SINIUS Brief operating instructions

Transcript of 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions...

Page 1: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

SINIUSBrief operating instructions

6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

Page 2: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

2 Dentsply Sirona

Dear Customer,You can use these brief operating instructions, valid for softwareversion 2.2 and higher, to quickly find the most importantcontrol procedures and functions of your dental treatmentcenter.

However, they do not describe all of its functions. Therefore,they cannot replace the complete operating instructions of yourdental treatment center, which contain many important descrip-tions and warnings. The care and cleaning of your treatmentcenter are described thoroughly in the complete operatinginstructions. A quick overview of the required measures isprovided in a document entitled "Care and Cleaning Plans forthe Treatment Center and the Instruments".

Your SINIUS® team *

Table of contentsSINIUS overview .................................................................... 3SINIUS CS overview ............................................................. 4SINIUS TS overview ............................................................. 5Power switch ........................................................................... 6Main switch .............................................................................. 6EasyTouch user interface .................................................. 7Touchscreen ............................................................................ 8Start program ......................................................................... 9Instrument program .......................................................... 12SIVISION program .............................................................. 16Fixed keys of dentist element ....................................... 19Setup Programs .................................................................. 23Cursor control ...................................................................... 26Programming chair programs ..................................... 28Assistant element .............................................................. 29Foot control .......................................................................... 304-way foot control ............................................................ 32Headrest switches ............................................................. 33SiroCam AF/AF+ intraoral camera ........................... 34SPRAYVIT M ......................................................................... 39SIROSONIC TL ...................................................................... 41Mini L.E.D. ............................................................................... 43Implantology and endodontic treatments ............. 44ApexLocator ......................................................................... 49

*SINIUS® is a registered trademark of Sirona Dental Systems GmbH.

6322858 KGBA SINIUS EN.book Seite 2 Montag, 10. April 2017 4:25 16

Page 3: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 3 Montag, 10. April 2017 4:25 16

SINIUS overview

* Maximum load from patient and accessories: 165kg

Dentist element instruments

* BL motor, BL ISO E motor, BL ISO C motor, or BL Implant motor

Assistant element instru-ments

A Patient chair* E Assistant elementB Headrest F Water unitC Foot control G Power switch and external

device connectionD Dentist element

Position 1 Positions 2, 3

Position 4 Position 5 Extra holder

SPRAYVIT M Motor* Motor* SIROSONIC TL SiroCam AF/AF+

High-speed handpiece

High-speed handpiece

SiroCam AF/AF+

SIROSONIC TL

Position 1 Position 2 Position 3 Position 4Mini L.E.D. SPRAYVIT M Spray aspirator Saliva ejectorSecond spray aspirator or saliva ejector

Dentsply Sirona 3

Page 4: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 4 Montag, 10. April 2017 4:25 16

SINIUS CS overview

* Maximum load from patient and accessories:: 165kg

Dentist element instruments

* BL motor, BL ISO E motor, BL ISO C motor, or BL Implant motor

Assistant element instru-ments

A Patient chair* E Assistant elementB Headrest F Water unitC Foot control G Power switch and external

device connectionD Dentist element

Position 1 Positions 2, 3 Position 4 Position 5SPRAYVIT M Motor* Motor* SIROSONIC TL

High-speed handpiece

High-speed handpiece

SiroCam AF/AF+

SIROSONIC TL

Position 1 Position 2 Position 3 Position 4Mini L.E.D. SPRAYVIT M Spray aspirator Saliva ejectorSiroCam AF/AF+

Second spray aspirator or saliva ejector

4 Dentsply Sirona

Page 5: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 5 Montag, 10. April 2017 4:25 16

SINIUS TS overview

* Maximum load from patient and accessories:: 165kg

Dentist element instruments

* BL motor, BL ISO E motor, BL ISO C motor, or BL Implant motor

Assistant element instru-ments

A Patient chair* E Assistant elementB Headrest F Water unitC Foot control G Power switch and external

device connectionD Dentist element

Position 1 Positions 2, 3

Position 4 Position 5 Extra holder

SPRAYVIT M Motor* Motor* SIROSONIC TL SiroCam AF/AF+

High-speed handpiece

High-speed handpiece

SiroCam AF/AF+

SIROSONIC TL

Position 1 Position 2 Position 3 Position 4Mini L.E.D. SPRAYVIT M Spray aspirator Saliva ejectorSiroCam AF/AF+

Second spray aspirator or saliva ejector

Dentsply Sirona 5

Page 6: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 6 Montag, 10. April 2017 4:25 16

Power switchThe power switch connects the treatment center to the power supply. During longer periods of disuse, the treatment center should be disconnected from the power supply. It then no longer consumes any energy.

Main switchSwitching on the treatment centerThe main switch switches the treatment center from standby mode to operational readiness.

Following switch-on, the operating system is booted and an automatic self-test is performed.

Switching the treatment center to standby mode Press and hold the main switch on the

dentist element until an acoustic signal sounds. Then release the key.

The treatment center then shuts down and switches itself to the standby mode.

The LED of the main switch goes out on the dentist element.

6 Dentsply Sirona

Page 7: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 7 Montag, 10. April 2017 4:25 16

EasyTouch user interface

SINIUS and SINIUS TS

SINIUS CS

* The illustration shows the Standard Start program operating mode

A Fixed keys (membrane keyboard)B Touchscreen (pressure-sensitive user interface)*C Fixed keys (membrane keyboard)D Main switch

Dentsply Sirona 7

Page 8: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 8 Montag, 10. April 2017 4:25 16

TouchscreenOperating conceptThe touchscreen displays virtual function keys according to the program selected. Required functions can be activated either by touching the function keys with your finger or via the cursor with the foot control.

Function keys for functions not included with the treatment center are not displayed on the touchscreen.

