LMG500 e 12.07 1 to 8 Channel Precision Power Meter LMG500 • Accuracy 0.03% • Bandwidth 10MHz (DC, 0.05Hz to 10MHz) • 3MSamples/s • Sampling absolutly gapless with evaluation of all sampling values, by this capturing of all inrush currents and signal changes • Harmonics and interharmonics up to 50kHz/1.5MHz • Flicker, interactions between network and load To improve your Motors, Transformers, Frequency Inverters, Power Electronics, Power Supplies, Lightings, Automotives in Efficiency, Reliability, Electro Magnetic Compatibilty and Life-Cycle Costs

Transcript of 1 to 8 Channel Precision Power Meter LMG500 - Beving...Precision Power Meters (German:...

  • LMG5

    00 e

    12.

    07

    1 to 8 Channel Precision Power Meter LMG500

    • Accuracy0.03%

    • Bandwidth10MHz(DC,0.05Hzto10MHz)

    • 3MSamples/s

    • Samplingabsolutlygaplesswithevaluationofallsamplingvalues,

    bythiscapturingofallinrushcurrentsandsignalchanges

    • Harmonicsandinterharmonicsupto50kHz/1.5MHz

    • Flicker,interactionsbetweennetworkandload

    ToimproveyourMotors,Transformers,

    FrequencyInverters,PowerElectronics,

    PowerSupplies,Lightings,Automotives

    inEfficiency,Reliability,

    ElectroMagneticCompatibiltyandLife-CycleCosts

  • • Verylowcapacityofmeasurementinputsagainstearth

  • Graphical display for wave form Line plot (trend display)

    Phas or diagram Representation of harmonics as bar graph

    Clearrepresentationofmeasuringprocess

    • Statusbarpermanentlydisplayedinallmenues• Measurementdisplayforoneorfourpowerchannels, alternativelywithsixor20values,40valuesormoretoscroll

    • Graphicaldisplayfor waveform,lineplot (trenddisplay), phasordiagramand bargraphfor harmonicanalysis

    Devicesettingseasilyandintuitivelyovermenues

    • Globalsettings• Twoindependentfiltersetstoprocesssynchronisationandmeasurementsignal• Manualorautomaticsettingofmeasurementranges

    Measurement display with six values Measurement display with 20 values

    Independent setting of synchronisation and measurement path

    Global settings, for example the star-delta conversion

    Selection of input sockets, scaling and measurement range

    Status bar to overview active measurements

  • DC

    AC

    DC

    AC

    M1

    M2Accu

    1Ø, 2Wire1 Watt Meter

    3Ø, 3Wire3 Watt Meter

    Transmission

    Combustion engineInv 1

    Inv 21Ø, 2Wire1 Watt Meter

    3Ø, 3Wire3 Watt Meter

    3 Ph

    ases

    3 Ph

    ases

    1 Ph

    ase

    1 Ph

    ase

    ChannelNo. Ch.1 Ch.2 Ch.3 Ch.4 Ch.5 Ch.6 Ch.7 Ch.8

    Group formation A B C D

    Possible wiring in the groups AtoD

    4Ø 4Wire 4Ø 4Wire4Ø 5Wire 4Ø 5Wire

    1Ø 2W 1Ø 2W 1Ø 2W 1Ø 2W 1Ø 2W 1Ø 2W 1Ø 2W 1Ø 2W3Ø 3Wire

    1Ø 2W3Ø 3W

    1Ø 2W3Ø 4Wire 3Ø 4W4Ø 4Wire 4Ø 4W

    3Ø 3W (Aron)/2Ø 3W 3Ø 3W (Aron)/2Ø 3W 3Ø 3W (Aron)/2Ø 3W 3Ø 3W (Aron)/2Ø 3W3Ø 3W (Aron)/2Ø 3W 1Ø 2W 1Ø 2W 3Ø 3W (Aron)/2Ø 3W 1Ø 2W 1Ø 2W

    Eight power measurement chan-nels, each of them sampled absolutely synchronously with 3MSamples/s, can be provided:

    •Eitherbyacoupled2nd device •OrbytheLMG500withits compact8channelenclosure.

