whisker

system. It was evident that the FP signal reported a neurometric curve that is strongly concordant

to the previously reported electrophysiological curve.

In the first step, the extent from which the FP collects fluorescence was quantified in

brain phantom slices and acute brain slices using GFP beads by means of direct measurements of

the detection volume. Since the collection of fluorescence occurs through the optical fiber, four

optical fibers with different diameter/NA combinations were investigated. It was shown that the

size and shape and of the detection volume pivots on the geometrical features of the optical fiber

in question, with low NA fibers allowing fluorescence to be collected from depths as far as ~300

µm from the tip of the optical fiber and large-diameter fiber permitting increased peripheral

detection along the lateral axis. The diameter/NA combination also affected the spatial detection

profile in the axial direction determining the magnitude by which each source will contribute to

the aggregate FP signal.

Page 109: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

109

Second, the spatial characterization of the detection volume was validated in-vivo in the

mouse visual cortex utilizing epi-fluorescent volumetric scanning as well as two-photon

microscopy. Optical scanning of the detection volume offered the spatial resolution, inherently

lacking in the FP system, to debrief the individual sources that constitute the aggregate FP signal

and the two-photon images verified the reliability of the sources. Compounding the sources by

applying the directly measured boundary of the detection volume and spatial profile enabled the

construction of a pseudo-FP statistic from its contributing sources in the form of a weighted sum.

In the third and final step, the quantification of the detection volume was generalized to

arbitrary optical fibers other than the four with which the investigation started, and a novel

prediction tool was delivered. This was achieved by developing two separate but interconnected

modules. First, a bi-directional, wavelength-specific and physically-plausible Monte Carlo

Simulation was developed to closely mimic the optical conditions under which the FP collects

neural dynamics. Second, an artificial neural network was trained to learn the relationship

between the geometrical features of an optical fiber and the associated shape of the detection

volume boundary. The FPMCS simulated the detection volume for optical fibers that were used

to train an artificial neural network and for new optical fibers that were needed to test the

network’s predictive power in extrapolating to un-seen geometric features. The artificial neural

network showcased strong prediction power for optical fibers with diameters less than 600 µm

and NAs between 0.15 and 0.66. This novel prediction tool was then translated to a MATLAB

based platform for ease of use and convenience.

Page 110: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

110

Future Work

Future work spans two aspects that address the main limitations of the research presented

in this dissertation. The first aspect is regarding the in-vivo validation of the detection volume. It

was shown that the overall ensemble statistic recorded by the FP system can be reconstructed

from the constituent sources. However, this ensemble statistic is a holistic measure that does not

reflect the temporal evolution of the neural activity. The reason, this approach was selected is

because the FP and the OS recordings were performed sequentially rather than simultaneously.

Hence, the temporal correspondence between both recorded signals was lost and a more holistic

measure had to be found. In the future, it’d be very insightful to compare both signals

temporally, which can be achieved if FP and OS signals are collected simultaneously from the

same location. The technical challenge will be how to combine the two optical paths whilst

minimizing interference. However, given the myriad of newly developed optics and abundant

optical tricks, this challenge might come with a possible solution.

The second aspect is related to the contributed novel prediction tool. So far, the tool

showed high performance in predicting the detection volume for optical fibers whose geometry

is close to the geometry of the optical fibers is in the training of the network. As such, having a

larger training set that is representative of more geometrical feature may lead to higher prediction

performance for optical fibers of large diameter and high NA. Furthermore, the tool was trained

on simulated detection volumes for blue excitation and green emission light only. In other words,

the tool is strictly wave-length dependent and is tailored to fluorescent sources that have a GFP-

like spectrum. In order to stay abreast with the quickly emerging spectra of state-of-the-art Ca2+

indicators (and other indicators), it would be highly advantageous to expand the prediction tool

to handle other wavelengths as well.

Page 111: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

111

Conclusion

Uncovering the neural basis of sensory integration, motor planning, cognition and

learning in the healthy and diseased states of the brain is the ultimate quest of neuroscience. It’s a

quest that has been, and still is, undertaken on multiple fronts with several investigational

approaches that vary in spatiotemporal resolution and hence provide unprecedented amounts of

information. Some approaches even combine the temporal precision of electrophysiology, with

the spatial single-cell resolution of two-photon microscopy and the coarse ensemble dynamics of

ECoG arrays in an attempt to elucidate neural circuit dynamics.

The past few decades have seen a vast expansion in new modalities that rely on light (in

the visible and invisible spectrum) to further our understanding of neural dynamics. Those tools

allow in-vivo reading from (recording or imaging) as well as writing to (modulating) neurons in

a way that simultaneously links structure and function. These agile advances in neural device

technology have been constantly reinforced by unceasing efforts to develop more robust calcium

(Ca2+) indicators, which serve as a proxy to neural activity and aim to relay information from

large ensembles with spatial and temporal resolution.

All these massive strides in microscopic technology, miniaturized optics, microbiology,

and biotechnology have certainly propelled the optical approach to the functional and structural

interrogation of neural circuits. However, technological advances are a double-sided sword

whose unfavorable side can be easily overseen given the extremely beneficial edge that offers

abundant features.

Whilst utilizing a new neural recording device or a novel imaging method to conduct

ground-breaking research, it is mandatory to Know Thy Device. In other words, new recording

and imaging methods shall not be used without a full understanding of their specifications as

Page 112: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

112

well as their limitations. This can be troublesome, given the wide array of options that are

currently available and that complicate the selection process.

Nowadays, selecting the recording or imaging method of choice relies heavily on the

scientific question and type of investigation as different methods are more suitable for certain

research endeavors than others. Furthermore, all-optical methods of in-vivo neural interrogation

come at the inevitable trade-off between cost, resolution, focality, targetability, flexibility in

behavioral paradigms and complexity. As such selecting the optimum method for an experiment

can be quite cumbersome. However, among all-optical methods, there is one that stands out for

its exceptional flexibility in paradigms of awake behavior, its ability to target superficial as well

very deep brain structures, its ease of operation, its versatility and low cost. This method is the

Fiber Photometry, which was systematically characterized in this dissertation in hope to inform

about its underlying mechanisms of action and to provide novel metrics of consideration when

utilizing this optical method for in-vivo interrogation of neural circuit dynamics.

Page 113: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

113

APPENDIX A

SYSTEM DESCRIPTION

In 2015, I was granted the outstanding opportunity to attend a Fiber Photometry

workshop at the Deisseroth lab at Stanford University. The two-and-a-half-day workshop

introduced us to the modern1 FP design reported in Gunaydin et al. 2014 and offered hands-on

experience in the assembly of the system. After a great learning experience at Stanford, I

returned to our lab to establish the new technique. I implemented a modified version of the FP

rig that allowed for simultaneous calcium recording and optogenetic excitation while

maintaining a compact and portable footprint. Since then, the FP system was used in different

experiments to recorded neural activity in our lab.

The goal of FP is to record calcium transients (dynamics) as reported by a genetically

encoded calcium indicator (GECI), like GCamP6, that is expressed in neurons of awake or

anesthetized subjects under single-photon illumination. This assumes that subjects have

undergone a surgical procedure to deliver the GCamP6 to a target region and have been

implanted with an optical fiber in the same region. In the following, the different components of

the FP system will be described in detail.

System components

The FP system components and the different input, output, light path, and system controls

are illustrated in Figure A-2A. Figure A-2B shows a simpler cartoon of the FP system and zooms

into the optical fiber - brain interface.

1 I’m calling it “modern” FP design as the fundamental theory of FP existed long before 2014 [65], [136], [169]

Page 114: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

114

Light source

Since GCaMP6 relies on a green fluorescent protein (GFP) reporter it requires to be

illuminated with blue light in the range of 450-490 nm. A Light Emitting Diode (LED) with λnom

= 470 nm is thus used to deliver the blue excitation light. A second light source, a violet LED

with λnom = 405 nm, is used as a reference light for control measurements and motion artifact

removal.

Figure A-1. GCaMP6f excitation spectrum. Adapted from [170].

Page 115: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

115

A)

B)

Figure A-2. Diagram of the Fiber Photometry system. A) Detailed system components showing control signals (red) from TDT to the

LED driver. The LEDs are the light source and generate the blue and violet lights (forward path). Green fluorescent light

from expressing neurons is detected (return path) and acquired by TDT (green). B) Enlarged view of the red box in A

showing fiber-brain interface.

Page 116: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

116

Optical path

The optical path can be split into forward and return paths. The forward-path addresses

the delivery of the excitation light (blue) from the source to the GCaMP6 expressing neurons in

the brain while the return path describes the collection of the fluorescence (emission light, green)

emitted from those neurons and its propagation to the photodetector. In the forward path, the

excitation light exits the LED and is collimated onto an excitation filter before it passes through a

dichroic mirror (DMLP425). Likewise, the reference light is collimated on an excitation filter

before it reaches the dichroic mirror. At the same dichroic mirror, the reference light is deflected

and forced to merge in the direction of the excitation light. Obviously, this dichroic mirror

transmits blue light and reflects violet. Both lights continue to propagate until they pass through

a second dichroic mirror (NFD01-532) before they are collimated onto the optical patch cable

that guides both lights to the optical fiber implanted in the brain. The lights will emanate from

the fiber and illuminate a volume of GCaMP6 expressing neurons inside the brain. When

actively firing, those neurons will emit green fluorescence which demarcates the return path. In

the return path, green light emitted by active neurons will be captured at the tip of the fiber and

guided along the patch cable back to the collimator. After the collimator, it will arrive at the

second dichroic mirror. The second dichroic mirror (NFD01-532) serves in the forward and

return paths as it transmits all wavelengths except those falling between 502-562 nm, which

correspond to green light. Once directed upwards by the dichroic mirror, the green fluorescent

light goes through an emission filter and then a convex lens. The convex lens focuses the

collected green light onto a single cell photodetector where it is transduced to an electrical signal

and sent to the data acquisition system.

