4.A.1 Discussion. - INFLIBNETshodhganga.inflibnet.ac.in/bitstream/10603/76847/11/11_chapter-4.pdfby...
Transcript of 4.A.1 Discussion. - INFLIBNETshodhganga.inflibnet.ac.in/bitstream/10603/76847/11/11_chapter-4.pdfby...
295
4.A.1 Discussion.
Assay method development and Validation of Ziprasidone Hydrochloride
4.A.1.1. HPLC method development and optimization
In the present study the main target for the development of chromatographic method was to get
the reliable method for the quantification of Ziprasidone hydrochloride from bulk drug and can
be applied for the degradable products ,as very few researcher as carried and reported
systematic study of HPLC and UV methods for estimation of Ziprasidone Hydrochloride
Monohydrate which were sophisticated, but time consuming. Thus ,the present study was aimed
for development of speedy and cost effective HPLC technique for determination of Ziprasidone
Hydrochloride Monohydrate as bulk and in dosage forms. The chromatographic method was
optimized by changing various parameters, such as the mobile phase composition , pH of the
buffer used in the mobile phase. Retention time and separation of peak of Ziprasidone
Hydrochloride Monohydrate were dependent on pH of the buffer and the percentage of
methanol. Various blends of solvent systems in varying proportions were tried as mobile phase
by G.Srinubabu,1-3
Different mobile phases were tried, but satisfactory separation and good
symmetrical peak were obtained with mobile phases having triethylamine buffer PH = 3.0 with
orthophosphuric acid .530 ml of buffer mixed with 150 ml acetonitrile and 320 ml of
methanol,the volume was 53:15:32,v/v/v ,it was filtered and degased . and Column used YMC
pack C18, 250 x 4.6 mm, 5µm found to comparatively better and gave the graph with better
gaussian shape at retention time 9.80 min .
4.A.1.2. Method validation
Validation of an analytical method is the process by which it is established by laboratory studies,
that the performance characteristics of the method meet the requirements for the intended
analytical application.4
Validation is required for any new or amended method to ensure that it is
capable of giving reproducible and reliable results, when used by different operators employing
the same equipment in the same or different laboratories. The type of validation programme
required depends entirely on the particular method and its proposed applications. The proposed
analytical method was validated for the parameters such as linearity, precision, accuracy,
specificity, limit of quantification [LOQ], limit of detection [LOD] ,Forced degradation and
robustness.5-7
296
4.A.1.3. System precision:
Precision is the measure of the degree of repeatability of an analytical method under normal
operation and is normally expressed as the percent relative standard deviation for a statistically
significant number of samples. The two most common precision measures are 'repeatability' and
'reproducibility'.These are expression of two extreme measure of precision which can be
obtained.
In the present finding Standard solution of Ziprasidone Hydrochloride Monohydrate were
prepared as per testing procedure and injected into the HPLC system in six replicates. The values
of % relative standard deviation (NMT 2.0%) for peak area obtained in six replicate injections
were reported in Table-3.A.1.1. thus showing that the equipment used for the study worked
correctly for the developed analytical method, and being highly repetitive 8.
Figure -4.A.1. A chromatogram of the Ziprasidone Hydrochloride standard
297
4.A.1.4. Linearity
A linear study identifies a specified concentration range where analytes response is linearly
proportional to the concentration. The present study reveals that standard curve found to be
linear over the concentration range of 1000-3000 ppm. The results are depicted in Table 3.A.1.2
of chapter -3. The equation of the standard curve relating the peak area to the Ziprasidone
concentration in this range was y = 480x + 560. The drug showed good linearity in the range of
1000-3000 ppm with coefficient of correlation value 0.999 for peak area (Table-3.A.1.2)
Figure-4. A.2. Linearity graph of Ziprasidone HCl
y = 480x + 560
R² = 0.993
0
500
1000
1500
2000
2500
3000
3500
0 1 2 3 4 5 6
298
4.A.1.5.Accuracy
Accuracy is popularly used to describe the measure of exactness of an analytical method, or the
close of an agreement between the value, which is accepted as a conventional, true value or as an
accepted reference value, and the value found.It is properly a qualitative concept and the correct
term is 'bias'.The bias of a method is an expression of how close the mean of asset of results
(produced by the method) is to the true value. Bias is usually determined by study of relevant
reference materials or by spiking studies 9
Accuracy , sometimes also referred to as recovery is an indicator of the trueness of the test
measurements. To determine the accuracy of the method three quality control samples were used.
The samples chosen were such to represent the entire range of the standard curve i e lower,
middle and higher concentration of the range.10
In the present finding recovery studies were
conducted after addition of standard drug solution at three different levels i.e. 50 %, 100 %, and
150 % to pre-analyzed sample solution and to check the accuracy of the method, Similar work
as been also reported by Chudasama, 11-12
. The results of the finding are reported in chapter 3
and in Table-3.A.1.4.
4.A.1.6. Robustness
The robustness of an analytical procedure refers to its capability to remain unaffected by small
and deliberate variations in method parameters. 13-14
The percentage recovery of Ziprasidone Hydrochloride Monohydrate was good under most
conditions and didn’t show any significant change when the critical parameters were modified.
The tailing factor for Ziprasidone Hydrochloride Monohydrate was always less than 2.0 and it
was well separated under all the changes carried out. Considering the modifications in the system
suitability parameters and the specificity of the method it can conclude that the method is robust.
Similar result for other drugs were reported by Ahmed et.al 15
4.A.1.7. Forced degradation :
Forced degradation studies are used to facilitate the development of analytical methodology, to
gain a better understanding of active pharmaceutical ingredient (API) and drug product (DP)
stability, and to provide information about degradation pathways and degradation products16
.
In present finding Forced degradation studies of Ziprasidone Hydrochloride Monohydrate were
carried under different stress condition such as acid, alkali , peroxide ,thermal and water
according to ICH guideline shown in table- 3.A.1.10 .22.97% degradation observed in 1.0 N
299
HCl. The peaks of the degradation products were well resolved.17
The degradation study showed
that Ziprasidone Hydrochloride was stable to chemical oxidation study, dry heat and alkali
hydrolysis where as it was highly susceptible to acid hydrolysis .
Figure -4. A.3.A chromatogram of the Ziprasidone Hydrochloride alkali
degradation
300
4.A.2. Assay method development and Validation of Duloxetine Hydrochloride.
4.A.2.1 HPLC method development and optimization
In present investigation the main target was for the development of chromatographic method to
get the reliable method for the quantification of Duloxetine hydrochloride from bulk drug and
which is applicable for the degradable products, for that number of systematic trials were
performed to optimize the chromatographic conditions for developing a sensitive, precise and
accurate RP-HPLC method for the analysis of Duloxetine HCl in pharmaceutical dosage forms.
Various blends of solvent systems in varying proportions were tried as mobile phase by O’Neil
and others 18-19
The present method contains mobile phase 0.1% Orthophosphuric acid buffer as
A and acetonitrile as mobile phase B , gradient program mention in expermential details and
column used Kromasil C18, 250 x 4.6 mm, 5µm found to comparatively better and gave the
graph with better gaussian shape at retention time 17.72 min.
4.A.2.2 Method validation
The validation of analytical procedures is based on the four most common types of analytical
procedures:- Identification tests, Quantitative tests for impurities' content,Limit tests for the
control of impurities, Quantitative tests of the active moiety in samples of drug substance or drug
product or other selected component(s) in the drug product 20
In the present study the developed method was validated, as described below, for various
parameters like linearity and range, accuracy, precision, ruggedness, system suitability,
specificity, LOQ,and LOD.
4.A.2.3 Precision
The precision of an analytical procedure expresses the closeness of agreement (degree
of scatter) between a series of measurements which is obtained from multiple sampling of the
same homogeneous sample under the prescribed conditions. Precision may be considered at three
levels: repeatability,intermediate precision and reproducibility.
Precision should be investigated using homogeneous, authentic samples. However, if it
is not possible to obtain a homogeneous sample, it may be investigated using artificially prepared
samples or a sample solution. The precision of an analytical procedure is usually expressed as the
variance, standard deviation or coefficient of variation of a series of measurements 21
301
In our study instrument precision was determined by performing repeatability test and the %RSD
(NMT 2.0%) values for Doluxtine hydrochloride and are reported in Chapter 3, Table-3.A.2.2
Figure -3.A.4. A chromatogram of the Doluxetine Hydrochloride standard
4.A.2.4. LOD and LOQ
The Quantitation limit of an individual analytical procedure is the lowest amount of analyte in a
sample which can be quantitatively determined with suitable precision and accuracy. The
Quantitation limit is a parameter of quantitative assays for low levels of compounds in sample
matrices and is used particularly for the determination of impurities or degradation products. 21
The detection limit of an individual analytical procedure is the lowest amount of analyte in a
sample which can be detected but not necessarily quantitated as an exact value.21
LOD and LOQ
of the drug were calculated using the following equations according to International Conference
on Harmonization (ICH) guidelines. 22
LOD = 3.3 × /S LOQ = 10 × /S where = the
standard deviation of the response and S = the slope of the regression equation and were found to
be 0.026 and 0.078 µg/ml.(Table-3.A.2.1). The present study reveals that the method is sensitive
for the determination of Doluxtine hydrochloride.
4.A.2.5. Linearity and Range
Linearity is the method's ability to obtain results which are either directly, or after mathematical
transformation proportional to the concentration of the analyte within a given range. The range of
the method is the interval between the upper and lower levels of an analyte that have been
302
determined with acceptable precision, accuracy and linearity. It is determined by calculating the
regression line using a mathematical treatment of the results (ie least mean squares) vs analyte
concentration. . 23
In present study linearity of the method was evaluated at five concentration levels by diluting the
standard stock solution. The calibration curve for Doluxetine hydrochloride was prepared by
plotting area v/s concentration. Calibration data for Doluxtine hydrochloride are given in Table
Table-3.A.2.3. The linearity plot of Doluxetine hydrochloride was found to be linear and
correlation coefficient 1.000 for Doluxetine hydrochloride i.e y = 747.3 x + 1412 .The present
study reveals that linearity observed were in the expected concentration range, demonstrating
suitability of the method for analysis, which may be attributed to the method linear in the
specified range for the analysis of Doluxtine hydrochloride .Our findings are in good aggrement
with the earlier reported work on other drugs24
Figure-4. A.5. Linearity graph of Duloxetine Hydrochloride
y = 747.3x + 1412.
R² = 1
0
500000
1000000
1500000
2000000
2500000
0 500 1000 1500 2000 2500 3000 3500
303
4.A.2.6. Accuracy
The accuracy of an analytical procedure expresses the closeness of agreement between the value,
which is accepted either as a conventional true value or an accepted reference value and the
value found. This is sometimes termed trueness.
Accuracy , sometimes also referred to as recovery is an indicator of the trueness of the test
measurements. To determine the accuracy of the method three quality control samples were used.
The samples chosen were such to represent the entire range of the standard curve i.e lower,
middle and higher concentration of the range 10
The recovery experiments were carried out by the standard addition method. The recoveries
obtained by the RPHPLC method for Doluxetine hydrochloride are depicted in Table 3.A.2.4.
The method was found to be accurate with % recovery 99% - 100.50% and found to be in
acceptable %RSD of not more than 2% at each level.
4.A.2.7. Robustness
The robustness of an analytical procedure is a measure of its capacity to remain unaffected by
small, but deliberate variations in method parameters and provides an indication of its reliability
during normal usage.21.
The present study reveals that the method found to be robust, which may
be attributed to the small but deliberate changes in the method and low value of relative standard
deviation. These parameters have no detrimental effect on the method performance. Similar
results are also reported by different research during their study. 21
4.A.2.8. Forced degradation
Stress testing of the active substance can help to identify the likely degradation products, which
can in turn help to establish the degradation pathways and the intrinsic stability of the molecule
and validate the stability indicating power of the analytical procedures used. The nature of the
stress testing depends on the individual active substance and the type of Pharmaceutical product
involved.25
In our present investigation forced degradation study were carried out by subjecting
the drug to acid and alkali hydrolysis, chemical oxidation, dry heat degradation and photolytic
(sun light) conditions was carried. . The peaks of the degradation products were well resolved.
The study reveals that chromatograms of alkali degraded sample showed 19.55 % degradation in
0.2 N NaOH. The Doluxetine hydrochloride was found to be stable to rest of the conditions like
oxidative stress degradation, dry heat degradation and acid hydrolysis. Our findings showed that
304
Doluxtine hydrochloride was stable to chemical oxidation study, dry heat and alkali hydrolysis
while it was highly susceptible to alkali hydrolysis which is in good aggrement with the work
published earlier.26
Figure-4. A.6. A chromatogram of the Doluxetine Hydrochloride alkali degradation
305
4.A.3. Assay method development and Validation of carmustine
4.A.3.1. HPLC method development and optimization
In our present work the main target was for the development of chromatographic method was to
get the reliable method for the quantification of Carmustine from bulk drug and which will be
also applicable for the degradable products. Few HPLC and UV methods have been reported for
estimation of Carmustine which are sophisticated, but time consuming. The present study was
aimed at development of speedy and cost effective HPLC technique for determination of
Carmustine as bulk and in dosage forms. The chromatographic method was optimized by
changing various parameters, such as the mobile phase composition , pH of the buffer used in the
mobile phase. Retention time and separation of peak of Carmustine were dependent on pH of the
buffer and the percentage of methanol. Various blends of solvent systems in varying proportions
were tried as mobile phase by Nataranjan et al 27
. The present study reveals that during use of
different mobile phases 0.01 M Potassium dihydrogen phosphate at pH 3.2 with Orthophosphuric
acid and acetonitrile in the ratio of (70:30,v/v) and column temperature 30 °C showed better peak
shap, finally mobile phase filtered and degased. The Column used YMC ODS-A C18, 250 x 4.6
mm, 5µm found to comparatively better and gave the graph with better gaussian shape at
retention time 14.70 min min .
4.A.3.2. Method validation
The Method was validated based on the International Conference on Harmonization (ICH)
Guidelines 28-30
. The method validation parameters checked were specificity, linearity,
accuracy, precision, limit of detection, limits of quantitation and robustness. The analytical
method was validated as per ICH guidelines with respect to parameters such as linearity,
precision, accuracy, specificity ,forced degradation and robustness.
4.A.3.3. System precision:
The precision of an analytical method is the closeness of a series of individual measurements of
an analyte when the analytical preocedure is applied repeatedly to multiple aliquots of a single
homogeneous volume of biological matrix. The precision is calculated as coefficient of variation
(C. V.), i.e., relative standard deviation (RSD). The measure RSD can be subdivided into three
categories: repeatability (intra-day precision) and reproducibility (between laboratories
precision).31
306
In the present finding Standard solution of Carmustine were prepared as per testing procedure
and injected into the HPLC system in six replicates.The study reveals that the values of %
relative standard deviation 0.20 (NMT 2.0%) for peak area obtained in six replicate injections
indicates that the equipment used for the study worked correctly for the developed analytical
method, and being highly repetitive. It is in good agrrement with shinde et.al and others32-33.
The
results are reported in chapter 3, Table-3.A.3.1.
Figure -4.A.7. A chromatogram of the carmustine standard
4.A.3.4. Linearity
The linearity of an analytical procedure is its ability (within a given range) to obtain test results,
which are directly proportional to the concentration (amount) of analyte in the sample. 34-35
The
equation of the standard curve relating the peak area to the carmustine concentration were in the
range of y = 751x + 1252 .In our findings the drug showed good linearity in the range of 1000-
3000 ppm with coefficient of correlation value 1.0000 for peak area (Table-3.A.3.2). . The
standard curve found to be linear over the concentration range of 1000-3000 ppm.
307
Figure-4. A.8. Linearity graph of Carmustine
4.A.3.5. Accuracy
Accuracy , sometimes also referred to as recovery is an indicator of the trueness of the test
measurements. In our present study to determine the accuracy of the method by taking three
samples were used. The samples chosen were such that they represent the entire range of the
standard curve i.e lower, middle and higher concentration of the range.
Recovery studies are carried after addition of standard drug solution at three different levels i.e.
50 %, 100 %, and 150 % to pre-analyzed sample solution to varify the accuracy of method
similar study for different drugs was also carried by Susan et al and others 36-37
. The results are
given in table-3.A.3.3, 3.A.3.4 and 3.A.3.5.
4.A.3.6. Robustness
In the present findings the percentage recovery of Carmustine was good under most conditions
and didn’t show any significant change when the critical parameters were modified. The tailing
factor for Carmustine was always less than 2.0 and it was well separated under all the changes
carried out. The modifications made in the system suitability parameters and the specificity of
the method shows that the method is robust. Robustness parameter flow plus,flow minus,pH plus
and pH minus results are given in table-3.A.3.6, 3.A.3.7 and 3.A.3.8.
y = 751x + 1252.
R² = 1
0
500000
1000000
1500000
2000000
2500000
0 500 1000 1500 2000 2500 3000 3500
308
4.A.3.7. Forced degradation :
In present investigation the forced degradation of Carmustine under different stress condition
such as acid, alkali , peroxide ,thermal and water according to ICH guidline shown in table-
3.A.3.10 .
21.55% degradation observed in 0.1 N NaOH. The peaks of the degradation products were well
resolved shown in fig-3A.17. The degradation study reveals that Carmustine was stable to
chemical oxidation study, dry heat and acid hydrolysis while it was highly susceptible to alkali
hydrolysis .
Figure-4. A.9. A chromatogram of the carmustine alkali degradation.
309
4.A.4. Assay method development and Validation of Irinotecan Hydrochloride
4.A.4.1. HPLC method development and optimization
In our present work the main target for the development of chromatographic method was to get
the reliable method for the quantification of Irinotecan hydrochloride from bulk drug and which
are also applicable for the degradable products. It was also observed that few HPLC and UV
methods have been reported for estimation of Irinotecan hydrochloride which are sophisticated,
but time consuming. The present study was aimed at development of speedy and cost effective
HPLC technique for determination of Irinotecan hydrochloride as bulk and in dosage forms. The
chromatographic method was optimized by changing various parameters, such as the mobile
phase composition, pH of the buffer used in the mobile phase. Retention time and separation of
peak of Irinotecan hydrochloride were dependent on pH of the buffer and the percentage of
methanol. HPLC study carried out by using various blends of solvent systems in varying
proportions were tried as mobile phase as reported earlier 41-42
Different mobile phases were
tried, but satisfactory separation and good symmetrical peak were obtained with the mobile
2.0gm Sodium dihydrogen phosphate and 1.0gm 1-Octane Sulphonic acid salt in 1000 ml 550 ml
buffer ,170 ml acetonitrile and 280 ml methanol mixed filtered and degased . The column
temperature was maintained at 45 °C have better peak shape. and Column used were Inertsil
ODS 3V C18(250X4.6)mm,5µ found to comparatively better and gave the graph with better
gaussian shape at retention time 13.33 min.
4.A.4.2. Method validation
Method validation of an analytical procedure is to demonstrate suitability for intended purpose.
To meet current pharmaceutical regulatory guidelines i.e. USP , ICH and EP , a number of
criteria such as specificity, linearity, precision, accuracy, sensitivity and robustness are
determined investigated in order to validate analytical methods.5
4.A.4.3. System precision:
Precision which was an important aspect of method validation were carried and used to ensure
adequate performance of the chromatographic system. Retention time (RT), number of
theoretical plates (N) and tailing factor (T) Standard solution of Irinotecan hydrochloride was
prepared as per testing procedure and injected into the HPLC system in six replicates.171
The
values of % relative standard deviation 0.14 (NMT 2.0%) for peak area obtained in six replicate
injections are reported in Table-3.A.4.1. In our finding the the low value of relative standard
310
deviation attributes that the method is robust.Similar results are also reported by different
research during their study.Thus showing that the equipment used for the study worked correctly
for the developed analytical method, and being highly repetitive .43
Figure -3.A.10. A chromatogram of the Irinotecan Hydrochloride standard
prepration.
4.A.4.4. Linearity
linearity is a highly successful parameter for method accuracy and defined as the method's
ability to obtain results which are either directly, or after mathematical
transformation proportional to the concentration of the analyte within a given range. It is
determined by calculating the regression line using a mathematical treatment of the results (i.e
least mean squares) vs analyte concentration.A linear study identifies a specified concentration
range where analytes response is linearly proportional to the concentration. The standard curve
found to be linear over the concentration range of 1000-3000 ppm. The equation of the standard
curve relating the peak area to the Irinotecan Hydrochloride concentration in this range was y =
813x + 1523 . The present study reveals that ,the drug showed good linearity in the range of
1000-3000 ppm with coefficient of correlation value 0.9999 for peak area (Table-3.A.4.2).
311
Figure- 4. A.11. Linearity graph of Irnotecan Hydrochloride
4.A.4.5. Accuracy
Accuracy methods have been described as a measure of the closeness of test results obtained by
a method to the true value. Accuracy indicates the deviation between the mean value found and
the true value. lt is determined by applying the method to samples to which known amounts of
analyte have been added. It is analysed against standard and blank solutions to ensure that no
interference exists. The accuracy is then calculated from the test results as a percentage of the
analyte recovered by the assay. Accuracy ,it is sometimes also referred to as recovery is an
indicator of the trueness of the test measurements. In our present investigation to determine the
accuracy of the method three quality control samples were used. The samples chosen were such
to represent the entire range of the standard curve i.e. lower, middle and higher concentration of
the range To check the accuracy of the method, recovery studies were conducted after addition
of standard drug solution at three different levels i.e. 50 %, 100 %, and 150 % to pre-analyzed
sample solution. Our findigs good aggrement with the earlier reported work on other drugs by
Herben et al 44-46
. The results of the findings are given in table-3.A.4.3 3.A.4.4 and 3.A.4.5.
y = 813x + 1523.
