Biomedical Instrumentation, Measurement and DesignELEC4623/ELEC9734/
Lectures 3 & 4Biopotential Electrodes and Tissue Equivalent Circuits
Session 2, 2009 2ELEC4623/ELEC9734
Biopotential Electrodes
Biopotentials are generated by nerve or muscle tissueThe most widely used of all biological transducers
l h d h lBioelectric phenomena are associated with almost every organ system
Session 2, 2009 ELEC4623/ELEC9734 3
Bioelectric signals sensed by biopotential electrodes
Bioelectric signal Abbreviation Biologic sourceElectrocardiogram ECG (EKG) Heart (body surface)C di l t H t (i t ll d)Cardiac electrogram - Heart (internally sensed)Electromyogram EMG MuscleElectroencephalogram EEG BrainEl t ti EOG E di l fi ldElectrooptigram EOG Eye dipole fieldElectroretinogram ERG Eye retinaAction potential - Nerve or muscleElectrogastogram EGG StomachGalvanic skin response GSR Skin
Session 2, 2009 ELEC4623/ELEC9734 4
First impressions
Biopotential electrodes must conduct current to sense voltage (V = I.Zin)Normally current is very small (but non-zero) and Zin very largey y in y gMust have small current (< pA) to prevent instrument from affecting the readingCurrent must flow across the interface between the body and electronic measuring circuitelectronic measuring circuit
Bioelectrode converts ionic current in body to electronic current in circuit via electrode-electrolyte interface
Session 2, 2009 ELEC4623/ELEC9734 5
Electrode – Electrolyte Interface
Electrode Electrolyte (neutral charge)Electrode Electrolyte (neutral charge)C+, A- in solution
C C+
Current flow
Electron flow opposite direction
CC
A-
C+
C+e-
e-
To conventional current flow
A-
C+ : Cation A- : Anion e- : electron
Session 2, 2009 ELEC4623/ELEC9734 6
Fairly common electrode materials: Pt, Carbon, …, Au, Ag,…Electrode metal is use in conjunction with salt, e.g. AgCl or polymer coats
Electrode – Electrolyte Interface
−−
−+
+↔
+↔
meAAneCC
m
n1)
2)Oxidation → Reduction ←
1) If electrode is made of atom with same material as cation, then (i) this material gets oxidised and enters the electrolyte ( ) g yas a cation, and (ii) electrons remain at the electrode and flow in the external circuit.
2) If anion can be oxidised at the electrode-electrolyte interface to form a neutral atom, electrons are given to the electrode.
Reactions can go in either directionC t fl f l t d t l t l t O id ti (L f )Current flow from electrode to electrolyte : Oxidation (Loss of e-)Current flow from electrolyte to electrode : Reduction (Gain of e-)Session 2, 2009 7ELEC4623/ELEC9734
Half cell potential (HCP)
When electrode is immersed in electrolyte it (usually) begins to oxidise (reduction could happen too, depending on materials)
Electron remains in metal electrodeElectron remains in metal electrodeCation moves into electrolyte.
As the reaction continues, a polarised double layer is set up at , p y pthe interface
Electrons in electrodeCations in Electrolyte
Eventually this electric field stops any further cations moving to electrolyte
The voltage difference when this equilibrium is reached is calledThe voltage difference when this equilibrium is reached is called the Half Cell Potential (HCP)
Session 2, 2009 ELEC4623/ELEC9734 8
Measuring half cell potential
Not possible to measure half cell potentialNot possible to measure half cell potential We try by introducing another material into the electrolyte to complete the circuit (hence the half in the term)But all we now measure is the difference between two differentBut all we now measure is the difference between two different HCPs!
