Welcome to the DDG Online Meeting • DDG Online Meeting number 24, the fourth and final in

2016; our 6th year of consecutive meetings

• Worldwide users group for dissolution chemists

• Free, on-line, interactive bulletin board

• Provides annual and regional meetings around the world

• Place for practical answers to everyday questions on dissolution

2

Page 3: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

DDG Mission The mission of the Dissolution Discussion Group (DDG) is

to provide an independent forum to freely discuss the practical issues which challenge the pharmaceutical industry and affect the day-to-day task of developing, validating, and performing dissolution tests and related chemical analyses.

3

Page 4: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

www.dissolution.com

4

Page 5: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

Previous DDG Online Meetings: Nov 10, 2016: USP Performance Testing of Ointments, Creams and Gels

Aug 11, 2016: EMEA Draft – Reflection Paper on Dissolution Specifications for Generic Oral Immediate Release Products

May 12, 2016: USP <1092> and its Impact on Dissolution Automation

Feb 25, 2016: Dissolution SOP’s; Do What You Say and Say What You Do

Nov 12, 2015: FDA Draft Guidance; Dissolution Testing and Specifications for BCS Class 1 and 3 Drugs

Aug 13, 2015: Breaking Bad…Dissolution Habits

May 14, 2015: Crosslinking of Gelatin Capsules and Dissolution

Feb 12, 2015: Dissolution Method Development Considerations: Part 6 – Intrinsic Dissolution

Nov 13, 2014: Dissolution Method Development Considerations: Part 5 – Alcohol Induced Dose Dumping

Aug 14, 2014: Dissolution Method Development Considerations: Part 4 – Time Point Estimation

May 15, 2014: Dissolution Method Development Considerations: Part 3 – Challenges for Poorly Soluble Compounds

Feb 06, 2014: Dissolution Method Development Considerations; Part 2 – Media Volume and Sink Conditions for App 1&2

Nov 08, 2013: Dissolution Method Development Considerations; Part 1 – Filtration Requirements

Aug 15, 2013: Operational Checks; Enhanced Mechanical Qualification Requirement - At Time of Use

May 16, 2013: Deaeration of Dissolution Media; What’s All The Hot Air About?.

Feb 14, 2013: Thinking Small - Challenges with Small Volume Dissolution

Nov 15, 2012: Suggestions for the ASTM E2503-07 Dissolution Apparatus Mechanical Qualification; 5-Year Review

Aug 16, 2012: Use of Enzymes for Dissolution Testing of Gelatin Capsules

May 24, 2012: The Role of Dissolution in Quality by Design (QbD)

Feb 16, 2012: The Dissolution Method Transfer

Nov 17, 2011: The Dissolution Laboratory Audit

Aug 18, 2011: Do f1 and f2 Tell the Whole Story?

May 19, 2011: Best Practices for Success with the USP Dissolution Performance Verification Test (PVT)

Mar 15, 2011: Mechanical Calibration: ASTM vs USP; Impact of ICH on Dissolution; QBD and PAT

5

Page 6: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

Today’s Panelists: Bryan Crist – Agilent Technologies, Inc.

Vivian Gray – Dissolution Technologies and VA Gray Consulting

Ken Boda – Agilent Technologies, Inc.

6

Page 7: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

USP Performance testing of Ointments, Creams and Gels USP <1724> Overview

Theory of Topical Performance Testing

Diffusion Cells: Vertical Diffusion Cell

Immersion Cell

Flow Thru Cell

Membranes

Troubleshooting

References

7

Page 8: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

New USP Chapters – In Process Revision Tests for Semisolid dosage forms

First developed as <725> Topical and Transdermal Drug Products – Product Performance Tests published in PF 35(3) [May-Jun 2009}

Transferred into three USP Chapters:

<3> Topical and Transdermal Drug Products – Quality Tests

<724> Drug Release

<1724> Semisolid Drug Products – Performance Tests

Page 9: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

USP <3> Topical and Transdermal Drug Products – Quality Tests USP PF 36(6) Nov-Dec 2010, In-Process Revision

official USP 35 2012 Contains Product Quality Tests in accordance with ICH Q6A –

Specifications: Test Procedures and Acceptance Criteria for New Drug Substances and new Drug Products: Chemical Substances

Universal Tests: Description, ID, Assay and Impurities

Specific Tests: Physicochemical Properties, Uniformity, Water Content, Microbial Limits, Antimicrobial and Antioxidant Preservative, Sterility, pH, Particle Size Opthalmic Dosage Forms

Topically Applied Semisolid Drug Products

Transdermal Delivery Systems

Page 10: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

USP <724> Drug Release Existing Chapter for Drug Release.

