- Development of Chemical Dosimeters ... - OSTI.GOV
Transcript of - Development of Chemical Dosimeters ... - OSTI.GOV
SUDAN AGADEMY OF SGIENGES(SAS)
ATOMIC EhTERGY RESEARCHES COORDII\rATI ONCOUNCIL
- Development of Chemical Dosimeters
J
A dissertation Submitted in a partial Fulfi l lment of theRequirement fbr Diploma Degree in Nuclear Science
(Chemistry)
Fareed Fadl Alla Mersani
Supervisor Dr K.S. Adam
By
Murch 2006
SUDAN ACADEMY OF SCIENCES(SAS)
ATOMIC ENERGY RESEARCHES COORDINATIONCOUNCIL
Development of Chemical Dosimeters
A dissertation Submitted in a partial Fulfillment of theRequirement for Diploma Degree in Nuclear Science
(Chemistry)
By
Fareed Fadl Alla Mergani
Supervisor Dr K.S. Adam
March 2006
-
I
-
I
-
CONTENTS
Subject Page
Dedication..... IAcknowledgement .. .. IIAbstract III
DosimefryLntroductionPrinciple of DosimetryDosimefiy Systemsprimary standard dosimeters
l-3-Z Reference standard dosimeters ... ..Transfer standard dosimeters ... ...Routine dosimetersMeasurement of absorbed dose .Calibration of Dosimetrv svstemTransit dose effects
Requirements of chemical dosimetersIntroductionDeveloping of chemical dosimetersClassification of Dosimetry methodsRequiremsnts of ideal chemical dosimetersTypes of chemical systemLiquid aqueous systemsPrerequisites of chemical Dosimety
Purification of water
Treaftnent of glass ware and irradiation cellsTreatrnsnt of plastic irradiation cellsUse of irradiation cells .Stabilizing AgentsDefinition of G value
Ch-3 Types of aqueous chemical dosimeters3-I Fricke dosimeter3-l-1 Principle
ch-1t- lt-2l-3l -3 - l
L-3-3t-3-4I-4l -5l-6
ch-22- l2-22-32-42-52-62-72-7- l202-7-22-7-32-7-42-82-9
I2
J
4445
668
1lt2
l415
l6t7t9
20202l
2 l2 l
2323243-l-2 Basic reactions
----
CONTENTS
Subject Page
- Dedication... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... IAcknowledgement ... '" ... ... ... ... ... ... '" ... ... ... ... '" ... '" ..... 11Abstract ... ... ... '" ... ... ... '" ... ... ... ... -..... ... ... ... ... ... ..... III-
--
Ch-l1-11-21-31-3-11-3-21-3-31-3-41-41-51-6
DosimetryIntroduction .Principle of Dosimetry '" '" .Dosimetry Systems .primary standard dosimeters '" .Reference standard dosimeters " .Transfer standard dosimeters '" .Routine dosimeters .Measurement of absorbed dose .Calibration ofDosimetry system '" .Transit dose effects .
1234445
668
Ch-22-12-22-32-42-52-62-72-7-1202-7-22-7-32-7-42-82-9
Requirements ofchemical dosimetersIntroduction ... . . . ... . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . .. . . . . . . . . . . 11Developing ofchemical dosimeters ... . .. .. . . . . . . . ... .. . . . . . 12Classification of Dosimetry methods. .. . . . . . . . . . . . . . . . . . . . .. ... 14Requirements of ideal chemical dosimeters ,. . . . 15
Types ofchemical system .. . . .. . . . . . . . . . .. 16Liquid aqueous systems. .. . . . . . . .. . . . . . . . .. . . . . . . . . . . . . . . . . . 17Prerequisites of chemical Dosimetry '" ... 19
Purification of water '" " .
Treatment ofglass ware and irradiation cells... . .. ... . 20Treatment ofplastic irradiation cells '" '" .. . .. 20Use of irradiation cells... ... ... ... ... ... 21Stabilizing Agents ... . . . .. . . . . . . . . . . .. . . . . . . . . .. .. . .. . . . . 21Defmition ofG value ... . . . . . . . .. .. . . . . .. . . . . . .. . . . . .. . . . 21
Ch-33-13-1-13-1-2
Types of aqueous chemical dosimetersFricke dosimeter '" '" '" ..Principle '" '" .Basic reactions '" .. . .. . . .. . .. .. . .. . .. . . . . . .. . ..
232324
35
I
t
L
L
f
h
3-l-3 Calculation of the absorbed dose . 243-l-4 Influence of various factors 263-2 Ceric sulphate (Ceric - Cerous) dosimeter 283-2-l Principle 293-2-2 Calculation of absorbed dose 293-2-3 lnfluence of various factors 303-3 Other aqueous qystems 303-3-l Aromatic solutes Systems 303-3-2 Chlorinated hydrocarbon Solutes 313-4 Potential liquid dosimeters ... .. 313-5 Gaseous chemical dosimeters ..... 323-6 Solid chemical dosimeters ... . . ... ... ...32
ConclusionReferences
LIST OF TABTES
Table (1) :Table (2) :
Various Radiation Dosimetrv standardsEvents in Radiolysis of aqueous solutions.
34
6t9
•
L
...
3-1-3 Calculation of the absorbed dose 243-1-4 Influence ofvarious factors 263-2 Ceric sulphate (Ceric - Cerous) dosimeter 283-2-1 Principle 293-2-2 Calculation of absorbed dose...... 293-2-3 Influence ofvarious factors... 303-3 Other aqueous systems 303-3-1 Aromatic solutes Systems '" 303-3-2 Chlorinated hydrocarbon Solutes 313-4 Potential liquid dosimeters , '" 313-5 Gaseous chemical dosimeters " 323-6 Solid chemical dosimeters 32
Conclusion , 34References ... 35
LIST OF TABLES
Table (1): Various Radiation Dosimetry standards 6Table (2): Events in Radiolysis of aqueous solutions... .. 19
ffiEfrTTffi
t*I.
tJ-
It -
rI
l -
To those who are donate me their live...To those who are lighten my way by candles...To those who are learned me and guided me...
To all of you ...
MotherTeachers
BrothersFriends
11 I offer you my bit hopping thatit will find even a bit
l- of Your consideration.
, -
trl
r{t -
/-
J.
Father
L
(1,-
SE
-
-
L
~rrrr~
r~
To those who are donate me their live...To those who are lighten my way by candles .To those who are learned me and guided me .
To all of you ...
FatherMother
TeachersBrothers
Friends
I offer you my bit hopping that it will find even a bitof your consideration.
I
L
- f -
a-
J,
L-
L
ACKNOI|TLEDGE|lJ|ENTI thank GOD too mush that I could complete this
L beneficial work after a lot of effort and constraintsthat faced me.
I
I deeply grateful to all those who helped me in1' achieving this research
I Many thanks for Sudanese Atomic EnergyCommission (S.A.E.C) Staff for their advises.
--
Also full thanks for my supervisor Dr K.S.ADAM-- for suggested idea and orientation and for his valuable
r gtidance through out the research.G
rL
a/-
a.
J
---~ -
1......
...
..
ACKNOWLEDGEMENTI thank GOD too mush that I could complete this
beneficial work after a lot of effort and constraintsthat faced me.
I deeply grateful to all those who helped me inachieving this research
Many thanks for Sudanese Atomic EnergyCommission (S.A.E.C) Staff for their advises.
Also full thanks for my supervisor Dr K.S.ADAMfor suggested idea and orientation and for his valuable
guidance through out the research.
11
L -
L -
L -
L- -
Affi AOTA chemical dosimeter is a systern that meazures the energy by virtue of
chemical changes from ionizing absorbed radiation produced unit when it
is exposed to ionizing radiation. In all chemical dosimeters radiation
induced chemical reaction produces at least one, initially absent species,
which is properties long lived enough to determine its quantity or the
change in the initial systems. Different types of chemical dosimeters were
discussed such as aqueous, gaseous and solid, but the great consideration
was given to aqueous systems because of their vital role in setting many
processes.
ilI
--
....'---
....'---
'---
....
....
-
A chemical dosimeter is a system that measures the energy by virtue of
chemical changes from ionizing absorbed radiation produced unit when it
is exposed to ionizing radiation. In all chemical dosimeters radiation1
induced chemical reaction produces at least one, initially absent species,
which is properties long lived enough to determine its quantity or the
change in the initial systems. Different types ofchemical dosimeters were
discussed such as aqueous, gaseous and solid, but the great consideration
was given to aqueous systems because of their vital role in setting many
processes.
III
ilI{APTER ONEl.DOSIMETRY' l -
I
I
a .
a-.j
aJ
a_-
t--'
----
-· -
· -· -'
· -· -"
· -'
~
· -"
~
· -
" -'
-"~
"I .J
.l-.-.
L....
'----'
-"
"-----"
Il..-........o
-"
Il..-........o
...lI.-..-.-A
...'-.~
...'-
.-I
APTERONEI-DOSIMETRY
1 -1 Introduction:
Use of ionizing Radiation has become increasingly important in a different
fields, industrial processing applications such as polymer cross linking,
polyrner degradation, polymer grafting, vulcanrzation, curing of coating,
scrubbing of gaseous effluents, sterilization of medical products, sewage
sludge hygienisation, delayed ripening of fruits, sprout inhibition and insect
population control. The absorbed doses employed in these applications range
from 10 Gy - more than 100 kcy). 2
In the use of ionizing Radiation, reproducible and an accurate irradiation of
the products to get desirable effects depends on having reliable dosimetry
systems. Dosimetry play a vital role in setting of process parameters to meet
variety of specifications, dose mapping in products, carrying out validation,
commissioning procedure as well as in the day today operation of the plant.
Measuring response of dosimeter to Radiation is generally much easier and
quicker to measure than any other parameter of the product that has been
irradiated. Documentation of dose is often required in order to ensure
operational safety quality control. 2
The effects of Radiation on the matter depend on the Radiation field, as
specified by the Radiation quantities and on the interactions between
Radiation and matter, as characterized by the interaction quantities.
Dosimetric quantities are products of radiometric quantities and interaction
coefficients. Radiation interacts with matter in a series of processes in which
particle energy is converted and finally deposited in matter. Measurement of
-
...
-
-
-
1 -1 Introduction:
Use of ionizing Radiation has become increasingly important in a different
fields, industrial processing applications such as polymer cross linking,
polymer degradation, polymer grafting, vulcanization, curing of coating,
scrubbing of gaseous effluents, sterilization of medical products, sewage
sludge hygienisation, delayed ripening of fruits, sprout inhibition and insect
population control. The absorbed doses employed in these applications range
from 10 Gy - more than 100 kGy). 2
In the use of ionizing Radiation, reproducible and an accurate irradiation of
the products to get desirable effects depends on having reliable dosimetry
systems. Dosimetry play a vital role in setting of process parameters to meet
variety of specifications, dose mapping in products, carrying out validation,
commissioning procedure as well as in the day today operation of the plant.
Measuring response of dosimeter to Radiation is generally much easier and
quicker to measure than any other parameter of the product that has been
irradiated. Documentation of dose is often required in order to ensure
operational safety quality control. 2
The effects of Radiation on the matter depend on the Radiation field, as
specified by the Radiation quantities and on the interactions between
Radiation and matter, as characterized by the interaction quantities.
Dosimetric quantities are products of radiometric quantities and interaction
coefficients. Radiation interacts with matter in a series of processes in which
particle energy is converted and finally deposited in matter. Measurement of
1
i\
the energy absorbed per unit mass in a medium exposed to ionizing
Radiation necessitates the introduction of a dosimeter into the medium.
Dosimeter is a device that, when irradiated, exhibits a quantifiable changes
in some property of the device which can be related to the absorbed dose in a
given material using appropriate analytical instrumentation and techniques.
Different tlpes of dosimeters like gaseous ionization chambers, thin films,
solids and liquids are used in Radiation dosimetry. There is a considerable
variation in their size and composition. Also several wall materials having
varying thickness are used to contain the dosimeter. 2
l-2 Principles of Dosimetry:
The dose in a medium is measured by replacing the medium by a dosimeter.
Normally, dosimeter will differ from the medium in both atomic number and
density and it therefore, constitutes a discontinuity, which will be referred to
as activity. The energy absorbed in the dosimeter is therefore not the same as
that absorbed by the medium. Under electronic equilibrium conditions, the
dose in the medium can be estimated using cavity (Bragg GraD theory. (2)
For irradiations using a photo-source, the dosimeter may be considered as
activity in the material of interest, and the interpretation of absorbed dose in
material as follows. If the sensitive region of the dosimeter is very thin
compared to the range of the highest energy secondary electrons, then most
of the energy deposited in the dosimeter and in the material surrounding it
results from secondary electrons produced outside the dosimeter (that is, in
the equilibrium layer of material). Thus the absorbed dose in the material 2
, Dr[, is given by:
Dm = 6l P)m ,n
(s I P)d
H
2
- ...
_ .J
-- ~,
-
- ...
- ...
- ...
- ...
- ,.,
- ..
- ...
.. -- ...
. ..
.. -
the energy absorbed per unit mass III a medium exposed to ionizing
Radiation necessitates the introduction of a dosimeter into the medium.
Dosimeter is a device that, when irradiated, exhibits a quantifiable changes
in some property of the device which can be related to the absorbed dose in a
given material using appropriate analytical instrumentation and techniques.
