Date : 05-Jun-2018 Feedback Page 1 for Steroid Hormones Distribution : 453 Laboratory : Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300 © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk Published at 8:57 on Thursday 14 June 2018 Quality Manager Pathology Laboratory Biochemistry Department Town County Postcode Country This Scheme is essentially web-based. We can alert you to information regarding the Scheme via email. The e-mail address (or addresses) we are currently using to contact your laboratory is shown below in red . If no e-mail address is displayed or the information shown is incorrect, please email us with an appropriate contact e-mail address as soon as possible, using the word 'feedback' in the title line. Based on the date information you have provided, the transit time from specimen dispatch [ ] to receipt [ ] was day(s), and the subsequent time to analysis [ ] in your laboratory was day(s). (Missing values indicate dates not provided. "0 days" represents same day). Any comments you made to us are shown below and have been acted upon where necessary Any specific comments applicable only to laboratory are shown below Any general comments applicable to all laboratories are shown below Please note that Free Testosterone now appears on our Schedule of Accreditation for ISO 17043 You have until the close of Distribution 454 (03/07/2018) to submit late results / request amendments to results for Distribution 453. Please enter any late results or requests to amend non-analytical errors for Distribution 453 on the web under the usual Results button ensuring the correct Distribution number has been selected. You should include your name and a valid reason. Amending results is at the discretion of the Director and is not an automatic entitlement. We are not accepting any further requests to amend results or to change methods for Distribution 452. Performance Surveillance and Performance Issues If you have a red banner for any analyte on the Performance Summary Icon (PSI) page, then you should log this in your Quality Management System. Birmingham Quality is here to help. If you are unsure as to why you have out-of-consensus problems or if you are having difficulties with your Root Cause Analysis, then please do contact us. It is the responsibility of the Participant to act on, investigate and hopefully resolve, all out-of-consensus performance. It is the responsibility of the Organiser to highlight in the regular report any out-of-consensus performance by the use of scores, symbols, graphs and ‘traffic lights’ etc. Likewise, for the non-numerical analytes, please check the Pie Charts and other graphical representations to see if you are out-of-consensus. Finlay MacKenzie Director, Birmingham Quality Report authorised on Wednesday 13 June 2018 by: Finlay MacKenzie Scheme Organiser Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. Please see http://www.ukas.com for full details of the accreditation status of our services Birmingham Quality is proud to offer EQA services that adhere to the Code of Practice and have the badge of quality of UK NEQAS of 60

Transcript of 453 Date : 05-Jun-2018 Feedback - Birmingham Quality · 2021. 1. 29. · H454 453E 3.15 P481 19.5...

  • Date : 05-Jun-2018

    Feedback

    Page 1

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    Quality ManagerPathology LaboratoryBiochemistry DepartmentTownCountyPostcodeCountry

    This Scheme is essentially web-based. We can alert you to information regarding the Scheme via email. The e-mail address (or addresses) we are currently using to contact your laboratory is shown below in red . If no e-mail address is displayed or the information shown is incorrect, please email us with an appropriate contact e-mail address as soon as possible, using the word 'feedback' in the title line. Based on the date information you have provided, the transit time from specimen dispatch [ ] to receipt [ ] was day(s), and the subsequent time to analysis [ ] in your laboratory was day(s). (Missing values indicate dates not provided. "0 days" represents same day). Any comments you made to us are shown below and have been acted upon where necessary Any specific comments applicable only to laboratory are shown below Any general comments applicable to all laboratories are shown below Please note that Free Testosterone now appears on our Schedule of Accreditation for ISO 17043 You have until the close of Distribution 454 (03/07/2018) to submit late results / request amendments to results for Distribution 453. Please enter any late results or requests to amend non-analytical errors for Distribution 453 on the web under the usual Results button ensuring the correct Distribution number has been selected. You should include your name and a valid reason. Amending results is at the discretion of the Director and is not an automatic entitlement. We are not accepting any further requests to amend results or to change methods for Distribution 452. Performance Surveillance and Performance Issues If you have a red banner for any analyte on the Performance Summary Icon (PSI) page, then you should log this in your Quality Management System. Birmingham Quality is here to help. If you are unsure as to why you have out-of-consensus problems or if you are having difficulties with your Root Cause Analysis, then please do contact us. It is the responsibility of the Participant to act on, investigate and hopefully resolve, all out-of-consensus performance. It is the responsibility of the Organiser to highlight in the regular report any out-of-consensus performance by the use of scores, symbols, graphs and ‘traffic lights’ etc. Likewise, for the non-numerical analytes, please check the Pie Charts and other graphical representations to see if you are out-of-consensus. Finlay MacKenzie Director, Birmingham Quality Report authorised on Wednesday 13 June 2018 by: Finlay MacKenzie Scheme Organiser

    Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. Please see http://www.ukas.com for full details of the accreditation status of our services

    Birmingham Quality is proud to offer EQA services that adhere to the Code of Practice and have the badge of quality of UK NEQAS

    of 60

  • Date : 05-Jun-2018

    Performance Summary Icons (click graph for details)

    Page 2

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    B s

    core

    C score

    Cortisol

    P-ABCPENR (DASH1)

    0 5 10 15 20 25

    -30

    -20

    -10

    0

    10

    20

    30

    B s

    core

    C score

    Urinary Free Cortisol

    P-ABCPENR (DASH1)

    0 25 50 75 100

    -25

    0

    25

    50

    75

    100

    B s

    core

    C score

    17OH-Progesterone

    P-ABCPENR (DASH1)

    0 20 40 60 80

    -60

    -40

    -20

    0

    20

    40

    60

    B s

    core

    C score

    Progesterone

    P-ABCPENR (DASH1)

    0 5 10 15 20 25

    -25-20-15-10-505

    10152025

    B s

    core

    C score

    Oestradiol [E2]

    P-ABCPENR (DASH1)

    0 10 20 30 40

    -40

    -30

    -20

    -10

    0

    10

    20

    30

    40

    B s

    core

    C score

    Oestradiol [high-level]

    P-ABCPENR (DASH1)

    0 10 20 30 40

    -40

    -30

    -20

    -10

    0

    10

    20

    30

    40

    B s

    core

    C score

    Testosterone [female]

    P-ABCPENR (DASH1)

    0 20 40 60 80

    -40

    -20

    0

    20

    40

    60

    80

    B s

    core

    C score

    Testosterone [male]

    P-ABCPENR (DASH1)

    0 5 10 15 20 25

    -30

    -20

    -10

    0

    10

    20

    30

    B s

    core

    C score

    Free Testosterone

    P-ABCPENR (DASH1)

    0 5 10 15 20 25 30

    -30

    -20

    -10

    0

    10

    20

    30

    B s

    core

    C score

    SHBG

    P-ABCPENR (DASH1)

    0 5 10 15 20

    -30

    -20

    -10

    0

    10

    20

    30

    B s

    core

    C score

    Androstenedione

    P-ABCPENR (DASH1)

    0 20 40 60 80 100 120 140

    -40

    -20

    0

    20

    40

    60

    80

    100

    120

    140

    B s

    core

    C score

    DHA sulphate

    P-ABCPENR (DASH1)

    0 5 10 15 20 25 30

    -40

    -30

    -20

    -10

    0

    10

    20

    30

    40

    B s

    core

    C score

    Aldosterone

    P-ABCPENR (DASH1)

    0 10 20 30 40 50 60

    -20

    -10

    0

    10

    20

    30

    40

    50

    60

    of 60

  • Date : 05-Jun-2018

    Participation summary

    Page 3

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    All our time periods are 'rolling' to give you current information. You may wish to keep you own log of Calendar Year or Financial Year time points if you require 'year-end' statements for your own internal use. Any analytes with out of consensus performance will be highlighted in red and can be clicked for further details.

    17OH-Progesterone

    You have no performance data for:

    You have out of consensus performance for:

    DHA sulphateFree TestosteroneYou have in consensus performance for:

    Testosterone [male]SHBGAndrostenedione

    Cortisol

    Testosterone [female]

    Analytical Performance over the last 6 months (rolling time window of 6 distributions)

    None

    Oestradiol [high-level] Urinary Free Cortisol

    Aldosterone Progesterone Oestradiol [E2]

    This scheme cycle is notionally every four weeks. Analytically, we assess you over a six month time window (6 Distributions). For return rates, late and amended results we assess you over a twelve month period (12 distributions).

    Satisfactory0 distributions accepted from the last 12

    ParticipationSatisfactoryLate ReturnsSatisfactory

    0 distributions from the last 1212 distributions out of a possible 12

    Amendments

    Rating Affected Distributions Distributions

    Participation and Return Rates

    You can judge, in association with your IQC and other QA measures, if your current performance is a blip or part of a trend.

