High Performance Liquid Chromatography. HPLC originally refered to: High Pressure Liquid...

Post on 18-Dec-2015

261 views 2 download

Tags:

Transcript of High Performance Liquid Chromatography. HPLC originally refered to: High Pressure Liquid...

High PerformanceLiquid

Chromatography

HPLC

originally refered to:High Pressure Liquid Chromatography

currently refers to:High Precision Liquid Chromatography

– high pressure to be able to use small particle size to allow proper separation at reasonable flow rates

“Applications of Liquid

Chromatography”

Effect of Particle Size

“Effect of particle size of packing and flow rate upon height H in liquid chromatography. Column dimensions: 30cm x 2.4mm. Solute: N,N-diethyl-n-aminoazobenzene. Mobile phase: mixture of hexane, methylene chloride, isopropyl alcohol.”

“Effect of sample size on plate

height”

Components

Solvent Reservoir and Degassing System

PumpsPrecolumnsSample Injection SystemColumnsTemperature ControlDetectors

“Schematic of an

apparatus for

HPLC”

Components

Solvent Reservoir and Degassing System– isocratic elution - single solvent separation

teachnique– gradiant elution - 2 or more solvents, varied

during separation

Improvement in separation efficiency by gradient elution. Column: 1m x 2.1mm id, precision-bore stainless; packing: 1% PermaphaseR ODS. Sample: 5L of chlorinated benzenes in isopropanol. Detector: UV photometer (254 nm). Conditions: temperature 60oC, pressure, 1200 psi.

Components

Pumps– 1000 - 6000 psi at rate of @ 3 mL/min– reproducible to 2%– screw driven syringe– reciprocating pump

“A reciprocating pump for HPLC.”

Components

Precolumns– remove impurities from solvent– saturates mobile phase with liquid of

stationary pahse before the analytical column

Components

Sample Injection System– sample valve– syringe

“A sampling loop for liquid chromatography”

Components

Columns– straight, 15 to 150 cm in length; 2 to 3 mm i.d.– packing - silica gel, alumina, Celite

Components

Temperature Control

Components

Detectors– single wavelength UV– tuneable UV– diode-array– refractive index– MS

Performances of LC Detectors

Ultraviolet detector cell for HPLC

Diode-ArrayAbsorption Spectra“Absorption spectra of the eluent from a mixture of three steroids taken at 5-second intervals.”