Different types of dosimeters like gaseous ionization chambers, thin films,
solids and liquids are used in Radiation dosimetry. There is a considerable
variation in their size and composition. Also several wall materials having
varying thickness are used to contain the dosimeter. 2
1-2 Principles of Dosimetry:
The dose in a medium is measured by replacing the medium by a dosimeter.
Normally, dosimeter will differ from the medium in both atomic number and
density and it therefore, constitutes a discontinuity, which will be referred to
as activity. The energy absorbed in the dosimeter is therefore not the same as
that absorbed by the medium. Under electronic equilibrium conditions, the
dose in the medium can be estimated using cavity (Bragg Gray) theory. (2)
For irradiations using a photo-source, the dosimeter may be considered as
activity in the material of interest, and the interpretation of absorbed dose in
material as follows. If the sensitive region of the dosimeter is very thin
compared to the range of the highest energy secondary electrons, then most
of the energy deposited in the dosimeter and in the material surrounding it
results from secondary electrons produced outside the dosimeter (that is, in
the equilibrium layer of material). Thus the absorbed dose in the material 2
, Dm, is given by:
Dm = (SIP)m Dd(SIP)d
2
J
L. . -
J
1
L -
1
I
Where (sf t ' )m ancl (SlP)d are mass col l is ion stoppirrg powcr l i r r the
st t r rounding nrater ia l ancl dosir lcter ' . rcspect ivel ; , . l )d ls absorbecl dose ip
the dosimeter. l f the sensi t ive rcgion ol- t l ie c losir letcr l ras a th ickncss uruclr
greater than the I 'ange of the highesl energy sccorrc larv c lcct lnns. lhor rn11sl
of the energy de posi tecl in i t resul ts 1l 'or l thc scconclarv elcctrons proclucecl
wi th in the dosinreter i t sel l , t l r r , rs , the absorbed dose in the rnater iz i l is q ivcnr 2O V :
I)rt = 0y4!!,, D,l(per t , P)c l
Where (1-rcnf I))nt ar'r4v l lrcnf1')r/ ' ,, ' . thc nrass absorpti.n coelflcierrts of thc
ntediutn, / t / a l td the dosimetcr tnertcr ia l , / , resl lect ively. I f thc scnsi t ivc
region of t l ie dosi tneter has a t l r ic l<ncss betwecn thc two l i rn i ts. thqr lhe
above two ecl t tat ions may l re cornbirrccl rv i th approl t r iatc u 'e isht i l tg thctors to
re f lec t the re la t i ve co t r t r ibu t io t rs o l ' cach tc rn t ( l Jur l in 's gcnera l ( 'av i ty ' thcur .y
l 'he col l is ion stopping powcrs ancl l l re crrcr .g) , absor. l ) t iorr cocl- l lc icnts ar-c
energy dependent: however lor hoiv atornic nr.rrr rbcr rrratcr ia ls arrc l f or t l rc
el lergy range 0.1 to l0 MeV. thc rat ios of ' thc stol ' r l t i r ru powcl 's l r rd cncluy
absorpt ion cocf ' l lc ients c lc l I )oI var '1, s igrr i l icant l_1, as a l i rnct iorr o l 'cncrs-r , .
A l tho t rgh th is de f i l t i l i o t r i s g ivcn s l r ' i c t l v lo r i r t rsorbcc l r losc a t a po i r r l i r r
I {adiat ion absorbing trrater, i t is gerreral ly avclagccl o l 'cr a l j r r i tc nrass ul 'u
given nrater ia l , thc absorbecl c lose bcirrg r"cacl b1,a c losinrctcr cal ibratc i r r
terms of energy i rnparted per r . rn i t rnass of 'a givcn r latcr iar l . ln l {acl iat iorr
processi t rg, the rcferetrce tnater ia i in nrosl cal ibrat ion is r ,vatcr , i t is i rnpo11alr
to real ize that the dosimetet ' is intendecl to givc nrcasurcrnerr t o l 'absor.bccl
dose averagecl t lver a smal l volurne. - l -he
s ignal Iorrrr th is c lc ls i r r rctcr carr
therefore be a tneasttre of encrgy absorbccl in that volrrrnc. vvhich can bc
related to the c lose in the proclLrct . 2
-
'-----
'------
.....
'---
'--
-
-
Where (S /1')//1 and (S'/ P)d are mass collision stopping power for the
surrounding material and dosimeter. respectively. Dd Is absorbed dose in
the dosimeter. I f the sensitive region of the dosimeter has a thickness much
greater than the range of the highest energy secondary electrons, then most
of the energy deposited in it results from the secondary electrons produced
within the dosil11eter it selC thus, the absorbed dose in the material is given
by: 2
Dill = (pen/ PJ!!2 Dd(pen/P)d
Where (I-ten/P)17I and (pen/ jJ)d arc the mass absorption coefficients of the
medium, 11/ and the dosimeter material d ,respectively. If the sensitive
region of the dosimeter has a thickness between the two limits, then the
above two equations may be combined with appropriate weighting f~tctors to
ref1ect the relative contributions of each term (Burlin's general Cavity theory
The collision stopping powers and the energy nbsorption coefficients arc
energy dependent; however for how atomic number materials and 1'01' the
energy range 0.1 to 10 MeV, the ratios of the stopping powers and energy
absorption coefficients do not vary signi ficantly as a function of energy.
Although this definition is given strictly for absorbed dose at a point ill
Radiation absorbing mater, it is generally averaged over a finite mass of a
given material, the absorbed dose being rend by a dosimeter calibrate in
terms of energy imparted per unit mass of a given material. In Radiation
processing, the reference material in 1110st calibration is water, it is important
to realize that the dosimeter is intended to give measurement of absorbed
dose averaged over a small volume. The signal form this dosimeter can
therefore be a measure of energy absorbed in that vulume. vvhich call be
related to the dose in the product. 2
-r'
' t
L _ -
L
- !
l -3 Dosi tnctry Systerns:
1 'hey are usecl to tneasut 'e absorbcd dosc. ' l -hcy
consist o1-thc dosirncters.
measurenrent instrunrents ancl thci l associzr tccl rc l -crcncc stanclalds. ancl
procedures Ibr the system's r"rsc. Dosir lctcrs nray bc c l iv idcd into forrr basic
classes according to the elccuracy of thc c losir lctrv s1'stcnrs ancl arcas i t l -
appl icat ions as lo l lows. 2
l-3- l Pr imary Starrdirrd Dosirncters:
Pr imary standard dosimelct 's arc establ ishccl ancl rnaintaincd by nat iorral
standards laborator ies for cal ibrat ion ol ' l {acl iut iorr cnvi lor . r rncnts ( l ie lc ls) ancl
other dosimetcr"s. these are ei thcr ionizat iorr charnbcrs or calor i r letcrs. ' l ' l rc
overal l l tncertai t r ly at thc staqc is 1 lo,"o (at()5ot 'o cs "starrc l i r rc l c lcviat iorr") . 2
l 3-2 l tefcrcnce Startdarr l Dosirnetcr 's' l 'hese
closi tneters usecl to cal iLr latc l tacl iat ion cnr, ' i rorrnrcnts ancl rout ine
closimeters, re l 'c t 'c t rcc statrc larc l c losirnctcrs nr[r) / a lso l rc Lrsccl as nrut i r rc
dos imcters . ' l ' hcsc are chcnr ica l c los i rnc t r ' l , sys tc rn rvhcrc rcs l )onsc no t on ly to
I tadiat ion br-r t to ot l tcr in l lucncing l i rcturs such as tcnr l - ' 'c1 'n11r lc ancl c losc: ra lc:
i s reproc lL rc ib lc anc l wc l l charac tc r iz .cc l . ' l ' hcy a rc c luss i l l cc l as 11 ,1)c : "A" a r rc l
type "B" . ' ] .hc type "A" c los i t t t c tc t ' s I i r r cxa tnp lc . I ; t ' i c l< r : . ( ' c r i c . r l i c l r ro r ru r tc .
ethanol-nronochlorotrctrzcnc sol t t t iotrs arc svstctrrs iv l rosc I {acl iat ion
c l ' re rn ica l response is c l ra rac tc l i s t i c o l ' a par t i cLr la r c l rc r r i ca l cornpos i t ion anc l
th is can bc guarantcccl , ccrtai t t lv to rv i th i r r a l -c l r , 1-rcrccnt. provic lcd a goocl
chern ica l p rac t icc i s f i r l l o rvcc l . ' l - ypc " l l " r los i rnc tc rs i r rc sys lcnrs u 'h ic l r
exh ib i ts h igh prec is ion bu1- havc to bc ca l ib ra tcc l aga i r rs t a s tanc la rc l h ighcr in
the ser ies . '1 -he i l l tac l ia t ion fcsponsc car r no t bc p rc r l i ca tcc l pLr rc l ; ,on thc
bas ic o f con ipos i t ion thc ovcra l l r rnccr ta i r r tv assoc ia lcc l rv i t l r rc l ' c rc r rcc
dosirrreters is about + 3o .2
==
• -6
'------
• ..A
• -6
· ~
"----
• -6
· ...
~--
· ....
- ....
· .... ...
1-3 Dosimetry Systems:
They are used to measure absorbed dose. They consist or the dosimeters.
measurement instruments and their associated relcrence standards, and
procedures for the system's use. Dosimcters may be divided into four basic
classes according to the accuracy of the dosimetry systems and areas or
applications as follows. 2
1-3-1 Primary Stalllhud Dosimetcrs:
Primary standard dosimeters arc established and maintaincd by national
standards laboratories for calibration of Radiation environments (fields) and
other dosimeters, these are either ionization chambers or calorimeters. The
overall uncertainly at the stage is ± I% (at 95% er "standard deviation"). 2
13-2 Reference Standard Dosimetcl's
These dosimeters used to calibrate Radiation environments and routine
dosimeters, reference standard dosimeters may <llso be used as routine
dosimeters. These are chemical dosimetry system where response not only to
Radiation but to other inlluencing factors such as temperature <lIld dose rate
is reproducible and well characterized. They arc classilied as type "A" and
type "13". The type "1\" dosimeters l()r example. Fricke, eeric. dichroll1ate.
ethanol-monochlorobenzene solutions arc svstems whose Radiation
chemical response is cltaracteristic of' a particular chcmiull composition and
this can be guaranteed, certainly to within a few percent, provided a good
chemical practice is followed. Type "13" dosimeters arc systems which
exhibits high precision but have to lx' cal ibrated against <1 standard higher in
the series. Their Radiation response can not be predicated purely on the
basic of composition the overall uncertainty associated with reference
dosimeters is about ± 3%. 2
r /
L-
L -
L
H
1-3-3 Transfcr Standnrd l )osi lne(e rs
Tl-rese are specizt l ly selcctecl c losinrctcrs usccl f i r r t rans{crr ing absorbccl c losc
inlbrmat ion f ronr accredi tcd or nat ional stanclarrc ls labur i r turv tc l an i r racl iat ion
lac i l i t y i t r o rdcr to cs tab l i shec l t raccab i l i t y {o r tha l lac i l i t l , . ' l ' l r csc c los in rc tc rs
shoulc l be usecl unclcr concl i t ions that are caref i r l lv control lccl b i r t l tc issLr ins
laboratory. f 'ratrsfer stiurclarcl closinrcte rs nray bc sclcctccl l i 'onr cit l icr
reference stanclard dositlctcrs or routinc closinrctcrs ancl shall havc
perforutancc character ist ics sucl t as long shcl l l i fb. casi ly cal i t r ratccl . stablc.
rugged, l tor tablc, t -nai l able, broacl absorbccl dosc langc. l lacl iat ion absor 'pt ion
propert ies s inr i lar to those of i r racl iat ion proclLrct . r 'c lat ivcly incxpcnsivc to
extremes of environnrental concl i t ions ,corrcctabIc st 's tcrnat ic crrors (c.g.
tetnperature, hr-rrnic l i ty , ctc) . proclLrciblc lots. r 'cpnrclrrc ib lc rcsl-ronsc ancl
snral l d imensions colxpal 'ccl to c l is tanccs ovcr rvhic l r atrsor i ' rccl c losc gracl icnts
becontc s isni f lcant. ctc. 2
l-3-4 l l.outinc l)osinrctcrs' they
nray be usecl lbr cpral i ty control arrc l l ) r 'occss nrorr i tor ins. I {out inc
dosimeters are systctns whosc pcr l i r r rnancc, part icLr l i t r ly l i , i th lcspcct to
environtnental inf l r tcrrc i t rg l i rctols, is not as goocl as t l rc l 'c l -crurcc systcnrs.
but whose case o1- usc ancl low cost nakcs thcrn i r lcal l i r r c lay l i i c la,r ,
mon i to r ing o f I {ad ia l io r r c loscs c lu r ing pr 'occsscs . ( ' r i l c r ia l i r r sc lcc l io r r o l '
rou t ine c los i r l c t ry sys tcn r inc luc lc su i tab i l i t y 'o f ' t l r t : c los in rc tc r ' { i r r t l r c
absorbed dose ratrgc ol ' intcrcst ancl l i r r usc wi th a s l )cci l ic l t roclLrct . sta["r i l i tv
anc l reproc iL rc ib i l i t y o l ' thc sys tcnr . casc o l ' sys tc rn ca l ib la t ion . t raccub i l i t l , o l '
systet l cal ibrat iorr to ni t t ional stanclarcls. abi l i tv to corr t r 'o l svstcn'r l 'csp() l rsc
lbr system unccrtaint ics. such as t l rosc causccl b1, 1" ' , . , ' ' , . ' 'a t r r rc uncl l r r r r r r ic l i t r , .
ease aud s in ip l i c i t y o f usc . ava i lab i l i t y o { - c los i rnc t t - r ' s ; i r r rcasonab l ) ' la lgc
- ~
- ~
- .....
