2D Color Analyzer CA-2500 Series
Instruction ManualBefore using this instrument,please read this manual.
Safety SymbolsThe following symbols are used in this manual to prevent accidents which may occur as a result of incorrect use of the instrument.
Denotes a sentence regarding safety warning or precaution.Read the sentence carefully to ensure safe and correct use.
Denotes a prohibited operation.The operation must never be performed.
Denotes an instruction.The instruction must be strictly adhered to.
Denotes an instruction.Disconnect AC adapter from AC outlet.
Denotes a prohibited operation.Never disassemble this instrument.
This symbol indicates A.C.
This symbol indicates D.C.
Trademarks● “Microsoft®”, “Windows®”, “Windows®XP”, “Windows® 7”, “Windows® 8”and “Excel®” are registered trademarksofMicrosoftCorporationintheU.S.A.andothercountries.
● “Intel®”and“Pentium®”areregisteredtrademarksofIntelCorporationintheU.S.A.andothercountries. Other company names and product names mentioned in this manual are the registered trademarks or
trademarks of their respective companies.
Notes on this manual● Nopartof thismanualmaybe reprintedor reproduced inany formorbyanymeanswithout thepermissionofKonicaMinoltaOptics,Inc.
● Thecontentsofthismanualaresubjecttochangewithoutnotice.● Everyefforthasbeenmadetoensuretheaccuracyofthecontentsofthismanual.However,shouldyouhaveanyquestionsorcomments,or findanerrorormissingsection,pleasecontactaKonicaMinolta-authorized service facility.
● KonicaMinoltaacceptsnoresponsibilityforconsequencesresultingfromfailuretofollowtheinstructionsoutlined in this manual, the condition above notwithstanding.
1
Safety Warnings and CautionsToensurecorrectuseof this instrument,readthefollowingpointscarefullyandadheretothem.Afteryouhave read this manual, keep it in a safe place where it can be referred to anytime a question arises.
Warning(Failuretoadheretothefollowingpointsmayresultindeathorseriousinjury.)
Donotusethisinstrumentinplaceswhereflammableorcombustiblegases(gasolineetc.)arepresent. Doing so may cause fire.
AlwaysusetheACadapterandpowercordsuppliedasastandardaccessoryoroptional(AC-A312),andconnectittoindoorACoutletofratedvoltage100-240V 50/60Hz.FailuretofolloweitherofthesemayresultindamagetoinstrumentorACadapter,fireorelectricshock.
Insertthepowerplugfullyandsecurely.Ifthepowerplugisnotpushedcompleteinsertingmaycausefireorelectricshock.
Ifthisinstrumentisnotusedforalongtime,disconnectACadapterfromACoutlet.AccumulateddirtorwateronprongsofACadapterplugmaycausefire.Besuretoremovebeforeuse.
Do not forcibly pull any part on power cord when unplugging since this may damage power cord, resultinginfireorelectricshock.Gentlydisconnectbyholdingplug.Also,donothandlepowerplug with wet hands. Doing so may cause electric shock.Dust could enter inside this instrument from air duct. Do not leave such dust accumulated while usingforthismaycausefire.ContactwiththenearestKonicaMinolta-authorizedservicefacilityforperiodical checkup.Donotforciblybend,twistorpullpowercord.Also,donotplaceheavyobjectonpowercord,ordamageormodifyone.Anyofthesemaycausefireorelectricshockduetodamagetopowercord.
DonotdisassembleormodifythisinstrumentorACadapter.Doingsomaycausefireorelectricshock.
Donotexposethis instrumentto liquidormetalobjectwhichmaycausefireorelectricshock.Shouldeitherofthesehappen,switchpoweroffimmediately,unplugACadapter,andcontactthenearestKonicaMinolta-authorizedservicefacility.ShouldthisinstrumentorACadapterbedamagedorsmokeoroddsmellbegenerated,donotkeepusingonewithoutcorrection.Doingsomaycause fire. Insuchcases,switchpoweroffimmediately,unplugACadapter fromACoutlet,andthencontact thenearestKonicaMinolta-authorized service facility.
Caution
UsethisinstrumentnearACoutletforeasypluggingorunplugginginusingACadapter.
Do not place this instrument on unstable or sloping surface which may drop or overturn it. Droppingoroverturningmayinjuresomeonearound.Takecarenottodropthisinstrumentwhencarrying.
Whencleaning,disconnectthepowerplug.Connectingmaycauseelectricshock.
(Failuretoadheretofollowingpointsmayresultininjuryordamagetothisinstrumentorotherproperty.)
2
IntroductionThankyouforpurchasingour2DColorAnalyzerCA-2500series.Thisinstrumentmeasuresunevennessofluminanceandcoloronvariousdisplayunitorprojector,backlightandcarinstrumentpanelin2dimension.Read this manual carefully before use."CA-2500"andCA-2500Sareappliedas representativeexamples for explanationand illustrationrespectively in this instruction manual. Only those with different specification are indicated model by model.
Packaging materialRetainaccompanyingpackagingmaterials(carton,protector,andplasticbag)forfutureuse.This isdelicatemeasurement instrument.Usepackagingmaterialssupplied inpurchasing incasethisinstrumentneeds tobe transferred forsuchpurposeasmaintenance inKonicaMinolta-authorizedservice facility. These packaging materials are useful for minimizing shock or vibration to this instrument insuchsituation.Shouldanyofthesepackagingmaterialsbelostorbroken,contactthenearestKonicaMinolta-authorizedservicefacility.
