Post on 16-Jul-2019
Jérémy Flamanc Product Manager Specialty Germanium CANBERRA Specialty Division jeremy.flamanc@canberra.com +33 6-76-49-38-35 (Mobile) +33 3-88-77-43-67 (Office)
Nuclear Physics experiments
Synchrotron detectors
Industrial systems
New Sealed HPGe probe
AREVA MINING original requirements: u BORE HOLE inspection for Uranium exploration u Rugged, Sealed, Easy to clean u Streamline, all attitude u LN2 free u Turnkey, Mobile & portable, Easy to deploy u High Energy resolution (MDA, mixture)
Outcome: a harsh environment compatible HPGe
New Sealed HPGe probe
Sealed Housing
HPGe capsule
Front end electronics
Electrical Cooler
Passive cooler
Watertight connector
Industrial composite cable
Comprehensive Acquisition station
Performances u Relative efficiency : 20% (coaxial HPGe Φ 50mm x L 50mm) u Energy resolutions FWHM: 1,2 keV @ 122keV / 2,2 keV @1332 keV u MDA in water: down to 0,1 Bq /L (depends on background) u Energy range: 50 keV – 10 MeV u Count rate up to 100kcps
Probe mechanical characteristics u Dimensions 80mm x 800 mm u Sealed (no vents) u Weight 10kg u Cable length: up to 100m u Operating temperature up to 35°C u Storage temperature up to 50°C
New Sealed HPGe probe
Analysis stations
Watertight case version
Flight case version
How to improve the mobility? u Low cooler power consumtion: 30W u Ultra high vacuum encapsulation of the HPGe detector
• Full thermal cyclings are not required • Increased sensor life time & reliability
New Sealed HPGe probe
Station with a 4h battery
Présentation CANBERRA Lingolsheim CONFIDENTIEL
Our users needs it for
Dismantling & Decommissioning u Wall, Soil u Collimators for high count rates u Easy to decontaminate
Water contamination u Fresh / sea water u Low MDA
Emergency response u Robotic system u Hand held
New Sealed HPGe probe
HPGe is the prefered solution when adressing: u High background environment u Mixture of radionuclides u Low MDA / faster counting time u Separation of natural and anthropogenic sources
Higher efficiency solutions
Rugged HPGe for higher efficiencies
Design for large crystals u Efficiency up to 175% u Air cooling (Fan) u Or Sealed Version
• Optional External water chiller • Temperature up to 60°C • Contaminated areas
u Compatible and exchangeable with existing portable LN2 dewar
Even Higher efficiency detectors
Cluster of largest HPGe crystals u Coincidence to increase the sensitivity: 1300% achievable u SW suite allows calculation of the ground dose u Comprehensive Database: GPS, raw spectra, monitoring
data and more
Extreme efficiencies solutions
5% 20% 175% 1300%
Summary
EFFICICENY RANGE
Summary
In situ / outdoor spectrometry with HPGe is possible Energy resolutions similar to lab environments harsh environmental conditions : immersion, bore hole, transportation with cold diode Optimization for mobile use: minimum down time, fast deployement, and easy decontamination, minimal maintenance Custom designs to meet operational footprint constraints The Buidling Blocks developped are applicable to multiple designs
Summary
Questions?
Meet us at CANBERRA booth
jeremy.flamanc@canberra.com