1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ......

47
dIXEL MONITORING UNIT XJ500 Instructions Manual

Transcript of 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ......

Page 1: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

dIXEL

MONITORING UNIT XJ500

I n s t r u c t i o n s M a n u a l

Page 2: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

WARNING:

This controlling and monitoring unit is compliant with standard EN 12830 if it is used together with probes that are compliant with standard EN 13485

Page 3: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Index of Contents

1 GENERAL DESCRIPTION 6 2 THINGS TO KNOW BEFORE THE INSTALLATION 7 3 THINGS REQUIRED FOR THE SYSTEM 8 4 CONNECTIONS 10 5 USER INTERFACE 12 6 PARAMETER PROGRAMMING 14

6.2 PARAMETER DESCRIPTION 14 7 START THE PROGRAMMING 19

7.1 HOW TO START 19 7.2 AUTOMATIC SEARCH OF THE INSTRUMENTS 20 7.3 MANUAL SEARCH OF THE INSTRUMENT 21 7.4 CONFIGURE THE INSTRUMENT 22 7.5 START THE RECORDING 23 7.6 RECORDING STOP 24

8 VIEW 24 8.1 VIEW THE ACTUAL DATA 24 8.2 VIEW THE DATA FROM THE ARCHIVE 25 8.3 VIEW ACTUAL ALARMS 26 8.4 VIEW ALARM FROM ARCHIVE 26

9 PRINT THE DATA 27 9.1 AUTOMATIC PRINTS 27 9.2 MANUAL PRINT 28 9.3 MANUAL PRINT OF THE ACTUAL DATA 30

10 ALARM MANAGEMENT 31 10.1 ALARM DEFINITIONS 31 10.2 SHOW ALARM EVENTS 31 10.3 HOW TO CHECK THE ALARM EVENTS IF HAPPENED DURING THE NIGHT 33 10.4 VIEW THE ACTUAL ALARMS 34 10.5 HOW TO SHOW THE ALARM EVENTS AFTER SOME MONTHS FROM THE BEGIN OF THE RECORDINGS 34 10.6 ALARM REPORT 34 10.7 ALARM FAX MESSAGE 34

11 FAX / SMS 35 11.1 FAX-SMS MANAGEMENT 35 11.2 THE “GENERAL” MENU 37

12 PASSWORD 38 13 SPECIAL FUNCTION 39

13.1 TIMER 1 AND TIMER 2 39 13.2 SYNCHRONISED DEFROST 41 13.3 STAND-BY FUNCTION 42

14 DEFAULT PARAMETER 43 15 TECHNICAL DATA 46 16 LIMITED LIFE TIME WARRANTY 46 17 IF SOMETHING DOES NOT RUN.. 47

Page 4: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.4 / 47

Parameter Flow

Page 5: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.5 / 47

SHORTCUTS

AUTOMATIC SEARCH OF THE INSTRUMENTS

1. Menu SETUP – DEVICE – ADD 2. Push Enter to start the search of the devices connected to the serial line. 3. Wait until the progressive address reaches 224. 4. At the end of the search push the ENTER for each device followed by the blinking NEW message.

CONFIGURATE THE RECORDING FEATURES OF EACH INSTRUMENT

1. Menu SEUP – DEVICE – CONFIGURE – SINGLE 2. Select the instrument with < or > (when INDEX parameter is selected) 3. Select the parameter with the o or n arrows. 4. Modify the value of the selected parameter with < or > (to change the name follow the instruction 7.3.1) 5. Push the MENU key and confirm with ENTER.

START THE RECORDINGS

1. From USER MENU select SYSTEM STATUS with o or n. 2. Push ENTER and then push two times the > key to select REC. 3. Push ENTER to confirm the operation. The REC led will turn on.

STOP THE RECORDINGS

1. From USER MENU select SYSTEM STATUS with o or n. 2. Push ENTER and then push two times the < to select OFF. 3. Push ENTER to confirm the operation. The REC led will turn off.

SHOW THE ACTUAL DATA 1. From USER MENU select VIEW with o or n. 2. Select DATA with o or n and push ENTER. 3. Select ACTUAL with o or n and push ENTER. 4. Select the device to display with < o >.

SHOW THE RECORDINGS

1. From USER MENU select VIEW with o or n. 2. Select DATA with o or n and push ENTER. 3. Select FROM ARCHIVE with o or n and push ENTER. 4. Check the START DATE and the END DATE. 5. Select the DEVICE and push ENTER. 6. Select the instrument with o or n and push ENTER to start the viewing.

PRINT THE RECORDINGS

1. From USER MENU select VIEW with o or n. 2. Select PRINT with o or n and push ENTER. 3. Select FROM ARCHIVE with o or n and push ENTER. 4. Check the START DATE and the END DATE. 5. Select FILTER, push ENTER, set the DATA to print and push ENTER to confirm. 6. Select DEVICE and push ENTER, select the instrument with o or n and push ENTER. Repeat this

operation for all the instrument to print. 7. Push MENU to return to the previous menu. 8. Select START PRINT with n and push ENTER.

Page 6: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.6 / 47

VERIFY THE ALARM ORIGIN

1. From the normal viewing push ENTER, from the programming menu push EXIT and then ENTER. 2. Select the instrument from the list with o or n push ENTER. 3. Push ENTER if you want to stop the buzzer of the instrument. 4. Select the available data to show with n. 5. Push MENU to select the list of instrument again or EXIT to return to exit the viewing.

1 GENERAL DESCRIPTION The XJ500 is a monitoring system which is specific for the control of all data, alarms and the functioning status of refrigeration plants, in food preparation and food conservation processes and also in distribution. This device can communicate with all the Dixell instruments with serial port both DixBUS and ModBUS protocol communication. THE APPLICATION The main characteristic of the XJ500 is that the same unit can work as a “Stand alone system” or be connected to a Personal Computer so it can be easily installed both in small shops and big markets. Typical applications are: food plants, for medical use (blood banks, medicine stores, etc), industrial processes, supermarkets, cold rooms, catering, fast food or normal restaurants, gasoline stations and convenience stores. THE DATA The XJ500 unit receives input data coming from Dixell instruments provided with serial output. For each instrument the monitoring program allows the reading and storage the following information: probe values, alarm and functioning status. It can be interfaced to refrigeration instruments, step controllers, temperature humidity and pressure units with Dixell serial output. THE RECORDING Basically the XJ500 was created to record room temperature, alarm status and functioning status of instruments connected to RS485 line for a year with a sampling time of 15 minutes in order to comply with the European Community standards for food controlling processes 89/108 CEE, 92/1 CEE, 92/2 CEE and 93/43 CEE. The unit can also manage other signals like “Bar” or “Psi” for pressure or “Rh” and any other unit of measurements interfaced through 4/20mA signals. THE ALARMS The XJ manages the “General alarms” defined as probe failure, high and low alarms, digital input activation and the “System alarms” defined as data memory overflow, “no link” with an instrument, “no link” with the RS485 and “no link” with the printer. When an alarm condition is detected, the unit can manage the local print and the fax messages. PRINT The temperature and alarm reports can be printed by enabling either the Automatic print or the Manual print. The printed report shows the temperatures, the interval time of the sampling and the alarms. The special HACCP function allows the user to print automatically, at a desired interval of time, the instant measured temperature. THE FAX/MODEM XJ500 Automatically sends fax or SMS messages in case of alarm situations to a service centre or to a selected telephone number. A list of 15 numbers is programmable by the User. THE INTERACTIVE FUNCTIONS The XJ500 is provided with manual and time-based procedures to manage the synchronised defrost of instruments and execute daily or weekly cycles with the auxiliary relays such as the blind and the light controls. The programming timers of each auxiliary relay and defrost, allow the User to set 90 steps made of on/off operations, 1 operation every 10 minutes. THE SYSTEM PROTECTIONS Access to programming can be protected by activating one of the four available passwords DATA RECORDING Through the RS485 line, the information coming from DIXELL instruments are collected by the XJ500 unit, which logs them. For each instrument, the User can decide what data has to be stored, such as: probe values, probe failure, high and low alarms and output status (defrost, comp. status etc.). The recording flow is a FIFO format (First In, First out) into a Flash memory that ensures a high speed and big data capacity. Before reaching maximum storage space, the XJ500 actives alarm signals. In this case the data should be Down-loaded so it is saved.

Page 7: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.7 / 47

MANAGEMENT OF THE RECORDED DATA It is possible to display all the data directly through the XJ500 LCD display or through the printed reports. The manual print allows the User to print the data available in the memory, the report is selectable in a range of times programmable by the User. In order to have a historical file the unit is also capable of executing daily or weekly prints. THE XView TOOL SOFTWARE Through this tool software it is possible to manage the same programming functions of the XJ500 from a local or remote Personal Computer in order to make the setup procedure easier. The tool software also allows to down load the recordings, executed by the XJ500, to display and print them in table or graph format. SERIAL OUTPUT RS232 MODBUS-RTU The XJ500 will maintain all the characteristics of the previous series, independently recording data and system diagnostics even in mixed networks in which DixBUS and ModBUS protocol instruments are used. Moreover RS232 serial output allows all system data to be transferred to a main ModBUS compatible monitoring system, in this way acting as a gateway.

2 THINGS TO KNOW BEFORE THE INSTALLATION THE UNIT As default the unit is based on 2 Mbyte of memory suitable for recording the data of the first probe (usually the room probe) of 12 instruments for a year. The interval time is defined in 15 minutes. The User can modify this default parameter programming or expand the memory up to 10 Mb of memory. THE PACKAGING Inside the packaging you will find: 1) the XJ500 unit. 2) the instruction manual. 3) 3 disks with the “XVIEW” software

PART FUNCTION CODE NUMBER Memory card of 8Mb Expansion flash memory J0VZZZZ100 RS232 LAP LINK cable Connection cable for PC, 1.8mt, 9 pin, female DD300000 10 Centronic Cable Connection cable for printer DD400000 10

OPTIONAL You can require: A) the LAP LINK cable for the connection to the Personal Computer (not suitable to connect the modem); B) the printer XJ50PR with 40 columns; C) the card memory of 8 Mb. THE FAX MODEM DEVICE Be sure that the modem type is CLASS 1 compatible, V21/V29, G3. THE GMS MODEM DEVICE The SMS modem to be used with XJ500 is the SIEMENS TC35 terminal. Dixell offer a kit, named TC35-KIT, which includes the modem, the power supply unit, the aerial with relevant cable and the connection cable to XJ500. THE PRINTER We suggest the use of a FX EPSON emulation, IMB emulation and HP emulation printers. THE TOOL PROGRAM The “XVIEW” runs under Windows 95/98 and Windows NT platforms.