R² = 1
0
500000
1000000
1500000
2000000
2500000
0 500 1000 1500 2000 2500 3000 3500
312
4.A.4.6. Robustness
Robustness is the ability of a method to tolerate variations of parameters without significant
change in the result47
.In the present findings recovery of Irinotecan hydrochloride was good
under most conditions and didn’t show any significant change when the critical parameters were
modified. The tailing factor for Irinotecan hydrochloride was always less than 2.0 and it was well
separated under all the changes carried out.48-51
Considering the modifications in the system
suitability parameters and the specificity of the method it can conclude that the method is robust.
Robustness parameter flow plus,flow minus, Temperature plus and Temperature minus results
are depicted in table-3.A.4.6, 3.A.4.7 and 3.A.4.8.
4.A.4.7. Forced degradation :
In present investigation the forced degradation of Irinotecan hydrochloride under different stress
condition such as acid, alkali , peroxide ,thermal and water according to ICH guidline reported
in table- 3.A.4.10 . The present study reveals that 15.88% degradation observed in 0.2 N NaOH.
Which is an indicative of development of new chromatographic analytical method to resolve the
problem. The chromatogram are shown in figure-4.A.8 .The peaks of the degradation products
were well resolved, shown in fig-3A.21 The degradation study thereby indicated that Irinotecan
hydrochloride were stable to chemical oxidation study, dry heat and alkali hydrolysis while it
was highly susceptible to alkali hydrolysis .Acid and thermal degradation study tried by Kono
et al and others 52-54
Figure -4. A.12. A chromatogram of the Irinotecan Hydrochloride alkali
degradation.
313
4.A.5. Assay method development and Validation of Cyclophosphamide
4.A.5.1. HPLC method development and optimization
In present investigation the main target for the development of chromatographic method was to
get the reliable method for the quantification of Cyclophosphamide from bulk drug and which
will be also applicable for the degradable products.55
It was also observed that few HPLC and
UV methods have been reported for estimation of Cyclophosphamide from plasma which are
sophisticated, but time consuming. The present study was aimed at development of speedy and
cost effective HPLC technique for determination of Cyclophosphamide as bulk and in dosage
forms. The chromatographic method was optimized by changing various parameters, such as the
mobile phase composition, pH of the buffer used in the mobile phase. Retention time and
separation of peak of Cyclophosphamide were dependent on pH of the buffer and the percentage
of methanol. Various blends of solvent systems in varying proportions were tried as mobile phase
by Alsarra et al and others 56-57
. Different mobile phases were tried, but satisfactory separation
and good symmetrical peak were obtained with the mobile phases containg 800 ml of water and
200 ml methanol mixed filtered and degased and column temperature was maintained at 25 °C
have better peak shape. The Column used Hypersil BDS C8, 250 x 4.6 mm, 5µm found to
comparatively better and gave the graph with a much better peak shape at retention time 6.21
min.
4.A.5.2. Method validation
Validation is required for any new or amended method to ensure that it is capable of giving
reproducible and reliable results, when used by different operators employing the same
equipment in the same or different laboratories.The analytical method was validated as per ICH
guidelines with respect to parameters such as linearity, precision, accuracy, specificity, Forced
degradation and robustness.
4.A.5.3. System precision:
"The precision of an analytical method is the degree of agreement among individual test results
obtained when the method is applied to multiple sampling of a homogenous sample. It is a
measure of the reproducibility of the whole analytical method (including sampling, sample
preparation and analysis) under normal operating circumstances. It can be determined by using
the method to assay a sample for a sufficient number of times to obtain statistically valid results
314
and is then expressed as the relative standard deviation. In the present study Standard solution of
Cyclophosphamide was prepared as per testing procedure and injected into the HPLC system in
six replicates. The values of % relative standard deviation 0.11 (NMT 2.0%) for peak area
obtained in six replicate injections are reported in Table-3.A.5.1. In our finding the the low value
of relative standard deviation attributes that the method is robust.Similar results are also reported
by different researchers Casale et al during their studyWhich is an indicative of the equipment
used for the study worked correctly for the developed analytical method, and being highly
repetitive. 58-59
Figure -3.A.13. A chromatogram of the Cyclophosphamide Standard
4.A.5.4. Linearity
Linearity is determined by calculating the regression line using a mathematical treatment of the
results (ie least mean squares) vs analyte concentration.
In our present investigation a linear study identifies a specified concentration range where
analytes response is linearly proportional to the concentration. The standard curve found to be
linear over the concentration range of 1000-3000 ppm. The equation of the standard curve
relating the peak area to the Cyclophosphamide concentration in this range was y = 699x + 1359
The drug showed good linearity in the range of 1000-3000 ppm with coefficient of correlation
value 0.9993 for peak area (Table-3.A.5.2). Our findings are in good aggrement with the work
published earlier by Compagnon et al 60
315
Figure : 4. A.14. Linearity graph of cyclophosphamide
4.A.5.5. Accuracy
Accuracy indicates the deviation between the mean value found and the true value. lt is
determined by applying the method to samples to which known amounts of analyte are added,
And are analysed against standard and blank solutions to ensure that no interference exists. The
accuracy is then calculated from the test results as a percentage of the analyte recovered by the
assay.. In the present study to determine the accuracy of the method three samples were used.
The samples chosen to represented the entire range of the standard curve i.e lower, middle and
higher concentration of the range. Recovery studies were also carried after addition of standard
drug solution at three different levels i.e. 50 %, 100 %, and 150 % to pre-analyzed sample
solution for conformation of methods accuracy61
. The results are given in table-3.A.5.3, 3.A.5.4
and 3.A.5.5
4.A.5.6. Robustness
"The robustness of an analytical procedure is a measure of its capacity to remain unaffected by
small, but deliberate variations in method parameters and provides an indication of its reliability
during normal usage".62
In present investigation the percentage recovery of Cyclophosphamide was good under most
conditions and didn’t showed no significant change when the critical parameters were modified.
y = 699x + 1359.
R² = 1
0
500000
1000000
1500000
2000000
2500000
0 500 1000 1500 2000 2500 3000 3500
316
The tailing factor for Cyclophosphamide was always less than 2.0 and well separated under all
the changes carried out. Considering the modifications in the system suitability parameters and
the specificity of the method it can conclude that the method is robust. Robustness parameter
flow plus,flow minus, Temperature plus and Temperature minus results are given in table-
3.A.5.6, 3.A.5.7 , 3.A.5.8 and 3.A.5.9.
4.A.5.7. Forced degradation :
Analytical specificity of a method may be defined as the ability to measure accurately and
specifically the analyte in the presence of components that may be expected to be present in the
sample matrix.Specificity for an assay ensures that the signal measured comes from the
substance of interest, and that there is no interference from excipient degradation products and/or
impurities. 63
In our present study the forced degradation of Cyclophosphamide under different stress condition
such as acid, alkali , peroxide ,thermal and water according to ICH guideline are reported in
table- 3.A.5.10 . However this is the first time reported that 12.44% degradation observed in 0.1
N NaOH. Typical chromatogram show in figure-4.A.10 The peaks of the degradation products
were well resolved. The degradation study thereby indicated that Cyclophosphamide was stable
to chemical Our findings are in good aggrement with the work published earlier for different
drug by Arayne et al and others 64-68
oxidation study, dry heat and alkali hydrolysis while it was
highly susceptible to alkali hydrolysis .
Figure -4. A.15. A chromatogram of the Cyclophosphamide alkali degradation.
317
4.A.6. Assay method development and Validation of Thiamine Hydrochloride
4.A.6.1. HPLC method development and optimization
In our present work the main target for the development of chromatographic method was to get
the reliable method for the quantification of Thiamine hydrochloride from bulk drug and which
will be also applicable for the degradable products. HPLC method by Thomas 69
and UV
methods have been reported for estimation of Thiamine hydrochloride which are sophisticated,
but time consuming. UV assay method reported by Yantih et al and others. 70-72
The present
study was aimed at development of speedy and cost effective HPLC technique for determination
of Thiamine hydrochloride as bulk and in dosage forms. The chromatographic method was
optimized by changing various parameters, such as the mobile phase composition , pH of the
buffer used in the mobile phase. Retention time and separation of peak of Thiamine
hydrochloride were dependent on pH of the buffer and the percentage of methanol. Various
blends of solvent systems in varying proportions were tried as mobile phase 73
. Different mobile
phases were tried, but satisfactory separation and good symmetrical peak were obtained with the
mobile phases 0.01 M Ammonium acetate buffer adjust PH = 4.5 with acetic acid . 100 ml of
buffer ,600 ml acetonitrile and 300 ml of methanol mixed filtered and degased .The Column used
Water Symmetry C18(250X4.6),5 m found to comparatively better and gave the graph with
better gaussian shape at retention time 12.17 min.
4.A.6.2. Method validation
Method validation of analytical procedures is based on the four most common types of analytical
procedures:- Identification tests,Quantitative tests for impurities' content,Limit tests for the
control of impurities, Quantitative tests of the active moiety in samples of drug substance or drug
product or other selected component(s) in the drug product. 20
The analytical method was validated as per ICH guidelines with respect to parameters such as
linearity, precision, accuracy, specificity, forced degradation and robustness.
4.A.6.3. System precision:
In the present findings Standard solution of Thiamine hydrochloride was prepared as per testing
procedure and injected into the HPLC system in six replicates. The values of % relative standard
deviation 0.08 (NMT 2.0%) for peak area obtained in six replicate injections are reported in
Table-3.A.6.1. thus showing that the equipment used for the study worked correctly for the
318
developed analytical method, and being highly repetitive. In our finding the low value of relative
standard deviation attributes that the method is robust.Similar results are also reported by
different researchers during their study Amidzic et al 74
.
Figure -3.A.16. A chromatogram of the Thiamine Hydrochloride standard
Prepration.
4.A.6.4. Linearity
Analytical linearity study identifies a specified concentration range where analytes response is
linearly proportional to the concentration. In our finding the standard curve found to be linear
over the concentration range of 1000-3000 ppm. The equation of the standard curve relating the
peak area to the Thiamine Hydrochloride concentration in this range was y = 733x+1389
The drug showed good linearity in the range of 1000-3000 ppm with coefficient of correlation
value 0.9999 for peak area (Table-3.A.6.2). Although the initial results obtained in this study are
very promising.
Figure- 4. A.17.Linearity graph of Thiamine Hydrochloride
y = 733x + 1389.
R² = 1
0
500000
1000000
1500000
2000000
2500000
0 500 1000 1500 2000 2500 3000 3500
319
4.A.6.5. Accuracy
Accuracy is popularly used to describe the measure of exactness of an analytical method, or the
close of an agreement between the value, which is accepted as a conventional, true value or as an
accepted reference value, and the value found.It is properly a qualitative concept and the correct
term is 'bias'. In present investigation to check the accuracy of the method, recovery studies were
carried after addition of standard drug solution at three different levels i.e. 50 %, 100 %, and 150
% to pre-analyzed sample solution for conformation of methods accuracy75
. The results are given
in table-3.A.6.3, 3.A.6.4 and 3.A.6.5. Hence this present study investigates the various abilities.
4.A.6.6. Robustness
In our finding the percentage recovery of Thiamine hydrochloride was good under most
conditions and didn’t show any significant change when the critical parameters were modified.
The tailing factor for Thiamine hydrochloride was always less than 2.0 and it was well separated
under all the changes carried out. Considering the modifications in the system suitability
parameters and the specificity of the method it can conclude that the method is robust.
Robustness parameter flow plus,flow minus, Temperature plus and Temperature minus results
are given in table-3.A.6.6, 3.A.6.7 , 3.A.6.8 and 3.A.6.9. Our findings are in good aggrement
with the work published earlier by Ditjen et al 76-77
4.A.6.7. Forced degradation :
Selectivity is the ability to measure accurately and specifically the analyte in the presence of
components that may be expected to be present in the sample matrix.
In present investigation the forced degradation of Thiamine hydrochloride under different stress
condition such as acid, alkali , peroxide ,thermal and water according to ICH guideline shown in
table- 3.A.6.10 .However this present study investigates the various abilities ,In present
investigation 11.89% degradation observed in 0.2 N NaOH. Typical chromatogram of degredable
compound are shown in figure-4.A.12 The peaks of the degradation products were well resolved.
The degradation study thereby indicated that Thiamine hydrochloride was stable to chemical
oxidation study, dry heat and acid hydrolysis while it was highly susceptible to alkali hydrolysis
shown in figure. Our findings are in good aggrement with the work published earlier by Ibrahim
et al 78
320
Figure- 4. A.18.A chromatogram of the Thiamine Hydrochloride alkali degradation
321
4.A.7. Assay method development and Validation of Topotecan Hydrochloride
4.A.7.1. HPLC method development and optimization
In present investigation the Topotecan is freely soluble in isopropyl alcohol. The drug can be
separated on a Inertsil ODS 3V C18(250X4.6)mm,5µ column in reverse phase mode The main
target for the development of chromatographic method was to get the reliable method for the
quantification of Topotecan hydrochloride from bulk drug and which will be also applicable for
the degradable products. We also observed a trend where few HPLC and UV methods have been
reported for estimation of Topotecan hydrochloride which are sophisticated, but time consuming.
The present study was aimed at development of speedy and cost effective HPLC technique for
determination of Topotecan hydrochloride as bulk and in dosage forms. The chromatographic
method was optimized by changing various parameters, such as the mobile phase composition ,
pH of the buffer used in the mobile phase. Retention time and separation of peak of Topotecan
hydrochloride were dependent on pH of the buffer and the percentage of methanol. Various
blends of solvent systems in varying proportions were tried as mobile phase by Hsiang et al and
others 79-80
. Different mobile phases were tried, but satisfactory separation and good symmetrical
peak were obtained with the mobile phases 0.02M Potassium dihydrogen phosphate buffer pH =
3.0 with orthophosphuric acid , 300 ml of buffer and 700 ml acetonitrile are mixed,filtered and
degassed.The column temperature were maintained at 35 °C and gave the graph with better
gaussian shape at retention time 13.26 min .
4.A.7.2. Method validation
The objective of validation of an analytical procedure is to demonstrate that it is suitable for its
intended purpose. To meet current pharmaceutical regulatory guidelines i.e. USP , ICH and EP ,
a number of criteria such as specificity, linearity, precision, accuracy, sensitivity and robustness
must be investigated in order to validate analytical methods.
4.A.7.3. System precision:
Precision was carefully tested ,System suitability test as an integral part of method development
was used to ensure adequate performance of the chromatographic system. Retention time (RT),
number of theoretical plates (N) and tailing factor (T) .Standard solution of Topotecan
hydrochloride was prepared as per testing procedure and injected into the HPLC system in six
replicates. The values of % relative standard deviation 0.10 (NMT 2.0%) for peak area obtained
322
in six replicate injections were reported in Table-3.A.7.1. In our finding the the low value of
relative standard deviation attributes that the method is robust.Similar results are also reported by
different research during their study Underberg et al 81
.The present study reveals that the
equipment used for the study worked correctly for the developed analytical method, and being
highly repetitive.
Figure -3.A.19. A chromatogram of the Topotecan Hydrochloride standard
prepration
4.A.7.4. Linearity
. Linearity is determined by calculating the regression line using a mathematical treatment of the
results (i.e least mean squares) vs analyte concentration. In the present findings a linear study
identifies a specified concentration range where analytes response is linearly proportional to the
concentration. The standard curve found to be linear over the concentration range of 1000-3000
ppm. The equation of the standard curve relating the peak area to the Topotecan Hydrochloride
concentration in this range was y = 812x+1544. This is an important observation. Therefore,in
this thesis, developments of new chromatographic analytical methods were established to solve
this problem. The drug showed good linearity in the range of 1000-3000 ppm with coefficient of
correlation value 0.9995 for peak area (Table-3.A.7.2).
323
Figure- 4. A.20.Linearity graph of Topotecan Hydrochloride
4.A.7.5. Accuracy
Analytical accuracy , sometimes also referred to as recovery is an indicator of the trueness of the
test measurements. To determine the accuracy of the method three quality control samples were
used. The samples chosen were such to represent the entire range of the standard curve i.e. lower,
middle and higher concentration of the range
In present investigation to check the accuracy of the method, recovery studies were conducted
after addition of standard drug solution at three different levels i.e. 50 %, 100 %, and 150 % to
pre-analyzed sample solution Hence these evidence support the hypothesis that result are good
aggriments with Pearson et al 82-83
. The results are given in table-3.A.7.3, 3.A.7.4 and 3.A.7.5
4.A.7.6. Robustness
The ICH guidelines also recommend that "one consequence of the evaluation of robustness
should be that a series of system suitability parameters (e.g. resolution tests) is established to
ensure that the validity of the analytical procedure is maintained whenever used".84
In the present findings the percentage recovery of Topotecan hydrochloride was good under most
conditions and didn’t show any significant change when the critical parameters were modified.
The tailing factor for Topotecan hydrochloride was always less than 2.0 and it was well separated
y = 812x + 1544.
R² = 1
0
500000
1000000
1500000
2000000
2500000
0 500 1000 1500 2000 2500 3000 3500
324
under all the changes carried out.85-88
Considering the modifications in the system suitability
parameters and the specificity of the method it can conclude that the method is robust.
Robustness parameter flow plus,flow minus, Temperature plus and Temperature minus results
are given in table-3.A.7.6, 3.A.7.7 , 3.A.7.8 and 3.A.7.9.
4.A.7.7. Forced degradation :
Selectivity is the ability to measure accurately and specifically the analyte in the presence of
components that may be expected to be present in the sample matrix.Our present work the forced
degradation of Topotecan hydrochloride under different stress condition such as acid, alkali ,
peroxide ,thermal and water according to ICH guidline shown in table- 3.A.7.10 .In present
findings 18.78% degradation observed in 0.5 N NaOH. Typical chromatogram of degredable
compound show in figure-4.A.14. The peaks of the degradation products were well resolved.89-92
The degradation study thereby indicated that Topotecan hydrochloride was stable to chemical
oxidation study, dry heat and alkali hydrolysis while it was highly susceptible to alkali hydrolysis
. These results facilitate the detection of that impurity in presence of the main drug by using the
developed method.
Figure- 4. A.21. A chromatogram of the Topotecan Hydrochloride alkali
degradation
325
4.A.8. Assay method development and Validation of Tenofovir Disoproxil fumarate
4.A.8.1. HPLC method development and optimization
In present investigation the main target for the development of chromatographic method was to
get the reliable method for the quantification of Tenofovir Disoproxil fumarate from bulk drug
and which will be also applicable for the degradable products. This is an important observation
few HPLC and UV methods have been reported for estimation of Tenofovir Disoproxil fumarate
which are sophisticated, but time consuming. The present study was aimed at development of
speedy and cost effective HPLC technique for determination of Tenofovir Disoproxil fumarate as
bulk and in dosage forms. The chromatographic method was optimized by changing various
parameters, such as the mobile phase composition , pH of the buffer used in the mobile phase.
Retention time and separation of peak of Tenofovir Disoproxil fumarate were dependent on pH
of the buffer and the percentage of methanol. Various blends of solvent systems in varying
proportions were tried as mobile phase by Sweetman et al 93-94
. Different mobile phases were
tried, but satisfactory separation and good symmetrical peak were obtained with the mobile
phases 0.01M Potassium dihydrogen phosphate buffer pH = 4.5 with orthophosphuric acid 900
ml buffer and 100 ml acetonitrile mixed, filtered anddegassed.The column temperature was
maintained at 25°C . Column used are Inertsil ODS 3V C18(150X4.6)mm,5µ found to
comparatively better and gave the graph with better gaussian shape at retention time 25.54 min .
4.A.8.2. Method validation
The Method was validated based on the International Conference on Harmonization (ICH)
Guidelines 28-30
. The method validation parameters checked were specificity, linearity, accuracy,
precision, limit of detection, limits of quantitation and robustness. The analytical method was
validated as per ICH guidelines with respect to parameters such as linearity, precision, accuracy,
specificity ,forced degradation and robustness
4.A.8.3. System precision:
In present investigation Tenofovir Disoproxil fumarate was prepared as per testing procedure and
injected into the HPLC system in six replicates. The values of % relative standard deviation 0.14
(NMT 2.0%) for peak area obtained in six replicate injections are reported in Table-3.A.8.1. In
our finding the the low value of relative standard deviation attributes that the method is
robust.Similar results are also reported by different researchers during their study. 95
326
thus showing that the equipment used for the study worked correctly for the developed analytical
method, and being highly repetitive 96
.
Figure -4.A.22. A chromatogram of the tenofovir Desoproxill fumarate standard
4.A.8.4. Linearity
Linearity corresponds to the capacity of the method to supply results directly proportional to the
concentration of the substance being determined within a certain interval of concentration A
linear study identifies a specified concentration range where analytes response is linearly
proportional to the concentration. In present findings the standard curve found to be linear over
the concentration range of 1000-3000 ppm. The equation of the standard curve relating the peak
area to the tenofovir Desoproxill fumarate concentration in this range was y = 712x+1388.
