1 CPAC NeSSI Focus Group May 3, 2004 Focus Group Agenda –Dave Veltkamp – Roll Out of NeSSI...

10
1 CPAC NeSSI Focus Group May 3, 2004 Focus Group Agenda Dave Veltkamp – Roll Out of NeSSI Knowledge Network website (30 min) • discussion John Mosher – Update on DOE activities (10 min) • discussion Rob Dubois – Some de facto preferences for SP76 technology (20 min) • slides and discussion

Transcript of 1 CPAC NeSSI Focus Group May 3, 2004 Focus Group Agenda –Dave Veltkamp – Roll Out of NeSSI...

Page 1: 1 CPAC NeSSI Focus Group May 3, 2004 Focus Group Agenda –Dave Veltkamp – Roll Out of NeSSI Knowledge Network website (30 min) discussion –John Mosher –

1

CPAC NeSSI Focus GroupMay 3, 2004

• Focus Group Agenda– Dave Veltkamp – Roll Out of NeSSI

Knowledge Network website (30 min)• discussion

– John Mosher – Update on DOE activities (10 min)

• discussion

– Rob Dubois – Some de facto preferences for SP76 technology (20 min)

• slides and discussion

Page 2: 1 CPAC NeSSI Focus Group May 3, 2004 Focus Group Agenda –Dave Veltkamp – Roll Out of NeSSI Knowledge Network website (30 min) discussion –John Mosher –

2

Some De Facto Preferences for SP76

TechnologyMay 3, 2004

CPAC, Seattle, WA

Page 3: 1 CPAC NeSSI Focus Group May 3, 2004 Focus Group Agenda –Dave Veltkamp – Roll Out of NeSSI Knowledge Network website (30 min) discussion –John Mosher –

3

de facto spacing & back panel marking

A preferred practice would be to leave spaces between substrate rows and columns in multiples of SP76 blocks. e.g. 1.53”, 3.06”, etc. By this means the back plate can be pegboard drilled and tube to block connectors may be built in standard multiples to span open spaces. Tagging the back panel as illustrated is also helpful to the End User.

Page 4: 1 CPAC NeSSI Focus Group May 3, 2004 Focus Group Agenda –Dave Veltkamp – Roll Out of NeSSI Knowledge Network website (30 min) discussion –John Mosher –

4

Consider substrate flow marking

Here is an example where one manufacturer marks his substrate to indicate the flow paths. This may not be feasible in some cases but is shown as one way to demonstrate flow.

Page 5: 1 CPAC NeSSI Focus Group May 3, 2004 Focus Group Agenda –Dave Veltkamp – Roll Out of NeSSI Knowledge Network website (30 min) discussion –John Mosher –

5

helium leak check channel useful

Some manufacturers provide this channel. We think that this is a good idea.

Page 6: 1 CPAC NeSSI Focus Group May 3, 2004 Focus Group Agenda –Dave Veltkamp – Roll Out of NeSSI Knowledge Network website (30 min) discussion –John Mosher –

6

SM component leak check port

Some manufacturers provide helium check ports on their components. We believe that this is a good idea. Do you plan to do this with your SM components?

Page 7: 1 CPAC NeSSI Focus Group May 3, 2004 Focus Group Agenda –Dave Veltkamp – Roll Out of NeSSI Knowledge Network website (30 min) discussion –John Mosher –

7

We like comprehensive SM component identification

Some surface mount components incorporate extensive indelible tagging. Having material of construction, serial number, manufacturer name and logo, model number, material traceability, pressure rating, directional arrows, inlet/outlet etc. are all very helpful to the End User.

Page 8: 1 CPAC NeSSI Focus Group May 3, 2004 Focus Group Agenda –Dave Veltkamp – Roll Out of NeSSI Knowledge Network website (30 min) discussion –John Mosher –

8

End User tagging & replaceable tags

Consider some way to add End User tagging e.g. AFL-001 (and variable information i.e. filter element size) to the SM component. Needs to be non-permanent. Adhesive tags may be ideal. Wired on tags are not acceptable.

Page 9: 1 CPAC NeSSI Focus Group May 3, 2004 Focus Group Agenda –Dave Veltkamp – Roll Out of NeSSI Knowledge Network website (30 min) discussion –John Mosher –

9

Combi-Valve Interfaces

A Standard Interface (close-coupled) between pneumatic actuator and electrical operator would be desirable.

combi-valve

Page 10: 1 CPAC NeSSI Focus Group May 3, 2004 Focus Group Agenda –Dave Veltkamp – Roll Out of NeSSI Knowledge Network website (30 min) discussion –John Mosher –

10

Further thoughts…• Enclosure Mounting. Use flexible metallic hoses

between substrate and enclosure bulkheads/tube entries– must be low cost/light weight– eliminates the need for any tube bending

• SAE fittings are useful for angle connectors esp. on solenoid/pneumatic valve connections

• low cost flow switch which operates in both vertical and horizontal positions

• require more items on substrate incl. close coupled or on substrate pumps, coolers, coalescers…

• as a de facto requirement use the "center position” hole as the INLET for surface mounted components