    You are not registered for:

    Testosterone [male] DHA sulphate

    Out of consensus for at least one specimen for:

    You have no specimen %bias etc. for:In consensus for all specimens for: Free Testosterone

    Analytical Performance for specimens from distribution 453 only

    Oestradiol [high-level] Aldosterone

    SHBG 17OH-Progesterone Progesterone

    Oestradiol [E2]

    None

    Urinary Free Cortisol Androstenedione Cortisol

    Testosterone [female]

    None

    of 60

  • Date : 05-Jun-2018

    Distribution Summary

    Page 4

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    %biasBResult A score C scorePool CB scoreTargetSpecimen ASpecimen

    -2.4

    374

    453E

    FT439

    453D

    17.9

    84

    453B

    +4.6

    40.7

    170

    184

    +8.5

    281

    Free Testosterone

    FT432

    453A

    273

    Progesterone

    453D

    C546

    18.7

    9262

    P480

    E523

    +13.6

    11413

    134

    258

    (nmol/L)

    5.10

    1.20

    25.3

  • Date : 05-Jun-2018

    Distribution Summary

    Page 5

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    %biasBResult A score C scorePool CB scoreTargetSpecimen ASpecimen

    1.9A407

    D326

    (umol/L)

    48

    4.2

    +61.8

    -13.3

    453C

    198

    297

    -13.349

    3.0

    1.7

    +1.3

    207.9

    (pmol/L)

    453E +44.0

    L103732

    G396

    523

    26.1

    56.9

    453E

    4.8

    453A

    L104

    +35.1

    27.9

    9.1

    453B

    A409

    >30.0+6.9

    9.2D335

    Androstenedione

    3.3

    453A

    L094

    973

    A411

    G391

    D324

    453D1.3

    +32.0

    251

    2.4

    2.2

    D327

    D328

    453C

    DHA sulphate

    2.2

    2.46 +10.5-2.3

    (nmol/L) 56.5SHBG

    +32.8

    30.8

    (nmol/L)

    +62.1215

    14.4

    G360 -4.8

    14.2453B

    453B

    453C

    1.8

    124

    4.5

    453D

    A410

    +0.9

    59.5

    453B +39.4

    -19.4

    453A

    51.9

    453D

    53

    453A

    Aldosterone+143.3

    -8.8

    +38.2453C

    116

    15.3

    481

    G416

    1.4 +0.7A408

    45

    If your laboratory is outside of the acceptable limits of performance for any its rolling time-window scores (A, B or C scores), this will be indicated by a red traffic light symbol. It is the responsibility of the laboratory to undertake an internal investigation to establish the underlying cause and put in place corrective and preventive action. Please do not wait to receive a formal notification of performance from the Scheme Organiser or the National Quality Assurance Advisory Panel (NQAAP) before logging the non-conformity and, where necessary, acting upon the data contained in your report. A green traffic light merely reflects that your laboratory is performing as well as the state-of-the-art allows; it does not necessarily mean that your assay / laboratory performance is good enough clinically. . The 'A-Score' is not available for Free Testosterone. Once we have accumulated more data points we will be able to calculate concentration-dependent degree of difficulty factors and put them in the same ‘currency’ as the A scores in other Schemes.

    You have until the close of Distribution 454 (03/07/2018) to submit late results / request amendments to results for Distribution 453. Please enter any late results or requests to amend non-analytical errors for Distribution 453 on the web under the usual Results button ensuring the correct Distribution number has been selected. You should include your name and a valid reason. Amending results is at the discretion of the Director and is not an automatic entitlement. We are not accepting any further requests to amend results or to change methods for Distribution 452.

    Distribution 454 will be dispatched on 12/06/2018 . Results are due back in Birmingham by, notionally, 23:59 on 03/07/2018. Where your results appear as "XPL", it is because you did not report a numerical value for that analyte, but you did provide an explanation as to why a result was not reported.

    of 60

  • Date : 05-Jun-2018

    Method Summary

    Page 6

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    arrow median

    Units Method

    A score

    Your Submethod

    A scoremedian

    A scoreYour Method All labwith trend

    pmol/LBeckman Access/Dxi

    94

    nmol/L

    DiaSorin Liaison

    Urinary Free Cortisol

    Beckman Access/Dxi

    74

    pmol/L

    67

    Beckman Access/Dxi

    116

    218

    pmol/L

    127

    53

    Beckman Access/Dxi

    323

    Oestradiol [high-level] 147

    212

    78

    Tandem Mass Spectrometry

    126

    Beckman Access/Dxi [2SF]

    nmol/L

    DHA sulphate

    Androstenedione

    220

    72

    nmol/L

    SHBG

    nmol/L

    128

    251

    umol/L

    nmol/L

    Progesterone

    44

    nmol/L

    124

    Aldosterone

    59

    Beckman Access/Dxi

    56

    Cortisol

    60Oestradiol [E2] 256

    187

    273

    Tandem Mass Spectrometry

    160

    Tandem Mass Spectrometry

    Free Testosterone

    45

    Beckman Access/Dxi

    84

    76

    171

    149

    17OH-Progesterone

    74nmol/L

    69

    nmol/L

    nmol/L

    194

    Siemens Immulite 2000/2500 [X]

    Beckman Access/Dxi

    121

    216

    Testosterone [female]

    Testosterone [male]

    117

    Our method update service is web-based and is accessed online via the 'edit' button on the 'Results and Reports page'. You can select from a dropdown of methods or select the default option from the major manufacturer’s products*. *If you are not using the system according to the manufacturer's instructions, please select the in-house category within your system's method principle.

    A s

    core

    Graphic Equalizer Plot of A scoresAll laboratories

    P-BWATSCR2CO

    RT

    UF

    C

    17P

    PR

    OG E2

    E2H

    I

    FT

    ES

    MT

    ES

    FR

    EE

    T

    SH

    BG

    AN

    DR

    DH

    AS

    ALD

    O

    0

    50

    100

    150

    200

    250

    300

    A s

    core

    Method Graphic Equalizer Plot of A scoresfor your methods

    P-BWATSCM2CO

    RT

    UF

    C

    17P

    PR

    OG E2

    E2H

    I

    FT

    ES

    MT

    ES

    FR

    EE

    T

    SH

    BG

    AN

    DR

    DH

    AS

    ALD

    O

    0

    50

    100

    150

    200

    250

    300

    of 60

  • Date : 05-Jun-2018

    Cumulative Summary

    Page 7

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    B s

    core

    C score

    Cortisol B score is -8.3 and C score is 12.9Your method is Beckman Access/Dxi

    P-ABCPENR (ABCPENR)

    0 5 10 15 20 25

    -30

    -20

    -10

    0

    10

    20

    30

    Your labYour methodOverall (n=254)Your method (n=22)

    B s

    core

    C score

    Urinary Free Cortisol B score is +62.0 and C score is 76.7Your method is Siemens Immulite 2000/2500 [X]

    P-ABCPENR (ABCPENR)

    0 25 50 75 100

    -25

    0

    25

    50

    75

    100

    Your labYour methodOverall (n=81)Your method (n=3)

    B s

    core

    C score

    17OH-Progesterone B score is -4.9 and C score is 17.9Your method is Tandem Mass Spectrometry

    P-ABCPENR (ABCPENR)

    0 20 40 60 80

    -60

    -40

    -20

    0

    20

    40

    60

    Your labYour methodOverall (n=70)Your method (n=46)

    B s

    core

    C score

    Progesterone B score is -1.7 and C score is 15.5Your method is Beckman Access/Dxi

    P-ABCPENR (ABCPENR)

    0 5 10 15 20 25

    -25-20-15-10

    -505

    10152025

    Your labYour methodOverall (n=240)Your method (n=21)

    B s

    core

    C score

    Oestradiol [E2] B score is +51.2 and C score is 48.5Your method is Beckman Access/Dxi

    P-ABCPENR (ABCPENR)

    0 10 20 30 40

    -40

    -30

    -20

    -10

    0

    10

    20

    30

    40

    Your labYour methodOverall (n=242)Your method (n=18)

    B s

    core

    C score

    Oestradiol [high-level] B score is +19.3 and C score is 14.7Your method is Beckman Access/Dxi

    P-ABCPENR (ABCPENR)

    0 10 20 30 40

    -40

    -30

    -20

    -10

    0

    10

    20

    30

    40

    Your labYour methodOverall (n=110)Your method (n=9)

    B s

    core

    C score

    Testosterone [female] B score is +58.6 and C score is 65.4Your method is Beckman Access/Dxi

    P-ABCPENR (ABCPENR)

    0 20 40 60 80

    -40

    -20

    0

    20

    40

    60

    80

    Your labYour methodOverall (n=213)Your method (n=13)

    B s

    core

    C score

    Testosterone [male] B score is -14.3 and C score is 8.3Your method is Beckman Access/Dxi

    P-ABCPENR (ABCPENR)

    0 5 10 15 20 25

    -30

    -20

    -10

    0

    10

    20

    30

    Your labYour methodOverall (n=216)Your method (n=15)

    B s

    core

    C score

    Free Testosterone B score is -6.8 and C score is 14.3Your method is Beckman Access/Dxi [2SF]

    P-ABCPENR (ABCPENR)

    0 5 10 15 20 25 30

    -30

    -20

    -10

    0

    10

    20

    30

    Your labYour methodOverall (n=38)Your method (n=1)

    B s

    core

    C score

    SHBG B score is -8.8 and C score is 4.8Your method is Beckman Access/Dxi

    P-ABCPENR (ABCPENR)

    0 5 10 15 20

    -30

    -20

    -10

    0

    10

    20

    30

    Your labYour methodOverall (n=158)Your method (n=14)

    of 60

  • Date : 05-Jun-2018

    Cumulative Summary

    Page 8

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    B s

    core

    C score

    Androstenedione B score is +0.7 and C score is 15.3Your method is Tandem Mass Spectrometry

    P-ABCPENR (ABCPENR)

    0 20 40 60 80 100 120 140

    -40

    -20

    0

    20

    40

    60

    80

    100

    120

    140

    Your labYour methodOverall (n=85)Your method (n=43)

    B s

    core

    C score

    DHA sulphate B score is +10.5 and C score is 4.2Your method is Tandem Mass Spectrometry

    P-ABCPENR (ABCPENR)

    0 5 10 15 20 25 30

    -40

    -30

    -20

    -10

    0

    10

    20

    30

    40

    Your labYour methodOverall (n=83)Your method (n=15)

    B s

    core

    C score

    Aldosterone B score is +61.8 and C score is 56.9Your method is DiaSorin Liaison

    P-ABCPENR (ABCPENR)

    0 10 20 30 40 50 60

    -20

    -10

    0

    10

    20

    30

    40

    50

    60

    Your labYour methodOverall (n=40)Your method (n=14)

    of 60

  • Date : 05-Jun-2018

    Analyte : Cortisol (nmol/L)

    Page 9

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    C546453B

    C522

    PoolSpec.