- .....
- .....
- ...
- ....
- ....
- ....
- -'"
- -'to
- ....
------- ..--- ...
- ..- ...
.....
.... --
1-3-3 Transfer Standard Dosimetcrs
These are specially selected dosimeters used for transICrring absorbed dose
information from accredited or national standards laboratory to an irradiation
facility in order to established traceability for that f~lcil ity. These dosimcters
should be used under conditions that are carefully controlled by the issuing
laboratory. Transfer standard dosimeters may be selected from either
reference standard dosimeters or routine dosimeters and shall have
performance characteristics such as long shelf life, easily calibrated, stable,
rugged, portable, mail able, broad absorbed dose range, Radiation absorption
properties similar to those of irradiation product, relatively inexpensive to
extremes of environmental conditions ,correctable svstcmatic errors (e.g.• L
temperature, humidity, etc), producible lots, reproducible response and
small dimensions compared to distances over which absorbed dose gradients
become significant, etc. 2
1-3-4 Routine Dosimeters
They may be used for quality control and process monitoring. Routine
dosimeters are systems whose performance, particularly with respect to
environmental influencing f~lctors, is not as good as the reference systems,
but whose case of use and low cost makes them ideal for day to day
monitoring of Radiation doses during processes. Criteria for selection or
routine dosimetry system include suitability or the dosimeter for the
absorbed dose range of interest and for use with a specific product. stability
and reproducibility of the system, ea5e of system calibration, traceability of
system calibration to national standards, ability to control system response
for system uncertainties, such as t!lo5C caused by temperature <lIld humidity,
ease and simplicity of use, availability of dosimeters in reasonably large
5
cluant i t ies, overal l in i t ia l ancl operat ional cost c l l -systcrn i r rc lucl i r rq c losirnctcrs
readout equipnretrt ,ancl labor recluirccl fbr cklsiurctcr rcaclout ancl
interpretat ic ln. t ' r - rggedncss of thc s1,st t - 'nr rcsistancc to r lanrai tc c l r r r i r rg roLrt inc
hand l ing and usc in a p rocess ins cnv i ronnrcn l . c tc . . 2
fable ( I ) : Var ioLrs l {adiat ion c losirnclry standarcls
DosimetersI {out i rre Dosir leters ( 5?6) R"f.*,,r*-- Str""k, ' ,1;
Dosi rnetcrs (3%r)l)r'i rrI ) os
Plastics Fr ic l<c !gr( ' a l rDve o last ics Clcr ic- Clerous
(CTA) f ihns [ ) i chronru lcRadio chronic d) '" f l l rns
' f I -
dosimetersl r thano l r lonoch lo r 'obcnz .c r re u Ia r r inc
l -4 Measurenrent of absorbed t lose or t losc ratc
The reference soLrrce is cal ibratecl by mcans ol-onc ol ' thc l r r inrary stanclarcls
(e.g. graphi te calor inretr ic or standarcl izecl re l -crcncc urcthor is " f i l ic l tc
dosirnetry") and the closirneter signal is therr cclnvcrlccl to absortrccl cJosc or'
dosc t 'a (c in rva tc t ' l r1 ' ( '11111 l r l r t r t t i t r t t s t ' t r t sc r l t r t r ( ' i r v i l v th t ' r r r r ' . ' l ' l r t ' r ' r ' { i ' l t ' t r c i '
soL l l ' oc i s t rscc l to i r t ' l c l ia tc t l t c t 'o t t t inc r los i r r rc tc r i s r r r r t le r i t l c r r t i c i r l cor r t l i t i o r rs .
and the cal ibrat ion is r . tsual ly cxpressccl in tcrnrs ol 'c l i rsc ur c lc lsc ratc in t [c
dosimeter mecl iut 'n. fhe roul i t re c losi lneter is thcn cxl tosccl tc lgct l tcr q, i t I thc
produc t in the i r rac l ia t ion , fac i l i t l , . a l t i r< lugh thc i r ra r l ia t io r r conr l i t ions ( i . c .
el le l 'gy spectrunr erncl gconrctry) ntay t re ntar.kcr l l ) , c l i l ' lcr-cnt l l .or t r thosc ol ' thc
cal ibrat ion. 2
I -5 Ca l ib ra t ion o l 'Dos in rc t l ' l svs le lns
Cal ibrat ion ver i f lcat ion is pcrfonnecl pcr iocl ical ly to corr f l r ru 1hc cont inrrcr l
val id i ty of thc cal ibrat ic ln curve. [ {ourt inc dosinrctry s l ,stc lns cztn bc
i rrrctcrs ( l-2'1,1,)zi t t iort c lr iur l rcr '
-. ~
. ~
quantities, overall initial and operational cost orsystcm including dosimeters
readout equipment ,and labor required ror dosimeter readmit and
interpretation, ruggedness or the system resistance to damage during routine
handling and use in a processing environment, etc .. 2
Table (1): Various Radiation dosimctry standards
'1lllary Standardsosil11eters (1-2(~(l)
nizatioll chambera lori Illeters
Dosimeters.-- ---
Routine Dosimeters (5%) Reference Standards PIDosimeters (3(Xl) !)
Plastics Fricke 10Dye plastics Ceric- Cemus C(eTA) films Dichromate
-- ._- ---
Radio chronic dye films TL Ethanol monoch lor-dosimeters obenzene alanine
1-4 Measuremcnt of absorbcd dose or dose ratc
The reference source is calibrated by means of onc of the primary standards
(e.g. graphite calorimetric or standardized reference methods "Fricke
dosimetry") and the dosimeter signal is then converted to absorbed dose or
dose rate in W\1ter b~' c()mputnti()n:-. bnsed on Cn\'i\\' \\w\)ry, TIl\' r\'!t'n'\\\'\'
source is lIsed to irradiate the routille dusillll'tl'l' is ulldn idclltic,t1 cUlldiliulls.
and the calibration is usually expressed in terms or dose or dose rate in the
dosimeter medium. The routine c10simeter is then exposed together with the
product in the irradiation, facility, although the irradi<ltion conditions (i.e.
energy spectrum and geometry) may be markedly di JTcrent II'om those or the
calibration. 2
1-5 Calibration of Uosimctry systcms
Calibration verification is perrormed periodically to conllrm the continued
validity of the calibration curve. Routine dosimetry systems can be
6
t_ .
-
t
t - ,
t
-
t
-
t
calibrated by irradiation at high dose Radiation dosimetry calibration
laboratory, an in- house calibration facility whose dose rates have been
demonstrated to be traceable to appropriate nationai standard or irradiation
of reference or transfer standard dosimeters with routine dosimeters in the
production irradiation facility. All possible factors that may affect the
response of dosimeters, including environmental conditions and variations of
such conditions within a processing facility should be known and taken in to
a count.
For each absorbed dose point, use number of dosimeters required to achieve
the desired confidence level. The number of dosimeters, n, required to
estimate the dosimeter response at a given absorbed- dose level is based on
the determination of a two sided confidence interval that is expected to
bracket the true mean response y6,100 (l- a)% of the time. In order to
determine the mean response, y6, within + 5oh at a95oh confidence level, the
number of dosimeters required for absorbed dose level is
t 7 s z
" =
(oLoEiC
Where .s is the estimate of the standard deviation of the response
distribution within a batch of random sample of dosimeters and r is the
student's distribution. 2
The number of sets of dosimeters required to determine the calibration curve
of the dosimetry system depends on the absorbed- dose range of utilization.
Use at least five sets or each factor of ten span of absorbed dose, or at least
four sets if the rang of utilization is less than a factor of ten. Position the
dosimeters in the calibration curve Radiation field in a defined, reproducible
location. The variation in absorbed dose rate within the volume occupied by
the dosimeter should be within+lo/o of the averase value.
•.........
•.........
.-~ ..
............
.-
I.......
.-
.........
.-
I.......
..-
-1--__
..
---
..
calibrated by irradiation at high dose Radiation dosimetry calibration
laboratory, an in- house calibration facility whose dose rates have been
demonstrated to be traceable to appropriate national standard or irradiation
of reference or transfer standard dosimeters with routine dosimeters in the
production irradiation facility. All possible factors that may affect the
response of dosimeters, including environmental conditions and variations of
such conditions within a processing facility should be known and taken in to
a count.
For each absorbed dose point, use number of dosimeters required to achieve
the desired confidence level. The number of dosimeters, n, required to
estimate the dosimeter response at a given absorbed- dose level is based on
the determination of a two sided confidence interval that is expected to
bracket the true mean response Yo,1 00 (1- a)% of the time. In order to
determine the mean response, Yo, within ± 5% at a 95% confidence level, the
number of dosimeters required for absorbed dose level is
Where S IS the estimate of the standard deviation of the response
distribution within a batch of random sample of dosimeters and t is the
student's distribution. 2
The number of sets of dosimeters required to determine the calibration curve
of the dosimetry system depends on the absorbed- dose range of utilization.
Use at least five sets or each factor of ten span of absorbed dose, or at least
four sets if the rang of utilization is less than a factor of ten. Position the
dosimeters in the calibration curve Radiation field in a defined, reproducible
location. The variation in absorbed dose rate within the volume occupied by
the dosimeter should be within ±1% of the average value.
7
I t
j
When using galnma- ray source or X-ray beam for calibration surround thedosimeter with a sufficient amount of material to achieve approximate
electron equilibrium conditions. The appropriate thickness of such materialdepends on the energy of the Radiation. 2
For measurement of absorbed dose in water, use material that have Radiation
absorption properties essentially equivarent to water for example 60co
source' 5nm of solid poly styrene (or equivalent polymeric material) shouldsurround the dosimeter in all directions. Monitor and control temperature
and humidity during calibration. 2
1-6 Transit Dose effects:
Transit dose effect occur when the timing of calibration irradiation dose not
take into account, the dose received during the movement of the dosimeters
into and out of the irradiation position for example the timer on gamma cell-type irradiation dose not start until the sample chamber reaches the fully
down (irradiate) position. Some dose is received as the drawer goes down
and after the irradiation as the drawer goes up that is not accounted for in thetimer setting. This transit dose can be significant for low dose irradiations
and should be determined experimentally and incorporated into a formula for
calculating timer settings. For determining transit dose, a series ofincremental inadration times that started as clos e to zero selected,, with
specific dosimetry system employed. Five different inadiation times were
used. The results analyzed using linear regression analysis. A graph of theresults looked similar to following diagram: 2
J
...-.
•- ..--
...
-...
--41
~
....-
...
--...
-...
...
When using gamma- ray source or X-ray beam for calibration surround the
dosimeter with a sufficient amount of material to achieve approximate
electron equilibrium conditions. The appropriate thickness of such material
depends on the energy of the Radiation. 2
For measurement of absorbed dose in water, use material that have Radiation
absorption properties essentially equivalent to water for example 60CO
source, 5nm of solid poly styrene (or equivalent polymeric material) should
surround the dosimeter in all directions. Monitor and control temperature
and humidity during calibration. 2
1-6 Transit Dose effects:
Transit dose effect occur when the timing of calibration irradiation dose not
take into account, the dose received during the movement of the dosimeters
into and out of the irradiation position for example the timer on gamma cell
type irradiation dose not start until the sample chamber reaches the fully
down (irradiate) position. Some dose is received as the drawer goes down
and after the irradiation as the drawer goes up that is not accounted for in the
timer setting. This transit dose can be significant for low dose irradiations
and should be determined experimentally and incorporated into a formula for
calculating timer settings. For determining transit dose, a series of
incremental irradiation times that started as close to zero selected, with
specific dosimetry system employed. Five different irradiation times were
used. The results analyzed using linear regression analysis. A graph of the
results looked similar to following diagram: 2
8
The intercept with Y- axis gives the value of the transit dose. To provide a
correction factor for all future irradiations, convert the transit dose by the
current dose rate. Subsequent timer settings can be calculated using the
formula:
Where D, is the target dose and r is the timer setting required achieving
that dose 2.
! - D, , at a given time , l, is given by
t-
Ir-T-
(.) 200a
100
100 200 300
Time of irradiation (min)
T- D ' - r ,D
l -I
Since the absorbed dose rate, D due to galnma ray emission by a Radioactive
nuclide source also varies exponentially with the decay time r, the dose rate,
9
-.
_A
-...
- ....
300
400 ~----------------~---,
300,-.>-
Cl--(l) 200rJ:J00
100 ...........................
0100 200
- ....
- ...
- ....
- ...
Time of irradiation (min)
.. ...
.. ...
The intercept with Y-axis gives the value of the transit dose. To provide a
correction factor for all future irradiations, convert the transit dose by the
current dose rate. Subsequent timer settings can be calculated using the
formula:
..... T=Dt--ID t
Where D t is the target dose and 1 is the timer setting required achieving
that dose 2.