Page 117: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

117

Photodetector

The photodetector is a transducer that detects light and converts it to an electric signal

that is proportional to the intensity of the detected light. Photodetectors have a wide range of

specifications that make one design more suitable for a given application. FP relies on the

measurement of minute changes in fluorescent intensity inside a highly scattering medium, like

the brain. Thus, high sensitivity and high gain are an indispensable feature of the FP

photodetector. Miniscule signals, like the one recorded by the FP, also need good shielding from

60Hz interference. While this could be remedied using signal processing techniques and filtering,

it is always better to avoid it during signal acquisition. Battery-operated photodetectors,

therefore, offer an important edge compared to their power-line operated counterparts. A battery-

operated femto-watt photodetector (Newport, Corporation, Irvine CA) that offers extremely high

gain (up to 1x1011) and can detect sub-pico-watts to 0.5 nano-watts optical signals is thus the

right choice.

System controls and data acquisition

A Tucker Davis Technology (TDT) BioAmp processor was used to control the LED

drivers for the excitation and reference lights. While most fluorescence applications use constant

illumination, the FP makes use of a very smart trick. Instead of driving the LEDs with a constant

current resulting in constant illumination, which could lead to tissue heating or photobleaching of

the fluorescent protein, it modulates the excitation and reference lights sinusoidally at different

frequencies, fex and fref respectively. The detected signal (electrical) is then demodulated at those

frequencies to extract physiologically relevant fluorescence emitted due to excitation light at fex

and fluorescence due to motion artifact or other non-physiologically relevant processes occurring

at fref. The modulation frequency can take on any value but must satisfy the following

constraints:

Page 118: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

118

• It must meet the Nyquist criterion for GCamP6 signals, i.e.: be several times higher than

the fastest GCamP6 event, which could reach up to ~15 Hz. As such it needs to be higher

than 100 Hz.

• It must fall within the bandwidth of the used photodetector.

• It must be spectrally separated from sources of optical and electrical noise. This means it

cannot take on values of 6 0Hz and its harmonics.

Accordingly, 210 Hz was chosen as fex chosen to be fast enough to easily recover the

signal of interest, to be near the peak responsiveness of the detector, and to be 30 Hz away from

harmonics of 60 Hz. Similarly, 530 Hz was chosen for fref.

Once demodulated, the detected signal represents the raw fluorescence emitted by the

GCamP6 expressing neurons. This signal is proportional to the firing activity of these neurons

and represents the aggregate ensemble dynamics. In other words, every time one or more

neurons fire there will be an increase in the detected fluorescent GCamP6 signal. This raw

GCamP6 signal can then be manipulated to calculate the amount of change in form or dF/F or a

Z-score. Figure A-3 shows a sample trace of the GCamP6 signal recorded with the FP from the

rat’s vibrissal sensory cortex during presentation of mechanical whisker deflections.

Figure A-3. Sample Ca2+ trace recorded with FP. Red ticks indicate the presentation of the

sensory stimulus. Top: recorded fluorescent green channel showing Ca2+ transients in

response to stimuli. Bottom: recorded violet channel.

Page 119: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

119

The previous section expounded the fundamental mechanics of FP, which can be

summarized by shining blue light onto GCamP6 expressing neurons and recording their

aggregate activity that is reflected as transients in the detected green fluorescent signal. A very

natural thought that follows here is, how does this aggregate (optical) neuronal signal compare to

classical multi-unit electrophysiological recordings? The next appendix will briefly summarize

two experiments I carried out to answer this thought. These two experiments inspired the main

question of my research and serve as the main motivation behind it.

Page 120: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

120

APPENDIX B

PILOT EXPERIMENTS

Experiment 1

Single-photon fiber photometry is a very simple and economic modality to record Ca2+

dynamics as a surrogate for neural activity. However, validation was necessary for the

implementation of the nascent technique. The system was implemented according to the material

provided in a workshop at Stanford University (Deisseroth lab, 2015) [70]. The rodent whisker

system lends itself as a very apt model to corroborate the viability of the device, by virtue of its

clear somato-topical organization that allows a one to one mapping of a single vibrissa to a

specific cortical column (barrel) in the vibrissal representation of the somatosensory cortex

(vS1). This means that mechanical deflections, which present the sensory stimulus, of a

particular whisker will result in a response in a well-defined cortical area. Cortical sensory

coding of stimulus parameters has been intensely investigated and was shown that changes in

sensory stimulus strength are coded in cortical response probability [119]–[121]. These studies

used classical electrophysiological (ephys) methods to record multi-unit activity. I was curious to

know how the aggregate FP signal will compare to the multi-unit ephys recoding and whether it

will be sensitive enough to detect changes in sensory stimulus strength.

Methods

Surgical Procedure

All animal care and experimental procedures were approved by the University of Florida

Institutional Animal Care and Use Committee. Long Evan rats (n = 4) were injected with ~1 µL

of GCaMP6f (AAV1.CamKII.GCaMP6f.WPRE.SV40) and chronically implanted with an optical

fiber (diameter = 400 µm, 0.5 NA) in vS1 (AP: -3.2, ML: -5, DV: 0.6 mm). After recovery and

Page 121: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

121

viral expression (~4 weeks), animals were lightly sedated, and a single whisker was

mechanically deflected by a piezo-electric element at different velocities Figure B-1A.

Whisker deflection

The deflection waveform was carefully designed to accommodate the slow dynamics of

calcium currents and consisted of three epochs: 1. a linear ramp that elicits a distinct cortical

response aligned to the onset of the stimulus; 2. an extended hold time (6 s) which allows

capturing the slow calcium response; 3. A slow return that prevents the masking of the original

response by a stimulus-offset response Figure B-1C. The whisker was deflected at six different

velocities 1500, 1000, 500, 250, 125 and 62 °/s, that were picked to satisfy the minimum criteria

for sensory detection reported in the literature [120], [121], [171]–[176].

Data collection and analysis

The FP system delivered blue light (λ = 475nm), modulated at fex = 210 Hz, for GCamP6f

excitation and violet light (λ = 405 nm), modulated at fref = 530 Hz, for motion artifact removal.

Green fluorescence (λ = 535 nm) emitted by GCamP6f was collected by the optical fiber and

routed to a single cell femto-watt photo-detector. The photo-detector converted the optical signal

into an analog signal that is fed to a Tucker Davis Technologies (TDT) data acquisition system

for storage and analysis. The TDT system controlled the delivery of the mechanical deflections

to the whisker and guaranteed synchrony between stimulus presentation and recorded FP signal.

Custom MATLAB® scripts were used written to demodulate, low pass filter and down-

sample the recorded signal. After correcting for motion artifact, by subtracting the violet channel

from the green channel, dF/F was calculated as the Z-score of the green channel to simplify

cross-animal comparison.

Page 122: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

122

Results

The aim of this experiment was to compare the optical readout from the FP with the

previously reported ephys readout to draw a conclusion regarding how the FP signal compares to

ephys data. The Ca2+ traces recorded in response to whisker stimulation are shown in Figure B-

2A. Large transients can be seen following the stimulus presentation. Some cases did not elicit a

response, reminiscent of cortical response variability. The cortical response probability was

calculated for each of the different deflection velocities and was summarized in the optical

neurometric curve in Figure B-2B, which strongly reflects the ephys neurometric curve for the

same velocities [119]–[121]. As such it was concluded that the optical signal recorded with the

FP was sensitive enough to detect changes in the sensory stimulus parameter and reliably

encoded the velocity of the stimulus.

Furthermore, the experiment assessed the reliability of the FP signal in a longitudinal

study. Since the brain constantly changes its microstructure on short and long timescales to adapt

to the environment, long term stable recordings are thus crucial to capture the dynamics of

processes like learning and memory consolidation. The major drawback of current

electrophysiological recording modalities is the short signal lifetime. Therefore, the longevity of

the FP signal was appraised. The stability of the signal in the coding of the sensory presentation

was quantified in terms of peak amplitude and latency (Figure B-2C) for up to 200 days.

Transients were classified as stimulus-evoked responses following the criteria in [177]. In Figure

B2-D the sensory-evoked responses recorded on the first and last day are shown with significant

similarity, proving reliable signal detection after five months that potentially outperforms

penetrating microelectrodes [178].

Page 123: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

123

A)

B)

C)

D)

Figure B-1. Setup of Experiment 1. Sprague-Dawley rats (n = 4) were injected with GCamP6f in

L4 of vS1 and chronically implanted with a fiber optic. B) Diagram of the FP system.

C) Untrimmed D1 or D2 whisker was inserted in a light-weight polyimide tube

(diameter: 137µm) and deflected 3-5mm from the base with 6 velocities (Vd) under

anesthesia. D) Picture of the implanted subject. [Picture courtesy of the author]

Page 124: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

124

Figure B-2. Results of Experiment 1. A) Top: Two-minute trace of stimulus-evoked, time-locked

Ca2+ transients in response to whisker deflection. Notice the absence of some

responses, resembling cortical response probability. Bottom: Control trace. B) Left:

Overlaid single response (blue) and miss (orange) trials (KS-test, p<0.01). Right:

Representative single Ca2+ transient. C) Ca2+ transient peak amplitudes (mean ±

s.e.m.) and latency for n=4 across all recording sessions at Vd=1000 °/s. D) Average

waveforms on first and last day of recording for n=4. The shaded area represents the

standard deviation. ρ is the Pearson’s correlation coefficient

Page 125: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

125

Experiment 2

The vagus nerve, the longest nerve in the human body, is the tenth cranial nerve and

predominantly provides parasympathetic innervation to the lungs, heart, and digestive tract.

Nevertheless, vagus also comprises a significant array of sensory afferents that report different

states of the body to the central nervous system (CNS) by means of neuromodulator transmission

[179].

Different techniques, like vagotomies, electrical stimulation, and pharmacological

inactivation, were used to investigate the functional role of the vagus nerve in conveying

modulatory information from the autonomic nervous system to the brain. Many recent studies

showed the various neuromodulator mediated effects of vagus on memory consolidation [180],

[181], learning [182], increased alertness (desynchronized EEG) [183], [184], improved

cognitive functions, targeted cortical plasticity [185]–[190] and accelerated motor rehabilitation

after traumatic brain injury in animal models [185], [191]. This pronounced enhancement of

different cognitive modalities with electrical stimulation of the vagus nerve has been linked to

adequate provision of the neuromodulator Acetylcholine (Ach) [185], [192]–[194].

