8 Validation Methods 2008-2

99
Validation methods 1 Update:24/020/08 Bioanalytical methods validation for pharmacokinetic studies P.L. Toutain Toulouse Feb. 2008 ECOLE NATIONALE VETERINAIRE T O U L O U S E

Transcript of 8 Validation Methods 2008-2

Page 1: 8 Validation Methods 2008-2

Validation methods 1Update:24/020/08

Bioanalytical methods validation for pharmacokinetic studies

Bioanalytical methods validation for pharmacokinetic studies

P.L. Toutain

Toulouse Feb. 2008

ECOLENATIONALEVETERINAIRE

T O U L O U S E

Page 2: 8 Validation Methods 2008-2

Validation methods 2Update:24/020/08

Validation methodsValidation methods

• Selective and sensitive analytical methods for the quantitative determination of drugs and their metabolites (analytes) are critical for successful performance of PK and bioequivalence studies

Page 3: 8 Validation Methods 2008-2

Validation methods 3Update:24/020/08

Validation methodsValidation methods

• Validation of analytical methods includes all the procedures recommended to demonstrate that a particular method, for a given matrix, is reliable and reproducible

Page 4: 8 Validation Methods 2008-2

Validation methods 4Update:24/020/08

Validation methodsValidation methods

1. A priori validation:– Pre-study validation for analytical

method development and method establishment

2. In-life validation

(Routine validation)

Page 5: 8 Validation Methods 2008-2

Validation methods 5Update:24/020/08

Regulatory requirementsRegulatory requirements

• G.L.P.– (e.g.; bioequivalence, Toxicokinetics)

• S.O.P. (standard operating procedure) – (from sample collection to reporting)– Record keeping– Chain of sample custody (chaîne des garanties)– Sample preparation– Analytical tools– Procedures for quality control and verification of

results

Page 6: 8 Validation Methods 2008-2

Validation methods 6Update:24/020/08

A priori validation makes sure the method

is suitable for its intended use

A priori validation makes sure the method

is suitable for its intended use

Page 7: 8 Validation Methods 2008-2

Validation methods 7Update:24/020/08

A priori validation: criteria to be validated

A priori validation: criteria to be validated

1. Calibration curve2. Accuracy3. Precision (repeatability, reproducibility)4. Limit of quantification (LOQ) 5. Limit of detection (LOD)6. Sensitivity7. Specificity/selectivity8. Stability of the analyte in the matrix under study9. Others (ruggedness, agreement,…)

Page 8: 8 Validation Methods 2008-2

Validation methods 8Update:24/020/08

1. Calibration curve1. Calibration curve

Page 9: 8 Validation Methods 2008-2

Validation methods 9Update:24/020/08

Definition It is the relationship between known concentrations and experimental response values

Goal Determine the unknown concentration of a sample

Calibration curveCalibration curve

Page 10: 8 Validation Methods 2008-2

Validation methods 10Update:24/020/08

Response: dependent variable(peak,area ..)

Y (

ob

serv

ed)

Y

X

y = ax + b

Independent variable:exactly knownconcentrations

Calibration curveCalibration curve

x1

y1

xn

Yn

Page 11: 8 Validation Methods 2008-2

Validation methods 11Update:24/020/08

Response: dependent variable

Y (

ob

serv

ed)

Y

X

y = ax + b

Independent variable:X

estimated concentration^

Calibration curveCalibration curve

x1

y1

xn

Yn

Page 12: 8 Validation Methods 2008-2

Validation methods 12Update:24/020/08

x

Response

x

Response

GOOD BAD

^ ^

Calibration curveCalibration curve

Page 13: 8 Validation Methods 2008-2

Validation methods 13Update:24/020/08

Calibration curveCalibration curve

• Construction

– 5 to 8 points over the analytical domain

– replicates are required to test linearity

• 3 to 5 replicates per levels

Page 14: 8 Validation Methods 2008-2

Validation methods 14Update:24/020/08

Standard calibration curveStandard calibration curve

Page 15: 8 Validation Methods 2008-2

Validation methods 15Update:24/020/08

Calibration curveCalibration curve

• The calibration curve should be prepared

in the same biological matrix (e.g. plasma )

as the sample in the intended study by

spiking with known concentration of the

analyte (or by serial dilution).

