The Toshiba FDX4343R X-ray Flat Panel Detector (FPD)
for radiographic applications delivers high sensitivity,
high resolution, low noise and superior image quality.
It has an active area of 43 x 43cm and features excellent
diagnostic images, while delivering the benefits of
lower radiation dose to the patient.
The FDX4343R achieves a new level of
functionality and reliability in
developing functional FPDs for
system manufacturers.
FDX4343RX-Ray FLAT PANEL DETECTOR
Digital Radiography
Excellent Sensitivity & Resolution
Excellent Image Quality
Excellent Reliability
Product Information PI No. PE-FDX4343R 2010-06-25
Sales & Marketing Department1385 Shimoishigami, Otawara-shi, Tochigi, 324-8550 JapanPHONE: +81-287-26-6531 FAX: +81-287-26-6059
http://www.toshiba-tetd.co.jp/
2010-08-1000 FXSS
Toshiba Electron Tubes & Devices Co., Ltd.meets the Environmental Management SystemStandard, ISO14001.
Toshiba Electron Tubes & Devices Co., Ltd.meets internationally recognized Standards forQuality Management System ISO9001,ISO13485.
Characteristics
Dimension
512
50143
495
1
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
MTF
Frequency (Ip/mm)
0 0.5 1 1.5 2 2.5 3 3.5
MTF of FDX4343R
High MTF!
Application General radiography
Scintillator CsI : Tl (direct vapor deposition) + reflective coating
Pixel Pitch 143×143μm
Pixels 3008 (H)×3072 (V) (approx. 9.2 million pixels)
Image Size 16.9 (H)×17.3 inchs (V) (43×43.9 cm)
A /D Conversion 14 bit
Output data 16 bit grayscales tiff
Preview image access time Approx. 4 seconds after X-ray exposure
MTF (2 lp /mm) 36 % typ.
DQE (0) 70 % typ.
Power (Voltage /Frequency) AC100~240V 50/60Hz
Operating environment Sensor Unit : 10-35℃ , 30-80% RH (non-condensing)
Dimensions Sensor Unit : 512×495×43mm
Weight Sensor Unit : 16kg
Standard components Sensor unit, Power unit, Cables
Certification •IEC60601-1,IEC60601-1-2 •MDD93/42/EEC(CEmarking)
FDX4343R Specification
Excellent Sensitivity & Resolution◦ Toshiba’s advanced and proven fine structured CsI/Tl and direct vapor
deposition technology deliver higher sensitivity and better resolution.◦ Reflection coating on CsI/Tl screen enables excellent Detective
Quantum Efficiency (DQE) and low noise.◦ A high Modulation Transfer Function (MTF) FPD is the result of these
outstanding technologies.
Excellent Image Quality◦ Achieves a raw image with a low-noise through the use of Toshiba’s
own circuit technology.◦ Images have a shorter lag time with an improved refresh function,
suitable for radiographic double exposure.◦ Driver software (attached to the product) provides excellent, clear
raw images, adopting real-time offset, gain and defect corrections.
Excellent Reliability◦ Excellent durability by using CsI/TI screen direct vapor deposition
method.◦ The structure is highly reliable and protected from degradation due to
the use of a unique moisture-proof sealing method for the CsI/TI screen.
* The information contained herein is presented only as a guide for the applications of our products. No Responsibility is assumed by TOSHIBA ELECTRON TUBES & DEVICES CO., LTD. (TETD) for any
infringements of patents or other rights of the third parties which may result from its use. No license is granted by implication or otherwise under any patent or patent rights of TETD or others.
* The information contained herein may be changed without prior notice. It is therefore advisable to contact TETD before proceeding with the design of equipment incorporating this product.
Excellent X-ray image CsI screen technology