Post on 16-Mar-2018
Sodium Analyzer Series 2131 Specifications 70-82-03-60 September 2009
Introduction
The 2131 is the most advanced online instrument
specifically developed for use in continuous measurement
and determination of sodium ion concentration. Common
application points in a boiler system for online sodium
measurement include: make-up water, condensate, boiler
feedwater and saturated steam/main steam. The 2131
sodium analyzer operates automatically and comes
complete with true fully automatic calibration.
Features
Wide Analysis Range (0.01 PPB - 10 PPM)
True fully automatic single or 2-point calibration
Low maintenance : no change of standards for up to
two months (one month for reagent)
Grab sample capability
No change of diffusion tubing
Automatic Range Selection
Process calibration-calibrate to sample
Two 4-20mA isolated current outputs
Protection of electrodes from “Hot sample”
Automatic temperature compensation
Automatically stores last 10 calibrations and alarm logs
No use of highly toxic reagents such as
monoethylamine or di-isopropylamine
2131 Sensor Unit
The sample flows through the heat exchanger, passes
through constant head (used to stabilize minor changes in
flow rate) and through the three solenoid valves. A vapor
entrainment mechanism adds alkaline buffer vapor to the
sample before it enters the sodium measuring flow cell and
flows to drain. Calibration and grab sample are controlled
automatically by the analyzer.
Sodium Analyzer Series 2131
Standard/grab sample liquid is brought into system via the
three 3-way solenoid valves. A thermistor is housed in the
flowcell and is used to detect the temperature of the
sample/standard/ grab sample.
Sensor flow diagram
HFS Catalog_Without Tab_HighRes.pdf 68 6/8/2011 12:41:07 PM
Sodium Analyzer Series 2131 2
Sensor Parts
General Specifications
Performance
Range 0.01 ppb - 10 ppm.
Accuracy ± 5% of reading or ± 0.1 ppb (whichever is greater) within ± 5°C of calibration temperature.
Reproducibility ± 5% of reading or ± 0.1 ppb (whichever is greater) at constant temperature.
Response Time 90% of 1-10 ppb step: less than 4 minutes 90% of 100-1 ppb step: less than 6 minutes
Temperature Compensation Automatic.
pH Compensation Alkaline vapor buffer: 0.88 SG (35% W/V) ammonia above 0.1 ppb sodium. 50% diethyl amine solution below 0.1 ppb sodium.
Single or 2-point calibration Manual and fully automatic calibration. Manual initiated by operator, automatic with programmable frequency in days. No manual changes of standards during calibration.
Current Outputs Two isolated 4-20 mA current (analog) outputs.
Digital Outputs Communication interface via RS-232 & RS-485 using MODBUS RTU protocol
Alarms Three voltage-free contacts, alarm points set from transmitter rated at 2A, 250VAC (non-inductive)
Power Wide range of input power supply 90VAC – 250VAC
CE Certification (on request) Meets low voltage and low electromagnetic compatibility directives
HFS Catalog_Without Tab_HighRes.pdf 69 6/8/2011 12:41:07 PM
Sodium Analyzer Series 2131 3
General Specifications
Operating Conditions
Sample Temperature: 41 – 131 °F (5 - 55°C); Flow 150 - 400 ml/min.
Ambient Temperature 32 - 131 °F (0 - 55°C).
Pressure 5 – 30 psig (0.35 – 2.07 bar)
Composition Sample should be filtered to 60 micron, free of film forming compounds.
Mechanical
Construction
Transmitter – high strength ABS, IP65 (NEMA 4x)
Wet section – high strength ABS and sheet metal, all wetted parts stainless steel or PVC, IP55 (NEMA 4)
Dimensions
Height=31”(78.74 cm) including drain
Width=22”(55.88 cm)
Depth=8.5”(21.59 cm)
Mounting
Four key-style holes:
Height=0.86”(2.18 cm), Top Width=0.281”(0.714 cm), Radius=0.280”(0.711 cm)
Horizontal: 14.7”(37.34 cm), Vertical: 11”(27.94 cm)
Weight 40 lbs. (18.14 kg)
Wiring Eight (8) conduit knockouts
Piping Sample Inlet: 1/4"(0.635 cm) OD Swagelok®
Drain: 3/8"(0.953 cm) OD tube
Finish Corrosion resistant.
Distance Up to 325 feet (100 meters) on request
HFS Catalog_Without Tab_HighRes.pdf 70 6/8/2011 12:41:07 PM
Sodium Analyzer Series 2131 4
For More Information
Learn more about how Honeywell’s 2131C Sodium
Analyzer can increase performance, reduce downtime
and decrease configuration costs, visit our website
www.honeywell.com/ps/hfs or contact your Honeywell
account manager.
Honeywell Process Solutions
1860 West Rose Garden Lane
Phoenix, Arizona 85027
Tel: 1-800-423-9883 or 1-800-343-0228
www.honeywell.com/ps
Warranty/Remedy
Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Contact your local
sales office for warranty information. If warranted goods are returned to Honeywell during the period of coverage, Honeywell
will repair or replace without charge those items it finds defective. The foregoing is Buyer's sole remedy and is in lieu of all
other warranties, expressed or implied, including those of merchantability and fitness for a particular purpose.
Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this
printing. However, we assume no responsibility for its use.
While we provide application assistance personally, through our literature and the Honeywell web site, it is up to the
customer to determine the suitability of the product in the application.
Specifications are subject to change without notice.
70-82-03-60 September 2009 © 2010 Honeywell International Inc.
