Post on 04-Feb-2022
2 CT Basics
The initial use of computed tomography (CT) for applications
in radiological diagnostics during the seventies sparked a
revolution in the field of medical engineering. And even
throughout the eighties, a CT examination lost little if any of
its special and exclusive character. In the meantime, however,
times have changed. Today computed tomography represents
a perfectly natural and established technology which has
advanced to become an indispensable and integral component
of routine work in clinics and medical practices.
The following brochure will give you insights into the history of
computed tomography as well as its present status. Nevertheless,
this is an on-going development process. And Siemens will
continue to consistently extend its leading position in CT
technology. With you. For you. And for your patients.
Innovations For People
3
Historical Outline 4
State of the Art 12
Sequential CT 12Spiral CT 13
Setup of a CT System 14
Scanning unit (gantry) 15X-ray components 15Data acquisition components 16
Scanner Parameters 18
Collimation 18Increment 19Pitch 20Rotation time 21mAs 22
Image Generation 23
A CT image is produced 23What does a CT image show? 24Windowing 24
Image Evaluation and Image Processing 25
Two-dimensional displays 26Three-dimensional displays 27
Clinical Use of CT 32
CT in general clinical use 32Advantages of spiral CT in clinical use 32CT-Angiography (CTA) 33When is a CT examination indicated? 33
The invention of computed tomography is considered to be thegreatest innovation in the field of radiology since the discoveryof X-rays. This cross-sectional imaging technique provideddiagnostic radiology with better insight into the pathogenesisof the body, thereby increasing the chances of recovery. In1979, G.N. Hounsfield and A.M. Cormack were awarded theNobel Prize in medicine for the invention of CT.
Today, CT is one of the most important methods of radiologicaldiagnosis. It delivers non-superimposed, cross-sectional imagesof the body, which can show smaller contrast differences thanconventional X-ray images. This allows better visualization ofspecific differently structured soft-tissue regions, for example,which could otherwise not be visualized satisfactorily.
Since the introduction of spiral CT in the nineties, computedtomography has seen a constant succession of innovations.The development of slipring technology allowed for acontinuously rotating gantry – the prerequisite for spiral CT.The first spiral CT scanner was a Siemens SOMATOM Plus system.Today this technology is widely used.
Historical Outline
4 CT Basics
11/08/1895
Discoverer: The physicistand later nobel prizewinner Wilhelm ConradRoentgen (1845–1923),dean at the JuliusMaximilian University ofWuerzburg and holder ofthe chair of physics
Discovery of X-rayradiation
1896
F.H. Williams
F.H. Williams succeeds intaking the first chest X-rayin Boston, and CarlSchleussner develops thefirst silver bromide coatedphotographic X-ray platesin Frankfurt a. Main,Germany
Diagnostic results cannow be archived; untilthen, fluorescent screensresulting in high radiationexposure had been used
M i l e s t o n e s i n t h e H i s t o r y o f C T
1903
E.A.O. Pasche builds acollimator for suppressingscattered radiation
Radiation protection forall parties involved for thefirst time; until then, bothpatients and physiciansalways had to face the“bare“ X-ray tube
E.A.O. Pasche
1913
Gustav Bucky develops thescattered radiation grid inBerlin, Germany
The engineer William D.Coolidge builds the first high-vacuum hot-cathode tube inMassachusetts, USA
Image quality is improvedby eliminating scatteredradiation before it strikesthe X-ray film. Performanceand durability of X-ray tubesgreatly enhanced
Gustav Bucky
William D. Coolidge
5
X-ray image ofhis wife’s hand
1930/31
Allesandro Vallebona developsstratigraphy, thus paving the wayfor tomography
