Data Sheet
PID-Analyzer
Ion Mobility Spectrometry
Photo Ionization Detector
Gas Chromatography
Volatile Organic Compounds (VOC) like BTEX are emanated from broad range of sources and have direct influences on life q
The PID-Analyzer is a multi-gas analyzer as gas-chromatograph (GC) with a Photo Ionization Detector (PID) for the read-out able to identify compounds. It has a high sensitive detection and identification of Volatile Organic Compounds such as Benzene, Toluene, Ethyl benzene and Xylene (BTEX). For the monitoring of Total-VOC the PID-Analyzer can be set-up as bare PID. No carrier gas is required which results into low operation costs and makes it suitable for mobile use. As mobile device the PID-Analyzer it is the smallest and most sensitive GC-PID device on the market.
uality and health. But also other compounds such as Ethylene oxide used in medical or pharmaceutical applications, light weight hydrocarbons such as Hexane in the petrochemie or solvents in the semiconductor production are of interest. In principal all compounds which can be ionized with an UV lamp (standard 10.6 eV UV lamp) can be measured with the PID-Analyzer.
Analysis Application
• Emission / Immission Control
• Ambient Air Monitoring
• Process Gas Monitoring
• Continuously Emission Control
• Semiconductor Fabrication
• Environmental Monitoring
• Workplace Safety
• Pharmaceutical Sterilization Process
Features
• Real-time analysis of
Volatile Organic Compounds (VOC)
• High sensitivity in lower ppb-range
• Superior selectivity and cross sensitivity
rejection
• Analysis directly in situ
(No Tedlar bag sampling / no lab analysis)
• Operation under ambient air pressure
• Low costs of operation, no consumables,
no carrier gas required
A high sensitive multi-gas GC-PID analyzer
Data Sheet
Technical DataPID-Analyzer
A high sensitive multi-gas GC-PID analyzer
The PID-Analyzer is a multi-gas trace analyzer for the detection and identification of Volatile Organic Compounds (VOCs) in the ppb-range wi thout any pre-concentration.It is a Photo Ionization Detector (PID) coupled with a gas chromatographic column (GC-column) for pre-separation.
The PID-Analyzer does not show any interference (quenching) from humidity and offers manual and automatic analysis.With this detector combination the disadvantage of conventional PIDs is overcome and therefore allows the identification of compounds. This is quite important for the risk assessment because only with identi f ication a related concentration can be measured. It is especially important for gas mixtures which can be disentangled by the GC-PID.
The PID-cell used in this system is a special designed PID with sensitivity in the ppb-range. Since no carrier gas is required the operational costs of the PID-Analyzer (GC-PID) are low. As mobile version the PID-Analyzer can be set-up as an autarkic and mobile working analyzer but can also be connected to an external computer via interface.
EX-protection (purge system) is available.
Calibration functions with an additional calibrator for SPAN/ZERO calibration is offered as well as Multi-Point-Sample (MPS) systems, dilution- or even gas probe chiller systems.
The PID-Analyzer enclosures can range from Class 1 Division 1 to general purpose. Also
Detection principle:
GC-PIDPhoto Ionization Detector (PID) with gas-chromatographic pre-separation as bare PID for Total-VOC monitoring
Ionization: UV-lamp Standard (10.6 eV)
Measurablecompounds(only as example;other compounds possible):
• Benzene• Toluene• Xylene• Styrene• a-Pinene• Ethylene oxide• Pyrrole
Sensitivity: Lower ppb range
Moisture: 0% to 90% RH
Data output:(depending on version)
• Integrated GUI• 4 - 20 mA (analog)• RS-232 (optional)• RS-485 (optional)• USB (optional)• LAN / WLAN (optional)• Digital in/out
Warm-up time: 20 minutes
Measuring time: 30 seconds to minutes (depending on GC setup)
Power: 100 V - 240 V / 50Hz - 60 Hzmobile: typically 8 hours with Li-ion battery
Carrier gas: No carrier gas required
Enclosure: 19'' rack mounted
Industrial enclosure (NEMA 4)
Mobile
Dimension: W:H:D:
483 mm135 mm420 mm
W:H:D:
500 mm500 mm210 mm
W:H:D:
280 mm100 mm280 mm
Weight: ~10 kg (~22 lbs)
18 kg (40 lbs.)
~7 kg (~15.5 lbs)
Operationtemperature:
0°C - 50°C(+32°F - 122°F)