MICRO-OHMMETER 6240 ENGLISH User Manual

Transcript of 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies...

Page 1: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

MICRO-OHMMETER 6240

E N G L I S H User Manual

Page 2: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol
Page 3: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

Statement of Compliance

Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments certifies that this instrument has been calibrated using standards and instruments traceable to international standards.

We guarantee that at the time of shipping your instrument has met its published specifications.

An NIST traceable certificate may be requested at the time of purchase, or obtained by returning the instrument to our repair and calibration facility, for a nominal charge.

The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer. For recalibration, please use our calibration services. Refer to our repair and calibration section at www.aemc.com.

Serial #: ________________________________

Catalog #: 2129.80

Model #: 6240

Please fill in the appropriate date as indicated:

Date Received: _________________________________

Date Calibration Due: _______________________

Chauvin Arnoux®, Inc.d.b.a AEMC® Instrumentswww.aemc.com

Page 4: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

Table of Contents

1. INTRODUCTION ................................................................................. 41.1 International Electrical Symbols ................................................51.2 DefinitionofMeasurementCategories .....................................51.3 ReceivingYourShipment ..........................................................51.4 OrderingInformation .................................................................6

1.4.1 AccessoriesandReplacementParts ............................6

2. PRODUCT FEATURES ......................................................................... 72.1 Description ................................................................................72.2 Applications ..............................................................................82.3 KeyFeatures ............................................................................82.4 ControlFeatures .......................................................................92.5 ButtonFunctions .....................................................................102.6 DisplaySymbols .....................................................................11

3. SPECIFICATIONS ............................................................................. 123.1 Electrical .................................................................................123.2 Mechanical ..............................................................................133.3 Environmental .........................................................................143.4 Safety ......................................................................................14

4. OPERATION ..................................................................................... 154.1 PrecautionsBeforeUse ..........................................................154.2 InstrumentConfiguration(SET-UPmode) ..............................154.3 ResistanceMeasurements .....................................................16

4.3.1 MeasurementReadings ..............................................174.3.2 ConnectionsandReadings .........................................174.3.3 Test Lead Connection .................................................18

4.4 RepetitiveMeasurements .......................................................184.5 MeasuringVeryLowResistance ............................................194.6 AutomaticRecording ..............................................................204.7 StoringResultsintoMemory ...................................................21

Page 5: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

Micro-Ohmmeter Model 6240 3

4.8 RecallingResultsfromMemory ..............................................224.9 ErasingMeasurementsfromMemory .....................................23

4.9.1 ErasingSelectedMeasurements ................................234.9.2 ErasingAllMeasurements ..........................................23

5. DATAVIEW® SOFTWARE ................................................................... 245.1 InstallingDataView® ................................................................245.2 ConnectingtheInstrumenttoyourComputer .........................285.3 OpeningtheMicro-ohmmeterControlPanel ..........................295.4 ConfiguringtheInstrument .....................................................305.5 DownloadingStoredTests ......................................................315.6 Saving/ExportingDatatoaSpreadsheetorPDFFile ............32

5.6.1 SavingDatatoSpreadsheetfromControlPanel ........325.6.2 ExportingDatatoaPDFfromDataView® ...................32

6. TROUBLESHOOTING ........................................................................ 336.1 ErrorMessages ......................................................................33

6.1.1 VoltagePresent ..........................................................336.1.2 Overrange ...................................................................346.1.3 NoisyMeasurement ....................................................346.1.4 Overheating ................................................................34

7. MAINTENANCE ................................................................................ 357.1 Warning..................................................................................357.2 Cleaning ..................................................................................357.3 Charging/RechargingtheBattery ...........................................367.4 BatteryandFuseReplacement ..............................................37RepairandCalibration ....................................................................38TechnicalandSalesAssistance ......................................................38Limited Warranty .............................................................................39WarrantyRepairs ............................................................................39

Page 6: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

4 Micro-Ohmmeter Model 6240

CHAPTER 1

INTRODUCTION

WARNING These safety warnings are provided to ensure the safety ofpersonnelandproperoperationoftheinstrument.• Thisinstrumentisprotectedfromaccidentalvoltagesofnot

more than50Vwith respect toearth inmeasurementCat-egoryIII.

• Donotattempttoperformanytestswiththisinstrumentuntilyouhavereadtheusermanual.

• Tests are to be carried out on de-energized circuits only!Never connect the unit to a live circuit.

• Make sure that none of the input terminals are connectedandthattheswitchissettoOFFbeforepluggingintheACpowertorechargethebatteryoftheinstrument.

• Makesuretheinternalbatteryisfullychargedpriortotesting.If the instrumenthasbeen leftunused forseveralmonths,rechargethebattery.

• If the case needs cleaning, do not use any alcohol or oilbased cleaners. Preferably use soapy water with a dampclothorsponge.Donotimmersethe6240micro-ohmmeterinwater.

• Use connection accessories of which the overvoltage cat-egoryand theservicevoltagearegreater thanorequal tothoseof themeasuring instrument (50VCAT III).Useonlyaccessoriesthatcomplywithsafetystandards(IEC61010-2-031).

• The test leads and measuring wires must be in goodconditionandshouldbereplacedifthereisanyevidenceofdeterioration(insulationsplit,burnt,etc.).

• Troubleshooting and metrological verification proceduresmustonlybeperformedbyqualified,approvedpersonnel,orthefactory.

Page 7: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

Micro-Ohmmeter Model 6240 5

1.1 International Electrical Symbols

Signifies that the instrument is protected by double or reinforced insulation.

This symbol on the instrument indicates a WARNING and that the operator must refer to the user manual for instructions before operating the instrument. In this manual, the symbol preceding instructions indicates that if the instructions are not followed, bodily injury, installation/sample and/or product damage may result.

Risk of electric shock. The voltage at the parts marked with this symbol may be dangerous.

When disposed of, this product must be sorted for the recycling of electrical and electronic equipment in accordance with WEEE directive 2002/96/EC.