Nevertheless, further refinement of VNS mechanisms can lead to more accurate control

of neuromodulator release, which governs specific cognitive conditions, and thereby improve or

expedite therapy. The aim experiment is to explore the effect VNS parameters on neuronal

ensembles in the medial prefrontal cortex (mPFC) using the optical readout from the Fiber

Photometry system. Specifically,

Vagus nerve stimulation is a technique that requires mastering fabrication of miniature

stimulating cuff electrodes and a highly precise surgical implantation procedure. I had the

pleasure of receiving hands-on training in two renowned labs at the University of Texas Dallas

(Dr. Kroener and Dr. Kilgard).

Page 126: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

126

Subjects were implanted with a vagus nerve cuff, injected with GCaMP6f and implanted

with an optical fiber in mPFC simultaneously. Standard stimulation was delivered 3-4 weeks

after surgery while mPFC Ca2+ dynamics were recorded using the FP. Vagally induced effects

were quantified in mPFC by monitoring changes in the rate of Ca2+ transients when the pulse

width was varied (Figure B-3B).

Methods

Surgical methods

Wild type (WT) Long-Evan rats (n = 3, 250 - 350 g) were anesthetized with 2 - 3 %

isoflurane. About 1 µl of GCamP6f (AAV1.CamKII.GCaMP6f.WPRE.SV40) was injected in

medial prefrontal cortex (mPFC AP: +3.00, ML: 1.00,DV: -3.2 mm) and a multimodal fiber-

optic cannula (NA = 0.5, diameter = 400 µm) was be implanted approximately 100 µm above the

injection site. The bipolar stimulation cuff needs to maintain smooth edges and be fully insulated

on the outside. The electrodes (bare wires on the inside) must provide 1 - 10 kΩ impedance.

Once built, tested for impedance and sterilized, the cuff is ready for implantation. Animals were

implanted with a vagus nerve cuff (built in house, ~3 kΩ impedance) during the same procedure.

Vagus nerve stimulation connector and fiber cannula were secured to the skull using dental

cement and four bone screws. Animals were then left to recover and express the virus for 3

weeks. Afterward, they were tethered to the optical patch cord and the VNS stimulator while

lightly sedated. Recording and stimulation commenced only after the animals regained

consciousness and were fully ambulatory in their home cage.

Vagus nerve cuff implantation

Blunt dissection of the cervical section of the vagus nerve mandates minimal damage to

the surrounding neck muscles and tissue to warrant apt animal recovery. Separation of the vagus

nerve from the carotid artery is followed by removal of all layers of thin connective tissue that

Page 127: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

127

constitutes the sheath around vagus. This step is crucial as it guarantees confined electric

stimulation to the nerve of interest and saves the surrounding tissue and other fine sensory nerves

from being influenced, which may pose a confounding factor. Once ~4 - 5 mm of the nerve is

fully exposed, the cuff is passed through a subcutaneous tunnel from the head down to the neck.

The exposed nerve is placed inside the cuff and the integrity of the procedure is tested by

observing a cessation of breathing (COB) or a ~15 % drop in oxygenation upon delivering of a

test stimulus (0.8 mA, 100 µs, 30 Hz, 10 s). It is important at this stage to watch for any adverse

effect to the stimulation, like jaw twitching, whisker deflection, head jerking, which are

indicative of having other fine nerves or tissue caught inside the cuff. In this case, the cuff must

be removed to allow for further cleaning of the nerve (Figure B-3A).

Vagus nerve stimulation

An isolated current stimulator (A-M systems model 4100) was programmed to deliver

standard VNS stimulation parameters (trains of 100 µs pulses at 30 Hz for 500 ms, inter-train-

interval of 5 s, for 2 minutes). The pulse width was varied to 300 and 500 µs to explore pulse

width effects (Figure B3-B).

Data collection and analysis

The FP system delivered blue light (λ = 475 nm), modulated at fex = 210 Hz, for

GCamP6f excitation and violet light (λ = 405 nm), modulated at fref = 530 Hz, for motion artifact

removal. Green fluorescence (λ = 535 nm) emitted by GCamP6f was collected by the optical

fiber and routed to a single cell femto-watt photo-detector. The photo-detector converted the

optical signal into an analog signal that is fed to a Tucker Davis Technologies (TDT) data

acquisition system for storage and analysis. The TDT system and the VNS stimulator are

synched via a ‘gating’ signal generated by TDT. Custom MATLAB® scripts were written to

demodulate, low pass filter and down-sample the recorded signal. After correcting for motion

Page 128: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

128

artifact, by subtracting the violet channel from the green, dF/F was calculated as the Z-score of

the green channel to simplify cross-animal comparison.

Results

The aim of this experiment was to explore the neuromodulatory effects on the excitatory

neurons in mPFC under electrical stimulation of the vagus nerve. The effect was quantified in

terms of changes in Ca2+ transients or spike rate. This means the number of Ca2+ peaks were

counted per unit time and compared under different pulse width stimulation. Figure B-3C shows

the recorded Ca2+ traces during epochs of stimulation (orange shading) and epoch where

stimulation was withheld. Visually inspecting the responses, it can be seen that the 300 and 500

µs stimulation pulses caused an increase in spiking rate. The increase in spiking rate was

quantified and is summarized in Figure B-3D, which demonstrates the significant increase in

excitability with 300 µs stimulation. A pulse width of 500 µs also resulted in a higher spiking

rate but not as significant as the 300 µs. It should be noted, that the inter-stimulus-trial of

2minutes may have been too short, possibly not allowing the system to return to baseline. An

observation that could explain the increased spiking rate during the no-stim epochs of the 300

and 500 µs stimulation compared to the 100 µs pulse width stimulation.

The results of this experiment were exciting as they provide evidence that the FP system

is capable of detecting slight changes in neuronal excitability due to modulatory effects mediated

by the electrical stimulation of the vagus nerve.

.

Page 129: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

129

Figure B-3 Methods and results of the experiment. A) Surgical procedure. Left: The exposed

vagus nerve. Inset: picture of the fabricated and tested cuff with stimulation

connector. Right: Process of placing the nerve inside the cuff. B) Left: Picture of

vagus nerve cuffed subject tethered to VNS stimulator and FP rig. Head-cap shows

VNS connector and fiber cannula cemented to skull. Right: VNS pattern as described

in [72]. Two-minute stimulation epochs that consisted of 500msec trains of 0.8mA,

30Hz pulses every 5seconds. C) Sample Ca2+ recording from a subject while being

stimulated with 100µs pulse width (top trace), 300µs (middle) and 500µs (bottom).

Asterisks indicate Ca2+ transients (spikes). Orange shading present stimulation

epochs, white intervals are no-stimulation epochs. D) Bar graph illustrating vagally

induced increase in Ca2+ spike rate (orange), compared to no-stimulation epochs

(blue). (t-test, ** p < 0.005; *** p < 0.0005, n = 3) [Pictures courtesy of the author]

Page 130: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

130

APPENDIX C

SUPPLEMENTARY MATERIAL

Mathematical Formulation of the Tuning Statistic and the LMM

The calcium trace C(t) (recorded with FP or extracted from OS data) is chopped into 8s

snippets y(t) that are aligned with the onset of the visual stimulus, θ. Eight visual stimuli were

presented randomly in 10 repeated trials.

1. Feature extraction from time series y(t) after the presentation of the stimulus

The following was extracted from every time signal 𝑦𝑖(𝑡) (snippet):

𝜏𝑖 = 𝑡 ∶ 𝑦𝑖(𝑡) = max(𝑦𝑖 (𝑡))

𝐴𝑖 = 𝑦𝑖(𝑡)𝛿(𝑡 − 𝜏𝑖)

𝜎𝑖 = √∑ |𝑦(𝑡) − |2𝑁

𝑡=1

𝑁 − 1

Where i is the trial number and 𝑖 ∈ (1,2, … ,10), N is the number of samples in the

snippet and is the temporal average of the snippet.

𝑖 = [𝐴𝑖

𝜏𝑖

𝜎𝑖

] , where 𝑖, is the feature vector associated with every trial i

2. Decision rule

From a signal detection theory perspective, the problem can be formulated as a binary

hypothesis test in which the null hypothesis, 𝐻𝑜 , expresses the observation when no stimulus is

presented, noise, n(t). The alternative hypothesis, 𝐻1, is when the observation contains a

stimulus-evoked response, ver(t).

𝐻𝑜: 𝑦𝑖(𝑡) = 𝑛(𝑡) , 𝑦𝑖(𝑡) ~ 𝒩(0, 𝜎𝑖)

𝐻1: 𝑦𝑖 (𝑡) = 𝑣𝑒𝑟(𝑡) + 𝑛(𝑡) , 𝑦𝑖(𝑡)~𝒩(𝑣𝑒𝑟(𝑡), 𝜎𝑖)

However, since we apply the hypothesis test only at time 𝜏𝑖 , then

𝑦𝑖(𝑡)~𝒩(𝑣𝑒𝑟(𝑡), 𝜎𝑖) and 𝐻1 can be written as: 𝐻1: 𝑦𝑖(𝑡) = 𝑣𝑒𝑟(𝑡) + 𝑛(𝑡) , 𝑦𝑖(𝑡)~𝒩(𝐴𝑖, 𝜎𝑖)

Page 131: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

131

As such the discriminability index d’ (also known as sensitivity or index of separation)

will evaluate to 𝑑′ = 𝐴𝑖

√1

2𝜎𝑖

2

The log-likelihood ratio LLR is a robust strategy for selecting the position of the

detection threshold. Taking the log of the ratio of the conditional probabilities 𝑃𝑦|𝐻0(𝑦|𝐻0) and

𝑃𝑦|𝐻1(𝑦|𝐻1) will result in the log-likelihood ratio LLR, which defines the detection threshold, γ.

𝐿𝐿𝑅(𝑦𝑖 = 𝐴𝑖) = 1

2(

𝐴𝑖

𝜎𝑖)

2

≥ γ (C − 1)

In the case of the ideal observer, the detection threshold is formally the ratio of the priors

𝑃0(𝐻0)

𝑃1(𝐻1), equally penalizing missed events and false positives. However, the LLR is also related to

d’ through the parameter c, which is the distance between the actual threshold and the threshold

of the ideal observer.