Status barThe status line is located at the bottom edge of the screen. If multiple user profiles are preselected, active user profiles A to F are displayed with the current time including seconds next to them. In addition, status messages are displayed to indicate, for example, the need to change the amalgam separator, add water sanitizing agent, charge the battery of the wireless foot control, refill the cleaning agent for chemical treatment of the suction hoses, as well as the number of days left until the next maintenance call or sanitization run and error messages.

Sub-screensSome programs are divided into a main program and sub-screens. The function keys for the basic functions are displayed in the main programs. The sub-screen key (two rectangles) leads to additional setting options.

Settings screens

In many cases, functions can not only be switched on or off, but also can be set. If a function key is pressed and held (> 2 s), the corresponding settings screen appears. This screen is superimposed on the current screen. The screen located in the back-

10:45:03 B

8 Dentsply Sirona

Page 9: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 9 Montag, 10. April 2017 4:25 16

ground is semitransparent and is tempo-rarily disabled for input.

Settings screens and sub-screens are auto-matically hidden after a certain period has elapsed. The Back key (return arrow) closes the opened screen immediately.

Background colors of keysGeneral functions are represented by gray keys. If the corresponding function is switched on or active, the key is displayed orange.

Keys that initiate a dialog change or lead to sub-screens and settings dialogs are displayed blue.

As long as a key remains activated, its active state is marked by a bold black border.

Start programStandard Start programIn the Standard Start program, the chair program function keys and, in the case of a motor-driven headrest, the Move headrest in/out function keys are displayed on the screen.

All other chair functions are listed sepa-rately on the Manual chair adjustment screen. This screen is accessed by pressing the Manual chair adjustment key.

Dentsply Sirona 9

Page 10: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 10 Montag, 10. April 2017 4:25 16

EasyMode Start programThe EasyMode Start program operating mode displays the function keys for the patient chair programs and manual chair adjustment, as well as the quick setting keys for adjusting the intensity of the instruments together on the same screen. The main keys for the treatment are there-fore always displayed on the touchscreen. It is not necessary to switch between the Start and Instrument programs. This makes the operation of the dental treatment center similar to other Dentsply Sirona dental treatment centers that are operated without a touchscreen.

You can set the operating mode you would like to work with in the Setup screen:

Chair program SMouth rinsing position with memory func-tion (freely assignable)

Chair program 0Entry/exit position (freely assignable)

Chair programs 1 and 2(freely assignable)

Move headrest in/out(if a motor-driven headrest is installed)

Tilt headrest(if a motor-driven headrest is installed)

Tilt patient couchCompensated motion of the seat and back-rest without compressing or stretching the patient

Adjusting the chair height

10 Dentsply Sirona

Page 11: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 11 Montag, 10. April 2017 4:25 16

Select user profileA stored user profile can be selected for up to six different users (A to F).

X-ray viewerAlso white screen on SIVISION monitor for SIVISION digital

Open manual chair setting programOnly in the Standard Start operating mode

Treatment functionActivation of the implantology/endodon-tics functions

Open sub-screenAccess to additional subordinate functions; see the following function keys:

Purge water linesManual process for purging instruments

Autopurge – Purge water lines automati-callyAutomatic process for purging water-conducting instruments

SanitizationStarts the treatment center sanitization program

Self-sufficient water supplyOperation of the treatment center with self-sufficient water supply

Extract residual waterExtraction of residual water by standard wet suction before cleaning the filter

External HF surgery unitActivation of the protection against inter-ference caused by an external HF unit

Apex measurement with file clampActivation of the ApexLocator for manual measurement with the file clamp, showing the distance display

Dentsply Sirona 11

Page 12: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 12 Montag, 10. April 2017 4:25 16

SPRAYVIT instrument lightActivates the instrument light on the SPRAYVIT when it is removed from the holder

Active lumbar functionAdjustment of active lumbar support

Instrument programIn the Standard Start program operating mode, the instrument settings can be made with static quick setting keys, program-mable quick setting keys, or function levels.

Static quick setting keysThe values shown on the quick setting keys can be selected by touching them briefly (< 1 s).

Speed quick setting keysFor setting specified speeds or increments

BL, BL ISO C, and BL Implant motors: min. 90 rpm, max. 40,000 rpm

BL ISO E motor: min. 2000 rpm, max. 40,000 rpm

Intensity quick setting keysFor setting preset or intermediate intensity values ranging from 1% to 100%

Setting speed and intermediate inten-sity values with quick setting keys Increase intensity: Press and hold (>1 s)

a quick setting key whose intensity value is greater than or equal to the value displayed in the first line.

Reduce intensity: Press and hold (>1 s) a quick setting key whose intensity value is less than the value displayed in the first line.

Memory key (SaveMode)After an instrument is deposited in its holder, the settings made in the instrument program are saved only if the Memory key is first pressed and held >2 sec.

12 Dentsply Sirona

Page 13: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 13 Montag, 10. April 2017 4:25 16

In the Setup program, you can switch to DropMode. The Memory key is then hidden and when an instrument is deposited, the settings made in the Instrument program are automatically saved.

Programmable quick setting keysThe key values can be changed individually with the programmable quick setting keys.

The speed/intensity shown in the first line can be increased or decreased by pressing and holding (>1 s) the left or right - or + quick setting key.

The programming mode is activated by holding down the middle quick setting key Set (> 2 s). The speed/intensity flashes on the touchscreen and a flashing bar appears behind the quick setting keys. Now press the quick setting key to which the set value is to be saved. A signal sounds as confirma-tion. Further settings such as activation of the coolant or Endo mode are also saved on the quick setting key.

The settings can be recalled by touching the respective quick setting key.

Dentsply Sirona 13

Page 14: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 14 Montag, 10. April 2017 4:25 16

Function levels

Briefly pressing the +/- keys (<1 s) adjusts the speed in larger increments.