1.2 Definition of Measurement CategoriesCAT II: Formeasurements performed on circuits directly connected to

theelectricaldistributionsystem.Examplesaremeasurementsonhouseholdappliancesorportabletools.

CAT III: For measurements performed in the building installation atthe distribution level such as on hardwired equipment in fixedinstallationandcircuitbreakers.

CAT IV: For measurements performed at the primary electrical supply(<1000V) such as on primary overcurrent protection devices,ripplecontrolunits,ormeters.

1.3 Receiving Your ShipmentUponreceivingyourshipment,makesurethatthecontentsareconsistentwiththepackinglist.Notifyyourdistributorofanymissingitems.Iftheequip-mentappearstobedamaged,fileaclaimimmediatelywiththecarrierandnotifyyourdistributoratonce,givingadetaileddescriptionofanydamage.Savethedamagedpackingcontainertosubstantiateyourclaim.

NOTE: Charge the instrument fully before use.

Page 8: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

6 Micro-Ohmmeter Model 6240

1.4 Ordering Information

Micro-ohmmeter Model 6240 ............................................Cat. #2129.80Includes extra large tool bag, set of two 10ft Kelvin clips (10A - Hippo), optical USB cable, US 115V power cord, one pad of measurement results forms, two spare fuses (12.5A), NiMH rechargeable battery pack, and USB stick supplied with product user manual and DataView® software.

1.4.1 Accessories and Replacement Parts

Kelvinclips(10A-Hippo),10ftcolor-codedleadswithspadelugterminations .................................................Cat. #1017.84

Kelvinclips(10A-Hippo),20ftcolor-codedleadswithspadelugterminations ................................................. Cat. #2118.70

KelvinProbes(1A-SpringLoaded),10ftwith4mmbananaplugtermination(includessetof5forkterminals) ............... Cat. #2118.73

KelvinProbes(1A-SpringLoaded),20ftwith4mmbananaplugtermination(includessetof5forkterminals) ............... Cat. #2118.74

KelvinProbesPistolGrip10ft(10A)SpringLoaded........... Cat. #2118.75

KelvinProbes10ft(10A)SpringLoaded ........................... Cat. #2118.77

KelvinClips10ft(1-10A) ..................................................... Cat. #2118.79

KelvinClips20ft(1-10A) ..................................................... Cat. #2118.80

Thermo-HygrometerModelCA846 .....................................Cat. #2121.24

OpticalUSBCable...............................................................Cat. #2135.41

ReplacementBattery6.5V,8.5AH .......................................Cat. #2129.91

Fuse,setof5,12.5A/500V,6.3x32mm .............................Cat. #2129.92

ExtraLargeClassicToolBag ..............................................Cat. #2133.73

Inverter–12VDCto120VAC200Wattforvehicleuse .........Cat. #2135.43

MeasurementResultForms,replacementsetoftwo ..........Cat. #2129.94

115VPowerCordUS ..........................................................Cat. #5000.14

Order Accessories and Replacement Parts Directly OnlineCheck our Storefront at www.aemc.com/store for availability

Page 9: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

Micro-Ohmmeter Model 6240 7

CHAPTER 2

PRODUCT FEATURES

2.1 DescriptionThe Micro-ohmmeter Model 6240 is used to perform low resistancemeasurementsfrom5μΩ to 400Ω.Therearesixrangeswithtestcurrentsfrom10mAto10A.

Thefrontendofthemicro-ohmmeteremploysafour-wireKelvinconfigu-ration,whicheliminatestestleadresistanceforameasurementaccuracyof0.25%.Abuilt-incircuitfiltersoutACsignals.

TheMicro-ohmmeterModel6240ispackagedinasealedfieldcasewellsuitedforshopandfielduse.Powerissuppliedbyalong-lifeNiMHbatterypackwithabuilt-inrecharger(110/220V).

Thelarge,easy-to-readliquidcrystaldisplayis2.25x4.00".Itdisplaysthevalueofresistance,currentorvoltagetest,polarityandbatterycharge.

Foroperatorsafetyandinstrumentprotection,themicro-ohmmeterisfuseprotectedattheinputs.Onefuse,accessiblefromthefrontpanel,protectsagainststoredenergyininductiveloads.

Enhanced internal circuitry protects against possible inductive kickbackwhenthecurrentisshutoff.

Abuilt-inthermalswitchprotectsthemicro-ohmmeteragainstoverheatingonthe10Arangewhenincontinuoususe.

Page 10: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

8 Micro-Ohmmeter Model 6240

2.2 ApplicationsSomeofthemorepopularandmostfrequentusesofthemicro-ohmmeterareinapplicationsfor:

• Checkingmetalliccoatingresistance,especiallyinaeronautics• Groundconnectionsandcontinuitymeasurement• Resistancemeasurementsonmotorsandsmalltransformers• Contactresistancemeasurementsonbreakersandswitches• Componentmeasurement• Electricalcableresistancemeasurement• Mechanicalbondtests• Wire to terminal connections• Aircraftandrailbonds

2.3 Key Features• Measuresfrom5µΩ to 400Ω• Testcurrentselectionfrom10mAto10A• Manualtemperaturecompensation(withDataView®software)• Frontpanelpolarityreversefunction• Storesupto99testresults• Operatorsafetybyautomaticdischargeofresidualchargeonthe

equipmentundertest• Instantaneous,continuousormultipletestoperation• Autostoreofmultipletestresults• Internal,rechargeablebatteriesconductupto850,10Atests• Abuilt-inbatterypackrechargerrechargesthebatteriesby

connectingtotheACline(110V-230V,50/60Hz)usingastandardline cord

• 4-Wiremeasurementwithautomaticcompensationofundesirablevoltagesandleadresistance

• Largemulti-functionbacklitdisplay• Directdisplayofthemeasurementwithunits,rangeandtest

current• Rugged,sealedcase

Page 11: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

Micro-Ohmmeter Model 6240 9

2.4 Control Features

°C°FmVmA

°

><

VµmΩ

AVG

OBJ. TEST

MEMMRCLRP

HOLD

COM

DCAC

2nd

I+I_

1

7

8

6

42 3 5

Figure 2-1

1. Kelvininputterminals

2. AClinerechargingreceptacle

3. Outputfuse-12.5A,500V,6x32mm

4. Largemulti-linebacklitliquidcrystaldisplay(see§2.6)

5. Opticalinterfaceconnectorforconnectiontoacomputer

6. Rangeselectionswitch

7. Start/Stopbutton

8. Program/functionbuttons(see§2.5)

Page 12: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

10 Micro-Ohmmeter Model 6240

2.5 Button Functions

MEM Stores the measurement at an address identified by an object number (OBJ) and a test number (TEST).Two presses on this button are required, one to select the location (use theandbuttons to change the location) and another to store the measurement.