Since the absorbed dose rate, D due to gamma ray emission by a Radioactive
nuclide source also varies exponentially with the decay time 1 , the dose rate,
~ .A
I
l~
rr~
r=
D t , at a given time, I, is given by
9
dose varies inversely with the dose rate and source activity, and is given by
/-^\ (rs)(75 f, = -:------11
e
Where (fs) is the timer setting necessary to deliver the required target dose
IL-
L -
D, = Dor-'' '
Where D,the dose rate is at a time,Do is the dose rate at some earlier
time(r=o). The timer setting,Zs necessary to deliver the targeted central
at timer, (rs), is the timer sitting at some earlier time,r=0 to deliver the
same tarset dose 2
t-
10
- ..
- ...
- .- ...
_. AI
_...
_. "'"
- AI
- ~
- ,lID
- ....
Where D( the dose rate is at a time, Do is the dose rate at some earlier
time (t = 0). The timer setting, TS necessary to deliver the targeted central
dose varies inversely with the dose rate and source activity, and is given by
(TS} = (T~)oe
Where (TS)/ is the timer setting necessary to deliver the required target dose
at timet, (TS)o is the timer sitting at some earlier time, t =0 to deliver the
same target dose 2
~ ..
L ....-....- 10~ ..
l ..
[
- 1
i ' {
r*I
I
\ -
t*L -
fltIAPTERT\VO2- RE,QUIREMENTS OF CHEMICAL DOSIMETERS
.....
.. ...
. .
- ..
- ..
............
- ..
~ -
l .
IIAPTERTWO2- REQUIREMENTS OF CHEMICAL DOSIMETERS
L- ' -
: rrl
r--L,
L T
t-:t,
t-:
2-1. Introduction:
The first chemical dosimeters were crude systems which changed color on
exposure to the relatively soft x- rays then being used medically. Barium
platinocyanide was the primary chemical ingredient of pastiiles, which were
developed by several investigators. After irradiation, this material was found
to change from its normal green color to orange, and then to various shades
of brown. (3)
The main difficulty encountered with pastilles was that the color reverted to
its original green on exposure to light, further more, the color standards in
the comparator charts gradually faded and thereby contributed to in accurate
measurements. Most important, these systems were highly energy dependent
because they used materials of high atomic number. There fore, changes in
the spectrum of the x-rays caused large variations in doses registered.
A few years Iater, other chemical systems were developed, for example,
Fruend 6 utilized the effect of x-rays on various iodine compounds. A
mixture of iodofonn was found to be sensitive to x- irradiation and was
employed in dosimetry, but the reaction was also produced by head, light
and spontaneous oxidation, and was not proportional to dosage. 3
According to Gruther et al rt i928 chloroform inadiated with x-rays librates
hydrochloric acid in amounts proportional to the Radiation energy absorbed.
Acid formation increased in water saturated chloroform solutions, it was also
found that irradiation of chloroform solutions by the gamma- rays of Radium
resulted in acid evaluation. 3'a
Numerous chemical systems have been described which undergo chemical
changes on exposure to penetrating x- and gamma- Radiations, such as the
t1
-,.......~-
.., ....
......---
.....
......
..' ....
- ....
-
2-1. Introduction:
The first chemical dosimeters were crude systems which changed color on
exposure to the relatively soft x- rays then being used medically. Barium
platinocyanide was the primary chemical ingredient of pastilles, which were
developed by several investigators. After irradiation, this material was found
to change from its normal green color to orange, and then to various shades
of brown. (3)
The main difficulty encountered with pastilles was that the color reverted to
its original green on exposure to light, further more, the color standards in
the comparator charts gradually faded and thereby contributed to in accurate
measurements. Most important, these systems were highly energy dependent
because they used materials of high atomic number. There fore, changes in
the spectrum of the x-rays caused large variations in doses registered.
A few years later, other chemical systems were developed, for example,
Fruend 6 utilized the effect of x-rays on various iodine compounds. A
mixture of iodoform was found to be sensitive to x- irradiation and was
employed in dosimetry, but the reaction was also produced by head, light
and spontaneous oxidation, and was not proportional to dosage. 3
According to Gruther et al in 1928 chloroform irradiated with x-rays librates
hydrochloric acid in amounts proportional to the Radiation energy absorbed.
Acid formation increased in water saturated chloroform solutions, it was also
found that irradiation of chloroform solutions by the gamma- rays of Radium
1 d · 'd 1 . 34resu te In aCl eva uatlOn. '
Numerous chemical systems have been described which undergo chemical
changes on exposure to penetrating x- and gamma- Radiations, such as the
11
^ J i
oxidation of ferrous to ferric ion compounds and other chemical systems'
Nearly all those chemical systems have been developed and used by
individual working in Radiation chemistry' 3
Physical methods, employing ionization in gases, have been used almost
exclusively for determining the dose of the x and ganlma Radiations utilized
in Radiation thereby and in Radiobiological research.
Because of rapid advancements in electronics, physical methods have
dominated the field of dosimetry, and chemical systems remained in disuse
until quite recentlY.3
2-2 Development of chemical dosimetric systems:
Renewed interest in chemical methods of dosimetry came with the advent of
nuclear reactions, atomic weapons' and the experimental use of increasingly
intense Radiation sources such as the betatron, telecobalt, and Van de Graaff
x-ray generators, and the recent emphasis in Radiobiological researches on
the chemical changes produced in cells and tissue fluids by ionizing
Radiations.3
These developments created an urgent need for dosimeters capable of
registering comparatively large doses of high- energy x- and gamma
Radiations, delivered at tremendously high dose rates up to 105r/min,
furthermore, the dosimeters required by the Armed forces for personnel use
must be small, inexpensive, rugged, thermo stable' direct- reading,
reproducible, and within X20% limits of error responsive to either prompt or
residual bomb gamma Radiation exposure under rigorous field conditions'
T2
- ....
--- ....
--- ...
- ""
-""
- .-
- ....
- ...
- ....
- ....
-,.
- .-
--- ...
--
- ....
oxidation of ferrous to ferric ion compounds and other chemical systems.
Nearly all those chemical systems have been developed and used by
individual working in Radiation chemistry. 3
Physical methods, employing ionization in gases, have been used almost
exclusively for determining the dose of the x and gamma Radiations utilized
in Radiation thereby and in Radiobiological research.
Because of rapid advancements in electronics, physical methods have
dominated the field of dosimetry, and chemical systems remained in disuse
until quite recently. 3
2-2 Development of chemical dosimetric systems:
Renewed interest in chemical methods of dosimetry came with the advent of
nuclear reactions, atomic weapons, and the experimental use of increasingly
intense Radiation sources such as the betatron, telecobalt, and Van de Graaff
x-ray generators, and the recent emphasis in Radiobiological researches on
the chemical changes produced in cells and tissue fluids by ionizing
Radiations. 3
These developments created an urgent need for dosimeters capable of
registering comparatively large doses of high- energy x- and gamma
Radiations, delivered at tremendously high dose rates up to 105r/min,
furthermore, the dosimeters required by the Armed forces for personnel use
must be small, inexpensive, rugged, thermo stable, direct- reading,
reproducible, and within ± 20% limits of error responsive to either prompt or
residual bomb gamma Radiation exposure under rigorous field conditions.
12
a-
t--i--t-T-t:t-t-r_t_
Considerable advanced have been made toward the development of chemical
dosimeter systems and practical dosimeters which can register these
relatively large doses of x- and gamma Radiation. Attention has been
redirected to chemical means of dosimetry because the trend in dose
evaluation is towards the measurement of energy absorption, the chemical
changes produced in tissue equivalent materials should be in many ways
preferable to methods employing iontzation in gases. The yields in aqueous
chemical systems, for example, are not greatly influenced by ambient
temperature changes or wide variations in the energy of the beam (0.1 to 1 .2
MeV) and further more, large integral doses can readily be measured with
dosimeters of small size.
The most difficult problem in chemical dosimetry has been to final stable,
reproducible systems, which register integral doses of x- and gamma-
Radiations in the range below lO0Rad.the ferrous-ferric sulfate system
accepted as the best chemical method for measuring x-ray and gamma ray s
in kilorontgen doses at dose rates up to 1000 Rad/min, and meets most of the
requirements of ideal system. An alternative method with sodium benzoate
or benzene has also been developed. It has advantages that the initial
solutions and the product of irradiation can be kept for longer period without
spontaneous decomposition, both these methods are not sufficiently sensitive
to register galnma- Radiation in the lower part of the dose range that is
biologically most interesting (0 to 1000r). Further more, because of gradual
spontaneous oxidation; the ferrous- ferric system is not sufficiently stable to
permit its instrumentation for long term use. 3
The solution of these two mayor problems in chemical dosimetry has been
accomplished by using systems which respond to Radiation by relatively
long- chain mechanism and by developing methods to control the length of
theses chain reactions while preserving adequate sensitivity in the system.
13
-.............._--
............ --
~._-
-.....i---.- _
r=r=~
~
r=[~
r=(--=
r=
Considerable advanced have been made toward the development of chemical
dosimeter systems and practical dosimeters which can register these
relatively large doses of x- and gamma Radiation. Attention has been
redirected to chemical means of dosimetry because the trend in dose
evaluation is towards the measurement of energy absorption, the chemical
changes produced in tissue equivalent materials should be in many ways
preferable to methods employing ionization in gases. The yields in aqueous
chemical systems, for example, are not greatly influenced by ambient
temperature changes or wide variations in the energy of the beam (0.1 to 1.2
MeV) and further more, large integral doses can readily be measured with
dosimeters of small size.
The most difficult problem in chemical dosimetry has been to final stable,
reproducible systems, which register integral doses of x- and gamma
Radiations in the range below 100Rad.the ferrous-ferric sulfate system
accepted as the best chemical method for measuring x-ray and gamma ray s
in kilorontgen doses at dose rates up to 1000 Rad/min, and meets most of the
requirements of ideal system. An alternative method with sodium benzoate
or benzene has also been developed. It has advantages that the initial
solutions and the product of irradiation can be kept for longer period without
spontaneous decomposition, both these methods are not sufficiently sensitive
to register gamma- Radiation in the lower part of the dose range that is
biologically most interesting (0 to 1000r). Further more, because of gradual
spontaneous oxidation; the ferrous- ferric system is not sufficiently stable to
permit its instrumentation for long term use. 3
The solution of these two mayor problems in chemical dosimetry has been
accomplished by using systems which respond to Radiation by relatively
long- chain mechanism and by developing methods to control the length of
theses chain reactions while preserving adequate sensitivity in the system.
13
2-3. Classification of dosimetrv methods:
There are three ways in which the dose received from a Radiation source
may be determined:
A. Direct:
In this method the energy absorbed is measured as the heat in to which it is
ultimately degraded. Since the method it self elaborate and tedious it is not
used in routine work but only to calibrate more convenient chemical
methods. Calibration is carried out by measuring the rate of change of
temperature with time, in air free water which has reached stationary state,
i.e., no net chemical range. Then the water is replaced by the dosimetric fluid
and the chemical change measured.
A suitable correction must be made for heating of the walls of the container
by electrons which do not enter the solution. This method has the advantage
that it can be applied to wide range of Radiations and is especially useful for
very mixed Radiations such as are obtained from reactors.
B. Semi- direct:
This method dose not involve direct measurement of energy but of some
physical quantity directly related to energy; either the charge-input or the
ionization due to the Radiation.
I-charge-input:
If the number of particles which are completely absorbed is measured then
(numberf s) x (energ,, f particle)Dose rate:
volume or weight in which absorbed
This method is useful for particle beams from accelerators or for B- particles
from external Radiation source. There are however many sources of error.
t4
--.... -.. ..:
.. -
.. -
----
--
-..;
-..;
-...;
--
----.....
2-3. Classification of dosimetry methods:
There are three ways in which the dose received from a Radiation source
may be determined:
A. Direct:
In this method the energy absorbed is measured as the heat in to which it is
ultimately degraded. Since the method it self elaborate and tedious it is not
used in routine work but only to calibrate more convenient chemical
methods. Calibration is carried out by measuring the rate of change of
temperature with time, in air free water which has reached stationary state,
i.e., no net chemical range. Then the water is replaced by the dosimetric fluid
and the chemical change measured.
A suitable correction must be made for heating of the walls of the container
by electrons which do not enter the solution. This method has the advantage
that it can be applied to wide range of Radiations and is especially useful for
very mixed Radiations such as are obtained from reactors.
B. Semi- direct:
This method dose not involve direct measurement of energy but of some
physical quantity directly related to energy; either the charge-input or the
ionization due to the Radiation.
I-charge-input:
If the number of particles which are completely absorbed is measured then
(number/ s) x (energy/particle)Dose rate =
volume or weight in which absorbed
This method is useful for particle beams from accelerators or for B- particles
from external Radiation source. There are however many sources of error,
14
such as back scatter from cell walls, absorption in the windows,
bremsstrahlung current- leakage and beam inhomogeneity.
II- Ionization
The number of ion pair is measured and dose rate : W, number of ion pairs
per unit maSS or volume where "W: mean energy to create an ion- pair in the
substance of interest. "
The disadvantages of this method are:
a- W has often to be assumed and.
b- The results for ionization in the gas phase must be related to those in
the iiquid phase, sine the ionization of liquids can not generally be
measured.
C-Indirect:
Dosimeters of this type all involved measurements of some chemical or
physical effect caused by Radiation and therefore all require calibration by a
method of type A) or B).
r :
1- Physical effects include crystal coloration; include photo-
conductivity and scintillation.