Pressing and holding the +/- keys (>1 s) enables fine adjustment.

When using function levels, you have two "storage locations" for each user profile (E1/E2) that can be used to save settings or call them up at the push of a button.

Functions of the instrument programDirection of rotationSwitches counterclockwise rotation ON/OFF

The key is marked orange when CCW rota-tion is activated.

Chip blowerUsed to dry treatment areas or blow off drilling chips by emitting air blasts from the treatment instrument

This key is displayed only when cursor control is activated

Boost functionIncreases the intensity of the SIROSONIC TL ultrasonic handpiece during treatment in 20 steps in reference to the final value

Endo functionActivates the endodontics function of the SIROSONIC TL ultrasonic handpiece (power limit)

14 Dentsply Sirona

Page 15: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 15 Montag, 10. April 2017 4:25 16

Opening sub-screensUsed to access additional subordinate functions; see the following function keys:

Coolant: cooling sprayUsed to cool the treatment area with spray

Coolant: airFor cooling the treatment area with air

NaCl saline solutionFor cooling the treatment area with sterile saline solution

Setting the automatic motor stop of the ApexLocatorWhen the Apex Stop key is pressed, the – and + keys and the Auto-Reverse key are displayed.

AutoReverseSwitches the ApexLocator auto-reverse function on or off

Instrument lightActivates and sets the instrument light

Direct starter / speed foot controlDirect starter: Switches the instrument on at the set speed or intensity

Speed foot control (highlighted orange): Regulates the instrument up to the maximum set speed or intensity depending on the position of the foot control pedal.

Apex acoustic signalsSwitches the apex acoustic signals on and off

Apex distance acoustic signalsSwitches the apex distance acoustic signals on or off

Dentsply Sirona 15

Page 16: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 16 Montag, 10. April 2017 4:25 16

SIVISION programThe treatment center can be connected to a PC via an Ethernet cable. The SIVISION and/or SIUCOM plus PC application enables communication between the treat-ment center and the PC. SIUCOM plus is required for PC control. The PC can be operated directly on the touchscreen and in conjunction with the SIROCAM AF/AF+ intraoral camera, with the foot control of the treatment center as well.

For details, please refer to the "Installation and Configuration of SIUCOM plus / SIVI-SION connect" manual.

The treatment center can be operated with the following PC versions:

• External PCThe treatment center can be connected to an external PC via an Ethernet cable. PC applications such as Media Player, Microsoft® PowerPoint®, SIDEXIS, or SI Video can also be controlled from the treatment center.

• Internal PCThe treatment center is operated as a standalone unit without an external PC. A mini PC is integrated in the base of the chair instead. It is required to display the camera images using the SI Video PC application. It is not possible to control additional PC applications with this version.

• Operation without PCThe use of SIVISION is not possible.

The SI Video application is used in treat-ment centers with external PC, on which SIDEXIS is not installed, or with an internal PC (standalone solution). In these cases, SI Video serves as a video application for the camera.

Starting PC communication All PC applications which were started

from the treatment center have been terminated.

SIUCOM plus has been started, e.g. via Autostart.

16 Dentsply Sirona

Page 17: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 17 Montag, 10. April 2017 4:25 16

Touch the SIVISION program change key next to the touchscreen on the left side.Or, if intraoral scans are desired:Remove the camera from the holder. The SIDEXIS or SI Video PC application starts immediately.

The SIVISION dialog is displayed on the touchscreen.

Select the PC application you would like to communicate with.

For PC applications that access files on the PC (e.g. Media Player, Microsoft Power-Point), the File Selection screen opens. Select the desired file by touching it.

A number of buttons which can be used to control the selected PC application then appear on the touchscreen.

The keys displayed on the touchscreen and their arrangement are preconfigured at the factory. The configuration can, however, be individually customized - see "Installation and configuration of SIUCOM plus / SIVISION connect". You can also define

Dentsply Sirona 17

Page 18: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 18 Montag, 10. April 2017 4:25 16

there which file folder the File Selection screen should access for each PC applica-tion.

The following overview shows the keys available for configuring the SIDEXIS and Media Player PC applications.

Communication with SIDEXIS

Media Player

Next image Contrast opti-mization filter

Tiled layout Relief display filter

Cascaded layout

Smooth image

Status over-view layout

Sharpen image

Full frame Undo

Zoom in/out Restore orig-inal image

Rotate image 90° left/right

Close current media window

Rotate image 180°

Close all media windows

Previous/next title

Mute

Stop playback Adjust volume

Start/interrupt playback

18 Dentsply Sirona

Page 19: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 19 Montag, 10. April 2017 4:25 16

Fixed keys of dentist elementProgram change keysThe program change keys allow the user to switch between the main programs Start program, Instrument program and SIVI-SION program in the Standard Start program operating mode.

In the EasyMode Start program operating mode, the program change keys Chair and Instrument can be used to switch to the relevant sub-screens.

Fixed keysMain switchSwitches the treatment center on/off [6].

Timer functionTriggers elapsed time, which is displayed in the status bar of the touchscreen.

Presetting the timerWhen the Timer key is pressed (>2 s), the Timer function settings screen appears.

Select one of the four timers to change its setting. Touch one of the selection keys along the bottom edge of the settings screen.

The selected timer is highlighted orange. Use the – and + keys to set the desired

time. Increments:From 00:05 to 1:00 = 5-s incrementsFrom 1:00 to 3:00 = 10-s incrementsFrom 3:00 to 9:30 = 30-s increments

Select whether the time loop and the acoustic signal should be activated/deactivated for the selected timer. Touch the Time loop and/or Acoustic signal key.

If a function is switched on, the corre-sponding key is highlighted orange.