Page 16: 8 Validation Methods 2008-2

Validation methods 16Update:24/020/08

Reference StandardReference Standard

• Calibration standards and quality control samples (QC)

• Authenticated analytical reference standard should be used to prepare (separately) solution of known concentration– certified reference standards

• Never from a marketed drug formulation

– commercially supplied reference standards

– other material of documented purity

Page 17: 8 Validation Methods 2008-2

Validation methods 17Update:24/020/08

Building the calibration curve: a regression problem

Building the calibration curve: a regression problem

Page 18: 8 Validation Methods 2008-2

Validation methods 18Update:24/020/08

Building the calibration curve: a regression problem

Building the calibration curve: a regression problem

• In statistics, regression analysis is a statistical technique which examines the relation of a dependent variable (response variable or dependent variable i.e. Y) that is for us the response of the analytical apparatus (peak, area..) to specified independent variables (explanatory variables or independent variable i.e. X) that is for us the concentration of calibrators .

Page 19: 8 Validation Methods 2008-2

Validation methods 19Update:24/020/08

Linear regression : see WikipediaLinear regression : see Wikipedia

• Linear regression - Wikipedia, the free encyclopedia

Page 20: 8 Validation Methods 2008-2

Validation methods 20Update:24/020/08

Linear regression : WikipediaLinear regression : Wikipedia

• In statistics, linear regression is a regression method that models the relationship between a dependent variable Y, independent variables Xi, i = 1, ..., p, and a random term ε. The model can be written as:

• where β0 is the intercept ("constant" term), the βis are the respective parameters of independent variables, and p is the number of parameters to be estimated in the linear regression.

Page 21: 8 Validation Methods 2008-2

Validation methods 21Update:24/020/08

Linear regression : WikipediaLinear regression : Wikipedia

• This method is called "linear" because the relation of the response (the dependent variable Y) to the independent variables is assumed to be a linear function of the parameters.

Page 22: 8 Validation Methods 2008-2

Validation methods 22Update:24/020/08

Linear regression : WikipediaLinear regression : Wikipedia

• It is often erroneously thought that the reason the technique is called "linear regression" is that the graph of Y = β0 + βx is a straight line or that Y is a linear function of the X variables. But if the model is (for example)

• the problem is still one of linear regression, that is, linear in x and x2 respectively, even though the graph on x by itself is not a straight line. In other words, Y can be considered a linear function of the parameters (α, β, and γ), even though it is not a linear function of x.

Page 23: 8 Validation Methods 2008-2

Validation methods 23Update:24/020/08

Statistical requirements to build a calibration curve

Statistical requirements to build a calibration curve

Page 24: 8 Validation Methods 2008-2

Validation methods 24Update:24/020/08

Statistical requirements to build a calibration curve

Statistical requirements to build a calibration curve

1. Standard concentration (X) are known without error

2. Variance of response (Y) should be constant over the analytical domain (homoscedasticity hypothesis); this equivalent to say that the random errors εi are homoscedastic i.e., they all have the same variance.

3. The random errors εi have expected value 0.

4. The random errors εi should be independent from Y and are uncorrelated.

These assumptions imply that least-squares estimates of the parameters are optimal in a certain sense

Page 25: 8 Validation Methods 2008-2

Validation methods 25Update:24/020/08

Regression can be used for prediction

Regression can be used for prediction

Page 26: 8 Validation Methods 2008-2

Validation methods 26Update:24/020/08

Regression can be used for predictionRegression can be used for prediction

• These uses of regression (calibration curve) rely heavily on the model assumptions being satisfied.

• Calibration curve is misused for these purposes where the appropriate assumptions cannot be verified to hold

• The misuse of regression is due to the fact that it take considerably more knowledge and experience to critique a model than to fit a model with a software.

