OrthopaedicsOne Articles Page of 318 372 66 Total ankle replacement (TAR) Contents Indications and Contraindications for TAR First-Generation Total Ankle Replacements Second-Generation of Total Ankle Replacements Results of First- and Second-Generation TARs Modern TAR Designs (Third-Generation TARs) The TAR Learning Curve Modern TAR Designs: Promising Results Future Insights References 66.1 Indications and Contraindications for TAR 66.1.1 Indications Total ankle replacement (TAR) is an alternative to ankle arthrodesis for the treatment of end-stage ankle osteroarthritis (OA) in select patients with advanced painful arthropathy of all three main etiologies: primary, post-traumatic, and secondary. One of the main advantages of TAR compared with ankle arthrodesis is preservation of functional range of motion (ROM) which is sacrificed in ankle fusion. Improved ROM allows patients to better perform activities of daily living and possibly regain athletic activities. The ideal candidate for TAR is an older (middle- to old-aged), reasonably mobile patient with no significant co-morbidities, a normal or low body mass index, adequate bone stock, a well-aligned and stable hindfoot, good soft tissues conditions, and no neurovascular impairment of the lower extremity. Another indication for TAR is bilateral end-stage ankle OA. Bilateral ankle fusion may not be the most appropriate treatment option given its significant detrimental effects on gait and functional status of patients. 1 This is particularly important to consider in patients with secondary OA due to hemochromatosis or hemophilia. Furthermore, in patients with previously performed subtalar, triple, and/or midfoot fusion the 2,3 tibiotalar fusion would completely “stiffen” the hindfoot, while the TAR may preserve functional motion. It has been shown that the clinical outcome of TAR when combined with hindfoot fusion is comparable to that of ankle replacement alone. 4 66.1.2 Contraindications

Transcript of 66 Total ankle replacement (TAR) - aofas.org · 66 Total ankle replacement (TAR) Contents...

OrthopaedicsOne Articles

Page of 318 372

66 Total ankle replacement (TAR)

Contents

Indications and Contraindications for TAR

First-Generation Total Ankle Replacements

Second-Generation of Total Ankle Replacements

Results of First- and Second-Generation TARs

Modern TAR Designs (Third-Generation TARs)

The TAR Learning Curve

Modern TAR Designs: Promising Results

Future Insights

References

66.1 Indications and Contraindications for TAR

66.1.1 Indications

Total ankle replacement (TAR) is an alternative to ankle arthrodesis for the treatment of end-stage ankle

osteroarthritis (OA) in select patients with advanced painful arthropathy of all three main etiologies: primary,

post-traumatic, and secondary. One of the main advantages of TAR compared with ankle arthrodesis is

preservation of functional range of motion (ROM) which is sacrificed in ankle fusion. Improved ROM allows

patients to better perform activities of daily living and possibly regain athletic activities.

The ideal candidate for TAR is an older (middle- to old-aged), reasonably mobile patient with no significant

co-morbidities, a normal or low body mass index, adequate bone stock, a well-aligned and stable hindfoot,

good soft tissues conditions, and no neurovascular impairment of the lower extremity.

Another indication for TAR is bilateral end-stage ankle OA. Bilateral ankle fusion may not be the most

appropriate treatment option given its significant detrimental effects on gait and functional status of patients.1

This is particularly important to consider in patients with secondary OA due to hemochromatosis or

hemophilia. Furthermore, in patients with previously performed subtalar, triple, and/or midfoot fusion the2,3

tibiotalar fusion would completely “stiffen” the hindfoot, while the TAR may preserve functional motion. It has

been shown that the clinical outcome of TAR when combined with hindfoot fusion is comparable to that of

ankle replacement alone.4

66.1.2 Contraindications

OrthopaedicsOne Articles

Page of 319 372

The relative contraindications for TAR are severe osteoporosis, history of osteomyelitis, diffuse

osteonecrosis, or significant bone defect on the tibial and/or talar site. Previous long-term therapy with

steroids or immunosuppressive substances may also reduce bone quality, resulting in compromised

osteointegration of prosthesis components. Further relative contraindications for TAR include heavy physical

work, medium level of sports activities (eg, tennis, jogging, and downhill ski), high body mass index,

diabetes, and smoking.

Significant preoperative varus or valgus deformity (>10°) has also been seen as a contraindication for TAR.

Doets et al found that preoperative deformity in the frontal plane is difficult to correct, causing instability and

subluxation of the bearing, which may result in the prosthesis failure. Wood and Deakin observed in their5

study including 200 STAR implants that preoperative varus or valgus deformity >15° may cause edge

loading of the mobile bearing. Therefore, they stated that this may be a relative contraindication for TAR.6 7

However, the preoperative hindfoot deformity should not be an absolute contraindication, as long as

additional realignment procedures (supramalleolar and/or calcaneal osteotomies, ligament reconstruction,

subtalar fusion) may correct the deformity. Karantana et al did not observe any differences in functional8-11

outcome and prosthesis component survivorship between patients with and without preoperative deformities

as long the deformity was addressed at the time of prosthesis implantation. Daniels et al demonstrated12,13

that correction of moderate to severe varus deformities is possible and results in good functional outcome

and stability of prosthesis components. Kim et al reported that the clinical outcome of TAR performed in14

ankles with preoperative varus deformity >10° is comparable with that of neutrally aligned ankles.15

However, the simultaneous surgical procedures addressing the preoperative deformity are necessary to

achieve good results.15

The absolute contraindications for TAR include:

Neuroarthropathy (Charcot foot)

Non-manageable hindfoot malalignment

Massive joint laxity (eg, patients with Marfan disease)

Highly compromised periarticular soft tissues (eg, in patients with posttraumatic OA who underwent

several previous surgeries)

Severe sensomotoric dysfunction of foot/ankle

Active soft-tissue or bony infection

Additionally, TAR should not be considered as the first-choice therapy in patients with a high level of

functional demand (eg, contact sports).

While many authors suggest that previous ankle infection is an absolute contraindication for TAR, we do16-20

not confirm this idea. In a series of 17 consecutive patients who underwent HINTEGRA TAR, we achieved

good functional results and did not observe any recurrence of infection (unpublished data). Eichinger et al

and Espinosa and Klammer also recommend TAR in patients with ankle OA due to previous infection.21,22

In several studies, avascular necrosis of the talus has been identified as an absolute contraindication for

TAR. However, it should be considered that some prosthesis designs offer the possibility to use the17,23-29

revision talar component. Also custom made components may be used replacing the whole body of talus.30

66.2 First-Generation Total Ankle Replacements

OrthopaedicsOne Articles

Page of 320 372

Since the 1970s, ankle arthrodesis has been recognized as having limitations regarding complication rate

and functional outcome. In most articles addressing TAR history, the study by Lord and31-33 22,31,32,34-38

Marotte is described as the first clinical study with TAR patients.39

However, Muir et al described in 2002 outcome results in a 71-year-old male who underwent talar dome

resurfacing with a custom Vitallium implant for post-traumatic OA in 1962. The clinical examination at40

40-year follow up showed mild hindfoot malalignment with slightly decreased ROM (25° plantar flexion),

AOFAS score of 85, no pain, and no activity limitation.40

In their study, Lord and Marotte used an inverted hip stem, which was implanted into the tibia. After the39

talus had been completely removed, they implanted a cemented acetabular cup in the calcaneus. This

procedure was performed in 25 consecutive patients and only seven patients reported satisfaction

postoperatively. Twelve of the 25 arthroplasties failed, and therefore the authors did not recommend the41

further use of this prosthesis design. At the time, the authors recognized the complexity of ankle41

biomechanics and concluded that a simple hinge prosthesis system with plantar flexion and dorsiflexion

would be insufficient to mimic the normal ankle joint. Overall, the majority of first-generation prostheses41

were eventually withdrawn from the market because of high failure rates with subsidence, continued patient

pain, or progressive alignment deformities.

The research published by these original investigators led to the development of the first-generation of

TARs. These included:

, a semi-constrained prosthesis type introduced in 1973St. Georg-Buchholz ankle prosthesis, a two-component, constrained implant with aImperial College of London Hospital prosthesis

polyethylene tibial component 42,43

(Howmedica prosthesis), one of the first early ankle prostheses where a specialIrvine Ankle TARtalus anatomy was regarded, with the prosthesis designers performed anatomical measurements of

32 tali to establish the morphology of talus 44

, a non-constrained so-called “ball-and-socket” (spherocentric) prosthesis thatRichard Smith TAR

was introduced in 1975 45

, a very constrained prosthesis type Conaxial Beck-Steffee ankle prosthesis 46

, a non-constrained cemented prosthesis including the high densityNewton ankle implantpolyethylene tibial and Vitallium talar components

, an unconstrained, two-component total ankle design.Bath-Wessex TAR, a highly congruent two-component design including a polyethylene tibial component withMayo TAR

cement fixation 47,48

, a single-axis, two-component TAR designOregon ankle prosthesis, a two-component semi-constrained cemented implant that includedThompson-Richard prosthesis

a polyethylene tibial component with a concave articular surface and a lip on each side 1,49

, developed by a bioengineer and an orthopaedic surgeon New Jersey or Cylindrical TAR 50

66.3 Second-Generation of Total Ankle Replacements

Based on research demonstrating the high complication and failure rates and lack of patient satisfaction with

the first-generation of TARs, a second-generation was developed, including:

OrthopaedicsOne Articles

Page of 321 372

The (Depuy), the first of a new generation of ankle prostheses, has beenAgility ankle prosthesis

used since 1984. The Agility has been approved by the FDA and is currently the most widely used51

ankle prosthesis in the United States, and with more than 20 years of implantations, it has the longest

follow up of any fixed-bearing TAR. The Agility ankle prosthesis is a semi-constrained,52

two-component prosthesis consisting of a titanium tibial and cobalt-chromium talar component. For

improved osseous integration, both components have a sintered titanium bead surface. As this

prosthesis is a two-component system, a modular polyethylene insert is locked into the tibial

component. In 2007, the Agility LP Total Ankle System was introduced with some modification of its

design. All improvements were designed after careful analysis of published data to improve the52

outcome and avoid mid- and long-term complications. The new prosthesis features include: a

redesigned broad-based talar component (to avoid subsidence of the tibial component, especially in

patients with nonunion of tibiofibular syndesmosis), the ability to mix and match component sizes to

match native anatomy, and a front-loading polyethylene (easier surgical technique for exchange of

insert).52

The is the first reported three-component prosthesis with aBuechel-Pappas ankle prosthesis

mobile bearing. It is the evolution of the first-generation New Jersey ankle prosthesis. In the first38,50

Buechel-Pappas design (Mark I), the anterior-posterior constraint between the tibial and mobile

bearing components was removed. This shallow sulcus design allowed more ROM without

compromising the intrinsic sagittal stability of the ankle replacement. Postoperative complications

included delayed wound healing, reflex sympathetic dystrophy, deep infection, mobile bearing

subluxation, talar component subsidence, severe bearing wear, malleolar fracture, and osteolysis.

Analysis of complications from using this prosthesis led to modifications resulting in the Mark II

Buechel-Pappas prosthesis. This new design (also known as the deep sulcus design) included two

fins, a thicker meniscal component, and deeper sulcus with a gap in the plastic.

The (STAR) was developed as a two-component,Scandinavian Total Ankle Replacementanatomic, unconstrained resurfacing ankle prosthesis with congruent parts covering the medial and

lateral facet joints. Since 1986, the tibial part of the STAR prosthesis has included a polyethylene53

component. This modification was performed to minimize rotational stress at the implant-bone53

interface. The current design of the STAR prosthesis is a congruent, cylindrical, three-component

prosthesis. Initial osseous integration of the prosthesis is secured by a single fin on the talar side and

by two cylindrical fins on the tibial side. Both metallic components have hydroxyapatite coated

surfaces. The STAR prosthesis, one of the most popular TARs used in Europe, has one of the longest

histories in ankle replacement surgery, with several modifications made during its clinical use.54

66.4 Results of First- and Second-Generation TARs

The majority of first-generation TAR designs were two-component prostheses that used cement fixation on

both the talar and tibial sides. The reason for cement fixation was simple: in the 1970s cementless fixation

had not been widely used in knee and hip prostheses and cement fixation led to acceptable early component

stability. However, an extremely high complication rate was observed with increased incidence of loosening,

wide osteolysis, subsidence, and mechanical failure of prosthesis components.

OrthopaedicsOne Articles

Page of 322 372

Cement fixation required a larger bone resection. Therefore, bone quality at the cement-bone interface was

not optimal, as the main load transfer occurred on the weaker methaphyseal bone. Most TARs included a

polyethylene concave tibial and a metal convex component for the talus, usually made using cobalt chrome

alloy. Both types of prostheses – constrained and unconstrained – were available at that time. One design

feature of most first-generation TAR designs was a tibial component that was significantly larger than the

talar. The idea was to allow physiological dorsi-/plantar flexion ROM as well as axial rotation. However, due

to low intrinsic stability, significant increases in shear forces occurred. This was especially true in patients

with chronic ligamental instability which resulted in prosthesis loosening early on.

Most clinical studies addressing outcomes in patients who underwent first-generation TAR were case reports

or studies including between 20 to 40 patients. Another critical factor was the short follow-up periods55

reported (mostly 5 years or less). Patient satisfaction with first generation TAR was reciprocally proportional

to the length of follow-up and varied between 19% and 81%.55

Generally, the clinical results of first-generation TAR were highly discouraging. The alarmingly high

prosthesis component failure rate along with other complications like wound healing and unacceptable

functional results were the reasons that foot and ankle surgeons were advised to use ankle fusion as the

primary treatment option for ankle OA. Failure analysis of first generation TARs showed that only29,48,56-60

significant improvements in prosthetic design, change of fixation (elimination of use of cement), and

improved anatomical access would change arthroplasty outcomes.61

An analysis of the main failure reasons of the first-generation TAR designs was crucial for the development

of the second-generation ankle prostheses. More conservative and sparing bone cuts and the elimination of

bone cement have significantly improved problems with component loosening. New biologic interfaces with

special porous coatings for bony ingrowth and/or adding of hydroxyapatite were investigated as another

possible method to ensure the primary prosthesis fixation. To reduce subsidence, second-generation62,63

ankle prostheses were designed to increase the surface area of the metallic components. Increased surface

area sought to decrease the average local contact pressure and pressure peaks during gait.

The three main second generation TAR designs – Agility, Buechel-Pappas, and STAR prostheses – have

been implanted with encouraging mid- and long-term results. Positive clinical results, high patient64

satisfaction, and acceptable survivorship of prosthesis components presented at national meetings and

published in orthopaedic literature led to rethinking that ankle fusion may not be the only one reasonable

treatment option for patients with severe ankle OA. The continued critical review of second-generation

implant failures and biomechanical studies provided important data that led to the development of modern

TAR designs.

66.5 Modern TAR Designs (Third-Generation TARs)

OrthopaedicsOne Articles

Page of 323 372

The was developed between 1994 and 1996 by Michel Bonnin. This TARSalto Total Ankle 65

represents the third-generation of cementless meniscal-bearing designs. The tibial component has a

flat surface toward the mobile bearing, allowing its free translation and rotation. The 3-mm medial rim

is designed to avoid insert impingement against the medial malleolus. For osseous integration, the

component has a keel and a fixation peg. The specific shape of the talar component mimics the

natural talar geometry with the anterior width being wider than the posterior, and the lateral flange

having a larger curvature radius than the medial. The mobile bearing is manufactured from

ultra-high-molecular-weight polyethylene (UHMWPE) and has full congruency with the talar

component in flexion and extension. All components are available in three sizes.65

The is an unconstrained, three-component system that provides inversion-eversionHINTEGRA TAR

stability and was designed in 2000 by Beat Hintermann, Greta Dereymaeker, Ramon Viladot, and

Patrice Diebold. The mobile bearing provides axial rotation and normal flexion-extension mobility.66-68

The HINTEGRA TAR includes two metallic components and an ultrahigh-density polyethylene mobile

bearing. The non-articulating surfaces have a porous coating with 20% porosity and are covered by

titanium fluid and hydroxyapatite. The tibial component has a flat, 4-mm thick loading plate with

pyramidal peaks against the tibia. Additional stability may be achieved by fixation with two screws.