Due to ICH, USP Apparatus 3 – Reciprocating Cylinder and Apparatus 4 – Flow Thru Cell were moved to <711> Dissolution

Contains Acceptance Tables and : Apparatus 5 - Paddle Over Disk

Apparatus 6 - Rotating Cylinder

Apparatus 7 - Reciprocating Holder: Disk Sample Holder

Angled Disk

Cylinder Holder

Acrylic Rod

Spring Holder

USP <1724> Semisolid Drug Products – Performance Tests, US Pharmacopeial Convention, USP 39, 2016

Page 11: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

USP <1724> Semisolid Drug Products - Performance Tests Official USP 36, April 2013

Performance tests for topical dosage forms including creams, ointments, gels and lotions through various diffusion cell apparatus.

In Vitro Performance Tests: Vertical Diffusion Cell – Franz Cell

Immersion Cell – Enhancer Cell

Flow Through Cell – Apparatus 4

Page 12: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

USP <1724> Semisolid Drug Products - Performance Tests

12

Page 13: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

THEORY OF THE DIFFUSION CELL

13

Page 14: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

Theory of the Diffusion Cell

The diffusion cell is a reliable and reproducible means of measuring drug release from semisolid dosage forms. A thick layer of the semisolid product under evaluation is placed in contact with a medium in a reservoir, and the latter acts as a receptor when the drug substance diffuses through the formulation, across the membrane, and into the reservoir. Diffusion occurs across an inert, highly permeable support membrane. The membrane is intended to keep the product and the receptor medium separate and distinct. Membranes should offer the least possible diffusional resistance and should not be rate controlling. Samples are withdrawn from the receptor chamber, typically at 1-h intervals over a 4–6 h period.

14

USP <1724> Semisolid Drug Products – Performance Tests, US Pharmacopeial Convention, USP 39, 2016

Page 15: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

Vertical Diffusion Cell Typically, amounts of the semisolid sample NLT 200 mg are used. Diffusive communication between the semisolid sample and the reservoir takes place through the support membrane. The membrane is intended to keep the drug product sample and receptor medium separate and distinct. A heating jacket or a suitable device should be used to maintain the temperature within the cell. The release rate experiment is carried out at 32 ± 1°, except in the case of vaginal drug products for which the temperature should be 37 ± 1°. Usually a set of 6 cell assemblies are operated together at one time in a single run.

15

USP <1724> Semisolid Drug Products – Performance Tests, US Pharmacopeial Convention, USP 39, 2016

Page 16: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

Vertical Diffusion Cell Sampling is generally performed over a 4–6 h time period, and the volume withdrawn is replaced with stock receptor medium. To achieve sink condition, the receptor medium must have a high capacity to dissolve the drug, and the drug concentration in the receptor medium at the end of the test ideally should be as low as possible. For each cell, the amount of drug released (mg/cm2) at each sampling time (t1, t2, etc.) is determined, and the cumulative amount released plotted versus time. The slope of the resulting line is a measure of the rate of drug release. The test is often conducted with a group of 6 or 12 cells per test run. The average of 6 slopes for each test and reference product is a measure of the drug release rate from the dosage form.

16

USP <1724> Semisolid Drug Products – Performance Tests, US Pharmacopeial Convention, USP 39, 2016

Page 17: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

Vertical Diffusion Cell

17

USP <1724> Semisolid Drug Products – Performance Tests, US Pharmacopeial Convention, USP 39, 2016

Page 18: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

Vertical Diffusion Cell

USP <1724> Semisolid Drug Products – Performance Tests, US Pharmacopeial Convention, USP 39, 2016

Page 19: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

Immersion Cell – Enhancer Cell

The immersion cell can be used with USP Apparatus 2 (see general chapter Dissolution <711>) with vessel volumes that vary from 100 mL up to 4 L, but the 150- or 200-mL vessels are the most commonly used.

A flat-bottom variation of the 150- or 200-mL vessel can be used to avoid the issue of dead space under the cell when it is used in a round-bottom vessel.

If analysts are going to use a 150- or 200-mL vessel with USP Apparatus 2, then the appropriate modifications must be made, including holders for the small-volume vessels and replacement of the standard paddle with the appropriate paddle.