    The current and the voltage paths of the power measure-ment channels are all isolated against each other and against

    earth. This enables a free structuring of measurement on various power applications. The channels are arranged within up to 4 groups (see listed table): channels 1 to 4 (device 1) into groupAandBandchannels5to8(device2)intogroupCandD.Eachgroupissynchro-nised by a group specific signal. The synchronisation with an externalsignalorby„line”isalso possible. ThesettingsofgroupAandBfor certain wirings are indepen-dent to the settings of group CandD.

    1.HybridAutomotiveDrives

    LMG500–Compactwith8Channels

    Optimisationoftheenergyma-nagement of hybrid automotive drives through analysing the power flow in various operation modes and conditions:

    1. Automotive drives through combustion engine with or without booster of the inverter fed 3-phase electrical machines M1 and M2.

    2. Energetic feedback of the braking energy into the battery. 3.Charging/rechargingofthebattery with combustion engine. Eight power measurement

    channels and the process signal interface for torque and revolutionacquireexactlysyn-chronously all data to precisely define efficiency.

    Flexibleuseofthepowermeasurementchannels

  • DC

    3f=50Hz

    AI3

    VU1

    AI2

    AI1

    VU2 VU3

    DC

    3f=50Hz

    AI7

    VU5

    AI6

    AI5

    VU6 VU7

    AI4 A

    I8

    VU4 VU8L2

    L3

    L1

    N

    2

    3

    1

    n

    I4-I8

    G~

    AC

    ACDC

    DC

    Pmech

    3Ø, 3 Wire3 Watt Meter

    3Ø, 3 Wire3 Watt Meter

    PmainsProtor

    Pstator

    1Ø, 2 Wire1 Watt Meter

    1Ø, 2 Wire1 Watt Meter

    Supply

    DC-Link

    G~

    AC

    ACDC

    DC

    Pmech

    3Ø, 3 Wire2 Watt Meter

    3Ø, 3 Wire2 Watt Meter

    3Ø, 3 Wire2 Watt Meter

    3Ø, 3 Wire2 Watt Meter

    PmainsProtor

    Pstator

    U1 U5

    U2

    U3

    U4

    u2

    v2

    v1w2

    w1 U6

    U7

    U8

    u3

    v3

    w3

    u1

    Mains Rectifier

    2.UninterruptiblePowerSupply withDC-Link

    Three channels per stator- and per rotor power, one channel per DC-link and per supply

    Determination of efficiency at different operation points and conditions

    LMG500–Compactwith8Channels

    3.DoubleFedAsynchronousMachines

    Stator power, rotor power, mains power, mains specific inverter power in each case with two watt meters in Aron circuit

    4.TransformerswithMultipleWindings

    Eight channel measurement at 12pulse rectifier transformer

    Wind generators without gearingbutwithanextendedrange for revolution speed. Withfixedfrequencyatstatorside and by setting the rotor frequency the double fed asyn-chronous machine can work as a generator with high efficiency at different wind speeds.

    A three winding transformer with two by 30° electrically shif-tedoutputscoilsfeedstwo6pulsrectifiers.Therebytheprimary winding suppresses harmonics, e. g. the 5th, 7th, 17th and 19th. The power measurement channels are configured as two groups with channels 1 and 5 in parallel. Therewith all measurement channels have the same phase reference and special rectifier transformers with (n.30°) deviating phase angles can be measured exactly.