The talar component is conically shaped with a smaller radius medially than laterally, mimicking the

normal anatomy of talus. It has 2.5-mm high rims on each side that ensure stable positioning and

guide the anteroposterior translation of the mobile bearing. The anterior shield of this component

increases primary bone support, especially in cases with weaker bone, and may prevent the

adherence of scar tissue and avoid restriction of ROM in cases with arthrofibrosis.67

The was developed by Pascal Rippstein, Peter Wood, and Chris Coetzee. ThisMobility Ankle System

is a three-component Buechel-Pappas type prosthesis with a short, conical tibial stem. The talar

component of the Mobility implant resurfaces the superior dome of the talus, while the medial and

lateral aspects of the talus remain untreated (unlike the Buechel-Pappas prosthesis). The talar

component has a central, longitudinal sulcus and two fins, enhancing its intrinsic stability. The

non-articulating surfaces are porous coated with a titanium spray.

The was developed in 1987 and first implanted in 1989 by a French design group.Ramses TAR 69,70

The Ramses TAR is a three-component, semi-constrained prosthesis with the high-density mobile

bearing. Initially, a cemented fixation of prosthesis was used between 1980 and 2000.70

Since its introduction in 1975, the has undergone many modifications toTNK total ankle replacement

address the material of the components (stainless steel, polyethylene, alumina ceramic), coating

(without/with hydroxyapatite), and fixation (cement/cementless fixation). Currently, this is the only

TAR design containing alumina ceramic components. While the studies by the designer reported

favorable results using the third-generation TNK prosthesis, independent studies addressing71,72

TAR results in patients with rheumatoid OA show less promising results.73,74

OrthopaedicsOne Articles

Page of 324 372

The is a further development of the Buechel-Pappas-typeAnkle Evolutive System (AES) TAR

prosthesis. This design has a modular stem and allows hemi-replacement of the medial tibiotalar and

talofibular joints. This prosthesis has been widely used in England and France, and has also75 76,77

been introduced in Norway. Recent studies of the AES TAR reported a high rate of osteolytic78

lesions. It is still unclear whether the osteolysis process is a result of failure of the hydroxyapatite79-82

coating of the metal components or failure of the mobile bearing. As a result of independently

published results showing high osteolysis rate, the AES prosthesis has been withdrawn from the

market.83

The was developed in the late 1990s by Leardini et al. This prosthesis is aBOX TAR

three-component implant with metal components fixed to the proximal talus and the distal tibia and

interposed UHMWPE meniscal bearing. The biomechanical development of this prosthesis type has

been well documented in the literature by its designers.84-86

The is a two-component prosthesis designed for cementless implantation between 1985ESKA ankle

and 1989. The following features were included to improve biomechanics of the replaced ankle:87,88

cementless implantation and porous-structured implant surface for faster osteointegration, shear force

reduction by shape design of both metallic components, and easy replacement of the polyethylene

without disturbing prosthesis anchoring. Because of the ridge-like shaping and its transverse87,88

anchoring peg in both metallic components, a lateral or in special cases, medial malleolar approach

has to be used for implantation.87

Other modern TAR designs include:

, a three-component prosthesis allowing rotation around each of theGerman Ankle Systemthree possible movement axes

, a non-constrained Buechel-Pappas type design with aAlphanorm Total Ankle Replacement90° tibial stem without inclination

, which has a titanium coating and is optionally available withTARIC Total Ankle Replacementan additional hydroxyapatite coating

, a fixed-bearing, two-component total ankle system with a modular stem systemINBONE TARfor both metallic components

66.6 The TAR Learning Curve

It has been shown that there is a steep learning curve associated with performing TAR. The following

intraoperative complications have been commonly reported:

Medial and/or lateral malleolar fractures

Laceration to the tendons (posterior tibial tendon, flexor digitorum/hallucis longus)

Nerve injures (deep/superficial peroneal nerve) 89-92

The influence of surgeon experience on complication rates in patient receiving the Agility prosthesis was

examined by Saltzman et al. The perioperative records of the first 10 cases of nine surgeons with different91

training levels were recorded. The authors did not identify any specific training method that significantly

decreased complication rates.91

OrthopaedicsOne Articles

Page of 325 372

Myerson and Mocked performed a retrospective radiographic and chart review of 50 arthroplasties

performed by the same surgeon using the Agility prosthesis. Patients were divided into two groups due to90

surgeon’s experience, each including 25 patients. The number of minor wound complications decreased

from six in the first group to two in the second group. Also, the number of intraoperative fractures was

different in favor of the second group (five vs. two fractures). The nerve or tendon lacerations (n=4) all

occurred in the first group. Regarding the overall decreased complication rate in the second group of 25

patients, the authors stated that there was a notable learning curve in TAR performance.90

A similar study with 50 patients who underwent Agility TAR was performed by Schuberth et al. In this93

study, patients were also divided into two 25-patient groups. There was a significantly decreased rate of the

following complications with increased surgeon experience: medial and lateral fractures, major revisions, and

malpositioning of prosthesis components. Schutte and Louwerens reported their initial results obtained in93

49 patients who received a STAR prosthesis. The following intraoperative complications were detected: six92

fractures of the medial and two of the lateral malleolus, three fractures of the distal tibia, one injury of the

peroneal nerve, and two malpositions of the tibial and two of talar components. Based on these numbers,

the authors concluded that TAR should be performed only by an experienced orthopaedic foot and ankle

surgeon.92

The aforementioned studies addressed intraoperative complications in patients receiving second-generation

TARs. Similar studies have been conducted for patient groups receiving third-generation TARs.89-92 89,94

Lee et al addressed the perioperative complications in the 25 initial patients to receive the HINTEGRA

prosthesis and compared these results with those from a subsequent 25 cases. In the first group,89

perioperative complications occurred in 60% of all cases, while in the second group, only five complications

(20%) were observed. All major complications (deep infection and aseptic loosening) occurred in the first

group. The rate of minor complications (fractures, minor wound problem, nerve/tendon injuries, and

heterotopic ossifications) significantly decreased in the second group. However, the authors were unable to

show a decrease in the number of malpositioning of prosthesis components as a result of increased surgeon

experience. The same working group compared the perioperative complication of the HINTEGRA total89

ankle system with the MOBILITY total ankle system. The authors did not find any differences in94

perioperative complications between the two total ankle systems, but medial malleolar fractures did occur

more frequently when using the MOBILITY prosthesis.94

Recently, Reuver et al addressed short-term results of TARs performed in low-volume arthroplasty centers.

In total, 64 TARs were performed using Salto implants between 2003 and 2007 at four low-volume95

centers. Fifty-five patients (59 ankles) were reviewed at a mean follow up of 36 months. Seven ankles had to

undergo revision surgery – two revision arthroplasties and five fusions – because of loosening, and two

cases of deep infection resulting in a survivorship of 86% at final follow up. Significant pain relief and

functional improvement were observed in this review, as assessed using VAS and AOFAS score. The

authors felt confident that results of TAR performed in low-volume centers are comparable to most

high-volume centers. However, the survival of implanted components was significantly lower, especially

regarding the relatively short follow up.95

In summary, despite the encouraging results reported by studies using modern third-generation TAR, we

believe, that TAR should be limited to foot and ankle orthopaedic surgeons with special training and

adequate experience in arthroplasty techniques.

OrthopaedicsOne Articles

Page of 326 372

See for more on the use of TAR in painful ankle arthrodesisSpecial Situations for Total Ankle Replacement

and simultaneous bilateral TAR.

66.7 Modern TAR Designs: Promising Results

Stengel et al performed a systematic review and meta-analysis to address the efficacy of TAR with

meniscal-bearing implants. The following inclusion criteria were defined for this study: a minimum sample96

size of 20 subjects, at least 1 year of follow up, and a clinically relevant study endpoint. In total, 18 studies

with 1,086 patients were included in the review. Most patients experienced significant functional

improvement (average 45.2 points using standardized 100-point ankle and hindfoot scores) and a mild

increase of ROM (mean increase = 6.3%, 95% CI, 2.2% - 10.5%). Weighted complication rates ranged from

1.6% (deep infection) to 14.7% (impingement). Secondary surgeries were necessary in 12.5% of all patients.

Ankle fusions were required in 6.3% of patients due to implant failure, resulting in 1- and 5-year survivorship

of 96.9% (95% CI, 94.9% - 98.8%) and 90.6% (95% CI, 84.1% - 97.1%), respectively. The data of this

meta-analysis showed that TARs using current three-component designs provide an acceptable benefit-risk

ratio. However, the results should be interpreted with caution due to non-optimal methodological quality,

sample sizes, and short follow ups.96

SooHoo et al compared reoperation rates following ankle fusion and TAR using California’s hospital

discharge database. A total of 4,705 ankle fusions and 480 TARs were included in the review during the97

10-year study period (1995 through 2004). It was shown that patients who underwent TAR had an increased

risk of periprosthetic infection. The rates of major revision surgery after TAR were 9% at 1 year and 23% at 5

years compared with 5% and 11% following ankle fusion. However, TAR was shown to have advantages in

terms of functional results.97

In another study from 2004, SooHoo and Kominski performed cost analyses of TAR compared to ankle

fusion. The authors performed a thorough literature review to identify possible outcomes and their98

probabilities following ankle fusion versus TAR. They found that TAR generated total expected lifetime

treatment costs of $16,568, which is $6,990 more than costs following ankle fusion. Furthermore, TAR had

an incremental cost-effectiveness ratio in the reference case of $18,419 for each quality-adjusted life year

gained.98

In 2007, Haddad et al performed a systematic literature review addressing the intermediate and long-term

outcomes of interest in TAR and ankle fusion. In total, 10 studies including 852 TAR patients and 3999

studies including 1,262 with ankle fusion were analyzed. The authors showed that 38% of patients had

excellent postoperative results, 30.5% had good results, 5.5% had fair results, and 24% had poor results.

The corresponding values in patients who underwent ankle fusion were 31%, 37%, 13%, and 13%,

respectively. The 5- and 10-year survivorship for TAR was 78% (95% CI, 69.0% - 87.6%) and 77% (95% CI,

63.3% - 90.8%), respectively. The revision rate in patients with TAR and ankle fusion was 7% (95% CI, 3.5%

- 10.9%) and 9% (95% CI, 5.5% - 11.6%), respectively. The main reasons for revision surgery were

component loosening and subsidence in the TAR group and non-union in the fusion group. The results of

this study suggest that the intermediate outcome is comparable in both procedures.99

In 2010, Gougoulias et al performed a systematic review of the literature to address the outcome of TAR

implants currently in use. In total, 13 Level IV peer-reviewed studies were included, reporting the outcome100

of 1,105 TARs:

OrthopaedicsOne Articles

Page of 327 372

234 Agility

344 STAR

153 Buechel-Pappas

152 HINTEGRA

98 Salto

70 TNK

54 Mobility implants

Postoperatively, a remarkable portion of patients still had residual pain (range, 27% - 60%). Also, superficial

wound complications and deep infection were often reported, with rates up to 14.7% and 4.6%, respectively.

Overall failure rate at 5 years had a wide range, 0% and 32%. In general, most patients experienced

significant functional improvement as assessed by the AOFAS score. However, the postoperative

improvement of ROM was relatively small (0° - 14°). Therefore, the patients should be informed that

significantly improved ROM is not one of the postoperatively expected benefits of TAR.100

Recently, Slobogean et al compared preference-based quality of life in patients with end-stage ankle OA

treated with TAR or ankle fusion. The quality of life of 107 subjects was assessed using health state101

values derived from SF-36 (SF-6D transformation). The mean baseline SF-6D health state value in the TAR

group was 0.67 (95% CI, 0.64 – 0.69) and in the ankle fusion group 0.66 (95% CI, 0.63 – 0.68). At 1 year

follow up, both groups had significant and comparable improvements with 0.73 for the TAR group (95% CI,

0.71 – 0.76) and 0.73 for the ankle fusion group (95% CI, 0.70 – 0.76).101

To date, there has been no clear evidence in the literature that three-component TAR designs are superior

compared with two-component designs. Also, the high-quality comparative studies addressing102,103

postoperative outcomes in patients who underwent TAR vs. ankle fusion are rare. There has been only one

prospective, controlled, comparative surgical trial performed including both patient cohorts. Future104

high-quality prospective, randomized, controlled studies would significantly help to establish the clinical

practice guidelines needed for surgeons to make a correct decision about treatment choice.104-108

66.8 Future Insights

TAR is increasingly gaining acceptance as a valuable option for treating patients with end-stage ankle OA.

Current reports of this procedure show consistently good to excellent mid-term results with significant pain

relief, good functional outcomes, and high patient satisfaction. The high failure rate of the100,109,110

first-generation ankle prostheses has been thoroughly analyzed and the TAR designs have been

significantly improved. Current fixation techniques without cement have become the gold standard using

“biological surfaces” (eg, introduction of hydroxyapatite in the 1990s) for better osseous integration of

metallic prosthesis components.110,111

OrthopaedicsOne Articles

Page of 328 372

1.

2.

3.

4.

5.

6.

7.

8.

9.

One of the main principles of TAR surgical technique is preserving adequate bone stock. It has been111

widely recognized, with regard to bone resection, that less is more. An extensive bone resection may

drastically limit the revision surgery in case of failure of the TAR, especially on the talar side. Also, the bone

resection for implantation of prosthesis components should consider the anatomical inner structure of bone

and especially trabecular microarchitecture for optimal load transfer. The optimal load transfer is very111

important because this would avoid pathologically increased pressure peaks, which may cause loosening

and subsidence of components. Therefore, the load transfer on a tibial component that incorporates

circumferential bony support may be superior to that of a stemmed component in the long term.110

Furthermore, the natural articular geometry of the ankle should be considered during the design of any ankle

prostheses.

Modern implants try to retain the radius of the curvature of the talus, resulting in improved and a more

natural ROM. The modern TAR is not only a resurfacing procedure of the osteoarthritic ankle but also a111

restoration of normal biomechanics of the entire hindfoot. If necessary, additional surgeries should be110

performed to achieve the appropriate ligamental and osseous balancing of the hindfoot. Failure to8-10,110

correct hindfoot alignment or undercorrection of hindfoot deformity can cause a significant increase of

translation forces and movements during gait. Especially in patients with remaining valgus misalignment, this

may lead to prosthesis failure because valgus misalignment is tolerated more poorly than varus.110

In conclusion, TAR with current devices, equipment, and techniques has improved considerably over the

past several decades to show that the ankle fusion is no longer the “gold standard” treatment for all patients

with severe end-stage ankle OA. Future biomechanical and clinical studies addressing the outcomes and

biomechanical properties of TAR should be continued with the aim of improving current TAR designs.

66.9 References

Barg,A., Knupp,M., and Hintermann,B.: Simultaneous bilateral versus unilateral total ankle

replacement: A patient-based comparison of pain relief, quality of life and functional outcome. J Bone

Joint Surg Br, 92:1659-1663, 2010.

Barg,A., Elsner,A., Hefti,D., and Hintermann,B.: Total Ankle Arthroplasty in Patients with Hereditary

Hemochromatosis. Clin Orthop Relat Res, 7-28-2010.

Barg,A., Elsner,A., Hefti,D., and Hintermann,B.: Haemophilic arthropathy of the ankle treated by total

ankle replacement: a case series. Haemophilia, 16:647-655, 7-1-2010.

Kim,B.S., Knupp,M., Zwicky,L., Lee,J.W., and Hintermann,B.: Total ankle replacement in association

with hindfoot fusion: Outcome and complications. J Bone Joint Surg Br, 92:1540-1547, 2010.