It also may require repositioning of any automated sampling device and/or manifold.

The water bath or vessel heater should be set to have the medium temperature at 32.0 ± 0.5° or 37.0 ± 0.5°.

Before loading the cells and placing the medium in the vessel, set the paddle height, which is 1.0 ± 0.2 cm above the surface of the membrane.

19

USP <1724> Semisolid Drug Products – Performance Tests, US Pharmacopeial Convention, USP 39, 2016

Page 20: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

Immersion Cell – Enhancer Cell All other operational parameters, such as level, vibration, wobble, etc., should be set at the same conditions defined for USP Apparatus 2.

The small-volume condition is qualified by first using the standard Apparatus 2 setup and Performance Verification Test, Apparatus 1 and 2 (see Dissolution <711>).

Cut the membrane to an appropriate size. If necessary, soak the membrane in the receptor medium for at least 30 min before loading. If the membrane is thick, a longer soaking time period may be necessary.

Prepare the immersion cell components as specified by the device manufacturer.

Fill the reservoir dosage area with the sample under test. Ensure that the reservoir is filled to the top in order to minimize the possibility of air bubble formation between the surface of the sample and the membrane.

A uniform surface can be obtained with the aid of a spatula.

The typical quantity of sample is between 300 mg and 2 g, depending on the type of immersion cell used.

An excess of sample is needed to obtain a steady-state drug release rate.

20

USP <1724> Semisolid Drug Products – Performance Tests, US Pharmacopeial Convention, USP 39, 2016

Page 21: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

Immersion Cell – Enhancer Cell

Using forceps or tweezers, remove the membrane from the soaking medium and place it over the top of the sample compartment. Ensure that the membrane is free of wrinkles. Assemble the immersion cell components as specified by the device manufacturer. Carefully place the completed assembly into the bottom of the dissolution vessel with the membrane facing up. The appropriate preheated medium may be preloaded in the vessel or can be added after immersion of the immersion cell to start the test. Samples from at least 5 time points should be obtained in the steady-state (linear) portion of the drug release profile. The data points are cumulative and expressed as concentration per surface area, typically per cm2, as a function of the square root of time. Sampling is generally performed over a 4–6 h time period. The slope of the line is the in vitro release rate of drug from the product. At the end of the test period, dismantle the cell and examine the contents for anything unusual that could explain any anomalous data (e.g., leaks, bubbles, etc.).

21

USP <1724> Semisolid Drug Products – Performance Tests, US Pharmacopeial Convention, USP 39, 2016

Page 22: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

Immersion Cell – Enhancer Cell

22

Page 23: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

Immersion Cell – Enhancer Cell

23

USP <1724> Semisolid Drug Products – Performance Tests, US Pharmacopeial Convention, USP 39, 2016

Page 24: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

Flow-Through Cell The membrane, which may be soaked in the receptor medium beforehand, is loaded in the membrane ring using the provided tool. The membrane should be large enough to overlap the top edge of the reservoir body with a diameter of 18 mm.

The sample is loaded into the reservoir. The other side of the tool can be used to hold the reservoir while loading the sample.

If necessary, the excess of sample can be removed using a spatula. Screw the membrane ring onto the sample reservoir. Ensure that the membrane is free of wrinkles while screwing.

Remove the semisolid sample adapter from the tool, and slide it into the cylindrical part of the 22.6-mm cell with the membrane facing downward.

Vertical positioning within the cell can be adjusted using the tablet holder scoring, if desired (see Figure 9).

If the lower position is chosen, release can be higher due to the proximity to the flow inlet. The prepared cell is inserted in a heating jacket.

24

USP <1724> Semisolid Drug Products – Performance Tests, US Pharmacopeial Convention, USP 39, 2016

Page 25: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

Flow-Through Cell The defined volume of release medium is introduced in the reservoir. Unless otherwise specified, the medium should be deaerated in order to minimize the risk of air bubbles. A deaeration procedure is described in Dissolution <711>, but other validated deaeration techniques can be used.

The reservoir can be adapted to the volume needed in order to achieve sink conditions and to ensure precision of the analytical method. Typical volumes range from 50 to 1000 mL, but values above and below this range also can be used as the formulation demands.

When the pump is switched on, the medium will be pumped through the cell. This represents the time zero of the test.