Page 27: 8 Validation Methods 2008-2

Validation methods 27Update:24/020/08

Assessing the calibration curve Assessing the calibration curve

the calibration curve (here a statistical model ) should be checked for two different things:

1. Whether the assumptions of least-squares are fulfilled

• Analysis (inspection) of residuals

2. Whether the model is valid and useful• Test of linearity• Back calculations

Page 28: 8 Validation Methods 2008-2

Validation methods 28Update:24/020/08

Validation of the calibration curveValidation of the calibration curve

• Homogeneity of variance

• Linearity

• Back calculations

Page 29: 8 Validation Methods 2008-2

Validation methods 29Update:24/020/08

Checking model assumptions Checking model assumptions

• The model assumptions are checked by calculating the residuals and plotting them.

• The residuals are calculated as follows :

fittedobserved YYsiduals Re

fitttedobserved YYsiduals Re

Page 30: 8 Validation Methods 2008-2

Validation methods 30Update:24/020/08

Inspection of residualsInspection of residuals

The following plots can be constructed to test the validity of the assumptions:

1. A normal probability plot of the residuals to test normality. The points should lie along a straight line.

2. Residuals against the explanatory variables, X. 3. Residuals against the fitted values, Y . 4. Residuals against the preceding residual.

• There should not be any noticeable pattern to the data in all but the first plot

Page 31: 8 Validation Methods 2008-2

Validation methods 31Update:24/020/08

Validation of the calibration curveValidation of the calibration curve

Homogeneity of variance

Page 32: 8 Validation Methods 2008-2

Validation methods 32Update:24/020/08

Problem of the homogeneity of variance Cochran's test

Homogeneous Non homogeneous"cone shaped"

Calibration curve: homogeneity of varianceCalibration curve: homogeneity of variance

Page 33: 8 Validation Methods 2008-2

Validation methods 33Update:24/020/08

Calibration curve: linearity & homogeneity of varianceInspection of a residuals plot

Calibration curve: linearity & homogeneity of varianceInspection of a residuals plot

If the linear model and the assumption of homoscedasticity are valid, the residual should be normally distributed and no trends should be apparent

Page 34: 8 Validation Methods 2008-2

Validation methods 34Update:24/020/08

Calibration curve: linearity & homogeneity of varianceInspection of a residuals plot

Calibration curve: linearity & homogeneity of varianceInspection of a residuals plot

The fact that the weighted residuals show a fan-like pattern, getting larger as X increase suggest heteroscedasticity and the use of a weighting procedure to reduce variance heterogeneity

Page 35: 8 Validation Methods 2008-2

Validation methods 35Update:24/020/08

• Heterogeneity of variance– Commonly observed

– Y has often a constant coefficient of variation

• Weighted regression– weighing factor proportional to the inverse of

variance (1/X, 1/X²…)

• After weighing, the coefficient of correlation (r) can be lower but accuracy and precision of prediction are better

Calibration curve: homogeneity of varianceCalibration curve: homogeneity of variance

Page 36: 8 Validation Methods 2008-2

Validation methods 36Update:24/020/08

Calibration curve: homogeneity of variance

Calibration curve: homogeneity of variance

Weighing factor=1/x2

Page 37: 8 Validation Methods 2008-2

Validation methods 37Update:24/020/08

Inspection of the residual plotInspection of the residual plot

Weighted residues Unweighted residues

Misfit evidenced by visual inspection of residuals despite the use of weighted regression: does the simple linear model holds???

Page 38: 8 Validation Methods 2008-2

Validation methods 38Update:24/020/08

Calibration curve : LinearityCalibration curve : Linearity

- Specific tests of linearity should be used

- The coefficient of correlation (r) cannot assess linearity except for r = 1

e.g.: r = 0.999 can be associated with a calibration curve which is not a straight line

