1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb...

45
1 International International Committee Committee of the Red Cross of the Red Cross Polypropylene Technology Polypropylene Technology Manufacture of Lower Limb Prostheses in PP Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT TF STATIC ALIGNMENT

Transcript of 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb...

Page 1: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

11

International Committee International Committee of the Red Crossof the Red Cross

Polypropylene TechnologyPolypropylene Technology

Manufacture of Lower Limb Prostheses in PPManufacture of Lower Limb Prostheses in PP

TF STATIC ALIGNMENTTF STATIC ALIGNMENT

Page 2: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

22

Page 3: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

33

Page 4: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

44

SWING PHASE: OSCILLATION OF THE SHIN

VISIBLE INTERNAL OR EXTERNAL ROTATION OF KNEE

MID SWING PHASE

Page 5: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

55

- INCORRECT INITIAL KNEE ROTATION

- INCORRECT SOCKET FITTING

- INCORRECT SOCKET ALIGNMENT

- INSUFFICIENT SUSPENSION OR INCORRECT LOCATION OF THE SILESIAN BELT

PROSTHESIS

SWING PHASE: OSCILLATION OF THE SHIN

VISIBLE INTERNAL OR EXTERNAL ROTATION OF KNEE

MID SWING PHASE

Page 6: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

66

- WEAK AND FLABBY STUMP

PATIENT

SWING PHASE: OSCILLATION OF THE SHIN

VISIBLE INTERNAL OR EXTERNAL ROTATION OF KNEE

MID SWING PHASE

Page 7: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

77 WALKING BASE IS TOO WIDE

SWING PHASE: ABDUCTED GAIT

Page 8: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

88

- PROSTHESIS IS TOO LONG

- INSUFFICIENT SOCKET ABDUCTION - INSUFFICIENT LATERAL FEMORAL SUPPORT - MEDIAL SHELF IS TOO HIGH

- EXCESSIVE SUSPENSION OR INCORRECT LOCATION OF THE SILESIAN BELT

PROSTHESIS

WALKING BASE IS TOO WIDE

SWING PHASE: ABDUCTED GAIT

Page 9: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

99

- SHORT STUMP

- ABDUCTORS CONTRACTURE - PAIN IN PERINEAL AREA

- INSECURE BALANCE

- BAD GAIT HABIT

PATIENT

WALKING BASE IS TOO WIDE

SWING PHASE: ABDUCTED GAIT

Page 10: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

1010

PROSTHESIS CURVES LATERALLY

SWING PHASE: CIRCUMDUCTION

Page 11: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

1111

- PROSTHESIS IS TOO LONG

- EXCESSIVE PLANTAR FLEXION

- INSUFFICIENT SUSPENSION (PISTONING)

- INCORRECT SOCKET FITTING

- EXCESSIVE SUSPENSION OR INCORRECT

LOCATION OF THE SILESIAN BELT

- INSUFFICIENT ABDUCTION OF THE SOCKET

- MEDIAL SHELF IS TOO HIGH

- ANTERIOR DISPLACEMENT OF THE SOCKET

- EXCESSIVE KNEE AXIS FRICTION

- EXCESSIVE STRENGTH OF THE EXTENSION AID

PROSTHESIS

PROSTHESIS CURVES LATERALLY

SWING PHASE: CIRCUMDUCTION

Page 12: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

1212

PROSTHESIS CURVES LATERALLY

SWING PHASE: CIRCUMDUCTION

- ABDUCTORS CONTRACTURE

- INSUFFICIENT HIP FLEXION

- WEAK QUADRICEPS

- FEAR OF TOUCHING THE GROUND

- PAIN IN PERINEAL AREA

- BAD GAIT HABIT

PATIENT

Page 13: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

1313

EXCESSIVE ELEVATION OF PELVIS AND SHOULDERON THE PROSTHETIC SIDE

SWING PHASE: ELEVATION OF THE PELVIS

Page 14: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

1414

- PROTHESIS IS TOO LONG

- EXCESSIVE PLANTAR FLEXION

- ANTERIOR DISPLACEMENT OF THE FOOT

- INSUFFICIENT SUSPENSION (PISTONING)