Note on useOperating environmentlDonotusethisinstrumentoutdoorsincestandardaccessoryACadapter(AC-A312)isdesignedforindooruse.lNeverdisassemblethisinstrumentforbeingcomposedofdelicateelectroniccomponents.lUsethisinstrumentatratedvoltageof100-240V .ConnectACpowercordtoACoutletwithratedvoltage100-240V 50/60Hz.Connectedvoltageshouldnotbeoutsidetherangeof±10%ofnominal.
l ThisinstrumentisclassifiedintoaPollutionDegree2asinstrumentusedmainlyinmanufacturingplant,laboratory,warehouseorequivalents.Usethisinstrumentinmetaldustfreeandnoncondensingpotentialenvironment.
lThis instrument iscategorized intoOvervoltageCategory Iasequipmentwhich ispoweredbyanACadaptor connected to commercially available power.
lIfprotectiveearthterminalisequippedwiththePCtocontrolthisinstrument,connectthePCterminaltoaprotectiveearthoutlet.Ifnot,electricshockmayoccurduetoearthleakage.
l Take care not to enter foreign substance like water or metal in this instrument. Operating in such state causes serious danger.
l Do not use this instrument under direct sunlight or near heater. The internal temperature of this instrument becomesmuchhigherthanambienttemperaturewhichmaybreakthisinstrument.Also,besuretokeepgoodventilationinusingthisinstrument.Nevercoverairductofthisinstrument.
lAvoidrapidchangeinambienttemperaturewhichmayformdewcondensation.lAvoidusingthisinstrumentinextremelydustyorhumidplace.lUsethis instrumentatambient temperaturebetween10ºCand30ºCandrelativehumidity70%or less
with no condensation. Operating this instrument outside specified temperature and humidity range may unsatisfy its original performance.
lDonotusethisinstrumentataltitudesabove2000meters(6560ft).lEven if theproduct isusedwithin thespecifiedoperating temperature/humidity range, thedisplayedvaluesmaybeaffectedby long-termconditionsofuse. Ifyouhaveanyquestionaboutspecification,pleasecontactthenearestKonicaMinolta-authorizedservicefacility.
This instrumentlDonotsubjectthisinstrumenttostrongimpactorvibration.lDonotforciblypull,bend,orapplystrongforcetopowercordforattachedACadapterorUSBcable.This
may result in snapping.lConnectthisunittopowersourcewithminimalnoise.lShouldbreakageorabnormalitybe foundduringoperation,switchpoweroff immediatelyandunplug.Then,referto"ErrorCheck"onpages29.
lShouldthisinstrumentbreakdown,donottrytodisassembleandrepairitbyyourself.ContactthenearestKonicaMinolta-authorizedservicefacility.
3
Lens and filter inside lens mount on bodyl Be sure that there is no dirt on surface of lens and filter inside lens mount on body before measurement. Correctmeasurementresultcannotbeobtainedifdirt,dustorhandsoilisleftoruncleaned.
l Do not touch surface of lens and filter inside lens mount on body with hand.l Lens and filter inside lens mount on body could become fogged up if ambient temperature fluctuates under
high humidity. Be careful not to generate such condition for correct measurement cannot be performed.
StorageBodyl Coverlensmountwithstandardaccessorymountcoverinstoringlensinremovedstate.l Do not store this instrument under direct sunlight or near heater. The internal temperature of this
instrument could become much higher than ambient temperature which may break this instrument.l Storethis instrumentatambient temperaturebetween0ºCand30ºCandrelativehumidity70%or lesswithnocondensationorbetween30ºCand35ºCandrelativehumidity55%orlesswithnocondensation.Storageunderhightemperatureandhumiditymaydeteriorateperformanceofthisinstrument.Foraddedsafety, we recommend storage with a drying agent at room temperature.
l Eveniftheproductisstoredwithinthespecifiedstoragetemperature/humidityrange,thedisplayedvaluesmaybeaffectedbylong-termconditionsofstorage.Ifyouhaveanyquestionaboutspecification,pleasecontactthenearestKonicaMinolta-authorizedservicefacility.
l Takecarenottoformcondensation.Avoidrapidchangeinambienttemperaturewhentransferringbodyfor storage.
l Putbodyinpackagingboxsuppliedwhenpurchasedoroptionalsoftcase(CA-A60)tostoreinsafeplace.
Lensl Coverlenswithlenscap,orlenshoodwithhoodcapforstorage.lWhenyouremovelensfrombodyforstorage,coverlenswithlenscapandmountcap.
Temperature/Humidity conditions under which the instrument is used and storedIt is recommended that the instrumentbeusedandstoredunderstandardconditions (Temperature:23ºC;Relativehumidity:40%),andthatareassubjecttohightemperatureand/orhumiditybeavoided.Inadditioninordertomaintainthemeasurement accuracy of this instrument, it is recommenced that it be inspected regularly about once a year. For details onhavingtheinstrumentinspected,pleasecontactthenearestKonicaMinolta-authorizedservicefacility.
Even if theproduct isusedwithin thespecifiedoperating temperature/humidity rangeorstoredwithin thespecifiedstoragetemperature/humidityrange,thedisplayedvaluesmaybeaffectedbylong-termconditionsofuseorstorage.Iftheinstrumentisleftunderthefollowinghigh-temperatureconditionsforalongperiodoftime,thedisplayedvaluesmaychangeasfollows:
Temperature:30ºC;Relativehumidity:70%;Periodundertheseconditions:720hours(30days)Accuracy:Luminance:±0.4%;Chromaticity:±0.003Inter-pointerror:Luminance:±0.2%;Chromaticity:±0.0003
Temperature:35ºC;Relativehumidity:55%;Periodundertheseconditions:336hours(14days)Accuracy:Luminance:±1%;Chromaticity:±0.006Inter-pointerror:Luminance:±0.5%;Chromaticity:±0.001
Thesedifferences indisplayvaluesaredue to the instrumentmaterialsand/orcomponentsbeingaffectedby thetemperatureandhumidityconditionsof long-termuseorstorage. Inparticular,optical filtersareeasilyaffectedbytemperatureorhumidity.Althoughmeasureshavebeentakentoimproveresistancetotemperature/humiditychanges,theaccumulatedeffectoflong-termuseorstoragemayaffectthedisplayedvalues.
4
CleaningBodyl Ifthisunitbecomesdirty,wipewithdryandsoftcloth.Donotuseorganicsolventlikebenzineorthinnerandotherchemicalagent forcleaning.Shouldnoneof thesemethodsbehelpful,contact thenearestKonicaMinolta-authorizedservicefacility.
Lens and filter inside lens mount on bodyl Shoulditbegottendirtordust,wipeoffwithdryandsoftclothorlenscleaningpaper.Donotuseorganicsolventlikebenzineorthinnerandotherchemicalagentforcleaning.Shouldnoneofthesemethodsbehelpful,contactthenearestKonicaMinolta-authorizedservicefacility.