Doets,H.C., Brand,R., and Nelissen,R.G.: Total ankle arthroplasty in inflammatory joint disease with

use of two mobile-bearing designs. J Bone Joint Surg Am, 88:1272-1284, 2006.

Wood,P.L. and Deakin,S.: Total ankle replacement. The results in 200 ankles. J Bone Joint Surg Br,

85:334-341, 2003.

Wood,P.L., Crawford,L.A., Suneja,R., and Kenyon,A.: Total ankle replacement for rheumatoid ankle

arthritis. Foot Ankle Clin, 12:497-508, 2007.

Kim,B.S. and Lee,J.W.: Total ankle replacement for the vaus unstable osteoarthritic ankle. Tech Foot

& Ankle, 9:157-167, 2010.

Brunner,S., Knupp,M., and Hintermann,B.: Total ankle replacement for the valgus unstable

osteoarthritic ankle. Tech Foot & Ankle, 9:174, 2010.

OrthopaedicsOne Articles

Page of 329 372

10.

11.

12.

13.

14.

15.

16.

17.

18.

19.

20.

21.

22.

23.

24.

25.

26.

27.

28.

29.

30.

31.

32.

Knupp,M., Stufkens,S.A., Bolliger,L., Brunner,S., and Hintermann,B.: Total ankle replacement and

supramalleolar osteotomies for malaligned osteoarthritis ankle. Tech Foot & Ankle, 9:175-181, 2010.

Deorio,J.K.: Total ankle replacement with subtalar arthrodesis: management of combined ankle ans

subtalar arthritis. Tech Foot & Ankle, 9:182-189, 2010.

Hobson,S.A., Karantana,A., and Dhar,S.: Total ankle replacement in patients with significant

pre-operative deformity of the hindfoot. J Bone Joint Surg Br, 91:481-486, 2009.

Karantana,A., Hobson,S., and Dhar,S.: The scandinavian total ankle replacement: survivorship at 5

and 8 years comparable to other series. Clin Orthop Relat Res, 468:951-957, 2010.

Daniels,T.R., Cadden,A.R., and Lim,K.: Correction of varus talar deformities in ankle joint

replacement. Oper Tech Orthop, 18:282-286, 2008.

Kim,B.S., Choi,W.J., Kim,Y.S., and Lee,J.W.: Total ankle replacement in moderate to severe varus

deformity of the ankle. J Bone Joint Surg Br, 91:1183-1190, 2009.

Thomas,R.H. and Daniels,T.R.: Ankle arthritis. J Bone Joint Surg Am, 85-A:923-936, 2003.

Guyer,A.J. and Richardson,G.: Current concepts review: total ankle arthroplasty. Foot Ankle Int,

29:256-264, 2008.

Newton,S.E., III: An artificial ankle joint. Clin Orthop Relat Res, 424:3-5, 2004.

van den,H.A., Van Bouwel,S., and Dereymaeker,G.: Total ankle replacement. Design evolution and

results. Acta Orthop Belg, 76:150-161, 2010.

Whalen,J.L., Spelsberg,S.C., and Murray,P.: Wound breakdown after total ankle arthroplasty. Foot

Ankle Int, 31:301-305, 2010.

Eichinger,S., Forst,R., and Kindervater,M.: Indications and alternatives for arthroplasty in young

patients. Orthopade, 36:311-324, 2007.

Espinosa,N. and Klammer,G.: Treatment of ankle osteoarthritis: arthrodesis versus total ankle

replacement. Eur J Trauma Emerg Surg, 36:epub ahead of print, 2010.

Anderson,T., Montgomery,F., and Carlsson,A.: Uncemented STAR total ankle prostheses. J Bone

Joint Surg Am, 86-A Suppl 1:103-111, 2004.

Bauer,G., Eberhardt,O., Rosenbaum,D., and Claes,L.: Total ankle replacement. Review and critical

analysis of the current status. Foot Ankle Surg, 2:119-126, 1996.

Clare,M.P. and Sanders,R.W.: Preoperative considerations in ankle replacement surgery. Foot Ankle

Clin, 7:709-720, 2002.

Jackson,M.P. and Singh,D.: Total ankle replacement. Current Orthopaedics, 17:292-298, 2003.

Murnaghan,J.M., Warnock,D.S., and Henderson,S.A.: Total ankle replacement. Early experiences

with STAR prosthesis. Ulster Med J, 74:9-13, 2005.

Myerson,M.S. and Won,H.Y.: Primary and revision total ankle replacement using custom-designed

prostheses. Foot Ankle Clin, 13:521-538, 2008.

Newton,S.E., III: Total ankle arthroplasty. Clinical study of fifty cases. J Bone Joint Surg Am,

64:104-111, 1982.

Tsukamoto,S., Tanaka,Y., Maegawa,N., Shinohara,Y., Taniguchi,A., Kumai,T., and Takakura,Y.:

Total talar replacement following collapse of the talar body as a complication of total ankle

arthroplasty: a case report. J Bone Joint Surg Am, 92:2115-2120, 9-1-2010.

Hintermann,B.: History of total ankle arthroplasty, in Hintermann,B. (ed), Total ankle arthroplasty:

Historical overview, current concepts and future perspectives. Wien New York , Springer, 2005, pp.

43-57.

Henne,T.D. and Anderson,J.G.: Total ankle arthroplasty: a historical perspective. Foot Ankle Clin,

7:695-702, 2002.

OrthopaedicsOne Articles

Page of 330 372

33.

34.

35.

36.

37.

38.

39.

40.

41.

42.

43.

44.

45.

46.

47.

48.

49.

50.

51.

52.

53.

54.

55.

56.

57.

Muir,D.C., Amendola,A., and Saltzman,C.L.: Long-term outcome of ankle arthrodesis. Foot Ankle

Clin, 7:703-708, 2002.

Bonasia,D.E., Dettoni,F., Femino,J.E., Phisitkul,P., Germano,M., and Amendola,A.: Total ankle

replacement: why, when and how? Iowa Orthop J, 30:119-130, 2010.

Giannini,S., Leardini,A., and O'Connor,J.J.: Total ankle replacement: review of the design and of the

current status. Foot Ankle Surg, 6:77-88, 2000.

Gougoulias,N.E., Khanna,A., and Maffulli,N.: History and evolution in total ankle arthroplasty. Br Med

Bull, 89:111-151, 2009.

van den Heuvel,A., Van Bouwel,S., and Dereymaeker,G.: Total ankle replacement. Design evolution

and results. Acta Orthop Belg, 76:150-161, 2010.

Vickerstaff,J.A., Miles,A.W., and Cunningham,J.L.: A brief history of total ankle replacement and a

review of the current status. Med Eng Phys, 29:1056-1064, 2007.

Lord,G. and Marotte,J.H.: Total ankle prosthesis. Technic and 1st results. Apropos of 12 cases. Rev

Chir Orthop Reparatrice Appar Mot, 59:139-151, 1973.

Muir,D.C., Amendola,A., and Saltzman,C.L.: Forty-year outcome of ankle "cup" arthroplasty for

post-traumatic arthritis. Iowa Orthop J, 22:99-102, 2002.

Lord,G. and Marotte,J.H.: Total ankle replacement (author's transl). Rev Chir Orthop Reparatrice

Appar Mot, 66:527-530, 1980.

Kempson,G.E., Freeman,M.A., and Tuke,M.A.: Engineering considerations in the design of an ankle

joint. Biomed Eng, 10:166-71, 80, 1975.

Freeman,M.A., Kempson,G.E., and Tuke,M.A.: Total replacement o fthe ankle with the ICLH

prosthesis. Int Orthop, 2:237-331, 1979.

Waugh,T.R., Evanski,P.M., and McMaster,W.C.: Irvine ankle arthroplasty. Prosthetic design and

surgical technique. Clin Orthop Relat Res, 114:180-184, 1976.

Kirkup,J.: Richard Smith ankle arthroplasty. J R Soc Med, 78:301-304, 1985.

Matejczyk,M.B., Greenwald,A.S., and Black,J.D.: Ankle implant systems: laboratory evaluation and

clinical correlation. Orthop Trans, 3:199, 1979.

Stauffer,R.N., Chao,E.Y., and Brewster,R.C.: Force and motion analysis of the normal, diseased, and

prosthetic ankle joint. Clin Orthop Relat Res, 127:189-196, 1977.

Stauffer,R.N. and Segal,N.M.: Total ankle arthroplasty: four years' experience. Clin Orthop Relat Res,

160:217-221, 1981.

Wood,P.L.R., Clough,T.M., and Jari,S.: Clinical comparison of two total ankle replacements. Foot

Ankle Int, 21:546-550, 2000.

Pappas,M., Buechel,F.F., and DePalma,A.F.: Cylindrical total ankle joint replacement: surgical and

biomechanical rationale. Clin Orthop Relat Res, 118:82-92, 1976.

Alvine,F.G.: The Agility ankle replacement: the good and the bad. Foot Ankle Clin, 7:737-753, 2002.

Cerrato,R. and Myerson,M.S.: Total ankle replacement: the Agility LP prosthesis. Foot Ankle Clin,

13:485-494, 2008.

Kofoed,H.: Scandinavian Total Ankle Replacement (STAR). Clin Orthop Relat Res, 424:73-79, 2004.

Gittins,J. and Mann,R.A.: The history of the STAR total ankle arthroplasty. Foot Ankle Clin, 7:809-816,

2002.

Saltzman,C.L.: Perspective on total ankle replacement. Foot Ankle Clin, 5:761-775, 2000.

Carlsson,A.S., Henricson,A., Linder,L., Hilsson,J.A., and Redlund-Johnell,I.: A 10-year analysis of 69

Bath and Wessex ankle replacements. Foot Ankle Surg, 7:39-44, 2001.

Kitaoka,H.B., Patzer,G.L., Ilstrup,D.M., and Wallrichs,S.L.: Survivorship analysis of the Mayo total

ankle arthroplasty. J Bone Joint Surg Am, 76:974-979, 1994.

OrthopaedicsOne Articles

Page of 331 372

58.

59.

60.

61.

62.

63.

64.

65.

66.

67.

68.

69.

70.

71.

72.

73.

74.

75.

76.

77.

78.

79.

Kitaoka,H.B. and Patzer,G.L.: Clinical results of the Mayo total ankle arthroplasty. J Bone Joint Surg

Am, 78:1658-1664, 1996.

Demottaz,J.D., Mazur,J.M., Thomas,W.H., Sledge,C.B., and Simon,S.R.: Clinical study of total ankle

replacement with gait analysis. A preliminary report. J Bone Joint Surg Am, 61:976-988, 1979.

Wynn,A.H. and Wilde,A.H.: Long-term follow-up of the Conaxial (Beck-Steffee) total ankle

arthroplasty. Foot Ankle, 13:303-306, 1992.

Hamblen,D.L.: Can the ankle joint be replaced? J Bone Joint Surg Br, 67:689-690, 1985.

Zerahn,B., Kofoed,H., and Borgwardt,A.: Increased bone mineral density adjacent to

hydroxy-apatite-coated ankle arthroplasty. Foot Ankle Int, 21:285-289, 2000.

Zerahn,B. and Kofoed,H.: Bone mineral density, gait analysis, and patient satisfaction, before and

after ankle arthroplasty. Foot Ankle Int, 25:208-214, 2004.

Rippstein,P.F.: Clinical experiences with three different designs of ankle prostheses. Foot Ankle Clin,

7:817-831, 2002.

Bonnin,M., Judet,T., Colombier,J.A., Buscayret,F., Graveleau,N., and Piriou,P.: Midterm results of the

Salto Total Ankle Prosthesis. Clin Orthop Relat Res, 424:6-18, 2004.

Valderrabano,V., Pagenstert,G.I., and Hintermann,B.: Total ankle replacement - three-component

prosthesis. Tech Foot & Ankle, 2:84-90, 2005.

Hintermann,B. and Barg,A.: The HINTEGRA total ankle arthroplasty, in Wiesel,S.W. (ed), Operative

Techniques in Orthopaedic Surgery Lippincott Williams & Wilkins, 2010, pp. 4022-4031.

Hintermann,B.: Surgical techniques, in Hintermann,B. (ed), Total ankle arthroplasty: Historical

overview, current concepts and future perspectives. Springer , Wien New York, 2005, pp. 105-126.

Mendolia,G. and Talus Group: The Ramses ankel replacement: design-surgical technique result,

results in first 38 cases. The French Orthopedic Web Journal (available at: www matrise-orthop com),

2007.

Mendolia,G., Coillard,J.Y., Cermolacce,C., and Determe,P.: Long-term (10 to 14 years) results of the

Ramses total ankle arthroplasty. Tech Foot & Ankle, 4:160-173, 2005.

Takakura,Y., Tanaka,Y., Kumai,T., Sugimoto,K., and Ohgushi,H.: Ankle arthroplasty using three

generations of metal and ceramic prostheses. Clin Orthop Relat Res, 424:130-136, 2004.

Tanaka,Y. and Takakura,Y.: The TNK ankle: short- and mid-term results. Orthopade, 35:546-551,

2006.

Nishikawa,M., Tomita,T., Fujii,M., Watanabe,T., Hashimoto,J., Sugamoto,K., Ochi,T., and

Yoshikawa,H.: Total ankle replacement in rheumatoid arthritis. Int Orthop, 28:123-126, 2004.

Nagashima,M., Takahashi,H., Kakumoto,S., Miyamoto,Y., and Yoshino,S.: Total ankle arthroplasty for

deformity of the foot in patients with rheumatoid arthritis using the TNK ankle

Hintermann,B.: Current designs of total ankle prostheses, in Hintermann,B. (ed), Total ankle

arthroplasty. Historical overview, current concepts and future perspectives. Wien , Springer, 2004, pp.

69-100.ystem: clinical results of 21 cases. Mod Rheumatol, 14:48-53, 2004.

Goldberg,A.J., Sharp,R.J., and Cooke,P.: Ankle replacement: current practice of foot & ankle

surgeons in the United kingdom. Foot Ankle Int, 30:950-954, 2009.

Besse,J.L., Colombier,J.A., Asencio,J., et al.: Total ankle arthroplasty in France. Orthop Traumatol

Surg Res, 96:291-303, 2010.

Fevang,B.T., Lie,S.A., Havelin,L.I., Brun,J.G., Skredderstuen,A., and Furnes,O.: 257 ankle

arthroplasties performed in Norway between 1994 and 2005. Acta Orthop, 78:575-583, 2007.

Morgan,S.S., Brooke,B., and Harris,N.J.: Total ankle replacement by the Ankle Evolution System:

medium-term outcome. J Bone Joint Surg Br, 92:61-65, 2010.

OrthopaedicsOne Articles

Page of 332 372

80.

81.

82.

83.

84.

85.

86.

87.

88.

89.

90.

91.

92.

93.

94.

95.

96.

97.

98.

99.

100.

Besse,J.L., Brito,N., and Lienhart,C.: Clinical evaluation and radiographic assessment of bone lysis of

the AES total ankle replacement. Foot Ankle Int, 30:964-975, 2009.

Koivu,H., Kohonen,I., Sipola,E., Alanen,K., Vahlberg,T., and Tiusanen,H.: Severe periprosthetic

osteolytic lesions after the Ankle Evolutive System total ankle replacement. J Bone Joint Surg Br,

91:907-914, 2009.

Rodriguez,D., Bevernage,B.D., Maldague,P., Deleu,P.A., Tribak,K., and Leemrijse,T.: Medium term

follow-up of the AES ankle prosthesis: High rate of asymptomatic osteolysis. Foot Ankle Surg,

16:54-60, 2010.

Smith,T.W. and Stephens,M.: Ankle arthroplasty. Foot Ankle Surg, 16:53, 2010.

Leardini,A.: Geometry and mechanics of the human ankle complex and ankle prosthesis design. Clin

Biomech (Bristol , Avon ), 16:706-709, 2001.

Leardini,A., O'Connor,J.J., Catani,F., and Giannini,S.: Mobility of the human ankle and the design of

total ankle replacement. Clin Orthop Relat Res, 424:39-46, 2004.

