P 6 S H U T T E R S
By RICK SPENCE
September 6, 2005
![Page 2: 1 P 6 S H U T T E R S By RICK SPENCE September 6, 2005.](https://reader030.fdocuments.net/reader030/viewer/2022032723/56649f515503460f94c74932/html5/thumbnails/2.jpg)
2
INTEGRAL SHUTTER defined:
A device that integrates a white-beam stop, two monochromatic-beam shutters, a safety stop and a collimator into one assembly.
![Page 3: 1 P 6 S H U T T E R S By RICK SPENCE September 6, 2005.](https://reader030.fdocuments.net/reader030/viewer/2022032723/56649f515503460f94c74932/html5/thumbnails/3.jpg)
3
P6 Integral Shutter @ 2-BM-A
![Page 4: 1 P 6 S H U T T E R S By RICK SPENCE September 6, 2005.](https://reader030.fdocuments.net/reader030/viewer/2022032723/56649f515503460f94c74932/html5/thumbnails/4.jpg)
4
P6 Shutters at the APS
P6-20 shutter is the original design. 1-BM-A P6-20 Integral Shutter. 2-BM-A P6-20 Integral Shutter. 12-BM-A P6-20 Integral Shutter
- locked in mono mode. 17-BM-A Originally a P6 shutter
- locked in mono mode.Significantly modified P6 shutter.
![Page 5: 1 P 6 S H U T T E R S By RICK SPENCE September 6, 2005.](https://reader030.fdocuments.net/reader030/viewer/2022032723/56649f515503460f94c74932/html5/thumbnails/5.jpg)
5
FIGURE 1
Typical FOE arrangement shown schematically. Copy from paper.
![Page 6: 1 P 6 S H U T T E R S By RICK SPENCE September 6, 2005.](https://reader030.fdocuments.net/reader030/viewer/2022032723/56649f515503460f94c74932/html5/thumbnails/6.jpg)
6
P6-20 INTEGRAL SHUTTER
White beamPhoton Stop
FirstMono Shutter
SecondMono Shutter
Movable SafetyStop & Collimator
![Page 7: 1 P 6 S H U T T E R S By RICK SPENCE September 6, 2005.](https://reader030.fdocuments.net/reader030/viewer/2022032723/56649f515503460f94c74932/html5/thumbnails/7.jpg)
7
P6 Components/function White Beam Photon Stop – absorbs high power x-rays
in mono mode. (2) Mono-beam Shutters – redundant protection,
absorbs mono beam. Safety Stop – blocks bremsstahlung radiation in mono
mode. Collimator – 18 x 120 mm aperture, passes white or
mono beam depending on mode setting. Kirk-key interlock system - guarantees proper
component operation.
![Page 8: 1 P 6 S H U T T E R S By RICK SPENCE September 6, 2005.](https://reader030.fdocuments.net/reader030/viewer/2022032723/56649f515503460f94c74932/html5/thumbnails/8.jpg)
8
Bremsstrahlung
White beam
Mono-beam
Mono ON
![Page 9: 1 P 6 S H U T T E R S By RICK SPENCE September 6, 2005.](https://reader030.fdocuments.net/reader030/viewer/2022032723/56649f515503460f94c74932/html5/thumbnails/9.jpg)
9
White ON
White beam
![Page 10: 1 P 6 S H U T T E R S By RICK SPENCE September 6, 2005.](https://reader030.fdocuments.net/reader030/viewer/2022032723/56649f515503460f94c74932/html5/thumbnails/10.jpg)
10
Mono OFFMono ON
White OFFWhite ON
![Page 11: 1 P 6 S H U T T E R S By RICK SPENCE September 6, 2005.](https://reader030.fdocuments.net/reader030/viewer/2022032723/56649f515503460f94c74932/html5/thumbnails/11.jpg)
11
Beam Direction
P6-20 INTEGRAL SHUTTER
MATERIALS USED
Tungsten inside stainless steel
housing Tungsten
Glid-cop
![Page 12: 1 P 6 S H U T T E R S By RICK SPENCE September 6, 2005.](https://reader030.fdocuments.net/reader030/viewer/2022032723/56649f515503460f94c74932/html5/thumbnails/12.jpg)
12
P6 Component Specs White Beam Photon Stop
Can withstand 7-GeV, 300 mA operation.
