Table of Contents
1 Introduction 31.1 Conventions and symbols used in these Operating Instructions ......................................... 3
2 Safety Information 42.1 Definition of signal warnings and symbols...................................................................... 42.2 Product safety information ............................................................................................ 4
3 Design and Function 73.1 Overview..................................................................................................................... 7
3.1.1 Components................................................................................................. 73.1.2 Operation keys.............................................................................................. 83.1.3 Display ........................................................................................................ 9
3.2 Basic principles for operation ........................................................................................ 10
4 Installation and Putting into Operation 124.1 Unpacking and delivery inspection................................................................................. 124.2 Installing components .................................................................................................. 134.3 Installing protective cover.............................................................................................. 154.4 Selecting a location...................................................................................................... 164.5 Connecting the balance ................................................................................................ 164.6 Setting up the balance .................................................................................................. 17
4.6.1 Switching on the balance ............................................................................... 174.6.2 Leveling the balance...................................................................................... 174.6.3 Setting date and time..................................................................................... 174.6.4 Adjusting the balance .................................................................................... 19
4.7 Adjustment (calibration) ............................................................................................... 194.7.1 Adjustment with internal weight....................................................................... 194.7.2 Adjustment with external weight ...................................................................... 204.7.3 Customer fine adjustment............................................................................... 21
4.8 Performing a simple weighing ....................................................................................... 224.9 Weighing below the balance ......................................................................................... 234.10 Transporting the balance .............................................................................................. 24
5 The Menu 255.1 What is in the menu ? .................................................................................................. 255.2 Description of menu topics............................................................................................ 26
5.2.1 Main menu................................................................................................... 265.2.2 Basic menu.................................................................................................. 265.2.3 Advanced menu............................................................................................ 275.2.4 Interface menu.............................................................................................. 29
6 Applications 346.1 Application piece counting ............................................................................................ 346.2 Application percent weighing......................................................................................... 376.3 Application check weighing........................................................................................... 396.4 Application statistics..................................................................................................... 416.5 Application totaling ...................................................................................................... 436.6 Application multiplication factor weighing ....................................................................... 446.7 Application division factor weighing ............................................................................... 466.8 Application density....................................................................................................... 48
6.8.1 Density determination of solids ....................................................................... 486.8.2 Density determination of liquids ...................................................................... 506.8.3 Formulae used to calculate density ................................................................. 50
7 Communication with Peripheral Devices 537.1 Function PC-Direct ....................................................................................................... 53
Table of Contents 1Jewelry Balances
8 Error and Status Messages 558.1 Error messages ........................................................................................................... 558.2 Status messages ......................................................................................................... 56
9 Maintenance 579.1 Cleaning and service.................................................................................................... 579.2 Cleaning the draft shield ............................................................................................... 589.3 Disposal ..................................................................................................................... 589.4 Firmware (Software) Updates ........................................................................................ 58
10 Technical Data 5910.1 General data ............................................................................................................... 5910.2 Model-specific data...................................................................................................... 61
10.2.1 Carat balances with readability of 0.001 ct / 0.1 mg ........................................ 6110.2.2 Gold balances with readability of 1 mg............................................................ 6310.2.3 Gold balances with readability of 10 mg.......................................................... 65
10.3 Dimensions................................................................................................................. 6810.3.1 Carat balances with readability of 0.001 ct / 0.1 mg ........................................ 6810.3.2 Gold balances with readability of 1 mg............................................................ 6910.3.3 Gold balances with readability of 10 mg.......................................................... 70
10.4 Interface specification ................................................................................................... 7110.4.1 RS232C interface .......................................................................................... 7110.4.2 MT-SICS interface commands and functions..................................................... 71
11 Accessories and Spare Parts 7211.1 Accessories................................................................................................................. 7211.2 Spare parts ................................................................................................................. 76
Index 79
Table of Contents2 Jewelry Balances
Introduction 3Jewelry Balances
1 IntroductionThank you for choosing a METTLER TOLEDO balance. The balances of the Jewelry line combine a largenumber of weighing possibilities with easy operation.The operating instructions are based on the initially installed terminal firmware (software) version V 1.01.
u www.mt.com/jewelry
1.1 Conventions and symbols used in these Operating Instructions
Key and / or button designations are shown in graphic or bold text (e.g. ).These symbols indicate an instruction:
§ prerequisites
1 steps2 ...ð results
This symbol indicates press key briefly (less than 1.5 s).
This symbol indicates press and hold key down (longer than 1.5 s).
This symbol indicates a flashing display.
Safety Information4 Jewelry Balances
2 Safety Information• Read and understand the instructions in this manual before you use the balance.• Keep this manual for future reference.• Include this manual if you pass on the balance to other parties.If the balance is not used according to the instructions in this manual or if it is modified, the safety of theuser may be impaired and Mettler-Toledo GmbH assumes no liability.
2.1 Definition of signal warnings and symbolsSafety notes are marked with signal words and warning symbols. These show safety issues and warnings.Ignoring the safety notes may lead to personal injury, damage to the instrument, malfunctions and falseresults.
WARNING for a hazardous situation with medium risk, possibly resulting in death or severeinjury if not avoided.
CAUTION for a hazardous situation with low risk, resulting in minor or moderate injury if notavoided.
NOTICE for a hazardous situation with low risk, resulting in damage to the balance, othermaterial damage, malfunctions and erroneous results, or loss of data.
Note (no symbol) for useful information about the product.
General hazard Electrical shock
NOTICE
2.2 Product safety informationYour balance represents state-of-the-art technology and complies with all recognized safety rules, however,certain hazards may arise in extraneous circumstances. Do not open the housing of the balance; it does notcontain any parts that can be maintained, repaired or replaced by the user. If you experience problems withyour balance, contact your authorized METTLER TOLEDO dealer or service representative.The balance has been tested for the experiments and intended purposes documented in the appropriatemanual. However, this does not absolve you from the responsibility of performing your own tests of theproducts supplied by us regarding their suitability for the methods and purposes you intend to use them for.
Intended useThis balance is designed to be used in analytical laboratories by qualified staff. Your balance is used forweighing. Use the balance exclusively for this purpose.Any other type of use and operation beyond the limits of technical specifications without written consentfrom Mettler-Toledo GmbH, is considered as not intended.
Site requirementsThe balance has been developed for indoor operation. Avoid the following environmental influences:• Conditions outside of the ambient conditions specified in the technical data• Powerful vibrations• Direct sunlight• Corrosive gas atmosphere
Safety Information 5Jewelry Balances
• Explosive atmosphere of gases, steam, fog, dust and flammable dust• Powerful electric or magnetic fields
Staff qualificationIncorrect use of the balance or the chemicals used in the analysis can lead to death or injury. The followingexperience is needed for operating the balance.• Knowledge and experience in working with toxic and caustic substances.• Knowledge and experience in working with standard laboratory equipment.• Knowledge and experience in working in accordance with general lab safety rules.
Responsibilities of the balance ownerThe balance owner is the person that uses the balance for commercial use or places the balance at thedisposal of his staff. The balance owner is responsible for product safety and the safety of staff, user(s) andthird party.The operator has the following responsibilities:• Know the rules for safety at the workplace that are in effect and enforce them.• Ensure that only qualified staff uses the balance.• Define the responsibilities for installation, operation, cleaning, troubleshooting and maintenance and
ensure that the tasks are done.• Train the staff in regular intervals and inform them about dangers.• Provide the necessary protective gear for the staff.
Shut down of the balance in emergency situations− Pull the plug from the electrical outlet.
Safety notes
WARNINGDanger of death or serious injury due to electric shock!Contact with parts that contain a live current can lead to injury and death. If the balancecannot be shut down in an emergency situation, people can be injured or the balance canbe damaged.1 Only use the supplied three-core power cable with equipment grounding conductor to
connect your balance.2 Check that the voltage printed on the balance is the same as your local power supply
voltage.
ð If this is not the case, under no circumstances connect the AC/DC adapter to thepower supply, but contact a METTLER TOLEDO representative.
3 Only connect the balance to a three-pin power socket with earthing contact.4 Only standardized extension cable with equipment grounding conductor must be used
for operation of the balance.5 Do not disconnection the equipment grounding conductor.6 Check the cables and the plug for damage and replace damaged cables and plugs.7 Make sure that the cables are arranged so that they cannot be damaged or interfere
with the operation.8 Keep all electrical cables and connections away from liquids.9 Make sure that the power plug is accessible at all times.
NOTICEEnvironmentOnly use indoors in dry locations.
Safety Information6 Jewelry Balances
NOTICEDanger of damaging the keyboard with pointed or sharp objects!Do not use pointed or sharp objects to navigate on the keyboard. This may damage thesurface of the keyboard.− Operate the keyboard with your fingers.
NOTICEDanger of damage to the balance!Never open the balance. The balance contains no user-serviceable parts.− In the event of problems, please contact a METTLER TOLEDO representative.
NOTICEDanger of damage to the balance due to incorrect parts!Using incorrect parts with the balance can damage the balance or cause the balance tomalfunction.− Only use parts supplied with the balance, listed accessories and spare parts from
Mettler-Toledo GmbH.
NOTICEDamage of the balance or softwareIn some countries, excessive mains voltage fluctuations and strong glitches may occur.This may affect the balance functions or damage the software.− Use a voltage regulator for stabilizing.
Finding more information
Refers to an external document.
Design and Function 7Jewelry Balances
3 Design and Function
3.1 Overview
3.1.1 Components
METT
LER T
OLEDO
METTLER TOLEDO
METTLER TOLEDO
7
0.001 ct / 0.1 mg 1 mg
10 mg
METTLER TOLEDO
8
8
8
1
2
3
12
3
7
4
4
5
6
3
4
6
2
1
6
6
6
8
910
1112
1 Display 2 Operation keys
3 Level indicator 4 Weighing pan
5 Draft shield element 6 Handle for operation of the draft shield door
7 Glass draft shield 8 Leveling foot
9 Legal for Trade (LFT) sealing 10 Socket for AC adapter
11 RS232C serial interface 12 Kensington slot for anti-theft purposes
Design and Function8 Jewelry Balances
3.1.2 Operation keys
1 2 3 4 5
Legend key functions
No. Key Press briefly (less than 1.5 s) Press and hold (longer than 1.5 s)
1 • Cancel or leave menu without saving• One step back in the menu
• Select the simple weighing application• Exit application
2 • Print display value• Transmit data• To navigate backwards in the menu or
menu selection• Decrease parameters in menu or appli-
cations
• Open the application list for selecting anapplication
3 • Zero/Tare• Switch on
• Switch off into standby mode
4 • With entries, scroll down• To navigate forward menu topics or
menu selections• To toggle between unit 1, recall value (if
selected), unit 2 (if different from unit 1)and the application unit (if any)
• Increase parameters in menu or appli-cations.
• Select adjustment (calibration)– with internal weight *– with external weight– Customer fine adjustment *
* On models with internal weight only
5 • Enter or leave menu selection• To enter application parameter digit and
switch to next parameter digit• To accept parameter in menu selection.
• Enter or leave menu (parameter settings)• To store parameter• To accept numeric inputs in appli-
cations.
Design and Function 9Jewelry Balances
3.1.3 Display
Application Icons Status Icons
Wei
ghin
g-in a
id
Weight Value Field Unit Field
Application IconsApplication weighing Application totaling
Application piece counting Application multiplication factor
Application percent weighing Application division factor
Application check weighing Application density
Application statistics Menu locked
NoteWhile an application is running, the corresponding application icon appears at the top of the display.
Status IconsIndicates stored value (Memory) Acoustic feedback for pressed keys
activated
Adjustments (calibration) started Weighing range 1 (Dual Range modelsonly)
Service reminder Weighing range 2 (Dual Range modelsonly)
Weight Value Field and Weighing-in aidIndicates negative values Brackets to indicate uncertified digits
(approved models only)
Indicates unstable values Marking of nominal or target weight
Indicates calculated values Marking of tolerance limit T+
Marking of tolerance limit T-
Unit Field
g gram ozt troy ounce tls Singapore taels
kg kilogram GN grain tlt Taiwan taels
mg milligram dwt pennyweight tola tola
ct carat mom momme baht baht
lb pound msg mesghal
oz ounce tlh Hong Kong taels
Design and Function10 Jewelry Balances
3.2 Basic principles for operation
Selecting simple weighing or terminate application
− Press and hold until WEIGH appears on the display.
ð The balance returns to the simple weighing mode.
WEIGH
0.00 g
I00.0%
NoteHow to perform simple weighing see [Performing a simple weighing Page 22].
Selecting an application
1 Press and hold F until APP.LIST (application list).
ð Last active application e.g. COUNT appears on the display.