    Pool C520:Pool C521, ratio 1:2453E453D

    Pooled male serum

    Pool description / Treatments / Additions

    Pooled male serumNormal serum [M]

    C491 Normal serum [M]

    C573453A

    453C

    C572 12.9

    The B limit is +/-

    Your B score is

    The A limit is15.0

    121

    20.0

    Your A score is-8.3

    The C limit is

    200

    Your C score is

    All methods Beckman Access/Dxi

    Specimen : 453A

    400 500 600 700 800Cortisol (nmol/L)

    0

    10

    20

    30

    40

    50

    60

    70

    no. of la

    bora

    tories

    67134

    Roche Cobas/Modular

    602

    Gen II

    22

    Meann

    517

    3.8

    Beckman Access/Dxi

    42

    984.4

    624

    5.9

    8.817

    28

    Siemens ADVIA Centaur 8.2

    Abbott Architect

    13

    109

    540

    62724

    255

    32Siemens Immulite 2000/2500

    SD

    10.8

    Tandem Mass Spectrometry

    53 3.1

    553

    65

    CV(%)

    55

    6.413

    658

    46

    All methods [ALTM] 658

    Your method mean

    (Tandem Mass Spectrometry)Standard Uncertainty

    Accuracy Index

    479

    %bias

    517

    Your result

    277

    Beckman Access/Dxi

    -27.2

    16

    Your specimen:

    Target

    Specimen : 453B

    130 160 190 220 250Cortisol (nmol/L)

    0

    20

    40

    60

    80

    100

    no. of la

    bora

    tories

    20434

    Roche Cobas/Modular

    169

    Gen II

    22

    Meann

    190

    3.6

    Beckman Access/Dxi

    13

    983.9

    157

    11.1

    6.86

    6

    Siemens ADVIA Centaur 8.5

    Abbott Architect

    13

    109

    non-numeric results

    158

    1586

    254

    22Siemens Immulite 2000/2500

    SD

    12.5

    Tandem Mass Spectrometry

    52

    1

    3.9

    203

    21

    CV(%)

    17

    8.113

    163

    13

    All methods [ALTM] 163

    Your method mean

    (Tandem Mass Spectrometry)Standard Uncertainty

    Accuracy Index

    170

    %bias

    190

    Your result

    49

    Beckman Access/Dxi

    +4.6Your specimen:

    5

    Target

    Specimen : 453C

    100 120 140 160 180Cortisol (nmol/L)

    0

    20

    40

    60

    80

    100

    no. of la

    bora

    tories

    15934

    Roche Cobas/Modular

    145

    Gen II

    22

    Meann

    141

    3.5

    Beckman Access/Dxi

    12

    984.0

    142

    8.9

    8.76

    6

    Siemens ADVIA Centaur 9.2

    Abbott Architect

    13

    109

    non-numeric results

    139

    1435

    254

    14Siemens Immulite 2000/2500

    SD

    7.4

    Tandem Mass Spectrometry

    52

    1

    4.2

    154

    11

    CV(%)

    15

    8.213

    143

    12

    All methods [ALTM] 143

    Your method mean

    (Tandem Mass Spectrometry)Standard Uncertainty

    Accuracy Index

    123

    %bias

    141

    Your result

    145

    Beckman Access/Dxi

    -14.1Your specimen:

    4

    Target

    Specimen : 453D

    200 240 280 320 360Cortisol (nmol/L)

    0

    20

    40

    60

    80

    100

    no. of la

    bora

    tories

    32434

    Roche Cobas/Modular

    281

    Gen II

    22

    Meann

    277

    3.7

    Beckman Access/Dxi

    25

    984.2

    273

    7.1

    9.110

    11

    Siemens ADVIA Centaur 6.2

    Abbott Architect

    13

    109

    265

    27410

    255

    21Siemens Immulite 2000/2500

    SD

    8.9

    Tandem Mass Spectrometry

    53 3.7

    296

    25

    CV(%)

    20

    8.913

    281

    25

    All methods [ALTM] 281

    Your method mean

    (Tandem Mass Spectrometry)Standard Uncertainty

    Accuracy Index

    256

    %bias

    277

    Your result

    104

    Beckman Access/Dxi

    -8.9Your specimen:

    9

    Target

    Specimen : 453E

    220 280 340 400 460Cortisol (nmol/L)

    0

    10

    20

    30

    40

    50

    60

    70

    no. of la

    bora

    tories

    36534

    Roche Cobas/Modular

    344

    Gen II

    22

    Meann

    293

    3.8

    Beckman Access/Dxi

    29

    984.3

    364

    7.0

    9.610

    16

    Siemens ADVIA Centaur 7.8

    Abbott Architect

    13

    109

    302

    36514

    255

    23Siemens Immulite 2000/2500

    SD

    11.3

    Tandem Mass Spectrometry

    53 3.3

    329

    39

    CV(%)

    28

    7.713

    374

    28

    All methods [ALTM] 374

    Your method mean

    (Tandem Mass Spectrometry)Standard Uncertainty

    Accuracy Index

    258

    %bias

    293

    Your result

    371

    Beckman Access/Dxi

    -31.1Your specimen:

    11

    Target

    of 60

  • Pool

    (exclusion)

    [Type]

    Distribution 448

    05-Dec-2017

    result target %bias

    Distribution 449

    23-Jan-2018

    result target %bias

    Distribution 450

    27-Feb-2018

    result target %bias

    Distribution 451

    27-Mar-2018

    result target %bias

    Distribution 452

    01-May-2018

    result target %bias

    Distribution 453

    05-Jun-2018

    result target %bias

    60 59 +1.7C551 [F,V]103 116 -10.9C552 [F,V]

    123 143 -14.1C546 [M,R,V]161 145 +11.0 145 148 -1.9C526 [M,R,V]

    149 161 -7.5C561 [F,V]170 163 +4.6C573 [M,V]

    146 182 -19.7C553 [F,V]160 186 -14.0C569 [F,V]187 194 -3.8C567 [F,V]

    206 225 -8.3C562 [F,V]221 234 -5.7C499 [M,V]

    275 250 +9.8C527 [M,R,V]266 257 +3.6C571 [M,V]

    213 262 -18.7C555 [F,V]282 275 +2.4C563 [M,V]

    256 281 -8.9C491 [M,V]261 293 -11.0C570 [F,V]

    267 295 -9.3C564 [M,V]256 302 -15.2C568 [F,V]

    312 339 -8.1C534 [M,V]263 340 -22.5C565 [F,V]

    346 349 -1.0C528 [M,R,V]239 350 -31.8C516 [F,V]318 366 -13.2C519 [M,V]

    258 374 -31.1C522 [F,V]296 386 -23.4C554 [F,V]

    599 561 +6.7C529 [M,R,V]479 658 -27.2C572 [M,V]

    1050 947 +10.9C530 [M,R,V]

    MethodmeanA scoreB scoreC score

    SF1+7.5

    126-8.313.2

    SF1-9.0

    135-9.612.4

    SF1-8.1

    122-8.511.4

    SF1-14.2

    135-9.812.0

    SF1-12.1

    120-8.511.6

    SF1-15.3

    121-8.312.9

    F female only poolM male only poolR RecoveryV no preservative

    Date : 05-Jun-2018

    Analyte : Cortisol (nmol/L)

    Page 10

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    A s

    core

    Cortisol Median and IQRs of A scores

    P-BWABCAT

    AB

    13

    SF

    1

    BO

    5

    CO

    10

    DC

    11

    MS

    2

    0

    50

    100

    150

    200

    250

    300

    All

    Meth

    od

    B s

    core

    Median and IQRs of B score

    P-BWABCBR

    AB

    13

    SF

    1

    BO

    5

    CO

    10

    DC

    11

    MS

    2

    -30

    -20

    -10

    0

    10

    20

    30

    All

    Meth

    od

    B s

    core

    C score

    Cortisol B score is -8.3 and C score is 12.9Your method is Beckman Access/Dxi

    P-ABCPENR

    0 5 10 15 20 25

    -30

    -20

    -10

    0

    10

    20

    30

    Your labYour methodOverall (n=254)Your method (n=22)

    C s

    core

    Median and IQRs of C score

    P-BWABCCR

    AB

    13

    SF

    1

    BO

    5

    CO

    10

    DC

    11

    MS

    2

    0

    5

    10

    15

    20

    25

    All

    Meth

    od

    of 60

  • Date : 05-Jun-2018

    Analyte : Cortisol (nmol/L)

    Page 11

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    1

    Gen II

    Beckman Access/Dxi

    13

    Roche Cobas/Modular

    All methods [ALTM]

    n

    109

    1

    3

    Tandem Mass SpectrometrySiemens Immulite 2000/2500

    13

    4

    2

    34

    1

    non-numeric results

    Not stated, please specify

    98

    OCD (J&J) VITROS

    Siemens ADVIA Centaur

    Tosoh AIA

    53

    255

    DiaSource4

    Randox Evolution

    22

    Gen I

    Abbott Architect

    24624

    10.8

    627

    SDMean

    977

    792

    8.8

    658

    28 4.4

    553

    625

    517

    622

    540 17

    569

    5.9

    490

    6.4

    65

    671

    CV(%)

    453A

    32

    3.146

    8.23.8

    42

    55

    602

    6157

    12.5

    158

    SDMean

    217

    182

    6.8

    163

    6 3.9

    203

    154

    190

    283

    158 6

    170

    11.1

    156

    8.1

    21

    204

    CV(%)

    453B

    22

    3.913

    8.53.6

    13

    17

    169

    5142

    7.4

    143

    SDMean

    202

    183

    8.7

    143

    6 4.0

    154

    142

    141

    177

    139 6

    140

    8.9

    115

    8.2

    11

    159

    CV(%)

    453C

    14

    4.212

    9.23.5

    12

    15

    145

    10273

    8.9

    274

    SDMean

    389

    348

    9.1

    281

    11 4.2

    296

    271

    277

    349

    265 10

    274

    7.1

    244

    8.9

    25

    324

    CV(%)

    453D

    21

    3.725

    6.23.7

    25

    20

    281

    14364

    11.3

    365

    SDMean

    483

    429

    9.6

    374

    16 4.3

    329

    363

    293

    371

    302 10

    343

    7.0

    293

    7.7

    39

    365

    CV(%)