The drug showed good linearity in the range of 1000-3000 ppm with coefficient of correlation
value 1.0000 The linearity was evaluated by linear regression analysis calculated by the least
square method (Table-3.A.8.2). Our findings are in good aggrement with the earlier reported
work on other drugs by Gish et al. 97
327
Figure- 4. A.23. Linearity graph of Tenofovir Disoproxil Fumarate
4.A.8.5. Accuracy
Analytical accuracy of a method may be defined as , an indicator of the trueness of the test
measurements. To determine the accuracy of the method three quality control samples were used.
The samples chosen were such to represent the entire range of the standard curve i.e. lower,
middle and higher concentration of the range 98
In the present findings to check the accuracy of the method, recovery studies were conducted
after addition of standard drug solution at three different levels i.e. 50 %, 100 %, and 150 % to
pre-analyzed sample solution. The results are given in table-3.A.8.3, 3.A.8.4 and 3.A.8.5
4.A.8.6. Robustness
In the present findings the percentage recovery of Tenofovir Disoproxil fumarate was good under
most conditions and didn’t show any significant change when the critical parameters were
modified. The tailing factor for Tenofovir Disoproxil fumarate was always less than 2.0 and it
was well separated under all the changes carried out. Considering the modifications in the system
suitability parameters and the specificity of the method it can conclude that the method is robust.
Robustness parameter flow plus,flow minus, Temperature plus and Temperature minus results
are given in table-3.A.8.6, 3.A.8.7 , 3.A.8.8 and 3.A.8.9. 99
y = 712.8x + 1388.
R² = 1
0
500000
1000000
1500000
2000000
2500000
0 500 1000 1500 2000 2500 3000 3500
328
4.A.8.7. Forced degradation :
In our finding the forced degradation of Tenofovir Disoproxil fumarate under different stress
condition such as acid, alkali , peroxide ,thermal and water according to ICH guidline shown in
table- 3.A.1.10 . Our results have shown that 12.77% degradation observed in 0.2N M NaOH
and 0.1N M HCl shown 10.44% The peaks of the degradation products were well resolved.
Typical chromatogram of degredable compound show in figure-4.A.16. The degradation study
thereby indicated that Tenofovir Disoproxil fumarate was stable to chemical oxidation study, dry
heat and alkali hydrolysis while it was highly susceptible to acid and alkali hydrolysis. Our
findings are in good aggrement with the work published earlier by Mangoankar et al .100-101
Figure- 4. A.24. A chromatogram of the tenofovir Desoproxill fumarate acid
degradation
329
Figure- 4. A.25. A chromatogram of the tenofovir Desoproxill fumarate alkali
degradation
330
4.A.9. Assay method development and Validation of Atazanavir
4.A.9.1. HPLC method development and optimization
In our present work the main target for the development of chromatographic method was to get
the reliable method for the quantification of Atazanavir from bulk drug and which will be also
applicable for the degradable products. It was also observed that few HPLC and UV methods
have been reported for estimation of Atzanavir which are sophisticated, but time consuming. The
present study was aimed at development of speedy and cost effective HPLC technique for
determination of Atazanavir as bulk and in dosage forms. The chromatographic method was
optimized by changing various parameters, such as the mobile phase composition , pH of the
buffer used in the mobile phase. Retention time and separation of peak of Atazanavir were
dependent on pH of the buffer and the percentage of methanol. Various blends of solvent systems
in varying proportions were tried as mobile phase by Cateau et al 102-103
. Different mobile phases
were tried, but satisfactory separation and good symmetrical peak were obtained with the mobile
phases 0.01M Potassium dihydrogen phosphate buffer pH = 3.2 with orthophosphuric acid .800
ml of buffer and 200 ml acetonitrile mixed filtered and degased. The column temperature was
maintained at 25°C . The Column used YMC ODS-A(150X4.6) mm, 3.5 found to
comparatively better and gave the graph with better gaussian shape at retention time 29.08 min .
4.A.9.2. Method validation
The objective of validation of an analytical procedure is to demonstrate that it is suitable for its
intended purpose. To meet current pharmaceutical regulatory guidelines i.e. USP , ICH and EP ,
a number of criteria such as specificity, linearity, precision, accuracy, sensitivity and robustness
must be investigated in order to validate analytical methods.
4.A.9.3. System precision:
Precision was carefully tested the precision of an analytical method is the degree of agreement
among individual test results obtained when the method is applied to multiple sampling of a
homogenous sample.Precision is a measure of the reproducibility of the whole analytical method
(including sampling, sample preparation and analysis) under normal operating circumstances.
Precision is determined by using the method to assay a sample for a sufficient number of times to
obtain statistically valid results . The precision is then expressed as the relative standard
deviation.
331
Standard solution of Atazanavir was prepared as per testing procedure and injected into the
HPLC system in six replicates. The values of % relative standard deviation 0.07 (NMT 2.0%) for
peak area obtained in six replicate injections were reported in Table-3.A.9.1. In our finding the
low value of relative standard deviation attributes that the method is robust.Similar results are
also reported by different research during their study thus showing that the equipment used for
the study worked correctly for the developed analytical method, and being highly repetitive
Loregian et al 104
.
Figure -4.A.26. A chromatogram of the atazanavir standard prepration
4.A.9.4. Linearity
Linearity is determined by calculating the regression line using a mathematical treatment of
the results (ie least mean squares) vs analyte concentration.In our present work a linear study
identifies a specified concentration range where analytes response is linearly proportional to the
concentration. In present findings standard curve found to be linear over the concentration range
of 1000-3000 ppm. The equation of the standard curve relating the peak area to the Atazanavir
concentration in this range was y = 1933x + 588.7 . The drug shown good linearity in the range
of 1000-3000 ppm with coefficient of correlation value 0.9996 for peak area (Table-3.A.9.2).
332
Figure- 4. A.27. Linearity graph of Atazanavir
4.A.9.5. Accuracy
Accuracy was measured as the percent of deviation from the nominal concentration Accuracy ,
sometimes also referred to as recovery is an indicator of the trueness of the test measurements.
To determine the accuracy of the method three quality control samples were used. The samples
chosen were such to represent the entire range of the standard curve i.e. lower, middle and higher
concentration of the range
It was sometimes also referred to as recovery is an indicator of the trueness of the test
measurements. To determine the accuracy of the method three quality control samples were used.
The samples chosen were such to represent the entire range of the standard curve i e lower,
middle and higher concentration of the range. In the present finding recovery studies were
conducted after addition of standard drug solution at three different levels .i.e. 50 %, 100 %, and
150 % to pre-analyzed sample solution105
. The results are given in table-3.A.9,3, 3.A.9,4 and
3.A.9,5
4.A.9.6. Robustness
Robustness is the ability of a method to tolerate variations of parameters without significant
change in the result In the present findings the percentage recovery of Atzanavir was good under
most conditions and didn’t show any significant change when the critical parameters were
y = 1933.x + 588.7
R² = 0.999
0
5000
10000
15000
20000
25000
30000
35000
0 2 4 6 8 10 12 14 16
333
modified. The tailing factor for Atzanavir was always less than 2.0 and it was well separated
under all the changes carried out. Considering the modifications in the system suitability
parameters and the specificity of the method it can conclude that the method is robust.
Robustness parameter flow plus,flow minus, Temperature plus and Temperature minus results
are given in table-3.A.9.6, 3.A.9.7 , 3.A.9.8 and 3.A.9.9.
4.A.9.7. Forced degradation :
In the present findings Different force degradation samples were analyzed and it was found that
the drug peaks in acid, alkali, peroxide, UV and photo degraded drug sample solutions have
passed the purity test. Purity angle for the selected drug components in all stress conditions were
found to be less than the threshold angle. This study confirms the specificity of the developed
method.The overall degradation in basic condition was found to be around 16.22% The peaks of
the degradation products were well resolved. Our findings are in good aggrement with the work
published earlier by Raymond et al 106-107
Typical chromatogram of degredable compound show
in figure-4.A.19.The degradation study thereby indicated that Atzanavir was stable to chemical
oxidation study, dry heat and alkali hydrolysis while it was highly susceptible to alkali
hydrolysis.
Figure- 4. A.28. A chromatogram of the Atazanavir alkali degradation.
334
4.B Discussion of Related substance method development and Validation
4.B.1 Related Substance method development and Validation of Carmustine
4.B.1.1. HPLC method development and optimization
In present investigation the main target for the development of chromatographic method was to
get the reliable method for the quantification of Carmustine from bulk drug and which will be
also applicable for the degradable products. The elution gradient and influence of mobile phase
pH were studied in order to optimize the analytical performance. A short analytical column and
elution gradient ,buffer and organic modifier were chose as the best compromise between
retention time of analyte. This suggests that the few HPLC and UV methods have been reported
for estimation of Carmustine which are sophisticated, but time consuming. The present study
was aimed at development of speedy and cost effective HPLC technique for determination of
Carmustine as bulk and in dosage forms. The chromatographic method was optimized by
changing various parameters, such as the mobile phase composition , pH of the buffer used in the
mobile phase. Retention time and separation of peak of Carmustine were dependent on pH of the
buffer and the percentage of methanol. Various blends of solvent systems in varying proportions
were tried as mobile phase by Kirtikar et al and others 108-109
. Different mobile phases were tried,
but satisfactory separation and good symmetrical peak were obtained with the mobile phases
0.01 M Potassium dihydrogen phosphate buffer PH = 3.2 with orthophosphuric acid as A and
Methanol as mobile phase B, gradient program mention in expermential details and column
temperature maintained at 30 °C. The Column used YMC ODS-A C18, 250 x 4.6 mm, 5µm
found to comparatively better and gave the graph with better gaussian shape at retention time
36.11min .
4.B.1.2. Method validation
The Method was validated based on the International Conference on Harmonization (ICH)
Guidelines 28-30
. The method validation parameters checked were specificity, linearity, accuracy,
precision, limit of detection, limits of quantitation and robustness. The analytical method was
validated as per ICH guidelines with respect to parameters such as linearity, precision, accuracy,
specificity ,Forced degradation and robustness
335
4.B.1.3. LOD and LOQ
Analytical LOD of a method may be defined as the lowest concentration in a sample that can be
detected, but not necessarily quantitated, under the stated experimental conditions. The limit of
detection is important for impurity tests..LOQ is the lowest concentration of analyte in a sample
that can be determined with acceptable precision and accuracy.lt is quoted as the concentration
yielding a signal-to-noise ratio of 1 0: 1 and is confirmed by analyzing a number of samples near
this value. In the present findings these data show that the method is sensitive for the
determination of carmustine . The LOD and LOQ were measured by using an equation and were
found to be 0.1ppm. (Table-3.B.1.1)
4.B.1.4. System precision:
Precision is the measure of the degree of repeatability of an analytical method under normal
operation and is normally expressed as the percent relative standard deviation for a statistically
significant number of samples. The two most common precision measures are 'repeatability' and
'reproducibility'.These are expression of two extreme measure of precision which can be
obtained.
In present investigation Standard solution of Carmustine was prepared as per testing procedure
and injected into the HPLC system in six replicates. The values of % relative standard deviation
1.94 (NMT 5.0%) for peak area obtained in six replicate injections were reported in Table-
3.B.1.2. In our finding the the low value of relative standard deviation attributes that the method
is robust.Similar results are also reported by different research during their study. Zuolian et al 110
thus showing that the equipment used for the study worked correctly for the developed
analytical method, and being highly repetitive .
336
Figure -4.B.1. A chromatogram of carmustine standard prepration
4.B.1.5. Linearity
Linearity is the ability of a method to elicit test results that are directly proportional to analyte
concentration within a given range. Range is the interval between the upper and lower levels of
analyte that have been demonstrated to be determined with precision, accuracy, and linearity
using the method as written. The accepted criteria for linearity is that the correlation coefficient
(R2) is not less than 0.999 for the least squares method of analysis of the line 111-112
In present investigation the linear study identifies a specified concentration range where analytes
response is linearly proportional to the concentration. The standard curve found to be linear over
the concentration range of 5.0-15.0 ppm. The equation of the standard curve relating the peak
area to the Carmustine concentration in this range was y = 4008x – 899 . The drug showed
good linearity in the range of 5.0-15.0 ppm with coefficient of correlation value 0.9996 for peak
area (Table-3.B.1.3). Therefore,in this thesis, developments of new chromatographic analytical
methods were established to solve this problem.
337
Figure -4.B.2. Linearity graph of Carmustine
Figure -4.B.3.Linearity graph of Impurity-A
y = 4008x 899
R² = 0.998
0
5000
10000
15000
20000
25000
30000
35000
40000
45000
50000
0 2 4 6 8 10 12 14 16
y = 2948x +2269
R² = 0.998
0
5000
10000
15000
20000
25000
30000
35000
40000
45000
50000
0 2 4 6 8 10 12 14 16
338
4.B.1.6. Accuracy
Accuracy of a method is expressed as percentage of analyte recovered by spiking samples in
placebo of the drug formulation . To document accuracy, a minimum of nine determinations over
a minimum of three concentration levels covering the specified range (for example, three
concentrations, three replicates for each) were collected. It was performed at 50, 100, and 150%
levels of label claim. The RSD of the replicates provides the analysis variation and gives an
indication of the precision of the test method.
In our present investigation to check the accuracy of the method, recovery studies were
conducted after addition of standard drug solution at three different levels i.e. 50 %, 100 %, and
150 % to pre-analyzed sample solution214
. The results are given in table-3.B.1.4.
4.B.1.7. Robustness
The assessment of the robustness of a method is not required yet by the ICH guidelines, but it
can be expected that in the near future it will become obligatory. The definition for
robustness/ruggedness applied is "The robustness/ruggedness of an analytical procedure is a
measure of its capacity to remain unaffected by small, but deliberate variations in method
parameters and provides an indication of its reliability during normal usage" .
In our present finding the percentage recovery of Carmustine was good under most conditions
and didn’t show any significant change when the critical parameters were modified. The tailing
factor for Carmustine was always less than 2.0 and it was well separated under all the changes
carried out. Considering the modifications in the system suitability parameters and the specificity
of the method it can conclude that the method is robust. Our findigs good aggrement with the
earlier reported work on other drugs by Yang et al 113
4.B.1.8. Forced degradation :
Specificity is the ability to assess unequivocally the analyte in the presence of components that
may be expected to be present, such as impurities, degradation products, and matrix components
114-116. It is a measure of the degree of interferences from such components, ensuring that a peak
response is due to a single component only. Specificity is measured and documented in a
separation by the resolution, plate count (efficiency), and tailing factor. Specificity of the current
method. In present investigation the forced degradation of Carmustine under different stress
condition such as acid, alkali , peroxide ,thermal and water according to ICH guideline shown in
339
table- 3.B.1.9. In present findings have shown that 23.19% degradation observed in 0.2N NaOH,
and 12.78 % degradation observed in 0.5 N HCl , UV degradation 10.78% and thermal
degradation 11.56 % were observed,the peaks of the degradation products were well resolved.
Typical chromatogram of degredable compound shown in figure-4.B.3, 4.B.4 4.B.5 .The
degradation study thereby indicated that Carmustine was stable to chemical oxidation study and
water while it was highly susceptible to alkali hydrolysis . Our findings are in good aggrement
with the work published earlier by Mahmood et al and others. 117-118
Figure -4.B.4. A chromatogram of the Carmustine alkali degradation
Figure -4.B.5. A chromatogram of the Carmustine acid degradation
340
Figure -4.B.6. A chromatogram of the Carmustine Thermal degradation
341
4.B.2. Related Substance method development and Validation of Irinotecan
Hydrochloride
4.B.2.1. HPLC method development and optimization
In our present work the elution gradient and influence of mobile phase pH were studied in order
to optimize the analytical performance. A short analytical column and elution gradient ,buffer
and organic modifier were chose as the best compromise between retention time of analyteThe
main target for the development of chromatographic method was to get the reliable method for
the quantification of Irinotecan hydrochloride from bulk drug and which will be also applicable
for the degradable products. We also observed a trend where Few HPLC and UV methods have
been reported for estimation of Irinotecan hydrochloride which are sophisticated, but time
consuming. The present study was aimed at development of speedy and cost effective HPLC
technique for determination of Irinotecan hydrochloride as bulk and in dosage forms. The
chromatographic method was optimized by changing various parameters, such as the mobile
phase composition , pH of the buffer used in the mobile phase. Retention time and separation of
peak of Irinotecan hydrochloride were dependent on pH of the buffer and the percentage of
methanol. Various blends of solvent systems in varying proportions were tried as mobile phase
by Ganjali et al and others 119-123
. Different mobile phases were tried, but satisfactory separation
and good symmetrical peak were obtained with the mobile phases 1.8 gm Potassium dihydrogen
phosphate in 1000 ml water and and adjust PH 3.3 with orthophosphuric acid as A and mixture of
600 ml of acetonitrile and 400ml of methanol as mobile phase B . The column temperature was
maintained at 30 °C, and Column used Water symmetry RP18 (250 x 4.6)mm , 5 found to
comparatively better and gave the graph with better Gaussian shape at retention time 13.20 min
.124
4.B.2.2. Method validation
The validation of analytical procedures is based on the four most common types of analytical
procedures:- Identification tests, Quantitative tests for impurities' content,Limit tests for the
control of impurities, Quantitative tests of the active moiety in samples of drug substance or drug
product or other selected component(s) in the drug product 20
In the present study the developed method was validated, as described below, for various
parameters like linearity and range, accuracy, precision, ruggedness, system suitability,
specificity, LOQ,and LOD.
342
4.B.2.3. LOD and LOQ
Limit of detection (LOD) is the lowest concentration of analyte in a sample that can be detected,
but not necessarily quantitated, under the stated experimental conditions .
In the present findings these data show that the method is sensitive for the determination of
Irinotecan hydrochloride . The LOD and LOQ were measured by using an equation and were
found to be 0.25ppm and 0.5ppm .(Table-3.B.2.1). Our findigs good agrrement with the earlier
reported work on other drugs by Dwyer et al. 125
4.B.2.4. System precision:
Precision is the measure of the degree of repeatability of an analytical method under normal
operation and is normally expressed as the RSD for a statistically significant number of samples.
RSD for the peak areas of the samples should not be greater than 1.5% .System suitability test as
an integral part of method development was used to ensure adequate performance of the
chromatographic system. Retention time (RT), number of theoretical plates (N) and tailing factor
(T). Standard solution of Irinotecan hydrochloride was prepared as per testing procedure and
injected into the HPLC system in six replicates. The values of % relative standard deviation of
Related Compound-A, Related Compound-B, 7-Desethyl Irinotecan Camptothecine, -Ethyl
Camptothecine and 7,11-Diethyl-10-hydroxycamptothecine are 1.61, 1.79, 0.50, 1.41 and 1.32
respectively. (NMT 5.0%) for peak area obtained in six replicate injections were reported in
Table-3.A.1.1. the present study reveals that that the equipment used for the study worked
correctly for the developed analytical method, and being highly repetitive 126
.
Figure -4.B.7. A chromatogram of the Irinotecan and impurity prepration.
343
4.B.2.5. Linearity
Linearity is the ability of a method to elicit test results that are directly proportional to analyte
concentration within a given range. Range is the interval between the upper and lower levels of
analyte that have been demonstrated to be determined with precision, accuracy, and linearity
using the method as written. The accepted criteria for linearity is that the correlation coefficient
(R2) is not less than 0.990 for the least squares method of analysis of the line.
127
In the present findings the standard curve found to be linear over the concentration range of 5.0-
15.0 ppm. The equation of the standard curve relating the peak area to the Irinotecan
Hydrochloride concentration in this range was y = 1190x – 885. The drug showed good linearity
in the range of 5-15 ppm with coefficient of correlation value of Related Compound-A, Related
Compound-B, 7-Desethyl Irinotecan Camptothecine, -Ethyl Camptothecine and 7,11-Diethyl-10-
hydroxycamptothecine are 0.9997, 0.9998, 0.9997, 0.9998, 0.9997 and 0.9998 resp.for peak area
(Table-3.B.2.3). Our findings good agrrement with the earlier reported work on other drugs. 128
Figure -4.B.8.Linearity graph of Irinotecan Hydrochloride
y = 1190.x 885
R² = 0.999
0
5000
10000
15000
20000
25000
30000
35000
40000
45000
50000
0 2 4 6 8 10 12 14 16
344
Figure -4.B.9.Linearity graph of Related Comp-A
Figure -4.B.10.Linearity graph of Related Comp-B
y = 982x 224
R² = 1.0
0
5000
10000
15000
20000
25000
30000
35000
40000
45000
50000
0 2 4 6 8 10 12 14 16
y = 1447.x 1075
R² = 0.999
0
5000
10000
15000
20000
25000
30000
35000
0 2 4 6 8 10 12 14 16
345
Figure -4.B.11.Linearity graph 7-Desethyl Irinotecan
Camptothecine
Figure -4.B.12.Linearity graph of 7-Ethyl Camptothecine
y = 1277.x 949
R² = 0.998
0
5000
10000
15000
20000
25000
30000
35000
40000
45000
50000
0 2 4 6 8 10 12 14 16
y = 1196x 273
R² = 0.999
0
5000
10000
15000
20000
25000
30000
35000
0 2 4 6 8 10 12 14 16
346
Figure -4.B.13.Linearity graph of 7,11-Diethyl-10-hydroxycamptothecine
4.B.2.6. Accuracy
Accuracy , sometimes also referred to as recovery is an indicator of the trueness of the test
measurements. In our present study to determine the accuracy of the method by taking three
samples were used. The samples chosen were such that they represent the entire range of the
standard curve i.e lower, middle and higher concentration of the range.