2 Select an application by multiple pressing .
3 To execute selected application press .APP.LIST
0.00 g
COUNT
FACTOR.D
Available applications
Display Remark Description
COUNT Piece counting see [Application piece counting Page 34]
PERCENT Percent weighing see [Application percent weighing Page 37]
CHECK Check weighing see [Application check weighing Page 39]
STAT Statistics see [Application statistics Page 41]
TOTAL Totaling see [Application totaling Page 43]
FACTOR.M Multiplication factor see [Application multiplication factor weighing Page 44]
FACTOR.D Division factor see [Application division factor weighing Page 46]
DENSITY Density see [Application density Page 48]
Entering the menu
1 Press and hold Menu to enter main menu. The first menu BASICis displayed (except menu protection is active).
2 Press repeatedly to change menu.
3 Press to confirm the selection.
406.2 g
ADVANCE
BASIC
NoteDetailed description of the menu see [The Menu Page 25].
Design and Function 11Jewelry Balances
Selecting menu topics
− Press . The next menu topic appears in the display. Each time is pressed, the balance switches to the next menu topic.
TIME
DATE
Changing settings in selected menu topic
1 Press . The display shows the current setting in the selectedmenu topic. Each time is pressed, the balance switches tothe next selection. After the last selection, the first is shown again.
2 Press to confirm the setting. For store the setting see sectionSaving settings and closing the menu.
ENVIRON.
STABLE
ENVIRON.
Changing settings in a submenu selectionThe same procedure as for menu topics.
Input principle of numerical values
1 Press to select a digit (cyclically from left to right) or a value(depending on the application). The selected digit or the selectedvalue is blinking.
2 For changing blinking digits or values, press to increase or Fto decrease.
3 Press and hold to accept the value.
01.2500
01. 6500
01.6500
Saving settings and closing the menu
1 Press and hold Menu to leave menu topic.
ð SAVE:YES appears on the display.
2 Press to toggle between SAVE:YES and SAVE:NO.
3 Press to execute SAVE:YES. Changes are saved.
4 Press to execute SAVE:NO. Changes are not saved.
SAVE:YES
SAVE:NO
Cancel
§ During menu operation
− To leave menu topic or menu selection without saving press C(one step back in the menu).
§ During application operation
− To cancel settings press C.
ð The balance returns to the previous active application.
ImportantIf no entry is made within 30 seconds, the balance reverts to last active application mode. Changes are notsaved. If changes are made, the balance asks "SAVE:NO".
Installation and Putting into Operation12 Jewelry Balances
4 Installation and Putting into Operation
WARNINGDanger of death or serious injury due to electric shock!The balance must be disconnected from the power supply when carrying out all setupand mounting work.
4.1 Unpacking and delivery inspection
NOTICEDanger of damage to the balance due to incorrect parts!Using incorrect parts with the balance can damage the balance or cause the balance tomalfunction.− Only use parts supplied with the balance, listed accessories and spare parts from
Mettler-Toledo GmbH.
1 Open the packaging and carefully remove all components.2 Check the delivered items.
The standard scope of delivery contains the following items:
Components Carat balances Gold balances
0.001 ct / 0.1 mg 1 mg 10 mgDraft shield 170 mm –
Weighing pan ø 90 mm – –
ø 120 mm – –
180 × 180 mm – –
Draft shield element –
Pan support – –
Protective cover
Universal AC adapter (with plug set)
Carat pan S 80 ø × 20 mm – –
M 90 ø × 30 mm – –
EC declaration of conformity
Operating instructions or User Manual;printed or on CD-ROM, depending oncountry of use
Installation and Putting into Operation 13Jewelry Balances
4.2 Installing componentsCarat balances with readability of 0.001 ct / 0.1 mg
Place the following components on the balance inthe specified order:1 Push the side glass doors back as far as will go.2 Place draft shield element (1).3 Place weighing pan (2).
NoteCleaning the draft shield see [Cleaning andservice Page 57].
METTLER TOLEDO
1
2
Gold balances with readability of 1 mg
Place the following components on the balance inthe specified order:1 Push the side glass doors back as far as will go.2 Place weighing pan (1).
NoteCleaning the draft shield see [Cleaning andservice Page 57].
METTLER TOLEDO
1
Installation and Putting into Operation14 Jewelry Balances
Gold balances with readability of 10 mg
Place the following components on the balance inthe specified order:1 Place draft shield element (1): carefully pull apart
the draft shield element to fix it under the retainingplate.
2 Insert pan support (2).3 Place weighing pan (3).
METTLER TOLEDO
1
2
3
Installation and Putting into Operation 15Jewelry Balances
4.3 Installing protective coverImportantMake sure using the correct protective cover, see [Accessories and Spare Parts Page 72].
Balances with readability of 0.1 mg / 1 mg Balances with readability of 10 mgInstall the protective cover according to the illus-trations below, using a screwdriver Philips No. 2.
METTLER TOLEDO
Install the protective cover according to the illus-trations below, using a screwdriver Torx TX20.
METTLER TOLEDO
Installation and Putting into Operation16 Jewelry Balances
4.4 Selecting a locationAn optimal location will ensure accurate and reliable operation of the balance. The surface must be able tosafely take the weight of the balance when fully loaded. The following local conditions must be observed:
ImportantIf the balance is not horizontal at the outset, it must be leveled during commissioning.
• The balance must only be used indoors and up to a maximumaltitude of 4,000 m above sea level.
• Before switching on the balance, wait until all parts are at roomtemperature (+5 to 40 °C). The humidity must be between 10% and 80% non-condensing.
• The power plug must be accessible at all times.• Firm, horizontal and vibration-free location.• Avoid direct sunlight.• No excessive temperature fluctuations.• No strong drafts.
4.5 Connecting the balance
WARNINGDanger of death or serious injury due to electric shock!Contact with parts that contain a live current can lead to injury and death. If the balancecannot be shut down in an emergency situation, people can be injured or the balance canbe damaged.1 Only use the supplied three-core power cable with equipment grounding conductor to
connect your balance.2 Check that the voltage printed on the balance is the same as your local power supply
voltage.
ð If this is not the case, under no circumstances connect the AC/DC adapter to thepower supply, but contact a METTLER TOLEDO representative.
3 Only connect the balance to a three-pin power socket with earthing contact.4 Only standardized extension cable with equipment grounding conductor must be used
for operation of the balance.5 Do not disconnection the equipment grounding conductor.6 Check the cables and the plug for damage and replace damaged cables and plugs.7 Make sure that the cables are arranged so that they cannot be damaged or interfere
with the operation.8 Keep all electrical cables and connections away from liquids.9 Make sure that the power plug is accessible at all times.
NOTICEDanger of damage to the AC/DC adapter due to overheating!If the AC/DC adapter is covered or in a container, it is not sufficiently cooled andoverheats.1 Do not cover the AC/DC adapter.2 Do not put the AC/DC adapter in a container.
The balance is supplied with an universal AC adapter and a country-specific plug. The AC adapter issuitable for use with the following voltage range:
Installation and Putting into Operation 17Jewelry Balances
100 – 240 V AC, 50/60 Hz.
− Connect the AC adapter to the connection socket on the back ofyour balance (see figure) and to the power line.ð The balance performs a display test (all segments in the
display light up briefly), WELCOME, Software version,Maximum load and Readability appears briefly.
The balance is ready for use.
4.6 Setting up the balance
4.6.1 Switching on the balanceBefore working with the balance, it must be warmed up in order to obtain accurate weighing results. Toreach operating temperature, the balance must be connected to the power supply for at least 30 minutes(0.1 mg models 60 minutes).
Switching on
§ The balance is in STANDBY mode. MT.GREEN appears on thedisplay.
− Press or remove any load from weighing pan or tap on theweighing pan.
The balance is ready for weighing or for operation with the last activeapplication.
ImportantApproved balances can only be switched on by pressing inselected countries.
0.00 g
MT.GREEN
4.6.2 Leveling the balanceImportantIf the balance is not horizontal at the outset, it must be leveled during commissioning.
ImportantThe balance must be leveled and adjusted each time it is moved to a new location.
1 Align the balance horizontally.2 Turning the two front leveling screws of the housing until the air
bubble is in the inner circle of the level indicator.ð The position of the air bubble illustrates which leveling screw
you need to turn (L = left leveling screw, R = right levelingscrew) and in which direction so that the air bubble moves tothe center.
ExampleIn this example, turn the left leveling screw counterclockwise.
4.6.3 Setting date and timeWhen you put your new instrument into operation for the first time, you should enter the current date andtime.
Note• These settings are retained even if you disconnect your instrument from the power supply.• A reset of the instrument will not change these settings.
Installation and Putting into Operation18 Jewelry Balances
• Set the current date according to the date format DATE.FRM in the menu ADVANCE..• Set the current time according to the time format TIME.FRM in the menu ADVANCE..
1 Press and hold Menu until menu BASIC appears on the display.
2 Press to open menu BASIC.
ð DATE appears.
3 Press to confirm.
4 Set current date. Press to select day, month or year; press to set current day, month or year.
5 Press and hold to confirm the settings.
ð DATE appears..
6 Set current time. Press to select TIME.
7 Press to confirm.
ð +1H appears.
8 Select SET.TIME by pressing .
9 Press to confirm.
10 Press to select hours or minutes; press to set currenthours or minutes.
11 Press and hold to confirm the settings.
ð TIME appears.
12 Press and hold to store the settings.
ð SAVE:YES appears.
13 Press to confirm.
BASIC
SET.TIME
DATE
TIME
06.05.12
13.30
SAVE:YES
DATE
TIME
+1H
Installation and Putting into Operation 19Jewelry Balances
4.6.4 Adjusting the balanceTo obtain accurate weighing results, the balance must be adjusted to match the gravitational acceleration atits location and depending on the ambient conditions. After reaching the operation temperature, adjusting isnecessary• before the balance is used for the first time.• after a change of the location.• at regular intervals during weighing service.
4.7 Adjustment (calibration)
NOTICEBefore adjusting the balance, it must be warmed up.
4.7.1 Adjustment with internal weightNoteOn models with internal weight only (see technical data).
§ Weighing pan is unloaded.
1 To carry out this operation press and hold CAL until ADJUSTappears.
2 Select ADJ.INT by pressing .
ð ADJ.INT appears on the display.
3 Press to execute internal adjustment.The balance adjusts itself automatically. The adjusting is finishedwhen the message ADJ.DONE appears briefly on the display. Thebalance returns to the last active application and is ready foroperation.
------
ADJ.DONE
0.00 g
ADJ.INT
Cal ADJ.EXT
Sample adjustment printout using internal weight:
- Internal Adjustment --
21.Jan 2012 12:56
METTLER TOLEDO
Balance Type JE703C
SNR 1234567890
Temperature 22.5 °C
Diff 3 ppm
Adjustment done
-----------------------
Installation and Putting into Operation20 Jewelry Balances
4.7.2 Adjustment with external weightImportantBecause of certification legislation, the approved models cannot be adjusted with an external weight *(depend on selected countries' certification legislation).* except OIML accuracy class I approved models.
1 Have required adjustment weight ready.
2 To carry out this operation press and hold CAL until ADJUSTappears.
3 Select ADJ.EXT by pressing .
ð ADJ.EXT appears on the display.
4 Unload weighing pan.5 Optional: If necessary, you can define a different weight value.
Press to change a digit (cyclically from left to right); press to change the blinking digit.
6 Press and hold to execute external adjustment.ð The required adjustment weight value flashes in the display.
7 Place adjustment weight in center of pan.ð The balance adjusts itself automatically.
8 When zero is flashing, remove adjustment weight.ð The adjusting is finished when the message ADJ.DONE appears
briefly on the display. The balance returns to the last active appli-cation and is ready for operation.
------
2000.00 g
ADJ.DONE
0.00 g
C
0.00 g
------
C
ADJ.INT
ADJ.EXT
Cal
4000.00 g
2000.00 g
Sample adjustment printout using external weight:
- External Adjustment --
21.Jan 2012 12:56
METTLER TOLEDO
Balance Type JE3002GE
SNR 1234567890
Temperature 22.5 °C
Nominal 2000.00 g
Actual 1999.99 g
Diff 5 ppm
Adjustment done
Signature
.......................
-----------------------
Installation and Putting into Operation 21Jewelry Balances
4.7.3 Customer fine adjustment
NOTICEThis function should be executed only by trained personnel.
The function customer fine adjustment ADJ.CF allows you to adjust the value of the internal adjustmentweight with your own adjustment weight. The adjustable range of the adjustment weight is possible only ina very small range. Customer fine adjustment impacts the function of internal adjustment. The customer fineadjustment can be deactivated at any time.
Note• This feature is available on models with internal weight only.• Because of certification legislation, approved models cannot be adjusted with customer fine adjustment
(depending on selected countries' certification legislation).• Use certificated weights.• Balance and test weight have to be on operating temperature.• Observe the correct environmental conditions.
Execute customer fine adjustment
§ The balance is under measuring condition.
1 Have required adjustment weight ready.2 Unload weighing pan.
3 To carry out this operation press and hold CAL until ADJUSTappears.
4 Select ADJ.CF by pressing .
ð ADJ.CF appears on the display.
5 Select EXECUTE.
6 Start Adjustment with .
ð SET REF. appears briefly.
ð The last saved value flashes on the display.
7 Select the target adjustment weight. Press to change a digit(cyclically from left to right); press to change the blinkingdigit.
8 Press and hold to confirm and execute ADJ.CF.
ð The required adjustment weight value flashes in the display.This could take some time.