  • du/dt Filter 1

    du/dt Filter 2

    du/dt Filter 3

    AC

    AC~

    I1

    I2

    I3

    U3

    U2

    U1

    MU4

    du/dt Filter 1

    du/dt Filter 2

    du/dt Filter 3

    I5

    I6

    I7U7U6U5

    AC

    AC~

    I1

    I2

    I3U3U2U1

    M

    BandwidthDC-10MHz•DelayTime<3ns•CapacityagainstEarth<30pF

    Power loss through differentiation measurement before and behind the filter

    Power loss through measurement across the filter

    L-L voltage U4 before the filter, voltage U1 across the filter and filter input current I1

    Menutocompensatethedelaytimeofexternalsensors

    Currents>30Aaremeasuredbymeansofexternalsensors. Wideband(>100kHz)currenttransducersformorethan100A,e.g.ZESZIMMERtypePSU,areused.Theerrorcausedthroughthe group delay of the current transducer can be corrected by assistance of the delay time menu inserting the necessary time adjustment. An outstanding tool with an easy to use menu.

    To increase efficiency of modern PWM frequency inverters fast swit-ching semiconductors are used for minimising the switching losses in theoutputstage.However,theextremelysteepvoltageedgescausecapacitive currents that stress bearings and isolation of the motors – this leads to an early break down. Motor filters (e. g. du/dt-filters) attenuate the voltage edges in rise and fall, but generate power losses by the filters’ transient oscillations (typ.>100kHz). ThehighbandwidthandtheverysmallgroupdelaytimeofU-/I-measuringchannels,

  • ˜I

    U

    ThepowermeasuredwiththeexcitingcurrentIandtheinducedvoltageUatthesensorwinding(coremagnetisationvoltage)directlyyieldsthecorelosseswithoutcopperlosses.WiththerectifiedvalueofthesensorvoltageU–ameasureforthevoltage-timeareaandtherewiththeinducedflux–,theexcitingcurrentIandthegeometriccoredata,thecharacteristiccurvese.g.P(Bpk),Bpk(Hpk) can be generated. With the high-performance formula editor the respective curve points are calculated measuring cycle by measuring cycle.

    DuetothelowearthcapacitanceoftheLMG500measuringinputs(

  • BandwidthDC-10MHz•DelayTime<3ns•CapacityagainstEarth<30pF

    Voltagepeaksbyreflexionoccuronlongconnection cables between frequency inverter and motor. They attain up to the double of the transmitted voltage pulse and stress isolation in addition. Onaccountofthehighbandwidthofthe LMG500 these voltage peaks are captured minutely.

    5.TransientVoltages withLongConnectingLines

    Voltage at inverter output as well as voltage and current at motor with long connection cables

    Easy to recognize: Double increase of peak value Upp

    The on-board power supplies of modern large scale aircraft are operatingwithfrequencyupto800Hz.BasicstandardslikeEUROCAEED-14DandABD0100.1.8arecreatedtodefinelimitvalues,astospecifythosepowersupplies.Fundamentalfrequencyfrom360Hzto800Hzmustbeassessedbytheirharmonicsinrangesupto150kHz.TherefortheLMG500isbestapplicable. ThespectralanalysisofcurrentsandvoltagesDCto1.5MHzissup-ported by a separate ZES ZIMMER application software . You will get graphical data representation in linear or logarithmic scaling. Thevaluescanalsobeexportedastablese.g.intoMSExcel.

    6.Avionics: MonitoringHarmonicsupto150kHz

    Representation of harmonics with ZES ZIMMER software TERM-L5

    HighBasicAccuracy–HighRangeDynamics

    ByassistanceoftheMaximumPowerPointTracker(MPPT)theoperatingpointofasolargenerator can always be kept in its optimum. While non steady solar radiation currents and power can vary between a few percent and nominal value in short time intervals.The tracking can only be optimised if measuring is consistently and without dropout e.g. that may occur while measuring range is switched. The high nominal accuracy of 0.03% allows a correct measuring of very low currents also in the 32A range.

    SolarTechnology

    Voltage, current and effective power high precise without range switching

    DC

    DC

    DC

    AC

    U

    I

    U

    I

    MPPT

    Solar cell 50/60Hz-Mains

  • U12

    U3

    U2U1

    I31

    I12

    I23U23 U31

    I1L1

    L2

    L3

    I2

    I3

    The high quality basic configuration of the LMG500 at reasonable price allows comfortable working. Already contained are RS232 interface, a printer interface, the powerful formula editor and 3.5”Floppydriveastosavedataanddeviceconfigurations.