Page 39: 8 Validation Methods 2008-2

Validation methods 39Update:24/020/08

Res

po

nse

Y

X

Calibration curve: linearityCalibration curve: linearity

Concentration

Test of linearity : Coefficient of correlation

r = 0.99does not prove linearity

Page 40: 8 Validation Methods 2008-2

Validation methods 40Update:24/020/08

Calibration curve: linearityCalibration curve: linearity

• Test of lack of fit

• Requires replicates

• Should be carried out after weighing

• ANOVA

Page 41: 8 Validation Methods 2008-2

Validation methods 41Update:24/020/08

YResponse

Test of lack-of-fit It is a comparison of 2 variances

Variance 1Mean estimated from

each set of data

Variance 2Mean estimated from the curve

?=

XConcentration

The case of very precise technique

Calibration curve: linearityCalibration curve: linearity

Page 42: 8 Validation Methods 2008-2

Validation methods 42Update:24/020/08

Calibration curve: linearityCalibration curve: linearity

• If no replicate

• Y = ax + b vs Y= ax + cx² + b

Y

X

Y

X

Test the significance of C

Page 43: 8 Validation Methods 2008-2

Validation methods 43Update:24/020/08

Calibration curve: linearityCalibration curve: linearity

• If non linearity

– use the 2nd degree polynom

– reduce the domain of the calibration curve

Page 44: 8 Validation Methods 2008-2

Validation methods 44Update:24/020/08

Calibration curve:Weight=1/X2 & quadratic component

Calibration curve:Weight=1/X2 & quadratic component

Page 45: 8 Validation Methods 2008-2

Validation methods 45Update:24/020/08

Calibration curve:Weight=1/X2 & quadratic

component

Calibration curve:Weight=1/X2 & quadratic

component

Linear &Weighted residues

Linear &Unweighted residues

Quadratic &Weighted residues

Page 46: 8 Validation Methods 2008-2

Validation methods 46Update:24/020/08

Page 47: 8 Validation Methods 2008-2

Validation methods 47Update:24/020/08

Coefficient of correlationCoefficient of correlation

Page 48: 8 Validation Methods 2008-2

Validation methods 48Update:24/020/08

Coefficient of correlationCoefficient of correlation

Page 49: 8 Validation Methods 2008-2

Validation methods 49Update:24/020/08

Coefficient of correlationCoefficient of correlation

Page 50: 8 Validation Methods 2008-2

Validation methods 50Update:24/020/08

Validation of the calibration curve:Back calculations

Validation of the calibration curve:Back calculations

• back calculation of the concentrations of calibration samples using the fitted curve coefficients

• The ULOQ calibrator must back-calculate to within ±15% of the nominal concentration.

• At least four out of six non-zero standards should meet the back-calculation criteria, including the LLOQ and ULOQ standards.

Page 51: 8 Validation Methods 2008-2

Validation methods 51Update:24/020/08

Calibration curve: ParallelismCalibration curve: Parallelism

• If samples should be diluted with blank plasma, parallelism should be investigated with QC samples

Page 52: 8 Validation Methods 2008-2

Validation methods 52Update:24/020/08

Freeze/thaw stabilityFreeze/thaw stability

• Avoid freeze and thaw cycles

• Enough aliquot samples should be to be prepared

Page 53: 8 Validation Methods 2008-2

Validation methods 53Update:24/020/08

Calibration curve: sensitivityCalibration curve: sensitivity

The sensitivity of an analytical method is itsability to give response to small changes inthe absolute amount of analyte present

1

2

3

High sensitivity

Concentration (X)

added quantity

Response (Y)measuredquantity

Page 54: 8 Validation Methods 2008-2

Validation methods 54Update:24/020/08

Long term freezer stabilityLong term freezer stability

• Required for some analytes and for retrospective investigations

• Re-assay QC after the study is completed

Page 55: 8 Validation Methods 2008-2

Validation methods 55Update:24/020/08

1

2

A1 A2 A2A1x x

Performance : The slope factor

Y

X^ ^

Calibration curve: sensitivityCalibration curve: sensitivity

Page 56: 8 Validation Methods 2008-2

Validation methods 56Update:24/020/08

Accuracy and precisionAccuracy and precision

Page 57: 8 Validation Methods 2008-2

Validation methods 57Update:24/020/08

Origin of the error :Accuracy and precision

Origin of the error :Accuracy and precision

• Systematic (not random)– bias

– impossible to be corrected accuracy

• Random– can be evaluated by statistics precision

Page 58: 8 Validation Methods 2008-2

Validation methods 58Update:24/020/08

Good PrecisionGood Accuracy

Poor PrecisionGood Accuracy

Good PrecisionPoor Accuracy

Poor PrecisionPoor Accuracy

Gold Standard

Silver Standard

Off-Base Model

Hit or Miss Model

Bias and precisionBias and precision

Page 59: 8 Validation Methods 2008-2

Validation methods 59Update:24/020/08

AccuracyAccuracy

Closeness of determined value to the true value.