Notes on transferl Usetheoriginalpackagingmaterialstominimizetheshockandvibrationtotheproductwhentransportingit.l Pack the product and all of its accessories together in one package before sending them to us for repair, inspectionand/orcalibration.
Maintenancel Periodical checkup is recommended annually to maintain measurement accuracy of instrument. For detailsoncheckup,contactthenearestKonicaMinolta-authorizedservicefacility.
Disposal Methodl MakesurethattheCA-2500,itsaccessoriesandthepackingmaterialsareeitherdisposedoforrecycled
correctly in accordance with local laws and regulations.
5
CONTENTSSafety Warnings and Cautions …………………1
Introduction …………………………………………2Note on use ………………………………………… 2
Operating environment …………………………………… 2This instrument …………………………………………… 2Lens and filter inside lens mount on body ……………… 3
Storage ……………………………………………… 3Body ………………………………………………………… 3Lens ………………………………………………………… 3
Cleaning ……………………………………………… 4Body ………………………………………………………… 4Lens and filter inside lens mount on body ……………… 4
Notes on transfer …………………………………… 4Maintenance ………………………………………… 4Disposal Method …………………………………… 4
Standard Accessory ………………………………6
Optional Accessory ………………………………7
System Configuration ……………………………8
Names and Functions of Parts ……………… 10<Body> ……………………………………………………… 10<Lens> ……………………………………………………… 11
Installing
Goods/Environment to Prepare on Customer's Side … 14Controlling PC ………………………………………14Mounting stage ………………………………………14Blackout curtain (quickly prepared darkroom) / Dark room ……………………………………………14
Installing ………………………………………… 15
Placing Lens …………………………………… 16Replacing lens ………………………………………17
Connecting AC Adapter ……………………… 18Connection procedure ……………………………18
ON ( | ) / OFF ( ) of Power Switch …………… 19ToON ……………………………………………………… 19To OFF ……………………………………………………… 19
Connecting with PC …………………………… 19Operation procedure ………………………………19
Measurement
Starting Up and Ending System …………… 22Starting up ……………………………………………22Connectionrecognitionerror …………………………… 22
Ending …………………………………………………22
Setting the measurement distance and the pint ring range index ……………………………… 23
Setting measurement distance …………………23Caution ……………………………………………………… 24
Example of reference position to read distance …25Howtoreaddistanceincaseofstandardlensortelephotolens … 25Howtoreaddistanceincaseofwidelens ……………… 25
Others
Outer Dimensions ……………………………… 28
Error Check ……………………………………… 29
Main Specification ……………………………… 30
6
Standard Accessory
Lens Capl Place on lens for protection in not using this
instrument.For standard or telephoto lensFor wide lens
Mount Capl Place on lens mount for protection when lens is
removed from body.For standard or telephoto lensFor wide lens
Lens Hood For standard or telephoto lens CA-A68
For wide lens CA-A69 For macro measurement CA-A70
Hood Capl Place on lens for protection in not using this
instrument. For standard lens, telephoto lens, or macro measurement CA-A71For wide lens CA-A75
Mount Cover CA-A61l Place for protection of lens mount of body when lens
is removed.
AC Adapter AC-A312l SuppliespowerfromACoutlet. Input: 100-240V 0.75A 50/60Hz Output:12V 3.0A
CA-A71 CA-A75
For wide lens
For standard ortelephoto lens
CA-A68 CA-A69 CA-A70
Attachedlenscap,mountcap,lenshoodandhoodcapcorrespondingwithlenstypeforeachmodel.
For standard or telephoto lens
For wide lens
7
USB Cable (3 m) IF-A18 l Usedforcommunicationbetweenthis instrument
andPC.
Calibration Data DVDl Suppliedwitheachlens.
Tripod CS-A3
Data Management Software CA-S25wl FordatamanagementofthisinstrumentfromPC.
Mura Evaluation Software CA-Mural Thisprogramisanadd-onsoftwareforCA-S25w. Itusesthe"muraquantitativeevaluationmethod"
developedbySonytoprovidethe"mura indexes"and "mura index images" ref lect ing humansensitivity.
Soft Case CA-A60l Usedtokeep this instrumentandaccessoriesor
carrythemwithhand.Neverusefortransfer.
Optional Accessory
Pan Head CS-A4
8
System ConfigurationCA-2500A (With all lenses)
Wide lens CA-A58 Hood cap(For wide)CA-A75
Lens hood(For wide)CA-A69
Standard lens CA-A57 Hood cap(For standard/telephoto, macro)CA-A71
Lens hood(For standard/telephoto)CA-A68
Telephoto lens CA-A59
Hood cap(For standard/telephoto, macro)CA-A71
Lens hood(For standard/telephoto)CA-A68
Lens hood(For macro)CA-A70
Mount cover
Soft caseCA-A60
TripodCS-A3
Pan headCS-A4
AC adapterAC-A312
USB cableIF-A18
CA-2500
Data management softwareCA-S25w
Mura Evaluation softwareCA-MuraPC
(on the market)
Highmagnificationmacro ringMacro 2
Lowmagnificationmacro ringMacro 1
CA-2500S (With standard lens)
Standard lens CA-A57 Hood cap(For standard/telephoto, macro)CA-A71
Lens hood(For standard/telephoto)CA-A68
Mount cover
Soft caseCA-A60
TripodCS-A3
Pan headCS-A4
AC adapterAC-A312
USB cableIF-A18
CA-2500
Data management softwareCA-S25w PC
(on the market)
Mura Evaluation softwareCA-Mura
CA-2500W (With wide lens)
Wide lens CA-A58 Hood cap (For wide)CA-A75 Lens hood
(For wide)CA-A69
Mount cover
Soft caseCA-A60
TripodCS-A3
Pan headCS-A4
AC adapterAC-A312
USB cableIF-A18
CA-2500
Data management softwareCA-S25w PC
(on the market)
Mura Evaluation softwareCA-Mura
CA-2500T (With telephoto lens)
Telephoto lens CA-A59
Hood cap(For standard/telephoto, macro)CA-A71
Lens hood(For standard/telephoto)CA-A68