2 Mono-beam Shutters Stopping power ~ 10 Watts
Safety Stop/Collimator Aperture = 18 mm (h) x 120 mm (w)
![Page 13: 1 P 6 S H U T T E R S By RICK SPENCE September 6, 2005.](https://reader030.fdocuments.net/reader030/viewer/2022032723/56649f515503460f94c74932/html5/thumbnails/13.jpg)
13
KIRK-KEY POSITIONS
White BeamMono Beam
![Page 14: 1 P 6 S H U T T E R S By RICK SPENCE September 6, 2005.](https://reader030.fdocuments.net/reader030/viewer/2022032723/56649f515503460f94c74932/html5/thumbnails/14.jpg)
14
P6 & P4 Integral Shutters
![Page 15: 1 P 6 S H U T T E R S By RICK SPENCE September 6, 2005.](https://reader030.fdocuments.net/reader030/viewer/2022032723/56649f515503460f94c74932/html5/thumbnails/15.jpg)
15
P6 vs. P4 SHUTTERS–Similarities:
The P6 shutter is very similar to the P4 shutter.
Both are integral shutters & include the same internal components.
The P6 and P4 integral shutters were the first shutter used for BM/ID beamlines at the APS, at Sectors 1 & 2.
![Page 16: 1 P 6 S H U T T E R S By RICK SPENCE September 6, 2005.](https://reader030.fdocuments.net/reader030/viewer/2022032723/56649f515503460f94c74932/html5/thumbnails/16.jpg)
16
P6 vs. P4 SHUTTERS -Differences:
P6 – Used in BM beamlines only, P4 – ID only. White Beam Photon Stop design
(P6 – block, P4 – hockey stick) Collimator opening size
(P6 -18x120mm, P4 -18x80mm) Design Geometry
![Page 17: 1 P 6 S H U T T E R S By RICK SPENCE September 6, 2005.](https://reader030.fdocuments.net/reader030/viewer/2022032723/56649f515503460f94c74932/html5/thumbnails/17.jpg)
17
Reliability of P6 Shutters Originally designed for 10 years. 1-BM and 2-
BM P6 shutters in operation for 10 years and counting.
Shutter Stats (response time) available thru EPICS for not only PS1/PS2 but for other beamline shutters as well.
To access from EPICS XFD Display: PSS/IOC Stats/pink tab –IOC Stats/ 01-BM Stats
![Page 18: 1 P 6 S H U T T E R S By RICK SPENCE September 6, 2005.](https://reader030.fdocuments.net/reader030/viewer/2022032723/56649f515503460f94c74932/html5/thumbnails/18.jpg)
18
P6 Shutter Stats via EPICS
![Page 19: 1 P 6 S H U T T E R S By RICK SPENCE September 6, 2005.](https://reader030.fdocuments.net/reader030/viewer/2022032723/56649f515503460f94c74932/html5/thumbnails/19.jpg)
19
Expected failures for P6 shutters (& other shutters)
Failure of seals, resulting in restricted operation and/or vacuum leaks.
Failure of air actuators. Failure of bellows, resulting in vacuum
leaks.
![Page 20: 1 P 6 S H U T T E R S By RICK SPENCE September 6, 2005.](https://reader030.fdocuments.net/reader030/viewer/2022032723/56649f515503460f94c74932/html5/thumbnails/20.jpg)
20
ACKNOWLEDGEMENTS
Mohan Ramanathan Deming Shu Ercan Alp Ali Mashayekhi Nadia Leyarovska Ray Monroe Kevin Beyer Paper: Design of integral shutters for beamlines at the APS.
By J.Chang, D.Shu, H.L.Nian, T.M.Kuzay and P.K.Job.Rev. Sci. Instrum. 66 (2), February 1995