Il-Dosimeters in which a chemical observed are commonly used r
2-4 Requirements of Ideal Chemical Dosimeters:
The ideal dosimeter for measuring high- energy x-ray or gamma ray
Radiations in the dose range between 50 and 5000Y should embody several
essential properties and have certain conditions characteristics; the first
requirement of chemical dosimeter is that the amount of chemical change is
proportional to the dose and independent of the dose rate (r)
15
.. ..:.
• .=
.. ....:
. ....:
. ..;;
- -=
- ..;;
-..:
-~
---
-...;
---
-...;
.. -
such as back scatter from cell walls, absorption in the windows,
bremsstrahlung current-leakage and beam inhomogeneity.
11- Ionization
The number of ion pair is measured and dose rate = Wx number of ion pairs
per unit mass or volume where "W= mean energy to create an ion- pair in the
substance of interest."
The disadvantages of this method are:
a- W has often to be assumed and.
b- The results for ionization in the gas phase must be related to those in
the liquid phase, sine the ionization of liquids can not generally be
measured.
C-Indirect:
Dosimeters of this type all involved measurements of some chemical or
physical effect caused by Radiation and therefore all require calibration by a
method of type A) or B).
1- Physical effects include crystal coloration; include photo
conductivity and scintillation.
II-Dosimeters in which a chemical observed are commonly used' 1
2-4 Requirements of Ideal Chemical Dosimeters:
The ideal dosimeter for measuring high- energy x-ray or gamma ray
Radiations in the dose range between 50 and 5000Y should embody several
essential properties and have certain conditions characteristics; the first
requirement of chemical dosimeter is that the amount of chemical change is
proportional to the dose and independent of the dose rate,(J)
15
8 - J
Thus if G value Radiation chemical yeld is defined as the number of
molecules of product formed or reagent destroyed per 100 eV of energy from
the ionizing Radiation
n( r \G(X) - " ' " ' (Jn i t : mol . j - '
t
Where 4*) : number of molecules or ions formed or destroyed.
€ : the energy imported to the matter of that system.
G- Value should be independent of dose and dose rate.
It is desirable that G- value should be independent of the temperature of
Radiation, quality of Radiation and the presence of air and the chemical
impurities, also the chemical changes should be readily measurable (Note.'
1000 Rad of gamma- rays will produce concentration change of 1 '01 G x 1 0-6
mol/L in a reaction in water for which the G-value is G. In addition to, the
dosimeter should be made form biological tissue equivalent materials in
respect to its density and Radiation absorption properties , should lend it self
to standard production method and have suffrcient stability to give shelf life
of at least one year prior to use, thereby permitting the manufacture of
accurate, highly reproducible dosimeters system. Finally the chemical
dosimeter should be easy to prepare and use.
Not a single dosimeter meets all the above mentioned requirements.
However, we can always select one which satisfied maximum requirements
and has known dependence of on the parameters like (LET) Liner Energy
Transfer, dose rate, temperature, etc.
2-5. Types of Chemical SYstems:
Determination of Ra<liation induced changes in gases by chemical analysis is
difficult technically, thus rendering such systems unsuitable for practical
L6
, J ;
L r i
q - j
-' -l
""'!~.......~........--=~========~~=======------~-------
• --=
... ...:
---
-----'---
•. -.-;
Thus if G value Radiation chemical yield is defined as the number of
molecules of product formed or reagent destroyed per 100 eV of energy from
the ionizing Radiation
O(X) n(x) TT . I .-\=--unzt : mo .}&
Where r(x) = number of molecules or ions formed or destroyed.
G = the energy imported to the matter of that system.
G- Value should be independent of dose and dose rate.
It is desirable that G- value should be independent of the temperature of
Radiation, quality of Radiation and the presence of air and the chemical
impurities, also the chemical changes should be readily measurable (Note:
1000 Rad of gamma- rays will produce concentration change of 1.01 G x 10-6
mollL in a reaction in water for which the G-value is G. In addition to, the
dosimeter should be made form biological tissue equivalent materials in
respect to its density and Radiation absorption properties, should lend it self
to standard production method and have sufficient stability to give shelf life
of at least one year prior to use, thereby permitting the manufacture of
accurate, highly reproducible dosimeters system. Finally the chemical
dosimeter should be easy to prepare and use.
Not a single dosimeter meets all the above mentioned requirements.
However, we can always select one which satisfied maximum requirements
and has known dependence of on the parameters like (LET) Liner Energy
Transfer, dose rate, temperature, etc.
2-5. Types of Chemical Systems:
Determination of Radiation induced changes in gases by chemical analysis is
difficult technically, thus rendering such systems unsuitable for practical
16
l- _:
a . l
-\_
+ _
-\-
chemical dosimetry. Solid systems generally are too insensitive to provide
direct- reading methods of dosimetry of value in Radiation biology or
Radiation thereby, unless the changes included in them are detected by
complicated electronic amplifying devices.
Emphasis is being placed on liquid systems for several reasons. They may be
prepared from reagents which are water- or tissue equivalent in respect to
density and Radiation absorption properties. The Radiation induces reaction
products are relatively stable and can be measured directly by color changes,
or indirectly by simple analyical procedures. Liquid chemical dosimeters
can be prepared in small containers within body cavity, in tumor areas, or at
positions close to high, Intensity sources. Finally aqueous chemical systems
absorb Radiation of various type and energies by mechanisms more like
those occurring in body tissues or fluid than do gaseous or solid systems.
2-6. Liquid aqueous systems :
Principle: In a dosirneter of this type, solute is present which can reacts
stochiometrically with one or more of the primary species whilst in
sufficiently low concentration so as not to influence the rate of energy
deposition.
To understand the mode of action of aqueous dosimeters it is necessary first
to consider the nature and distribution of the species. These differ some what
for Radiations of high and low linear Energy Transfer (LET).
Immediately after the passage of through water of a fast charged particle
those molecules closed to the Radiation track are ionized whilst those further
away are raised to an excited level.
H2O -+ HzO- * e- f HzO*
_ t !
t *
\
t -I
1 1I I
_...:.
..~'..::::
-. ..:.
-~
.. ...:
~.
.. ...:
.....:
..... ...:
--.......
chemical dosimetry. Solid systems generally are too insensitive to provide
direct- reading methods of dosimetry of value in Radiation biology or
Radiation thereby, unless the changes included in them are detected by
complicated electronic amplifying devices.
Emphasis is being placed on liquid systems for several reasons. They may be
prepared from reagents which are water- or tissue equivalent in respect to
density and Radiation absorption properties. The Radiation induces reaction
products are relatively stable and can be measured directly by color changes,
or indirectly by simple analytical procedures. Liquid chemical dosimeters
can be prepared in small containers within body cavity, in tumor areas, or at
positions close to high, Intensity sources. Finally aqueous chemical systems
absorb Radiation of various type and energies by mechanisms more like
those occurring in body tissues or fluid than do gaseous or solid systems.
2-6. Liquid aqueous systems:
Principle: In a dosimeter of this type, solute is present which can reacts
stochiometrically with one or more of the primary species whilst in
sufficiently low concentration so as not to influence the rate of energy
deposition.
To understand the mode of action of aqueous dosimeters it is necessary first
to consider the nature and distribution of the species. These differ some what
for Radiations of high and low linear Energy Transfer (LET).
Immediately after the passage of through water of a fast charged particle
those molecules closed to the Radiation track are ionized whilst those further
away are raised to an excited level.
17
b. -:
\
! , . J
\ ,
I ' :
\-
The Hzo- ion reacts with H2o to give oH' Radicals whilst H2o* may
decompose or be deactivated, all within 1O-ttg. Formerly, it was considered
that the electron reacted rapidly to form a hydrogen atom and a hydroxide
ion, but it is now known that the electron persists up to about 10-as before
reacting in this way. After not less than 10-'1s the electron becomes solvated
by the water molecules and in these conditions is usually represented by
e a 9
H:O* + HzO -+ H2O* + OHo
H2O* -+ Ho + OHo
In the period i0-lrs to 10-7s there is competition between combination of
Radicals to give "molecular products" and diffusive escape from the spur or
track. The recombination reactions Ho * OHo-+ HzO, or e- + OHo -> OH
are omitted because they lead to no chemical change in a solute, combination
reactions are:
e-uo *e uo ---+ H2 + 2OH-
Ho + Ho _-+ H2
e-ro+ Hu--- H: + OH-
OHo + OHo --' H:Oz
Radiation of high LET favors the formation of molecular products over
diffusive escape and also the intra -track reactions:
OHo+ H2O2* H2O+ H2O
e-uq* HzOz-- OH-+ OHo
Since most of the "molecular products" have been produced after the lapse of
10-r7s the Radical combination process will not be interfered with by solutes
which can react with e- or OH or both, provided the half-lives of these latter
reactions are larger than about l0-7s. These half-lives vary with nature and
g :
^.
u t :
--_t-_
w :
\
u :
3/ ,:
t :
I
'!r r -:
lcr :
lr, .:
\ . . :
L, .:
t , :
t.r-t . j
t,-"r:t:t:
18
E .
... .:..
.. ~
.. ...:.
-~
... ..:
--~
~, -
_....a..1
r[~
[-~
. ..::
[~
[-~
, ..,
L
The H20+ ion reacts with H20 to give OHO Radicals whilst H20* may
decompose or be deactivated, all within 10-12g. Formerly, it was considered
that the electron reacted rapidly to form a hydrogen atom and a hydroxide
ion, but it is now known that the electron persists up to about 10-4s before
reacting in this way. After not less than 10-IIS the electron becomes solvated
by the water molecules and in these conditions is usually represented by
e aq
H20+ + H20 ~ H20+ + OHo
H20* ~ HO + OHo
In the period 10-1JS to 10-7s there is competition between combination of
Radicals to give "molecular products" and diffusive escape from the spur or
track. The recombination reactions HO + OHo~ H20, or e- + OHo ~ OH
are omitted because they lead to no chemical change in a solute, combination
reactions are:
OHO + OHo ~ H20 2
Radiation of high LET favors the formation of molecular products over
diffusive escape and also the intra -track reactions:
OHo+ H20 r -+ H20+ H20
e-aq+ H202~ OH-+ OHo
Since most ofthe "molecular products" have been produced after the lapse of
10-17s the Radical combination process will not be interfered with by solutes
which can react with e- or OH or both, provided the half-lives of these latter
reactions are larger than about 10-7s. These half-lives vary with nature and
18
J
concentration of solute but frequently the rate constant of the bimolecular
reaction of reactive solutes with these Radicals is about 1010M-rs-t so that, if
- l 0 - r0l0_ , (
^ , (= l0_ ,sso lute '
The G vaiue of the primary species may be treated as being sensibly
constant. The time scale of these processes is illustrated in table (2).
Therefore, we usually write the following reactions fin stochiometry
equations G "H") is commonly written for G (e ) + G (H) because H and e-uo
are frequently stochiometrically equivalents. 1
2-7. Prerequisites of chemical dosimetry:
Radiolytic reactions are extremely sensitive to trace impurities. Hence
water, glassware's and inadiation cells used for dosimetry must be free from
trace of amount of organic and inorganic impurities 2
T9
Table (2): Events in the radiolysis of aqueous solutions
Time(s)Events in the aqueous solutions
-10 - ' Ionization and excitation. HrO_+HrO-+ s- + H,O".4 x10 - ' Formation of the hydroxyl Radical; HzO- + HzO-+HzO- + OH.- 10 - ' ' Dissociation of H"O" ---+ H+OH.-10 - " Solvation of the electron: e ---+ e- 10-''/trr Relaxation time of the ion- atmosphere of e- aq in an aqueous solution of
ionic strength:p-10 - ' Combination; (2e- ae r Hz + 2OH'; euo-* H-+ H2 and 2OH -- H2O2). And
Reactions (e-uo*OH---+OH-andH+OH--HzO) .reactions virtually completein low LET systems. Intra lack reactions (OH+ H2O2---+ HzO + H2O and e-uo+ HzOz ---OH+ OH-). In high LET systems will also be complete in aboutthis time.
-10- '" / [s] Time in which 50oh of Radicals will have reacted with reactive solute Spresent in concentration IS].
concentration of solute but frequently the rate constant of the bimolecular
reaction of reactive solutes with these Radicals is about 1OloM-lS-l so that, if
10-10
10-7( (>::::, 10-5ssolute
The G value of the pnmary speCIes may be treated as being sensibly
constant. The time scale of these processes is illustrated in table (2).
Therefore, we usually write the following reactions [in stochiometry
..:: equations G "H") is commonly written for G (e-) + G (H) because Hand e-aq
are frequently stochiometrically equivalents. 1
Table (2): Events in the radiolysis of aqueous solutions
Events in the aqueous solutionsTime(s)
~1O- Ionization and excitation, H20~H20++ e- + H2O°.4 xlO-1
'> Formation of the hydroxyl Radical; H20' + H20~H20 +OH.~lO-u Dissociation of Ho0° ~ H+OH.~10-11 Solvation of the electron; e- ~ e- aa.~lO-lU/1l Relaxation time of the ion- atmosphere of e- aq in an aqueous solution of
ionic strength=1l~1O-/ Combination; (2e- aq ~ H2 + 20K; eaq-+ H~ H2 and 20H ~ H20 2). And
Reactions (e-aq+OH~OH-andH+OH~H20) .reactions virtually completein low LET systems. Intra lack reactions (OH+ H202~ H20 + H20 and e-aq+ H20 2~OH+ OH} In high LET systems will also be complete in aboutthis time.
~lO-lU/[S] Time in which 50% of Radicals will have reacted with reactive solute Spresent in concentration [S].