Page 8: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.8 / 47

3 THINGS REQUIRED FOR THE SYSTEM The XJ500 can be connected in three different configurations: - Without modem and without the PC; - With a modem via RS232 to send data and alarm messages; - With a local PC to manage big systems with lots of instruments.

Assure that all the Dixell instruments that have to be connected to the XJ500 supervisor system are provided with the serial output.

BEFORE STARTING THE SUPERVISOR To run the XJ500 unit read these important information: 1) A shielded cable with 3 wires must reach all the places where the instruments are fitted (eg BELDEN 8772); 2) The wires of the RS485 line must be connected with the interface modules XJRS485; 3) Each XJRS485 interface module must be connected with the corresponding instrument through the 5 pins TTL cable

supplied with the instrument itself; 4) Each instrument must be addressed with its unique number, by programming the parameters Ad2 or Adr ;(Ad1 must be

equal to 0)

THE RS485 SERIAL LINE - The cable of the RS485 line can be a three wires plus shield starting from a minimum of 0,5mm2. (e.g. BELDEN 8772) - From the office where the XJ will be installed, the cable must reach all the instruments controlled by the supervisor. - If possible keep this cable away from power lines or contactors. - It is really important not connect the cable shield to the plant electrical ground. - Draw a map of the followed path can be useful when a line check is necessary. - Do not make loops or other type of connections.

To keep balanced the RS485 the beginning and the end of line must be closed with a resistor. This resistor is already mounted inside the instruments and it can be activated by following these two different applications schemes. The XJ500 is connected at the beginning of the line:

In this case close the Jumper inside the monitoring unit and set the end of the line with a jumper on the terminals 18 - 19. These terminals are indicated on the label of XJRS485. 2) The XJ500 is connected in the middle of the line:

The jumper of the XJ500 must be removed, but to set the beginning and the end of the line it necessary to close the jumper (terminals 18 - 19) of the first and the last instrument connected.

Page 9: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.9 / 47

3) When the last instrument in the net is with built in RS485 with 2 cables (XR100, XJA, XJP etc.) if there are problems in the serial communication it could be necessary to connect an 100ohm resistor in parallel to the serial line to keep balanced the RS485. Do not make loops in the connections

CONNECTING THE INSTRUMENT 1. Each XJRS485 interface module is provided with 3 terminals for the input / output data. 2. Decide which of the 3 wires must be connected to the “+” terminal, to the “-“ terminal and to the “Gnd”, proceed always

respecting the polarity. 3. Some series (XR100, XJA, XJP, XT/XH/XP “R” format) with serial output do not need the XJRS485.

THE TTL CONNECTION 1. Each instrument must be connected to its RS485 interface through a 5 pins TTL cable. 2. If powered with 12Vac, the instrument can be supplied directly from the terminals 7 and 8 of the XJRS485, with a

maximum power of 5VA. 3. In any case keep the TTL cable away from any power lines or any contactors. 4. The series ordered with serial output RS485, does not use the interface XJRS485 and the TTL cable. 5. In the instruments with built in RS485 with two cables must be connected the “+” and the “-“. The GND doesn’t have to

be connected.

Page 10: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.10 / 47

SERIAL ADDRESSING OF THE INSTRUMENTS 1. Each instrument needs a unique address in order to be recognised by the XJ500. 2. This address can be programmed by the Adr parameter in the ModBUS instruments and by Ad2 parameter in the

DixBUS instrument. Ad1 must be set equal to 0 in the instruments with DixBUS protocol communication. Refer to the Installing and Operating Manual of the instrument.

3. To make the description procedure easier, it is better divide the installed instruments in categories having progressive number Adr (Ad2) for each device.

COMPATIBILITY OF THE INSTRUMENTS

Instrument series Standard serial connection Protocol

compatible i XR100/500 with serial output RS 485 ModBUS-RTU V3.xx

XR700 with serial output RS 485 ModBUS-RTU V3.4 WING Series RS 485 ModBUS-RTU V3.xx

XW700 Series TTL + XJ485 ModBUS-RTU V3.4 XM400 Series RS 485 ModBUS-RTU V3.xx XM600 Series RS 485 ModBUS-RTU V3.4

XB Series RS 485 DixBUS V2.00 & V3.xx XC400 Series TTL + XJRS485 DixBUS V2.00 & V3.xx

XC700/800/900M Series RS 485 ModBUS-RTU V3.xx XT/XH/XP Series format C e D TTL + XJRS485 DixBUS V2.00 & V3.xx

XT/XH/XP Series format R RS 485 DixBUS V2.00 & V3.xx XT100 Series RS 485 ModBUS-RTU V3.4

XJP – XJA Series RS 485 ModBUS-RTU V3.xx

4 CONNECTIONS

4.1.1 THE PANEL WITH THE TERMINALS

Page 11: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.11 / 47

4.1.2 THE RS485 CONNECTION AND THE MODEM SUPPLY

4.1.3 CONNECTION DIAGRAM

Page 12: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.12 / 47

4.1.4 THE INTERNAL CONNECTIONS FOR THE POWER SUPPLY AND THE CLOCK BACK-UP

5 USER INTERFACE THE MEANING OF THE LED

SYSTEM The “Net” led signals the working status of the RS485 line; The “REC” led signals the working status of the recording; The “DATA” led signals if there is a queue in the data buffer of the printer or the fax; The “ALARM” led signals the alarm conditions “General alarms” are coming from the instrument. and “System alarms” are

coming from the XJ500.

NET: RS485 line activation.

REC: registrations active

DATA: there is a print or a fax queue.

ALARM: it is ON for “General alarms”. It is blinking for “System alarms”

ALARM: it is ON when the alarm contacts are closed during an alarm condition.

AUX2: it is ON when the relay 2 contacts are closed.

AUX1: it is ON when the relay 1 contacts are closed.

MODEM: it is ON only for one seconds when the modem contatcts are open in order to cut the supply to the modem and reset it.

Page 13: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.13 / 47

OUTPUTS The “Alarm” led signals the working status of the alarm relay; The “Modem” led signals the working status of the modem reset; The “Aux1” led signals the working status of the auxiliary output 1; The “Aux2” led signals the working status of the auxiliary output 2;

KEY FUNCTION WHEN BROWSING

KEY FUNCTION WHEN EDITING

EXIT: to exit the programming without saving.

UP KEY: to scroll up the menu list.

AUX: to show the status of the auxiliary outputs.

MENU: to enter the programming; Back to the previous menu.

DOWN KEY: to scroll down the menu list.

DELETE: no function.

RIGHT key: to increase the parameter value.

LEFT KEY: to decrease the parameter value.

ENTER: to enter a menu or confirm a new programming.

EXIT: to exit the programming without saving.

UP KEY: to change the value of the blinking character. AUX: no function.

MENU: to exit to the previous programming menu;

DOWN KEY: to change the value of the blinking character.

DELETE: to delete the blinking character.

RIGHT KEY: to shift right the blinking prompt.

LEFT key: to shift left the blinking prompt.

ENTER: to enter the editing menu; to confirm the programming.

Page 14: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.14 / 47

6 PARAMETER PROGRAMMING Pushing the MENU key the <USER MENU> list is displayed as follow:

< USER MENU> VIEW

PRINT SYSTEM STATUS

The arrow placed before the parameter or the submenu indicates the selection.

The Down arrow on the low Right side, indicates that there are one or more parameters or sub menu reachable by pushing the Down key.

The Up arrow on the up right corner, indicates that there are one or more parameters reachable by pushing the Up key.

6.1.1 < USER MENU > The User Menu contains all the parameters and submenus. By pushing the MENU key the following items will be showed: VIEW is used to display the data and alarm both for Actual and From Archive; PRINT is used to print the data and the alarm both for Actual and From Archive; SYSTEM STATUS is the parameter that enables the RS485 (ACQ) and the recording (REC) functions; SETUP is used to define the whole parameter programming.

6.2 PARAMETER DESCRIPTION

6.2.1 USER MENU - VIEW This menu allows to show the alarm status and the probe values on the display of the unit. The View function can also display the Actual situation of the plant or the file of recording. ALARM to display all the devices and the active alarms: Actual shows the on line events; From Archive shows the recorded alarms;

DATA displays the probe values read by the devices: Actual shows the on line data; From archive shows the recorded data;

By entering “From archive”, the following filter mask allow to define the time range and the instrument information that have to be viewed:

< SETTINGS > START DATE (First recording to END DATE) END DATE (START TIME to last recording) DEVICE (List of instruments)

Page 15: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.15 / 47

6.2.2 USER MENU - PRINT ALARM to print the active alarms: Actual prints the on line alarm events (alarm-all); From archive prints the recorded alarms;

DATA to print the probe values: Actual prints the on line data (data or alarm & data); From archive prints the recorded data;

< SETTINGS > START DATE (First recording to STOP TIME) END DATE (START TIME to last recording) DEVICE (List of instruments) FILTERS (menu of the information to print) START VIEW

Selection options for Filter menu

< SET FILTERS > Enable Probe1

Enable Probe2 Enable Probe3 Alarm Status

Yes No No Yes No\

SETUP DEVICE(S) prints the settings information of each device (Model, Name, Address, etc.); XJUNIT SETUP prints the settings of the XJ500; AUX ½ OUTS SETUP prints the settings of the auxiliary relay cycles; DEVICE TIMER SETUP prints the settings of the Synchronised Defrost and the Stand-by cycles.