Not long after, Bernard Ziedsesdes Plantes develops planigraphy,thereby further refining thetomographic technique andcreating geometrically perfectimages for the first time
To some extent this helpedovercome the disadvantageof superimposition in the two-dimensional display of three-dimensional objects whererelevant structural informationcan be lost; the requiredplanigraphic depth had tobe guessed
Allesandro Vallebona
Bernard Ziedsesdes Plantes
1972
Godfrey N. Hounsfield
In London, Godfrey N.Hounsfield’s developmentof computed tomographymarks the beginning ofa new era in diagnosticimaging
With the aid of computedtomography it waspossible for the firsttime to produce non-superimposed imagesof an object slice. Usingso-called convolvers,digitized scan data arereconstructed into animage of a slice plane
1974
First CT systemfrom a medicalequipmentmanufacturer
SIRETOM
1976
OPTI X-raytube with highheat capacity:1.0 MHU
5 s scan time
Whole body CT
SOMATOM
Instant image reconstruction
1978
ECG-synchronizedCT image
512 x 512 displaymatrix
SOMATOM 2
SIRETOM CT system
6
1981/82
512 x 512 reconstructionmatrix
50 cm FOV (display field);graphically selectable andcontinuously variable
3 s scan time for a full360° rotation
1 mm slice width
0.5 mm high-contrastresolution
OPTI X-ray tube with1.35 MHU
Automatic scanning(AutoScan)
Graphically specified scanrange
SOMATOM DR3
1983/84
± 25° gantry tilt angle
70 cm gantry opening
OPTI X-ray tube with1.75 MHU
Integrated MPRfunction (Multi PlanarReformatting)
SOMATOM DRH
1985/86
Modern applications:DYNAMIC CT and 3Dwith MPR and SSD(Shaded SurfaceDisplay)
3D Shaded SurfaceDisplay (SSD)
SOMATOM CR
3D OberflächendarstellungSchlüsselbeinfraktur,
SOMATOM DRH
Coronales MPR,Femurkopf/Azetabulum,
SOMATOM DRH
7
1987/88
Shorter examinationtimes
Increased patient comfort
DURA X-ray tube with4.3 MHU
40 kW DURA generator
Integrated gantry cooling:Heat exchanger
1024 x 1024 display anddocumentation matrix
Real-time MPR
Selection of five slicethicknesses (1–10 mm)
SOMATOM Plus
1 s/360° slipringtechnology
Continuous rotation oftube and detector
1989
Continuous volumemeasurement
Data acquisition duringone breath hold
24 s/24 cm continuousvolume acquisition
First spiral CT scanner in routine operation
1990
32 s continuousSPIRAL scan
QUANTILLARC 3detector
SOMATOM Plus-S
Spiral CT acquisition withcontinuous tabletop feed
Left: 3D Shaded SurfaceDisplay of the lung, high-resolution, spiral CT
Right: Spiral CT, lung,high-resolution, 2 mm,SOMATOM Plus
1991
Compact slipring scanner
Monitor with 4-quadrant display(5122 ea.) for multitasking
High-efficiency QUANTILLARC 4detector
SOMATOM AR
IntuitiveWindows™user interface,mouse-controlled
SOMATOM AR
8 CT Basics
Intuitive, mouse-controlled Windows™ user interface
More volume inthe same time
1994
Larger volumes acquired faster,shorter breath hold, increased patientcomfort, improved thin-sliceresolution
0.75 s scan time for a full 360°rotation, 0.5 s for a quick scan
100 s/130 cm multiple spiralacquisition with 1:1 pitch
Selection of six-slice thicknesses(1–10 mm)
± 30° remote-controlled gantry tilt
QUANTILLARC 6 inert gas detector
40–55 kW modularly upgradeableDURA generator
DURA X-ray tube with 5.3 MHU
SOMATOM Plus 4
Subsecond spiral CT
Advantages of subsecondSPIRAL scans
1992
Integrated CT-Angiography
DURA-S X-ray tube with 4.3 MHU
Prospective and retrospectiveoverlapping SPIRAL reconstruction
SPIRAL XP: 40 s (pitch variable from1:1 to 2:1)
Maximum Intensity Projection (MIP)and Minimum Intensity Projection(MinIP) incl. editing function for CTA
CT-Angio (MIP)of the renalarteries,SOMATOMPlus-S
CT-Angio (MIP)of the circleof Willis,SOMATOMPlus-S
9
Improved thin-sliceresolution
The same volumein less time
Subsecond spiral CT, long MPR,abdomen/pelvis, SOMATOM Plus 4
40 cm/30 s
30 cm/30 s
0,75 s SPIRAL 1,0 s SPIRAL
0.75 s SPIRAL 1.0 s SPIRAL
22 s 0.75 s SPIRAL
30 s 1.0 s SPIRAL
1996/97
UFC Ultra Fast CeramicDetector
