322-332

download 322-332

of 10

Transcript of 322-332

  • 8/13/2019 322-332

    1/10

    ERRORS DUE TO ENVIRONMENTAL INPUTS

    environmental input : apparently real input to ameasurement system caused by a change in theenvironmental conditions surrounding themeasurement system.

    The static and dynamic characteristics specifiedfor measuring instruments are only valid forparticular environmental conditions, must bereproduced as closely as possible during

    calibration exercises The magnitude is quantified by the two

    constants: sensitivity drift and zero drift

  • 8/13/2019 322-332

    2/10

    EXAMPLE

    Suppose we are given a small closed box and told thatit may contain either a mouse or a rat. We are also toldthat the box weighs 0.1kg when empty. If we put thebox onto bathroom scales and observe a reading of 1.0

    kg, this does not immediately tell us what is in the boxbecause the reading may be due to one of three things:

    (a) a 0.9 kg rat in the box (real input)

    (b) an empty box with a 0.9 kg bias on the scalesdue to a temperature change (environmentalinput)

    (c) a 0.4 kg mouse in the box together with a 0.5 kgbias (real + environmental inputs).

  • 8/13/2019 322-332

    3/10

    The magnitude of any environmental input

    must be measured before the value of the

    measured quantity (the real input) can be

    determined from the output reading of an

    instrument.

    Difficult to avoid environmental inputs

  • 8/13/2019 322-332

    4/10

    WEAR IN INSTRUMENT COMPONENTS

    Systematic errors can frequently develop over

    a period of time because of wear in

    instrument components.

    Solution: Recalibration

  • 8/13/2019 322-332

    5/10

    CONNECTING LEADS

    Common source of error : failure to take proper account ofthe resistance of connecting leads.

    In resistance thermometer, thermometer is separated fromother parts of the measurement system by 100 metres. The

    resistance of such a length of 20 gauge copper wire is 7,and there is a further complication that such wire has atemperature coefficient of 1m/C.

    Connecting leads should: Adequate cross-section so that their resistance is minimized

    Adequately screened if they are to be subject to electrical ormagnetic fields that could otherwise cause induced noise.

    The routing of cables needs careful planning: changing theroute of the cables between the transducers and thecontrol room, the magnitude of noise can be reduced.

  • 8/13/2019 322-332

    6/10

    REDUCTION OF SYSTEMATICERRORS

  • 8/13/2019 322-332

    7/10

    The prerequisite for the reduction of

    systematic errors is a complete analysis of the

    measurement system that identifies all

    sources of error.

    Simple faults within a system can usually be

    readily and cheaply rectified once they have

    been identified.

  • 8/13/2019 322-332

    8/10

    CAREFUL INSTRUMENT DESIGN

    Careful instrument design : reducing the

    sensitivity of an instrument to environmental

    inputs to as low a level as possible.

    Errors due to the way in which an instrument

    is designed are not always easy to correct,

    Solution: high cost of redesign or accept the

    reduced measurement accuracy

  • 8/13/2019 322-332

    9/10

    METHOD OF OPPOSING INPUTS

    Compensates the effect of an environmental

    input by introducing an equal and opposite

    environmental input that cancels it out.

    Example: Milivoltmeter

  • 8/13/2019 322-332

    10/10