6.2.3 SYSTEM STATUS (PARAMETER) This parameter enables the Acquisition and the Recording functions, the options are OFF, ACQ and REC. By setting ACQ (acquisition) the XJ500 actives the RS485 network in order to collect and display the data coming from the instruments while by setting REC (recording) the unit also records the data into its memory. To change the parameter value push ENTER then > or < keys and then ENTER again to confirm.

6.2.4 USER MENU – SETUP

USER MENU – SETUP – TIME/DATE The Time/date function sets the clock of the XJ500. Remember that if a power failure occurs, the clock back-up is 12 hours. HOUR values from 0 to 23; MINUTE values from 0 to 59; DAY (WEEK) values from SUN to SAT; DAY (MONTH) values from 1 to 31; MONTH values from 1 to 12; YEAR from 1996 to 2999; FORMAT EURopean or AMErican. DLS start month daylight saving time start month DLS start day daylight saving time start day DLS end month daylight saving time end month DLS end day daylight saving time end day

Page 16: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.16 / 47

USER MENU – SETUP – XJUNIT This menu defines the language, the system identification, the acquisition times and the special function: LANGUAGE selectable between two languages. IDENTIFICATION SYSTEM NAME defines the identification name; SYSTEM ID defines identification number; RELEASE defines the software version.

CONTRAST Allows to increases or decreases the display contrast with the < or > keys (0/255). ACQUISITION CLEAR ARCHIVE removes the recording file from the XJ memory; AUTOSTART automatically restarts the SYSTEM STATUS as programmed in this parameter after a power failure (OFF =

the system restarts with the previous status, ACQ = the system restarts with the Acquisition, REC = the system restarts with the Recording);

STORE INTERVAL (1..255 min) to define the interval time between two recording ; SAVE MODE to define the type of recording method: Cont (every Store Interval time) or Int actives the recording only after

a probe value variation to save memory space (not yet available); SAVE TEMP (Yes/No) enables the recording of the probe values; SAVE ALARM (Yes/No) enables the recording of the instrument alarm; SAVE DEV STAT (Yes/No) enables the recording of the instrument status.

SPECIAL FUNCTION This menu allows the User to carry out On/Off operations with the auxiliary relays of the XJ500. These daily cycles called Timer 1and Timer 2 can be used for blinds and light managing or for time based controls. A third timer can manage the synchronised defrosts for one or groups of instruments. TIMER 1, TIMER 2 define two independent On/Off daily cycles based on a step of 10 minutes: TIME (0..24 hours, step=10 min) sets the hour and minutes of the operation; STATUS (Off/On) sets the status of the relay at that hour.

TIMER DEVICE sets the activation/deactivation of some functions (defrost, instrument stand-by, instrument On/Off, light output, Aux output and energy saving cycle) for a single or a group of instruments. It is a daily cycle based on a step of 10 minutes for a total of 144 possible daily commands. CONFIRM SETTINGS allows to confirm the changes; TIME (0..24 hours, step=10 min) sets the defrost times; FUNCTION defines the status of the following functions: Defrost (DEF): starts a defrost cycle Stand by ON (ST-BY ON): sets the selected instruments in Stand by (the instruments are not considered by XJ500) Stand by OFF (ST-BY OFF): sets out the selected instrument from Stand by Device ON (DEV ON): turn the selected instruments on Device OFF (DEV OFF): turn the selected instruments off Light ON (LIGHT ON): turn the instrument light output on (if available) Light OFF (LIGHT OFF): turn the instrument light output off. Energy saving ON (ES ON): starts the energy saving cycle on the selected devices Energy saving OFF (ES OFF): Ends the energy saving cycle on the selected devices Aux output ON (AUX ON): turn the instrument Aux output on (if available) Aux output OFF (AUX OFF): turn the instrument Aux output off. ADD define a list of max 20 instruments for each operation selected from the previous menus; ERASE deletes an instrument from the list; VIEW shows the programmed cycles.

Page 17: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.17 / 47

AUX1 and AUX2 WORKING MODE define the weekly cycle of the Aux1 and Aux2 relays based on the following parameters:

< AUX 1 WORKING MODE> SUNDAY Off, manual, timer1, timer2

MONDAY Off, manual, timer1, timer2 TUESDAY Off, manual, timer1, timer2

USER MENU – SETUP – DEVICE (S) The Device(s) menu allows to detect, configure or remove the instruments connected the RS485 line. To use this parameters it is always necessary have the Acquisition or the Recording not active (System Status =Off). ADD allows the User to detect one or more instruments connected to the RS485: MANUAL you are required to enter the address (Adr) of the new instrument. It must be the same Adr or Ad2 parameter

values already programmed in that instrument; AUTOMATIC the XJ500 will automatically start searching all the instruments from 00 up to the address Adr=224.

REMOVE deletes one or more instruments from the network control. This function deletes also all the configuration files. Single deletes only the selected instrument from the list; All Device(s) deletes all the instrument.

CONFIGURE allows to customise the information relative to the instruments where: SINGLE Index returns the progressive on the total found; Address shows the parameter Adr Name (15 chars) is the identification name of the instrument; Cat (16 names) defines the category at which the instrument belongs (the category names can be programmed only by

using the PC software); Enable (Yes/No) enables whole the instrument control; Sinc Clock (Yes/No) enables the synchronism with the XJ500 clock; Save Alarm (Yes/No) enables the recording of the alarms; Save Status (Yes/No) enables the recording of the functioning status; Auto print (Yes/No) enables the automatic prints of the data of this instrument; UMD defines the unit of measurement; Save Input 1..8 enable the recording of the corresponding input probe (Yes/No).

ALL DEVICES Sinc Clock (Yes/No) enables the clock synchronism with the XJ500 unit; Clock Interval (0..24 hours) defines the time interval between two clock synchronism.

USER MENU – SETUP – ALARM This menu allows to set the alarm delay times and the alarm procedures that will be enabled in case of alarm activation. ALARM CONTROL (Yes/No) enables the whole alarm control. If set to “No” the alarm control is disabled. TIME ESCL. PO (0..255 min) delays an alarm after activating the Acquisition or the Recording functions. E.g: after a power

failure, after the first system start-up etc. TIME CONFIRM (1..255 min) sets the time period between two alarm prints or alarm fax transmissions. PRINT ENABLE NO = no alarm reports; INT = the alarm report is printed after the TIME CONF delay parameter; CONT = the alarm event is printed immediately. TX FAX ENABLE (Yes/No) enables the alarm fax transmission (after the TIME CONFIRM delay). The User can also select

which are the alarms that will enable a fax transmission: LINK 485 for communication breakdown; DIGIT. INPUT for digital input errors; PROBE 1..3 for probe alarms;

Page 18: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.18 / 47

COMP 1-6for maximum compressor running hours of the XC controllers.

USER MENU – SETUP – PRINT This menu allows to enable the printer and define the Automatic prints. PRINT ENABLE (Yes/No) enables the whole printer utility control. If no printer is connected set PRINT ENABLE = No. PRINT MODE (Man = only manual, Day = daily or Wek = weekly) selects the automatic print. PRINT DAY (Sunday .. Saturday) sets the day of the week of the printing; PRINT HOUR (0..24 hours) sets the hour of the printing; PRINT MIN (0..59 minutes) sets the minute of the printing; PRINTER selects the printer type EPSON emulation (normally the continuous form); IBM emulation for deskjet printers; HP emulation for laser printers. COLUMNS selects the printer type: 40, 80 (standard chars) or 132 (compress chars). LINES (62..68) sets the number of line of the paper form if it is a special model; HACCP INT. (OFF and 1 … 24h) select the time interval between two HACCP print: OFF = print disabled HACCP START (00:00 – 23:50) set the starting hour of the HACCP print.

HACCP print is a particular print mode that allows the user to have, automatically and with the desired time interval, a print of the instant values measure by the probes of all the instruments at that time and the status of instruments output.

USER MENU – SETUP – FAX-SMS This menu allows to set all the fax-SMS features and the telephone number list called during an alarm condition. PHONE BOOK is made of 15 programmable phone numbers. Each telephone number is defined by an index from 1 to 16, the referring name and the days of the week when the number is enabled to receive the call:

< SETTINGS > INDEX 1 to 16 SERVICE FAX AREA CODE 12 chars NUMBER 12 chars REF 14 chars SUNDAY Yes / No

N.B. NOT USE THE 16TH INDEX

CALLING GROUPS define the allocation of the phone book number to one of the 4 calling groups. GROUP 1..4 To each groups is possible to assign up to 4 different numbers selected from the phone book. XJ500 will

send a message to all the numbers present in group 1. If none of them is available (for example they are not allowed for that particular day), it will go to group 2 and so on.

FAX AT INTERVAL Enable the sending of a Fax/SMS about the working status of XJ500; INTERVAL (OFF…24) defines the interval (in hours) between the sending of the fax/SMS. Ik set as OFF the function is

disabled. STARTING TIME (00:00…23:50) defines the hour when to send the first fax/SMS about XJ500 status. Starting from this

time, the fax/sms will be sent following the interval selected in the previous parameter. GENERAL defines the general modem features: RING NUMBER before answering (1-15); TIME DELAY BETWEEN TWO CALLS (1..15 min); NUM RETRY (0 - 3) number of retries if the line is not free. SPEED is the modem transmission speed (2400, 4800, 9600, 19200, 38400 baud).

Page 19: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.19 / 47

USER MENU – SETUP - PASSWORD This menu is made of 6 passwords of four chars each one. Only the owner of the Admin password is enabled to define the other 4 User passwords and the corresponding access levels. ERASE : it deletes an existing password; MODIFY : it sets a new password or it can change the old value: ADMIN : it is the Administrator password. The owner can access the whole system programming and all the 4 User

passwords are submitted to this. USER1..4 :these User passwords are enabled only if Admin is already active. REMOTE CON. : it is the password for remote connection.