MR Retrieves stored data except for the OFF and SET-UP positions (this function is independent of the selector setting of the switch). Data is viewed using theandbuttons.

± I Reverses the direction of the current and displays the average.

DISPLAY Displays the current or voltage on the terminals.

AUTO>2s Activates automatic recordings.

Used in the SET-UP and memory mode, selects a function or incre-ments a flashing parameter.

Used in the SET-UP and memory mode, accesses the function to be modified.

CLR Erases the memory

Page 13: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

Micro-Ohmmeter Model 6240 11

2.6 Display SymbolsThedisplayincorporatestwolinesofcharacterstodisplaytestresults,aswellasalibraryofsymbolstoassisttheoperatorindeterminingconditionsataglance.ThesymbolsthatcanappearareshowninFigure2-2.

°C°FmVmA

°

><

VµmΩ

AVG

OBJ. TEST

MEMMRCLRP

HOLD

COM

DCAC

2nd

I+I_

Figure 2-2

Indicates the measurement is noisy; accuracy is not guaranteed

Indicates the battery charge condition; the segments represent energy

Indicates that power down is deactivated

Not used

MEM Displayed measurement about to be stored in memory

MR Memory recall

CLR Indicates the memory is being erased

Indicates internal overheating

+ I Indicates the direction of current

- I Indicates the direction of current

Indicates that a measurement has been stopped

OBJ. First position locator for data stored in memory

TEST Second position locator for data stored in memory

°C / °F Not used

Memory utilization indicator

AVG Displays average

Not used

> Indicates an overrange

R(+I)+R(-I)2

Page 14: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

12 Micro-Ohmmeter Model 6240

CHAPTER 3

SPECIFICATIONS

3.1 ElectricalSpecifications are given for an ambient temperature of 23°C ± 3°C, rela-tive humidity of 45 to 55%, supply voltage of 6V ± 0.2V and magnetic field of <40A/m.

RESISTANCE

Measurement Range

5.0 to 3999µΩ

4.0 to39.99mΩ

40.0 to 399.9mΩ

400 to 3999mΩ

4.0 to39.99Ω

40.0 to399.9Ω

Resolution 1µΩ 10µΩ 100µΩ 1mΩ 10mΩ 100mΩ

Accuracy ± 0.25% ± 2ct

Measurement current

10.2A ± 2% (1)

1.02A ± 2%

102mA ± 2%

10.2mA ± 2% (2)

Open voltage 4 to 6V

Inductance 0.5H Max

(1) With nominal value of 10.2A, the measurement current is at least 10A whatever the charge condition of the battery.(2) The current is 10mA only up to 300Ω. If the battery is low, it can fall to as low as 8mA.

VOLTAGE MEASUREMENT INDICATION

Measurement Range

0.01 to3.999mV

4.0 to39.99mV

40.0 to399.9mV

0.40 to3.999V

4.0 to4.70V

Resolution 1µV 10µV 100µV 1mV 10mV

CURRENT MEASUREMENT INDICATION

Measurement Range

5.0 to 39.99mA

40.0 to399.9mA

0.40 to3.999A

4.0 to 11.00A

Resolution 10µA 100µA 1mA 10mA

Page 15: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

Micro-Ohmmeter Model 6240 13

Influences on the Resistance Measurement

Quantities of Influence Range of useVariation of the measurement

Typical Maximum

Temperature -10 to +55°C 0.1% /10°C 0.5% /10 °C + 2cts

Relative Humidity 10 to 85% RH @ 45°C 0.1 % 0.5% + 2cts

Supply Voltage 5 to 7V 2cts 0.2% / V + 2cts

Series Mode Rejection, 50/60Hz*

U (AC) = (R measured x I measurement)

< 0.2% 2% + 1ct

Common Mode Rejection, 50/60Hz AC

0 to 50 V AC > 80 dB > 60 dB

*Example: If the measured resistance is 1mΩ and the measurement current is 10A, an alternating voltage of 1mV RMS in series with the resistance to be measured will induce an error of not more than 2%.

Power Supply:6V,8.5AhrechargeableNiMHbatterypackBattery Life: Rangedependent

Range Number of measurements*

10A 850

1A 3500

100mA 4500

10mA 5000

In Standby or Off battery life 4 to 6 months

*Established for measurements lasting 5s, every 25s.

3.2 Mechanical

Dimensions: 10.70x9.76x7.17"(272x248x182mm)

Weight: 10lbs(4.5kgapprox)

Case Protection: ABSplasticpolycarbonate:watertighttoIP54(cover closed),waterresistanttoIP53(coveropen).

Color: Safetyyellowcasewithgrayfaceplate

Page 16: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

14 Micro-Ohmmeter Model 6240

3.3 Environmental

Operating Temperature:14°to132°F(-10°to55°C),10to85%RH

Storage Temperature:-40°to158°F(-40°to70°C);10to90%RH

Altitude:<2000m

NOTE: For long-term storage (>1 year) with the battery installed, the stor-age range is -4 to 86°F (-20 to 30°C); 85% RH, otherwise the battery will deteriorate. For short-term storage (1 month) the temperature can reach up to 122°F (50°C).