𝐿𝐿𝑅(𝑦𝑖 = 𝐴𝑖) = 𝑐 𝑑′ (C - 2)

It’s become custom for the value of c to be chosen empirically and with the aid of visual

inspection [195], [196]. As such, the decision rule 𝜑(𝑥𝑖), decides in favor of one hypothesis over

the other based on 1-the detection threshold derived from combining Equation C-1 and C-2, and

2- the value of 𝜏𝑖, which is derived from the physiological characteristics of the visually evoked

response [107], [144], [144], [149], [177], [197]–[199].

𝜑(𝑥𝑖) = 1,

𝐴𝑖

𝜎𝑖≥ 2 𝑎𝑛𝑑 𝜏𝑖 ∈ [0.02,2]𝑠

0, 𝑜𝑡ℎ𝑒𝑟𝑤𝑖𝑠𝑒

3. Tuning curve derivation:

𝑇∙(𝜃𝑘) = ∑ 𝜑(𝑥𝑖)

𝑖 ∈ 𝛤

, 𝛤 = 𝑚: 𝜃𝑚 = 𝜃𝑘

𝜃∙ ∶ 1,2, … ,8 → 0°, 45°, … 315° and 𝛤 is the set of trials with trial number m such that

the orientation of the presented stimulus 𝜃𝑚 is equal 𝜃𝑘 .

Page 132: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

132

4. The LMM:

𝑇𝐹𝑃(𝜃𝑘) ≈ 𝑂𝑆(𝜃𝑘)

𝑂𝑆(𝜃𝑘) = ∑

𝐷

𝑑=1

∑ 𝑤𝑛,𝑑(𝑥, 𝑦)𝑇𝑛,𝑑(𝜃𝑘)

𝑁

𝑛=1

Page 133: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

133

Figure C-1. Surgical procedure and evaluation. Monitoring of the viral expression. The top row

corresponds to subject 1 and bottom to subject 2. Left to right: V1 targeting,

placement of optically clear cranial window and head plate. Expression spread 1

week after the day of injection. Expression spread 3 weeks after the day of injection.

[Pictures courtesy of author and Rebeca Castro.]

Page 134: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

134

Figure C-2. Quantification of the visually evoked responses (VER). Quantification of the peak

amplitude and latency of the recorded VERs. Each row represents the recording

session from one of the four anatomical locations (A-E). The left column shows

latency and peak amplitude for VERs recorded with OS. Middle column shows

latency and peak amplitude for VERs recorded with FP. Right column compares the

latency of FP and OS VERs.

Page 135: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

135

LIST OF REFERENCES

[1] R. Hooke, Micrographia. 1665.

[2] A. Chvátal, “Discovering the Structure of Nerve Tissue: Part 2: Gabriel Valentin, Robert

Remak, and Jan Evangelista Purkyně,” J. Hist. Neurosci., vol. 24, no. 4, pp. 326–351,

Oct. 2015.

[3] B. J. Ford, “Enlightening Neuroscience: Microscopes and Microscopy in the Eighteenth

Century,” in Brain, Mind and Medicine: Essays in Eighteenth-Century Neuroscience, H.

Whitaker, C. U. M. Smith, and S. Finger, Eds. Boston, MA: Springer US, 2007, pp. 29–

41.

[4] G. Valentin, “Über den Verlauf und die letzten Enden der Nerven,” Nova Acta Phys-Med

Acad Leopoldina (Breslau), 1836.

[5] “History of Neuroscience.” [Online]. Available:

https://faculty.washington.edu/chudler/hist.html. [Accessed: 05-Jun-2019].

[6] F. Barnard, “Micrographia,” 2008.

[7] F. López-Muñoz, J. Boya, and C. Alamo, “Neuron theory, the cornerstone of

neuroscience, on the centenary of the Nobel Prize award to Santiago Ramón y Cajal,”

Brain Res. Bull., vol. 70, no. 4, pp. 391–405, Oct. 2006.

[8] J. Bernstein, “Über den zeitlichen Verlauf der negativen Schwankung des

Nervenstroms.,” 1868.

[9] T. Lenoir, “Models and Instruments in the Development of Electrophysiology, 1845-

1912,” Hist. Stud. Phys. Biol. Sci., vol. 17, no. 1, pp. 1–54, 1986.

[10] J. C. Eccles, R. M. Eccles, and A. Lundberg, “The convergence of monosynaptic

excitatory afferents on to many different species of alpha motoneurones,” J. Physiol., vol.

137, no. 1, pp. 22–50, 1957.

[11] K. M. Szostak, L. Grand, and T. G. Constandinou, “Neural Interfaces for Intracortical

Recording: Requirements, Fabrication Methods, and Characteristics,” Front. Neurosci.,

vol. 11, Dec. 2017.

[12] E. Marg and J. E. Adams, “Indwelling multiple micro-electrodes in the brain,”

Electroencephalogr. Clin. Neurophysiol., vol. 23, no. 3, pp. 277–280, Sep. 1967.

[13] O. Heimstädt, “Das Fluoreszenzmikroskop,” Z. Für Wiss. Mikrosk., vol. Bad 28, pp. 330–

337, 1911.

[14] “Milestones timeline: Nature Milestones in Light Microscopy.” [Online]. Available:

https://www.nature.com/milestones/milelight/timeline.html. [Accessed: 25-Dec-2018].

Page 136: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

136

[15] D. Magde, E. Elson, and W. W. Webb, “Thermodynamic Fluctuations in a Reacting

System---Measurement by Fluorescence Correlation Spectroscopy,” Phys. Rev. Lett., vol.

29, no. 11, pp. 705–708, Sep. 1972.

[16] D. M. Egelman and P. Read Montague, “Calcium Dynamics in the Extracellular Space of

Mammalian Neural Tissue,” Biophys. J., vol. 76, no. 4, pp. 1856–1867, Apr. 1999.

[17] A. Brading, E. Bülbring, and T. Tomita, “The effect of sodium and calcium on the action

potential of the smooth muscle of the guinea-pig taenia coli,” J. Physiol., vol. 200, no. 3,

pp. 637–654, 1969.

[18] D. Geduldig and D. Junge, “Sodium and calcium components of action potentials in

Aplysia giant neurone,” J. Physiol., vol. 199, no. 2, pp. 347–365, 1968.

[19] S. Ebashi and M. Endo, “Calcium and muscle contraction,” Prog. Biophys. Mol. Biol.,

vol. 18, pp. 123–183, Jan. 1968.

[20] H. Reuter, “The dependence of slow inward current in Purkinje fibres on the extracellular

calcium-concentration,” J. Physiol., vol. 192, no. 2, pp. 479–492, 1967.

[21] B. Katz and R. Miledi, “The timing of calcium action during neuromuscular

transmission,” J. Physiol., vol. 189, no. 3, pp. 535–544, 1967.

[22] F. A. Dodge and R. Rahamimoff, “Co-operative action of calcium ions in transmitter

release at the neuromuscular junction,” J. Physiol., vol. 193, no. 2, pp. 419–432, 1967.

[23] R. Miledi and C. R. Slater, “The action of calcium on neuronal synapses in the squid,” J.

Physiol., vol. 184, no. 2, pp. 473–498, 1966.

[24] H. C. Lüttgau, “The action of calcium ions on potassium contractures of single muscle

fibres,” J. Physiol., vol. 168, no. 3, pp. 679–697, 1963.

[25] I. Tasaki and M. Shimamura, “Further Observations on Resting and Action Potential of

Intracellularly Perfused Squid Axon,” Proc. Natl. Acad. Sci., vol. 48, no. 9, pp. 1571–

1577, Sep. 1962.

[26] B. Frankenhaeuser and A. L. Hodgkin, “The action of calcium on the electrical properties

of squid axons,” J. Physiol., vol. 137, no. 2, pp. 218–244, 1957.

[27] A. L. Hodgkin and A. F. Huxley, “A quantitative description of membrane current and its

application to conduction and excitation in nerve,” J. Physiol., vol. 117, no. 4, pp. 500–

544, 1952.

[28] G. Grynkiewicz, M. Poenie, and R. Y. Tsien, “A new generation of Ca2+ indicators with

greatly improved fluorescence properties.,” J. Biol. Chem., vol. 260, no. 6, pp. 3440–

3450, Mar. 1985.

Page 137: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

137

[29] A. Minta, J. P. Kao, and R. Y. Tsien, “Fluorescent indicators for cytosolic calcium based

on rhodamine and fluorescein chromophores.,” J. Biol. Chem., vol. 264, no. 14, pp.

8171–8178, May 1989.

[30] D. A. Williams, K. E. Fogarty, R. Y. Tsien, and F. S. Fay, “Calcium gradients in single

smooth muscle cells revealed by the digital imaging microscope using Fura-2,” Nature,

vol. 318, no. 6046, pp. 558–561, Dec. 1985.

[31] R. Y. Tsien, “A non-disruptive technique for loading calcium buffers and indicators into

cells,” Nature, vol. 290, no. 5806, pp. 527–528, Apr. 1981.

[32] T. Badea, J. Goldberg, B. Mao, and R. Yuste, “Calcium imaging of epileptiform events

with single-cell resolution,” J. Neurobiol., vol. 48, no. 3, pp. 215–227, 2001.

[33] H. Hirase, L. Qian, P. Barthó, and G. Buzsáki, “Calcium Dynamics of Cortical Astrocytic

Networks In Vivo,” PLOS Biol., vol. 2, no. 4, p. e96, Apr. 2004.

[34] G.-F. Tian et al., “An astrocytic basis of epilepsy,” Nat. Med., vol. 11, no. 9, p. 973, Sep.

2005.

[35] X. Wang et al., “Astrocytic Ca 2+ signaling evoked by sensory stimulation in vivo,” Nat.

Neurosci., vol. 9, no. 6, p. 816, Jun. 2006.

[36] L. E. Enright, S. Zhang, and T. H. Murphy, “Fine Mapping of the Spatial Relationship

between Acute Ischemia and Dendritic Structure Indicates Selective Vulnerability of

Layer V Neuron Dendritic Tufts within Single Neurons in Vivo,” J. Cereb. Blood Flow

Metab., vol. 27, no. 6, pp. 1185–1200, Jun. 2007.