Dentsply Sirona 19

Page 20: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 20 Montag, 10. April 2017 4:25 16

Select another timer for adjustment or close the settings screen with the Return key.

All settings are automatically saved when the screen is closed.

Starting the timer Press the Timer fixed key briefly. The last timer used is started immedi-

ately. The set and elapsed time are displayed in the status bar. The Timer function screen remains visible.

If you wish to use a different timer, touch one of the timers at the lower edge of the screen.

Shock positioningImmediately moves the patient chair to a position for shock positioning of the patient.

The position for shock positioning preset at the factory can be reprogrammed.

Move the patient chair to the desired position.

Press and hold the shock positioning key >2 s.

Operating lightSwitches the operating light on/off. If the operating light key is pressed for >2 s, the settings screen appears. For the LEDview, the light as well as the distance from which the no-touch sensor should react to move-ments can be adjusted.

In the LEDview Plus, the color temperature of the light can also be adjusted.

20 Dentsply Sirona

Page 21: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 21 Montag, 10. April 2017 4:25 16

Composite functionSwitches the composite function for the operating light on/off.

This function is used to prevent premature curing of composite fillings.

Tumbler fillingStarts the tumbler filling function.

When the Tumbler filling key is pressed (>2 s), the filling time and water tempera-ture settings screen appears.

Touch the Link tumbler filling to mouth rinsing position key.

If the key is marked orange, the tumbler filling function will automatically be switched on for the duration of the preset filling time when the mouth rinsing position (S) chair program is activated.

Touch the Tumbler temperature control key.

If the key is highlighted orange, the tumbler temperature control function is activated.

Set the tumbler filling time with the + and – keys.

Flushing the cuspidorStarts the flushing of the cuspidor bowl.

If the Flushing key is pressed >2 s, the Flushing Time settings screen appears.

Set the flushing time with the + and – keys.

Touch the Link flushing to mouth rinsing position S key.

Dentsply Sirona 21

Page 22: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 22 Montag, 10. April 2017 4:25 16

If the key is marked orange, flushing is automatically activated for the duration of the flushing time set when the chair is moved to the mouth rinsing position S.

Freely selectable functione.g. call key

This function can be preset as a button or as a switch in the Setup program.

Freely available relay 230 V, 6 A(connected by service engineer)

Freely selectable functionThis function can be preset as a button or as a switch in the Setup program.

Freely available relay 230 V, 6 A(connected by service engineer)

Clean keyDeactivates the complete user interface of the dentist element except for the main switch.

This is used to make sure that no unwanted functions can be accidentally triggered while cleaning the surface. Press the key again >3 s or step on the foot control to reactivate the user interface.

Setup keyUsed for individual configuration of the treatment center by the user and for reading out messages by the service engi-neer.

When an instrument is removed from the holder, this key opens the corresponding setup program, e.g. to set water tempering of SPRAYVIT.

22 Dentsply Sirona

Page 23: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 23 Montag, 10. April 2017 4:25 16

Setup ProgramsVarious treatment center functions can be individually configured via the Setup settings. The treatment center can thus be adapted to match the user's personal method of treatment.

All instruments are in their holders. Press and hold the Setup fixed key for

>2 s. The Setup program appears on the

touchscreen.

You can adapt the basic settings of the treatment center in six sub-screens. The selected function is highlighted by an orange frame.

EasyTouch user interface

Switch the key sound on/off

Calibrate the touchscreen

Adjusting the touchscreen brightness

Date and time

Set the date

Set the time

Switch to 12 or 24 hour display

Dentsply Sirona 23

Page 24: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 24 Montag, 10. April 2017 4:25 16

Control options

Pre-select the number of user profiles

Set cursor control (without/with cursor control, cursor control with program change)

Select the operating mode of the Start program

Switch the foot control function for intra-oral camera focus on/off

Link the spray aspirator to the 4-way foot control

Link the headrest tilt to the 4-way foot control

Set the hash key function on the assistant element

Set bell/hash fixed key to pushbutton or switch actuation

Show/hide White screen key on the SIVI-SION monitor (only possible when no X-ray viewer is available)

Set the cleaning agent mixture for chem-ical suction hose cleaning

Adjust temperature of the tumbler heater

Link tumbler heater to chair programs

Instruments

Pre-select how instrument settings are to be saved

Switch afterblow on/off

Show/hide the key for the external HF surgical unit

Set spray temperature

24 Dentsply Sirona

Page 25: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 25 Montag, 10. April 2017 4:25 16

Network connectionThe network connection must be config-ured for communication between the treat-ment center and the PC. The IP address (Internet protocol address), the subnet mask and the gateway (transition point to other networks) can be configured at the treatment center.

The treatment center can be integrated into the practice network by means of a static network address or via DHCP (Dynamic Host Configuration Protocol).

The network configuration is described in detail in the "SINIUS installation instruc-tions".

Service domainThe Service domain is intended to be used only by service engineers.

Dentsply Sirona 25

Page 26: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 26 Montag, 10. April 2017 4:25 16

Cursor controlThe touchscreen and the fixed keys of the dentist element can also be operated hand-free via the foot control.

Current cursor positionIf the cursor control is activated, see "Setting cursor control" [24], the current position of the cursor is displayed by an orange bar between the pair of keys on the touchscreen or the fixed keys.

Orange and blue barsA blue bar indicates which functions are assigned to the left or right key of the foot control. E.g. in the Start program, the keys are assigned to the mouth rinsing position (S) and the entry/exit position (0) (see example on the left), but in the Instrument program to motor and turbine with spray or chip blower.

If the cursor control without program change is activated, the blue bars also can be selected with the cursor.

If the cursor control with program change is activated, the blue bars are skipped for faster navigation. If the orange cursor is at the top starting point in this mode, pressing it again switches the program.