    453E

    23

    3.328

    7.83.8

    29

    28

    344

    Randox Evolution

    Siemens ADVIA Centaur

    51

    13

    Gen I

    All methods

    CO10

    n

    4

    37

    1103

    12Tandem Mass SpectrometryDC11

    AB13

    MS2

    254

    SF1RX4

    101

    Roche Cobas/Modular BO5

    Beckman Access/Dxi

    Siemens Immulite 2000/2500

    Gen II

    22Abbott Architect

    52

    108

    211

    Interquartile rangeMedian

    240

    41

    125

    53

    186 245

    97

    269117

    53

    97

    32100

    82

    47

    41

    A score

    32

    42

    87151

    110102

    91113

    31

    214

    74

    -1.6

    +1.1

    +19.5

    Interquartile rangeMedian

    +21.1

    -3.0

    -5.8

    +2.4

    +17.2 +22.5

    -0.2

    +22.9-8.5

    -1.3

    -8.8

    -4.2+6.0

    -10.1

    -0.2

    -2.6

    B score

    -4.2

    -6.5

    -4.2+11.9

    -7.1-10.1

    -0.9+8.5

    -3.4

    +17.9

    -3.1

    4.2

    9.6

    6.6

    Interquartile rangeMedian

    20.2

    3.4

    11.6

    4.8

    6.1 7.3

    11.1

    20.510.6

    4.5

    6.1

    2.710.0

    5.5

    4.3

    3.4

    C score

    2.8

    3.6

    10.213.7

    6.99.6

    10.711.5

    3.0

    19.4

    5.7

    of 60

  • Date : 05-Jun-2018

    Cortisol (nmol/L)

    Page 12

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    Specim

    en A

    ccura

    cy Index

    Distribution

    Specimen Accuracy Indexfor each specimen in time-window

    P-ABCATS (ABC1)

    448 449 450 451 452 453

    0

    50

    100

    150

    200

    250

    300

    A s

    core

    Distribution

    A score by distributionYour current A score is 121

    P-ABCAT (ABC1)

    442 443 444 445 446 447 448 449 450 451 452 453

    0

    50

    100

    150

    200

    250

    300

    All

    Meth

    od

    Specim

    en %

    bia

    s

    Distribution

    Specimen %biasfor each specimen in time-window

    P-ABCBRS (ABC1)

    448 449 450 451 452 453

    -30

    -20

    -10

    0

    10

    20

    30

    B s

    core

    Distribution

    B score by distributionYour current B score is -8.3

    P-ABCBR (ABC1)

    442 443 444 445 446 447 448 449 450 451 452 453

    -30

    -20

    -10

    0

    10

    20

    30

    All

    Meth

    od

    Diffe

    rence (

    nm

    ol/L)

    Concentration (nmol/L)

    Difference (result - target)for each specimen in time-window

    P-BA1FS1A (ABC1)

    0 200 400 600 800 1000

    -100-80-60-40-20

    020406080

    100

    C s

    core

    Distribution

    C score by distributionYour current C score is 12.9

    P-ABCCR (ABC1)

    442 443 444 445 446 447 448 449 450 451 452 453

    0

    5

    10

    15

    20

    25

    All

    Meth

    od

    Specim

    en %

    bia

    s

    Concentration (nmol/L)

    Specimen %biasfor each specimen in time-window

    P-BA1FS1P (ABC1)

    0 200 400 600 800 1000

    -30

    -20

    -10

    0

    10

    20

    30

    B s

    core

    C score

    Cortisol B score is -8.3 and C score is 12.9Your method is Beckman Access/Dxi

    P-ABCPENR (ABC1)

    0 5 10 15 20 25

    -30

    -20

    -10

    0

    10

    20

    30

    Your labYour methodOverall (n=254)Your method (n=22)

    Specim

    en A

    ccura

    cy Index

    Concentration (nmol/L)

    Specimen Accuracy Indexfor each specimen in time-window

    P-YSTMOD (ABC1)

    0 200 400 600 800 1000

    0

    50

    100

    150

    200

    250

    300 ALTMMLTM

    CV

    %

    Concentration (nmol/L)

    Cortisol Between-laboratory agreementBeckman Access/Dxi

    P-PROFILE (ABC1)

    0 200 400 600 800 1000

    0

    5

    10

    15

    20

    25

    of 60

  • Date : 05-Jun-2018

    Analyte : Urinary Free Cortisol (nmol/L)

    Page 13

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    U428453B

    PoolSpec.

    453D

    Normal urine [M]

    Pool description / Treatments / Additions

    Pool U463 [M] + 200 nmol/L CortisolPool U421 [M] + 138 nmol/L CortisolNormal paediatric urineU453

    U464453A

    453C

    U463 76.7

    The B limit is +/-

    Your B score is

    The A limit is20.0

    323

    20.0

    +62.0Your A score is

    The C limit is

    200

    Your C score is

    All methods Siemens Immulite 2000/2500 [X]

    Specimen : 453A

    0 33 66 99 132Urinary Free Cortisol (nmol/L)

    0

    10

    20

    30

    40

    no. of la

    bora

    tories

    Abbott Architect

    60

    48

    11.9

    3

    Mean

    Tandem Mass Spectrometry

    n

    116

    46

    Roche Cobas/Modular [X]53

    4

    80

    SD

    43.8

    18

    3

    4 7.7

    21

    CV(%)

    36Siemens Immulite 2000/2500 [X]

    All methods [ALTM] 36

    Your method mean

    (Tandem Mass Spectrometry)Standard Uncertainty

    Accuracy Index

    72

    %bias

    60

    Your result

    500

    Siemens Immulite 2000/2500 [X]

    +101.7Your specimen:

    1

    Target

    Specimen : 453B

    172 210 248 286 324Urinary Free Cortisol (nmol/L)

    0

    5

    10

    15

    20

    25

    30

    no. of la

    bora

    tories

    Abbott Architect

    271

    249

    7.0

    3

    Mean

    Tandem Mass Spectrometry

    n

    264

    46

    Roche Cobas/Modular [X]252

    17

    80

    SD

    9.8

    18

    3

    10 3.8

    24

    CV(%)

    241Siemens Immulite 2000/2500 [X]

    All methods [ALTM] 241

    Your method mean

    (Tandem Mass Spectrometry)Standard Uncertainty

    Accuracy Index

    312

    %bias

    271

    Your result

    203

    Siemens Immulite 2000/2500 [X]

    +29.6Your specimen:

    3

    Target

    Specimen : 453C

    102 150 198 246 294Urinary Free Cortisol (nmol/L)

    0

    5

    10

    15

    20

    25

    30

    no. of la

    bora

    tories

    Abbott Architect

    234

    199

    11.9

    3

    Mean

    Tandem Mass Spectrometry

    n

    310

    46

    Roche Cobas/Modular [X]204

    22

    80

    SD

    16.0

    18

    3

    11 5.4

    32

    CV(%)

    184Siemens Immulite 2000/2500 [X]

    All methods [ALTM] 184

    Your method mean

    (Tandem Mass Spectrometry)Standard Uncertainty

    Accuracy Index

    284

    %bias

    234

    Your result

    389

    Siemens Immulite 2000/2500 [X]

    +54.6Your specimen:

    4

    Target

    Specimen : 453D

    0 48 96 144 192Urinary Free Cortisol (nmol/L)

    0

    5

    10

    15

    20

    25

    30

    no. of la

    bora

    tories

    Abbott Architect

    103

    84

    11.0

    3

    Mean

    Tandem Mass Spectrometry

    n

    330

    46

    Roche Cobas/Modular [X]92

    7

    80

    SD

    36.9

    18

    3

    5 5.4

    31

    CV(%)

    66Siemens Immulite 2000/2500 [X]

    All methods [ALTM] 66

    Your method mean

    (Tandem Mass Spectrometry)Standard Uncertainty

    Accuracy Index

    133

    %bias

    103

    Your result

    500

    Siemens Immulite 2000/2500 [X]

    +101.1Your specimen:

    1

    Target

    of 60

  • Pool

    (exclusion)

    [Type]

    Distribution 448

    05-Dec-2017

    result target %bias

    Distribution 449

    23-Jan-2018

    result target %bias

    Distribution 450

    27-Feb-2018

    result target %bias

    Distribution 451

    27-Mar-2018

    result target %bias

    Distribution 452

    01-May-2018

    result target %bias

    Distribution 453

    05-Jun-2018

    result target %bias

    (

  • Date : 05-Jun-2018

    Analyte : Urinary Free Cortisol (nmol/L)

    Page 15

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    18

    Tandem Mass Spectrometry

    80

    Roche Cobas/Modular [X]Siemens Immulite 2000/2500 [X]

    46

    All methods [ALTM]

    3

    n

    3

    Abbott Architect 4

    43.8

    SD

    453A

    Mean

    3660

    21

    CV(%)

    7.7

    11.9

    116

    4

    53

    48

    10

    9.8

    SD

    453B

    Mean

    241271

    24

    CV(%)

    3.8

    7.0

    264

    17

    252

    249

    11

    16.0

    SD

    453C

    Mean

    184234

    32

    CV(%)

    5.4

    11.9

    310

    22

    204

    199

    5

    36.9

    SD

    453D

    Mean

    66103

    31

    CV(%)

    5.4

    11.0

    330

    7

    92

    84

    Immunotech RIA

    MS2

    16

    Siemens Immulite 2000/2500 [X]

    All methods

    48

    n

    DC1133

    AB13

    81

    IB1IT1

    Tandem Mass Spectrometry

    Roche Cobas/Modular [X] BO8

    IBL Elisa

    3

    3Abbott Architect

    63

    233

    212

    Interquartile rangeMedian

    293

    44

    493

    196 268

    296487

    453

    225

    28

    222

    407 370

    38

    235480

    A score

    281

    72

    +5.3

    +32.6

    +27.8

    Interquartile rangeMedian

    +64.2

    +0.1

    +885.0

    +25.1 +44.9

    +64.5+522.9

    +238.9

    +31.0

    -4.1

    +30.0

    +216.4 +143.6

    -1.0

    +33.0+439.1

    B score

    +54.8

    +8.1

    8.2

    46.3

    44.9

    Interquartile rangeMedian

    96.6

    5.8

    1390.4

    42.0 60.8

    97.2605.8

    251.2

    46.0

    4.4

    43.6

    123.6 100.6

    5.4

    47.0491.9

    C score

    81.7

    10.4

    of 60

  • Date : 05-Jun-2018

    Urinary Free Cortisol (nmol/L)