Lens hood(For macro)CA-A70High
magnificationmacro ringMacro 2
Low magnificationmacro ringMacro 1
Mount cover
Soft caseCA-A60
TripodCS-A3
Pan headCS-A4
AC adapterAC-A312
USB cableIF-A18
CA-2500
Data management softwareCA-S25w PC
(on the market)
Mura Evaluation softwareCA-Mura
9
CA-2500SW (With standard and wide lenses)Wide lens CA-A58 Hood cap (For wide)
CA-A75Lens hood(For wide)CA-A69
Standard lens CA-A57 Hood cap(For standard/telephoto, macro)CA-A71
Lens hood(For standard/telephoto)CA-A68
Mount cover
Soft caseCA-A60
TripodCS-A3
Pan headCS-A4
AC adapterAC-A312
USB cableIF-A18
CA-2500
Data management softwareCA-S25w PC
(on the market)
Mura Evaluation softwareCA-Mura
CA-2500ST (With standard and telephoto lenses)
Wide lens CA-A58 Hood cap (For wide)CA-A75 Lens hood
(For wide)CA-A69
Telephoto lens CA-A59
Hood cap(For standard/telephoto, macro)CA-A71
Lens hood(For standard/telephoto)CA-A68
Lens hood(For macro)CA-A70
Mount cover
Soft caseCA-A60
TripodCS-A3
Pan headCS-A4
AC adapterAC-A312
High magnificationmacro ringMacro 2
Low magnificationmacro ringMacro 1
USB cableIF-A18
CA-2500
Data management softwareCA-S25w PC
(on the market)
Mura Evaluation softwareCA-Mura
CA-2500WT (With wide and telephoto lenses)
Standard lens CA-A57 Hood cap(For standard/telephoto, macro)CA-A71
Lens hood(For standard/telephoto)CA-A68
Telephoto lens CA-A59
Hood cap(For standard/telephoto, macro)CA-A71
Lens hood(For standard/telephoto)CA-A68
Lens hood(For macro)CA-A70
Mount cover
Soft caseCA-A60
TripodCS-A3
Pan headCS-A4
AC adapterAC-A312
High magnificationmacro ringMacro 2
Low magnificationmacro ringMacro 1
USB cableIF-A18
CA-2500
Data management softwareCA-S25w PC
(on the market)
Mura Evaluation softwareCA-Mura
StandardaccessoryOptional accessory
10
Names and Functions of Parts<Body>
Power switch Switchesthis instrumenton/off. [|]sideforON; [○]sideforOFF.GreenLEDlightsupwhileswitchedon. ………………………………………… (p.19)
AC adapter input terminalTowhichaccessoryACadapterisconnected. ………………………… (p.18)USB connector TowhichUSBcableisconnectedwhenusedwithPC. ……………… (p.19)Air duct Airduct toair-cool interior.Donotcoverhere.Air-cooling fanbuilt in this
instrument operates while switched on. Lens mount To install lens on.Lens attaching portion To which lens hood is attached.Screw hole for fixing Usedtosetthisinstrumentontripodorjig.
Lens attaching portion
USB connector
Lens mount
AC adapter input terminal
Screw hole for fixing (Tripod screw)
Air duct
Screw hole for fixing (ISO screw)
Handle
Power switch
11
<Lens>
Body and lens arranged in a pair are calibrated. Pair those with identical numbers.CA-2500A, CA-2500S, CA-2500SW, CA-2500ST
CA-2500A, CA-2500W, CA-2500SW, CA-2500WT
CA-2500A, CA-2500T, CA-2500ST, CA-2500WT
Lens hood Protects lens. Remove this once when replacing lens.Focus ring Adjustsfocus.Aperture ring Beingfixed,thisshouldnotbeused.Donotoperatebyremovingorrevolving
screw for this would deteriorate the performance.Macro ring Placed between telephoto lens and body when utilizing telephoto lens for
macro.Youcannotplace2typesofmacroringsimultaneously.
Focus ring
Distance indicator
Lens hood
Lens
(Low magnification) (High magnification)
Macro ringTelephoto lens
Wide lens
Standard lens
Aperture ring
12
13
Installing
14
Installing
Goods/Environment to Prepare on Customer's SideGetpreparedforbelowgoods/environmentbeforeusingthisinstrument.
Controlling PCOS : Windows®XPProfessionalSP3,Windows®XPProfessionalx64EditionSP2 Windows®7Professional32bit(x86),Windows®7Professional64bit(x64) Windows®8Pro32bit(x86),Windows®8Pro64bit(x64) (English,Japanese,orChinese(simplifiedfont)versionoftheaboveOS) ThePCmustmeetthesystemrequirementsspecifiedbytheapplicable
OSorlistedbelow,whicheverishigher.CPU : EquivalenttoPentium®4,2.8GHzorbetterMemory : 1024MBormoreHarddiskdrive : Atleast80MBofavailablediskspaceisrequiredonthesystemdrive
(drivewheretheOSisinstalled). To install the calibration data, the following disk space is required for
eachlens: •Forastandardlens:approximately540MB •Forawideanglelens:approximately470MB •Foratelephotolens:approximately1.3GB
To save measurement data on the hard disk, you need to reserve the followingadditionalspaceontheharddisk:
•Forexample,approximately110MBisrequiredtosave10piecesofmeasurementdatawith980×980resolution.
•About1MBisrequiredforconfigurationfilessuchasthespotsettingfile and measurement condition file.
Display : Displayhardwarecapableofdisplaying1280×768pixels /65,536colors(HighColor:16-bitcolor)orbetter
Other : CD-ROMdriveRequiredforinstallation1DVD-ROMdriveRequiredforinstallation of calibration data
* Inplaceof theabove twodrives,youcanuseasingledrive that iscapableofreadingbothCD-RandDVD-Rmedia.
AUSBportisrequiredforconnectionoftheinstrument.
Inaddition,followingitemsareuseful.
Mounting stageTable on which this instrument is installed. Below are recommended conditions.lCanchangemeasurementdistanceaccordingtomeasurementobject.l Painted with black to prevent the entrance of light diffusely reflecting.
Blackout curtain (quickly prepared darkroom) / Dark room Neededtoblockthisinstrumentandmeasurementobjectfromoutsidelight.Get prepared for blackout curtain for quickly prepared dark room or dark room.