Half cell potentials are thus measured with respect to a hydrogen electrode based on the reaction:hydrogen electrode based on the reaction:
2 2H H +2 2 2H H e+ −↔ +
I b l l d i l i
Session 2, 2009 9ELEC4623/ELEC9734
Its absolute electrode potential is estimated to be 4.44 ± 0.02 V at 25 °C
Standard half cell potentials for common electrode materials (at 25 °C)
Metal and Reaction Potential E°, VAl → Al3+ + 3e- -1.706
Zn → Zn2+ + 2e- -0.763
Cr → Cr3+ + 3e- -0.744Cr → Cr 3e 0.744
Fe → Fe2+ +2e- -0.409
Cd → Cd2+ + 2e- -0.401
Ni → Ni2+ +2e- -0.230
Pb → Pb2+ + 2e 0 126Pb → Pb2+ + 2e- -0.126
H2 → 2H+ + 2e- 0 (by definition)
Ag+Cl- → AgCl + e+ +0.233
2Hg+2Cl- → Hg2Cl2 + 2e- +0.268
Cu → Cu2+ + 2e- +0.340
Cu → Cu+ + e- +0.522
Ag → Ag+ + e- +0.799
Au → Au3+ + 3e- +1.420
Au → Au+ + e+ +`1.680
Session 2, 2009 ELEC4623/ELEC9734 10
Nernst equation and ionic availability
Ionic activity is defined as the availability of an ionic species in solution to enter into a reaction
In dilute solutions ionic activity is approximately equal to ionic concentration
At higher concentration, intermolecular effects are significant, so activity of ions is less than their concentration
If ionic activities across a membrane are a1 and a2Nernst equation may be rewritten as:
aRT ⎡ ⎤1
2
ln aRTEnF a
⎡ ⎤= − ⎢ ⎥
⎣ ⎦Session 2, 2009 ELEC4623/ELEC9734 11
Nernst equation and HCP
−++↔+ neDCBA δγβα
For the general oxidation-reduction reaction:
γβ
The Nernst equation for half cell potential is:
⎥⎦
⎤⎢⎣
⎡+= βα
δγ
BA
DC
aaaa
nFRTEE ln0
where E0 : Standard Half Cell Potential
E : Half Cell Potential
I i A i i ( ll i )a : Ionic Activity (generally same as concentration)
n : Number of valence electrons involved
Session 2, 2009 12ELEC4623/ELEC9734
PolarisationStandard HCP measured only when zero current flow across interface WhenStandard HCP measured only when zero current flow across interface. When current flows, difference between observed HCP and equilibrium zero current HCP is known as overpotential:
• V is the total overpotential or polarisation potential of the electrode
ACRp VVVV ++=
Vp is the total overpotential or polarisation potential of the electrode• VR is the ohmic (resistive) overpotential caused by current flow across
electrode/electrolyte resistance (this resistance can vary with current but not necessarily linearly)V i h i i l hi h l f h i h• VC is the concentration overpotential which results from changes in the distribution of ions in the electrolyte in the vicinity of the electrode-electrolyte interface when a current flows.
• VA is the activation overpotential which arises from the fact that reactions in 1) and 2) in slide 7 involve the expenditure of different activation energies which are not identical in both directions (oxidation and reduction different).
Session 2, 2009 ELEC4623/ELEC9734 13
Polarisable and nonpolarisable electrodes
Perfectly polarisable electrodes Would behave like capacitors in that no net charge would actually cross the electrode-electrolyte interface. Majority of overpotentialy j y pfrom concentration potential.These are best approximated, by noble or inert metals (e.g. platinum) which do not easily oxidize.
Nonpolarisable electrodes Require that the current passes freely across the interface without any expenditure of energy, thus no overpotentials.any expenditure of energy, thus no overpotentials. The electrode, which closest approaches a nonpolarizableelectrode, is the silver/silver chloride (Ag/AgCl) electrode.
Session 2, 2009 ELEC4623/ELEC9734 14
The Silver/Silver Chloride Electrode (Ag/AgCl)Ag electrode with thin layer of AgCl deposit. g y g pElectrolyte commonly contains KCl.
−+ +↔ eAgAg ↓↔+ −+ AgClClAgSAg Cl
a a K+ −× =Under equilibrium
K is called the
AgCl not very soluble and precipitates at rate KsPrecipitation: formation of solid in solution during a reaction
+↔ eAgAg ↓↔+ AgClClAg
⎤⎡ KRT
KS is called the solubility product
⎥⎥⎦
⎤
⎢⎢⎣
⎡+=
−Cl
sAg a
KnFRTEE ln0
RT RT[ ]0 ln lnAg s Cl
RT RTE E K anF nF −⎡ ⎤= + − ⎣ ⎦
First two terms are constants, third term d d i f Cl hi h i ldepends on concentration of Cl- which is also stable due to high concentration in biological fluids, so Ag/AgCl electrode is stable
Session 2, 2009 15ELEC4623/ELEC9734
The electrical noise evident is of such a frequency and magnitude as to seriously
Electrical Noise from Ag/AgCl electrodesFigure shows electrical noise generated for Ag/AgCl electrodes against purely frequency and magnitude as to seriously
interfere with the recording of a high quality ECG signal.
for Ag/AgCl electrodes against purely metallic Ag electrodes.