    Page 16

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    Specim

    en A

    ccura

    cy Index

    Distribution

    Specimen Accuracy Indexfor each specimen in time-window

    P-ABCATS (ABC1)

    448 449 450 451 452 453

    0

    50

    100

    150

    200

    250

    300

    A s

    core

    Distribution

    A score by distributionYour current A score is 323

    P-ABCAT (ABC1)

    442 443 444 445 446 447 448 449 450 451 452 453

    0

    50

    100

    150

    200

    250

    300

    All

    Meth

    od

    Specim

    en %

    bia

    s

    Distribution

    Specimen %biasfor each specimen in time-window

    P-ABCBRS (ABC1)

    448 449 450 451 452 453

    -25

    0

    25

    50

    75

    100

    B s

    core

    Distribution

    B score by distributionYour current B score is +62.0

    P-ABCBR (ABC1)

    442 443 444 445 446 447 448 449 450 451 452 453

    -25

    0

    25

    50

    75

    100

    All

    Gro

    up

    Meth

    od

    Diffe

    rence (

    nm

    ol/L)

    Concentration (nmol/L)

    Difference (result - target)for each specimen in time-window

    P-BA1FS1A (ABC1)

    0 200 400 600 800 1000

    -100-80-60-40-20

    020406080

    100

    C s

    core

    Distribution

    C score by distributionYour current C score is 76.7

    P-ABCCR (ABC1)

    442 443 444 445 446 447 448 449 450 451 452 453

    0

    25

    50

    75

    100

    All

    Gro

    up

    Meth

    od

    Specim

    en %

    bia

    s

    Concentration (nmol/L)

    Specimen %biasfor each specimen in time-window

    P-BA1FS1P (ABC1)

    0 200 400 600 800 1000

    -25

    0

    25

    50

    75

    100

    B s

    core

    C score

    Urinary Free Cortisol B score is +62.0 and C score is 76.7Your method is Siemens Immulite 2000/2500 [X]

    P-ABCPENR (ABC1)

    0 25 50 75 100

    -25

    0

    25

    50

    75

    100

    Your labYour methodOverall (n=81)Your method (n=3)

    Specim

    en A

    ccura

    cy Index

    Concentration (nmol/L)

    Specimen Accuracy Indexfor each specimen in time-window

    P-YSTMOD (ABC1)

    0 200 400 600 800 1000

    0

    50

    100

    150

    200

    250

    300 ALTMMLTM

    CV

    %

    Concentration (nmol/L)

    Urinary Free Cortisol Between-laboratory agreementSiemens Immulite 2000/2500 [X]

    P-PROFILE (ABC1)

    0 200 400 600 800 1000

    0

    10

    20

    30

    40

    50

    of 60

  • Date : 05-Jun-2018

    Analyte : 17OH-Progesterone (nmol/L)

    Page 17

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    H456453B

    H459

    PoolSpec.

    Pool H454 + 25.5 nmol/L 17OHP + 50 nmol/L 11-Deoxycorticosterone453E453D

    Normal serum [F]

    Pool description / Treatments / Additions

    Pool H454 + 5 nmol/L 11-DeoxycorticosteronePool H454 + 50 nmol/L 11-DeoxycorticosteronePool H454 + 25.5 nmol/L 17OHP + 5 nmol/L 11-DeoxycorticosteroneH458

    H455453A

    453C

    H454 17.9

    The B limit is +/-

    Your B score is

    The A limit is15.0

    67

    20.0

    -4.9Your A score is

    The C limit is

    200

    Your C score is

    All methods Tandem Mass Spectrometry

    Specimen : 453A

    0.0 0.9 1.8 2.7 3.617OH-Progesterone (nmol/L)

    0

    5

    10

    15

    20

    25

    30

    no. of la

    bora

    tories

    CIS Coated Tube

    1.87

    1.43

    21.8

    0.68

    Mean

    Tandem Mass Spectrometry

    n

    22.9

    39

    2.971.05

    non-numeric results0.26

    61

    4

    SD

    38.9

    3

    3DiaSource 7

    0.56

    CV(%)

    1.21IBL Elisa

    All methods [ALTM] 1.21

    Your method mean

    (Tandem Mass Spectrometry)Standard Uncertainty

    Accuracy Index

    1.3

    %bias

    1.21

    Your result

    36

    Tandem Mass Spectrometry

    +7.8Your specimen:

    0.06

    Target

    Specimen : 453B

    0.0 0.9 1.8 2.7 3.617OH-Progesterone (nmol/L)

    0

    5

    10

    15

    20

    no. of la

    bora

    tories

    CIS Coated Tube

    1.66

    1.38

    18.8

    0.47

    Mean

    Tandem Mass Spectrometry

    n

    17.8

    40

    2.641.03

    non-numeric results0.23

    62

    3

    SD

    35.4

    3

    3DiaSource 7

    0.49

    CV(%)

    1.20IBL Elisa

    All methods [ALTM] 1.20

    Your method mean

    (Tandem Mass Spectrometry)Standard Uncertainty

    Accuracy Index

    1.4

    %bias

    1.20

    Your result

    78

    Tandem Mass Spectrometry

    +16.7Your specimen:

    0.05

    Target

    Specimen : 453C

    0.0 0.9 1.8 2.7 3.617OH-Progesterone (nmol/L)

    0

    5

    10

    15

    20

    25

    30

    no. of la

    bora

    tories

    CIS Coated Tube

    1.72

    1.53

    35.2

    0.43

    Mean

    Tandem Mass Spectrometry

    n

    15.7

    37

    2.731.14

    non-numeric results0.45

    59

    6

    SD

    46.6

    3

    3DiaSource 7

    0.71

    CV(%)

    1.29IBL Elisa

    All methods [ALTM] 1.29

    Your method mean

    (Tandem Mass Spectrometry)Standard Uncertainty

    Accuracy Index

    1.4

    %bias

    1.29

    Your result

    40

    Tandem Mass Spectrometry

    +8.5Your specimen:

    0.10

    Target

    Specimen : 453D

    8 16 24 32 4017OH-Progesterone (nmol/L)

    0

    5

    10

    15

    20

    no. of la

    bora

    tories

    CIS Coated Tube

    14.5

    23.3

    10.6

    1.7

    Mean

    Tandem Mass Spectrometry

    n

    7.4

    44

    22.613.3

    2.7

    66

    SD

    20.1

    3

    3DiaSource 7

    4.7

    CV(%)

    25.2IBL Elisa

    All methods [ALTM] 25.2

    Your method mean

    (Tandem Mass Spectrometry)Standard Uncertainty

    Accuracy Index

    28.6

    %bias

    25.2

    Your result

    76

    Tandem Mass Spectrometry

    +13.6Your specimen:

    0.54

    Target

    Specimen : 453E

    8 16 24 32 4017OH-Progesterone (nmol/L)

    0

    5

    10

    15

    20

    no. of la

    bora

    tories

    CIS Coated Tube

    14.7

    23.5

    11.0

    1.6

    Mean

    Tandem Mass Spectrometry

    n

    7.7

    44

    21.013.6

    2.8

    66

    SD

    20.9

    3

    3DiaSource 7

    4.9

    CV(%)

    25.3IBL Elisa

    All methods [ALTM] 25.3

    Your method mean

    (Tandem Mass Spectrometry)Standard Uncertainty

    Accuracy Index

    31.7

    %bias

    25.3

    Your result

    140

    Tandem Mass Spectrometry

    +25.2Your specimen:

    0.56

    Target

    of 60

  • Pool

    (exclusion)

    [Type]

    Distribution 448

    05-Dec-2017

    result target %bias

    Distribution 449

    23-Jan-2018

    result target %bias

    Distribution 450

    27-Feb-2018

    result target %bias

    Distribution 451

    27-Mar-2018

    result target %bias

    Distribution 452

    01-May-2018

    result target %bias

    Distribution 453

    05-Jun-2018

    result target %bias

    0.5 0.39 +29.6H451 [F,V]0.5 0.46 +9.4H449 [F,V]0.5 0.50 -0.3H452 [F,V]

    0.2 0.52 -61.5H445 [F,V]0.5 0.70 -28.8H444 [F,V]

    1.4 1.20 +16.7H455 [F,V]1.3 1.21 +7.8H454 [F,V]1.4 1.29 +8.5H456 [F,V]

    1.5 1.50 -0.3H450 [F,V]1.4 1.83 -23.5H447 [M,V]

    2.2 2.34 -5.8H453 [M,V]2.0 2.59 -22.8H446 [M,V]

    2.2 3.08 -28.6H423 [F,V]2.8 3.57 -21.7 3.8 3.57 +6.3H404 [F,V]

    4.6 4.77 -3.6H405 [F,V]4.3 5.41 -20.5H448 [F,V]

    5.8 7.2 -19.6 6.9 7.1 -3.0H406 [F,V]7.8 8.4 -7.3H407 [F,V]

    9.5 11.4 -16.9 10.1 9.8 +3.4H408 [F,R]11.6 12.2 -5.1 15.5 12.7 +22.3H409 [F,V]14.4 15.9 -9.2H411 17.3 17.9 -3.4 17.8 18.7 -4.7H410

    28.6 25.2 +13.6H458 [F,V]31.7 25.3 +25.2H459 [F,V]

    41.6 49.0 -15.1H396 [F,V]

    MethodmeanA scoreB scoreC score

    MS2-20.4

    63-12.6

    7.0

    MS2-31.4

    74-14.9

    8.7

    MS2+6.5

    74-13.310.6

    MS2-5.7

    70-12.511.6

    MS2+4.9

    64-9.713.8

    MS2+14.4

    67-4.917.9

    F female only poolM male only poolR RecoveryV no preservative

    Date : 05-Jun-2018

    Analyte : 17OH-Progesterone (nmol/L)