15
Installing
Installing
Tosetthisinstrumentontripodorjig,usescrewholeforfixingatthebottomofthisinstrument.2typeholesareavailableasfollows;Tripodscrewhole:Tosetontripod.Screwdepthis6.5mm.ISOscrewhole:Tosetonjig.UserISO5mmscrewandscrewdepthis6.5mm.
Forfurtherdetail,seep.28.
Warning for safety Donotusethisinstrumentinplaceswhereflammableorcombustiblegases(gasolineetc.)arepresent. Doing so may cause fire.
Caution for safety Do not place this instrument on unstable or sloping surface which may drop or overturn it. Dropping or overturning
mayinjuresomeonearound.Takecarenottodropthisinstrumentwhencarrying.
*Whenstandardlensandhoodareplaced
ISO screw hole
Optical axis
Tripod screw hole
UNC 1/4 Depth 6.5
95.2
186.2
68.2
125.2
16
Installing
Placing LensBelowshows lens typeaccompanying respectiveCA-2500series.Lenshood isservedasstandardcorresponding to lens type.CA-2500A Standard,wideandtelephotolenses,and2macroringsCA-2500S StandardlensCA-2500W WidelensCA-2500T Telephotolensand2macroringsCA-2500SW StandardandwidelensesCA-2500ST Standardandtelephotolenses,and2macroringsCA-2500WT Wideandtelephotolenses,and2macrorings
Operation procedure
1. Remove mount cover from lens mount on body.
2. Place lens on lens mount before rotating clockwise.lBesuretoinsertlensstraightly.Ifforciblyput
aslant, lens mount could be broken. lRevolvelensuntilfixed.
For CA-2500A, CA-2500T, CA-2500ST, or CA-2500WTInusingamacroringforeitherofCA-2500A,CA-2500T,CA-2500ST,orCA-2500WT,sandwich it between lens and lens mount.
3. Align 2 pins of lens hood and 2 holes of body. Then, insert lens hood as if it covers lens. Finally, rotate screw portion of lens hood clockwise.lSelect appropriate lens hooddepending
on lens type. Inappropriate lenshoodmaycause the lens head to be positioned back from the lens hood, which leads to incorrect measurement.
lHoldthecircumferenceoflenshood.lRotatescrewportionoflenshooduntilfixed.
Mount cover
Lens
Macro ring
Lens hood
Lens hood pin hole
17
Installingl Checkwhetherbodyandlenshaveidenticalserialnumbers. Serialnumbercanbefoundonlabelofbodyandlenssiderespectively. Topursuehigh-accuracymeasurement,calibrationwasdoneonmaker's
side for each unit before shipment combining with each accessory lens forCA-2500series.Thismeansthatattachedcalibrationcoefficientdatatoeachunitisvalidexclusivelyforthepackagedcombinationofunitandlens.
l SetlenstypetouseinsoftwareCA-S25w. (Fordetail,seetheinstructionmanualfordatamanagementsoftwareCA-S25w.) This enables the application of appropriate calibration coefficient data depending on usage condition
which would finally give correct measurement result.
Replacing lensFollow reverse procedure to above to replace lens. First, remove lens hood and then lens.
XXXXXX1
WARNING:WARNUNG:Niemals das gehause offnen.
Do not remove covers.警告:カバーを開けないでください。
12 4
18
Installing
Connecting AC Adapter UseaccessoryACadapterforpowersource.
Warning (Failuretoadheretothefollowingpointsmayresultindeathorseriousinjury.) AlwaysusetheACadapterandpowercordsuppliedasastandardaccessoryoroptional(AC-A312),and
connect it to indoorACoutletofratedvoltage100-240V 50/60Hz.Failuretofolloweitherof thesemay
result in damage to unit, fire or electric shock.
Ifthisinstrumentisnotusedforalongtime,disconnectACadapterfromACoutlet.AccumulateddirtorwateronprongsofACadapterplugmaycausefireandshouldberemoved.Besuretoremovebeforeuse.
Do not forcibly pull any part on power cord when unplugging since this may cause fire or electric shock. Gentlydisconnectbyholdingplug.Also,donothandlepowerplugwithwethands.Doingsomaycauseelectric shock.
Donotforciblybend,twistorpullpowercord.Also,donotplaceheavyobjectonpowercord,ordamageormodifyone.Anyofthesemaycausefireorelectricshockduetodamagetopowercord.
DonotdisassembleormodifythisinstrumentorACadapter.Doingsomaycausefireorelectricshock.
ShouldthisinstrumentorACadapterbedamagedorsmokeoroddsmellbegenerated,donotkeepusingonewithoutcorrection.Doingsomaycausefire.Switchpoweroffimmediately,unplugACadapterfromACoutletinsuchcase.Ifoperatedonbattery,removebatteryandcontactthenearestKonicaMinolta-authorizedservice facility.
Caution (Failuretoadheretofollowingpointsmayresultininjuryordamagetoinstrumentorotherproperty.)
UsethisinstrumentnearACoutletforeasypluggingorunplugginginusingACadapter.
Connection procedure 1. Be sure that power switch is OFF ( [ ]
mark pressed).
2. Connect AC adapter plug to AC adapter input terminal of body.
3. Plug AC adapter to outlet (AC 100-240 V , 50/60 Hz).l InsertACadapterplugallthewayseatedinACoutlet.
DC・IN
19
Installing
ON ( | ) / OFF ( ) of Power SwitchTo ON
1. Press [ | ] side of power switch to ON.
To OFF
2. Press [ ] side of power switch to OFF after completion of measurement.
Connecting with PCBeforeconnectingCA-2500bodywithPC,installsoftwareCA-S25winPC.(Fordetail,seetheInstallationGuidefordatamanagementsoftwareCA-S25w.)UseaccessoryUSBcable(3m)IF-A18forthispurpose.Noproblemwouldoccur ifUSBcable ispluggedorunpluggedwhilepower ison,but instructionhereindicates the case when power is off.
Operation procedure1. Switch power off [ ] .
2. Connect USB cable to USB connector on body.
3. Insert all the way to the end. Check whether the connection is secure.
Communication interfaceof this instrument complieswithUSB2.0.AlwaysholdplugportionofUSBcable inunplugging.Neverhold cord portion to unplug. MatchtheUSBcableplugshapewiththeslotofeitherbodyorPCbeforeinserting.AnyUSBport isusableonPCif ithasmultiple,butabnormaloperationcouldbedetectedonPC in case that it isusedtogetherwithotherUSBdevicethanCA-2500.