In present investigation three solutions were prepared for this study having a concentration of
0.1 mg/ml of Irinotecan hydrochloride and three different concentrations of Irinotecan
hydrochloride related compounds: 50%, 100%, and 150%. Low concentration of Irinotecan
hydrochloride related compounds relative to Irinotecan hydrochloride was employed to check if
this low concentration of Irinotecan hydrochloride related compounds can be recovered in the
presence of high concentration of Irinotecan hydrochloride Percentage recovery of Irinotecan
hydrochloride related compounds at these levels was found consistent as can be seen in table
3.B.2.4, 3.B.2.5, 3.B.2.6, 3.B.2.7 and 3.B.2.8.Our findings similar with other researchers. 129
y = 2913.x 94.77
R² = 0.999
0
5000
10000
15000
20000
25000
30000
35000
40000
45000
50000
0 2 4 6 8 10 12 14 16
347
4.B.2.7. Robustness
The robustness of the HPLC method was evaluated according to the variety of conditions such as
small changes in the percentage of mobile phase and buffer concentration, pH and flow rate of
the mobile phase .
In present investigation the percentage recovery of Irinotecan hydrochloride was good under
most conditions and should no significant change when the critical parameters were modified.
The tailing factor for Irinotecan hydrochloride was always less than 2.0 and it was well separated
under all the changes carried out. Considering the modifications in the system suitability
parameters and the specificity of the method it can conclude that the method is robust.130-131
4.B.2.8. Forced degradation :
In present investigation the forced degradation of Irinotecan hydrochloride under different stress
condition such as acid, alkali , peroxide ,thermal and water according to ICH guideline shown in
table- 3.A.1.10 . Our findings shown that 23.56% degradation observed in 0.01 N NaOH. The
peaks of the degradation products were well resolved. Typical chromatogram of degredable
compound show in figure-4.B.12. The degradation study thereby indicated that Irinotecan
hydrochloride was stable to chemical oxidation study, dry heat and alkali hydrolysis while it was
highly susceptible to acid hydrolysis . Our findings are in good aggrement with the work
published earlier by Amidzic et al and others.132-135
Figure -4.B.14. A chromatogram of the Irinotecan Hydrochloride alkali degradation
348
4.B.3. Related Substance method development and Validation of Cyclophosphamide
4.B.3.1. HPLC method development and optimization
In our present work the main target for the development of chromatographic method was to get
the reliable method for the quantification of Cyclophosphamide from bulk drug and which will
be also applicable for the degradable products. We also observed a trend where few HPLC and
UV methods have been reported for estimation of Cyclophosphamide which are sophisticated,
but time consuming. The present study was aimed at development of speedy and cost effective
HPLC technique for determination of Cyclophosphamide as bulk and in dosage forms. The
chromatographic method was optimized by changing various parameters, such as the mobile
phase composition , pH of the buffer used in the mobile phase. Retention time and separation of
peak of Cyclophosphamide were dependent on pH of the buffer and the percentage of methanol.
Various blends of solvent systems in varying proportions were tried as mobile phase by Danesi et
al 136-137
. Different mobile phases were tried, but satisfactory separation and good symmetrical
peak were obtained with the mobile phases 900 ml of water 100 ml acetonitrile as A and 300 ml
of water 700 ml acetonitrile as mobile phase B, gradient program mentioned in experimental
details and column temperature was maintained at 30 °C. The Column used YMC Pack ODS
C18, 250 x 4.6 mm, 5µm found to comparatively better and gave the graph with better gaussian
shape at retention time 24.96 min .138
4.B.3.2. Method validation
The analytical method was validated as per ICH guidelines with respect to parameters such as
linearity, precision, accuracy, specificity, limit of quantification [LOQ], limit of detection
[LOD],Forced degradation and robustness.
4.B.3.3. LOD and LOQ
The Quantitation limit of an individual analytical procedure is the lowest amount of analyte in a
sample which can be quantitatively determined with suitable precision and accuracy. The
Quantitation limit is a parameter of quantitative assays for low levels of compounds in sample
matrices and is used particularly for the determination of impurities or degradation products. The
detection limit of an individual analytical procedure is the lowest amount of analyte in a sample
which can be detected but not necessarily quantitated as an exact value.21
LOD and LOQ of the
drug were calculated using the following equations according to International Conference on
349
Harmonization (ICH) guidelines. 22
LOD = 3.3 × /S LOQ = 10 × /S where = the standard
deviation of the response and S = the slope of the regression equation .
In our present work these data show that the method is sensitive for the determination of
Cyclophosphamide . The LOD and LOQ were measured by using an equation and were found to
be 0.1ppm. (Table-3.B.3.1). Our findigs good aggrement with the earlier reported work on other
drugs by Vainchtein et al139
4.B.3.4. System precision:
"The precision of an analytical method is the degree of agreement among individual test results
obtained when the method is applied to multiple sampling of a homogenous sample. Precision is
a measure of the reproducibility of the whole analytical method (including sampling, sample
preparation and analysis) under normal operating circumstances. Precision is determined by
using the method to assay a sample for a sufficient number of times to obtain statistically valid
results . The precision is then expressed as the relative standard deviation. In the present
investigation Standard solution of Cyclophosphamide was prepared as per testing procedure and
injected into the HPLC system in six replicates. The values of % relative standard deviation of
Monochlorocyclophosphamide is 1.81 and Impurity-A is 2.27 (NMT 5.0%) for peak area
obtained in six replicate injections were reported in Table-3.B.3.2. the present study reveals that
the equipment used for the study worked correctly for the developed analytical method, and
being highly repetitive. 140-143
.
Figure -4.B.15. A chromatogram of the Cyclophosphamide and impurity
prepration.
350
4.B.3.5. Linearity
A linear study identifies a specified concentration range where analytes response is linearly
proportional to the concentration.In present investigation the standard curve found to be linear
over the concentration range of 5.0-15.0 ppm. The equation of the standard curve relating the
peak area to the Cyclophosphamide concentration in this range was y = 1378x +323 . The drug
showed good linearity in the range of 5.0-15.0 ppm with coefficient of correlation value for
Monochlorocyclophosphamide is 0.9998 and Impurity-A is 0.9998 (Table-3.B.3.3). Our
findings are good aggrement with the earlier reported work on other drugs by Ibrahim et al and
others. 144-147
Figure -4.B.16.Linearity graph of Cyclophosphamide
y = 1378x +323
R² = 0.999
0
5000
10000
15000
20000
25000
30000
35000
40000
45000
50000
0 2 4 6 8 10 12 14 16
351
Figure -4.B.17.Linearity graph of Monochlorocyclophosphamide
Figure -4.B.18.Linearity graph of Impurity-A
y = 1606x +497
R² = 0.999
0
5000
10000
15000
20000
25000
30000
35000
40000
45000
50000
0 2 4 6 8 10 12 14 16
y = 1743x + 639
R² = 0.999
0
5000
10000
15000
20000
25000
30000
35000
0 2 4 6 8 10 12 14 16
352
4.B.3.6. Accuracy
Accuracy is popularly used to describe the measure of exactness of an analytical method, or the
close of an agreement between the value, which is accepted as a conventional, true value or as an
accepted reference value, and the value found.It is properly a qualitative concept and the correct
term is 'bias'.The bias of a method is an expression of how close the mean of asset of results
(produced by the method) is to the true value. Bias is usually determined by study of relevant
reference materials or by spiking studies
To determine the accuracy of the method three quality control samples were used. The samples
chosen were such to represent the entire range of the standard curve i.e. lower, middle and higher
concentration of the range Our findings shown that recovery studies were conducted after
addition of standard drug solution at three different levels i.e. 50 %, 100 %, and 150 % to pre-
analyzed sample solution148-150
. The results are given in table-3.B.3.4 for
Monochlorocyclophosphamide and for Impurity-A in table-3.B.3.5.
4.B.3.7. Robustness
"The robustness of an analytical procedure is a measure of its capacity to remain unaffected by
small, but deliberate variations in method parameters and provides an indication of its reliability
during normal usage".
In present investigation the percentage recovery of Cyclophosphamide was good under most
conditions and didn’t showed no significant change when the critical parameters were modified.
The tailing factor for Cyclophosphamide was always less than 2.0 and well separated under all
the changes carried out. Considering the modifications in the system suitability parameters and
the specificity of the method it can conclude that the method is robust. Considering the
modifications in the system suitability parameters and the specificity of the method it can
conclude that the method is robust.151
4.B.3.8. Forced degradation :
In present investigation the forced degradation of Cyclophosphamide under different stress
condition such as acid, alkali , peroxide ,thermal and water according to ICH guideline are
reported in table- 3.B.3.10. Our results have shown that 21.44% degradation observed in 0.5 N
353
NaOH. Typical chromatogram of degredable compound show in figure-4.B.16. The peaks of the
degradation products were well resolved. The degradation study thereby indicated that
Cyclophosphamide was stable to chemical oxidation study, dry heat and acid hydrolysis while it
was highly susceptible to alkali hydrolysis .152-154
Figure -4.B.19. A chromatogram of the Cyclophosphamide acid degradation
Figure -4.B.20. A chromatogram of the Cyclophosphamide alkali degradation.
354
4.B.4. Related Substance method development and Validation of Thiamine
Hydrochloride
4.B.4.1. HPLC method development and optimization
Our present investigation the main target for the development of chromatographic method was
to get the reliable method for the quantification of Thiamine hydrochloride from bulk drug and
which will be also applicable for the degradable products. Few HPLC and UV methods have
been reported for estimation of Thiamine hydrochloride which are sophisticated, but time
consuming. The present study was aimed at development of speedy and cost effective HPLC
technique for determination of Thiamine hydrochloride as bulk and in dosage forms.155-157
The
chromatographic method was optimized by changing various parameters, such as the mobile
phase composition , pH of the buffer used in the mobile phase. Retention time and separation of
peak of Thiamine hydrochloride were dependent on pH of the buffer and the percentage of
methanol. Various blends of solvent systems in varying proportions were tried as mobile phase
by Rivory et al 158-160
. Different mobile phases were tried, but satisfactory separation and good
symmetrical peak were obtained with the mobile phases 0.1 M Ammonium acetate buffer PH =
4.5 with acetic acid , 800 ml buffer and 200 ml water as A and mobile phase-B as 200 ml of
water and 800 ml acetonitrile mixed filtered and degased and final mobile phase is a homogenus
mixture of 700 ml of A and 300 ml of B and column temperature was maintained at35 °C . The
Column used Xterra C18, 250 x 4.6 mm, 5µm found to comparatively better and gave the graph
with better Gaussian shape at retention time 16.27 min . Hence this present study investigates the
various abilities these shown in ICH Harmonised Tripartite Guideline. 161
4.B.4.2. Method validation
The objective of validation of an analytical procedure is to demonstrate that it is suitable for its
intended purpose. To meet current pharmaceutical regulatory guidelines i.e. USP , ICH and EP ,
a number of criteria such as specificity, linearity, precision, accuracy, sensitivity and robustness
must be investigated in order to validate analytical methods.
4.B.4.3. LOD and LOQ
Limit of detection (LOD) is the lowest concentration of analyte in a sample that can be detected,
but not necessarily quantitated, under the stated experimental conditions . LOD can be
determined by preparing a solution that is expected to produce a response that is approximately 3
to 10 times of the base line noise. The solution is injected three times, and the signal and the
355
noise for each injection are recorded. Each signal to noise ratio (S/N) is then calculated, and
averaged.
In the present study the LOD and LOQ were measured by using an equation were found to be
0.1ppm .(Table-3.B.4.1). Our findigs good aggrement with the earlier reported work on other
drugs by Sawada et al 162-164
4.B.4.4. System precision:
Precision was carefully tested ,System suitability test as an integral part of method development
was used to ensure adequate performance of the chromatographic system. Retention time (RT),
number of theoretical plates (N) and tailing factor (T) Standard solution of Thiamine
hydrochloride Hydrochloride was prepared as per testing procedure and injected into the HPLC
system in six replicates. The values of % relative standard deviation for Related Compound-B is
1.75 and Related Compound-C is 2.53 (NMT 5.0%) for peak area obtained in six replicate
injections were reported in Table-3.B.4.2. thus showing that the equipment used for the study
worked correctly for the developed analytical method, and being highly repetitive 165-166
.
Figure -4.B.21. A chromatogram of the Thiamine Hydrochloride and impurity
preparation.
356
4.B.4.5. Linearity
In present investigation linearity of the method was determined by constructing calibration
curves. Standard solutions of Thiamine at different concentrations level (50%, , 80%, 100%,
120% and 150%) were used for this purpose. Before injection of the solutions, the column was
equilibrated for at least 30 min with the mobile phase. Each measurement was carried out in six
replicates to verify the reproducibility of the detector response at each concentration level.167-168
A linear study identifies a specified concentration range where analytes response is linearly
proportional to the concentration. The standard curve found to be linear over the concentration
range of 5.0-15.0 ppm. The equation of the standard curve relating the peak area to the Thiamine
Hydrochloride concentration in this range was y = 3990x – 569 . In our investigation first time
reported that the drug showed good linearity in the range of 5.0-15.0 ppm with coefficient of
correlation value for Thiamine HCl is 0.9995, Related Compound-B is 0.9993 and Related
Compound-C is 0.9992 for peak area (Table-3.B.4.3).
Figure -4.B.22.Linearity graph of Thiamine Hydrochloride
y = 3990x 569
R² = 1.0
0
5000
10000
15000
20000
25000
30000
35000
40000
45000
50000
0 2 4 6 8 10 12 14 16
357
Figure -4.B.23.Linearity graph of Related compound-B
Figure -4.B.24.Linearity graph of Related compound-C
y = 2873x 101
R² = 0.999
0
5000
10000
15000
20000
25000
30000
35000
0 2 4 6 8 10 12 14 16
y = 2088x + 367
R² = 0.999
0
5000
10000
15000
20000
25000
30000
35000
0 2 4 6 8 10 12 14 16
358
4.B.4.6. Accuracy
Accuracy is generally assessed by analyzing a sample with known concentration and comparing
the measured value with the true value. The measured value was obtained by recovery
test.Accuracy , sometimes also referred to as recovery is an indicator of the trueness of the test
measurements. To determine the accuracy of the method three samples were used. The samples
chosen were such to represent the entire range of the standard curve i.e. lower, middle and higher
concentration of the range. In present investigation recovery studies were conducted after
addition of standard drug solution at three different levels i.e. 50 %, 100 %, and 150 % to pre-
analyzed sample solution. The results are given in table-3.B.4.4for Related Comp-B and table-
3.B.4.5 for Related Comp-C.169-172
.4.B.4.7. Robustness
Robustness is the ability of a method to tolerate variations of parameters without significant
change in the result. The robustness of the HPLC method was evaluated according to the variety
of conditions such as small changes in the percentage of mobile phase and buffer concentration,
pH and flow rate of the mobile phase . In present findings the percentage recovery of Thiamine
hydrochloride was good under most conditions and didn’t show any significant change when the
critical parameters were modified. The tailing factor for Thiamine hydrochloride was always less
than 2.0 and it was well separated under all the changes carried out. Considering the
modifications in the system suitability parameters and the specificity of the method it can
conclude that the method is robust.
4.B.4.8. Forced degradation :
Analytical specificity of a method may be defined as a measure of the degree of interferences
from such components, ensuring that a peak response is due to a single component only. . In
present investigation the forced degradation of Thiamine hydrochloride under different stress
condition such as acid, alkali , peroxide ,thermal and water according to ICH guidline shown in
table- 3.B.4.10. Our results have shown that 16.86% degradation observed in 0.5 N NaOH.
Typical chromatogram of degredable compound show in figure-4.B.21. The peaks of the
degradation products were well resolved. The degradation study thereby indicated that Thiamine
hydrochloride was stable to chemical oxidation study, dry heat and acid hydrolysis while it was
highly susceptible to alkali hydrolysis . findings are in good aggrement with the work published
earlier by Takasago et al 173-176
359
Figure -4.B.25. A chromatogram of the Thiamine Hydrochloride acid degradation
Figure -4.B.26. A chromatogram of the Thiamine Hydrochloride alkali degradation
360
4.B.5. Related Substance method development and Validation of Topotecan
Hydrochloride
4.B.5.1. HPLC method development and optimization
In present investigation the main target for the development of chromatographic method was to
get the reliable method for the quantification of Topotecan hydrochloride from bulk drug and
which will be also applicable for the degradable products. However few HPLC and UV methods
have been reported for estimation of Topotecan hydrochloride which are sophisticated, but time
consuming. The present study was aimed at development of speedy and cost effective HPLC
technique for determination of Topotecan hydrochloride as bulk and in dosage forms. The
chromatographic method was optimized by changing various parameters, such as the mobile
phase composition , pH of the buffer used in the mobile phase. Retention time and separation of
peak of Topotecan hydrochloride were dependent on pH of the buffer and the percentage of
methanol. Various blends of solvent systems in varying proportions were tried as mobile phase
by Faouzi et al 177-178
. Different mobile phases were tried, but satisfactory separation and good
symmetrical peak were obtained with the mobile phases 0.02M Potassium dihydrogen phosphate
buffer pH = 3.0 with orthophosphuric acid as A and 300 ml of water and 700 ml acetonitrile as
mobile phase B mixed filtred and degassed. The column temperature was maintained at 35°C ,
and Column used Inertsil ODS 3V C18(250X4.6)mm,5µ found to comparatively better and gave
the graph with better gaussian shape at retention time 20.37 min .
4.B.5.2. Method validation
The validation of analytical procedures is based on the four most common types of analytical
procedures:- Identification tests,Quantitative tests for impurities' content,Limit tests for the
control of impurities, Quantitative tests of the active moiety in samples of drug substance or drug
product or other selected component(s) in the drug product 20
In the present study the developed method was validated, as described below, for various
parameters like linearity and range, accuracy, precision, ruggedness, system suitability,
specificity, LOQ,and LOD.
361
4.B.5.3. LOD and LOQ
Limit of detection (LOD) is the lowest concentration of analyte in a sample that can be detected,
but not necessarily quantitated, under the stated experimental conditions . LOD can be
determined by preparing a solution that is expected to produce a response that is approximately 3
to 10 times of the base line noise.
In the present findings these data show that the method is sensitive for the determination of
Topotecan hydrochloride . The LOD and LOQ were measured by using an equation and were
found to be 0.1ppm .(Table-3.B.5.1). Our findigs good aggrement with the earlier reported work
on other drugs 179
4.B.5.4. System precision:
Precision is the measure of the degree of repeatability of an analytical method under normal
operation and is normally expressed as the percent relative standard deviation for a statistically
significant number of samples. The two most common precision measures are 'repeatability' and
'reproducibility'.These are expression of two extreme measure of precision which can be
obtained.
Our findings are Standard solution of Topotecan hydrochloride was prepared as per testing
procedure and injected into the HPLC system in six replicates. The values of % relative standard
deviation for Impurity-A is 1.74 and Impurity-B is 1.62 (NMT 5.0%) for peak area obtained in
six replicate injections were reported in Table-3.B.5.2. thus showing that the equipment used for
the study worked correctly for the developed analytical method, and being highly repetitive. In
our finding the low value of relative standard deviation attributes that the method is
robust.Similar results are also reported by different research during their study Draviam et al 180-
181.
362
Figure -4.B.27. A chromatogram of the Topotecan Hydrochloride Standard
prepration.
4.B.5.5. Linearity
Linearity is a highly successful parameter for method accuracy and is defined as the ability of a
method to elicit test results that are directly proportional to analyte concentration within a given
range. Range is the interval between the upper and lower levels of analyte that have been
demonstrated to be determined with precision, accuracy, and linearity using the method as
written. The accepted criteria for linearity is that the correlation coefficient (R2) is not less than
0.999 for the least squares method of analysis .In our finding the linear study identifies a
specified concentration range where analytes response is linearly proportional to the
concentration. The standard curve found to be linear over the concentration range of 5.0-15.0
ppm. The equation of the standard curve relating the peak area to the Topotecan Hydrochloride
concentration in this range was y = 2300x – 448
In present findings the drug showed good linearity in the range of 5.0-15.0 ppm with
coefficient of correlation value for Topotecan is 0.9999, Impurity-A is 0.9992 and Impurity-B is
0.9997 for peak area (Table-3.B.5.3)
363
Figure -4.B.28.Linearity graph of Topotecan Hydrochloride
Figure -4.B.29.Linearity graph of Impurity-A
y = 2300x + 448
R² = 0.999
0
5000
10000
15000
20000
25000
30000
35000
0 2 4 6 8 10 12 14 16
y = 1710x 291
R² = 0.999
0
5000
10000
15000
20000
25000
30000
35000
0 2 4 6 8 10 12 14 16
364
Figure -4.B.30.Linearity graph of Impurity-B
4.B.5.6. Accuracy
Accuracy of a method is expressed as percentage of analyte recovered by spiking samples in
drug formulation . The RSD of the replicates provides the analysis variation and gives an
indication of the precision of the test method. It was sometimes also referred to as recovery is an
indicator of the trueness of the test measurements. To determine the accuracy of the method three
quality control samples were used. The samples chosen were such to represent the entire range of
the standard curve i.e. lower, middle and higher concentration of the range
In present investigation to check the accuracy of the method, recovery studies were conducted
after addition of standard drug solution at three different levels i.e. 50 %, 100 %, and 150 % to
pre-analyzed sample solution. The results are given in table-3.B.5.4for Impurity-A and table-
3.B.5.5 for Impurity-B.182-183
y = 1602x 652
R² = 0.999
0
5000
10000
15000
20000
25000
30000
35000
0 2 4 6 8 10 12 14 16
365
4.B.5.7. Robustness :
The robustness of an analytical procedure is a measure of its capacity to remain unaffected by
small, but deliberate variations in method parameters and provides an indication of its reliability
during normal usage..The present study reveals that the method found to be robust, which may
be attributed to the small but deliberate changes in the method and low value of relative standard
deviation. These parameters have no detrimental effect on the method performance. Similar
results are also reported by different research during their study.
