JointDualPhotopeak ImageReconstructioninLu-177SPECT

2018EANM

HongkiLim,JeffreyFessler,YuniDewarajaUniversityofMichigan,AnnArbor

EANMDisclosure Statement

1) I or one of my co-authors hold a position as an employee, consultant, assessor or advisor for a pharmaceutical, deviceor biotechnology company. If yes, please specify name/position/company:Nothing to disclose

2) I or one of my co-authors receive support from a pharmaceutical, device or biotechnology company. If yes, pleasespecify name/position/company/which project and whether support is in kind or monetary:Yuni Dewaraja is paid consultants forMIM Software Inc., Cleveland, USA

3) I or one of my co-authors hold property rights/patents for (radio)pharmaceuticals, medical devices or medicalconsulting firms. If yes, please specify name/position/company:Nothing to disclose

4)Ioroneofmyco-authorshavewrittenarticlesfor(radio)pharmaceutical,medicaldevice,biotechnologyorconsultingcompaniesduringthelast5years.Ifyes,pleasespecifyname/position/company/article/journalandco-authors:Nothing to disclose

2/21JointdualphotopeakimagereconstructioninLu-177SPECT

HongkiLim(Universityof Michigan)

EnergySpectrumof177Lu

3/21JointdualphotopeakimagereconstructioninLu-177SPECT

HongkiLim(Universityof Michigan)

1. Ljungberg et al. "MIRD pamphlet no. 26" J Nucl Med 57.1 (2016): 151-62.

q Generalrecommendation1

• MEcollimator+208keV

q Anotheroption

• MEcollimator+113&208keV

0 300Re

lativeCo

unts

Energy(keV)

Simulationwithpointsourceinwaterphantom&MEcollimator

Totalevents PrimaryEvents Total- Primary

EnergySpectrumofLu-177

3/21

q Generalrecommendation

• MEcollimator+208keVàClinicallyadoptedoption

q Anotheroption

• MEcollimator+113&208keV

0 300Re

lativeCo

unts

Energy(keV)

Simulationwithpointsourceinwaterphantom&MEcollimator

Totalevents PrimaryEvents Total- Primary

208

JointdualphotopeakimagereconstructioninLu-177SPECT

HongkiLim(Universityof Michigan)

EnergySpectrumofLu-177

3/21

q Generalrecommendation

• MEcollimator+208keV

q Anotheroption

• MEcollimator+113&208keVàGoodtousewhen#ofcountsis

insufficient(imagingatlatertimepointorpretherapydiagnosticimaging)

0 300Re

lativeCo

unts

Energy(keV)

Simulationwithpointsourceinwaterphantom&MEcollimator

Totalevents PrimaryEvents Total- Primary

113 208

JointdualphotopeakimagereconstructioninLu-177SPECT

HongkiLim(Universityof Michigan)

EnergySpectrumofLu-177

4/21

q Comparedto208-keV,113-keV has

! ~60%ofYield! ~70%PrimaryCounts(cps/MBq)forMEcollimator

0 300Re

lativeCo

unts

Energy(keV)

Simulationwithpointsourceinwaterphantom&MEcollimator

Totalevents PrimaryEvents Total- Primary

113 208

JointdualphotopeakimagereconstructioninLu-177SPECT

HongkiLim(Universityof Michigan)

Optionsforusingbothphotopeaks- 113&208keV

5/21

q Twooptions1) Reconstructtwoimagesseparately andmakesum

𝑦#~Poisson 𝐴#𝑥# + 𝑠# ,𝑥/# = argmin𝑥#𝑓(𝑥#; 𝑦#, 𝑠#)𝑦8~Poisson 𝐴8𝑥8 + 𝑠8 ,𝑥/8 = argmin𝑥8𝑓(𝑥8; 𝑦8, 𝑠8)𝑥/ = 𝑥/#+𝑥/8

2)Reconstructoneimagejointly

𝑦 =𝑦#𝑦8 ~Poisson( 𝐴#𝐴8

𝑥 +𝑠#𝑠8 )