20
21
Measurement
22
Measurement
Starting Up and Ending SystemStarting up
Switchpoweronasinstructed.Ifnot,erroneousoperationcouldoccur.
1. Switch power on for body. (Press [ | ] side.) l Motorsoundsindicatingcontrollerinsidemachinestartsinitialization.l Fan inside body starts to run. l15-minutewarm-upat least(to leaveinthisstateafterswitchingon.) isrecommendedforaccurate
measurement.
2. Start up Windows by turning on controlling PC.
3. Connect PC and CA-2500 with USB cable.lPCrecognizesthatCA-2500hasbeenconnected.
4. Select "CA-S25w" from Start → [Program].lCA-S25wstartstorun.
Connection recognition error Sometimes,PCfailstorecognizeitsconnectionwithCA-2500ifCA-2500wasswitchedonafterconnectedwithUSBcabletoPCwhichhadalreadystartedup.SuchconditioncouldnotallowcontrolofCA-2500fromsoftware.Startupthisinstrumentaccordingtotheprocedureasaboveagain.l IncasethatPChasalreadystarted,unplugUSBcableonceandthenconnectafterswitchingCA-2500
on. lSwitchCA-2500onbeforePCifUSBcablecannotbeunpluggedforsomereasonconcerningsetting. Shouldconnectionerroroccur,unplugUSBcableonceandconnectagain.Then,bothcanstartup
normally.
Ending
1. Select "End CA-S25w" from "File" of menu bar.lCA-S25wends.lYoucanfinishthroughstandardapplicationoperationofOS.
2. Finish Windows and switch off PC.
3. Switch off body. (Press [ ] side.)
23
Measurement
Setting the measurement distance and the pint ring range index Setting measurement distanceFivetypesofobjective lensesareavailableforCA-2500dependingonthesizeofmeasuringobjects.Therelationbetweenthesizeofmeasuringobjectsandthemeasurementdistanceforeachlensisasfollows:
CA-2500A, CA-2500S, CA-2500SW, CA-2500ST CA-2500A, CA-2500W, CA-2500SW, CA-2500WTStandard lens Wide lens
Size of measurement object (One-side length of square) Measurement distance Size of measurement object
(One-side length of square) Measurement distance
98mmapprox. 250mm 145mmapprox. 200mm
210mmapprox. 500mm 410mmapprox. 500mm
440mmapprox. 1000mm 850mmapprox. 1000mm
890mmapprox. 2000mm 1770mmapprox. 2000mm
CA-2500A, CA-2500T, CA-2500ST, CA-2500WTTelephoto lens With low-magnification macro ring (Macro 1)
Size of measurement object (One-side length of square) Measurement distance Size of measurement object
(One-side length of square) Measurement distance
115mmapprox. 900mmapprox. 57mmapprox. 500mm
275mmapprox. 2000mmapprox. With high-magnification macro ring (Macro 2)420mmapprox. 3000mmapprox. Size of measurement object
(One-side length of square) Measurement distance
27mmapprox. 300mm
Optical lens features in varying light intensity and illuminance distribution of imaging surface if focus ring is rotated. Therefore, lens calibration has been done lens by lens according to the distance from lens to measurementobject(distanceindicatoroffocusring).Eachhasitsowncalibrationcoefficient.Correctcalibrationcoefficientcanbeappliedbyselectingwhatkindof lens isused forwhich focusingdistance,insoftwareCA-S25w.
CA-2500A, CA-2500S, CA-2500SW, CA-2500STStandard lens
Distance With distance label On indicator
0.25m ○ Decimalpointindicatorposition(SeeFigure1-1.)
Intermediateindicatorbetweenpreviousandnextreferencepositions(Center)(SeeFigure1-2.)
0.3m ○ Decimal point indicator position
Intermediateindicatorbetweenpreviousandnext referencepositions(Center)
0.5m ○ Decimal point indicator position
Intermediateindicatorbetweenpreviousandnextreferencepositions(Center)(SeeFigure1-3.)
1 m ○ Centerof"1"
Intermediateindicatorbetweenpreviousandnext referencepositions(Center)
○ Endoffocus-ringtravel([ ]side)
24
Measurement
CA-2500A, CA-2500S, CA-2500SW, CA-2500STWide lens
Distance With distance label On indicator
0.2m ○ Indicatorposition(SeeFigure3-1.)
0.24m ○ Indicatorposition
0.3m ○ Indicatorposition(SeeFigure3-2.)
0.5m ○ Indicatorposition
1 m ○ Indicatorposition
○ Indicatorposition
CA-2500A, CA-2500S, CA-2500SW, CA-2500STWide lens
Distance With distance label On indicator
0.9m ○ Decimalpointindicatorposition(SeeFigure2-1.)
Intermediateindicatorbetweenpreviousandnext referencepositions(Center)
1 m ○ Centerof"1"
Intermediateindicatorbetweenpreviousandnextreferencepositionsin1:2(short-distanceside)(SeeFigure2-2.)Intermediateindicatorbetweenpreviousandnextreferencepositionsin2:1(long-distanceside)(SeeFigure2-3.)
1.5 m ○ Decimal point indicator position
Intermediateindicatorbetweenpreviousandnextreferencepositionsin1:2(short-distanceside)Intermediateindicatorbetweenpreviousandnextreferencepositionsin2:1(long-distanceside)
3m ○ Centerof"3"
Intermediateindicatorbetweenpreviousandnextreferencepositionsin1:2(short-distanceside)Intermediateindicatorbetweenpreviousandnextreferencepositionsin2:1(long-distanceside)
○ Endoffocus-ringtravel([ ]side)
With low-magnification macro ring (Macro 1)
Distance With distance label Precise position on focus ring
○ Endoffocus-ringtravel([ ]side)
With high-magnification macro ring (Macro 2)
Distance With distance label Precise position on focus ring
○ Endoffocus-ringtravel([ ]side)
Caution TheCA-2500has28calibrationpointsasshown in the tableabove,atwhich themeasuringaccuracyisensured.Whenusing the instrumentatapointother than thecalibrationpoints,use thecalibrationcoefficient of the closest calibration point. Doing so may result in a decrease in measuring accuracy.