9 Place required adjustment weight in center of pan.10 Remove adjustment weight when zero is flashing.
11 Wait until ADJ.DONE briefly appears.
ð The adjusting is finished when the message ADJ.DONE appearsbriefly on the display. The balance returns to the last active appli-cation and is ready for operation.
ð If the error message WRONG ADJUSTMENT WEIGHT appears, theweight is not within the allowed value range and could not beaccepted. ADJ.CF could not be executed.
NoteStoring the adjustment is not required.
200.0000 g
ADJ.DONE
0.00 g
------
149.9995 g
------
149.9995 g
C
C
ADJ.CF
EXECUTE
SET REF.
ADJUST
------
Cal
0.00 g
Installation and Putting into Operation22 Jewelry Balances
Deactivate customer fine adjustment1 To carry out this operation press and hold CAL until ADJUST appears.
2 Select ADJ.CF by pressing .
ð ADJ.CF appears on the display.
3 Select RESET.
4 Start RESET by pressing .
ð NO? appears.
5 Select YES? and confirm with .
ð The adjusting is finished when the message ADJ.DONE appears briefly on the display. The balancereturns to the last active application and is ready for operation with initial adjustment.
4.8 Performing a simple weighing1 Press to zero the balance.
Note: If your balance is not in the weighing mode, press and holdthe key down until WEIGHING appears in the display. Releasethe key. Your balance is in the weighing mode and set to zero.
2 Place weighing sample on the weighing pan.3 Wait until the instability detector disappears and the stability
beep sounds.4 Read the result.
1250.38
0.00
1182.03 C
¢¢¢¢¢¢¢¢
¢¢¢¢¢¢
# 0.00
Zeroing1 Unload the balance.
2 Press to set the balance to zero. All weight values aremeasured in relation to this zero point.
Use the zeroing key before you start with a weighing.
121.01 g¢¢¢¢
# 0.01 g
0.00 g
C
Taring§ If you are working with a weighing container, first set the balance
to zero.1 Place empty container on the weighing pan. The weight is
displayed.
2 Press to set the balance to zero.
ð 0.00 g appears in the display.
3 Place weighing sample into the weighing container.ð The result appears in the display.
C
121.01 g
0.00 g
0.00 g
95.97 g
#
C
Switching weight unitsThe key can be used at any time to toggle between weight unitUNIT 1, RECALL value (if selected) and weight unit UNIT 2 (ifdifferent from weight unit 1) and the application unit (if any).
0.78 oz
22.00 g
¢¢¢¢
*
22.00 g¢¢¢¢
M
Installation and Putting into Operation 23Jewelry Balances
Recall / Recall weight valueRecall stores stable weights with an absolute display value bigger than 10d.Requirement: The function RECALL must be activated in the menu.
1 Load weighing sample. The display shows weight value andstores stable value.
2 Remove weighing sample. When the weight is removed thedisplay shows zero.
3 Press . The display shows last stored stable weight value for5 seconds together with asterisk (*) and memory (M) symbols.After 5 seconds the display goes back to zero. This can berepeated unlimited times.
Delete last weight valueAs soon a new stable weight value is displayed, the old recall valuebecomes replaced by the new weight value. When pressing
, the recall value is set to 0.
Note: If the power is switched off, the recall value is lost. The recallvalue can not be printed.
121.01 g¢¢¢¢
0.00 g
0.00 g
M
121.01 g
0.00 g
C
C
Weighing with the weighing-in aidThe weighing-in aid is a dynamic graphic indicator which shows theused amount of the total weighing range. You can thus recognize ata glance when the load on the balance approaches the maximumload.
¢¢¢¢¢¢¢¢
0% 100%
Print / Transmit dataPressing the key transmits the weighing results over the interfacee.g. to a printer or a PC.
Switching off into standby mode− Press and hold the key until STANDBY appears on the display.
Release the key.ð MT.GREEN appears on the display.
0.00 g
MT.GREEN
Note• Once your balance has been switched off, it is in energy saver mode STANDBY. In this case your
balance needs no warm-up time in the standby mode and is immediately ready for weighing.If you wish to perform a weighing, you only need to place the sample on the weighing pan and thebalance immediately displays the result. There is no need to switch it on with the key (with approvedbalances only possible in selected countries).
• To completely switch off the balance, disconnect it from the power supply.
4.9 Weighing below the balanceYour balance is equipped with a hanger for carrying out weighings below the work surface (weighing belowthe balance).
NOTICEDamage to balanceDo not place the balance on the pan support location bolt.
Installation and Putting into Operation24 Jewelry Balances
1 Press and hold the key.2 Disconnect the balance from the power supply.3 Disconnect all interface cables.
1 Remove weighing pan, pan support and draft shield element ifpresent.
2 Turn the balance carefully on its side.3 Remove the cap. Keep it for later use.4 Turn the balance to its normal position and simply reinstall all
components in the reverse order.
4.10 Transporting the balanceSwitching off the balance1 Press and hold the key.2 Disconnect the balance from the power supply.3 Disconnect all interface cables.
Transport over short distancesTo move the balance over a short distance to a new location, follow the instructions below.
NOTICEDanger of damage to the glass draft shield!Glass parts of the balance can be damaged.− Never lift the balance by the glass draft shield. The draft shield is not sufficiently
fastened to the balance.
Transporting over long distancesThe complete original packaging must be used for transportation or shipment of the balance over longdistances or if it cannot be ensured that the balance will be transported upright.
The Menu 25Jewelry Balances
5 The Menu
5.1 What is in the menu ?The menu allows you to match your balance to your specific weighing needs. In the menu you can changethe settings of your balance and activate functions. The main menu has 4 different menus and thesecontains different topics, each of which allows you various selection possibilities.
For menu PROTECT see [Main menu Page 26].
Menu BASIC
Topic Explanation Description
DATE Setting the current date. [see Page 26]
TIME Setting the current time. [see Page 26]
1/10 D Setting display increment (1/10d function) [see Page 26]
UNIT 1 Specification of the 1st weight unit in which the balance should showthe result.
[see Page 26]
UNIT 2 Specification of the 2nd weight unit in which the balance shouldshow the result.
[see Page 27]
SET ID Setting an identification. [see Page 27]
PRT.MENU Printing the settings. [see Page 27]
RESET Call up of the factory settings. [see Page 27]
Menu ADVANCE.
Topic Explanation Description
ENVIRON. Matching the balance to the ambient conditions. [see Page 27]
ADJ.LOCK Switching the adjustment function on or off. [see Page 28]
DATE.FRM Setting the date format. [see Page 28]
TIME.FRM Preselection of the time format. [see Page 28]
RECALL Switching the application "Recall" for storing stable weights on oroff.
[see Page 28]
STANDBY Setting the time after which the balance should be switched offautomatically.
[see Page 28]
B.LIGHT Switching on or off the display backlight. [see Page 28]
A.ZERO Switching the automatic zero correction (Autozero) on or off. [see Page 29]
SRV.ICON Switching the service reminder (service icon) on or off. [see Page 29]
SRV.D.RST Reset service date and hours (service reminder) [see Page 29]
Menu INT.FACE
Topic Explanation Description
RS232 Matching the serial interface RS232C to a peripheral unit. [see Page 29]
HEADER Setting the header for printout of individual values. [see Page 30]
SIGN.L Setting the footer for printout of individual values. [see Page 30]
LN.FEED Setting line feed for printout of individual values. [see Page 30]
ZERO.PRT Setting the auto print function for printing zero. [see Page 30]
COM.SET Setting the data communication format of the serial interfaceRS232C.
[see Page 30]
BAUD Setting the transfer speed of the serial interface RS232C. [see Page 31]
BIT.PAR. Setting the character format (Bit/Parity) of the serial interfaceRS232C.
[see Page 31]
The Menu26 Jewelry Balances
Topic Explanation Description
STOPBIT Setting the character format (stop bit) of the serial interface RS232C. [see Page 32]
HD.SHK Setting the transfer protocol (Handshake) of the serial interfaceRS232C.
[see Page 32]
RS.TX.E.O.L. Setting the end of line format of the serial interface RS232C. [see Page 32]
RS.CHAR Setting the char set of the serial interface RS232C. [see Page 32]
INTERVL. Selection of the time interval for the simulated print key press. [see Page 32]
5.2 Description of menu topicsIn this section you will find information regarding the individual menu topics and the available selections.
5.2.1 Main menuSelecting the submenu.
BASIC The small BASIC menu for simple weighing is displayed.
ADVANCE. The extended ADVANCE. menu for further weighing settings isdisplayed.
INT.FACE The menu INT.FACE for all interface parameter settings forperipheral devices e.g. printer is displayed.
PROTECT Menu protection. Protection of balance configurations againstunmeant manipulation.
OFF Menu protection is off. (Factory setting)
ON Menu protection is on. The menu BASIC, ADVANCE. andINT.FACE are not displayed. This is indicated with in thedisplay.
5.2.2 Basic menuDATE – DateSetting the current date according to date format.
Important: A reset of the balance will not change this setting.
TIME – TimeSetting the current time according to time format
+1H Set the current time forwards by 1 hour. (Factory setting)-1H Set the current time backwards by 1 hour.
SET.TIME Enter the current time.
Important: A reset of the balance will not change this setting.
1/10 D – Display increment 1/10 dThis menu topic allows you to reduce the readability of the display.
Important: This menu topic is not available with models which are approved and e=d.
OFF 1/10 D display increment is switched off (full resolution)(Factory setting)
ON 1/10 D switched on (low resolution)
Important: A reset of the balance will not change this setting.
UNIT 1 – Weight unit 1Depending on requirements, the balance can operate with the following units (depending on the model)
The Menu 27Jewelry Balances
• Only those weight units allowed by the appropriate national legislation are selectable.• With approved balances, this menu topic has a fixed setting and cannot be changed.
Units:g Gram dwt Pennyweight
kg Kilogram mom Momme
mg Milligram msg Mesghal
ct Carat tlh Tael Hong Kong
lb Pound tls Tael Singapore
oz Ounce (avdp) tlt Tael Taiwan
ozt Ounce (troy) tola Tola
GN Grain baht Baht
UNIT 2 – Weight unit 2If it is required to show the weighing results in weighing mode in an additional unit, the desired secondweight unit can be selected in this menu topic (depending on the model). Units see UNIT 1.
Important: Only those weight units allowed by the appropriate national legislation are selectable.
SET ID – Set identificationThis menu topic allows you to set your own desired identification to the balance for the convenience ofasset management or other purposes. The ID can be printed with other balance information. One ID can beset and max 7 alphanumeric characters are possible (blank, 0…9, A…Z).
SET ID Set identificationThe setting starts from left to right and the display prompts theconfigurable position by flashing corresponding place.
§ SET ID is selected.
1 Search through (blank, 0…9, A…Z) by pressing .
2 After selecting the character, press to confirm and moveto the next place. To store press and hold .
Important: A reset of the balance will not change this setting.
PRT.MENU – Print menuThis menu topic allows you to execute a printout of the menu settings if a printer is connected. This topic isonly visible if PRINTER mode is selected.
§ PRT.MENU appears on the display and a printer is properly connected.
− To execute a printout press .
RESET – Reset balance settingsThis menu topic allows you to call-up the factory settings.
To toggle between YES? and NO? press .
Important: A reset of the balance will not change DATE, TIME, 1/10 D, and SET ID settings.
5.2.3 Advanced menuENVIRON. – Environment settingsThis setting can be used to match your balance to the ambient conditions.
STD. Setting for an average working environment subject to moderatevariations in the ambient conditions. (Factory setting)
UNSTAB. Setting for a working environment where the conditions arecontinuously changing.
STABLE Setting for a working environment which is practically free fromdrafts and vibrations.
The Menu28 Jewelry Balances
ADJ.LOCK – Adjustment (calibration) lockUnder this menu topic you can lock function of the Cal key.
OFF The adjustment lock is switched off. The adjustment function ison. The Cal key is activ. (Factory setting)
ON The adjustment lock is switched on. The adjustment function isoff. The Cal key has no function.
DATE.FRM – Date formatThis menu topic allows you to preselect the date format.The following date formats are available:
Display examples Printing examplesDD.MM.Y 01.02.09 01.02.2009
MM/DD/Y 02/01/09 02/01/2009
Y-MM-DD 09-02-01 2009-02-01
D.MMM Y 1.FEB.09 1.FEB 2009
MMM D Y FEB.1.09 FEB 1 2009
Factory setting: DD.MM.Y
TIME.FRM – Time formatThis menu topic allows you to preselect the time format.The following date formats are available:
Display examples24:MM 15:04
12:MM 3:04 PM
24.MM 15.04
12.MM 3.04 PM
Factory setting: 24:MM
RECALL – RecallThis menu topic allows you to switch the RECALL function on or off. When it is switched on recall stores thelast stable weight if the absolute display value was bigger than 10d.
OFFON
RECALL switched off. (Factory setting)RECALL switched on.
Important: The recall value is displayed with an asterisk and cannot be printed.
STANDBY – Automatic standbyIf the automatic standby function is activated, the balance automatically switches itself after a pre selectedtime of inactivity into the energy saver mode STANDBY (e.g. with no key being pressed and no changes ofweight occurring).
