ANDE Course Electric, Magnetic and Electromagnetic Techniques
Part 1
October 15, 2009 C.V.Krishnamurthy
Electromagnetic Spectrum
10/16/2009
2
Statics, Dynamics and Waves
Quasistatics
Frequency, f Lightdc
Dynamics
Statics
Statics: f = 0; dc (time derivatives vanish)Dynamics: No restriction; complete Maxwells equations;
Electromagnetic wavesQuasistatics: Low-frequency extension of statics, or
low-frequency approximation of dynamics; Non-radiative
Overview
Electric fields Capacitive
Conductive
Magnetic fields MFL
Barkhausen Noise
Electromagnetic fields Eddy current (low frequencies)
Radiative (high frequencies)
10/16/2009
3
Faradays Law
E dl d
dtC B dSS dS
SC
B
Voltage around C, also known as electromotive
force (emf) around C (but not really a force),
2 2Wb m m , or Wb.Magnetic flux crossing S,
Time rate of decrease of magnetic flux crossing S,
Right-hand screw Rule The magnetic flux crossing the surface S is to be evaluated toward that side of S a right-hand screw advances as it is turned in the sense of C.
C
A loop of wire coinciding along the imaginary contour C will result in a current flowing in the wire.
Lenzs Law States that the sense of the induced emf is such that any current it produces, if the closed path were a loop of wire, tends to oppose the change in the magnetic flux that produces it. Thus the magnetic flux produced by the induced current and that is bounded by C must be such that it opposes the change in the magnetic flux producing the induced emf.
B
Et
BvF qLorentz force on a current carrying wire is :
The emf is known as motional emf.
Amperes Law
H dl C J dS
d
dtS D dSS
dS
SC
J, D
Magnetomotive force (only by analogy with electromotive force),
A m m, or A.
Current due to flow of charges crossing S,
2 2A m m , or A.
Displacement flux, or electric flux, crossing S,
2 2C m m , or C.
Time rate of increase of displacement flux crossing S, or, displacement current crossing S,
C s, or A.
Right-hand screw rule applies.
DH J
t
J dS=S1 I but J dS =S2
0
D dS =S1 0 but D dS S2
0
d
dtD dS must be I
S2
so that H dlC is unique.
C
S1
S2
I(t)
10/16/2009
4
Gausss Law and
B dS = 0SD dS S dvV
Magnetic flux emanating from a closed surface S = 0.
B 0Solenoidal property of magnetic field lines.
aaa
J t 0
J dS +d
dt dv 0
VSCurrent due to flow of charges emanating from a closed surface S = Time rate of decrease of charge enclosed by S.
ContinuityEquation
LAW OF CONSERVATION OF CHARGE
D
Displacement flux emanating from a closed surface S = charge contained in the volume bounded by S = charge enclosed by S.
Maxwells Equations
B
Et
DH J
t
aaa
J t 0
0 B
D
Faradays Law
Amperes Circuital Law
Gauss Law for the Electric Field
Continuity Equation
Gauss Law for the Magnetic Field
E dl = C
d
dtB dSS
H dl =C J dS+
d
dtS D dSS
10/16/2009
5
Radiative and Non-RadiativeElectromagnetic Fields
Non-radiative case:Electric field lines are perpendicular to the Electrodes. Electric and Magnetic
fields are independent
Radiative case: Electric and Magnetic fields are inter-dependent and
mutually perpendicular to the propagation direction
E
H
k
Transmitter Receiver
Direction of propagation
E
H
10/16/2009
6
t
EE
t
DJH
t
H
t
BE
2
22
t
E
t
EE
t
EE
2
)]([)(~
;0~~ 22 tEEEkE
f
2
Propagation constant
e.g., for an Al alloy, 7 mm at f 9 kHz
EM Wave in a ConductorMaxwells Equations
EEE 2)(
Use
To get the wave equation
Eddy current testing is typically performed at frequencies lower than 10 MHz, with probe coil and defect dimensions less than 1 cm. In the air medium outside the metal testpiece, the characteristic length is the wavelength . At usual eddy current test frequencies, is in the range of kilometers, so the quasistatic approximation is always applicable in the air medium.
i
i
2
We get the diffusion equation !