Reggiani,B., Leardini,A., Corazza,F., and Taylor,M.: Finite element analysis of a total ankle

replacement during the stance phase of gait. J Biomech, 39:1435-1443, 2006.

Rudiger,J.: Ankle replacement by the cementless ESKA endoprosthesis. Tech Foot & Ankle,

4:125-136, 2005.

Rudiger,J., Menzinger,F., and Grundei,H.: 14 Jahre Erfahrungen mit der zementfreien

ESKA-Sprunggelenksendoprothese. Fuss Sprungg, 2:65-75, 2004.

Lee,K.B., Cho,S.G., Hur,C.I., and Yoon,T.R.: Perioperative complications of HINTEGRA total ankle

replacement: our initial 50 cases. Foot Ankle Int, 29:978-984, 2008.

Myerson,M.S. and Mroczek,K.: Perioperative complications of total ankle arthroplasty. Foot Ankle Int,

24:17-21, 2003.

Saltzman,C.L., Amendola,A., Anderson,R., et al.: Surgeon training and complications in total ankle

arthroplasty. Foot Ankle Int, 24:514-518, 2003.

Schutte,B.G. and Louwerens,J.W.: Short-term results of our first 49 Scandanavian total ankle

replacements (STAR). Foot Ankle Int, 29:124-127, 2008.

Schuberth,J.M., Patel,S., and Zarutsky,E.: Perioperative complications of the Agility total ankle

replacement in 50 initial, consecutive cases. J Foot Ankle Surg, 45:139-146, 2006.

Lee,K.T., Lee,Y.K., Young,K.W., Kim,H.J., Park,S.Y., Kim,J.S., and Kim,K.C.: Perioperative

complications of the MOBILITY total ankle system: comparison with the HINTEGRA total ankle

system. J Orthop Sci, 15:317-322, 2010.

Reuver,J.M., Dayerizadeh,N., Burger,B., Elmans,L., Hoelen,M., and Tulp,N.: Total ankle replacement

outcome in low volume centers: short-term followup. Foot Ankle Int, 31:1064-1068, 2010.

Stengel,D., Bauwens,K., Ekkernkamp,A., and Cramer,J.: Efficacy of total ankle replacement with

meniscal-bearing devices: a systematic review and meta-analysis. Arch Orthop Trauma Surg,

125:109-119, 2005.

SooHoo,N.F., Zingmond,D.S., and Ko,C.Y.: Comparison of reoperation rates following ankle

arthrodesis and total ankle arthroplasty. J Bone Joint Surg Am, 89:2143-2149, 2007.

SooHoo,N.F. and Kominski,G.: Cost-effectiveness analysis of total ankle arthroplasty. J Bone Joint

Surg Am, 86-A:2446-2455, 2004.

Haddad,S.L., Coetzee,J.C., Estok,R., Fahrbach,K., Banel,D., and Nalyshyk,L.: Intermediate and

long-term outcomes of total ankle arthroplasty and ankle arthrodesis. A systematic review of the

literature. J Bone Joint Surg Am, 89:1899-1905, 2007.

Gougoulias,N., Khanna,A., and Maffulli,N.: How successful are current ankle replacements?: a

systematic review of the literature. Clin Orthop Relat Res, 468:199-208, 2010.

OrthopaedicsOne Articles

Page of 333 372

101.

102.

103.

104.

105.

106.

107.

108.

109.

110.

111.

Slobogean,G.P., Younger,A., Apostle,K.L., Marra,C.A., Wing,K., Penner,M., Daniels,T., and

Glazebrook,M.: Preference-based quality of life of end-stage ankle arthritis treated with arthroplasty or

arthrodesis. Foot Ankle Int, 31:563-566, 2010.

Dahabreh,Z., Gonsalves,S., Monkhouse,R., and Harris,N.J.: Extrusion of metal radiological marker

from a total ankle replacement insert: a case report. J Foot Ankle Surg, 45:185-189, 2006.

Younger,A., Penner,M., and Wing,K.: Mobile-bearing total ankle arthroplasty. Foot Ankle Clin,

13:495-504, 2008.

Saltzman,C.L., Mann,R.A., Ahrens,J.E., et al.: Prospective controlled trial of STAR total ankle

replacement versus ankle fusion: initial results. Foot Ankle Int, 30:579-596, 2009.

Pinzur,M.: FootForum: clinical practice guidelines. Foot Ankle Int, 31:275-276, 2010.

Pinzur,M.S.: FootForum: experimental surgery. Foot Ankle Int, 30:472-473, 2009.

Pinzur,M.: FootForum: experimental surgery revisited. Foot Ankle Int, 31:102, 2010.

Pinzur,M.: FootForum: the power of evidence. Foot Ankle Int, 31:468, 2010.

Chou,L.B., Coughlin,M.T., Hansen S Jr, Haskell,A., Lundeen,G., Saltzman,C.L., and Mann,R.A.:

Osteoarthritis of the ankle: the role of arthroplasty. J Am Acad Orthop Surg, 16:249-259, 2008.

Hintermann,B.: Ankle osteoarthritis: Five take-home points regarding total ankle arthroplasty in the

rest of the world. 26th Annual Summer Meeting of the American Orthopaedic Foot & Ankle Society,

2010.

Saltzman,C.L.: Ankle osteoarthritis: Five take-home points regarding total ankle arthroplasty in USA.

26th Annual Summer Meeting of the American Orthopaedic Foot & Ankle Society, 2010

66.10 AES Total Ankle Replacement

This article is taken wholly from, or contains information that was originally published by the

American Orthopaedic Foot and Ankle Society.

more

Section editors and authors for Orthopaedia may have edited its content or added new information. The use

of information from the American Orthopaedic Foot and Ankle Society should not be construed as support

for or endorsement by that organization for any new information added by Orthopaedia members, or for any

editing of the original content.

Learn about the AOFAS

The Ankle Evolutive System (AES) TAR is a further development of the Buechel-Pappas-type prosthesis.

This design has a modular stem and allows hemi-replacement of the medial tibiotalar and talofibular joints.1

This prosthesis type has been widely used in England and France and has also been introduced in2,3

Norway. While some studies show promising functional results, this prosthesis has experienced significant4

problems relating to osteolysis and has been withdrawn from the market.17

OrthopaedicsOne Articles

Page of 334 372

One of the first reports addressing the outcome in patients who underwent AES TAR is a study by Patsalis.5

This study included 15 AES TAR patients with an average short-term follow up of 8.5 months. Three

malleolar fractures were observed intraoperatively. Two replaced ankles had to be revised, while the

remaining 13 patients showed significant functional improvement as assessed by the AOFAS score.5

In 2005, a short-term study by the designers of the AES prosthesis outlined the suggested surgical

technique and reported good preliminary results.6

Henricson and Ågren addressed the influence of preoperative hindfoot alignment on secondary surgery after

TAR. The patient cohort included 109 STAR, 62 Buechel-Pappas, and 22 AES TARs. The mean follow-up7

time in the whole patient group was 4.2 years.Two cases with instability requiring additional surgery (one

case with preoperative varus and one case with preoperative valgus deformity) were observed in the AES

TAR group. No revision surgeries addressing prosthesis loosening were necessary.7

Brooke et al presented two cases with valgus malalignment following TAR using an AES prosthesis. They8

performed a fibula lengthening osteotomy to regain the anatomical alignment and to avoid pathological

increases in wear. In both cases, good clinical and radiographic results were achieved. Kharwadkar and8

Harris published a report of two cases using the AES tibial component as a revision component in patients

with failed STAR prostheses. The hybrid AES-STAR revision procedures were performed at 4 and 7 years9

following the primary STAR prosthesis implantation. The mid-term results were satisfactory, with no

restriction of daily activities.9

Dahabreh et al published a case report with a patient having revision surgery due to extrusion of a metal

radiological marker. The exchange of polyethylene insert was performed 9 months after the initial TAR10

using an AES implant. During revision surgery, the insert was found to be intact without fracture. Morgan10

et al presented 2.5 years' follow-up results in a female patient who had poliomyelitis as a child and had been

treated with the AES prosthesis because of a painful, degenerate ankle with preoperatively significant varus

deformity. The patient showed a satisfactory functional outcome with a well-aligned replacement ankle and11

no evidence of loosening or osteolysis around prosthesis components.11

Anders et al reported their mid-term results in 94 patients who underwent AES TAR between 2002 and

2007. One patient was deceased, leaving 93 ankles for evaluation at a mean follow up of 3.5 years. There12

were five intraoperative malleolar fractures, which were all secured with screws. One patient was revised

after 5.5 years due to loosening of both metallic components, and two patients with loosening of the tibial

component were pending revisions. In an additional three cases osteolysis, around the component was

seen. Two ankles were revised due to fixed varus or valgus deformity. In one patient, an ankle fusion was

performed because of a fracture of the distal tibia. Two patients were revised for deep infection. Overall, the

cumulative 5-year survivorship with revision for any reason was 90%. In summary, the authors stated that

mid-term results in their patient cohort were promising.12

Morgan et al presented the outcomes of 38 consecutive patients who were treated with AES TAR between

2002 and 2004. All patients were reviewed clinically and radiographically at a minimum follow up of 413

years. Most patients presented with significantly improved function and pain relief. Two replaced ankles were

converted to ankle fusion resulting in a cumulative 6-year survivorship of 94.7% (95% CI, 80.3% - 98.7%).

Ten patients presented with edge-loading, of whom nine had corrective surgery. In nine patients significant

osteolysis around the prosthesis components was seen. Because of non-progressive symptoms no further

revision surgeries were suggested.

OrthopaedicsOne Articles

Page of 335 372

1.

2.

3.

4.

5.

6.

7.

Despite high patient satisfaction, the authors reported some concerns about an observed high rate of

osteolysis. The reported high rate of osteolysis in patients who underwent AES TAR has been confirmed13

by other studies. Besse et al reported mid-term results of their prospective study including 51 AES implants

performed from 2003 to 2006. All patients were reviewed at a mean follow-up of 40 months. Eighty-two14

percent of all patients had good functional outcome showing a significant postoperative improvement of

AOFAS score. In two patients, replaced ankles were converted to ankle fusion because of talar component

subsidence and mechanical dislocation. Although the functional outcome and patient satisfaction were

comparable to other results published in studies using other third generation prostheses, significant

osteolysis with the AES TAR was more frequent with risk of subsidence. As a consequence, the authors

stopped implantation of this prosthesis type and recommended a preventive grafting for severe osteolysis.14

The comparably high osteolysis rate was also seen in a study by Koivu et al reviewing 130 consecutive AES

implants performed between 2002 and 2008. Radiolucent lines or osteolytic lesions were seen on plain15

radiographs in 48 ankles (37%). Marked osteolytic lesions were found in 27 ankles (21%). The talar

component migrated in nine ankles; in an additional two ankles, a shift of the tibial component was observed.

Of the 27 ankles with marked osteolysis, 16 underwent revision surgery resulting in a revision rate of 15.5%.

The contents of the osteolysis cavities were used for microbiological and histological analysis. The

histological findings were interpreted as a foreign-body reaction. The authors concluded that the use of AES

implants should be avoided until the reason and extent for the problem of osteolysis has been solved.15

Rodriguez et al observed a high frequency of delayed appearance of osteolysis (77%) in 18 ankles replaced

with AES prosthesis at the mean follow-up of 39.4 months.16

In summary, recent studies of the AES TAR reported a high rate of osteolytic lesions. It is still unclear13-16

whether the osteolysis process is a result of failure of the hydroxyapatite coating of the metal components or

failure of the mobile bearing. As a result of independently published results showing high osteolysis rate, the

AES prosthesis has been withdrawn from the market.17

66.10.1 References

Hintermann,B.: Current designs of total ankle prostheses, in Hintermann,B. (ed), Total ankle

arthroplasty. Historical overview, current concepts and future perspectives. Wien , Springer, 2004, pp.

69-100.

Goldberg,A.J., Sharp,R.J., and Cooke,P.: Ankle replacement: current practice of foot & ankle

surgeons in the United kingdom. Foot Ankle Int, 30:950-954, 2009.

Besse,J.L., Colombier,J.A., Asencio,J., Bonnin,M., Gaudot,F., Jarde,O., Judet,T., Maestro,M.,

Lemrijse,T., Leonardi,C., and Toullec,E.: Total ankle arthroplasty in France. Orthop Traumatol Surg

Res, 96:291-303, 2010.

Fevang,B.T., Lie,S.A., Havelin,L.I., Brun,J.G., Skredderstuen,A., and Furnes,O.: 257 ankle

arthroplasties performed in Norway between 1994 and 2005. Acta Orthop, 78:575-583, 2007.

Patsalis,T.: Die AES-Sprunggelenksprothese: Indikation, Technik und erste Ergebnisse. Fuss

Sprungg, 2:38-44, 2004.

Asencio,J. and Leonardi,C.: Ankle Evolutive System prosthesis: a simple, accurate, and reliable

specific concept for primary and revision surgery. Tech Foot & Ankle, 4:119-124, 2005.

Henricson,A. and Agren,P.H.: Secondary surgery after total ankle replacement. The influence of

preoperative hindfoot alignment. Foot Ankle Surg, 13:41-44, 2007.

OrthopaedicsOne Articles

Page of 336 372

8.

9.

10.

11.

12.

13.

14.

15.

16.

17.

Brooke,B.T., Harris,N.J., and Morgan,S.S.: Fibula lengthening osteotomy to correct valgus

mal-alignment following total ankle arthrplasty. Foot Ankle Surg,epub ahead of print, 2009.

Kharwadkar,N. and Harris,N.J.: Revision of STAR total ankle replacement to hybrid AES-STAR total

ankle replacement-a report of two cases. Foot Ankle Surg, 15:101-105, 2009.

Dahabreh,Z., Gonsalves,S., Monkhouse,R., and Harris,N.J.: Extrusion of metal radiological marker

from a total ankle replacement insert: a case report. J Foot Ankle Surg, 45:185-189, 2006.

Morgan,S.S., Brooke,B., and Harris,N.J.: Is there a role for total ankle replacement in polio patients?:

A case report and review of the literature. Foot Ankle Surg,epub ahead of print, 2009.

Anders,H., Kaj,K., Johan,J., and Urban,R.: The AES total ankle replacement: A mid-term analysis of

93 cases. Foot Ankle Surg, 16:61-64, 2010.

Morgan,S.S., Brooke,B., and Harris,N.J.: Total ankle replacement by the Ankle Evolution System:

medium-term outcome. J Bone Joint Surg Br, 92:61-65, 2010.

Besse,J.L., Brito,N., and Lienhart,C.: Clinical evaluation and radiographic assessment of bone lysis of

the AES total ankle replacement. Foot Ankle Int, 30:964-975, 2009.

Koivu,H., Kohonen,I., Sipola,E., Alanen,K., Vahlberg,T., and Tiusanen,H.: Severe periprosthetic

osteolytic lesions after the Ankle Evolutive System total ankle replacement. J Bone Joint Surg Br,

91:907-914, 2009.

Rodriguez,D., Bevernage,B.D., Maldague,P., Deleu,P.A., Tribak,K., and Leemrijse,T.: Medium term

follow-up of the AES ankle prosthesis: High rate of asymptomatic osteolysis. Foot Ankle Surg,

16:54-60, 2010.

Smith,T.W. and Stephens,M.: Ankle arthroplasty. Foot Ankle Surg, 16:53, 2010.

66.11 Alphanorm Total Ankle Replacement

This article is taken wholly from, or contains information that was originally published by the

American Orthopaedic Foot and Ankle Society.

more

Section editors and authors for Orthopaedia may have edited its content or added new information. The use

of information from the American Orthopaedic Foot and Ankle Society should not be construed as support

for or endorsement by that organization for any new information added by Orthopaedia members, or for any

editing of the original content.

