Introductionto
SEC (GPC) Detection
André M. StriegelNational Institute of Standards & Technology
Why Do We Need Detectors in SEC?
We want to know what is eluting from our column.
We want to know when this takes place.
We want to determine certain physicochemical properties of our analyte(s).
We want to monitor system performance.
So…
9/22/2016
2
Course Module Objective
What detection method(s) do I need,and why?
What Detector(s) Do I Need, And Why?
To answer these questions, we need to know:
1. What property/properties of our analyte(s) do we want to determine (e.g., molar mass averages/distribution, long-chain branching, chemical composition, etc.)?
2. What level of quantitation, if any, do we desire?
3. Does a single detection method exist that can measure this property directly and quantitatively? If so, which method and how?
4. If not, can that information be obtained by using multiple detectors? If so, which methods and how?
9/22/2016
3
Should Be Aware Of…
Limitations of the separation method w/r to isolating property of interest. (Beyond objective of this module).
Limitations of the detection method(s) w/r to measuring and quantitating property of interest.
Today’s Menu
What detection methods are out there?
What do they measure?
How do some of them combine with each other?
9/22/2016
4
Classifying SEC Detectors
For purposes of discussion, we will classify SEC detectors as Chemical and Physical.
Chemical detectors provide information in an additive fashion.
Physical detectors provide information in a synergistic fashion
Some detectors can be either Chemical or Physical, depending on mode of operation or on how data are employed.
Chemical Detectors
UV
MS• ESI• MALD/I• ICP• IMS
IR/(Raman)
NMR
Fluorescence
Conductivity
Etc.
For the most part, these are methods you learned about in Instrumental Analysis (or related course) and function the way you were taught.
Fairly straightforward, when operated in Chemical mode (some can be Physical methods, as well).
Therefore, we will not spend much time on these.
9/22/2016
5
Chemical Detectors
For UV, IR/Raman, and fluorescence, need “active” (absorbent, fluorescent) groups on molecules.
For conductivity, need charge-bearing groups.
If these groups don’t exist, but you wish to use one or more of these detection methods, will need to derivatize(preferably quantitatively and w/o degradation) your polymer.
Chemical Detectors combine with each other additively.
SEC/UV SEC/IR SEC/FL
Kirkland & Antle, J. Chromatogr. Sci. 15 (1977) 137 Nomura et al., Macromolecules 35 (2002) 5388 Moon, Hoye, Macosko, J. Polym. Sci. A 38 (2000) 989
ex = 358 nm, em = 405 nm
9/22/2016
6
SEC/DRI/UV of St-MMA Copolymer
20 22 24 26
0.0
0.2
0.4
0.6
0.8
1.0
UV Signal Conc. of St only
Rel
ativ
e D
etec
tor
Sig
nal
(ar
bit
rary
un
its)
Retention volume (mL)
DRI Signal Conc. of both St and MMA
5.0x104 1.0x105 1.5x105 2.0x105 2.5x105 3.0x105 3.5x105
0.5
1.0
1.5
2.0
2.5
3.0
18
20
22
24
26
28
30
Dif
fere
nti
al W
eig
ht
Fra
ctio
n (
dW
f /d
M)
Molar Mass, M (g/mol)
MMD % Styrene
Weig
ht P
ercent S
tyrene (%
St)
• Random copolymer of styrene (St) and methyl methacrylate (MMA).
• Styrene absorbs at 260 nm, but methyl methacrylate does not.
• UV detector @ 260 nm will respond only to styrene content in polymer.
• Differential refractometer will respond to both styrene and methyl methacrylate in polymer.
Haidar Ahmad & Striegel, Anal. Bioanal. Chem. 396 (2010) 1589
SEC/ICP-MS
Precautions
Minimize use of organic solvents.
Careful attention to ionic strength of mobile phase.
Minimize salt content of mobile phase.
Careful sample preparation.
Zoorob, McKiernan, Caruso, Mikrochim. Acta 128 (1998) 145
9/22/2016
7
SEC/ICP-MS
Zoorob, McKiernan, Caruso, Mikrochim. Acta 128 (1998) 145 Wrobel et al., Anal. Chem. 75 (2003) 761
Physical Detectors Concentration-sensitive
• UV ( absorptivity conc)• DRI ( n/c conc)• EMD
Viscometry
Light scattering• Static ( Mw (n/c)2 conc)
o Low-angle
o Multi-angle
• Dynamic
IR
NMR
Less interested in what the detectors measure, per se, than in what the measurements can tell us about the analyte(s).