    Page 18

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    A s

    core

    17OH-Progesterone Median and IQRs of A scores

    P-BWABCAT

    DE

    1

    IB1

    MS

    2

    0

    50

    100

    150

    200

    250

    300

    All

    Meth

    od

    B s

    core

    Median and IQRs of B score

    P-BWABCBR

    DE

    1

    IB1

    MS

    2

    -60

    -40

    -20

    0

    20

    40

    60

    All

    Meth

    od

    B s

    core

    C score

    17OH-Progesterone B score is -4.9 and C score is 17.9Your method is Tandem Mass Spectrometry

    P-ABCPENR

    0 20 40 60 80

    -60

    -40

    -20

    0

    20

    40

    60

    Your labYour methodOverall (n=70)Your method (n=46)

    C s

    core

    Median and IQRs of C score

    P-BWABCCR

    DE

    1

    IB1

    MS

    2

    0

    20

    40

    60

    80

    All

    Meth

    od

    of 60

  • Date : 05-Jun-2018

    Analyte : 17OH-Progesterone (nmol/L)

    Page 19

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    3

    Tandem Mass Spectrometry

    66

    DiaSourceIBL Elisa

    44

    All methods [ALTM]

    6

    7

    n

    non-numeric results

    3

    CIS Coated Tube

    38.9

    SDMean

    1.21

    453A

    1.870.68

    0.56

    CV(%)

    2.97

    21.8

    22.9

    0.26

    1.05

    1.43 35.4

    SDMean

    1.20

    453B

    1.660.47

    0.49

    CV(%)

    2.64

    18.8

    17.8

    0.23

    1.03

    1.38 46.6

    SDMean

    1.29

    453C

    1.720.43

    0.71

    CV(%)

    2.73

    35.2

    15.7

    0.45

    1.14

    1.53 20.1

    SDMean

    25.2

    453D

    14.51.7

    4.7

    CV(%)

    22.6

    10.6

    7.4

    2.7

    13.3

    23.3 20.9

    SDMean

    25.3

    453E

    14.71.6

    4.9

    CV(%)

    21.0

    11.0

    7.7

    2.8

    13.6

    23.5

    IBL Elisa

    4

    All methods

    n

    465

    CI3

    70

    DE1IB1

    Tandem Mass Spectrometry MS2

    DiaSource 7CIS Coated Tube 156

    175

    A score

    Interquartile range

    56

    Median

    17437

    308

    46

    176

    183299 294

    75143

    171

    74

    -20.6

    +13.7

    B score

    Interquartile range

    0.0

    Median

    -8.6-4.9

    +84.4

    -7.7

    -16.7

    +4.8+79.3 +74.6

    +7.6-14.2

    -18.1

    +1.0

    28.8

    41.4

    C score

    Interquartile range

    9.3

    Median

    46.26.8

    110.2

    8.1

    40.5

    57.3107.3 97.4

    16.532.4

    34.5

    17.3

    of 60

  • Date : 05-Jun-2018

    17OH-Progesterone (nmol/L)

    Page 20

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    Specim

    en A

    ccura

    cy Index

    Distribution

    Specimen Accuracy Indexfor each specimen in time-window

    P-ABCATS (ABC1)

    448 449 450 451 452 453

    0

    50

    100

    150

    200

    250

    300

    A s

    core

    Distribution

    A score by distributionYour current A score is 67

    P-ABCAT (ABC1)

    442 443 444 445 446 447 448 449 450 451 452 453

    0

    50

    100

    150

    200

    250

    300

    All

    Meth

    od

    Specim

    en %

    bia

    s

    Distribution

    Specimen %biasfor each specimen in time-window

    P-ABCBRS (ABC1)

    448 449 450 451 452 453

    -60

    -40

    -20

    0

    20

    40

    60

    B s

    core

    Distribution

    B score by distributionYour current B score is -4.9

    P-ABCBR (ABC1)

    442 443 444 445 446 447 448 449 450 451 452 453

    -60

    -40

    -20

    0

    20

    40

    60

    All

    Meth

    od

    Diffe

    rence (

    nm

    ol/L)

    Concentration (nmol/L)

    Difference (result - target)for each specimen in time-window

    P-BA1FS1A (ABC1)

    0 20 40 60 80 100 120

    -20

    -15

    -10

    -5

    0

    5

    10

    15

    20

    C s

    core

    Distribution

    C score by distributionYour current C score is 17.9

    P-ABCCR (ABC1)

    442 443 444 445 446 447 448 449 450 451 452 453

    0

    20

    40

    60

    80

    All

    Meth

    od

    Specim

    en %

    bia

    s

    Concentration (nmol/L)

    Specimen %biasfor each specimen in time-window

    P-BA1FS1P (ABC1)

    0 20 40 60 80 100 120

    -60

    -40

    -20

    0

    20

    40

    60

    B s

    core

    C score

    17OH-Progesterone B score is -4.9 and C score is 17.9Your method is Tandem Mass Spectrometry

    P-ABCPENR (ABC1)

    0 20 40 60 80

    -60

    -40

    -20

    0

    20

    40

    60

    Your labYour methodOverall (n=70)Your method (n=46)

    Specim

    en A

    ccura

    cy Index

    Concentration (nmol/L)

    Specimen Accuracy Indexfor each specimen in time-window

    P-YSTMOD (ABC1)

    0 20 40 60 80 100 120

    0

    50

    100

    150

    200

    250

    300 ALTMMLTM

    CV

    %

    Concentration (nmol/L)

    17OH-Progesterone Between-laboratory agreementTandem Mass Spectrometry

    P-PROFILE (ABC1)

    0 20 40 60 80 100 120

    0

    10

    20

    30

    40

    50

    of 60

  • Date : 05-Jun-2018

    Analyte : Progesterone (nmol/L)

    Page 21

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    P481453B

    P483

    PoolSpec.

    Normal serum [F]453E453D

    Pool 487:Pool 488, ratio 3:1

    Pool description / Treatments / Additions

    Normal serum [F]Normal serum [F]Normal serum [F]P482

    P480453A

    453C

    P493 15.5

    The B limit is +/-

    Your B score is

    The A limit is15.0

    84

    15.0

    -1.7Your A score is

    The C limit is

    200

    Your C score is

    All methods Beckman Access/Dxi

    Specimen : 453A

    14 18 22 26 30Progesterone (nmol/L)

    0

    20

    40

    60

    80

    no. of la

    bora

    tories

    1.123.7

    Beckman Access/Dxi Alt Calib

    23.1

    20.9

    21

    Meann

    18.1

    Roche Cobas/Modular

    Beckman Access/Dxi

    6.9

    Randox Evolution43

    Siemens Immulite 2000/2500

    7.41.2

    OCD (J&J) VITROS

    104Siemens ADVIA Centaur

    Abbott Architect

    18.0

    3

    23.3

    22.6

    23.5

    238

    7.137

    SD

    8.5

    7

    54 5.2

    1.7

    2.0

    CV(%)

    4.524.1

    1.2

    1.3

    All methods [ALTM] 23.1

    Your method mean

    (ALTM)Standard Uncertainty

    Accuracy Index

    19.5

    %bias

    18.1

    Your result

    171

    Beckman Access/Dxi

    -15.7Your specimen:

    0.2

    Target

    Specimen : 453B

    6 9 12 15 18Progesterone (nmol/L)

    0

    20

    40

    60

    80

    100

    120

    no. of la

    bora

    tories

    0.611.7

    Beckman Access/Dxi Alt Calib

    11.7

    12.5

    21

    Meann

    10.0

    Roche Cobas/Modular

    Beckman Access/Dxi

    2.4

    Randox Evolution43

    Siemens Immulite 2000/2500

    8.50.7

    OCD (J&J) VITROS

    104Siemens ADVIA Centaur

    Abbott Architect

    10.3

    3

    12.1

    12.6

    11.2

    238

    9.337

    SD

    7.7

    7

    54 5.5

    1.1

    0.9

    CV(%)

    5.311.9

    0.3

    0.8

    All methods [ALTM] 11.7

    Your method mean

    (ALTM)Standard Uncertainty

    Accuracy Index

    12.0

    %bias

    10.0

    Your result

    22

    Beckman Access/Dxi

    +2.9Your specimen:

    0.1

    Target

    Specimen : 453C

    14 17 20 23 26Progesterone (nmol/L)

    0

    20

    40

    60

    80

    no. of la

    bora

    tories

    0.819.5

    Beckman Access/Dxi Alt Calib

    19.4

    23.6

    21

    Meann

    17.9

    Roche Cobas/Modular

    Beckman Access/Dxi

    9.2

    Randox Evolution43

    Siemens Immulite 2000/2500

    9.61.0

    OCD (J&J) VITROS

    104Siemens ADVIA Centaur

    Abbott Architect

    16.7

    3

    20.2

    22.9

    17.8

    238

    9.037

    SD

    7.0

    7

    54 4.9

    1.7

    1.4

    CV(%)

    4.018.6

    1.5

    1.7

    All methods [ALTM] 19.4

    Your method mean

    (ALTM)Standard Uncertainty

    Accuracy Index

    21.9

    %bias

    17.9

    Your result

    129

    Beckman Access/Dxi

    +13.1Your specimen:

    0.1

    Target

    Specimen : 453D

    18 21 24 27 30Progesterone (nmol/L)

    0

    10

    20

    30

    40

    50

    60

    70

    no. of la

    bora

    tories

    0.924.3

    Beckman Access/Dxi Alt Calib

    23.7

    23.4

    21

    Meann

    20.3

    Roche Cobas/Modular

    Beckman Access/Dxi

    4.1

    Randox Evolution43

    Siemens Immulite 2000/2500

    11.31.1

    OCD (J&J) VITROS

    104Siemens ADVIA Centaur

    Abbott Architect

    19.4

    3

    24.8

    24.4

    21.4

    238

    7.837

    SD

    7.8

    7

    54 4.4

    1.8

    1.9

    CV(%)

    3.822.6

    0.8

    2.3

    All methods [ALTM] 23.7

    Your method mean

    (ALTM)Standard Uncertainty

    Accuracy Index

    23.4

    %bias

    20.3

    Your result

    15

    Beckman Access/Dxi

    -1.3Your specimen:

    0.2

    Target

    Specimen : 453E

    32 37 42 47 52Progesterone (nmol/L)