A.OFF Automatic standby deactivated.
A.ON Automatic standby activated (Factory setting).
60 Time in minutes of inactivity for activating standby function. Setting range: 2…720 minutes.
B.LIGHT – BacklightUnder this menu topic, the display backlight can be switched off or on.
B.L. ON Backlight is always on. (Factory setting)
B.L. OFF Backlight is always off.
The Menu 29Jewelry Balances
A.ZERO – Automatic zero settingThis menu topic allows you to switch the automatic zero setting on or off.
ON A.ZERO switched on (factory setting). The automatic zerosetting continuously corrects possible variations in the zeropoint that might be caused through small amounts of contami-nation on the weighing pan.
OFF A.ZERO switched off. The zero point is not automaticallycorrected. This setting is advantageous for special applications(e.g. evaporation measurements).
Important: With approved balances, this setting is not available (only available in selected countries).
SRV.ICON – Service reminderThis menu topic allows you to switch the service reminder on or off.
ON Service reminder switched on. You will be informed to callservice for recalibration. This will be indicated by the flashingservice icon: . (Factory setting)
OFF Service reminder switched off.
SRV.D.RST – Service date resetThis menu topic allows you to reset service date.Important: This menu topic is only available if SRV.ICON setting ON was selected.
To toggle between YES? and NO? press .
5.2.4 Interface menuRS232 – RS232C interfaceAt this menu topic you can select the peripheral device connected to the RS232C interface and specify howthe data is transmitted.
PRINTER Connection to a printer. (Factory setting)NoteOnly one printer possible.
Refer to your printer documentation for recommended printersettings.
PRT.STAB If the key is pressed, the next stable weight value will beprinted. (Factory setting)
PRT.AUTO Every stable weight value will be printed, without pressing the key.
PRT.ALL If the key is pressed, the weight value will be printedregardless of stability.
PC-DIR. Connection to a PC: the balance can send data (as aKeyboard) to the PC used for PC applications e.g. Excel.
Note• The balance sends the weight value without the unit to the
PC.• Not available on Win7.
PRT.STAB If the key is pressed, the next stable weight value will besent followed by an enter. (Factory setting)
PRT.AUTO Every stable weight value will be sent followed by an enter,without pressing the key.
The Menu30 Jewelry Balances
PRT.ALL If the key is pressed, the weight value will be sent followedby an enter regardless of stability.
HOST Connection to a PC, Barcode Reader etc.: the balance can senddata to the PC and receive commands or data from the PC.Note The balance sends the complete MT-SICS answer to thePC (see chapter "MT-SICS interface commands and functions".
SND.OFF Send mode switched off. (Factory setting)
SND.STB If the key is pressed, the next stable weight value will besent.
SND.CONT All weight value updates will be sent regardless of stability,without pressing the key.
SND.AUTO Every stable weight value will be sent, without pressing the key.
SND.ALL If the key is pressed, the weight value will be sentregardless of stability.
2.DISP Connection of an optional auxiliary display unit.Important: The transmission parameters cannot be selected.Settings are automatically set.
HEADER – Options for the printout header of individual valuesThis menu topic allows you to specify the information that is to be printed at the top of the printout for everyindividual weighing results (after pressing ).Important: This menu topic is only available if PRINTER setting was selected.
NO The header is not be printed. (Factory setting)
DAT/TIM Date and time are printed.
D/T/BAL Date, time and balance information (Balance type, SNR,Balance ID) are printed.
Note: Balance ID only if set.
SIGN.L – Options for the printout footer for signature line of individual valuesThis menu topic allows you to set a footer for signature at the bottom of the printout for every individualweighing result (after pressing ).Important: This menu topic is only available if PRINTER setting was selected.
OFF The signature footer is not be printed. (Factory setting)
ON The signature footer is printed.
LN.FEED – Options for complete the printout of individual valuesThis menu topic allows you to specify the number of blank lines to complete the printout (line feed) forevery individual weighing result (after pressing ).Important: This menu topic is only available if PRINTER setting was selected.
0 Possible numbers of blank lines: 0 to 99. (Factory setting =0)
ZERO.PRT – Options for PRT.AUTOThis menu topic allows you to specify the auto print function PRT.AUTO for printing zero YES or NO.
OFF Zero is not be printed (Zero +/- 3d). (Factory setting)
ON Zero is always printed.
Important: This menu topic is only available if PRT.AUTO function of the PRINTER or PC-DIR. was selected.
COM.SET – Options for the data communication format (RS232C) (HOST)This menu topic allows you to set the data format depending on which peripheral device is connected. Important: This menu topic is only available if HOST setting was selected.
The Menu 31Jewelry Balances
MT-SICS The MT-SICS data transfer formats is used. (Factory setting) For more information see section “MT-SICS interface commandsand functions”.
SART The following Sartorius commands are supported:K Ambient conditions: very stableL Ambient conditions: stableM Ambient conditions: unstableN Ambient conditions: very unstableO Block keysP Print key (print, auto print; activate or block)R Unblock keysS Restart/self-testT Tare keyW Calibration/adjustment *)
Z Internal calibration/adjustment **)
f1_ Function key (CAL)s3_ C keyx0_ Perform internal calibration **)
x1_ Print balance/scale modelx2_ Print weighing cell serial numberx3_ Print software version*) May be inaccessible on verified balances/scales**) Only on models with built-in motorized calibration weight
Functionality mappingHOST settings: Sartorius printer settings:SND.OFF not applicable
SND.STB manually print with stability
SND.ALL manually print without stability
SND.CONT automatically print without stability
SND.AUTO similar applicable to automatically printwhen load is changed
BAUD – Baud rate RS232CThis menu topic allows you to match the data transmission to different serial RS232C receivers. The baudrate (data transfer rate) determines the speed of transmission via the serial interface. For problem-free datatransmission the sending and receiving devices must be set at the same value.The following settings are available: 600 bd, 1200 bd, 2400 bd, 4800 bd, 9600 bd (Factory setting), 19200 and 38400 bd.
Important• Not visible for 2nd display.• Each device has separate settings.
BIT.PAR. – Bit/Parity RS232CAt this menu topic you can set the character format for the attached RS232C serial peripheral device.
8/NO 8 data bits/no parity (Factory setting)
7/NO 7 data bits/no parity
The Menu32 Jewelry Balances
7/MARK 7 data bits/mark parity
7/SPACE 7 data bits/space parity
7/EVEN 7 data bits/even parity
7/ODD 7 data bits/odd parity
Important• Not visible for 2nd display.• Each device has separate settings.
STOPBIT – Stop bits RS232CAt this menu topic you can set the stop bits of the transmitted data to different RS232C serial receivers.
1 BIT 1 Stop bit (Factory setting)
2 BITS 2 Stop bits
HD.SHK – Handshake RS232CThis menu topic allows you to match the data transmission to different RS232C serial receivers.
XON.XOFF Software handshake (XON/XOFF) (Factory setting)
RTS.CTS Hardware handshake (RTS/CTS)
OFF No handshake
Important• Not visible for 2nd display.• Each device has separate settings.
RS.TX.E.O.L. – End of line RS232CAt this menu topic you can set the end of line character of the outgoing transmitted data to different RS232Cserial receivers.
CR LF <CR><LF> Carriage Return followed by Line feed (ASCII-Codes013 + 010) (Factory setting)
CR <CR> Carriage Return (ASCII-Code 013)
LF <LF> Line feed (ASCII-Code 010)
TAB <TAB> Horizontal tab (ASCII-Code 009) (only visible if PC-DIR. is selected)
Important• Not visible for 2nd display.• Each device has separate settings.
RS.CHAR – Char set RS232CAt this menu topic you can set the character set of the transmitted data to different RS232C serial receivers.
IBM.DOS Char set IBM/DOS (Factory setting)
ANSI.WIN Char set ANSI/WINDOWS
Important• Not visible for 2nd display.• Each device has separate settings.
INTERVL. – Print key simulationAt this menu topic you can activate a simulation of the key. INTERVL. simulates a print key press everyx seconds.
The Menu 33Jewelry Balances
Range: 0 to 65535 seconds0 sec: Disables the print key simulation
Factory setting: 0 sec
Important: The executed action is according to the configuration of the print key, (see interface setting).
Applications34 Jewelry Balances
6 Applications
6.1 Application piece counting
The Piece counting application allows you to determine the number of pieces put on theweighing pan. All pieces must be of approximately equal weight, since the number ofpieces is determined on the basis of average weight.
1 Call-up APP.LIST by pressing and holding F.
2 Select application COUNT by scrolling with .
3 Activate function COUNT by pressing .
APP.LIST
COUNT
Piece counting first requires the setting of a reference weight, there are 4 possibilities:
DA B C
• A Setting the reference by multiple pieces with fix reference values.
• B Setting the reference by multiple pieces with variable reference values.
• C Setting the reference for 1 piece in weighing mode.
• D Setting the reference for 1 piece in manual mode.
Setting the reference by multiple pieces with fix reference values
FIX 10PCSC
10PCS
------
FIX 10PCSA
1 Select a number of reference pieces by scrolling with . Possible numbers* are 5, 10, 20 and 50. * with approved balances in selected countries: min 10.
2 Press to tare. If using: place empty container on the weighing pan first or tare again.3 Add the selected number of reference pieces to container.4 Press to confirm.
Applications 35Jewelry Balances
Setting possibility
B Setting the reference by multiple pieces with variablereference values1 Select VAR.REF by scrolling with . Press to confirm.
2 Select the number of reference pieces. Possible numbers are 1 to999. With approved balances in selected countries: min 10
3 To select a digit, press (cyclically from left to right).ð The selected digit is blinking.
4 To change the digit, press .
5 Press to zero/tare. If using: place empty container onthe weighing pan first or zero/tare again.
6 Add the selected number of reference pieces to container.
7 Press and hold to confirm.
C
------
VAR.REF
25PCS
B
015 PCSREF.
025 PCSREF.
Setting the reference for one piece in weighing mode
0.00 g
1PCS
2.74 g
1
C
---------
PCS.WGT
C
1 Select PCS.WGT by scrolling with .
2 Press to tare. If using: place empty container on the weighing pan first or tare again.3 Add one reference piece to container. The weight of one piece is displayed.4 Press to confirm.
ImportantWith approved balances, this setting is not available in selected countries.
Applications36 Jewelry Balances
Setting possibility
D Setting the reference for one piece in manual mode1 Select PCS.WGT by scrolling with .
2 Press to confirm.3 Enter the final reference one piece weight.
4 To select a digit, press (cyclically from left to right).ð The selected digit is blinking.
5 To change the digit, press .
6 Press and hold to confirm.
Important: With approved balances, this setting is not available inselected countries.
.68 g*
0 PCS
0002 .74 g*
0002
PCS.WGT
D
Important: If without any key press within 60 seconds, the balance return to the previous active application.Press C to cancel and returns to the previous active application.
On completion of the setting procedure, your balance is ready for piece counting.
Note• The RECALL value is displayed with an asterisk (*) and icon M and can not be printed.
• Take into account minimum values: min. reference weight = 10d (10 digits), min. piece weight* = 1d(1 digit)!* with approved balances in selected countries: min 3e
• The current reference weight remains stored until the reference setting is changed.
Exit current applicationTo exit the current application and to return to simple weighing mode, press and hold (longer than1.5 s).
Applications 37Jewelry Balances
6.2 Application percent weighing
The Percent weighing application allows you to check a sample weight as percentage toa reference target weight.
1 Call-up APP.LIST by pressing and holding F.
2 Select application PERCENT by scrolling with .
3 Activate function PERCENT by pressing .
APP.LIST
PERCENT
Percent weighing first requires the setting of a reference weight that should corresponds to 100%,there are 2 possibilities:
A B
• A Setting the reference in manual mode (enter 100%).• B Setting the reference in weighing mode (weigh 100%).
Setting possibility
A Setting the reference by manual mode (enter 100%)1 Press to activate manual mode.
2 To select a digit, press (cyclically from left to right).ð The selected digit is blinking.
3 To change the digit, press .
4 Press and hold to confirm the value. 0.000%
0175 .28 g
A SET 100%
Setting the reference by weighing mode (weigh 100%)
I00.0%
175.28 g
¢¢¢¢
SET 100%
C
0.00 g
B
1 Press to tare the balance and to activate the weighing mode. If needed: place emptycontainer on the weighing pan and tare again.
2 Load the reference weight (100%).Note: Reference weight must be at least +/- 10d.
3 Press to confirm.
Important: if without any key press within 60 seconds or by pressing C, the balance returns to the previousactive application.
On completion of the weighing-in procedure, your balance is ready for percent weighing.
Applications38 Jewelry Balances
Switching between percent and weight display
,
128.25 g
73.17%
128250 mg
128.25 gM
− Press key at any time to switch the display between percent display, weighing unit UNIT 1, RECALLvalue (if activated) and weighing unit UNIT 2 (if different from UNIT 1).
Note• The recall value is displayed with an asterisk (*) as well as icon M and can not be printed.
• The current set weight remains stored until it is redetermined.