2-7. Prerequisites of chemical dosimetry:
Radiolytic reactions are extremely sensitive to trace impurities. Hence
water, glassware's and irradiation cells used for dosimetry must be free from
trace of amount of organic and inorganic impurities 2
19
l -
II
t _
2-7-1. Purification of water:
Commercially available distilled water contains traces of organic and
inorganic impurities which have been shown to cause significant variations
in the response of these systems. Pure water is necessary for reproducible
results. Distilled water is purified by redistillation from alkaline
pennanganate and from sulfuric acid followed by a third redistillation and by
collection of the condensed steam into closed Pyrex glass containers. This
procedure gives a supply of water of low conductivity which is sufficiently
free of all impurities. This pure water is used in the preparation of other
reagents and for final rinsing of all glass ware.
2-7-2 Treatment of glass wares and irradiation cells:
Glass wares and irradiation cells used for dosimetry arc generally made from
Borosilicate or Silica glass, they are filled with or dipped in 1:1 mixture of
concentrated sulphuric and nitric acids for 24 hours, then washed
successively with tap water and distilled water. Then they are filled distiiled
water and exposed to a dose of approximately lKGy.
Inadiation cells purified in this way are kept filled with distilled water or
with dosimetric solution. when not in use. '
2-7-3. Treatment of plastic irradiation cells:
Use of material like Teflon, Perspex, ploy ethylene, ploy styrene and poly
propylene as container for a chemical dosimeter require a special procedure
for cleaning, storage and use. The inner walls of the inadiation tube should
be inert. i.e. it should not liberate any impurities during irradiation or pre and
post irradiation storage of dosimetric solution. The procedure followed as
follows:
20
..
... -
...
...
... -
... -
.... -
--
--
..... -
-. -
- -
--
L~
2-7-1. Purification of water:
Commercially available distilled water contains traces of orgamc and
inorganic impurities which have been shown to cause significant variations
in the response of these systems. Pure water is necessary for reproducible
results. Distilled water is purified by redistillation from alkaline
permanganate and from sulfuric acid followed by a third redistillation and by
collection of the condensed steam into closed Pyrex glass containers. This
procedure gives a supply of water of low conductivity which is sufficiently
free of all impurities. This pure water is used in the preparation of other
reagents and for final rinsing of all glass ware.
2-7-2 Treatment of glass wares and irradiation cells:
Glass wares and irradiation cells used for dosimetry are generally made from
Borosilicate or Silica glass, they are filled with or dipped in 1: 1 mixture of
concentrated sulphuric and nitric acids for 24 hours, then washed
successively with tap water and distilled water. Then they are filled distilled
water and exposed to a dose of approximately lKGy.
Irradiation cells purified in this way are kept filled with distilled water or
with dosimetric solution, when not in use. 2
2-7-3. Treatment of plastic irradiation cells:
Use of material like Teflon, Perspex, ploy ethylene, ploy styrene and poly
propylene as container for a chemical dosimeter require a special procedure
for cleaning, storage and use. The inner walls of the irradiation tube should
be inert. i.e. it should not liberate any impurities during irradiation or pre and
post irradiation storage of dosimetric solution. The procedure followed as
follows:
20
E-I
l r -
!
! F
I
r i -II
\-
! , - -
L
- -
I -
- _
t -
- -
l -
b s
t
\ - -
l -
t
Clcaning: Plast ic tubes afc f i l lecl u, i th l0% I INIOT I 'or 24 hours. - fhen
they arc l inscc l w i th tap watc r i lnc l then c l i s t i l l ec l rva tc r ' . ' l ' o rna l<c the tubc
free fi 'otr any fr,rrther clctachable irrpLrrity lronr the inner lvall, the tLrbes
are f i l lecl wi th c losimetr ic soh-r t ion ancl exposecl to a c lose ol ' .10 Gy.
Storagc. 'Plast ic t t rbes or bags are I l l lcc l r . r , i th c losimctr ic solut iorr n,hcrr
not ir-r use.
2-7-4. IJse of i r rad ia t ion cc l ls :
When glass or plast ic i rracl iat iorr cel ls are to be usecl fbr dosirnetry. thc
o ld dos imctr ic so lu t ion in thern is c l iscarc lec l . ' fhe ce l is are r insed at least
two times with liesl-r closirnetric solution. I-hey are then ll l led rvith tht:
solut ion, stopper arrcl i rracl iatecl. 2
2-8 . S tab i l i z ing Age n ts :
Al l mater i i i ls shor,r lc l bc reagent gracle or analyt ical ly pr-rre substances, so
statecl by the tnanulactr , r ret ' . I rur thcr pr-rr i l rcat ion is nracle Lry l t 'ac: t ic lnal
rec l i s t i i l a t ion anc l /o r by rcc rys ta l l i za t ion . A l l reagents shoLr lc l be l i cp t in
clean, c losecl , I )yrex containers. ' I 'he rcagcrr t rvhich have been l 'ouncl tcr
provide aclequate thernral stabi l i ty ancl s iurul taneously to rc i tc lcr '
chloro{bnn ancl / or tetrachloroethylene- c lye systc.nr lbr exanrple relat ivcl) '
c lose- rate ancl ternperature inciepenclent. l
2 -9 Daf in i t ion o f C va luc :
G va lue l {ac l ia t ion chcnr ica l y ic ld i s thc l ' L rnc l ln rcn la l quunt i t r t i vc
character ist ic o{ ' l {acl iat ion inclr-rce c l re act ion in a chcmical c losin-re ler . The
react ior-r chernical y ie lc l , G 1. ; o l 'an ent i ty, x, is the c1r-rot ient of 'n(X) b1'r ,
where n(x) is t l ' ie mean amolrnt of sr-rbstance of that ent i ty proclucecl ,
destroyed or changed in s1,stem b1, t l ie eticrgl, inrpartecl,
t-
rt"-t_r-r
2I
~
i...--
-c
-~
-.
----~--
.. -
.. -
.. -
il.. -
l -
-L-
r=r=c[~
['-
Cleaning: Plastic tubes are filled with 10% HN03 for 24 hours. Then
they are rinsed with tap water and then distilled water. 'ro make the tube
free from any further detachable impurity from the inner wall, the tubes
are filled with dosimetric solution and exposed to a dose of 40 Gy.
Storage: Plastic tubes or bags are filled with dosimetric solution when
not in use.
2-7-4. Use of irradiation cells:
When glass or plastic irradiation cells are to be used for dosimetry, the
old dosimetric solution in them is discarded. The cells are rinsed at least
two times with fresh dosimetric solution. They are then filled with the
solution, stopper and irradiated. 2
2-8. Stabilizing Agents:
All materials should be reagent grade or analytically pure substances, so
stated by the manufacturer. Further purification is made by fractional
redistillation and/or by recrystallization. All reagents should be kept in
clean, dosed, Pyrex containers. The reagent which have been found to
provide adequate thermal stability and simultaneously to render
chloroform and/ or tetrachloroethylene- dye system for example relatively
dose- rate and temperature independent. 3
2-9 Definition of G value:
G value Radiation chemical yield is the fundamental quantitative
characteristic of Radiation induced reaction in a chemical dosimeter. The
reaction chemical yield, G (x) of an entity, x, is the quotient of n(X) by;"
where n(x) is the mean amount of substance of that entity produced,
destroyed or changed in system by the energy imparted,
21
t['L , .
J
l E .
^L
-A
l L l -II
-A
j-r-
l - -
-L
l -
L
\ -
JIL , -
ll
I\ _
_t.-
I*
,A-
i\*
l-
I^^
L .
e , t o the t l l a t t e l . o f ' t ha t sys te ln , t hus :
G(X) = n(x) f r : (Jn i t : n to l , /
O r i t i s c le f i nec l as thc r t L r t r t be l o l ' i ons , 1 r ' cc I {a r l i ca l s , u ton rs , r . } r ( ) l c r ru l cs
fbrrned or c l isapPearec l \vhcn thc syst .enr has absor-bcc l 100 ev,o l -e .e l .g '
f t "orn the ion iz ing rac l ia t ion 2 .
22
~ -
~ -
... -
.. -
L
L
,-
E, to the matter of that systcm, thus:
G( X) = n(x)/dJnir : mol.! I
Or it is defined as the number of ions, frcc R~ldic~J!s, ~ltoms; molecules
formed or disappeared when the system has ~lbsorbcd 100 ev of energy
from the ionizing radiation 2 .
22
l -
-t.l.aI
t . -
1-
l _
-I
l -
1-
l _
.t-
l -
THAPTERTHREE)- 3. TYPES OF AQUEOUS CHEMICAL DOSIMETERS
l-
t _
\ _
! -
l . '
A-
.. -
l _
IHAPTER THREE~ ---L.
l _
l _
l _
.. -
... -
.. -
. -
3- TYPES OF AQUEOUS CHEMICAL DOSIMETERS
-..L.- .
L I
; :
l , r
t r
t ]
3-1. Fricke dosimeter:
This dosimeter "ferrous- ferric sulfate system", which described by Fricke
and Morse then developed by Miller t, is presently accepted by numerous
workers as the best chemical method for measuring x and garnma rays, and it
meets most if the requirements of the ideal system'(3)
This dosimeter is the best one for low medium dose-rate. It is based on the
oxidation of ferrous to ferric ions in aerated 0.8N H2SO4 and the ferric ion
concentration may be measured as follows:-
i) By measuring of light absorption at 3050 A'. The extinction :
(2200t9) M-r at 20'C and show the considerable temperature
dependence, namely @e I e)dT :0.007.
ii) Alternatively the ortho-phenanthroline complex with Fe*2 may be
formed and its light absorption at 5100 A0 measured: -11000 M-lcm-r.
iii) Other techniques involve potentiometric titration, labeling and
measurement of the light absorption of the Fe SCN*2 complexes at
4600 A0.
3-1-1. Principle:
When an air saturated dilute ferrous sulfate or ferrous ammonium sulfate in
solution in 0.8 N H2SO4 is exposed to ionizing Radiation, ferrous ions are
oxidized to ferric ions in a series of quantitative and irreversible reactions
Fe 2* -+ Fe3*
Concentration of ferric ions gives the measure of absorbed dose. The
concentration is measured by spectrophotometric analysis means which is
generally more convenient than chemical titration methods for it is rapid,
l *
;
t--t-_t^tt.r_itii-_i_t+
I
23
l .--
l _
~
~
~
~
r:r:~
~
l~
~;---.
3-1. Fricke dosimeter:
This dosimeter "ferrous- ferric sulfate system", which described by Fricke
and Morse then developed by Miller 5, is presently accepted by numerous
workers as the best chemical method for measuring x and gamma rays, and it
meets most if the requirements of the ideal system-(3)
This dosimeter is the best one for low medium dose-rate. It is based on the
oxidation of ferrous to ferric ions in aerated 0.8N H2S04 and the ferric ion
concentration may be measured as follows:-
i) By measuring of light absorption at 3050 Aa. The extinction
(2200±9) M-J at 20°C and show the considerable temperature
dependence, namely (dEI E)dT = 0.007.
ii) Alternatively the ortho-phenanthroline complex with Fe+2 may be
formed and its light absorption at 5100 Aa measured: ~11000 M-Jcm- I.
iii) Other techniques involve potentiometric titration, labeling and
measurement of the light absorption of the Fe SCN+2 complexes at
4600 Aa.
3-1-1. Principle:
When an air saturated dilute ferrous sulfate or ferrous ammonium sulfate in
solution in 0.8 N H2S04 is exposed to ionizing Radiation, ferrous ions are
oxidized to ferric ions in a series of quantitative and irreversible reactions
Fe 2+ -+ Fe3+
Concentration of ferric ions gives the measure of absorbed dose. The
concentration is measured by spectrophotometric analysis means which is
generally more convenient than chemical titration methods for it is rapid,
23
t *
L !
i. _-
l q
L !
II
l . !
acQurate, and expedient for the analysis for low ferric concentration andsmall quantities of solution.(1)
3-l-Z Basic reactions:
s -uc+Huq-Ho (3 l )
In 0'8N HzSo+, this reaction is complete in less than 10-es. In deaeratedsolution at pH<1 and in the absence of organic impurities the onlyparticipating reactions are:
H+ Fe2* *HzO - Hz*oH +Fe3* (3.2)
oH'+ Fe 2* __- oH-+ Fe3* (3 3)
H2O'+ Fe2n --- OH + OH- + Fe3* (3.4)
HO2+Fe2* * Hoz-* Fe3* (3.5)
Hence for time (t) : co
G 1r'":*y- : Gs *Gon -t2Gnzoz*3G'oz.
