Software Development For Correction of Gradient- Nonlinearity Distortions in MR Images T.S. Lee,...

Post on 28-Dec-2015

219 views 0 download

Tags:

Transcript of Software Development For Correction of Gradient- Nonlinearity Distortions in MR Images T.S. Lee,...

Software Development For Software Development For Correction of Gradient-Correction of Gradient-

Nonlinearity Distortions in MR Nonlinearity Distortions in MR ImagesImages

T.S. Lee, K.E. Schubert

Computer Science

CSUSB

R.W. Schulte

Radiation Medicine

LLUMC

Functional Proton Radiosurgery Functional Neurosurgery

– Trigeminal Neuralgia

– Parkinson’s DiseaseParkinson’s Disease

– Brain regions (< 1 cm)

Proton Radiosurgery– Accurate to less than 1 mm

MRI for target localization– Distinguish tissue types

– 512512 images

– 262,144 pixels/study

– Gradient nonlinearity distortions (~2mm)

Example MR Phantom Images

Sagittal Plane

Coronal Plane

Axial Plane

Bad SlidesPa

rtia

l Ph

anto

m

Part

ial

Phan

tom

No

Phan

tom

Off

Cen

ter

Phan

tom

Finding EdgesFinding Edges

DisconnectsDisconnects

Remove Extraneous FeaturesRemove Extraneous Features

Can We Fix it?A

ir B

ubbl

eL

eaky

Sl

ice

Example Edge Images

Axial Plane

Coronal Plane

Sagittal Plane

MidplanesMidplanes

Calculate Midpoints Fit Midplane

Ideal shape, size, and orientation of phantom’s faces

Stack Midpoints

Ideal PlanesShift ±½ the phantom dimensionPerpendicular to face

Distortion ModelingMagnetic Field of CylinderSum of spherical harmonics:

5

4322

3

2222

31

220

44

2223

2222

21

220

44

2223

2222

21

220

,,

,,

,,

zKzyxKzyxK

zKzyzKzzyxfZ

yzKyzyxKyyxK

yzKyyxKyzyxfY

xzKxzyxKxyxK

xzKxyxKxzyxfX

xzz

zzz

yyy

yyy

xxx

xxx

Distortion ModelingMeasured

Corrected

Applying Distortion CorrectionAxial Plane

Applying Distortion CorrectionCoronal Plane

Applying Distortion CorrectionSagittal Plane

ResultsTheoretical undistorted points vs. corrected pointsStandard deviations of correction

+X face: Standard Deviation = 0.23503 mm

-X face: Standard Deviation = 0.25168 mm

+Y face: Standard Deviation = 0.15322 mm

-Y face: Standard Deviation = 0.15823 mm

+Z face: Standard Deviation = 0.13548 mm

-Z face: Standard Deviation = 0.14173 mm

Conclusions3 range 0.4 – 0.8 mm

–1-2 pixels on each image–Originally 2mm (5-6 pixels)

Accurate localization of anatomical targets

Future WorkFurther verification and testingClinical trialsFDA approvalTreatment on humans