𝑥/ = argmin𝑥𝑓(𝑥; 𝑦, 𝑠)

• 𝒚#(𝒚8) ismainenergywindowand𝒔#(𝒔8) isscatterenergywindowacquisitionaround208(113)-keV• 𝑓(𝑥)denotesnegativePoisson likelihood function• Forscatterestimation,weusedtriple-energy-window(TEW)technique

JointdualphotopeakimagereconstructioninLu-177SPECT

HongkiLim(Universityof Michigan)

Optionsforusingbothphotopeaks- 113&208keV

5/21

àMorealignedwithactualPoissonnatureofmulti-windowacquisition

• 𝒚#(𝒚8) ismainenergywindowand𝒔#(𝒔8) isscatterenergywindowacquisitionaround208(113)-keV• 𝑓(𝑥)denotesnegativePoisson likelihood function• Forscatterestimation,weusedtriple-energy-window(TEW)technique

q Twooptions1) Reconstructtwoimagesseparately andmakesum

𝑦#~Poisson 𝐴#𝑥# + 𝑠# ,𝑥/# = argmin𝑥#𝑓(𝑥#; 𝑦#, 𝑠#)𝑦8~Poisson 𝐴8𝑥8 + 𝑠8 ,𝑥/8 = argmin𝑥8𝑓(𝑥8; 𝑦8, 𝑠8)𝑥/ = 𝑥/#+𝑥/8

2)Reconstructoneimagejointly

𝑦 =𝑦#𝑦8 ~Poisson( 𝐴#𝐴8

𝑥 +𝑠#𝑠8 )

𝑥/ = argmin𝑥𝑓(𝑥; 𝑦, 𝑠)

JointdualphotopeakimagereconstructioninLu-177SPECT

HongkiLim(Universityof Michigan)

Whatfactorsneedtobeconsideredforjointreconstruction?

6/21

q SystemModel𝑦#𝑦8 ~Poisson(

𝐴#𝐴8

𝑥 +𝑠#𝑠8 ),

where𝑎EFG = 𝛼G𝑝EFG𝑔EFG

𝑎EFG: elementofsystemmodel(𝐴G) at𝑖-th rowand𝑗-th column in𝑘-th energywindow

𝛼G:Relativecameradetection efficiency × branchingratio𝑝EFG :depth-dependentandenergy-dependenteffectofattenuation𝑔EFG:depth-dependentandenergy-dependent3D-collimator detectorresponse

JointdualphotopeakimagereconstructioninLu-177SPECT

HongkiLim(Universityof Michigan)

Experimentalsetting

7/21

q 2measurements+1simulation+1patientscan

I. Spherephantommeasurement(4scans)

II. Liver-torsophantommeasurement(4scanswithhigh/low-counts)

III. XCATphantomsimulation (10realizationswithhigh/low-counts)

IV. Patientscan

JointdualphotopeakimagereconstructioninLu-177SPECT

HongkiLim(Universityof Michigan)

Physicalphantommeasurement1

8/21

q Spherephantommeasurement(4scanstoassessnoise)- Sphere-to-backgroundratiowas6:1 ,30minutesacquisition- Mainwindowwidth:20%,scatterwindowwidth:10%

208-keVmainwindow 208-keVscatterwindow

113-keVmainwindow 113-keVscatterwindow

JointdualphotopeakimagereconstructioninLu-177SPECT

HongkiLim(Universityof Michigan)

Physicalphantommeasurement2

9/21

q Liver-torsophantommeasurement- 4scanstoassessnoise:

4scanof35 min(mimickingDay3post177Lu-DOTATATE)4scansof3 min(mimickingDay3pre-therapydiagnosticimaging)

- Lesion-to-liverratiowas4:1

35 minutesscan

208-keVmainwindow

113-keVmainwindow

208-keVmainwindow

113-keVmainwindow

3 minutesscanJointdualphotopeakimagereconstructioninLu-177SPECT

HongkiLim(Universityof Michigan)