    IEEE488interface (Orderno.L50-O1) Interpretation of the complete SCPI,aswellastheLMG500specific command set. The data transfer rate yields up to 1Mbyte/sec.

    USBandEthernetinterface(Orderno.L50-O2) USBforserialaccesstoPC/Notebook,Ethernetlinktonetwork.

    Processsignalinterface,digitalandanaloguein-andoutputs(Orderno.L50-O3) To monitor further process ma-gnitudes like revolution, torque etc. With assistance of the

    formula editor efficiency and other magnitudes can be dedu-ced and be applied as control parameters.

    Flickermeter(Orderno.L50-O4) ComplianttoEN61000-4-15. The evaluation of the voltage fluctuations by currents up to 16AcomplianttoEN61000-3-3,by currents up to 75A compliant toEN61000-3-11.

    Eventtriggering (Orderno.L50-O5) Operatesinthebackgroundofthe„normalmode“andcanbeconditionedformax.fourmeasurement values u, i, p co-

    ming from different measuring channels. At a sampling rate of 3Msamples/s peaks and breaks are identified down to a reso-lutionof330nsec.Displayinthe scope menu with 50% pre-/post-trigger. The sampling va-lues are available via interface. Furtherdescriptionseebelow.

    Startodeltaconversion (Orderno.L50-O6) for 3phase-3wire systems. Furtherdescriptionseebelow.

    Harmonicsupto99th forU,I,P,QundS(Orderno.L50-O8) Current,voltageandpowerareanalysedupto50kHzonfundamentals ranging from

    0.1Hzto1.2kHz.Evaluationofinter-harmonics is possible by dividing the given fundamental to a lower one using it as refe-rence. With use of the sampling values the harmonic analysis upto1.5MHzonaexternalPCis given.

    CEHarmonics(Orderno.L50-O9)Uptothe40th, for currents upto16AincompliancewithEN61000-3-2,forcurrentsfrom16Ato75AincompliancewithEN61000-3-12

    DSPModules(Orderno.L50-O10) Necessarytooperateparticularoptions.

    HighQualityBasicConfiguration

    OptionsandAccessoriesforOperatingExtensions

    Conditioning of event triggering

    Option:EventtriggeringOrderno.L50-O5

    If trigger conditions are set this option operates in the background of the normal measuringmode.Bydetectionofgiventriggerconditionsthescopedisplaywillbe„frozen”(display„finish”inthestatusline).However,thenormalmeasuringproceedswithoutanygaps,evaluatingcompletelyallsamplevalues.Fourlogicalconnectabletriggerevents,whichareselectedviathemenu,canbedefinedtocontroltheU-andImeasuringinputs.Ineachtriggereventyoucanascertain:Valuelarger/smaller,insideof/outsideofawindow,eventtime330ns...10s.Bythisandthesamplingwith3Msam-ples/sparticularswellsandsagswillbedetected.Viatheprinterinterfacethealignedscope picture can be printed, also the 2 million samples representing the event can be transferredviathedatainterface,onrequest.WiththesoftkeybuttonRUNthescopedisplay is switched again to the current measuring mode until a new event is detected.

    Option:Startodeltaconversionfor3phase-3wiresystem Orderno.L50-O6

    3phase-3wire system: measurement of line to line voltages and line currents

    At 3phase-3wire systems only thelinetolinevoltagesU12, U23andU31 and the line cur-rents I1, I2 and I3 are available for measurement. Byassistanceofthestartodelta conversion the line to line voltages can be converted into those not directly acces-sible phase voltages (line to neutral voltages, i. e. the phase voltages of the load as if star-connected) and the assigned active powers can be defined. In analogy the line currents can beconvertedintothe„linked”

    currents (line to line currents of the delta connected load). Outofthecalculated„linked”values other magnitudes are de-

    duced, as well as harmonics can be assessed. Unbalanceofnetandload,aswell as distorted wave forms are

    handled correctly by the star to delta conversion.

