05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency...

34
Implantable Implantable Cardioverter Cardioverter - - Defibril Defibril l l ators ators Ratko Magjarević University of Zagreb Faculty of Electrical Engineering and Computing Croatia [email protected]

Transcript of 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency...

Page 1: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

Implantable Implantable CardioverterCardioverter--DefibrilDefibrilllatorsators

Ratko Magjarević

University of Zagreb

Faculty of Electrical Engineering and Computing

Croatia

[email protected]

Page 2: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 2

Ventricular Fibrillation

• Ventricular fibrillation – asynchroneouscontractions of the myocardial cells

• Cardiac output near zero

• Irreversible damage to the brain if not handeledwithin approx. 5 min

• Therapy: defibrillation

– Defibrillation Principle: the electrical shock stops ALL

heart muscle fibers enter the refractory period at the same time

– Normal heart rhythm may resume

Page 3: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 3

Ventricular Fibrillation

Page 4: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 4

Defibrillators

• External defibrillators – electric shock is carried out by passing a current pulse through large electrodes placed transthoracally

• Capacitive discharge defibrillators – the capacitor is charged and the stored energy is up to 400 J (E = CU2 / 2 )

• For maximal energy, the voltage ranges from 2 kV to 9 kV!

Page 5: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 5

Defibrillation in Numbers

• Typical resistance of the chest (as

measured through the defibrillator pads) is

50 ohm

• Typical capacitance of the high voltage capacitors is 80 µF

• E = CU2 / 2 → for E = 250 J, U = 2500 V

Page 6: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 6

Capacitive discharge defibrillators

a) Block diagram of a capacitive

discharge defibrillator

b) The waveform of the outputpulse

Page 7: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 7

Defibrillator currents

Waveforms of currents from different different defibrillator typs:a) AC

defibrillator, b) capacitive discharge, c) serial resonant circuit, d)

rectangular, e) with delay line, f) biphasic

Page 8: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 8

Biphasic Defibrillators

Energy levels of biphasic defibrillators are approx. twice lower than those of

the monophasic apparatus.

Page 9: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 9

Defibrillator – Pacemaker - Monitor

Portability of emergency equipment is necessary due to very short times

available for interventions.

Reduction in size and weight while remaining the reliability is one of major

concerns in new developments.

Page 10: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 10

Defibrillator electrodes

a) Spoon shaped internal electrodes for direct application on heart during

surgery

b) A paddle type electrode for use on the chest

Page 11: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 11

Automatic External Defibrillator

Recent advances in technology have allowed people with little training to use

automatic external defibrillators (AEDs) in an emergency when medical

professionals are not present

Page 12: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 12

AED procedure

Page 13: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 13

Implantable

Cardioverter -

Defibrillators

Cardioversion - applying electrical shock to change an abnormal heartbeat into a normal one

Defibrillation - applying electrical shock to stop cardiac fibrillation and obtain sinus rithm

Page 14: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 14

Cardioversion

Page 15: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 15

Defibrillation

Page 16: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 16

Characteristics of ICD

• Volume 40 cm3

• Mass 70 g

• Defibrillation energy 30 J

• Detection accuracy 90%

Page 17: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 17

ICD Cross- section

Page 18: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 18

Battery

• Small internal impedance – capacitors are

charged by a current of 2-3 A

• Battery capacity equivalent to approx. 200 electric shocks

• or in time measures, approx. 9 years

Page 19: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 19

Defibrillation High Voltage

• Batery voltage (3 V) is transformed to 750 V

High Voltage Capacitor

• Aluminium Electrolytic Capacitor

• Capacitor volume – approx. 30% of the

ICD volume

Page 20: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 20

Therapy• Ventricular tachycardia

• Ventricular fibrillation

• Supraventricular arrhythmias

Implantation

• Local anesthesia

• Short postoperative care (1 day in

hospital)

• 55.000 implantations per year (USA)

Page 21: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 21

Atrium & Ventricle

� Bradycardia sensing

� Bradycardia pacing

Ventricle

� VT prevention

� Antitachycardia pacing

� Cardioversion

� Defibrillation

Therapies Provided by Today’s

Dual-Chamber ICDs

Page 22: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 22

Recent ICD Technologies:

Integrated Atrial Therapies

Atrium & Ventricle

• Bradycardiasensing

• Bradycardiapacing

Atrium

� Atrial tachyarrhythmia prevention

� Antitachycardia pacing

� Cardioversion

Ventricle

� VT prevention

� Antitachycardia pacing

� Cardioversion

� Defibrillation

Page 23: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 23

Page 24: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 24

Intracardial ECG Signal Analysis

• Tachycardia diagnosis is based on:

– PR interval analysis

– HR analysis

Page 25: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 25

Implantable leads and electrodes

Page 26: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 26

Implantable leads and electrodes

Page 27: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 27

Large devices –

Abdominal site• First human implants

• Thoracotomy, multiple incisions

• General anesthesia

• Long hospital stays

• Complications from major surgery

• Perioperative mortality up to 9%

• Nonprogrammable therapy

• High-energy shock only

• Device longevity ≈ 1.5 years

• Fewer than 1,000 implants/year

1980

Page 28: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 28

Small devices –

Pectoral site• First-line therapy for VT/VF

patients

• Transvenous, single incision

• Local anaesthesia; conscious sedation

• Short hospital stays

• Few complications

• Perioperative mortality < 1%

• Programmable therapy options

• Single- or dual-chamber therapy

• Battery longevity up to 9 years

Today

Page 29: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 29

ICD Evolution

Page 30: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 30

Programator

Page 31: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 31

Programator

• Radio-frequency telemetry and

programming

• Analysis of the signals

• Therapy

– pacing

– electrical shocks

• ICD performance testing

Page 32: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 32

GUI of a Programator

Page 33: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 33

Conclusions

• Recent trials demonstrated that ICDs reduce SCD and overall mortality compared to AA drugs for VT/VF patients and a subset of high-risk, post-MI patients

• Rapid improvements in ICD technology have resulted in smaller devices, simplified implant procedures, improved device longevity, expanded therapy options and reduced implant costs

• Up to 54% of ICD patients could benefit from dual-chamber therapy (traditional pacemaker indications and other indications)

Page 34: 05 Defibrillators Cardioverters · Defibrillator – Pacemaker - Monitor Portability of emergency equipment is necessary due to very short times available for interventions. Reduction

ICD lecture 2011 34

References:• M. Schaldach: Advances in Pacemaker Technology,

New York Univ Press, Monographs in BiomedicalEngineering Series, 1994

• Brown,B.H. et alt., “Medical Physics and BiomedicalEngineering”., IoP Publishing, London, reprinted 2001.

• Webster,J.G. (Ed.), “Medical Instrumentation, Applicationand Design.” 2nd ed., J. Wiley & Sons, Inc., New York, 1995.

• Nelson, C.V., Geselowitz D.B., ur.: “The TheoreticalBasis of Electrocardiography”. Claredon Press, 1976.

• Webster,J.G. (Ed.), “Bioinstrumentation”. John Wiley & Sons, Inc., New York, 2003