The acceptance criteria is mean value 15% deviation from true value.

At LOQ, 20% deviation is acceptable.

Page 60: 8 Validation Methods 2008-2

Validation methods 60Update:24/020/08

Accuracy Accuracy

The accuracy is calculated using the followingequation :

Accuracy (%) = 100 x Found value - Theoretical value

Theoretical value

The accuracy at each concentration level mustbe lower than 15% except a LOQ (20%)

Page 61: 8 Validation Methods 2008-2

Validation methods 61Update:24/020/08

AccuracyAccuracy

• Determination– by replicate analysis of the sample

containing known amount of analyte

– 5 samples for at least 3 levels

– The mean value should be within 15% of the actual value except at LOQ where it should not deviate by more than 20%

Page 62: 8 Validation Methods 2008-2

Validation methods 62Update:24/020/08

PrecisionPrecision

The closeness of replicate determinations of a sample by an assay.

The acceptance criteria is 15% CV. At LOQ, 20% deviation is acceptable.

Page 63: 8 Validation Methods 2008-2

Validation methods 63Update:24/020/08

Repeatability (r)

Agreement between successive measurements on the same sample under the same conditions

Reproducibility (R)

The closeness of agreement between resultsobtained with the same method under different conditions

PrecisionPrecision

Page 64: 8 Validation Methods 2008-2

Validation methods 64Update:24/020/08

Precision… Considered at 3 Levels

Precision… Considered at 3 Levels

• Repeatability

• Intermediate Precision

• Reproducibility

Page 65: 8 Validation Methods 2008-2

Validation methods 65Update:24/020/08

Repeatability Repeatability

• Express the precision under the same

operating conditions over a short

interval of time.

• Also referred to as Intra-assay precision

– (within day)

Page 66: 8 Validation Methods 2008-2

Validation methods 66Update:24/020/08

Intermediate PrecisionIntermediate Precision

• Express within-laboratory variations.

• Between days variability

• Known as part of Ruggedness in USP

Page 67: 8 Validation Methods 2008-2

Validation methods 67Update:24/020/08

ReproducibilityReproducibility

• Definition: Ability reproduce data

within the predefined precision

• Repeatability test at two different labs

Page 68: 8 Validation Methods 2008-2

Validation methods 68Update:24/020/08

Precision: measurementPrecision: measurement

• Should be measured using a minimum of 5 determinations per concentration– A minimum of 3 concentrations in the

range of expected concentrations

– The precision at each concentration should not exceed 15% except for the LOQ (20%)

Page 69: 8 Validation Methods 2008-2

Validation methods 69Update:24/020/08

Precision: measurementPrecision: measurement

• for a single measurement : CV(%)

• for intra-day and inter-day precision ANOVA

Page 70: 8 Validation Methods 2008-2

Validation methods 70Update:24/020/08

Precision: data analysisPrecision: data analysis

• Single level of concentration with repetitione.g. 12, 13, 12, 14, 13, 14 µg/mL– mean : 13.0 µg/mL– SD: 0.8944 µg/mL– CV% = SD/mean * 100 = 6.88%

• CV% is also known as the relative standard deviation or RSD

Page 71: 8 Validation Methods 2008-2

Validation methods 71Update:24/020/08

Precision: data analysisPrecision: data analysis• Several levels of concentration and several days

day 1 levels (µg/mL) 0.5 5 20

Repetitions 0.4 5.2 20.5

0.5 5.1 21.0

0.4 4.9 19.8

0.6 5.2 18.8

day 2 and 3 : same protocol

ANOVA

Page 72: 8 Validation Methods 2008-2

Validation methods 72Update:24/020/08

Precision: the statistical modelPrecision: the statistical model

• The statistical model (for each concentration level)