In the present findings the percentage recovery of Topotecan hydrochloride was good under most
conditions and didn’t show any significant change when the critical parameters were modified.
The tailing factor for Topotecan hydrochloride was always less than 2.0 and it was well separated
under all the changes carried out. Considering the modifications in the system suitability
parameters and the specificity of the method it can conclude that the method is robust.184
4.B.5.8. Forced degradation :
Selectivity is the ability to measure accurately and specifically the analyte in the presence of
components that may be expected to be present in the sample matrix. In our present investigation
Forced degradation study was carried out by subjecting the drug to stress condition such as acid,
alkali , peroxide ,thermal and water according to ICH guidline shown in table- 3.B.5.10. Our
findings shown that 18.69% degradation observed in 0.2N NaOH. Typical chromatogram of
degredable compound show in figure-4.B.26 .The peaks of the degradation products were well
resolved. The degradation study thereby indicated that Topotecan hydrochloride was stable to
chemical oxidation study, dry heat and acid hydrolysis while it was highly susceptible to alkali
hydrolysis .185-186
366
Figure -4.B.31. A chromatogram of the Topotecan HydrochlorideAcid degradation.
Figure -4.B.32. A chromatogram of the Topotecan Hydrochloride Alkali
degradation.
367
4.B.6. Related Substance method development and Validation of
TenofovirDisoproxil fumarate
4.B.6.1. HPLC method development and optimization
In present investigation the main target for the development of chromatographic method was to
get the reliable method for the quantification of TenofovirDisoproxil fumarate from bulk drug
and which will be also applicable for the degradable products. We also observed a trend where
few HPLC and UV methods have been reported for estimation of TenofovirDisoproxil fumarate
which are sophisticated, but time consuming. The present study was aimed at development of
speedy and cost effective HPLC technique for determination of TenofovirDisoproxil fumarate as
bulk and in dosage forms.265
The chromatographic method was optimized by changing various
parameters, such as the mobile phase composition , pH of the buffer used in the mobile phase.
Retention time and separation of peak of TenofovirDisoproxil fumarate were dependent on pH of
the buffer and the percentage of methanol. Various blends of solvent systems in varying
proportions were tried as mobile phase by Alderden-Los et al 187-189
. Different mobile phases
were tried, but satisfactory separation and good symmetrical peak were obtained with the mobile
phases 0.02M Potassium dihydrogen phosphate buffer pH = 3.0 with ortho-phosphuric acid as A
and 300 ml of water and 700 ml acetonitrile as mobile phase B mixed filtered and degassed.The
column temperature was maintained at 35 °C , and Column used YMC ODS -A
C18(150X4.6)mm,5µ found to comparatively better and gave the graph with better gaussian
shape at retention time 47.00 min .
4.B.6.2. Method validation
The analytical method was validated as per ICH guidelines with respect to parameters such as
linearity, precision, accuracy, specificity, limit of quantification [LOQ], limit of detection [LOD]
,Forced degradation and robustness.However,"The precision of an analytical method is the
degree of agreement among individual test results obtained when the method is applied to
multiple sampling of a homogenous sample. Precision is a measure of the reproducibility of the
whole analytical method (including sampling, sample preparation and analysis) under normal
operating circumstances. Precision is determined by using the method to assay a sample for a
sufficient number of times to obtain statistically valid results . The precision is then expressed as
the relative standard deviation.
368
4.B.6.3. LOD and LOQ
Analytical LOD and LOQ of a method may be defined as the lowest concentration in a sample
that can be detected, but not necessarily quantitated, under the stated experimental conditions.
The limit of detection is important for impurity tests. In the present findings,these data show that
the method is sensitive for the determination of TenofovirDisoproxil fumarate . The LOD and
LOQ were measured by using an equation and were found to be 0.1ppm .(Table-3.B.6.1). Our
findings are in good aggrement with the work published earlier of Weller et al.190-191
4.B.6.4. System precision:
In our present investigation Standard solution of TenofovirDisoproxil fumarate was prepared as
per testing procedure and injected into the HPLC system in six replicates. The values of %
relative standard deviation for Mono poc PMPA is 1.58, Moc poc PMPA is 2.00 , Ipr poc pmpa is
1.16 , Dec poc PMPA is 0.44 , n-poc poc PMPA is 0.80 , Mixed Dimer is 0.50 and Dimer is 0.70
(NMT 5.0%) for peak area obtained in six replicate injections were reported in Table-3.B.6.2.In
our finding the low value of relative standard deviation attributes that the method is
robust.Similar results are also reported by different research during their study. thus showing that
the equipment used for the study worked correctly for the developed analytical method, and
being highly repetitive 192-193
.
Figure -4.B.33. A chromatogram of Tenofovir Disoproxil Fumarate Sample
4.B.6.5. Linearity
Linearity is the ability of a method to elicit test results that are directly proportional to analyte
concentration within a given range. Range is the interval between the upper and lower levels of
analyte that have been demonstrated to be determined with precision, accuracy, and linearity
using the method as written. The accepted criteria for linearity is that the correlation coefficient
369
(R2) is not less than 0.990 for the least squares method of analysis. In present investigation a
linear study identifies a specified concentration range where analytes response is linearly
proportional to the concentration. The standard curve found to be linear over the concentration
range of 5.0-15.0 ppm. The equation of the standard curve relating the peak area to the Tenofovir
Disoproxil Fumarate concentration in this range was y = 3931x – 591 . Our findings are in good
aggrement with the work published earlier by Rossella et al 194
In present investigation the drug shown good linearity in the range of 5.0-15 ppm with
coefficient of correlation value for Tenofovir is 0.9998 Mono poc PMPA is 0.9993, Moc poc
PMPA is 0.9992, Ipr poc pmpa is 0.9993, Dec poc PMPA is 0.9993, n-poc poc PMPA is 0.9992,
Mixed Dimer is 0.9999 and Dimer is 0.9999 for peak area (Table-3.B.6.3, 3.B.6.4)
Figure -4.B.34.Linearity graph of Tenofovir Disoproxil Fumarate
y = 3931x 591
R² = 0.998
0
5000
10000
15000
20000
25000
30000
35000
40000
45000
50000
0 2 4 6 8 10 12 14 16
370
Figure -4.B.35.Linearity graph of Mono
poc PMPA
Figure -4.B.36.Linearity graph of Mono poc PMPA
y = 1798x +519
R² = 0.999
0
5000
10000
15000
20000
25000
30000
35000
40000
45000
50000
0 2 4 6 8 10 12 14 16
y = 1961x + 414
R² = 0.999
0
5000
10000
15000
20000
25000
30000
35000
0 2 4 6 8 10 12 14 16
371
Figure -4.B.37.Linearity graph of Ipr poc pmpa
Figure -4.B.38.Linearity graph of Dec poc PMPA
y = 2330x +573
R² = 0.999
0
5000
10000
15000
20000
25000
30000
35000
40000
45000
50000
0 2 4 6 8 10 12 14 16
y = 1645x + 411
R² = 0.999
0
5000
10000
15000
20000
25000
30000
35000
0 2 4 6 8 10 12 14 16
372
Figure -4.B.39.Linearity graph of n-poc poc PMPA
Figure -4.B.40.Linearity graph of Mixed Dimer
y = 1402x +214
R² = 0.998
0
5000
10000
15000
20000
25000
30000
35000
40000
45000
50000
0 2 4 6 8 10 12 14 16
y = 1678x +168
R² = 0.998
0
5000
10000
15000
20000
25000
30000
35000
40000
45000
50000
0 2 4 6 8 10 12 14 16
373
Figure -4.B.41.Linearity graph of Dimer
4.B.6.6. Accuracy
Accuracy indicates the deviation between the mean value found and the true value. lt is
determined by applying the method to samples to which known amounts of analyte have been
added. These should be analysed against standard and blank solutions to ensure that no
interference exists. To determine the accuracy of the method three samples were used. The
samples chosen were such to represent the entire range of the standard curve i.e. lower, middle
and higher concentration of the range in present study investigates the various abilities to check
the accuracy of the method, recovery studies were conducted after addition of standard drug
solution at three different levels i.e. 50 %, 100 %, and 150 % to pre-analyzed sample solution[5-
8]. The results are given in table-3.B.6.5for Mono poc PMPA , table-3.B.6.6for Ipr poc pmpa,
table-3.B.6.7for Dec poc PMPA, table-3.B.6.8for n-poc poc PMPA, table-3.B.6.9 for Mixed
Dimer and table-3.B.6.10for Dimer. 195
y = 1585x 238
R² = 0.999
0
5000
10000
15000
20000
25000
30000
35000
40000
45000
50000
0 2 4 6 8 10 12 14 16
374
4.B.6.7. Robustness
In the present findings the percentage recovery of Tenofovir Disoproxil fumarate was good under
most conditions and didn’t show any significant change when the critical parameters were
modified. The tailing factor for TenofovirDisoproxil fumarate was always less than 2.0 and it
was well separated under all the changes carried out. Considering the modifications in the system
suitability parameters and the specificity of the method it can conclude that the method is
robust.196
4.B.6.8. Forced degradation :
Purity angle for the selected drug components in all stress conditions were found to be less than
the threshold angle. In our present investigation the forced degradation of TenofovirDisoproxil
fumarate under different stress condition such as acid, alkali , peroxide ,thermal and water
according to ICH guidline shown in table- 3.B.6.14. Our results have shown that15.74%
degradation observed in 0.5N NaOH and 10.89% degradation observed in UV Light degradation.
Typical chromatogram of degredable compound show in figure-4.B.36. The peaks of the
degradation products were well resolved. The degradation study thereby indicated that
TenofovirDisoproxil fumarate was stable to chemical oxidation study, dry heat and acid
hydrolysis while it was highly susceptible to alkali hydrolysis. Our findings are in good
aggrement with the work published earlier by Vainchtein et al and others.197-200
Figure -4.B.42. A chromatogram of the Tenofovir Disoproxil Fumarate acid
degradation
375
Figure -4.B.43. A chromatogram of the Tenofovir Disoproxil Fumarate Peroxide
degradation
Figure -4.B.44. A chromatogram of the Tenofovir Disoproxil Fumarate alkali
degradation.
376
Figure -4.B.45. A chromatogram of the Tenofovir Disoproxil Fumarate Thermal
degradation.
Figure -4.B.46. A chromatogram of the Tenofovir Disoproxil Fumarate Photolytic
degradation.
377
4.B.7. Related Substance method development and Validation of Atazanavir
4.B.7.1. HPLC method development and optimization
In present investigation the main target for the development of chromatographic method was to
get the reliable method for the quantification of Atazanavir from bulk drug and which will be
also applicable for the degradable products. Few HPLC and UV methods have been reported for
estimation of Atazanavir which are sophisticated, but time consuming.201
The present study was
aimed at development of speedy and cost effective HPLC technique for determination of
Atazanavir as bulk and in dosage forms. The chromatographic method was optimized by
changing various parameters, such as the mobile phase composition , pH of the buffer used in the
mobile phase. Retention time and separation of peak of Atzanavir were dependent on pH of the
buffer and the percentage of methanol. Various blends of solvent systems in varying proportions
were tried as mobile phase by D'Avolio et al 202-204
. Different mobile phases were tried, but
satisfactory separation and good symmetrical peak were obtained with the mobile phases 0.01M
Potassium dihydrogen phosphate buffer pH = 3.2 with orthophosphuric acid as A and acetonitrile
as mobile phase B mixed filtered and degased. The Column used YMC ODS-A(150X4.6) mm,
3.5 µ , column temperature 25 °C found to comparatively better and gave the graph with better
Gaussian shape at retention time 19.79 min .
4.B.7.2. Method validation
Method validation of analytical procedures is based on the four most common types of analytical
procedures:- Identification tests,Quantitative tests for impurities' content,Limit tests for the
control of impurities, Quantitative tests of the active moiety in samples of drug substance or drug
product or other selected component(s) in the drug product.
The analytical method was validated as per ICH guidelines with respect to parameters such as
linearity, precision, accuracy, specificity, forced degradation and robustness.
4.B.7.3. LOD and LOQ
Limit of detection (LOD) is the lowest concentration of analyte in a sample that can be detected,
but not necessarily quantitated, under the stated experimental conditions . LOD can be
determined by preparing a solution that is expected to produce a response that is approximately 3
to 10 times of the base line noise. The solution is injected three times, and the signal and the
378
noise for each injection are recorded. Each signal to noise ratio (S/N) is then calculated, and
averaged. The concentration of the solution is used for determination of the detection limit if the
average S/N ratio is between 3 and 10. If it is not between 3 and 10, the solution concentration is
modified as necessary and the experiment is repeated. LOD may be expressed as: LOD = 3.3 /S
Where is the standard deviation of the response, and S is the slope of the calibration curve.
Limit of quantitation can be determined in the same manner but using the formula 10 /S.
In present investigation these data show that the method is sensitive for the determination of
Atzanavir . The LOD and LOQ were measured by using an equation and were found to be
0.1ppm .(Table-3.B.7.1).Our findigs good aggrement with the earlier reported work on other
drugs by Zhou et al 205-206
4.B.7.4. System precision:
Precision is a measure of the reproducibility of the whole analytical method (including sampling,
sample preparation and analysis) under normal operating circumstances. Precision is determined
by using the method to assay a sample for a sufficient number of times to obtain statistically
valid results (ie between 6 - 1 0). Precision is determined by using the method to assay a sample
for a sufficient number of times to obtain statistically valid results . The precision is then
expressed as the relative standard deviation as
std dev x 100%
%RSD = --------------------
mean
Standard solution of Atazanavir was prepared as per testing procedure and injected into the
HPLC system in six replicates. The values of % relative standard deviation is 1.24 (NMT 2.0%)
for peak area obtained in six replicate injections were reported in Table-3.B.7.2. In our finding
the the low value of relative standard deviation attributes that the method is robust.Similar results
are also reported by different research during their study by Carpenter et al and others 208-210
379
Figure -4.B.47. A chromatogram of the Atazanavir Sample degradation sample
4.B.7.5. Linearity
Linearity is the ability of a method to elicit test results that are directly proportional to analyte
concentration within a given range. Range is the interval between the upper and lower levels of
analyte that have been demonstrated to be determined with precision, accuracy, and linearity
using the method as written. The accepted criteria for linearity is that the correlation coefficient
(R2) is not less than 0.990 for the least squares method of analysis of the line
211
In present investigation a linear study identifies a specified concentration range where analytes
response is linearly proportional to the concentration. The standard curve found to be linear over
the concentration range of 5.0-15.0 ppm. The equation of the standard curve relating the peak
area to the Atazanavir concentration in this range was y = 2856x – 265 . Our findigs good
aggrement with the earlier reported work on other drugs by McCormick et al and others.212-213
The drug showed good linearity in the range of 5.0-15.0 ppm with coefficient of correlation
value 0.9999 for peak area of Atzanavir and 0.9997 for peak area of Impurity-A (Table-3.B.7.3)
380
Figure -4.B.48.Linearity graph of Atzanavir
Figure -4.B.49.Linearity graph of Impurity-A
y = 2856x +265
R² = 0.999
0
5000
10000
15000
20000
25000
30000
35000
40000
45000
50000
0 2 4 6 8 10 12 14 16
y = 1934x + 592
R² = 0.999
0
5000
10000
15000
20000
25000
30000
35000
0 2 4 6 8 10 12 14 16
381
4.B.7.6. Accuracy
Accuracy , sometimes also referred to as recovery is an indicator of the trueness of the test
measurements. To determine the accuracy of the method three quality control samples were used.
The samples chosen were such to represent the entire range of the standard curve i.e. lower,
middle and higher concentration of the range. In our findings the accuracy of the method,
recovery studies were conducted after addition of standard drug solution at three different levels
i.e. 50 %, 100 %, and 150 % to pre-analyzed sample solution280-284
. The results are given in
table-3.B.7.4.Hence this present study investigates the various abilities.214-216
4.B.7.7. Robustness
In the present findings the percentage recovery of Atazanavir was good under most conditions
and didn’t show any significant change when the critical parameters were modified. The tailing
factor for Atzanavir was always less than 2.0 and it was well separated under all the changes
carried out. Considering the modifications in the system suitability parameters and the specificity
of the method it can conclude that the method is robust.217
4.B.7.8. Forced degradation :
In our present investigation Forced degradation study was carried out by subjecting the drug to
different stress condition such as acid, alkali , peroxide ,thermal and water according to ICH
guidline shown in table- 3.B.7.9.Our results have shown that 12.86% degradation observed in 0.5
N NaOH. Typical chromatogram of degredable compound show in figure-4.B.43 The peaks of
the degradation products were well resolved. The degradation study thereby indicated that
Atzanavir was stable to chemical oxidation study, dry heat and acid hydrolysis while it was
highly susceptible to alkali hydrolysis .218-219
382
Figure -4.B.50. A chromatogram of the Atazanavir Alkali degradation sample
383
4.B.8. Enantiomer content method Development and validation of Irinotecan
Hydrochloride
4.B.8.1. HPLC method development and optimization
In present investigation the main target for the development of chromatographic method was to
get the reliable method for the quantification of Irinotecan HCl from bulk drug. The elution
gradient and influence of mobile phase were studied in order to optimize the analytical
performance. A short analytical column and isocratic and organic modifier were chose as the best
compromise between retention time of analyte.Method development of Irinotecan HCl
enantiomer content also done by Sweetman et al.220
For separating conditions HPLC as an
effective tool. In HPLC three important variables influencing separation are the organic modifier
, basic modifier and Temperature.As a means of quantifying resolution, the peak capacity of the
separation was determined. 221-222
Enantiomer content method development of Irinotecan HCl
from bulk drug and which is applicable for accurate quantification. Initially, the effort for the
development of Irinotecan HCl from bulk drug HPLC method of R and S isomer seperation. For
this purpose, we have used Chiralpak ADH, 250 x 4.6 mm, 5µm and Chiralcel, 250 x 4.6 mm,
5µm. Out of these used HPLC column, Chiralcel, 250 x 4.6 mm, 5µm found to comparatively
better and gave the graph with better gaussian shape at retention time for R and S isomer is 10.64
and 15.22 min. To improve the shape and width of the graph, for the above columns different
solvent used such as n-hexane and IPA(70:30),n-hexane and IPA(68:32),.Finally tried for459 ml
n-hexane and 505 ml ethanol mixed and degased and column temperature was maintained at30
°C gave better peak shape. Method developed for Irinotecan HCl accurate short. and
enantioselective method.
4.B.8.2. Method validation
The objective of validation of an analytical procedure is to demonstrate that it is suitable for its
intended purpose. To meet current pharmaceutical regulatory guidelines i.e. USP , ICH and EP ,
a number of criteria such as specificity, linearity, precision, accuracy, sensitivity and robustness
must be investigated in order to validate analytical methods.
384
4.B.8.3. LOD and LOQ
This is the lowest concentration in a sample that can be detected, but not necessarily
quantitated, under the stated experimental conditions. The limit of detection is important for
impurity tests. The limit of detection is generally quoted as the concentration yielding a signal-
to-noise ratio of 2:1 and is confirmed by analyzing a number of samples near this value using
the following equation. The signal-to-noise ratio is determined by:s = H/h Where H = height of
the peak corresponding to the component. h = absolute value of the largest noise fluctuation from
the baseline of the chromatogram of a blank solution.In our present finding, LOQ determination
at 0.1 ppm level (Table-3.B.8.1)and it should good agreement with Perry et al 223-224
4.B.8.4. Method Precision
The system suitability was determined by injecting racemic mixture containing equal quantity
of (R) and (S)-enantiomers. Since the enantiomers form a critical pair of peaks in the
chromatogram, the qualification criteria was resolution between two enantiomers, shown to be
not less than 2 and tailing factor should not exceed 1.5. "The precision of an analytical method
is the degree of agreement among individual test results obtained when the method is applied to
multiple sampling of a homogenous sample. Precision is a measure of the reproducibility of the
whole analytical method (including sampling, sample preparation and analysis) under normal
operating circumstances. Precision is determined by using the method to assay a sample for a
sufficient number of times to obtain statistically valid results . The precision is then expressed as
the relative standard deviation.