Exit current applicationTo exit the current application and to return to simple weighing mode, press and hold (longer than1.5 s).
Applications 39Jewelry Balances
6.3 Application check weighing
The Check weighing application allows you to check the deviation of a sample weightwithin a tolerance limit to a reference target weight.
1 Call-up APP.LIST by pressing and holding F.
2 Select application CHECK by scrolling with .
3 Activate function CHECK by pressing .
APP.LIST
CHECK
Step 1: Check weighing first requires the setting of a referenceweight that should corresponds to the nominal weight, there are 2possibilities:1A Setting the reference in manual mode (enter nominal weight).1B Setting the reference in weighing mode (weigh nominal weight).
Step 2: Check weighing needs the upper and lower limits:2 Setting the upper and lower limits in percentage.
1A
21B
Step 1, setting possibility:
1A Setting the reference by manual mode (enter nominalweight)
1 Press to activate manual mode.2 Select the reference target weight.
3 To select a digit, press (cyclically from left to right).ð The selected digit is blinking.
4 To change the digit, press .
5 Press and hold to confirm the nominal weight.
SET NOM .
0010.00 g
0020.28 g
1A
Step 1, setting possibility:
1B Setting the reference by weighing mode (weigh nominalweight)
1 Press to tare the balance and to activate the weighingmode. If using: place empty container on the weighing pan first ortare again.
2 Load the nominal weight.3 Press to confirm the nominal weight.
1B SET NOM .
0.00 g
20.28 g
C
Applications40 Jewelry Balances
Step 2:
2 Setting the upper and lower limits (in percentage):
1 Press to start setting.
2 Press to confirm the default limit of +/- 2.5% or enter the limitvalue.
3 To select a digit, press (cyclically from left to right).ð The selected digit is blinking.
4 To change the digit, press .
5 Press and hold to confirm the limits.
SET.T+/- %
002.50 %
20.28 g
2
Important• If without any key press within 60 seconds, the balance returns to the previous active application. Press
C to cancel and returns to the previous active application.
• The nominal weight must be at least 10 digit.
On completion of the setting procedure, your balance is ready for checkweighing.
Weighing-in-AidThe Weighing-in-Aid helps you quickly determine the position of thesample weight regarding the tolerance.
1 Lower limit2 Target weight3 Upper limit
20.15 g
Exit current applicationTo exit the current application and to return to simple weighing mode, press and hold (longer than1.5 s).
Applications 41Jewelry Balances
6.4 Application statistics
The Statistics application allows you to generate statistics of a series of weighing values.1 to 999 values are possible.
1 Call-up APP.LIST by pressing and holding F.
2 Select application STAT. by scrolling with .
3 Activate function STAT. by pressing .
APP.LIST
STAT
Memory clear questionIf the memory is already cleared (sample counter is 0) the memoryclear question will not be displayed.
1 To continue the last statistics press to confirm CLR.M:NO.
2 For a new statistical evaluation clear the memory. Press toselect CLR.M:YES and press to confirm.
CLR. M:NO
CLEARED
Weighing the first sample weight1 Press to zero/tare the balance if needed.2 Load the first sample weight.
3 Press . The display shows the sample count - 1 - and thecurrent weight is stored as sample and the weight is printed out.Note: When the sample counter is displayed you may press C toundo (drop) this sample.
4 Unload the first sample weight.
Weighing further sample weightsThe same procedure as for the first sample weight.• 1…999 samples are possible.• The next value will be accepted if the sample weight is in the
range 70% – 130% of the current average value. OUT OF RANGEwill be displayed if the sample is not accepted.
46.36 g
0.00 g
- 1 -
C
999
C
0.00 g
Results
− Press if the numbers of sample are greater than or equal to 2.ð Results are displayed and printed.
Applications42 Jewelry Balances
Displayed results1 Press to show the next
statistical value.
2 Press C to cancel displayingresults and to continueweighing next sample.
0.5 secondsnumber of samples
5 N
average X 50.530 g
standard deviation S.DEV 3.961 g
relative standard deviation S.REL 7.84 %
lowest value (minimum) MIN 46.36 g
highest value (maximum) MAX 55.81 g
different between the minimumand the maximum
DIFF 9.45 g
sum of all values SUM 252.65 g
Exit current applicationTo exit the current application and to return to simple weighing mode, press and hold (longer than1.5 s).
Applications 43Jewelry Balances
6.5 Application totaling
The TOTALING application allows you to weigh in different samples to add their weightvalues and to totalize them. 1 to 999 samples are possible.
1 Call-up APP.LIST by pressing and holding F.
2 Select application TOTAL by scrolling with .
3 Activate function TOTAL by pressing .
APP.LIST
TOTAL
Memory clear questionIf the memory is already cleared (sample counter is 0) the memoryclear question will not be displayed.
1 To continue the totaling evaluation press to confirmCLR.M:NO.
2 For a new totaling evaluation clear the memory. Press toselect CLR.M:YES and press to confirm.
CLR. M:NO
CLEARED
Weighing in the sample weight1 If using a container: place empty container on the weighing pan
and press to zero or tare the balance.2 Load the first sample weight.
3 Press . The display shows the sample count - 1 - and thecurrent weight is stored.Note: When the sample counter is displayed you may press C toundo (drop) this sample.
4 Unload the first sample weight. The display shows zero.
Weighing in further sample weightsThe same procedure as for the first sample weight.• 1…999 samples are possible.
46.36 g
- 1 -
999
0.00 g
0.00 g
C
C
Results
− Press if the numbers of sample are greater than or equal to 2.ð Results are displayed and printed.
Displayed results1 Press briefly to show the
totalized value.
2 Press C briefly to cancel.
0.5 secondsnumber of samples
879 N
totalized value TOTAL 8789.79 g
Exit current applicationTo exit the current application and to return to simple weighing mode, press and hold (longer than1.5 s).
Applications44 Jewelry Balances
6.6 Application multiplication factor weighing
The Multiplication factor weighing application allows you to multiply the weight value (ingrams) by a predefined factor (result = factor * weight) and have it calculated to apredefined number of decimal places.
1 Call-up APP.LIST by pressing and holding F.
2 Select application FACTOR.M by scrolling with .
3 Activate function FACTOR.M by pressing .
APP.LIST
FACTOR.M
1 Setting the factor value1 Press to execute SET.F.MUL. Either the factor 1 appears as
default value or the factor that was saved most recently.
2 To select a digit, press (cyclically from left to right).ð The selected digit is blinking.
3 To change the digit, press .
4 Press and hold to confirm the selected factor (no automaticacceptance).
Important: Zero for multiplication factor value is outside the allowedrange, the error message FACTOR OUT OF RANGE will be displayed.
SET.F.MUL
001.0000
009.8100
2 Setting the step valueSET.STEP appears in the display, and the program changesautomatically to allow the display increments to be entered. Thesmallest possible display increment appears as default value, or thelast value that was saved.
1 Press to execute SET.STEP.
2 To select a digit, press (cyclically from left to right).ð The selected digit is blinking.
3 To change the digit, press .
4 Press and hold to confirm the selected step (no automaticacceptance).
Important: The allowed range for the step depends on the factor andthe resolution of the balance. If it is outside the allowed range theerror message STEP OUT OF RANGE will be displayed.
SET.STEP
001.0
005.0
Important: If without any key press within 60 seconds, the balance return to the previous active application.Press C to cancel and returns to the previous active application.
On completion of the setting procedure, your balance is ready for multiplication factor weighing.
Applications 45Jewelry Balances
Weighing procedure1 Press to zero/tare.2 Load sample weight on weighing pan.3 Read the result. The appropriate calculation is then made using
the weight of sample and the selected factor, the result beingdisplayed with the selected display step. Note: No units are displayed.
4 Unload sample weight.
Toggling between displaying the calculated value and themeasured weight You can use the key to toggle between the calculated Value,weight value UNIT 1, RECALL value (if selected) and weight valueUNIT 2 (if different from UNIT 1).
C
9810
1000.02 g
0
C
Exit current applicationTo exit the current application and to return to simple weighing mode, press and hold (longer than1.5 s).
Applications46 Jewelry Balances
6.7 Application division factor weighing
The Division factor weighing divide a predefined factor by the weight value (in grams)(result = factor / weight) and have it rounded to a predefined number of decimal places.
1 Call-up APP.LIST by pressing and holding F.
2 Select application FACTOR.D by scrolling with .
3 Activate function FACTOR.D by pressing .
APP.LIST
FACTOR.D
1 Setting the factor value1 Press to execute SET.F.DIV. Either the factor 1 appears as
default value or the factor that was saved most recently.
2 To select a digit, press (cyclically from left to right).ð The selected digit is blinking.
3 To change the digit, press .
4 Press and hold to confirm the selected factor (no automaticacceptance).
Important: Zero for division factor value is outside the allowedrange, the error message FACTOR OUT OF RANGE will be displayed.
SET.F.DIV
001.0000
009.8100
2 Setting the step valueSET.STEP appears in the display, and the program changesautomatically to allow the display increments to be entered. Thesmallest possible display increment appears as default value, or thelast value that was saved.
SET.STEP appears in the display, and the program changesautomatically to allow the display increments to be entered. Thesmallest possible display increment appears as default value, or thelast value that was saved.
1 Press to execute SET.STEP.
2 To select a digit, press (cyclically from left to right).ð The selected digit is blinking.
3 To change the digit, press .
4 Press to confirm the selected step (no automatic acceptance).
Important: The allowed range for the step depends on the factor andthe resolution of the balance. If it is outside the allowed range theerror message STEP OUT OF RANGE will be displayed.
SET.STEP
0.0001
0.0050
Important: If without any key press within 60 seconds, the balance return to the previous active application.Press C to cancel and returns to the previous active application.
On completion of the setting procedure, your balance is ready for division factor weighing.
Applications 47Jewelry Balances
Weighing procedure1 Press to zero/tare.2 Load sample weight on weighing pan.3 Read the result. The appropriate calculation is then made using
the weight of sample and the selected factor, the result beingdisplayed with the selected display step. Note: No units are displayed. To avoid a division by zero, thefactor division is not calculated at zero.
4 Unload sample weight.
Toggling between displaying the calculated value and themeasured weight You can use the key to toggle between the calculated Value,weight value UNIT 1, RECALL value (if selected) and weight valueUNIT 2 (if different from UNIT 1).
C
0.010
1000.02 g
0.000
C
Exit current applicationTo exit the current application and to return to simple weighing mode, press and hold (longer than1.5 s).
Applications48 Jewelry Balances
6.8 Application density
The Density application allows you to determine the density of solid bodies and liquids.Determination of the density uses Archimedes' principle according to which a bodyimmersed in a fluid undergoes an apparent loss in weight which is equal to the weight ofthe fluid it displaces.To determine the density of solid bodies, we recommend you to work with the optionaldensity kit which contains all the attachments and aids needed for convenient and precisedensity determination. To determine the density of liquids, you additionally need a sinkerwhich you can also obtain from your METTLER TOLEDO dealer.
Note for performing of density determinations• You can also use the hanger for weighing below the balance which belongs to your
balance.• If a METTLER TOLEDO printer is attached to your balance, the settings will be automat-
ically recorded.
We recommend you to consult the operating instructions enclosed with thedensity kit.
1 Call-up APP.LIST by pressing and holding F.
2 Select application DENSITY by scrolling with .
3 Activate function DENSITY by pressing .
APP.LIST
DENSITY
Setting the method for density determination1 Select:
SOLID, the function for the density determination of solids, orLIQUID, the function for the density determination of liquids with asinker.
2 Press to confirm the selection
SOLID
LIQUID
Exit current applicationTo exit the current application and to return to simple weighing mode, press and hold (longer than1.5 s).
6.8.1 Density determination of solidsRequirement: the method SOLID is set.
Setting the parameter of the auxiliary liquid1 Select the auxiliary liquid by scrolling with : H-2-O for distilled
water, ETHANOL or FREE for a freely definable auxiliary liquid.
2 Press to confirm the selection.
H-2-0
ETHANOL
FREE
Applications 49Jewelry Balances
If you have selected water or ethanol as the auxiliary liquid1 Enter the current temperature of the auxiliary liquid (read off on
thermometer).2 Set the value in °C. The temperature ranges from 10 °C to 30.9
°C.
3 To select a digit, press (cyclically from left to right).ð The selected digit is blinking.
4 To change the digit, press .
5 Press and hold to confirm the value.
Note: the densities of distilled water and ethanol in the range 10 °Cto 30.9 °C are stored in the balance.
20.0
21.8
If you have selected a freely definable auxiliary liquid1 Enter the density of the auxiliary liquid in g/cm3 at the current
temperature (read off on thermometer).
2 To select a digit, press (cyclically from left to right).ð The selected digit is blinking.
3 To change the digit, press .
4 Press and hold to confirm the value.
01.00200
01.00000
Important: if without any key press within 60 seconds or by pressing C, the balance returns to the previousactive application.
On completion of the settings, your balance is ready for performing the density determination ofliquids.Note: taring the balance is possible at any time.
The balance prompts you: PRESS ENTER TO START.