XCATsimulation

10/21

q XCATcoupledwithSIMIND2 simulation(10realizationstoassessnoise)- Realisticactivitydistributionfor177Lu-DOTATATE

2. Ljungberg M. Monte Carlo Calculation in Nuclear Medicine: Application in Diagnostic Imaging, 2nd edn, 2012.

OrganRelativeactivity

concentration

Lesioninliver 1.00

Spleen 0.59

Kidney 0.48

Liver 0.16

Gallbladder 0.02

Renalpelvis 0.02

Post-therapyDay1count-level Pre-therapyDay1count-level

208-keVmain 208-keVscatter

113-keVmain 113-keVscatter

208-keVmain 208-keVscatter

113-keVmain 113-keVscatter

JointdualphotopeakimagereconstructioninLu-177SPECT

HongkiLim(Universityof Michigan)

177Lu-DOTATATEPatientscan

11/21

q Day5 acquisition- 1clinicalscanwith208-keV (Scannedfor25minutes)- 3researchscanswithboth photopeaks (Scannedfor2 minutestomimicpre-therapy)

25minutesscanatDay5

2minutesscanatDay5

208-keVmainwindow 208-keVscatterwindow

113-keVmainwindow 113-keVscatterwindow

208-keVmainwindow 208-keVscatterwindow

JointdualphotopeakimagereconstructioninLu-177SPECT

HongkiLim(Universityof Michigan)

ReconstructionandEvaluationmethods

12/21

I. Activity-recovery

II. Contrast-recovery

III. Image-ensemble-noise(variabilityacrossmultipleacquisitions)

q Reconstructiondetails(forbothsingleandjoint)• AppliedGaussiansmoothingtoTEWscatterestimate• EMalgorithm+50iteration(1subset)

q Evaluationmetrics

JointdualphotopeakimagereconstructioninLu-177SPECT

HongkiLim(Universityof Michigan)

Calibrationfactorforabsolutequantification

13/21

1. NISTrecommendation3 (0.9%uncertainty)• 3mL177Luina10mLSchottvial:CRC-15Rdialsetting449x10

2. TransferringCalibrationtoanewgeometry(10mLsyringe)• Activity insyringe=activityinvial- residual

3. Withthesyringe inthedosecalibratoradjustsetting• Settingtogetcorrectreadingfor3mLinsyringe:480x10

Syringe tapedtoSourceholder

q Imagingforcalibrationfactor• planar with20%mainwindow&10%TEWSC

q 208keV(head1):12.9cps/MBq(head2):13.4cps/MBq

à (within1%ofmanuf.specifiedvalue for5/8”xtal)q 113keV(head1): 9.2cps/Mbq (70%ofabove)

(head2):9.3cps/MBq

‘NISTgeometry’fordosecalibrator

Transfertosyringeforfillingphantomandcamerasensitivity

3. Bergeron DE, Cessna JT. Nucl Med Commun. 2018 Jun;39(6):500-504

JointdualphotopeakimagereconstructioninLu-177SPECT

HongkiLim(Universityof Michigan)

Result:Spherephantommeasurement

14/21

Single208-keV Single113-keV Joint

Single208-keV Single113-keV Joint

0 200 400 600x-axis

0

5

10

Activ

ity (B

q/m

l)

104

TRUESIR 208SIR 113JIR

Profileacrosscoldspot

Profile

JointdualphotopeakimagereconstructioninLu-177SPECT

HongkiLim(Universityof Michigan)

Result:Spherephantommeasurement

15/21

0

2

4

6

8

10

12

14

0 50 100 150

ContrastRecoverytoNoiseRatio

SphereSize(ml)

SinglePhotopeakRecon@208-keV

SinglePhotopeakRecon@113-keV

JointPhotopeakRecon

0

20

40

60

80

100

0 20 40 60 80 100 120

ContrastRecovery

SphereSize(ml)