In present investigation Table-3.B.8.2 showed Method precision data in which % RSD of six
replicate were mentioned.In our finding the low value of relative standard deviation attributes
that the method is robust.Similar results are also reported by different research during their study
225-227
385
Figure -4.B.51. A chromatogram of the Irinotecan Hydrochloride System suitability
prepration
Figure -4.B.52. A chromatogram of the Irinotecan Hydrochloride standard
prepration
4.B.8.5. Linearity
In present investigation a linear study identifies a specified concentration range where analytes
response is linearly proportional to the concentration. The standard curve found to be linear over
the concentration range of 1000-3000 ppm. The equation of the standard curve relating the peak
area to the Irinotecan Hydrochloride concentration in this range was y = 2491x – 130 and y =
2773x + 360
Calibration curves showed high linearity over the concentration range with correlation
coefficient 0.999 (Table-3.B.8.3). In our finding the low value of relative standard deviation
386
attributes that the method is robust.Similar results are also reported by different research during
their study Shirkhedkar et al228-229
Figure -4.B.53. Linearity graph of Irinotecan Hydrochloride ( R- Isomer)
Figure -4.B.54. Linearity graph of Irinotecan Hydrochloride ( R- Isomer)
y = 2491.x 130
R² = 0.999
0
2000
4000
6000
8000
10000
12000
14000
16000
18000
20000
0 2 4 6 8 10 12 14 16
y = 2773x + 360
R² = 0.999
0
2000
4000
6000
8000
10000
12000
14000
16000
18000
20000
0 2 4 6 8 10 12 14 16
387
4.B.8.6. Accuracy
Analytical accuracy , sometimes also referred to as recovery is an indicator of the trueness of the
test measurements. To determine the accuracy of the method three quality control samples were
used. The samples chosen were such to represent the entire range of the standard curve i.e. lower,
middle and higher concentration of the range
In present investigation to check the accuracy of the method, recovery studies were conducted
after addition of standard drug solution at three different levels i.e. 50 %, 100 %, and 150 % to
pre-analyzed sample solution. The recovery was determined using impurity spiked with sample
at three different concentration level show in table-3.B.8.4 (82-101%).
4.B.8.7. Robustness
"The robustness/ruggedness of an analytical procedure is a measure of its capacity to remain
unaffected by small, but deliberate variations in method parameters and provides an indication of
its reliability during normal usage"
In the present findings the robustness study flow rate change,pH change result showen in
table. Finally both the temperature and the egree of organic modifier were explored as a possible
means of improving resolution. It was found that increasing the temperature did not improve
peak capacity.230
Figure -4.B.55. A chromatogram of the Irinotecan Hydrochloride sample
prepration.
388
4.B.9. Enantiomer content method Development and validation of Topotecan
Hydrochloride
In present investigation the main target for the development of chromatographic method was to
get the reliable method for the quantification of Topotecan HCl from bulk drug. The elution
isocratic and influence of mobile phase were studied in order to optimize the analytical
performance 231
. Analytical column and elution depand on organic modifier, acidic or basic
modifier ,were chose as the best compromise between retention time of analyte. For separating
conditions HPLC as an effective tool. In HPLC three important variables influencing separation
are the organic modifier and Temperature.As a means of quantifying resolution, the peak
capacity of the separation was determined. Method development of Topotecan HCl enantiomer
content done by Venkataramanan et al 232-233
Enantiomer content method development of
Topotecan HCl from bulk drug and which is applicable for accurate quantification. Initially, we
took the effort for the development of Topotecan HCl from bulk drug HPLC method of R and S
isomer seperation. For this purpose, we have used Chiralpak ADH, 250 x 4.6 mm, 5µm and
Chiralcel, 250 x 4.6 mm, 5µm. Out of these used HPLC column, Chiralpak ADH, 250 x 4.6 mm,
5µm found to comparatively better and gave the graph with better gaussian shape at retention
time for R and S isomer is 9.50 and 14.15 min. To improve the shape and width of the graph, for
the above columns different solvent used such as acetonitrile and methanol (62:38), acetonitrile
and methanol (65:35),.Finally tried for 700 mlMethanol, 300 ml acetonitrile and 2.0 ml diethyl
amine mixed and degassed. The column temperature was maintained at 30 °C gave better peak
shape. Method developed for Topotecan HCl accurate short. and enantioselective method.
4.B.9.1. Method validation
The objective of validation of an analytical procedure is to demonstrate that it is suitable for its
intended purpose. To meet current pharmaceutical regulatory guidelines i.e. USP , ICH and EP , a
number of criteria such as specificity, linearity, precision, accuracy, sensitivity and robustness
must be investigated in order to validate analytical methods.
4.B.9.2. LOD and LOQ
Limit of detection (LOD) is the lowest concentration of analyte in a sample that can be detected,
but not necessarily quantitated, under the stated experimental conditions. LOD can be
389
determined by preparing a solution that is expected to produce a response that is approximately 3
to 10 times of the base line noise.
In the present findings,the limits of detection (LOD) and LOQ were determined by analyzing
Working Standard .LOD and LOQ were found to be 0.1ppm .(Table-3.B.9.1).
4.B.9.3. Method Precision
"The precision of an analytical method is the degree of agreement among individual test
results obtained when the method is applied to multiple sampling of a homogenous sample.
Precision is a measure of the reproducibility of the whole analytical method (including sampling,
sample preparation and analysis) under normal operating circumstances. Precision is determined
by using the method to assay a sample for a sufficient number of times to obtain statistically
valid results . The precision is then expressed as the relative standard deviation.
In present investigation Standard solution of Topotecan was prepared as per testing procedure
and injected into the HPLC system in six replicates. The values of % relative standard deviation
of R-Isomer was 1.81 and S-Isomer was 2.27 (NMT 5.0%) for peak area obtained in six
replicate injections are reported in Table-3.B.9.2 showed Method precision data in which % RSD
of six replicate were mentioned.234
Figure -4.B.56. A chromatogram of the Topotecan Hydrochloride system suitability
solution.
390
4.B.9.4. Linearity
Linearity is the ability of a method to elicit test results that are directly proportional to analyte
concentration within a given range. In present study investigates a linear study identifies a
specified concentration range where analytes response is linearly proportional to the
concentration. The standard curve found to be linear over the concentration range of 1000-3000
ppm. The equation of the standard curve relating the peak area to the Topotecan Hydrochloride
concentration in this range was y = 1256x +220 and
y = 1715x +449. Calibration curves showed high linearity over the concentration range with
correlation coefficient 0.999 (Table-3.B.9.3). Our findings are in good aggrement with the work
published earlier by Delahunty et al 235
Figure -4.B.57. Linearity graph of Topotecan Hydrochloride ( R-Isomer)
y = 1256x + 220
R² = 0.999
0
2000
4000
6000
8000
10000
12000
14000
16000
18000
20000
0 2 4 6 8 10 12 14 16
391
Figure -4.B.58. Linearity graph of Topotecan Hydrochloride( S-Isomer)
4.B.9.5. Accuracy
Accuracy , sometimes also referred to as recovery is an indicator of the trueness of the test
measurements. To determine the accuracy of the method three quality control samples were used.
The samples chosen were such to represent the entire range of the standard curve i.e lower,
middle and higher concentration of the range
In present investigation the recovery was determined using impurity spiked with sample at
three different concentration level show in table-3.B.9.4 (83-101%).Our findigs good aggrement
with the earlier reported work on other drugs by Massaki et al and others 236-237
4.B.9.6. Robustness
The ICH guidelines also recommend that "one consequence of the evaluation of robustness
should be that a series of system suitability parameters (e.g. resolution tests) is established to
ensure that the validity of the analytical procedure is maintained whenever used".
In the present findings, robustness study flow rate change result showen in table-3.B.9.5.
Finally both the temperature and the organic modifier were explored as a possible means of
improving resolution. It was found that increasing the temperature did not improve peak
capacity.238
y = 1715x + 449
R² = 0.999
0
2000
4000
6000
8000
10000
12000
14000
16000
18000
20000
0 2 4 6 8 10 12 14 16
392
Figure -4.B.59. A chromatogram of the Topotecan Hydrochloride Sample
prepration.
393
4.B.10. Enantiomer content method Development and validation of
TenofovirDisoproxil fumarate
In present investigation the main target for the development of chromatographic method was to
get the reliable method for the quantification of Tenofovir Disoproxil fumarate from bulk drug.
The elution gradient and influence of mobile phase were studied in order to optimize the
analytical performance. A short analytical column and organic modifier were chose as the best
compromise between retention time of analyte. For separating conditions HPLC as an effective
tool. In HPLC three important variables influencing separation are the organic modifier and
Temperature.As a means of quantifying resolution, the peak capacity of the separation was
determined. Method development of TenofovirDisoproxil fumarate enantiomer content done by
Unnam et al.239-240
Enantiomer content method development of Tenofovir Disoproxil fumarate
from bulk drug and which is applicable for accurate quantification.308-312
Initially, the effort for
the development of TenofovirDisoproxil fumarate from bulk drug HPLC method of R and S
isomer seperation. For this purpose, we have used Chiralpak ADH, 250 x 4.6 mm, 5µm and
Chiralcel, 250 x 4.6 mm, 5µm. Out of these used HPLC column, Chiralpak AD-RH (150 X 4.6)
mm, 5!. found to comparatively better and gave the graph with better gaussian shape at
retention time for R and S isomer is 2.31 and 3.39 min. To improve the shape and width of the
graph, for the above columns different solvent used such as acetonitrile and methanol (45:55),
acetonitrile and methanol (50:50),.Finally tried for Methanol,acetonitrile and diethyl amine
(850:150:0.8,v/v/v) and column temperature was maintained at 35 °C have better peak shape.
Method developed for TenofovirDisoproxil fumarate accurate short. and enantioselective
method.Therefore,in this thesis, developments of new chromatographic analytical methods were
established to solve this problem.
4.B.10.1. Method validation
The objective of validation of an analytical procedure is to demonstrate that it is suitable for its
intended purpose. To meet current pharmaceutical regulatory guidelines i.e. USP , ICH and EP ,
a number of criteria such as specificity, linearity, precision, accuracy, sensitivity and robustness
must be investigated in order to validate analytical methods.
4.B.10.2. LOD and LOQ
Limit of detection (LOD) is the lowest concentration of analyte in a sample that can be detected,
but not necessarily quantitated, under the stated experimental conditions . LOD can be
394
determined by preparing a solution that is expected to produce a response that is approximately 3
to 10 times of the base line noise. The solution is injected three times, and the signal and the
noise for each injection are recorded. Each signal to noise ratio (S/N) is then calculated, and
averaged. The concentration of the solution is used for determination of the detection limit if the
average S/N ratio is between 3 and 10. If it is not between 3 and 10, the solution concentration is
modified as necessary and the experiment is repeated. LOD may be expressed as: LOD = 3.3 /S
Where is the standard deviation of the response, and S is the slope of the calibration curve.
Limit of quantitation can be determined in the same manner but using the formula 10 /S.
In present investigation the limits of detection (LOD) and LOQ were determined by analyzing
Working Standard .
LOD and LOQ were found to be 0.1 ppm .(Table-3.B.10.1).Similar findings were reported by
researchers in different drugs. 242-243
4.B.10.3.Method Precision
"The precision of an analytical method is the degree of agreement among individual test results
obtained when the method is applied to multiple sampling of a homogenous sample. Precision is
a measure of the reproducibility of the whole analytical method (including sampling, sample
preparation and analysis) under normal operating circumstances. Precision is determined by
using the method to assay a sample for a sufficient number of times to obtain statistically valid
results . The precision is then expressed as the relative standard deviation.
Table-3.B.10.2 showed Method precision data in which % RSD of Assay of six replicate were
mentioned.In our finding the low value of relative standard deviation attributes that the method is
robust.Similar results are also reported by different research during their study by Weller et al
and others.244-248
395
Figure -4.B.60. A chromatogram of the Tenofovir Disoproxil Fumarate System
suitability prepration.
4.B.10.4.Linearity
In the present findings a linear study identifies a specified concentration range where analytes
response is linearly proportional to the concentration. The standard curve found to be linear over
the concentration range of 1000-3000 ppm. The equation of the standard curve relating the peak
area to the Tenofovir Disoproxill Fumarate concentration in this range was y = 1430x +313 and
y = 1244x +302. 249-253
Calibration curves showed high linearity over the concentration range
with correlation coefficient 0.999 (Table-3.B.10.3).
Figure -4.B.61. Linearity graph of Tenofovir Disoproxill Fumarate( R-Isomer)
y = 1296x + 326
R² = 0.999
0
2000
4000
6000
8000
10000
12000
14000
16000
18000
20000
0 2 4 6 8 10 12 14 16
396
Figure -4.B.62. Linearity graph of Tenofovir Disoproxill Fumarate( S-Isomer)
4.B.10.5.Recovery
Accuracy , sometimes also referred to as recovery is an indicator of the trueness of the test
measurements. To determine the accuracy of the method three quality control samples were used.
The samples chosen were such to represent the entire range of the standard curve i.e. lower,
middle and higher concentration of the range
In present investigation the recovery was determined using impurity spiked with sample at
three different concentration level show in table-3.B.10.4 (82-101%).Our findigs good
aggrement with the earlier reported work on other drugs by Manogaokar et al. 254-255
4.B.10.6.Robustness
In our present work the robustness study flow rate change,pH change result showen in table.
Finally both the temperature and the egree of organic modifier were explored as a possible means
of improving resolution. It was found that increasing the temperature did not improve peak
capacity.256
y = 1387x + 352
R² = 0.999
0
2000
4000
6000
8000
10000
12000
14000
16000
18000
20000
0 2 4 6 8 10 12 14 16
397
Figure -4.B.63. A chromatogram of the Tenofovir Disoproxil Fumarate sample
prepration.
398
References:
1]. G.Srinubabu, B. Sudha Rani, J.V.L.N. Seshagiri Rao. Spectrophotometric Determination of
Ziprasidone in Pharmaceutical Formulations. E-Journal of Chemistry Vol. 3, No.10, pp 9-12,
January 2006.
2]. Brunton LL, Lazo JS, Parker KL, Goodman and Gilman’s the pharmacological basis of
therapeutics, 11th ed, McGraw-Hill, New York.
3]. B. Sudha Rani and P.Venkata Reddy. Estimation of Ziprasidone Hydrochloride Monohydrate
in Bulk and Capsules by Reverse Phase HPLC. E-Journal of Chemistry Vol. 3, No.12, pp 169-
172, July 2006.
4] Validation of compedial Assays-Guidelines'Pharmacopeial Convention, Rockvilie, MD,
1985.
5]. ICH guidelines on analytical method validation, In: proceedings of the international
conference of harmonization,Geneva, (1996). ICH Guidelines, Q2B,Analytical Method
Validation.
6] N. Mallikarjun Rao, et al, RP-HPLC method development and validation for estimation of
Voglibose in bulk and tablet dosage form. Int. J. Res. Pharm. Sci. 1(2), 190-194, (2010).
7]. N. Sreekanth, et al, RP-HPLC method development and validation of Rimonabant
Hydrochloride in bulk and pharmaceutical dosage form. International journal of pharmacy and
pharmaceutical science, 1(1), 199-205, 12 (2009).
8]. Y. Anand Kumar, M.Anitha, A.Hemanth, S. Srinivas. Development of rapid UV
spectrophotometric method for the estimation of Ziprasidone Hydrochloride in bulk and
formulations. Digest Journal of Nanomaterials and Biostructures Vol. 5, No 1, March 2010, p.
279 – 283
9].The Fitness for Purpose of Analytical Methods: A Laboratory Guide to Method Validation and
Related Topics (1998)
10] Jatin Khurana... A Thesis Submitted to the faculty of the Graduate School of the Creighton
University in partial fulfillment of the requirements for the Degree of Master of Science in the
Department of Pharmacy Sciences. Omaha, NE (December 2009).
11]. J. D. Chudasama, K. P. Channabasavaraj, C. B. Pandya, T. T. Mani. Development and
validation of a rapid RP-HPLC method for the estimation of Ziprasidone Hydrochloride
399
Monohydrate in bulk and its capsule dosage forms. International Journal of Pharmaceutical
Sciences Review and Research Volume 4, Issue 3, September – October 2010; Article 031.
12]. Zhang G, Terry AV, Bartlett MG, Simultaneous determination of five antipsychotic drugs in
rat plasma by high performance liquid chromatography with ultraviolet detection, J Chromatogr
B Analyt Technol Biomed Life Sci, 856, 2007, 20-8.
13] ICH - International Conference on Harmonisation. (2005). Topic Q2B, Validation of
Analytical Methods: Methodology. The Third International Conference on Harmonization of
Technical Requirements for Registration of Pharmaceuticals for Human Use Yokohama-Japan.
1.8: 71-76.
14] Vander, Y., Heyden, A., Nijhuis, J., Smeyers-Verbeke, B.G.M., Vandeginste, D.L. (2007).
Talanta. 71: 1424-1429.
15 ] Ahmed B. Eldin, Abdalla A. Shalaby and Magda Youssif El Maamly ,”Acta Pharmaceutica
Sciencia 52: 229-238 (2010) .
16] Kung-Tien Liu,Hang-Hsing Yang, Yi-Chih Hisa,An-Shoei Yang.. Journal of Food and Drug
Analysis, Vol. 16, No. 5, 2008, Pages 28-38.
17]. Suckow RF, Feina M, Correllb CU, Cooper TB, Determination of plasma ziprasidone using
liquid chromatography with fluorescence detection. Journal of Chromatography, 799(2), 2004,
201-8.
18]. O’Neil MJ.The Merck Index. 14th Ed., Merck & Co, Inc., Whitehouse Station, NJ, 2006;
3465.
19]. Turcotte JE, Debonnel G, de Montigny C, Hébert C, Blier P.
Assessment of the Serotonin and Norepinephrine Reuptake Blocking Properties of Duloxetine in
Healthy Subjects.Neuropsychopharmacology. 2001; 24: 511 521.doi:10.1038/sj.npp.1395628.
20]. ICH-Guidelines Q2A, Validation of Analytical Procedures: Definition and
terminology,Geneva, March 1995. (CPMP III/5626/94).
21 P.V.Pawar , P.D.Gaikwad,V.H.Bankar, S.P.Pawar International Journal of Pharma.Research &
Development IJPRD/2011/PUB/ARTI/VOV-3/ISSUE-1/MARCH/017 (Online-journal).
22] ICH-Topic Q1A (R2). Stability Testing of New Drug Substances and Products. Geneva;
August 2003. (CPMP/ICH/2736/99).
400
23]. Chalon SA, Granier LA, Vandenhende FR, Bieck PR, Bymaster FP, Joliat MJ, Hirth C,
Potter WZ.Duloxetine Increases Serotonin and Norepinephrine Availability in Healthy Subjects:
A Double-Blind,Controlled Study.Neuropsychopharmacology. 2003; 28: 1685.
24]. Mishra L.In: Drug Today.Lorina Publications Inc., Delhi, India, 2006; 1: 489.
25]. WHO-Stability Testing of Active Substances and Pharmaceutical Products. Geneva; April
2006. (QAS/06.179).
26] Lantz RJ, Gillespie TA, Rash TJ, Kuo F, Skinner M, Kuan HY, Knadler MP.
Metabolism, excretion and pharmacokinetics of duloxetine in healthy human subjects.Drug
Metab Disp. 2003; 31: 1142–1150.
27].Nataranjan S and Raman B, Asian J Chem, 2008, 20(3), 1833-1840
28 ICH, Q2B- Validation of Analytical Procedures: Methodology, International Conference on
Harmonization, Nov, 1996.
29]. International Conference on Harmonization, ICH Q1 A (R2); Stability Testing of New Drug
Substances and Products 2003.
30]. ICH Steering Committee (1996) International conference on harmonization (ICH) of
technical requirements for registration of pharmaceuticals for human use, validation of analytical
procedure methodology, Geneva.
31].S.Gopalakrishnana*, E.Vadivela, P.Krishnavenia and B.Jeyashreeb
Research Journal of Pharmaceutical, Biological and Chemical Sciences RJPBCS 1(4) October –
December 2010 Page No. 200-207
32].R.Shinde sachin, I Bhoir Suvarna, S.Pawar Namdev, B.Yadav Suman and M. Bhagwat
Ashok.E-Journal of Chemistry Vol.4 No.11,ppq-13,January 2007
33 ].Susan Budavari, Maryadele J O’ Neil, Ann Smith, Patricia E, Heckelman, Joanne F and
Kinneary, The Merck Index, Merck Research Laboratories, 12th
Edn., 1994,1103
34 ]. ICH-Guidelines Q2B, Validation of Analytical Procedures: Methodology.Geneva,
November 1996. (CPMP/ICH/281/95)
35]. Nash RA, Wachter AH, A Pharmaceutical Process Validation, 3rd edn,2003,480-483
36].Susan Budavari, Maryadele J O’ Neil, Ann Smith, Patricia E, Heckelman, Joanne F and
Kinneary, The Merck Index, Merck Research Laboratories, 12th
Edn., 1994,1103
37].Rajeshwari, Sankar K R, Rao G G and Raju A L, Asian J Chem, 2005,17,1259-1263
38].Ramkrishna N V S, J Pharm Biomed Anal, 2005,39, 1006-1013
401
39]. Anderson PM, Schroeder G, Skubitz KM. Oral glutamine reduces the duration and severity
of stomatitis after cytotoxic cancer chemotherapy. Cancer 1998;83:1433–9.
40].Thevis M, Opfermann G and Schanzer w, Biomed Chromatogr.,2001,15,393-402
41]. A. Sparreboom, P. de Bruijn, M. J. A. de Jonge, W. J. Loos, G. Stoter, J. Verweij and K.
Nooter, J.Chromatogr. B Biomed. Sci. Appl. 712 (1998) 225
42]. P. de Bruijn, J. Verweij, W. J. Loos, K. Nooter, G. Stoter, and A. Sparreboom, J. Chromatogr.
B Biomed. Sci. Appl. 698 (1997) 277
43]. available online: http://en.wikipedia.org/wiki/Irinotecan
44]. V. M. M. Herben, J. H. M. Schellens, M. Swart, G. Gruia, L. Vernillet, J. H. Beijnen, and W.
W.ten B. Huinink, J. Liq. Chromatogr. Relat. Technol. 21 (1998) 1541
45]. A. Sparreboom, P. de Bruijn, M. J. A. de Jonge, W. J. Loos, G. Stoter, J. Verweij and K.