− Press to start. Tare/Zero is executed. PRESS EN
The balance prompts you to weigh the solid in air WEIGH IN AIR.
1 Load the solid.
2 Press to initiate the measurement.
WEIGH IN AIR
The balance prompts you to weigh the solid in the auxiliary liquidWEIGH IN LIQUID.
1 Load the solid.
2 Press to initiate the measurement.
WEIGH IN LIQUID
The balance now shows the determined density of the solid in g/cm3.
Note• This result has already been corrected for the air buoyancy. The
buoyancy caused by the two immersed wires (ø 0.6 mm) can beneglected.
• By pressing C, the balance returns to PRESS ENTER TO START.
5.988*
Result
− Press .ð The result will be printed.
Applications50 Jewelry Balances
6.8.2 Density determination of liquidsRequirement: the method LIQUID is set.
Setting the displacement volume of your sinkerPress and hold to confirm the default value of 10.0 cm3 orchange it if needed:
1 To select a digit, press (cyclically from left to right).ð The selected digit is blinking.
2 To change the digit, press .
3 Press and hold to confirm the value.
0100.000
0120.000
Important: if without any key press within 60 seconds or by pressing C, the balance returns to the previousactive application.
On completion of the settings, your balance is ready for performing the density determination ofliquids.Note: taring the balance is possible at any time.
The balance prompts you: PRESS ENTER TO START.
− Press to start. PRESS EN
The balance prompts you to weigh the sinker in air WEIGH IN AIR.
1 Position the sinker.
2 Press to initiate the measurement.
WEIGH IN AIR
The balance prompts you to weigh the sinker in the liquid WEIGH INLIQUID.
1 Pour the liquid into the beaker. Make sure that the sinker isimmersed by al least 1 cm in the liquid, and that there are no airbubbles in the container.
2 Press to initiate the measurement.
WEIGH IN LIQUID
The balance now shows the determined density of the liquid at thecurrent temperature (read off on the thermometer).
Note
• This result has already been corrected for the air buoyancy. Thebuoyancy caused by the immersed wire (ø 0.2 mm) of the sinkercan be neglected.
• By pressing C, the balance returns to PRESS ENTER TO START.
1.000*
Result
− Press .ð The result will be printed.
6.8.3 Formulae used to calculate densityThe DENSITY"application is based on the formulae listed below.
Applications 51Jewelry Balances
Formulae for determining the density of solids with compensation for air density
= Density of the sample
A = Weight of the sample in airB = Weight of the sample in the auxiliary liquidV = Volume of the sample
= Density of the auxiliary liquid
= Density of air (0.0012 g/cm3)
= Weight correction factor (0.99985), to take the atmospheric buoyancy of the adjustment weightinto account
Formula for determining the density of liquids with compensation for air density
= Density of the liquid
P = Weight of the displaced liquidV = Volume of the sinker
= Density of air (0.0012 g/cm3)
= Weight correction factor (0.99985), to take the atmospheric buoyancy of the adjustment weightinto account
Applications52 Jewelry Balances
Density of C2H5OH according to the "American Institute of Physics Handbook".
Communication with Peripheral Devices 53Jewelry Balances
7 Communication with Peripheral Devices
7.1 Function PC-DirectThe numerical value displayed at the balance can be transferred to the cursor position in Windows Appli-cations (e.g. Excel, Word) as by typing with the keyboard.
Note: The units will not be transferred.
Requirements• PC with one of the Microsoft Windows® operating system 32bit/64bit: XP (SP3), Vista (SP2), Win 7
(SP1) or Win 8.• Serial interface RS232 or USB.• Administrator rights for installing software (for USB not required).• Windows application (e.g. Excel).• Balance to PC connection with cable RS232 or USB.
Settings at the balance
NOTICEDisconnect the USB connection from the balance prior to changing settings.
Balance interface settings, see Interface menu:
• Topic RS232 or USB: set PC-DIR. and select the most appropriate option for the desired weighingresult.
• Topic RS.TX.E.O.L./RS E.O.L. or USB E.O.L./USB E.O.L:– set <TAB> to write into the same row (e.g. in Excel).– set <CR><LF> to write into the same column (e.g. in Excel).
• Save changes.
Installing SerialPortToKeyboardOperation of PC-Direct via serial port RS232 requires the installation of SerialPortToKeyboard on your hostcomputer. The file SerialPortToKeyboard.exe can be found on the CD-ROM in the folder Software. If youhave any questions please contact a METTLER TOLEDO representative.
Download SerialPortToKeyboard
1 Insert the product CD in the CD/DVD drive of the host computer.
2 Tap Software.
3 Download and unpack the .exe file from the CD on yourcomputer.
4 Click Save to download to your specified location.
5 Right-click on the downloaded install program SerialPort-ToKeyboard.exe and select Run as Administrator from the menu.
6 If a safety warning appears, allow Windows to install.
7 Click Next and follow the installer's instructions.
Communication with Peripheral Devices54 Jewelry Balances
Settings at the PCSettings for SerialPortToKeyboard
1 Select the serial port COM to be used for connection with thebalance.
2 Set the Baud Rate to 9600.
3 Activate ConnectNote• The window can be minimized.• Closing of the window terminates the session.
➀
➁
➂
Checking operation1 Start SerialPortToKeyboard (RS232)
2 Start Excel (or another application) at the PC.3 Activate a cell in Excel.
According to your selected PC-DIR. option, the displayed values will appear e.g. in the column one after theother one in the different rows.
Error and Status Messages 55Jewelry Balances
8 Error and Status Messages
8.1 Error messagesError messages in the display draw your attention to incorrect operation or that the balance could notexecute a procedure properly.
Error message Cause Rectification
NO STABILITY No stability. Ensure more stable ambientconditions. If not possible, checksettings for environment.
WRONG ADJUSTMENT WEIGHT Wrong adjustment weight on panor none at all.
Place required adjustment weightin center of pan.
REFERENCE TOO SMALL Reference for piece counting toosmall.
Increase reference weight.
EEPROM ERROR - PLEASECONTACT CUSTOMER SERVICE
• EEPROM (memory) error.• Excessive mains voltage
fluctuation or strong glitchesoccurred.
Please contact METTLER TOLEDOcustomer service.
WRONG CELL DATA - PLEASECONTACT CUSTOMER SERVICE
Wrong cell data. Please contact METTLER TOLEDOcustomer service.
NO STANDARD ADJUSTMENT -PLEASE CONTACT CUSTOMERSERVICE
No standard calibration. Please contact METTLER TOLEDOcustomer service.
PROGRAM MEMORY DEFECT -PLEASE CONTACT CUSTOMERSERVICE
Program memory defect. Please contact METTLER TOLEDOcustomer service.
TEMP SENSOR DEFECT - PLEASECONTACT CUSTOMER SERVICE
Temperature sensor defect. Please contact METTLER TOLEDOcustomer service.
WRONG LOAD CELL BRAND -PLEASE CONTACT CUSTOMERSERVICE
Wrong load cell brand. Please contact METTLER TOLEDOcustomer service.
WRONG TYPE DATA SET -PLEASE CONTACT CUSTOMERSERVICE
Wrong type data set. Please contact METTLER TOLEDOcustomer service.
BATTERY BACKUP LOST - CHECKDATE TIME SETTINGS
Backup battery is empty. Thisbattery ensures that the date andtime are not lost when the balanceis disconnected from power.
Connect the balance to the powersupply for charging the battery(e.g. during the night) or contactMETTLER TOLEDO customerservice.
Overload - The weight on the panexceeds the weighing capacity ofthe balance.
Reduce the weight on theweighing pan.
Underload Check that the weighing pan ispositioned correctly.
ABOVE INITIAL ZERO RANGE Wrong weighing pan or pan is notempty.
Mount correct weighing pan orunload weighing pan.
BELOW INITIAL ZERO RANGE Wrong weighing pan or pan ismissing.
Mount correct weighing pan.
MEM.FULL Memory full. Clear the memory and start a newevaluation.
Error and Status Messages56 Jewelry Balances
Error message Cause Rectification
FACTOR OUT OF RANGE Factor is outside the allow range. Select a new factor.
STEP OUT OF RANGE Step is outside the allowed range. Select a new step.
OUT OF RANGE Sample weight is outside theallowed range.
Unload the pan and load a newsample weight.
NOTICEDamage of the balance or softwareIn some countries, excessive mains voltage fluctuations and strong glitches may occur.This may affect the balance functions or damage the software.− Use a voltage regulator for stabilizing.
8.2 Status messagesStatus messages are displayed by means of small icons. The status icons indicate the following:
Status Icon Signification
Service Reminder Your balance is due for servicing. Contact your dealer's customerservice department as soon as possible to have a technician service your balance.See menu topic SRV.ICON.
Maintenance 57Jewelry Balances
9 Maintenance
9.1 Cleaning and service
Every now and then, clean the weighing pan, draft shield element, bottom plate, draft shield (depending onthe model) and housing of your balance. Your balance is made from high-quality, durable materials andcan therefore be cleaned using a damp cloth or with a standard cleaning agent.To thoroughly clean the draft shield glass panels, remove the draft shield from the balance. Whenreinstalling the draft shield, ensure that it is in the correct position.
Please observe the following notes:
WARNINGDanger of death or serious injury due to electric shock!Contact with parts that contain a live current can lead to injury and death. If the balancecannot be shut down in an emergency situation, people can be injured or the balance canbe damaged.1 Disconnect the balance from the power supply prior to cleaning and maintenance.2 Only use METTLER TOLEDO power cable, if these need to be replaced.3 Make sure that no liquid enters into the balance, terminal or AC/DC adapter.4 Do not open the balance, terminal or AC/DC adapter.
These contain no user-serviceable parts.
NOTICEDanger of damage to the balance due to inappropriate cleaning methods!The balance is made from high quality, resistant materials and can be damaged bycertain cleaning agents, solvents or abrasives. If liquids enter the housing they candamage the balance.1 Use only water and a mild detergent to clean the balance or terminal.2 Wipe off any spills immediately.3 Make sure that no liquid enters the interior of the balance.
ImportantContact a METTLER TOLEDO representative to find about the service options available – regularmaintenance by an authorized service engineer will ensure consistent weighing accuracy over the longterm and extend the service life of the balance.
Maintenance58 Jewelry Balances
9.2 Cleaning the draft shieldIt is possible to remove the sliding glass doors for cleaning or for replacing.
Removing or inserting sliding glass doors
1 Remove the handle first.2 Remove the sliding glass doors.3 Install the handle after insertion of the glass door.
ImportantFront and rear glass panels cannot be removed.
9.3 DisposalIn conformance with the European Directive 2012/19/EU on Waste Electrical andElectronic Equipment (WEEE) this device may not be disposed of in domestic waste. Thisalso applies to countries outside the EU, per their specific requirements. Please dispose of this product in accordance with local regulations at the collecting pointspecified for electrical and electronic equipment. If you have any questions, please contactthe responsible authority or the distributor from which you purchased this device. Shouldthis device be passed on to other parties (for private or professional use), the content ofthis regulation must also be related.Thank you for your contribution to environmental protection.
9.4 Firmware (Software) UpdatesIn the interest of its customers, METTLER TOLEDO continuously updates its internal software (firmware) forthe balances. Inquire at your local METTLER TOLEDO office about upgrade and update options.
Technical Data 59Jewelry Balances
10 Technical Data
10.1 General data
WARNINGDanger of death or serious injury due to electric shock!Contact with parts that contain a live current can lead to injury and death.1 Only use an approved AC/DC adapter with a current-limited SELV output.2 Ensure correct polarity
Standard power supply
AC adapter: Primary: 100 – 240 V, ±10%, 50/60Hz, 0.3 ASecondary: 12 V DC, 0.84 A (with electronic overloadprotection)
Balance power supply: 12 V DC, 0.84 ACan be used up to 2000 m height above mean sea level.
NOTICEIf the balance is used above 2000 m mean sea level, the optional power supply must beused.
Optional power supply
AC adapter: Primary: 100 – 240 V, ±10%, 50/60HzSecondary: 12 V DC ±3%, 2.5 A (with electronic overloadprotection)
Cable for AC adapter: 3-core, with country-specific plugBalance power supply: 12 V DC ±3%, 2.25 A, maximum ripple: 80 mVpp
Can be used up to 4000 m height above mean sea level.