HFS Catalog_Without Tab_HighRes.pdf 71 6/8/2011 12:41:07 PM
Sodium Analyzer Series 2131C Specifications 70-82-03-61 September 2009
Introduction
The 2131C is the most advanced online instrument
specifically developed for use in continuous
measurement and determination of sodium ion
concentration. Common application points in a boiler
system for online sodium measurement include: make-
up water, condensate, boiler feedwater and saturated
steam/main steam.
Features
Wide Analysis Range (0.01 PPB - 10 PPM)
Low maintenance
Very low reagent & standard consumption
No change of diffusion tubing
Automatic Range Selection
Process calibration-calibrate to sample
Two 4-20mA isolated current outputs
Protection of electrodes from “Hot sample”
Automatic temperature compensation
Automatically stores last 10 calibrations and alarm
logs
No use of highly toxic reagents such as
monoethylamine or diisopropylamine
Sodium Analyzer
2131C Sensor Unit
During calibration the 3-way valve is activated, a
standard of known sodium concentration passes
through a heat exchanger to equilibrate the solution’s
temperature to that of the sample. The standard then
enters the constant head unit where flow is stabilized.
Alkaline buffer vapor is added via a vapor
entrainment mechanism before the standard enters
the sodium measuring flow cell and flows to drain.
During normal operating conditions the standard
solution is replaced by the sample.
HFS Catalog_Without Tab_HighRes.pdf 72 6/8/2011 12:41:07 PM
Sodium Analyzer Series 2131C 2
Sensor Parts
General Specifications
Performance
Range 0.01 ppb - 10 ppm.
Accuracy ± 5% of reading or ± 0.1 ppb (whichever is greater) within ± 5°C of calibration temperature
Reproducibility ± 5% of reading or ± 0.1 ppb (whichever is greater) at constant temperature
Response Time 90% of 1-10 ppb step: less than 4 minutes 90% of 100-1 ppb step: less than 6 minutes
Temperature Compensation Automatic
pH Compensation Alkaline vapor buffer: 0.88 SG (35% W/V) ammonia above 0.1 ppb sodium 50% diethylamine solution below 0.1 ppb sodium
Single or 2-point calibration Manual calibration
Current Outputs Two isolated 4-20 mA current (analog) outputs
Digital Outputs Communication interface via RS-232 & RS-485 using MODBUS RTU protocol
Alarms Three voltage-free contacts, alarm points set from transmitter rated at 2A, 250VAC (non-inductive)
Power Wide range of input power supply 90VAC – 250VAC
CE Certification (on request)
Meets low voltage and low electromagnetic compatibility directives
HFS Catalog_Without Tab_HighRes.pdf 73 6/8/2011 12:41:07 PM
Sodium Analyzer Series 2131C 3
General Specifications
Operating Conditions
Sample Temperature: 41 – 131 °F (5 - 55°C); Flow 150 - 400 ml/min
Ambient Temperature 32 - 131 °F (0 - 55°C)
Pressure 5 – 30 psig (0.35 – 2.07 bar)
Composition Filter to 60 micron, free of film forming compounds
Mechanical
Construction
Transmitter – high strength ABS, IP65 (NEMA 4x)
Wet section – high strength ABS and sheet metal, all wetted parts stainless steel or PVC, IP55 (NEMA 4)
Dimensions
Transmitter
Height=10.92”(27.74 cm)
Width=7”(17.78 cm)
Depth=3.54”(8.99 cm)
Wet section
Height=18.25”(46.36 cm)
Width=11.25”(28.58 cm)
Depth=7.63”(19.38 cm)
Mounting
Four key-style holes:
Height=0.86”(2.18 cm), Top Width=0.281”(0.714 cm), Radius=0.280”(0.711 cm)
Horizontal: 14.7”(37.34 cm), Vertical: 11”(27.94 cm)
Weight 23 lbs. (10.44 kg)
Wiring Eight (8) conduit knockouts
Piping Sample Inlet: 1/4"(0.635 cm) OD Swagelok®
Drain: 3/8"(0.953 cm) OD tube
Finish Corrosion resistant.
Distance Up to 325 feet (100 meters) on request
HFS Catalog_Without Tab_HighRes.pdf 74 6/8/2011 12:41:07 PM
Sodium Analyzer Series 2131C 4
For More Information
Learn more about how Honeywell’s 2131C Sodium
Analyzer can increase performance, reduce downtime
and decrease configuration costs, visit our website
www.honeywell.com/ps/hfs or contact your Honeywell
account manager.
Honeywell Process Solutions
1860 West Rose Garden Lane
Phoenix, Arizona 85027
Tel: 1-800-423-9883 or 1-800-343-0228
www.honeywell.com/ps
Warranty/Remedy
Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Contact your local
sales office for warranty information. If warranted goods are returned to Honeywell during the period of coverage, Honeywell
will repair or replace without charge those items it finds defective. The foregoing is Buyer's sole remedy and is in lieu of all
other warranties, expressed or implied, including those of merchantability and fitness for a particular purpose.
Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this
printing. However, we assume no responsibility for its use.
While we provide application assistance personally, through our literature and the Honeywell web site, it is up to the
customer to determine the suitability of the product in the application.
Specifications are subject to change without notice.
Swagelok is a registered product of Swagelok Co.
70-82-03-61 September 2009 © 2010 Honeywell International Inc.
HFS Catalog_Without Tab_HighRes.pdf 75 6/8/2011 12:41:07 PM