While Physical Detectors can be used alone, biggest advantage comes from synergistic combination of detectors.
9/22/2016
8
DRI, UV, IR, LS Response Depends On…
Solvent
Temperature
Wavelength
Chemical identity of analyte
Monomer ratio (copolymers)
DP (oligomers)
Why Use Physical Detectors?
Need “absolute” (i.e., calibrant-independent) M averages and/or MMD.
Need architectural information (LCB, SCB, conformation, topology, etc.).
Did analyte aggregate in solution?
Tacticity info.
Etc.
9/22/2016
9
Determining “Absolute” Molar Mass
Most accurate and precise method: Apply peak-position (calibrant-relative) calibration.
IF calibrants and analytes have the same architecture ANDchemical repeat unit.
Otherwise, can perform universal calibration using on-line viscometer and concentration-sensitive detection (e.g, DRI, UV).
Use on-line static light scattering detection (LALS, MALS), in combination with concentration-sensitive detection (e.g., DRI, UV).
Absolute M by Static Light Scattering (SLS)
Fundamental equation of SLS is the Rayleigh-Gans-Debyeapproximation:
where
and
AN
cnnK
40
220
2* 4
...2
sin3
161
)(
1 22,2
2
zGRP
R(): Excess Rayleigh scattering ratio
Mw: Weight-average molar mass
c: Concentration
A2: Second virial coefficient of solution
n0: RI of solvent at experimental temperature &wavelength
n/c: Specific refractive index increment ofsolution
0: Vacuum wavelength of incident radiation
NA: Avogadro’s cell phone number
P(): Particle scattering factor
: Angle of observation
RG,z: z-average radius of gyration
: Wavelength of radiation in medium ( 0/n0)
)(21)()(
2*
PcMAPMcK
Rww
9/22/2016
10
Obtaining Mw From a Dual-Detector System: SEC/SLS/DRI
If , then
We thus recognize that, from an SEC experiment with on-line DRI and static LS detection, we can obtain the molar mass at each slice by taking the ratio of the static LS and DRI signals at each slice.
signal
signalw DRI
SLS
c
RM
)(wM
cK
R
*
)(
Obtaining Mw From a Dual-Detector System: SEC/SLS/DRI
At each elution slice i, we can take the ratio of the signals from these two detectors, SLS and DRI (after correcting for interdetector delay or split ratio, as appropriate), to obtain the weight-average molar mass Mw as a function of elution volume and, consequently, obtain the molar mass distribution.
Retention Volume
Det
ecto
r R
espo
nse SLS
DRI
slice i
iwi
i
isignal
isignal Mc
R
DRI
SLS,
,
, )(
9/22/2016
11
Molar Mass Averages from SEC/SLS/DRI
Using static LS and DRI as detection methods in SEC, we can measure Mw,i, the molar mass at each slice i eluting from the SEC column(s).
If ci is the concentration of species of molar mass Mw,i eluting in slice i, then we can calculate the various molar mass averages of the polymer sample using the Meyerhoff equation
We can also combine the data from the concentration-sensitiveand static LS detectors to obtain the molar mass distribution.
i
xiwi
i
xiwi
Mc
McM
1,
,
when x = 0, = n; x = 1, = w; x = 2, = z
Static LS:A Molar-Mass-Sensitive Detector
As noted, the signal from a static LS detector is proportional to molar mass:
R() Mw
One application of this feature is the detection of polymeric aggregates in solution.
When polymers aggregate (for whatever reason), the molar mass of the aggregate will be greater than the molar mass of an individual polymer chain.
Therefore, the static LS signal from the aggregate will be greater than the signal produced by the unaggregated polymer.
9/22/2016
12
SLS Sensitivity to Aggregation
AUX: DRI
Aggregation of Pullulan in DMAc/0.5% LiCl, 80 oC
Striegel & Timpa, in Strategies in Size-ExclusionChromatography, Potschka & Dubin, eds. (1996) p. 366
The aggregation of pullulan in solution is easily observed with static LS. On the other hand, there is no indication of aggregation from the
refractometer signal.
Dynamic Light Scattering (DLS)
a.k.a. Quasi-Elastic LS (QELS) &Photon Correlation Spectroscopy (PCS)
9/22/2016
13
Static vs. Dynamic LS
In static light scattering (SLS), we measure the time-averaged fluctuations of the scattered light.
In dynamic light scattering (DLS, or QELS), we measure the time-dependent fluctuations of the scattered light.