PERMISSION : it sets the following access levels of the User passwords: Permission Value Default System status Yes/No No Setup Yes/No No Alarm Yes/No No XJUNIT Yes/No No Print Yes/No No Fax-Modem Yes/No No Device(S) Yes/No No Time/Date Yes/No No

7 START THE PROGRAMMING THE DISPLAY After turning on the power supply, the XJ500 lights the LCD green display alternating the date, the time and the memory free space:

DIXELL XJ500 No Name

STAND – BY FREE MEMORY 100%

The display probably will show “DATE and TIME INVALID” therefore execute immediately the “SETTING CLOCK” procedure explained below.

7.1 HOW TO START To make easier the set-up operations, it will be defined a simply starting procedure. In this way the User becomes familiar with the menu flow and the programming: 1. assign an identification name to the system and then set the date and time of the clock 7.1.1 and 7.1.2; 2. detect the instruments connected to the RS485 line 7.2; 3. configure each instrument 7.4; 4. enable the Acquisition and the Recording 7.5;

7.1.1 SETTING THE IDENTIFICATION NAME push MENU key and under <USER MENU>: 1. select SETUP, push ENTER; 2. select SYSTEM, push ENTER;

Page 20: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.20 / 47

3. select IDENTIFICATION, push ENTER; < XJUNIT>

LANGUAGE CONTRAST IDENTIFICATION

4. select SYSTEM NAME, push ENTER; 5. push o or n keys to type the name, push < or > keys to shift left / right the prompt (see editing table at page 12); 6. push ENTER to confirm the new data; 7. push MENU to skip to the previous menu; 8. select SYSTEM ID, push ENTER; 9. push o or n to change the number, push < or > to shift the prompt; 10. push ENTER to confirm.

7.1.2 SETTING THE CLOCK

The internal clock back-up time is 12 hours. To extend this time connect an external back-up battery as showed in wiring connections.

Under USER MENU: 1. select SETUP, push ENTER; 2. select TIME/DATE, push ENTER; 3. select HOUR: by pushing < or > the value changes; 4. select MINUTE, by pushing < or > the value changes; 5. select DAY(of the week), by pushing < or > the value changes from 1 = Sunday to 7 = Saturday; 6. select DAY (of the month), by pushing < or > the value changes; 7. select MONTH, by pushing < or > the value changes; 8. select YEAR, by pushing < or > the value changes;

< TIME / DATE > HOUR 18

MINUTE 53 DAY (week) Mon

10. push ENTER to confirm the data.

7.2 AUTOMATIC SEARCH OF THE INSTRUMENTS It is the procedure to find automatically the instrument connected to the serial line. Under USER MENU: 1. select SETUP, push ENTER; 2. select DEVICE(S), push ENTER; 3. select ADD, push ENTER; 4. select AUTOMATIC, push ENTER:

< AUTOMATIC ADD >

[ENTER TO START]

5. push ENTER key to start, the Address parameter increases from 00 up to 224.

Page 21: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.21 / 47

< RECOGNISING > CHECK ADDRESS 1 DEVICE FOUND 1

Please Wait…. 6. Check the “Device Found” parameter:

The number of devices should be the same as the number of instruments connected to the RS485. If it is not equal, probably the line is interrupted or some Adr (Ad2) parameters are not properly programmed.

6. At the end of the searching each instrument has to be confirmed, an indicator shows the device progressive number

1/11:

< CONFIRMATION > ADDRESS 01 NEW MODEL XR570C PROBE(s) 2

7. The “NEW” label indicates that the instrument found is not present in the actual setup therefore push the ENTER key

to confirm it. If the “NEW” label does not appear, the instrument is still present therefore push DELETE otherwise it overwrites the existing instrument.

By pushing the EXIT key during this procedure only the confirmed devices will be saved into the memory.

If no key buttons are pressed within 40 seconds, the XJ500 exits from the programming. Only the confirmed devices will be saved into the memory.

If you need to restart a new and complete configuration, remember to REMOVE all the devices to avoid having conflicts between old and new devices.

7.3 MANUAL SEARCH OF THE INSTRUMENT This procedure allows the User to add a single instrument to the existing set-up. In this case it is necessary to know the Adr (Ad2) parameter of the instrument in order to set the Family and Device with the right values.

< DEVICE SETUP > ADR 00

set the two values and then push ENTER to start the search.

7.3.1 DELETE THE INSTRUMENT To remove the instruments, enter the USER MENU - SETUP - DEVICES - REMOVE menu select SINGLE and push ENTER. If you select SINGLE (it removes only one device at time), you will be required to select it from the list of available devices actually in the setup. By selecting ALL, be careful because you will remove the complete setup.

Page 22: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.22 / 47

7.4 CONFIGURE THE INSTRUMENT With these parameters it is possible to customize the instrument information. Under USER MENU: select SETUP, push ENTER; select DEVICE(s), push ENTER; select CONFIGURE, push ENTER; select SINGLE, push ENTER:

< CONFIGURE > INDEX 1/11

ADDRESS 01 NAME No Name

INDEX is a counter which returns the device position of the total found (in this case the first of 11 instruments found). The available parameters are:

PARAMETER FUNCTION Index * Device progressive number Address * Serial Address Name Identification device name CAT Category Enable Enabling supervision Sinc Clock Enable XJ500 clock synchronism Read Alarm Enabling the alarm reading Read Status Enabling the status reading UDM Unit of measurement Read Input 1 Enabling input probe 1 Read Input 2 Enabling input probe 2 Read Input 3 Enabling input probe 3 * Parameter not modifiable

7.4.1 SELECT AN INSTRUMENT FROM THE LIST

From the INDEX parameter, push < or > to select the instrument that you want configure.

7.4.2 SELECT THE PARAMETER TO CHANGE

Push the n or o keys to select the other parameters.

7.4.3 CHANGE THE NAME OF AN INSTRUMENT select NAME and push ENTER; push o or n keys to change the character; push MENU + n to toggle between small and capital letters; push MENU + o to select the numbers; push > or < to shift the prompt; push DELETE to delete the character or a blank space; push ENTER to confirm the new data;

Page 23: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.23 / 47

7.4.4 CHANGE THE VALUE OF THE OTHER PARAMETERS

select the parameter with o or n; change the value by pushing < or >; After having changed one or more parameters of the instruments, to confirm the new settings push MENU key:

< SAVE SETTINGS >

[ENTER] TO CONFIRM.

the system requires 20 seconds to update the files.

7.5 START THE RECORDING The SYSTEM STATUS parameter actives the RS485 network and the recording: ACQ (ACQUISITION) allows to collect the data without recording; REC (RECORDING) allows to collect and record the data. The sampling time is programmable under SETUP – SYSTEM – ACQUISITION - STORE INTERVAL (default 15 min).

7.5.1 START THE ACQUISITION Select SYSTEM STATUS and push ENTER; push the > key until “ACQ” appears and push ENTER to confirm:

< USER MENU > VIEW

PRINT SYSTEM STATUS ACQ

When the Acquisition is enabled only the green led “NET” turns to on. The RS485 line is now working and it is also possible to display all the data collected under the VIEW - DATA - ACTUAL menu.

7.5.2 START THE RECORDINGS Under USER MENU: select SYSTEM STATUS and push ENTER; push the > key and select “REC”, then push ENTER to confirm; The SYSTEM STATUS label changes from “ACQ” to “REC” and the green led “REC” turns on:

< USER MENU > VIEW

PRINT SYSTEM STATUS REC

The XJ500 unit is now recording all the data read by the RS485 line.

REMEMBER that the XJ records the data which have been enabled under the CONFIGURE menu and under the ACQUISITION menu. The XJ500 records the data only if the SYSTEM STATUS parameter is set to “REC”.

Page 24: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.24 / 47

7.6 RECORDING STOP

7.6.1 POWER FAILURE In case of power failure the unit stops immediately the recording. However before turning completely off, the XJ500 records the “Black-Out” alarm message. After the power supply restarts, the recording file is verified; the time passed between the last and the new recordings is considered as Black-out duration time. During the black-out event the unit does not record any data and when the power supply is restored, the recordings restart automatically. In this situation both the instrument and the modem are switched off. During this situation the Alarm Relay contacts are closed to signal the alarm event.

REMEMBER that the internal battery of the clock lasts maximum 12 hours without power supply. After this time the unit needs to set the clock parameters.

If you intend to extend the time duration of the clock battery, it is possible to connect a 12Vdc external battery to the back-up terminals (see chapter 4). To avoid turning off the unit, it is necessary to connect an external Back-up system which supplies both the XJ500 and the modem when the supply fails.

7.6.2 WHEN THE RECORDINGS ARE MANUALLY STOPPED When somebody stops manually the recordings by means the parameter SYSTEM STATUS = OFF, the XJ500 saves the date, the time and the “STOP” warning message. This same message will be recorded even in the case that the recordings are stopped from a remote connection executed by the XVIEW program.

8 VIEW This menu allows the User to display and verify the data and the alarms which are available instantaneously on the RS485 line (ACTUAL), or those that have already been recorded (FROM ARCHIVE). The ACTUAL menu is enabled only if SYSTEM STATUS parameter is not set to OFF.

8.1 VIEW THE ACTUAL DATA Under USER MENU: Select VIEW menu, push ENTER; Select DATA, push ENTER; Select ACTUAL, push ENTER:

< ACTUAL DATA 1/11 > No Name OK

PB1: -18.3 °C PB2: - 23.4 °C <STATUS> STA : ON DEF : OFF CC: OFF ES:OFF <OUTPUTS> CP1: ON compressor output

select another instrument with by pushing < or >.

Page 25: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.25 / 47

DESCRIPTION FUNCTION

No name_1 Name assigned by default. The User can change it under the DEVICE - CONFIGURATION menu.

Label on the right side of the name of the instrument.

OK: the instrument is working properly. ALARM: if there is an alarm condition. NoLNK: the communication is absent. STBY: instrument in stand-by.

Pb1, Pb2, Pb3 Values of the probe 1,2 and 3. STA Digital input Status (no alarm) DEF Defrost status (Yes / No) CC Fast Freeze status (Yes / No) ES Energy Saving status (Yes / No) DI Digital input status (active / not active)

Remember that the XJ as default, records only the first probe. To enable the recording of the other probes enter the SETUP - DEVICE(s) - CONFIGURE - SINGLE.