G 1u zy: GHz*Gn
Reaction ( a ) is slow compared with the others when t: 0G 1. . ' * )o:Gu * Gou* G noz
In practice the half -life of reaction (3.a) may be increased by using verydilute soiutions of ferrous sulphate and thus G 1re+:)0 measured. In an aeratedsolution reaction (3.2) is completely superseded by reaction (3.6) followedby reaction (3.5)
U + Oz -) HOz (3.6)
Hence:
G 1n. :*;- : 3Gn +Gos+3Gnozt2GHzOz
c (Hz) :G (Hz)
G 1E.:*; o :GH*Gon*Gsoz
3-1-3. Calculation of the absorbed dose..
t:It-tlt-fIIt.it_r-L *
[ _
-! *
I
24
l .!!
accurate, and expedient for the analysis for low ferric concentration andsmall quantities of solution. (l)
3-1-2 Basic reactions:
- - 0e aq+ H aq ~ H (3.1 )
Hence:
G (Fe3+) 0=GH+GOH+GH023-1-3. Calculation of the absorbed dose:
G (Fe 3+)00 = 3GH+GoH+3GH02+2GH202
G (H2) =G (H2)
(3.2)
(3.3)
(3.4)
(3.5)
H+ Fe2++H20 ---+ H2+OH-+Fe3+
OHo+ Fe 2+ ~ OH-+ Fe3+
H20 2+ Fe2+~ OHo+ OH- + Fe3+
H02+Fe2+ ---+ H02-+ Fe3+
G (Fe 3+)00 = GH+GOH +2GH202+3GH02,
G (H 2) = GH2+GH
Reaction ( & ) is slow compared with the others when t= 0
G (Fe 3+)0 =GH+ GOH + G H02
In practice the half -life of reaction (3.4) may be increased by using verydilute solutions of ferrous sulphate and thus G (Fe+3)0 measured. In an aeratedsolution reaction (3.2) is completely superseded by reaction (3.6) followedby reaction (3.5)
Hence for time (t) = (fJ
In 0.8N H2S04 , this reaction is complete in less than 10-9s. In deaeratedsolution at pH<l and in the absence of organic impurities the onlyparticipating reactions are:
r----
l A
r~
~
~
~
LLr[
[
[
rr
24
GyM
-,
--
l-
I
I
!
I - -
L-
!
!
Ferric ion has two absorption peaks at 3004 nm and 224 nm, the one at 224
nm being more intense, the sensitivity of this analyical method can be
doubled by measuring the optical unit density at 224 nm ferric ion peak.
Another advantage of measuring at 224 nm peak is that the molar linear
absorption coefficient increase with rising temperature by only 0.13% C.
However, at this lower wavelength, the absorption due to ferrous is not
whol ly negl ig ib le.
In addition, impurities from plastic container are reported to be more
troublesome at 224nm. Hence 304nm is used for measurement of Fe*3 ion
concentration, at temperature 25 oC.
Molar absorption coefficient should be determined by measuring of optical
density of solutions of different known concentrations of ferric ions.
Absorbance of irradiated dosimeters is measured against the Inadiated
solution form the standard flask, optical density of the unirradiated reference
blanks in the dosimetric tubes is also measured against solution in standard
flask. Absorbance of the irradiated solution is corrected for reference blank
reading. The absorbed dose is calculated from the following equation 2:
D_Pe I Gu,
Where:
M
€
I
: the difference in absorbance (optical
irradiated and unirradiated solution.
: molar linear absorption coefficient.
: optical path length 0.01 m
density) between the
p : density of solution: 1024 Kg m3
G(' , : G 10.* ' ) : 1 .61 x 10-6 mol J- l .
Substituting the above values gives:
25
-
"
J-
Ferric ion has two absorption peaks at 3004 nm and 224 nm, the one at 224
nm being more intense, the sensitivity of this analytical method can be
doubled by measuring the optical unit density at 224 nm ferric ion peak.
Another advantage of measuring at 224 nm peak is that the molar linear
absorption coefficient increase with rising temperature by only 0.13% C.
However, at this lower wavelength, the absorption due to ferrous is not
wholly negligible.
In addition, impurities from plastic container are reported to be more
troublesome at 224nm. Hence 304nm is used for measurement of Fe+3 ion
concentration, at temperature 25°C.
Molar absorption coefficient should be determined by measuring of optical
density of solutions of different known concentrations of ferric ions.
Absorbance of irradiated dosimeters is measured against the Irradiated
solution form the standard flask, optical density of the unirradiated reference
blanks in the dosimetric tubes is also measured against solution in standard
flask. Absorbance of the irradiated solution is corrected for reference blank
reading. The absorbed dose is calculated from the following equation 2:
D= M GyP& I G(x)
Where:
M = the difference in absorbance (optical density) between the
irradiated and unirradiated solution.
& = molar linear absorption coefficient.
= optical path length 0.01 m
p = density of solution= 1024 Kg m-3
G(x) = G(F/3) = 1.61 x10-6 mol r l
.
Substituting the above values gives:
25
rb,-
--
{
\a-
a-
\
L-
\,
\,
I
I
L
tt
L
n _ 0.0977 LA_ _
G lFe'* lx e\ / o t 2 5 " C
3-l-4. Influence of various factors :
I) Oxygen Supply:
It is necessary that the Fricke dosimeter should have an adequate supply ofoxygen since, when all the oxygen has been used up H can no longer formH2o and G (Fe*3)is conespondingly diminished, when oxygen is completelyexhausted ,reaction mechanism change reducing G(Fe ,*) to 0.g5 Mol J-.
Fe 2* +H" + H* -+ Fe3* + 11,
Here one atom oxidizes only one ferrous ion to ferric instead of three.Therefore, G (Fe*3) :2 GH26,* esFr
II) Ferrous ion concentration:
For concentration of ferrous ion < l0-1 M reaction (3.4) becomes slow inrelation to irradiation time, so that if the analyical procedures are rapid themeasured G (Fe*3) witl lie between G (Fen3)6and G (F"*r)_. Solutions of thisconcentration are also inconvenient in that all the ferrous ion will be destrovby a dose of 7000 Rad. (r)
III) Sulphuric acid concentration:
The concentration of 0.4M is always used; chemically all that is necessarv isthat the pH of the solution shall be less than 1.5.
L
26
w
...
-----
\...
t.
D = O.0977MG (Fe 3
+ )X £ at 2SoC
3-1-4. Influence of various factors:
I) Oxygen Supply:
It is necessary that the Fricke dosimeter should have an adequate supply ofoxygen since, when all the oxygen has been used up H can no longer formH20 and G (Fe+3)is correspondingly diminished, when oxygen is completelyexhausted ,reaction mechanism change reducing G(Fe 3+) to 0.85 Mol f.
Fe 2+ +HO
+ H+ --...., Fe3++ H2Here one atom oxidizes only one ferrous ion to ferric instead of three.
Therefore, G (Fe+3) = 2 GH202+ GOH
11) Ferrous ion concentration:
For concentration of ferrous ion < 10-4 M reaction (3.4) becomes slow inrelation to irradiation time, so that if the analytical procedures are rapid themeasured G (Fe+3) will lie between G (Fe+3)oand G (Fe+3}Xl. Solutions of thisconcentration are also inconvenient in that all the ferrous ion will be destroyby a dose of 7000 Rad. (1)
Ill) Sulphuric acid concentration:
The concentration of O.4M is always used; chemically all that is necessary isthat the pH of the solution shall be less than 1.5.
26
-
-.L
rI
-,4-
-L
-.4-
-./-
-.4-
-,/-
0.4 M sulphuric acid solution is used because Fricke consider that the
electron concentration of this medium would approximate to that of the
living cell.
IV) Purity of reagents and cleanliness of glass ware:
Organic substances, RH, compete with ferrous for OH Radicals
OHo+ B11 - HzO + Ro
And the resultant organic Radical, Ro, may either react with 02 to give Ro2'
which oxidizes three ferrous ions, so that G (Fe'*) increases.
Ro + Oz- ROzo
e .g. RH: alcohol.
RO2o+Fe 2* -) Fe3* +ROz-
ROOH+ Fe2' -+ pe3+ +ROO+OH-
ROo + Fe 2* -+ Fe 3* +RO--+ ROH
or reduce Fe3n when G 1Fe3n)- decreases. The former predominates when
RH is alcohol and the latter when RH is allyl thiourea, or styrene.
A test of organic impurities can be made by adding chloride ion (-10 3- M
NaCl) when the rapid reaction:
OH" +Cl -+OH- +Clo
occur, since chloride atoms react much more rapidly with Fe2* than RH, any
change of G(Fe3*) caused by addition of chloride is indicative of the
presence of organic impurity. Great care is necessary to ensure that pure
water and pure chemicals reagents are used in very clean glass containers 1
V) Effect of Dose- rate:
Irrrrr
27
• L
• L
· "
· .
.. 1
0.4 M sulphuric acid solution is used because Fricke consider that the
electron concentration of this medium would approximate to that of the
living cell.
IV) Purity of reagents and cleanliness of glass ware:
Organic substances, RH, compete with ferrous for OH Radicals
OHO+ RH ~ H20 + RO
And the resultant organic Radical, RO, may either react with O2 to give R02°
which oxidizes three ferrous ions, so that G (Fe3+) increases.
RO '+ 02~ R02°
e.g. RH = alcohol.
R02o+Fe 2+ ~ Fe3++R02-
ROOH+ Fe2+ ~ Fe3++ROo+OH-
RO° + Fe 2+ ~ Fe 3+ +RO-~ ROH
Or reduce Fe3+ when G (Fe3+)oo decreases. The former predominates when
RH is alcohol and the latter when RH is allyl thiourea, or styrene.
A test of organic impurities can be made by adding chloride ion (~l 0 3- M
NaCl) when the rapid reaction:
OHO +Cl -~OH- +Clo
Occur, since chloride atoms react much more rapidly with Fe2+than RH, any
change of G(Fe3+) caused by addition of chloride is indicative of the
presence of organic impurity. Great care is necessary to ensure that pure
water and pure chemicals reagents are used in very clean glass containers 1
V) Effect of Dose- rate:
27
At dose-rate of about 108 Rad,4nin Radicals from adjacent tract with the
solute and as consequence G (Fe'*) diminishes as the dose-rate is further
increased. Practical
a. High LET, e.g. Po particle - up to 1020 e v 1-1 S-r
b. Low LET, e.g. X- or r- rays- up to 1026 e v 1-r S I
VI) Effect of linear Energy Transfer (L.E.T):
From the mechanism of energy loss one would be expect no great difference
in the number of e-uo , Ho and OHo produced by ionizing particles per 100
eV, as either phase or particle is changed. However, as the LET is altered by
changing the energy or mass of the particle or by changing from water vapor
to liquid, also the true G-s26 is constant the fraction of e-,q, Ho and OHo
escaping will vary between limits of 100% and 0o/o it is equivalent to saylng
that the molecular product yields increase, tending to 112 G-Hzo whilst Gg
and G6s decrease to zero it follows that G 1F*3; will decrease as LET
increase.
VII) Temperature:
By increasing temperature during irradiation G(F*') increase slightly The
temperature at which the spectrophotometric measurements are dose is very
important as the molar linear absorption coefficient increase by about 0.69%
per C for the 304 nm peak. For this reason it is important to know the
temperature at which the spechrophotometric measurements are done.
3-2 Ceric Sulphate (Ceric- Cerous) dosimeter.'
This dosimeter makes use of the reduction of ceric sulphate of by Radiation.
This dosimeter is standardtzed against Fricke dosimeter. With great care
precision of + lo/o can be obtained. (1)
28
At dose-rate of about 108 Rad linin Radicals from adjacent tract with the
solute and as consequence G (Fe3+) diminishes as the dose-rate is further
increased. Practical
a. High LET, e.g. Po particle - up to 1020 e v 1-1 S-l
b. Low LET, e.g. X- or r- rays- up to 1026 e v 1-1 S-l
VI) Effect of linear Energy Transfer (L.E.T):
From the mechanism of energy loss one would be expect no great difference
in the number of e-aq , HO and GHO produced by ionizing particles per 100
eV, as either phase or particle is changed. However, as the LET is altered by
changing the energy or mass of the particle or by changing from water vapor
to liquid, also the true G-H20 is constant the fraction of e-aq , HO and GHO
escaping will vary between limits of 100% and 0% it is equivalent to saying
that the molecular product yields increase, tending to 1/2 G-H20 whilst GH
and GOH decrease to zero it follows that G (F+3) will decrease as LET
mcrease.
VII) Temperature:
By increasing temperature during irradiation G(F+3) increase slightly The
temperature at which the spectrophotometric measurements are dose is very
important as the molar linear absorption coefficient increase by about 0.69%
per C for the 304 nm peak. For this reason it is important to know the
temperature at which the spechrophotometric measurements are done.
3-2 Ceric Sulphate (Ceric- Cerous) dosimeter:
This dosimeter makes use of the reduction of ceric sulphate of by Radiation.
This dosimeter is standardized against Fricke dosimeter. With great care
precision of± 1% can be obtained. (1)
28
3-2-l Principle:
when dilute aerated solution of ceric sulphate in 0.gN H2so4 is exposed to
ionizing Radiation, ceric ions are reduced to cerous. Decrease in ceric ions
concentration is measured spectrophotometrically to evaluate the absorbed
dose.
Ce4* ---) Ce3*
The basic reactions dosimeters are about + 3oA. Are dosimeters is about +
3%.
H (or HO2) + Ce ---) H*(or ,1n + Oz) * Ce3* (3.7)
H2o, + Cea* ----) H*+ Ho2 + Ce3* (3.g)
OH * Ce3* -) OH- + ge4* (3.9)
G (Ce3*) = GH + 2GH2O2_ GOH_ (3. 10)
The yield of Ce is independent of oxygen since H hand H2o both reduce
ceric ions, the stoichiometry equations indicates that G(ce) is largelydetermined by the magnitude of 4 Gn:oz- 2Gnzand is therefore:
a. Strongly dependent on LET
b' Small and consequently very useful for measurement of hieh dose-rate.