    Calculated values (linked values) of the star connected windings (Wiring: 3+1, U∆ I∗ -> U∗ I∗)

    Calculated values (linked values) of the delta connected windings (Wiring: 3+1, U∆ I∗ -> U∆ I∆)

  • VoltagemeasuringrangesU* Nominalvalue/V 3 6 12.5 25 60 130 250 400 600 1000 Maximumtrmsvalue/V 3.6 7.2 14.4 30 66 136 270 560 999 1001 Maximumpeakvalueforfullscale/V 6 12 25 50 100 200 400 800 1600 3200 Inputimpedance >4.5MΩ ||

  • Accuracy ± (% of measuring value + % of measuring range)DC 0.05Hz..45Hz 45Hz..65Hz 65Hz..3kHz 3kHz..15kHz 15kHz..100kHz 100kHz..500kHz 500kHz..1MHz 1MHz .. 3MHz 3MHz .. 10MHz

    Voltage U* 0.02+0.06 0.02+0.03 0.01+0.02 0.02+0.03 0.03+0.06 0.1+0.2 0.5+1.0 0.5+1.0 3+3 f/1MHz*1.2 + f/1MHz*1.2Usensor 0.02+0.06 0.015+0.03 0.01+0.02 0.015+0.03 0.03+0.06 0.2+0.4 0.4+0.8 0.4+0.8 f/1MHz*0.7 + f/1MHz*1.5 f/1MHz*0.7 + f/1MHz*1.5

    Current I* (20mA .. 5A) 0.02+0.06 0.015+0.03 0.01+0.02 0.015+0.03 0.03+0.06 0.2+0.4 0.5+1.0 0.5+1.0 f/1MHz*1 + f/1MHz*2 -I* (10A .. 32A) 0.1+0.2 0.3+0.6 f/100kHz*0.8 + f/100kHz*1.2 - - -I HF 0.03+0.06 0.2+0.4 0.5+1.0 0.5+1.0 f/1MHz*1 + f/1MHz*2 -I sensor 0.03+0.06 0.2+0.4 0.4+0.8 0.4+0.8 f/1MHz*0.7 + f/1MHz*1.5 f/1MHz*0.7 + f/1MHz*1.5

    Power U* / I* (20mA .. 5A) 0.032+0.06 0.028+0.03 0.016+0.02 0.028+0.03 0.048+0.06 0.24+0.3 0.8+1.0 0.8+1.0 f/1MHz*3.2 + f/1MHz*2.5 -

    U* / I* (10A .. 32A) 0.104+0.13 0.32+0.4 f/100kHz*1 + f/100kHz*1.1 - - -

    U* / I HF 0.048+0.06 0.24+0.3 0.8+1.0 0.8+1.0 f/1MHz*3.2 + f/1MHz*2.5 -

    U* / I sensor 0.048+0.06 0.24+0.3 0.72+0.9 0.72+0.9 f/1MHz*3 + f/1MHz*2.3 f/1MHz*1.5 + f/1MHz*1.4

    U sensor / I* (20mA .. 5A) 0.024+0.03 0.024+0.03 0.048+0.06 0.32+0.4 0.72+0.9 0.72+0.9 f/1MHz*1.4 + f/1MHz*1.8 -

    U sensor / I* (10A .. 32A) 0.104+0.13 0.4+0.5 f/100kHz*1 + f/100kHz*1 - - -

    U sensor / I HF 0.048+0.06 0.32+0.4 0.72+0.9 0.72+0.9 f/1MHz*1.4 + f/1MHz*2 -

    U sensor / Isensor 0.048+0.06 0.32+0.4 0.64+0.8 0.64+0.8 f/1MHz*1.12 + f/1MHz*1.5 f/1MHz*1.12 + f/1MHz*1.5

