eXact® Micro 10 Methodology for Poolside Testing

www.sensafe.com / its@sensafe.com

2

What makes a good test? 1. Test is acceptable or compliant

2. Test is appropriate for the staff technical ability

3. Test is robust, reagents and equipment are reliable & stable in summer heat

4. Test is not affected by interferences

5. Test is accurate when staff performs the test correctly

eXact® Photometers

• 0.01 ppm (mg/L) precision

• 4 ml built-in, built-in cell

• Automatic countdown timing

• 20 Test memory for each direct read test• Incorporates eXact® Strip technology

patented EZ-3 method (Fill, Dip and Read)

• No color matching

• Lightweight and waterproof (IP-67) meter

• CE certified

3

EZ-3™ Reagent Delivery Method

4

eXact® Micro 10 Instructions

5

I n d e x :

P a r a m e t e r P a g e M E N U

T o t a l A l k a l i n i t y 4 A L 1p H 5 P H 2F r e e C h l o r i n e ( D P D - 1 ) 6 C L 3C o m b i n e d C h l o r i n e ( D P D - 3 ) 7 C L 3B r o m i n e 8 C L 3T o t a l C h l o r i n e ( D P D - 4 ) 8 C L 3P h o s p h a t e 9 P O 4C a l c i u m H a r d n e s s 1 0 C A 5C o p p e r 1 0 C U 8C h l o r i d e ( S a l t ) 1 1 C H 6C y a n u r i c A c i d 1 2 C Y 7H y d r o g e n P e r o x i d e C o n v e r s i o n 1 5 C Y 7

A b o u t Y o u r P h o t o m e t e r 3T i p s f o r B e s t A c c u r a c y 1 3B a t t e r y I n s t a l l a t i o n 1 4W a r r a n t y ( 2 y e a r ) 1 4R e o r d e r I n f o r m a t i o n 1 5U S E P A C o m p l i a n c e 2 , 1 6K i t C o m p o n e n t s 1 6

V i s i t u s o n l i n e a t s e n s a f e . c o m / m i c r o 1 0 f o r u p - t o - d a t ep r o d u c t i n f o r m a t i o n & N E W t e s t s a v a i l a b l e .

I D E A L F O R T E S T I N G P O O L S A N D S P A S .A C C E P T E D B Y M O S T H E A L T H D E P A R T M E N T SF O R C O M M E R C I A L P O O L A N D S P A T E S T I N G .

T h e e X a c t ® M i c r o 1 0 A d v a n c e d P h o t o m e t e r S y s t e m h a s b e e n d e s i g n e d f o r u s e w i t h t h e e X a c t ® S t r i p M i c r o r e a g e n t d e l i v e r y s y s t e m .

®

A d v a n c e d P h o t o m e t e r S y s t e m

U S E P A , D I N , & I S O C o m p l i a n t f o r F r e e , C o m b i n e d , & T o t a l C h l o r i n e T e s t i n g( 4 5 0 0 - C L G , D I N S t a n d a r d 3 8 4 0 8 G 4 , I S O 7 3 9 3 / 2 )

®

M a n u f a c t u r e d B y : I n d u s t r i a l T e s t S y s t e m s , I n c .

P h o n e : - I N S I D E T H E U . S .- O U T S I D E T H E U . S .F a x :

I T S @ S E N S A F E . C O MW W W . S E N S A F E . C O M

w w w . p o o l c h e c k o n l i n e . c o m

M i c r o 1 0 i s M a n u f a c t u r e d a n d t e s t e d i n

a n I S O 9 0 0 1 F a c i l i t y

T o p r e v e n t w a r r a n t y i s s u e s , y o u m u s t r e g i s t e r y o u r o w n e r s h i p w i t h i n 3 0 d a y s o f p u r c h a s e a n d i n c l u d e t h e s e r i a l n u m b e r ( o n b o t t o m o f M i c r o 1 0 M e t e r ) a n d v a l i d e - m a i l a d d r e s s . V i s i t o u r w e b s i t e s e n s a f e . c o m / m i c r o 1 0 t o r e g i s t e r . B y r e g i s t e r i n g , y o u r 2 - y e a r w a r r a n t y i s a c t i v a t e d a n d y o u w i l l b e u p d a t e d w h e n n e e d e d r e g a r d i n g t h e M i c r o 1 0 .