Same image quality withsignificantly reducedradiation dose
Improved MPR – coronal/sagittal/oblique
Real-time image display(1 image/rotation)
Volume Rendering Technique(VRT)
Lightning UFC™
1998
Multislice spiral scanningwith 4 slices per rotation
Fastest rotation timeof 0.5 s
Virtually isotropicresolution
HeartView CT: First useof Cardio CT in routineoperation, temporalresolution of up to 125 ms
SureView™:Reconstruction algorithmfor multislice spiralscanning
SOMATOM Volume Zoom
1999
Intuitive user interface forall medical applications andproducts
syngo software
10
2001
CARE Dose
11
2000
Designed to be thesmallest and most cost-effective spiral CT scanner
Plug-and-play installation
Do-it-yourself applicationtraining and serviceconcept supported byinteractive CD-ROMs
SOMATOM Smile
2002
Multislice scanning with 16 slicesper rotation
Isotropic resolution of up to0.6 mm
Routine submillimeter slices(16 x 0.75 mm)
Cardio CT with extremely highimage quality – temporalresolution of up to 105 ms withHeartView CT
Reconstruction algorithm withcone beam correction
SOMATOM Sensation 16
Dose modulation forreduced radiation
Sequential CT
A cross-sectional image is produced by scanning atransverse slice of the body from different angularpositions while the tube and detector rotate 360°around the patient with the table being stationary.The image is reconstructed from the resultingprojection data.
If the patient moves during the acquisition, the dataobtained from the different angular positions are nolonger consistent. The image is degraded by motionartifacts and may be of limited diagnostic value. Thetomographic technique is suitable only to a limitedextent for the diagnosis of anatomical regions withautomatism functions such as the heart or the lung.
State of the Art
12 CT Basics
Patienttable
Gantry
Detectors
Tubes
0
0
Spiral CT
Spiral CT is often referred to as “volume scanning“. This impliesa clear differentiation from conventional CT and the tomographictechnique used there. Spiral CT uses a different scanningprinciple. Unlike in sequential CT, the patient on the table ismoved continuously through the scan field in the z directionwhile the gantry performs multiple 360° rotations in the samedirection. The X-ray thus traces a spiral around the body andproduces a data volume. This volume is created from a multitudeof three-dimensional picture elements, i.e. voxels.
The table movement in the z direction during the acquisitionwill naturally generate inconsistent sets of data, causing everyimage reconstructed directly from a volume data set to bedegraded by artifacts. However, special reconstruction principles– interpolation techniques which generate a planar set of datafor each table position – produce artifact-free images. Thus itis possible to reconstruct individual slices from a large datavolume by overlapping reconstructions as often as required.
Software applications enable the clinical use of spiral CTeven for regions which are subject to involuntarymovements.
13
Direction of patienttransport
Path of rotation tubeand detector
z, mm
t, s
A CT system comprises several components.
These basically include:
• The scanning unit, i.e. the gantry, with tubeand detector system
• The patient table
• The image processor for image reconstruction
• The console
The console represents the man-machine interfaceand is designed to be multifunctional. It is the controlunit for all examination procedures, and is also usedto evaluate the examination results. To enhance theworkflow, Siemens has developed a double consolecapable of performing both functions at the sametime.
Setup of a CT System
14 CT Basics
Scanning unit (gantry)
A CT scanning system consists of an X-ray unit, whichfunctions as a transmitter, and a data acquisition unit,which functions as a receiver. In commercial CTsystems these two components are housed in a ring-shaped unit called the gantry.