Cursor returnOnce a function is activated with the cursor control, the orange cursor typically returns to the starting position of the program, for instance after switching on the X-ray image viewer, the operating light, or cuspidor flushing. The cursor position will not change for functions that are operated by permanent shifting of the 4-way foot control plate, e.g., manual chair adjustment.

S

0

S

0

S 0

26 Dentsply Sirona

Page 27: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 27 Montag, 10. April 2017 4:25 16

Start program cursor path

The cursor path runs between the pairs of keys, moving from top to bottom and from left to right, usually in multiple loops. The cursor path can be traversed between the starting and end points either in a forward or a reverse direction.

If no further cursor position is available on the touchscreen, the cursor jumps out of the touchscreen. The cursor path is then continued among the fixed keys on the EasyTouch user interface.

Instrument program cursor path

In Instrument programs, all quick setting keys are selected simultaneously. This is indicated by a horizontal orange bar located behind the quick setting keys. The speed or intensity is then set by actuating the 4-way foot control plate to the left or right briefly (values on quick setting keys) or for a longer time (intermediate values).

Dentsply Sirona 27

Page 28: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 28 Montag, 10. April 2017 4:25 16

Programming chair programsThe four factory-set chair programs can be individually reprogrammed for each of the six user profiles (A to F):

• Mouth rinsing position S

• Entry/exit position 0

• 1 and 2

The Start program is displayed on the touchscreen.

1. Move the patient chair to the desired treatment position.

2. For motor-driven headrest: Tilt the headrest to the desired treatment position.

3. Switch the operating light on or off (this is set during programming).

4. Press and hold (>2 s) the desired program key (S, 0, 1 or 2).

An acoustic signal sounds. Your settings are now stored under the desired program key.

28 Dentsply Sirona

Page 29: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 29 Montag, 10. April 2017 4:25 16

Assistant element

Tumbler fillingon/off

Flushing the cuspidoron/off

Operating lightPress the key repeatedly to set the following switching positions:

1. Switched on

2. Composite function

3. Switched off

Chair program SMouth rinsing position with memory func-tion (programmable)

Chair program 0Entry/exit position (programmable)

Hash keyThe Hash key on the assistant element can be configured in the Setup program:

• Switch the X-ray viewer or the white screen function on the SIVISION monitor on/off or

• Control the bell or hash relay as a key or a switch

Freely available relay 230 V, 6 A(connected by service engineer)

Dentsply Sirona 29

Page 30: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 30 Montag, 10. April 2017 4:25 16

Foot controlFoot pedalActivating the instrument An instrument is removed. Step on the foot pedal. The instrument starts running. If appropriate, intensity control relative

to pedal movement (if speed foot control has been preset).

When the intraoral camera is removed from the holder, the camera image focuses and the display switches to a still or live image.

4-way foot control plateCursor control The cursor control is switched on. Shift 4-way foot control plate.

Cursor directionsA: To the front – cursor upwardsB: To the back – cursor downwardsC: To the left – key function

at left of cursorD: To the right – key function

at right of cursor

Cursor navigation along the cursor path and triggering of the corresponding key function.

CW/CCW rotation for the electric motor The cursor control switched off. An electric motor is removed from the

holder. Shift the 4-way foot control plate to the

right or to the left. The clockwise/counterclockwise rota-

tion of the electric motor is activated.

Left buttonActivating rinsing position S All instruments are in place. Press the left button. The chair moves to mouth rinsing posi-

tion S.

Spray, air or NaCl on/off An instrument (motor, turbine, SIRO-

SONIC TL) is removed. Press the left button. Spray, air or NaCl is switched on/off.

S

0

S

0

B

CA

DB

CA

D

S

0

S

0

S

30 Dentsply Sirona

Page 31: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 31 Montag, 10. April 2017 4:25 16

If the intraoral camera is removed, the video still image is saved in SIDEXIS; the live image is displayed in the next quadrant in SI Video.

Right buttonActivating entry/exit position 0 All instruments are in place. Press the right button. The chair moves to entry/exit position 0.

Activating the chip blower An instrument (motor, turbine) is

removed. Press the right button. The chip blower remains activated as

long as the button is pressed.

The boost function can be selected in ultra-sound mode. When the intraoral camera is removed, it is possible to toggle between single image and quad image in SI Video.

Setting the wireless foot control on the treatment centerThe wireless foot control must be assigned to the treatment center via by registering it. This prevents malfunctions caused by neighboring wireless foot controls.

The treatment center and wireless foot control are ready for operation.

All instruments are in their holders. Simultaneously press and hold the left

and right buttons of the foot control (> 2 s).

An acoustic signal sounds. The following message appears on the touch screen:

Confirm registration with the OK button. The message is hidden. The wireless

foot control is assigned to the treatment center.

S

0

0

S

0

S 0

Dentsply Sirona 31

Page 32: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 32 Montag, 10. April 2017 4:25 16

4-way foot controlAdjusting the chair height Slide the 4-way foot control upward or

downward. The chair height is set.

Tilting the patient couch Slide the 4-way foot control to the left

or right. Tilt the patient couch up or down.

Headrest tilt via 4-way foot controlIf a motor-driven headrest is used, you can replace control of the Tilt patient couch function by the headrest tilt function on the 4-way foot control in the Setup program Control options.

Suction via 4-way foot controlThe Setup program control options can be set so that when the suction hose is removed from its holder, the suction flow of the spray aspirator can be interrupted and switched back on by moving the 4-way foot control at the base of the chair in any direction.

WARNING: Due to the lack of suction flow, fluid may run back out of the spray aspi-rator and into the patient's mouth.

Always remove the spray aspirator from the patient's mouth before switching it off.

32 Dentsply Sirona

Page 33: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 33 Montag, 10. April 2017 4:25 16

Headrest switchesInclining the headrest Slide the upper 4-way switch to the left

or right.