R e v i s i o n 0 5 / 1 9 / 1 4 S t a n d a r d K i t

S t a r t e r K i t

6

2

e X a c t ®

1

2

e X a c t ®

R 0 5 1 9 1 4

A L 1P H 2C L 3

C L 3

P O 4

C A 5C H 6C Y 7

C Y 7

C U 8

2

+ 2

1 2

1 2

1 2

1

11

1 1

1 E x p e c t e d M e t e r3

F

E

D

C

B

A

GH

+–

I T E M P I C T U R E DL E T T E R

C O M P O N E N TN A M E

A M i x i n g C a p / C e l l C o v e r

B C E L L ( B u i l t - i n P l a s t i c , 4 m l )

C L C D D i s p l a y

D R E A D B u t t o n

E M E N U B u t t o n

F Z E R O / O N B u t t o n

G B a t t e r y C o v e r w i t h l o o p f o r l a n y a r d

H A A A B a t t e r i e s ( x 4 ) N O T I N C L U D E D

e X a c t ® S t r i p M i c r o 1 0 h a s b e e n d e s i g n e d t o o f f e r t h e u s e r a m o r e “ G r e e n ” a n d c o s t - e f f e c t i v e a l t e r n a t i v e t o t e s t i n g .I n s t e a d o f u s i n g a 1 0 m l w a t e r s a m p l e , e X a c t ® S t r i p M i c r o 1 0 u s e s a 4 m l w a t e r s a m p l e , w h i c h u s e s u p t o 6 0 %l e s s c h e m i c a l p e r t e s t . T h e a c c u r a c y o f t h e m e t e r i s m a i n t a i n e d b y d e s i g n i n g t h e p h o t o c e l l w i t h a 2 0 m m p a t h l e n g t h .

( U s e d f o r C y a n u r i cA c i d a n d I r o n t e s t s

a n d w h e n t e s t i n go u t d o o r s )

M e a s u r e m e n t M e t h o d : P h o t o m e t r i cL i g h t S o u r c e : L i g h t E m i t t i n g D i o d e ( L E D )W a v e l e n g t h : 5 2 5 n mT r a n s m i s s i o n E q u i v a l e n t : 1 0 0 - 0 . 0 0 % TP h o t o m e t r i c P r e c i s i o n : + / - 0 . 1 / 0 . 0 1 % T ( a s t r a n s m i s s i o n )A u t o m a t i c R a n g i n g : A v a i l a b l e f o r C h l o r i n eD i s p l a y : 3 - d i g i t c u s t o m i z e d l i q u i d c r y s t a l d i s p l a y w i t h a n n u n c i a t o r sC E L L P a t h l e n g t h : 2 0 m m

C e l l C h a m b e r : C u s t o m - m o l d e d , p r o p r i e t a r y , P E T p l a s t i c f u s e d i n t o c h a m b e r , n o n - r e m o v a b l eS a m p l e R e q u i r e d : 4 m l ( 0 . 1 3 o z )O p e r a t i n g T e m p e r a t u r e R a n g e : 0 - 5 0 ° C ( 3 2 ° - 1 2 2 ° F )P o w e r S u p p l y : ( 4 ) A A A a l k a l i n e b a t t e r i e s N O T I N C L U D E DB a t t e r y L i f e : > 2 0 0 0 t e s t s w i t h a l k a l i n e b a t t e r i e sE l e c t r o m a g n e t i c C o m p l i a n c e : E m i t t e d I n t e r f e r e n c e - E N 6 1 3 2 6( E M C ) I m m u n i t y t o I n t e r f e r e n c e - E N 6 1 3 2 6W a t e r p r o o f R a t i n g : E x c e e d s I P 6 7W e i g h t : I n s t r u m e n t : 1 4 0 g ( 5 o z )D i m e n s i o n s : I n s t r u m e n t : 5 ( W ) x 3 . 5 ( D ) x 1 6 . 5 ( H ) c m ; ( 2 x 1 . 4 x 6 . 3 7 5 i n )