X-ray components
• Tube
Manufacturers of CT systems use X-ray tubes withvariable focal spot sizes. This makes sense, becausevolumes for which good low-contrast resolution isessential need to be scanned with a large focal spotand high power, whereas high resolution imageswith thin slices require a small focal spot. Tubesused in modern CT scanners have a power ratingof 20–60 kW at voltages of 80 to 140 kV. Thesystems can, however, be operated at maximumpower for a limited time only. These limits aredefined by the properties of the anode and thegenerator. To prevent overloading of the X-ray tube,the power must be reduced for long scans. Thedevelopment of multi-row detector systems haspractically excluded this limitation, since thesedetector systems make much more efficient useof the available tube power.
• Shielding
Each CT scanner is equipped with grids, collimatorsand filters to provide shielding against scatteredradiation, to define the scan slice and to absorb thelow-energy portion of the X-ray spectrum. In thisway, both the patient and the examiner areprotected.
15
•••••••
Data acquisition components
• Detector
The detector system plays a special role in the interaction ofthe CT components. It converts the incident X-rays of varyingintensity to electric signals. These analog signals are amplifiedby downstream electronic components and converted todigital pulses. Over time, certain materials have proven veryeffective in the utilization of X-rays. For example, Siemensuses UFC (Ultra Fast Ceramic) Detectors which, due to theirexcellent material properties, dramatically improve imagequality.
16 CT Basics
Scintillation crystal
X-ray radiation
v
Light
UFC Detector
Photodiode
• Multi-row detector
Multi-row detectors utilize radiation delivered from theX-ray tube more efficiently than single-row detectors. Bysimultaneously scanning several slices, the scan time can bereduced significantly or the smallest details can be scannedwithin practicable scan times. In the adaptive array detectorsused by Siemens, the rows inside the detector are very narrow,becoming wider as you move toward its outer edges in thez direction (longitudinal axis of the body). A combinationof collimation and electronic interconnection providesconsiderable flexibility in the selection of slice thicknesses.At the same time the space required by the detector septa,and therefore the unused space, is minimized.
17
Adaptive array detector
Ultra Fast Ceramic Detector (UFC)
5 mm 2.5 mm 1.5 1 5 mm2.5 mm1 1.5
2 x 0.5 mm 4 x 1 mm 4 x 2.5 mm 4 x 5 mm
Variable slice collimation through primary collimation
and electronic signal combination
18 CT Basics
Collimation
The radiation beam emitted by the X-ray tube is shapedusing special diaphragms also referred to as ”collimators”.A distinction can be made between two types of collimators.The source collimator is located directly in front of the radiationsource – i.e. the X-ray tube. It reduces the radiation beam toform the maximum required fan beam, thus also determiningthe emitted dose.
The detector collimator, which is positioned directly in frontof the detectors, is primarily used to shield the detector againstscattered radiation, thus preventing image artifacts. Thecollimation and focal size determine the quality of the sliceprofile.
From the data volume from a multislice scanner images can bereconstructed with slice thickness equal to or larger than thedetector collimation. For example, a 5 mm collimation allowsimages to be reconstructed with a slice thickness of 5 mmor more. The widest range of possibilities in the selection ofcollimation and reconstructed slice thicknesses is offered inspiral CT using multidetector systems.
Scanner Parameters
19
Increment
The increment determines the distance between imagesreconstructed from a data volume. If an appropriate incrementis used, overlapping images can be reconstructed. In sequentialCT, overlapping images are obtained only if the table feedbetween two sequences is smaller than the collimated slicethickness. This, however, increases the patient dose.
In spiral CT the increment is freely selectable as a reconstructionparameter, i.e. by selecting the increment the user canretrospectively and freely determine the degree of overlapwithout increasing the dose. Overlapping reconstructions offerthe advantage of better image quality due to lower noise andeasier and more accurate diagnosis of small structures.
An illustrative example: A 100 mm range was acquired in thespiral mode with 10 mm collimation. After the acquisition,slices of 10 mm thickness can be reconstructed at any point ofthis range. If an increment of 10 mm is used, contiguous slicesof 10 mm thickness are reconstructed every 10 mm.(see figure 1).