Moving the headrest in/out Slide the upper 4-way switch upward or

downward.

Tilting the patient couch Slide the lower 4-way switch to the left

or right.

Adjusting the chair height Slide the lower 4-way switch upward or

downward.

Tilting the headrest via quick mechan-ical adjustment Hold the headrest securely in place

before unlocking it. Press buttons A together. The headrest is thus disconnected from

the motor drive and can be tilted manu-ally.

A A

Dentsply Sirona 33

Page 34: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 34 Montag, 10. April 2017 4:25 16

SiroCam AF/AF+ intraoral camera

The SIVISION digital video system enables intraoral and extraoral images to be made. The SiroCam AF/AF+ intraoral camera generates digital image data that can be transmitted via a USB 2.0 port to a connected PC and stored there. This PC can then display the images on the SIVI-SION monitor of the treatment center. The SIDEXIS or SI video can be used as a PC video application.

The video images are an excellent option for improving patient communication.

Focusing the intraoral cameraThe SiroCam AF/AF+ intraoral camera focuses the image at the touch of a button. The focus is continuously variable.

Direct the camera to the subject to be acquired.

Press the Autofocus A key.

In the treatment center setup, you can configure whether the camera image can be focused using the foot control, see "Switch the foot control function for intra-oral camera focus on/off" [24].

Using the camera with SIDEXISBesides displaying and processing X-ray images, SIDEXIS can also be used as a video application for the SiroCam AF/AF+ intraoral camera. SIDEXIS displays live and still images in separate windows. Still images are stored in the patient database.

Switching the intraoral camera on/off The PC is in operation and the SIVISION

connect or SIUCOM plus PC application is started.

Remove the intraoral camera from the holder.

The SIVISION dialog is displayed on the touchscreen. SIDEXIS is started and the live image appears on the SIVISION monitor.

A Lens windowB Key for automatic

focus (Auto focus)C Connecting cable

SiroCam AF

CBA

34 Dentsply Sirona

Page 35: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 35 Montag, 10. April 2017 4:25 16

Generating a still image The live image is displayed on the SIVI-

SION monitor. Step on the foot pedal. The display switches from live to still

image. Press the foot pedal again. The live image is again displayed on the

SIVISION monitor.

In the treatment center setup, you can configure whether the camera image can be focused using the foot control, see "Switch the foot control function for intra-oral camera focus on/off" [24]. In this case, the foot pedal must be pressed down fully to switch between live and still images.

Saving an image A patient registered in SIDEXIS. The still image to be saved is displayed

on the SIVISION monitor. Press the left button of the foot switch. An acoustic signal sounds. The still

image is saved.

Displaying and processing stored imagesImages stored on the PC can be redis-played and processed on the SIVISION monitor. To do this, use the keys for controlling SIDEXIS located on the touch-screen of the treatment center (see "Communication with SIDEXIS" [ 18]).

S

0

S

0

S

Dentsply Sirona 35

Page 36: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 36 Montag, 10. April 2017 4:25 16

Using the camera with SI VideoUsing SI Video, up to four still images can be generated with the intraoral camera. These can be displayed on the SIVISION monitor either as a sequence of single images or simultaneously as a quad image. The active quadrant is marked with an orange square. The quadrants are displayed in a clockwise sequence.

The generated still images remain until the treatment center or the PC is disconnected from the power supply.

Remove the intraoral camera from the holder.

The live image is displayed on the SIVI-SION monitor as a single image. The SIVISION dialog is displayed on the touchscreen.

When the intraoral camera is put in the holder, the live image is no longer displayed on the monitor. The generated still images remain displayed.

Operation of SI Video with the camera removedWith the camera removed, SI Video is controlled via the foot control and the SIVI-SION program keys.

Switching between live and still image The live image is displayed on the SIVI-

SION monitor Step on the foot pedal. The display switches from live to still

image. Press the foot pedal again. The live image is again displayed on the

SIVISION monitor.

In the treatment center setup, you can configure whether the camera image can be focused using the foot control, see "Switch the foot control function for intra-

S

0

36 Dentsply Sirona

Page 37: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 37 Montag, 10. April 2017 4:25 16

oral camera focus on/off" [24]. In this case, the foot pedal must be pressed down fully to switch between live and still images.

Selecting the next quadrantIn order to produce another still image, another quadrant must be selected. If a still image is already displayed in the desired quadrant, it can be replaced by a new still image.

The intraoral camera must be removed from its holder to allow for operation via the foot control.

Press the left button of the foot switch.Or touch the Select next quadrant key on the touchscreen.

The orange square highlights the selected quadrant. When the intraoral camera is removed, the live image is displayed.

Switching between quad image and single imageWhen you change from quad to single image, the previously marked quad image is displayed as a single image. Conversely, the displayed single image is marked in the quad image.

The intraoral camera must be removed from its holder to allow for operation via the foot control.

Press the right button of the foot control.Or touch the Quad image key on the touchscreen.

The mode changes between quad and single image.

S

0

S

S

0

0

Dentsply Sirona 37

Page 38: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 38 Montag, 10. April 2017 4:25 16

Operation of SI Video with the camera in placeWith the camera in place, SI Video is controlled via the SIVISION program buttons.

Switching between quad image and single imageIn single image mode, the still image in the next quadrant is displayed. In quad image mode, the next quadrant is highlighted.

Quad imageDisplay quad image or single image. Up to four single images are simultaneously displayed on the SIVISION monitor in quad image mode.

Deleting imagesAll generated still images are deleted.

38 Dentsply Sirona

Page 39: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 39 Montag, 10. April 2017 4:25 16

SPRAYVIT MA = air button

B = water button

Available option: SPRAYVIT M with reverse key assignment

Switching the SPRAYVIT M instrument light on/off in the Start sub-program Go to the Start sub-screen.