7

eXact® Strip Micro AL - Part #486641

8

9

Micro 10 Video – Total Alkalinity

10

eXact® Strip Micro pH - Part #486639

11

12

Micro 10 Video - pH

13

eXact® Strip Micro CL (DPD-1)Part #486637

14

15

Micro 10 Video – Free Chlorine

16

eXact® Strip Micro CL (DPD-3)Part #486638

17

18

Micro 10 Video – Combined Chlorine

19

eXact® Strip Micro PO4 - Part #486814

20

21

Micro 10 Video – Phosphate

22

eXact® Strip Micro CA - Part #486629

23

24

Micro 10 Video – Calcium Hardness

25

eXact® Strip Micro CU - Part #486632

26

27

Micro 10 Video – Copper

28

eXact® Strip Micro CH - Part #481657-II

29

30

20:1 Dilution Method for Salt Test

Micro 10 Video – 20:1 Dilution Method

31

32

Micro 10 Video – Chloride (Salt)

33

eXact® Reagent Micro CY-II Part #481652-II

34

35

Micro 10 Video – Cyanuric Acid

36

eXact® Strip Micro HP - Part #486648

37

38

39

Micro 10 Video – Hydrogen Peroxide

40

Ready Snap® Verification Solutions

41

• Ampoule (solution) used to verify system (operator, reagent, and photometer) is performing correctly

• Available for Total Alkalinity, pH, Calcium Hardness, Copper, Cyanuric Acid, Phosphate, and Salt,

• No dilution needed, just snap ampoule open, fill cell with sample, and run test

42

For US Inquiries and Re-Orders:Industrial Test Systems, Inc.

Phone: INSIDE THE U.S.OUTSIDE THE U.S. Fax:

ITS@SENSAFE.COMwww.poolcheckonline.comWWW.SENSAFE.COM

Part No. Description Price®

For European & Middle East Inquiries and Re-Orders:ITS Europe, LTD

Tel: Fax:

SALES@SENSAFE.COMwww.poolcheckonline.comWWW.ITSEUROPE.CO.UK

Contact Information

Reorder Information

I T S E u r o p e L t d .

I n n o v a t o r s o f W a t e r Q u a l i t y T e s t i n g

P a r a m e t e r T e s tM e n uI t e m

D e s i r e dV a l u e

A c c e p t a b l eR a n g e

B o r d e r l i n eR a n g e

A L 1 T o t a l A l k a l i n i t y 5 2 p p m 4 2 - 6 2 3 8 - 6 7P H 2 p H 7 . 5 p H 7 . 3 - 7 . 7 7 . 2 - 7 . 8 P O 4 P h o s p h a t e 2 p p m 1 . 7 - 2 . 3 1 . 6 - 2 . 4C A 5 C a l c i u m a s C a C O 3 2 2 2 p p m 1 9 2 - 2 5 2 1 8 0 - 2 6 4C Y 7 C y a n u r i c A c i d 4 0 p p m 3 0 - 5 0 2 6 - 5 4C U 8 C o p p e r a s C u + 2 0 . 2 2 p p m 0 . 1 7 - 0 . 2 7 0 . 1 4 - 0 . 3 0C U 8 T o t a l C o p p e r a s C u + 2 0 . 6 4 p p m 0 . 5 6 - 0 . 7 2 p p m 0 . 5 3 - 0 . 7 5 p p m

R051914

Assigned Value Chart for READY SNAP® 1P (lot 4813) Solution:

43

P a r a m e t e r T e s tM e n uI t e m

D e s i r e dV a l u e

A c c e p t a b l eR a n g e

B o r d e r l i n eR a n g e

A L 1 T o t a l A l k a l i n i t y 5 2 p p m 4 2 - 6 2 3 8 - 6 7P H 2 p H 7 . 5 p H 7 . 3 - 7 . 7 7 . 2 - 7 . 8 P O 4 P h o s p h a t e 2 p p m 1 . 7 - 2 . 3 1 . 6 - 2 . 4C A 5 C a l c i u m a s C a C O 3 2 2 2 p p m 1 9 2 - 2 5 2 1 8 0 - 2 6 4C Y 7 C y a n u r i c A c i d 4 0 p p m 3 0 - 5 0 2 6 - 5 4C U 8 C o p p e r a s C u + 2 0 . 2 2 p p m 0 . 1 7 - 0 . 2 7 0 . 1 4 - 0 . 3 0C U 8 T o t a l C o p p e r a s C u + 2 0 . 6 4 p p m 0 . 5 6 - 0 . 7 2 p p m 0 . 5 3 - 0 . 7 5 p p m

R051914

Assigned Value Chart for READY SNAP® 1P (lot 4813) Solution:

1. Turn on the eXact® Micro

three times with clean

2.test method procedure you want verified (for example

3.Snap®

Verification Standard by twisting the top 180° (see Image 1

4 . S q u e e z e p l a s t i c a m p o u l e

s e e I m a g e 2

5.sample followed with a quick shake of the meter to

remaining water drops.

6. S q u e e z e p l a s t i c a m p o u l e

s e e I m a g e 2

7 .y o u d o n o r m a l l y u s i n g t h e c o r r e c t p r o c e d u r e ( s e e I m a g e 4 a n d r e f e r t o p h o t o m e t e r m a n u a l

8 .

V a l u e c h a r t p r o v i d e d b e l o w a n d o n b a c k ( s e e I m a g e 5 .

9 . I f y o u r v a l u e i s w i t h i n t h e

a r e o p e r a t i n g t h e e X a c t ®

f o r t h i s M E N U t e s t m e t h o d .

10. If your value is Borderline, review the Instruction Manual and verify that you are following the proper procedure.

Step 1 through Step 8.

3

74

5

N O T E : CA5 test

result was 220

Micro 10 Video – Ready Snap 1P

44

45

eXact® Micro 10 for Poolside Testing

U s e t h e E Z - 3 ™ M e t h o d !

FAST, ACCURATERES ULTS IN

E Z - 3 ™ M e t h o d

ITS ManufacturingITS Manufacturing• Manufacturing system based on Good Manufacturing

Practices (GMP) • Temperature and humidity controlled environment • Quality Control laboratory• 3 Research and Development laboratories• Latest manufacturing and packaging equipment, with

2011 investment of $500,000 in new equipment• Currently adding 10,000 square feet of additional

manufacturing space to complete Dec 2014

46

ITS Quality ControlITS Quality Control• Material is QC tested during filter paper impregnation

and after final assembly• Testing is performed with water standards prepared

with ACS grade chemicals and Industry Accepted procedures

• Standards are verified by Industry Accepted method using colorimetric, titrimetric and gravimetric methods

• All reagent strip lots are confirmed to be match listed specifications

47

Shelf LifeShelf Life

• Accelerated Shelf Life Study–Product is placed in 55° C ovens for fixed times that simulate 6 month, 12 month, 18 month, 24 month, and 30 month product aging. The material is then tested for performance to estimate shelf life.

• Real Time Shelf Life Study–Product is stored at room temperature and evaluated at six month intervals to confirm real shelf life. This study is repeated for every product at documented frequencies to update shelf life dating given to products

48

49

eXact® Micro 10 Methodology for Poolside Testing

www.sensafe.com / its@sensafe.com