Nooter, J.Chromatogr. B, Biomed. Sci. Appl. 712 (1998) 225
46]. M. R. Ganjali, M. Hariri, S. Riahi, P. Norouzi, and M. Javaheri, Int. J. Electrochem. Sci. 4
(2009)295
47]. S. Sawada, S. Okajima, R. Aiyama, K. Nokota, T. Furuta, T. Yokokura, E. Sugini, K.
Yamaguchi, and K. Miyasaka, Chem. Pharm. Bull., 39, 1446 (1991)
48]. Y. Kawato, M. Aonuma, Y. Hirota, H. Kuga, and K. Sato, Cancer Res., 51, 4187(1991)
49]. G. Nishimura, T. Satou, Y. Yoshimitsu, Y. Kurosaka, and T. Fujimura, Jpn. J. Cancer
Chemother., 22, 93 (1995)
50]. Y. Ohe, Y. Sasaki, T. Shinkai, K. Eguchi, T. Tamura, A. Kojima, H. Kunikane,
H. Okamoto, A. Karato, H. Ohmatsu, F. Kanzawa, and N. Saijo, J. Natl. Cancer
Inst. 84, 972 (2004)
51]. K. Takasago, Y. Kasai, Y. Kitano, K. Mori, K. Kakihata, M. Hirohashi, and
M. Nonura, Folia Pharmacol. Jpn., 105, 447 (1995)
52]. A. Kono and Y. Hara, Jpn. J. Cancer Chemother., 18, 2175 (1991)
53]. R. Mick, E. Gupta, E Vokes, and M. Ratain, J. Clin. Oncol., 14, 2012 (1996)
54]. H. Sumoyoshi, Y. Fujiwara, T. Ohune, N. Yamaoka, K. Tamura, and M. Yamakido,
J. Chromatogr. B, 670, 309 (1995)
55]Alsarra IA, Alarifi MN (2004). Validated liquid chromatographic determination of 5-
fluorouracil in human plasma. J. Chromatogr B.,804: 435–439.
402
56]Arayne MS, Sultana N, Zuberi MH, Siddiqui FA (2010). Simultaneous Determination of
Metformin, Cimetidine, Famotidine, and Ranitidine in Human Serum and Dosage Formulations
Using HPLC with UV Detection. J. Chromatogr. Sci., 48: 721-725.
57]Bhaskar M, L’Aurelle J, Bei N, Dolly P, Brock M, Pamala J, Natalia T (2010). Quantitative
High-Performance Liquid Chromatography Electrospray Ionization Tandem Mass Spectrometry
Analysis of Bis-N7-Guanine DNA DNA Cross-Links in White Blood Cells of Cancer Patients
Receiving Cyclophosphamide Therapy. Anal Chem., 82: 3650–3658
58]Casale F, Canaparo R, Muntoni E, Serpe L, Zara GP, Della Pepa C,Berno E, Costa M, Eandi
M (2002). Simultaneous HPLC determination of 5-fluorouracil and its metabolites in plasma of
cancer
patients. Biomed. Chromatogr. 16: 446-452.
59]Ciccolini J, Mercier C, Blachon MF, Favre R, Durand AA, Lacarelle B(2004). A simple and
rapid high-performance liquid chromatographic (HPLC) method for 5-fluorouracil (5-FU) assay
in plasma and possible detection of patients with impaired dihydropyrimidine dehydrogenase
(DPD) activity. J. Clin. Pharm. Therap., 29: 307-315.
60]Compagnon EL, Thiberville N, Moore C, Thudlez C, Lacroix (1996).Simple high-
performance liquid chromatographic method for the quantitation of 5-fluorouracil in human
plasma. J. Chromatogr. B.,
677: 380-383.
61]Dafeng C, Jingkai G, Wanhui L, Paul FJ, Qingguang D (2003). Sensitive liquid
chromatographic assay for the simultaneous determination of 5-fluorouracil and its prodrug,
tegafur, in beagle dog plasma. J.Chromatogr. B., 795: 377-382.
62]. Mahmood Ahmad, Muhammad Usman, Asadullah Madni, M. Zubair, Qamar-uz-Zaman, M.
Saleem Qureshi, Abubakar Munir, Mahtab Ahmad and Arshad Mahmood A fast and simple
HPLC-UV method for simultaneous determination of three anti-cancer agents in plasma of breast
cancer patients and its application to clinical pharmacokinetics, African Journal of Pharmacy and
Pharmacology Vol. 5(7), pp. 915 - 922, July 2011
63].Alsarra IA, Alarifi MN (2004). Validated liquid chromatographic determination of 5-
fluorouracil in human plasma. J. Chromatogr B.,804: 435–439.
403
64].Arayne MS, Sultana N, Zuberi MH, Siddiqui FA (2010). Simultaneous Determination of
Metformin, Cimetidine, Famotidine, and Ranitidine in Human Serum and Dosage Formulations
Using HPLC with UV Detection. J. Chromatogr. Sci., 48: 721-725.
65].Bhaskar M, L’Aurelle J, Bei N, Dolly P, Brock M, Pamala J, Natalia T (2010). Quantitative
High-Performance Liquid Chromatography Electrospray Ionization Tandem Mass Spectrometry
Analysis of Bis-N7-Guanine DNA DNA Cross-Links in White Blood Cells of Cancer Patients
Receiving Cyclophosphamide Therapy. Anal Chem., 82: 3650–3658
66].Casale F, Canaparo R, Muntoni E, Serpe L, Zara GP, Della Pepa C,Berno E, Costa M, Eandi
M (2002). Simultaneous HPLC determination of 5-fluorouracil and its metabolites in plasma of
cancer patients. Biomed. Chromatogr. 16: 446-452.
67].Ciccolini J, Mercier C, Blachon MF, Favre R, Durand AA, Lacarelle B (2004). A simple and
rapid high-performance liquid chromatographic (HPLC) method for 5-fluorouracil (5-FU) assay
in plasma and possible detection of patients with impaired dihydropyrimidine dehydrogenase
(DPD) activity. J. Clin. Pharm. Therap., 29: 307-315.
68].Compagnon EL, Thiberville N, Moore C, Thudlez C, Lacroix (1996). Simple high-
performance liquid chromatographic method for the quantitation of 5-fluorouracil in human
plasma. J. Chromatogr. B.,677: 380-383.
69]. Thomas S, Kumar R, Sharma A, Issarani R, Prakash B, Staility-indicating HPLC method
for determination of vitamins B1, B2, B3, and B6 in pharmaceutical dosage form, Indian J.
O. Chem. Technol. 2008;15:598-603.
70]. Yantih N, Musadad A., Ibrahim S., Emran R., Determination of Seven Water-Soluble
Vitamins in Tablet Simultaneously by Reversed-Phase High Performance Liquid
278
Chromatography, the International Seminar on Pharmaceutics; 2007, Institute Technology
Bandung, Indonesia.
71]. Amidzic R, Brboric J, Cudina O, Vladimirov S, RP-HPLC determination of vitamin B1,
B3, B6, Folic Acid, and B12 in multivitamin tablets, J. Serb. Chem. Soc.I 2005; 70 (10):
1229-1235.
72]. Thomas S, Kumar R, Sharma A, Issarani R, Prakash B, Staility-indicating HPLC method
for determination of vitamins B1, B2, B3, and B6 in pharmaceutical dosage form, Indian J.
O. Chem. Technol. 2008;15:598-603.
404
73]. Yantih N, Musadad A., Ibrahim S., Emran R., Determination of Seven Water-Soluble
Vitamins in Tablet Simultaneously by Reversed-Phase High Performance Liquid
Chromatography, the International Seminar on Pharmaceutics; 2007, Institute Technology
Bandung, Indonesia.
74]. Amidzic R, Brboric J, Cudina O, Vladimirov S, RP-HPLC determination of vitamin B1,
B3, B6, Folic Acid, and B12 in multivitamin tablets, J. Serb. Chem. Soc.I 2005; 70 (10):
1229-1235.
75]. International Conference on Harmonization, Validation of analytical procedures: text and
methodology Q2 (R1) (2005).
76]. United States Pharmacopeial Convention, The United States Pharmacopeia 32–The
National Formulary 27, Rockville: The United States Pharmacopeial Convention (2009).
77]. Ditjen POM Depkes RI, Farmakope Indonesia, ed. IV, Depkes RI, Jakarta, 1995, pp
1016-1017.
78]. Ibrahim S, Berbagai pendekatan pengujian kelinearan kurva baku pada metode analisis
instrumental, Acta Pharm. Ind. 2005; 30 (1): 30-34. African Journal of Pharmacy and
Pharmacology Vol. 2(2). pp. 029-036, April, 2008
79]. Y.H. Hsiang, R. Hertzberg, S. Hecht et al., J. Biol.Chem., 260(27) (1985) 14 873–14 878.
80]. J. Fassberg, V.J. Stella, J. Pharm. Sci. 81 (7) (1992)676–684.
81]. W.J.M. Underberg, R.M.J. Goossen, B.R. Smith, J.Pharm. Biomed. Anal. 8 (1990) 681–683.
82]. S.D. Pearson, L.A. Trissel, Am. J. Hosp. Pharm. 50
(1993) 1405–1409.
83].M.F. Aignasse, P. Prognon,M. Stachowicz, Int. J.Pharm. 113 (1995) 241–246.
84]. M.A. Faouzi, T. Dine, M. Luyckx, Int. J. Pharm. 105 (1994) 89–93.
85]. S.Budhavari, The Merck Index (monograph#3521), 14, 598,(2006).
86]. S.Budhavari, The Merck Index (monograph#3565), 14, 606,(2006).
87]. S.Budhavari, The Merck Index(monograph#9146), 14, 1573(2006)
88]. C.Sean Sweetman, Martindale-The Complete Drug Reference, 34,632, (2005).
89]. C.Sean Sweetman, Martindale-The Complete Drug Reference, 34,654, (2005).
90]. Indian pharmacopoeia vol.2, 1071, (2007).
91]. Indian pharmacopoeia vol.2, 1075, (2007).
92]. Indian pharmacopoeia vol.2, 1782, (2007).
405
93]. C.Sean Sweetman, Martindale-The Complete Drug Reference, 34,632, (2005).
94]. C.Sean Sweetman, Martindale-The Complete Drug Reference, 34,654, (2005).
95]. Indian pharmacopoeia vol.2, 1071, (2007).
96]. Indian pharmacopoeia vol.2, 1075, (2007).
97]. Indian pharmacopoeia vol.2, 1782, (2007).
98]. Notari, Stefania; Bocedi, Alessio; Ippolito, Giuseppe; Narciso,Pasquale; Pucillo, Leopoldo
Paolo; Tossini, Gianna; Donnorso, Raffaele Perrone; Gasparrini, Francesco; Ascenzi, Paolo,
Journal of Chromatography, B: 831(1-2), 258-266,(2006).
99]. Gish R G, Trinh H, Leung N, Chan F K L, Fried M L, Wright T L, Wang C, Anderson
J, Mondou E , Snow A, Sorbel J, Rousseau F, Corey L (2005) Safety and antiviral
activity of emtricitabine (FTC) for the chronic hepatitis B infection: a two-year study.
J Hepatol 43(1): 60-6.
100]. Mangoankar K and Desai A (2008) Simultaneous estimation of emtricitabine,
tenofovir disoproxil fumarate and efavirenz from tablets by reverse phase highperformance
liquid chromatography method. Indian Drugs 45(3): 188-92.
101]. Appala N R, Rao V J, Vanitha P K, Mukilteo K and Srinivasu K (2008)
Simultaneous estimation of tenofovir disoproxil, emtricitabine and efavirenz in tablet
dosage form by RP- HPLC. Orient J Chem 24(2): in press.
102]. The Merck Index, XIV Edition, 142, 2008.
103]. Cateau E, Tournier N, Dupuis A, Gwenael Le M and Venisse N, J Pharm Biomed
Anal., 2005, 39(3-4), 791-795.
104]. Loregian A, Pagni S, Ballarin E, Sinigalia E and Saverio G P and Palu G, J Pharm
Biomed Anal., 2006, 42(4), 500-505.
105]. Choi S Ok, Rezk N L and Kashuba Angela D M, J Pharm Biomed Anal., 2007, 43(4),
1562-1567.
106]. Raymond Suckow F, Mira Fein, Christoph Corell U and Thomas Cooper B, Journal of
chromatography B, 2004,799, 201-208.
107]. Sachse, Julia, Hartter, Sebastian, Hiemke and Christoph, Therauptic drug monitoring,
2005, 27(2), 158-162 .
108]. Kirtikar KR and Basu BD : Indian Medicinal plants (1991 Reprint). Part III. B.S.M.P Singh
Publisher, Dehradun, India (1918).
406
109]. Mac Donell AA and Keith AB: Vedic Index of Names and Subjects (3rd Reprint, 1967)
Vol.II, Motilal Banarasi Dass, Delhi-Patna-Varanasi (1912).
110].Li Li, Zhao Zuolian, Bi Xueyan, Wu Haixia (Handan Institute for Drug Control, Handan,
Hebei, 056001, P. R. China 2005-10
111]. Martindale, The Extra Pharmacopoeia, 34 th Edition, 580.
112]. Budavari S.Eds,Merck Index,13th edition,2001,5109, Merck Research lab.
113]. Yang X, Hu Z, Chan SY, Goh BC, Duan W, Chan E, Zhou S,J Chromatogr B
Analyt Technol Biomed Life Sci., 2005,821,221
114]. Poujol S, Pinguet F, Malosse F, Astre C, Ychou M, Culine S, Bressolle F. Clin
Chem, 2003 ,49,1900.
115]. de Jong FA, Mathijssen RH, de Bruijn P, Loos WJ, Verweij J, Sparreboom A.J
Chromatogr B Analyt Technol Biomed Life Sci, 2003,795, 383.
116]. Schoemaker NE, Rosing H, Jansen S, Schellens JH, Beijnen JH. Ther Drug Monit,
2003,25,120.
117]. Mahmood Ahmad, Muhammad Usman, Asadullah Madni, M. Zubair, Qamar-uz-Zaman,
M. Saleem Qureshi, Abubakar Munir, Mahtab Ahmad and Arshad Mahmood A fast and simple
HPLC-UV method for simultaneous determination of three anti-cancer agents in plasma of breast
cancer patients and its application to clinical pharmacokinetics, African Journal of Pharmacy and
Pharmacology Vol. 5(7), pp. 915 - 922, July 2011
118].Alsarra IA, Alarifi MN (2004). Validated liquid chromatographic determination of 5-
fluorouracil in human plasma. J. Chromatogr B.,804: 435–439.
119]. M. R. Ganjali, T. Razavi, F. Faridbod, S. Riahi, and P. Norouzi, Current Pharm. Anal. 5
(2009) 28
120]. M. R. Ganjali, T. Razavi, R. Dinarvand, S. Riahi, and P. Norouzi, Int. J. Electrochem. Sci. 3
(2008)
1543
121]. M. Javanbakht, A. Mohammadi, M. R. Ganjali, P. Norouzi, F. Faridbod, and H. Pirelahi, J.
Chinese Chem. Soc. 54 (2007) 1495
122]. M. R. Ganjali, B. Vesimohammadi, S. Riahi, and P. Norouzi, Int. J. Electrochem. Sci. 4
(2009) 740
407
123]. F. Faridbod, M. R. Ganjali, S. Labbafi, S. Riahi, and P. Norouzi, Int. J. Electrochem. Sci. 4
(2009)772
124]. Innocenti F, Undevia SD, Iyer L. et al. Genetic variants in the UDPglucuronosyltransferase
1A1 gene predict the risk of severe neutropenia of irinotecan. J Clin Oncol. 2004;22(8):1382– 8.
125]. O'Dwyer PJ and Catalano RB.Uridinediphosphateglucuronosyltransfe rase (UGT) 1A1 and
irinotecan. pharmacogenomics arrives in cancer therapy. J Clin Oncol. 2006;24(28):4534– 8.
126].Bhaskar M, L’Aurelle J, Bei N, Dolly P, Brock M, Pamala J, Natalia T (2010). Quantitative
High-Performance Liquid Chromatography Electrospray Ionization Tandem Mass Spectrometry
Analysis of Bis-N7-Guanine DNA DNA Cross-Links in White Blood Cells of Cancer Patients
Receiving Irnotecan Therapy. Anal Chem., 82: 3650–3658
127].Casale F, Canaparo R, Muntoni E, Serpe L, Zara GP, Della Pepa C,Berno E, Costa M, Eandi
M (2002). Simultaneous HPLC determination of 5-fluorouracil and its metabolites in plasma of
cancer patients. Biomed. Chromatogr. 16: 446-452.
128].Ciccolini J, Mercier C, Blachon MF, Favre R, Durand AA, Lacarelle B (2004). A simple
and rapid high-performance liquid chromatographic (HPLC) method for 5-fluorouracil (5-FU)
assay in plasma and possible detection of patients with impaired dihydropyrimidine
dehydrogenase (DPD) activity. J. Clin. Pharm. Therap., 29: 307-315.
129].Compagnon EL, Thiberville N, Moore C, Thudlez C, Lacroix (1996). Simple high-
performance liquid chromatographic method for the quantitation of 5-fluorouracil in human
plasma. J. Chromatogr. B.,677: 380-383.
130]. Thomas S, Kumar R, Sharma A, Issarani R, Prakash B, Staility-indicating HPLC method
for determination of vitamins B1, B2, B3, and B6 in pharmaceutical dosage form, Indian J.
O. Chem. Technol. 2008;15:598-603.
131]. Yantih N, Musadad A., Ibrahim S., Emran R., Determination of Seven Water-Soluble
Vitamins in Tablet Simultaneously by Reversed-Phase High Performance Liquid
Chromatography, the International Seminar on Pharmaceutics; 2007, Institute Technology
Bandung, Indonesia.
132]. Amidzic R, Brboric J, Cudina O, Vladimirov S, RP-HPLC determination of vitamin B1,
B3, B6, Folic Acid, and B12 in multivitamin tablets, J. Serb. Chem. Soc.I 2005; 70 (10):
1229-1235.
133]. International Conference on Harmonization, Validation of analytical procedures: text and
408
methodology Q2 (R1) (2005).
134]. United States Pharmacopeial Convention, The United States Pharmacopeia 32–The
National Formulary 27, Rockville: The United States Pharmacopeial Convention (2009).
135]. Ditjen POM Depkes RI, Farmakope Indonesia, ed. IV, Depkes RI, Jakarta, 1995, pp
1016-1017.
136]Di Paolo A, Danesi R, Ciofi L, Vannozzi F, Bocci G, Lastella M, Amatori F, Martelloni BM,
Ibrahim T, Amadori D, Falcone A, del Tacca M(2005). Improved analysis of 5-fluorouracil and
5,6-dihydro-5-
fluorouracil by HPLC with diode array detection for determination of cellular dihydropyrimidine
dehydrogenase activity and pharmacokinetic profiling . Ther. Drug Monit., 27: 362-368
137]Eiji M, Kozo Y, Ryuichi K, Ken-ichiro Y (1997). Determination of S-1(combined drug of
tegafur, 5-choloro-2,4-dihydroxypyridine and potassium oxonate) and 5-fluorouracil in human
plasma and urine using high-performance liquid chromatography and gas chromatography-
negative ion chemical ionization mass spectrometry.J. Chromatogr. B., 691: 95.
138]Kristein M (2009). Applied Therapeutics the Clinical Use of Drugs,Neoplastic disorders and
their treatment: General principles, Lippincott Williums & Wilkins.
139]Vainchtein LD, Rosing H, Schellens JH, Beijnen JH (2010). A new,validated HPLC-MS/MS
method for the simultaneous determination of the anti-cancer agent capecitabine and its
metabolites: 5'-deoxy-5-fluorocytidine, 5'-deoxy-5-fluorouridine, 5-fluorouracil and 5-
fluorodihydrouracil, in human plasma. Biomed. Chromatogr., 24: 374-86.
140]Wattanatorn W, McLeod HL, Cassidy J, Kendle KE (1997). Highperformanceliquid
chromatographic assay of 5-fluorouracil in human erythrocytes, plasma and whole blood. J.
Chromatogr. B., 692: 233-237.
141]Wei G, Xiao S, Si D, Liu C (2008). Improved HPLC method for doxorubicin quantification
in rat plasma to study the pharmacokinetics of micelle-encapsulated and liposome-encapsulated
doxorubicin formulations. Biomed. Chromatogr., 22: 1252-8.
142]Yuan G, Rong L, Duanyun S, Changxiao L (2010). Determination of 5-fluorouracil in
human plasma by high-performance liquid chromatography (HPLC). Tran. Tianjin Uni., 16: 167-
173.
409
143]Zhou QY, Chowbay B (2002). Determination of doxorubicin and its metabolites in rat serum
and bile by LC: application to preclinicalpharmacokinetic studies. J. Pharm. Biomed. Anal., 30:
1063-1074.
144]. Ibrahim S, Berbagai pendekatan pengujian kelinearan kurva baku pada metode analisis
instrumental, Acta Pharm. Ind. 2005; 30 (1): 30-34. African Journal of Pharmacy and
Pharmacology Vol. 2(2). pp. 029-036, April, 2008
145]. Indian pharmacopoeia vol.2, 1071, (2007).
146]. Indian pharmacopoeia vol.2, 1075, (2007).
147]. Indian pharmacopoeia vol.2, 1782, (2007).