If an increment of 5 mm is used, slices of 10 mm thickness arereconstructed every 5 mm. The slices overlap by 50%. With anappropriate increment an overlap of 90% can be achieved.Modern CT systems allow the reconstruction of slices witharbitrary increments (see figure 2). A clinical useful overlapis 30%–50%.
Increment
Overlapping slices
Slice thickness
Figure 2
Figure 1
20 CT Basics
Pitch
An important factor in spiral scanning is the table feed perrotation. The larger the table feed, the faster (i.e. with fewerrotations) a body region can be scanned. However, if the tablefeed is too large, image quality will be impaired. In this contextthe term “pitch“ is used. For single-row systems the definition
pitch = table feed per rotation/collimation
is generally accepted. Experience has shown that a good imagequality can be obtained with a pitch between 1 and 2. It shouldalso be noted that the dose can be significantly reduced insingle-row systems if a pitch factor > 1 is applied. In the contextof multi-row systems, pitch cannot yet be defined clearly. Theambiguity involved becomes apparent in the following example(SOMATOM Sensation 4):
Collimation 4 x 2.5 mm, table feed 10 mm
First possibility: pitch = 10 mm/4 x 2.5 mm = 1
Second possibility: pitch = 10 mm/2.5 mm = 4
To avoid misunderstandings, we use “feed per rotation“ ratherthan “pitch“ on the user interface.
15 s 30 s
The first pitch scans the same volume in less time
21
Rotation time
Rotation time is the time interval needed for a complete 360°rotation of the tube-detector system around the patient. Itaffects the spiral scan length and thus the coverage of the scanrange during a certain period of time. Ultra modern CT systemsrequire only 0.4 seconds for one rotation. A short rotation timehas the following advantages:
• A longer spiral scan can be acquired in the same scan time
• The same volume and the same slice thickness can be scannedin less time
• Motion artifacts are eliminated
• Savings on contrast media through shorter examination times
• Reduced patient discomfort, since less contrast mediumis required
For shorter examination times or fast acquisition of largeanatomical regions a subsecond rotation time is recommended.This applies especially, for instance, to constantly moving organssuch as the heart.
22 CT Basics
mAs
The mAs value (e.g. 100 mAs) is the product of thetube current (e.g. 200 mA) and the rotation time(e.g. 0.5 s).
In multi-row CT systems we simplify this equation byusing what is commonly called “effective mAs”. Thisis the product of the tube current and the exposuretime for one slice (rotation x collimation/feed perrotation).
The selected mAs and tube voltage determine thedose. The mAs value selected depends on the typeof examination. Higher mAs values reduce the imagenoise, thus improving the detectability of lowercontrasts. For visualizations of soft tissue, i.e. regionsof low contrast, a higher dose and larger slice thicknessare required. The abdomen and brain are typicalregions of soft-tissue contrast. Visualizations of bonesor the lungs, i.e. regions of high contrast, as well ascontrast studies of vessels require lower doses andthinner slices.
Siemens CT scanners also feature the CARE Dosetechnical measures package (CARE = CombinedApplications to Reduce Exposure), which wasdeveloped to reduce patient exposure to radiation.This package guarantees shorter examination times,the lowest possible exposure to radiation, and imagesof excellent quality.
Ultra modern computer technology “monitors” thepatient during the entire examination period. Duringeach rotation, the radiation is continuously measuredand modulated according to the current attenuationlevel. CARE Dose thus makes it possible to vary theradiation dose depending on the patient’s anatomyand thus reduce it by as much as 56%.
Scanner parameters determine the image quality.Optimal performance of spiral CT systems can beachieved only with an optimal combination ofparameters.
23
A CT image is produced
Acquisition:
In the simplest case, the object (here a round cylinder) is linearlyscanned by a thin, needle-like beam. This produces a sortof shadow image (referred to as ”attenuation profile” or”projection”), which is recorded by the detector and the imageprocessor. Following further rotation of the tube and the detectorby a small angle, the object is once again linearly scanned fromanother direction, thus producing a second shadow image. Thisprocedure is repeated several times until the object has beenscanned for a 180° rotation.