[37] J. Koenigsknecht-Talboo et al., “Rapid Microglial Response Around Amyloid Pathology

after Systemic Anti-Aβ Antibody Administration in PDAPP Mice,” J. Neurosci., vol. 28,

no. 52, pp. 14156–14164, Dec. 2008.

[38] S. D. Atkin et al., “Transgenic mice expressing a cameleon fluorescent Ca2+ indicator in

astrocytes and Schwann cells allow study of glial cell Ca2+ signals in situ and in vivo,” J.

Neurosci. Methods, vol. 181, no. 2, pp. 212–226, Jul. 2009.

[39] K. V. Kuchibhotla, C. R. Lattarulo, B. T. Hyman, and B. J. Bacskai, “Synchronous

Hyperactivity and Intercellular Calcium Waves in Astrocytes in Alzheimer Mice,”

Science, vol. 323, no. 5918, pp. 1211–1215, Feb. 2009.

[40] F. Wong et al., “Axonal and neuromuscular synaptic phenotypes in WldS, SOD1G93A

and ostes mutant mice identified by fiber-optic confocal microendoscopy,” Mol. Cell.

Neurosci., vol. 42, no. 4, pp. 296–307, Nov. 2009.

[41] N. D. Wong et al., “Thoracic Aortic Calcium Versus Coronary Artery Calcium for the

Prediction of Coronary Heart Disease and Cardiovascular Disease Events,” JACC

Cardiovasc. Imaging, vol. 2, no. 3, pp. 319–326, Mar. 2009.

Page 138: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

138

[42] T. Komiyama et al., “Learning-related fine-scale specificity imaged in motor cortex

circuits of behaving mice,” Nature, vol. 464, no. 7292, pp. 1182–1186, Apr. 2010.

[43] D. H. O’Connor, S. P. Peron, D. Huber, and K. Svoboda, “Neural Activity in Barrel

Cortex Underlying Vibrissa-Based Object Localization in Mice,” Neuron, vol. 67, no. 6,

pp. 1048–1061, Sep. 2010.

[44] J. T. Russell, “Imaging calcium signals in vivo: a powerful tool in physiology and

pharmacology,” Br. J. Pharmacol., vol. 163, no. 8, pp. 1605–1625, Aug. 2011.

[45] J. L. Chen, O. A. Pfäffli, F. F. Voigt, D. J. Margolis, and F. Helmchen, “Online

correction of licking-induced brain motion during two-photon imaging with a tunable

lens: Online correction of licking-induced brain motion,” J. Physiol., vol. 591, no. 19, pp.

4689–4698, Oct. 2013.

[46] M. D. Al-Amri, M. El-Gomati, and M. S. Zubairy, Eds., Optics in Our Time. Springer

International Publishing, 2016.

[47] R. DESCARTES, LA DIOPTRIQUE. 1637.

[48] C. Huygens, Treatise On Light. 1690.

[49] I. Newton, Opticks: or a treatise of the reflections, refractions, inflections and colours of

light. 1704.

[50] J. C. Maxwell, A Treatise on Electricity and Magnetism. .

[51] A. Einstein, “Über einen die Erzeugung und Verwandlung des Lichtes betreffenden

heuristischen Gesichtspunkt,” Ann. Phys., vol. 322, no. 6, pp. 132–148, 1905.

[52] C. J. Davisson and L. H. Germer, “Reflection of Electrons by a Crystal of Nickel,” Proc.

Natl. Acad. Sci., vol. 14, no. 4, pp. 317–322, Apr. 1928.

[53] “Illumination Fundamentals.” 2000.

[54] The Handbook of Optics. 1995.

[55] A. Ryer, Light Measurement Handbook. International Light Inc., 1998.

[56] L. Grosenick, J. H. Marshel, and K. Deisseroth, “Closed-Loop and Activity-Guided

Optogenetic Control,” Neuron, vol. 86, no. 1, pp. 106–139, Apr. 2015.

[57] O. Yizhar, L. E. Fenno, T. J. Davidson, M. Mogri, and K. Deisseroth, “Optogenetics in

Neural Systems,” Neuron, vol. 71, no. 1, pp. 9–34, Jul. 2011.

[58] C. K. Kim, A. Adhikari, and K. Deisseroth, “Integration of optogenetics with

complementary methodologies in systems neuroscience,” Nat. Rev. Neurosci., vol. 18,

no. 4, pp. 222–235, Apr. 2017.

Page 139: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

139

[59] M. Häusser, “Optogenetics: the age of light,” Nat. Methods, vol. 11, no. 10, pp. 1012–

1014, Oct. 2014.

[60] V. Emiliani, A. E. Cohen, K. Deisseroth, and M. Häusser, “All-Optical Interrogation of

Neural Circuits,” J. Neurosci., vol. 35, no. 41, pp. 13917–13926, Oct. 2015.

[61] M. L. Rein and J. M. Deussing, “The optogenetic (r)evolution,” Mol. Genet. Genomics,

vol. 287, no. 2, pp. 95–109, Feb. 2012.

[62] A. M. Aravanis et al., “An optical neural interface: in vivo control of rodent motor cortex

with integrated fiberoptic and optogenetic technology,” J. Neural Eng., vol. 4, no. 3, p.

S143, 2007.

[63] G. Yona, N. Meitav, I. Kahn, and S. Shoham, “Realistic Numerical and Analytical

Modeling of Light Scattering in Brain Tissue for Optogenetic Applications,” eNeuro, vol.

3, no. 1, Feb. 2016.

[64] H. Dana et al., “High-performance GFP-based calcium indicators for imaging activity in

neuronal populations and microcompartments,” bioRxiv, p. 434589, Oct. 2018.

[65] H. Lütcke et al., “Optical recording of neuronal activity with a genetically-encoded

calcium indicator in anesthetized and freely moving mice,” Front. Neural Circuits, vol. 4,

2010.

[66] C. Grienberger and A. Konnerth, “Imaging Calcium in Neurons,” Neuron, vol. 73, no. 5,

pp. 862–885, Mar. 2012.

[67] G. Cui et al., “Concurrent activation of striatal direct and indirect pathways during action

initiation,” Nature, vol. 494, no. 7436, pp. 238–242, Feb. 2013.

[68] H. Adelsberger, A. Zainos, M. Alvarez, R. Romo, and A. Konnerth, “Local domains of

motor cortical activity revealed by fiber-optic calcium recordings in behaving nonhuman

primates,” Proc. Natl. Acad. Sci., vol. 111, no. 1, pp. 463–468, Jan. 2014.

[69] G. Cui et al., “Deep brain optical measurements of cell type–specific neural activity in

behaving mice,” Nat. Protoc., vol. 9, no. 6, pp. 1213–1228, Jun. 2014.

[70] L. A. Gunaydin et al., “Natural Neural Projection Dynamics Underlying Social

Behavior,” Cell, vol. 157, no. 7, pp. 1535–1551, Jun. 2014.

[71] F. Fuhrmann et al., “Locomotion, Theta Oscillations, and the Speed-Correlated Firing of

Hippocampal Neurons Are Controlled by a Medial Septal Glutamatergic Circuit,”

Neuron, vol. 86, no. 5, pp. 1253–1264, Jun. 2015.

[72] K. Schulz et al., “Simultaneous BOLD fMRI and fiber-optic calcium recording in rat

neocortex,” Nat. Methods, vol. 9, no. 6, pp. 597–602, Jun. 2012.

Page 140: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

140

[73] J. Muir et al., “In Vivo Fiber Photometry Reveals Signature of Future Stress

Susceptibility in Nucleus Accumbens,” Neuropsychopharmacology, vol. 43, no. 2, pp.

255–263, Jan. 2018.

[74] C. A. Siciliano and K. M. Tye, “Leveraging calcium imaging to illuminate circuit

dysfunction in addiction,” Alcohol, vol. 74, pp. 47–63, Feb. 2019.

[75] N. A. Steinmetz et al., “Aberrant Cortical Activity in Multiple GCaMP6-Expressing

Transgenic Mouse Lines,” eNeuro, vol. 4, no. 5, p. ENEURO.0207-17.2017, Sep. 2017.

[76] W. E. Allen et al., “Global Representations of Goal-Directed Behavior in Distinct Cell

Types of Mouse Neocortex,” Neuron, vol. 94, no. 4, pp. 891-907.e6, May 2017.

[77] M. Sato, Y. Motegi, S. Yagi, K. Gengyo-Ando, M. Ohkura, and J. Nakai, “Fast varifocal

two-photon microendoscope for imaging neuronal activity in the deep brain,” Biomed.

Opt. Express, vol. 8, no. 9, pp. 4049–4060, Sep. 2017.

[78] D. Barson, A. S. Hamodi, G. Lur, J. A. Cardin, M. Crair, and M. J. Higley, “Multiscale

Optical Imaging of Cortical Activity in Mouse,” in Optics in the Life Sciences Congress

(2017), paper JTu4A.13, 2017, p. JTu4A.13.

[79] T. A. Zolnik et al., “All-optical functional synaptic connectivity mapping in acute brain

slices using CaMPARI,” J. Physiol., p. n/a-n/a, Nov. 2016.

[80] M. J. Siniscalchi, V. Phoumthipphavong, F. Ali, M. Lozano, and A. C. Kwan, “Fast and

slow transitions in frontal ensemble activity during flexible sensorimotor behavior,” Nat.

Neurosci., 2016.

[81] S. R. Schultz, C. Copeland, A. Foust, P. Quicke, and R. Schuck, “Advances in two

photon scanning and scanless microscopic technologies for functional neural circuit

imaging,” bioRxiv, p. 036632, Jan. 2016.

[82] T. H. Murphy et al., “High-throughput automated home-cage mesoscopic functional

imaging of mouse cortex,” Nat. Commun., vol. 7, p. 11611, Jun. 2016.

[83] W. A. Liberti Iii et al., “Unstable neurons underlie a stable learned behavior,” Nat.

Neurosci., vol. 19, no. 12, pp. 1665–1671, Dec. 2016.

[84] N. Li, K. Daie, K. Svoboda, and S. Druckmann, “Robust neuronal dynamics in premotor

cortex during motor planning,” Nature, vol. 532, no. 7600, pp. 459–464, Apr. 2016.

[85] M. W. Howe and D. A. Dombeck, “Rapid signalling in distinct dopaminergic axons

during locomotion and reward,” Nature, vol. 535, no. 7613, pp. 505–510, Jul. 2016.