In a highly conducting testpiece, the time derivative terms in the wave equation reduce to
in the frequency range for eddy current testing andconvert the wave equation into the diffusion equation,where the characteristic length is the skin depth
For time harmonic fields, Fourier transform leads to
Low Frequency Effects
10/16/2009
7
Inductive ReactanceThe reduction of current flow in a circuit due to induction is called inductive reactance.
The direction of the magnetic field can be determined by taking your right hand and pointing your thumb in the direction of the current. Your fingers will then point in the direction of the magnetic field.
It can be seen that the magnetic field from one loop of the wire will cut across the other loops in the coil and this will induce current flow (shown in green) in the circuit.
The induced current working against the primary current results in a reduction of current flow in the circuit.
It should be noted that the inductive reactance will increase if the number of winds in the coil is increased since the magnetic field from one coil will have more coils to interact with.
According to Lenz's law, the induced current must flow in the opposite direction of the primary current.
Induction Effects
Current carrying wire produces primary
magnetic field
Primary magnetic field induces currents in a
nearby conductor
Eddy currents generate secondary magnetic field which will oppose the
primary magnetic field
Eddy Current Generation
10/16/2009
8
Secondary magnetic field opposing the primary magnetic field alters
the current flow in the wire leading to inductive reactance
Near a flaw, eddy current flow is disturbed, changing the secondary
magnetic field, which in turn modifies the inductive reactance
Eddy Current Changes
The impedance of an eddy current probe may be affected by the following factors:
variations in operating frequency
variations in electrical conductivity and the magnetic permeability of a object or structure, caused by structural changes such as grain structure, work hardening, heat treatment, etc.
changes in liftoff or fill factor resulting from probe wobble, uneven surfaces, and eccentricity of tubes caused by faulty manufacture or damage
the presence of surface defects such as cracks, and subsurface defects such as voids and nonmetallic inclusions
dimensional changes, for example, thinning of tube walls due to corrosion, deposition of metal deposits or sludge, and the effects of denting
the presence of supports, walls, and brackets
the presence of discontinuities such as edges
ECT measures Impedance changes
10/16/2009
9
mmHz H/mm % IACS
Skin Depth
If the eddy current circuit is balanced in air and then placed on a piece of aluminum, the resistance component will increase and the inductive reactance of the coil decreases.
If a crack is present in the material, fewer eddy currents will be able to form and the resistance will go back down and the inductive reactance will go back up.
Changes in conductivity will cause the eddy current signal to change in a different way.
When a probe is placed on a magnetic material such as steel, the reactance increases. This is because the magnetic permeability of the steel concentrates the coil's magnetic field. This increase in the magnetic field strength completely overshadows the magnetic field of the eddy currents.
The presence of a crack or a change in the conductivity will produce a change in the eddy current signal similar to that seen with aluminum.
Impedance Plane Diagram
10/16/2009
10
Flaw Detection Using ECT
Response at 50 kHz Response at 300 kHz
A B C
Rea
ctan
ce
Rea
ctan
ce
Resistance Resistance
ECT probe movement
Flaw Detection using ECT in Tubes
10/16/2009
11
Electrical conductivity of a metal depends on several factors, such as its chemical composition and the stress state of its crystalline structure.
It can be used for sorting metals, checking for proper heat treatment, and inspecting for heat damage.
The technique can be used to easily sort magnetic materials from nonmagnetic materials but it is difficult to separate the conductivity effects from the magnetic permeability effects, so conductivity measurements are limited to nonmagnetic materials.
The technique usually involves nulling an absolute probe in air and placing the probe in contact with the sample surface. The thickness of the specimen should generally be greater than three standard depths of penetration.
Generally large pancake type, surface probes are used to get a value for a relatively large sample area.