Protection and standards
Overvoltage category: IIDegree of pollution: 2Protection: Protected against dust and waterStandards for safety and EMC: See Declaration of ConformityRange of application: For use only in closed interior rooms
Environmental conditions
Height above mean sea level: Depending on the power adapter (2000 - 4000 m)Except for China: max. 2000 m
Ambient temperature: Operating condition for ordinary lab application: +10 to 30 °C(operability guaranteed between +5 to 40 °C)Storage condition: -25 to 70 °C
Relative air humidity: 10% up to 80% at 31 °C, linearly decreasing to 50% at40 °C, non-condensing
Warm-up time: At least 30 minutes (0.1 mg models 60 minutes) afterconnecting the balance to the power supply
Technical Data60 Jewelry Balances
Materials
Housing: Top housing: Plastic (ABS)Bottom housing: Die-cast aluminum, lacquered
Weighing pan: Pan ø 90 mm: Stainless steel X2CrNiMo 17-12-2 (1.4404)All others: Stainless steel X5CrNi 18-10 (1.4301)
Draft shield element: 0.1 mg models: Stainless steel X5CrNi 18-10 (1.4301)Draft shield: Plastic (ABS), glassIn-use-cover: Plastic (PET)
Technical Data 61Jewelry Balances
10.2 Model-specific data
10.2.1 Carat balances with readability of 0.001 ct / 0.1 mgTechnical data
JE503C JE503CE
Limit valuesMaximum capacity 505 ct / 101 g 505 ct / 101 g
Readability 0.001 ct / 0.1 mg 0.001 ct / 0.1 mg
Repeatability (at nominal load) 0.1 mg 0.1 mg
Linearity deviation 0.2 mg 0.2 mg
Sensitivity temperature drift 2 ppm/°C 2 ppm/°C
Typical valuesRepeatability (at nominal load) 0.08 mg 0.08 mg
Linearity deviation 0.06 mg 0.06 mg
Minimum sample weight (U=1%, k=2) 16 mg 16 mg
Minimum sample weight OIML 10 mg 10 mg
Settling time 2 s 2 s
Adjustment Int. Cal Ext. Cal
Interfaces 1 RS232 1 RS232
Balance dimensions (W × D × H) 210×344×279 mm 210×344×279 mm
Weighing pan dimensions ø 90 mm ø 90 mm
Usable height of draft shield 170 mm 170 mm
Weight of balance 4.6 kg 4.4 kg
Weights for routine testingOIML CarePac
Weights#11123002100 g F2, 5 g E2
#11123002100 g F2, 5 g E2
ASTM CarePacWeights
#11123102100 g 1, 5 g 1
#11123102100 g 1, 5 g 1
JE703C JE703CE
Limit valuesMaximum capacity 700 ct / 140 g 700 ct / 140 g
Readability 0.001 ct / 0.1 mg 0.001 ct / 0.1 mg
Repeatability (at nominal load) 0.1 mg 0.1 mg
Linearity deviation 0.2 mg 0.2 mg
Sensitivity temperature drift 2 ppm/°C 2 ppm/°C
Typical valuesRepeatability (at nominal load) 0.08 mg 0.08 mg
Linearity deviation 0.06 mg 0.06 mg
Minimum sample weight (U=1%, k=2) 16 mg 16 mg
Minimum sample weight OIML 10 mg 10 mg
Settling time 2 s 2 s
Adjustment Int. Cal Ext. Cal
Interfaces 1 RS232 1 RS232
Technical Data62 Jewelry Balances
JE703C JE703CEBalance dimensions (W × D × H) 210×344×279 mm 210×344×279 mm
Weighing pan dimensions ø 90 mm ø 90 mm
Usable height of draft shield 170 mm 170 mm
Weight of balance 4.6 kg 4.4 kg
Weights for routine testingOIML CarePac
Weights#11123002100 g F2, 5 g E2
#11123002100 g F2, 5 g E2
ASTM CarePacWeights
#11123102100 g 1, 5 g 1
#11123102100 g 1, 5 g 1
JE1103C JE1103CE
Limit valuesMaximum capacity 1100 ct / 220 g 1100 ct / 220 g
Readability 0.001 ct / 0.1 mg 0.001 ct / 0.1 mg
Repeatability (at nominal load) 0.1 mg 0.1 mg
Linearity deviation 0.2 mg 0.2 mg
Sensitivity temperature drift 2 ppm/°C 2 ppm/°C
Typical valuesRepeatability (at nominal load) 0.08 mg 0.08 mg
Linearity deviation 0.06 mg 0.06 mg
Minimum sample weight (U=1%, k=2) 16 mg 16 mg
Minimum sample weight OIML 10 mg 10 mg
Settling time 2 s 2 s
Adjustment Int. Cal Ext. Cal
Interfaces 1 RS232 1 RS232
Balance dimensions (W × D × H) 210×344×279 mm 210×344×279 mm
Weighing pan dimensions ø 90 mm ø 90 mm
Usable height of draft shield 170 mm 170 mm
Weight of balance 4.6 kg 4.4 kg
Weights for routine testingOIML CarePac
Weights#11123001200 g F2, 10 g F1
#11123001200 g F2, 10 g F1
ASTM CarePacWeights
#11123001200 g F2, 10 g F1
#11123101200 g 1, 10 g 1
Technical Data 63Jewelry Balances
10.2.2 Gold balances with readability of 1 mgTechnical data
JE203G JE203GE
Limit valuesMaximum capacity 220 g 220 g
Readability 1 mg 1 mg
Repeatability (at nominal load) 1 mg 1 mg
Linearity deviation 2 mg 2 mg
Sensitivity temperature drift 2 ppm/°C 2 ppm/°C
Typical valuesRepeatability (at nominal load) 0.7 mg 0.7 mg
Linearity deviation 0.6 mg 0.6 mg
Minimum sample weight (U=1%, k=2) 140 mg 140 mg
Minimum sample weight OIML 20 mg 20 mg
Settling time 1.5 s 1.5 s
Adjustment Int. Cal Ext. Cal
Interfaces 1 RS232 1 RS232
Balance dimensions (W × D × H) 210×319×289 mm 210×319×289 mm
Weighing pan dimensions ø 120 mm ø 120 mm
Usable height of draft shield 170 mm 170 mm
Weight of balance 4.6 kg 4.4 kg
Weights for routine testingOIML CarePac
Weights#11123001200 g F2, 10 g F1
#11123001200 g F2, 10 g F1
ASTM CarePacWeights
#11123101200 g 1, 10 g 1
#11123101200 g 1, 10 g 1
JE303G JE303GE
Limit valuesMaximum capacity 320 g 320 g
Readability 1 mg 1 mg
Repeatability (at nominal load) 1 mg 1 mg
Linearity deviation 2 mg 2 mg
Sensitivity temperature drift 2 ppm/°C 2 ppm/°C
Typical valuesRepeatability (at nominal load) 0.7 mg 0.7 mg
Linearity deviation 0.6 mg 0.6 mg
Minimum sample weight (U=1%, k=2) 140 mg 140 mg
Minimum sample weight OIML 20 mg 20 mg
Settling time 1.5 s 1.5 s
Adjustment Int. Cal Ext. Cal
Interfaces 1 RS232 1 RS232
Balance dimensions (W × D × H) 210×319×289 mm 210×319×289 mm
Weighing pan dimensions ø 120 mm ø 120 mm
Technical Data64 Jewelry Balances
JE303G JE303GEUsable height of draft shield 170 mm 170 mm
Weight of balance 4.6 kg 4.4 kg
Weights for routine testingOIML CarePac
Weights#11123001200 g F2, 10 g F1
#11123001200 g F2, 10 g F1
ASTM CarePacWeights
#11123101200 g 1, 10 g 1
#11123101200 g 1, 10 g 1
JE503G JE503GE
Limit valuesMaximum capacity 520 g 520 g
Readability 1 mg 1 mg
Repeatability (at nominal load) 1 mg 1 mg
Linearity deviation 2 mg 2 mg
Sensitivity temperature drift 3 ppm/°C 3 ppm/°C
Typical valuesRepeatability (at nominal load) 0.7 mg 0.7 mg
Linearity deviation 0.6 mg 0.6 mg
Minimum sample weight (U=1%, k=2) 140 mg 140 mg
Minimum sample weight OIML 20 mg 20 mg
Settling time 1.5 s 1.5 s
Adjustment Int. Cal Ext. Cal
Interfaces 1 RS232 1 RS232
Balance dimensions (W × D × H) 210×319×289 mm 210×319×289 mm
Weighing pan dimensions ø 120 mm ø 120 mm
Usable height of draft shield 170 mm 170 mm
Weight of balance 4.6 kg 4.4 kg
Weights for routine testingOIML CarePac
Weights#11123007500 g F2, 20 g F1
#11123007500 g F2, 20 g F1
ASTM CarePacWeights
#11123107500 g 1, 20 g 1
#11123107500 g 1, 20 g 1
Technical Data 65Jewelry Balances
10.2.3 Gold balances with readability of 10 mgTechnical data
JE1002G JE1002GE
Limit valuesMaximum capacity 1200 g 1200 g
Readability 10 mg 10 mg
Repeatability (at nominal load) 10 mg 10 mg
Linearity deviation 20 mg 20 mg
Sensitivity temperature drift 3 ppm/°C 3 ppm/°C
Typical valuesRepeatability (at nominal load) 7 mg 7 mg
Linearity deviation 6 mg 6 mg
Minimum sample weight (U=1%, k=2) 1.4 g 1.4 g
Minimum sample weight OIML 500 mg 500 mg
Settling time 1 s 1 s
Adjustment Int.Cal Ext.Cal
Interfaces 1 RS232 1 RS232
Balance dimensions (W × D × H) 200×319×100 mm 200×319×100 mm
Weighing pan dimensions 180×180 mm 180×180 mm
Weight of balance 3.8 kg 3.2 kg
Weights for routine testingOIML CarePac
Weights#111230081000 g F2, 50 g F2
#111230081000 g F2, 50 g F2
ASTM CarePacWeights
#111231081000 g 1, 50 g 1
#111231081000 g 1, 50 g 1
JE2002G JE2002GE
Limit valuesMaximum capacity 2200 g 2200 g
Readability 10 mg 10 mg
Repeatability (at nominal load) 10 mg 10 mg
Linearity deviation 20 mg 20 mg
Sensitivity temperature drift 3 ppm/°C 3 ppm/°C
Typical valuesRepeatability (at nominal load) 7 mg 7 mg
Linearity deviation 6 mg 6 mg
Minimum sample weight (U=1%, k=2) 1.4 g 1.4 g
Minimum sample weight OIML 500 mg 500 mg
Settling time 1 s 1 s
Adjustment Int.Cal Ext.Cal
Interfaces 1 RS232 1 RS232
Balance dimensions (W × D × H) 200×319×100 mm 200×319×100 mm
Weighing pan dimensions 180×180 mm 180×180 mm
Weight of balance 3.8 kg 3.2 kg
Technical Data66 Jewelry Balances
JE2002G JE2002GE
Weights for routine testingOIML CarePac
Weights#111230092000 g F2, 100 g F2
#111230092000 g F2, 100 g F2
ASTM CarePacWeights
#111231092000 g 1, 100 g 1
#111231092000 g 1, 100 g 1
JE3002G JE3002GE
Limit valuesMaximum capacity 3200 g 3200 g
Readability 10 mg 10 mg
Repeatability (at nominal load) 10 mg 10 mg
Linearity deviation 20 mg 20 mg
Sensitivity temperature drift 3 ppm/°C 3 ppm/°C
Typical valuesRepeatability (at nominal load) 7 mg 7 mg
Linearity deviation 6 mg 6 mg
Minimum sample weight (U=1%, k=2) 1.4 g 1.4 g
Minimum sample weight OIML 500 mg 500 mg
Settling time 1 s 1 s
Adjustment Int.Cal Ext.Cal
Interfaces 1 RS232 1 RS232
Balance dimensions (W × D × H) 200×319×100 mm 200×319×100 mm
Weighing pan dimensions 180×180 mm 180×180 mm
Weight of balance 3.8 kg 3.2 kg
Weights for routine testingOIML CarePac
Weights#111230092000 g F2, 100 g F2
#111230092000 g F2, 100 g F2
ASTM CarePacWeights
#111231092000 g 1, 100 g 1
#111231092000 g 1, 100 g 1
JE4002G JE4002GE
Limit valuesMaximum capacity 4200 g 4200 g
Readability 10 mg 10 mg
Repeatability (at nominal load) 10 mg 10 mg
Linearity deviation 20 mg 20 mg
Sensitivity temperature drift 3 ppm/°C 3 ppm/°C
Typical valuesRepeatability (at nominal load) 7 mg 7 mg
Linearity deviation 6 mg 6 mg
Minimum sample weight (U=1%, k=2) 1.4 g 1.4 g
Minimum sample weight OIML 500 mg 500 mg
Technical Data 67Jewelry Balances
JE4002G JE4002GESettling time 1 s 1 s
Adjustment Int.Cal Ext.Cal
Interfaces 1 RS232 1 RS232
Balance dimensions (W × D × H) 200×319×100 mm 200×319×100 mm
Weighing pan dimensions 180×180 mm 180×180 mm
Weight of balance 3.8 kg 3.2 kg
Weights for routine testingOIML CarePac
Weights#111230102000 g F2, 200 g F2
#111230102000 g F2, 200 g F2
ASTM CarePacWeights
#111231102000 g 4, 200 g 4
#111231102000 g 4, 200 g 4
JE5002G JE5002GE
Limit valuesMaximum capacity 5200 g 5200 g
Readability 10 mg 10 mg
Repeatability (at nominal load) 10 mg 10 mg
Linearity deviation 20 mg 20 mg
Sensitivity temperature drift 3 ppm/°C 3 ppm/°C
Typical valuesRepeatability (at nominal load) 7 mg 7 mg
Linearity deviation 6 mg 6 mg
Minimum sample weight (U=1%, k=2) 1.4 g 1.4 g
Minimum sample weight OIML 500 mg 500 mg
Settling time 1 s 1 s
Adjustment Int.Cal Ext.Cal
Interfaces 1 RS232 1 RS232
Balance dimensions (W × D × H) 200×319×100 mm 200×319×100 mm
Weighing pan dimensions 180×180 mm 180×180 mm
Weight of balance 3.8 kg 3.2 kg
Weights for routine testingOIML CarePac
Weights#111230115000 g F2, 200 g F2
#111230115000 g F2, 200 g F2
ASTM CarePacWeights
#111231115000 g 4, 200 g 4
#111231115000 g 4, 200 g 4
Technical Data68 Jewelry Balances
10.3 Dimensions
10.3.1 Carat balances with readability of 0.001 ct / 0.1 mg
Models:
JE503CJE503CEJE703CJE703CEJE1103CJE1103CE
Technical Data 69Jewelry Balances
10.3.2 Gold balances with readability of 1 mg
Models:JE203GJE203GEJE303GJE303GEJE503GJE503GE
279
289
172.6
142
110
95
89
103
95
101
199
222
192
313
154
200
210
180
319
95
120
420.6
Technical Data70 Jewelry Balances
10.3.3 Gold balances with readability of 10 mg
Models:JE1002GJE1002GEJE2002GJE2002GEJE3002GJE3002GEJE4002GJE4002GEJE5002GJE5002GE
100
95 101
199
78
95
319
180
180
200
313
222
154
Technical Data 71Jewelry Balances
10.4 Interface specification
10.4.1 RS232C interfaceEach balance is equipped with an RS232C Interface as standard for the attachment of a peripheral device(e.g. printer or computer).