[86] T. Deneux et al., “Accurate spike estimation from noisy calcium signals for ultrafast

three-dimensional imaging of large neuronal populations in vivo,” Nat. Commun., vol. 7,

p. 12190, Jul. 2016.

Page 141: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

141

[87] S. D. Antic, R. M. Empson, and T. Knopfel, “Voltage imaging to understand connections

and functions of neuronal circuits,” J. Neurophysiol., p. jn.00226.2016, Apr. 2016.

[88] V. Alvarez, “Clues on the coding of reward cues by the nucleus accumbens,” 2016.

[Online]. Available: http://www.pnas.org/content/113/10/2560.full. [Accessed: 19-Jan-

2018].

[89] A. M. Packer, L. E. Russell, H. W. P. Dalgleish, and M. Häusser, “Simultaneous all-

optical manipulation and recording of neural circuit activity with cellular resolution in

vivo,” Nat. Methods, vol. 12, no. 2, pp. 140–146, Feb. 2015.

[90] E. J. O. Hamel, B. F. Grewe, J. G. Parker, and M. J. Schnitzer, “Cellular Level Brain

Imaging in Behaving Mammals: An Engineering Approach,” Neuron, vol. 86, no. 1, pp.

140–159, Apr. 2015.

[91] V. Szabo, C. Ventalon, V. De Sars, J. Bradley, and V. Emiliani, “Spatially Selective

Holographic Photoactivation and Functional Fluorescence Imaging in Freely Behaving

Mice with a Fiberscope,” Neuron, vol. 84, no. 6, pp. 1157–1169, Dec. 2014.

[92] B. A. Simms and G. W. Zamponi, “Neuronal Voltage-Gated Calcium Channels:

Structure, Function, and Dysfunction,” Neuron, vol. 82, no. 1, pp. 24–45, Apr. 2014.

[93] J. P. Rickgauer, K. Deisseroth, and D. W. Tank, “Simultaneous cellular-resolution optical

perturbation and imaging of place cell firing fields,” Nat. Neurosci., vol. 17, no. 12, pp.

1816–1824, Dec. 2014.

[94] H. Ma et al., “Wide-field in vivo neocortical calcium dye imaging using a convection-

enhanced loading technique combined with simultaneous multiwavelength imaging of

voltage-sensitive dyes and hemodynamic signals,” Neurophotonics, vol. 1, no. 1, pp.

015003–015003, 2014.

[95] D. S. Greenberg, D. J. Wallace, and J. N. D. Kerr, “Imaging Neuronal Population

Activity in Awake and Anesthetized Rodents,” Cold Spring Harb. Protoc., vol. 2014, no.

9, p. pdb.top083535, Sep. 2014.

[96] C. Koelbl, M. Helmstaedter, J. Lübke, and D. Feldmeyer, “A Barrel-Related Interneuron

in Layer 4 of Rat Somatosensory Cortex with a High Intrabarrel Connectivity,” Cereb.

Cortex, p. bht263, Sep. 2013.

[97] A. J. Sadovsky, P. B. Kruskal, J. M. Kimmel, J. Ostmeyer, F. B. Neubauer, and J. N.

MacLean, “Heuristically optimal path scanning for high-speed multiphoton circuit

imaging,” J. Neurophysiol., vol. 106, no. 3, pp. 1591–1598, Sep. 2011.

[98] A. Holtmaat et al., “Long-term, high-resolution imaging in the mouse neocortex through

a chronic cranial window,” Nat. Protoc., vol. 4, no. 8, pp. 1128–1144, Jul. 2009.

Page 142: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

142

[99] J. N. D. Kerr, D. Greenberg, and F. Helmchen, “Imaging input and output of neocortical

networks in vivo,” Proc. Natl. Acad. Sci. U. S. A., vol. 102, no. 39, pp. 14063–14068,

Sep. 2005.

[100] G. Miesenböck and I. G. Kevrekidis, “Optical Imaging and Control of Genetically

Designated Neurons in Functioning Circuits,” Annu. Rev. Neurosci., vol. 28, no. 1, pp.

533–563, 2005.

[101] H. H. Yang, F. St-Pierre, X. Sun, X. Ding, M. Z. Lin, and T. R. Clandinin, “Subcellular

Imaging of Voltage and Calcium Signals Reveals Neural Processing In Vivo,” Cell, vol.

166, no. 1, pp. 245–257, Jun. 2016.

[102] N. C. Flytzanis et al., “Archaerhodopsin variants with enhanced voltage-sensitive

fluorescence in mammalian and Caenorhabditis elegans neurons,” Nat. Commun., vol. 5,

p. 4894, Sep. 2014.

[103] Y. Mishina, H. Mutoh, and T. Knöpfel, “Transfer of Kv3.1 Voltage Sensor Features to

the Isolated Ci-VSP Voltage-Sensing Domain,” Biophys. J., vol. 103, no. 4, pp. 669–676,

Aug. 2012.

[104] L. Jin, Z. Han, J. Platisa, J. R. A. Wooltorton, L. B. Cohen, and V. A. Pieribone, “Single

Action Potentials and Subthreshold Electrical Events Imaged in Neurons with a

Fluorescent Protein Voltage Probe,” Neuron, vol. 75, no. 5, pp. 779–785, Sep. 2012.

[105] W. Akemann, H. Mutoh, A. Perron, Y. K. Park, Y. Iwamoto, and T. Knöpfel, “Imaging

neural circuit dynamics with a voltage-sensitive fluorescent protein,” J. Neurophysiol.,

vol. 108, no. 8, pp. 2323–2337, Jul. 2012.

[106] A. Lundby, W. Akemann, and T. Knöpfel, “Biophysical characterization of the

fluorescent protein voltage probe VSFP2.3 based on the voltage-sensing domain of Ci-

VSP,” Eur. Biophys. J., vol. 39, no. 12, pp. 1625–1635, Nov. 2010.

[107] H. Dana et al., “Sensitive red protein calcium indicators for imaging neural activity,”

eLife, vol. 5, p. e12727, Mar. 2016.

[108] M. Inoue et al., “Rational design of a high-affinity, fast, red calcium indicator R-

CaMP2,” Nat. Methods, vol. 12, no. 1, pp. 64–70, Jan. 2015.

[109] A. Badura, X. R. Sun, A. Giovannucci, L. A. Lynch, and S. S. H. Wang, “Fast calcium

sensor proteins for monitoring neural activity,” Neurophotonics, vol. 1, no. 2, p. 025008,

Oct. 2014.

[110] X. R. Sun et al., “Fast GCaMPs for improved tracking of neuronal activity,” Nat.

Commun., vol. 4, p. 2170, Jul. 2013.

[111] T.-W. Chen et al., “Ultrasensitive fluorescent proteins for imaging neuronal activity,”

Nature, vol. 499, no. 7458, pp. 295–300, Jul. 2013.

Page 143: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

143

[112] J. Akerboom et al., “Optimization of a GCaMP Calcium Indicator for Neural Activity

Imaging,” J. Neurosci., vol. 32, no. 40, pp. 13819–13840, Oct. 2012.

[113] L. Tian et al., “Imaging neural activity in worms, flies and mice with improved GCaMP

calcium indicators,” Nat. Methods, vol. 6, no. 12, pp. 875–881, Dec. 2009.

[114] M. Z. Lin and M. J. Schnitzer, “Genetically encoded indicators of neuronal activity,” Nat.

Neurosci., vol. 19, no. 9, pp. 1142–1153, Aug. 2016.

[115] H. C. Jones and R. F. Keep, “Brain fluid calcium concentration and response to acute

hypercalcaemia during development in the rat.,” J. Physiol., vol. 402, no. 1, pp. 579–593,

1988.

[116] T. J. B. Simons, “Calcium and neuronal function,” Neurosurg. Rev., vol. 11, no. 2, pp.

119–129, Jun. 1988.

[117] N. Rocherfort, J. Hongbo, and A. Konnerth, “Calcium imaging in the living brain:

prospects for molecular medicine,” Cell, 2008.

[118] K. S. Girven and D. R. Sparta, “Probing Deep Brain Circuitry: New Advances in in Vivo

Calcium Measurement Strategies,” ACS Chem. Neurosci., vol. 8, no. 2, pp. 243–251, Feb.

2017.

[119] M. C. Stüttgen, J. Rüter, and C. Schwarz, “Two Psychophysical Channels of Whisker

Deflection in Rats Align with Two Neuronal Classes of Primary Afferents,” J. Neurosci.,

vol. 26, no. 30, pp. 7933–7941, Jul. 2006.

[120] M. C. Stüttgen and C. Schwarz, “Psychophysical and neurometric detection performance

under stimulus uncertainty,” Nat. Neurosci., vol. 11, no. 9, pp. 1091–1099, Sep. 2008.

[121] C. A. Gollnick, D. C. Millard, A. D. Ortiz, R. V. Bellamkonda, and G. B. Stanley,

“Response reliability observed with voltage-sensitive dye imaging of cortical layer 2/3:

The Probability of Activation Hypothesis,” J. Neurophysiol., p. jn.00547.2015, Feb.

2016.

[122] D. C. Ng et al., “On-chip biofluorescence imaging inside a brain tissue phantom using a

CMOS image sensor for in vivo brain imaging verification,” Sens. Actuators B Chem.,

vol. 119, no. 1, pp. 262–274, Nov. 2006.

[123] J. Pichaandi, J.-C. Boyer, K. R. Delaney, and F. C. J. M. van Veggel, “Two-Photon

Upconversion Laser (Scanning and Wide-Field) Microscopy Using Ln3+-Doped NaYF4

Upconverting Nanocrystals: A Critical Evaluation of their Performance and Potential in

Bioimaging,” J. Phys. Chem. C, vol. 115, no. 39, pp. 19054–19064, Oct. 2011.

[124] M. Pisanello et al., “The Three-Dimensional Signal Collection Field for Fiber

Photometry in Brain Tissue,” Front. Neurosci., vol. 13, 2019.

Page 144: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

144

[125] C. J. Engelbrecht, W. Göbel, and F. Helmchen, “Enhanced fluorescence signal in

nonlinear microscopy through supplementary fiber-optic light collection,” Opt. Express,

vol. 17, no. 8, pp. 6421–6435, Apr. 2009.