9/22/2016
14
Dynamic Light Scattering (DLS)
From the measured time-dependent fluctuations in scattered light, we determine the translational diffusion coefficient, D, of the analyte.
Using D, we can calculate RH, the hydrodynamic or Stokes radius of the analyte:
Can think of RH as the radius of a homogeneous hard sphere that has the same translational diffusion coefficient as the analyte.
D
TkR B
H06
kB: Boltzmann’s constant
T: Absolute temperature of the experiment
0: Solvent viscosity at the experimental T
4.5 5.0 5.5 6.0 6.5 7.0 7.5
0.0
0.5
1.0
90o S
LS
(n
orm
aliz
ed)
Retention volume (mL)
1.0 mL min-1
PS 800K, Mw/M
n = 1.03
THF, 35 oC
1
10
100
Hyd
rod
ynam
ic radiu
s, RH
(nm
)
SEC/DLS of Narrow PS Standard
Adapted from Striegel, J. Chromatogr. A, 1359 (2014) 147
Predicted values based on RH vs. Mrelationship for PS in THF @ 25oC. SeeWernert et al., Anal. Chem. 82 (2010) 2668
9/22/2016
15
Viscometry
On-Line ViscometersTwo basic types: Single-Capillary and Differential
J Lesec, in Encyclopedia of Chromatography,2nd ed., J Cazes, Ed. (2005)
9/22/2016
16
On-Line ViscometersTwo basic types: Single-Capillary and Differential
J Lesec, in Encyclopedia of Chromatography,2nd ed., J Cazes, Ed. (2005)
Viscometry: General Principles For laminar flow within a cylinder, driven by constant pressure
gradient, the pressure drop P across a capillary is related to the length L and radius r of the tube, as well as to the viscosity of the solution flowing through the tube and to the volumetric flow rate Qthrough the tube. Whew!
The relation is given by Poiseuille’s Equation (a.k.a. the Hagen-Poiseuille Equation):
Because L, r, and Q are usually known quantities in an SEC experiment, can be directly related to P. The latter is measured using a differential pressure transducer, which measures the difference in pressure between the inlet and outlet of the cylinder.
4
8
r
QLP
9/22/2016
17
Single-Capillary Viscometers
Differential pressure transducer
measures pressure drop across tube.
At constant flow rate Q, the viscosity ηi of each slice i eluting from the SEC columns is calculated using Poiseuille’s law.
The specific viscosity of the solution, sp, is then calculated via
Pi: Pressure drop along capillary for slice i. Vi: Voltage signal corresponding to pressure drop of slice i
P0: Average solvent baseline value of pressure. V0: Average solvent baseline value of voltage
0
0
0
0
00
0 1V
VV
P
PP iisp
Single-Capillary Viscometers
REM: The intrinsic viscosity [η] is defined as
We recognize [η] as the ratio of the signal from the viscometer, which gives us ηsp, to the signal from the differential refractometer, which gives us c:
csp
c
0lim][
][
cDRI
VISC sp
signal
signal
9/22/2016
18
Obtaining [η] From a Dual-Detector System: SEC/VISC/DRI
At each elution slice i, we can take the ratio of the signals from these two detectors, VISC and DRI (after correcting for interdetector delay or split ratio, as appropriate), to obtain the intrinsic viscosity [η] as a function of elution volume and, consequently, as a function of molar mass M.
As we will see later, this [η] vs. M relationship can be quite informative.
Retention Volume
Det
ecto
r R
espo
nse VISC
DRI
slice i
ii
isp
isignal
isignal
cDRI
VISC][,
,
,
IR and NMRas
Physical Detectors
9/22/2016
19
IR as a Physical Detector: SCB of Polyolefins Monitor IR absorbance of CH3 and
combined CH2CH3 groups in polyolefins.
TCB, ODCB, and perchloroethylene are transparent in IR region of interest.
Ratio of absorbances corresponds to amount of SCB, expressed as methyl chain ends per one thousand total carbon atoms, or CH3/1000C.
For high levels of SCB, or for SCB in propylene-ethylene copolymers, calibrate using PP and PE homopolymers.