Learn about the AOFAS

The Alphanorm TAR was developed by Professor Tilmann in Germany and has been used since 1996. The1

Alphanorm prosthesis is a non-constrained Buechel-Pappas type design with a 90° tibial stem without

inclination. The prosthesis is made out of cobalt chrome alloy with a titanium coating. The medial and lateral2

talar surfaces are not replaced during component implantation, thus preventing inversion and eversion.

Before developing the Alphanorm TAR for use, the designer had experience with the TRP prosthesis, New

Jersey prosthesis, and STAR prosthesis between 1976 and 1998.3,4

OrthopaedicsOne Articles

Page of 337 372

1.

2.

3.

4.

To date, there are no published studies available from Professor Tilmann or others addressing clinical

outcomes or surgical technique of the Alphanorm TAR.

66.11.1 References

Tillmann,K.: Endoprostheses of the ankle joint. Indications, development, current status and trends.

Orthopade, 32:179-186, 2003.

Gougoulias,N.E., Khanna,A., and Maffulli,N.: History and evolution in total ankle arthroplasty. Br Med

Bull, 89:111-151, 2009.

Tillmann,K., Schaar,M., Schaar,B., and Fink,B.: Ergebnisse von OSG-Endoprothesen bei

rheumatoider Arthritis: Eine klinische und pedobarographische Untersuchung. Fuss Sprungg, 1:56-65,

2003.

Feldman MH, Rockwood J (2004) Total ankle arthroplasty: a review of 11 current ankle implants. Clin

Podiatr Med Surg, 21, 393--406.

66.12 BOX Total Ankle Replacement

This article is taken wholly from, or contains information that was originally published by the

American Orthopaedic Foot and Ankle Society.

more

Section editors and authors for Orthopaedia may have edited its content or added new information. The use

of information from the American Orthopaedic Foot and Ankle Society should not be construed as support

for or endorsement by that organization for any new information added by Orthopaedia members, or for any

editing of the original content.

Learn about the AOFAS

The BOX TAR was developed from a collaborative effort of the Rizzoli Orthopaedic Institute (Drs. Giannini,

Catani, Leardini) in Italy and Oxford University (Dr. O'Connor) in England in the late 1990s. This

non-constrained, mobile-bearing prosthesis is a three-component implant with metal components fixed to the

proximal talus and the distal tibia and interposed UHMWPE meniscal bearing. The upper UHMWPE surface

is concave, not flat is in previous designs. The biomechanical development of this prosthesis type has been

well documented in the literature by its designers. Original research studies by the designers of this1-3

prosthesis focused on movement and stability of the ankle and sought to provide detailed understanding of

the role of the ligaments in controlling and limiting joint movement.

OrthopaedicsOne Articles

Page of 338 372

In 2001 Leardini developed a geometrical model of the intact human ankle complex that established the

basic principles for design of a new total ankle prostheses. Seven lower leg specimens were prepared for1

passive flexion analysis using a stereophotogrammetric system. It was shown that the geometry of the

articular surface of the ankle is strictly related to that of the ligaments. Therefore, the author stated that the

ideal TAR design should be based on ligament compatibility and the careful reconstruction of ligaments

should be performed in any foot and ankle surgery to recreate the normal kinematics and mechanics of the

ankle. A four-bar linkage model was developed from this initial study that also showed that both rolling and1

sliding motions take place at the talocrural joint.

In 2004 Leardini et al described the BOX TAR design and rationale, including compatibility to the

physiological function of surrounding ligaments. The designers aimed to reproduce physiological ankle2

mobility with the following design features:

Spherical convex tibial component

Talar component with radius of curvature in the sagittal plane longer than that of the natural talus

Fully conforming meniscal bearing.

Preliminary results from their study of two female patients demonstrated the feasibility of their surgical

technique. Reggiani et al developed a finite element model to address the BOX TAR analysis during the2

stance phase of gait. Overall kinematics, contact pressures, and ligament forces were analyzed during3

both, passive (eg, virtually unloaded) and active (eg,stance phase of gait) conditions. The authors showed

that this prosthesis design could allow the necessary ROM and constrain the motion of the prosthetic

components, especially the mobile bearing.3

Affatato et al have used a four-station knee joint simulator to address meniscal wear of the BOX TAR mobile

bearing. The knee wear simulator was able to reproduce load-motion patterns comparable to those of a4

replaced ankle. Tests of three specimens showed a linear penetration of 0.0178, 0.0081, and 0.0339 mm per

million-cycle. The linear penetration observed in this study was comparable to that found in

ceramic-to-polyethylene or metal-to-polyethylene couplings.4

Two years later, Affatato et al performed a comparative study of wear behavior in the BOX TAR between an

in vitro simulation and retrieved prostheses. Three retrieved mobile bearings were available from revision5

surgeries performed 24, 24, and 9 months after the initial TAR surgery. Visual and microscopic observations,

analyses, and Raman crystallinity-based measurements showed similarity between the patterns generated

experimentally using a four-station knee joint simulator and those seen in retrievals with similar wear

duration.5

Ingrosso et al performed gait analysis in BOX TAR patients by a stereophotogrammetric system with eight

M2-cameras. The study included 10 patients with a follow-up of 6 and 12 months after the surgery. Normal6

patterns and ROMs were observed in all 10 patients at both follow-ups.6

Giannini et al presented short-term results in 51 patients who were treated with a BOX TAR. The minimum7

follow-up in this study was 24 months. All patients showed significant functional improvement as assessed

by the AOFAS score. A revision arthroplasty had to be performed In one patient because of lateral

impingement, resulting in a 3-year cumulative survivorship of 97%. The main indication for this procedure7

was stage III ankle OA (with subtotal or total disappearance or deformation of joint space) with preserved or

restored ankle anatomy.8

OrthopaedicsOne Articles

Page of 339 372

1.

2.

3.

4.

5.

6.

7.

8.

In summary, the design process and biomechanical properties of the BOX TAR have been documented in

detail. The designers claim that it maintains complete congruency during the entire arc of motion and1-3

closely resembles normal ankle biomechanics. Studies addressing the wear behavior in this prosthesis

design have significantly contributed to the understanding of modern three-part prostheses. The4,5

short-term results are promising, with high satisfaction among treated patients, good functional results, and a

low revision rate. However, all aforementioned studies were performed by one of designers of this7

prosthesis; therefore, further study by independent groups should be performed. Long-term results have to

provide evidence on their clinical success.

66.12.1 References

Leardini,A.: Geometry and mechanics of the human ankle complex and ankle prosthesis design. Clin

Biomech (Bristol , Avon ), 16:706-709, 2001.

Leardini,A., O'Connor,J.J., Catani,F., and Giannini,S.: Mobility of the human ankle and the design of

total ankle replacement. Clin Orthop Relat Res, 424:39-46, 2004.

Reggiani,B., Leardini,A., Corazza,F., and Taylor,M.: Finite element analysis of a total ankle

replacement during the stance phase of gait. J Biomech, 39:1435-1443, 2006.

Affatato,S., Leardini,A., Leardini,W., Giannini,S., and Viceconti,M.: Meniscal wear at a

three-component total ankle prosthesis by a knee joint simulator. J Biomech, 40:1871-1876, 2007.

Affatato,S., Taddei,P., Leardini,A., Giannini,S., Spinelli,M., and Viceconti,M.: Wear behaviour in total

ankle replacement: a comparison between an in vitro simulation and retrieved prostheses. Clin

Biomech (Bristol , Avon ), 24:661-669, 2009.

Ingrosso,S., Benedetti,M.G., Leardini,A., Casanelli,S., Sforza,T., and Giannini,S.: GAIT analysis in

patients operated with a novel total ankle prosthesis. Gait Posture, 30:132-137, 2009.

Giannini,S., Romagnoli,M., O'Connor,J.J., Malerba,F., and Leardini,A.: Total ankle replacement

compatible with ligament function produces mobility, good clinical scores, and low complication rates:

an early clinical assessment. Clin Orthop Relat Res, 468:2746-2753, 2010.

Giannini,S., Buda,R., Faldini,C., Vannini,F., Romagnoli,M., Grandi,G., and Bevoni,R.: The treatment

of severe posttraumatic arthritis of the ankle joint. J Bone Joint Surg Am, 89 Suppl 3:15-28, 2007.

66.13 ESKA Ankle Prosthesis

This article is taken wholly from, or contains information that was originally published by the

American Orthopaedic Foot and Ankle Society.

more

OrthopaedicsOne Articles

Page of 340 372

1.

2.

Section editors and authors for Orthopaedia may have edited its content or added new information. The use

of information from the American Orthopaedic Foot and Ankle Society should not be construed as support

for or endorsement by that organization for any new information added by Orthopaedia members, or for any

editing of the original content.

Learn about the AOFAS

The ESKA TAR is a non-constrained, fixed-bearing, two-component prosthesis that was designed in

Germany for cementless implantation between 1985 and 1989. It has a shallow groove on the talar1,2

component that is congrous to the ultra-high-molecular-weight polyethylene (UHMWPE) bearing fixed to the

tibial prosthesis. The following features were included to improve the biomechanics of the ankle

replacement:

Cementless implantation and porous-structured implant surface for faster osteointegration

Shear force reduction and rotational force control by shape design of both metallic components

Easy replacement of the polyethylene without disturbing prosthesis anchoring.1,2

Because of the ridge-like shaping and its transverse anchoring peg in both metallic components, a lateral

approach with fibular osteotomy or in special cases, a medial malleolar approach is used for component

implantation.3

In 2001, the prosthesis designer, Dr. Rudigier, published short-term results of 56 patients treated with the

ESKA prosthesis since 1990. Forty of 56 patients were reviewed at a minimum follow up of 1 year. All4

patients experienced significant pain relief, improved ankle ROM and ability to walk, and showed functional

improvement as assessed by the Kofoed ankle score. In two cases, deep infection led to prosthesis removal

and conversion to ankle fusion. Another patient showed painful progressive ossification of the joint capsule

resulting in ankle fusion. The same patient cohort was again reviewed at a longer follow up and the results4

were published in another two studies. The authors reported a significant improvement of the Kofoed3,5

ankle score from 37.6 points preoperatively to 90.4 points postoperatively. Of 12 TARs performed 10-155

years before, 8 (67%) remained in situ (2 early deep infections, 2 aseptic loosenings). In 20 implants

inserted 5-10 years before, 3 were revised, leaving 17 (85%) in situ functioning well. None of the implants

with a follow-up of 1-5 years failed. No results from independent authors have been published.

In summary, the limited studies performed by ESKA prosthesis designers reported favorable mid-term

results with improved ankle ROM, pain relief, and ability to walk long distances postoperatively. However,3-5

the authors conceded that this surgery should be limited to only highly experienced foot and ankle surgeons

due to the demanding lateral approach with fibular osteotomy. As with the BOX TAR, published results have

come from only the prosthesis designers. Thus additional, independent analyses of ESKA efficacy are

needed.

66.13.1 References

Rudiger,J.: Ankle replacement by the cementless ESKA endoprosthesis. Tech Foot & Ankle,

4:125-136, 2005.

Rudiger,J., Menzinger,F., and Grundei,H.: 14 Jahre Erfahrungen mit der zementfreien

ESKA-Sprunggelenksendoprothese. Fuss Sprungg, 2:65-75, 2004.

OrthopaedicsOne Articles

Page of 341 372

3.

4.

5.

1.

Rudigier,J.: Ankle replacement by the cementless ESKA endoprosthesis. Tech Foot & Ankle,

4:125-136, 2005.

Rudigier,J., Grundei,H., and Menzinger,F.: Prosthetic replacement of the ankle in posttraumatic

arthrosis: 10-year experience with the cementless ESKA ankle prosthesis. Eur J Trauma, 27:66-74,

2001.

Rudigier,J., Menzinger,F., and Grundei,H.: 14 Jahre Erfahrungen mit der zementfreien

ESKA-Sprunggelenksendoprothese. Fuss Sprungg, 2:65-75, 2004.

66.14 German Ankle System

This article is taken wholly from, or contains information that was originally published by the

American Orthopaedic Foot and Ankle Society.

more

Section editors and authors for Orthopaedia may have edited its content or added new information. The use

of information from the American Orthopaedic Foot and Ankle Society should not be construed as support

for or endorsement by that organization for any new information added by Orthopaedia members, or for any

editing of the original content.

Learn about the AOFAS

The German Ankle System is a three-component prosthesis allowing rotation around each of the three

possible movement axes. The talar component of this prosthesis includes side borders to keep the mobile1

bearing in position, which may prevent inlay dislocation. Both metallic components have BONIT© coating

(porous coating with titanium plasma sprayed surface and an additional layer of calcium phosphate) to

ensure the proper osteointegration. The system includes an option for computer-assisted implantation

surgery.

Richter et al performed the robotic-guided surgical approach on a cadaver and compared the results with

those obtained using a HINTEGRA prosthesis. The authors found that the German Ankle Prosthesis had1

smaller effect on resulting forces and torques during partial weight-bearing passive ankle motion than the

HINTEGRA prosthesis. However, to date no clinical results from patients treated with this prosthesis type1

are available, raising some concern regarding the interpretation of obtained data in the study by the

designer.

66.14.1 Reference

Richter,M., Zech,S., Westphal,R., Klimesch,Y., and Gosling,T.: Robotic cadaver testing of a new total

ankle prosthesis model (German Ankle System). Foot Ankle Int, 28:1276-1286, 2007.

OrthopaedicsOne Articles

Page of 342 372

66.15 HINTEGRA Total Ankle Replacement

This article is taken wholly from, or contains information that was originally published by the

American Orthopaedic Foot and Ankle Society.

more

Section editors and authors for Orthopaedia may have edited its content or added new information. The use

of information from the American Orthopaedic Foot and Ankle Society should not be construed as support

for or endorsement by that organization for any new information added by Orthopaedia members, or for any

editing of the original content.

Learn about the AOFAS

The HINTEGRA Total Ankle Replacement (TAR) is an unconstrained, three-component system that provides

inversion-eversion stability and was designed in 2000 by Beat Hintermann, Greta Dereymaeker, Ramon

Viladot, and Patrice Diebold. The mobile bearing provides axial rotation and normal flexion-extension

mobility. The HINTEGRA TAR includes two metallic components and an ultrahigh-density polyethylene1-3

mobile bearing. The non-articulating surfaces have a porous coating with 20% porosity and are covered by

titanium fluid and hydroxyapatite.

The tibial component has a flat, 4-mm thick loading plate with pyramidal peaks against the tibia. Additional

stability may be achieved by fixation with two screws. The talar component is conically shaped with a smaller

radius medially than laterally, mimicking the normal anatomy of talus. It has 2.5-mm high rims on each side

that ensure stable positioning and guide the anteroposterior translation of the mobile bearing. The anterior

shield of this component increases primary bone support, especially in cases with weaker bone, and may

prevent the adherence of scar tissue and avoid restriction of ROM in cases with arthrofibrosis.2

The HINTEGRA TAR has been used since 2000 in Europe, since 2004 in Canada and Korea, and since

2005 in Brazil. The use of this prosthesis is also documented in National Arthroplasty Registers of4-7

Finnland, Sweden, Norway, and New Zealand.8 9 10 11

In the current literature there are numerous studies addressing clinical outcome and biomechanical

properties of the HINTEGRA TAR, published by the designer. In 2004, Hintermann et al published the first

study reporting HINTEGRA design rationale, surgical technique, and short-term results of the first

consecutive 122 ankles in 166 patients. All patients were reviewed at a mean follow up of 19 months.12

Eight ankles had to undergo revision surgery, four because of loosening of at least one component, one

because of dislocation of the mobile bearing, and three for other reasons. All revisions were successful.