Typical flow rates are 16 mL/min and 24 mL/min, but flow rate is a method-development parameter and must be optimized accordingly. The flow passing through the cell ensures both agitation and renewal of the receptor medium at the interface with the membrane.

Sampling can be performed either manually or automatically directly from the medium reservoir, thus ensuring no interference with the flow cell and its contents. An automated fraction collector may be appropriate for release periods longer than 6 h.

After quantification, plot the amount of drug release per surface area versus the square root of time, with the slope of the line representing the in vitro release rate.

25

USP <1724> Semisolid Drug Products – Performance Tests, US Pharmacopeial Convention, USP 39, 2016

Page 26: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

Flow-Through Cell

26

USP <1724> Semisolid Drug Products – Performance Tests, US Pharmacopeial Convention, USP 39, 2016

Page 27: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

Troubleshooting Many of these issues are discussed in the USP <1724> Semisolid Drug Products – Performance Testing Chapter:

Deaeration

Sample Preparation in Various Cells

Starting the Test

Variable Test Results

Membrane Selection

Calculation

27

Page 28: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

Comparison…

Page 29: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

SOURCES FOR MEMBRANES

VanKel Technology Group

Page 30: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

Ideal Membrane Characteristics Minimize resistance to drug transport

High porosity

Minimal thickness

No drug binding

Page 31: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

Examples of Membranes Used in In Vitro Tests Regenerated Cellulose (Dialysis membrane)

Silastic membrane - polydimethyl siloxane

Fluoropore FHLP hydrophobic membrane

Ethylene vinyl acetate (Cotrans 9702)

Skin (human, rat, rabbit, porcine, mice)

Cuprophan

Carbosil

Polysulfone

Page 32: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

3-M Drug Delivery Systems: http://www.3m.com/3M/en_US/drug-delivery-systems-us/

Contact Us: 3M Drug Delivery Systems

Contact us today to discuss how we can make your drug delivery needs a reality.

Call Us

US: (1) 800-643-8086 UK: (44) 1509 613626 Japan: (81) 3 6409 3553 China: (86) 21 62753535 Argentina: (54) 11 4469 8259

We will respond to your email request in 24 - 48 business hours.

32

Page 33: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

Porex Corporation http://www.porex.com/products/medical-devices/drug-delivery-pharma/

AMERICAS: Porex Corporation 500 Bohannon Road Fairburn, GA 30213 USA Tel: +1 770 964 1421 Fax: +1 770 969 0954 E-mail: [email protected] EMEA: Porex Technologies GmbH Strangenhäuschen 30 52070 Aachen Germany Tel: +49 241 9105250 Fax: +49 241 91052516 E-mail: [email protected] ASIA PACIFIC: Porex Technologies Sdn Bhd Lot PT 74, Jln Hulu Tinggi 26/6, Sek. 26, Hicom Industrial Park, 40400 Shah Alam, Selangor, Malaysia Tel: +603 51913308 Fax: +603 51923308 E-mail: [email protected]

33

Page 34: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

References: USP <1724> Performance Testing of Ointments, Creams and Gels; USP 39, Rockville, MD, USA

Development in release testing of topical dosage forms: Use of the Enhancer Cell; Pankaj Rl Rege, Vinod D. Vilvalam, Charles C. Collins. Journal of Pharmaceutical and Biomedical Analysis

Measurement of Drug Release from Topical Gels Using Two Types of Apparatus; Hani M. Fares and Joel L. Zats. Pharmaceutical Technology

Comparison of Transdermal Diffusion of Hydrocortisone using the Traditional “Franz Cell” Apparatus and the “Enhancer Cell”; Pradeepkumar Sanghvi and Charles C. Collins

34

Page 35: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

GENERAL DISCUSSION…

Page 36: 10 November, 2016 Moderated by Bryan Crist Agilent ... · 10 November, 2016 Moderated by Bryan Crist Agilent Technologies 1 . ... practical issues which challenge the pharmaceutical

Mark your calendars for the next Online DDG meeting:

Thursday, 09 February, 2017. Topic to be announced.

We look for your suggestions for current hot topics in the field of dissolution such as drug release from novel dosage forms, FDA Guidance, Apparatus Qualification, etc. We do want to hear from you so please continue to suggest topics on the DDG site.

Meetings are held quarterly on Tuesdays from 10:30-11:30 Eastern US Time. Please consult the DDG site for recorded sessions, future topics and updated information.