Schematic Item Specification
15
9 6
IN
GND
OUT
HAND
SHAKE
IN
DATA
OUT
RxD
TxD
CTS
RTS
OUT
POWER
SUPPLY
+12V
2nd display mode only
Interface type Voltage interface according to EIA RS-232C/DIN66020 CCITT V24/V.28)
Max. cable length 15 m
Signal level Outputs: +5 V … +15 V (RL = 3–7 kΩ) –5 V … –15 V (RL = 3–7 kΩ) Inputs: +3 V … +25 V –3 V … -25 V
Connector Sub-D, 9-pole, female
Operating mode Full duplex
Transmission mode Bit-serial, asynchronous
Transmission code ASCII
Baud rates 600, 1200, 2400, 4800, 9600, 19200,38400 (software selectable)
Bits/parity 7-bit/none, 7-bit/even, 7-bit/odd, 8-bit/none(software selectable)
Stop bits 1 stop bit
Handshake None, XON/XOFF, RTS/CTS (softwareselectable)
End-of-line <CR><LF>, <CR>, <LF> (software selectable)
Power supply for2nd display
+ 12 V, max 40 mA (software selectable, 2nddisplay mode only)
10.4.2 MT-SICS interface commands and functionsMany of the instruments and balances used have to be capable of integration in a complex computer ordata acquisition system.To enable you to integrate balances in your system in a simple manner and utilize their capabilities to thefull, most balance functions are also available as appropriate commands via the data interface.All new METTLER TOLEDO balances launched on the market support the standardized command set"METTLER TOLEDO Standard Interface Command Set" (MT-SICS). The commands available depend on thefunctionality of the balance.For further information please contact your METTLER TOLEDO representative.For further information please refer to the Reference Manual MT-SICS downloadable from the Internet under
u www.mt.com/jewelry
Accessories and Spare Parts72 Jewelry Balances
11 Accessories and Spare Parts
11.1 AccessoriesAccessories
Printers
RS-P25 printer with RS232 connection to instrument 11124300Paper roll (length: 20 m), set of 5 pcs 00072456Paper roll (length: 13 m), self-adhesive, set of3 pcs
11600388
Ribbon cartridge, black, set of 2 pcs 00065975
RS-P26 printer with RS232 connection to instrument (withdate and time)
11124303
Paper roll (length: 20 m), set of 5 pcs 00072456Paper roll, self-adhesive (length: 13 m), set of3 pcs
11600388
Ribbon cartridge, black, set of 2 pcs 00065975
RS-P28 printer with RS232 connection to instrument (withdate, time and applications)
11124304
Paper roll (length: 20 m), set of 5 pcs 00072456Paper roll, self-adhesive (length: 13 m), set of3 pcs
11600388
Ribbon cartridge, black, set of 2 pcs 00065975
P-52RUE dot matrix printer with USB, RS232, and ethernetconnections
30237920
Paper roll, set of 5 pcs 00072456Paper roll, self-adhesive, set of 3 pcs 11600388Ribbon cartridge, black, set of 2 pcs 00065975
P-56RUE thermal printer with RS232, USB and ethernetconnections, simple printouts, date and time, label printing(limited).
30094673
Paper roll, white (length: 27 m), set of 10 pcs 30094723Paper roll, white, self-adhesive (length: 13 m), setof 10 pcs
30094724
Paper roll, white, self-adhesive labels (550labels), set of 6 pcs
30094725
P-58RUE thermal printer with RS232, USB and ethernetconnections, simple printouts, date and time, label printing,balance applications: statistics, formulation, totaling,
30094674
Paper roll, white (length: 27 m), set of 10 pcs 30094723Paper roll, white, self-adhesive (length: 13 m), setof 10 pcs
30094724
Paper roll, white, self-adhesive labels (550labels), set of 6 pcs
30094725
Accessories and Spare Parts 73Jewelry Balances
Cables for RS232 interface
RS9 – RS9 (m/f): connection cable for PC, length = 1 m 11101051
RS9 – RS25 (m/f): connection cable for PC, length = 2 m 11101052
RS232 - USB converter cable – Cable with converter toconnect a balance (RS232) to a USB port
64088427
Density determination
Density kit ME-DNY-4 for balances with readability of0.01 mg / 0.1 mg
30029886
Glass beaker, height 100 mm, ø 60 mm 00238167
Sinker for density of liquids in conjunction with density kit 00210260Calibrated (sinker + certificate) 00210672Recalibrated (new certificate) 00210674
Calibrated thermometer with certificate 11132685
Carat pans
Carat pan XS, ø 50 mm / heigh 20 mm (set of 10 units) 12102565
Accessories and Spare Parts74 Jewelry Balances
Carat pan S, ø 80 mm / heigh 20 mm (set of 10 units) 12102645
Carat pan M, ø 90 mm / heigh 30 mm (set of 10 units) 12102646
Carat pan L, ø 90 mm / heigh 45 mm (set of 10 units) 12102647
Weighing pans
Set of weighing pan ø 160 mm with pan support forbalances with readability of 10 mg and 100 mg using draftshield
30042896
Draft shields
Draft shield low with sliding doors, usable heigh 170 mm.• for balances 0.1 mg or 1 mg• for balances 10 mg or 100 mg, weighing pan ø 160 mm
is needed (#30046407)
30042884
Auxiliary displays
METTLER TOLEDO
000000
RS232 auxiliary display AD-RS-J7 12122380
Protective covers
Protective cover for models with readability of 0.01 mg /0.1 mg
30037742
Protective cover for models with readability of 1 mg …100 mg
30042890
Accessories and Spare Parts 75Jewelry Balances
Dust covers
Dust cover for models without draft shield 30029051
Dust cover for models with draft shield low (170 mm) 30029050
Anti-theft devices
Anti-theft steel cable 11600361
Software
LabX direct balance (simple data transfer) 11120340
Transport cases
Transport case for models with draft shield (170 mm) 30046405Transport case for models without draft shield 30046406
Miscellaneous
Tool kit, contains brush, tweezer and glove 30046403
AC/DC universal adapter (EU, USA, AU, UK) 100–240 VAC,50/60 Hz, 0.5 A, 12 VDC 1 A
11120270
AC/DC adapter (without power cable) 100–240 V AC, 0.8 A,50/60 Hz, 12 V DC 2.5 A
11107909
Accessories and Spare Parts76 Jewelry Balances
Country-specific 3-Pin power cable with groundingconductor.
Power cable AU 00088751Power cable BR 30015268Power cable CH 00087920Power cable CN 30047293Power cable DK 00087452Power cable EU 00087925Power cable GB 00089405Power cable IL 00225297Power cable IN 11600569Power cable IT 00087457Power cable JP 11107881Power cable TH, PE 11107880Power cable US 00088668Power cable ZA 00089728
Adjustment weights
OIML / ASTM Weights (with calibration certificate) see http://www.mt.com/weights
11.2 Spare partsJE balances with readability of 0.1 mg / 1 mg
Drawing Pos Description Part No.
METTLER TOLEDO
6
7
2
5
4
3
11 Sliding top door for draft shield high or
low with mounted handle (170 mm)30037733
2 Pair of handles for sliding doors ofdraft shield
30037736
3 Pair of sliding side doors for draftshield low with mounted handles (leftand right), 170 mm
30042885
4 Weighing pan ø 120 mm incl. pansupport, 1 mg
30042889
5 Weighing pan ø 90 mm incl. pansupport, 0.1 mg
30037737
6 Draft shield element, 0.1 mg 12122043
7 Pair of leveling foot 30037744
Accessories and Spare Parts 77Jewelry Balances
JE balances with readability of 10 mg
Drawing Pos Description Part No.
METTLER TOLEDO
2
3
1
1
1 Weighing pan 180 × 180 mm withpan support, 10 mg
30042895
2 Draft shield element 180 × 180 mm,10 mg
30042897
3 Pair of leveling foot 30037744
Miscellaneous
Drawing Pos Description Part No.Bottom plate for draft shield 30037739
Bottom plate for models without draftshield
30042901
Accessories and Spare Parts78 Jewelry Balances
Index 79Jewelry Balances
IndexNumerics1/10 d 26
AAC adapter 59Accessories 72Adjusting 19Adjustment 21Adjustment lock 28Advanced menu 25Application check weighing 39Application density 48Application division factor weighing 46Application icons 9Application multiplication factor weighing 44Application percent weighing 37Application piece counting 34Application statistics 41Application totaling 43Applications 10Auto print 30Automatic standby 28Automatic zero setting 29Autozero 29Average (Statistics) 41
BBacklight 28Basic menu 25Basic principles for operation 10Baudrate 31Below-the-balance weighing 23Bit/Parity 31
CCancel 11Change settings 11Char set 32Check weighing 39Cleaning 57Closing the menu 11Components 8Conventions and symbols 3Customer fine adjustment 21
DData communication format 30Date 17, 26Date format 28Delivery inspection 12Density 48Density kit 48Density table for ethanol 52Dimensions 68Display 30Display increment 26Display panel 9Disposal 58Division factor weighing 46Draft shield 57, 58
EEnd of line 32Environment 27Environmental conditions 59Error messages 55Ethanol 52External weight 20
FFactor weighing 44, 46Function PC-Direct 53
GGeneral safety information 4
HHandles 58Handshake 32Header 30Host 30
IIcons 9Identification 27In use cover
see protective cover 15Input principle 11Installing
Components 13Handles 58
Index80 Jewelry Balances
Intended use 4Interface
MT-SICS 71Interface menu 25, 29Interface RS232C 29, 71Internal weight 19Interval 32Introduction 3
LLevel indicator 17Leveling 17Line feed 30Liquid 48Liquids 50Local conditions 16Location 16Lock adjustment 28
MMain menu 26Manual adjustment with external weight 20Manual adjustment with internal weight 19Materials 60Menu 26Menu Advanced 25Menu Basic 25Menu Interface 25, 29Menu operation 10Menu protection 26Menu topic 11, 26MT-SICS 71Multiplication factor weighing 44
NNumerical values 11
OON / OFF 17, 23Operating temperature 17Overview 8
PPC-DIR 29PC-Direct 53Percent weighing 37Performing a simple weighing 22Piece counting 34
Power cable 16Power supply 59Power supply voltage 16Print 23Print menu 27Printer 29Protect 26Protection and standards 59Protective cover 15
RRecall 23, 28Reset 27RS232C interface 29, 71
SSafety Information
General 4Intended use 4Signal words 4Warning Symbols 4
Saving settings 11Select menu 10Select menu topic 11Selecting weighing application 10Service 29, 57Service date reset 29Service icon 29Service reminder 29Set identification 27Setting Date and Time 17Setting up the balance 12Signal words 4Signature line 30Sinker 48, 50Solids 48Spare parts 77Standard deviation (Statistics) 41Standby 28Statistics 41Status icons 9Status messages 56Stop bit 32Submenu 11Switching
On 17Switching off 23
Index 81Jewelry Balances
Switching weight units 22Symbols 3Symbols and conventions 3
TTaring 22Technical data 59Technical data model-specific 61Time 17, 26Time format 28Topic 11, 26Totaling 43Transmit data 23Transport over long distances 24Transport over short distances 24Transporting the balance 24
UUnit 26, 27Unpacking 12
WWarm-up time 17Warning Symbols 4Weighing application 10Weighing-in aid 23Weight unit 22, 26, 27
ZZero print 30Zero setting 22Zeroing 29
Index82 Jewelry Balances