(a) Normal Ag/AgCl electrode (b) Ag/AgCl Electrode with AgCl rubbed off (c)AgCl redeposited
Session 2, 2009 16ELEC4623/ELEC9734
Electrode-electrolyte interface equivalent circuitThe current voltage relationship are often non linear and dependent on theThe current voltage relationship, are often non-linear and dependent on the magnitude of the current passing through the electrodeElectrode characteristics are different for high and low current densities, are frequency dependent and are also dependent on the current waveformsCapacitance comes from the double layer distribution of charge that occurs at the electrode-electrolyte interfaceQuestion!!!High frequency: f→∞ Zt t→??? 2High frequency: f ∞ Ztot ???Low frequency f→0, Ztot→???.
0.25cm2 Metallic silver electrode impedance
C it f l t d l t l t i t fCd : capacitance of electrode-electrolyte interfaceRd : resistance of electrode-electrolyte interfaceRs : resistance of electrolyte and wiresEhc : half cell potential for electrode
Session 2, 2009 17ELEC4623/ELEC9734
The skin – cross section
Session 2, 2009 18ELEC4623/ELEC9734
The skin electrode interface
Session 2, 2009 19ELEC4623/ELEC9734
Equivalent circuit of skin electrode interface
? ?? ?
The skin: an additional interface!Session 2, 2009 20ELEC4623/ELEC9734
Equivalent circuit of skin electrode interface
??
The skin: an additional interface!Session 2, 2009 21ELEC4623/ELEC9734
Equivalent circuit of skin electrode interface
The skin: an additional interface!Session 2, 2009 22ELEC4623/ELEC9734
Equivalent circuit of skin electrode interface
The skin: an additional interface!Session 2, 2009 23ELEC4623/ELEC9734
Equivalent circuit of skin electrode interface
The skin: an additional interface!Dominant elements
Session 2, 2009 24ELEC4623/ELEC9734
Impedance vs. frequency
Session 2, 2009 25ELEC4623/ELEC9734
Motion Artifacts
Mechanical disturbance of the distribution of charge at the electrode/electrolyte interface, will momentarily change the HCP until equilibrium is restoredHCP until equilibrium is restored.
Motion artifact mainly affects polarisable electrodes, but minimal for non polarisable electrodes (Why?)minimal for non-polarisable electrodes (Why?)
Mechanical disturbances can also influence Ese (Vep). Removing h ll h l h ff dthe stratum corneum will thus also have an effect on reducing
sensitivity to motion artefacts by short-circuiting this source(don’t abrade too much – irritation!)
Session 2, 2009 26ELEC4623/ELEC9734
Motion artifact(a) Metallic Ag electrodes in(a) Metallic Ag electrodes in agitated physiological saline solution
(b) Same electrodes with an AgCl surface film in agitated physiological saline solution
(c) Output from amplifier used for recordings, when electrodes are replaced by a 1/5-kΩare replaced by a 1/5-kΩresistor
Heavy lines under curvesHeavy lines under curves indicate periods of agitation of saline solution
Session 2, 2009 27ELEC4623/ELEC9734
Motion artifact
Skin stretch is known to be one of the largest generators of motion artifact (Talhouet and Thakor, Physiol. Meas. 17 (1996) 81–93.)
Session 2, 2009 28ELEC4623/ELEC9734
Motion artifact – skin stretch
Session 2, 2009 29ELEC4623/ELEC9734
Motion artifact – skin stretch
Session 2, 2009 30ELEC4623/ELEC9734
Research – motion artifact is a big issue in telehealth6
2
4
V)
-2
0
plitu
de (m
V
-6
-4
EC
G A
mp
0 5 10 15 20 25 30-10
-8 Artifact Mask
Session 2, 2009 31ELEC4623/ELEC9734
0 5 10 15 20 25 30Time (s)
Example of movement artifact in a sample ECG recording. Also shown is the artifact detection determinedby the algorithm developed in UNSW.