Most patients experienced significant functional improvement postoperatively, as assessed by the AOFAS

score; 68% were pain free at latest follow up. The clinically and radiographically measured ROM was 39°

and 37°, respectively. Tibial components were stable in all reviewed ankles; in two ankles a slight migration

of the talar component was observed. Similar results were published in a study by Valderrabano and12

Hintermann, including 125 HINTEGRA implants.13

OrthopaedicsOne Articles

Page of 343 372

In 2006, Hintermann et al presented mid-term results of 271 HINTEGRA TARs performed in 261 patients

between 2000 and 2004. The mean follow-up was 36 months with a range between 12 and 64 months.14

Intraoperatively, four malleolar fractures occurred, which all healed within 6 weeks. Five ankles (1.8%) had

to be converted to ankle fusion. Thirteen talar and two tibial components were later revised. In total, 39

revision surgeries (eg, open arthrolysis, lengthening of Achilles tendon, ligament reconstruction) were

necessary to address late postoperative complications.14

Kim et al compared the outcome and complications of HINTEGRA TAR with and without hindfoot fusion. In15

total 60 ankles with HINTEGRA implants and subtalar or triple fusion were compared to a control group of

288 ankles treated with TAR alone. The mean follow up was 39.5 months. The authors found that the clinical

outcome of TAR when combined with hindfoot fusion (subtalar or triple fusion) is comparable to that of ankle

replacement alone. Therefore, the authors recommended that hindfoot fusion should be performed

simultaneously with TAR in cases when indicated.15

Recently, Barg et al published clinical observation studies addressing clinical and radiological outcomes in

patients who underwent HINTEGRA TAR because of end-stage osteoarthritis (OA) due to hemophilia or16

hemochromatosis. In both patient cohorts, favorable outcomes with significant pain relief and functional17

improvement were observed, demonstrating TAR as a reliable option treatment.16,17

Valderrabano et al addressed the sporting and recreational activity of patients with end-stage ankle arthritis

before and ankle TAR. The authors recommended specific Sports Frequency Score. A clinical evaluation18

was performed preoperatively and at a mean follow up of 2.8 years in 147 patients (152 ankles). All patients

experienced significant functional improvement as assessed by the AOFAS score. After TAR, patients with

sports activities had a higher AOFAS score. The three most frequent sports activities after TAR were hiking,

biking, and swimming.18

Daniels et al published their results for 32 TARs in patients with significant preoperative varus talar

deformities. In all patients, cementless, mobile-bearing, three-part component prostheses had been used:19

26 HINTEGRA, four Mobility, and two STAR implants. A satisfactory radiographic correction was obtained in

the most cases (30 ankles) at a mean follow up of 17 months. In 24 ankles, additional procedures after TAR

were required to obtained a plantigrade foot.19

Lee et al addressed perioperative complications of HINTEGRA TAR in their 50 initial cases. The same20

author presented two case reports addressing HINTEGRA TAR in patients following revascularization of

avascular necrosis of the talar body. The authors stated that in patients with avascular necrosis of the21

talus, healing of necrotic bone by creeping substitution TAR is a valuable option.21

Kim et al described clinical outcome of TAR in 23 patients with moderate to severe varus deformity and

compared results to those in 22 patients with neutral alignment. All patients were reviewed at a mean22

follow-up of 27 months and showed substantial pain relief and functional improvement in both groups as

assessed by VAS and AOFAS score, respectively. Failure of the implant with conversion to ankle fusion

occurred in one case in each group. The authors stated that the TAR is a valuable treatment for severe

ankle OA, including patients with hindfoot misalignment. However, the appropriate additional procedures to

address the deformity are necessary to obtain good clinical results including prosthesis stability and ROM.22

OrthopaedicsOne Articles

Page of 344 372

1.

2.

3.

4.

5.

6.

7.

Recently, Bai et al compared clinical outcome and revision rate after TAR between patients with

posttraumatic and primary OA. Sixty-seven consecutive TARs were performed using HINTEGRA23

prosthesis in 65 patients between 2005 and 2007. All patients were divided into two groups: posttraumatic

OA group (37 ankles) and primary OA group (30 ankles). At a mean follow-up of 38 months the clinical

(AOFAS score, ROM) and radiographic outcome were comparable. However, the incidence of postoperative

complications was significantly higher in the posttraumatic OA group.23

Valderrabano et al published a series of studies addressing muscle biomechanics and muscle rehabilitation

in patients with severe ankle OA who underwent TAR. A prospective study performed including 1524-27

patients who were reviewed preoperatively and postoperatively in 3-month intervals up to 1 year showed that

TAR surgery may improve muscle biomechanics as assessed by torque measurement and EMG intensity.24

However, in most patients the muscle rehabilitation was not complete at a follow up of 1 year.24

Müller et al used a Heidelberger foot and ankle analysis model to address the 3D kinematics and foot and

ankle shape in 12 patients who underwent HINTEGRA TAR. The authors detected some decreased ROM28

after ankle replacement compared with contralateral ankles free of degenerative changes. However, these

differences did not affect the gait kinematics, showing TAR may be able to preserve ROM which, in turn,

may avoid or at least decelerate the degenerative changes in adjacent joints.28

Lee et al investigated static and dynamic postural balance after TAR using HINTEGRA in 30 patients and

compared the results to an age- and sex-matched control group. The authors showed that patients who29

underwent TAR have a higher degree of dynamic postural imbalance. Some motor control deficits were also

detected in the TAR group. The authors stated that more intensive postoperative balance training may

decrease the observed deficits.29

In conclusion, the mid-term results in patients with HINTEGRA TAR are favorable. However, most12,14

clinical studies have been published by the designer. Comparable studies by independent authors and

studies with longer follow-ups should be performed.

66.15.1 References

Valderrabano,V., Pagenstert,G.I., and Hintermann,B.: Total ankle replacement - three-component

prosthesis. Tech Foot & Ankle, 2:84-90, 2005.

Hintermann,B. and Barg,A.: The HINTEGRA total ankle arthroplasty, in Wiesel,S.W. (ed), Operative

Techniques in Orthopaedic Surgery Lippincott Williams & Wilkins, 2010, pp. 4022-4031.

Hintermann,B.: Surgical techniques, in Hintermann,B. (ed), Total ankle arthroplasty: Historical

overview, current concepts and future perspectives. Springer , Wien New York, 2005, pp. 105-126.

Kim,B.S., Choi,W.J., Kim,Y.S., and Lee,J.W.: Total ankle replacement in moderate to severe varus

deformity of the ankle. J Bone Joint Surg Br, 91:1183-1190, 2009.

Gougoulias,N.E., Khanna,A., and Maffulli,N.: History and evolution in total ankle arthroplasty. Br Med

Bull, 89:111-151, 2009.

Goldberg,A.J., Sharp,R.J., and Cooke,P.: Ankle replacement: current practice of foot & ankle

surgeons in the United kingdom. Foot Ankle Int, 30:950-954, 2009.

Besse,J.L., Colombier,J.A., Asencio,J., Bonnin,M., Gaudot,F., Jarde,O., Judet,T., Maestro,M.,

Lemrijse,T., Leonardi,C., and Toullec,E.: Total ankle arthroplasty in France. Orthop Traumatol Surg

Res, 96:291-303, 2010.

OrthopaedicsOne Articles

Page of 345 372

8.

9.

10.

11.

12.

13.

14.

15.

16.

17.

18.

19.

20.

21.

22.

23.

24.

25.

26.

27.

Skytta,E.T., Koivu,H., Eskelinen,A., Ikavalko,M., Paavolainen,P., and Remes,V.: Total ankle

replacement: a population-based study of 515 cases from the Finnish Arthroplasty Register. Acta

Orthop, 81:114-118, 2010.

Henricson,A., Skoog,A., and Carlsson,A.: The Swedish Ankle Arthroplasty Register: An analysis of

531 arthroplasties between 1993 and 2005. Acta Orthop, 78:569-574, 2007.

Fevang,B.T., Lie,S.A., Havelin,L.I., Brun,J.G., Skredderstuen,A., and Furnes,O.: 257 ankle

arthroplasties performed in Norway between 1994 and 2005. Acta Orthop, 78:575-583, 2007.

Hosman,A.H., Mason,R.B., Hobbs,T., and Rothwell,A.G.: A New Zealand national joint registry review

of 202 total ankle replacements followed for up to 6 years. Acta Orthop, 78:584-591, 2007

Hintermann,B., Valderrabano,V., Dereymaeker,G., and Dick,W.: The HINTEGRA ankle: rationale and

short-term results of 122 consecutive ankles. Clin Orthop Relat Res, 424:57-68, 2004.

Valderrabano,V. and Hintermann,B.: HINTEGRA-Sprunggelenkprothese: Präliminäre Resultate der

ersten 125 Fälle. Fuss Sprungg, 2:7-16, 2004.

Hintermann,B., Valderrabano,V., Knupp,M., and Horisberger,M.: The HINTEGRA ankle: short- and

mid-term result. Orthopade, 35:533-545, 2006.

Kim,B.S., Knupp,M., Zwicky,L., Lee,J.W., and Hintermann,B.: Total ankle replacement in association

with hindfoot fusion: Outcome and complications. J Bone Joint Surg Br, 92:1540-1547, 2010.

Barg,A., Elsner,A., Hefti,D., and Hintermann,B.: Haemophilic arthropathy of the ankle treated by total

ankle replacement: a case series. Haemophilia, 16:647-655, 7-1-2010.

Barg,A., Elsner,A., Hefti,D., and Hintermann,B.: Total Ankle Arthroplasty in Patients with Hereditary

Hemochromatosis. Clin Orthop Relat Res, 7-28-2010.

Valderrabano,V., Pagenstert,G., Horisberger,M., Knupp,M., and Hintermann,B.: Sports and recreation

activity of ankle arthritis patients before and after total ankle replacement. Am J Sports Med,

34:993-999, 2006.

Daniels,T.R., Cadden,A.R., and Lim,K.: Correction of varus talar deformities in ankle joint

replacement. Oper Tech Orthop, 18:282-286, 2008.

Lee,K.B., Cho,S.G., Hur,C.I., and Yoon,T.R.: Perioperative complications of HINTEGRA total ankle

replacement: our initial 50 cases. Foot Ankle Int, 29:978-984, 2008

Lee,K.B., Cho,S.G., Jung,S.T., and Kim,M.S.: Total ankle arthroplasty following revascularization of

avascular necrosis of the talar body: two case reports and literature review. Foot Ankle Int,

29:852-858, 2008.

Kim,B.S., Choi,W.J., Kim,Y.S., and Lee,J.W.: Total ankle replacement in moderate to severe varus

deformity of the ankle. J Bone Joint Surg Br, 91:1183-1190, 2009.

Bai,L.B., Lee,K.B., Song,E.K., Yoon,T.R., and Seon,J.K.: Total ankle arthroplasty outcome

comparison for post-traumatic and primary osteoarthritis. Foot Ankle Int, 31:1048-1056, 2010.

Valderrabano,V., Nigg,B.M., von,T., V, Frank,C.B., and Hintermann,B.: J. Leonard Goldner Award

2006. Total ankle replacement in ankle osteoarthritis: an analysis of muscle rehabilitation. Foot Ankle

Int, 28:281-291, 2007.

Valderrabano,V., von Tscharner,V., Nigg,B.M., Goepfert,B., Frank,C.B., and Hintermann,B.:

Unterschenkel-Muskelatrophie bei Arthrose des oberen Sprunggelenks und deren Rehabilitation nach

Implantation einer Sprunggelenksprothese. Fuss Sprungg, 5:33-43, 2007.

Valderrabano,V., Hintermann,B., von,T., V, Gopfert,B., Dick,W., and Nigg,B.M.: Muscle biomechanics

in total ankle replacement. Orthopade, 35:513-520, 2006.

Valderrabano,V., Nigg,B.M., von,T., V, Stefanyshyn,D.J., Goepfert,B., and Hintermann,B.: Gait

analysis in ankle osteoarthritis and total ankle replacement. Clin Biomech (Bristol , Avon ),

22:894-904, 2007.

OrthopaedicsOne Articles

Page of 346 372

28.

29.

1.

2.

3.

4.

5.

Muller,S., Wolf,S., and Doderlein,L.: [Three-dimensional analysis of the foot following implantation of

a HINTEGRA ankle prosthesis: evaluation with the Heidelberg foot model]. Orthopade, 35:506-512,

2006.

Lee,K.B., Park,Y.H., Song,E.K., Yoon,T.R., and Jung,K.I.: Static and dynamic postural balance after

successful mobile-bearing total ankle arthroplasty. Arch Phys Med Rehabil, 91:519-522, 2010.

66.16 INBONE Total Ankle Replacement

This article is taken wholly from, or contains information that was originally published by the

American Orthopaedic Foot and Ankle Society.

more

Section editors and authors for Orthopaedia may have edited its content or added new information. The use

of information from the American Orthopaedic Foot and Ankle Society should not be construed as support

for or endorsement by that organization for any new information added by Orthopaedia members, or for any

editing of the original content.

Learn about the AOFAS

The INBONE TAR is a fixed-bearing, two-component total ankle system. A special feature of this ankle1,2

design is a modular stem system for both metallic components. The tibial stem has the ability to be extended

by adding more modular segments. The stem of the talar component may be short and limited to the talar

body. However, in cases where subtalar fusion has to be performed or for greater stability, the talar stem

may be extended across the subtalar joint.

To date, there are limited reports in the literature on clinical and radiographic outcomes in patients who

received INBONE TAR. The designer of the prosthesis has performed more than 240 INBONE TARs and3,4

data have been collected in preparation for publication. There are also no biomechanical studies available5

that address the kinematics and biomechanical properties of this prosthesis design.

66.16.1 References

Deorio,J.K. and Easley,M.E.: Total ankle arthroplasty. Instr Course Lect, 57:383-413, 2008.

Cracchiolo,A., III and Deorio,J.K.: Design features of current total ankle replacements: implants and

instrumentation. J Am Acad Orthop Surg, 16:530-540, 2008.

Reiley,M.A.: Total ankle arthroplasty with bone defects. Foot Ankle Spec, 2:32-34, 2009.

Reiley,M.A.: INBONE total ankle replacement. Foot Ankle Spec, 1:305-308, 2008.

Ellis,S. and Deorio,J.K.: The INBONE total ankle replacement. Oper Tech Orthop, 20:201-210, 2010.

OrthopaedicsOne Articles

Page of 347 372

66.17 Mobility Ankle System

This article is taken wholly from, or contains information that was originally published by the

American Orthopaedic Foot and Ankle Society.

more

Section editors and authors for Orthopaedia may have edited its content or added new information. The use

of information from the American Orthopaedic Foot and Ankle Society should not be construed as support

for or endorsement by that organization for any new information added by Orthopaedia members, or for any

editing of the original content.

Learn about the AOFAS

The Mobility Ankle System, developed by Pascal Rippstein, Peter Wood, and Chris Coetzee, is a

three-component Buechel-Pappas type prosthesis with a short, conical tibial stem. The talar component of

the Mobility implant resurfaces the superior dome of the talus, while the medial and lateral aspects of the

talus remain untreated (unlike the Buechel-Pappas prosthesis). The talar component has a central,

longitudinal sulcus and two fins, enhancing its intrinsic stability. The nonarticulating surfaces are porous

coated with a titanium spray.

Goldberg et al reported results of their questionnaire-based survey sent to all Consultant members of the

British Orthopaedic Foot & Ankle Society. The Mobility prosthesis was the most commonly used prosthesis1

among 62% of all surgeons in the United Kingdom.1 The use of this prosthesis has also been documented in

the Swedish Ankle Arthroplasty Register and New Zealand Ankle Arthroplasty Register. The Swedish2 3

Ankle Arthroplasty Register included a total of 531 total ankle replacements (TARs) with 23 Mobility

prostheses implanted since 2005. No peri- or postoperative complications with the Mobility prosthesis were2

observed in the Swedish cohort. The New Zealand Ankle Arthroplasty Register included a total of 202 TAR2

with a mean follow-up of 6 years.3 Twenty-nine patients in this registry underwent Mobility prosthesis; no

failures were observed in this study.3

Recently, Wood et al published early results from their prospective study that included 100 Mobility implants

performed in 96 patients between 2003 and 2005.4 At a minimum follow up of 5 years, a total of five ankles

(5%) had to undergo revision surgery -- two ankle fusions and three revision arthroplasties -- resulting in 3-

and 4-year survivorship of 97% (95%CI, 91% -- 99%) and 93.6% (95%CI, 84.7% -- 97.4%), respectively. In

14 ankles a radiolucent line or osteolytic cavity was observed. However, only in five ankles was it more than

10 mm in width. The authors presented encouraging short-term results which are comparable to those

obtained using other modern three-component prostheses.4

OrthopaedicsOne Articles

Page of 348 372

1.