Display:
The various attenuation profiles are further processed in theimage processor. In the case of simple backprojection, eachattenuation profile in the scanning direction is added up inthe image memory. This results in a blurred image due to thedisadvantage of this simple backprojection, i.e. each objectnot only contributes to its own display, but also influencesthe image as a whole. This already becomes visible after 3projections. To avoid this problem, each attenuation profile issubjected to a mathematical high-pass filter (also referred toas ”kernel”) prior to the backprojection. This produces overshootand undershoot at the edges of the object. The mathematicaloperation is referred to as ”convolution”. The convolvedattenuation profiles are then added up in the image memoryto produce a sharp image.
Image Generation
X-raytube
Translation
Rotation
Detector andmeasuringelectronicsAttenuation
profile(simplified)
Backprojection
without convolution with convolution
0 projections
1 projection
3 projections
N° projections
N° projections
Profilesection
Collimator
Ka 2000
24 CT Basics
What does a CT image show?
The CT image does not show these µ values directly,but the CT numbers according to Hounsfield:
CT number =1000 (µ – µwater) / µwater,
CT numbers are measured in HU = Hounsfield units. The CTnumber of water and air is defined as 0 HU and –1000 HUrespectively; this scale has no limit in the positive rangeof values. Medical scanners typically work in a range of–1024 HU to +3071 HU.
Windowing
In the CT image, density values are represented as gray scalevalues. However, since the human eye can discern only approx.80 gray scale values, not all possible density values can bedisplayed in discernible shades of gray. For this reason, thedensity range of diagnostic relevance is assigned the wholerange of discernible gray values. This process is calledwindowing.
To set the window, it is first defined which CT number thecentral gray scale value is to be assigned to. By setting thewindow width, it is then defined which CT numbers above andbelow the central gray value can still be discriminated by varyingshades of gray, with black representing tissue of the lowestdensity and white representing tissue of the highest density.
Lung window Soft-tissue window
The Hounsfield scale
CT no.HU8070605040302010
0
Kidneys
Blood
Pancreas
4020
5030
Liver
6050
7050
CT no.HU
1000800600400200
0-200-400-600-800
-1000
Water
Air
Lungs
Fat
Compact bone
Spongybone
+4-4
20050
-990-1000
-550-950
-80-100
250
25
The first obvious results of any CT examination arethe axial cross-sectional images. Since these imagesare already available in digital form on a storagemedium, they can be processed immediately by theprocessor. The evaluation of geometrical parameterssuch as distance, area, angle and volume as well asdensity measurements are part of clinical routinetoday. The tissue density is determined using the CTnumber averaged over a defined evaluation area,the so-called Region Of Interest (ROI). Geometricalparameters can be defined more accurately than inconventional radiography, since the problems ofsuperimposition and distortion do not exist in CT.
Image Evaluation and Image Processing
Examples of 2D post processing capabilities offeredby modern CT systems:
• Display of the CT numbers of arbitrary pixels inthe image
• Display of the CT number profiles along arbitrary intervals in the image
• Zoom and shift of image segments
• Filtering of images
• Addition, subtraction or other possibilities of imageoverlapping
The terms “two-dimensional“ (2D) and “three-dimensional“ (3D) refer to the image content.Views showing entire volumes are referred to as“3D displays”.
26 CT Basics
Two-dimensional displays
CT mainly uses the transverse plane as the imagingplane. Therefore, views of other orientations usuallyhave to be reconstructed from the original images.This is done by Multi Planar Reformation (MPR).
With MPR, a series of axial images are combined toform a stack. By aligning the same columns and rowsof all images of a series, the computer reconstructscontiguous images for any arbitrary plane. Theexaminer can then interactively page through theseimages, using the mouse for example, and evaluatethem (iMPR). By paging forwards and backwards,he finds the image which most clearly shows theanatomical and pathological detail of interest.4-quadrant displays with axial, sagittal, coronary,and oblique slice orientation are standard todayand provide a good overview.