Take a SPRAYVIT M multifunctional syringe from the holder of the dentist or assistant element.

The SPRAYVIT instrument light key is displayed in the Start sub-screen.

Touch the Instrument light SPRAYVIT key.

When the key is highlighted orange, the instrument light at the SPRAYVIT M is turned on if it is the only instrument to have been removed.

Dentsply Sirona 39

Page 40: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 40 Montag, 10. April 2017 4:25 16

Activating/deactivating and setting the water temperature control in the SPRAYVIT M setup program Remove the SPRAYVIT M at the dentist

or assistant element and press and hold the Setup fixed key >2 s.

Touch the Water temperature key. If the key is highlighted orange, the

water tempering function is activated. Use the – and + keys to set the water

temperature.

40 Dentsply Sirona

Page 41: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 41 Montag, 10. April 2017 4:25 16

SIROSONIC TLThe SIROSONIC TL scaler is used to remove plaque and for endodontic treat-ment.

Selecting the intensity with the quick setting keys

The SIROSONIC TL scaler is removed from its holder.

Touch one of the quick setting keys in the bottom line briefly (<1 s).

The quick setting key is highlighted orange. The selected intensity is displayed in percent in the first line.

Setting intermediate intensity values with quick setting keys Increase intensity: Press and hold (>1 s)

a quick setting key whose intensity value is greater than or equal to the value displayed in the first line.

Reduce intensity: Press and hold (>1 s) a quick setting key whose intensity value is less than the value displayed in the first line.

Increasing the intensity by increments of 20 (boost function)The boost function allows you to increase the intensity during treatment in incre-ments of 20 in relation to the final value. After an intensity of 80 is reached, only the maximum value of 100 can be selected.

Switching on the endodontics func-tionThe intensity of the endodontics function is limited for safety reasons, e.g. in order to prevent broken needles. The intensity can be set only from 1e to 5e.

The SIROSONIC TL scaler is removed from its holder.

The Ultrasonic screen appears on the touchscreen.

Dentsply Sirona 41

Page 42: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 42 Montag, 10. April 2017 4:25 16

Touch the Endo key. The key is highlighted orange. Instead of

the ultrasonic intensity values, the touchscreen displays the endodontic intensity values.

Please note that the endodontics intensity values of 1e to 5e do not match the values of 1 to 5 in the scaler mode.

42 Dentsply Sirona

Page 43: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 43 Montag, 10. April 2017 4:25 16

Mini L.E.D.

The status control lamp (D) signals the following states:

The mini LED has three operating modes, which you can select with the Mode H key:

Quick curing modeThe mini LED operates at full power for 10 seconds.

Pulse curing modeThe mini LED runs at full power in "Pulse" mode (1 s on, 0.25 s off).

"Soft start" modeSoft start in 10 s, then full power for 10 s

A Glare shield F Pulse curing mode lightB Fiber optics G "Soft start" mode lampC On/off button H Mode buttonD Status control lamp I HandpieceE Quick curing mode light J Supply cable

green Normal operationred, flashing Overheating protection

Dentsply Sirona 43

Page 44: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 44 Montag, 10. April 2017 4:25 16

Implantology and endodontic treatments

The treatment functions support implant and endodontic treatments.

It requires the use of a precisely control-lable brushless BL, BL ISO C, or BL Implant motor.

The basic functionality of the treatment functions is described below. For more information, please refer to your complete SINIUS Operating Instructions.

Treatment selection The Start program is displayed on the

touchscreen. Touch the Treatment key. The Treatment selection screen is

displayed.

Touch the button for the desired endodontic (left) or implantology (right) treatment.

The Start program is shown. The Treat-ment key is marked orange. The burr drive that corresponds to the treatment is displayed in the status bar with an orange dot. Please refer to the following section "Assignment of bur drives" for information on the significance of the empty or solid dots.

Take the bur drive from the instrument holder which is marked with an orange dot in the status line.

44 Dentsply Sirona

Page 45: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 45 Montag, 10. April 2017 4:25 16

Depending on the type of treatment selected, the Endodontics program or the Implantology program is displayed on the touchscreen.

In the Endodontics and Implantology treat-ment programs, the key assignment of the foot control is indicated by blue and orange cursor bars even when cursor control is switched off. The orange cursor can be moved with the 4-way foot control plate only when cursor control is switched on. For details, see "Cursor control" [24].

Assignment of bur drives

If you would like to use a different bur drive for the selected treatment type, you can change this setting.

Before selecting the treatment, touch the upper or lower Assign bur drive key.Upper key: Endodontic treatmentLower key: Implantology treatment

The next usable burr drive is marked using the orange dot. The required burr drive is assigned to endodontics or implantology treatment.

Switching the treatment function offIf the Treatment function is activated, the Implantology program or the Endodontics program will be displayed on the touch-screen instead of the Motor program when the bur drive assigned to the selected treatment is removed from its holder. In order for the Motor program to be displayed again the next time the bur drive

Endodontics program Implantology program

Empty gray dot

Instrument cannot be used for this treatment

Solid gray dot

Burr drive can be assigned to this treatment

Solid orange dot

Burr drive is assigned to this treatment

Dentsply Sirona 45

Page 46: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 46 Montag, 10. April 2017 4:25 16

is removed, the Treatment function must be switched off beforehand.

Touch the Treatment key again. If the key turns gray, the treatment func-

tion is switched off. The Motor screen is activated when a bur drive is removed from its holder.

Implantology

Calibrating the bur driveCalibration must be performed each time you start treatment and each time you change or lubricate the contra-angle hand-pieces.

Attach the contra-angle handpiece that you would like to use for the implanto-logy treatment to the electric motor.

Insert the tool in the contra-angle hand-piece. This is therefore also taken into account for measurements.