SinglePhotopeakRecon@208-keV

SinglePhotopeakRecon@113-keV

JointPhotopeakRecon

0

20

40

60

80

100

120

0 20 40 60 80 100 120

ActivityRecovery

SphereSize(ml)

SinglePhotopeakRecon@208-keV

SinglePhotopeakRecon@113-keV

JointPhotopeakRecon

JointdualphotopeakimagereconstructioninLu-177SPECT

HongkiLim(Universityof Michigan)

Result:Liver-torsophantommeasurement(axialview)

16/21

Single208-keV Single113-keV Joint

Single208-keV Single113-keV Joint

HighCountsà

LowCountsà

JointdualphotopeakimagereconstructioninLu-177SPECT

HongkiLim(Universityof Michigan)

Result:Liver-torsophantommeasurement

17/21

Noise ContrastRecovery CRNR29mlovoid 16mlsphere 29mlovoid 16mlsphere

208-keV 11.4 75.3 70.4 6.6 6.2113-keV 16.4 72.4 66.6 4.4 4.1Joint 9.5 74.4 69.1 7.8 7.2

Noise ContrastRecovery CRNR29mlovoid 16mlsphere 29mlovoid 16mlsphere

208-keV 36.4 65.2 63.7 1.8 1.8113-keV 48.5 58.7 51.8 1.2 1.1Joint 29.3 62.1 58.1 2.1 2.0

35 minutesscan

3 minutesscan

JointdualphotopeakimagereconstructioninLu-177SPECT

HongkiLim(Universityof Michigan)

Result:XCATphantomsimulation

18/21

Noise ContrastRecovery CRNR29mlovoid 16mlsphere 29mlovoid 16mlsphere

208keV 10.1 76.3 92.9 7.5 9.2113keV 12.9 73.3 90.4 5.7 7.0Joint 8.0 75.0 91.8 9.4 11.5

Noise ContrastRecovery CRNR29mlovoid 16mlsphere 29mlovoid 16mlsphere

208keV 65.2 75.8 91.7 1.2 1.4113keV 85.5 68.7 91.5 0.8 1.1Joint 52.2 72.8 91.6 1.4 1.8

Post-therapyDay1count-level

Pre-therapycount-level

208-keV 113-keV Joint

208-keV 113-keV Joint

Liver

Kidneys

JointdualphotopeakimagereconstructioninLu-177SPECT

HongkiLim(Universityof Michigan)

Result:Patientscan

19/21

25minutes scan,208-keVSPECT/CT

Single208-keVShortScan

Single113-keVShortScan

JointShortScan

Recon(unit:Bq/ml) RMSD* Noise*208-keV 1.47e4 332.9113-keV 1.90e4 278.6Joint 1.45e4 260.4*RMSDrelativetolongscan*Imageensemblenoisebetween3repeatshortscans

Spleen

Liver

Tumor

Single208-keVLongScan

JointdualphotopeakimagereconstructioninLu-177SPECT

HongkiLim(Universityof Michigan)

Summary

20/21

q Jointreconstructionw/E-dependentsystemmodel&TEWSCiseasytoimplementq Comparedtosinglephotopeak(208-keV)reconstruction,jointreconstruction- Gavecomparableactivity/contrastrecovery- Improvednoisesignificantlyevenforhigh-countsdataà Importantforvoxel-leveldosimetry- Improvedcontrastrecoverytonoiseratio: 16.7-27.3%

q Jointreconstructioncanbebeneficialforreducingscantime

ExperimentsSpherephantom Liverphantom XCAT

Count-level High Low High Low

Improvement(%) 20.9-23.2 16.7-17.5 17.3-21.5 25.4-25.8 24.7-27.3

Improvement(%)ofContrastRecoverytoNoiseRatiooversingle208-keV

JointdualphotopeakimagereconstructioninLu-177SPECT

HongkiLim(Universityof Michigan)

Thank You

Weacknowledge• UofMichiganRadiologySeedGrant• JeremyNiedbala,Gerrid Rosebushforallhelpinmeasurementsandcalibration