25
Measurement
Example of reference position to read distanceHow to read distance in case of standard lens or telephoto lensThering(focusring)onwhichnumbervaluesarewrittenrotatesincaseofstandardlensortelephotolens.If the"distanceindicator" includesdecimalpoint,"." isdeterminedasreferenceposition;andif itdoesnotincludedecimalpoint,thecenterofthenumberor[ ]isdeterminedasreferenceposition.
Standard lens
Telephoto lens
How to read distance in case of wide lensThering(focusring)onwhichindicatorposition(thewhitecircleinthefiguresbelow)iswrittenrotatesincaseofwidelens.Settheindicatorpositiontothelineofnumbervalue.
Wide lensFigure 3-1Example when setting to "0.2"
Figure 3-2Example when setting to "0.3"
0.30.5 0.24 0.21 0.30.5 0.24 0.21
Figure 1-1Example when setting to "0.25"
0.30.25 0.30.25 0.5 1
Figure 1-3 Example when setting to "diverting point between 0.5 and 1 in 1:1"
Figure 1-2Example when setting to "diverting point between 0.25 and 0.3 in 1:1"
Setinreferenceofthe1/2position between decimal point of0.5andthecenterof1.
Setinreferenceofthe1/2position between decimal pointsof0.25and0.3.
10.9 1.510.9 1.51
Figure 2-3 Example when setting to "diverting point between 1 and 1.5 in 2:1"
Figure 2-2 Example when setting to "diverting point between 1 and 1.5 in 1:2"
Figure 2-1 Example when setting to "0.9"
Setinreferenceofthe1/3position between the center of 1 and decimal point of 1.5.
Setinreferenceofthe2/3position between the center of 1 and decimal point of 1.5.
26
Measurement
27
Others
28
Others
Outer Dimensions
WARNING:WARNUNG:Niemals das gehause offnen.
Do not remove covers.警告:カバーを開けないでください。
XXXXXX1
192.1
223
Reference position
113160
97
72.7
Tripod socketUNC 1/4, Depth : 6.5
Positioning hole Depth : 4.5
Ø5 positioning hole Depth : 4.5
M5, Depth : Ø6.5
Rubber foot diameter : Ø15
186.2
128.961.2
68.2 6060
81.295.2111.2125.2
160
52.8
91
5
164.
3211.
3
12.7
47
9
29
Others
Error CheckShoulderrorbe found in this instrument, trycorrectiveactionsshown in the following table. If thisdoesnothelp, this instrumenthaspossiblybeenbroken.Pleasecontact thenearestKonicaMinolta-authorizedservice facility with indicated error number.
ErrorNo.
Symptom ItemtoCheck CorrectiveAction
1 Power does not turn on.Does not normally start up.
HasACadapterbeenpluggedin outlet?
ConnectACadapter.
HasACadapterbeenconnected with body?
ConnectACadapter.
HaveyouconnectedcorrectAC adapter?
UseourspecifiedstandardaccessoryAC adapter and power cord or optionally offeredACadapterAC-A312.*WhenyoudisconnectawrongadapterandreconnectanAC-A312adapter,waitat least 5 minutes before turning the power switch on after turning it off.
IstheACpowerwithinrating? BesuretheACpoweriswithin±10%ofthe nominal voltage.
2 Time for machine initialization after switched on varies.
Itcouldvarydependingonthe machine position when switched on.Itshouldnotberegardedas failure if it takes only a few seconds.
Iflongtimeisneeded,switchpoweroffand contactthenearestKonicaMinolta-authorized service facility.
3 PCcannotcommunicate withCA-2500.
CheckwhetherPCcorrectlyrecognizesCA-2500.
PCcouldnotproperlyrecognizethisinstrumentdependingonPCtypeorhow CA-2500wasstartedup.RestartPC, software,andCA-2500accordingtoeach instruction.
Checkwhetherdevicedrivernormallyoperates with device driver manager.
Ifnot,uninstallsoftwareonceforreinstallation.SeetheinstructionmanualforCA-S25w.
HasUSBcablebeenconnected securely?
ConnectthisinstrumentandPCsecurely.
CheckthestandardforUSB. CannotusewiththosesupportingexceptforUSB2.0.UsePCwithwhichUSB2.0isusable.
30
Others
Main SpecificationModel CA-2500S CA-2500W CA-2500T
Light receptor CCD imagesensor (monochrome);2/3-inch;Effectivenumberofpixels:1,000x1,000pixels;EquippedwithXYZfilter(closelymatchesCIE1931color-matchingfunction)andNDfilter
Lens InterchangeableStandard,wide,andtelephotolenses;low-magnificationandhigh-magnificationmacrorings(forusewithtelephotolens)
Measurementpoints(Resolution) 980x980(490x490or196x196selectable byusingDataManagementSoftwareCA-S25w)
Colorindicationmodes XYZ,Lvxy,Lvu’v’, T uv,Dominantwavelength,Excitationpurity,Lv contrast
Display modes Pseudocolor,Chromaticitydiagram,Spot,Crosssection,Colordifference
Measurementsizes(lengthpersideofsquare)(*1)
Standardlens Widelens Telephoto lens Withlow-magnificationmacro ring
Withhigh-magnificationmacro ring
Approx.98mm or more (depending on the distance)
Approx.145mm or more (depending on the distance)
Approx.115mm or more (depending on the distance)
Approx.57mm(Fixed)
Approx.27mm(Fixed)
Measurablesizefortypicalmeasurementdistances(size/distance)
98mm/250mm Approx.
145mm/200mmApprox.
115 mm/900mmApprox.
57mm/500mmApprox.(Fixed)
27mm/300mmApprox.(Fixed)
210mm/500mmApprox.
410mm/500mmApprox.
275mm/2,000mm Approx.
440mm/1,000mm Approx.
850mm/1,000mm Approx. 420mm/3,000mm
Approx.890mm/2,000mm Approx.
1,770mm/2,000mm Approx.