Y = μ+ day + – μ: general mean

– day: an effect (day, technician, or any factor = inter )

– : error-random = intra

Page 73: 8 Validation Methods 2008-2

Validation methods 73Update:24/020/08

ANOVAANOVA

• Allows an estimation of the 2 variance terms

– inter-day mean square (BMS)

– intra-day mean square (WMS)

Page 74: 8 Validation Methods 2008-2

Validation methods 74Update:24/020/08

Repeatability and reproducibilityRepeatability and reproducibility

• SD for repeatability

r = Var(e)

• SD for reproducibility

R = ²(day) + ²(r)

variance for reproducibility is the sum of the variance for repeatability and the inter-day variance

Inter-day intra-day

Page 75: 8 Validation Methods 2008-2

Validation methods 75Update:24/020/08

Precision: ANOVAPrecision: ANOVA

• CV intra : 5%

• CV inter : 8%

CV inter CV intra

Page 76: 8 Validation Methods 2008-2

Validation methods 76Update:24/020/08

The limit of quantification (LOQ)The limit of quantification (LOQ)

• LOQ is the lowest amount of analytes in a sample which can be determined with defined precision and accuracy

• LOQ : 20%

Page 77: 8 Validation Methods 2008-2

Validation methods 77Update:24/020/08

Limit of quantification (LOQ)Limit of quantification (LOQ)

• The lowest standard on the calibration curve is the LOQ if:– no interference is present in the blanks

at retention time of the analyte for this concentration

– the response (analyte peak) has a precision of 20% and accuracy 80-120%

Page 78: 8 Validation Methods 2008-2

Validation methods 78Update:24/020/08

Estimation of chromatographic baseline noiseEstimation of chromatographic baseline noise

Np-pNp

W : Peak width1

Baseline noise

Largest variationof the baseline noise (N )p-p

Most important deviation (N )p

Sample chromatogram

Blanc chromatogram

(a)

(b)

Page 79: 8 Validation Methods 2008-2

Validation methods 79Update:24/020/08

Three analytical areasThree analytical areas

1 2 3

Xb

not detected Area of detection

Area of quantification

or CV<20%

LOD LOQ

Page 80: 8 Validation Methods 2008-2

Validation methods 80Update:24/020/08

•The recovery of an analyte in an assay is the detector response obtained from an amount of the analyte added to and extracted from the biological matrix, compared to the detector response obtained for the true concentration of the pure authentic standard

•The recovery allows to determine the percent of lost drug during sample preparation

•Minimal extraction ratio required to ensure a good repeatability

Recovery: definitionRecovery: definition

Page 81: 8 Validation Methods 2008-2

Validation methods 81Update:24/020/08

•Absolute recovery is evaluated using low, medium, and high QC samples and at least three times for each level

•The extraction recovery of the analyte (s) and internal standard(s) should be higher than 70%, precise, and reproducible.

Recovery: DeterminationRecovery: Determination

Page 82: 8 Validation Methods 2008-2

Validation methods 82Update:24/020/08

•Recommended to be a close analog of the analyte of interest

•Advantages and limits

Recovery: Internal standardRecovery: Internal standard

Page 83: 8 Validation Methods 2008-2

Validation methods 83Update:24/020/08

RecoveryRecovery

)_tan,/___(_

)_,/___(_covRe

solutiondardsmLngxofareaPeak

extractplasmamLngxofareaPeakery 100

Page 84: 8 Validation Methods 2008-2

Validation methods 84Update:24/020/08

Specificity / Selectivity (1)Specificity / Selectivity (1)

• Specificity : for an analyte– ability of the method to produce a

response for a single analyte• metabolites

• enantiomers

• Selectivity: for a matrix

Page 85: 8 Validation Methods 2008-2

Validation methods 85Update:24/020/08

Specificity / Selectivity (2)Specificity / Selectivity (2)

• Analyses of blank samples from different subjects (n=6)

• Blanks should be tested for interference using the proposed extraction procedure and other chromatographic conditions