2.

3.

Naal et al addressed habitual physical activity and sports participation before and after TAR in 137

consecutive patients with 155 implants. One hundred one ankles were available for the review at a mean5

follow up of 44 months, 54 of them were Mobility implants. The percentage of patients who were active in

sports did not change after the TAR (62.4% preoperatively and 66.3% postoperatively). The most common

sports activities after TAR were swimming, cycling, and fitness/weight training. The authors registered high

functional improvement in their patient cohort as assessed by AOFAS scores. No association was found

between sports participation, increased physical activity level, and the appearance of periprosthetic

radiolucencies.5

Thermann et al reported a single case of a 58-year-old patient who underwent Mobility TAR and tibialis

posterior tendon transfer for ankle osteoarthritis (OA) and drop foot deformity. Postoperative radiographs6

showed a medial malleolus fracture despite intraoperative pinning with the Kirschner wires. However, the

fracture completely healed after 8 weeks. Three years after the primary procedure, the patient was

asymptomatic and had a stable ankle joint with 5° dorsiflexion and 20° plantar flexion.6

Goldberg et al reported two cases of early failure in patients who underwent Mobility TAR due to component

malposition. In both cases, the talar component was inserted back to front, as was the polyethylene insert,7

requiring ankle fusion in the first case and revision arthroplasty in the second case. The authors emphasized

the need for adequate training of foot and ankle surgeons for performing TARs. They also suggested that

designers have a responsibility to improve marketing and education for surgeons using their products to

minimize incidents of incorrect use.7

Espinosa et al generated two finite element models of TAR prostheses (Agility and Mobility) to investigate

the misalignment of prosthesis components on joint contact pressures. The authors have shown that the8

highly congruent mobile-bearing design of Mobility prosthesis may result in more evenly distributed contact

pressures than the less congruent two-component Agility prosthesis. However, in both design the

misalignment of prosthesis components may lead to pathologically increased contact stresses.8

Bell and Fisher investigated polyethylene wear in Mobility and Buechel-Pappas prostheses using a modified

knee prosthesis simulator. The authors showed that wear in the two models was comparable and that the9

wear rate for both designs significantly increases with the inclusion of an anterior/posterior displacement in

the kinematic inputs, simulating malalignment of prosthesis components.9

In conclusion, the Mobility TAR has been in general use since 2003 in Europe, Australia, New Zealand,

South Africa, and Canada. In the United States a multicenter FDA trial including this prosthesis type is

running.10

66.17.1 References

Goldberg,A.J., Sharp,R.J., and Cooke,P.: Ankle replacement: current practice of foot & ankle

surgeons in the United kingdom. Foot Ankle Int, 30:950-954, 2009.

Henricson,A., Skoog,A., and Carlsson,A.: The Swedish Ankle Arthroplasty Register: An analysis of

531 arthroplasties between 1993 and 2005. Acta Orthop, 78:569-574, 2007.

Hosman,A.H., Mason,R.B., Hobbs,T., and Rothwell,A.G.: A New Zealand national joint registry review

of 202 total ankle replacements followed for up to 6 years. Acta Orthop, 78:584-591, 2007.

OrthopaedicsOne Articles

Page of 349 372

4.

5.

6.

7.

8.

9.

10.

Wood,P.L., Karski,M.T., and Watmough,P.: Total ankle replacement: the results of 100 mobility total

ankle replacements. J Bone Joint Surg Br, 92:958-962, 2010.

Naal,F.D., Impellizzeri,F.M., Loibl,M., Huber,M., and Rippstein,P.F.: Habitual physical activity and

sports participation after total ankle arthroplasty. Am J Sports Med, 37:95-102, 2009.

Thermann,H., Gavriilidis,I., Longo,U.G., and Maffulli,N.: Total ankle arthroplasty and tibialis posterior

tendon transfer for ankle osteoarthritis and drop foot deformity. Foot Ankle Surg,epub ahead of print,

2009.

Goldberg,A.J., Sharp,B., and Cooke,P.: Early failure in total ankle replacements due to component

malposition: a report of two cases. Foot Ankle Int, 30:783-787, 2009.

Espinosa,N., Walti,M., Favre,P., and Snedeker,J.G.: Misalignment of total ankle components can

induce high joint contact pressures. J Bone Joint Surg Am, 92:1179-1187, 2010.

Bell,C.J. and Fisher,J.: Simulation of polyethylene wear in ankle joint prostheses. J Biomed Mater Res

B Appl Biomater, 81:162-167, 2007.

Deorio,J.K. and Easley,M.E.: Total ankle arthroplasty. Instr Course Lect, 57:383-413, 2008.

66.18 Ramses Total Ankle Replacement

This article is taken wholly from, or contains information that was originally published by the

American Orthopaedic Foot and Ankle Society.

more

Section editors and authors for Orthopaedia may have edited its content or added new information. The use

of information from the American Orthopaedic Foot and Ankle Society should not be construed as support

for or endorsement by that organization for any new information added by Orthopaedia members, or for any

editing of the original content.

Learn about the AOFAS

The Ramses TAR was developed in 1987 and first implanted in 1989 by a French designer group. The1,2

Ramses TAR is a three-component, semi-constrained prosthesis with a high-density mobile bearing. Initially,

cemented fixation of the prosthesis was used between 1980 and 2000.

A total of approximately 350 TARs using the Ramses TAR have been performed by the inventing group.

Mendolia et al reported long-term results in 69 patients who underwent Ramses TAR between 1989 and

1993. In seven cases, the replaced ankle had to be converted to an ankle arthrodesis (four ankles for pain2

without loosening and three ankles for clinical and radiographic loosening). In an additional five cases, a

second surgery was performed (two cases with mobile bearing ankle replacement and three cases with

revision arthroplasties).2

OrthopaedicsOne Articles

Page of 350 372

1.

2.

3.

4.

5.

Delagoutte performed a retrospective analysis of 110 TARs between 1991 and 1998. This was a3

multicenter study including 22 hospitals. Three different TAR types were used in this study: Ramses (n=66,

60%), LCS (n=36, 33%), and STAR (n=8, 7%) prosthesis. The follow up varied in this study between 3 and

37 months. In general, the postoperative functional results were less favorable, with no postoperative

improvement in dorsiflexion. In two patients with Ramses prostheses, revision surgery was necessary to

address prosthesis loosening.3

Use of the Ramses prosthesis has been listed in 11 of 202 TARs mentioned in the New Zealand national

joint registry. Two failures in this patient cohort were reported.4 4

In conclusion, the mid- and long-term results in patients who underwent Ramses TAR are not satisfactory

due to high revision rates of 18% and 34% at 2 and 10 years, respectively. The number of pain free patients

also has equaled the number of patients with remaining pain.5

66.18.1 References

Mendolia,G. and Talus Group: The Ramses ankel replacement: design-surgical technique result,

results in first 38 cases. The French Orthopedic Web Journal (available at: www matrise-orthop com),

2007.

Mendolia,G., Coillard,J.Y., Cermolacce,C., and Determe,P.: Long-term (10 to 14 years) results of the

Ramses total ankle arthroplasty. Tech Foot & Ankle, 4:160-173, 2005.

Delagoutte,J.P.: Retrospective analysis of 110 ankle prostheses. Eur J Orthop Surg Traumatol,

12:198-205, 2002.

Hosman,A.H., Mason,R.B., Hobbs,T., and Rothwell,A.G.: A New Zealand national joint registry review

of 202 total ankle replacements followed for up to 6 years. Acta Orthop, 78:584-591, 2007.

Michael,J.M., Golshani,A., Gargac,S., and Goswami,T.: Biomechanics of the ankle joint and clinical

outcomes of total ankle replacement. J Mech Behav Biomed Mater, 1:276-294, 2008.

66.19 Salto Total Ankle Prosthesis

This article is taken wholly from, or contains information that was originally published by the

American Orthopaedic Foot and Ankle Society.

more

Section editors and authors for Orthopaedia may have edited its content or added new information. The use

of information from the American Orthopaedic Foot and Ankle Society should not be construed as support

for or endorsement by that organization for any new information added by Orthopaedia members, or for any

editing of the original content.

Learn about the AOFAS

OrthopaedicsOne Articles

Page of 351 372

The Salto Total Ankle Prosthesis was developed between 1994 and 1996 by Michel Bonnin. This total1

ankle replacement (TAR) is the third generation of cementless meniscal-bearing designs. The tibial

component has a flat surface that faces the mobile bearing, allowing free translation and rotation. The 3 mm

medial rim is designed to avoid insert impingement against the medial malleolus. For osseous integration,

the component has a keel and a fixation peg. The specific shape of the talar component mimics the natural

talar geometry - the anterior width is wider than the posterior width and the lateral flange has a larger

curvature radius than the medial. The mobile bearing is manufactured from UHMWPE and has full

congruency with the talar component in flexion and extension. All components are available in three different

sizes.1

In 2008, Leszko et al performed an in vivo kinematics study of the Salto TAR in 20 patients using

fluoroscopy and a 3D-to-2D registration technique. The motion of the prosthesis was described in terms of2

clinical rotations and as rotation about the helical axis. Among the clinical rotations, the dorsi-/plantarflexion

was the most dominant with a mean ROM of 9.2 degrees. The anterior-posterior translation of the mobile

bearing was measured at 1.5 mm and 2.3 mm for gait and step-up, respectively.2

The first clinical study on the Salto prosthesis was published in 2004 by the prosthesis designer. Bonnin et1

al implanted 98 consecutive Salto prostheses between 1997 and 2000. Ninety-three implants in 91 patients

were reviewed clinically and radiographically at a mean follow-up of 35 months. Most patients were pain free

and showed a significant functional improvement as assessed by AOFAS score (from 32.2 to 83.1 points).

Two prostheses had to be converted to ankle fusion, resulting in survivorship of over 95% at 68 months. In1

the following 2 years, 22 additional Salto implants were performed, as published by Weber et al, who

included a total of 115 implants in their study. At a mean follow up of 22 months, four ankles had to undergo3

revision surgery.3

Recently, Bonnin et al presented survivorship at 7 to 11 years in a retrospective review of 98 TARs

performed between 1997 and 2000. Six replaced ankles had to be converted to ankle fusion and an4

additional 18 ankles underwent reoperation without ankle fusion (10-year survivorship 65% with 95% CI of

50% to 80%). The most common complications requiring additional surgery were bone cysts (11 ankles),

fracture of polyethylene (5 ankles), and unexplained pain (3 ankles).4

Furthermore, Bonnin et al addressed the sports activity level of 145 patients who underwent Salto TAR

between 1997 and 2005. Ankle function was assessed using the Foot Function Index and Foot and Ankle5

Ability Measurement. In most cases, replaced ankles were reported to have normal (15.2%) or nearly normal

(60.7%) function. Participation in sports and recreational activities was analyzed only in the osteoarthritis

(OA) group (100 patients), but not in patients with rheumatoid arthritis. While most patients participated in

some sport activities, a return to impact sports was rarely possible.5

In conclusion, the mid-term results of the Salto TAR are promising. However, the only clinical reports come4

from the prosthesis designer. In the United States, a modified design of the Salto TAR is approved for use in

patients with ankle OA. This fixed-bearing design includes a titanium tibial component with a highly6

conforming polyethylene articulating insert. However, to date no clinical studies reporting results of this6

prosthesis are available.

66.19.1 References

OrthopaedicsOne Articles

Page of 352 372

1.

2.

3.

4.

5.

6.

Bonnin,M., Judet,T., Colombier,J.A., Buscayret,F., Graveleau,N., and Piriou,P.: Midterm results of the

Salto Total Ankle Prosthesis. Clin Orthop Relat Res, 424:6-18, 2004.

Leszko,F., Komistek,R.D., Mahfouz,M.R., Ratron,Y.A., Judet,T., Bonnin,M., Colombier,J.A., and

Lin,S.S.: In vivo kinematics of the salto total ankle prosthesis. Foot Ankle Int, 29:1117-1125, 2008.

Weber,M., Bonnin,M., Columbier,J.A., and Judet,T.: Erste Ergebnisse der

SALTO-Sprunggelenkendoprothese: Eine französische Multizenterstudie mit 155 Implantaten. Fuss

Sprungg, 2:29-37, 2004.

Bonnin,M., Gaudot,F., Laurent,J.R., Ellis,S., Colombier,J.A., and Judet,T.: The Salto Total Ankle

Arthroplasty: Survivorship and Analysis of Failures at 7 to 11 years. Clin Orthop Relat Res, 7-1-2010.

Bonnin,M.P., Laurent,J.R., and Casillas,M.: Ankle function and sports activity after total ankle

arthroplasty. Foot Ankle Int, 30:933-944, 2009.

Yalamanchili,P., Donely,B., Casillas,M., Ables,A., and Lin,S.S.: Salto Talaris total ankle replacement.

Oper Tech Orthop, 18:277-281, 2008.

66.20 Special Situations for Total Ankle Replacement

Takedown of Painful Ankle Fusion and Total Ankle Replacement

Simultaneous Bilateral Total Ankle Replacement

66.20.1 Simultaneous Bilateral Total Ankle Replacement

This article is taken wholly from, or contains information that was originally published by the

American Orthopaedic Foot and Ankle Society.

more

Section editors and authors for Orthopaedia may have edited its content or added new information. The use

of information from the American Orthopaedic Foot and Ankle Society should not be construed as support

for or endorsement by that organization for any new information added by Orthopaedia members, or for any

editing of the original content.

Learn about the AOFAS

Some patients may present degenerative changes on both ankles concurrently. In such patients, an1

appropriate treatment may be bilateral total ankle replacement (TAR).

OrthopaedicsOne Articles

Page of 353 372

1.

2.

3.

There is limited information in the current literature about the outcome of simultaneous bilateral TAR.

Karantana et al published a case series including five consecutive simultaneous bilateral TARs performed

between 2002 and 2006 using the STAR prosthesis. The mean follow-up in this patient cohort was 462

months, with a range between 26 and 65 months. Two patients had delayed wound healing. In one patient, a

stress fracture of medial malleolus was seen 10 weeks postoperatively and treated nonoperatively in an

Aircast splint for 6 weeks. At the latest follow-up, all patients experienced significant pain relief and

presented with good functional outcome and reported excellent satisfaction relating to their experience.2

Barg et al recently compared pain relief, quality of life, and functional outcome of 23 simultaneous bilateral

TARs with that of 46 matched unilateral TARs. After 4 months, patients with simultaneous bilateral TAR had3

a higher pain level and worse functional outcome and quality of life as assessed by using AOFAS and SF-36

scores, respectively. However, the observed differences disappeared at the 1- and 2-year follow-up.3

In summary, the two available clinical studies suggest that simultaneous bilateral TAR under one anesthetic

can be offered to patients with bilateral severe ankle OA. However, further studies including more patients2,3

with clinical and radiographic review at longer follow-up should be performed to prove the safety and efficacy

of this procedure.

References

Valderrabano,V., Horisberger,M., Russell,I., Dougall,H., and Hintermann,B.: Etiology of ankle

osteoarthritis. Clin Orthop Relat Res, 467:1800-1806, 2009.

Karantana,A., Martin,G.J., Shandil,M., and Dhar,S.: Simultaneous bilateral total ankle replacement

using the S.T.A.R.: a case series. Foot Ankle Int, 31:86-89, 2010.

Barg,A., Knupp,M., and Hintermann,B.: Simultaneous bilateral versus unilateral total ankle

replacement: A patient-based comparison of pain relief, quality of life and functional outcome. J Bone

Joint Surg Br, 92:1659-1663, 2010.