Monrabal, Sancho-Tello, LCGC Asia Pacific, September 2009
NMR as a Physical Detector:Tacticity Distribution of PMMAs
Ute et al., Macromol. Chem. Phys. 202 (2001) 3081
For more info on SEC-NMR, seetalk by Wolf Hiller on Thursday
9/22/2016
20
Obtaining Architectural Information(LCB & Conformation)
byCombining Physical Detectors
RL Whistler, in Industrial Gums, 3rd ed., RL Whistler & JN BeMiller, Eds. (1993)
Two polymers, one linear and one branched,composed of the same chemical repeat units and having the same molar
mass as each other, will occupy different hydrodynamic volumes in solutionat identical solvent/temperature conditions.
9/22/2016
21
Conformation & Mark-Houwink Plotsof Linear vs. Branched Polymers
Lo
g R
adiu
s o
f G
yrat
ion
, RG
Log Molar Mass, M
Linear
Branched
For any given M > Mx,
RG,Linear
> RG,Branched
Mx
Lo
g In
trin
sic
Vis
cosi
ty, []
Log Molar Mass, M
Linear
Branched
For any given M > Mx,
[]Linear
> []Branched
Mx
Conformation & Mark-Houwink PlotsPolymer Architecture
0.33 0.44
0.44 < 0.5
RG ~ M
Lo
g R
ad
ius
of
Gyr
atio
n, R
G
Log Molar Mass, M
0.5 < 0.6
[] ~ Ma
Lo
g In
trin
sic
Vis
cosi
ty, []
)
Log Molar Mass, M
0.5 < a 0.8
0.33 < a 0.5
0 a 0.33
Striegel, Yau, Kirkland, & Bly, Modern Size-Exclusion Liquid Chromatography, 2nd ed. (2009)
9/22/2016
22
Relation Between Slopes of Conformation & Mark-HouwinkPlots and Polymer Architecture/Conformation
Architecture/Conformation a
Rigid rod 1 2
Linear random coil (good) 0.5 < 0.6 0.5 < a 0.8
Linear random coil () 0.5 0.5
Random branching (good) 0.44 < 0.5 0.33 < a 0.5
Random branching () 0.44 0.33
Hard sphere 0.33 0
“good” and “” correspond to values at good and theta solvent/temperature conditions, respectively
Striegel, Yau, Kirkland, & Bly, Modern Size-Exclusion Liquid Chromatography, 2nd ed. (2009)
Polymer Conformation & Dimensionless Size
Radius of Gyration (from MALS)
31
10
3
AN
MR
i
cmiG Rrn
R 2
1
1
.
Viscometric Radius (from MALS+VISC)
Random coil of homopolymers
Hard sphere ≈ 1.3
Rigid rod0.3 – 0.4More extended
structure
More compact structure
ri: location individual atom or group of atoms
Rcm: location of polymer center of mass
n: degree of polymerization
[]: intrinsic viscosity
M: molar mass
GR
R
9/22/2016
23
Dimensionless Size & Polymeric Architecture
Architecture RG,z/RH,z
Homogeneous hard sphere 0.778
Gaussian “soft” sphere(e.g., dendrimers generation ≥ 10)
0.977
Monodisperse linear random coil-conditionsGood conditions
1.5041.78
Polydisperse linear random coil-conditionsGood conditions
1.732.05
Regular stars with uniform arm length-conditions, f = 4Good conditions, f » 1
1.3331.079
Regular stars with polydisperse arm length-conditions, f = 4Good conditions, f » 1
1.5341.225
Burchard, Adv. Polym. Sci. 1999, 143, 113
Obtain RG,z from MALS
Obtain RH,z from DLS
Combined “dimensionless ratio” is additional info not provided by either detector, individually (i.e., detector synergism).
SEC/MALS/DLS/DRI: as f(ret. vol.)
4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0
DR
I re
spo
nse
Retention volume (mL)
0.0
0.5
1.0
1.5
2.0
RG /R
H
PS1.7M, Mw/M
n = 1.10
THF, 35 oC1 mL min-1
AM Striegel, unpublished data
9/22/2016
24
Final Thoughts There is virtually no limit to how many detectors can be combined and
how this may be done (destructive detectors and backpressure issues, etc. notwithstanding).
Example 1: Ludlow et al. (J. Chromatogr. A 857 (1999) 89) performed SEC/UV/1H-NMR/ESI-MS/FT-IR to analyze polymer additives (antioxidants and plasticizers).
Example 2: Rowland & Striegel (Anal. Chem. 84 (2012) 4812) performed SEC/MALS/DLS/VISC/UV/DRI to study copolymers and blends. Determined M averages, MMD, radii, chemical heterogeneity, change in conformation as a f(M), and chemical (monomeric sequence) basis for conformational changes.
So…