1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the...
-
Upload
lindsay-oconnor -
Category
Documents
-
view
217 -
download
0
Transcript of 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the...
![Page 1: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/1.jpg)
1
ECG Signal Make-up
![Page 2: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/2.jpg)
2
Biopotential Requirements High Input Impedance
– Minimum loading of the signal being measured– Minimum 10M
Input circuit must have isolation and protection circuits to protect the patient
Output Impedance must be low compared to the load impedance
Proper bandwidth range. Not too narrow and not too wide
Gain at least 1000 times For bipolar electrodes, use differential amplifier
![Page 3: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/3.jpg)
3
ECG Functional Blocks
1. Protection circuit
2. Lead selector
3. Calibration signal
4. Preamplifier
5. Isolation circuit
6. Driven right leg system
7. Driver amplifier
8. Memory System
9. Microcomputer
10. Recorder - Printer
![Page 4: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/4.jpg)
4
The Electrocardiogram
Outline The ECG Specific Requirements Functional Blocks Frequent Problems Some Solutions
![Page 5: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/5.jpg)
5
Heart Dipole
![Page 6: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/6.jpg)
6
Cardiac Vector
![Page 7: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/7.jpg)
7
Eindhoven’s Triangle
![Page 8: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/8.jpg)
8
Wilson’s Central Terminal Connection
![Page 9: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/9.jpg)
9
Augmented Leads
![Page 10: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/10.jpg)
10
Comparing Augmented leads to Standard Leads
![Page 11: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/11.jpg)
11
Precordial Leads
![Page 12: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/12.jpg)
12
Problems Frequently Encountered Frequency Distortion Saturation or Cutoff Distortion Ground Loops Open Lead Wires Artifacts from Large Electric Devices Interference From Electric Devices Electromagnetic Interference
![Page 13: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/13.jpg)
13
Frequency Distortion
![Page 14: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/14.jpg)
14
Saturation - cutoff Distortion
![Page 15: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/15.jpg)
15
Ground Loops
![Page 16: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/16.jpg)
16
Ground Loops - 2
![Page 17: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/17.jpg)
17
Artifacts from large electric devices
![Page 18: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/18.jpg)
18
(a) 60 Hz Line Interference(b) EMG interference
![Page 19: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/19.jpg)
19
Mechanism of electric-field pickup from Power line
![Page 20: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/20.jpg)
20
Current flows from the power line into the body
![Page 21: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/21.jpg)
21
Magnetic-field Pickup by ECG
![Page 22: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/22.jpg)
22
Voltage protection scheme at input
![Page 23: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/23.jpg)
23
Voltage limiting device
![Page 24: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/24.jpg)
24
Electrostatic Shield - to eliminate electric-field interference
![Page 25: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/25.jpg)
25
Driven-Right-Leg System
![Page 26: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/26.jpg)
26
Equivalent Ct. of a driven right leg system
![Page 27: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/27.jpg)
27
Bio-Signals Charachteristics
![Page 28: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/28.jpg)
28
ECG Amplifier
![Page 29: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/29.jpg)
29
![Page 30: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/30.jpg)
30
![Page 31: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/31.jpg)
31
![Page 32: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/32.jpg)
32
![Page 33: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/33.jpg)
33
![Page 34: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/34.jpg)
34
![Page 35: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/35.jpg)
35
![Page 36: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/36.jpg)
36
![Page 37: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/37.jpg)
37
![Page 38: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/38.jpg)
38
![Page 39: 1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M n Input circuit.](https://reader035.fdocuments.net/reader035/viewer/2022062801/56649e7a5503460f94b7aea1/html5/thumbnails/39.jpg)
39