322-332
-
Upload
amalia-eka-rakhmania -
Category
Documents
-
view
217 -
download
0
Transcript of 322-332
-
8/13/2019 322-332
1/10
ERRORS DUE TO ENVIRONMENTAL INPUTS
environmental input : apparently real input to ameasurement system caused by a change in theenvironmental conditions surrounding themeasurement system.
The static and dynamic characteristics specifiedfor measuring instruments are only valid forparticular environmental conditions, must bereproduced as closely as possible during
calibration exercises The magnitude is quantified by the two
constants: sensitivity drift and zero drift
-
8/13/2019 322-332
2/10
EXAMPLE
Suppose we are given a small closed box and told thatit may contain either a mouse or a rat. We are also toldthat the box weighs 0.1kg when empty. If we put thebox onto bathroom scales and observe a reading of 1.0
kg, this does not immediately tell us what is in the boxbecause the reading may be due to one of three things:
(a) a 0.9 kg rat in the box (real input)
(b) an empty box with a 0.9 kg bias on the scalesdue to a temperature change (environmentalinput)
(c) a 0.4 kg mouse in the box together with a 0.5 kgbias (real + environmental inputs).
-
8/13/2019 322-332
3/10
The magnitude of any environmental input
must be measured before the value of the
measured quantity (the real input) can be
determined from the output reading of an
instrument.
Difficult to avoid environmental inputs
-
8/13/2019 322-332
4/10
WEAR IN INSTRUMENT COMPONENTS
Systematic errors can frequently develop over
a period of time because of wear in
instrument components.
Solution: Recalibration
-
8/13/2019 322-332
5/10
CONNECTING LEADS
Common source of error : failure to take proper account ofthe resistance of connecting leads.
In resistance thermometer, thermometer is separated fromother parts of the measurement system by 100 metres. The
resistance of such a length of 20 gauge copper wire is 7,and there is a further complication that such wire has atemperature coefficient of 1m/C.
Connecting leads should: Adequate cross-section so that their resistance is minimized
Adequately screened if they are to be subject to electrical ormagnetic fields that could otherwise cause induced noise.
The routing of cables needs careful planning: changing theroute of the cables between the transducers and thecontrol room, the magnitude of noise can be reduced.
-
8/13/2019 322-332
6/10
REDUCTION OF SYSTEMATICERRORS
-
8/13/2019 322-332
7/10
The prerequisite for the reduction of
systematic errors is a complete analysis of the
measurement system that identifies all
sources of error.
Simple faults within a system can usually be
readily and cheaply rectified once they have
been identified.
-
8/13/2019 322-332
8/10
CAREFUL INSTRUMENT DESIGN
Careful instrument design : reducing the
sensitivity of an instrument to environmental
inputs to as low a level as possible.
Errors due to the way in which an instrument
is designed are not always easy to correct,
Solution: high cost of redesign or accept the
reduced measurement accuracy
-
8/13/2019 322-332
9/10
METHOD OF OPPOSING INPUTS
Compensates the effect of an environmental
input by introducing an equal and opposite
environmental input that cancels it out.
Example: Milivoltmeter
-
8/13/2019 322-332
10/10