3.4 SafetyElectricalsafetyasperEN61010-1(Ed.2of2001),EN61557(Ed.97)parts1and4.Degreeofpollution:2Protection:MeasurementCATIII,50Vwithrespecttoearth,500Vdiffer-entialbetweenterminals,and300VCATIIonthechargerinput

Electromagnetic Compatibility:TheinstrumentsatisfiestheCEMandDBTdirectivesrequiredfortheCEmarkingandproductstandardEN61326-1(Ed.97)+A1(Ed.98)Emissions in residential environmentImmunityinindustrialenvironment

*Specifications are subject to change without notice

Page 17: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

Micro-Ohmmeter Model 6240 15

CHAPTER 4

OPERATION

NOTE: Charge the instrument fully before use.

4.1 Precautions Before Use • Never use test leadsormeasuringwires if there isanyevidenceof

deterioration(insulationsplit,burnt,etc).• Never exceedthesafetyvaluesindicatedinthespecifications.• Never connecttheunittoalivecircuit.• Never disconnecttheconnectionwiresbeforethe icondisappears

fromthedisplay.

4.2 Instrument Configuration (SET-UP mode)TheSET-UPmodeisusedtomodifytheinstrument’sconfiguration.ThiscanalsobeperformedbyusingtheDataView® softwarethatcamewiththeinstrument(see§5).ToconfiguretheModel6240directlyfromthedisplay,turntherotaryswitchtotheSET-UPposition.Thefollowingscreenwillappear:

MEM

• The display order of the parameters that can be modified is:- Clear all memory (see§4.9.2)- Time- Date- AutomaticPowerOff- DisplayoftheInternalParameters(e.g.serialnumber,software

version,dateoflastcalibration,lightingofallsegmentsondisplay)

Page 18: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

16 Micro-Ohmmeter Model 6240

To display and modify parameters:• Tomovethroughparameters,pressthe button.• Tochange thedisplayedparametersorviewadditionalones,press

andholdthe button.

• Tomodifytheparameters,usethe button.

• Toaccept thechanges,pressandhold the button>2suntil thedisplaystopsblinking.

4.3 Resistance Measurements

WARNING: Before performing the resistance test, verify that the sample under test is not energized.

1. Cleanallsurfacesbeforeconnectingtestleads.2. Connectthetwoleadstothefourterminals,thenthetwoKelvinclips

totheobjecttobetested.3. Settherangeselectorswitchtothedesiredrangeforthetest.Ifthe

anticipated resistance is not known, begin with the highest range(400Ω) and successively lower the range selection until adequateresolutionisachieved.

4. PresstheSTART/STOPbuttontostartthemeasurement.

NOTE: The START/STOP needs to be pressed each time the range is changed. The range selection may be changed while the instrument is on.

HOLD

5. Press theSTART/STOP button again to stop themeasurement ordisconnect one of theKelvin clips.The lastmeasurementmade isdisplayedalongwiththe symbol.

Page 19: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

Micro-Ohmmeter Model 6240 17

NOTE: If the measurement was stopped by disconnecting a Kelvin clip, simply connect it to another object to start the next measurement; It is not necessary to press the START/STOP button.

6. Todisplaythevoltageontheterminalsoftheresistanceinsteadofthemeasurementcurrent,presstheDISPLAYkey.

Current results are displayedbefore pressing DISPLAY button

Voltage results are displayedafter pressing DISPLAY button

4.3.1 Measurement ReadingsWhentestingresistivesamples,themeterreadingwillstabilizewithinthefirstfewhundredmilliseconds.Oninductiveloads(e.g.small transform-ers),themeasurementreadingmaytakefromseveralsecondstoafewminutestostabilizeanddependsgreatlyonthetypeofequipmentandtheimpedanceoftheequipmentundertest.

4.3.2 Connections and ReadingsTheModel6240generatesacurrent (I)from the internal voltage source (V). Avoltmeter measures the voltage dropVxat theKelvinprobecontactpoints tothe resistance tobemeasured (Rx)anddisplays the resistance measurement(Rx)directlyusingtheformulaRx=Vx/I.

The result is not affected by the otherresistances encountered in the currentloop (Ri, Rf, Rc), as long as the totalvoltagedropinducedacrossRx remains belowthevoltagesuppliedbythesourcewhichisbetween5to6V.Themaximumadmissible lead resistance level is Rf≈(V-Vx)/I.TheuseofKelvinprobeshelps,astheyeliminatetheeffectoftheleadresistance(Rf).

Ri Rf

Rf

Rf

RfC2

P2

P2

C1

Rc

Rx

Rc

V

IVx

+

Ri=UnitinternalresistanceRf = Lead resistanceRc = Contact resistanceRx=Resistancetobemeasured

Page 20: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

18 Micro-Ohmmeter Model 6240

4.3.3 Test Lead ConnectionThemeasurementleadsareconnectedusingthefourbindingpostsontheleftsideofthefrontpanelasshownbelow.ConnectthetworedleadstoterminalsC1andP1.ConnectthetwoblackleadstoterminalsC2andP2.Anydropinthevoltageontheloadterminalsismeasuredbetweenthetwo“voltage”(V)leads,P1andP2.Thecurrentleads(C1andC2)candelivercurrentfrom10mAto10A.

UI

4.4 Repetitive Measurements1. ConnecttheKelvinclipstothefirstobject.PressSTART/STOP.

2. Disconnecttheclips.Themeasurementstopsandresultisdisplayed.

3. Connect the clips to the second object. Themeasurement restartsautomatically.Repeatasnecessary.

4. Afterthelastmeasurement,pressSTART/STOPagain.

NOTE: This operation is valid only if all of the objects to be measured have substantially the same value and all of the measurements are made in the same range.

Page 21: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

Micro-Ohmmeter Model 6240 19

4.5 Measuring Very Low ResistanceWhenmeasuringverylowresistivevaluesintheµΩrange,thepresenceofstrayDCcurrentsmayaffecttheaccuracyofthemeasurements.ThesecurrentscanbepresentduetoavarietyofreasonsincludingchemicalorthermalEMFinsamplesmadeofdissimilarmetals.