    Measuringaccuracyadditional measurement uncertainty in the ranges 10A to 32A: ±I2trms.30µA/A2 Accuracies based on: 1. sinusoidal voltage and current 4. definition of power range as the product of 2. ambienttemperature23°C currentandvoltagerange,0≤ IλI ≤ 1, (λ=Power factor=P/S) 3. warm up time 1h 5. calibration interval 12 month Othervalues Allothervaluesarederivedfromthecurrent,voltageandactivepowervalues.Accuraciesforderivedvaluesdependonthe functionalrelationship(e.g.S=I*U,∆S/S = ∆I/I + ∆U/U)

    Isolation All current and voltage inputs isolated against each other, against remaining electronic and against earth max.1000V/CATIIIresp.600VCATIV

    Synchronization Themeasurementissynchronizedonthesignalsperiod.Thereisachoicetodeterminetheperiodfrom„line“,„extern“, u(t),i(t)aswellastheirenvelopes,combinedwithsettablefilters.Bythisverystablereadingsareachieved,evenatsignalsof pulse width modulated frequency inverters and amplitude modulated electronic ballasts

    Harmonicanalysis MeasuringofcurrentandvoltagewithevaluationinfullcompliancewithEN61000-3-2/-12, (optionCEHarmL50-O9) measurementaccordingtoEN61000-4-7

    Harmonicanalysis Analysis of current, voltage (incl. phase angle) and power up to 99thharmonics,intotal100harmonicsincludingDCcomponent. (optionHarm100L50-O8) Fundamentalintherangefrom0.1Hzto1.2kHz.Analysisupto10kHz(50kHzwithoutantialiasingfilter). Byintegerdivider(1...128)anewreferencefundamentalcanbecreatedastodetectinterharmonics. ExternallyonPCupto1.5MhzwithZESZIMMERsoftware.

    Flickermeasuring(optionL50-O4) FlickermeteraccordingtoEN61000-4-15withevaluationinfullcompliancewithEN61000-3-3/-11Transients(optionL50-O5) Detectingandrecordingoftransients>330ns

    Scopefunction(standard) Graphical representation of sampled values versus time

    Plotfunction(standard) Time(Trend)diagramofmax.4readings,minimalresolution50ms,respectively10msin50Hzhalf-wave(flicker)mode

    Stardeltaconversion(optionL50-O6) Sums and differences between channels on sample basis

    Computerinterfaces RS232 (standard) and IEEE488.2(optionL50-O1), additional USB2.0TypBand Ethernet10/100Base-TRJ45(option L50-O2)available.Onlyoneinterfacecanbeusedatthesametime Remote control All functions can be remote-controlled, keyboard lock for measuring parameters Outputdata Outputofallreadings,dataformatsBIN/ASCII,SCPIcommandset Transferrate RS232:max.115200Baud,IEEE488.2:max.1MByte/s

    Floppydrive(standard) 3.5’’,1.44MB

    Printerinterface(standard) ParallelPC-Printerinterfacewith25-pinSUB-Dsocket,printingmeasuringvalues,tablesandgraphics tomatrix,inkjetorlaserprinters

    Processingsignalinterface 2x25pinSUB-Dsocketwith: (optionL50-O3) • 8analoginputsforprocessmagnitudes(24Bit,±10V) • 8analogoutputs(14Bit,±10V) • 8digitalinputs • 8digitaloutputs • 2inputforfrequency(0.1Hz...500kHz)androtationdirection • in-andoutputsareisolatedagainstotherelectronics(testvoltage500V)

    OtherdataDimensions/Weight • Benchcase1to4channelsW433mmxH148mmxD506mm/about12kg • Benchcase1to8channelsW433mmxH298mmxD506mm/about23kg • Accessories:bracketsfor19‘‘rack,84PU,3HU,D464mm Protectionclass EN61010(IEC61010,VDE0411),protectionclassI Electromagneticcompatibility EN61326 Protectionsystem IP20inaccordancetoEN60529 Operating/storagetemperature 0...40°C,-20...50°C Climaticclass NormalenvironmentconditionsaccordingtoEN61010 Powersupply 100...240V,50...60Hz,max.150W(4channeldevice),max.300W(8channeldevice)