To sort materials using an impedance plane device, the signal from the unknown sample must be compared to a signal from a variety of reference standards such as the IACS (International Annealed Copper Standard).
Electrical Conductivity Measurements
Ait
AEJ
Induced eddy currents
On the impedance plane, thickness variations exhibit the same type of eddy current signal response as a subsurface defect, except that the signal represents a void of infinite size and depth. The phase rotation pattern is the same, but the signal amplitude is greater.
The depth of penetration of the eddy currents must cover the entire range of thicknesses being measured. Typically, a frequency is selected that produces about one standard depth of penetration at the maximum thickness.
But at lower frequencies the probe impedance is more sensitive to changes in electrical conductivity. Any variations of conductivity over the region of interest have to be at a sufficiently low level for reliable measurements..
Metal thickness measurements
10/16/2009
12
The thickness of nonmetallic coatings on metal substrates can be determined simply from the effect of liftoff on impedance. The coating serves as a spacer between the probe and the conductive surface. Thicknesses between 0.5 and 25 m can be measured to an accuracy between 10% for lower values and 4% for higher values.
Contributions to impedance changes due to conductivity variations should be phased out, unless it is known that conductivity variations are negligible, as normally found at higher frequencies.
Thickness of Nonmetallic Coatings on Metal Substrates
Schematic representations of defect profile curves. (a) Small probe diameter. (b) Large probe diameter.
Imaging with ECT
10/16/2009
13
Recent developments
Typical MWM sensor and MWM-Arrays: (a) MWM sensor, (b) scanning five-element MWM-Array, (c) eight-element MWM-Array for detection on fatigue initiation, (d) four-element MWM-Rosette for detection and monitoring of fatigue cracks at fasteners (note that (c) and (d) are examples of MWM-Arrays designed for permanent mounting).
Meandering Winding Magnetometer (MWM) is a novel eddy current sensor that can measure absolute magnetic and conducting properties of ferrous and nonferrousalloys on flat and curved surfaces
Features of MWM Sensors
.
MWM sensors and MWM-Arrays can be permanently mounted for crack detection and monitoring in difficult-to-access fatigue-critical locations on operating equipment, e.g. in fuel tanks on aircraft, or between layers in a lapjoint
MWM can detect precrack fatigue damage in austenitic
stainless steels.
MWM provides the capability for continuous on-line monitoring of
crack initiation and growth during fatigue tests of coupons,
components, and full-scale test articles.
10/16/2009
14
Example of MWM Sensor for Fatigue Damage Assessment
Detection of fatigue damage in Type 304 stainless steel by MWM bi-directional
permeability measurements.
Representative measurement grids relating the magnitude and phase of the sensor trans-impedance to the (a) lift-off and magneticpermeability for 4340 alloy steel and (b) lift-off and electrical conductivity for titanium and Type 304 stainless steel.
Pulsed ECT
(a) Schematic representation of the model for theoretical calculations. Coil impedances of two cases (with and without coating) are calculated and used to predict the time-domain current differences.
(b) The step-function voltage that was applied to excite the coils in the PEC measurements and the resulting current difference between the two cases.
Comparison of measured data and theoretical calculations using the PEC method: (a) non-magnetic coatings on magnetic base metal
(zinc on steel) and (b) magnetic coatings on non-magnetic base metal
(nickel on copper).
10/16/2009
15
Advantages are:
Crack detection (Sensitive to small cracks and other defects)
Detects surface and near surface defects
Material thickness measurements
Coating thickness measurements
Conductivity measurements for:Material identificationHeat damage detectionCase depth determinationHeat treatment monitoring
Equipment is very portable
Test probe does not need to contact the part
Inspects complex shapes and sizes of conductive materials
Advantages and Limitations of ECTSome of the limitations of eddy current inspection include:
Only conductive materials can be inspected
Surface must be accessible to the probe
Skill and training required is more extensive than other techniques
Surface finish and and roughness may interfere
Reference standards needed for setup
Depth of penetration is limited
Flaws such as delaminations that lie parallel to the probe coil winding and probe scan direction are undetectable