Youcaneliminatetheeffectsofthisbyreversingthedirectionofcurrentflow(seeexamplebelow)andusingtheaverageofforwardandreversereadings.

ThepresenceofACinterferenceinthesampleundertestmaycausethemeasuredvalueonthedisplaytofluctuate.Thisinterferencemaybecomemorenoticeableinthepresenceofstrongelectricfields.Theeffectsofthisinterferencemaybereducedbytwistingtheleadsaroundeachother.

1. Reversethedirectionofthecurrentbypressingthe±Ibuttonandtheinstrumentdisplaystheaverage:

R(+I)+R(-I)

RAVG = 2

2. TodisplaythevaluesR(+I)andR(-I),presstheDISPLAYbutton.

A

µ Ω

AVG

HOLD

I_+I

R(+I)+V

R(+I)-I

R(-I)-V

R(-I)

Press the DISPLAY button to cycle through the values

Page 22: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

20 Micro-Ohmmeter Model 6240

4.6 Automatic Recording1. Beforemakingmeasurements,presstheAUTO >2sbutton.

2. Whenthebuttonispressed,theMEMsymbolflashesandthesmalldisplayindicatesthefirstfreeOBJ : TESTnumber(e.g.01:01).Themaindisplayindicates .

3. Usethe buttontoswitchbetweentheOBJ and TESToption,thenusethe buttontochangethenumberofthetestorobject.

4. Whenfinished,presstheMEMbuttonfor>2s.Automaticrecordingisactivated(theMEMsymbolflashes).

OBJ. TEST

+I Use the arrow buttonsto change the valuesof the OBJ and TEST

When finished, pressthe MEM button for >2s

> 2s

MEM

5. PresstheSTART/STOPbuttontobeginmeasuring. Ateachnewmeasurement,thetestnumberisincrementedandthe

measurementisrecorded.

6. PresstheSTART/STOPbuttonagaintostopautomaticrecording.

µ Ω

OBJ. TEST

MEMP

HOLD

+I

Page 23: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

Micro-Ohmmeter Model 6240 21

4.7 Storing Results into MemoryDatastorageisorganizedintoobjects(OBJ),eachofwhichcancontainseveraltests(TEST).OBJcorrespondstotheobjecttestedandeachtestcorrespondstoameasurementmadeontheobject.Theinstrumentcanstore99measurements.

1. StartandStopameasurement(see§4.3).

2. PresstheMEMbutton.

3. Whenthebuttonispressed,theMEMsymbolflashesandthesmalldisplayindicatesthefirstfreeOBJ : TESTnumber(e.g.01:01).Themaindisplayindicates .

4. Usethe buttontoswitchbetweentheOBJ and TESToption,thenusethe buttontochangethenumberofthetestorobject.

5. Whenfinished,presstheMEMbuttonfor>2s.Themeasurement isstored.

OBJ. TEST

MEM

+I Use the arrow buttonsto change the valuesof the OBJ and TEST

When finished, pressthe MEM button for >2s

> 2s

6. Iftheuserselectsamemoryaddressthatisalreadyoccupied, appearsonthemainscreen.Tooverwritethelocation,presstheMEM buttonfor>2s.

Page 24: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

22 Micro-Ohmmeter Model 6240

OBJ. TEST

MEM

+I

either

> 2s

or

Indicates address is already occupied Overwrites the Exits without Memory Recording

Displayshownifmemoryis full: DisplayshownIfmemoryis empty:

OBJ. TEST

MEMP

2nd

+I

P

+I

MR

OBJ. TEST

4.8 Recalling Results from Memory

1. Makesurethereisnomeasurementinprogress.

2. PresstheMRbutton.

3. Usethe and buttonstoselectthedesiredOBJandTEST.

4. PresstheMRbuttonagaintoexitthememoryfunction.

Use the arrow buttonsto change the valuesof the OBJ and TEST

Page 25: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

Micro-Ohmmeter Model 6240 23

4.9 Erasing Measurements from Memory

4.9.1 Erasing Selected Measurements

1. PresstheCLRbutton.

2. Usethe and buttonstoselectthedesiredOBJandTEST.

3. PresstheCLRbuttonfor>2stodeletetheselectedrecord.

OBJ. TEST

MEM

+IUse the arrow buttons to selectthe desired OBJ and TEST

When finished, pressthe CLR button for >2s

> 2s µ Ω

OBJ. TEST

+I

CLR

4.9.2 Erasing All Measurements

1. TurntherotaryswitchtoSet-up.Thedefaultdisplayappears:

MEM

2. Press for>2s. willblink.

3. Press again. willdisplay.

4. Press againfor>2s.Allmeasurementsareerasedandthescreenreturnstothedefaultdisplay.

Page 26: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

24 Micro-Ohmmeter Model 6240

CHAPTER 5

DATAVIEW® SOFTWARE

5.1 Installing DataView®

DO NOT CONNECT THE INSTRUMENT TO THE PC BEFORE INSTALLING THE SOFTWARE AND DRIVERS.

NOTE:When installing, theusermusthaveAdministrativeaccessrightsduringtheinstallation.Theusersaccessrightscanbechangedaftertheinstallationiscomplete.DataView®mustbereinstalledforeachuserinamulti-usersystem.

USB Flash Drive Install

1. InserttheUSBstickintoanavailableUSBport(waitfordrivertobeinstalled).

2. If Autorun is enabled then anAutoPlay window should appear asshown.

Page 27: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

Micro-Ohmmeter Model 6240 25

NOTE: If Autorun is disabled, it will be necessary to open Windows Explorer, then locate and open the USB stick drive labeled “DataView” to view the files on the drive.

3. IntheAutoPlaywindow,selectOpen Folder to view Files.

4. Double-clickonSetup.exefromtheopenedfolderviewtolaunchtheDataViewsetupprogram.

NOTE: If installing onto a Vista based computer the User Account Control dialog box will be displayed. Select the Allow option to proceed.

5. A Set-up window,similartotheonebelow,willappear.

Figure 5-1

Page 28: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

26 Micro-Ohmmeter Model 6240

Thereareseveraldifferentoptionstochoosefrom.Someoptions(*)requirean internet connection.