8.1.1 SCROLL THE LIST OF THE INSTRUMENTS

To display the other instruments push the > or < keys, a progressive number on the right side of the title shows the position of the device of the total found (1/11).

8.2 VIEW THE DATA FROM THE ARCHIVE Under the USER MENU: select VIEW, push ENTER; select DATA, push ENTER; select FROM ARCHIVE, push ENTER;

< DATA LOGGED > START DATE

END DATE DEVICE

8.2.1 SET THE START DATE (eg. 08:30, 04 Monday / 01 / 1999) select START DATE and push ENTER; select the parameter with o or n keys; change the value with the < or > keys:

< START DATE > HOUR 08

MINUTE 30 DAY 04

4. after having programmed HOUR = 08, MINUTE = 30, DAY week = 2, DAY month = 4 MONTH = 1 and YEAR = 1999, press ENTER key to confirm the date;

Page 26: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.26 / 47

8.2.2 SET THE STOP DATE (eg. 11:45, 16 (Saturday) / 01 / 1999) after having programmed HOUR = 11, MINUTE = 45, DAY week = 7, DAY month = 16, MONTH = 1 and YEAR = 1999, press ENTER key to confirm the date;

8.2.3 SELECT THE INSTRUMENT TO VIEW Select DEVICE, push ENTER

< SELECT DEVICE > CABINET 1

CABINET 2 COLD ROOM

push o or n key to select the instrument; push ENTER to confirm and back to the previous menu;

Remember that as default, the START DATE is always the first recording executed by the XJ500 and the STOP DATE is always last available recording.

8.3 VIEW ACTUAL ALARMS If there is an active alarm: push MENU KEY to enter the USER MENU; select VIEW – ALARM – ACTUAL:

< ACTUAL ALARM 1/4 > No Name 1

Pb1: HIGH Pb2: -14.3

To display other instruments push the > or < keys, a progressive number on the right of the title shows the position. NAME MEANING STATUS

Name Instrument name

Alarm BkOut (Black Out) Stop (manual stop) NoLnk (no link) NoEnb (not enabled)

Er1 Bad error OK / Lost Er2 Clock error OK / Lost

IN1..IN3 Probe inputs OK HIGH alarm LOW alarm ERROR probe failure

8.4 VIEW ALARM FROM ARCHIVE As for the DATA showing, also for the recorded alarms it is necessary to set the interval time between the START DATE and the END DATE which are normally set to the first and last recordings of the archive. Further you must select the instrument under the DEVICE menu.

Page 27: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.27 / 47

< SETTINGS > START DATE

END DATE DEVICES

Remember that as default, the START DATE is always the first recording executed by the XJ500 and the STOP DATE is always last available recording.

Push Enter when DEVICES is selected, the display will show the list of available instruments:

< SELECT > CABINET 1

CABINET 2 COLD ROOM

push o or n to select the instrument; push ENTER to confirm the selection.

< CABINET 2 > from: 03 / 03 / 1999 15:30

to: 03 / 03 / 1999 18:30 Pb1: HIGH

push o or n to show other information if available; push the > or < to show the successive date of the recordings.

REMEMBER that this archive does not record the maximum or minimum temperature values detected during the alarm condition.

9 PRINT THE DATA The print operations can be carried out with a MANUAL or AUTOMATIC procedure. The MANUAL can be executed at any time by the User while the AUTOMATIC must be programmed before starting the recording.

If you do not connect a printer, remember to set the parameter PRINT ENABLE = No, this to avoid having problems with the print queue.

9.1 AUTOMATIC PRINTS The automatic print allows the User to have a daily or weekly reports of the data collected by the instruments. All the setting operations must be carried out before starting the recordings. The report always shows the probe values and the alarms.

The automatic print can be selected between a daily print or a weekly print, it depends on the PRINT MODE parameter.

The SYSTEM STATUS parameter must be Off then under USER MENU: - select SETUP, push ENTER; - select PRINT, push ENTER; - select PRINT ENABLE, modify it to YES;

Page 28: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.28 / 47

- select PRINT MODE, modify it to Day (daily) or Week (weekly); - select PRINT DAY, set the day; - select PRINT HOUR, set the hour; - select PRINT MIN, set the minutes; select PRINT COLUMNS, set it with the printer columns (40 for small printer, 80 standard printer, 132 for compress

characters); select LINES, set the corresponding number of lines of the paper form (if needed); press ENTER key to save and exit.

< SETUP PRINT > PRINT ENABLE Yes

PRINT MODE Wek PRINT DAY Tue

9.2 MANUAL PRINT The manual print allows the User to print all or a part of the recorded data, alarm and status during the normal functioning of the unit. Under the USER MENU; select PRINT, push ENTER; select DATA, push ENTER; select FROM ARCHIVE, push ENTER:

< SETTINGS> START DATE

STOP DATE DEVICE(s)

FILTER START PRINT...

9.2.1 SET THE START DATE Select START TIME and push ENTER:

< START DATE > HOUR : 16

MINUTES : 59 DAY : 8

the HOUR parameter is selected; change the value with the < or > keys; select MINUTES with n; change the value with the < or > keys; select DAY(week) with n; change the value with the < or > keys; select DAY(month) with n; change the value with the < or > keys; select MONTH with n;

Page 29: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.29 / 47

change the value with the < or > keys; select YEAR with n; change the value with the < or > keys; press ENTER to confirm the new data;

9.2.2 SET THE STOP DATE Enter this menu and set or verify the last recording date.

If you do not select any START or END DATE, the Xj500 considers the interval time between the first and the last recording of the entire archive.

9.2.3 SELECT THE INSTRUMENT TO PRINT Select DEVICE and push ENTER: select the device and push ENTER:

SINGLE selects one instrument at time; ALL DEVICE selects all the devices;

push ENTER on SINGLE;

< SELECT DEVICE > Cabinet_1

Cabinet_2 Cabinet_3

push o or n arrows to select the instrument from the list of available; push ENTER to confirm it; push MENU key to exit to the previous menu.

9.2.4 SET THE FILTER Select FILTER and push ENTER, these the data which can be printed:

< SET FILTERS > Enable Probe1

Enable Probe2 Enable Probe3

Yes No No

Alarm Status

Yes No

Select with o or n keys; Enable with the < and > keys the data to print (toggle yes/no); push ENTER to confirm.

9.2.5 START PRINT Select START PRINT and push ENTER.

9.2.6 RESET THE PRINTING QUEUE If a manual print report is launched, the User can stop the print-out by pushing the DELETE key and then ENTER to confirm the operation.

Page 30: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.30 / 47

This procedure is valid if executed when the User is not working with the menu and the DATA led is still on. TYPICAL REPORT

DESCRIPTION At 12:50 everything is working properly; At 13:05 somebody turned manually the recording Off; At 13:12 somebody turned the recording On again; At 13:12 the probe P1 of the device 1 was in high alarm “H”; At 13:27 the probe P1 is still in alarm; At 13:42 the device 1 is in defrost; the probe P2 of the “Username 2” is faulty (open or short-circuited); At 13:57 the probe P2 of the “Username 2” is still faulty; At 14:12 the “User name 1” lost the communication with the XJ (NoLnk); the probe P2 of the “Username 2” is still faulty; At 14:27 there has been a power supply black-out (BkOut);

9.3 MANUAL PRINT OF THE ACTUAL DATA The option USER MENU – PRINT DATA – ACTUAL allows to print all the data read instantaneously by the RS485 line:

< PRINT ACTUAL DATA > ACTUAL

FROM ARCHIVE DATA (ALL)

By means of the < or > keys, the User can choice to print only the DATA or, with the ALL option, print the data, the alarms, and the functioning status.

9.3.1 PRINT THE SETUP OF THE INSTRUMENT This printing shows all the information of the parameters of the CONFIGURE menu of the instruments (not the parameters of the instrument itself).

9.3.2 PRINT THE SETUP OF THE XJ500 This printing shows all the information of the parameters of the XJ500 relative to the recordings.

9.3.3 PRINT THE SETUP OF THE AUX1 AND AUX2 RELAY CYCLES This printing shows the time table of the daily cycles followed by the auxiliary relay AUX1 and AUX2 of the menu “Special Function”.

Page 31: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.31 / 47

9.3.4 PRINT THE SETUP OF THE DEVICE TIMER This printing shows the time table of the “Defrost Synchronism” and the Stand-by functions.

10 ALARM MANAGEMENT

10.1 ALARM DEFINITIONS The Historical Alarm Archive contains the recordings executed with the TIME INTERVAL parameter. This file can be consulted in any time by entering the VIEW – ALARM – FROM ARCHIVE menu. The information contained in this archive can be used also several months after the date of the recording. The Event Alarm Archive contains the last 100 alarms with the full description of the event such as: start and stop dates, origin of the alarm, probe temperature, setpoint of the instrument. This alarm information can be showed at any time just pushing ENTER directly of the keyboard when the XJ500 is not under the key MENU activity. This is a FIFO file format and, after 100 alarm events, the first recording will be pushed out from the archive. The information contained in this archive are the same used to send the fax alarm message and to print alarm report.

THINGS YOU MUST KNOW Enable the whole alarm management under: USER MENU –SETUP - ALARM - ALARM CONTROL (default not enabled). Check the alarm management of a single instrument under USER MENU – SETUP - DEVICE(S) – CONFIGURE - SAVE ALARM.

The alarm print out can be enabled under USER MENU –SETUP - ALARM – PRINT ENABLE.

Enable the alarm fax transmission under: USER MENU –SETUP - ALARM - TX FAX ENABLE. After activating the SYSTEM STATUS parameter (ACQ or REC), no alarms are detected for the time set in TIME ESC PO parameter value. During an alarm situation with the "DATA" led turned on, do not overload the unit with manual prints or other operations (the XJ500 is trying to print or send a fax).

The alarm relay stays on for all the alarm duration time and it can never be stopped.

10.2 SHOW ALARM EVENTS When an alarm occurs the XJ500 displays a warning message, the buzzer is on and the alarm relay contacts are closed.