- INCORRECT SOCKET FITTING

- ANTERIOR DISPLACEMENT OF THE SOCKET

- EXCESSIVE KNEE AXIS FRICTION

- EXCESSIVE STRENGTH OF THE EXTENSION AID

PROSTHESIS

EXCESSIVE ELEVATION OF PELVIS AND SHOULDER

ON THE PROSTHETIC SIDE

SWING PHASE: ELEVATION OF THE PELVIS

Page 15: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

1515

- INSUFFICIENT HIP FLEXION

- WEAK QUADRICEPS MUSCLES

- FEAR OF TOUCHING THE GROUND

- BAD GAIT HABIT

PATIENT

SWING PHASE: ELEVATION OF THE PELVIS

EXCESSIVE ELEVATION OF PELVIS AND SHOULDER

ON THE PROSTHETIC SIDE

Page 16: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

1616

LATERAL INCLINATION OF THE TRUNKON THE PROSTHETIC SIDE

SWING PHASE: LATERAL TRUNK BENDING

Page 17: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

1717

- PROSTHESIS IS TOO SHORT

- EXCESSIVE ABDUCTION OF THE SOCKET

- INSUFFICIENT LATERAL FEMORAL

SUPPORT

- MEDIAL SHELF IS TOO HIGH

PROSTHESIS

LATERAL INCLINATION OF THE TRUNKON THE PROSTHETIC SIDE

SWING PHASE: LATERAL TRUNK BENDING

Page 18: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

1818

- WEAK ABDUCTOR MUSCLES

(TRENDELENBURG SYNDROME) OR

ABDUCTORS CONTRACTURE

- SHORT STUMP

- HYPER SENSITIVE STUMP

- INSUFFICIENT BALANCE

- BAD GAIT HABIT

PATIENT

SWING PHASE: LATERAL TRUNK BENDING

LATERAL INCLINATION OF THE TRUNKON THE PROSTHETIC SIDE

Page 19: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

1919

HEEL CONTACT: OSCILLATION OF THE FOOT

VISIBLE OSCILLATION OF THE FOREFOOT

Page 20: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

2020

HEEL CONTACT: OSCILLATION OF THE FOOT

- HARD HEEL CUSHION

- EXCESSIVE EXTERNAL FOOT ROTATION

- INCORRECT SOCKET ALIGNMENT

PROSTHESIS

VISIBLE OSCILLATION OF THE FOREFOOT

Page 21: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

2121

HEEL CONTACT: OSCILLATION OF THE FOOT

- SHORT STUMP

- WEAK AND FLABBY STUMP

- EXCESSIVE HEEL STRIKE

PATIENT

VISIBLE OSCILLATION OF THE FOREFOOT

Page 22: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

2222

HEEL CONTACT WITH THE GROUND IS TOO SHORT

HEEL CONTACT: EARLY HEEL ROLLOVER

Page 23: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

2323

- ANTERIOR DISPLACEMENT OF THE SOCKET

- INSUFFICIENT FLEXION OF THE SOCKET

- EXCESSIVE DORSAL FLEXION

- HARD HEEL CUSHION

PROSTHESIS

HEEL CONTACT WITH THE GROUND IS TOO SHORT

HEEL CONTACT: EARLY HEEL ROLLOVER

Page 24: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

2424

- WEAK GLUTEAL MUSCLES

PATIENT

HEEL CONTACT: EARLY HEEL ROLLOVER

HEEL CONTACT WITH THE GROUND IS TOO SHORT

Page 25: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

2525

SLAP OF THE FOREFOOT ON THE GROUND

HEEL CONTACT: FOOT SLAP

Page 26: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

2626

- SOFT HEEL CUSHION

- EXCESSIVE PLANTAR FLEXION

PROSTHESIS

SLAP OF THE FOREFOOT ON THE GROUND

HEEL CONTACT: FOOT SLAP

Page 27: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

2727

- WEAK HIP EXTENSORS, NOT PROVIDING SUFFICIENT ANTERIOR KNEE STABILITY

PATIENT

HEEL CONTACT: FOOT SLAP

SLAP OF THE FOREFOOT ON THE GROUND

Page 28: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

2828

DECELERATION PHASE: TERMINAL IMPACT

VISIBLE KNEE SHOCK

Page 29: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

2929

DECELERATION PHASE: TERMINAL IMPACT

- INSUFFICIENT FRICTION OF KNEE AXIS

- EXCESSIVE STRENGTH OF THE EXTENSION AID

- EXTENSION BUFFER IS TOO HARD

PROSTHESIS

VISIBLE KNEE SHOCK

Page 30: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

3030

DECELERATION PHASE: TERMINAL IMPACT

- EXCESSIVE ACCELERATION FORCE

- BAD GAIT HABIT

PATIENT

VISIBLE KNEE SHOCK