• DataView, Version x.xx.xxxx-InstallsDataView®ontothePC.

• *Adobe Reader -Links to theAdobe®website todownload themostrecentversionofAdobe®Readertothecomputer.Adobe® Reader is required for viewing PDF documents supplied with DataView®.

• *DataView Updates - Links to the online DataView® softwareupdatestocheckfornewsoftwareversionreleases.

• *Firmware Upgrades -Links to theonlinefirmwareupdates tocheckfornewfirmwareversionreleases.

• Documents - Shows a list of instrument related documentsthatyoucanview. Adobe®Reader is required forviewingPDFdocumentssuppliedwithDataView®.

6. DataView, Version x.xx.xxxxoptionshouldbeselectedbydefault.SelectthedesiredlanguageandthenclickonInstall.

7. TheInstallation Wizardwindowwillappear.ClickNext.

8. Toproceed,acceptthetermsofthelicenseagreementandclickNext.

9. IntheCustomer Informationwindow,enteraNameandCompany,thenclickNext.

10. IntheSetup Typewindowthatappears,selectthe“Complete” radio buttonoption,thenclickNext.

11. IntheSelect Featureswindowthatappears,selecttheinstrument’scontrolpanelthatyouwanttoinstall,thenclickNext.

NOTE: The PDF-XChange option must be selected to be able to generate PDF reports from within DataView®.

Page 29: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

Micro-Ohmmeter Model 6240 27

Figure 5-2

12. IntheReady to Install the Programwindow,clickonInstall.

13. If the instrumentselected for installation requires theuseofaUSBport,awarningboxwillappear,similartoFigure5-3.ClickOK.

Figure 5-3

NOTE: The installation of the drivers may take a few moments. Windows may even indicate that it is not responding, however it is running. Please wait for it to finish.

14.Whenthedriversarefinishedinstalling,theInstallation Successful dialogboxwillappear.ClickonOK.

15.Next,theInstallation Wizard Completewindowwillappear.ClickonFinish.

16. A Questiondialogboxappearsnext.ClickYestoreadtheprocedureforconnectingtheinstrumenttotheUSBportonthecomputer.

Page 30: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

28 Micro-Ohmmeter Model 6240

NOTE: The Set-up window remains open. You may now select another option to download (e.g. Adobe® Reader), or close the window.

17. Restartyourcomputer,thenconnecttheinstrumenttotheUSBportonthecomputer.

18.Onceconnected,theFound New Hardwaredialogboxwillappear.Windowswillcompletethedriverinstallationprocessautomatically.

ShortcutsforDataView®andeachinstrumentcontrolpanelselectedduringtheinstallationprocesshavebeenaddedtoyourdesktop.

NOTE: If you connected your instrument to the computer before installing the software and drivers, you may need to use the Add/Remove Hard-ware utility to remove the instrument driver before repeating the process.

5.2 Connecting the Instrument to your ComputerTheModel 6240 is suppliedwith aUSB interface cable (Cat. #2135.41) necessary for connecting the instrument to the computer.This cable is equippedwithatype-AUSBconnectorononeendandanopticalconnec-torontheotherend,thusprovidingtheabilitytointerfacewithcomputersthathaveUSBconnectors.

To connect the instrument to your computer:• Connect the optical connector end of the cable to the optical

communicationsportonthefrontpaneloftheModel6240.• Connectthetype-AUSBendofthecabletoanavailableUSBport

onyourcomputer.Thefirsttimethiscableisconnectedtothecomputer,Windowswillinstallthedriversforit.Oncethishasfinished,thesystemisreadytocommuni-cateusingtheDataView®softwarewiththemicro-ohmmeter.

Page 31: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

Micro-Ohmmeter Model 6240 29

5.3 Opening the Micro-ohmmeter Control Panel

Toopenthecontrolpanel:

1. Double-clicktheMicro-ohmmeter Icon thatwascreatedduringinstallation,locatedonthedesktop.TheMicro-ohmmeterControlPanelwillopen.

Figure 5-4

2. To establish a communication link with the instrument, go toInstrument > Connectinthemainmenu.

3. TheCommunicationdialogboxwill appear.Makesure that thecommunicationportdisplayedinthedialogboxmatchestheportthattheserialcableispluggedinto.

Figure 5-5

4. Oncethepropercommunicationparametershavebeenspeci-fied,clickOK.

Page 32: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

30 Micro-Ohmmeter Model 6240

5.4 Configuring the InstrumentToconfiguretheinstrumentperformthefollowingsteps.

1. Double-clicktheMicro-ohmmetericononyourdesktop.

2. Go to Instrument > Configure from the main menu of theMicro-ohmmeterControlPanel.

NOTE: Ifaconnectionhasnotbeenpreviouslyestablished,theCommunicationdialogboxwillappearallowingyoutoselecttheCommunicationPortandInstrumentModel.

3. Once identification is complete, the Configuration dialog boxwillappearon thescreen,allowingyou toconfigure theDateFormatandAutoPowerOFFoptions.

Figure 5-6

Thissetupdialogboxalsocontainsfivecommandbuttons:• Set Clock:Programsthecomputer’stimeanddateintothecon-

figurationoftheinstrument.• Delete Tests: Displaysadialogboxwhichallowsthedeletionof

selected tests.• Write to Instrument: Programs themicro-ohmmeter using the

currentsettings.• Read from Instrument: Reads thecurrent configurationof the

micro-ohmmeterattachedviathecommunicationscable.• Cancel: Closes theConfigurationdialogboxandbringsup the

ControlPanel.

Onceallofthefieldshavebeenconfigured,clickontheWrite to Instrument buttontoconfiguretheinstrumentandclosetheConfigurationdialogbox.

Page 33: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

Micro-Ohmmeter Model 6240 31

TIP: While connected, both the DataView® and Micro-ohmmeter windows may be open. They have similar main menu options. To determine which window is active, look at the top left corner of the window. It will either display “DataView” or “Micro-ohmmeter”. Use the Windows taskbar at the bottom of your screen to access the appropriate window.