3-2-2 Calculation of absorbed dose:
when the dosimeter is used for calibration pu{pose, spectrophotometry isused to measure the decrease in ceric ion concentration .as the starting Ce*3
ion concentration is high, both unirradiated and irradiated solutions have to
be diluted to obey Beer Lambert's law .ceric ions have an absorption
maximum at 320nm .the molar linear absorption coefficient e (ce*1) is
561mzmo1-Iat 25c0 e (ce*a) at this wavelength is only 0.27m2 mol-r.so
interference to cerous ions can be neglected.
The absorbed dose is then calculated as follows 2:
29
3-2-1 Principle:
When dilute aerated solution of ceric sulphate in 0.8N H2S04 is exposed to
ionizing Radiation, ceric ions are reduced to cerous. Decrease in ceric ions
concentration is measured spectrophotometrically to evaluate the absorbed
dose.
C 4+ C 3+e ---+ e
The basic reactions dosimeters are about ± 3%. Are dosimeters is about ±
3%.
H (or H02) + Ce ---+ H+(or H+ + O2) + Ce3+ (3.7)
H20 2 + Ce 4+ ---+ H++ H02 + Ce3+ (3.8)
OH + Ce3+ ~ OH- + Ce4+ (3.9)
G (Ce3+) = GH + 2GH20 2 - GOH- (3.10)
The yield of Ce is independent of oxygen since H hand H20 both reduce
ceric ions, the stoichiometry equations indicates that G(Ce) is largely
determined by the magnitude of 4 GH20r 2GH2 and is therefore:
a. Strongly dependent on LET
b. Small and consequently very useful for measurement of high dose
rate.
3-2-2 Calculation of absorbed dose:
When the dosimeter is used for calibration purpose, spectrophotometry is
used to measure the decrease in ceric ion concentration .as the starting Ce+3
ion concentration is high, both unirradiated and irradiated solutions have to
be diluted to obey Beer Lambert's law .ceric ions have an absorption
maximum at 320nm .the molar linear absorption coefficient [> (Ce+4) is
56lm2mor lat 25Co[> (Ce+4) at this wavelength is only 0.27m2 morl.so
interference to cerous ions can be neglected.
The absorbed dose is then calculated as follows 2:
29
Dose,Gy - ., -Y,G\Ce'. )t.p.l
Where:
M : charge in absorbance.
^ / ^ r - \u\Le' ) : Radiation chemical yield of cerous ions : 0.25umol J-lor
{2.4( l00eVr- ' }
t :molar linear absorption coefficient of ce*a ions (561m2mol-1
p :density of the dosimetric solution, Kg--'
I : path length of the dosimetric solution within the spectrophotometric cell
3-2-3 influence of various factors:
This system is influenced by the same and specific factors which alter the
response of Fricke system,with following exceptions (a)oxygen content does
not effects the yield; (b) dose rate effects are minimal; (c)trace amounts of
organic impurities produced very erratic yields; and (d) the system has
definite spectral dependence at energies below 100kvp t.
3-3 Other systems:
3-3-1-aromatic solutes :
Many aromatic compounds such as benzene and benzoic acid undergo
hydroxylation on irradiation in aqueous solution and the amount of product
formed can be related to the dose .for calcium benzoate in water the product
is salicylic acid, the concentration of which can be measured bv its
fluorescence at 400040 excited by 2900,4.0 tignt. 1t;The advantage of these dosimeters is that the solution needed not to be
acidic. The disadvantages include:
a- G (PhOH) is low; being of the order of 1.6
b- Products other than phenol are produced.
30
----------_....................................~~~======--===., --.
Dose, Gy = ( 3)GCe + c.p.!
Where:
M = charge in absorbance.
G(Ce3+) = Radiation chemical yield of cerous IOns = O.25/lmol rlor
{2.4(l OOey)-I}
[; =molar linear absorption coefficient ofCe+4 ions (561m2morl
p =density of the dosimetric solution, Kgm-3
I = path length of the dosimetric solution within the spectrophotometric cell'
3-2-3 influence of various factors:
This system is influenced by the same and specific factors which alter the
response of Fricke system ,with following exceptions (a)oxygen content does
not effects the yield; (b) dose rate effects are minimal; (c)trace amounts of
organic impurities produced very erratic yields; and (d) the system has
definite spectral dependence at energies below 100kvp 1.
3-3 Other systems:
3-3-1-aromatic solutes:
Many aromatic compounds such as benzene and benzoic acid undergo
hydroxylation on irradiation in aqueous solution and the amount of product
formed can be related to the dose .for calcium benzoate in water the product
is salicylic acid, the concentration of which can be measured by its
fluorescence at 4000A0 excited by 2900A0 light. (l)
The advantage of these dosimeters is that the solution needed not to be
acidic. The disadvantages include:
a- G (PhOH) is low; being of the order of 1.6
b- Products other than phenol are produced.
30
c- G (PhoH) shows fairly marked dependences on (phH), LET and dose
rate.
d- The system is too inadequately calibrated to be of great use.
3-3-2 Chlorinated hydrocarbon solutes:
These systems (two -phase) are better thermal stability and higher Radiation
sensitivity; these features permit the preparation of stable dosimeters capable
of registering gammaray exposures covering abroad range (1.0 to t06 Rad).
The chemical reactions are complex, of ten involving chain reactions
consequently the G-values are high so that they can be used to measure low
doses, but they are very sensitive to impurities. It has claimed that the
aqueous tetrachloroethylene system can be used to measure doses in the
range 0.5 to 3.0 Rad with20o/o precision. I
3-4 Potential liquid dosimeters:
Generally, there are many liquids which on irradiation generate intermediate
Radicals and excited molecules which can be reacting with solutes having
strong absorption spectra. These solutions should show indirect action and
might yield useful dosimeter after proper calibration.
Iodine can be regarded as a suitable solute in that; its concentration may be
measured spectrophotometrically, but in aromatic solvents back reaction
occurs which militate against its use.r
Also Diphenyl picryl hydrazyl (DPPH) considered as a suitable solute, sinceit has a high excitation coefficient and G (-DpH) is independent of (DppH)
Radical. over considerable range, however, the mechanism is complex and
G is very dependent on the solvent, being 0.76 inbenzene and high as 35 in
chloroform.
Some inorganic solutes which might be used are (FeCl3, cuCl2) in NN-dimethyl form amide I
31
----------------------,
c- G (PhOH) shows fairly marked dependences on (PhH), LET and dose
rate.
d- The system is too inadequately calibrated to be of great use.
3-3-2 Chlorinated hydrocarbon solutes:
These systems (two -phase) are better thermal stability and higher Radiation
sensitivity; these features permit the preparation of stable dosimeters capable
of registering gamma ray exposures covering abroad range (l.0 to 106 Rad).
The chemical reactions are complex, of ten involving chain reactions
consequently the G-values are high so that they can be used to measure low
doses, but they are very sensitive to impurities. It has claimed that the
aqueous tetrachloroethylene system can be used to measure doses in the
range 0.5 to 3.0 Rad with 20% precision. 1
3-4 Potential liquid dosimeters:
Generally, there are many liquids which on irradiation generate intermediate
Radicals and excited molecules which can be reacting with solutes having
strong absorption spectra. These solutions should show indirect action and
might yield useful dosimeter after proper calibration.
Iodine can be regarded as a suitable solute in that; its concentration may be
measured spectrophotometrically, but in aromatic solvents back reaction
occurs which militate against its use. 1
Also Diphenyl picryl hydrazyl (DPPH) considered as a suitable solute, since
it has a high excitation coefficient and G (-DPH) is independent of (DPPH)
Radical. Over considerable range, however, the mechanism is complex and
G is very dependent on the solvent, being 0.76 in benzene and high as 35 in
chloroform.
Some inorganic solutes which might be used are (FeCI3, CuCb) in NN
dimethyl form amide 1
31
3-5 Gaseous chemical dosimeters:
These are less important than the liquid phase t1pe, since the dose rate in a
gaseous system may be obtained from ionization current measurement and
W-values (mean energy to create an ion-pair in the substance of interest) or,
if an internal particle source is used under conditions that the rafige of the
particle is considerably less than the containing vessel dimensions, from the
product of the disintegration rate and the mean energy per particle.
The most useful gaseous dosimeter is the use of nitrous-oxide; the
mechanism of which is:
NzO ---) N + NO or N2 * Q,
N + N2O --+ Nz * NO,
O+N2O -> 2NOorNz+Oz
2NO + Oz
G value is independent of temperature in the range -80 "C to 200 oC the
extent of reaction by measuring the amount of NOz
spectrophotometrically, the disadvantage of this dosimeter is that aback
titration becomes important at dose >104 Rad 1.
3-6 Dosimeters based on solids:
These are all rely on the fact that some intermediate is trapped, or causes
some chemical change in a solute or in the gel material itseli which may
be detected calorimetrically, either directiy or after subsequent chemical
development. Such systems are easy to handle and have the possibility of
being made three dimensional; furthermore, back reactions can only be
slow, because of the immobility of the solutes.
In order to obtain reproducible results such dosimeters must be very
carefully prepared and highly standardized procedures.') Ft<l ,/
N20 ----+
N+N2O ----+
0+N2O ----+
2NO+02 ----+
3-5 Gaseous chemical dosimeters:
These are less important than the liquid phase type, since the dose rate in a
gaseous system may be obtained from ionization current measurement and
W-values (mean energy to create an ion-pair in the substance of interest) or,
if an internal particle source is used under conditions that the range of the
particle is considerably less than the containing vessel dimensions, from the
product of the disintegration rate and the mean energy per particle.
The most useful gaseous dosimeter is the use of nitrous-oxide; the
mechanism of which is:
N + NO or N2 + 0,
N2 +NO,
2NO orN2 + O2
2N02
G value is independent of temperature in the range -80 QC to 200 QC the
extent of reaction by measurmg the amount of N02
spectrophotometrically, the disadvantage of this dosimeter is that aback
titration becomes important at dose>104 Rad 1.
3-6 Dosimeters based on solids:
These are all rely on the fact that some intermediate is trapped, or causes
some chemical change in a solute or in the gel material itself, which may
be detected calorimetrically, either directly or after subsequent chemical
development. Such systems are easy to handle and have the possibility of
being made three dimensional; furthermore, back reactions can only be
slow, because of the immobility of the solutes.
In order to obtain reproducible results such dosimeters must be very
carefully prepared and highly standardized procedures.
32
The optical density-dose relation is linear and thus these systems haveobvious potentialities as dosimeters.l
33
t
~----------------======~~~,
The optical density-dose relation is linear and thus these systems have
obvious potentialities as dosimeters. 1
33
CONCLUSION
Based on this review on development of chemical dosirneters,one can
draw the following conclusions remarks:
1-Chernical dosimetcrs are stil l under developing,but an aqueous
chemical dosirneters are preferable because :
2- 'fhey can be prcpared in small containers within body cavitics which
pennits the actual measurcment of Radiation dosc.
3-They are nrore convenient in Itadiation thcrapy.,'/
4-They rnay preparcd hom reagents which arc watcr or tissue eclr-rivalents
in respect to density and Itadiation absorption properlies.
5-The Radiation induced reaction products are relatively stablc and can
be measured directly by color changes or indirectly by simple analytical
procedures.
34
CONCLUSION
Based on this review on development of chemical dosimeters,one can
draw the following conclusions remarks:
I-Chemical dosimeters are still under developing,but an aqueous
chemical dosimeters are preferable because :
2- They can be prepared in small containers within body cavities which
permits the actual measurement of Radiation dose.
3-They are more convenient in Radiation therapy.
4-They may~repared from reagents which are water or tissue equivalents
in respect to density and Radiation absorption properties.
5-The Radiation induced reaction products are relatively stable and can
be measured directly by color changes or indirectly by simple analytical
procedures.
34
flHAPTERFOUR
4- REFERENCES
IHAPTER FOUR
4- REFERENCES
II.EFtrITENCIIS
l - F.S Dainton: Chemical Dosimctry- I ladiat ion Dosimctry
(G.W.lteed, ed), Acadernic press, Nerv York I-ondon (1964).
2- P.G Benny: Iladiation I)osimctry, Radiation Tcchnology
Development Section, Ilhabha Atornic l{esearch Centre, ' l 'rornbay,
Murnbai (lndia)
3- Gordan I-.8 and Gcrald J. i- I . : l tadiat ion Dosimctry, IJoston,
Massachusetts ( 1956).
4- Gunther, P., Von der FI orst, Il.D., and Cronheirn, C.lt., Z.
Elektrochem.34, 6i6 ( 1928).
5- Mi l le r , N. , J :Chcm.Phys.18,79 (1950) .
6- Freund, L . : Wien K l in .Wochschr . 17,412 ( 1904) .
35
- - ------------------
REFERENCES
1- F.S Dainton: Chemical Dosimetry- Radiation Dosimctry
(G.W.Reed, ed), Academic press, New York London (1964).
2- P.G Benny: Radiation Dosimetry, Radiation Technology
Development Section, Bhabha Atomic Research Centre, Trombay,
Mumbai (India)
3- Gm'dan L.B and Gerald 1.1-1.: Radiation Dosimetry, Boston,
Massachusetts (1956).
4- Gunther, P., Von der H orst, H.D., and Cronheim, C.F., Z.
Elektrochem.34, 616 (1928).
5- Miller, N., J :Chem.Phys.18, 79 (1950).
6- Freund, L.: Wien Klin .Wochschr.17, 412 (1904).
35