[126] D. Kupferschmidt, “Parallel, but Dissociable, Processing in Discrete Corticostriatal

Inputs Encodes Skill Learning,” 2017.

[127] J. A. González, L. T. Jensen, P. Iordanidou, M. Strom, L. Fugger, and D. Burdakov,

“Inhibitory Interplay between Orexin Neurons and Eating,” Curr. Biol., vol. 26, no. 18,

pp. 2486–2491, Sep. 2016.

[128] J. Muir et al., “In Vivo Fiber Photometry Reveals Signature of Future Stress

Susceptibility in Nucleus Accumbens,” Neuropsychopharmacology, vol. 43, no. 2, pp.

255–263, Jan. 2018.

[129] A. Eban-Rothschild, G. Rothschild, W. J. Giardino, J. R. Jones, and L. de Lecea, “VTA

dopaminergic neurons regulate ethologically relevant sleep–wake behaviors,” Nat.

Neurosci., vol. 19, no. 10, pp. 1356–1366, Oct. 2016.

[130] Y. Chen, Y.-C. Lin, T.-W. Kuo, and Z. A. Knight, “Sensory Detection of Food Rapidly

Modulates Arcuate Feeding Circuits,” Cell, vol. 160, no. 5, pp. 829–841, Feb. 2015.

[131] C. K. Kim et al., “Simultaneous fast measurement of circuit dynamics at multiple sites

across the mammalian brain,” Nat. Methods, vol. 13, no. 4, pp. 325–328, Apr. 2016.

[132] G. A. Matthews et al., “Dorsal Raphe Dopamine Neurons Represent the Experience of

Social Isolation,” Cell, vol. 164, no. 4, pp. 617–631, Feb. 2016.

[133] Y. Mandelblat-Cerf et al., “Bidirectional Anticipation of Future Osmotic Challenges by

Vasopressin Neurons,” Neuron, vol. 93, no. 1, pp. 57–65, Jan. 2017.

[134] E. S. Calipari et al., “In vivo imaging identifies temporal signature of D1 and D2 medium

spiny neurons in cocaine reward,” Proc. Natl. Acad. Sci., vol. 113, no. 10, pp. 2726–

2731, Mar. 2016.

[135] Q. Guo et al., “Multi-channel fiber photometry for population neuronal activity

recording,” Biomed. Opt. Express, vol. 6, no. 10, p. 3919, Oct. 2015.

[136] H. Adelsberger, O. Garaschuk, and A. Konnerth, “Cortical calcium waves in resting

newborn mice,” Nat. Neurosci., vol. 8, no. 8, pp. 988–990, Aug. 2005.

[137] A. Stroh et al., “Making Waves: Initiation and Propagation of Corticothalamic Ca2+

Waves In Vivo,” Neuron, vol. 77, no. 6, pp. 1136–1150, Mar. 2013.

[138] H. Adelsberger, C. Grienberger, A. Stroh, and A. Konnerth, “In Vivo Calcium

Recordings and Channelrhodopsin-2 Activation through an Optical Fiber,” Cold Spring

Harb. Protoc., vol. 2014, no. 10, p. pdb.prot084145, Oct. 2014.

Page 145: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

145

[139] J. L. Chen, D. J. Margolis, A. Stankov, L. T. Sumanovski, B. L. Schneider, and F.

Helmchen, “Pathway-specific reorganization of projection neurons in somatosensory

cortex during learning,” Nat. Neurosci., vol. 18, no. 8, pp. 1101–1108, Aug. 2015.

[140] F. Schmid et al., “Assessing sensory versus optogenetic network activation by combining

(o)fMRI with optical Ca2+ recordings,” J. Cereb. Blood Flow Metab., p.

0271678X15619428, Nov. 2015.

[141] D. Mu et al., “A central neural circuit for itch sensation,” Science, vol. 357, no. 6352, pp.

695–699, Aug. 2017.

[142] J. Clarkson et al., “Definition of the hypothalamic GnRH pulse generator in mice,” Proc.

Natl. Acad. Sci., vol. 114, no. 47, pp. E10216–E10223, Nov. 2017.

[143] J. Sawinski, D. J. Wallace, D. S. Greenberg, S. Grossmann, W. Denk, and J. N. D. Kerr,

“Visually evoked activity in cortical cells imaged in freely moving animals,” Proc. Natl.

Acad. Sci., vol. 106, no. 46, pp. 19557–19562, Nov. 2009.

[144] K. Sohya, K. Kameyama, Y. Yanagawa, K. Obata, and T. Tsumoto, “GABAergic

Neurons Are Less Selective to Stimulus Orientation than Excitatory Neurons in Layer

II/III of Visual Cortex, as Revealed by In Vivo Functional Ca2+ Imaging in Transgenic

Mice,” J. Neurosci., vol. 27, no. 8, pp. 2145–2149, Feb. 2007.

[145] K. Ohki, S. Chung, Y. H. Ch’ng, P. Kara, and R. C. Reid, “Functional imaging with

cellular resolution reveals precise micro-architecture in visual cortex,” Nature, vol. 433,

no. 7026, p. 597, Feb. 2005.

[146] M. Kaschube, M. Schnabel, S. Löwel, D. M. Coppola, L. E. White, and F. Wolf,

“Universality in the Evolution of Orientation Columns in the Visual Cortex,” Science,

vol. 330, no. 6007, pp. 1113–1116, Nov. 2010.

[147] S. Schuett, T. Bonhoeffer, and M. Hübener, “Mapping Retinotopic Structure in Mouse

Visual Cortex with Optical Imaging,” J. Neurosci., vol. 22, no. 15, pp. 6549–6559, Aug.

2002.

[148] Q. Wang and A. Burkhalter, “Area map of mouse visual cortex,” J. Comp. Neurol., vol.

502, no. 3, pp. 339–357, 2007.

[149] S. Han, W. Yang, and R. Yuste, “Two-Color Volumetric Imaging of Neuronal Activity of

Cortical Columns,” Cell Rep., vol. 27, no. 7, pp. 2229-2240.e4, May 2019.

[150] M. M. Mansy, H. Kim, and K. G. Oweiss, “Spatial detection characteristics of a single

photon fiber photometry system for imaging neural ensembles*,” in 2019 9th

International IEEE/EMBS Conference on Neural Engineering (NER), 2019, pp. 969–972.

[151] E. A. Pnevmatikakis et al., “Simultaneous Denoising, Deconvolution, and Demixing of

Calcium Imaging Data,” Neuron, vol. 89, no. 2, pp. 285–299, Jan. 2016.

Page 146: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

146

[152] A. Giovannucci et al., “CaImAn an open source tool for scalable calcium imaging data

analysis,” p. 45.

[153] Y. Sych, M. Chernysheva, L. T. Sumanovski, and F. Helmchen, “High-density multi-

fiber photometry for studying large-scale brain circuit dynamics,” Nat. Methods, vol. 16,

no. 6, p. 553, Jun. 2019.

[154] S. L. Resendez and G. D. Stuber, “In vivo Calcium Imaging to Illuminate Neurocircuit

Activity Dynamics Underlying Naturalistic Behavior,” Neuropsychopharmacology, vol.

40, no. 1, p. 238, Jan. 2015.

[155] K. Ronayne, “Optoelectronic Systems and Applications for In Vivo Fiber Photometry,”

2016.

[156] H. E. Anderson, A. K. Fontaine, J. H. Caldwell, and R. F. Weir, “Imaging of electrical

activity in small diameter fibers of the murine peripheral nerve with virally-delivered

GCaMP6f,” Sci. Rep., vol. 8, no. 1, p. 3219, Feb. 2018.

[157] M. Murayama and M. Larkum, “Fiber-Optic Calcium Monitoring of Dendritic Activity In

Vivo,” Cold Spring Harb. Protoc., vol. 2012, no. 2, p. pdb.prot067835, Feb. 2012.

[158] L. Wang, S. L. Jacques, and L. Zheng, “MCML—Monte Carlo modeling of light

transport in multi-layered tissues,” Comput. Methods Programs Biomed., vol. 47, no. 2,

pp. 131–146, Jul. 1995.

[159] M. Abdellah, “In Silico Brain Imaging Physically-plausible Methods for Visualizing

Neocortical Microcircuitry,” Infoscience, 2017. [Online]. Available:

https://infoscience.epfl.ch/record/232444. [Accessed: 21-Jun-2019].

[160] M. Abdellah, A. Bilgili, S. Eilemann, J. Shillcock, H. Markram, and F. Schürmann, “Bio-

physically plausible visualization of highly scattering fluorescent neocortical models for

in silico experimentation,” BMC Bioinformatics, vol. 18, no. 2, p. 62, Feb. 2017.

[161] W. Bruno and F. Helmchen, Optical Imaging of Neocortical Dynamics Book.pdf. 2014.

[162] B. Gysbrechts et al., “Measurement of the optical properties of rat brain tissue using

contact spatially resolved spectroscopy,” in Biophotonics: Photonic Solutions for Better

Health Care IV, 2014, vol. 9129, p. 912936.

[163] M. Azimipour, R. Baumgartner, Y. Liu, S. L. Jacques, K. W. Eliceiri, and R. Pashaie,

“Extraction of optical properties and prediction of light distribution in rat brain tissue,” J.

Biomed. Opt., vol. 19, no. 7, p. 075001, Jul. 2014.

[164] Y. Liu, S. L. Jacques, M. Azimipour, J. D. Rogers, R. Pashaie, and K. W. Eliceiri,

“OptogenSIM: a 3D Monte Carlo simulation platform for light delivery design in

optogenetics,” Biomed. Opt. Express, vol. 6, no. 12, pp. 4859–4870, Dec. 2015.

[165] S. L. Jacques, “Optical properties of biological tissues: a review,” 2013.

Page 147: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

147

[166] S. I. Al-Juboori, A. Dondzillo, E. A. Stubblefield, G. Felsen, T. C. Lei, and A. Klug,

“Light Scattering Properties Vary across Different Regions of the Adult Mouse Brain,”

PLOS ONE, vol. 8, no. 7, p. e67626, Jul. 2013.

[167] T. Vo-Dinh, Biomedical Photonics Handbook. 2003.