An extension of the MPR technique allows theinteractive combination of thin slices into slices (slabs)of any thickness. This technique is used for bettervisualization of structures, such as vessels, whichextend across several slices. This technique hasbecome known as Sliding-Thin-Slab (STS).
Advantages of two-dimensional displays
• Direct, accurate display of CT numbers
• Easy orientation in the volume
• Unambiguous interpretation of image values
• Interactive evaluation on the monitor
• Basis for 3D images
sagittal coronary transversal
27
Three-dimensional displays
For 3D visualizations the position and viewing directionwith respect to the volume of interest must beindicated. Along this viewing direction through thedata volume a spatial image is reconstructed fromthe CT numbers pixel by pixel. Such virtual views areespecially suitable for structures which clearly standout against their surrounding area, such as the skeletalsystem or contrast-filled vessels.
Shaded SurfaceDisplay (SSD)
Maximum IntensityProjection (MIP)
28 CT Basics
• Shaded Surface Display (SSD)
For threshold based surface displays a CT number,e.g. 150 HU, is predefined as a threshold. All pixels,i.e. voxels, which exceed this threshold valuecontribute to the result image. From the viewer‘sposition these are all those pixels along each beamwhich first exceed the threshold. The surface isthen reconstructed from the totality of pixels andilluminated by an artificial light source to create ashading effect (Surface Shaded Display, SSD). Theshading effect intensifies the depth impressionfor the viewer. With this technique, however, theoriginal density information from the CT numbersis lost.
With SSD it must be noted that the gray scale valuesare no longer related to the original density of thestructures. For example, if several structures whichexceed the threshold value are superimposed inthe viewing direction, only the structure closestto the front surface of the monitor is displayed.This also applies if the structures behind it haveconsiderably higher CT numbers.
It also needs to be taken into account that SSDdisplays always depend on the threshold valueselected. For example, the display of vascularstenoses can be distorted by selecting aninappropriate threshold. If the threshold value isincreased, a higher degree of stenosis results, if itis decreased, stenoses may be obscured. In this waycalcifications and contrast medium in the vesselscan no longer be differentiated. Therefore, SSDimages are hardly suitable for diagnosis. They are,however, used to document results or for 3Ddisplays.
CT numberHU
600
400
200
55 10
Lumen filledwith CM
Calcification
CT numberprofile alongthe search beam
160 HUthreshold Soft tissue
y, cm
Observer’s position
Virtual lightsource
29
• Maximum Intensity Projection (MIP)
Maximum Intensity Projections (MIP) are based onthe voxels with the highest density, i.e. CT number.Along a virtual beam extending from the viewerthrough the 3D image volume, the voxel with thehighest density value is displayed in the resultingMIP image. Each MIP image is therefore a 2Dprojection image. Running several MIP imagesin fast sequence can create a realistic spatialimpression. For this, an image series is createdby varying the viewing angle in small steps.
In contrast to SSD, with this technique a minimumof density information is retained. Moreover, theprojection is always a combination of those voxelsfrom the entire volume that have the highestdensity, irrespective of whether these voxels arelocated further towards the front or the back of theimage. Unlike in SSD images, calcifications andcontrast medium are clearly differentiated in MIPimages.
Alternative to the MIP image, it is also possibleto display pixels with the lowest intensity in theprojection image. These images are called MinIPimages. They are used to display structures suchas the bronchial tree.
CT numberHU
600
400
200
55 10
Lumen filledwith CM
Calcification
CT numberprofile alongthe search beam
160 HUthreshold Soft tissue
y, cm
Observer’s position
30 CT Basics
• Volume Rendering Technique (VRT)
Volume Rendering Technique (VRT) refers to theprocess of reconstructing a 3D model from a 2Dimage stack. VR techniques go beyond SSD and MIPtechniques in their basic approach and performance.They are not limited to a certain threshold ormaximum density value. Instead, all densityvalues along a virtual beam which have a suitableweighting can contribute to the result image. Incontrast to MIP and SSD, the entire Hounsfield scalecan be included in VRT. Each CT number is assignedan opacity and color via freely selectable andinteractively modifiable transfer functions. Thismakes it possible to simultaneously display anextremely wide variety of tissue structures of variousdensity or HU value in a single volume data set.