    0

    10

    20

    30

    40

    50

    60

    70

    no. of la

    bora

    tories

    1.642.5

    Beckman Access/Dxi Alt Calib

    41.7

    39.0

    21

    Meann

    34.6

    Roche Cobas/Modular

    Beckman Access/Dxi

    10.1

    Randox Evolution43

    Siemens Immulite 2000/2500

    9.81.9

    OCD (J&J) VITROS

    104Siemens ADVIA Centaur

    Abbott Architect

    32.5

    3

    42.6

    42.2

    44.5

    238

    6.737

    SD

    7.0

    7

    54 4.5

    2.8

    2.9

    CV(%)

    3.842.3

    3.3

    3.4

    All methods [ALTM] 41.7

    Your method mean

    (ALTM)Standard Uncertainty

    Accuracy Index

    40.7

    %bias

    34.6

    Your result

    25

    Beckman Access/Dxi

    -2.4Your specimen:

    0.2

    Target

    of 60

  • Pool

    (exclusion)

    [Type]

    Distribution 448

    05-Dec-2017

    result target %bias

    Distribution 449

    23-Jan-2018

    result target %bias

    Distribution 450

    27-Feb-2018

    result target %bias

    Distribution 451

    27-Mar-2018

    result target %bias

    Distribution 452

    01-May-2018

    result target %bias

    Distribution 453

    05-Jun-2018

    result target %bias

    (0.8)(P510) [F,V](0.7)(P509) [F,V]

    (1.1)(P504) [F,V](0.9)(P503) [F,V]

    (1.1)(P507) [F,V]1.4 1.2 +16.5P511 [V]

    2.0 2.1 -3.9P502 [F,V]4.9 3.9 +25.2P508 [F,V]

    5.4 4.0 +36.1P479 [F,V]9.9 8.7 +14.3P465 [F,V]

    12.0 11.7 +2.9P480 [F,V]10.5 15.1 -30.6P489 [F,R,V]12.7 15.3 -17.0 15.6 15.5 +0.4P468 [F,V]

    21.9 19.4 +13.1P481 [F,V]21.2 20.1 +5.4P467 [F,V]

    19.8 22.9 -13.4P512 [F,V]21.2 22.9 -7.5P462 [F,V]

    19.5 23.1 -15.7P493 [F,V]23.4 23.7 -1.3P482 [F,V]

    24.8 26.8 -7.4P513 [F,V]22.2 29.0 -23.4P490 [F,R,V]

    27.8 29.8 -6.7P514 [F,V]27.3 30.8 -11.3P466 [F,V]

    31.0 32.4 -4.3P515 [F,V]34.9 35.3 -1.0P516 [F,V]

    40.7 41.7 -2.4P483 [F,V]69.2 64.7 +6.9P506 [F,V]

    100.5 86.3 +16.5P505 [F,V]59.4 86.3 -31.1P492 [F,R,V]

    MethodmeanA scoreB scoreC score

    SF1-22.7

    82-10.0

    9.2

    SF1+6.5

    66-7.88.2

    SF1+20.9

    64-4.713.2

    SF1-6.6

    72-6.39.1

    SF1+9.7

    78-3.214.0

    SF1-0.7

    84-1.715.5

    F female only poolR RecoveryV no preservative

    Date : 05-Jun-2018

    Analyte : Progesterone (nmol/L)

    Page 22

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    A s

    core

    Progesterone Median and IQRs of A scores

    P-BWABCAT

    AB

    13

    SF

    1

    BO

    5

    CO

    10

    DC

    11

    0

    50

    100

    150

    200

    250

    300

    All

    Meth

    od

    B s

    core

    Median and IQRs of B score

    P-BWABCBR

    AB

    13

    SF

    1

    BO

    5

    CO

    10

    DC

    11

    -25-20-15-10

    -505

    10152025

    All

    Meth

    od

    B s

    core

    C score

    Progesterone B score is -1.7 and C score is 15.5Your method is Beckman Access/Dxi

    P-ABCPENR

    0 5 10 15 20 25

    -25-20-15-10

    -505

    10152025

    Your labYour methodOverall (n=240)Your method (n=21)

    C s

    core

    Median and IQRs of C score

    P-BWABCCR

    AB

    13

    SF

    1

    BO

    5

    CO

    10

    DC

    11

    0

    5

    10

    15

    20

    25

    All M

    eth

    od

    of 60

  • Date : 05-Jun-2018

    Analyte : Progesterone (nmol/L)

    Page 23

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    OCD (J&J) VITROS

    Beckman Access/Dxi

    Siemens ADVIA Centaur

    All methods [ALTM]

    n

    37104

    Randox Evolution

    54

    238

    Beckman Access/Dxi Alt Calib4

    7

    3Roche Cobas/Modular

    21

    Siemens Immulite 2000/2500

    3

    Abbott Architect

    1.218.0

    8.5

    4.5

    SDMean

    22.6

    7.1

    7.4

    1.124.1

    20.9

    18.1

    23.7

    23.3 1.2

    23.5

    1.7

    2.0

    453A

    CV(%)

    5.21.3

    6.9

    23.1

    0.310.3

    7.7

    5.3

    SDMean

    12.6

    9.3

    8.5

    0.611.9

    12.5

    10.0

    11.7

    12.1 0.7

    11.2

    1.1

    0.9

    453B

    CV(%)

    5.50.8

    2.4

    11.7

    1.516.7

    7.0

    4.0

    SDMean

    22.9

    9.0

    9.6

    0.818.6

    23.6

    17.9

    19.5

    20.2 1.0

    17.8

    1.7

    1.4

    453C

    CV(%)

    4.91.7

    9.2

    19.4

    0.819.4

    7.8

    3.8

    SDMean

    24.4

    7.8

    11.3

    0.922.6

    23.4

    20.3

    24.3

    24.8 1.1

    21.4

    1.8

    1.9

    453D

    CV(%)

    4.42.3

    4.1

    23.7

    3.332.5

    7.0

    3.8

    SDMean

    42.2

    6.7

    9.8

    1.642.3

    39.0

    34.6

    42.5

    42.6 1.9

    44.5

    2.8

    2.9

    453E

    CV(%)

    4.53.4

    10.1

    41.7

    Beckman Access/Dxi Alt Calib

    BO5

    53

    Siemens Immulite 2000/2500

    Randox Evolution

    All methods

    105

    n

    RX4

    MS2

    Siemens ADVIA Centaur

    43

    Tandem Mass SpectrometryDC11

    37

    AB13

    7

    240

    SF1SF4

    Roche Cobas/Modular

    OCD (J&J) VITROS AM12

    Beckman Access/Dxi

    CO10

    3

    21

    3

    Abbott Architect

    43

    72

    88

    Interquartile rangeMedian

    106

    35

    171

    76

    82 104

    150

    112149

    134

    39

    2957

    31

    71

    125

    A score

    117

    32

    14277

    51126

    14465

    59

    84

    45

    +2.6

    +4.0

    -2.0

    Interquartile rangeMedian

    +6.0

    +1.0

    -14.1

    -5.0

    -4.5 +3.1

    -14.0

    +8.7-17.1

    +6.8

    +3.1

    -0.4+2.4

    +1.4

    -7.7

    +4.7

    B score

    +1.7

    -1.2

    -16.6+5.9

    +4.6-18.5

    -15.9+4.4

    -8.5

    -3.6

    +1.5

    5.8

    8.8

    15.9

    Interquartile rangeMedian

    11.8

    4.9

    11.1

    8.6

    14.1 16.8

    9.2

    12.68.7

    31.5

    5.4

    4.18.8

    4.3

    8.5

    25.2

    C score

    19.4

    4.6

    6.611.9

    6.76.9

    7.210.7

    8.0

    11.6

    6.1

    of 60

  • Date : 05-Jun-2018

    Progesterone (nmol/L)

    Page 24

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    Specim

    en A

    ccura

    cy Index

    Distribution

    Specimen Accuracy Indexfor each specimen in time-window

    P-ABCATS (ABC1)

    448 449 450 451 452 453

    0

    50

    100

    150

    200

    250

    300

    A s

    core

    Distribution

    A score by distributionYour current A score is 84

    P-ABCAT (ABC1)

    442 443 444 445 446 447 448 449 450 451 452 453

    0

    50

    100

    150

    200

    250

    300

    All

    Meth

    od

    Specim

    en %

    bia

    s

    Distribution

    Specimen %biasfor each specimen in time-window

    P-ABCBRS (ABC1)

    448 449 450 451 452 453

    -25-20-15-10

    -505

    10152025

    B s

    core

    Distribution

    B score by distributionYour current B score is -1.7

    P-ABCBR (ABC1)

    442 443 444 445 446 447 448 449 450 451 452 453

    -25-20-15-10

    -505

    10152025

    All

    Meth

    od

    Diffe

    rence (

    nm

    ol/L)

    Concentration (nmol/L)

    Difference (result - target)for each specimen in time-window

    P-BA1FS1A (ABC1)

    0 10 20 30 40 50 60 70 80 90 100

    -5-4-3-2-1012345

    C s

    core

    Distribution

    C score by distributionYour current C score is 15.5

    P-ABCCR (ABC1)

    442 443 444 445 446 447 448 449 450 451 452 453

    0

    5

    10

    15

    20

    25

    All M

    eth

    od

    Specim

    en %

    bia

    s

    Concentration (nmol/L)

    Specimen %biasfor each specimen in time-window

    P-BA1FS1P (ABC1)

    0 10 20 30 40 50 60 70 80 90 100

    -25-20-15-10

    -505

    10152025

    B s

    core

    C score

    Progesterone B score is -1.7 and C score is 15.5Your method is Beckman Access/Dxi

    P-ABCPENR (ABC1)

    0 5 10 15 20 25

    -25-20-15-10

    -505

    10152025

    Your labYour methodOverall (n=240)Your method (n=21)

    Specim

    en A

    ccura

    cy Index

    Concentration (nmol/L)

    Specimen Accuracy Indexfor each specimen in time-window

    P-YSTMOD (ABC1)

    0 10 20 30 40 50 60 70 80 90 100

    0

    50

    100

    150

    200

    250

    300 ALTMMLTM

    CV

    %

    Concentration (nmol/L)

    Progesterone Between-laboratory agreementBeckman Access/Dxi

    P-PROFILE (ABC1)

    0 10 20 30 40 50 60 70 80 90 100

    0

    5

    10

    15

    20

    25

    of 60

  • Date : 05-Jun-2018

    Analyte : Oestradiol [E2] (pmol/L)

    Page 25

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    E522453B

    E524

    PoolSpec.