5.5 Downloading Stored TestsTheDownloadcommandinitiatesthetransferoftestmeasurementsstoredintheinstrument.Todownloadstoredtestsperformthefollowingsteps:

1. Select the download command (either from the Instrument menuortheDownloadbutton).

2. ThesoftwarewillbuildalistoftestsstoredintheinstrumentandtheSelect Testsdialogboxwillappear,allowingyoutoselectwhichteststodownload.

Figure 5-7

3. HighlightthedesiredtestsusingtheShiftorCtrlkeysandselecttheDownloadbutton.

NOTE: For additional help on using the software, use DataView’s “Help” Menu, which is located on the menu bar.

Page 34: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

32 Micro-Ohmmeter Model 6240

5.6 Saving/Exporting Data to a Spreadsheet or PDF FileMeasurementvaluesstoredinadatabasecanbesavedtoaspreadsheet(.csvfile,whichcanbeopenedinMicrosoft®Excel)orexportedtoaPDFfile(whichcanbeopenedusingfreeAdobe®Readersoftware).

NOTE: Exporting to a PDF file can only be performed from within DataView itself, not the Micro-ohmmeter Control Panel.

5.6.1 Saving Data to a Spreadsheet from the Control Panel

To save data to a spreadsheet:1. Withadatabaseopen,gotoFile > Save As.2. In theSaveAsdialogbox thatappears, choosea location to

savethefilefromthe“Savein”drop-downmenu.3. Select CSV (Comma delimited)fromthe“Saveastype”drop-

downmenu,thenclickSave.

NOTE: To be able to export the data to a PDF from the Micro-ohmmeter Control Panel, first go to File > Create Dataview Report, save as a .dvb, then open the saved .dvb file from within DataView and perform the steps below.

5.6.2 Exporting Data to a PDF from DataView®

To export data to a PDF file:1. Withadatabaseopen,gotoFile > Generate PDF.2. In the Print dialog box that appears, make sure that PDF-

XChange 3.0isselectedfromthedrop-downmenu,thenclickOK.

3. WhenthePDFiscomplete,theSaveAsdialogboxwillappear.ChoosealocationtosaveittoandclickSave.

NOTE:DatacanalsobeexportedtoaspreadsheetfromwithinDataViewbygoingtoFile>ExporttoSpreadsheet.

Page 35: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

Micro-Ohmmeter Model 6240 33

CHAPTER 6

TROUBLESHOOTING

6.1 Error Messages

6.1.1 Voltage Present

Anerrormessagewillappearwhenanexternalvoltageispresentonthedevicebeingmeasured.(e.g.C1C2orC1P1).Beforeameasurementispossiblethevoltagemustberemovedfromthetestobject.

V

P

DC

+I or

2C

2P

1P

1C

> 7 V

Avoltagegreaterthan20VappliedacrosstheC1-C2terminalwillcausetheFuseonthefrontpaneltoblow.(See§7.4forfusereplacement)

A

Ω

HOLD

+I

2C

2P

1P

1C

> 20 V

Page 36: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

34 Micro-Ohmmeter Model 6240

6.1.2 Overrange

The>symbolindicatesanoverrangecondition.Switchtoahighermea-surementrangeandrestartthemeasurementuntilthe>symbolnolongerappears.

A

> µ Ω

+I

This symbolindicates an

overrange

6.1.3 Noisy Measurement

The symbol indicatesnoiseinthemeasurement.Themeasurementaccuracyisnotguaranteed.

A

µ Ω

+I

This symbolindicates noisymeasurement

6.1.4 Overheating

Internal overheating canoccur if ameasurement in the10A range lastfor severalminutes.Nomeasurement is possible until the temperaturesymbol and indicationarenolongervisible.

° +I

Page 37: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

Micro-Ohmmeter Model 6240 35

CHAPTER 7

MAINTENANCE

Use only factory specified replacement parts.AEMC® will not be heldresponsible for anyaccident, incident, ormalfunction followinga repairdoneotherthanbyitsservicecenterorbyanapprovedrepaircenter.

7.1 Warning • Toavoidelectricalshock,donotattempttoperformanyservicing

unlessyouarequalifiedtodoso.

• Donotperformanyservicewhilethemicro-ohmmeterisonanycircuit.

• Toavoidelectricalshockand/ordamagetotheinstrument,donotgetwaterorotherforeignagentsintotheelectronicmodule.

• Makesuretheinternalbatteryisfullychargedpriortotesting.Iftheinstrumenthasbeenleftunusedforseveralmonths,rechargethebattery.

• Werecommendrechargingthemicro-ohmmetereverymonthtoensureafullbatterychargewhenused.

• When replacing the fuse, install only the fusewhich is a directreplacement.

7.2 Cleaning

WARNING: Disconnect the instrument from any source of electricity.

• If thecaseneedscleaning,donotuseanyalcoholoroilbasedcleaners. Preferably use soapy water with a damp cloth orsponge.

• Dry immediatelyaftercleaning.Avoidwaterpenetration into theelectronicmodule.

• Makesurethemicro-ohmmeterandallleadsaredrybeforefur-theruse.

Page 38: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

36 Micro-Ohmmeter Model 6240

7.3 Charging/Recharging the Battery

AC POWER SELECTION

TheModel6240mayberechargedfrom110to230VAC(50/60Hz).Theinstrumentincludesa110Vlinecord,whichprovidesthechargingvoltagefortherechargeablebattery.

CHARGING THE BATTERY

• TheModel6240shouldbechargedtoafullchargebeforeusingitforthefirsttime.

• Chargingtofullcapacitymaytakeupto6hoursforacompletelydischargedbattery.

• Ifthebatterysymbolshowsasanemptycell,thebatteryneedstoberecharged.

The battery will only charge in the OFF mode.• IntheONmode,thecharginghaltsandthebatterylevelindication

isshowninthebatterysymboltothetopleftofthedisplay.• IntheOFFmode,indicationisasdescribedinthechargingindica-

torsectionbelow.