148]. Appala N R, Rao V J, Vanitha P K, Mukilteo K and Srinivasu K (2008) Simultaneous
estimation of tenofovir disoproxil, emtricitabine and efavirenz in tablet dosage form by RP-
HPLC. Orient J Chem 24(2): in press.
149]. The Merck Index, XIV Edition, 142, 2008.
150]. Cateau E, Tournier N, Dupuis A, Gwenael Le M and Venisse N, J Pharm Biomed Anal.,
2005, 39(3-4), 791-795.
151]. Loregian A, Pagni S, Ballarin E, Sinigalia E and Saverio G P and Palu G, J Pharm Biomed
Anal., 2006, 42(4), 500-505.
152]. Choi S Ok, Rezk N L and Kashuba Angela D M, J Pharm Biomed Anal., 2007, 43(4),
1562-1567.
153]. Rebiere H, Mazel B, Civade C and Bonnet Pierre-Antoine, J Chromatography B, 2007,
850(1-2), 376-383.
154].M. Palumbo, C. Sissi, B. Gatto, S. Moro, and G. Zagotto, J. Chromatogr. B, 764 , 121
(2001)
155]. Ditjen POM Depkes RI, Farmakope Indonesia, ed. IV, Depkes RI, Jakarta, 1995, pp 1016-
1017.
156]. Ibrahim S, Berbagai pendekatan pengujian kelinearan kurva baku pada metode analisis
instrumental, Acta Pharm. Ind. 2005; 30 (1): 30-34
157]. S. Sawada, S. Okajima, R. Aiyama, K. Nokota, T. Furuta, T. Yokokura, E. Sugini,
K. Yamaguchi, and K. Miyasaka, Chem. Pharm. Bull., 39, 1446 (1991)
158]. Rivory LP, Robert J.,J Chromatogr B Biomed Appl, 1994,661,133.
410
159]. Barilero I, Gandia D, Armand JP, Mathieu-Boue A, Re M, Gouyette A, Chabot GG.J
Chromatogr, 1992,575, 275.
160]. Validation of Analytical Methods: Definitions and Terminology, ICH Harmonised
Tripartite Guideline, ICH Topic Q2A, 1994.
161]. Validation of Analytical Procedures,Methodology, Step 4, Consensus,ICH Harmonised
Tripartite Guideline, Guideline, 1996.
162].S. Sawada, S. Okajima, R. Aiyama, K. Nokota, T. Furuta, T. Yokokura, E. Sugini, K.
Yamaguchi, and K. Miyasaka, Chem. Pharm. Bull., 39 , 1446 (1991)
163]. Y. Kawato, M. Aonuma, Y. Hirota, H. Kuga, and K. Sato, Cancer Res., 51 , 4187 (1991)
164].K. Takasago, Y. Kasai, Y. Kitano, K. Mori, K. Kakihata, M. Hirohashi, and M. Nonura,
Folia Pharmacol. Jpn., 105 , 447 (1995)
165].A. Kono and Y. Hara, Jpn. J. Cancer Chemother., 18 , 2175 (1991)
166].R. Mick, E. Gupta, E Vokes, and M. Ratain, J. Clin. Oncol., 14 , 2012 (1996)
167].H. Sumoyoshi, Y. Fujiwara, T. Ohune, N. Yamaoka, K. Tamura, and M. Yamakido, J.
Chromatogr. B, 670 , 309 (1995)
168].A. Sparreboom, P. Bruijn, M. Jonge, W. Loos, J. Stoter, J. Verweij, and K. Nooter, J.
Chromatogr. B, 712 , 225 (1998)
169].J. Escoriaza, A. Aldaz, C. Castellanos, E. Calvo, and J. Giraldez, J. Chromatogr. B, 740 ,
159 (2000)
170]. Y. Kawato, M. Aonuma, Y. Hirota, H. Kuga, and K. Sato, Cancer Res., 51, 4187 (1991)
171]. G. Nishimura, T. Satou, Y. Yoshimitsu, Y. Kurosaka, and T. Fujimura, Jpn. J. Cancer
Chemother., 22, 93 (1995)
172]. Y. Ohe, Y. Sasaki, T. Shinkai, K. Eguchi, T. Tamura, A. Kojima, H. Kunikane, H. Okamoto,
A. Karato, H. Ohmatsu, F. Kanzawa, and N. Saijo, J. Natl. Cancer Inst. 84, 972 (2004)
173]. K. Takasago, Y. Kasai, Y. Kitano, K. Mori, K. Kakihata, M. Hirohashi, and M. Nonura,
Folia Pharmacol. Jpn., 105, 447 (1995)
174]. A. Kono and Y. Hara, Jpn. J. Cancer Chemother., 18, 2175 (1991)
175]. R. Mick, E. Gupta, E Vokes, and M. Ratain, J. Clin. Oncol., 14, 2012 (1996)
176]. H. Sumoyoshi, Y. Fujiwara, T. Ohune, N. Yamaoka, K. Tamura, and M. Yamakido, J.
Chromatogr. B, 670, 309 (1995)
177]. M.A. Faouzi, T. Dine, M. Luyckx, J. Pharm. Biomed.Anal. 13 (1995) 1363–1372.
411
178]. A.E. Ganning, U. Brunk, G. Dallner, Hepatology 4 (3)(1984) 541–547.
179]. W.N. Waugh, L.A. Trissel, V.J. Stella, Am. J. Hosp.Pharm. 48 (1991) 1520–1524.
180]. E.J. Draviam, K.H. Pearson, J. Kerkay, Anal. Lett. 15(B21 and 22) (1982) 1729–1750.
181]. R.P. Snell, J. AOAC Int. 76 (3) (1993) 531–534.
182].Dafeng C, Jingkai G, Wanhui L, Paul FJ, Qingguang D (2003). Sensitive Ahmad et al. 921
liquid chromatographic assay for the simultaneous determination of 5-fluorouracil and its
prodrug, tegafur, in beagle dog plasma. J.Chromatogr. B., 795: 377-382.
183].Di Paolo A, Danesi R, Ciofi L, Vannozzi F, Bocci G, Lastella M, Amatori F, Martelloni BM,
Ibrahim T, Amadori D, Falcone A, del Tacca M (2005). Improved analysis of 5-fluorouracil and
5,6-dihydro-5-fluorouracil by HPLC with diode array detection for determination of cellular
dihydropyrimidine dehydrogenase activity and pharmacokinetic profiling . Ther. Drug Monit.,
27: 362-368
184].Kristein M (2009). Applied Therapeutics the Clinical Use of Drugs,Neoplastic disorders
and their treatment: General principles,Lippincott Williums & Wilkins.Larson RR, LMB,
Khazaeli, Kenneth DH (2003). Development of anHPLC method for simultaneous analysis of
Five Antineoplastic Agents. App. Occup. Environ. Hygiene, 18: 109-119.
185].Loreto AC, Luisa S, Jose ML (1999). High performance chromatographic validated assay of
doxorubicin in rat plasma and tissues. J. Chromatogr. B., 721: 271-278.
186].Mark AN, Farshad HS, Christine MC, Shereeni V, Matshela JM, Mihail O, Diedre R, Susan
B, Diane F, David JS, Rakesh G (2002). AnHPLC method for the measurement of 5-fluorouracil
in human plasma with a low detection limit and a high extraction yield. Inter J.Mol. Med., 10:
513-516.
187]. Ter Heine, Rob; Alderden-Los, Carolien G.; Rosing, Hilde;Hillebrand, Michel J. X.; van
Gorp, Eric C. M.; Huitema,Alwin D. R.; Beijnen, Jos H. Rapid Communications in Mass
Spectrometry, 21(15), 2505-2514 (2007).
188]. Rebiere, Herve; Mazel, Bernard; Civade, Corinne; Bonnet, Pierre-Antoine, Journal of
Chromatography B: 850(1-2), 376-383 (2007).
189]. Choi, Sun Ok; Rezk, Naser L.; Kashuba, Angela D. M.,Journal of Pharmaceutical and
Biomedical Analysis, 43(4),1562-1567, (2007).
190]. Weller, Dennis R.; Brundage, Richard C.; Balfour, Henry H.;Vezina, Heather E. Journal of
Chromatography, B: 848(2),369-373 (2007).
412
191]. Manogaokar.K, and Desai. A. Indian Drugs 45(3), 188-192,(2008).
192]. Mangoankar K and Desai A (2008) Simultaneous estimation of emtricitabine,tenofovir
disoproxil fumarate and efavirenz from tablets by reverse phase highperformance liquid
chromatography method. Indian Drugs 45(3): 188-92.
193].Ping Z, Alekha KD (1999). A simple HPLC method using a microbore column for the
analysis of doxorubicin. J. Pharm. Biomed. Anal., 20:543-548.
194].Rossella P, Massimo B, Simona G, Christopher AJ (2005). Hydrophilic interaction liquid
chromatography–APCI–mass spectrometry determination of 5-fluorouracil in plasma and tissues.
J. Pharm. Biomed. Anal., 38: 738-745.
195].Stephen PA, Garg MB, Dunstan HR (1997). Simultaneous Determination of
Dihydrofluorouracil and 5-Fluorouracil in Plasma by High-Performance Liquid
Chromatography. Anal. Biochem., 246: 79-85.
196].US Food and Drug Administration, Fed Regist. 57:17997-18001 (2009). Available from:
http//www.fda.gov/cder/ogd/ Accessed on 2009 Nov24.
197].Vainchtein LD, Rosing H, Schellens JH, Beijnen JH (2010). A new,validated HPLC-
MS/MS method for the simultaneous determination of the anti-cancer agent capecitabine and its
metabolites: 5'-deoxy-5-fluorocytidine, 5'-deoxy-5-fluorouridine, 5-fluorouracil and 5-
fluorodihydrouracil, in human plasma. Biomed. Chromatogr., 24: 374-86.
198].Wattanatorn W, McLeod HL, Cassidy J, Kendle KE (1997). Highperformance liquid
chromatographic assay of 5-fluorouracil in human erythrocytes, plasma and whole blood. J.
Chromatogr. B., 692: 233-237.
199].Wei G, Xiao S, Si D, Liu C (2008). Improved HPLC method for doxorubicin quantification
in rat plasma to study the pharmacokinetics of micelle-encapsulated and liposome-encapsulated
doxorubicin formulations. Biomed. Chromatogr., 22: 1252-8.
200].Yuan G, Rong L, Duanyun S, Changxiao L (2010). Determination of 5-922 Afr. J. Pharm.
Pharmacol.fluorouracil in human plasma by high-performance liquid chromatography (HPLC).
Tran. Tianjin Uni., 16: 167-173.
201]. Weller Dennis R, Brundage Richard C, Balfour Henry H and Vezina Heather E, J
Chromatography B, 2007, 848(2), 369-373.
202]. D'Avolio A, Siccardi M, Sciandra M, Lorena B, Bonora S, Trentini L and Di Perri G,
J Chromatogr B, 2007, 859(2), 234-240.
413
203]. Cattaneo D, Maggiolo F, Ripamonti D and Perico N, J Chromatographic Science,2008,
46(6), 485-489.
204]. Seshachalam U, Rao Narasimha D V L, Haribabu B and Chandrasekhar K
B,Chromatographia, 2007, 65(5/6), 355-358.
205].Zhou QY, Chowbay B (2002). Determination of doxorubicin and its metabolites in rat
serum and bile by LC: application to preclinical pharmacokinetic studies. J. Pharm. Biomed.
Anal., 30: 1063-1074.
206].Anderson SH, Charles TJ, Nicol AD (1985). Thiamine deficiency at a district general
hospital: report of five cases. Q. J. Med. 55: 15-32.
207].Berger NA (1985). Poly (ADP-ribose) in the cellular response to DNA damage. Radiat. Res.
101: 4-15.
208].Carpenter KJ, Lewin WJ (1985). A reexamination of the composition of diets associated
with pellagra. J. Nutr. 115: 543-552.
209].FAO/WHO (1974). Handbook on Human Nutritional Requirements, eds. Passmore R, Nicol
BM, Rao M, Narayana Beaton GH, de Mayer EM,
210].FAO Nutritional Studies WHO Monograph Series, Rome, Italy. 6 : 28 FAO/WHO (2002).
Human Vitamin and Mineral Requirements, WHO,Rome, Italy.
211].McCormick DB (1988). Thiamin. In: Modern Nutrition in Health and Disease, 6th edition.
15.Shils ME, Young VR, eds. Philadelphia: Lea and Febiger, pp. 355-361.
212].McCormick DB (1988). Vitamin B6. In: Modern Nutrition in Health and Disease, 6th
edition. Shils ME, Young VR, eds. Philadelphia: Lea and Febiger. pp. 376-382.
213].McCormick DB (1988). Niacin. In: Modern Nutrition in Health and Disease, 6th edition.
Shils ME, Young VR, eds. Philadelphia: Lea and Febiger. pp.370-375.
214] .McCormick DB (1994). Riboflavin. In: Modern Nutrition in Health and Disease, 8th
edition. Shils ME, Olson JA, Shike M, eds. Philadelphia: Lea and Febiger. pp. 366-375.
215].McCormick DB (1996). Co-enzymes, Biochemistry of. In: Encyclopedia of Molecular
Biology and Molecular Medicine, Meyers RA, ed.Weinheim: VCH. 1:396-406. 036 Afr. J.
Pharm. Pharmacol.
216].McCormick DB (1997). Co-enzymes, Biochemistry. In: Encyclopedia of Human Biology
2nd edition. Dulbecco R, ed.-in-chief. San Diego: Academic Press, pp. 847-64.
414
217].Nichols HK, Basu TK (1994). Thiamin status of the elderly: dietary intake and thiamin
pyrophosphate response. J. Am. Coll. Nutr. 13: 57-61.
218].Olaniyi AA (2000). Essential Medicinal Chemistry, 2nd Edition, Shaneson, Ibadan, p. 324.
Patterson JI, Brown RR, Linkswiler H, Harper AE (1980). Excretion of tryptophan-niacin
metabolites by young men: effects of tryptophan, leucine, and vitamin B6 intakes. Am. J. Clin.
Nutr. 33: 2157-67.
219].Shibata K, Matsuo H (1989). Effect of supplementing low protein diets with the limiting
amino acids on the excretion of N1-methylnicotinamide and its pyridones in rat. J. Nutr. 119:
896-901
220]. Sweetman et al. Martindale: The Complete Drug Reference. 36th edition.2009.
Pharmaceutical Press.
221]. British National Formulary. 58th edition.2009. British Medical Association and Royal
Pharmaceutical Society of GreatBritain.
222]. Electronic Medicines Compendium.(Accessed October 2009).
223]Perry MC. The Chemotherapy Source Book. 4th edition. 2007. Lippincott, Williams and
Wilkins
224]. Notari, Stefania; Bocedi, Alessio; Ippolito, Giuseppe;Narciso,Pasquale; Pucillo, Leopoldo
Paolo; Tossini, Gianna;Donnorso, Raffaele Perrone; Gasparrini, Francesco; Ascenzi, Paolo,
Journal of Chromatography, B: 831(1-2), 258-266,(2006).
225]. Ter Heine, Rob; Alderden-Los, Carolien G.; Rosing, Hilde; Hillebrand, Michel J. X.; van
Gorp, Eric C. M.; Huitema, Alwin D. R.; Beijnen, Jos H. Rapid Communications in Mass
Spectrometry, 21(15), 2505-2514 (2007).
226]. Rebiere, Herve; Mazel, Bernard; Civade, Corinne; Bonnet, Pierre- Antoine, Journal of
Chromatography B: 850(1-2), 376-383 (2007).
227]. Raju N A, Begum S (2008) Simultaneous RP-HPLC Method for the Estimation of the
Emtricitabine, Tenofovir Disoproxil fumarate and Efavirenz in Tablet Dosage form.Research J
Pharm and Tech 1(4):522-25.
228]. Shirkhedkar A A, Bhirud C H, Surana S J (2009) Application of UVSpectrophotometric
Methods for Estimation of Tenofovir Disoproxil Fumarate in Tablets. Pak J Pharm Sci 22(1): 27-
29.
229]. Sparidans R W, Crommentuyn K M, Schellens J H, Beijnen J H (2003) Liquid-
415
Chromatography assay for the antiviral nucleotide analogue tenofovir in plasma using
derivitization with chloroacetaldehyde. J Chromatogr B 791: 227-33.
230]. Sentenac S, Fernandez C, Thuillier A, Lechat P and Aymard G (2003) Sensitive
determination of tenofovir in human plasma samples using reversed-phase
liquidchromatography. J Chromatogr B 793(2): 317-24.
231] C. Persiani, P. Cukor, J. Chromatogr. 109 (1975) 413–417.
232] R. Venkataramanan, G.J. Burckart, R.J. Ptachcinski,Am. J. Hosp. Pharm. 43 (1986) 2800–
2802.
233] A. Kearney, K. Patel, N. Palepu, Int. J. Pharm. 127(1996) 229–237.
234]. Kandagal P B, Manjunatha D H, Seetharamappa J and Kalanur S S (2008) RP-HPLC
method for the determination of tenofovir in pharmaceutical formulations and spiked human
plasma. Anal Lett 41(4): 561-70.
235]. Delahunty T, Bushman L, Fletcher C V (2006) Sensitive assay for determining plasma
tenofovir concentrations by LC/MS/MS. J Chromatogr B 830: 6-12.
236]. Massaki T, Yuichi K, Naoya O, Atsushi H, Kazuhide B, Tsuguhiro K (2007) Determination
of plasma tenofovir concentration using a conventional LC-MS method. Biol Pharm Bull 30:
1784-86.
237]. King T, Bushman L, Kiser J, Anderson P L, Michelle R, Delahunty T and Fletcher C
V (2006) Liquid chromatography-tandem mass spectrometric determination of tenofovir-
diphosphate in human peripheral blood mononuclear cells. J Chromato B 843(2,7): 147-56.
238]. Droste J A H, Aarnoutse R E, Burger D M (2007) Determination of Emtricitabine in
Human Plasma using HPLC with Fluorometric Detection. J Liq Chromatogr Related Technol
30(18): 2769-78.
239]. Unnam S, Bodepudi H, Kottapalli C B (2007) Development and Validation of Stability
indicating liquid chromatographic method for determination of emtricitabine and related
impurities in drug substances. J Sep Sci 30: 999-1004.
240]. Rezk N L , Crutchley R D , Kashuba A D M ( 2005) Simultaneous quantification of
emtricitabine and tenofovir in human plasma using high performance liquid chromatography
after solid phase extraction. J Chromatogr B 822: 201-8.
241]. Gomes N A, Vaidya V V, Pudage A, Joshi S S and Parekh S A (2008) Liquid
chromatography–tandem mass spectrometry (LC–MS/MS) method for simultaneous
416
determination of tenofovir and emtricitabine in human plasma and its application to a
bioequivalence study. J Pharm Biomed Anal 48(3): 918-26.
242]. Joshi M, Nikalje A P, Shahed M, Dehghan M (2009) HPTLC method for the simultaneous
estimation of emtricitabine and tenofovir in tablet dosage form. Indian JPharm Sci 71(1): 95-97.
243]. International Conference on Harmonization. ICH Harmonised Tripartite Guidelines–
Validation of Analytical Procedures: Methodology, (1997) Fed. Regist.
244]. Weller Dennis R, Brundage Richard C, Balfour Henry H and Vezina Heather E, J
Chromatography B, 2007, 848(2), 369-373.
245]. D'Avolio A, Siccardi M, Sciandra M, Lorena B, Bonora S, Trentini L and Di Perri G,J
Chromatogr B, 2007, 859(2), 234-240.
246]. Cattaneo D, Maggiolo F, Ripamonti D and Perico N, J Chromatographic Science,2008,
46(6), 485-489.
247]. Seshachalam U, Rao Narasimha D V L, Haribabu B and Chandrasekhar K
B,Chromatographia, 2007, 65(5/6), 355-358.
248]. Aarti Raja, John lebbos, Peter Kirkpatric, Atazanavir Sulphate, Nature Review Drug
Discovery 2003; 2: 857-8.
249]. Lovgian Arianna, Pagni silvana, Ballanin Elisa, Palu Giologio, Parisi Saverio Giuseppe.
Simple activation of the HIV protease inhibition atazanavir in human plasma by HPLC with UV
detection. Journal of pharmaceutical and biomedical analysis 2006; 42: 500-5.
250]. Trbat, verdier, Avienx, Allin, Michelet, Danicle. Simultaneous quantitative assay of
atazanavir and 6 other HIV protease inhibitors by isocratic reversed phase liquid chromatography
in human plasma. Therapeutic drug monitoring 2005; 27:265-9.
251]. E.Dailly, F.Raffi, P.Jolliet. Determination of atazanavir and other anti-retroviral drugs
plasms levels by High Performance Liquid Chromatography with UV- detection. Journal of
chromatography B 2004; 813: 353-8.
252]. U.Seshachalam, D.V.L Narasimha Rao,B.Haribasu, K.B. Chandrasekhar. Determination of
atazanavir in the presence of its degradation products by a stability indicating L.C. method.
Chromotographia 2007; 65:355-8
253]. ICH, Q2 (R1) validation of analytical procedures: text and methodology, International
conference on harmonization; Nov.1996.
417
254]. Manogaokar.K, and Desai. A. Indian Drugs 45(3), 188-192,(2008).
255]. Budawari S (2001) The Merck Index, 13th Edition, Merck and Co. Inc. Whitehouse
Station. NJ, p 630, 1631-32.
256]. Martindale (2002) The Complete Drug Reference, 33 rd Edition, Pharmaceutical Press,
London, p 620, 642.