    LMG500applicationsoftware (Nameofsoftwareisequalwithordernumber,pleaserequestdetaileddatasheets)

    LMGControl Individual configuration of measurement, using all features of the LMG500, spectral analysis, remote of LMG500, storageinMSExcelformatpossible

    PQA-SOFT Softwareespeciallydesignedforpowerqualityanalysis(e.g.EN50160),easyconfiguringofmeasurementinafewsteps

    SYS61K-3-SOFT ControlandevaluationsoftwarefortestsystemsofharmonicsandflickeraccordingtoEN61000-3-2/-3/-11/-12

    TechnicalData

  • L45-Z06 L50-Z29-Hall

    HST12-1

    LMG-MAK3

    NDL5

    ZES ZIMMER Electronic Systems GmbHTabaksmühlenweg 30, D-61440 Oberursel/Ts.Tel. +49 6171 628750 Fax +49 6171 52086

    www.zes.com • [email protected]

    Subject to technical changes, especially to improve the product, at any time without prior notification.

    Adapterfor3-phasemeasurementsOrderno.LMG-MAK3

    • CEE-Plug,5pins,16A,2msupplycord • CEE-Socket,5pins,16A,forEUT • SocketforsupplyingthemeterLMG500/LMG450 • 4mmsafetysockets,measuringaccesstocurrentandvoltage • Safetyacc.IEC61010:300V/CATIII

    AutonomousDataLoggerNDL5 Orderno.NDL5

    • Autonomouslongtimedataloggingonharddisk for LMG500/LMG450/LMG95, configuration with ZES ZIMMER software • CommunicationviaInternet/Ethernet, even when recording • Uninterruptiblepowersupply forNDL5andLMG500integrated • W320mmxH50mmxD307mm

    „PlugN‘Measure“currentsensors

    Orderno.PSU600-L50 Orderno.L45-Z06,-Z07,-Z10,-Z11, -Z16,-Z17,-Z26,-Z28,-Z32,L50-Z29 all to be used with LMG500 by means of adapter L50-Z14 (Please request for special datasheets)

    MeasurementAccesoriesandExtensions

    PSU600-L50 (with Adapter L50-Z14)

    PrecisionhighvoltagedividerOrderno. HST3-1,HST6-1,HST9-1,HST12-1 HST3-2,HST6-2,HST9-2,HST12-2 HST3-3,HST6-3,HST9-3,HST12-3

    Precisionhighvoltagedividerfor3/6/9/12kVupto1MHz -1 for single-earthed high voltage measurements, single phase -2 for earth-free high voltage measurements, single phase -3 for single-earthed high voltage measurements, three phase Accuracy:0.05%(45-65Hz),0.3%(DC-100kHz)

    CalibrationcertificateOrderno.KR-L50

    Calibrationwithcertificate,traceableaccordingtoISO9000

    CalibrationandservicepackageforextendedwarrantyOrderno.L50-KSP

    With the purchase of the calibration and service package the warranty will beextendedeveryyearforfurther12 months according to your wishes and technical applications. The necessary access is the calibration accordingtoISO9000atfirstdeliveryofthe device. After 12 months the device has to be sent back to ZES ZIMMER for a further calibration and if necessary for adjustment. Along with the calibration the appropriate maintenance work is accomplished.Duringthewarranty

    periodandextendedwarrantyperiodallincidental repair work is accomplished free of charge. Repairs of failures through abrasion and faulty handling areexceptedfromthewarranty.Requirementforextendedwarrantyand its continuation is the calibration at first delivery and the annual due in time calibration in continuity. The requiredextensionalwaysneedsourwrittenacknowledgement.Onthiswayawarranty time of 10 years or longer can be achieved.