    Pool E520:Pool E521, ratio 1:3453E453D

    Normal serum [F]

    Pool description / Treatments / Additions

    Stripped serumPool E520:Pool E521, ratio 3:1Pool E520:Pool E521, ratio 1:1E523

    E521453A

    453C

    E520 48.5

    The B limit is +/-

    Your B score is

    The A limit is15.0

    256

    20.0

    +51.2Your A score is

    The C limit is

    200

    Your C score is

    All methods Beckman Access/Dxi

    Specimen : 453A

    100 130 160 190 220Oestradiol [E2] (pmol/L)

    0

    10

    20

    30

    40

    50

    no. of la

    bora

    tories

    10.318

    OCD (J&J) VITROS

    162

    163

    19

    Meann

    170

    33

    Beckman Access/Dxi

    16106 9.8

    14

    41.015

    Roche Cobas/Modular179

    5

    Abbott Architect

    12

    4

    160

    138

    Siemens ADVIA Centaur

    240

    Tandem Mass Spectrometry149

    SD

    12.9

    151

    48 9.4

    24.7

    21

    CV(%)

    Siemens Immulite 2000/2500 377.8

    70

    All methods [ALTM] 162

    Your method mean

    (ALTM)Standard Uncertainty

    Accuracy Index

    157

    %bias

    170

    Your result

    15

    Beckman Access/Dxi

    -3.2Your specimen:

    2

    Target

    Specimen : 453B

    0 32 64 96 128Oestradiol [E2] (pmol/L)

    0

    5

    10

    15

    20

    no. of la

    bora

    tories

    Abbott Architect

    Roche Cobas/Modular

    39

    Siemens ADVIA Centaur

    8

    Mean

    22

    n

    36

    170

    28Beckman Access/Dxi

    51

    non-numeric results4

    38.5

    64

    45

    SD

    53.0

    23127.3

    14 27.0

    21

    CV(%)

    1146

    All methods [ALTM] 39

    Your method mean

    (ALTM)Standard Uncertainty

    Accuracy Index%bias

    36

    Your result

    Beckman Access/Dxi

    Your specimen:

    3

    Target

  • Pool

    (exclusion)

    [Type]

    Distribution 448

    05-Dec-2017

    result target %bias

    Distribution 449

    23-Jan-2018

    result target %bias

    Distribution 450

    27-Feb-2018

    result target %bias

    Distribution 451

    27-Mar-2018

    result target %bias

    Distribution 452

    01-May-2018

    result target %bias

    Distribution 453

    05-Jun-2018

    result target %bias

  • Date : 05-Jun-2018

    Analyte : Oestradiol [E2] (pmol/L)

    Page 27

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    Randox Evolution

    non-numeric results

    Beckman Access/Dxi

    Siemens Immulite 2000/2500

    All methods [ALTM]

    170

    n

    533

    Roche Cobas/Modular

    48

    240

    OCD (J&J) VITROS3

    14

    106Siemens ADVIA Centaur

    19

    Tandem Mass Spectrometry

    4

    Abbott Architect

    24.7

    12.9

    179

    SDMean

    138

    149

    41.0

    18 10.3

    129

    170

    9.8

    160

    3712

    15

    163 16

    21

    453A

    7.8

    CV(%)

    9.470

    151

    162

    14

    11

    53.0

    SD

    38.5

    Mean

    127.3

    453B

    21

    CV(%)

    27.0

    28

    36 46

    45

    51

    39

    8.410

    14.9

    143

    SDMean

    10426.1

    20 13.9

    112

    151

    12.8

    135

    114

    14

    132 17

    20

    453C

    CV(%)

    10.239

    134

    10.08

    20.8

    104

    SDMean

    6639.3

    16 15.7

    86

    108

    18.7

    113

    77

    14

    101 19

    21

    453D

    CV(%)

    12.542

    101

    13.3

    65

    34.4

    19

    SDMean

    37

    40

    44.7

    27.768

    6788

    5

    453E

    16

    19 28.6

    23

    CV(%)

    18.530

    68

    Method development [MDV2]

    BO5

    46

    Siemens Immulite 2000/2500

    Randox Evolution

    All methods

    108

    n

    RX4

    MS2

    Siemens ADVIA Centaur

    43

    Tandem Mass SpectrometryDC11

    34

    AB13

    5

    242

    SF1MDV2

    Roche Cobas/Modular

    OCD (J&J) VITROS AM12

    Beckman Access/Dxi

    CO10

    3

    18

    14

    Abbott Architect

    62

    87

    142

    Interquartile rangeMedian

    103

    49

    161

    96

    135 143

    141

    123126

    124

    51

    3667

    46

    89

    112

    A score

    103

    46

    112108

    60103

    13783

    75

    100

    60

    +1.0

    +6.6

    +17.2

    Interquartile rangeMedian

    +20.5

    -3.6

    +24.7

    -13.6

    +6.5 +18.1

    +0.1

    +21.9+12.4

    -11.8

    -1.8

    -8.6+10.7

    -3.3

    -15.8

    -15.9

    B score

    -21.0

    -7.7

    -19.1+20.4

    +2.1+8.7

    -14.5+15.7

    -18.5

    +13.7

    -1.8

    15.5

    18.9

    38.1

    Interquartile rangeMedian

    37.8

    12.4

    36.9

    18.4

    37.3 42.2

    30.9

    38.130.8

    19.6

    12.4

    9.215.3

    10.7

    17.7

    18.8

    C score

    16.7

    11.4

    14.620.8

    14.423.9

    16.417.4

    16.5

    34.3

    14.6

    of 60

  • Date : 05-Jun-2018

    Oestradiol [E2] (pmol/L)

    Page 28

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    Specim

    en A

    ccura

    cy Index

    Distribution

    Specimen Accuracy Indexfor each specimen in time-window

    P-ABCATS (ABC1)

    448 449 450 451 452 453

    0

    50

    100

    150

    200

    250

    300

    A s

    core

    Distribution

    A score by distributionYour current A score is 256

    P-ABCAT (ABC1)

    442 443 444 445 446 447 448 449 450 451 452 453

    0

    50

    100

    150

    200

    250

    300

    All

    Meth

    od

    Specim

    en %

    bia

    s

    Distribution

    Specimen %biasfor each specimen in time-window

    P-ABCBRS (ABC1)

    448 449 450 451 452 453

    -40

    -30

    -20

    -10

    0

    10

    20

    30

    40

    B s

    core

    Distribution

    B score by distributionYour current B score is +51.2

    P-ABCBR (ABC1)

    442 443 444 445 446 447 448 449 450 451 452 453

    -40

    -30

    -20

    -10

    0

    10

    20

    30

    40

    All

    Meth

    od

    Diffe

    rence (

    pm

    ol/L)

    Concentration (pmol/L)

    Difference (result - target)for each specimen in time-window

    P-BA1FS1A (ABC1)

    0 200 400 600 800 1000 1200 1400

    -100-80-60-40-20

    020406080

    100

    C s

    core

    Distribution

    C score by distributionYour current C score is 48.5

    P-ABCCR (ABC1)

    442 443 444 445 446 447 448 449 450 451 452 453

    0

    10

    20

    30

    40

    All

    Meth

    od

    Specim

    en %

    bia

    s

    Concentration (pmol/L)

    Specimen %biasfor each specimen in time-window

    P-BA1FS1P (ABC1)

    0 200 400 600 800 1000 1200 1400

    -40

    -30

    -20

    -10

    0

    10

    20

    30

    40

    B s

    core

    C score

    Oestradiol [E2] B score is +51.2 and C score is 48.5Your method is Beckman Access/Dxi

    P-ABCPENR (ABC1)

    0 10 20 30 40

    -40

    -30

    -20

    -10

    0

    10

    20

    30

    40

    Your labYour methodOverall (n=242)Your method (n=18)

    Specim

    en A

    ccura

    cy Index

    Concentration (pmol/L)

    Specimen Accuracy Indexfor each specimen in time-window

    P-YSTMOD (ABC1)

    0 200 400 600 800 1000 1200 1400

    0

    50

    100

    150

    200

    250

    300 ALTMMLTM

    CV

    %

    Concentration (pmol/L)

    Oestradiol [E2] Between-laboratory agreementBeckman Access/Dxi

    P-PROFILE (ABC1)

    0 200 400 600 800 1000 1200 1400

    0

    10

    20

    30

    40

    50

    of 60

  • Date : 05-Jun-2018

    Analyte : Oestradiol [high-level] (pmol/L)

    Page 29

    for Steroid Hormones

    Distribution : 453

    Laboratory :

    Birmingham Quality is part of the University Hospitals Birmingham NHS Foundation Trust and provides this UK NEQAS service from PO Box 3909, Birmingham B15 2UE, UK To contact us, email [email protected] or phone us on +44 (0)121 414 7300

    © Data in UK NEQAS / Birmingham Quality reports is confidential. For this Scheme, the Organiser is Finlay MacKenzie Birmingham Quality is a UKAS accredited proficiency testing provider No. 7860. www.birminghamquality.org.uk

    Published at 8:57 on Thursday 14 June 2018

    HE219453B

    PoolSpec.

    453D

    Pool HE199 [F] + 8352 pmol/L Oestradiol

    Pool description / Treatments / Additions

    Pooled pregnancy serumNormal serum [F]Normal serum [F]HE220

    HE210453A

    453C

    HE198 14.7

    The B limit is +/-

    Your B score is

    The A limit is15.0

    216

    15.0

    +19.3Your A score is

    The C limit is

    200

    Your C score is

    All methods Beckman Access/Dxi

    Spe