Touch the Cal key on the touchscreen. The button flashes orange. Hold down the foot pedal throughout

the duration of the calibration. The Calkey continues to flash. If the burr

drive is calibrated, the key permanently remains highlighted orange. Calibration is then completed.

Setting the speed and torque Use the – and + keys to set the speed

and torque of the tool. You can also hold down the keys for this purpose.

The selected speed is displayed in the first line in rpm (revolutions per minute). The torque is displayed in the second line in Ncm (Newton centimeters).

Setting the direction of rotation Touch the CCW Rotation key on the

touchscreen. For counterclockwise rotation: The CCW

rotation key is highlighted orange and an orange CCW arrow appears. Coun-

46 Dentsply Sirona

Page 47: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 47 Montag, 10. April 2017 4:25 16

terclockwise rotation is performed without torque limitation.For clockwise rotation: The CCW rota-tion key turns gray and the orange CCW rotation arrow disappears.

Activating/deactivating and setting the NaCl flowThe peristaltic pump can be switched on/off by touching the NaCl key. If the key is highlighted orange, the pump can be acti-vated using the foot control.

The NaCl flow rate set of the peristaltic pump is permanently displayed by a bar in the bottom line of the touchscreen in the Implantology program.

The flow rate can be set by pressing and holding the NaCl key (>2 s).

Selecting a stepIndividual settings can be made and saved for each implantology work step, e.g. predrilling, final drilling, tapping, etc. At the end of each work step, the required settings are made immediately available by selecting the next step.

Select the required implantology work step. Touch the previous step or next step key.

The selected work step is displayed on the touchscreen. The settings saved in the work step are preset.

Dentsply Sirona 47

Page 48: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 48 Montag, 10. April 2017 4:25 16

Endodontics

File selectionA list of files for this endodontic treatment is displayed in the Endodontics Program .

Use the ↑ and ↓ keys to select the file you would like to use.

The selected file is highlighted orange.

Calibrating the treatment centerCalibration must be performed each time you start treatment and each time you change or lubricate the contra-angle hand-pieces.

Attach the contra-angle handpiece that you would like to use for the endodontic treatment to the electric motor.

Insert the file in the contra-angle hand-piece. This is therefore also taken into account for measurements.

Touch the Cal key on the touchscreen. The button flashes orange. Hold down the foot pedal throughout

the duration of the calibration. The Cal key continues to flash. If the

burr drive is calibrated, the key perma-nently remains highlighted orange. Cali-bration is then completed.

Setting the direction of rotation Touch the CCW Rotation key on the

touchscreen. For counterclockwise rotation: The CCW

rotation key is highlighted orange and an orange CCW arrow appears. Coun-terclockwise rotation is performed without torque limitation.For clockwise rotation: The CCW rota-tion key turns gray and the orange CCW rotation arrow disappears.

48 Dentsply Sirona

Page 49: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 49 Montag, 10. April 2017 4:25 16

ApexLocatorThe ApexLocator can be used to measure the working length of the root canal file in endodontic treatments using electrical impedance.

The ApexLocator can be used as follows:

• For measurement during treatment using the endodontic treatment option with electronic torque limitation

Distance display in the Endodontics program

• For measurement during treatment with the motor, without electronic torque limi-tation

Distance display in the Motor program

• For manual measurement using a file clamp

Distance display in the Start sub-screen

Dentsply Sirona 49

Page 50: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 50 Montag, 10. April 2017 4:25 16

The measured root canal depth is shown in the distance display on the touchscreen. A bar with 11 display levels shows the distance from the root canal file to the physiological apex. The root canal is divided into four colored sections in the distance display.

Gray regionThe tip of the root canal file is in the middle region of the root canal.

Blue regionThe tip of the root canal file is near the apex.

Green regionThe tip of the root canal file has reached the physiological apex.

Red regionThe tip of the root canal file has pierced the physiological apex. The instrument over-shoot is displayed.

In addition to the graphical distance display on the touchscreen, the position of the file in the root canal can also be indi-cated as an acoustic signal.

50 Dentsply Sirona

Page 51: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

6322858 KGBA SINIUS EN.book Seite 51 Montag, 10. April 2017 4:25 16

Automatic motor stop and auto-reverse functionYou can set the motor so that it stops auto-matically when the physiological apex is reached. If you want to use the automatic motor stop, this can be preset in the sub-screen. The automatic motor stop can be combined with the Auto-Reverse function.

Attach a file to the handpiece. Short the electrical measurement

system. Plug in the files and hold them directly against the mucosal electrode. This cancels out any inaccuracies caused by skips in impedance in the measurement set-up (standardization).

If a constant signal is heard and the distance display appears with no bars, the standardization was successful. If not, check the electrical connections for signs of damage.

Place the mucosal electrode in the patient's mouth and perform the treat-ment. Activate the bur using the foot pedal.

The measured root canal depth is displayed by a colored bar in the distance display.

For more details, please refer to your complete SINIUS Operating Instructions.

Dentsply Sirona 51

Page 52: 6322858 KGBA SINIUS EN - Sironatd.sirona.com/pdf/6322858.pdf · SINIUS Brief operating instructions 6322858 KGBA SINIUS EN.book Seite 1 Montag, 10. April 2017 4:25 16

Dentsply SironaSirona Dental Systems GmbH Fabrikstrasse 31 64625 Bensheim Germanydentsplysirona.com

Su

bje

ct t

o t

ech

nic

al c

han

ges

an

d e

rro

rsP

rin

t n

o. D

356

1.20

2.0

1.06

.02

Ord

er n

o. 6

3 2

2 8

58

03

.20

17C

han

ge

no

. 123

221

6322858 KGBA SINIUS EN.book Seite 52 Montag, 10. April 2017 4:25 16