Typical electrode/skin impedanceDried exudation of the Sterculia Urens tree
F and E are synthetic conductive adhesives,C and D are Karaya gum
and other species of Sterculia (native to India)
C and D are Karaya gum conductive adhesive, A and G are wet paste (gel) electrodes with stainless-steel conductor,G is wet paste Ag/AgClelectrode and
h b hH is the same as G but with a light abrasion of the skin
Typical values of electrode-skin impedance for eight different electrode types measured at 0.05 Hz, 1 minute after application
Session 2, 2009 32ELEC4623/ELEC9734
Metal plate electrodes
Large surfaceAncient (still used!)
l d k h lMetal disk with stainless steel, platinum or gold coating
dFoam pad version is disposable, others may be reused
(a) Metal-plate electrode used for application to limbs. (b) Metal-disk electrode applied with surgical tape. (c) Disposable foam-pad electrodes, often used with ECG
Session 2, 2009 33ELEC4623/ELEC9734
Suction Electrodes
No straps or adhesives requiredPrecordial (chest) ECGPrecordial (chest) ECGCan only be used for short periods (why?)S ll h hSmall contact area so high source impedance compared with metal plate electrode
Session 2, 2009 34ELEC4623/ELEC9734
Floating electrodes
Insulating
Metal disk (no direct contact with skin)
Reusable (metal type)
Double sided
Insulatingpackage
Reusable (metal type)
Double-sidedadhesive-tapering
Electrolyte gel in cavity
(a) (b)( ) ( )Snap coated with Ag/AgCl External snap
Plastic cup Plastic diskGel-coated sponge
( )
TackFoam padCapillary loops
Dead cellular material
Germinating layerDisposable Ag/AgCl
Minimize motion artefacts (how?)(c)
g y
Session 2, 2009 35ELEC4623/ELEC9734
Flexible electrodes
Body contours are often irregularRegularly shaped rigid electrodes may not alwayselectrodes may not always work.Example: infants Material :
l l h lPolymer or nylon with silver Carbon filled silicon rubber (Mylar film)
(a) Carbon-filled silicone rubber electrode. (b) Flexible thin-film neonatal electrode.(c) Cross-sectional view of the thin-film electrode in (b).
Session 2, 2009 36ELEC4623/ELEC9734
Internal electrodes
Needle and wire electrodes for percutaneous measurement of biopotentials
Percutaneous electrodePercutaneous = electrode crosses the skin
(a) Insulated needle electrode(b) Coaxial needle electrode(b) Coaxial needle electrode(c) Bipolar coaxial electrode(d) Fine-wire electrode connected
to hypodermic needle, before being inserted
(e) Fine-wire electrode after insertion(e) Fine wire electrode after insertion(f) Cross-sectional view of skin
and muscle, showing coiled fine-wire electrode in place
Session 2, 2009 37ELEC4623/ELEC9734
Fetal ECG Electrodes
Electrodes for detecting fetal electrocardiogram during labor, by means of intracutaneous needles (a) Suction electrode. (b) Cross-sectional view of suction electrode in place, showing penetration of probe through epidermis. (c) Helical electrode, which is attached to fetal skin by corkscrew type action.
Session 2, 2009 38ELEC4623/ELEC9734
Standards for pregelled electrodes (AAMI)Direct-current offset voltage. “A pair of electrodes connected gel to g p ggel after 1 min of stabilization must exhibit offset voltages no greater than 100mV.
Combined offset stability and internal noise. “A pair of electrodes connected gel to gel after 1min of stabilization shall generate aconnected gel to gel after 1min of stabilization shall generate a voltage no greater than 150μV P-P in the pass band of 0.15 to 100 Hz.”
Alternating current impedance. “For a pair of electrodes connected gel g p p gto gel the impedance at 10Hz will not exceed 3kΩ. The average for 12 pairs will not exceed 2kΩ.
Defibrillation overload recovery. “The absolute value of polarisation potential of a pair of electrodes connected gel to gel shall not exceedpotential of a pair of electrodes connected gel to gel shall not exceed 100 mV, 5 s after each of four capacitor discharges of 10μF charged to 200V”.
Bias current tolerance. “The observed dc voltage offset change across g gan electrode pair connected gel to gel shall not exceed 100mV when subjected to a continuous 200nA dc current over a period, not exceed 8 hours, as recommended by the manufacturer.”
Session 2, 2009 39ELEC4623/ELEC9734
Practical issuesElectrode and lead wire should ideally be of the same materialElectrode and lead wire should ideally be of the same material
Why? (Hint: Galvani potential)
Use the same material for each electrode, if more than one is used
Use tape to hold electrode onGive some slack to allow for movementLead wire should be VERY flexible
Input impedance of the amplifier must be much higher than the impedance of the source (body) and electrode equivalent impedances, otherwise distortion of the signal will occur. (Geddes et al. 1966, The relationship between input impedance and electrode area)relationship between input impedance and electrode area).
Session 2, 2009 40ELEC4623/ELEC9734