Measurementluminancerange(includingNDfilteruse)
0.05-100,000cd/m2 0.05-100,000cd/m2 0.25-100,000cd/m2 0.25-100,000cd/m2 0.5-100,000cd/m2
Measurementtime(*2)Single:Approx.5sec.ormore;4-timeintegration:Approx.6sec.ormore;16-timeintegration:Ap-prox.10sec.ormore;64-timeintegration:Approx.25sec.ormore;256-timeintegration:Approx.80sec.ormore
Accuracy(*3)
Luminance ±3% ±3% ±3% ±3% ±3%
Chromaticity ±0.005 ±0.005 ±0.005 ±0.005 ±0.005
Temperature/humiditydrift(withintheoperatingtemperature/humidityrange)Luminance:±2%ofchangecomparedtoreferencetemperatureandrelativehumidityof23°Cand40%Chromaticity:±0.004ofchangecomparedtoreferencetemperatureandrelativehumidityof23°Cand40%
Repeatability (*4)
Luminance 0.5% 0.5% 0.5% 0.5% 0.5%
Chromaticity 0.001 0.001 0.001 0.001 0.001
Inter-pointerror(*5)
Luminance(*6) ±2% ±2% ±2% ±2% ±2%
Chromaticity(*6) ±0.002 ±0.002 ±0.002 ±0.002 ±0.002
Luminance(*7) ±3% ±3% ±3% ±3% ±3%
Chromaticity(*7) ±0.003 ±0.003 ±0.003 ±0.003 ±0.003
Other functions Measurementsync(Synchronizationfrequencyselectable),UsercalibrationIntegrationfunction
Interface USB2.0orhigherOperating temperature and humidity range(*8) 10-30°C,Relativehumidity70%orless/Nocondensation
Storagetemperatureandhumidityrange(*8)
0-30°C,Relativehumidity70%orless/Nocondensation,30-35°C,Relativehumidity55%orless/Nocondensation
SizeBody only 160(W)×164(H)×192(D)mm(Heightincludinghandle:211mm)Whenlensandlenshoodareattached 224(D)mm 219(D)mm 224(D)mm 230(D)mm 237(D)mm
Weight 3.5kgapprox.(whenstandardlensandlenshoodareattached)
Power source ACadapter100-240V ,0.75A, 50/60Hz
Standardacces-sories
LensHood CA-A68 CA-A69 CA-A68 CA-A70
HoodCap CA-A71 CA-A75 CA-A71
CalibrationdataDVD Includedwitheachlens.
Other MountCover,ACAdapterAC-A312,ACcable,USBCableIF-A18,DataManagementSoftwareCA-S25w,SDKCA2D-SDK
Optional accessories SoftCaseCA-A60,TripodCS-A3,PanHeadCS-A4,Calibrationcertificate,MuraEvaluationSoftwareCA-Mura
31
Others
*1: Errorinangleofview:7%
*2: Measurementtimediffersdependingonbrightnessofmeasurementobject,PCperformance,anddataprocessingcontents. ThespecificationsabovewereobtainedunderKonicaMinolta'smeasurementconditionsshownbelow: PCCPU : Intel®Core™i7-37703.40GHz Memory :8GB Dataprocessing :Pseudocolordisplayofluminancedata Resolution :980x980 Shutterspeed :Ymeasurement:1/64sec.,XZmeasurement:1/32sec. Measurementsubjectbrightness: Standard/widelens:Approx.80cd/m2,Telephotolens:Approx.300cd/m2
Low-magnificationmacroringandtelephotolens:Approx.400cd/m2, High-magnificationmacroringandtelephotolens:Approx.600cd/m2
*Themeasurementtimebecomeslongerwhentheobjectisdark.Thelongestmeasurementtimeisapprox.14secondswith1-timeintegration,approx.45secondswith4-timeintegration,approx.3minuteswith16-timeintegration,approx.11minuteswith64-timeintegrationandapprox.44minuteswith256-timeintegration
*3:ThespecificationsabovewereobtainedunderKonicaMinolta'smeasurementconditionsshownbelow: Measurementsubjectbrightness: Standard/widelens:Approx.35cd/m2, Telephotolens:Approx.140cd/m2
Low-magnificationmacroringandtelephotolens:Approx.250cd/m2, High-magnificationmacroringandtelephotolens:Approx.250cd/m2
Distance : Minimumdistanceofeachlens Evaluation : Basedontheaverageobtainedwithin10%rangeatthecenterofthescreen Temperature : 23°C±2°C Relativehumidity : 40%±10% Measuringlight : White,referencelightsource Integration : 64times(Normalmode)
*4:ThespecificationsabovewereobtainedunderKonicaMinolta'smeasurementconditionsshownbelow: Resolution : 196x196 Shutterspeed : Ymeasurement:1/64sec.,XZmeasurement:1/32sec. Gain : Normal(x1), Lightintensitylevel: Midpointofappropriatelightintensityrange Evaluation : Basedonthemaximumrepeatability(2 )ofallpixels Temperature : 23°C±2°C Relativehumidity : 40%±10% Measurementsubject: White,referencelightsource Integration : 64times(Normalmode)
*5:ThespecificationsabovewereobtainedunderKonicaMinolta'smeasurementconditionsshownbelow: Measurementsubjectbrightness: Standard/widelens:Approx.40cd/m2,Telephotolens:Approx.160cd/m2
Low-magnificationmacroringandtelephotolens:Approx.200cd/m2, High-magnificationmacroringandtelephotolens:Approx.350cd/m2
Distance :Calibrationdistanceofeachlens Resolution :196x196 Evaluation(*6) :Basedonthemaximum/minimumvaluesobtainedinasquareatthecenterofthescreenmeasuring60%oftheheight
and width of the entire screen (*7) :Basedonthemaximum/minimumvaluesobtainedintheentirescreen
Temperature :23°C±2°C Relativehumidity :40%±10% Measurementsubject :White,referencelightsource Integration :64times(Normalmode)
*8:Even if the instrument isstoredwithin thespecifiedusage(orstorage) temperaturehumidity range, thedisplayedvaluemaychangedependingonlong-periodusageorstorageconditions.
• Windows®,Excel®andWord®areregisteredtrademarksoratrademarksofMicrosoftCorporationintheUnitedStatesandothercountries.• Intel® andPentium® are a registered trademarks or trademarksofIntelCorporationintheUnitedStatesandothercountries.• Thespecificationsanddrawingsgivenherearesubjecttochangewithoutpriornotice.• Forspecification,pleasecontactyournearestKonicaMinoltasalesofficeordealer.