66.20.2 Takedown of Painful Ankle Fusion and Total Ankle

Replacement

This article is taken wholly from, or contains information that was originally published by the

American Orthopaedic Foot and Ankle Society.

more

OrthopaedicsOne Articles

Page of 354 372

Section editors and authors for Orthopaedia may have edited its content or added new information. The use

of information from the American Orthopaedic Foot and Ankle Society should not be construed as support

for or endorsement by that organization for any new information added by Orthopaedia members, or for any

editing of the original content.

Learn about the AOFAS

Ankle fusion as a treatment option for end-stage ankle osteoarthritis (OA) has been reported as the “gold

standard”. Most patients who undergo ankle fusion experience significant early pain relief postoperatively.1

However, many patients with an ankle fusion develop long-term degenerative changes in adjacent joints.1-3

This progressive development of degenerative changes may require addition fusion surgery, eg, subtalar2,4

fusion.5

A nonunion or malunion of an ankle fusion is an uncommon but severe post-surgical complication. In1

patients with painful nonunited or malunited ankle arthrodesis, takedown of the arthrodesis and subsequent

implantation of a total ankle prosthesis may be a treatment option.

Newton performed three total ankle replacements (TARs) using the Newton Ankle Implant in patients with

nonunion of a prior ankle fusion. The initial ankle fusion was performed because of avascular necrosis of6,7

the talus. In one patient, four unsuccessful fusion attempts were performed. All three implants failed,

requiring one prosthesis revision and two lower leg amputations. Therefore, the author stated that TAR

should not be performed in patients with a previously failed ankle fusion.6,7

In 2004, Greisberg et al published a retrospective study of 23 procedures in 22 patients to review takedown

of ankle fusion and conversion to TAR using the Agility prosthesis. At a mean follow-up of 30 months, 188

patients (19 ankles) were available for clinical and radiological review. In three patients, a lower leg

amputation was performed because of significant residual pain. In the remaining 16 ankles, a significant

functional improvement was detected as assessed by the AOFAS score. Thus, in this study a remarkable

rate of postoperative complications with a failure rate of 42.1% was reported. However, the authors

suggested that this procedure is a viable alternative to amputation, primarily in patients with a definable

source of pain and who have not had previous malleolar resection during the initial ankle fusion.8

Between 1999 and 2004, Hintermann et al performed 29 conversions of painful ankle fusion to TAR in 27

patients using the HINTEGRA TAR. At a mean follow up of 56 months, most patients (82.7%) were9,10

satisfied with results and showed significant pain relief and functional improvement as assessed using VAS

and AOFAS scores, respectively. One ankle had to be revised because of persistent pain and significant

loosening of the talar component. All but one tibial component were stable; the talar component was found to

have migrated in four ankles but was asymptomatic in two of them.9,10

Atkinson et al published a case report addressing clinical outcome and gait analysis following conversion of

tibiotalocalcaneal fusion to TAR using HINTEGRA implants. Two years after the conversion to TAR, the11

patient was subjectively delighted with her increased mobility and functional improvement. The objective

results of her gait analysis, which included comfortable walking pace, stride length, and cadence, showed a

trend toward normalization of gait mechanics.11

In 2010, Barg and Hintermann published the detailed technique of this procedure. The absolute12

contraindications for the procedure include:

OrthopaedicsOne Articles

Page of 355 372

1.

2.

3.

4.

5.

6.

7.

8.

9.

10.

11.

Clubfoot deformity

* Non-manageable hindfoot deformity

* Highly comprised soft tissues and large scars on the medial side

* Severe vascular and/or neurological deficiency

* Active osteomyelitis and/or deep infection

* Chronic pain syndrome existing over years

* Neuropathic disorders (Charcot arthropathy).12

This procedure may be used with some restrictions (relative contraindications) in the following cases:

* Previous fibulectomy

* Long-standing immobilization

* More than 3 cm shortening of the affected leg

* High demands for physical and sports activities

* Diabetic syndrome without polyneuropathy.12

In conclusion, current studies show favorable mid-term results after conversion to TAR in patients with

painful ankle arthrodesis. However, this procedure is technically more demanding that primary TAR and9-11

should be limited to foot and ankle surgeons with adequate experience in primary TAR. Additionally, careful

preoperative planning and selection of patients are very important factors in achieving a good clinical

outcome.

References

Nihal,A., Gellman,R.E., Embil,J.M., and Trepman,E.: Ankle arthrodesis. Foot Ankle Surg, 14:1-10,

2008.

Coester,L.M., Saltzman,C.L., Leupold,J., and Pontarelli,W.: Long-term results following ankle

arthrodesis for post-traumatic arthritis. J Bone Joint Surg Am, 83-A:219-228, 2001.

Plaass,C., Knupp,M., Barg,A., and Hintermann,B.: Anterior double plating for rigid fixation of isolated

tibiotalar arthrodesis. Foot Ankle Int, 30:631-639, 2009.

Fuchs,S., Sandmann,C., Skwara,A., and Chylarecki,C.: Quality of life 20 years after arthrodesis of the

ankle. A study of adjacent joints. J Bone Joint Surg Br, 85:994-998, 2003.

SooHoo,N.F., Zingmond,D.S., and Ko,C.Y.: Comparison of reoperation rates following ankle

arthrodesis and total ankle arthroplasty. J Bone Joint Surg Am, 89:2143-2149, 2007.

Newton,S.E., III: Total ankle arthroplasty. Clinical study of fifty cases. J Bone Joint Surg Am,

64:104-111, 1982.

Newton,S.E.: An artificial ankle joint. Clin Orthop Relat Res, 142:141-145, 1979.

Greisberg,J., Assal,M., Flueckiger,G., and Hansen,S.T., Jr.: Takedown of ankle fusion and conversion

to total ankle replacement. Clin Orthop Relat Res, 424:80-88, 2004.

Hintermann,B., Barg,A., Knupp,M., and Valderrabano,V.: Conversion of painful ankle arthrodesis to

total ankle arthroplasty. J Bone Joint Surg Am, 91:850-858, 2009.

Hintermann,B., Barg,A., Knupp,M., and Valderrabano,V.: Conversion of painful ankle arthrodesis to

total ankle arthroplasty. Surgical technique. J Bone Joint Surg Am, 92 Suppl 1 Pt 1:55-66, 2010.

Atkinson,H.D., Daniels,T.R., Klejman,S., Pinsker,E., Houck,J.R., and Singer,S.: Pre- and

postoperative gait analysis following conversion of tibiotalocalcaneal fusion to total ankle arthroplasty.

Foot Ankle Int, 31:927-932, 2010.

OrthopaedicsOne Articles

Page of 356 372

12.

1.

2.

3.

Barg,A. and Hintermann,B.: Takedown of painful ankle fusion and total ankle replacement using a

3-component ankle prosthesis. Tech Foot & Ankle, 9:190-198, 2010.

66.21 TARIC Total Ankle Replacement

This article is taken wholly from, or contains information that was originally published by the

American Orthopaedic Foot and Ankle Society.

more

Section editors and authors for Orthopaedia may have edited its content or added new information. The use

of information from the American Orthopaedic Foot and Ankle Society should not be construed as support

for or endorsement by that organization for any new information added by Orthopaedia members, or for any

editing of the original content.

Learn about the AOFAS

The TARIC TAR was developed by Schill in Germany and has been used since 2006. The TARIC1

prosthesis has a titanium coating and is optionally available with an additional hydroxyapatite coating. The

tibial component has two fixation pegs. Before creating the TARIC TAR, the designer had experience with2

the TRP and STAR prostheses.3

Currently, there are no studies available addressing clinical outcomes or surgical technique of the TARIC

TAR.

66.21.1 References

Schill,S., Rehart,S., and Fink,B.: Endoprothetik am rheumatischen oberen Sprunggelenk: Historie und

Zukunftsperspektive. Fuss Sprungg, 4:98-105, 2006.

Gougoulias,N.E., Khanna,A., and Maffulli,N.: History and evolution in total ankle arthroplasty. Br Med

Bull, 89:111-151, 2009.

Schill,S., Biehl,C., and Thabe,H.: Ankle prostheses. Mid-term results after Thompson-Richards and

STAR prostheses. Orthopade, 27:183-187, 1998.

66.22 TNK Total Ankle Replacement

OrthopaedicsOne Articles

Page of 357 372

This article is taken wholly from, or contains information that was originally published by the

American Orthopaedic Foot and Ankle Society.

more

Section editors and authors for Orthopaedia may have edited its content or added new information. The use

of information from the American Orthopaedic Foot and Ankle Society should not be construed as support

for or endorsement by that organization for any new information added by Orthopaedia members, or for any

editing of the original content.

Learn about the AOFAS

The TNK prosthesis was developed by Dr. Takakura in Japan in 1975. It is currently the only total ankle

prosthesis with alumina ceramic components. The third generation of this device has shown promising

results in reports by the designer, but these have not been results have not been duplicated by others.

In 1988, Takakura et al reported a comparative study of cemented metal and uncemented ceramic TNK

ankle prostheses.1 Both types were two-component prostheses with high-density polyethylene fixed to the

tibial component. Before 1980, the authors implanted 30 cemented stainless steel prostheses (first

generation); after 1980 they used 9 cemented and 30 uncemented ceramic prostheses (second generation).

The mean follow-up time for cemented and uncemented total ankle replacements (TARs) was 8.1 and 4.1

years, respectively.

Patients who underwent TAR with an uncemented prosthesis were 67% more satisfied than those who

underwent the procedure with a cemented prosthesis (27%). Five metal TARs and one ceramic TAR had to

be revised (five arthrodeses and one revision arthroplasty).1

The initial encouraging results were not maintained in another study by the same working group, published

in 2004. Significant loosening and subsidence of the prosthesis were observed in most patients with the2

first-generation TAR (stainless steel prosthesis). The second-generation TAR (ceramic prosthesis) was

implanted in 60 ankles between 1980 and 1991 (in 12 ankles, cement fixation was used). However,

loosening and significant subsidence occurred in most patients within 5 years after the surgery.

In 1991, the ceramic prosthesis (second generation) was modified by adding bead-formed alumina coated

with hydroxyapatite (third generation). Between 1991 and 2001, this prosthesis type was used in 70 ankles.

The mean follow up in this patient group was 5 years. In three patients, the replaced ankle had to be2,3

fused (one case with deep infection, two cases with severe subsidence of talar component). Overall results

were excellent or good in the most patients (52 ankles). The patient satisfaction rate was higher in patients

with osteoarthritis (OA) compared with patients with rheumatoid OA.2,3

OrthopaedicsOne Articles

Page of 358 372

The TNK prosthesis has been used for treatment of rheumatoid OA with poor clinical and radiographic

outcomes, as shown in two studies. Nishikawa et al reviewed 26 patients (six patients received bilateral4,5

TAR) with rheumatoid OA who were treated with a TNK prosthesis between 1984 and 2000. At a mean

follow up of 72 months 27 ankles in 21 patients were reviewed. Three ankles had to undergo revision

surgery. Nine patients reported their outcome as excellent or good, but patients complained about residual

pain in 13 ankles. A high rate of radiolucency was observed in this patient cohort: migration of the tibial

component in 13 ankles and collapse of the talus in nine.4

Nagashima et al reported results of 21 TARs performed in 19 patients between 1998 and 2002 using the

TNK prosthesis. Hybrid-type fixation (talus component cemented, tibial component not cemented) was used5

in 15 of 21 cases, while both components were cemented in the remaining cases. Early postoperative

complications included two patients with delayed wound healing and one patient with deep infection. In 11 of

21 ankles, significant radiolucency lines were observed. Most patients experienced pain relief and functional

improvement; however, postoperative improvement of ROM directly correlated with preoperative ROM.5

Shinomiya et al performed 20 TARs using TNK prosthesis in 18 selected patients with rheumatoid OA

between 1988 and 1996. All patients were reviewed at a mean follow up of 8 years, with a range of 5 to 126

years. All patients experienced substantial pain relief and showed functional improvement superior to those

who underwent ankle arthrodesis in the same period. No conversions to ankle fusion or revision

arthroplasties were necessary in this cohort. However, a radiolucent line was observed in all replaced ankles

at the final follow up. The authors stated that TAR may be useful in young patients with rheumatoid OA,

considering their better postoperative quality of life.6

Shi et al developed a special hydroxyapatite augmentation for bone atrophy in TAR for rheumatoid patients

receiving a TNK prosthesis. This specially designed hydroxyapatite coating was used in 16 ankles (147

patients) and results were reviewed at a mean follow up of 23 months. More than half of all ankles showed a

radiolucency zone between hydroxyapatite and the tibial component on radiographs at the final follow up.

However, no significant changes of coating position were registered; also, no significant subsidence was

noted. The authors suggested that this new technique using hydroxyapatite may increase the primary

implant fixation to the bone, especially in patients with rheumatoid OA and showing significant bone atrophy.7

Recently, Tsukamoto et al reported a case using a custom-made alumina ceramic total talar component for

treatment of collapse of the talar body in one patient who received a TNK prosthesis. The revision surgery8

was performed in a 56-year-old female patient 4 years after the initial TAR. The authors achieved substantial

pain relief and functional improvement especially regarding walking ability. This case report demonstrated a

feasible revision treatment in patients with failed TAR.8

In summary, since its introduction in 1975, the TNK prosthesis has undergone many modifications to

address the material of the components (stainless steel, polyethylene, alumina ceramic), coating

(without/with hydroxyapatite), and fixation (cement/cementless fixation). Currently, this is the only TAR

design with alumina ceramic components. While the studies by the designer reported favorable results using

the third-generation TNK prosthesis, independent studies addressing TAR results in patients with2,3

rheumatoid OA show less-promising results.4,5

66.22.1 References

OrthopaedicsOne Articles

Page of 359 372

1.

2.

3.

4.

5.

6.

7.

8.

Takakura,Y., Tanaka,Y., Sugimoto,K., Tamai,S., and Masuhara,K.: Ankle arthroplasty. A comparative

study of cemented metal and uncemented ceramic prostheses. Clin Orthop Relat Res, 252:209-216,

1990.

Takakura,Y., Tanaka,Y., Kumai,T., Sugimoto,K., and Ohgushi,H.: Ankle arthroplasty using three

generations of metal and ceramic prostheses. Clin Orthop Relat Res, 424:130-136, 2004.

Tanaka,Y. and Takakura,Y.: [The TNK ankle: short

- and mid-term results]. Orthopade, 35:546-551, 2006.

Nishikawa,M., Tomita,T., Fujii,M., Watanabe,T., Hashimoto,J., Sugamoto,K., Ochi,T., and

Yoshikawa,H.: Total ankle replacement in rheumatoid arthritis. Int Orthop, 28:123-126, 2004.

Nagashima,M., Takahashi,H., Kakumoto,S., Miyamoto,Y., and Yoshino,S.: Total ankle arthroplasty for

deformity of the foot in patients with rheumatoid arthritis using the TNK ankle system: clinical results

of 21 cases. Mod Rheumatol, 14:48-53, 2004.

Shinomiya,F., Okada,M., Hamada,Y., Fujimura,T., and Hamada,D.: Indications of total ankle

arthroplasty for rheumatoid arthritis: evaluation at 5 years or more after the operation. Mod

Rheumatol, 13:153-159, 2003.

Shi,K., Hayashida,K., Hashimoto,J., Sugamoto,K., Kawai,H., and Yoshikawa,H.: Hydroxyapatite

augmentation for bone atrophy in total ankle replacement in rheumatoid arthritis. J Foot Ankle Surg,

45:316-321, 2006.

Tsukamoto,S., Tanaka,Y., Maegawa,N., Shinohara,Y., Taniguchi,A., Kumai,T., and Takakura,Y.:

Total talar replacement following collapse of the talar body as a complication of total ankle

arthroplasty: a case report. J Bone Joint Surg Am, 92:2115-2120, 9-1-2010.