ConnecttheModel6240to110VACusingthepowercordprovided(charg-ingstartsautomaticallyiftheinstrumentisintheOFFmode).

NOTE: Measurement can be obtained while the power cord is connected but the charging process will be stopped until the instrument is turned OFF again.

CHARGING INDICATORS

OFF Mode:• bAttonthesmalldisplayand###onthemaindisplay,signifies

fast charging in progress.Where ### is the percent of batterycharge(onlyintheOFFposition).

• bAttonthesmalldisplayandFULLonthemaindisplay,signifiesthat charging is complete.At this point a low charge current isappliedtomaintainthebatterycharge.

Page 39: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

Micro-Ohmmeter Model 6240 37

ON Mode:• bAttshownwheninstrumentisturnedonandbatteryislow.There

willbea5seconddelaythen4beepsbeforetheinstrumentshutsdown.

• bAtt shown during high current measurement indicates a lowbatterycondition(measurementwillbepossibleonlowercurrentrangeonly).

Battery Indicator Status Battery Charge (C)C>87.5%

87.5%<C<62.5%62.5%<C<37.5%37.5%<C<12.5%

12.5%<C

Blink C=0(batteryIndicatornotinitialized)Automaticshutdown C = 0

If the instrument is turned on and the battery voltage is >5V, then thenormaluseofthedeviceispermitted.

7.4 Battery and Fuse Replacement

NOTE: Make sure that no terminals are connected and that the switch is in the OFF position before removing the instrument from the case.

BATTERY

WARNING:Replacingthebatterycausesdatalossfrommemory.• Thebatteryshouldbereplacedbyanauthorizedrepairfacilityrec-

ognizedbyAEMC® Instruments.SeetheRepairandCalibrationsectionforreturninstructions.

• The battery is accessible using a Phillips Head screwdriver toremovethefourscrewsonthebottomsideofthecaseandpullingthechassisoutfromthetop.

FUSE• The fuse is locatedon the front panel between theC1andP1

inputterminals.• FuseF1, is a 6.3x32mm, fast acting, 12.5A/500V, low internal

resistance,protectingthecurrentsourcefromoutsidevoltagesonenergizedspecimens.

Page 40: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

38 Micro-Ohmmeter Model 6240

Repair and Calibration

Toensurethatyourinstrumentmeetsfactoryspecifications,werecommendthatitbescheduledbacktoourfactoryServiceCenteratone-yearintervalsforrecalibration,orasrequiredbyotherstandardsorinternalprocedures.

For instrument repair and calibration:Youmust contactourServiceCenter foraCustomerServiceAuthorizationNumber(CSA#).Thiswillensurethatwhenyourinstrumentarrives,itwillbetrackedandprocessedpromptly.PleasewritetheCSA#ontheoutsideoftheshipping container. If the instrument is returned for calibration,weneed toknowifyouwantastandardcalibration,oracalibrationtraceabletoN.I.S.T.(Includescalibrationcertificateplusrecordedcalibrationdata).

Ship To: ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 15 Faraday Drive Dover,NH03820USA Phone:(800)945-2362(Ext.360) (603)749-6434(Ext.360) Fax: (603)742-2346or(603)749-6309 E-mail:[email protected]

(Orcontactyourauthorizeddistributor)Costsforrepair,standardcalibration,andcalibrationtraceabletoN.I.S.T.areavailable.NOTE: You must obtain a CSA# before returning any instrument.

Technical and Sales Assistance

Ifyouareexperiencinganytechnicalproblems,orrequireanyassistancewiththeproperoperationorapplicationofyourinstrument,pleasecall,mail,faxore-mailourtechnicalsupportteam:

ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 200FoxboroughBoulevard Foxborough,MA02035USA Phone:(800)343-1391 (508)698-2115 Fax: (508)698-2118 E-mail:[email protected] www.aemc.com

NOTE: Do not ship Instruments to our Foxborough, MA address.

Page 41: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

Micro-Ohmmeter Model 6240 39

Limited Warranty

TheModel6240iswarrantedtotheownerforaperiodofoneyearfromthedateoforiginalpurchaseagainstdefectsinmanufacture.ThislimitedwarrantyisgivenbyAEMC®Instruments,notbythedistributorfromwhomitwaspur-chased.Thiswarrantyisvoidiftheunithasbeentamperedwith,abusedorifthedefectisrelatedtoservicenotperformedbyAEMC®Instruments.

Full warranty coverage and product registration is available on our website at www.aemc.com/warranty.html.

Please print the online Warranty Coverage Information for your records.

What AEMC® Instruments will do:Ifamalfunctionoccurswithintheone-yearperiod,youmayreturntheinstrumenttousforrepair,providedwehaveyourwarrantyregistrationinformationonfileoraproofofpurchase.AEMC®Instrumentswill,atitsoption,repairorreplacethefaultymaterial.

REGISTER ONLINE AT: www.aemc.com

Warranty Repairs

What you must do to return an Instrument for Warranty Repair: First, request aCustomerServiceAuthorizationNumber (CSA#) by phoneorbyfaxfromourServiceDepartment(seeaddressbelow),thenreturntheinstrumentalongwith thesignedCSAForm.Pleasewrite theCSA#on theoutsideoftheshippingcontainer.Returntheinstrument,postageorshipmentpre-paidto:

Ship To: ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 15FaradayDrive•Dover,NH03820USA Phone:(800)945-2362(Ext.360) (603)749-6434(Ext.360) Fax: (603)742-2346or(603)749-6309 E-mail:[email protected]

Caution:Toprotectyourselfagainstin-transitloss,werecommendyouinsureyourreturnedmaterial.

NOTE: You must obtain a CSA# before returning any instrument.

Page 42: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

Notes:

Page 43: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol
Page 44: 6240 · 2017-02-02 · Micro-Ohmmeter Model 6240 5 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation. This symbol

04/16

99-MAN 100324 v15

Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments15FaradayDrive•Dover,NH03820USA•Phone:(603)749-6434•Fax:(603)742-2346