10.2.1 HOW TO VIEW THE ALARM EVENTS When there is an alarm condition, press the ENTER button to show the list of the last 100 alarm events. If you are working under the MENU key activity, push the EXIT key to return to the main display. 1. Push ENTER key to show the last 100 events:

DIXELL XJ500 SMALL MARKET DEVICE ALARM TIME: 17:10 DATE 03:03:1999

2. If the password is active, you are required to type the right value to silence the buzzer:

Page 32: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.32 / 47

ENTER PASSWORD

* * * *

3. Then it is possible to show the ALARM EVENTS ordered by time:

< ALARM LIST > CABINET 5 On (last event)

CABINET 1 End CABINET 6 Stop

The first label identifies the instrument name which is in alarm condition (eg. CABINET5) while the successive label corresponds to the alarm status:

Message on the display Alarm situation

End Automatically Recovered Stop Recovered by manual stop of the recording NoLk Recovered by a no link Bout Recovered by a Black-Out Stby Recovered by a stand-by

4. To show all the information of the event, push the o or n arrows to select the item and then push ENTER:

< CABINET 5 > Pb1: HIGH TEMPERATURE fm: 03:03:1999 12:20 to: Active

5. Push the n arrow to see the temperature of the probe and the setpoint value at the time of the alarm activation:

< CABINET 5 > to: Active Probe: 10.8 Set: 5.0°C

6. Push the < or > arrows to display the information relative to other alarm events of the archive.

10.2.2 SILENCE THE BUZZER OF THE INSTRUMENT If the alarm is still active, push one time the ENTER key to silence the buzzer of the instrument you are looking at the display in that moment.

Page 33: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.33 / 47

10.3 HOW TO CHECK THE ALARM EVENTS IF HAPPENED DURING THE NIGHT If an alarm event starts and recovers during the night time or during an holiday and nobody acknowledged it (nobody has touch the ENTER key to see what was happening in that moment), the display of the XJ500 shows the following blinking message: “CHECK ALARM”:

DIXELL XJ500 SMALL MARKET CHECK ALARM TIME: 17:10 DATE 03:03:1999

In this case the User must push the ENTER key to find the list of alarm events and the instrument or the instruments affected by an unchecked alarm :

< ACTUAL ALARMS> CABINET 5 On

CABINET 1 End CABINET 6

10.3.1 FIRST LOOK

DIXELL XJ500 SMALL MARKET DEVICE ALARM TIME: 17:10 DATE 03:03:1999

To identify the problem have a look at the 4 leds under the label “System” combined with the alarm MESSAGE:

Message on the display Buzzer Alarm led

DEVICE ALARM (Probe failure, high and low, digital input) On Always on

NO LINK ALARM (Instrument not linked) Cycling Blinking with Net led

FAX ALARM (Fax transmission failed) Cycling Blinking with

Data led

PRINTER ALARM (Print report failed) On Blinking with

Data led

CLOCK ALARM (the clock alarm) Cycling Blinking with

Rec led

CHECK ALARMS (Acknowledge alarm) Off Off

Page 34: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.34 / 47

10.4 VIEW THE ACTUAL ALARMS Under USER MENU – VIEW – ALARM – ACTUAL the User can verify the exact origin of the problem and the number of devices in alarm:

< ACTUAL ALARM > No Name 1/11

Pb1: HIGH Pb2: -14.3

If no alarm is present this menu does not shows any device.

10.5 HOW TO SHOW THE ALARM EVENTS AFTER SOME MONTHS FROM THE BEGIN OF THE RECORDINGS

See the VIEW – ALARM – FROM ARCHIVE cap 8.3. IMPORTANT NOTE

During a Manual or Automatic print the alarm warning procedure will be postponed until the previous printing or fax queue is finished. In any case the alarm information is stored.

If the alarm duration is lower than the sampling interval time STORE INTERVAL, It will be not recorded because it is not considered a dangerous alarm.

10.6 ALARM REPORT When a new alarm event intervenes, the User can decide the type of print with the PRINT ENABLE parameter: if PRINT ENABLE = NO the alarm print is not enabled. if PRINT ENABLE = CON each alarm event correspond to an immediate print out of the event itself. if PRINT ENABLE = INT the print is delayed for the time set in TIME CONFIRM parameter. Therefore if e.g. TIME CONFIRM = 60 minutes, the local print out will be executed ONLY if there have been changes in the alarm conditions during the last hour. Remember that the print and the fax have the same report.

With PRINT ENABLE = CON it is possible that you fill find one report page for each alarm event. This can happen because some printers gives automatically the form feed command (eg laser and desk jet).

If the alarm print report can not be printed, after two minutes the printing queue will be erased after two minutes. In this case the message PRINTER ERROR will appear on the display.

10.7 ALARM FAX MESSAGE If the fax device is enabled under USER MENU - SETUP - ALARM - TX FAX ENABLE, the XJ500 will provide to send a warning message to the selected priority phone numbers (see the CALLING NUMBER list). The XJ unit sends the alarm report every interval time defined in “TIME CONFIRM”. Therefore if e.g. TIME CONFIRM = 60 minutes, the fax will be sent ONLY if there have been changes in the alarm conditions during the last hour. When it is necessary to send a fax message, the XJ500 unit selects the first phone number programmed in “CALLING NUMBER” . If the line is not free, the unit will attempt the connection for the NUMBER OF RETRY, the interval time is defined by TIME DELAY BETWEEN TWO CALLS. If no connection has been established, the unit selects the second number of and tries again.

Page 35: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.35 / 47

If the alarm fax message can not be sent for any reason, the fax will be cancel after the last retry and the FAX ERROR message will appear on the display.

10.7.1 THE FAX REPORT The alarm report sent through the fax shows the information coming from the plant situation and collected by the Alarm Event Archive:

dIXEL XJ500 ALARM MESSAGE Date: 06/03/99 Time:17:32 System Name: SUPERMARKET

06 CABINET5 PB1 HIGH ALARM fm: 06/03/99 17:00 to 06/03/99 17:15 PB1: -10.6 °C SET:-18°C

02 CABINET1 PB2 FAILURE fm: 06/03/99 17:00 to: Active

05 CABINET3 NO LINK ALARM fm: 06/03/99 17:00 to 06/03/99 17:25

Black out restored at 06/03/99 16:34

The report shows the Adr address, the name of the instrument, and the description of the alarm event. THE MODEM RESET At the end of a fax transmission or when the communication is lost, the relay contacts of the modem output will be instantaneously opened in order to cut the supply of the modem device and obtain a complete reset. This operation can be executed only by following the modem connection exposed in chapter 4.

11 FAX / SMS NOTE In order to use the SMS service, a GSM modem with a valid SIM card enables to SMS sending has to be connected to XJ500 The fax transmission and the modem call are based on a list of 15 programmable telephone number list called PHONE BOOK. From this list the User can assign up to 4 phone numbers to each of the 4 calling groups. XJ500 will send a fax/SMS to each number present in group 1. If none of this is not possible (for example the numbers of group 1 aren’t enabled for that day of the week), the system will try with the numbers present in group 2. If this is not possible too, it’ll pass to group 3 and so on

11.1 FAX-SMS MANAGEMENT The fax-SMS procedure sends alarm warning messages to the selected telephone number.

THINGS YOU MUST KNOW The whole alarm control is enabled under SETUP – ALARM – ALARM CONTROL (default=no); Set the new telephone numbers under SETUP – FAX-SMS – PHONE BOOK; Set the features of the phone call under the FAX-SMS - GENERAL menu;

Page 36: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.36 / 47

Set, under SETUP – FAX-SMS – CALLING NUMBER, the priority list of the 4 phone numbers to call;

the 4 calling groups for the Fax/SMS are selected in SETUP – FAX-SMS – CALLING GROUPS The whole alarm fax transmission is enabled under : CONFIGURATION – ALARM – FAX ENABLE;

11.1.1 PROGRAM THE PHONE BOOK Under USER MENU – CONFIGURATION – FAX-SMS:

< SETUP FAX-SMS > PHONE BOOK

CALLING GROUPS FAX AT INTERVAL

- select PHONE BOOK, push ENTER; - select ADD and push ENTER (see the next table for the parameters);

< PHONE BOOK > INDEX: 1/16

SERVICE FAX AREA CODE: 0039 NUM:

- Select SERVICE and then use < or > to change from FAX to SMS and vice versa;

- Select AREA CODE and push ENTER; - use o or n to change the value; - use < or > keys to shift the prompt; - Push ENTER key to confirm; - Repeat this operations for NUM and REF; To enable the day of the week use < or > keys;

CONFIRM THE PROGRAMMING

After programming the data, push the ENTER key when INDEX parameter is selected within 30 seconds. - push MENU key to go to the previous menu. PARAMETER MEANING INDEX Progressive number of the list SERVICE Type of service (FAX or SMS) AREACODE Country and city number NUM Telephone number REF Phone identification name SUN..SAT Enabling the day of the week

11.1.2 MODIFY THE PHONE BOOK select MODIFY, push ENTER; follow the same steps described for the ADD procedure; push MENU key to return to the previous menu.

Page 37: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.37 / 47

Example of PHONE BOOK programming PARAM FIRST SECOND Index 1 / 16 2 / 16 Service FAX SMS Area code 0039 437 0039 338 Num 012012 1491258 Ref Maintenance Administrator Sunday Yes No - - - Saturday Yes Yes

N.B. NOT USE THE 16TH INDEX

11.1.3 ERASE THE PHONE BOOK select ERASE, push ENTER; select the number with < or > key; push DELETE to erase it; push MENU or Exit key;

11.1.4 THE CALLING GROUPS MENU The CALLING GROUPS list defines the four numbers enabled to receive the FAX/SMS messages of each of the 4 calling groups. The numbers are selected from the PHONE BOOK list:

< CALLING GROUPS > GROUP 1

GROUP 2 GROUP 3

Select a calling group and press ENTER The list of number belonging to the group will be displayed;

< NUMBER LIST > TEL1: ASSIGNED

TEL2: ASSIGNED TEL3: FREE

Select one of the FREE labels, push ENTER; The PHONE BOOK list will be showed; Select the number with < or > then push ENTER; Push MENU to return to the previous menu. To erase a number, push the DELETE key.