CT numberHU
600
400
200
55 10
Lumen filledwith CM
Calcification
CT numberprofile alongthe search beam
160 HUthreshold Soft tissue
y, cm
Observer’s position
• Virtual endoscopy (VE)
A special type of VRT is perspective VolumeRendering (pVR), which is used mainly to generatevirtual endoscopic views. This technique is used toobtain a perspective view of the display region.Virtual endoscopy is mainly used for anatomicalcavities: These include, for example, the bronchialtree, large vessels, the colon and paranasal sinuses.But VE is also used for areas not directly accessiblefor conventional endoscopy, such as the cisternaeof the brain or the gastrointestinal region.
When the endoscope is inserted in a cavity displayedwith the perspective Volume Rendering Technique,this gives the user the impression of “flying through”the displayed region (virtual flight).
Advantages of three-dimensional displays
• Realistic display of volumes
• Presentation of the entire volume in onesingle image
• Improved recognition of diagnosticallyrelevant details
• Helpful for more precise surgical planning
• CT image data as a basis for three-dimensional models
• Possible free rotation of 3D objects
31
CT in general clinical use
The use of spiral CT has significantly shortened scantimes compared to sequential CT. This is of greatadvantage when examining patients who, due to thenature of their illness, are unable to cooperate. Motionartifacts caused by different respiratory conditionsduring the acquisition are reduced considerably,because in spiral CT the entire volume is scannedfaster and without gaps. Multiple scans due tobreathing during the acquisition are no longernecessary. The patient dose is therefore reduced.
Advantages of spiral CT in clinical use
• Complete coverage of organs in a single respiratoryposition
• Short scan times (resulting in fewer motion artifactsand a lower contrast medium requirement)
• Additional diagnostic information due to improvedresolution (thinner slices) and 3D visualization inroutine operation
• Special cost-effective applications based onspiral CT
Clinical Use of CT
32 CT Basics
CT-Angiography (CTA)
CT-Angiography (CTA) enables the display of vascularstructures aided by injections of contrast medium.The introduction of the multislice scanner has madeit possible to display the entire vascular system withmaximum contrast enhancement in extremely shortscan times. Image post processing enables gooddisplay of the entire vascular system. Even smallvascular exits and origins (branches) and embolismsor dissection membranes can be displayed. Thephysician can retrospectively select any projectionand generate three-dimensional images, e.g. forsurgical planning.
33
34 CT Basics
HeadHead, general/brainOrbitaSella turcicaPetrous bonesParanasal sinusesCircle of Willis3D cranial, facial bone
NeckCervical soft tissueCarotids
ThoraxMediastinumThorax high resolutionThoracic vesselsPulmonary vesselsHeart
SpineCervical spineThoracic spineLumbar spine
When is a CT examination indicated?
Here are several examples of CTexaminations:
• Head • Abdomen• Neck • Extremities• Thorax • Spine
35
Abdomen/pelvisLiverCT-Arterioportography (CTAP)PancreasKidneys, biphaseAdrenal glandsRenal arteriesAbdominal vesselsSmall pelvisVessels, pelvic/lower extremity
ExtremitiesShoulder jointHip jointWrist boneKnee jointFoot
Order No. A91100-M2100-A844-1-7600
Printed in Germany
BKW 63844 WS 11025.
Siemens AG, Medical SolutionsComputed TomographySiemensstr. 1, D-91301 ForchheimGermanyTelephone: ++49 9191 18-0
The information in this document contains general
descriptions of technical possibilities which do not
necessarily apply to each individual case. Any
performance features individually required must
therefore be stipulated upon conclusion of the relevant
agreement.
Siemens reserves the right to modify designs and
specifications without prior notice. Please contact your
local Siemens Sales Representative for the most current
information.
Original images always lose a certain amount of detail
when reproduced.
Siemens AG, Medical SolutionsHenkestr. 127, D-91052 ErlangenGermanyTelephone: ++49 9131 84-0www.SiemensMedical.com