• Results should be compared with those obtained with aqueous solution of the analyte at a concentration near the LOQ

• Blank plasma and pre-dose samples should be without interference

Page 86: 8 Validation Methods 2008-2

Validation methods 86Update:24/020/08

Specificity / Selectivity (3)Specificity / Selectivity (3)

• If more than 10% of the blank samples exhibit significant interference, the method should be changed to eliminate interference

Page 87: 8 Validation Methods 2008-2

Validation methods 87Update:24/020/08

Definition The drug must keep all its properties during the investigations

Stability at room temperature An experiment should cover 6 to 24h

Stability in frozen biological samples : (-20°C or -80°C) Stability sample should allow assay from day 0 to day 20

Stability during a freeze / thaw cycle Samples should be frozen and submitted to three freeze / thaw cycles Aliquotage is better than repeated freeze / thaw cycles

StabilityStability

Page 88: 8 Validation Methods 2008-2

Validation methods 88Update:24/020/08

In life validationIn life validation

Page 89: 8 Validation Methods 2008-2

Validation methods 89Update:24/020/08

In life validationIn life validation

– should be generated for each run

– no replicate

– should be validated

• back calculation

• quality control (QC)

Page 90: 8 Validation Methods 2008-2

Validation methods 90Update:24/020/08

In life validationIn life validation

• Validation performed in each batch (day) of study samples to be analyzed

• Validation of the routine calibration curve

QC samples

Page 91: 8 Validation Methods 2008-2

Validation methods 91Update:24/020/08

In life validation: validation of the calibration curve

In life validation: validation of the calibration curve

• Prepare routine calibration in the matrix of interest– calibration samples, n6 including blank

• Validation of the routine calibration curve– QC samples

– 3 concentration levels

– 3 QC per level

Page 92: 8 Validation Methods 2008-2

Validation methods 92Update:24/020/08

In life validation: calibration curveIn life validation: calibration curve

• separately prepared QC samples should be analyzed with test samples

• QC in duplicate at 3 different concentrations (one <=3X LOQ, one in midrange and one close to the high end of range) should be incorporated in each run

Page 93: 8 Validation Methods 2008-2

Validation methods 93Update:24/020/08

In life validation: calibration curveIn life validation: calibration curve

• Decision rule– at least 4 of 6 QC should be within 20%

of their respective nominal value

– 2 out of 6 QC may be outside the 20% of their respective nominal value but not at the same level

Page 94: 8 Validation Methods 2008-2

Validation methods 94Update:24/020/08

Y

X

Significant : Origin ?NS : Keep the intercept as an empirical parameter

Intercept : Test hypothesis that the line goes through the origin

Calibration curveCalibration curve

Page 95: 8 Validation Methods 2008-2

Validation methods 95Update:24/020/08

In life validation:Robustness/Stability assay of a drug

In life validation:Robustness/Stability assay of a drug

1.00

0.80

1.20

1.40

1.60

1.80

0.60

0 4 8 12 16 20

Time (days)

+ 2 SD

- 2 SD

Mean

Calculated concentration(mg/ml)

Page 96: 8 Validation Methods 2008-2

Validation methods 96Update:24/020/08

In life validation: the QCIn life validation: the QC

• to evaluate accuracy

• to evaluate precision

• to confirm LOQ

• to evaluate robustness of the method

• to confirm sample stability

Page 97: 8 Validation Methods 2008-2

Validation methods 97Update:24/020/08

ReferencesReferences

See Guidance for Industry (main guidances in the world)

• Bioanalytical Method Validation

– FDA May 2001: Bioanalytical Method Validation

– ICH 1995

– EMEA: no specific document

• Published Workshop Reports

• Shah, V.P. et al, Pharmaceutical Research: 1992; 9:588-592

• Shah, V.P. et al, Pharmaceutical Research: 2000; 17: 1551-1557

Page 98: 8 Validation Methods 2008-2

Validation methods 98Update:24/020/08

To see this guidance

Page 99: 8 Validation Methods 2008-2

Validation methods 99Update:24/020/08

To see this guidanceTo see this guidance