11.2 THE “GENERAL” MENU This menu allows to set the features of a call phone. RING NUMBER (1 to 15) selects the number of rings before answering to a call from the XVIEW program; TIME WAIT CALL (1 to 15 min) selects the time interval between two calls of the same number made by the XJ500 during an alarm condition.

Page 38: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.38 / 47

NUM RETRY (1 to 3) sets the number of calls before passing to the successive number programmed in the CALLING NUMBER list (see the SETUP FAX). SPEED define the connection speed (2400…4800…9600…19200…38400 baud)

12 PASSWORD The password menu allows the User to limit the free access to the XJ500 programming. There are 6 available passwords of 4 characters each one. The User 1 to User 4 passwords are submitted to the Administrator which can enable or disable the corresponding permission. PASS NAME PARAMETER ACCESS ADMIN Full access USER1 Only those enabled in Permission USER2 Only those enabled in Permission USER3 Only those enabled in Permission USER4 Only those enabled in Permission REMOTE CON Only for remote connection

To limit the access follow these steps: Set the ADMIN password. Set the name of USER1. Set the Permission of USER1. Repeat the step 2 and 3 for the other passwords if necessary.

12.1.1 PASSWORD MANAGEMENT The ADMIN password allows the “Administrator” to control the whole access to the XJ programming. Only the Administrator will be able to set, change and erase the other USER passwords or assign the corresponding Permissions. As default no password is programmed. As soon the Admin password is set, after pushing the MENU key the display shows:

ENTER PASSWORD * * * *

you are required to ENTER your password. If you try to enter a menu not enabled (under PERMISSION) the “ACCESS DENIED” message will be displayed. Set/modify the password Under USER MENU – SETUP – PASSWORD - MODIFY: Select the password to set (ex ADMIN)), push ENTER; Set the password name with o, n, < and > keys; Re-enter the password name to confirm the value; Push ENTER key to save it;

The “PASSWORD ACTIVE” message will appear on the display.

12.1.2 ERASE A PASSWORD USER MENU – SETUP – PASSWORD - ERASE: Select ERASE, push ENTER;

Page 39: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.39 / 47

Select the password to erase, push ENTER; The message “PASSWORD ERASED” will appear on the display.

By erasing the ADMIN password, the USER1 to USER4 passwords will be automatically erased.

12.1.3 THE PERMISSION MENU USER MENU – SETUP – PASSWORD - MODIFY - PERMISSION: Select one of the USER password and push ENTER;

< SYSTEM SETUP > SYSTEM STATUS Yes

SETUP No ALARM Yes

Select the menu with o, n keys; Enable the menu with < or > key; At the end of the whole programming push ENTER key to save the new data.

THE PERMISSION TABLE FUNCTION PERMISSION SYSTEM STATUS Yes / No SETUP menu Yes / No ALARM menu Yes / No PRINT menu Yes / No FAX / MODEM menu Yes / No DEVICE(s) menu Yes / No TIME/DATE menu Yes / No

12.1.4 PASSWORD FOR REMOTE CONNECTION By setting this password the remote connection via modem is enabled. The XVIEW program needs the password name of the local system to be connected otherwise the connection will be denied.

13 SPECIAL FUNCTION 13.1 TIMER 1 AND TIMER 2 The XJ500 unit is provided with two auxiliary relays which can be related to a daily or a weekly cycle. Two different timers are programmable in order to define the cycle executed by the relays. Possible applications for this function are the light and the blind controls.

HOW PROCEED 1) Define the ON/OFF cycle for Timer 1 and Timer 2 2) Select the relation between the relay outputs and the timers.

DEFINE THE TIMER FOR THE CYCLE TIMER 1 and TIMER 2 are daily cycles made of 144 On/Off operations with a step of 10 minutes. Each timer can be program independently and used during the week both for Aux 1 and Aux 2 relays. Example:

Page 40: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.40 / 47

INTERVAL TIME RELAY STATUS from 8:00 to 12:30 on from 12:30 to 16:00 off from 16:00 to 20:00 on

Under SETUP-SYSTEM-SPECIAL FUNCTION-TIMER: with the o key select TIME = 08:00; push > and keep it pushed: the STATUS label changes from OFF to ON; by keeping the > pushed, push also < until the TIME reaches 12:20; release all keys; push o to set TIME = 12:30; select OFF with < and keep it pushed; by keeping < pushed, push also the o until the TIME = 16:00; release all the keys; push > and keep it pushed: the STATUS = ON; by keeping the > pushed, push also the o until TIME = 20:00; press ENTER to confirm the programming:

13.1.1 SET THE RELATION BETWEEN THE TIMER AND THE RELAY OUTPUT Under the parameters WORKING MODE, the User is allowed to define day by day the operations of the relay output:

WORKING MODE STATUS Selectable Function OFF the relay will not execute any operation. TIMER 1 the relay will execute the cycle set in Timer 1. TIMER 2 the relay will execute the cycle set in Timer 2. MANUAL the relay can be switched ON and OFF by mean of the AUX key button

Next example shows the procedure to define a weekly cycle executed by the auxiliary relay 1:

WORKING MODE Day Relay AUX 1 Sunday OFF Monday Timer 1 Tuesday Timer 1 Wednesday Timer 2 Thursday Timer 1 Friday Timer 1 Saturday MANUAL

Page 41: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.41 / 47

Under SETUP - SYSTEM - SPECIAL FUNCTION - AUX1 WORKING MODE; select Sunday = OFF; push one time n for Monday then push > to select TIMER 1; push one time n for Tuesday then push > to select TIMER 1; push one time n for Wednesday then push > to select TIMER 2; push one time n for Thursday then push > to select TIMER 1; push one time n for Friday then push > to select TIMER 1; push one time n for Saturday then push > to select MANUAL; push ENTER to confirm the new data. The same method is used to set the AUX2 weekly cycle. Remember that Timer 1 and Timer 2 are independent and they can be used for both the relays.

13.1.2 SET THE MANUAL CONTROL FROM THE XJ500 KEYBOARD Suppose that every Saturday the AUX1 relay must be manually turned on and off by the user. To make this operation set under SETUP - SYSTEM - SPECIAL FUNCTION - AUX1 WORKING MODE = MANUAL (Saturday). When the instrument is not in programming mode, by pushing the AUX key the User can check which is the active function selected for the auxiliary relay. If no function is selected the DISABLE message will be showed:

< AUX STATUS > AUX1: OFF AUX2: MANUAL

If for that day the MANUAL option is selected, it is possible without entering the programming to manage the ON/OFF status: Select the AUX1 with o or n; Push < or > to change the status; Push ENTER to confirm. By repeating this operation you can switch manually the relay.

13.2 SYNCHRONISED DEFROST This function allows to set a synchronised defrost cycle both for one instrument and groups of instruments. The synchronised defrost is based on a daily timer which is programmable with a step of 10 minutes for a total of 144 possible defrosts, and moreover the User can select from 1 to 10 instruments for each defrost operation. The defrost timer follows the Real Time Clock of the XJ500 which send the defrost command to the selected instruments. Each instrument will carry out the defrost by following the local evaporator condition and the parameter programming. For example: if the evaporator temperature is too high the defrost will not be executed.

If the XJ500 loses the clock synchronism, each instrument will automatically restart working with its parameter programming defrost cycle.

13.2.1 THE PARAMETER OF THE DEVICE TIMER TIME sets the hour and the minutes of each defrost with a step of 10 minutes; FUNCTION enables the defrost function at the selected TIME; ADD adds to a list up to 10 devices to one defrost command; ERASE deletes the instruments from a list; VIEW allows to show the defrost setting.

Page 42: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.42 / 47

CONFIRM SETTING after changing the previous parameters, select this label with n key and then push ENTER key to confirm the new programming. SET THE DEFROST TIMER Under USER MENU – SETUP – SYSTEM – SPECIAL FUNCTION - DEFROST TIMER:

< SPECIAL FUNCTION > TIME 00:00

FUNCTION Def ADD (to select the instruments)

set the defrost TIME with < or > keys; select FUNCTION with n, change it from OFF to DEF with >; select ADD (a instrument) with n; push ENTER to display the list of available instruments: select the instrument with < or > key; press ENTER to confirm the one you want add to the list (max 10 devices); push MENU to reach the previous menu; TO CONFIRM the entire operation, push MENU then push ENTER (otherwise DELETE).

VIEW THE DEFROST SETTINGS To view the list of instrument that are enabled to defrost, select the TIME parameter with < and > keys until the FUNCTION shows “DEF”. At this point select the VIEW parameter and push ENTER, you will be able to verify the list of instruments by pushing > or < key.

13.2.2 EXAMPLE OF PROGRAMMING The following programming shows as the devices can defrost under XJ500 clock control:

13.3 STAND-BY FUNCTION The Stand-by function allows to stop automatically the recordings and the alarm management of an instrument for a period of time (such as the night time). The User is allowed to program this interval time under the “Device Timer”.

< SPECIAL FUNCTION > CONFIRM SETTINGS.....

TIME 20:00 FUNCTION St-by ON

For example a display cabinet is not used during the night time because the food is stocked in another cold room. Suppose that this cabinet from the 20:00 (08:00 pm) to 08:00 (08:00 am) will be in stand-by, therefore: select TIME and set 20:00 (to 08:00 am) with \ key; select FUNCTION with n key; set ST-BY ON with > key;

Page 43: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.43 / 47

select ADD with n key and push ENTER;

< ADD DEVICE > INDEX 1/6

ADDRESS: 00:01 NAME: CABINET_1

select the instrument with > or < keys; confirm the selection with ENTER; Stop time push MENU and with o key select TIME; set 08:00 (08:00 am) with > key; select FUNCTION with n key; set ST-BY OFF with > key; select ADD with n key and push ENTER; select the same instrument as the point (6) with > or < keys; confirm the selection with ENTER; push MENU and then MENU again:

< SAVE SETTINGS >

[ENTER] TO CONFIRM.

