39_Intermed_Catalogo_iX5-Ing.pdf

3
 Neonatal, Pediatric and Adult Lung Ventilator 

Transcript of 39_Intermed_Catalogo_iX5-Ing.pdf

  • Neonatal,Pediatric and Adult

    Lung Ventilator

  • REF

    . 806

    .004

    81 R

    000

    VENTILATION MODES

    VCV

    (Volume-Controlled)

    PCV

    (Pressure-Controlled)

    TCPLV

    (Continuous Flow, Time cycled

    and Pressure Limited)

    NIV

    (Non-Invasive Ventilation)

    A/C

    SIMV / CPAP

    A/C

    SIMV / CPAP

    A/C

    SIMV / CPAP

    A/C

    SIMV/CPAP

    POWER SUPPLY

    Full Range Power Supply

    (automatic switching)

    Internal Battery (Lead-acid)

    External Power Supply Source

    Voltage: 100V 240V

    Current: 0,3A 0,7A

    Frequency: 50Hz 60Hz

    Up to 180 minutes

    12V 14V

    PNEUMATIC SUPPLY

    Intake gases (Air and Oxygen)

    Pressure Range:

    Maximum demand flow:

    Connection Standard

    NBR 12188

    250kPa 600kPa

    50L/min at 350kPa

    PHYSICAL FEATURES

    Height: 300 mm

    Width: 290 mm

    Weight: 12kg

    Depth: 400 mm

    Pedestal Height: 1100 mm

    CLASS I EQUIPMENT / INTERNALLY ENERGIZED

    EQUIPMENT B-TYPE APPLIED PART / IP24 /

    CONTINUOUS OPERATION

    EQUIPMENT IS SUBJECT TO CHANGE WITHOUT PRIOR

    NOTIFICATION

    PICTURES ARE FOR ILLUSTRATIVE PURPOSE ONLY

    ALWAYS USE INTERMED ORIGINAL ACESSORIES

    ALWAYS CONSULT THE OPERATION MANUAL

    ACCESSORIES OF VENTILATION SYSTEM

    Pedestal with casters

    Articulated Arm

    Pediatric / Adult Patient Circuit

    Neonatal Patient Circuit

    Pediatric / Adult Patient Circuit

    without drain

    Neonatal Patient Circuit without

    drain

    Adult Flow Sensor

    Pediatric Flow Sensor

    Neonatal Flow Sensor

    Expiratory Flow Sensor

    Exhalation Valve Diaphragm

    Exhalation Valve

    TGI / NIV Connector

    Micro Mist Nebulizer Kit

    O2 Hose (4 meter long)

    Air Hose (4 meter long)

    Code 179.01100

    Code 179.01000

    Code 100.21000

    Code 100.19000

    Code 100.22000

    Code 100.20000

    Code 136.00310

    Code 136.00311

    Code 136.00347

    Code 179.00734

    Code 130.01362A

    Code 177.00437

    Code 100.00202

    Code 303.00115

    Code 408.00002

    Code 408.00003

    ALARMS

    Audio Pause Key

    PHYSIOLOGICAL ALARMS

    Low Inspiratory Pressure

    High Inspiratory Pressure

    Low Tidal Volume

    High Tidal Volume

    Low Minute Volume

    High Minute Volume

    AUTOMATIC ALARMS

    Low PEEP

    High PEEP

    Patient Circuit Leakage

    Patient Circuit Occlusion

    TECHNICAL ALARMS

    Low O2 Pressure

    High O2 Pressure

    Low Air Pressure

    High Air Pressure

    Gas Fail

    Up to 120 seconds

    Low FIO2High FIO2Low Respiratory Rate

    High Respiratory Rate

    Apnea (with Backup cycle)

    High CPAP

    Auto PEEP

    Autocycling

    Internal Battery in Use

    Low Internal Battery

    Internal Battery Fail

    Power Fail

    Inoperative Ventilator

    GRAPHICS

    Up to 5 simultaneously curves

    Loop Pressure x Volume

    Loop Flow x Volume

    TRENDS (UP TO 72 HOURS)

    Peak Pressure

    PEEP

    Tidal Volume

    FIO2

    VIGILANCE SCREEN (COGNITIVE)

    Curve Pressure x Time

    Curve Flow x Time

    Curve Volume x Time

    Respiratory Rate

    Dynamic Compliance

    Inspiratory Resistance

    MONITORS

    MONITORED PARAMETERS

    Peak Pressure

    PEEP

    Mean Pressure

    Plateau Pressure

    Total Respiratory Rate

    Spontaneous Respiratory Rate

    Inhaled Oxygen Fraction (FIO2)

    RESPIRATORY MECHANIC

    Dynamic Complacence

    Static Complacence

    Occlusion Pressure (P0.1)

    LOG SCREEN

    (UP TO 72 HOURS)

    Alarms

    Events

    Tidal Volume

    Minute Volume

    Inspiratory Peak Flow

    Expiratory Peak Flow

    Inspiratory Time

    Expiratory Time

    I:E Ratio

    Inspiratory Resistance

    Expiratory Resistance

    Auto PEEP

    OFF; 0,3 a 10,0 cmH2O

    OFF; 0,5 a 10,0 cmH2O

    OFF; 0,5 a 10,0 cmH2O

    OFF; 0,2 a 2,0 L/min

    OFF; 0,5 a 5,0 L/min

    OFF; 2,0 a 15,0 L/min

    CONTROLS

    BASIC PARAMETERS

    O2 Concentration

    Controlled Volume

    Inspiratory Flow (Controlled)

    Inspiratory Flow (Demand)

    Types of Wave forms

    Controlled Pressure (in PCV)

    Inspiratory Pressure (in NIV)

    Support Pressure

    Pressure Slope

    End Flow (Expiratory Sensitivity)

    Limit Pressure

    Inspiratory Time

    Inspiratory Pause

    Respiratory Rate

    PEEP / CPAP

    Inspiratory Sensitivity by Pressure

    Inspiratory Sensitivity by Flow

    COMPLEMENTARY RESOURCES

    Standby Mode

    100% O2Manual Cycle

    Inspiratory Pause (up to 15 s)

    21 to 100%

    Pediatric:

    Adult:

    Neonatal:

    Pediatric:

    Adult:

    Up to 180 L/min

    Square/ Decreasing / Sinusoidal

    Up to 80 cmH2O (above PEEP)

    Up to 40 cmH2O

    Up to 80 cmH2O (above PEEP)

    25 to 100%

    10 to 75%

    Up to 80 cmH2O

    0,10 to 15,0 s

    0,0 to 2,0 s

    0 to 180 min-1

    0 to 50 cmH2O

    Neonatal:

    Pediatric:

    Adult:

    Neonatal:

    Pediatric:

    Adult:

    Expiratory Pause (up to 15 s)

    Sigh

    Nebulization

    Tracheal Gas Insufflation (TGI)

    TECHNICAL FEATURES

    10 to 500 mL

    100 to 2000 mL

    2 to 30 L/min

    2 to 50 L/min

    2 to 120 L/min

    Tel.: +55 (11) 4615.9300

    iX5 ext ing1iX5 int ing2iX5 ext ing3