Page 31: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

3131

HEEL RAISE PHASE: LUMBAR HYPER LORDOSIS

VISIBLE ANTERIOR LUMBAR BEND

Page 32: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

3232

HEEL RAISE PHASE: LUMBAR HYPER LORDOSIS

- INSUFFICIENT FLEXION OF THE SOCKET

- PROSTHESIS IS TOO LONG

- ISCHIAL SEAT IS TOO HIGH OR ANTERIORLY INCLINED

PROSTHESIS

VISIBLE ANTERIOR LUMBAR BEND

Page 33: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

3333

HEEL RAISE PHASE: LUMBAR HYPER LORDOSIS

- HIP FLEXORS CONTRACTURE

- ISCHIUM PAIN

- POSTURAL PROBLEM

- BAD GAIT HABIT

PATIENT

VISIBLE ANTERIOR LUMBAR BEND

Page 34: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

3434

MID SWING PHASE: VAULTING

RAISE OF THE BODY BY EARLY EXCESSIVE PLANTAR FLEXION ON THE SOUND SIDE

Page 35: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

3535

MID SWING PHASE: VAULTING

- PROSTHESIS IS TOO LONG

- EXCESSIVE PLANTAR FLEXION

- INSUFFICIENT SUSPENSION (PISTONING)

- INCORRECT SOCKET FITTING

- MEDIAL SHELF IS TOO HIGH

- ANTERIOR DISPLACEMENT OF THE SOCKET

- EXCESSIVE FRICTION IN KNEE AXIS

- EXCESSIVE STRENGTH OF THE EXTENSION AID

PROSTHESIS

RAISE OF THE BODY BY EARLY EXCESSIVE PLANTAR FLEXION

ON THE SOUND SIDE

Page 36: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

3636

MID SWING PHASE: VAULTING

- FEAR OF TOUCHING GROUND

- VERY SENSITIVE OR PAINFUL STUMP

- BAD GAIT HABIT

PATIENT

RAISE OF THE BODY BY EARLY EXCESSIVE PLANTAR FLEXION

ON THE SOUND SIDE

Page 37: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

3737

ACCELERATION PHASE: UNEVEN AND RAPID HEEL RAISE

EXCESSIVE AND RAPID ELEVATION OF THE HEEL

Page 38: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

3838

ACCELERATION PHASE: UNEVEN AND RAPID HEEL RAISE

- INSUFICIENT FRICTION OF THE KNEE AXIS

- INSUFFICIENT STRENGTH OF THE EXTENSION AID

- POSTERIOR DISPLACEMENT OF THE SOCKET

PROSTHESIS

EXCESSIVE AND RAPID ELEVATION OF THE HEEL

Page 39: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

3939

ACCELERATION PHASE: UNEVEN AND RAPID HEEL RAISE

- GAIT IS TOO FAST

- EXCESSIVE HIP FLEXION - BAD GAIT HABIT

PATIENT

EXCESSIVE AND RAPID ELEVATION OF THE HEEL

Page 40: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

4040

TRANSVERSAL VIEW: UNEVEN STEP LENGTH

SHORT STEP ON PROSTHETIC SIDE

Page 41: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

4141

TRANSVERSAL VIEW: UNEVEN STEP LENGTH

- ANTERIOR DISPLACEMENT OF THE SOCKET

- EXCESSIVE FLEXION OF THE SOCKET

- PISTONING ACTION

- EXCESSIVE FRICTION OF THE KNEE AXIS

- EXCESSIVE STRENGTH OF THE EXTENSION AID

PROSTHESIS

SHORT STEP ON PROSTHETIC SIDE

Page 42: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

4242

TRANSVERSAL VIEW: UNEVEN STEP LENGTH

- SHORT, WEAK OR PAINFUL STUMP

- GAIT FEAR

- LIMITED HIP FLEXION -EXTENSION

- WEAK QUADRICEPS MUSCLES

PATIENT

SHORT STEP ON PROSTHETIC SIDE

Page 43: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

4343SHORT STEP ON SOUND SIDE

TRANSVERSAL VIEW: UNEVEN STEP LENGTH

Page 44: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

4444

- INSUFFICIENT FLEXION OF THE SOCKET

- EXCESSIVE PLANTAR FLEXION

- INSUFFICIENT FLEXION OF THE KNEE

- EXCESSIVE STRENGTH OF THE EXTENSION AID

PROSTHESIS

SHORT STEP ON SOUND SIDE

TRANSVERSAL VIEW: UNEVEN STEP LENGTH

Page 45: 1 International Committee of the Red Cross Polypropylene Technology Manufacture of Lower Limb Prostheses in PP TF STATIC ALIGNMENT.

4545

- SHORT, WEAK OR PAINFUL STUMP

- HIP EXTENSORS CONTRACTURE

- LIMITED HIP FLEXION - EXTENSION

- INSUFFICIENT STABILITY

- PROBLEMS WITH SOUND LEG

PATIENT

TRANSVERSAL VIEW: UNEVEN STEP LENGTH

SHORT STEP ON SOUND SIDE