[168] M. Azimipour, F. Atry, and R. Pashaie, “Effect of blood vessels on light distribution in

optogenetic stimulation of cortex,” Opt. Lett., vol. 40, no. 10, pp. 2173–2176, May 2015.

[169] B. A. Flusberg, E. D. Cocker, W. Piyawattanametha, J. C. Jung, E. L. M. Cheung, and M.

J. Schnitzer, “Fiber-optic fluorescence imaging,” Nat. Methods, vol. 2, no. 12, pp. 941–

950, Dec. 2005.

[170] V. Venkatachalam and A. E. Cohen, “Imaging GFP-Based Reporters in Neurons with

Multiwavelength Optogenetic Control,” Biophys. J., vol. 107, no. 7, pp. 1554–1563, Oct.

2014.

[171] T. V. Gerdjikov, C. G. Bergner, M. C. Stüttgen, C. Waiblinger, and C. Schwarz,

“Discrimination of Vibrotactile Stimuli in the Rat Whisker System: Behavior and

Neurometrics,” Neuron, vol. 65, no. 4, pp. 530–540, Feb. 2010.

[172] D. J. Pinto, J. C. Brumberg, and D. J. Simons, “Circuit Dynamics and Coding Strategies

in Rodent Somatosensory Cortex,” J. Neurophysiol., vol. 83, no. 3, pp. 1158–1166, Mar.

2000.

[173] C. Waiblinger, D. Brugger, and C. Schwarz, “Vibrotactile Discrimination in the Rat

Whisker System is Based on Neuronal Coding of Instantaneous Kinematic Cues,” Cereb.

Cortex, vol. 25, no. 4, pp. 1093–1106, Apr. 2015.

[174] D. R. Ollerenshaw, H. J. V. Zheng, D. C. Millard, Q. Wang, and G. B. Stanley, “The

Adaptive Trade-Off between Detection and Discrimination in Cortical Representations

and Behavior,” Neuron, vol. 81, no. 5, pp. 1152–1164, Mar. 2014.

[175] A. Pitas, A. L. Albarracín, M. Molano-Mazón, and M. Maravall, “Variable Temporal

Integration of Stimulus Patterns in the Mouse Barrel Cortex,” Cereb. Cortex, p. bhw006,

Feb. 2016.

[176] C. P. J. De Kock, R. M. Bruno, H. Spors, and B. Sakmann, “Layer- and cell-type-specific

suprathreshold stimulus representation in rat primary somatosensory cortex,” J. Physiol.,

vol. 581, no. 1, pp. 139–154, May 2007.

[177] T.-W. Chen et al., “Ultrasensitive fluorescent proteins for imaging neuronal activity,”

Nature, vol. 499, no. 7458, pp. 295–300, Jul. 2013.

[178] P. R. Patel et al., “Chronic in vivo stability assessment of carbon fiber microelectrode

arrays,” J. Neural Eng., vol. 13, no. 6, p. 066002, 2016.

Page 148: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

148

[179] R. H. Howland, “Vagus Nerve Stimulation,” Curr. Behav. Neurosci. Rep., vol. 1, no. 2,

pp. 64–73, Jun. 2014.

[180] S. A. Hays, R. L. Rennaker, and M. P. Kilgard, “Targeting Plasticity with Vagus Nerve

Stimulation to Treat Neurological Disease,” Prog. Brain Res., vol. 207, pp. 275–299,

2013.

[181] R. A. Jensen, “Modulation of memory storage processes by peripherally acting

pharmacological agents.,” in Proceedings of the Western Pharmacology Society, 1996,

vol. 39, p. 85.

[182] K. B. Clark, D. C. Smith, D. L. Hassert, R. A. Browning, D. K. Naritoku, and R. A.

Jensen, “Posttraining Electrical Stimulation of Vagal Afferents with Concomitant Vagal

Efferent Inactivation Enhances Memory Storage Processes in the Rat,” Neurobiol. Learn.

Mem., vol. 70, no. 3, pp. 364–373, Nov. 1998.

[183] B. A. Malow, J. Edwards, M. Marzec, O. Sagher, D. Ross, and G. Fromes, “Vagus nerve

stimulation reduces daytime sleepiness in epilepsy patients,” Neurology, vol. 57, no. 5,

pp. 879–884, 2001.

[184] P. Rizzo et al., “Chronic vagus nerve stimulation improves alertness and reduces rapid

eye movement sleep in patients affected by refractory epilepsy,” Sleep, vol. 26, no. 5, pp.

607–611, 2003.

[185] J. A. Nichols, A. R. Nichols, S. M. Smirnakis, N. D. Engineer, M. P. Kilgard, and M.

Atzori, “Vagus nerve stimulation modulates cortical synchrony and excitability through

the activation of muscarinic receptors,” Neuroscience, vol. 189, pp. 207–214, Aug. 2011.

[186] J. A. Shetake, N. D. Engineer, W. A. Vrana, J. T. Wolf, and M. P. Kilgard, “Pairing tone

trains with vagus nerve stimulation induces temporal plasticity in auditory cortex,” Exp.

Neurol., vol. 233, no. 1, pp. 342–349, Jan. 2012.

[187] N. Khodaparast et al., “Vagus Nerve Stimulation Delivered During Motor Rehabilitation

Improves Recovery in a Rat Model of Stroke,” Neurorehabil. Neural Repair, vol. 28, no.

7, pp. 698–706, Sep. 2014.

[188] N. Khodaparast et al., “Vagus Nerve Stimulation During Rehabilitative Training

Improves Forelimb Recovery After Chronic Ischemic Stroke in Rats,” Neurorehabil.

Neural Repair, vol. 30, no. 7, pp. 676–684, Aug. 2016.

[189] D. T. Pruitt et al., “Vagus Nerve Stimulation Delivered with Motor Training Enhances

Recovery of Function after Traumatic Brain Injury,” J. Neurotrauma, vol. 33, no. 9, pp.

871–879, May 2015.

[190] S. A. Hays et al., “Vagus nerve stimulation during rehabilitative training enhances

recovery of forelimb function after ischemic stroke in aged rats,” Neurobiol. Aging, vol.

43, pp. 111–118, Jul. 2016.

Page 149: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

149

[191] D. C. Smith et al., “Electrical Stimulation of the Vagus Nerve Enhances Cognitive and

Motor Recovery following Moderate Fluid Percussion Injury in the Rat,” J.

Neurotrauma, vol. 22, no. 12, pp. 1485–1502, Dec. 2005.

[192] E. L. Newman, K. Gupta, J. R. Climer, C. K. Monaghan, and M. E. Hasselmo,

“Cholinergic modulation of cognitive processing: insights drawn from computational

models,” Front. Behav. Neurosci., vol. 6, Jun. 2012.

[193] N. J. Woolf and L. L. Butcher, “Cholinergic systems mediate action from movement to

higher consciousness,” Behav. Brain Res., vol. 221, no. 2, pp. 488–498, Aug. 2011.

[194] V. Parikh, R. Kozak, V. Martinez, and M. Sarter, “Prefrontal Acetylcholine Release

Controls Cue Detection on Multiple Timescales,” Neuron, vol. 56, no. 1, pp. 141–154,

Oct. 2007.

[195] K. G. Oweiss, Statistical Signal Processing for Neuroscience and Neurotechnology.

Academic Press, 2010.

[196] R. G. Gallager, Detection, decisions, and hypothesis testing. 2012.

[197] M. L. Andermann, A. M. Kerlin, and C. Reid, “Chronic cellular imaging of mouse visual

cortex during operant behavior and passive viewing,” Front. Cell. Neurosci., vol. 4, 2010.

[198] M. Leinweber, D. R. Ward, J. M. Sobczak, A. Attinger, and G. B. Keller, “A

Sensorimotor Circuit in Mouse Cortex for Visual Flow Predictions,” Neuron, vol. 95, no.

6, pp. 1420-1432.e5, Sep. 2017.

[199] C. M. Niell and M. P. Stryker, “Modulation of Visual Responses by Behavioral State in

Mouse Visual Cortex,” Neuron, vol. 65, no. 4, pp. 472–479, Feb. 2010.

Page 150: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

150

BIOGRAPHICAL SKETCH

May was born in Cairo, Egypt and lived a considerable amount of her childhood in

Dresden, Germany. She finished her primary education in an international German school in

Cairo where she received the Abitur (Germany’s higher education entrance qualification

requirement). After that, she pursued a Bachelor of Science degree in systems and biomedical

engineering at the Cairo University in Egypt, which she graduated with honored distinction. May

worked for three years as a software engineering at two healthcare and medical software solution

companies.

In 2012 May moved to the US to start graduate school. She was admitted to Dr. Karim

Oweiss’ lab at Michigan State University in 2013, where she started her Ph.D. in electrical and

computer engineering. One year later, she moved with Dr. Oweiss to the University of Florida

and transferred to the Department of Biomedical Engineering.

May work in the Oweiss lab was very versatile and spanned numerous neuroscience

techniques like stereotaxic cranial surgery, injections and implants, post-mortem histological

analysis, neural data analysis, neural device design as well as peripheral nerve surgery, cuff

fabrication and electrical stimulation. She also served as the lab manager for three years.

May’s doctoral research focused on recording in-vivo neural ensemble activity as

reported by genetically encoded Ca2+ indicator using the fiber photometry system. She then

concentrated her efforts on deciphering the nature of neural signal recorded by the Fiber

Photometry system and conducted a systematic characterization of the device.

Over the course of her doctoral tenure, May also served as a graduate teaching assistant

for three graduate and undergraduate courses with the biomedical engineering department and

co-instructed the bioelectrical systems class with the Department of Electrical and Computer

Engineering at UF. May had numerous leadership experiences and received a certificate in

Page 151: A SYSTEMATIC CHARACTERIZATION OF FIBER PHOTOMETRY … · Abduraham Siddiqi, Kyle Sheller. I would like to express my sincere gratitude to mentor, Dr. Yifei Dai, for offering me the

151

engineering leadership by the College of Engineering at UF in 2018. In 2017, she was elected

president of the Women in Science and Engineering (WiSE) student organization and has been

an active member since then. Her strong passion to encourage grade school students to learn

about STEM topics keeps her constantly involved in outreach programs and events.