Push ENTER key to save the whole changes or any other key to abort.

The recordings for this instrument(s) will follow this daily cycle:

14 DEFAULT PARAMETER Parameter Description Limits Default User

USER MENU – SETUP – TIME DATE HOUR Hour of the XJ500 clock 0 ÷ 24 hours 00 MINUTES Minute of the XJ500 clock 0 ÷ 59 min 00 DAY Day of the XJ500 clock 1 ÷ 31 s 0 MONTH Month of the XJ500 clock 1 ÷ 12 0 YEAR Year of the XJ500 clock 1995 ÷ 2080 0 FORMAT EURopean or AMErican EUR / AME EUR DLS start month Daylight saving time starting month 1 ÷ 12 0 DLS start day Daylight saving time starting day 1 ÷ 31 0

Page 44: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.44 / 47

DLS end month Daylight saving time end month 1 ÷ 12 0 DLS end day Daylight saving time end day 1 ÷ 31

USER MENU – SETUP – XJ UNITA – ACQUISITION

CLEAR ARCHIVE WARNING! It deletes the whole recording.

AUTO START Restart mode after a black-out Off - Acq -Rec OFF STORE INTERVAL Recording interval 1 ÷ 255 15 MODE Type of recording CONT / INT INT SAVE TEMP Enable the data recording YES ÷ NO YES SAVE ALARM Enable the alarm recording YES ÷ NO YES SAVE DEV STAT Enable the status recording YES ÷ NO YES

USER MENU – SETUP - XJ UNIT – SPECIAL FUNCTION TIMER 1 Timer to control the AUX 0:24, On:Off OFF TIMER 2 Timer to control the AUX 0:24, On:Off OFF

DEVICE TIMER Timer to control some instrument functions

0:24, Off, Def, Stand-by ON, Stand-by OFF, DEV ON, DEV

OFF, LIGHT ON, LIGHT OFF, ES

ON, ES OFF, AUX ON, AUX OFF

OFF

AUX1 WORKING MODE Type of function followed during the week

Off, Manual, Timer1 , Timer2 OFF

AUX2 WORKING MODE Type of function followed during the week

Off, Manual, Timer1 , Timer2 OFF

USER MENU – DEVICES – SINGLE – CONFIGURE INDEX Progressive number of the instrument on the total found NAME Name assigned 16 character No name CAT Category 16 names ENABLE Enable the serial control YES ÷ NO YES SINC CLOCK Clock synchronism with XJ YES ÷ NO NO SAVE ALARM To log the alarm condition YES ÷ NO SAVE STATUS To log the relay status YES ÷ NO AUTO PRINT To be in the automatic prints YES ÷ NO UDM Unit of measurement SAVE INPUT To log the probe values YES ÷ NO

USER MENU – SETUP – DEVICES – ALARM

ALARM CONTROL To enable the whole alarm management

TIME ESC PO To disable the alarm after the power on of the unit 0 255 min 60

TIME CONFIRM Interval time to prints or send the fax 1 255 min 15

PRINT ENABLE Alarm type No = alarm prints disabled; Int = alarm prints each TIME

No, Int, Cont NO

Page 45: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.45 / 47

CONFIRM delay time; Cont = print immediately.

TX FAX ENABLE Enable the fax transmission to send alarm messages YES ÷ NO NO

USER MENU – SETUP - PRINT PRINT ENABLE To enable the printer YES ÷ NO NO

PRINT MODE To select the automatic prints (man = only manual) Man, Day, Wek MAN

DAY Day of printing (only if Wek) SUN ÷ SAT TUE HOURS Hour of printing 00 ÷ 24 12 MINUTES Minute of printing 00 ÷ 59 0

FILTERS Select the data to print ALL, DATA, ALARM ALARM

PRINTER To select the printer type EPSON, IBM ,HP HP COLUMNS 80 normal, 132 compress 40, 80 , 132 80 LINES Lines of the paper modules 62 ÷ 68 64 HACCP INT Interval between two HACCP print OFF, 1 ÷ 24h OFF HACCP START HACCP printing start time 00:00 ÷ 23:50 00:00

USER MENU – SETUP - FAX / MODEM – PHONE BOOK

INDEX Progressive list number N.B. NOT USE THE 16TH INDEX 1 ÷ 16

SERVICE to choose whether the number is a FAX or SMS one FAX ÷ SMS SMS

AREA CODE Number of the country / city 12 Chars 0 TEL Telephone number 12 Chars 0 RIF Reference name 14 Chars SUN..SAT To enable the day YES ÷ NO NO

USER MENU –SETUP – PASSWORD - MODIFY ADMIN Unlimited password 4 Chars NULL UTENTE 1..4 Password with access limits 4 Chars NULL

REMOTE CONNECT. Password to enable the remote connection 4 Chars NULL

Page 46: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.46 / 47

15 TECHNICAL DATA INPUTS

Power supply: 24 Vac o 110/230 Vac ±10%. Serial line RS485 : three screw terminals (gnd), (-) and (+). Length of the RS 485 line: 1000 Mt.

OUTPUT Printer: 25 pins Centronic connector. RS232: 9 pins connector. Alarm relay: 8A/230 Vac SPST. Auxiliary relay 1 and 2 : 8A/230 Vac SPST. Reset modem: 8A/230 Vac SPST.

MONITORING Maximum number of instruments: 225. Type of recording: FIFO (first in first out). Internal clock Back-up : 12 hours. Type of memory: Flash. Sampling time (S.T.): programmable from 1 up to 255 minutes. Example of programming with 1 year of recording data: - standard 2Mbyte of memory: 12 instruments with S.T. = 15 min. - expanded memory 8 Mbytes: 60 instruments with S.C. = 15 min.

ENVIRONMENTAL CONDITION Operating temperature: 0÷40°C / 32÷122°F. Storage temperature: -20÷70°C / 4÷158°F. Relative humidity: Lower than 90% (not condensing).

MECHANICAL DATA Housing : anti-shock polystyrene. Dimensions: 165x235x80mm. Mounting: desktop or wall mounting with bracket. Keyboard: polycarbonate with 9 keys.

USER INTERFACE LCD display: 4 lines with 20 characters each line. Colour: green back lighted. Back light safety: off after 1 minute. Legible area: 70 x 22 mm. Character height: 5 mm.

16 LIMITED LIFE TIME WARRANTY I. LIMITED ONE YEAR WARRANTY ON THE XJ500 Dixell warrants the XJ500 unit to be free from defects in material or workmanship. If within one year of the date of manufacture the XJ500 malfunctions or fails as a result of a defect in material or workmanship, DIXELL will repair or, as its option, replace it. II. EXCLUSIONS and LIMITATIONS 1) the XJ500 is not properly installed; 2) any failure or malfunction is a result from improper or negligent operation, accident, misuse, unauthorised alteration or

improper repair or maintenance; the warranty does not cover shipping charges and delivery expenses.

Page 47: 1592010001 XJ500 v3 4 GB r1.1 16 04 2003 · Wait until the progressive address reaches 224. 4. ... Push MENU to return to the previous menu. 8. Select START PRINT with n and push

Dixell Instruction manual rel.1.1 – 16/04/03 - cod. 1592010001

1592010001 XJ500 v3_4 GB r1.1 16_04_2003.doc XJ500 Pag.47 / 47

17 IF SOMETHING DOES NOT RUN..

PROBLEM REASON POSSIBLE SOLUTION The display shows:

INSTRUMENT ALARM One or more instruments are in alarm condition. Push ENTER and verify the event alarm list.

The display shows: CHECK ALARM

There was an alarm condition not checked (eg an alarm during the night). Push ENTER key and verify the event alarm list.

The display shows: CLOCK ALARM.

The power supply has failed for more than 12 hours. Set the clock parameter again.

The search does not find the instrument or the instruments.

The wires of the serial line are disconnected or exchanged. The Adr instrument parameter, is not properly programmed.

1. Check the continuity of the serial RS485 line. 2. Check the Adr parameter programming. 3. The instrument belongs to a release not compatible (manufactured before November 1998).

The display shows: PRINTER ALARM.

The paper is out.. The cable of the printer is not properly connected. The printer is not present.

1. Load the paper. 2. Be sure that the printer is on and the cable is OK. 3. If you do not connect the printer be sure that the parameter SETUP–ALARM –PRINTALM = NO.

The unit does not print the alarm report.

The parameter SETUP – ALARM – PRINTALM = NO. The TIME CONFIRM parameter is too big.

Check these parameters.

The manual prints does not correspond to the required information.

The printing FILTERS are not enabled. The parameters SAVE INPUT, SAVE ALM and SAVE STATUS are not enabled.

Check these parameters.

FAX ALARM. The fax communication was not completed. The fax is not connected. The fax is not present.

1. If possible verify that the phone line to the remote fax is OK and not out of order. 2. Check the modem connection. 3. If you do not use the fax device, be sure that the parameter SETUP – ALARM – FAX ENABLE = NO.

You can not silence the buzzer.

The password is incorrect. There is an alarm repetition.

1. Be sure of the password value. 2. Check the alarm event list to .

The keyboard is locked. There is a REMOTE CONNETION. Wait the end of the connection.

NO DEVICES ON SETUP No instrument detected. Enter the: SETUP – DEVICES – ADD and start the search.

EMPTY SETUP No instrument detected. Enter the: SETUP – DEVICES – ADD and start the search.

ACCESS DENIED Your password is not the right one. Check with the Administrator of the system.

It prints one alarm description for page.

The parameter PRINT ENABLE = CON the printer is a LASER or DeskJet model.

These type of printers automatically execute the form feed. Set the PRINT ENABLE = INT and increase the TIME CONFIRM parameter.

The XJ500 does not answer to the remote call. The parameter RING NUMBER is too high. Set this parameter to lower value.

Dixell s.r.l. Z.I. Via dell’Industria, 27 - 32010 Pieve d’Alpago (BL) ITALY tel. +39 - 0437 - 98 33 - fax +39 - 0437 - 98 93 13 http://www.dixell.com E-mail: [email protected]